This product is required to comply with the European Union’s
Waste Electrical & Electronic Equipment (WEEE) Directive
2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more
recycling/disposal companies in each EU Member State, and
this product should be disposed of or recycled through them.
Further information on Thermo Fisher Scientific compliance
with these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may
assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/WEEERoHS
RadEye G20 DB-067 E 0-9
Iw/Tr/Ff 26.11.2012
Page 12
SAFETY INSTRUCTIONS
The RadEye G20 is suited to perform highly accurate dose
measurements. It is however not intended for use as a legal
personal dose meter.
The RadEye B20 is not well suited for pulsed radiation. There
will be a significant under estimation at dose rates during
pulse of more than 0.2 R/h.
In case the dose rate measuring range is exceeded, the value of
the upper end of the dose rate measuring range is used for the
dose integration.
The instrument must not be used in explosive atmosphere.
Do not use the unit if error messages appear on the screen.
The earphone connector at the bottom side of the instrument
must be exclusively used by equipment that is specified for
use with RadEye G20.
The RadEye G20 is not suited for pulsed photon radiation.
0-10 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 13
1. Introduction
The instruments of the RadEye G20 series are modern compact survey meters for sensitive gamma and x-ray dose rate
measurements. The instrument incorporates a sensitive GM
pancake tube detector allowing the dose rate measurement
from background levels. ER-versions extend the measuring
range up to 100 mSv/h (10 rem/h respectively 10 R/h).
RadEye G20-10 and RadEye G20-ER10 are energy compensated according to the SI-standard units of ambient dose
equivalent H*(10) (= “deep dose”) and show a flat response
from 17 keV to 1.3 MeV. RadEye G20 and RadEye G20-ER
are energy compensated according to roentgen units (40 keV
to 1.3 MeV). In this context it is important to note, that users
may select “R/h, Sv/h or rem/h” as a display unit, with the
understanding that the energy response characteristic is not
altered by this selection. Within the RadEye G20 family the
relation 1 R = 1 rem =10 mSv is used.
The last 1600 mean and maximum values of the count rate or
dose rate are recorded internally and can be read out via serial
interface. Additionally the RadEye G20 logs the last 250
alarms, errors and changes of the configuration. All events can
be read out via serial interface.
RadEye G20 DB-067 E 1-1
Iw/Tr/Ff 26.11.2012
Page 14
A real time clock is provided to add a time stamp to all buffer
data. The characteristic feature of the RadEye G20 is the use
of sophisticated low power technology components and microprocessor based fully automatic self checks. No maintenance is required.
All or selected menu functions described in 3.1 can be con-
figured to be invisible and inaccessible by the user.
Thus the instrument can be configured to both, an ex-
tremely simple mode allowing just LCD-illumination and
alarm acknowledgment to a very versatile mode for the
more experienced user.
Alarm LED showing
radiation exceeding
thresholds.
Key on top side to
acknowledge alarms if
worn in a holster
Powerful graphical
display
Rugged plastic
enclosure
Easy to use, four key,
menu driven user
interface
Earphone
connector
1-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 15
2. Installation and start-up
2.1Scope of delivery
The RadEye G20 instruments are delivered in a box together
with two AAA cells and an operating manual.
Ample space is provided for a data cable or a test adapter.
.
Ordering information for accessories see chapter 11
RadEye G20 DB-067 E 2-1
Iw/Tr/Ff 26.11.2012
Page 16
2.2 Inserting the batteries
The two AAA-Alkaline cells as delivered with the instrument
allow about to 900 h with RadEye G20/G20-10 respectively
700h with RadEye G20-ER/G20-ER10 of normal operation.
AAA size rechargeable batteries can be used as well.
• Switch off the measuring instrument.
• Remove rubber sleeve.
• Open the cover of the battery compartment.
Use of a coin is recommended.
• Exchange the batteries according to the shown polarity.
• Close the compartment cover hooks first, care for the rubber
seal being in its groove.
• Switch on the unit again. (see chapter 2.4)
The instrument continues working in the operating mode set last
(see chapter 3.2).
2-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 17
The measured values in the history memory remain stored.
The real time clock for time stamp of history values and logbook continue operation, if batteries are exchanged without
delays. If Real Time Clock is set, actual time and date is displayed for 3 s (see chapter 2.4).
After power on the RadEye first display zero dose rate. After
the first detector pulse the measurement will be started.
If correct timing information of logbook and history is required,
the Real Time Clock should be set by means of the PC-program
or via menu function “Set Time/Date”.
To keep RTC running during battery exchange, batteries must
be exchanged without delays.
RTC will always be reset, if instrument is stored without batteries for more than 10 seconds.
History data and measurement parameters are stored permanently, even if batteries are removed.
Even without RTC information the relative time information of
logbook and history is corrected to the actual PC clock time
during read out unless the batteries had not been removed between the event and the data read out.
RadEye G20 DB-067 E 2-3
Iw/Tr/Ff 26.11.2012
Page 18
2.3 Mounting of the protection sleeve
The rubber protection sleeve improves ruggedness to mechanical shocks.
For mounting of the sleeve first put the instrument into the bottom of the sleeve. Then pull lower edges of the sleeve, one after
the other into its right position.
First step Second step
2-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 19
2.4 Switching the unit on
(example refers to G20-10, calibrated in rem/h)
To switch on the RadEye G20, keep the “On/▼” button pressed
for at least one second. The sound generator (beeper) is initiated.
After switching the unit on, the RadEye G20 starts working
with the parameters previously selected (operation mode, calibration factor, alarm thresholds etc.).
Calibration date
Checksum of the
Battery voltage
If the real time clock is set, actual time and date is displayed for
3s.
Firmware version
firmware
The RadEye G20 first displays 0 µrem/h. After the first detector
pulse the measurement will be started.
RadEye G20 DB-067 E 2-5
Iw/Tr/Ff 26.11.2012
Page 20
With the setting of the history recording interval of 120 s, the
last 52 hours of operation will be stored in the history memory.
The parameters (except calibration parameters) can be changed
with the help of RadEye buttons and the optional PC-program
“RadEye.exe” V1.30 and the cable 42540/29 or /26.
The calibration parameters can only be changed in the factory
using special software tools.
