Thermo Scientific SL16R, Thermo Scientific SL16 Instruction Manual

Thermo Fisher Scientific
Thermo Scientific SL16 Thermo Scientific SL16R
Instruction Manual
50120986-1 June 2009
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
BIOLiner
TM
, BIOShieldTM, HIGHPlateTM, HIGHConicTM, AutoLockTM, FIBERLite TM and CLINIConic
TM
re either registered trademarks or trademarks of Thermo Fisher Scientific..
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
Release history: 50120986-1 printed in June 2009.
For Research Use Only. Not for use in diagnostic procedures.
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96 It is marked by the symbol opposite:
Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU Member States for the recycling and disposal of this device. For information on recycling and disposal companies in Germany and on the products of Thermo Fisher Scientific, which fall under the RoHS Directive (Restriction of the use of certain hazardous substances in electrical and electronic equip­ment), please visit the website www.thermo.com/WEEERoHS.
Table of Contents
Preface .......................................................................................................................................iii
Scope of supply ........................................................................................................................iii
Proper use .................................................................................................................................iii
Accident Prevention ................................................................................................................ iv
Precautions ............................................................................................................................... iv
Chapter 1 Introduction and Description ...............................................................................................1-1
Characteristics ..........................................................................................................1-2
Technical Data .........................................................................................................1-3
Directives, standards and guidelines ......................................................................................... 1-4
Functions and features ............................................................................................................. 1-5
Mains Supply ...........................................................................................................1-5
Rotor selection .........................................................................................................1-6
T
Chapter 2 Before use ................................................................................................................................2-1
Before setting up ......................................................................................................2-2
Transporting the centrifuge ......................................................................................2-2
Location ...................................................................................................................2-2
Aligning the centrifuge .............................................................................................2-3
Mains connection .....................................................................................................2-3
Storage .....................................................................................................................2-4
Chapter 3 Control panel ...........................................................................................................................3-1
Keys .........................................................................................................................3-2
Control panel ...........................................................................................................3-2
Chapter 4 Operation .................................................................................................................................4-1
Switch on centrifuge .................................................................................................4-2
Lid Opening .............................................................................................................4-2
Close lid ...................................................................................................................4-2
Rotor installation ......................................................................................................4-2
Entering parameters ..................................................................................................4-3
Saving programs .......................................................................................................4-6
Centrifuging .............................................................................................................4-7
Temperature adaptation during standstill .................................................................4-8
Short-term centrifuging ............................................................................................4-8
Removing the rotor ..................................................................................................4-8
Aerosol-tight rotors ...................................................................................................4-9
Switching the centrifuge off ......................................................................................4-9
Thermo Scientific SL 16 / 16R i
Table of Contents
Chapter 5 System menu ...........................................................................................................................5-1
Chapter 6 Maintenance and care .......................................................................................................... 6-1
Intervals ................................................................................................................... 6-2
Cleaning .................................................................................................................. 6-2
Disinfection ............................................................................................................. 6-3
Disinfection with bleaching agents ........................................................................... 6-4
Decontamination ..................................................................................................... 6-4
Autoclaving .............................................................................................................. 6-4
Service of Thermo Fisher Scientific .......................................................................... 6-4
Chapter 7 Troubleshooting ..................................................................................................................... 7-1
Mechanical emergency lid release ............................................................................. 7-2
Troubleshooting by user .......................................................................................... 7-3
When to contact customer service ............................................................................ 7-6
Chemical Compatibility Chart ..............................................................................A-1
Warranty ...................................................................................................................B-1
Contact Information ................................................................................................C-1
ii SL 16 / 16R Thermo Scientific
Preface
P
Before starting to use the centrifuge, read through this instruction manual carefully and follow the instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it is forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the sellers warranty.
Scope of supply
If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.
Proper use
This centrifuge is an IVD accessory, and therefore subject to the Directive 98/79/EC of the
European Parliament and the Council of October 27, 1998 on in vitro diagnostic medical devices.
Article Number Quantity Check
Centrifuge Thermo Scientific SL16 1
Power supply cable 1
50120986 Instruction manual 1
70009824 Anti-corrosion oil 1
This centrifuge is a laboratory product designed to separate components by generation of Relative
Centrifugal Force. It separates human samples (e.g. blood, urine and other body fluids) collected in appropriate containers, either alone or after addition of reagents or other additives.
