Thermo Scientific PARTISOL 2000 Service Manual

Page 1
SERVICE MANUAL
PARTISOL® MODEL 2000
AIR SAMPLER
April 2000
Revision B
R&P Part Number 42-002521
25 Corporate Circle
Albany, NY 12203 USA
phone 518/452-0065
fax 518/452-0067
the new microweighing technology
Page 2
Page 3
Service Manual, Partisol Model 2000 Air Sampler
Section Revision List
As R&P instrumentation changes, so do our Operating and Service manuals. How­ever, these changes may affect only one aspect of an instrument, while leaving the instrument as a whole unchanged. To explain these individual changes to customers, R&P will update only those sections of its Operating and Service manuals that are affected by the instrument updates or improvements. As each manual section changes, so does its revision number, which is located at the top right corner of each page of each section.
To help customers keep track of the changes to the Partisol Model 2000 Air Sampler and its service manual, following is a list of the manual sections with their respective revision numbers:
Section Number and Description Revision Number
Section 1: Introduction B.000
Revision B.000
Section 2: Troubleshooting Guides B.000
Section 3: Service Procedures B.000
Section 4: Electrical Schematics B.000
Section 5: R&P Return Policy B.000
Appendix A: Hardware Connections B.000
Index B.000
PAGE I
Page 4
Service Manual, Partisol Model 2000 Air Sampler
Revision Description List
To help customers keep track of the changes to the Partisol Model 2000 Air Sampler and its service manual, following is a list of the manual sections with their respective revision descriptions. This list contains all changes made to the manual since the last update:
Section Number and Description Revision Description
Section 1: Introduction -Updated Overview of
Section 2: Troubleshooting Guides -All troubleshooting guides
Section 3: Service Procedures -All service procedures
Revision B.000
Manual section.
have been updated to reflect changes to electronics.
have been updated to reflect changes to electronics.
Section 4: Electrical Schematics -All electrical schematics
have been updated.
Section 5: R&P Return Policy - Changes have been made
to update return procedures.
Appendix A: Hardware Connections -Updated hardware
connections.
Index - Newly added section.
PAGE II
Page 5
Service Manual, Partisol Model 2000 Air Sampler
Table of Contents
SECTION 1: INTRODUCTION ............................................................................. 1-1
1.1. Overview of Manual ................................................................. 1-1
S
ECTION 2: TROUBLESHOOTING GUIDES ........................................................... 2-1
2.1. Using the Troubleshooting Guides........................................... 2-1
2.1.1. What equipment do I need? ............................................. 2-1
2.1.2. Where do I start? .............................................................. 2-1
2.1.3. I have chosen a flow chart to begin with. Now what? ...... 2-2
2.1.4. How do I check a cable’s continuity? ................................ 2-2
2.1.5. How do I locate a test point on an electronic board? ........ 2-2
2.1.6. How do I find pin 1? .......................................................... 2-2
2.1.6.1. D-Type Connectors .................................................... 2-3
2.1.6.2. Ribbon Cable Connectors........................................... 2-3
2.1.6.3. Molex Connectors ....................................................... 2-3
2.1.6.4. Round Connectors ...................................................... 2-4
2.2. Troubleshooting Guides ........................................................... 2-5
2.2.1. Ambient Temperature Error (T) Flow Chart ...................... 2-6
2.2.2. Display Troubleshooting Flow Charts — A and B ............. 2-7
2.2.3. Display Backlight Troubleshooting Flow Chart ................. 2-9
2.2.4. Electronics Compartment Error (I) Flow Chart................ 2-10
2.2.5. Flow Out of Range Error (F) Flow Charts — A and B ..... 2-11
2.2.6. Keypad Troubleshooting Flow Chart .............................. 2-13
2.2.7. Ambient Pressure Error (P) Flow Chart .......................... 2-13
2.2.8. Power Supply Flow Charts ............................................. 2-14
2.2.9. System Leak Troubleshooting Flow Chart...................... 2-18
2.2.10. Power Fail Error (E) Flow Chart ..................................... 2-18
2.2.11. User Analog Output Troubleshooting Flow Chart ........... 2-19
Revision B.000
S
ECTION 3: SERVICE PROCEDURES ................................................................ 3-1
3.1. Maintenance Procedures ......................................................... 3-1
3.1.1. Exchanging the Large In-Line Filter.................................. 3-1
3.1.2. Exchanging the Fuses ...................................................... 3-3
3.1.3. Cleaning the Air Intake Filter ............................................ 3-5
3.1.4. Inspecting “V” Seals ......................................................... 3-6
3.1.5. Leak Check....................................................................... 3-7
