As R&P instrumentation changes, so do our Operating and Service manuals. However, these changes may affect only one aspect of an instrument, while leaving the
instrument as a whole unchanged. To explain these individual changes to customers,
R&P will update only those sections of its Operating and Service manuals that are
affected by the instrument updates or improvements. As each manual section changes,
so does its revision number, which is located at the top right corner of each page of each
section.
To help customers keep track of the changes to the Partisol Model 2000 Air Sampler
and its service manual, following is a list of the manual sections with their respective
revision numbers:
Section Number and DescriptionRevision Number
Section 1: IntroductionB.000
Revision B.000
Section 2: Troubleshooting GuidesB.000
Section 3: Service ProceduresB.000
Section 4: Electrical SchematicsB.000
Section 5: R&P Return PolicyB.000
Appendix A: Hardware ConnectionsB.000
IndexB.000
PAGEI
Page 4
Service Manual, Partisol Model 2000 Air Sampler
Revision Description List
To help customers keep track of the changes to the Partisol Model 2000 Air Sampler
and its service manual, following is a list of the manual sections with their respective
revision descriptions. This list contains all changes made to the manual since the last
update:
Section Number and DescriptionRevision Description
NDEX ............................................................................................................. I
PAGEIV
Page 7
Service Manual, Partisol Model 2000 Air Sampler
Section 1: Introduction
The information contained in this manual will allow the user to do preliminary
troubleshooting to fix some problems or to give the maximum amount of information
available to the factory.
1.1.OVERVIEWOF MANUAL
Section 1 : Introduction
This section provides an overview of the Service Manual.
Section 2 : Troubleshooting Guides
This section contains troubleshooting guides in the form of flow charts to aid the user
when performing diagnostics.
Section 3 : Service Procedures
This section contains the service procedures (maintenance and calibration) required
to maintain consistent operation of the Partisol sampler. In addition, this section
contains procedures describing installation of system components if exchange is
necessary.
Revision B.000
Section 4 : Electronic Schematic Diagrams
This section contains schematic diagrams and interconnect diagrams of the major
electronic components in the instrument.
Section 5 : R&P Return Policy
This section contains information regarding the return of whole systems or individual
components for warranty or billable repair.
Appendix A: Hardware Connections
This section contains the pin assignments of the output connectors on the Partisol
Sampler.
SECTION 1: INTRODUCTIONPAGE 1-1
Page 8
This page left intentionally blank.
Page 9
Service Manual, Partisol Model 2000 Air Sampler
Section 2: Troubleshooting Guides
This section contains troubleshooting guides in the form of flow charts to aid the user
when performing diagnostics. This section should be used in conjunction with the
electrical schematics located in Section 4. Read Section 2.1 before using any of the
troubleshooting guides.
2.1.USINGTHE TROUBLESHOOTING GUIDES
The troubleshooting guides give the user a logical succession of steps to follow when
diagnosing a problem. Following is a list of the questions most commonly asked by
users with regard to the guides.
Revision B.000
2.1.1.W
HATEQUIPMENTDO I NEED?
The main piece of equipment needed is a multimeter (that measures DC and AC
voltages and resistance) with at least a 4 digit resolution.
2.1.2.WHEREDO I START?
Look at the main screen of the Partisol Model 2000 Air Sampler. Any active errors
will be indicated in the upper left corner. Possible error codes are:
OKGreat! You have no errors!
EPower Failure — A system power failure occurred during sampling.
FFlow Out of Range — The measured sample flow rate through the
sample filter deviated by more than 7% from its setpoint for more than
1 minute.
TAmbient Temperature Sensor — The ambient temperature sensor
was not installed correctly or indicated an invalid value. The ambient
temperature must be between -40°C and 60°C to be valid.
PAmbient Pressure Sensor — The ambient pressure sensor indicated
an invalid value. The ambient pressure must be greater than 0.4
atmospheres to be valid.
IElectronics Compartment Temperature — The temperature of the
instrument electronics compartment was outside of its usual operating range. The electronics temperature must be between -5°C and
60°C or the electronics may not operate reliably.
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-1
Page 10
Service Manual, Partisol Model 2000 Air Sampler
2.1.3.I HAVECHOSENAFLOWCHARTTOBEGINWITH. NOW WHAT?
Now that you have chosen a particular flow chart, begin in the upper left corner of the
chart and follow the “flow” of instructions until you can go no further. These
instructions will bring you through a logical succession of steps to help you determine
where the problem is and what to do about it. Document the result of each step as it
is completed. This information is important if R&P is contacted.
Revision B.000
2.1.4.H
OWDO I CHECKACABLE’SCONTINUITY?
Throughout the flow charts are instructions which indicate that a cable’s continuity
should be checked. You must check a cable’s continuity to verify that a particular
cable has not been damaged.
Follow these steps to check a cable’s continuity:
1) Set the multimeter so it will measure resistance (W).
2) Determine which cable you are checking and locate its corresponding connectors.
3) Refer to the interconnect diagrams in Section 4 to determine
which pins are connected from one connector to the other. (For
example, is pin 1 on one connector connected to pin 1 on the
other connector or to a different pin?)
4) Place one lead of the multimeter on the first pin on one of the
connectors and place the other lead on the corresponding pin of
the other connector. The resistance should be very close to zero
if the continuity is correct.
5) Repeat steps 2-4 for each set of pins being checked.
2.1.5.HOWDO I LOCATEATESTPOINTONANELECTRONICBOARD?
Refer to the schematic diagrams for the electronics in Section 4. A drawing of the
interface board is included in this section and test points are labeled.
2.1.6.HOWDO I FINDPIN 1?
There are four types of connectors used on the Partisol Monitor, “D”-type connectors,
ribbon cable connectors, 2-, 4- and 10-pin Molex connectors and round 3-, 4- and 6pin connectors.
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-2
Page 11
Service Manual, Partisol Model 2000 Air Sampler
2.1.6.1.D-TYPE CONNECTORS
There is only one “D”-type connector used on the Partisol Monitor. It is a female, 9pin, “D”-type connector labeled “RS-232” on the front panel.
Revision B.000
2.1.6.2.R
IBBON CABLE CONNECTORS
All ribbon cable connector pins are counted the same, regardless of the total number
of pins. However, the location of pin 1 may vary. The location of pin 1 is noted within
the silk-screen on the corresponding electronic board. For example purposes, a 24pin ribbon cable connector is displayed below.
2.1.6.3.M
OLEX CONNECTORS
The Molex connectors used on the Partisol Monitor are white and have 2 pins (such
as the hub valve connector (P5) on the underside of the interface board), 4 pins (such
as the AC input connector (P2) on the interface board), or 10 pins (such as the 24V/
5V transformer connector (P3) on the interface board). The easiest way to determine
which pin is pin 1 on a Molex connector is to determine what color wire corresponds
to pin 1. You can find this information in the interconnection diagrams in Section 4.
Once you determine the proper wire color, then you can examine the connector and
find pin 1.
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-3
Page 12
Service Manual, Partisol Model 2000 Air Sampler
2.1.6.4.ROUND CONNECTORS
There are three round connectors used on the back of the Partisol Monitor: a 3-pin
round connector is used for the ambient temperature sensor, a 4-pin round connector
is used for the user output, and a 6-pin round connector is used for the anemometer.
