Thermo Scientific RadEye GF, RadEye GF-10 Operating Instructions Manual

Page 1
Operating Instructions DB-083 E
RadEye GF-Series
GF GF-10
Gamma Survey Meters
Page 2
Page 3
REVISIONS SHEET:
Rev. Rev.
state
06.04.11 RM&SI-EH Leder First edition
A 15.09.11 RM&SI-EH Leder cpl. A Complete revision for
B
12.09.12 RM&SI-EH Pij
C 07.06.13 RM&SI-EH Leder 11-2 A Energy range added D 20.07.15 RM&SI-EH
*) Category C: editorial correction I: clearing improvement A: substantial amendment
Explanations must be given, at least with Category A.
Dept. resp.
ECN80271
Name Rev.
page
4-14 5-4 7-1 8-1 9-1 10-1 12-1
Pij cpl. A from V3.06
Cat.
Explanation
*)
Firmware V2.06
A > V3.01
New document struc­ture Change menu language Calibration expiration Trouble shooting Maintenance Spare parts Optional accessories Annex
RadEye GF-Series DB-083 E 0-1 Le/Ff 20.07.2015
Page 4
CONTENTS:
1. Introduction ..................................... 1-1
2. Installation and start-up .................. 2-1
2.1 Scope of delivery ...................................................... 2-1
2.2 Inserting the battery .................................................. 2-2
2.3 Mounting of the protection sleeve ............................ 2-4
2.4 Switching the unit on ................................................ 2-5
3. Configuration ................................... 3-1
3.1 Menu functions ......................................................... 3-1
3.1.1 Main menu ......................................................... 3-4
3.1.2 Measuring unit ................................................... 3-5
3.1.3 Operation mode ................................................. 3-5
3.1.4 Scaler parameter ................................................ 3-6
3.1.5 Settings .............................................................. 3-6
3.1.6 Alarm indication ................................................ 3-7
3.1.7 Bluetooth ........................................................... 3-7
0-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 5
Ratemeter .................................................................. 3-8
3.2
3.2.1 Trend indication ................................................. 3-9
3.3 Scaler ...................................................................... 3-10
3.4 Alarm thresholds ..................................................... 3-11
3.5 Setting alarm thresholds .......................................... 3-11
4. Operation ......................................... 4-1
4.1 Audible single pulse indication and finder mode...... 4-1
4.2 Alarm indication ....................................................... 4-2
4.3 Additional information ............................................. 4-3
4.4 Key Lock................................................................... 4-3
4.5 Process description for detection of radiation sources4-4
4.6 Earphone ................................................................... 4-5
4.7 Alarm latching .......................................................... 4-6
4.8 Show alarm ............................................................... 4-6
4.9 Text Info ................................................................... 4-7
4.10 Display rotation ..................................................... 4-7
RadEye GF-Series DB-083 E 0-3 Le/Ff 20.07.2015
Page 6
Set Date and Time ................................................. 4-8
4.11
4.12 Battery type ........................................................... 4-8
4.13 BTcom cover ......................................................... 4-9
4.14 Source test ........................................................... 4-10
4.15 Change menu language ....................................... 4-12
4.16 Adjust display contrast ........................................ 4-12
4.17 Graphic display ................................................... 4-12
5. Functional test ................................. 5-1
5.1 Functional test ........................................................... 5-1
5.2 Failure indication ...................................................... 5-1
5.3 Overload indication ................................................... 5-3
5.4 Low Battery .............................................................. 5-4
5.5 Calibration expiration date ........................................ 5-5
6. PC configuration .............................. 6-1
6.1 Ways to connect to a PC ........................................... 6-1
6.1.1 Wireless connection via Bluetooth .................... 6-1
6.1.2 Cable connection ............................................... 6-2
0-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 7
Installation of the optional RadEye.EXE software ... 6-3
6.2
6.3 Starting RadEye.exe .................................................. 6-3
6.3.1 RadEye Device Parameters ............................... 6-4
6.3.2 Measurements .................................................... 6-4
6.3.3 Creating a Measurement File ............................. 6-5
6.3.4 Select serial interface ......................................... 6-8
6.4 Configuration ............................................................ 6-8
6.4.1 General configuration ...................................... 6-10
6.4.2 Text .................................................................. 6-11
6.4.3 Mode settings ................................................... 6-12
6.4.4 Source test ....................................................... 6-13
6.4.5 Alarm setting ................................................... 6-14
6.4.6 Bluetooth ......................................................... 6-15
6.4.7 Menu item ........................................................ 6-16
6.5 History .................................................................... 6-17
6.6 Logbook .................................................................. 6-18
7. Trouble Shooting ............................. 7-1
7.1 RadEye doesn’t power on ......................................... 7-1
7.1.1 Press “On” for at least one second ..................... 7-1
7.1.2 Check battery ..................................................... 7-1
7.2 Reading data from the start up screen ....................... 7-2
7.3 The RadEye doesn’t show the menu items as
described in the manual ............................................ 7-2
RadEye GF-Series DB-083 E 0-5 Le/Ff 20.07.2015
Page 8
The RadEye shows an error message in the
7.4
display’s top line ....................................................... 7-3
7.5 The RadEye is not found in the software .................. 7-3
7.6 RadEye lost date and time settings ........................... 7-5
7.7 RadEye’s battery status is lower than expected ........ 7-6
7.8 History data shows the wrong time and date ............ 7-6
8. Maintenance ..................................... 8-1
8.1 Recommended maintenance ..................................... 8-1
8.1.1 Source test ......................................................... 8-1
8.1.2 Cleaning the instrument ..................................... 8-1
8.1.3 Remove batteries for storage ............................. 8-2
9. Spare Parts ...................................... 9-1
9.1 Available spare parts ................................................. 9-1
9.2 Recommended spare parts ........................................ 9-1
10.
Optional Accessories ................. 10-1
10.1 RadEye Holster (425067046) .............................. 10-1
10.2 Earphone for RadEye series (425067037)........... 10-1
0-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 9
Ba-133-Adapter for RadEye (425067072) .......... 10-2
10.3
10.4 Desktop holder (425067060) ............................... 10-2
10.5 IR connection cable serial (4254029) .................. 10-2
10.6 IR connection cable USB (4254026) .................. 10-3
10.7 Bluetooth battery cover (425067087) ................. 10-3
10.8 RadEye car- and wall holder with accumulator
charging circuit .................................................... 10-4
10.8.1 Accessories for dash board mounting of car holder 42506/7065 (including charging function) holder 42506/7060 (mechanical holder only) .. 10-8
10.8.2 Accessories for data transmission using the
car adaptor ..................................................... 10-10
10.8.3 Accessories for AC – operation of the
car adaptor ..................................................... 10-10
10.9 RadEye Area Monitor with alarm contacts
(4250680) .......................................................... 10-11
10.10 RadEye inductive charger ................................. 10-11
10.10.1 LED indicators ........................................... 10-12
11.