3.1.6. Testing the Batteries......................................................... 3-9
3.1.6.1. Testing the Round Coin Cell Battery (BT1) ................ 3-9
3.1.6.2. Testing the Battery-Backed RAM Socket Battery ..... 3-10
3.2. Calibration Procedures .......................................................... 3-11
3.2.1. Interface Board Calibration ............................................. 3-11
3.2.2. Analog Input Calibration ................................................. 3-14
3.2.3. Ambient Air Temperature Calibration ............................. 3-17
PAGE III
Page 6
Service Manual, Partisol Model 2000 Air Sampler
Table of Contents (continued)
3.2.4. Ambient Pressure Calibration ......................................... 3-18
3.2.5. Mass Flow Controller Calibration.................................... 3-19
3.3. Remove and Replace Procedures ......................................... 3-22
3.3.1. Removing/Replacing the Display.................................... 3-22
3.3.2. Removing/Replacing the Fan ......................................... 3-24
3.3.3. Removing/Replacing the Mass Flow Controller Assy ..... 3-25
3.3.4. Removing/Replacing the Electronics Heater/Fan Assy .. 3-26
3.3.5. Removing/Replacing the Interface/CPU Board Assy ...... 3-27
3.3.6. Removing/Replacing the Keypad ................................... 3-29
3.3.7. Removing/Replacing the Pump ...................................... 3-30
3.3.8. Removing/Replacing a Valve ......................................... 3-32
S
ECTION 4: ELECTRICAL SCHEMATICS ............................................................. 4-1
4.1. Electrical Schematics ............................................................... 4-1
4.1.1. Interface Board ................................................................. 4-2
4.1.2. Interconnect Diagrams ................................................... 4-10
Revision B.000
S
ECTION 5: R&P RETURN POLICY ................................................................. 5-1
5.1. R&P Return Policy — Instrument Components ....................... 5-1
5.2. R&P Return Policy — Systems ................................................ 5-3
5.3. Continental US Shipping Instructions ....................................... 5-4
5.4. International Shipping Instructions ........................................... 5-4
A
PPENDIX A: HARDWARE CONNECTIONS ......................................................... A-1
I
NDEX ............................................................................................................. I
PAGE IV
Page 7
Service Manual, Partisol Model 2000 Air Sampler
Section 1: Introduction
The information contained in this manual will allow the user to do preliminary troubleshooting to fix some problems or to give the maximum amount of information available to the factory.
1.1. OVERVIEW OF MANUAL
Section 1 : Introduction This section provides an overview of the Service Manual.
Section 2 : Troubleshooting Guides This section contains troubleshooting guides in the form of flow charts to aid the user when performing diagnostics.
Section 3 : Service Procedures This section contains the service procedures (maintenance and calibration) required to maintain consistent operation of the Partisol sampler. In addition, this section contains procedures describing installation of system components if exchange is necessary.
Revision B.000
Section 4 : Electronic Schematic Diagrams This section contains schematic diagrams and interconnect diagrams of the major electronic components in the instrument.
Section 5 : R&P Return Policy This section contains information regarding the return of whole systems or individual components for warranty or billable repair.
Appendix A: Hardware Connections This section contains the pin assignments of the output connectors on the Partisol Sampler.
SECTION 1: INTRODUCTION PAGE 1-1
Page 8
This page left intentionally blank.
Page 9
Service Manual, Partisol Model 2000 Air Sampler
Section 2: Troubleshooting Guides
This section contains troubleshooting guides in the form of flow charts to aid the user when performing diagnostics. This section should be used in conjunction with the electrical schematics located in Section 4. Read Section 2.1 before using any of the troubleshooting guides.
2.1. USING THE TROUBLESHOOTING GUIDES
The troubleshooting guides give the user a logical succession of steps to follow when diagnosing a problem. Following is a list of the questions most commonly asked by users with regard to the guides.
Revision B.000
2.1.1. W
HAT EQUIPMENT DO I NEED?
The main piece of equipment needed is a multimeter (that measures DC and AC voltages and resistance) with at least a 4 digit resolution.
2.1.2. WHERE DO I START?
Look at the main screen of the Partisol Model 2000 Air Sampler. Any active errors will be indicated in the upper left corner. Possible error codes are:
OK Great! You have no errors! E Power Failure — A system power failure occurred during sampling. F Flow Out of Range — The measured sample flow rate through the
sample filter deviated by more than 7% from its setpoint for more than 1 minute.