2-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 21
3. Configuration
3.1Menu functions
To enter the operating menu, press the “Menu” key.
Scrolling through the single menu options is effected by the
“▲/Info” and “▼/On” keys.
The display returns to its initial default setting in case no key has
been activated for more than 10 seconds.
A “
9”to be found behind some menu options means that the
respective function is active.
The menu consists of a main menu and several sub menus. Some
menu options are disabled for some operation modes and display
modes.
Using the PC-Software and an interface adapter, any of the functions can be hidden. This allows the user to be given only the
functions necessary to accomplish his measurement duties, thus
simplifying the handling considerably.
To select a menu option, release the “◄/Menu” key as the respective menu option has been reached.
RadEye G20 DB-067 E 3-1
Iw/Tr/Ff 26.11.2012
Page 22
The meaning of the Menu key may change with the selected
menu. The meaning is shown on the bottom of the display.
Change: Edit values
Off, On: Switching a function on and off
Select: Select a default display mode
Yes: Confirmation of an action
Exit: Exit menu
In Change menu the “▲/▼” keys are used to change a digit of an
alarm value.
3.1.1 Menu structure
MENU
DISPLAY
Switch off RadEye is switched off. Time and stored data are
Backlight If selected the backlight is always on. Otherwise the
Measuring
unit
mode
Scaler
parameter
Alarm Dose
Rate
Alarm Dose Allows setting of Alarm 1 and 2 for dose. (rem, R,
SUBMENU
DISPLAY
rem/h Select Dose rate (rem/h)
R/h Select Dose rate (R/h)
Sv/h Select Dose rate (Sv/h)
Ratemeter Select Ratemeter mode Operation
Scaler Select Scaler mode
PresetTimeMode Scaler mode with fixed time
PresetCountMo
de
Set time/count Edit preset time and preset count
Auto restart Auto restart of scaler measurements
Allows setting of Alarm 1 and 2 for dose rate mode
DESCRIPTION OF THE FUNCTION
ACTIVATED BY THE “◄/MENU” KEY
maintained
backlight extinguishes after 10s
Scaler mode with fixed count number
(rem/h, R/h, Sv/h).
Sv).
3-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 23
MENU
DISPLAY
Clear Dose Clears the accumulated dose.
Settings
Language Change the menu language of RadEye
Alarm indication
Show alarm Display of alarms stored in alarm log.
Text info Displays the text stored in the RadEye
Bluetooth
BTcom Enable or disable the BTcom cover
SUBMENU
DISPLAY
Battery type Selection of battery type: rechargeable or non-
Autosend If activated, the instrument sends a data string peri-
Single Pulse Enabling and disabling of single pulse indication.
Finder Enabling and disabling of audible radiation intensity.
Set Time/Date Set date and time of the real time clock.
Source test Test the instrument with a small source
LED Enabling and disabling of the optical alarm
Sound Enabling and disabling of the acoustic alarm and the
Vibrator Enabling and disabling of the vibration alarm
Master Combo: Enable automatic connection to another
Remote/PC Combo: Remote device
DESCRIPTION OF THE FUNCTION
ACTIVATED BY THE “◄/MENU” KEY
rechargeable batteries. Correct low battery warning
depends on this setting.
odically to the infrared port. This is used for radio
transmission units.
Activation is done by audio keys (right button and top
button). Alternative to “Finder”.
Activation is done by audio keys (right button and top
button). Alternative to “single Pulse”.
audible indication of keystrokes
BTcom cover
PC: RadEye can be connected to a PC
RadEye G20 DB-067 E 3-3
Iw/Tr/Ff 26.11.2012
Page 24
3.1.2 Main menu
The main menu offers the following displays:
The above illustration depicts all menu options for the main menu
possible for dose rate mode.
3.1.3 Measuring unit
This submenu is used to select the measuring unit for ratemeter
and scaler measurements.
Attention:
Whether the instrument measures dose rate (G20 or G20-ER) or
dose equivalent (G20-10 or G20-ER10) depends on the hardware
of the energy filter and cannot be changed by the selected unit.
I. e. true R/h energy response is achieved only for RadEye G20
3-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 25
and RadEye G20-ER. However, for convenience reasons, the user
may select this unit for RadEye G20-10 and RadEye G20-ER10
with the understanding that the measured quantity is still dose
equivalent H*(10) and that the relation to the calibration is set as
1 R = 1 rem = 10 mSv.
3.1.4 Operation mode
This submenu is used to select between ratemeter and scaler
mode. A “9” shows the active configuration.
3.1.5 Scaler parameter
This submenu is used to set the parameters for scaler measurements. “PresetTimeMode” is used to measure a defined time.
“PresetCntMode” is used to measure a defined number of counts.
With “Auto restart” the RadEye G20 restarts automatically the
measurement. “Set Time/Count” edit the preset time and count.
RadEye G20 DB-067 E 3-5
Iw/Tr/Ff 26.11.2012
Page 26
3.1.6 Settings
In this submenu it is possible to select between audible single
pulse indication and finder mode. A “9” shows the active configuration.
With “Autosend” the RadEye sends a data string periodically to
the infrared port. “Batt.type” is used to set the correct threshold
for warning message “LOW BATT”. “Set Time/Date” sets time
and date of the RTC. Submenu “Language” is used to change the
langenage of RadEye menu.
3.1.7 Alarm indication
The submenu alarm indication is used to select the LED, sound
and vibrator for alarming. A “9” shows the active configuration.
3-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 27
beat
3.2 Ratemeter
The standard display set by factory is indication of the actual dose
rate.
Heart
Battery
indication
Bar graph
overview
Alarm
threshold 1
Pressing the ▲/Info key shows upon the
st
- 1
click: the accumulated dose (rem, R or Sv) is shown. It
displays the accumulated dose and the remaining
time until the dose alarm 1 is reached, assuming
the current dose rate will persist.
nd
- 2
click: the mean value and max value of the measured
dose rate (R/h, Sv/h or rem/h), together with the
time indicated since the last reset.
rd
- 3
click: standard display
Measurement
value
Alarm
threshold 2
RadEye G20 DB-067 E 3-7
Iw/Tr/Ff 26.11.2012
Page 28
After 10 seconds or after 3
rd
indication is activated again.
click standard display with dose rate
st
1
click 2nd click
3.2.1 Trend indication
A trend indication is given, if the measured dose rate is increasing
or decreasing.
