ThermoWorks 232-040 User Manual

Therma 20 Plus Therma 22 Plus
ThermaCheck Plus
Operating Instructions
INSTRUMENT OPERATION - Where applicable (Therma 20 Plus and Therma 22 Plus) connect a probe via the socket located at the top of the instrument, screw in the collar of the plug, finger tight, to ensure a secure connection (do not over tighten). Switch the instrument on and apply the tip of the probe to the substance or medium to be measured. The instrument reading may take a few seconds to stabilize, depending on the nature of the measurement and sensitivity of the probe.
HOLD - Press the HOLD button to stop measurements and freeze the display temporarily, press again to continue measuring.
AUTO-OFF - The unit will switch off automatically after 10 minutes. To disable auto-off press the HOLD button while switching on the instrument (this is reset when the instrument is switched off).
OPEN CIRCUIT/FAULTY PROBE - If the probe has developed an 'Open Circuit' fault the instrument will display 'Err'.
SENSORS - The ThermaCheck Plus incorporates a fixed thermistor probe. The Therma 20 Plus operates with an interchangeable thermistor probe. The Therma 22 Plus operates with either a interchangeable thermistor or type T thermocouple probe. These instruments are designed to be compatible with ThermoWorks manufactured probes.
BATTERY REPLACEMENT - The instrument will display a low battery symbol to warn that the batteries will need changing soon. The instrument continues to measure accurately but after further usage the instrument will display 'flat bat' and shutdown. Replace the three AAA batteries located under the rear case half that is held in place by six screws. Check the seal for damage and carefully replace the rear case half and re-fit and tighten the screws.
MAX/MIN - Press the MAX/MIN button to display the values. The instrument retains the max/min values when switched off. To reset, press the MAX/MIN button for three seconds, 'rst' will be displayed.
BACKLIGHT - Press the HOLD button for two seconds. Please Note: the backlight will not operate if the low battery symbol is being displayed. (Backlight timeout = 20 seconds).
Loading...
+ 2 hidden pages