Technical data ............................. 11-1
11.1 RadEye GF /GF10 ............................................... 11-1
RadEye GF-Series DB-083 E 0-7 Le/Ff 20.07.2015
Page 10
RadEye inductive charger ................................... 11-5
11.2
11.3 Firmware revisions ............................................ 11-12
V 2.02 ........................................................................... 11-12
V 2.05 ........................................................................... 11-12
V 2.06 ........................................................................... 11-12
V 3.01 ........................................................................... 11-12
V 3.06 ........................................................................... 11-13
12.
Annex ........................................... 12-1
12.1 Factory Settings ................................................... 12-1
13.
Service contact ........................... 13-1
0-8 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 11
TABLE OF FIGURES:
Diagram 11-1: Energy dependence in direction of max.
response, perpendicular to ref. mark, ambient
equivalent (RadEye GF) ................................................. 11-6
Diagram 11-2: Energy dependence in direction of max. ...........
response, perpendicular to ref. mark, ambient
equivalent (RadEye GF-10) ........................................... 11-6
Diagram 11-3: Angular response, horizontal plane
(RadEye GF) .................................................................. 11-7
Diagram 11-4: Angular response, horizontal plane
(RadEye GF-10) ............................................................. 11-8
Diagram 11-5: Angular response, vertical plane
(RadEye GF) .................................................................. 11-9
Diagram 11-6: Angular response, vertical plane
(RadEye GF-10) ........................................................... 11-10
Diagram 11-7: Statistical standard deviation of the measured
value with constant dose rates
(RadEye GF / GF-10) ................................................... 11-11
RadEye GF-Series DB-083 E 0-9 Le/Ff 20.07.2015
Page 12
SAFETY INSTRUCTIONS
The dose rate measuring range of the RadEye GF-10 is limited to 3 Sv/h (300 R/h for RadEye GF).
Correct dose integration is performed only, if the instrument does not display over load “OVERLOAD”.
The device is not intended for use as a legal personal dose meter.
The RadEye GF/GF-10 is not well suited for pulsed radiation. There will be a significant under estimation at dose rates dur­ing pulse of more than 1.7 R/h / 17 mSv/h.
Do not use the unit if error messages appear on the screen.
The earphone connector at the bottom side of the instrument must be exclusively used by equipment that is specified for use with RadEye GF / GF-10.
The instrument must not be used in explosive atmosphere.
0-10 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 13
WEEE Compliance:
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Fisher Scientific compliance with these Directives, the recyclers in your country, and in­formation on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Di­rective are available at www.thermo.com/WEEERoHS
RadEye GF-Series DB-083 E 0-11 Le/Ff 20.07.2015
Page 14
Page 15
1. Introduction
The pocket sized RadEye is a sensitive and rugged device to measure the ambient equivalent dose rate of gamma and X-ray radiation. . In this context, it is important to note, that users may select “R/h”, “Sv/h” “Gy/h” or “rem/h” as a display unit, with the understanding that the energy response characteristic is not altered by this selection. Within the RadEye GF/GF-10 the relation 1 R = 1 rem =10 mSv = 8.77 mGy/h is used.
The characteristic feature of the RadEye GF/GF-10 is the use of sophisticated low power technology components and mi­croprocessor based fully automatic self-checks. No mainte­nance is required.
The last 1600 mean and maximum values of the dose rate are recorded internally and can be read out via serial interface.
Additionally the device logs the last 250 alarms, errors and changes of the configuration. All events can be read out via serial interface.
A real time clock is provided to add a time stamp to all buffer data.
RadEye GF-Series DB-083 E 1-1 Le/Ff 20.07.2015
Page 16
All or selected menu functions described in 3.1 can be con­figured to be invisible and inaccessible by the user. Thus, the instrument can be configured to both, an extremely simple mode allowing just LCD-illumination and alarm ac­knowledgment to a very versatile mode for the more experi­enced user.
Alarm LED showing radiation exceeding thresh­olds. Visible from top if worn in a holster and from display
Rugged plastic enclosure
Key on top side to acknowledge alarms if worn in a holster
Powerful graph­ical display
Easy to use, four key, menu driven user interface
Earphone connector
1-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 17
2. Installation and start-up
2.1 Scope of delivery
The RadEye is delivered in a folding paper box together with two AAA cells and an operating manual.
A rubber protection sleeve, a holster and an earphone may be added optionally.
Ordering information for the accessories see chapter 10.
RadEye GF-Series DB-083 E 2-1 Le/Ff 20.07.2015
Page 18
2.2 Inserting the battery
The two AAA-Alkaline cells as delivered with the instrument allow up to 900 h of normal operation. AAA size rechargeable batteries can be used as well.
Switch off the measuring instrument.
If supplied, remove rubber sleeve.
Open the cover of the battery compartment.
Use of a coin is recommended.
Exchange the batteries according to the shown polarity.
Close the compartment cover hooks first, care for the rub-
ber seal being in its groove.
Switch on the unit again. (see chapter 2.4).
2-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 19
The instrument continues working in the operating mode set last. The measured values in the history memory remain stored. The real time clock for time stamp of history values and log­book continue operation if batteries are exchanged immediate­ly. If Real Time Clock is set, actual time and date is displayed for 3 s. RTC will always be reset, if instrument is stored without bat­teries for more than 10 seconds. History data and measurement parameters are stored perma­nently, even if batteries are removed.
RTC information is only required, if the instrument switched off and on during operation and if correct timing information for history and logbook entries is needed.
RadEye GF-Series DB-083 E 2-3 Le/Ff 20.07.2015
Page 20
2.3 Mounting of the protection sleeve
The rubber protection sleeve improves ruggedness to mechan­ical shocks.
For mounting of the sleeve first put the instrument into the top of the sleeve. Then pull lower edges of the sleeve, one after the other into its right position.
First step Second step
Front view Rear view
2-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 21
2.4 Switching the unit on
GF or GF
-10
To switch on the RadEye, keep the ON button pressed for at least one second. The sound generator (beeper) is initiated.
Calibration date
Checksum of
The RadEye starts working with the parameters previously selected (operation mode, calibration factor, alarm thresholds etc.). If the real time clock is set, actual date and time is displayed for 3s. The RadEye first displays 0.00 µSv/h. After the first detector pulse the measurement is started.
Firmware version Battery voltage
the firmware
RadEye GF-Series DB-083 E 2-5 Le/Ff 20.07.2015
Page 22
The first alarm thresholds for dose rate is around 200 - 500 % above the typical normal background and allow sensitive, but false alarm free dose rate monitoring with a reasonable fast response time. The alarm thresholds for the dose are set to the maximum possible values, thus being deactivated. With the setting of the history recording interval of 120 s, the last 52 hours of operation will be stored in the history memory.
These default values are reasonable for the majority of appli­cations. For special applications, the parameters (except cali­bration parameters) can be changed with the help of the op­tional PC-program “RadEye.EXE” from V1.30 and the cable 42540/29 or 42540/26. Furthermore, additional monitoring modes can be activated.
The calibration parameters can only be changed in the factory using special software tools.
2-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 23
3. Configuration
3.1 Menu functions
To enter the operating menu, press the “Menu/” key. Scrolling through the single menu options is effected by the / arrow keys. The display returns to its initial default setting in case no key has been activated for more than 10 seconds.
A “”behind some menu items identifies the corresponding function as active.
The menu consists of a main menu and several sub menus.
