Thermo Scientific Orion A324, A324 Instruction Manual

English
pH /
Thermo Scientific Orion Star A324 Portable pH/ISE Meter
ISE
Instruction Sheet
Preparation
Power Source
power adapter.
d. Connect the power adapter to the meter and power outlet.
2. Batteries (included with meter) a. Select four AA alkaline batteries. b. Confirm that the meter is powered off. c. Remove the battery compartment cover – loosen the
screws holding the battery cover, release the top portion of the battery cover from the meter (use a coin or your finger) and release the bottom portion of the battery cover.
d. Orientate the batteries as shown in the battery
compartment housing and insert batteries.
e. Replace the battery compartment cover and screws.
Electrodes and Other Connections
DO/ RDO
ATC/ COND
Orion Star A324 Portable
Use coin or finger to lift.
Remove screws
REF
USB
POWER
1. Prepare the pH electrode or ion selective electrode (ISE) and any other applicable electrodes according to the directions in the electrode user guide.
2. Connect the appropriate items as labeled on the meter and as shown in the figure above:
For additional information on meter setup and operation, refer to the reference guide. The reference guide is on the included CD and available at
www.thermoscientific.com/water.
BNC
RS232
Orion Star A324 Portable pH/ISE Meters
Display Information
Display Icon Description
Shown when the meter is running on AC power.
Shown when the meter has batteries installed.
Indicates data is being sent to a computer or printer.
Indicates data is being sent to the data log.
Shown when an alarm is set and the alarm value is reached.
Indicates the meter is set to be interfaced with a printer or computer via the RS232 port.
Indicates the meter is set to be interfaced with a printer or computer via the USB port.
Displays the time and date entered in the setup menu.
C
25.0˚
ATC
HOLD
M
100
pH
ready
7.000 pH
0.0mV
BUFFERS: 1.68, …
STANDARDS:
0.10 1.0 10
cal
view log
setup
Displays the current temperature based on the temperature probe reading or entered temperature value. Shows the origin of the temperature as MAN (entered temperature) or ATC (temperature probe).
Shown when is pressed and the displayed measurement is frozen.
Indicates a calibration was successfully completed.
Indicates the pH or ion selective electrode condition as good (two bars), fair (one bar) or bad (slash through it), based on the last saved calibration and measurement stability.
Indicates a method is in use and the number of the method being used.
Indicates the type of measurement and determines the type of calibration that will be performed.
Specifies the stability of the electrode as stabilizing or ready.
Shown when the meter is in AUTO-READ mode. The icon will blink while the reading is stabilizing and stop blinking when the reading is stable and the measurement is locked on the display.
Displays the measurement value based on the last saved calibration and current electrode reading. Units are shown to the right of the value.
Shows the raw millivolt reading of the electrode.
Shows the buffer values used for the last saved calibration. pH mode only.
Shows the standard values used for the last saved calibration. ISE mode only.
Shows the operator assigned sample ID number.
Shows the operator assigned user ID number.
Displays the action that will be performed when f1 is pressed.
Displays the action that will be performed when f2 is pressed.
Displays the action that will be performed when f3 is pressed.
2
Orion Star A324 Portable pH/ISE Meters
Keypad Display Information
Press the f1, f2 and f3 function keys to perform the action shown above each key on the display.
Press to turn the meter on.
When the meter is on, press and quickly release to turn the display backlight on or off or hold down to turn the meter off.
In the measurement mode, press to take a measurement.
In the setup, calibration and other menus, press to escape the current menu and return to the measurement mode.
In the measurement mode, press to enter the setup menu.
In the setup, calibration and other menus, press to scroll up through a list of options.
In the continuous measurement mode, press to freeze the displayed measurement and press again to unfreeze the measurement .
In the setup, calibration and other menus, press to scroll left through a lis t of options.
In the measurement mode, press to change the displayed measurement mode. Options are pH, mV, RmV (relative mV ), ORP and ISE.
In the setup, calibration and other menus, press to scroll right through a list of options.
In the measurement mode, press to log or print a measurement.
In the setup, calibr ation and other menus, press to scr oll down through a lis t of options.
Keypad
1. Press to power the meter on. When the meter is on, press and quickly release to turn the
backlight on or off or press and hold
2. Press
to exit any meter function and return to the measurement mode.
3. The f1, f2, and f3 function keys perform a variety of meter operations. The menu-specific operation is
shown above each key. For example, press f1 in the measurement mode to start a calibration.
4. The
, , or keys are used as navigation keys (up, right, down, left) when selecting from a fixed list or grid of meter options. In the measurement mode, these keys are used to access the setup menu, change the measurement mode, manually log or print a measurement and hold (freeze) a displayed measurement.
for about three seconds to power the meter off.