The dose rate measuring range of the RadEye GF-10 is limited
to 3 Sv/h (300 R/h for RadEye GF).
Correct dose integration is performed only, if the instrument
does not display over load “OVERLOAD”.
The device is not intended for use as a legal personal dose
meter.
The RadEye GF/GF-10 is not well suited for pulsed radiation.
There will be a significant under estimation at dose rates during pulse of more than 1.7 R/h / 17 mSv/h.
Do not use the unit if error messages appear on the screen.
The earphone connector at the bottom side of the instrument
must be exclusively used by equipment that is specified for
use with RadEye GF / GF-10.
The instrument must not be used in explosive atmosphere.
0-10 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 13
WEEE Compliance:
This product is required to comply with the European Union’s
Waste Electrical & Electronic Equipment (WEEE) Directive
2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more
recycling/disposal companies in each EU Member State, and
this product should be disposed of or recycled through them.
Further information on Thermo Fisher Scientific compliance
with these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may
assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/WEEERoHS
RadEye GF-Series DB-083 E 0-11
Le/Ff 20.07.2015
Page 14
Page 15
1. Introduction
The pocket sized RadEye is a sensitive and rugged device to
measure the ambient equivalent dose rate of gamma and X-ray
radiation. . In this context, it is important to note, that users
may select “R/h”, “Sv/h” “Gy/h” or “rem/h” as a display unit,
with the understanding that the energy response characteristic
is not altered by this selection. Within the RadEye GF/GF-10
the relation 1 R = 1 rem =10 mSv = 8.77 mGy/h is used.
The characteristic feature of the RadEye GF/GF-10 is the use
of sophisticated low power technology components and microprocessor based fully automatic self-checks. No maintenance is required.
The last 1600 mean and maximum values of the dose rate are
recorded internally and can be read out via serial interface.
Additionally the device logs the last 250 alarms, errors and
changes of the configuration. All events can be read out via
serial interface.
A real time clock is provided to add a time stamp to all buffer
data.
RadEye GF-Series DB-083 E 1-1
Le/Ff 20.07.2015
Page 16
All or selected menu functions described in 3.1 can be configured to be invisible and inaccessible by the user.
Thus, the instrument can be configured to both, an extremely
simple mode allowing just LCD-illumination and alarm acknowledgment to a very versatile mode for the more experienced user.
Alarm LED
showing radiation
exceeding thresholds.
Visible from top if
worn in a holster
and from display
Rugged plastic
enclosure
Key on top side to
acknowledge alarms
if worn in a holster
Powerful graphical display
Easy to use, four
key, menu driven
user interface
Earphone connector
1-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 17
2. Installation and start-up
2.1Scope of delivery
The RadEye is delivered in a folding paper box together with
two AAA cells and an operating manual.
A rubber protection sleeve, a holster and an earphone may be
added optionally.
Ordering information for the accessories see chapter 10.
RadEye GF-Series DB-083 E 2-1
Le/Ff 20.07.2015
Page 18
2.2 Inserting the battery
The two AAA-Alkaline cells as delivered with the instrument
allow up to 900 h of normal operation.
AAA size rechargeable batteries can be used as well.
• Switch off the measuring instrument.
• If supplied, remove rubber sleeve.
• Open the cover of the battery compartment.
Use of a coin is recommended.
• Exchange the batteries according to the shown polarity.
• Close the compartment cover hooks first, care for the rub-
ber seal being in its groove.
• Switch on the unit again. (see chapter 2.4).
2-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 19
The instrument continues working in the operating mode set
last.
The measured values in the history memory remain stored.
The real time clock for time stamp of history values and logbook continue operation if batteries are exchanged immediately. If Real Time Clock is set, actual time and date is displayed
for 3 s.
RTC will always be reset, if instrument is stored without batteries for more than 10 seconds.
History data and measurement parameters are stored permanently, even if batteries are removed.
RTC information is only required, if the instrument switched
off and on during operation and if correct timing information
for history and logbook entries is needed.
RadEye GF-Series DB-083 E 2-3
Le/Ff 20.07.2015
Page 20
2.3 Mounting of the protection sleeve
The rubber protection sleeve improves ruggedness to mechanical shocks.
For mounting of the sleeve first put the instrument into the top
of the sleeve. Then pull lower edges of the sleeve, one after
the other into its right position.
First step Second step
Front view Rear view
2-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 21
2.4Switching the unit on
GF or GF
-10
To switch on the RadEye, keep the ON button pressed for at
least one second. The sound generator (beeper) is initiated.
Calibration
date
Checksum of
The RadEye starts working with the parameters previously
selected (operation mode, calibration factor, alarm thresholds
etc.).
If the real time clock is set, actual date and time is displayed
for 3s. The RadEye first displays 0.00 µSv/h. After the first
detector pulse the measurement is started.
Firmware version Battery voltage
the firmware
RadEye GF-Series DB-083 E 2-5
Le/Ff 20.07.2015
Page 22
The first alarm thresholds for dose rate is around 200 - 500 %
above the typical normal background and allow sensitive, but
false alarm free dose rate monitoring with a reasonable fast
response time. The alarm thresholds for the dose are set to the
maximum possible values, thus being deactivated. With the
setting of the history recording interval of 120 s, the last 52
hours of operation will be stored in the history memory.
These default values are reasonable for the majority of applications. For special applications, the parameters (except calibration parameters) can be changed with the help of the optional PC-program “RadEye.EXE” from V1.30 and the cable
42540/29 or 42540/26. Furthermore, additional monitoring
modes can be activated.
The calibration parameters can only be changed in the factory
using special software tools.
2-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 23
3. Configuration
3.1Menu functions
To enter the operating menu, press the “Menu/◄” key.
Scrolling through the single menu options is effected by the
▲/▼ arrow keys.
The display returns to its initial default setting in case no key
has been activated for more than 10 seconds.
A “”behind some menu items identifies the corresponding
function as active.
The menu consists of a main menu and several sub menus.