Using the PC-Software and an interface adapter, any of the functions can be hidden. This allows the user to be given only the functions necessary to accomplish his measurement duties, thus simplifying the handling considerably.
The / keys are used to scroll through the menu.
To select a menu option, release the ◄/Menu key as the re- spective menu option has been reached.
RadEye GF-Series DB-083 E 3-1 Le/Ff 20.07.2015
Page 24
The meaning of the
/Menu key may change with the select-
ed menu. The meaning is shown on the bottom of the display.
Change Off, On Select Yes Exit
: Edit Alarm values : Switching a function on and off : Select a default display mode
: Confirmation of an action
: Exit menu In Change menu the / keys are used to change a digit of an alarm value.
3-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 25
MENU DISPLAY
Switch off RadEye is switched off. Time and stored
Backlight If selected the backlight is always on.
Measuring unit rem/h Select Dose rate (rem/h)
Operation mode Ratemeter Select Ratemeter mode
Scaler parameter PresetTimeMode Scaler mode with fixed time
Alarm Dose Rate Allows setting of Alarm 1 and 2 for dose
Alarm Dose Allows setting of Alarm 1 and 2 for dose. Clear Dose Clears the accumulated dose.. Settings Battery type Selection of battery type: rechargeable or
SUBMENU
DISPLAY
R/h Select Dose rate (R/h) Sv/h Select Dose rate (Sv/h)
Gy/h Select Dose rate (Gy/h)
Scaler Select Scaler mode
PresetCountMode Scaler mode with fixed count number Set time/count Edit preset time and preset count Auto restart Auto restart of scaler measurements
Autosend If activated, the instrument sends a data
Single Pulse Enabling and disabling of single pulse
Finder Enabling and disabling of audible radiation
Set Time/Date Set time and date of the real time clock. Source test Test the counter tube with a small source
DESCRIPTION OF THE FUNCTION
ACTIVATED BY THE LEFT KEY
data are maintained
Otherwise the backlight extinguishes after 10s
rate mode (rem/h, R/h, Sv/h).
non-rechargeable batteries. Correct low battery warning depends on this setting.
string periodically to the infrared port. This is used for radio transmission units.
indication. Activation is done by audio keys (right button and top button). Alterna­tive to “Finder”.
intensity. Activation is done by audio keys (right button and top button). Alternative to “single Pulse”.
RadEye GF-Series DB-083 E 3-3 Le/Ff 20.07.2015
Page 26
MENU DISPLAY
Alarm indication LED Enabling and disabling of the optical alarm
Show alarm Display of alarms stored in alarm log. Text info Displays the text stored in the RadEye Bluetooth Master Combo: Enable automatic connection to
BTcom Enable or disable the BTcom cover
SUBMENU
DISPLAY
Language Change menu language of RadEye Contrast Adjust the LC display contrast
Sound Enabling and disabling of the acoustic
Vibrator Enabling and disabling of the vibration
LCD LED Enabling and disabling backlight illumina-
Remote/PC Combo: Remote device
DESCRIPTION OF THE FUNCTION
ACTIVATED BY THE LEFT KEY
alarm and the audible indication of key­strokes
alarm
tion on alarm
another BTcom cover
3.1.1 Main menu
The menu offers the following displays:
The above illustration depicts all menu options available or possible.
3-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 27
3.1.2 Measuring unit
This submenu is used to select the measuring unit for rateme­ter and scaler measurements. The native units are as follows: RadEye GF: x [R/h] and dose: Hx[R]
RadEye GF-10: *(10) [Sv/h] and dose: H*(10)[Sv]
Please Note: If the native unit is changed by menu, the displayed measure­ment values are converted. This is not a conversion with ener- gy correction!
3.1.3 Operation mode
This submenu is used to select between ratemeter and scaler mode. The “
RadEye GF-Series DB-083 E 3-5 Le/Ff 20.07.2015
” show the active configuration.
Page 28
3.1.4 Scaler parameter
This submenu is used to set the parameters for scaler meas­urements. Preset TimeMode is used to measure a defined time. Preset CntMode is used to measure a defined number of counts. With “Auto restart” the RadEye restarts automatically the measurement. “Set Time/Count” edit the preset time and count.
3.1.5 Settings
This submenu is used to select between audible single pulse indication and finder mode. The “” show the active configu­ration. With “Autosend” the RadEye sends a data string periodically to the infrared port. “Batt.type” is used to set the correct threshold for warning message “LOW BATT”.
3-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 29
“Source test” performs a test for the counter tube with a small source (see chapter 4.14). “Language” changes the menu language between “English”, “German” “French” and Russian. A “ configuration.
Set the display contrast with menu item “Contrast”.
3.1.6 Alarm indication
The submenu alarm indication is used to select the LED, LCD LED, sound and vibrator for alarming. A “” shows the active configuration.
” shows the active
3.1.7 Bluetooth
“Master”: Enable automatic connection to another Btcom cov­er module.
“Remote/PC”: This module can be connected to the Master module “BTcom”: Enable or disable the BTcom cover.
RadEye GF-Series DB-083 E 3-7 Le/Ff 20.07.2015
Page 30
Please read see also chapter 6.4.6. For more information
in
dicat
ion
value
threshold 2
please read the manual DB-093 E.
3.2 Ratemeter
The standard display is indication of the actual dose rate.
Battery
Hearbeat
Bar graph overview
Alarm threshold 1
Alarm
Pressing the up /Info – key shows upon the
- 1st click: the accumulated dose Additionally the remaining time until the dose alarm 1 is reached, assuming the current dose rate will persist.
- 2nd click: the mean value and max value of the measured dose rate, together with the time indicated since the last reset.
- 3rd click: standard display
Measurement
3-8 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 31
rd
After 10 seconds or after 3
click standard display with dose
rate indication is activated again.
1st click 2nd click
3.2.1 Trend indication
A trend indication is given if the measured dose rate is increas­ing or decreasing.
For accurate measurements, only readings without trend indi­cation should be used. A trend indication is not given at dose rates of less than 8 µSv/h.
RadEye GF-Series DB-083 E 3-9 Le/Ff 20.07.2015
Page 32
3.3 Scaler
To enter the scaler mode select in the submenu “Operation mode” menu item „Scaler“.
The scaler mode starts with the parameter set in submenu „Scaler Parameter“.
Preset time
Elapsed time
Action of /Info key
To start the measurement, press the /Info key. While the measurement is in progress “Stop” is blinking.
Measure­ment value
3-10 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 33
3.4 Alarm thresholds
There are two alarm thresholds, each allocated to dose and dose rate.
In order to avoid dose alarms while using the instrument ex­clusively as a rate meter, the dose alarms can be set to the maximum level.
Configuration of the alarms is possible via infrared interface (chapter 6.4) or via the menu. As to the alarm activation, please also read chapter 4.2.
3.5 Setting alarm thresholds
The menu options
Alarm Dose
Rate and
Alarm Dose
allow the alarm thresholds to be modified. For this setting, the user has 255 seconds time. Changing the value is effected by pressing the left (Change) button if the corresponding “Alarm” is selected:
To increment the number, press the /▼ keys. To go on to the next digit or to quit the edit mode, menu use / keys. Once the last number has been set, quit the editing mode by pressing the “Exit”-key. Then, the value set is saved and after 10 s the unit returns to the basic display.