Increasing dose rate Decreasing dose rate
For accurate measurements, only readings without trend indication should be used.
A trend indication is not given at count rates of less than 1 cps.
3-8 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 29
3.3 Scaler mode
To enter the scaler mode select in the submenu „Scaler“.
The scaler mode starts with the parameter set in submenu „Scaler
Parameter“.
Preset time
Measurement value
Elapsed time
Action of
▲/Info key
To start the measurement, press the ▲/Info key. While the measurement is in progress “Stop” is blinking.
RadEye G20 DB-067 E 3-9
Iw/Tr/Ff 26.11.2012
Page 30
3.4 Alarm thresholds
There are two alarm thresholds each allocated to dose rate and
dose.
In order to avoid dose alarms while using the instrument exclusively as a rate meter, the dose alarms can be set to the maximum
level.
Configuration of the alarms is possible via infrared interface or
via the menu.
As to the alarm activation, please also read chapter 4.2.
3.5 Setting alarm thresholds
The menu options Alarm Dose Rate and Alarm Dose allow
the alarm thresholds to be modified. For this setting, the user has
255 seconds time. Changing the value is effected by pressing the
“◄/Menu” (Change) button if the corresponding “Alarm” is selected:
To increment the number press the “▲/▼” arrow keys. To go on
to the next digit or to quit the edit mode, menu use “◄/►” keys.
Once the last number has been set, quit the editing mode by pressing the “Exit”-key. Then, the value set is saved and after 10s the
unit returns to the basic display.
3-10 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 31
Example:
The dose rate alarm threshold must be changed.
Press “Menu” key, and ▲/▼keys
until
Alarm Dose Rate is selected.
Then enter change menu by pressing “Change” key.
Edit value by pressing ▲/▼, select
digit by ◄/► keys
Pre unit “µ” and “m” may be chosen as well.
Leaving the last digit with the right
arrow key:
Confirm storage of edited value
pressing “yes”
The set points of the actual dose rate alarm thresholds are seen at
the marks on the intensity bar scale. The upper mark shows alarm
threshold 1, the lower mark shows alarm threshold 2
RadEye G20 DB-067 E 3-11
Iw/Tr/Ff 26.11.2012
Page 32
Page 33
4. Operation
4.1Audible single pulse indication and finder mode
With the single pulse indication being selected, each pulse of
the detector generates a short audible signal emitted by the
beeper.
An audible alarm signal caused by exceeding the alarm threshold is
not given while single pulse indication or finder mode is active.
The single pulse indication must be enabled in the menu.
It is activated and deactivated by two times pressing the right or top
button:
Indication of activated
single pulse indication
RadEye G20 DB-067 E 4-1
Iw/Tr/Ff 26.11.2012
Page 34
Finder mode:
In the Finder mode, the sound frequency depends on the pulse rate
of the detector. The more detector pulses above the count rate that
is present at the moment of the activation of the finder mode are
captured by the detector, the higher the tone.
The finder mode must be enabled in the menu.
It is activated and deactivated by two times pressing the right or top
button:
Indication of activated
finder mode
4.2 Alarm indication
Each time the first alarm threshold is exceeded, the alarm devices
beeper, LED and the vibrator become active, if they are enabled.
Alarm 1: LED slowly blinking, two frequency alarm tone
Alarm 2: LED quick blinking, continuous alarm tone
Dose Alarm: LED constantly on, continuous alarm tone, vibrator
slow.
4-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 35
The alarm tone and vibrator are acknowledged by a short key
depression, the LED remains pulsing. The alarm is extinguished,
when the first alarm threshold is remained under.
4.3 Additional information
In addition to the basic display, using the menu options, requests
for further information can be started such as e.g. the setting of
the enabling and disabling of the various alarm indicators.
4.4 Key Lock
Pressing the “On/▼” key for at least 5 seconds, locks the key
pad:
RadEye G20 DB-067 E 4-3
Iw/Tr/Ff 26.11.2012
Page 36
It is recommended to lock the keys when wearing the unit in the
holster. Thus reduced battery time because of additional power
consumption by illumination of the LCD or other unintended
operations is avoided.
Unlocking is performed according to the LCD instructions upon
pressing any key:
Press left key first, then lower key and then right key.
4.5 Earphone
For alarming in noisy environments or for undercover investigations an earphone is available. See ordering information page 2-1.
The earphone is connected at the bottom of the instrument after
opening the rubber protection seal.
4-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 37
4.6 Alarm latching
Via PC-program it is possible to configure an alarm latching.
Alarm latching means the alarm is stored for the configured time
from 0 s (= off) to 9999 s.
A latched alarm is indicated by an “M” near by the alarm indicator. When the alarm is acknowledged the RadEye returns to normal operations.
Latching indication
RadEye G20 DB-067 E 4-5
Iw/Tr/Ff 26.11.2012
Page 38
4.7 Show alarm
t
m
By selecting the menu point “Show alarm” the alarms stored in
the alarm log are displayed together with the time of alarm and
the actual time.
Actual date and time of
the Real Time Clock
Stored alarm
By pressing the ▲ and ▼ arrow keys you can scroll through the
alarm log. By pressing the right button the display returns to the
menu.
4-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Date and time of the
s
ored alar
Page 39
4.8 Text Info
Via PC program “RadEye.exe” it is possible to place text information in the RadEye. This text information can be displayed
with the menu function “Text Info”.
4.9 Display rotation
If enabled, a short press on the “On/▼” key flips the display. If
the “On/▼” is pressed again or if “Menu/◄” button is pressed,
the screen flips into the normal orientation. This feature is very
useful in conjunction with the sample changer 42506/901001 in
order to achieve a good reading angle when sitting at a table.
RadEye G20 DB-067 E 4-7
Iw/Tr/Ff 26.11.2012
Page 40
4.10 Set Date and Time
Setting of date and time takes place via menu function Setting →
Set Date/Time.
The setting uses the format YYYY-MM-DD hh:mm. The fields
are selected with the keys ◄/►. With the keys ▲/▼ the Year,
month, day, hour or minute can be changed. After setting the date
and time the real time clock is set and is used for alarm logbook
and history.