Using the PC-Software and an interface adapter, any of the
functions can be hidden. This allows the user to be given only
the functions necessary to accomplish his measurement duties,
thus simplifying the handling considerably.
The ▲/▼ keys are used to scroll through the menu.
To select a menu option, release the ◄/Menu key as the re-
spective menu option has been reached.
RadEye GF-Series DB-083 E 3-1
Le/Ff 20.07.2015
Page 24
The meaning of the
◄/Menu key may change with the select-
ed menu. The meaning is shown on the bottom of the display.
Change
Off, On
Select
Yes
Exit
: Edit Alarm values
: Switching a function on and off
: Select a default display mode
: Confirmation of an action
: Exit menu
In Change menu the ▲/▼ keys are used to change a digit of
an alarm value.
3-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 25
MENU DISPLAY
Switch off RadEye is switched off. Time and stored
Backlight If selected the backlight is always on.
Measuring unit rem/h Select Dose rate (rem/h)
Operation mode Ratemeter Select Ratemeter mode
Scaler parameter PresetTimeMode Scaler mode with fixed time
Alarm Dose Rate Allows setting of Alarm 1 and 2 for dose
Alarm Dose Allows setting of Alarm 1 and 2 for dose.
Clear Dose Clears the accumulated dose..
Settings Battery type Selection of battery type: rechargeable or
PresetCountMode Scaler mode with fixed count number
Set time/count Edit preset time and preset count
Auto restart Auto restart of scaler measurements
Autosend If activated, the instrument sends a data
Single Pulse Enabling and disabling of single pulse
Finder Enabling and disabling of audible radiation
Set Time/Date Set time and date of the real time clock.
Source test Test the counter tube with a small source
DESCRIPTION OF THE FUNCTION
ACTIVATED BY THE LEFT KEY
data are maintained
Otherwise the backlight extinguishes after
10s
rate mode (rem/h, R/h, Sv/h).
non-rechargeable batteries. Correct low
battery warning depends on this setting.
string periodically to the infrared port. This
is used for radio transmission units.
indication. Activation is done by audio
keys (right button and top button). Alternative to “Finder”.
intensity. Activation is done by audio keys
(right button and top button). Alternative to
“single Pulse”.
RadEye GF-Series DB-083 E 3-3
Le/Ff 20.07.2015
Page 26
MENU DISPLAY
Alarm indication LED Enabling and disabling of the optical alarm
Show alarm Display of alarms stored in alarm log.
Text info Displays the text stored in the RadEye
Bluetooth Master Combo: Enable automatic connection to
BTcom Enable or disable the BTcom cover
SUBMENU
DISPLAY
Language Change menu language of RadEye
Contrast Adjust the LC display contrast
Sound Enabling and disabling of the acoustic
Vibrator Enabling and disabling of the vibration
LCD LED Enabling and disabling backlight illumina-
Remote/PC Combo: Remote device
DESCRIPTION OF THE FUNCTION
ACTIVATED BY THE LEFT KEY
alarm and the audible indication of keystrokes
alarm
tion on alarm
another BTcom cover
3.1.1 Main menu
The menu offers the following displays:
The above illustration depicts all menu options available or
possible.
3-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 27
3.1.2 Measuring unit
This submenu is used to select the measuring unit for ratemeter and scaler measurements.
The native units are as follows:
RadEye GF: x [R/h] and dose: Hx[R]
RadEye GF-10: *(10) [Sv/h] and dose: H*(10)[Sv]
Please Note:
If the native unit is changed by menu, the displayed measurement values are converted. This is not a conversion with ener-
gy correction!
3.1.3 Operation mode
This submenu is used to select between ratemeter and scaler
mode. The “
RadEye GF-Series DB-083 E 3-5
Le/Ff 20.07.2015
” show the active configuration.
Page 28
3.1.4 Scaler parameter
This submenu is used to set the parameters for scaler measurements. Preset TimeMode is used to measure a defined time.
Preset CntMode is used to measure a defined number of
counts. With “Auto restart” the RadEye restarts automatically
the measurement. “Set Time/Count” edit the preset time and
count.
3.1.5 Settings
This submenu is used to select between audible single pulse
indication and finder mode. The “” show the active configuration.
With “Autosend” the RadEye sends a data string periodically
to the infrared port. “Batt.type” is used to set the correct
threshold for warning message “LOW BATT”.
3-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 29
“Source test” performs a test for the counter tube with a small
source (see chapter 4.14).
“Language” changes the menu language between “English”,
“German” “French” and Russian. A “
configuration.
Set the display contrast with menu item “Contrast”.
3.1.6 Alarm indication
The submenu alarm indication is used to select the LED, LCD
LED, sound and vibrator for alarming. A “” shows the active
configuration.
” shows the active
3.1.7 Bluetooth
“Master”: Enable automatic connection to another Btcom cover module.
“Remote/PC”: This module can be connected to the Master
module “BTcom”: Enable or disable the BTcom cover.
RadEye GF-Series DB-083 E 3-7
Le/Ff 20.07.2015
Page 30
Please read see also chapter 6.4.6. For more information
in
dicat
ion
value
threshold 2
please read the manual DB-093 E.
3.2 Ratemeter
The standard display is indication of the actual dose rate.
Battery
Hearbeat
Bar
graph
overview
Alarm
threshold 1
Alarm
Pressing the up ▲/Info – key shows upon the
- 1st click: the accumulated dose
Additionally the remaining time until the
dose alarm 1 is reached, assuming the current
dose rate will persist.
- 2nd click: the mean value and max value of the measured
dose rate, together with the time
indicated since the last reset.
- 3rd click: standard display
Measurement
3-8 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 31
rd
After 10 seconds or after 3
click standard display with dose
rate indication is activated again.
1st click 2nd click
3.2.1 Trend indication
A trend indication is given if the measured dose rate is increasing or decreasing.
For accurate measurements, only readings without trend indication should be used.
A trend indication is not given at dose rates of less than
8 µSv/h.
RadEye GF-Series DB-083 E 3-9
Le/Ff 20.07.2015
Page 32
3.3 Scaler
To enter the scaler mode select in the submenu “Operation
mode” menu item „Scaler“.
