Thermo Scientific Wellwash User Manual

Page 1
Thermo Scientific
Wellwash
User Manual
Rev. 1.1, Cat. No. N15777
Page 2
Page 3
Page 4
Page 5
Page 6
Page 7
Page 8
Page 9
or
Page 10
Page 11
Page 12
Page 13
A
B
or
Page 14
Venting holes Intake tube with liquid filter
Sensor cable
Page 15
Float
Page 16
Page 17
Page 18
Power supply connector
USB port for service
USB port for memory
device
Mains (I/0) switch
Page 19
Page 20
Page 21
Page 22
Wash head on wash head arm
Priming vessel
Keypad
Display
Plate carrier
Transparent aerosol cover
Liquid bottles
Plate clamp
Bottle stand (option)
Power supply connector
USB port for service
USB port for memory
device
ON/OFF switch
Page 23
Connectors for liquid sensor cables, from top to bottom:
W
Connectors for liquid container tubes, from left to right:
W
ON/OFF (I/0) switch
USB port for memory device
Left
Right
Dispense valve
Dispense pump
Aspirate pump
Wash head
Liquid level sensor
Liquid filter
Buffer container
Waste container
Liquid level sensor
Page 24
Speed designation
Speed
Low
5 Hz, amplitude 2.5 mm
Medium
10 Hz, amplitude 1.5 mm
High
15 Hz, amplitude 1 mm
1x8
1x12
Page 25
Use the Left, Right, Up and Down arrow keys to navigate. You can
Press the OK button to select, edit, or accept the highlighted item.
Use the F1-F3 keys to select the corresponding action from the info text
Press the FILE key, for example, to save the active protocol in the Main
Use the HELP key for more detailed instructions.
Press the PRIME button to prime the instrument.
PRIME, START, and
Arrow keys and
Menu bar
Active row blue
Info text bar
Function keys
File and Help keys
Numeric keys with letters
Page 26
Press the START button to start the execution of the currently selected
Press the STOP button to terminate protocol execution.
Use the character keys to enter numerical data and text.
Main
Settings
Maintenance
Protocol
Prime parameters
Clean
Head / Plate
Sensors
Calibration
Well
Wash head configuration
Service
Steps
System
Layout
Reports
Page 27
*
An asterisk in front of a protocol name means that the protocol has not yet been saved.
A yellow caution symbol on the Head / Plate row means that the current protocol has a
Info text bar
Page 28
1. Ensure that the buffer solution in the buffer bottle A is correct and
2. Select a protocol. Refer to “Selecting a protocol”.
3. Ensure that the protocol parameters are correct.
4. Insert the 96-well plate to be washed onto the plate carrier and
5. Press the PRIME button to prime the instrument. Refer to “Priming
6. Select the strips if they are not already selected. Refer to “Strip
7. Press the START button.
8. The microplate is processed according to the predefined protocol.
9. Check that the liquid channels work properly during dispensing and
10. If you need to abort the run, press the STOP button.
11. Remove the plate after the protocol has ended.
12. Prime the instrument with distilled deionized water at the end of the
Page 29
Use the F1-F3 function keys reserved for ready-made, demo or favorite
F1-F3 function keys
Page 30
2. Select the ready-made protocol from the protocol list and press the OK
1x8
96-well plate
Landscape
1x12
96-well plate
Portrait
Example protocol list
Page 31
A1
Page 32
Press the PRIME button to fill the liquid tubing completely from the buffer bottle intake
1. To select the columns/strips of the plate, use the number keys.
Bottle to wash head 19 ml
Page 33
Go to the Layout row in the Main menu and press the OK button.
Select/unselect columns/strips with the OK button and move the cursor using
Page 34
1. Press the START button to start the ru n.
2. Remove the plate after the protocol has ended.
Press the STOP button to cancel a run.
Page 35
Main
Settings
Maintenance
Protocol
Prime parameters
Clean
Head / Plate
Sensors
Calibration
Well
Wash head configuration
Service
Steps
System
Layout
Reports
*
An asterisk in front of a protocol name means that the protocol has not yet been saved.