4.11 Battery type
Selection of battery type: Rechargeable or non-rechargeable batteries. It is needed for correct low battery warning. Type “Alk.”
has a threshold of 2.1V. Type “NiMH” has a threshold of 2.43V.
See also chapter 6.5.
Battery type “Alkaline” Battery type “NiMH”
4-8 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 41
4.12 BTcom cover
For detailed installation and operation procedure refer to BTcom
cover manual DB-076 E which is delivered together with the
Bluetooth battery lid.
The BTcom cover applies to the RadEye switched off.
With menu item “BTcom”, the user may switch on / off the
BTcom cover. “9” means BTcom cover is active
With “Master”, the BTcom cover connects automatically to another BTcom cover. In this case, the RadEye cannot be connected
to the PC. For more information please read the manual
DB-093 E.
By selecting “Remote/PC”, the BTcom cover is able to connect to
a PC.
If the BTcom cover is active but not connected to any PC, a Bluetooth symbol is flashing. If the BT-communication is established,
the symbol is always on.
RadEye G20 DB-067 E 4-9
Iw/Tr/Ff 26.11.2012
Page 42
BTcom cover
active
4.13 Source test
Source test is started with menu item “Source test” in submenu
“Settings”.
Source test starts with the parameters previously set with radeye.exe. Furthermore it is necessary to set the actual date and
time.
Start the source test without source. The RadEye check the background value. If the background is lower than 0.15 µSv/h the user
is asked to place the source. High range test is only performed
with RadEye G20-ER and G20-ER10.
4-10 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 43
t
Date and time of last successful source tes
Start source test
Low range test
High range test
After placing the source, source test is started automatically.
During measurement the display shows the actual measurement
value. After measurement the deviation is calculated and displayed. The test is successful when the deviation is between -20%
and +20%. During source test the alarm LED is permanently on.
Measured deviation
Terminate
source test with
“►” key
by placing the
source
Test result
Actual measurement value
RadEye G20 DB-067 E 4-11
Iw/Tr/Ff 26.11.2012
Remaining
time
Page 44
At the end of source test, is displayed. If every test is successful
the actual timestamp is stored and displayed together with the
message “Ready”. If source test fails, the message “Inspection” is
displayed.
Source test is successful Source test fails. RadEye
should be sent to inspection
4.14 Change menu language
Via menu item “Language” in submenu “Settings” the may
change the language of the RadEye menu.
4-12 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 45
5. Measurement units
Direction of maximum response
To detect hidden radiation sources, the alarm threshold must be
set to the lowest value that does not produce fail alarms. This is
typical a value of 0.2 to 0.3 µSv/h (20 – 30 µrem/h, µR/h).
Due to it’s almost 100 times higher efficiency for gamma emitters, the RadEye PRD is more suited for this task, than the
RadEye G20.
For measurements at spots with high local contamination respectively dose rate, the extension adapters 42506/7075, /7076 and
/7077 in conjunction with the adapter 42506/7078 can be used.
RadEye G20 DB-067 E 5-1
Iw/Tr/Ff 26.11.2012
Page 46
The user may as well select R/h as a displayed measuring unit
with the understanding that 1 R/h is simply set equal to 1 Rem/h
and that consequently energy dependant deviations to the exposure rate (and air kerma rate) will occur.
The RadEye G20-10 and G20-ER10 is factory calibrated to the
ambient dose equivalent rate H*(10) in a depth 10 mm for Cs-137
gamma radiation (662 keV).
For details of the response curves see technical data section.
5-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 47
6. Functional test
When the instrument is switched on, it will show zero dose rate
until the first detector pulse occurs.
The radiation meter performs continuous self-check routines. A
complete failure of the detector during operation will be indicated
after 1 minute on the LCD and be announced by the beeper. The
same applies to the battery voltage.
6.1 Functional test
To carry out a simple test, shortly press any key. A short audible
pulse has to be released and the LC display is illuminated for
some seconds.
The heart symbol next to the battery indicator must be “beating”.
This indicates that the cyclic tasks as calculating measurement
values and checking for alarm thresholds are active.
RadEye G20 DB-067 E 6-1
Iw/Tr/Ff 26.11.2012
Page 48
6.2 Failure indication
In case of a failure the beeper generates a sharp single pulse every
32 s.
The corresponding failure message is displayed in the LCD:
Error high voltage generation
No detector pulse within 128 seconds
EEPROM with calibration data shows
EEPROM Read or EEPROM Write
error.
A Watchdog Error indicates, that the
micro controller has problems to work
on it’s tasks in a given timeframe.
Reasons are strong electromagnetic
pulses, firmware errors or hardware
issues.
The BTcom cover failed to initialize
the Bluetooth module
6-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 49
6.3 Overload indication
If the count rate exceeds the maximum count rate, or if the dose
rate exceeds the maximum dose rate an overload indication is
given.
Maximum dose rate
RadEye G20
RadEye G20-10
RadEye G20-ER
RadEye G20-ER10
200 mR/h
200 mrem/h
2 mSv/h
10 R/h
10 rem/h
100 mSv/h
The dose value is marked with an overload indication, if a dose
rate overload occurred since the last dose reset.
Overload indication
in dose display
RadEye G20 DB-067 E 6-3
Iw/Tr/Ff 26.11.2012
Page 50
6.4 Response test with test adapter
The test-adapter 42549/48 offers a convenient way to verify the
radiation response performance of the RadEye G20. This “check
source” contains 50 g of natural Lu2O3 (which corresponds to
approximately 2400 Bq Lu-176). Due to the low resulting gamma
dose rate this yields a net dose rate of the RadEye G20-10 of approximately 0.25 µSv/h. The operational advantage of the testadapter is given by the fact that each and every individual adapter
has the same emission rate, and that due to the long life time
(3.6E10 years), which exceeds the age of the universe significantly, neither source specimen, nor half life corrections need to
be performed. Furthermore the specific radioactivity (approximately 48 Bq/g) of natural Lu2O3 is well below the threshold
values according to IATA, respectively NRC (USA) or
StrSchV(Germany) in respect to the scope of dangerous goods
definition and transport regulations of radioactive material.
At a typical background count rate < 0.1 µSv/h, it is sufficient to
verify the response within 300 seconds for a 10 % statistical uncertainty.