As general-purpose centrifuge, it is designed to also run other containers filled with chemicals,
environmental samples and other non-human body samples.
This centrifuge should be operated by trained specialists only.
Thermo Scientific SL 16 / 16R iii
P
Accident Prevention
Accident Prevention
Prerequisite for the safe operation of the Thermo Scientific SL16 / 16R is a work environment in compliance with standards, directives and trade association safety regulations and proper instruction of the user:
The safety regulations contain the following basic recommendations:
Maintain a radius of at least 30 cm around the centrifuge.
Implementation of special measures which ensure that no one can approach the centrifuge for
longer than absolutely necessary while it is running.
The mains plug must be freely accessible at all times. Pull out the power supply plug or
disconnect the power supply in an emergency.
Precautions
In order to ensure safe operation of the Thermo Scientific SL16 / 16R, the following general safety regulations must be followed:
Do not manipulate the safety devices.
The centrifuge should be operated by trained specialists only.
The centrifuge is to be used for its intended purpose only.
Plug the centrifuge only into sockets which have been properly grounded.
Do not move the centrifuge while it is running.
Use only rotors and accessories for this centrifuge which have been approved by Thermo Fisher
Scientific. Exceptions to this rule are commercially available glass or plastic centrifuge tubes, provided they have been approved for the speed or the RCF value of the rotor.
Do not use rotors which show any signs of corrosion and/or cracks.
Do not touch the mechanical components of the rotor and do not make any changes to the
mechanical components.
Use only with rotors which have been properly installed. Follow the instructions on the Autolock
in section “Rotor installation” on page 4-2.
Use only with rotors which have been loaded properly. Follow the instructions given in the rotor
manual.
Never overload the rotor. Follow the instructions given in the rotor manual.
Never start the centrifuge when the lid is open.
Never open the lid until the rotor has come to a complete stop and this has been confirmed in
the display.
iv SL 16 / 16R Thermo Scientific
P
Precautions
The lid emergency release may be used in emergencies only to recover the samples from the
centrifuge, e.g. during a power failure (see section “Mechanical emergency lid release” on
page 7-2).
Never use the centrifuge if parts of its cover panels are damaged or missing.
Do not touch the electronic components of the centrifuge or alter any electronic or mechanical
components.
Please observe the safety instructions.
Please pay particular attention to the following aspects:
Location: well-ventilated environment, set-up on a level and rigid surface with adequate load-bearing
capacity.
Rotor installation: make sure the rotor is locked properly into place before operating the
centrifuge.
Especially when working with corrosive samples (salt solutions, acids, bases), the accessory parts and
vessel have to be cleaned carefully.
Always balance the samples.
Centrifuging hazardous substances:
Do not centrifuge explosive or flammable materials or substances which could react violently with
one another.
The centrifuge is neither inert nor protected against explosion. Never use the centrifuge in an
explosion-prone environment.
Do not centrifuge inflammable substances.
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms without
suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the "Laboratory Bio-safety Manual" of the World Health Organization WHO), aerosol-tight biological seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
If toxins or pathogenic substances have gotten into the centrifuge or its parts, appropriate
disinfection measures have to be taken (see “Disinfection” on page 6-3).
Highly corrosive substances which can cause material damage and impair the mechanical stability
of the rotor, should only be centrifuged in corresponding protective tubes.
IF A HAZARDOUS SITUATION OCCURS, TURN OFF THE POWER SUPPLY TO THE CENTRIFUGE AND LEAVE THE AREA IMMEDIATELY.
Normal use
The centrifuge is to be used for separating materials of different density or particle size suspended in a liquid.
g
Maximum sample density at maximum speed: 1,2
Thermo Scientific SL 16 / 16R v
---------- -
cm
3
P
Precautions
The symbol on the left refers to general hazards. ATTENTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur.
The symbol on the left refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe.
vi SL 16 / 16R Thermo Scientific
Introduction and Description
“Characteristics” on page 1-2
“Technical Data” on page 1-3
“Functions and features” on page 1-5
“Rotor selection” on page 1-6
1
Thermo Scientific SL 16 / 16R 1-1
1
Introduction and Description
Characteristics
Characteristics
The Thermo Scientific SL16 / 16R is an in-vitro diagnostics device according to the In-Vitro Diagnostics Directive 98/79/EC.
