Thermo Scientific Safe 2020, Maxisafe 2020 Operating Instructions Manual

Valid: 01.2009 / 50118069
Operating Instructions Safety Cabinets
Safe 2020 Maxisafe 2020
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
These operating instructions are protected by copyright. Rights resulting the­reof, particularly reprint, photomechanical or digital postprocessing or reproduction, even in part, are only allowed with the written consent of Thermo Electron LED GmbH. This regulation does not apply to reproductions for in-plant use.
Trademarks
All trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers.
Thermo Electron LED GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
Contents
1. General notes ......................................................................................... 6
1.1 General safety instructions ............................................................. 6
1.2 Warranty ......................................................................................... 7
1.3 Explanation of symbols................................................................... 7
1.3.1 Symbols used in the operating instructions ............................. 7
1.3.2 Symbols on the device ............................................................. 8
1.4 Use of the device ............................................................................ 8
1.4.1 Correct use ............................................................................... 8
1.4.2 Incorrect use: ............................................................................ 9
1.5 Standards and safety regulations ................................................... 9
2. Delivery ................................................................................................. 10
2.1 Standard components................................................................... 10
2.2 Acceptance inspection.................................................................. 10
2.3 Transport lock and device packaging ........................................... 10
3. Installation ............................................................................................ 11
3.1 Ambient conditions ....................................................................... 11
3.2 Room ventilation ........................................................................... 11
3.3 Correct location ............................................................................. 12
3.4 Installation in line .......................................................................... 12
3.5 Transport ....................................................................................... 13
4. Unit description .................................................................................... 14
4.1 Overall view .................................................................................. 14
4.2 Safety system ............................................................................... 17
4.3 Filter system ................................................................................. 18
4.4 Controls and display ..................................................................... 20
4.5 Sample chamber access .............................................................. 21
4.6 Device interfaces .......................................................................... 22
4.7 Sample chamber illumination ....................................................... 23
4.8 UV lamp unit ................................................................................. 24
4.9 Working area ................................................................................. 24
5. Start-up.................................................................................................. 25
5.1 Initial operation ............................................................................. 25
5.2 Installing unit and accessories ..................................................... 25
5.3 Levelling the cabinet ..................................................................... 27
5.4 Power supply connection ............................................................. 28
5.5 RS 232 interface connection ........................................................ 30
5.6 UV connection .............................................................................. 30
5.7 Installation test .............................................................................. 31
6. Handling and control ........................................................................... 32
6.1 Display .......................................................................................... 32
6.1.1 Functions of the display components ..................................... 32
6.1.2 Display during the calibration routine ..................................... 33
6.1.3 Display in OFF mode .............................................................. 33
6.1.4 Display in work mode ............................................................. 33
6.1.5 Power interruption .................................................................. 33
6.1.6 Display and functions after a power failure ............................ 34
6.1.7 Failure messages ................................................................... 34
6.2 Description of the operating modes ............................................. 35
6.3 Operation system ......................................................................... 37
6.3.1 Basic functions ....................................................................... 37
6.3.2 Moving the front window to the working position ................... 39
6.3.3 Silencing the audible alarm signal ......................................... 39
6.3.4 Switching the illumination on and off ...................................... 40
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
Contents
6.3.5 Activating and deactivating the internal power supply ........... 40
6.3.6 Displaying the UV disinfection time ........................................ 40
6.3.7 Activating and deactivating the potential-free contact
(optional) ................................................................................. 40
6.3.8 Switching the cabinet to OFF mode ....................................... 41
6.3.9 Setting the time ....................................................................... 41
6.3.10 Displaying the downflow velocity ........................................... 42
6.3.11 Displaying the operating hours of the HEPA filters ................ 43
6.3.12 Displaying the Performance Factor ........................................ 43
6.3.13 Setting and activating the timer .............................................. 44
6.3.14 Deactivating the timer ............................................................. 45
6.3.15 Setting the UV disinfection time ............................................. 46
6.3.16 Starting the UV disinfection .................................................... 46
6.3.17 Cancelling the UV disinfection ............................................... 47