RadEye GF-Series DB-083 E 3-11 Le/Ff 20.07.2015
Page 34
Example: The dose rate alarm threshold must be changed.
Press “Menu” key,
Alarm Dose Rate
/ keys until
is selected
Then enter change menu by press­ing “Change” key.
Edit value by pressing up / down keys select digit by / keys Pre unit “µ”, “m” may be chosen as well.
Leaving the last digit with key: Confirm storage of edited value pressing “yes”
The actual alarm thresholds are seen at the marks on the inten­sity bar scale. The upper mark shows alarm threshold 1, the lower mark shows alarm threshold 2.
3-12 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 35
4. Operation
4.1 Audible single pulse indication and finder mode
With the single pulse indication being selected, each pulse of the detector generates a short audible signal emitted by the beeper.
An audible alarm signal caused by exceeding the alarm thresh­old is not given while single pulse indication of finder mode is active.
The single pulse indication must be enabled in the menu. It is activated and deactivated by two times pressing the or top button:
Indication of active single pulse.
RadEye GF-Series DB-083 E 4-1 Le/Ff 20.07.2015
Page 36
Finder mode:
In Finder mode, the sound frequency depends on the pulse rate of the detector. The more detector pulses above the count rate that is present at the moment of the activation of the finder mode are captured by the detector, the higher the tone.
The finder mode must be enabled in the menu. It is activated and deactivated by two times pressing the or top button:
Indication of active finder mode.
4.2 Alarm indication
Each time the first alarm threshold is exceeded, the alarm de­vices beeper, LED and the vibrator become active, if they are enabled.
Alarm 1: LED slowly blinking, two frequency alarm tone Alarm 2: LED quick blinking, continuous alarm tone Dose Alarm: LED constantly on, continuous alarm tone, vi-
brator slow.
4-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 37
The alarm tone and vibrator are acknowledged by a short key depression, the LED remains pulsing. The alarm is extin­guished, when the first alarm threshold is remained under.
4.3 Additional information
In addition to the basic display, using the menu options, re­quests for further information can be started such as e.g. the setting of the enabling and disabling of the various alarm indi­cators.
4.4 Key Lock
If “Key Lock” is enabled, pressing the “/On” key for at least 5 seconds locks the key pad:
RadEye GF-Series DB-083 E 4-3 Le/Ff 20.07.2015
Page 38
It is recommended to lock the keys when wearing the unit in the holster. Thus reduced battery time because of additional power consumption by illumination of the LCD or other unin­tended operations is avoided.
Unlocking is performed according to the LCD instructions upon pressing any key:
Press left key first, then lower key and then right key.
4.5 Process description for detection of radiation sources
Direction of maximum response
To detect hidden radiation sources, the dose rate alarm thresh­old must be set to the lowest value that does not produce fail alarms. This is typical a value of 0.3 µSv/h to 0.5 µSv/h.
4-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 39
Due to its almost 100 times higher efficiency, the RadEye PRD is more suited for this task, than the RadEye GF/GF10.
Subsequently, the operator passes the object while keeping thereby the distance between the monitor and the object to be checked as small as possible. With an alarm occurring, the operator should move the in­strument while observing the display in order to localize the position of the source. Please note that the alarm can occur up to a few seconds after passing the nearest distance to the source. If the alarm disappears after a few seconds, this alarm is to be considered as a statistically released false alarm. In case a radiation source has been detected, the radiation in­tensity is indicated. (See chapter 3.2).
4.6 Earphone
For alarming in noisy environments or for undercover investi­gations an earphone is available. See ordering information page 2.1.
The earphone is connected at the bottom of the instrument after opening the rubber protection seal.
RadEye GF-Series DB-083 E 4-5 Le/Ff 20.07.2015
Page 40
4.7 Alarm latching
alarm
number
Via PC-program it is possible to configure a alarm latching. Alarm latching means the alarm is stored for the configured time from 0 s (= off) to 9999 s. A latched alarm is indicated by an “M” near by the alarm indi­cator. When the alarm is acknowledged the RadEye returns to normal operations.
Latching indication
4.8 Show alarm
By selecting the menu point “show alarm” the alarms stored in the alarm log are displayed together with the time of alarm and the actual time.
Consecutive
Stored alarm
Timestamp of
4-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
the stored
Page 41
By pressing the up and down arrow keys you can scroll through the alarm log. Pressing the “ returns to the menu.
” button the display
4.9 Text Info
Via PC program “RadEye.exe” it is possible to place text in­formation in the RadEye. This text information can be dis­played with the menu function “Text Info”.
4.10 Display rotation
If enabled, a short press on the lower centre key (arrow down) flips the display. If the lower centre key is pressed again or if the menu button is pressed, the screen flips into the normal orientation.
RadEye GF-Series DB-083 E 4-7 Le/Ff 20.07.2015
Page 42
4.11 Set Date and Time
Setting of date and time takes place via menu function Setting Set Date/Time.
The format is YYYY-MM-DD hh:mm. The fields are selected with the keys /. With the keys /the Year, month, day, hour or minute can be changed. After setting the date and time the real time clock is set and is used for alarm logbook and history.
4.12 Battery type
Selection of battery type: Rechargeable or non-rechargeable batteries. It is needed for correct low battery warning. Type “Alk.” has a threshold of 2.1V. Type “NiMH” has a threshold of 2.35V. See also chapter 5.4.
Battery type “Battery” Battery type “Accumulator”
4-8 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 43
active
4.13 BTcom cover
For detailed installation and operation procedure refer to BTcom cover manual DB-076 E which is delivered together with the Bluetooth battery lid. The BTcom cover applies to the RadEye switched off.
With menu item “BTcom”, the user may switch on / off the BTcom cover. “” means BTcom cover is active. With “Master”, the BTcom cover connects automatically to another BTcom cover. In this case, the RadEye cannot be con­nected to the PC. For more information please read the manual DB-093 E. By selecting “Remote/PC”, the BTcom cover is able to con­nect to a PC.
If the BTcom cover is active but not connected to any PC, a Bluetooth symbol is flashing. If the BT-communication is established, the symbol is always on.
BTcom cover
RadEye GF-Series DB-083 E 4-9 Le/Ff 20.07.2015
Page 44
4.14 Source test
ful
source test
Source test is started with menu item “Source test” in submenu “Settings”.
Source test starts with the parameters previously set with radeye.exe. Furthermore it is necessary to set the actual date and time. Start the source test without source. The RadEye check the background value. If the background is lower than 0.15 µSv/h the user is asked to place the source.
Date and time of last success-
Start source test
Low range test High range test
by placing the source
After placing the source, source test is started automatically.
4-10 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Terminate source test with “” key
Page 45
During measurement the display shows the actual measure­ment value. After measurement the deviation is calculated and displayed. The test is successful when the deviation is lower between -20% and +20%. During source test the alarm LED is active.
Measured devi­ation
Test result
Actual meas­urement value
At the end of source test, is displayed. If every test is success­ful the actual timestamp is stored and displayed together with the message “Ready”. If source test fails, the message “Inspec­tion” is displayed.