The scaler mode starts with the parameter set in submenu
„Scaler Parameter“.
Preset
time
Elapsed
time
Action of
▲/Info key
To start the measurement, press the ▲/Info key. While the
measurement is in progress “Stop” is blinking.
Measurement value
3-10 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 33
3.4 Alarm thresholds
There are two alarm thresholds, each allocated to dose and
dose rate.
In order to avoid dose alarms while using the instrument exclusively as a rate meter, the dose alarms can be set to the
maximum level.
Configuration of the alarms is possible via infrared interface
(chapter 6.4) or via the menu.
As to the alarm activation, please also read chapter 4.2.
3.5 Setting alarm thresholds
The menu options
Alarm Dose
Rate and
Alarm Dose
allow the alarm thresholds to be modified. For this setting, the
user has 255 seconds time. Changing the value is effected by
pressing the left (Change) button if the corresponding “Alarm”
is selected:
To increment the number, press the ▲/▼ keys. To go on to
the next digit or to quit the edit mode, menu use ◄/► keys.
Once the last number has been set, quit the editing mode by
pressing the “Exit”-key. Then, the value set is saved and after
10 s the unit returns to the basic display.
RadEye GF-Series DB-083 E 3-11
Le/Ff 20.07.2015
Page 34
Example:
The dose rate alarm threshold must be changed.
Press “Menu” key,
Alarm Dose Rate
▲/▼ keys until
is selected
Then enter change menu by pressing “Change” key.
Edit value by pressing up / down
keys
select digit by ◄/► keys
Pre unit “µ”, “m” may be chosen as
well.
Leaving the last digit with ► key:
Confirm storage of edited value
pressing “yes”
The actual alarm thresholds are seen at the marks on the intensity bar scale. The upper mark shows alarm threshold 1, the
lower mark shows alarm threshold 2.
3-12 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 35
4. Operation
4.1Audible single pulse indication and finder mode
With the single pulse indication being selected, each pulse of
the detector generates a short audible signal emitted by the
beeper.
An audible alarm signal caused by exceeding the alarm threshold is not given while single pulse indication of finder mode is
active.
The single pulse indication must be enabled in the menu.
It is activated and deactivated by two times pressing the ► or
top button:
Indication of active single
pulse.
RadEye GF-Series DB-083 E 4-1
Le/Ff 20.07.2015
Page 36
Finder mode:
In Finder mode, the sound frequency depends on the pulse rate
of the detector. The more detector pulses above the count rate
that is present at the moment of the activation of the finder
mode are captured by the detector, the higher the tone.
The finder mode must be enabled in the menu.
It is activated and deactivated by two times pressing the ► or
top button:
Indication of active
finder mode.
4.2 Alarm indication
Each time the first alarm threshold is exceeded, the alarm devices beeper, LED and the vibrator become active, if they are
enabled.
Alarm 1: LED slowly blinking, two frequency alarm tone
Alarm 2: LED quick blinking, continuous alarm tone
Dose Alarm: LED constantly on, continuous alarm tone, vi-
brator slow.
4-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 37
The alarm tone and vibrator are acknowledged by a short key
depression, the LED remains pulsing. The alarm is extinguished, when the first alarm threshold is remained under.
4.3 Additional information
In addition to the basic display, using the menu options, requests for further information can be started such as e.g. the
setting of the enabling and disabling of the various alarm indicators.
4.4 Key Lock
If “Key Lock” is enabled, pressing the “▼/On” key for at least
5 seconds locks the key pad:
RadEye GF-Series DB-083 E 4-3
Le/Ff 20.07.2015
Page 38
It is recommended to lock the keys when wearing the unit in
the holster. Thus reduced battery time because of additional
power consumption by illumination of the LCD or other unintended operations is avoided.
Unlocking is performed according to the LCD instructions
upon pressing any key:
Press left key first, then lower key and then right key.
4.5 Process description for detection of radiation sources
Direction of maximum response
To detect hidden radiation sources, the dose rate alarm threshold must be set to the lowest value that does not produce fail
alarms. This is typical a value of 0.3 µSv/h to 0.5 µSv/h.
4-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 39
Due to its almost 100 times higher efficiency, the RadEye
PRD is more suited for this task, than the RadEye GF/GF10.
Subsequently, the operator passes the object while keeping
thereby the distance between the monitor and the object to be
checked as small as possible.
With an alarm occurring, the operator should move the instrument while observing the display in order to localize the
position of the source. Please note that the alarm can occur up
to a few seconds after passing the nearest distance to the
source.
If the alarm disappears after a few seconds, this alarm is to be
considered as a statistically released false alarm.
In case a radiation source has been detected, the radiation intensity is indicated. (See chapter 3.2).
4.6 Earphone
For alarming in noisy environments or for undercover investigations an earphone is available. See ordering information
page 2.1.
The earphone is connected at the bottom of the instrument
after opening the rubber protection seal.
RadEye GF-Series DB-083 E 4-5
Le/Ff 20.07.2015
Page 40
4.7 Alarm latching
alarm
number
Via PC-program it is possible to configure a alarm latching.
Alarm latching means the alarm is stored for the configured
time from 0 s (= off) to 9999 s.
A latched alarm is indicated by an “M” near by the alarm indicator. When the alarm is acknowledged the RadEye returns to
normal operations.
Latching indication
4.8 Show alarm
By selecting the menu point “show alarm” the alarms stored in
the alarm log are displayed together with the time of alarm and
the actual time.
Consecutive
Stored
alarm
Timestamp of
4-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
the stored
Page 41
By pressing the up and down arrow keys you can scroll
through the alarm log. Pressing the “
returns to the menu.
►” button the display
4.9 Text Info
Via PC program “RadEye.exe” it is possible to place text information in the RadEye. This text information can be displayed with the menu function “Text Info”.
4.10 Display rotation
If enabled, a short press on the lower centre key (arrow down)
flips the display. If the lower centre key is pressed again or if
the menu button is pressed, the screen flips into the normal
orientation.