Several rotors with commercial tubes can be used.
The set speed is reached in seconds. The maintenance-free induction motor ensures quiet and low-vibration operation even at high speeds, and guarantees a very long lifetime.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, running time, temperature, and running profile (acceleration and braking curves). You can choose between the display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is running.
With the help of the "Pulse"-key , you can also centrifuge a sample for just a few seconds, if called for.
The Thermo Scientific SL16 / 16R is equipped with various safety features:
The housing and rotor chamber consist of steel plate, the interior of armour steel, while the front
panel is made of high-impact resistant plastic.
The lid is equipped with a view port and a lock.
The lid of the centrifuge can only be opened while the centrifuge is switched on and the rotor
has come to a complete stop. The centrifuge cannot be started until the lid has been closed properly.
The integrated rotor detection systems ensures that no inadmissible speed settings can be
preselected.
Electronic Imbalance recognition
Lid emergency release: For emergencies only, e.g. during power failures (see “Mechanical
emergency lid release” on page 7-2
1-2 SL 16 / 16R Thermo Scientific
1
Introduction and Description
Technical Data
The technical data of the Thermo Scientific SL16 / 16R is listed in the following table.
Table 1- 1 . Technical Data Thermo Scientific SL16 / 16R
Feature Value
Environmental conditions - use in interior spaces
- altitudes of up to 2,000 m above sea level
- max. relative humidity 80% up to 31 °C; decreasing linearly up to 50% relative humidity at 40 °C.
permissible ambient temperature +2° C to +35° C +2° C to +35° C
Overvoltage category II II
Pollution degree 2 2
Heat dissipation refrigerated ventilated
230V 120V 230V 120V
4778 BTU/h 4096 BTU/h 3447 BTU/h 3901 BTU/h
Technical Data
IP 20 20
running time unlimited unlimited
maximum speed n
minimum speed n
maximum RCF value at n
maximum kinetic energy <62,5 kJ <51,7 kJ
noise level at maximum speed <63 dB(A) <63 dB(A)
temperature setting range -10 °C to +40 °C
Dimensions refrigerated ventilated
Height Height lid open Width Depth Table top height
weight without rotor 91,5 kg 57,5 kg
max
max
max
15200 rpm 15200 rpm
300 rpm 300 rpm
25701xg 25701xg
360 mm 870 mm 623 mm 605 mm 310 mm
360 mm 870 mm 440 mm 605 mm 310 mm
Thermo Scientific SL 16 / 16R 1-3
1
Introduction and Description
Directives, standards and guidelines
Directives, standards and guidelines
Table 1-2. Directives, standards and guidelines
Tension / Frequency
230V 50/60Hz 2006/95/EC Low Voltage Directive:
2006/42/EC & 98/37/EC Machine Directive:
2004/108/EC EMC Directive
98/79/EC In-vitro-Diagnostika (IvD)
Produced and inspected according to the following standards and guidelines
EN 61010-1, 2
EN 61010-2-020, 2
EN 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011B
EN 61000-6-2
EN ISO 13485
230V 60Hz
120V 60Hz
100V 60Hz
The centrifuges are produced and inspected according to the following standards and guidelines
UL 61010-1, 2
CAN/CSA-C22.2 No. 61010-1, 2
IEC 61010-2-20, 2
(Pollution degree 2, Overvoltage category II)
EN 61010-2-101
Emitted interference FCC Part 15 CLASS A
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
100V 50Hz IEC 61010-1 2
IEC 61010-2-020 2
IEC 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011A
EN 61000-6-2
EN ISO 13485
nd
Edition
nd
Edition
nd
Edition
nd
Edition
nd
Edition
nd
Edition
nd
Edition
1-4 SL 16 / 16R Thermo Scientific
1
Introduction and Description
Functions and features
Functions and features
The following table gives an overview of the important functional and performance characteristics of the Thermo Scientific SL16 / 16R.
Table 1- 3 . Functions and features
Component / Function Description / Features
Structure / Housing Galvanized steel chassis with armoured plating.