For a faster response verification Ba-133 check sources (or other
gamma sources) in the range of 100 – 400 kBq are suggested.
6-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 51
6.5 Low Battery warning
If the battery voltage is below the configured threshold (see chapter 4.11) the following warning appears:
The beeper generates a single pulse every 32s. This acoustic
warning can be suppressed by pressing the alarm acknowledge
key. After 8h this warning comes up again. The battery needs to
be changed. However, the RadEye can still be operated for several hours.
If the battery voltage falls below 2V, the battery symbol starts
flashing. An acoustic warning is generated every 8s and can’t be
acknowledged. The battery needs to be changed as soon as possible.
RadEye G20 DB-067 E 6-5
Iw/Tr/Ff 26.11.2012
Page 52
6.6 Calibration expiration date
If the expiration date is enabled and date is expired, the following
warning message is shown at start up of the RadEye:
If the expiration date is enabled the RTC is not set (e.g. after
change of batteries) the following warning message is shown at
start up.
After acknowledge the RadEye works as usual.
The expiration date can be changed with the tools “CalRadEye”
or “RadEyeSetCalibDate”.
6-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 53
7. PC configuration
7.1Ways to connect to a PC
A RadEye unit can be connected to a PC either by either serial
interface, USB interface or by Bluetooth
7.1.1 Wireless connection via Bluetooth
Details about the Bluetooth communication are provided
through the manual DB-076 E delivered with the optional
Bluetooth battery cover. Please note that the connected PC
must be equipped with a Bluetooth interface as well.
TM 1
.
1
Bluetooth is a trademark of Bluetooth SIG, Inc., Bellevue, Wash-
ingston, U.S.A.
RadEye G20 DB-067 E 7-1
Iw/Tr/Ff 26.11.2012
Page 54
7.1.2 Cable connection
It is recommended to use the optional RadEye desktop holder
to align the RadEye’s IR window with the IR window of a
connection cable.
The optional serial adapter cable is used to connect to a RS232
port of a PC.
This serial port is selected directly in the RadEye.EXE
software (see also chapter 7.4.4).
Alternatively, the optional USB connection cable is used to
connect to an USB port.
7-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 55
Using the USB connection requires the installation of driver
software that will generate a virtual serial port that can be
selected in the RadEye.EXE software. The driver is delivered
on the RadEye.EXE software installation CD ROM.
Installation of a connection cable
Using the hardware provided and without exerting too much
force, attach the connections of the adapter cable to the PC and
to the RadEye desktop holder.
7.2 Connection to a PC
The connection of the RadEye to a PC requires an adapter
cable.
Adapter cable Order No. 42540/29 is used to connect to a
RS232 port of a PC.
Adapter cable Order No. 42540/26 is used to connect to an
USB port. Using the USB port requires installation of driver
software first.
Using the hardware provided and without exerting too much
force, attaches the connections of the adapter cable to the PC
and to the RadEye desktop holder.
RadEye G20 DB-067 E 7-3
Iw/Tr/Ff 26.11.2012
Page 56
The following steps will show how to install the USB-driver
under Windows XP.
Power on your computer and boot to Windows.
Run the driver setup program “RadEye-USB.exe”. Click Next
to continue and start the installation.
Wait until the InstallShield Wizard informs you that driver is
successfully installed.
Click the Finish button to close the InstallShield program. If
you have plugged the adapter into the PC while running the
setup installation, please unplug and replug the adapter for the
system to detect the device.
7-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 57
7.3 Installation of the optional RadEye.EXE soft-
ware
Open the file “Setup.EXE” on the optional RadEye.EXE
Software installation CD Rom and follow the installation
guideline. The installation generates a RadEye.EXE Icon on
the desktop.
7.4 Starting the program
The icon can be used to start the software RadEye.exe by double click.
Once the RadEye.EXE program has been started, device parameters are displayed on the screen.
7.4.1 RadEye G20 Device Parameters
The Frame "RadEye G20" contains the unit's serial number
and version number of the software. Click on the Read button,
the Parameters of RadEye G20 will be read out from device
and shown in the Frame.
RadEye G20 DB-067 E 7-5
Iw/Tr/Ff 26.11.2012
Page 58
Figure 7-1: Main Window
7.4.2 Measurements
After pressing the button Read Values the current Count rate,
dose rate and dose are displayed in this frame.
With button Start the measured value is read from the unit at a
certain polling interval. Use the scroll bar to define the polling
interval. You can select a value between 1 and 3600 seconds.
The dose rate can be displayed numerically and graphically.
Click on the Graphics…. A diagram is displayed that gives a
representation of the dose rate values versus the time.
7-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 59
The current measured value is added at the right-hand side,
and the diagram is shifted to the left. Up to 100 measured values can be represented. The graduation of the ordinate is
automatically adjusted to the measured values supplied by the
unit.
Figure 7-2: Measurement graphics
Via menu “Configuration” the user is able to display the dose
and to personalize the diagram.
RadEye G20 DB-067 E 7-7
Iw/Tr/Ff 26.11.2012
Page 60
7.4.3 Creating a Measurement File
The actual measured values Count rate and Dose rate that are
shown in the "Measurement" display field can be stored in a
measurement file.
To do this, open the File menu, click on Open Logfile..., and
enter path and name. The file is stored with the file name extension ".log".
If another file of the same name exists, the system asks
whether that file shall be overwritten or the new measured data
shall be appended to the existing data.
Once you have pressed OK to confirm the entries, the measurement logfile is created and the polling measured data is
stored in the scan interval you have defined. An open measurement file is indicated by the name and the path of the
measurement logfile that appears in the top right-hand corner
of the window.
To terminate data storage, open the File menu and select the
Close Logfile menu item. No further data is recorded.
Open the File / View Logfile... menu to view the measurement
logfile.
For training and demonstration a prior recorded logfile can be
replayed by opening Replay logfile.
7-8 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 61
With the buttons Start and Stop together with the polling in-
terval the replay can be controlled.
Close Replay switches back to accessing measurement values
via infrared interface.
The measurement logfile is an ANSI text file with columns
that are separated by <TAB>. This enables this file to be read
easily into other programs (such as Excel) where the data can
be processed.