A yellow caution symbol on the Head / Plate row means that the current protocol has a
Info text bar
Page 36
It is possible to protect protocols from accidental editing or deleting. A locked symbol is
1. Press the FILE key in the Main menu to create a new protocol. Refer to
2. Select New and press the OK button.
3. Select the Head / Plate row. The default wash head shown is according to
Page 37
4. Select the wash head.
5. Select Well. The Well type window opens. Refer to “Well”.
6. Select the well bottom shape of the plate or strips.
7. Select Well offset to adjust the well offset value. For more information on
8. Select the Steps row in the Main menu. Refer to “Steps”.
Page 38
9. Press the F3 (New step) key to open the step list box.
10. Select the step.
11. Press the FILE key in the Main menu. Note that the strip selection is not
12. Select Save As. The Save Protocol As dialog opens.
13. Enter the protocol name, for example, Test1, by using the character keys.
Page 39
1. Select the Protocol row in the Main menu and press the OK button.
2. Select the Create a new protocol row and press the OK button.
3. Enter a name for the protocol and press the OK button.
Page 40
4. In the Main menu, set the desired protocol parameters and steps.
5. When ready, press the FILE key and select Save to save your protocol.
1. Select the Steps row in the Main menu and press the OK button.
2. Press the F3 (New step) key to open the step list box.
3. Select the new step and press the OK button.
4. A flashing insert bar with the selected step appears. Select the position
Page 41
5. Press the OK button to edit the new step. For more information on steps,
6. Save the changes you make in each step using the FileSave functions.
7. Press the F3 (New step) key to add further new steps to the protocol.
8. Use the Up and Down arrow keys to navigate in the protocol.
1. Select the step you want to delete and press the F1 (Delete) key.
2. Press the OK button to confirm the deletion.
Page 42
When you press the OK button, the wash head/plate list box appears.
Select the wash head and press the OK button. Refer to “Wash head configuration” and
Page 43
When you press the OK button, the Well type window appears.
Use the Up and Down arrow keys to adjust the value.
Press the F3 (Preview) key to move the plate carrier and align the wash
Press the OK button to accept the value and/or the F2 key to close the
Flat
C
U
V
Star A
Star B
Page 44
Bottom
Default
height
Thermo Scientific™
Max. overall
Flat
Star
3.1 mm
3.6 mm
439454 Nunc™ solid F96
441653 Nunc starwell strip plate
14.5 mm ± 0.2 mm
Aspirating
Aspirating
Dispensing
Dispensing
Page 45
Select the height parameter and press the OK button to activate the height
Press the F3 (Preview) key to move the plate carrier and align the wash
The wash parameters are grouped into two windows. Use the Right and
Dispense height end
Dispense height start
Well bottom
Plate bottom 0 mm
Page 46
Plate / Well shape
Default aspirate height
96 / Flat
2.6 mm
96 / C
2.9 mm
96 / U
4.2 mm
96 / V
5.4 mm
96 / Star
3.1 mm
Page 47
Page 48
Page 49
Page 50
Page 51
Page 52
Page 53
Select the Sensors row in the Settings menu and press the OK button.
Page 54
Select the Wash head configuration row in the Settings menu and press the OK
Page 55
Select the System row in the Settings menu and press the OK button.
Page 56
Page 57
2. Empty all bottles and fill the buffer bottle with cleaning reagent.
3. Select the soak time using the Right arrow key and number keys.
4. Press the F2 key to abort the soaking if you want a shorter soak time
Page 58
5. Empty all bottles and fill the buffer bottle with deionized distilled
6. Empty all bottles and press the OK button.
7. Press the OK button when the procedure is ready.
Page 59
1. Select the Clean wash head row and press the OK button.
2. Change the wash buffer to cleaning agent or deionized distilled
3. Select the soak time using the Right arrow key and number keys.
4. Press the F2 key to stop the soaking if you want a shorter soak time
Page 60
5. Press the OK button when the procedure is ready.
2. Change the wash buffer to cleaning agent or deionized distilled
Page 61
3. Press the F2 key to abort the soaking and to continue the
4. Press the OK button when the procedure is ready.
2. Fill the priming vessel manually if needed. Press the OK button.
Page 62
3. Press the F2 key to end the procedure.
2. Press the OK button to empty the priming vessel.
3. Press the OK button when the procedure is done.
Page 63
Page 64
Select the Calibration row in the Maintenance tab and press the OK button.