Source test is successful Source test fails. RadEye
should be sent to inspection
Remaining time
RadEye GF-Series DB-083 E 4-11 Le/Ff 20.07.2015
Page 46
4.15 Change menu language
Dose
rate
Maximum scale
Via menu item “Language” in submenu “Settings” the user may change the language of the RadEye menu.
4.16 Adjust display contrast
Here, the user can adjust the display contrast. This function is called in sub menu “Settings” “Contrast”.
4.17 Graphic display
If enabled, one or more short presses on the “On/” key dis­play a graphical view of the dose rate. This display is only available in ratemeter mode.
Alarm threshold 1
4-12 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 47
5. Functional test
The radiation meter performs continuous self-check routines. A complete failure of the detector will be detected in around 2 minutes and will be indicated on the LCD and be announced by the beeper. The same applies to the battery voltage.
5.1 Functional test
To carry out a simple test, shortly press any key. A short audi­ble pulse has to be released and the LC display is illuminated for some seconds.
The heart symbol next to the battery indicator must be “beat­ing”. This indicates that the cyclic tasks as calculating meas­urement values and checking for alarm thresholds are active.
5.2 Failure indication
In case of a failure, the beeper generates a sharp single pulse every 32 s. The corresponding failure message is displayed in the LCD:
RadEye GF-Series DB-083 E 5-1 Le/Ff 20.07.2015
Page 48
Error high voltage generation.
No detector pulse within the last 30 minutes.
EEPROM with calibration data shows EEPROM Read or EEPROM Write error.
A Watchdog Error indicates that the microcontroller has problems to work on its tasks in a given timeframe. Reasons are strong electromagnetic pulses, firmware errors or hardware issues.
The BTcom cover failed to initialize the Bluetooth module.
5-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 49
5.3 Overload indication
For dose rates of more than 3 Sv/h (300 R/h, 300 rem/h,
2.6 Gy/h) an overload indication is given:
The dose value is marked with an overload indication, if a dose rate overload occurred since the last dose reset.
Overload indication in dose display
RadEye GF-Series DB-083 E 5-3 Le/Ff 20.07.2015
Page 50
5.4 Low Battery
If the battery voltage is below the configured threshold (see chapter 4.12) the following warning appears:
The beeper generates a single pulse every 32s. This acoustic warning can be suppressed by pressing the alarm acknowledge key. After 8h this warning comes up again. The battery needs to be changed. However, the RadEye can still be operated for several hours.
If the battery voltage falls below 2V, the battery symbol starts flashing. An acoustic warning is generated every 8s and can’t be acknowledged. The battery needs to be changed as soon as possible.
5-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 51
5.5 Calibration expiration date
If the expiration date is enabled and date is expired, the fol­lowing warning message is shown at start up of the RadEye:
If the expiration date is enabled the RTC is not set (e.g. after change of batteries) the following warning message is shown at start up.
After acknowledge the RadEye works as usual. The expiration date can be changed with the tools “CalRadEye” or “RadEyeSetCalibDate”.
RadEye GF-Series DB-083 E 5-5 Le/Ff 20.07.2015
Page 52
Page 53
6. PC configuration
6.1 Ways to connect to a PC
A RadEye unit can be connected to a PC either by either serial interface, USB interface or by Bluetooth TM 1.
6.1.1 Wireless connection via Bluetooth
Details about the Bluetooth communication are provided through the manual DB-076 E delivered with the optional Bluetooth battery cover. Please note that the connected PC must be equipped with a Bluetooth interface as well.
1
Bluetooth is a trademark of Bluetooth SIG, Inc., Bellevue, Wash-
ington, U.S.A.
RadEye GF-Series DB-083 E 6-1 Le/Ff 20.07.2015
Page 54
6.1.2 Cable connection
It is recommended to use the optional RadEye desktop holder to align the RadEye’s IR window with the IR window of a connection cable.
The optional serial adapter cable is used to connect to a RS232 port of a PC.
This serial port is selected directly in the RadEye.EXE soft­ware (see also chapter 6.3.4).
Alternatively, the optional USB connection cable is used to connect to an USB port.
Using the USB connection requires the installation of driver software that will generate a virtual serial port that can be se-
6-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 55
lected in the RadEye.EXE software. The driver is delivered on the RadEye.EXE software installation CD ROM.
Installation of a connection cable Using the hardware provided and without exerting too much force, attach the connections of the adapter cable to the PC and to the RadEye desktop holder.
6.2 Installation of the optional RadEye.EXE
software
Open the file “Setup.EXE” on the optional RadEye.EXE Software installation CD Rom and follow the installation guideline. The installation generates a RadEye.EXE Icon on the desktop.
6.3 Starting RadEye.exe
Double click on the RadEye.exe symbol at the desktop Once the RadEye.EXE program has been started, device parameters are read out and displayed on the screen.
RadEye GF-Series DB-083 E 6-3 Le/Ff 20.07.2015
Page 56
6.3.1 RadEye Device Parameters
The Frame "RadEye GF(-10)" contains the unit's serial num­ber and version number of the software. Click on the Read button, the Parameters of RadEye will be read out from device and shown in the Frame.
Figure 6-1: Main Window
6.3.2 Measurements
After pressing the button Read Values the current Count rate, dose rate and dose are displayed in this frame.
6-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 57
With button Start the measured value is read from the unit at a certain polling interval. Use the scroll bar or the buttons be­neath the "polling interval" field to define the polling interval. You can select a value between 1 and 3600 seconds. The dose rate can be displayed numerically and graphically. Click on the Graphics…. A diagram is displayed that gives a representation of the dose rate values versus the time. The current measured value is added at the right-hand side, and the diagram is shifted to the left. Up to 100 measured val­ues can be represented. The graduation of the ordinate is au­tomatically adjusted to the measured values supplied by the unit. With menu item “Configuration” the user may adjust the y-axis of the graph.
Figure 6-2: Measurement graphics
6.3.3 Creating a Measurement File
The actual measured values Dose and Dose rate that are shown in the "Measurement" display field can be stored in a meas­urement file.
RadEye GF-Series DB-083 E 6-5 Le/Ff 20.07.2015
Page 58
To do this, open the File menu, click on Open Logfile..., and enter path and name. The file is stored with the file name ex­tension ".log".
If another file of the same name exists, the system asks wheth­er that file shall be overwritten or the new measured data shall be appended to the existing data.
Once you have pressed OK to confirm the entries, the meas­urement log file is created and the polling measured data is stored in the scan interval you have defined. An open meas­urement file is indicated by the name and the path of the measurement log file that appears in the top right-hand corner of the window.
To terminate data storage, open the File menu and select the Close Logfile menu item. No further data is recorded.
Open the File / View Logfile... menu to view the measurement log file.
For training and demonstration a prior recorded log file can be replayed by opening Replay logfile. With the buttons Start and Stop together with the polling in­terval the replay can be controlled.
Close Replay switches back to accessing measurement values via infrared interface.
6-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 59
The measurement log file is an ANSI text file with columns that are separated by <TAB>. This enables this file to be read easily into other programs (such as Excel) where the data can be processed.