Revision B.000
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-4
Page 13
Service Manual, Partisol Model 2000 Air Sampler
2.2.TROUBLESHOOTING GUIDES
The troubleshooting guides included in the section should be used in conjunction with
the following revisions of the schematics included in Section 4 of this manual:
SchematicSchematicSchematic
DescriptionPart Number Revision
Interface Board81-005496 B
Interconnection Diagrams81-005912 C
Revision B.000
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-5
Page 14
S
ECTION 2: TROUBLESHOOTING GUIDESPAGE 2-6
Error code
“T”
This error code occurs if
the ambient temperature
is out of range.
The temperature must be
between -40°C and 60°C
to be valid.
AMBIENT TEMPERATURE ERROR TROUBLESHOOTING FLOW CHART
Note:
1. The voltage at this test point is related to the
Replace cable
R&P P/N 51-005633
No
Continuity OK?
Check continuity of cable
between P54 and ambient
temperature sensor back
panel connector (P52) (see
note 2).
Yes
Check continuity of
cable between P4
(interface board) and
P54 (see note 3).
Replace cable
R&P P/N 51-006183
No
Continuity OK?
Yes
Replace ambient
temperature sensor cable
R&P P/N 51-004352-0001
2. Pins for continuity check of cable P54 to P52:
3. Pins for continuity check of cable P54 to P4:
temperature according to the following
equation:
1°K = 10mVDC (nominal)
P54P52
Pin 1Pin 1
Pin 2Pin 2
-----Pin 3
P54P4
Pin 1Pin 1
Pin 2Pin 2
Service Manual, Partisol Model 2000 Air Sampler
2.2.1.A
MBIENT TEMPERATURE ERROR (T) FLOW CHART
Ensure ambient temperature
probe is properly attached to
the back of the instrument.
Go to the statistics screen
and observe the ambient
temperature reading.
Place mult imeter ac ross test
points AMBTEMP (blue) and
AGND (black) on interface
board (see note 1).
Yes
Voltage =
~+15V?
Place multimeter across
test point +15V (red) and
AGND (black) on interface
board.
No
Voltage
nominal?
Yes
No
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Revision B.000
Page 15
S
ECTION 2: TROUBLESHOOTING GUIDESPAGE 2-7
Display blank or
incomplete
Power s ystem
off then on.
Blank or
incomplete?
Incomplete
Check continuity of cable
P16A (interfac e board) to
P1 (display) (see note 1
on she et B).
Continui ty OK?
Yes
Install a working
display and recheck.
R&P P/N 10-002254
This section describes the service procedures required to maintain consistent operation of the Partisol Sampler. Additional maintenance procedures (such as cleaning the
PM-10 inlet) are contained in the Operating Manual. This section provides recommended intervals for maintenance and calibration procedures. However, when
determining your own intervals for maintenance and calibration, you must take into
account the quality assurance guidelines of your specific monitoring organization and
any special local considerations.
Maintenance ProceduresIntervalReference
Exchange large in-line filter6 monthsSection 3.1.1
Exchange fusesAs neededSection 3.1.2
Clean air intake filters6 monthsSection 3.1.3
Inspect “V” sealsduring filter exchangeSection 3.1.4
System leak test3 monthsSection 3.1.5
Test batteries6 monthsSection 3.1.6
Rebuild vacuum pump18 monthsRebuild kit
Revision B.000
Calibration ProceduresIntervalReference
Interface board calibration1 yearSection 3.2.1
Analog input calibration1 yearSection 3.2.2
Ambient air temperature calibration1 yearSection 3.2.3
Ambient pressure calibration1 yearSection 3.2.4
Mass flow controller calibration1 yearSection 3.2.5
3.1.MAINTENANCE PROCEDURES
This section contains maintenance procedures for the Partisol Sampler. Additional
maintenance procedures (such as cleaning the PM-10 inlet) are contained in the
Operating Manual.
3.1.1.EXCHANGINGTHE LARGE IN-LINE FILTER
The large in-line filter prevents particulate matter from entering the sampler’s flow
controller and causing damage.
Follow these steps to exchange the large in-line filter:
1) Turn off the sampler and remove the pump compartment cover
(Figure 3-1).
Figure 3-1. Partisol
Sampler with door open.
Revision B.000
2) Locate the large in-line filter (inside the black filter holder) to the
right of the valve manifold (Figure 3-2).
Figure 3-2. Location of the
large in-line filter.
SECTION 3: SERVICE PROCEDURESPAGE 3-2
Page 31
Service Manual, Partisol Model 2000 Air Sampler
Hose clip
Large in-line filter
3) Remove the in-line filter from its holder and examine the hose
clips which secure the hoses to each end of the filter (Figure 3-3).
Figure 3-3. Large in-line
filter removed from its black
filter holder.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
4) Using a pair of needle nose pliers, squeeze the tabs on one of the
hose clips and remove the corresponding hose from the end of
the filter. Repeat this procedure for the other hose clip and hose.
5) Place the hoses and clips onto the new filter using the same
method used in step 4. Replace the filter into its holder.
6) Replace the pump compartment cover.
7) Turn on the sampler and perform a system leak check (Section
3.1.5).
3.1.2.EXCHANGINGTHE FUSES
The Partisol Sampler contains two fuses: a 5A and a 4A fuse. Both fuses are installed
into in-line fuse holders located between the transformer and the interface board. The
5A fuse is used for the 8.3V @ 5.4A voltage from the transformer and the 4A fuse is
used for the 24.5V @ 3.6A voltage from the transformer. The fuse must be removed
from the fuse holder to be identified.
Follow these steps to exchange the 5A and/or 4A fuse:
1) Turn off the sampler and disconnect the unit’s power cord from its
AC source.
2) Remove the three screws and open the electronics compartment
door (Figure 3-1).
3) Locate the in-line fuse holder (Figure 3-4). Grasp the holder on
both sides and unscrew it until the two halves come apart (Figure
3-5).
4) Remove the fuse from the fuse holder and determine its rating.
The fuse rating should be stamped on one of the ends of the fuse.
Figure 3-4. Inside the
electronics compartment.
Revision B.000
Figure 3-5. Inside the
electronics compartment
with an open fuse holder
highlighted.
SECTION 3: SERVICE PROCEDURESPAGE 3-4
Page 33
Service Manual, Partisol Model 2000 Air Sampler
Fan
5) Place a new fuse with the proper rating into the holder and screw
the two halves of the fuse holder together to secure.
6) Close the electronics compartment door and secure it with the
three screws that you previously removed.
7) Connect the power cord to the AC source and resume normal
operation.
Revision B.000
Figure 3-6. Location of fan
inside the pump compartment.
3.1.3.C
LEANINGTHE AIR INTAKE FILTER
The Partisol Sampler contains an air intake filter attached to the fan in the pump
compartment. This filter cleans the air flow drawn into the instrument enclosure by the
fan.
Tools Needed: Phillips screwdriver
Mild soap solution (if required)
Materials:Air intake filter (R&P P/N 10-002799)
Follow these steps to clean the air intake filter:
1) Turn off the sampler and remove the pump compartment cover
(Figure 3-1).
2) Locate the air intake fan on the left side of the pump compartment
(Figure 3-6).
SECTION 3: SERVICE PROCEDURESPAGE 3-5
Page 34
Service Manual, Partisol Model 2000 Air Sampler
3) Snap off the fan cover on the outside of the enclosure. This cover
contains the air intake filter and a screen.
4) Take out the filter and screen and brush any dirt off. If they are
covered with a sticky coating, wash both the filter and with a mild
soap solution and water. If either part is damaged, they should be
replaced.
5) If the filter and/or screen was washed in the liquid solution, allow
them to dry before reinstalling them into the cover.