RadEye GF-Series DB-083 E 4-7
Le/Ff 20.07.2015
Page 42
4.11 Set Date and Time
Setting of date and time takes place via menu function Setting
→ Set Date/Time.
The format is YYYY-MM-DD hh:mm. The fields are selected
with the keys ◄/►. With the keys ▲/▼the Year, month, day,
hour or minute can be changed. After setting the date and time
the real time clock is set and is used for alarm logbook and
history.
4.12 Battery type
Selection of battery type: Rechargeable or non-rechargeable
batteries. It is needed for correct low battery warning. Type
“Alk.” has a threshold of 2.1V. Type “NiMH” has a threshold
of 2.35V. See also chapter 5.4.
Battery type “Battery” Battery type “Accumulator”
4-8 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 43
active
4.13 BTcom cover
For detailed installation and operation procedure refer to
BTcom cover manual DB-076 E which is delivered together
with the Bluetooth battery lid.
The BTcom cover applies to the RadEye switched off.
With menu item “BTcom”, the user may switch on / off the
BTcom cover. “” means BTcom cover is active.
With “Master”, the BTcom cover connects automatically to
another BTcom cover. In this case, the RadEye cannot be connected to the PC. For more information please read the manual
DB-093 E.
By selecting “Remote/PC”, the BTcom cover is able to connect to a PC.
If the BTcom cover is active but not connected to any PC, a
Bluetooth symbol is flashing. If the BT-communication is
established, the symbol is always on.
BTcom cover
RadEye GF-Series DB-083 E 4-9
Le/Ff 20.07.2015
Page 44
4.14 Source test
ful
source test
Source test is started with menu item “Source test” in submenu
“Settings”.
Source test starts with the parameters previously set with
radeye.exe. Furthermore it is necessary to set the actual date
and time.
Start the source test without source. The RadEye check the
background value. If the background is lower than 0.15 µSv/h
the user is asked to place the source.
Date and time of last success-
Start source test
Low range test
High range test
by placing the
source
After placing the source, source test is started automatically.
4-10 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Terminate
source test with
“►” key
Page 45
During measurement the display shows the actual measurement value. After measurement the deviation is calculated and
displayed. The test is successful when the deviation is lower
between -20% and +20%. During source test the alarm LED is
active.
Measured deviation
Test result
Actual measurement value
At the end of source test, is displayed. If every test is successful the actual timestamp is stored and displayed together with
the message “Ready”. If source test fails, the message “Inspection” is displayed.
Source test is successful Source test fails. RadEye
should be sent to inspection
Remaining
time
RadEye GF-Series DB-083 E 4-11
Le/Ff 20.07.2015
Page 46
4.15 Change menu language
Dose
rate
Maximum scale
Via menu item “Language” in submenu “Settings” the user
may change the language of the RadEye menu.
4.16 Adjust display contrast
Here, the user can adjust the display contrast. This function is
called in sub menu “Settings” “Contrast”.
4.17 Graphic display
If enabled, one or more short presses on the “On/▼” key display a graphical view of the dose rate. This display is only
available in ratemeter mode.
Alarm
threshold 1
4-12 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 47
5. Functional test
The radiation meter performs continuous self-check routines.
A complete failure of the detector will be detected in around 2
minutes and will be indicated on the LCD and be announced
by the beeper. The same applies to the battery voltage.
5.1 Functional test
To carry out a simple test, shortly press any key. A short audible pulse has to be released and the LC display is illuminated
for some seconds.
The heart symbol next to the battery indicator must be “beating”. This indicates that the cyclic tasks as calculating measurement values and checking for alarm thresholds are active.
5.2 Failure indication
In case of a failure, the beeper generates a sharp single pulse
every 32 s.
The corresponding failure message is displayed in the LCD:
RadEye GF-Series DB-083 E 5-1
Le/Ff 20.07.2015
Page 48
Error high voltage generation.
No detector pulse within the last 30
minutes.
EEPROM with calibration data shows
EEPROM Read or EEPROM Write
error.
A Watchdog Error indicates that the
microcontroller has problems to work
on its tasks in a given timeframe.
Reasons are strong electromagnetic
pulses, firmware errors or hardware
issues.
The BTcom cover failed to initialize
the Bluetooth module.
5-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 49
5.3 Overload indication
For dose rates of more than 3 Sv/h (300 R/h, 300 rem/h,
2.6 Gy/h) an overload indication is given:
The dose value is marked with an overload indication, if a
dose rate overload occurred since the last dose reset.
Overload indication
in dose display
RadEye GF-Series DB-083 E 5-3
Le/Ff 20.07.2015
Page 50
5.4 Low Battery
If the battery voltage is below the configured threshold (see
chapter 4.12) the following warning appears:
The beeper generates a single pulse every 32s. This acoustic
warning can be suppressed by pressing the alarm acknowledge
key. After 8h this warning comes up again. The battery needs
to be changed. However, the RadEye can still be operated for
several hours.
If the battery voltage falls below 2V, the battery symbol starts
flashing. An acoustic warning is generated every 8s and can’t
be acknowledged. The battery needs to be changed as soon as
possible.
5-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 51
5.5 Calibration expiration date
If the expiration date is enabled and date is expired, the following warning message is shown at start up of the RadEye:
If the expiration date is enabled the RTC is not set (e.g. after
change of batteries) the following warning message is shown
at start up.
After acknowledge the RadEye works as usual.
The expiration date can be changed with the tools
“CalRadEye” or “RadEyeSetCalibDate”.
RadEye GF-Series DB-083 E 5-5
Le/Ff 20.07.2015
Page 52
Page 53
6. PC configuration
6.1Ways to connect to a PC
A RadEye unit can be connected to a PC either by either serial
interface, USB interface or by Bluetooth TM 1.
6.1.1 Wireless connection via Bluetooth
Details about the Bluetooth communication are provided
through the manual DB-076 E delivered with the optional
Bluetooth battery cover. Please note that the connected PC
must be equipped with a Bluetooth interface as well.
