
INSTRUCTIONS
Pub. No. MAN0015751
Pub. Part No. 2162579.1
Rev A.0
Zeba Spin Desalting C olumns and Plates,
7K MWCO
Number Description
89877 Zeba Spin Desalting Columns, Micro (75µL), 25 columns, for 2-12µL samples
89878 Zeba Spin Desalting Columns, Micro (75µL), 50 columns, for 2-12µL samples
89882 Zeba Spin Desal ting Columns, 0.5mL, 25 columns, for 30-130µL samples
89883 Zeba Spin Desalting Columns, 0.5mL, 50 columns, for 30-130µL samples
89889 Zeba Spin Desalting Columns, 2mL, 5 columns, for 200-700µL samples
89890 Zeba Spin Desalting Columns, 2mL, 25 columns, for 200-700µL samples
89891 Zeba Spin Desalting Columns, 5mL, 5 columns, for 500-2000µL samples
89892 Zeba Spin Desalting Columns, 5mL, 25 columns, for 500-2000µL samples
89893 Zeba Spin Desalting Columns, 10mL, 5 columns, for 700-4000µL samples
89894 Zeba Spin Desalting Columns, 10mL, 25 columns,
89807 Zeba Spin Desalting Plates, 96-well, 2 plates
89808 Zeba Spin Desalting Plates, 96-well, 4 plates
Note: These products are recommended for processing compounds > 7000Da. The resin slurry is
Storage: Upon receipt store at 4°C. Product shipped at ambient temperature.
Introduction
The Thermo Scientific™ Zeba™ Spin Desalting Columns contain a high-performance resin that offers exceptional desalting
or buffer-exchange for protei n s amples. Samples volumes between 2µL-4mL and containing as low as 20µg of protein/mL
can be processed with unsurpassed protein recovery and ≥ 95% retention of salts and other small molecules (< 1000Da).
These columns require no chromatography system. The s pin-column method eli mina te s waiting for samples to emerge by
gravity flow and the subsequent monitoring o f fractions for prot ein recovery.
Important Product Information
• The Zeba Spin Desalting Columns and Plates contain a size-exclusion chromatographic resin to separate proteins from
smal l molecules. As wit h all size exclusion-based separation, the amount of small molecule removed and protein
recovered are affected by the nature of the molecules and volume of sample. The sample volumes recommended provide
exceptional removal of a variety of small molecules (typically > 95% for molecules < 1000Da); however, proteins and
small molecules often behave differently than predicted because of a variety of factors such as hydrophobicity,
secondary structure and interactions. Therefore, some opti mization of sample volume might be required to achieve
optimal performance for each specific sample. In general, reducing the sample volume added to the column increases
small molecule removal, and increasing sample volume maximizes protein recovery.
for 700-4000µL samples
supplied in 0.05% sodium azide.
• Also available are Zeba Spin Desalting Columns and Plates with a 40K MWCO, which ena ble removal of salts and other
small molecules < 2000Da and recovery of proteins and other macromolecules > 40,000Da.

Table 1. Centrifugation times and volumes for the buffer, stacker and sample.
Wash/equilibration Buffer Volume
Optional Stacker Volume (µL)*
Storage Sol ution
Removal
*When using the indicated sample volumes, use a stacker to achieve the highest recovery. The stacker is a volume
of wash/equilibration buffer applied after the added sample has completely entered the desalting resin bed.
Procedure for Desalting or Buffer Exchange
Additional Materials Required
• For 75µL and 0.5mL spin columns: Bench-top microcentrifuge (1500 × g) and 1.5mL microcentrifuge tubes
Note: Use a centrifuge that can be adjusted to 1500 × g, such as the Thermo Scientific™ Sorvall™ Legend Micro 17
Microcentrifuge
• For 2 and 5mL spin columns: Centrifuge (1000 × g) and 15mL conical tubes
• For 10mL spin columns: Centrifuge (1000 × g) and 50mL conical tubes
• For the desalting plates: Varia ble-speed centrifuge with rotor and carrier capable of handling stacked plates
(height = 4.4cm) at 1000 × g
• Wash/equilibration buffer
Note: Use the same wash/equilibration (stacker) buffer as is desired for the final sample solution. Equilibrating the
desalting resin before sample loading is necessary to ensure proper buffer exchange.
Procedure for Protein Desalting
Note: See Table 1 for centrifugat ion times and volumes for the buffer, stacker and sample for each device. For maximum
protein recovery, add a stacker on top of the applied sample for volumes below the volume specified in Table 1.