6) Reinstall the filter and screen into the cover and remount the
assembly.
7) Replace the pump compartment cover.
8) Turn on the sampler and resume normal operation.
Revision B.000
3.1.4.INSPECTING “V” SEALS
There are two “V” seals used in the hub unit of the Partisol sampler: one mounted
above and one mounted below the filter cassette. Also, there are two “V” seals used
in each satellite unit. These seals ensure that the sample flow passes through the
sample filter and seals the filter from leaks. These seals must be kept clean and
inspected frequently for damage.
1) Open the filter exchange mechanism as if to perform a filter exchange. Remove any sample filter cassette installed.
2) Locate the two “V” seals, one mounted above and one mounted
below the filter cassette. Examine each for damage. Wipe seals
with a clean dry cloth, if required.
Follow these steps to perform a leak test on the hub and satellite units:
1) Install a new 47mm filter into the sampling station (hub or satellite)
being tested.
2) From the Partisol Main screen, press <F5:Setup> to enter the
Setup screen. Press <F5:Audit> from the Setup screen to enter
the Audit screen (Figure 3-7).
3) Remove the size-selective inlet from the sampling station being
checked. Refer to the Operating Manual for the maintenance
procedure for the PM-10 inlet, if needed.
4) Install the supplied flow audit adapter on the end of the sample
tube of the sampling station being checked. Ensure that the flow
audit adapter’s valve is in the open position (Figure 3-8).
5) Press either <F2:Hub1>, <F3:Sat2>, <F4:Sat3>, or <F9:Sat4>
(<Shift><F4>) depending on which sampling station is currently
being checked.
SECTION 3: SERVICE PROCEDURESPAGE 3-7
Page 36
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-8. Flow audit
adapter installed.
Revision B.000
6) Turn on the pump by pressing <F5:Pump> or <F10:Pump>
(<Shift><F5>).
7) Close the valve on the flow audit adapter.
8) Open the mass flow controller access door on the pump compartment cover (Figure 3-1).
9) Close the flow of the flow controller assembly by turning the
manual shut off valve that is attached to the large air filter on the
left side of the mass flow controller assembly (labeled “A” in
Figure 3-9).
10) Record the reading on the vacuum gauge on the mass flow controller assembly.
11) Shut off the flow to the pump by turning the other manual shut off
valve (the “pump” valve), which is located on the bottom of the
mass flow controller assembly (labeled “B” in Figure 3-9).
12) Record the reading on the vacuum gauge 10 seconds after the
pump valve is closed. This reading should not drop below half of
the original reading (recorded in step 10) during this 10 second
period. If the reading does drop below half of the original reading, go to Section 2.
13) Slowly open the two valves on the mass flow controller that were
closed.
SECTION 3: SERVICE PROCEDURESPAGE 3-8
Page 37
Service Manual, Partisol Model 2000 Air Sampler
A
B
Vacuum Gauge
Figure 3-9. Mass flow
controller.
Revision B.000
14) Slowly open the valve of the flow audit adapter assembly and
remove it from the sampling station being checked. Replace the
size-selective inlet.
15) Remove the 47mm filter from the sampling station being checked
and discard it.
16) Perform steps 1-14 for each sampling station.
17) When all sampling stations have been checked, close the mass
flow controller access door.
3.1.6.TESTINGTHE BATTERIES
There are two batteries used in the Partisol Sampler: a round coin cell battery and a
battery-backed RAM socket battery. Both batteries are located on the CPU board.
3.1.6.1.TESTINGTHE ROUND COIN CELL BATTERY (BT1)
The round coin cell battery has an approximate 6-month storage capacity when the unit
is unpowered and stored at a temperature of 20° C.
SECTION 3: SERVICE PROCEDURESPAGE 3-9
Page 38
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure 3-10. Setup screen.
Setup
Stations:3Std Temp:25.0
Run Type:BASICStd Pres:1.000
Set Flow:16.7Ave Temp:99.9
A/O Type:STATAve Pres:9.999
Curr Date:03-Dec-99Basic Cyc:0
Curr Time:16:21:36Intv Mode:SAMP
EditCalibDiagRS232Audit>
Soft Function Keys in View Data Mode
EditCalibDiagRS232Audit>
LCD +LCD -<
Soft Function Keys in Edit Mode
---+++Bksp>
+/-<
Follow these steps to check the round battery:
1) From the Partisol Main screen, press <F5:Setup> to enter the
Setup screen (Figure 3-10).
Unshifted
Shifted
Unshifted
Shifted
2) Press <F1:Edit> to enter the Edit Mode.
3) Change the time or date to a new value and press <Enter>.
4) Turn off the instrument, wait approximately 20 seconds and turn
the power back on.
5) From the Partisol Main screen, press <F5:Setup> to enter the
Setup screen.
6) If the time or date is the same value as entered in step 3, the
battery is functioning properly. If the time or date shows a different value than the value entered in step 3, the battery must be
replaced. Contact R&P.
The battery-backed RAM socket battery has an approximate 2-1/2 year storage
capacity if the unit is unpowered and stored at temperatures between 0° and 40° C.
SECTION 3: SERVICE PROCEDURESPAGE 3-10
Page 39
Service Manual, Partisol Model 2000 Air Sampler
Follow these steps to check the socket battery:
1) From the Partisol Main screen, press <F3:Storage> to enter the
Filter Data screen.
2) Scan the records on the Filter Data and Interval Data screens for
random characters or incorrect data. If any discrepancies are
found, contact R&P.
3.2.CALIBRATION PROCEDURES
This section contains calibration procedures for the Partisol Sampler. Calibrations
should be performed according to the table on page 3-1.
IMPORTANT:
R&P strongly advises that these calibrations be performed in
the order presented.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
3.2.1.INTERFACE BOARD CALIBRATION
The interface board is located on the bottom panel of the electronics compartment (the
area behind the keypad in the hub unit). See Figure 3-11 for interface board test point
locations.
Tools Needed: Phillips screwdriver
Multimeter with four-digit resolution
Materials:None
Follow these steps to calibrate the interface board:
1) Turn off the sampler.
2) Remove the three screws and open the electronics compartment
door (Figure 3-1).
3) Locate the test point labeled “+6V” (red) near the back of the
interface board (Figure 3-11). Place the positive lead of a multimeter on the +6V test point.
4) Locate the ground test point labelled “PGND” (black) to the left of
the +6V test point (Figure 3-11). Place the ground lead on the
PGND test point.
SECTION 3: SERVICE PROCEDURESPAGE 3-11
Page 40
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-11.
Interface board
test point
locations.
Revision B.000
SECTION 3: SERVICE PROCEDURESPAGE 3-12
Page 41
Service Manual, Partisol Model 2000 Air Sampler
5) Ensure that the sampler is turned on and that the display backlight
is on. The backlight must be on during the +6V calibration. If the
backlight is off, press any key on the keypad to turn it on.
6) Locate the R21 pot on the rear section of the interface board. It is
located tp the right of the +6V test point. R21 is a blue vertical
pot, and its designation, “R21,” is silk-screened on the board.
7) Adjust the R21 pot with a potentiometer adjustment tool until the
multimeter reads 6.00 VDC ± 0.05 V.
8) Locate the test point labeled “+10V” (red) in the center of the
interface board. Place the positive lead of the multimeter on the
+10V test point.
9) Locate the ground test point labelled “AGND” (black) near the
front right corner of the interface board. Place the ground lead on
the AGND test point.
Revision B.000
10) Locate the R44 pot in the center of the interface board. R44 is a
blue vertical pot, and its designation, “R44,” is silk-screened on
the board.