1
Bluetooth is a trademark of Bluetooth SIG, Inc., Bellevue, Wash-
ington, U.S.A.
RadEye GF-Series DB-083 E 6-1
Le/Ff 20.07.2015
Page 54
6.1.2 Cable connection
It is recommended to use the optional RadEye desktop holder
to align the RadEye’s IR window with the IR window of a
connection cable.
The optional serial adapter cable is used to connect to a RS232
port of a PC.
This serial port is selected directly in the RadEye.EXE software (see also chapter 6.3.4).
Alternatively, the optional USB connection cable is used to
connect to an USB port.
Using the USB connection requires the installation of driver
software that will generate a virtual serial port that can be se-
6-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 55
lected in the RadEye.EXE software. The driver is delivered on
the RadEye.EXE software installation CD ROM.
Installation of a connection cable
Using the hardware provided and without exerting too much
force, attach the connections of the adapter cable to the PC and
to the RadEye desktop holder.
6.2 Installation of the optional RadEye.EXE
software
Open the file “Setup.EXE” on the optional RadEye.EXE
Software installation CD Rom and follow the installation
guideline. The installation generates a RadEye.EXE Icon on
the desktop.
6.3 Starting RadEye.exe
Double click on the RadEye.exe symbol at the desktop Once
the RadEye.EXE program has been started, device parameters
are read out and displayed on the screen.
RadEye GF-Series DB-083 E 6-3
Le/Ff 20.07.2015
Page 56
6.3.1 RadEye Device Parameters
The Frame "RadEye GF(-10)" contains the unit's serial number and version number of the software. Click on the Read
button, the Parameters of RadEye will be read out from device
and shown in the Frame.
Figure 6-1: Main Window
6.3.2 Measurements
After pressing the button Read Values the current Count rate,
dose rate and dose are displayed in this frame.
6-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 57
With button Start the measured value is read from the unit at a
certain polling interval. Use the scroll bar or the buttons beneath the "polling interval" field to define the polling interval.
You can select a value between 1 and 3600 seconds.
The dose rate can be displayed numerically and graphically.
Click on the Graphics…. A diagram is displayed that gives a
representation of the dose rate values versus the time.
The current measured value is added at the right-hand side,
and the diagram is shifted to the left. Up to 100 measured values can be represented. The graduation of the ordinate is automatically adjusted to the measured values supplied by the
unit. With menu item “Configuration” the user may adjust the
y-axis of the graph.
Figure 6-2: Measurement graphics
6.3.3 Creating a Measurement File
The actual measured values Dose and Dose rate that are shown
in the "Measurement" display field can be stored in a measurement file.
RadEye GF-Series DB-083 E 6-5
Le/Ff 20.07.2015
Page 58
To do this, open the File menu, click on Open Logfile..., and
enter path and name. The file is stored with the file name extension ".log".
If another file of the same name exists, the system asks whether that file shall be overwritten or the new measured data shall
be appended to the existing data.
Once you have pressed OK to confirm the entries, the measurement log file is created and the polling measured data is
stored in the scan interval you have defined. An open measurement file is indicated by the name and the path of the
measurement log file that appears in the top right-hand corner
of the window.
To terminate data storage, open the File menu and select the
Close Logfile menu item. No further data is recorded.
Open the File / View Logfile... menu to view the measurement
log file.
For training and demonstration a prior recorded log file can be
replayed by opening Replay logfile.
With the buttons Start and Stop together with the polling interval the replay can be controlled.
Close Replay switches back to accessing measurement values
via infrared interface.
6-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 59
The measurement log file is an ANSI text file with columns
that are separated by <TAB>. This enables this file to be read
easily into other programs (such as Excel) where the data can
be processed.
The first line of the measurement log file contains the unit
name, the file name, and the path. Serial number and Device
identification are specified on the second line.
Date and time of the measurement are specified in the columns
under the field names "mm.dd.yy" and "hh:mm:ss". The time
setting corresponds to the PC system time.
The "Counter" column contains the numerical value of the
counter measurement. The "Unit" column informs about the
unit. The "Dose Rate" column contains the numerical value of
the dose rate measurement. The "Unit" column informs about
the unit (Sv/h rem/h or R/h).
RadEye GF-Series DB-083 E 6-7
Le/Ff 20.07.2015
Page 60
6.3.4 Select serial interface
Via Configuration / Com settings... menu another window is
opened from which you may select the corresponding interface.
If the RadEye is connected via BTcom cover select “Connected via Bluetooth”. “Read RadEye2 is used to check the connection to the device. “Search RadEye” scans all Com-Ports to
detect a RadEye. This action may need some time, depending
on the number of available Com-ports.
6.4 Configuration
The configuration of the RadEye can be protected with a
password.
6-8 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 61
Figure 6-3: Password Setting
No password is set during the first installation of the software.
A reminder window is activated, if the configuration window
is opened.
Figure 6-4: Password reminder
On the Configuration / RadEye … menu, the following parameters can be modified:
• Unit
• Time interval of the history
• Temperature display
RadEye GF-Series DB-083 E 6-9
Le/Ff 20.07.2015
Page 62
• Acoustic indication
• Alarm level for all basic displays
• Signaling types
• Additional surveillances
• Active menu functions
• Settings for source test
6.4.1 General configuration
Figure 6-5: General configuration
On the “Generals” tab, the user may set the time interval for
the history memory (1...43200s), the kind of the acoustic rate
indication (single pulse, finder – see chapter 4.1), temperature
display, type of battery and keyboard functions.
6-10 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 63
6.4.2 Text
Figure 6-6: Text
On the tab “Text” it is possible to store a text in the RadEye.
In the field “Text Info” the user can define 4 lines text, which
can be displayed on the LCD by the RadEye (see 4.10).
In the “Text Field” it is possible store a text with up to 200
characters in the RadEye. This text can not be displayed by the
RadEye.
RadEye GF-Series DB-083 E 6-11
Le/Ff 20.07.2015
Page 64
6.4.3 Mode settings
Figure 6-7: Mode settings
On the “Mode” tab, the user may set the operation mode, the
physical unit and parameters for scaler mode.