The first line of the measurement logfile contains the unit
name, the file name, and the path. Serial number and Device
identification are specified on the second line.
Date and time of the measurement are specified in the columns
under the field names "mm.dd.yy" and "hh:mm:ss". The time
setting corresponds to the PC system time.
The "Counter" column contains the numerical value of the
counter measurement. The "Unit" column informs about the
unit. The "Dose Rate" column contains the numerical value of
the dose rate measurement. The "Unit" column informs about
the unit (rem/h, Sv/h or R/h).
RadEye G20 DB-067 E 7-9
Iw/Tr/Ff 26.11.2012
Page 62
7.4.4 Select serial interface
Via Configuration / Com settings... menu another window is
opened from which you may select the corresponding interface.
Figure 7-3: Select serial port
If the connection shall be established with the optional Bluetooth battery cover, then option “Connected via Bluetooth”
must be activated. Please refer also the manual db076e delivered with the cover.
“Read RadEye” is used to check the connection to the device.
“Search RadEye” scans all Com-Ports to detect a RadEye.
This action may need some time, depending on the number of
available Com-ports.
7-10 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 63
7.5 Configuration
The configuration of the RadEye can be protected with a
password.
Æ
Figure 7-4: Password Setting
No password is set during the first installation of the software.
A reminder window is activated, if the configuration window
is opened.
Figure 7-5: Password reminder
RadEye G20 DB-067 E 7-11
Iw/Tr/Ff 26.11.2012
Page 64
On the Configuration / RadEye G20… menu, the following
parameters can be modified:
• Unit
• Time interval of the history
• Temperature display
• Acoustic indication
• Alarm level for all basic displays
• Signaling types
• Additional surveillances
• Active menu functions
• Settings for source test
7.5.1 General configuration
Figure 7-6: General configuration
7-12 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 65
On the “Generals” tab, the user may set the used filter, the
time interval for the history memory (1...43200s), the kind of
the acoustic rate indication (single pulse, finder 4.1), type of
battery, menu language and temperature.
7.5.2 User defined text
Figure 7-7: Menu selection
On the tab “Text” it is possible to store a text in the RadEye.
In the field “Text Info” the user can define 4 lines text, which
can be displayed on the LCD by the RadEye (see 4.10).
In the field “Text Field” it is possible to store a text with up to
200 characters in the RadEye. This text can not be displayed
by the RadEye.
RadEye G20 DB-067 E 7-13
Iw/Tr/Ff 26.11.2012
Page 66
7.5.3 Mode setting
Figure 7-8: Mode setting
On the “Mode“ tab, the user may set the operation mode, the
physical unit, parameters for background measurement and
parameters for scaler mode.
7-14 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 67
7.5.4 Source test
Figure 7-9: Source test parameter
This tab is used to provide the function “source test” with the
right data.
Nuclide: This name is prompted in function “source test”
Reference dose rate: dose rate of this nuclide at reference date
Reference Date: reference date for this source
Half life: Half life time of this nuclide.
Last source test: date and time of the last successful source
test
RadEye G20 DB-067 E 7-15
Iw/Tr/Ff 26.11.2012
Page 68
7.5.5 Alarm setting
Figure 7-10: Alarm setting
The “Alarm” tab offers the user the possibility of making the
alarm settings for the single basic displays. Furthermore, the
options on this tab allow beeper and additional alarm monitoring processes to be configured.
7-16 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 69
7.5.6 Bluetooth
Figure 7-11: Bluetooth
Here the user may configure the BTcom battery cover. The
frame “Device” informs about firmware version, serial number
and the physical address of the BTcom cover (not the RadEye
B-20!).
The Parameter “Blue LED on” relates to the blue LED at the
backside of the battery cover that shows the connection status.
Setting “Power off at battery low” the BTcom cover powers
off automatically if battery low status is detected to increase
battery life. “Secure connection” uses the PIN code that can
be edited in the right field.
With “Autoconnect” the BTcom cover connects automatically
to another paired BTcom cover. In this state, the BTcom cover
cannot be connected with the PC.
For more information please read the manual which is delivered together with the BTcom cover.
RadEye G20 DB-067 E 7-17
Iw/Tr/Ff 26.11.2012
Page 70
7.5.7 Menu configuration
Figure 7-12: Menu selection
On the “Functions” tab, the functions available to the user can
be selected to speed up access to frequently needed functions.
For example, if LED alarm is always active and rate level
indication is never used, these functions can be hidden for the
user.
Once the settings of the parameters are finished, these parameters have to be sent to the RadEye G20 by clicking on the
Write to RadEye G20 button.
Parameters including the selected menu configuration can be
saved with File / Save a
configuration can be reloaded using the menu File / Open…
and sent to RadEye G20.
s… as parameter file (*.cfg). A saved
7-18 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 71
7.6 History
Via the Options / History menu, the values stored in the RadEye G20 data memory can be read out, represented in a x/t
diagram and saved to the hard disk of the computer. These
data subsequently can be read in and further processed in a
spreadsheet program.
Time interval of History storing can be set from 1s to 43200 s
(12 hours). 1600 measurement values can be stored.
The following figure depicts for example the curve of the dose
rate over the last two days at a time resolution of 120 s.
Clearly various levels resulting from different locations and
points with high peaks can be recognized.
Every scaler measurement is stored as well. You can click on a
specific date point and the measured value, date and time as
well as the measuring time will be displayed.
RadEye G20 DB-067 E 7-19
Iw/Tr/Ff 26.11.2012
Page 72
For ratemeter data the blue line shows the mean values, the red
line the maximum value within the time interval.
If batteries are removed, the time reference is lost. In the diagram time information is only provided for history values with
valid timing entries.
Figure 7-13: History read out
Clicking at the graph, each individual, stored measured value
can be displayed:
Figure 7-14: Single history value indication for Ratemeter data
7-20 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 73
7.7 Logbook
Changes in configuration, occurring alarms and errors are
logged in a buffer.
These saved events can be read out via Options / Logbook… .
The logbook is shown as a table, and can be saved to PC hard
disc or printed.
The logbook has a maximum of 250 data sets. Several events
at the same time are saved as one record. At the display every
event is shown in one line for better overview. The date and
time of the PC is used for time relation.
