ThermoWorks R-PRO User Manual

IR Pro - Operating Instructions
Professional Infrared Thermometer (with type K socket) – The thermometer is a non-contact
infrared thermometer. Please remember to keep away from children and don’t use it for safety related applications.
(Default Screen)
Simply aim the thermometer at the measure target with Lens and press Meas. key to display the surface temperature. The Distance:Spot is 50:1. The two laser points are the reference for the target spot size. Please make sure the target area is within the field of view.
FUNCTION
Press Emissivity key for setting the emissivity
Press Emissivity key , then press Up key or Down key to set the emissivity, then press Mode key to confirm it. The emissivity can be changed from 0.10 (10E) to 1 (100E).
Press Mode key for scrolling more display function as follows.
** The thermometer will automatically shut off if left idle for more than 60sec.
In MAX, MIN, DIF, AVG mode:
In all modes: First hold on the Meas. key
Here will show the emissivity data. (The default emissivity is 0.95.)
Press Mode key for the Maximum (MAX), Minimum (MIN), Different between MAX and MIN (DIF) and Average (AVG)
modes. During the measurement, the special modes reading will be displayed beside the mode icon.
Press Up key or Down key key to change the High Alarm (HAL) or Lo Alarm (LAL), then press Meas. key to confirm it. For example: When the reading 26.9°C < LAL 27°C, the Low icon will flash and you will hear a beep sound.
Connect the thermocouple with Thermocouple socket and put the probe in/on the target, the thermometer will display the temperature automatically without pressing any button. To see the minimum or maximum data during the probe measurement, please hold down the Up key or Down key .
After measuring high temperatures, the probe may remain HOT for a while.
Press Up key for LOCK mode ON/OFF. The lock mode is particularly useful for continuous monitoring of temperatures for up to 60 minutes.
Press Down key to change display units from °C to °F and back
and press Up key for backlight function ON/OFF.
and press Down key for laser function ON/OFF.
1.800.393.6434
www.thermoworks.com
CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE TWO LASER BEAMS–PERMANENT EYE DAMAGE MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN. STORAGE & CLEANING The thermometer should be stored at room temperature between –20 to +65°C (-4~149°F). The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
HI
LOW
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements are possible ‘Battery Low’: battery needs to be replaced,
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries. Please note: It is important to turn
the instrument off before replacing the battery otherwise the thermometer may malfunction. Dispose of used battery promptly and keep away from children.
SPECIFICATIONS
Measurement Range -60 to +1000°C (-76 to +1832°F) -64 to +1400°C (-83.2 to +1999°F)
Operating Range 0 to +50°C (32 to +122°F)
Accuracy (Tobj=15~35°C, Tamb=25°C) ±1.0°C (1.8°F) +/-1% of reading or 1°C (1.8°F) whichever is greater (Test under
Emissivity Range 0.95 default – adjustable 0.1 to 1 step .01
Resolution (-9.9~999.9°C): 0.1°C/0.1°F, otherwise 1°C/1°F
Response Time 1sec (90%)
Distance:Spot 50:1
Battery Life Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.)
Dimensions 47.0 x 197.0 x 203.3mm (1.8_7.7_8.0 inch)
Weight 386.1 grams (13.62 oz) including batteries (AAA x 2pcs)
Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 10°C (18°F).
EMC/RFI:Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per meter, but the
performance of the instrument will not be permanently affected.
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and LAL.
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0oC (32°F) or +50°C (122°F). The thermometer should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature.
Error 5~9, for all other error messages it is necessary to reset the thermometer. To reset it, wait for auto power off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message remains please contact ThermWorks for further assistance.
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.
measurements are still possible
Non-contact Infrared Scan function Thermocouple Probe Scan function (K type; probe not included.)
(Tobj=-33~1000°C, Tamb=23±3°C) ±1.5% of reading or
1.5°C (2.7°F) whichever is greater
Tamb=23±6°C)
‘Battery Exhausted’: measurements are not possible
CERTIFICATE OF CONFORMANCE
The manufacturer of this instrument has implemented a quality assurance system under ISO 9001:2000 cer tified quality system and fully follow ISO GUM (Guide to the Expression of Uncertainty in Measurement) to evaluate the uncertainty of temperature standards, guarantee performance as below;
Calibration
Temperature
-30°C ±1.5°C ±0.3°C 0.3°C 0°C ±1.5°C ±0.4°C 0.1°C
400°C ±6.0°C ±2.0°C 0.2°C 900°C ±13.5°C ±3.0°C 0.4°C
Note: Assume the operation ambient temperature under 23 ± 3°C
Furthermore, we certify that this infrared thermometer has been inspected and found to comply with published specifications. This device has been calibrated by temperature standards that are traceable to NIST (U.S. National Institute of Standards and Technology) or NML (National Measurement Laboratory) of Taiwan, and the calibration procedure corresponds with generally accepted regulations and standards.
Max Error Target Accuracy Target Stability
ThermoWorks, Inc.
Utah, U.S.A.
www.thermoworks.com
Loading...