6-12 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 65
6.4.4 Source test
Figure 6-8: Source test parameter
This tab is used to provide the function “source test” with the
right data.
Nuclide: This name is prompted in function “source test”
Reference dose rate: dose rate of this nuclide at reference date
Reference Date: reference date for this source
Half life: Half life time of this nuclide.
Last source test: date and time of the last successful source
test
RadEye GF-Series DB-083 E 6-13
Le/Ff 20.07.2015
Page 66
6.4.5 Alarm setting
Figure 6-9: Alarm setting
The tab “Alarm” offers the user the possibility of making the
alarm settings for dose rate and dose.
With “alarm thresholds” “read only” the thresholds cannot be
modified by RadEye.
Alarm indicator “Beeper”, “LED” and “Vibrator” has the same
consequence as the menu item “Alarm indication”.
6-14 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 67
6.4.6 Bluetooth
Figure 6-10: Bluetooth
Here the user may configure the BTcom battery cover. The
frame “Device” informs about firmware version, serial number
and the physical address of the BTcom cover (not the
RadEye).
The Parameter “Blue LED on” relates to the blue LED at the
backside of the battery cover that shows the connection status.
Setting “Power off at battery low” the BTcom cover powers
down automatically if battery low status is detected to increase
battery life of the RadEye. “Secure connection” uses the PIN
code that can be edited in the right field.
For more information please read the manual which is delivered together with the BTcom cover.
RadEye GF-Series DB-083 E 6-15
Le/Ff 20.07.2015
Page 68
6.4.7 Menu item
Figure 6-11: Menu selection
On the tab “Functions”, the functions available to the user can
be selected to speed up access to frequently needed functions.
For example, if LED alarm is always active and rate level
indication is never used, these functions can be hidden for the
user.
Once the setting of the parameters is finished, these parameters have to be sent to the RadEye by clicking on the Write to
RadEye ... button.
Parameters including the selected menu configuration can be
saved with File / Save as… as parameter file (*.cfg). A saved
configuration can be reloaded using the menu File / Open…
and sent to RadEye.
6-16 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 69
6.5 History
Via the Options / History menu, the values stored in the
RadEye data memory can be read out, represented in an x/t
diagram and saved to the hard disk of the computer. These
data subsequently can be read in and further processed in a
spreadsheet program.
Time interval of History storing can be set from 1s to 43200 s
(12 hours). 1600 measurement values can be stored.
The following figure depicts for example the curve of the dose
rate over the last two days at a time resolution of 120 s. Clearly various levels resulting from different locations and points
with high peaks can be recognized.
The blue line shows the mean values, the red line the maximum value within the time interval.
If batteries are removed, the time reference is lost. In the diagram, time information is only provided for history values
with valid timing entries.
Figure 6-12: History read out
RadEye GF-Series DB-083 E 6-17
Le/Ff 20.07.2015
Page 70
Figure 6-13: Single history value indication
Clicking at the graph, each individual, stored measured value
can be displayed.
6.6 Logbook
Changes in configuration, occurring alarms and errors are
logged in a buffer.
These saved events can be read out via Options / Logbook… .
The logbook is shown as a table, and can be saved to PC hard
disc or printed.
The logbook has a maximum of 250 data sets. Several events
at the same time are saved as one record. At the display every
event is shown in one line for better overview. The date and
time of the PC is used for time relation.
6-18 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 71
Figure 6-14: Logbook
RadEye GF-Series DB-083 E 6-19
Le/Ff 20.07.2015
Page 72
Page 73
7. Trouble Shooting
7.1 RadEye doesn’t power on
7.1.1 Press “On” for at least one second
Keep the button “” pressed until the LED and the sound is
active. If the button is pressed only shortly, then the start up
routine is not activated. The LED may flash then only very
shortly.
7.1.2 Check battery
If the RadEye does not respond to a long press (> 1s) of the
“” button, then please use the following step by step procedure:
A) Remove the current batteries
B) Clean the battery contacts
C) Check the required orientation of the batteries according to
the printed “+” symbols in the battery compartment.
D) Insert two -“AAA” with sufficient capacity: Rechargeable
AAA batteries should be completely loaded, non recharge able batteries should supply together at least 2.3 Volt under
load.
E) Press the “
7.1.1).
” button for more than one second (see also
RadEye GF-Series DB-083 E 7-1
Le/Ff 20.07.2015
Page 74
7.2Reading data from the start up screen
GF or GF
-10
The start up screen shows important data like the last calibration, the current voltage and the firmware version.
This screen stays active as long as the “” button is kept
pressed during the power on of the RadEye device.
7.3 The RadEye doesn’t show the menu items as
described in the manual
The items available in the RadEye menu can be disabled or
enabled via the RadEye.EXE software. The factory setting has
all menu functions enabled. Please refer to chapter 6.4, menu
settings. Next, please check the firmware version of your
RadEye (see chapter 2.4) and compare your version number
with the revision table of this manual at the beginning of the
document. The revision table shows the introduction of new
functions together with the according firmware version.
7-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 75
7.4 The RadEye shows an error message in the
display’s top line
Please refer also to chapter 5.2. If one of the following failures
is indicated, then the RadEye must be repaired by the Thermo
Scientific bench repair team:
HV-Error
Detect Err
EEPROM Err
Watchdog*
Please contact our Customer Service for more details.
*) Please see the failure description in the above referenced
chapter. If the failure was related to a strong electromagnetic
pulse scenario and the failure went away afterwards, then a
repair may not be necessary. A diligent observation of the
RadEye proper operation after the failure is strongly recommended.
7.5 The RadEye is not found in the software
If the connection between the RadEye and the computer cannot be established with a selected serial port (COM port), then
the program shows the bellow message first:
RadEye GF-Series DB-083 E 7-3
Le/Ff 20.07.2015
Page 76
The RadEye.EXE software offers to search for a connection to
the RadEye unit on other available serial ports.
