Thermo Scientific Revco RDE30086FD, RDE30086FA, RDE40086FA, RDE40086FD, RDE50086FA Data Sheet

...
Thermo Scientific Revco RDE Series Ultra-Low Temperature Freezers
Thermo Scientific™ Revco™ RDE Series Ultra-Low Temperature Freezers
• H-drive hydrocarbon refrigeration system provides sustainable performance with reliability you can count on
• Low peak variation helps ensure sample temperature protection throughout entire cabinet
• Fast door opening recovery in as little as 11 minutes
Long warm-up time from -80°C to -50°C in up to 5 hours
• Four capacities: 300, 400, 500 and 600 2-inch boxes
• Decrease storage footprint up to 25%
• 5-year standard warranty
Designed to keep your samples safe
• Worry-free storage; with low peak variation, samples are protected throughout the entire cabinet.
• Backed by our Global Service and Support Team
1. Typical freezer data based on internal testing with freezer setpoint at -80°C and ambient
temperature at 20°C. Performance data varies across freezer size and electrical specifications.
2. All data based on internal data unless otherwise noted. Data on file.
1
Limit environmental impact and lower HVAC costs
• Uses as low as 10.7 kWh/day, depending on size
• Less heat emission lowers HVAC costs
• SNAP compliant, natural hydrocarbon refrigerants
• Environmentally-friendly, water­blown foam insulation is compliant with SNAP
• Manufactured in an award-winning, zero-waste-to-landfill facility
User-friendly and quiet
• New HIC touch screen user interface
• Quiet enough to reside directly in the laboratory; at just 50 dBA, the sound is comparable to a home refrigerator.
2
Dependable storage with operational savings and sustainability
User-friendly HIC interface and cabinet
• Easy to use: plug in and go; no setup required
• H-drive Information Center (HIC) combines easy-to-read temperature with simple capacitive touch buttons
• Advanced security features: alarm settings, setpoint security passcode, and padlock­compatible doors
• Egonomic design: single-hand door operation, eye-level display, and shelf-level placement that do not require floor-level loading of racks
• Five inner doors minimize ambient temperature during sample retrieval
Specifications & Ordering Information
Biologics
Molecular Biologics
Get Connected
Upright Freezers (-50°C to -86°C)
General Purpose Model
RDE30086FA
RDE30086FD
RDE40086FA
RDE40086FD
RDE50086FA
RDE50086FD
RDE60086FA
RDE60086FD
*Allowanc e for han dle and h inge, ad d 4.2 in (10.7cm) to cabinet wid th **Doo rway wi dth cle aran ce is 34.5 in (87.6cm) for RDE50086 and RD E60086
Capacity
cu. ft. (L)
14.9
(422)
19.4
(549)
24.1
(682)
28.8
(816)
Cryoboxes
(2 in.)
300
400
500
600
Electrical
115V/60Hz 12.6/20 (NEMA 5-20)
208-230V/60Hz 5.6/15 (NEMA 6-15)
115V/60Hz 13.0/20 (NEMA 5-20)
208-230V/60Hz 5.9/15 (NEMA 6-15)
115V/60Hz 12.5/20 (NEMA 5-20)
208-230V/60Hz 5.0/15 (NEMA 6-15)
115V/60Hz 12.7/20 (NEMA 5-20)
208-230V/60Hz 6.2/15 (NEMA 6-15)
Amps / Breaker
(Plug t ype)
Performance Data
Model Series
RDE30086
RDE40086
RDE50086
RDE60086
Energy Consumption*
(kW-hr/day)
10.7 +6.9/-3.4 11 248 52
12.5 +5.2/-3.7 16 257 51
12.9 +4.0/-2.9 20 260 50
13.8 +3.9/-3.1 24 297 50
Max Shelf
Weight lbs.
(kg)
125
(57)
165
(75)
205
(93)
245
(111)
Interior Dimensions
H x W x D in. (cm)
51.2 x 17.8 x 28.3
(130.1 x 45.2 x 71.9)
51.2 x 23.1 x 28.3
(130.1 x 58.8 x 71.9)
51.2 x 28.8 x 28.3
(130.1 x 73.0 x 71.9)
51.2 x 34.4 x 28.3
(130.1 x 87.3 x 71.9)
Peak Variation
from - 80°C
Setpoint (°C)**
Exterior Dimensions
H x W* x D in. (cm)
78.0 x 23.0 x 38.5
(198.1 x 58.4 x 97.7)
78.0 x 28.3 x 38.5
(198.1 x 71.9 x 97.7)
78.0 x 34.0 x 38.5**
(198.1 x 86.4 x 97.7)
78.0 x 39.6 x 38.5**
(198.1 x 100.6 x 97.7)
60 Second Door
Opening
Recovery
Time (minutes)
Warm-Up Time (-80°C to -5 0°C)
(minutes)***
Shipping
Weight
lbs. (kg)
712
(323)
687
(312)
734
(333)
820
(372)
Noise
Level
(dBA)
*Energ y Consumptio n: Typical data - inter nal test ing with freeze r setpo int at -80°C and 20° C ambie nt temperatur e **Peak Va riatio n from Se tpoint: Typical dat a - internal testing with freeze r setpoint at -80 °C and 20°C ambie nt temperatur e ***Warm Up Tim e: Typical fr eezer d ata base d on inte rnal testing wi th empt y freezer setpoint at -8 0°C and a mbient tempe rature a t 20°C