The first line of the measurement log file contains the unit name, the file name, and the path. Serial number and Device identification are specified on the second line.
Date and time of the measurement are specified in the columns under the field names "mm.dd.yy" and "hh:mm:ss". The time setting corresponds to the PC system time.
The "Counter" column contains the numerical value of the counter measurement. The "Unit" column informs about the unit. The "Dose Rate" column contains the numerical value of the dose rate measurement. The "Unit" column informs about the unit (Sv/h rem/h or R/h).
RadEye GF-Series DB-083 E 6-7 Le/Ff 20.07.2015
Page 60
6.3.4 Select serial interface
Via Configuration / Com settings... menu another window is opened from which you may select the corresponding inter­face.
If the RadEye is connected via BTcom cover select “Connect­ed via Bluetooth”. “Read RadEye2 is used to check the con­nection to the device. “Search RadEye” scans all Com-Ports to detect a RadEye. This action may need some time, depending on the number of available Com-ports.
6.4 Configuration
The configuration of the RadEye can be protected with a password.
6-8 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 61
Figure 6-3: Password Setting
No password is set during the first installation of the software. A reminder window is activated, if the configuration window is opened.
Figure 6-4: Password reminder
On the Configuration / RadEye … menu, the following pa­rameters can be modified:
Unit
Time interval of the history
Temperature display
RadEye GF-Series DB-083 E 6-9 Le/Ff 20.07.2015
Page 62
Acoustic indication
Alarm level for all basic displays
Signaling types
Additional surveillances
Active menu functions
Settings for source test
6.4.1 General configuration
Figure 6-5: General configuration
On the “Generals” tab, the user may set the time interval for the history memory (1...43200s), the kind of the acoustic rate indication (single pulse, finder – see chapter 4.1), temperature display, type of battery and keyboard functions.
6-10 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 63
6.4.2 Text
Figure 6-6: Text
On the tab “Text” it is possible to store a text in the RadEye. In the field “Text Info” the user can define 4 lines text, which can be displayed on the LCD by the RadEye (see 4.10). In the “Text Field” it is possible store a text with up to 200 characters in the RadEye. This text can not be displayed by the RadEye.
RadEye GF-Series DB-083 E 6-11 Le/Ff 20.07.2015
Page 64
6.4.3 Mode settings
Figure 6-7: Mode settings
On the “Mode” tab, the user may set the operation mode, the physical unit and parameters for scaler mode.
6-12 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 65
6.4.4 Source test
Figure 6-8: Source test parameter
This tab is used to provide the function “source test” with the right data.
Nuclide: This name is prompted in function “source test” Reference dose rate: dose rate of this nuclide at reference date Reference Date: reference date for this source Half life: Half life time of this nuclide. Last source test: date and time of the last successful source
test
RadEye GF-Series DB-083 E 6-13 Le/Ff 20.07.2015
Page 66
6.4.5 Alarm setting
Figure 6-9: Alarm setting
The tab “Alarm” offers the user the possibility of making the alarm settings for dose rate and dose. With “alarm thresholds” “read only” the thresholds cannot be modified by RadEye. Alarm indicator “Beeper”, “LED” and “Vibrator” has the same consequence as the menu item “Alarm indication”.
6-14 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 67
6.4.6 Bluetooth
Figure 6-10: Bluetooth
Here the user may configure the BTcom battery cover. The frame “Device” informs about firmware version, serial number and the physical address of the BTcom cover (not the RadEye). The Parameter “Blue LED on” relates to the blue LED at the backside of the battery cover that shows the connection status. Setting “Power off at battery low” the BTcom cover powers down automatically if battery low status is detected to increase battery life of the RadEye. “Secure connection” uses the PIN code that can be edited in the right field. For more information please read the manual which is deliv­ered together with the BTcom cover.
RadEye GF-Series DB-083 E 6-15 Le/Ff 20.07.2015
Page 68
6.4.7 Menu item
Figure 6-11: Menu selection
On the tab “Functions”, the functions available to the user can be selected to speed up access to frequently needed functions. For example, if LED alarm is always active and rate level indication is never used, these functions can be hidden for the user. Once the setting of the parameters is finished, these parame­ters have to be sent to the RadEye by clicking on the Write to
RadEye ... button. Parameters including the selected menu configuration can be saved with File / Save as… as parameter file (*.cfg). A saved configuration can be reloaded using the menu File / Open… and sent to RadEye.
6-16 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 69
6.5 History
Via the Options / History menu, the values stored in the RadEye data memory can be read out, represented in an x/t diagram and saved to the hard disk of the computer. These data subsequently can be read in and further processed in a spreadsheet program. Time interval of History storing can be set from 1s to 43200 s (12 hours). 1600 measurement values can be stored. The following figure depicts for example the curve of the dose rate over the last two days at a time resolution of 120 s. Clear­ly various levels resulting from different locations and points with high peaks can be recognized. The blue line shows the mean values, the red line the maxi­mum value within the time interval. If batteries are removed, the time reference is lost. In the dia­gram, time information is only provided for history values with valid timing entries.
Figure 6-12: History read out
RadEye GF-Series DB-083 E 6-17 Le/Ff 20.07.2015
Page 70
Figure 6-13: Single history value indication
Clicking at the graph, each individual, stored measured value can be displayed.
6.6 Logbook
Changes in configuration, occurring alarms and errors are logged in a buffer. These saved events can be read out via Options / Logbook… . The logbook is shown as a table, and can be saved to PC hard disc or printed. The logbook has a maximum of 250 data sets. Several events at the same time are saved as one record. At the display every event is shown in one line for better overview. The date and time of the PC is used for time relation.
6-18 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 71
Figure 6-14: Logbook
RadEye GF-Series DB-083 E 6-19 Le/Ff 20.07.2015
Page 72
Page 73
7. Trouble Shooting
7.1 RadEye doesn’t power on
7.1.1 Press “On” for at least one second
Keep the button “” pressed until the LED and the sound is active. If the button is pressed only shortly, then the start up routine is not activated. The LED may flash then only very shortly.
7.1.2 Check battery
If the RadEye does not respond to a long press (> 1s) of the “” button, then please use the following step by step proce­dure: A) Remove the current batteries B) Clean the battery contacts C) Check the required orientation of the batteries according to the printed “+” symbols in the battery compartment. D) Insert two -“AAA” with sufficient capacity: Rechargeable AAA batteries should be completely loaded, non recharge­ able batteries should supply together at least 2.3 Volt under load. E) Press the “
7.1.1).
” button for more than one second (see also
RadEye GF-Series DB-083 E 7-1 Le/Ff 20.07.2015
Page 74
7.2 Reading data from the start up screen
GF or GF
-10
The start up screen shows important data like the last calibra­tion, the current voltage and the firmware version.
This screen stays active as long as the “” button is kept pressed during the power on of the RadEye device.
7.3 The RadEye doesn’t show the menu items as described in the manual
The items available in the RadEye menu can be disabled or enabled via the RadEye.EXE software. The factory setting has all menu functions enabled. Please refer to chapter 6.4, menu settings. Next, please check the firmware version of your RadEye (see chapter 2.4) and compare your version number with the revision table of this manual at the beginning of the document. The revision table shows the introduction of new functions together with the according firmware version.