If there is a faulty connection between the unit and the computer, the program shows the error message “Data transfer to
instrument fails”.
In this case, you must check whether:
• The RadEye has been positioned correctly in the desktop
holder (with rubber boot) and the RadEye is turned on.
• The infrared transmission windows are clean.
• The correct serial interface has been selected.
7-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 77
It is recommended to use the Windows system tools to determine the occupation of the currently active serial ports. Virtual
serial ports are labeled accordingly for identification.
´
Device Manager showing the virtual driver
for USB „Prolific“ at COM 4
Open next the serial interface step window in the RadEye.EXE
software: Configuration / Com settings... menu. The used
serial port has to be selected next.
7.6 RadEye lost date and time settings
The date and time information is lost, each time the batteries
are removed or below the required voltage for operation. The
simplest way to program the date and time is to use the “synchronize” function of the optional RadEye.EXE software.
Please refer to chapter 6.4.1. Else the current date and time can
be set also through the RadEye’s menu “Settings” / “Set Date /
Time”.
RadEye GF-Series DB-083 E 7-5
Le/Ff 20.07.2015
Page 78
7.7 RadEye’s battery status is lower than expected
Please check if the correct battery type is selected under “settings” / “Batt. type” and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batteries provide 1.5V.
Rechargeable batteries offer also a more steep drop from acceptable battery voltage to “battery low” than Alkaline types.
7.8 History data shows the wrong time and date
See chapter 7-5.
7-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 79
8. Maintenance
8.1Recommended maintenance
The RadEye GF(-10) is basically maintenance free.
8.1.1 Source test
It is recommended to perform a source test routine with the
optional Ba-133 test adapter on a regular basis. The source test
routine is described in chapter 4.14.
The rubber boot should be removed and cleaned separately.
The RadEye unit shall be cleaned with a moist tissue (only
water as cleaner). The rubber boot shall be washed in water.
Do not use aggressive chemicals like organic solvents.
RadEye GF-Series DB-083 E 8-1
Le/Ff 20.07.2015
Page 80
8.1.3 Remove batteries for storage
It is highly recommended to remove both AAA batteries from
the battery compartment, if the RadEye unit is stored for a
longer period (> 1 week) to avoid damage through battery
leakage.
The clip at the backside of the holster allows convenient carrying of the RadEye at the belt.
RadEye not included
10.2 Earphone for RadEye series (425067037)
The usage of the earphone is recommended, if the RadEye is
switched into FINDER mode and disturbance through noisy
environment shall be compensated.
The earphone is plugged into the RadEye’s socket at bottom of
the unit. The cable length is 1.2m.
RadEye GF-Series DB-083 E 10-1
Le/Ff 20.07.2015
Page 84
10.3 Ba-133-Adapter for RadEye (425067072)
The shape of the adapter insures that the Ba-133 is placed
close to the reference point of the RadEye unit.
Please see chapter 4.14 and chapter 6.4.4 about the usage of
the test adapter.
10.4 Desktop holder (425067060)
Alignment between the RadEye’s IR data interface window
and the IR interface of a PC connection cable is insured with
the usage of the desktop holder. Please see chapter 6.1.2 for
details.
10.5 IR connection cable serial (4254029)
The IR plug of the connection cable can be fixed easily by
screws at the desktop holder. The PC connection plug is a 9
pin sub D type for serial COM ports.
10-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 85
Please see chapter 6.1.2 for details. Cable length = 1.2 m
10.6 IR connection cable USB (4254026)
The IR plug of the connection cable can be fixed easily by
screws at the desktop holder. The PC connection plug is a
USB port.
Please see chapter 6.1.2 for details. Cable length = 1.2 m
10.7 Bluetooth battery cover (425067087)
The standard battery lid is replaced with the Bluetooth battery
cover. Please see chapter 6.1.1 for details.
RadEye GF-Series DB-083 E 10-3
Le/Ff 20.07.2015
Page 86
10.8 RadEye car- and wall holder with accumulator
charging circuit
Safety instructions
1. Only use NiMH – accumulators, no primary batteries
(Alkaline –manganese, zinc – carbon ...)
2. Only use two accumulators of the same type (manufac-
turer, rated capacity ...)
3. Only charge accumulators having the same discharging
state
4. Never use defective accumulators (penetrating/emerging
electrolyte, damaged housing...)
10-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 87
RadEye car holder : 42506/7065
The RadEye car holder 42506/7065 serves as supporting device and power supply unit for the operation of a RadEye instrument in a vehicle. For convenient mounting to the dashboard one of the mechanical adaptors 42506/7061 ... 64 is
required.
Together with a power supply unit 42506/7066, this support is
also suitable for a cost attractive stationary ambient monitoring.
Via the alarm contact, power consuming external alarm indicator can be operated.
RadEye GF-Series DB-083 E 10-5
Le/Ff 20.07.2015
Page 88
The RadEye car- and wall holder combines several functions.
(1) Mechanical holding device for instruments of the
RadEye-family
(2) Accumulator charger unit with temperature controlled
charging of NiMH accumulators in the device
(3) Infrared interface for direct connection to the serial in-
terface of computers
(4) Alarm contact for connecting alarm indicators up to a
power of 24W (at 24V, see RadEye area monitor
42506/80)
Before plugging the RadEye into the Car holder the rubber
protection plug at the phone jack has to be opened:
Carefully remove the protection sleeve and open the phone
jack
10-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 89
Remount the protection sleeve (see chapter 2.3). A small
screw driver for the protection plug might be helpful. The
phone jack must left be open, and the protection plug must be
perpendicular to the protection sleeve.
Then insert the RadEye into the car holder. Make sure that the
protection plug is not mechanically interfering with the charging pin of the car holder.