1. Select the Calibrate current wash head row and press the OK
Page 65
2. Press the OK button to start the calibration. The dispense pump
3. Insert the first tare weighed plate onto the plate carrier and press the
4. Weigh the dispensed plate and enter the mass of the dispensed
5. Insert the second tare weighed plate onto the plate carrier and press
Page 66
6. Weigh the dispensed plate and enter the mass of the dispensed
7. Press the F1 key to continue calibration.
8. Continue dispensing and weighing the third and fourth plate in the
Page 67
9. Press the F1 key to continue calibration.
10. Calibration is now complete. Press the F1 key to accept the new
11. Press the OK button when the procedure is done.
12. Verify the new performance by dispensing a plate, for example, with
Page 68
Page 69
1. Insert the USB memory device (Figure 35).
2. Go to the Protocol row in the Main menu and press the OK button. The list of
3. Select the protocol you want to export.
4. Press the FILE key to open the File menu.
Page 70
5. Select Export and press the OK button.
1. Insert the USB memory device (Figure 35).
2. Go to the Protocol row in the Main menu and press the OK button. The list of
3. Press the FILE key to open the File menu.
Page 71
4. Select Import and press the OK button.
5. Select the protocol you want to import and press the OK button. You can import
1. Insert the USB memory device (Figure 35).
2. Go to the Protocol row in the Main menu and press the OK button. The list of
3. Select the protocol you want to export.
Page 72
4. Press the FILE key to open the File menu.
5. Select ASCII Export and press the OK button.
2. Select the protocol you want to delete.
3. Press the FILE key to open the File menu.
Page 73
4. Select Delete and press the OK button.
5. Select Yes and press the OK button to confirm the deletion.
Page 74
Page 75
Perform the operational check. See “Performing the operational
Prime the instrument with deionized distilled water after use. See
Keep the instrument free of dust. See “Instrument care”.
Wipe away spilled saline solutions, solvents, acids or alkaline
If any surfaces have been contaminated with biohazardous material,
Clean the case of the instrument periodically. See “Cleaning of the
Clean the plate carrier if needed. See “Cleaning the plate carrier".
Check the operation of the plate clamp.
Clean the priming vessel if needed. See “Cleaning the priming
Clean the liquid bottles regularly. See “Cleaning liquid bottles”.
Clean or replace the liquid filter (Figure 37) of the intake tube if
Check the performance of the liquid sensors. See “Checking the
Clean the wash head daily and replace if needed. See “Cleaning
Check the dispensing accuracy and calibrate if needed. See
 
Check the aspirate height. See “Checking the aspirate height”.
Replace the tubings if needed. Contact service.
Ensure proper shutdown. See Chapter 9: “Shutdown”.
Decontaminate the instrument when relocating the ins trument or
Service the instrument regularly. See “Cleaning of the instrument”
Page 76
Page 77
Page 78
Cannot be
Can be
Liquid filter
Sensor cable Intake tube
Nut
Page 79
Autoclavable
Un-autoclavable
Liquid bottles
Aerosol cover
Liquid bottle caps
Liquid filter
Intake tubes
Liquid level sensor assemblies
Priming vessel
Sensor nut
Tube connectors
1x8 wash heads
1x12 wash heads
Page 80
Page 81
1. Fill the buffer bottle completely with liquid and close the cap. Place the bottle at the
2. Disconnect the liquid bottle tube from the tube fitting on the instrument.
3. Attach a 20 ml plastic syringe (Cat. No. N02942, supplied with the instrument) to the
4. Gently aspirate liquid from the liquid bottle to fill the tube by “filling” the syringe. Leave
5. Close (pinch) the tube, continue pinching the tube and then disconnect the tube from
6. Reattach the tube to the fitting on the instrument side and release your grip.
7. Press the PRIME button on the instrument keypad and prime the channel.
Page 82
Page 83
Page 84
Page 85
Pollution degree
2 (see “Safety specifications”)
Method of disposal
Electronic waste
Contaminated waste
(Infectious waste)
Page 86
Page 87
Altitude
Up to 2000 m
Temperature
+5°C to +40°C
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C
Mains supply fluctuations
±10% from nominal
Installation category (overvoltage category)
Pollution degree
2 according to IEC 60664-1 (see Note 2)
Page 88
2
Computer command not recognized.
Check the syntax.
3
Invalid computer command argument.
Check the syntax.
4
Plate position error.
Check the carrier movement. Contact authorized technical service.
5
Head position error.
Check the head movement. Contact authorized technical service.
6
Dispense pump rotation error.
Pump is jammed. Contact authorized technical service.
7
Dispense pump time out.
Pump is jammed. Contact authorized technical service.
8
Plate was not detected.
Insert the plate properly.
9
Plate sensor does not wor k.
Contact authorized technical service.
10
Non-volatile parameters lost.
Contact authorized technical service.
11
Attempt to reset the serial number.
12
No more memory for storing user data.
Memory is full, delete unused protocols.
13
Error(s) during startup.
Contact authorized technical service.
20
USBwiz error.
Contact authorized technical service.
21
No firmware on the USBwiz chip.
Contact authorized technical service.
22
XY offset adjustment failed.
23
More than one command has the same hash value.
– 25
Parameter memory not found.
Contact authorized technical service.
26
Parameter memory erase failure.
Contact authorized technical service.
27
Parameter memory write failure.