6.3.18 Activating the stop watch ....................................................... 47
6.4 Pilot switch .................................................................................... 48
6.4.1 Moving the front window ......................................................... 48
6.4.2 Switching the device to OFF mode ........................................ 49
7. Operation .............................................................................................. 50
7.1 Hygiene preparations for the sample chamber ............................ 50
7.2 Preparing the sample chamber .................................................... 50
7.3 Response to failure messages ..................................................... 50
7.4 Work rules ..................................................................................... 51
8. Shut-down............................................................................................. 52
8.1 Interrupting an operation .............................................................. 52
8.2 Shutting the unit down .................................................................. 52
9. Cleaning and decontamination .......................................................... 53
9.1 Decontamination procedure ......................................................... 53
9.2 Wipe/spray disinfection................................................................. 53
9.3 UV disinfection after a wipe/spray disinfection ............................ 54
9.3.1 UV disinfection using the integral UV lamps .......................... 54
9.3.2 UV disinfection using a mobile UV device ............................. 55
9.3.3 Changing the UV disinfection time ............................................ 55
9.4 Sterilization with formaldehyde..................................................... 56
9.5 Cleaning the exterior surfaces...................................................... 56
9.6 Cleaning the floorpan ................................................................... 56
10. Maintenance ......................................................................................... 57
10.1 Inspection...................................................................................... 57
10.2 Service .......................................................................................... 57
10.2.1 UV lamps ................................................................................ 57
10.2.2 Sample chamber illumination ................................................. 58
10.2.3 Replacing the front window seal ............................................ 58
10.3 Retrofitting and repairs ................................................................ 59
11. Disposal ................................................................................................ 60
11.1 Disposal procedure ...................................................................... 60
12. Technical data ...................................................................................... 62
13. Device log ............................................................................................. 69
14. Certificate of decontamination ........................................................... 70
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
Figures
Fig. 1 Device locations in a room ........................................................... 12
Fig. 2 Lift points ...................................................................................... 13
Fig. 3-a Overall view / model Safe 2020 ................................................... 15
Fig. 3-b Overall view / model Maxisafe 2020 ............................................16
Fig. 4-a Filter system with downflow filter and exhaust
air filter / model Safe 2020 ...........................................................18
Fig. 4-b Filter system with primary filter, downflow filter and exhaust
air filter / model Maxisafe 2020 ....................................................19
Fig. 5 Controls and indicators ................................................................ 20
Fig. 6-a Access through front cover ..........................................................21
Fig. 6-b Access through front window .......................................................21
Fig. 7-a Supply interfaces .......................................................................... 22
Fig. 7-b Sample chamber illumination ....................................................... 23
Fig. 8 UV lamp unit .................................................................................24
Fig. 9 Working area on the workplate, armrests .................................... 24
Fig. 10-a Stand installation / model Safe 2020 ........................................... 25
Fig. 10-b Stand installation / model Maxisafe 2020 .................................... 26
Fig. 10-c Drain valve installation ................................................................. 27
Fig. 11 RS 232 interface connection ........................................................ 30
Fig. 12 UV connection .............................................................................. 30
Fig. 13 Indicator and operating panel ...................................................... 33
Fig. 14 Indicator and operating panel ...................................................... 37
Fig. 15 Basic functions of the pilot switch ................................................48
Fig. 16 Sitting posture .............................................................................. 52
Fig. 17 Front window seal replacement ...................................................58
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
1. General notes
The following are the addresses of the international Thermo Sales Organisations.
1.1 General safety instructions
These safety instructions describe the safety features of the Safe 2020 and Maxisafe 2020 series. The safety cabinet has been manufactured in keeping with the latest technological developments and has been tested before delivery for its correct function. It may, however, present potential hazards if it is not used according to the intended purpose or outside of operating parameters. Therefore, the following procedures must always be observed:
The safety cabinet must be operated only by trained and authorized personnel.
• For any operation of this unit, the operator must prepare clear and concise written instructions in the language of the operating and cleaning personnel based on these operating instructions, applicable safety data sheets, plant hygiene guidelines, and technical regulations, in particular:
• which decontamination measures are to be applied for the cabinet and
accessories,
• which protective measures apply while specific agents are used,
• which measures are to be taken in the case of an accident.
• Repairs to the device must be carried out only by trained and authorized expert personnel.
• The contents of the operating instructions are subject to change without further notice.
