As R&P instrumentation changes, so do our Operating and Service manuals. However, these changes may affect only one aspect of an instrument, while leaving the
instrument as a whole unchanged. To explain these individual changes to customers,
R&P will update only those sections of its Operating and Service manuals that are
affected by the instrument updates or improvements. As each manual section changes,
so does its revision number, which is located at the top right corner of each page of each
section.
To help customers keep track of the changes to the Partisol Model 2000 Air Sampler
and its service manual, following is a list of the manual sections with their respective
revision numbers:
Section Number and DescriptionRevision Number
Section 1: IntroductionB.000
Revision B.000
Section 2: Troubleshooting GuidesB.000
Section 3: Service ProceduresB.000
Section 4: Electrical SchematicsB.000
Section 5: R&P Return PolicyB.000
Appendix A: Hardware ConnectionsB.000
IndexB.000
PAGEI
Service Manual, Partisol Model 2000 Air Sampler
Revision Description List
To help customers keep track of the changes to the Partisol Model 2000 Air Sampler
and its service manual, following is a list of the manual sections with their respective
revision descriptions. This list contains all changes made to the manual since the last
update:
Section Number and DescriptionRevision Description
NDEX ............................................................................................................. I
PAGEIV
Service Manual, Partisol Model 2000 Air Sampler
Section 1: Introduction
The information contained in this manual will allow the user to do preliminary
troubleshooting to fix some problems or to give the maximum amount of information
available to the factory.
1.1.OVERVIEWOF MANUAL
Section 1 : Introduction
This section provides an overview of the Service Manual.
Section 2 : Troubleshooting Guides
This section contains troubleshooting guides in the form of flow charts to aid the user
when performing diagnostics.
Section 3 : Service Procedures
This section contains the service procedures (maintenance and calibration) required
to maintain consistent operation of the Partisol sampler. In addition, this section
contains procedures describing installation of system components if exchange is
necessary.
Revision B.000
Section 4 : Electronic Schematic Diagrams
This section contains schematic diagrams and interconnect diagrams of the major
electronic components in the instrument.
Section 5 : R&P Return Policy
This section contains information regarding the return of whole systems or individual
components for warranty or billable repair.
Appendix A: Hardware Connections
This section contains the pin assignments of the output connectors on the Partisol
Sampler.
SECTION 1: INTRODUCTIONPAGE 1-1
This page left intentionally blank.
Service Manual, Partisol Model 2000 Air Sampler
Section 2: Troubleshooting Guides
This section contains troubleshooting guides in the form of flow charts to aid the user
when performing diagnostics. This section should be used in conjunction with the
electrical schematics located in Section 4. Read Section 2.1 before using any of the
troubleshooting guides.
2.1.USINGTHE TROUBLESHOOTING GUIDES
The troubleshooting guides give the user a logical succession of steps to follow when
diagnosing a problem. Following is a list of the questions most commonly asked by
users with regard to the guides.
Revision B.000
2.1.1.W
HATEQUIPMENTDO I NEED?
The main piece of equipment needed is a multimeter (that measures DC and AC
voltages and resistance) with at least a 4 digit resolution.
2.1.2.WHEREDO I START?
Look at the main screen of the Partisol Model 2000 Air Sampler. Any active errors
will be indicated in the upper left corner. Possible error codes are:
OKGreat! You have no errors!
EPower Failure — A system power failure occurred during sampling.
FFlow Out of Range — The measured sample flow rate through the
sample filter deviated by more than 7% from its setpoint for more than
1 minute.
TAmbient Temperature Sensor — The ambient temperature sensor
was not installed correctly or indicated an invalid value. The ambient
temperature must be between -40°C and 60°C to be valid.
PAmbient Pressure Sensor — The ambient pressure sensor indicated
an invalid value. The ambient pressure must be greater than 0.4
atmospheres to be valid.
IElectronics Compartment Temperature — The temperature of the
instrument electronics compartment was outside of its usual operating range. The electronics temperature must be between -5°C and
60°C or the electronics may not operate reliably.
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-1
Service Manual, Partisol Model 2000 Air Sampler
2.1.3.I HAVECHOSENAFLOWCHARTTOBEGINWITH. NOW WHAT?
Now that you have chosen a particular flow chart, begin in the upper left corner of the
chart and follow the “flow” of instructions until you can go no further. These
instructions will bring you through a logical succession of steps to help you determine
where the problem is and what to do about it. Document the result of each step as it
is completed. This information is important if R&P is contacted.
Revision B.000
2.1.4.H
OWDO I CHECKACABLE’SCONTINUITY?
Throughout the flow charts are instructions which indicate that a cable’s continuity
should be checked. You must check a cable’s continuity to verify that a particular
cable has not been damaged.
Follow these steps to check a cable’s continuity:
1) Set the multimeter so it will measure resistance (W).
2) Determine which cable you are checking and locate its corresponding connectors.
3) Refer to the interconnect diagrams in Section 4 to determine
which pins are connected from one connector to the other. (For
example, is pin 1 on one connector connected to pin 1 on the
other connector or to a different pin?)
4) Place one lead of the multimeter on the first pin on one of the
connectors and place the other lead on the corresponding pin of
the other connector. The resistance should be very close to zero
if the continuity is correct.
5) Repeat steps 2-4 for each set of pins being checked.
2.1.5.HOWDO I LOCATEATESTPOINTONANELECTRONICBOARD?
Refer to the schematic diagrams for the electronics in Section 4. A drawing of the
interface board is included in this section and test points are labeled.
2.1.6.HOWDO I FINDPIN 1?