Chamber Stainless steel
Drive Induction drive without carbon brushes
Keys and display Easy-to-clean keypad and display surface
Controls Microprocessor-controlled
Internal memory The most recent data is saved
Functions RCF-selection, temperature control, pretemp with cooling equipped devices
Acceleration / braking profiles 9 acceleration and 10 braking curves
Rotor recognition Automatic
Imbalance recognition Electronic, contingent on rotor and speed
Lid lock Automatic lid closing and locking starting from an initial hold position
Mains Supply
The following table contains an overview of the electrical connection data for the Thermo Scientific SL16 / 16R. This data is to be taken into consideration when selecting the mains connection socket.
Table 1- 4 . Electrical connection data of the Thermo Scientific SL16 / 16R
Cat. No.
7500 4030 refrigerated 230 V 50 / 60 Hz 8 A 1400 W 15 AT 16 AT
7500 4031 refrigerated 120 V 60 Hz 12 A 1200 W 15 AT 15 AT
7500 4000 ventilated 230 V 50 / 60 Hz 6 A 1010 W 15 AT 16 AT
7500 4001 ventilated 120 V 60 Hz 9,5 A 850 W 15 AT 15 AT
Mains voltage
Frequency Rated current
Power consumption
Equipment fuse
Building fus e
Thermo Scientific SL 16 / 16R 1-5
1
Introduction and Description
Rotor selection
Rotor selection
The Thermo Scientific SL16 / 16R is supplied without a rotor. Various rotors are available to choose from.
TX-400 7500 3629
with round buckets 7500 3655
TX-200 7500 3658
with round buckets 7500 3659
BIOShield
M-20 7500 3624
Microliter rotor 30x2 sealed 7500 3652
CLINIConic
8x50 sealed 7500 3694
FIBERLite
HIGHConic
TM
720 7500 3621
TM
30x15 7500 3623
TM
F15-6x100 7500 3698
TM
II 7500 3620
The technical data of the rotors and the corresponding adapters and reduction sleeves for various commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: http://www.thermo.com
1-6 SL 16 / 16R Thermo Scientific
Before use
“Before setting up” on page 2-2
“Transporting the centrifuge” on page 2-2
“Location” on page 2-2
“Aligning the centrifuge” on page 2-3
“Mains connection” on page 2-3
“Storage” on page 2-4
2
Thermo Scientific SL 16 / 16R 2-1
2
Before use
Before setting up
Before setting up
1. Check the centrifuge and the packaging for any shipping damage. Inform the shipping company and Thermo Fisher Scientific immediately if any damage is discovered.
2. Remove the packaging.
3. Check the order for completeness (see “Scope of supply” on page iii.). If the order is incomplete, please contact Thermo Fisher Scientific.
Transporting the centrifuge
Due to its weight (see “Technical Data” on page 1-3), the centrifuge should be carried by several
people.
Always lift the centrifuge at both sides.
Location
Figure 2-1. Lifting the centrifuge at both sides.
The centrifuge can be damaged by impacts.
Transport the centrifuge upright and if at all possible in its packaging.
WARNING Always lift the centrifuge on both sides. Never lift the centrifuge by its front or the back panel. Always remove the rotor before moving the centrifuge.
The centrifuge should only be operated indoors.
The set-up location must fulfil the following requirements:
A safety zone of at least 30 cm must be maintained around the centrifuge.
People and hazardous substances must be kept out of the safety zone while centrifuging.
2-2 SL 16 / 16R Thermo Scientific
The supporting structure must be stable and free of resonance, for example a level laboratory
bench.
The supporting structure must be suitable for horizontal setup of the centrifuge.
The centrifuge should not be exposed to heat and strong sunlight.
WARNING UV rays reduce the stability of plastics. Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
The set-up location must be well-ventilated at all times.
Aligning the centrifuge
The alignment of the centrifuge must be checked every time after moving it to a different location. The supporting structure must be suitable for horizontal setup of the centrifuge.
2
Before use
Aligning the centrifuge
VORSICHT If the centrifuge isn’t level, imbalances can occur and the centrifuge can be damaged. Do not place anything under the feet to level the centrifuge.
Mains connection
Main Power Switch
Mains plug
Figure 2-2. Mains connections
1. Turn off the power supply switch on the back (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
Thermo Scientific SL 16 / 16R 2-3
Loading...
+ 43 hidden pages