11) Adjust the R44 pot with a potentiometer adjustment tool until the
multimeter reads 10.000 VDC ± 0.002 V.
12) Locate the test point labelled “+2.5V” (red) near the right side of
the interface board. Place the positive lead of the multimeter on
the +2.5V test point.
13) Locate the test point labelled “PWM0” (green) to the right of the
+2.5V test point. Place the ground lead on the PWM0 test point.
14) Locate the R53 pot below the PWM0 test point on the interface
board. R53 is a blue vertical pot, and its designation, “R53,” is
silk-screened on the board.
15) Adjust the R53 pot with a potentiometer adjustment tool until the
multimeter reads 0.000 VDC ± 0.002 V.
16) Go to Section 3.2.2.
SECTION 3: SERVICE PROCEDURESPAGE 3-13
Page 42
Service Manual, Partisol Model 2000 Air Sampler
3.2.2.ANALOG INPUT CALIBRATION
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
Please note that the following procedure must be performed after the interface board
calibration and before the temperature, pressure and flow calibrations.
1) If the Main screen is not currently displayed, press <Esc> until the
Main screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an
analog input calibration. If the hardware is currently in the Run
Operating Mode, press <F4: Run/Stp> to enter the Stop Operating
Mode. The current operating mode is displayed in the upper
right-hand corner of the Main screen.
Revision B.000
Figure 3-12. Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib>
to access the Calibration screen (Figure 3-13).
4) Plug the 6-pin end of the Analog Input Calibration Cable supplied
with the sampler into the socket labelled “ANEMOMETER” on the
back of the Partisol hub unit (Figure 3-14).
5) Plug the 4-pin end of the Analog Input Calibration Cable into the
socket labelled “USER OUTPUT” on the back of the Partisol hub
unit.
6) Ensure that the electronics compartment door (Figure 3-1) is open.
SECTION 3: SERVICE PROCEDURESPAGE 3-15
Page 44
Service Manual, Partisol Model 2000 Air Sampler
7) Locate the test point labelled “PWM1” (green) on the middle-right
side of the interface board (Figure 3-11). Attach the positive lead
from the multimeter on the PWM1 test point.
8) Locate the ground test point labeled “AGND” (black) near the front
right corner of the interface board. Place the ground lead on the
AGND test point.
9) Using the arrow key, position the cursor so that it is in the location
labelled “A/O” on the Calibration screen.
10) Press <F1:Edit> to enter the Edit Mode. Type in a number between
0.050 and 0.150 volts and press <ENTER>.
11) Observe the number displayed in the row labelled “A/I” in the
column labelled “Calc.” Ensure that this number does not vary
more than ±0.005 volts after watching it for 5 seconds. If this
number is not stable, choose a new number for “A/O” between
0.050 and 0.150 volts.
Revision B.000
12) Read the voltage displayed on the multimeter.
13) Using the arrow keys, position the cursor so that it is in the row
labelled “A/I” and the column labelled “Act” (actual).
14) Press <F1:Edit> to enter the Edit Mode. Type the voltage read
from the multimeter (to a 3-digit accuracy; i.e., “0.xxx”) in this
position and press <ENTER>. This calculates the “Offset.”
15) Ensure that the number now displayed in the row labelled “A/I”
and the column labelled “Calc” matches the number displayed on
the multimeter within ±0.005 volts.
16) Using the arrow key, position the cursor so that it is in the location
labelled “A/O.”
17) Press <F1: Edit> to enter the Edit Mode. Type in a number between 4.800 and 4.900 volts and press <ENTER>.
18) Observe the number displayed in the row labelled “A/I” in the
column labelled “Calc.” Ensure that this number does not vary
more than ±0.005 volts after watching it for 5 seconds. If this
number is not stable, choose a new number for “A/O” between
4.800 and 4.900 volts.
19) Read the voltage displayed on the multimeter.
20) Using the arrow keys, position the cursor so that it is in the row
labelled “A/I” and the column labelled “Act.”
SECTION 3: SERVICE PROCEDURESPAGE 3-16
Page 45
Service Manual, Partisol Model 2000 Air Sampler
21) Press <F1:Edit> to enter the Edit Mode. Type the voltage read
from the multimeter (to a 3-digit accuracy; i.e., “0.xxx”) in this
position and press <ENTER>. This calculates the “Span.”
22) Ensure that the number now displayed in the row labelled “A/I”
and the column labelled “Calc” matches the number displayed on
the multimeter within ±0.005 volts.
23) Remove the multimeter leads from the interface board and the
Analog calibration cable from the back of the Partisol hub.
24) After the analog input calibration has been performed successfully, the temperature, pressure and flow calibrations must be
performed.
NOTE: If the instrument has been reset and you have recorded the value of
“Offset” and “Span” for the flow, you may enter these numbers directly when in
the Edit Mode.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
25) Close the electronics compartment door and secure with the three
screws previously removed.
26) Go to Section 3.2.3.
3.2.3.AMBIENT AIR TEMPERATURE CALIBRATION
Tools Needed: Temperature measurement device
Materials:None
Follow these steps to calibrate the ambient air temperature:
1) If the Main screen is not currently displayed, press <Esc> until the
screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an
analog input calibration. If the hardware is currently in the Run
Operating Mode, press <F4: Run/Stp> to enter the Stop Operating
Mode. The current operating mode is displayed in the upper
right-hand corner of the Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib>
to access the Calibration screen (Figure 3-13).
4) Determine the current temperature (°C) at the ambient temperature
sensor mounted on the sample tube of the hub unit using an
external thermometer, [°C = 5/9 x (°F - 32)].
SECTION 3: SERVICE PROCEDURESPAGE 3-17
Page 46
Service Manual, Partisol Model 2000 Air Sampler
5) Press <F1: Edit> to enter the Edit Mode, and move the cursor to
the “Act” (actual) column in the row labelled “Temp.”
6) Enter the current ambient temperature (°C), using the <F6: +/-> key
to enter negative temperatures if necessary. Press <ENTER> to
leave the Edit Mode. Check that the temperature shown in the
“Calc” column now matches the current temperature within ±2° C.
7) Upon receiving the actual temperature, the system’s microprocessor automatically computes “Span” for the ambient temperature.
Note this number for future reference.
NOTE: If the instrument has been reset and you have recorded the value of
“Span” for the ambient temperature, you may enter it directly in the “Span”
column when in the Edit Mode.
8) Go to Section 3.2.4.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
3.2.4.AMBIENT PRESSURE CALIBRATION
Tools Needed: None
Materials:None
Follow these steps to calibrate the ambient pressure:
1) If the Main screen is not currently displayed, press <Esc> until the
screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an
ambient pressure calibration. If the hardware is currently in the
Run Operating Mode, press <F4: Run/Stp> to enter the Stop
Operating Mode. The current operating mode is displayed in the
upper right-hand corner of the Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib>
to access the Calibration screen (Figure 3-13).
4) Determine the current ambient station pressure in Atmospheres
(absolute pressure, not corrected to sea level).
• To convert from mm Hg @ 0 °C to Atmospheres, multiply by 0.001316.
• To convert from millibars to Atmospheres, multiply by 0.000987.
• To convert from inches Hg @ 32 °F to Atmospheres, multiply by 0.03342.
SECTION 3: SERVICE PROCEDURESPAGE 3-18
Page 47
Service Manual, Partisol Model 2000 Air Sampler
5) Press <F1: Edit> to enter the Edit Mode, and move the cursor to
the “Act” (actual) column in the row labelled “Pres.”