7-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 75
7.4 The RadEye shows an error message in the display’s top line
Please refer also to chapter 5.2. If one of the following failures is indicated, then the RadEye must be repaired by the Thermo Scientific bench repair team:
HV-Error Detect Err EEPROM Err Watchdog*
Please contact our Customer Service for more details.
*) Please see the failure description in the above referenced chapter. If the failure was related to a strong electromagnetic pulse scenario and the failure went away afterwards, then a repair may not be necessary. A diligent observation of the RadEye proper operation after the failure is strongly recom­mended.
7.5 The RadEye is not found in the software
If the connection between the RadEye and the computer can­not be established with a selected serial port (COM port), then the program shows the bellow message first:
RadEye GF-Series DB-083 E 7-3 Le/Ff 20.07.2015
Page 76
The RadEye.EXE software offers to search for a connection to the RadEye unit on other available serial ports.
If there is a faulty connection between the unit and the com­puter, the program shows the error message “Data transfer to instrument fails”.
In this case, you must check whether:
The RadEye has been positioned correctly in the desktop
holder (with rubber boot) and the RadEye is turned on.
The infrared transmission windows are clean.
The correct serial interface has been selected.
7-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 77
It is recommended to use the Windows system tools to deter­mine the occupation of the currently active serial ports. Virtual serial ports are labeled accordingly for identification.
´
Device Manager showing the virtual driver
for USB „Prolific“ at COM 4
Open next the serial interface step window in the RadEye.EXE software: Configuration / Com settings... menu. The used serial port has to be selected next.
7.6 RadEye lost date and time settings
The date and time information is lost, each time the batteries are removed or below the required voltage for operation. The simplest way to program the date and time is to use the “syn­chronize” function of the optional RadEye.EXE software. Please refer to chapter 6.4.1. Else the current date and time can be set also through the RadEye’s menu “Settings” / “Set Date / Time”.
RadEye GF-Series DB-083 E 7-5 Le/Ff 20.07.2015
Page 78
7.7 RadEye’s battery status is lower than expected
Please check if the correct battery type is selected under “set­tings” / “Batt. type” and change the type in case. Rechargeable batteries supply a voltage of 1.2V each, while Alkaline batter­ies provide 1.5V.
Rechargeable batteries offer also a more steep drop from ac­ceptable battery voltage to “battery low” than Alkaline types.
7.8 History data shows the wrong time and date
See chapter 7-5.
7-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 79
8. Maintenance
8.1 Recommended maintenance
The RadEye GF(-10) is basically maintenance free.
8.1.1 Source test
It is recommended to perform a source test routine with the optional Ba-133 test adapter on a regular basis. The source test routine is described in chapter 4.14.
Level of usage Period between Source tests
Intensive, changing climate 3 month Normal, daily usage 6 month Low 12 month
8.1.2 Cleaning the instrument
The rubber boot should be removed and cleaned separately. The RadEye unit shall be cleaned with a moist tissue (only water as cleaner). The rubber boot shall be washed in water. Do not use aggressive chemicals like organic solvents.
RadEye GF-Series DB-083 E 8-1 Le/Ff 20.07.2015
Page 80
8.1.3 Remove batteries for storage
It is highly recommended to remove both AAA batteries from the battery compartment, if the RadEye unit is stored for a longer period (> 1 week) to avoid damage through battery leakage.
8-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 81
9. Spare Parts
9.1 Available spare parts
Item Part Number
Front foil RadEye GF-10 red 42506703066 Front foil RadEye GF yellow 42506703065 Battery cover complete 425067033 Rubber shock protection (rubber boot) Battery IEC-LR03 Micro 1,5V ENERGIZER E92 (1 pack = 2 batteries)
9.2 Recommended spare parts
42506703018
SM164600012
Battery IEC-LR03 Micro 1,5V ENERGIZER E92 (1 pack = 2 batteries)
RadEye GF-Series DB-083 E 9-1 Le/Ff 20.07.2015
SM164600012
Page 82
Page 83
10. Optional Accessories
10.1 RadEye Holster (425067046)
The clip at the backside of the holster allows convenient carry­ing of the RadEye at the belt.
RadEye not included
10.2 Earphone for RadEye series (425067037)
The usage of the earphone is recommended, if the RadEye is switched into FINDER mode and disturbance through noisy environment shall be compensated.
The earphone is plugged into the RadEye’s socket at bottom of the unit. The cable length is 1.2m.
RadEye GF-Series DB-083 E 10-1 Le/Ff 20.07.2015
Page 84
10.3 Ba-133-Adapter for RadEye (425067072)
The shape of the adapter insures that the Ba-133 is placed close to the reference point of the RadEye unit.
Please see chapter 4.14 and chapter 6.4.4 about the usage of the test adapter.
10.4 Desktop holder (425067060)
Alignment between the RadEye’s IR data interface window and the IR interface of a PC connection cable is insured with the usage of the desktop holder. Please see chapter 6.1.2 for details.
10.5 IR connection cable serial (4254029)
The IR plug of the connection cable can be fixed easily by screws at the desktop holder. The PC connection plug is a 9 pin sub D type for serial COM ports.
10-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 85
Please see chapter 6.1.2 for details. Cable length = 1.2 m
10.6 IR connection cable USB (4254026)
The IR plug of the connection cable can be fixed easily by screws at the desktop holder. The PC connection plug is a USB port.
Please see chapter 6.1.2 for details. Cable length = 1.2 m
10.7 Bluetooth battery cover (425067087)
The standard battery lid is replaced with the Bluetooth battery cover. Please see chapter 6.1.1 for details.
RadEye GF-Series DB-083 E 10-3 Le/Ff 20.07.2015
Page 86
10.8 RadEye car- and wall holder with accumulator charging circuit
Safety instructions
1. Only use NiMH – accumulators, no primary batteries
(Alkaline –manganese, zinc – carbon ...)
2. Only use two accumulators of the same type (manufac-
turer, rated capacity ...)
3. Only charge accumulators having the same discharging
state
4. Never use defective accumulators (penetrating/emerging
electrolyte, damaged housing...)
10-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 87
RadEye car holder : 42506/7065
The RadEye car holder 42506/7065 serves as supporting de­vice and power supply unit for the operation of a RadEye in­strument in a vehicle. For convenient mounting to the dash­board one of the mechanical adaptors 42506/7061 ... 64 is required.
Together with a power supply unit 42506/7066, this support is also suitable for a cost attractive stationary ambient monitor­ing. Via the alarm contact, power consuming external alarm indica­tor can be operated.
RadEye GF-Series DB-083 E 10-5 Le/Ff 20.07.2015
Page 88
The RadEye car- and wall holder combines several functions.
(1) Mechanical holding device for instruments of the
RadEye-family
(2) Accumulator charger unit with temperature controlled
charging of NiMH accumulators in the device
(3) Infrared interface for direct connection to the serial in-
terface of computers
(4) Alarm contact for connecting alarm indicators up to a
power of 24W (at 24V, see RadEye area monitor 42506/80)
Before plugging the RadEye into the Car holder the rubber protection plug at the phone jack has to be opened:
Carefully remove the protection sleeve and open the phone jack
10-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 89
Remount the protection sleeve (see chapter 2.3). A small screw driver for the protection plug might be helpful. The phone jack must left be open, and the protection plug must be perpendicular to the protection sleeve.