Figure 7-15: Logbook
RadEye G20 DB-067 E 7-21
Iw/Tr/Ff 26.11.2012
Page 74
Page 75
8. Trouble Shooting
8.1 RadEye doesn’t power on
8.1.1 Press “On” for at least one second
Keep the button “T” pressed until the LED and the sound is
active. If the button is pressed only shortly, then the start up
routine is not activated. The LED may flash then only very
shortly.
8.1.2 Check battery
If the RadEye does not respond to a long press (> 1s) of the
“T” button, then please use the following step by step procedure:
A) Remove the current batteries
B) Clean the battery contacts
C) Check the required orientation of the batteries according to
the printed “+” symbols in the battery compartment.
D) Insert two -“AAA” with sufficient capacity: Rechargeable
AAA batteries should be completely loaded, non rechargeable
batteries should supply together at least 2.3 Volt under load.
E) Press the “T” button for more than one second (see also
8.1.1)
RadEye G20 DB-067 E 8-1
Iw/Tr/Ff 26.11.2012
Page 76
8.2 Reading data from the start up screen
The start up screen shows important data like the last calibration, the current voltage and the firmware version.
This screen stays active as long as the “T” button is kept
pressed during the power on of the RadEye device.
8.3 The RadEye doesn’t show the menu items as
described in the manual
The items available in the RadEye menu can be disabled or
enabled via the RadEye.EXE software. The factory setting has
all menu functions enabled. Please refer to chapter 7.5.7, menu
settings. Next, please check the firmware version of your
RadEye (see chapter 2.4) and compare your version number
with the revision table of this manual at the beginning of the
document. The revision table shows the introduction of new
functions together with the according firmware version.
8-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 77
8.4 The RadEye shows an error message in the
display’s top line
Please refer also to chapter 6.2. If one of the following failures
is indicated, then the RadEye must be repaired by the Thermo
Scientific bench repair team:
HV-Error
Detect Err
EEPROM Err
Watchdog*
Please contact our Customer Service for more details.
*) Please see the failure description in the above referenced chapter.
If the failure was related to a strong electromagnetic pulse scenario
and the failure went away afterwards, then a repair may not be necessary. A diligent observation of the RadEye proper operation after
the failure is strongly recommended.
8.5 The RadEye is not found in the software
If the connection between the RadEye and the computer can’t
be established with a selected serial port (COM port), then the
program shows the bellow message first:
RadEye G20 DB-067 E 8-3
Iw/Tr/Ff 26.11.2012
Page 78
The RadEye.EXE software offers to search for a connection to
the RadEye unit on other available serial ports.
If there is a faulty connection between the unit and the computer, the program shows the error message “Data transfer to
instrument fails”
In this case, you must check whether:
• The RadEye has been positioned correctly in the desktop
holder (with rubber boot) and the RadEye is turned on.
• The infrared transmission windows are clean
• The correct serial interface has been selected.
8-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 79
It is recommended to use the Windows system tools to determine the occupation of the currently active serial ports. Virtual
serial ports are labeled accordingly for identification.
´
Device Manager showing the virtual driver
for USB „Prolific“ at COM 4
Open next the serial interface step window in the RadEye.EXE
software: Configuration / Com settings... menu. The used
serial port has to be selected next.
8.6 RadEye lost date and time settings
The date and time information is lost, each time the batteries
are removed or below the required voltage for operation. The
simplest way to program the date and time is to use the “synchronize” function of the optional RadEye.EXE software.
Please refer to chapter 7.5.1. Else the current date and time can
be set also through the RadEye’s menu “Settings” / “Set Date /
Time”
RadEye G20 DB-067 E 8-5
Iw/Tr/Ff 26.11.2012
Page 80
8.7 RadEye’s battery status is lower than expected
Please check if the correct battery type is selected under “settings” / “Batt. type” and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batteries provide 1.5V.
Rechargeable batteries offer also a more steep drop from acceptable battery voltage to “battery low” than Alkaline types.
8.8 History data shows the wrong time and date
See chapter 8.6.
8-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 81
9. Maintenance
9.1Recommended maintenance
The RadEye is basically maintenance free.
9.1.1 Source test
It is recommended to perform a source test routine with the
optional Lutetium test adapter on a regular basis. The source
test routine is described in chapter 4.13.
Level of usage Period between Source tests
Intensive, changing climate 3 month
Normal, daily usage 6 month
Low 12 month
9.1.2 Cleaning the instrument
The rubber boot should be removed and cleaned separately.
The RadEye unit shall be cleaned with a moist tissue (only
water as cleaner). The rubber boot shall be washed in water.
Do not use aggressive chemicals like organic solvents.
RadEye G20 DB-067 E 9-1
Iw/Tr/Ff 26.11.2012
Page 82
9.1.3 Remove batteries for storage
It is highly recommended to remove both AAA batteries from
the battery compartment, if the RadEye unit is stored for a
longer period (> 1 week) to avoid damage through battery
leakage.
This test adapter contains 50g Lutetiumoxide. Typical net dose
rate for RadEye G20: 25µrem/h (25µR/h, 0.25Sv/h)
Please see chapter 4.13 and chapter 9.1.1 about the usage of the
test adapter.
11-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 87
11.4 Desktop holder (425067060)
Alignment between the RadEye’s IR data interface window and
the IR interface of a PC connection cable is insured with the usage of the desktop holder. Please see chapter 7.1.2 for details.
11.5 IR connection cable serial (4254029)
The IR plug of the connection cable can be fixed easily by screws
at the desktop holder. The PC connection plug is a 9 pin sub D
type for serial COM ports.
Please see chapter 7.1.2 for details. Cable length = 1.2 m
RadEye G20 DB-067 E 11-3
Iw/Tr/Ff 26.11.2012
Page 88
11.6 IR connection cable USB (4254026)
The IR plug of the connection cable can be fixed easily by screws
at the desktop holder. The PC connection plug is a 9 pin sub D
type for serial COM ports.
Please see chapter 7.1.2 for details. Cable length = 1.2 m
11-4 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 89
11.7 Bluetooth battery cover (425067087)
The standard battery lid is replaced with the Bluetooth battery
cover. Please see chapter 7.1.1 for details.