6) Enter the current ambient pressure (Atmospheres) and press
<ENTER> to leave the Edit Mode. Check that the pressure shown
in the “Calc” column now matches the current temperature within
± 0.02 atm.
7) Upon receiving the actual pressure, the system’s microprocessor
automatically computes “Span” for the ambient pressure. Note
this number for future reference.
NOTE: If the instrument has been reset and you have recorded the value of
“Span” for the ambient pressure, you may enter it directly in the “Span” column
when in the Edit Mode.
8) Go to Section 3.2.5.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
3.2.5.MASS FLOW CONTROLLER CALIBRATION
The temperature and pressure calibrations described above must be performed before
the mass flow controller calibration is undertaken. In addition, the leak check
discussed in Section 3.1.5 must also be undertaken before the following instructions
are executed.
Follow these steps to calibrate the mass flow controller:
1) If the Main screen is not currently displayed, press <Esc> until the
screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an
analog input calibration. If the hardware is currently in the Run
Operating Mode, press <F4: Run/Stp> to enter the Stop Operating
Mode. The current operating mode is displayed in the upper
right-hand corner of the Main screen.
3) Remove the size-selective inlet from the hub.
4) Install a filter cassette containing a new 47 mm filter into the filter
holder of the hub unit. This filter will be thrown away at the end
of this flow calibration.
SECTION 3: SERVICE PROCEDURESPAGE 3-19
Page 48
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure 3-15. FTS Flow
Calibration screen.
BasicFTS Flow CalibrationStop
Cur Flow:0.1l/min
Act Flow:0.1l/minValve
FTS Pres:0.00in/H2O
5) While in the Main screen, press <F5: Setup> and then <F2: Calib>
to enter the Calibration screen. From the Calibration screen,
press <F6: FTS> (<SHIFT> <F1>) to enter the FTS Flow Calibration
screen (Figure 3-15).
Unshifted
Shifted
Unshifted
Shifted
6) If you are using the Streamline FTS Flow Transfer Standard to
calibrate the flow of the Partisol sampler, enter the FTS calibration constants “m” and “b” on the FTS Flow Calibration screen.
Execute the following steps to enter these constants:
• Press <F1:Edit>
• Scroll down to the “FTS Const m” field and enter the m constant from the FTS
calibration sheet
• Scroll down to the “FTS Const b” field and enter the b constant from the FTS
calibration sheet
• Press <Enter>
7) Attach the Streamline FTS Flow Transfer Standard to the hub’s
sample tube (Figure 3-16). Other flow meters may require the use
of the flow audit adapter to attach the meter to the sample tube.
The flow audit adapter should be installed with its valve open
(Figure 3-8).
8) Do not turn the pump on. If it is currently on, press <F7: Pump>
(<SHIFT> <F2>) to turn it off.
SECTION 3: SERVICE PROCEDURESPAGE 3-20
Page 49
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-16. Streamline
FTS Flow Transfer Standard
attached to the Partisol.
Revision B.000
9) Press <F1: Edit> to enter the Edit Mode, and move the cursor to
the row labelled “Act Flow.”
10) Enter a zero (“0”) in this field and press <ENTER> to leave the Edit
Mode. This causes the microprocessor to compute the “Offset,”
which is the zero offset for the mass flow sensor. Record this
number for future reference.
11) Turn on the pump by pressing <F7: Pump> (<SHIFT> <F2>), and
then press <F2: Hub1> to allow the sample flow to pass through
the sample tube of the hub. The valve column should display
“H1.”
12) Use the <F8: DecrFlw> (<SHIFT> <F3>) (continuously decrease
flow), <F9: HoldFlw> (<SHIFT> <F4>) (maintain current flow rate),
and <F10: IncrFlw> (<SHIFT> <F5>) (continuously increase flow)
keys to control the flow rate. The flow rate displayed in the row
labeled “Cur Flow” should be approximately 16.7 l/min.
NOTE: Pressing <F8:DecrFlw> (or <F10:IncrFlw>) will cause the flow to
continuously decrease (or increase) until <F9:HoldFlw> is pressed. If the flow
is not stopped, it will decrease (or increase) to a maximum point and will begin
to increase (or decrease) because of the circular gear used on the flow controller.
13) Determine the flow in actual l/min by using the external flow meter.
If you are using the Streamline FTS Flow Transfer Standard, enter
the pressure (inches H2O) in the FTS Pres field. Press <F1:Edit>,
SECTION 3: SERVICE PROCEDURESPAGE 3-21
Page 50
Service Manual, Partisol Model 2000 Air Sampler
enter the value in the FTS Pres field using the keypad, and then
press <Enter>. The sampler will calculate the volumetric flow in l/
min.
14) If you are using the Streamline FTS Flow Transfer Standard, the
sampler will automatically enter the calculated flow into the “ACT
FLOW” field and compute “Span,” which is the span slope for the
mass flow sensor. Record this number for future reference.
15) If you are using a different flow measurement device, you must
manually enter the volumetric flow in the “ACT FLOW” field by
pressing <F1:Edit> to enter the Edit Mode, moving the cursor to
the “ACT FLOW” field and entering the flow determined by the
external flow meter, followed by <Enter>. This causes the microprocessor to compute “Span,” which is the span slope for the
mass flow sensor. Record this number for future reference.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
NOTE: If the instrument has been reset and you have recorded the value of
“Offset” and “Span” for the flow, you may enter these numbers directly when in
the Edit Mode.
15) Return to the Main screen by pressing <ESC> three times.
16) Restore the sampling hardware to its original state by removing
the flow metering hardware and re-installing the size-selective
inlet on the sample tube of the hub. Remove the filter cassette
from the hub unit, and discard the filter installed in it.
3.3.REMOVEAND REPLACE PROCEDURES
3.3.1.REMOVING/REPLACINGTHE DISPLAY
Tools Needed:Phillips screwdriver
Slotted screwdriver
Materials:Display (R&P P/N 10-002254)
Static-shielded bag
Follow these steps to remove the display from the sampler:
1) Turn off the instrument and disconnect the power cord from its AC
source.
2) Remove the three screws and open the electronics compartment
door (Figure 3-1).
SECTION 3: SERVICE PROCEDURESPAGE 3-22
Page 51
Service Manual, Partisol Model 2000 Air Sampler
Display cover
Contrast
knob
RS232 connector
Circuit breaker
Ground wire
Figure 3-17. Back of
electronics compartment
door.
Revision B.000
3) Examine the back of the electronics compartment door. Remove
the screw securing the ground wire to the corner of the display
cover (Figure 3-17).
4) Remove the remaining screws that secure the display cover.
5) Note the orientation of the display as it is mounted on the back of
the electronics compartment door (the display connector is on
the hinge side of the door). You will need this information to
install a new display.
6) Disconnect the cable connecting the contrast knob to the display.
7) Disconnect the cable connecting P16A of the interface board and
the display at the display connector.
8) Disconnect the cable connecting P16B of the interface board and
the display at the interface board connector.
9) Remove the four round standoffs from the corners of the display.
10) Place display in a static-shielded bag.
SECTION 3: SERVICE PROCEDURESPAGE 3-23
Page 52
Service Manual, Partisol Model 2000 Air Sampler
Follow these steps to install the display into the sampler:
1) Mount the display into position on the electronics compartment
door. The ribbon cable connector should be located on the hinge
side of the door.
2) Replace the four round standoffs onto the corners of the display.
3) Connect the cable connecting P16B of the interface board and the
display.