Then insert the RadEye into the car holder. Make sure that the protection plug is not mechanically interfering with the charg­ing pin of the car holder.
RadEye GF-Series DB-083 E 10-7 Le/Ff 20.07.2015
Page 90
10.8.1 Accessories for dash board mounting of car holder 42506/7065 (including charging function) holder 42506/7060 (mechanical holder only)
Goose neck adaptor kit 42506/7061
Pivot arm adaptor kit 42506/7062
10-8 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 91
Knuckle joint adaptor kit 42506/7063
Goose neck adaptor kit with suction cup 42506/7064
RadEye GF-Series DB-083 E 10-9 Le/Ff 20.07.2015
Page 92
10.8.2 Accessories for data transmission using the car adaptor
RS232 connecting cable 2m: SM1685 35223
USB 1.1 to RS232 adapter cable: SM1685 35251
USB 2.0 to RS232 adapter cable: SM1685 35255
10.8.3 Accessories for AC – operation of the car adaptor
Power supply with ciga­rette lighter socket 42506/7066 for mains operation of the car holder 42506/7065
10-10 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 93
10.9 RadEye Area Monitor with alarm contacts (4250680)
A RadEye GF can be used as installed radiation monitor if used as RadEye Area Monitor. The RadEye is permanently powered and the alarm status is visible as light and with fur­ther accessories also as sound alarm (425068010). More in­formation is provided in the manual DB-062 E.
10.10 RadEye inductive charger
The RadEye holder serves as supporting device and, in combi­nation with the special battery lid 42506/7034, power supply unit for the operation of a RadEye instrument in a vehicle. For convenient mounting to the dashboard one of the mechanical adaptors 42506/7061 … 64 is required.
RadEye GF-Series DB-083 E 10-11 Le/Ff 20.07.2015
Page 94
Note: The charging function of this device requires the battery latch 42506/7034. Without the above mentioned part no charg­ing will be achieved.
Safety instructions
1. Only use NiMH – accumulators, no primary batteries (Al-
kaline – manganese, zinc – carbon ...).
2. Only use two accumulators of the same type (manufactur-
er, rated capacity ...).
3. Only charge accumulators having the same discharging
state.
4. Never use defective accumulators (penetrating/emerging
electrolyte, damaged housing...).
5. The bottom of the housing (aluminum heat sink) gets
warm (50°C, 120°F max.).
6. Not for use in open convertibles.
7. Do not put cards with magnet stripes (f.e. credit cards,
parking cards, phone cards, etc.) near the mounted warn­ing symbol. The magnetic field could erase the data from your card.
8. Clearance of at least 50cm (20”) between charger and car
radio is recommended.
10.10.1 LED indicators State LED (green)
Voltage supply on On Voltage supply off Off
10-12 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 95
11. Technical data
11.1 RadEye GF /GF10
Radiation type: Gamma and X-ray radiation
Measured quantity: Ambient equivalent dose rate:
RadEye GF: x [R/h] and dose: Hx[R] RadEye GF-10: *(10) [Sv/h] and dose: H*(10)[Sv] Please Note:
If the native unit (see above) is changed by menu, the displayed measurement values are converted. This is not a con­version with energy correction!
Measuring range: Dose Rate: 3 Sv/h, 300 R/h, 300 rem/h or 2.6 Gy/h Dose: 10 Sv
Overload display: more than 3 Sv/h or 300 R/h
overload indication up to 10 Sv/h
Linearity error: max. ± 10 % in the measuring range
RadEye GF-Series DB-083 E 11-1 Le/Ff 20.07.2015
Page 96
Sensitivity: RadEye GF: ~1.3 cps/mR/h RadEye GF-10: ~0.13 cps/µSv/h photon radiation 660 keV (Cs-137)
Alarm threshold: Two alarm thresholds for dose and dose rate each. Default setting: see chapter 2.4 Audible alarm intensity: at least85 dB at a distance of 30
cm
Time to alarm: typ.: 215 s for background to 3 µSv/h
typ.: 40 s for background to 10 µSv/h typ.: 15 s for background to 50 µSv/h
Energy range: 50 keV – 3 MeV error less than ±30% for dose and dose rate measurement
Energy dependence: see the following diagrams
Direction of max. response: perpendicular to the device’s
longitudinal axis, on the reference mark on the back side of the unit
Reference point: on the axis of the direction of max. response, 14 mm behind reference mark.
Angular dependence: see the following diagrams
11-2 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 97
Working temperature: -20°C ... + 50°C Relative humidity: 10 ... 93 % at 35°C non condensing Operating voltage: 1.8 ... 3.3 V,
Battery low voltage starting from 2.1 V
Degree of protection: IP 65 according to EN 60 529 EMC: Disturbance emission : EN 61000-6-3
Immunity : EN 61000-6-2 Mechanical shock: Drop onto a concrete surface
0,5m without protection sleeve 1,5m with protection sleeve
Size: 96 mm x 61 mm x 31 mm without rubber protection
Weight: around 160 g including two Alkali Manganese AAA cells LR 3 around 190g including AAA cells and protection sleeve
Internal memory: The last 1600 measured values are
saved and can be read out via
PC program. Max- and mean value of
dose rate. The time interval is factory preset to 120s by default. Logbook with 250 entries for changes of configuration, occurring alarms and errors.
RadEye GF-Series DB-083 E 11-3 Le/Ff 20.07.2015
Page 98
Averaging filters: Dose Rate filter type: Advanced Digital Filter (ADF) Digital RC-Filter with time constant
1s.....100s, depending on dose rate
and dose rate changing. Power consumption: ≈ 1 mA: normal operation without
alarm signals and LCD illumination 30 mA with illuminated LC display 18 mA LED alarm 40 mA acoustic alarm 30 mA vibrator alarm
Battery service life: 900h using two alkaline AAA cells depending on the operating mode 600h using 800mAh NiMH accu
11-4 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Page 99
11.2 RadEye inductive charger
Operating voltage 11,5…15VDC Current consumption 150…200mA With supply
from 14 V, charging cur­rent 50 mA
Accumulator charging
40…52 mA current Charge time RadEye on,
20…24h Accumulator
800mAh, discharged Backlight switched off *) Charge time RadEye switched off
18…20h Accumulator
800mAh, discharged Ambient temperature -20…+50°C Operation
–40…+70°C Storage
0…40°C Accumulator
charge
Ambient pressure 300…2000 hPa
*) During operation with backlit LCD, the battery charge is sustained only.
RadEye GF-Series DB-083 E 11-5 Le/Ff 20.07.2015
Page 100
Diagram 11-1: Energy dependence in direction of max. response, perpendicular to ref. mark, ambient equivalent (RadEye GF)
Diagram 11-2: Energy dependence in direction of max. response, perpendicular to ref. mark, ambient equivalent (RadEye GF-10)
11-6 DB-083 E RadEye GF-Series Le/Ff 20.07.2015
Loading...