Contact authorized technical service.
50
File open error. File not found.
53
The file already exists.
Use a different file name.
54
The file does not exist.
55
The media is full.
Delete unused protocols.
56
End of file.
57
Other file error (none of the previous).
58
Firmware update aborted by user.
59
File read failed.
60
File not found.
61
USB memory stick not present.
Insert a USB memory stick.
62
File write failed.
70
Not enough memory.
Delete unused protocols.
71
Adding the step not successful.
80
Invalid step parameter(s).
81
Too many step parameters.
82
At least one of the step parameters is missing.
83
Step list is not empty.
84
Creating a new step list failed.
85
No strips are selected.
Select strips.
86
Invalid SOF parameter.
91
USB device timeout.
Try with another USB memory stick.
Page 89
Dispensing volume is
Tubing is old or clogged.
position.
Use Maintenance > Clean > Soak wash head.
Define the correct aspirate height.
Dispensing volume is
Filter is blocked.
Calibrate the wash head.
service.
Volume is too low.
Tips are deformed. Leakage in tubing.
Change the wash head. Clean or replace the tubing.
Wrong volume is
Dispensing valve is leaking
high or too low.
Component failure. Contact authorized technical
Wrong volume is
Aspiration tube is not
Overpressure in the bottle.
Connect the aspiration tube to the wash head. Do not cover the venting holes in the bottle caps.
Page 90
Wrong parameters are used.
Check the parameters.
Plate is processed
Reagents are mixed.
A non-active valve is leaking.
Component failure. Contact authorized technical
Keys of the keypad do not function.
Keypad is broken.
Contact authorized technical service.
Display does not show
anything at all.
Display is broken.
Contact authorized technical service.
Liquid is dispensed
The plate orientation is
incorrect.
Check the layout and place the plate in the correct
wash head required by the layout and the protocol.
Liquid is spilled during
Well volume or shake speed
Check the tubings and O-rings, and replace if
Decrease the well volume or shake speed.
Priming failed.
Tubings incorrectly installed.
liquid bottles.
Check the tubings.
Tips are dripping.
Air leak.
Valve is leaking.
Check that the tubes and wash head plugs are
Contact authorized technical service.
5165000
Wellwash 1x8 100240V 50/60Hz
5165040
Wellwash 1x12 100240 V 50/6 0H z
Page 91
N15777
Wellwash User Manual, CD
1
N11165
Wellwash and Wellwash Versa Quick Reference Guide
1
N09541
Priming vessel
1
N10800
1x8 wash head
1
N10801
1x12 wash head
1
N10805
2 liter wash bottle, Wellwash (incl. cap with tubing and s ens or)
1
N10806
2 liter waste bottle, Wellwash (incl. cap with tubing and sensor)
1
N10821
Spare bottle 2 l with solid cap
1
N10811
4 liter wash bottle A, round (incl. cap with tubing and sensor)
1
N10810
4 liter waste bottle, square (incl. cap with tubing and sensor)
1
N10822
Spare bottle 4 l, round with solid cap
N10814
9 liter waste bottle, rectangular (incl. cap with tubing and sensor)
1
N10815
Aerosol cover, Wellwash
1
N10817
Bottle stand 1x2 configuration (2x2 l bottles), Wellwash
1
N10820
Bottle stand for own bottle
1
N12785
Spare cap for storage bottle, 2 pcs
1
N04001
USB A-B device cable 1.8 m *
1
N11116
Accessories Kit Incl. declogging tools normal/thin, syringe 20 ml, and silicon grease for O-rings
1
N12409
Wash head tubing connector Kit, 2 pcs each Incl. dispensing and aspiration tubing connectors (see Figure 44)
1 N12407
Liquid filter, 5 pcs
1
2305290
Serial cable F9/F25
1
SP-00245
PM Kit Wellwash ( incl. bott l e and wash hea d tubi ng, va lve tu bing, liqui d filt er, wash head plugs, O-rings, track slider, and cable ties)
1
Page 92
Page 93
A) I confirm that the returned items have not been contaminated by body fluids, toxic,
B) I confirm that the returned items have been decontaminated and can be handled without
Page 94
decontamination
Removal or neutralization of radiologic, bacteriological, chemical or other
disinfection
The destruction of pathogenic bacteria, usually with an antiseptic chemical
EIA
Enzyme immunoassay. An immunoassay using a color-changing enzyme-
ELISA
Abbreviation for enzyme-linked immunosorbent assay.
remote control
Running mode allowing a remote computer to operate the washer.
wash head
Interchangeable heads allowing either 1x8 or 1x12 processing.
USB
Universal serial bus.
Loading...