ThermoWorks P770 User Manual

Operation Manual
SERIE
Precision Measuring Instruments P700, P705, P710, P715, P750, P755, P770, P755-LOG, P770-LOG
DOSTMANN electronic
P700
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 1
Retractable stand
Mounting device for probe handle
Measuring port 1
Measuring port 2 (P705, P715, P755, P770, P755-LOG, P770-LOG)
USB PC interface port
Big LC-Display
Keyboard
Unit diagram
2
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 2
Operation Manual
Summary
1. Handling
1.1. General advices
1.2. Setting to work
1.3. Switch on/off
1.4. Menu
1.4.1. Measuring unit switching (UNIT)
1.4.2. Probe selection (Prob)
1.4.3. Difference temperature (Lin2)
1.4.4. Calibration (CAL)
1.4.4.1. Calibration combination probe (temperature/humidity)
1.4.5. Channel activation (Chnl)
1.4.6. Enter Area for volume flow (ArEA)
1.4.7. Memory setup (Lo6)
1.5. Recalling memory data (HOLD/MAX/MIN/AVG)
1.6. Measuring rate (FAST-modus)
1.7. AUTO-OFF-function
1.8. Special-functions
1.8.1 Ohm/microvolt/volt/hertz display
1.8.2 Zero adjustment
1.8.3 Deactivation of the 2nd channel (toggle)
2. Power supply / Changing the battery
3. Error codes / troubleshooting
4. Technical data
5. Interface protocol
6. Connector layout
7. Data of probe-calibration
8. Guarantee
3
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 3
1.1 General advice
• For cleaning the instrument please do not use abrasive cleaner but a dry or wet piece of cloth.
• Please store the measuring instrument in a dry and clean place.
• Avoid any force like shocks or pressure to the instrument.
• Do not use force to connect the probe or interface plugs in. The interface plug is different from the probe plug.
• If no sensor is connected to the instrument while switching on „open“ shows on the display (Please refer to chapter error codes/troubleshooting).
• A retractable stand on the back of the instrument allows it to be used as a bench top instrument.
1.2 Operation
Before switching on the instrument, connect the probe/s to the instrument and insert the battery (Please refer to chapter 6. Power supply / battery changing). A number on the instrument's housing marks each port.
1.3 Switching on and off
By operating the ON/OFF-key the instrument swit­ched on or off. After switching on the instrument indicates a full segment test for 1,5 sec., afterwards the instrument shows the calibration-code and the adjusted measurement category for channel 1. (Only 2-channel instruments: After additionally 1,5 sec. the instrument shows the calibration-code and the adjusted measurement category for channel 2.) Then it starts operating in measurement mode in­dicating the actual measurement value (for example: temperature). All instruments display the measure­ment value of channel 1 on the top display line (big display line). Directly below you will see the trending bargraph. The second channel of our 2-channel instruments you will see on the bottom line (small display line).
Fig 1:
Example of channel information after switching on.:
channel 1
Line 1: [oFF P = calibration code of channel 1 according to standard linearization (DIN), probe selec­tion of channel 1 = Pt100.
channel 2
Line 2: [oFF P = calibration code of channel 2 according to standard linearization (DIN), probe selec­tion of channel 2 = Pt100.
1. Handling
4
1
2
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 4
Note: On all instruments you can select the mea­surement category (chapter 1.4.2. Prob) according to the model version. Instruments that are delivered only with one probe the correct measurement cate­gory is preset. In other case please check chapter
1.4.2. (Prob) for the correct set-up.
1.4 Menu
The adjustments of the instruments like the mea­surement value, calibration of Up and down keys probes, deactivation of channels and so on are resulting from the structure of the menu. You will reach into the main menu by pushing [ENTER/MENUE]. With the up and down keys [