4) Connect the cable connecting P16A of the interface board and the
display.
5) Connect the cable connecting the contrast knob to the display
(Figure 3-17).
6) Replace the display cover onto the standoffs and secure with
three of the screws, leaving the lower left corner loose.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
7) Place the ground wire on the loose corner of the display cover and
secure with the remaining screw.
8)Close and secure the electronics compartment door.
9)Connect the unit’s power cord to an AC source.
3.3.2.R
EMOVING/REPLACINGTHE FAN
Tools Needed:Slotted screwdriver
Phillips screwdriver
Materials:Fan (R&P P/N 10-002251)
Follow these steps to remove the fan from the sampler:
1) Turn off the instrument and disconnect the power cord from the
AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Remove the fan cover, screen and filter from the outside of the fan
(on the outside of the enclosure).
4) Disconnect the fan connector (P9) from the underside of the
interface board.
3) Remove the four screws which secure the fan to the enclosure
and remove the fan assembly.
SECTION 3: SERVICE PROCEDURESPAGE 3-24
Page 53
Service Manual, Partisol Model 2000 Air Sampler
Follow these steps to install the fan into the sampler:
1) Hold the replacement fan and outside guard mounting frame in the
proper position on the enclosure. The fan should be positioned
so that the air is blown INTO the enclosure. There is an arrow on
the side of the fan that indicates the air flow direction.
2) Secure the fan using the four screws previously removed.
3) Replace the screen, filter and fan cover by pressing it onto the fan
mounting frame from the outside of the enclosure.
4) Connect the fan connector (P9) to the underside of the interface
board.
5) Replace the pump compartment cover.
6) Connect the power cord to the AC source and resume normal
operation.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
3.3.3.REMOVING/REPLACINGTHE MASS FLOW CONTROLLER ASSY
Tools Needed:Adjustable wrench
Phillips screwdriver
Pliers
Materials:Mass flow controller assembly
(R&P P/N 55-002663)
Follow these steps to remove the mass flow controller assembly from the
sampler:
1) Turn the instrument off and disconnect the unit’s power cord from
its AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Locate the hose that attaches the mass flow sensor to the large inline filter. Using a pair of pliers, squeeze the clip that secures the
hose to the mass flow sensor and remove the hose (Figure 3-3).
4) Locate the hose attached to the manual valve on the bottom of the
assembly. Examine the hose clip that secures the hose to the
valve.
5) Using a pair of pliers, squeeze the clip and remove the hose from
valve.
SECTION 3: SERVICE PROCEDURESPAGE 3-25
Page 54
Service Manual, Partisol Model 2000 Air Sampler
6) Locate the wires that are attached to the mass flow controller
motor. Follow the wires until you find a Molex connector. Unplug
this connector.
7) Disconnect the mass flow sensor connector (P21A) from the
interface board.
8) Remove the four nuts that secure the mass flow controller assembly to the enclosure. Remove the assembly.
Follow these steps to install the mass flow controller assembly into the
sampler:
1) Place the mass flow controller assembly in its proper position in
the enclosure. Ensure that the assembly is oriented so that the
vacuum gauge is visible.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
2) Replace the four nuts that secure the mass flow controller assembly to the enclosure.
3) Reconnect the mass flow sensor connector (P21A) on the interface board.
4) Reconnect the mass flow controller motor connector.
5) Using a pair of pliers, squeeze the clip that secures the hose to the
mass flow sensor and replace the hose onto the mass flow sensor fitting.
6) Using a pair of pliers, squeeze the clip that secures the hose to the
manual valve and replace the hose onto the valve.
7) Perform a system leak test (Section 3.1.5).
8) Perform a mass flow controller calibration (Section 3.2.5).
Follow these steps to remove the electronics compartment heater/fan
assembly from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from
its AC source.
SECTION 3: SERVICE PROCEDURESPAGE 3-26
Page 55
Service Manual, Partisol Model 2000 Air Sampler
2) Remove the three screws and open the electronics compartment
door (Figure 3-1).
3) Locate the electronics compartment heater/fan assembly and note
its orientation. You will need this information later when installing a new heater/fan assembly.
4) Disconnect the cable connecting the electronics heater to the
interface board (P11).
5) Disconnect the cable connecting the electronics fan to the interface board (P101).
6) While supporting the assembly, remove the two screws that secure the heater/fan assembly to the back of the electronics
compartment.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
Follow these steps to install the electronics compartment heater/fan
assembly into the sampler:
1) While supporting the assembly in the proper position on the back
of the electronics compartment, secure the two screws. Ensure
that the orientation is correct (the fan should be on top).
2) Reconnect the cable connecting the electronics heater to the
interface board (P11).
3) Reconnect the cable connecting the electronics fan to the interface board (P101).
4) Close the electronics compartment door and secure it with its
three screws.
5) Connect the unit’s power cord to its AC source and resume normal operation.
Inside the electronics compartment of the Partisol Sampler, the interface board is
mounted on top of the CPU board. These boards are attached so that they can be
installed and removed as a unit. The user should not attempt to separate these boards.
The boards should be returned to R&P as a set if service or replacement is required.
SECTION 3: SERVICE PROCEDURESPAGE 3-27
Page 56
Service Manual, Partisol Model 2000 Air Sampler
Tools Needed:Phillips screwdriver
Materials:CPU board (R&P P/N 50-002407)
Follow these steps to remove the interface/CPU board assembly from
the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from
its AC source.
2) Remove the three screws and open the electronics compartment
door (Figure 3-1).
Revision B.000
Slotted screwdriver
Labeling materials
Interface board (R&P P/N 50-005496)
Static-shielded bag
3) Remove all wires and cables connected to the interface and CPU
boards.
NOTE: While removing cables, be sure to label each connector with its location
for easy re-assembly.
4) Remove the 1/8-inch silicone rubber hose from the pressure
transducer on the right side of the interface board.
5) Locate the screws that secure the assembly to the floor of the
electronics compartment. Remove the screws.
6) Lift up the assembly approximately 2 inches from the floor of the
electronics compartment and remove it from the compartment.
7) Place the boards assembly in a static-shielded bag.
Follow these steps to install the interface/CPU board assembly into the
sampler:
1) Position the board assembly over the standoffs on the floor of the
electronics compartment.
2) Replace the screws and secure the assembly to the floor of the
electronics compartment.
3) Replace the 1/8-inch silicone rubber hose from the pressure
transducer on the right side of the interface board.
4) Replace all wires and cables connected to the CPU and interface
boards. Be sure to follow the location labels that you made earlier
when you removed the cables.
SECTION 3: SERVICE PROCEDURESPAGE 3-28
Page 57
Service Manual, Partisol Model 2000 Air Sampler
5) Close the electronics compartment door and secure it with the
tree screws.
6) Connect the unit’s power cord to an AC source.
7) Perform the calibration procedures listed in Section 3.2.
3.3.6.REMOVING/REPLACINGTHE KEYPAD
The keypad of the Partisol Sampler is integrated into the door to the electronics
compartment. Therefore, to replace the keypad, the entire electronics compartment
door must be replaced.
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
Tools Needed:Adjustable wrench
Materials:Keypad (R&P P/N 55-002345)
Revision B.000
Phillips screwdriver
Static-shielded bag
Follow these steps to remove the keypad from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from
its AC source.
2) Remove the three screws and open the electronics compartment
door (Figure 3-1).
3) Remove the display from the electronics compartment door (Section 3.3.1).
4) Remove the two screws that secure the RS232 connector to the
electronics compartment door (Figure 3-17). Remove the connector.