• Concerning translations into foreign languages, the German version of these operating instructions is binding lease.
• Keep these operating instructions close to the unit so that safety instructions and important information are always accessible.
• Should you encounter problems that are not detailed adequately in these operating instructions, please contact Thermo Electron LED GmbH immediately for your own safety.
Postal address Germany
Thermo Electron LED GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold
Enquiries from Germany: Phone
Sales 0800 1 536376 Service 0800 1 112110
Fax
Sales/Service 0800 1 112114
E-Mail
info.labequipment.de@thermofisher.com
International enquiries Tel. + 49 (0) 6184 / 90-6940 Fax + 49 (0) 6184 / 90-6772 E-Mail
info.labequipment@thermofisher.com
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
1. General notes
1.2 Warranty
Thermo Electron LED GmbH warrant the operational safety and functions of the safety cabinet only under the condition that:
• the device is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions,
• the device is not modified,
• only original spare parts and accessories that have been approved by Thermo Electron LED GmbH are used,
• inspections and maintenance are performed at the specified intervals,
• an installation test is performed prior to the initial operation of the device and that a repeat test is performed on the occasion of all inspections and repairs.
The warranty is valid from the date of delivery of the device to the operator.
1.3 Explanation of symbols
1.3.1 Symbols used in the operating instructions
WARNING!
is used if non-observance may cause serious or even lethal injuries.
CAUTION!
is used if non-observance may cause medium to minor injuries or damage.
NOTE
is used for hints and useful information.
RECYCLING!
Valuable raw materials can be reused.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
1. General notes
1.3.2 Symbols on the device
Observe operating instructions (switchbox ceiling)
Biohazard (left device front)
Cytostatic agents (left device front)
Checked safety (right side of device)
T5A note (sample chamber fusing)
RS 232 interface (connection label)
Armrest installation (right side of light dome)
1.4 Use of the device
1.4.1 Correct use
Model Safe 2020:
The safety cabinet is a laboratory device for installation and operation in microbiological and biotechnical laboratories of safety levels 1, 2, and 3. It has been designed as a Class II microbiological safety cabinet, in accordance with EN 12469 / 2000. Depending on the hazard level of the agents involved, the operator must prepare in writing appropriate decontamination procedures for the device and the accessories used in the sample chamber.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
1. General notes
Model Maxisafe 2020:
The safety cabinet also meets the requirements from DIN 12980 / 2005-06 and can be used for the production of cytostatic agents. For cytostatic agent applications, use the lowered one-piece working plate only.
Prior to the initial operation of the cabinet, the operator must perform an installation test. The test result must be documented by a test report. The cabinet must only be released for operation if it is in compliance with the operating parameters specified by Thermo Electron LED GmbH.
After any changes to the installation conditions or after any modification to the technical system, a repeat test must be performed and the test result must be documented by a test report that shows that all operating parameters are in compliance with those specified by Thermo Electron LED GmbH.
1.4.2 Incorrect use:
The safety cabinet must not be used in laboratories that do not comply with the requirements of safety levels 1, 2, or 3. The unit must not be operated as a Class II safety cabinet, if:
• no repeat test is performed after changes to the installation conditions or after modifications to the technical system,
• the alarm system of the device has issued a failure message and the
cause for the failure has not been repaired.
The alarm system must not be tampered with or disabled. If alarm system components heve been removed or disabled for service or repairs, the unit must only be released for operation if all alarm system components are functioning again properly.
The filters installed in the device are not capable of separating gaseous substances. Therefore, do not work with or store substances in the device:
• which in quantity or concentration are toxic,
• if a reaction with other substances may result in hazardous toxic concentrations or formation of toxic gases,
• that may form combustible or explosive mixtures in combination with air.
1.5 Standards and safety regulations
The device complies with the safety requirements of the following standards and guidelines:
• IEC 61010-1 / EN 61010-1
• EN 12469 / 2000
• DIN 12980 / 2005-06 (model Maxisafe 2020 only)
• Low Voltage Directive
• EMC Directive
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
10
2. Delivery
2.1 Standard components
Delivery for the safety cabinet includes the following:
Model Safe 2020:
• safety cabinet,
• drain valve.