] you can choose your required menu item. Push [ESC] to be back in the measuring mode.
Menu structure
1.4.1 Measuring unit switching °C and °F
respective %rH, td or g/m
3
/ [Unit]
Unit = Measuring unit
Measuring unit temperature
(°C=Celsius, °F=Fahrenheit)
Measuring unit humidity
(%rH=relative humidity, td=dew point, g/m
3
=abso-
lute humidity)
Important!
5
Unit
°C
°F
m/s
%rh
g/m
3
°C td
°F td
Pa
hPa
Prob Lin2 CAL Chnl ArEA Lo6
P
J
K
L
N
R
S
T
RH
D
Pr
H
T1-T2 oFF
oP1
oP2
OFF c OFF
ON m ON
m
3
/s
up and down keyskey [ESC]key
[ENTER/MENUE]
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 5
To change the measuring unit push [ENTER/MENUE]. Use the up and down keys [] to select Unit. Push again [ENTER/MENUE]. On the left corner of the display appears a small 1, which indicates the selected channel. For changing the channel use the up and down keys [

] Push [ENTER/MENUE] to
confirm. On the right corner of the display appears
°C or °F, or %rH, td or gm
3
(depending on the selec-
ted probe). Use the up and down keys [

] again to adjust the requested measuring unit and push [ENTER/MENUE] to confirm. Push [ESC] to be back in the measuring mode.
1.4.2 Probe selection / [Prob]
Push [ENTER/MENUE] to change a probe. Use the up and down keys [

] to select Prob. Push
[ENTER/MENUE] to confirm.
On the left corner of the display appears a small 1, which indicates the selected channel. For changing the channel use the up and down keys [

]. Push again [ENTER/MENUE] to confirm. On the right cor­ner of the display appears the active probe. Use the up and down keys [

] to change the probe. Push [ENTER/MENUE] to confirm the requested probe. Push [ESC] to be back in the measuring mode.
Measurement Probe selection LC-Display Suitable for
(Prob) these types
of measuring
instruments:
Temperature Pt100 (RTD) P700/P705/
P750/P755/P770
P755-LOG/P770-LOG
Temperature Fe-CuNi Typ J for all P700-
instruments
Temperature NiCr-Ni Typ K for all P700-
instruments
Temperature Fe-CuNi Typ L for all P700-
instruments
Temperature NiCrSi-NiSi Typ N for all P700-
instruments
Temperature Pt13Rh-Pt Typ R P700/P705/
P750/P755/P770
P755-LOG/P770-LOG
Temperature Pt10Rh-Pt Typ S P700/P705/
P750/P755/P770
P755-LOG/P770-LOG
Unit

MENUE
Prob Lin2 CAL PrEn Chnl ArEA Lo6
Unit

MENUE
Prob Lin2 CAL PrEn Chnl ArEA Lo6
6
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 6
Measurement Probe selection LC-Display Suitable for
(Prob) these types
of measuring
instruments:
Temperature Cu-CuNi Typ T for all P700-
instruments
Humidity %rF P750/P755/P770
P755-LOG/P770-LOG
Flow m/s P750/P755/P770
P755-LOG/P770-LOG
Pressure Pa P750/P755/P770
P755-LOG/P770-LOG
Thermal flow m/s P750/P755/P770 probe P755-LOG/P770-LOG
Note: Please check the probe selection to be sure that the correct probe is entered. If a measurement value is changed with the ENTER-key at the menu Prob, the standard calibration will be automatically used.
Indications for the combination probes (temperature and humidity):
Please adjust relatively humidity as measurement value for the channel on which the combination probe is plugged. If you want that the measured temperature of the plugged combination probe is shown as well, you have to deactivate the channel on which no probe is plugged on (see chapter
1.4.6./Chnl).
1.4.3 Difference temperature (only 2-channel instruments) / [Lin2]
To display the difference temperature push [ENTER/MENUE]. Use the up and down keys [

] to select Lin2. Push again [ENTER/MENUE]. On the right corner of the display appears a T
1-T2
. Use the
up and down keys [