5) Remove the four wires attached to the inside of the circuit breaker/
power switch by pulling them straight off.
6) Depress the tabs on both sides of the circuit breaker/power switch
and push through the hole in the electronics compartment door.
7) Remove the contrast adjustment knob from the electronics compartment door.
8) Disconnect the keypad cable from the mating cable on the right
side wall of the electronics compartment.
9) Remove the three nuts securing the electronics compartment door
to the enclosure.
SECTION 3: SERVICE PROCEDURESPAGE 3-29
Page 58
Service Manual, Partisol Model 2000 Air Sampler
10) Place the defective keypad into a static-shielded bag for return to
R&P.
Follow these steps to install the keypad into the sampler:
1) Place the electronics compartment door in its proper position on
the enclosure. Tighten the three nuts that secure the door to the
enclosure.
2) Connect the keypad cable to the mating cable on the right side
wall of the electronics compartment.
3) Replace the contrast adjustment knob onto the electronics compartment door.
4) Push the circuit breaker/power switch through the hole in the
electronics compartment door.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
5) Replace the four wires onto the circuit breaker/power switch.
6) Replace the two screws that secure the RS232 connector to the
electronics compartment door.
7) Replace the display onto the electronics compartment door (Section 3.3.1).
8) Close the electronics compartment door and secure.
9) Connect the unit’s power cord to an AC source and resume normal operation.
3.3.7.REMOVING/REPLACINGTHE PUMP
Tools Needed:Adjustable wrench
Phillips screwdriver
Materials:Pump
(R&P P/N 54-005949-0120 or 54-005949-0240)
Follow these steps to remove the pump from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from
its AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Remove the black, pneumatic tubing from the fittings of the pump.
Depress the collar on the fitting surrounding each tube with a flat
SECTION 3: SERVICE PROCEDURESPAGE 3-30
Page 59
Service Manual, Partisol Model 2000 Air Sampler
blade screwdriver and then pull the tubing straight out of its
fitting (Figure 3-18).
4) Remove the pump’s ground wire from the left side of the enclosure.
5) Disconnect the cable connecting the pump to the interface board
(P10). The cable is connected on the underside of the interface
board.
6) Remove the four nuts securing the pump to the floor of the pump
compartment. You will need to lift or tip the unit sideways to
access these nuts.
7) Lift the pump and remove it from the sampler.
Follow these steps to install the pump into the sampler:
Revision B.000
Figure 3-18. Pump compartment.
1) Place the pump into the enclosure in its proper position and
secure the pump with the four nuts. You will need to lift or tip the
unit sideways to access these nuts.
2) Reconnect the cable connecting the pump to the interface board
(P10). The cable is connected on the underside of the interface
board.
3) Replace the pump’s ground wire to the left side of the enclosure.
4) Reinsert the black tubing into the pump fittings. Press the tubing
firmly into each fitting to insure proper seating.
SECTION 3: SERVICE PROCEDURESPAGE 3-31
Page 60
Service Manual, Partisol Model 2000 Air Sampler
5) Replace the pump compartment cover.
6) Connect the unit’s power cord to an AC source and resume normal operation.
Revision B.000
4
Always wear appropriate
antistatic devices when
working with the system’s
electronics.
3.3.8.R
EMOVING/REPLACINGA VALVE
Tools Needed:Flat screwdriver
Phillips screwdriver
Materials:Valve (R&P P/N 59-002456)
Follow these steps to remove the valve from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from
its AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Locate the valve manifold below the electronics compartment and
determine which valve is being replaced (Figure 3-19).
4) Disconnect the cable connecting the valve to the interface board
(P5, P6, P7 or P8). The cable is connected on the underside of the
interface board.
5) Remove the upper left and lower right screws from the valve.
These are the screws that secure the valve to the manifold.
6) Remove the valve from the enclosure.
Follow these steps to install the valve into the sampler:
1) Place the valve onto the manifold in its proper orientation (the
valve plug should be toward the pump). Tighten the upper left
and lower right screws.
2) Reconnect the cable connecting the valve to the interface board
(P5, P6, P7 or P8). The cable is connected on the underside of the
interface board.
3) Replace the pump compartment cover.
3) Connect the unit’s power cord to its AC source and resume normal operation.
SECTION 3: SERVICE PROCEDURESPAGE 3-32
Page 61
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-19. Valve manifold.
Revision B.000
SECTION 3: SERVICE PROCEDURESPAGE 3-33
Page 62
This page left intentionally blank.
Page 63
Service Manual, Partisol Model 2000 Air Sampler
Section 4: Electrical Schematics
This section contains schematic diagrams and interconnect diagrams of the major
electronic components in the Partisol Model 2000 Sampler. Schematic diagrams for
the industrial microcomputer system are not included here, because they are proprietary to the manufacturers of these boards.
Refer to this section when shipping items to R&P. Always contact the R&P Service
Department before returning any unit components.
Service Department, R&P Co.
Phone: 518-452-0065
Fax: 518-452-0067
Email: SUPPORT@RPCO.COM
5.1.R&P RETURN POLICY — INSTRUMENT COMPONENTS
There are three types of returns:
·Warranty return for exchange
·Billable return for exchange
·Billable return for repair.
Revision B.000
Warranty Return for Exchange
If an instrument component becomes defective during the warranty period, follow this
procedure:
1. Contact the R&P Service Department. Have the following information
available:
·System serial number
·Nature of the problem
·Troubleshooting steps taken and results.
2. R&P service personnel will determine whether a component exchange is
necessary and whether the component is covered under warranty.
3. If R&P service personnel determine that a component exchange is
necessary and that it is covered under warranty, you will be issued a return
number.
4. A replacement component will be shipped as soon as possible.
5. Once you have received the replacement component and installed it into
your unit, you should ship the defective component back to R&P in the
same box that the replacement component arrived in, according to the
instructions in Section 5.3 or 5.4. Be sure to write the return number on
the box or in the return papers.
Billable Return for Exchange
If an instrument component becomes defective after the warranty period, follow this
procedure:
1. Contact the R&P Service Department. Have the following information
available:
·System serial number
·Nature of the problem
·Troubleshooting steps taken and results.
SECTION 5: R&P RETURN POLICYPAGE 5-1
Page 82
Service Manual, Partisol Model 2000 Air Sampler
2. R&P service personnel will determine whether a component exchange is
necessary and whether the component is covered under warranty.
3. If the component exchange is billable, R&P service personnel will inform
you of the exchange price.
4. You will be issued a return number.
5. As soon as R&P receives your purchase order and appropriate invoice,
the company will ship a replacement component to you.
6. Once you have received the replacement component and installed it into
your unit, you should ship the defective component back to R&P in the
same box that the replacement component arrived in, according to the
instructions in Section 5.3 or 5.4. Be sure to write the return number on
the box or in the return papers.
Billable Return for Repair
If an instrument component becomes defective after the warranty period, the component can be returned to R&P for repair. This procedure is generally less expensive than
a billable return for exchange but can result in a longer downtime for the instrument.
If you would like to return your instrument component for repair, follow this
procedure:
1. Contact the R&P Service Department. Have the following information
available:
·System serial number
·Nature of the problem
·Troubleshooting steps taken and results.
2. R&P service personnel will determine whether the component is defective and whether the component is covered under warranty.
3. If R&P service personnel determine that the component is billable and
you wish to return it for repair, they will inform you of the exchange price.
The billable repair will not exceed this price.