Model Maxisafe 2020:
• safety cabinet
• armrests
• stand
All models:
• Device documentation: — operating instructions, — factory test report.
Optional components and accessories are listed as separate items in the delivery document.
2.2 Acceptance inspection
After the device has been delivered, immediately check the device:
• for completeness,
• for possible damage.
If the delivery is incomplete or if you detect any transport damage to the device, contact the forwarding agency and Thermo Electron LED GmbH immediately.
2.3 Transport lock and device packaging
Do not transport the device over large distances without transport lock and original device packaging.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
11
3. Installation
3.1 Ambient conditions
The operational safety and correct function of the unit depend on the location where it is to be operated. The safety cabinet must be operated only at locations that meet the ambient conditions listed below.
Location requirements:
• The electrical system of the device has been designed for an operating height of up to 2000 m above sea level.
• The mains power supply outlets should be out of normal reach to prevent accidental shut-off. Ideally, the outlets should be installed above the safety cabinet.
• The flooring of the location must be adequately strong and not flammable.
• The room in which the device is installed must be of adequately height. For units not connected to an exhaust system, the distance between the exhaust air opening and the room ceiling must be at least 200 mm (8 in).
• The location must be equipped with an appropriate ventilation system (see Section 3.2.).
• The temperature within the room must be between 15 °C and 40 °C (49 °F and 104 °F).
• The relative humidity in the vicinity of the device must not exceed 90 %.
NOTE - Ambient conditions
If ambient conditions vary from those descri­bed above, please contact Thermo Electron LED GmbH for assistance in installing the de­vice.
NOTE - Temporary storage
If the device is stored only temporarily (up to four weeks), the ambient temperature may be between -20 °C and +60 °C (-4 °F and +140 °F) at a relative air humidity of up to 90 %. For lon­ger storage periods, the location requirements apply.
3.2 Room ventilation
The room ventilation should preferably be a ventilation system that complies with the national requirements for the application.
• The inlet air and exhaust air openings of the room ventilation must be located so that drafts are prevented from impairing the function of the safety cabinet air system.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
12
Fig. 1
Locations in a room
3. Installation
3.3 Correct location
Choose a draft-free location where the safety cabinet does not interfere with the room traffic. Fig. 1: This figure shows preferred locations for safety cabinets and unsuiable locations, not in accordance with the safety requirements. Unsuitable locations: The locations [1], [2], and [3] are not suitable because they are exposed to drafts from windows and doors. Location [5] is undesirable because it is in range of plant traffic and within the exhaust air range of a ventilation system [4]. Preferred locations [6], [7], and [8] are correct because they are in a draft-free section of the room and not exposed to plant traffic.
3.4 Installation in line
When several devices are to be installed in line, please observe the following:
• Make sure that vibrations cannot be transferred between adjacent units.
• Exterior surfaces of the cabinets must always be accessible for cleaning and disinfection.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
13
3. Installation
3.5 Transport
Model Safe 2020: Fig. 2: To prevent tilting, always transport the device using
a suitable carrier, even for a transport within a building, and separate it from the stand.
CAUTION – Danger of tipping over!
For transport, lift the device only using the lift points shown in the illustration. Do not load the floorpan with the weight of the device frame!
CAUTION – Contusion hazard!
When lifting the safety cabinet, never place fingers or hands under the base of the device!
Model Maxisafe 2020:
If required, the device stand must be removed from the stand to be reinstalled onto the stand for the initial start-up after the transport (see Section 5.2).
Fig. 2
Lift points
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
14
4. Unit description
4.1 Overall view
Model Safe 2020:
Fig. 3a: Plenum assembly [5] with plenum for downflow blower [24] and plenum for exhaust air blower [2]. The downflow filter and the exhaust air filter are installed directly to the relevant blower. The exhaust air is released into the environment around the device through the opening [1].
• At the side of the plenum, the optional media supply lines [25] are routed into the sample chamber [20].
• Electrical box [4] with power supply cable [3]. An RS 232 connection to a PC and two fuse holders are installed at the front of the electrical box.
• Front cover [6] with integral, electrically movable front window [7], operated by:
• the pilot switch [26]. The gas struts [22] secure the front cover in the open state.
• The light dome [23] with two tubes is part of the front cover assy.