] to adjust the requested sel­ection. Push [ENTER/MENUE] to confirm. Push [ESC] to be back in the measuring mode.
Note: Both channel have to be activated for showing the difference temperature.
1.4.4 Calibration function / [CAL]
Despite high quality manufacturing techniques, each probe is slightly different from specified standards. To eliminate inaccuracies caused by exchanging or ageing of probes, the instrument offer easy calibra­tion functions which guarantee that the system accuracy is always as good as if the instrument was specifically calibrated to the individual probes in our laboratory.
Important!
Important!
Important!
7
Unit

MENUE
Prob Lin2 CAL PrEn Chnl ArEA Lo6
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 7
The instruments offer three calibration options:
1) [OFF]:Standard characteristic curve (e.g. Pt100­resistance according DIN IEC 60751)
2) [OP1]:Calibration by code (2 x four digit code) is equivalent to a 2-point calibration The code is marked clearly by a label on each stan­dard probe.
3) [OP2]:Calibration by physical standard references (1-point, 2-point or 3-point calibration)
CAL = calibration
Push [ENTER/MENUE] to calibrate the instrument with sensor. Use the up and down keys [

] to select CAL. Push again [ENTER/MENUE]. On the left corner of the display appears a small 1, which indicates the selected channel. For changing the channel use the up and down keys [

]. Push
[ENTER/MENUE] to confirm.
Use the up and down keys [

] to select the re­quested calibration option. Push [ENTER/MENUE] to confirm.
Standard linearization according DIN IEC 60751 / [oFF
Use the up and down keys [

] to select [oFF. Push [ENTER/MENUE] to confirm. Push [ESC] to be back in the measuring mode.
Calibration by code / oP1
Use the up and down keys [

] to select oP1. Push [ENTER/MENUE] to confirm. On the bottom of the display appears a very small 1, after this number a four-digit number (Hex-Code/0..F) is displayed. For changing the number use the up key [
]. For stepping to the next number use the
down key [
]. If the requested number is complete then push [ENTER/MENUE] to confirm. Now on the bottom of the display appears a very small 2, after this number a second four-digit number is displayed. For changing the number please follow the manual as before. Push [ESC] to be back in the measuring mode.
Note: After confirming oP1 by pushing [ENTER/ MENUE] the function oP1 (calibration by code) is activated, even though you leave the menu by press­ing [ESC].

CAL
oFF oP1 oP2
Important!
8
Unit

MENUE
Prob Lin2 CAL PrEn Chnl ArEA Lo6
DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 8
Display-indication with active calibration code(OP1): The CAL -segment and the small 1 on the left indica-
tes to the user that oP1 is activated.
Calibration by physical standard references / oP2
Use the up and down keys [

] to select oP2. Push [ENTER/MENUE] to confirm. On the bottom of the display appears 1 P. For chang­ing between a 1-Point [1 P], 2-Point [2 P] or 3-Point [3 P] - calibration use the up and down keys [

].
Example of a 1-Point calibration:
Push [ENTER/MENUE] to confirm. On the display appears Go. After the displayed measuring value is stabile push [ENTER/MENUE]. App. 2 seconds later in the first display line appears P1 (measuring value
1); in the second display line appears dP. (standing for decimal point).
Use the up and down keys [

] to select the num­ber of digits after decimal point:
dP. = two decimal digits dP . = one decimal digit
(decimal point is jumping one digit to the right)
Push [ENTER/MENUE] to confirm. On the second line of the display appears Si -. Use the up and down keys [

] to select the requested sign.
Si = reference standard shows
a negative measuring value (below 0,00C°)
Si = reference standard shows
a positive measuring value (above 0,00C°)
Push [ENTER/MENUE] to confirm. On the second line of the display appears Fd 0. Use the up and down keys [

] to select the re­quested range:
Fd 0 = below 100,00°C Fd 1 = above 100,00°C
Push [ENTER/MENUE] to confirm. On the display appears 00.00. For changing the value use the up key [
] . For stepping to the next number use the
down key [
]. If the requested value is complete then push [ENTER/MENUE] to confirm. Push [ESC] to be back in the measuring mode.
Note: A break of the physical calibration cannot be done by the [ESC]-button. Switching off the in­strument can only do a break of the physical cali­bration.