There are four types of connectors used on the Partisol Monitor, “D”-type connectors,
ribbon cable connectors, 2-, 4- and 10-pin Molex connectors and round 3-, 4- and 6pin connectors.
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-2
Service Manual, Partisol Model 2000 Air Sampler
2.1.6.1.D-TYPE CONNECTORS
There is only one “D”-type connector used on the Partisol Monitor. It is a female, 9pin, “D”-type connector labeled “RS-232” on the front panel.
Revision B.000
2.1.6.2.R
IBBON CABLE CONNECTORS
All ribbon cable connector pins are counted the same, regardless of the total number
of pins. However, the location of pin 1 may vary. The location of pin 1 is noted within
the silk-screen on the corresponding electronic board. For example purposes, a 24pin ribbon cable connector is displayed below.
2.1.6.3.M
OLEX CONNECTORS
The Molex connectors used on the Partisol Monitor are white and have 2 pins (such
as the hub valve connector (P5) on the underside of the interface board), 4 pins (such
as the AC input connector (P2) on the interface board), or 10 pins (such as the 24V/
5V transformer connector (P3) on the interface board). The easiest way to determine
which pin is pin 1 on a Molex connector is to determine what color wire corresponds
to pin 1. You can find this information in the interconnection diagrams in Section 4.
Once you determine the proper wire color, then you can examine the connector and
find pin 1.
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-3
Service Manual, Partisol Model 2000 Air Sampler
2.1.6.4.ROUND CONNECTORS
There are three round connectors used on the back of the Partisol Monitor: a 3-pin
round connector is used for the ambient temperature sensor, a 4-pin round connector
is used for the user output, and a 6-pin round connector is used for the anemometer.
Revision B.000
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-4
Service Manual, Partisol Model 2000 Air Sampler
2.2.TROUBLESHOOTING GUIDES
The troubleshooting guides included in the section should be used in conjunction with
the following revisions of the schematics included in Section 4 of this manual:
SchematicSchematicSchematic
DescriptionPart Number Revision
Interface Board81-005496 B
Interconnection Diagrams81-005912 C
Revision B.000
SECTION 2: TROUBLESHOOTING GUIDESPAGE 2-5
S
ECTION 2: TROUBLESHOOTING GUIDESPAGE 2-6
Error code
“T”
This error code occurs if
the ambient temperature
is out of range.
The temperature must be
between -40°C and 60°C
to be valid.
AMBIENT TEMPERATURE ERROR TROUBLESHOOTING FLOW CHART
Note:
1. The voltage at this test point is related to the
Replace cable
R&P P/N 51-005633
No
Continuity OK?
Check continuity of cable
between P54 and ambient
temperature sensor back
panel connector (P52) (see
note 2).
Yes
Check continuity of
cable between P4
(interface board) and
P54 (see note 3).
Replace cable
R&P P/N 51-006183
No
Continuity OK?
Yes
Replace ambient
temperature sensor cable
R&P P/N 51-004352-0001
2. Pins for continuity check of cable P54 to P52:
3. Pins for continuity check of cable P54 to P4:
temperature according to the following
equation:
1°K = 10mVDC (nominal)
P54P52
Pin 1Pin 1
Pin 2Pin 2
-----Pin 3
P54P4
Pin 1Pin 1
Pin 2Pin 2
Service Manual, Partisol Model 2000 Air Sampler
2.2.1.A
MBIENT TEMPERATURE ERROR (T) FLOW CHART
Ensure ambient temperature
probe is properly attached to
the back of the instrument.
Go to the statistics screen
and observe the ambient
temperature reading.
Place mult imeter ac ross test
points AMBTEMP (blue) and
AGND (black) on interface
board (see note 1).
Yes
Voltage =
~+15V?
Place multimeter across
test point +15V (red) and
AGND (black) on interface
board.
No
Voltage
nominal?
Yes
No
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Revision B.000
S
ECTION 2: TROUBLESHOOTING GUIDESPAGE 2-7
Display blank or
incomplete
Power s ystem
off then on.
Blank or
incomplete?
Incomplete
Check continuity of cable
P16A (interfac e board) to
P1 (display) (see note 1
on she et B).
Continui ty OK?
Yes
Install a working
display and recheck.
R&P P/N 10-002254
This section describes the service procedures required to maintain consistent operation of the Partisol Sampler. Additional maintenance procedures (such as cleaning the
PM-10 inlet) are contained in the Operating Manual. This section provides recommended intervals for maintenance and calibration procedures. However, when
determining your own intervals for maintenance and calibration, you must take into
account the quality assurance guidelines of your specific monitoring organization and
any special local considerations.
Maintenance ProceduresIntervalReference
Exchange large in-line filter6 monthsSection 3.1.1
Exchange fusesAs neededSection 3.1.2
Clean air intake filters6 monthsSection 3.1.3
Inspect “V” sealsduring filter exchangeSection 3.1.4
System leak test3 monthsSection 3.1.5
Test batteries6 monthsSection 3.1.6
Rebuild vacuum pump18 monthsRebuild kit
Revision B.000
Calibration ProceduresIntervalReference
Interface board calibration1 yearSection 3.2.1
Analog input calibration1 yearSection 3.2.2
Ambient air temperature calibration1 yearSection 3.2.3
Ambient pressure calibration1 yearSection 3.2.4
Mass flow controller calibration1 yearSection 3.2.5
3.1.MAINTENANCE PROCEDURES
This section contains maintenance procedures for the Partisol Sampler. Additional
maintenance procedures (such as cleaning the PM-10 inlet) are contained in the
Operating Manual.
3.1.1.EXCHANGINGTHE LARGE IN-LINE FILTER
The large in-line filter prevents particulate matter from entering the sampler’s flow
controller and causing damage.