• Side panels [9] with two sealed access openings [19]. These can be equipped with media valves [13].
• Internal outlets [10] for power supply of accessories and adapter [11] for mobile UV device.
• Optional UV lamp unit [8] consisting of two UV lamps per side.
• Optional stands [12], adjustable in height and with fixed height
• Workplate segments [15] with optional arm rests [16]. A one-piece workplate and special workplates are available options.
• Lockable drain valve [17] for installation into floorpan.
• Display [21] with alarm system indicators.
• Test hoses for the supply unit [18] at the left side of the sample chamber and for the exhaust unit [14] at the right side of the sample chamber.
NOTE – Test hoses
Do not remove the two test hoses for checking downflow and exhaust air.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
15
4. Unit description
Fig. 3a
Overall view / Model Safe 2020
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
16
4. Unit description
Model Maxisafe 2020:
Fig. 3b: Stand [5] with integral prefilter housing [6].
• Filter inserts [2] for the prefilter housing. The number of supplied filter inserts depends on the width of the device. To protect from liquid residues, the filter inserts are oriented slightly sloping from the device backpanel to form a drain duct to the floorpan [3]. The filter plates [4] have tongues and grooves and are inserted in an overlapping pattern from right to left.
• Membrane sleeve or opening for filling with aerosol for testing the filter of the downflow unit [1] at the left side of the sample chamber and of the exhaust air unit [7] at the right side of the sample chamber.
Fig. 3b
Overall view / Model Maxisafe 2020
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
17
4. Unit description
4.2 Safety system
The safety system comprises a combination of protective and alarm systems that ensure maximum personal and material protection.
Safety systems:
Vacuum-sealed air system A vacuum-sealed air system in combination with HEPA filters for downflow and exhaust air forms the basis of the safety system for personal and material protection.
• Personal protection
Air aspired from the exterior along the entire working opening at a constant high velocity prevents that:
• agents may leak through the working opening of the chamber. As the exterior air pressure around the unit exceeds the pressure of the internal air system (vacuum sealing), it is ensured that:
• agents cannot be released to the exterior in the case of a leak in the
cabinet housing.
• Material protection
A steady airflow within the air system ensures that:
• a constant downflow allows the HEPA filters to remove contaminants
so that the samples are always surrounded by ultrapure air,
• harmful particles are not carried over through the sample chamber
(protection from cross-contamination).
HEPA filters The downflow (i.e. the air circulating within the device) and the exhaust air (air that is released to the exterior) are cleaned by HEPA filters (HEPA = High Efficiency Particulate Air Filter). For version Maxisafe 2020, a prefiltering system is used to increase the fil­ter efficiency and to protect the exhaust air and downflow systems.
• Safety lockout
To protect from UV radiation, the optional UV disinfection routine can be run only if the front opening is closed. During UV disinfection, the front opening safety lockout is activated and prevents harmful UV radiation from being emitted from the sample chamber.
Warning system:
• Airflow monitoring
Airflow monitoring determines the velocity of the airflow in the sample chamber as well as the inflow velocity of the air aspired from the exterior through the working opening. As soon as airflow velocities move above or below a specified safety value, a signal is transmitted to the alarm system.
• Visual and audible alarm system
The warning system constantly monitors the safety-relevant device functions:
• Inflow velocity of the air aspired from the exterior,
• downflow velocity,
• working position of the front window. If the warning system detects changes to one of these device functions, it issues:
• an audible and a visual alarm signal.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
18
4. Unit description
Position monitoring The position sensors monitor the position of the front cover as well as the movement of the front window; it will indicate when the front window is in the working position.
Performance Factor The Performance Factor (PER) is a value that indicates the safety state of the safety cabinet. This value is calculated from data determined by the safety system and from values captured empirically by service personnel during safety checks. This data is entered into a parameter list of the control software and interconnected. The result can be indicated by the display.
4.3 Filter system
Model Safe 2020: Fig. 4a: The filter system consists of two HEPA filters [2] and
[5] for the circulating air and for the exhaust air and of a coarse filter for the aspired air.