4. You will be issued a return number.
5. Ship the defective component back to R&P, according to the instructions
in Section 5.3 or 5.4. Be sure to write the return number on the box or in
the return papers. Please include the name and phone number of the
person who should be contacted with the final repair price.
6. R&P service personnel will call this person to tell them the final repair
price, once the repair is complete.
7. As soon as R&P receives your purchase order for the repair, the company
will ship your repaired component to you.
Revision B.000
SECTION 5: R&P RETURN POLICYPAGE 5-2
Page 83
Service Manual, Partisol Model 2000 Air Sampler
5.2.R&P RETURN POLICY — SYSTEMS
There are two types of returns:
·Warranty return for evaluation/repair
·Billable return for evaluation/repair.
Warranty Return for Evaluation/Repair
If an instrument becomes defective during the warranty period, follow this procedure:
1. Contact the R&P Service Department. Have the following information
available:
·System serial number
·Nature of the problem
·Troubleshooting steps taken and results.
2. R&P service personnel will determine whether it is necessary to ship the
system to R&P and whether the system is covered under warranty.
3. If R&P service personnel determine that it is necessary to ship the system
to R&P and that the system is covered under warranty, you will be issued
a return number.
4. R&P service personnel will specify which items should be returned.
These items should be returned to R&P according to the instructions in
Section 5.3 or 5.4.
NOTE: Send the system back to R&P in the original shipping boxes. If
you discarded the original shipping boxes and the system becomes
damaged during a return shipment to R&P, the warranty could be voided.
5. Be sure to write the return number on the box or in the return papers.
Revision B.000
Billable Return for Evaluation/Repair
If an instrument becomes defective after the warranty period, follow this procedure:
1. Contact the R&P Service Department. Have the following information
available:
·System serial number
·Nature of the problem
·Troubleshooting steps taken and results.
2. R&P service personnel will determine whether it is necessary to ship the
system to R&P and whether the system is covered under warranty.
3. If R&P service personnel determine that it is necessary to ship the system
to R&P, you will be issued a return number.
4. R&P service personnel will specify which items should be returned.
These items should be returned to R&P according to the instructions in
Section 5.3 or 5.4.
NOTE: Send the system back to R&P in the original shipping boxes. If
you discarded the original shipping boxes and the system becomes
damaged during a return shipment to R&P, additional costs could be
levied.
SECTION 5: R&P RETURN POLICYPAGE 5-3
Page 84
Service Manual, Partisol Model 2000 Air Sampler
5. Be sure to write the return number on the box or in the return papers.
Please include the name and phone number of the person who should be
contacted with the final repair price.
6. When R&P service personnel receive your instrument, they will repair
and calibrate the unit. They will determine a final repair price and contact
you with this information.
7. As soon as R&P receives your purchase order for the repair, R&P service
personnel will return your system.
5.3.CONTINENTAL US SHIPPING INSTRUCTIONS
The following instructions explain how to return merchandise to Rupprecht &
Patashnick Co., Inc. when shipping within the Continental US. It is important that
R&P receives and repairs defective exchanged items in a timely manner.
Revision B.000
NOTE: You are responsible for the shipping costs of
returned goods.
Please read this before returning any material:
1. Do not ship your order collect.
2. Clearly mark your return merchandise paperwork and packaging with the
return number to receive credit for your return.
5.4.INTERNATIONAL SHIPPING INSTRUCTIONS
The following instructions explain how to return merchandise to Rupprecht &
Patashnick Co., Inc. when shipping from places outside the Continental US (internationally). It is important that R&P receives and repairs defective exchanged items in
a timely manner. Please read this before returning any material:
NOTE: To avoid customs delays, it is very important that you clearly
mark your shipping paperwork with the following statement:
COUNTRY OF ORIGIN: USA
You are responsible for the shipping costs of returned goods. Do not
ship your order collect.
SECTION 5: R&P RETURN POLICYPAGE 5-4
Page 85
Service Manual, Partisol Model 2000 Air Sampler
1. Clearly mark your return merchandise paperwork and packaging with the
return number to receive credit for your return. Mark your paperwork:
2. R&P will instruct you on the method of return, as follows:
— For small items with no rush; ship regular air mail - parcel post.
— For intermediate or rush shipments using UPS, FEDEX or DHL,
have the shipper handle customs. List “Rupprecht & Patashnick Co.,
Inc.” as the “cosignee” and yourself as the “shipper.”
— For large items (rush or otherwise) use an air cargo service terminat-
ing at JFK International Airport. Shipping to a different airport will
result in significant lost time and added expense. Prepare your “Bill
of Lading” as follows:
List “Rupprecht & Patashnick Co., Inc.” as the “cosignee” and
yourself as the “shipper.” Address the boxes as follows:
Revision B.000
“COUNTRY OF ORIGIN: USA.”
Rupprecht & Patashnick Co., Inc.
c/o Air Express International Phone: (718) 244-4861
J. F. Kennedy International Airport
Building 89
Jamaica, NY 11430
Attn.: Julia Harris
— Avoid shipping directly to the Albany Airport. If you must ship
directly to the Albany Airport, prepare your “Bill of Lading” as
follows:
List “Rupprecht & Patashnick Co., Inc.” as the “cosignee” and
yourself as the “shipper.” List the following customs broker as
“notify”:
Notify: James E. Curran & Co., Inc.
29 Dartmouth St.
Albany, NY 12209
SECTION 5: R&P RETURN POLICYPAGE 5-5
Page 86
Service Manual, Partisol Model 2000 Air Sampler
3. For non-parcel post shipments over $2,500, a customs declaration letter
must accompany the shipment. It should be worded as follows:
“I, __________, declare that to the best of my knowledge and belief, the
articles herein specified are products of the USA; that they were exported
from the USA, from Albany, New York; and that they are returned
without having been advanced in value or improved in condition by any
process of manufacture or other means.”
Indicate on the declaration letter the items being returned and the purpose
for return (“For Credit,” “For Replacement” or “For Repair”).
Revision B.000
SECTION 5: R&P RETURN POLICYPAGE 5-6
Page 87
Service Manual, Partisol Model 2000 Air Sampler
Appendix A: Hardware Connections
The Partisol Model 2000 Air Sampler contains three connectors for communicating
analog and serial information with external devices:
User OutputThis 4-pin connector, located on the back panel of the
Revision B.000
sampler, transmits an analog voltage level in accordance
with the currently active sampling station and existence of
any status conditions. The pin layout, as viewed from the
back of the enclosure, is shown in Figure A-1 and is
defined below:
AnemometerThis 6-pin connector, located on the back panel of the
sampler, connects the sampler to a wind vane/anemometer or to an external device, such as a data
logger that provides a 0 to 5 VDC control signal to pin 1. The
pin layout, as viewed from the back of the enclosure, is
shown in Figure A-1 and is defined below:
RS-232This 9-pin connector, located on the keypad panel of the
sampler, allows two-way serial communication between
the hub and an external device. The 9 pins in the RS-232
connector are defined in the following manner:
Each hub unit is supplied with one cable connector for the 4-pin user output connector
(R&P part number 06-002453) and one cable connector for the 6-pin anemometer
connector (R&P part number 06-002454). The internal construction of these two
connectors is displayed in Figure A-2.
To connect a personal computer (PC) with your sampler, you will need a straight 9to-9 pin cable, which may be purchased from R&P as optional equipment (R&P part
number 07-000587).
APPENDIX A: HARDWARE CONNECTIONSPAGE A-2
Page 89
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure A-2. Internal
construction of 4- and 6-pin
cable connectors.