11.8 RadEye car- and wall holder with accumulator
charging circuit
Safety instructions
1. Only use NiMH – accumulators, no primary batteries (Alkaline –manganese, zinc – carbon ...)
2. Only use two accumulators of the same type (manufacturer,
rated capacity ...)
3. Only charge accumulators having the same discharging state
4. Never use defective accumulators (penetrating/emerging
electrolyte, damaged housing...)
RadEye G20 DB-067 E 11-5
Iw/Tr/Ff 26.11.2012
Page 90
RadEye car holder : 42506/7065
The RadEye car holder 42506/7065 serves as supporting device
and power supply unit for the operation of a RadEye instrument
in a vehicle. For convenient mounting to the dashboard one of the
mechanical adaptors 42506/7061 ... 64 is required.
Together with a power supply unit 42506/7066, this support is
also suitable for a cost attractive stationary ambient monitoring.
Via the alarm contact, power consuming external alarm indicator
can be operated.
11-6 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 91
The RadEye car- and wall holder combines several functions.
(1) Mechanical holding device for instruments of the Rad-
Eye-family
(2) Accumulator charger unit with temperature controlled
charging of NiMH accumulators in the device
(3) Infrared interface for direct connection to the serial inter-
face of computers
(4) Alarm contact for connecting alarm indicators up to a
power of 24W (at 24V, see RadEye area monitor
42506/80)
Before plugging the RadEye into the Car holder the rubber protection plug at the phone jack has to be opened:
Carefully remove the protection sleeve and open the phone jack
Remount the protection sleeve (see chapter 2.3). A small screw
driver for the protection plug might be helpful. The phone jack
must left be open, and the protection plug must be perpendicular
to the protection sleeve.
RadEye G20 DB-067 E 11-7
Iw/Tr/Ff 26.11.2012
Page 92
Then insert the RadEye into the car holder. Make sure that the
protection plug is not mechanically interfering with the charging
pin of the car holder.
11-8 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 93
11.8.1 Accessories for dash board mounting of
car holder 42506/7065 (including charging function)
holder 42506/7060 (mechanical holder only)
Goose neck adaptor kit 42506/7061
Pivot arm adaptor kit 42506/7062
RadEye G20 DB-067 E 11-9
Iw/Tr/Ff 26.11.2012
Page 94
Knuckle joint adaptor kit 42506/7063
Goose neck adaptor kit with suction cup 42506/7064
11-10 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 95
11.8.2 Accessories for data transmission using the car adaptor
RS232 connecting
cable 2m:
SM1685 35223
USB 1.1 to RS232
adapter cable:
SM1685 35251
USB 2.0 to RS232
11.8.3 Accessories for AC – operation of the car adaptor
adapter cable:
SM1685 35255
Power supply with cigarette lighter socket
42506/7066
for mains operation of the
car holder 42506/7065
RadEye G20 DB-067 E 11-11
Iw/Tr/Ff 26.11.2012
Page 96
11.9 RadEye inductive charger
The RadEye holder serves as supporting device and, in combination with the special battery lid 42506/7034, power supply unit
for the operation of a RadEye instrument in a vehicle. For convenient mounting to the dashboard one of the mechanical adaptors 42506/7061 … 64 is required.
Note: The charging function of this device requires the battery
latch 42506/7034. Without the above mentioned part no charging
will be achieved.
11-12 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Page 97
Safety instructions
1. Only use NiMH – accumulators, no primary batteries (Alka-
line – manganese, zinc – carbon ...)
2. Only use two accumulators of the same type (manufacturer,
rated capacity ...)
3. Only charge accumulators having the same discharging state
4. Never use defective accumulators (penetrating/emerging elec-
trolyte, damaged housing...)
5. The bottom of the housing (aluminum heat sink) gets warm
(50°C, 120°F max.).
6. Not for use in open convertibles.
7. Do not put cards with magnet stripes (f.e. credit cards, park-
ing cards, phone cards, etc.) near the mounted warning symbol. The magnetic field could erase the data from your card.
8. Clearance of at least 50cm (20”) between charger and car
radio is recommended.
11.9.1 LED indicators
State LED (green)
Voltage supply on On
Voltage supply off Off
RadEye G20 DB-067 E 11-13
Iw/Tr/Ff 26.11.2012
Page 98
Page 99
12. Technical data
12.1 RadEye G20-10 and RadEye G20-ER10
Radiation type:
Measured
quantities:
Measuring range:
Overload display:
Linearity error:
Alarm threshold:
Audible alarm
intensity:
Gamma, X-ray radiation
Ambient Dose Equivalent H*(10) and Rate
Maximum
dose rate
RadEye
G20-10
RadEye
G20-ER10
overload indication up to 10 Sv/h
(1000 rem/h)
max. ± 10 % in the measuring range
Two alarm thresholds for dose and dose rate
each. Default setting: see chapter 2.4
85 dB at a distance of 30 cm
2 mSv/h
200 mrem/h
100 mSv/h
10 rem/h
RadEye G20 DB-067 E 12-1
Iw/Tr/Ff 26.11.2012
Page 100
Response time
(to reach 90 %):
Photon Energy
range:
Energy depend-
ence:
Direction of max.
response:
Reference point:
Angular depend-
ence:
Working tempera-
ture:
Storage
temperature:
Relative humidity:
Operating voltage:
Degree of protec-
tion:
EMC:
Mechanical shock:
Size:
typ.: 10 s for background to 100μrem/h
(1μSv/h)
typ.: 5 s for background to 300μrem/h (3μSv/h)
typ.: 2 s for background to 1mrem/h (10μSv/h)
17 keV – 1.3 MeV: error less than ±30%
17 keV – 3 MeV: IEC 60846-1 (2009)
for dose and dose rate measurement
see Diagram 12-1
perpendicular to the center of the detector
surface
on the axis of the direction of max.
response, 10 mm depth
see Diagram 12-3 to Diagram 12-6
-20°C ... + 50°C
-25°C … + 50°C
< 93 % at 35°C non condensing
1,8 ... 3.3 V
IP 32 according to EN 60 529
Disturbance emission : EN 61000-6-3
Immunity : EN 61000-6-2
Drop onto a concrete surface
0,8 m with protection sleeve
120 mm x 70 mm x 70 mm
Without rubber protection
12-2 DB-067 E RadEye G20
Iw/Tr/Ff 26.11.2012
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.