RadEye GF-Series DB-083 E 10-7
Le/Ff 20.07.2015
Page 90
10.8.1 Accessories for dash board mounting of
car holder 42506/7065 (including charging function)
holder 42506/7060 (mechanical holder only)
Goose neck adaptor kit 42506/7061
Pivot arm adaptor kit 42506/7062
10-8 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 91
Knuckle joint adaptor kit 42506/7063
Goose neck adaptor kit with suction cup 42506/7064
RadEye GF-Series DB-083 E 10-9
Le/Ff 20.07.2015
Page 92
10.8.2 Accessories for data transmission using the
car adaptor
RS232 connecting
cable 2m:
SM1685 35223
USB 1.1 to RS232
adapter cable:
SM1685 35251
USB 2.0 to RS232
adapter cable:
SM1685 35255
10.8.3 Accessories for AC – operation of the car adaptor
Power supply with cigarette lighter socket
42506/7066 for mains
operation of the car holder
42506/7065
10-10 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 93
10.9 RadEye Area Monitor with alarm contacts
(4250680)
A RadEye GF can be used as installed radiation monitor if
used as RadEye Area Monitor. The RadEye is permanently
powered and the alarm status is visible as light and with further accessories also as sound alarm (425068010). More information is provided in the manual DB-062 E.
10.10 RadEye inductive charger
The RadEye holder serves as supporting device and, in combination with the special battery lid 42506/7034, power supply
unit for the operation of a RadEye instrument in a vehicle. For
convenient mounting to the dashboard one of the mechanical
adaptors 42506/7061 … 64 is required.
RadEye GF-Series DB-083 E 10-11
Le/Ff 20.07.2015
Page 94
Note: The charging function of this device requires the battery
latch 42506/7034. Without the above mentioned part no charging will be achieved.
Safety instructions
1. Only use NiMH – accumulators, no primary batteries (Al-
kaline – manganese, zinc – carbon ...).
2. Only use two accumulators of the same type (manufactur-
er, rated capacity ...).
3. Only charge accumulators having the same discharging
state.
4. Never use defective accumulators (penetrating/emerging
electrolyte, damaged housing...).
5. The bottom of the housing (aluminum heat sink) gets
warm (50°C, 120°F max.).
6. Not for use in open convertibles.
7. Do not put cards with magnet stripes (f.e. credit cards,
parking cards, phone cards, etc.) near the mounted warning symbol. The magnetic field could erase the data from
your card.
8. Clearance of at least 50cm (20”) between charger and car
radio is recommended.
10.10.1 LED indicators
State LED (green)
Voltage supply on On
Voltage supply off Off
10-12 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 95
11. Technical data
11.1 RadEye GF /GF10
Radiation type: Gamma and X-ray radiation
Measured quantity: Ambient equivalent dose rate:
RadEye GF:
x [R/h] and dose: Hx[R]
RadEye GF-10:
*(10) [Sv/h] and dose: H*(10)[Sv]
Please Note:
If the native unit (see above) is changed
by menu, the displayed measurement
values are converted. This is not a conversion with energy correction!
Alarm threshold:Two alarm thresholds for dose and dose rate each.
Default setting: see chapter 2.4
Audible alarm intensity:at least85 dB at a distance of 30
cm
Time to alarm:typ.: 215 s for background to 3 µSv/h
typ.: 40 s for background to 10 µSv/h
typ.: 15 s for background to 50 µSv/h
Energy range:50 keV – 3 MeV error less than ±30% for dose and dose rate measurement
Energy dependence: see the following diagrams
Direction of
max. response: perpendicular to the device’s
longitudinal axis, on the reference
mark on the back side of the unit
Reference point:on the axis of the direction of max. response,
14 mm behind reference mark.
Angular dependence: see the following diagrams
11-2 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 97
Working temperature: -20°C ... + 50°C
Relative humidity: 10 ... 93 % at 35°C non condensing
Operating voltage: 1.8 ... 3.3 V,
Battery low voltage starting from 2.1 V
Degree of protection: IP 65 according to EN 60 529
EMC: Disturbance emission : EN 61000-6-3
Immunity : EN 61000-6-2
Mechanical shock:Drop onto a concrete surface
0,5m without protection sleeve
1,5m with protection sleeve
Size:96 mm x 61 mm x 31 mm without rubber protection
Weight:around 160 g including two Alkali Manganese AAA cells LR 3
around 190g including AAA cells and
protection sleeve
Internal memory:The last 1600 measured values are
saved and can be read out via
PC program. Max- and mean value of
dose rate. The time interval is factory
preset to 120s by default.
Logbook with 250 entries for changes
of configuration, occurring alarms and
errors.
RadEye GF-Series DB-083 E 11-3
Le/Ff 20.07.2015
Page 98
Averaging filters:Dose Rate filter type: Advanced Digital Filter (ADF)
Digital RC-Filter with time constant
1s.....100s, depending on dose rate
and dose rate changing.
Power consumption: ≈ 1 mA: normal operation without
alarm signals and LCD illumination
≈ 30 mA with illuminated LC display
≈ 18 mA LED alarm
≈ 40 mA acoustic alarm
≈ 30 mA vibrator alarm
Battery service life:≈ 900h using two alkaline AAA cells depending on the operating mode
≈ 600h using 800mAh NiMH accu
11-4 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Page 99
11.2 RadEye inductive charger
Operating voltage 11,5…15VDC
Current consumption 150…200mA With supply
from 14 V, charging current 50 mA
Accumulator charging
40…52 mA
current
Charge time
RadEye on,
20…24h Accumulator
800mAh, discharged
Backlight switched off *)
Charge time
RadEye switched off
18…20h Accumulator
800mAh, discharged
Ambient temperature -20…+50°C Operation
–40…+70°C Storage
0…40°C Accumulator
charge
Ambient pressure 300…2000 hPa
*) During operation with backlit LCD, the battery charge
is sustained only.
RadEye GF-Series DB-083 E 11-5
Le/Ff 20.07.2015
Page 100
Diagram 11-1: Energy dependence in direction of max.
response, perpendicular to ref. mark, ambient
equivalent (RadEye GF)
Diagram 11-2: Energy dependence in direction of max.
response, perpendicular to ref. mark, ambient
equivalent (RadEye GF-10)
11-6 DB-083 E RadEye GF-Series
Le/Ff 20.07.2015
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.