1. Remove the column’s bottom closure or the plate’s bottom sealing material. Loosen cap (do not remove cap).
2. Place the column into a collection tube or plate on top of a wash plate and centrifuge to remove the storage solution.
3. Discard flow-through and replace the column back into the collection device.
4. Add wash /equilibration buffer on top of the resin. Centrifuge device and discard flow-through. Repeat this step two
additional times.
Note: After each spin, the resin should appear white and free of liquid. If liquid is present, make sure you are using the
correct centrifugation speed and time. Incomplete centrifugation may result in poor sample recovery or sample dilution.
5. Blot the bottom of the column or plate to remove excess liquid. Transfer device to a new collection tube or plate.
6. Apply sample on top of the resin. If needed, add a stacker as soon as the sample has entered the resin. Adding a stacker is
optional but recommended for dilute pr otein solutions or small sample volumes to ensure maximum sample recovery.
7. Centrifuge and retain flow-throug h that contains sample. Discard spin column or plate.
2-12 30-130 200-700 500-2000 700-4000 20-100
3 15 40 100 200 10
1000 1500 1000 1000 1000 1000
1 1 2 2 2 2
Centrifugation
Time (min)

Sample or buffer does not
Ensure that c entrifuge is in prop er working co ndition
Ensure bottom closure i s removed
Ensure top cap is loosened
Improper sample loading
Apply sample directly to center of the resin bed; touch tip
Avoid contact with sides of column
Improper centrifugation
For fixed-angle rotors, place colu mn in the same orientation
Do not exceed recommended centrifugation speed or time
Sample was not completely in
Portion of protein still
Use a stacker to recover more protein (for most samples,
Check for protein solubility in the final b uffer or solution
Protein bound to resin matrix
Use alternative resin, such as Polyacrylamide Spin
Wash/equilibration buffer was
Before adding the sample make sure the wash/equilibration
Troubleshooting
flow through resin
solution before adding to the
column
to resin to expel all sample
each time and do not exceed recommended centrifuge speed
Centrifuge sample at 14,000 × g for 10 minutes before
adding to the column
remaining in spin column
the majority of protein is recovered without a stacker)
equilibration buffer
Desalting Colu mns (see Related Thermo Scientific Products
section)
sample is dilute
not adequately removed
buffer was adequately removed by centrifugation (i.e.,
column appears uniformly white with no solvent streaks)
Related Thermo Scientific Products
89849 Pierce Polyacrylamide Spin Desalting Columns, 7K MWCO, 0.7mL, 25 columns
89862 Pierce Polyacrylamide Spin Desalting Columns, 7K MWCO, 0.7mL, 50 columns
Products are warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale, as set forth in
the Product documentation, specifications and/or accompanying pa ckage inserts (“Document ation”). No claim of suitability for use in applica tions regulated
by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the Documentation, this
warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This warranty does not extend to
anyone other than Buyer. Any model or sample furnished to Buyer is merely illustrative of the general type and quali ty of goods and does not represent that
any Product will conform to such model or sample.
NO OTHER WARRANTIES, EXPRESS OR IMPLIED, ARE GRANTED, INCLUDING WITHOUT LIMITATION, IMPLIED W ARRANTIES OF
MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE, OR NON INFRINGEMENT. BUYE R’S EXCLUSIVE REMEDY FOR NONCONFORMING PRODUCTS DURING THE WARRANTY PERIOD IS LIMITED TO REPAIR, REPLACEMENT OF OR REFUND FOR THE NONCONFORMING PRODUCT(S) AT SELLER’S SOLE OP T ION. THERE IS NO OBLIGATION TO REPAIR, REPLACE OR REFUND FOR PRODUCTS
AS THE RESULT OF (I) ACCIDENT, DISASTER OR EVE NT OF FORCE MAJEURE, (II) MISUSE, FAULT OR NEGLIGENCE OF OR B Y BUYER,
(III) USE OF THE PRODUCTS IN A MANNER FOR WHICH THEY WERE NOT DESIGNED, OR (IV) IMPROPER STORAGE AND HANDLING OF
THE PRODUCTS.
Unless otherwise expressly stated on the Product or in the documentation accompanying the Product, the Product is intended for research only and is not to
be used for any other purpose, including without limitation, unauth or ized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeuti c uses, or
any type of consumption by or application to humans or animals.
Current product instructions are available at thermofisher.com. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2016 Thermo Fisher Scientific Inc. All rights reserved. Unless ot herwise indicated, all trademarks are the property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.