HEPA filters: Room air [10] is drawn into the sample chamber through the working opening. In the air duct, room air and the downflow within the chamber [7] are then blended to make up the blend air [9]. The blend air is then:
• filtered proportionally by the downflow filter [5] and supplied as ultrapure air [6] evenly into the sample chamber of the device,
• filtered by the exhaust air filter [2] and released as ultrapure air [1] to the exterior of the device.
Inlet air protection: The air duct between the sample chamber and the device plenum has an inlet air protection [8] below the working surface to prevent coarse particles from entering the plenum where they may impair blower [3] und [4] and filter functions.
Fig. 4a
Filter system with downflow
filter and exhaust air filter /
model Safe 2020
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
19
4. Unit description
Model Maxisafe 2020: Fig. 4b: The filter system consists of a set of HEPA primary
filter [10] (the number of filters depends on the sample chamber width) and two HEPA filters [2] and [5] for downflow air and exhaust air.
HEPA primary filter:
The compact size of the filter inserts allows filter replacement with minimal contamination hazard. The use of the primary filter protects the downflow and exhaust air filters considerably. The filter inserts used are HEPA filters of Class H14 (according to DIN EN 1822). Alternatively, activated carbon filters can be inserted into the housing to protect from smells. Room air [11] is drawn into the sample chamber through the working opening. In the air duct above the prefilter, room air and the downflow within the chamber [7] are then blended to make up the blend air [8]. The blend air is then:
• aspirated into the prefilter [10], filtered and fed into the air duct [9].
HEPA filters: From the air duct, the primary filtered air is:
• filtered proportionally by the downflow filter [5] and supplied as ultrapure air [6] evenly into the sample chamber of the device,
• filtered by the exhaust air filter [2] and released as ultrapure air [1] to the exterior of the device.
Fig. 4b
Filter system with primary
filter, downflow filter and exhaust air filter / model
Maxisafe 2020
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
20
4. Unit description
4.4 Controls and display
Fig. 5: The safety cabinet is equipped with two separate control
elements that operate independently of each other:
• pilot switch [1].
• operating panel [2],
The status indicators of the display [2] indicate control operations initiated with the control elements.
Pilot switch: Use to control all basic functions required for the operation.
Operating panel: The operating panel [2] has a display panel for text or number value output as well as 12 LEDs for visual indication of the current operating state of the device. All device functions can be controlled via the keys on the operator panel.
Fig. 5
Controls and indicators
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
21
4. Unit description
4.5 Sample chamber access
The sample chamber of the device is accessible via two modes:
Front cover: Fig. 6a: Manual opening of the front cover [2] allows access
to the complete sample chamber width with an opening height C. It is generally needed for decontamination and introduction of larger accessories.
NOTE – Front cover lockout
The front cover is equipped with a safety lockout and can only be opened when the front window is completely closed.
Front window: Fig. 6b: The electrically operated front window [1] is made of
multi-layer safety glass and integral to the front cover frame. It can be raised to a maximum opening height B (480 mm). To access the sample chamber during the work process, the front window must stay in the work position with opening height A (200 mm).
CAUTION – Front window movement!
Do not attempt to move the front window manually as otherwise the motor drive may be damaged.
Lowering the front window when the device is deenergized:
The safety feature (optional) is backed up by a battery. Should a power failure occur, the pilot switch (see Section 6.4) can be used to lower the front window completely.
Fig. 6a
Access through front cover
Fig. 6b
Access through front window
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
22
4. Unit description
4.6 Device interfaces
Fig. 7a: The standard equipment includes the outlets [10] for internal power supply
as well as the openings [8] on both sides for routing of cables and hoses. All other supply connections are available as options. Power supply connection: The connection to the power supply system is achieved through a cable with grounding plug [2] at the rear of the electrical box. Contact connection: The front of the box has an RS 232 interface [3] for the connection to a PC as well as two fuse holders for 5A miniature fuses: [4] for (L), [5] for (N). Internal power supply: There are two electrical outlets (5A) [10] and one UV disinfection adapter (optional) [9] located in the side walls. Media valves: There are two sealed feed throughs [8] on each side. These may be used for installation of media valves [7]. Media supply lines: Additionally, media can be supplied into the sample chamber through three pipes. The inlets [1] are located on top of the housing, the outlets [6] are placed at the sample chamber backpanel.
Fig. 7a
Supply interfaces
Loading...
+ 49 hidden pages