T Ambient Temperature Sensor — The ambient temperature sensor
was not installed correctly or indicated an invalid value. The ambient temperature must be between -40°C and 60°C to be valid.
P Ambient Pressure Sensor — The ambient pressure sensor indicated
an invalid value. The ambient pressure must be greater than 0.4 atmospheres to be valid.
I Electronics Compartment Temperature — The temperature of the
instrument electronics compartment was outside of its usual operat­ing range. The electronics temperature must be between -5°C and 60°C or the electronics may not operate reliably.
SECTION 2: TROUBLESHOOTING GUIDES PAGE 2-1
Page 10
Service Manual, Partisol Model 2000 Air Sampler
2.1.3. I HAVE CHOSEN A FLOW CHART TO BEGIN WITH. NOW WHAT?
Now that you have chosen a particular flow chart, begin in the upper left corner of the chart and follow the “flow” of instructions until you can go no further. These instructions will bring you through a logical succession of steps to help you determine where the problem is and what to do about it. Document the result of each step as it is completed. This information is important if R&P is contacted.
Revision B.000
2.1.4. H
OW DO I CHECK A CABLES CONTINUITY?
Throughout the flow charts are instructions which indicate that a cable’s continuity should be checked. You must check a cable’s continuity to verify that a particular cable has not been damaged.
Follow these steps to check a cable’s continuity:
1) Set the multimeter so it will measure resistance (W).
2) Determine which cable you are checking and locate its corre­sponding connectors.
3) Refer to the interconnect diagrams in Section 4 to determine which pins are connected from one connector to the other. (For example, is pin 1 on one connector connected to pin 1 on the other connector or to a different pin?)
4) Place one lead of the multimeter on the first pin on one of the connectors and place the other lead on the corresponding pin of the other connector. The resistance should be very close to zero if the continuity is correct.
5) Repeat steps 2-4 for each set of pins being checked.
2.1.5. HOW DO I LOCATE A TEST POINT ON AN ELECTRONIC BOARD?
Refer to the schematic diagrams for the electronics in Section 4. A drawing of the interface board is included in this section and test points are labeled.
2.1.6. HOW DO I FIND PIN 1?
There are four types of connectors used on the Partisol Monitor, “D”-type connectors, ribbon cable connectors, 2-, 4- and 10-pin Molex connectors and round 3-, 4- and 6­pin connectors.
SECTION 2: TROUBLESHOOTING GUIDES PAGE 2-2
Page 11
Service Manual, Partisol Model 2000 Air Sampler
2.1.6.1. D-TYPE CONNECTORS
There is only one “D”-type connector used on the Partisol Monitor. It is a female, 9­pin, “D”-type connector labeled “RS-232” on the front panel.
Revision B.000
2.1.6.2. R
IBBON CABLE CONNECTORS
All ribbon cable connector pins are counted the same, regardless of the total number of pins. However, the location of pin 1 may vary. The location of pin 1 is noted within the silk-screen on the corresponding electronic board. For example purposes, a 24­pin ribbon cable connector is displayed below.
2.1.6.3. M
OLEX CONNECTORS
The Molex connectors used on the Partisol Monitor are white and have 2 pins (such as the hub valve connector (P5) on the underside of the interface board), 4 pins (such as the AC input connector (P2) on the interface board), or 10 pins (such as the 24V/ 5V transformer connector (P3) on the interface board). The easiest way to determine which pin is pin 1 on a Molex connector is to determine what color wire corresponds to pin 1. You can find this information in the interconnection diagrams in Section 4. Once you determine the proper wire color, then you can examine the connector and find pin 1.
SECTION 2: TROUBLESHOOTING GUIDES PAGE 2-3
Page 12
Service Manual, Partisol Model 2000 Air Sampler
2.1.6.4. ROUND CONNECTORS
There are three round connectors used on the back of the Partisol Monitor: a 3-pin round connector is used for the ambient temperature sensor, a 4-pin round connector is used for the user output, and a 6-pin round connector is used for the anemometer.
Revision B.000
SECTION 2: TROUBLESHOOTING GUIDES PAGE 2-4
Page 13
Service Manual, Partisol Model 2000 Air Sampler
2.2. TROUBLESHOOTING GUIDES
The troubleshooting guides included in the section should be used in conjunction with the following revisions of the schematics included in Section 4 of this manual:
Schematic Schematic Schematic Description Part Number Revision
Interface Board 81-005496 B Interconnection Diagrams 81-005912 C
Revision B.000
SECTION 2: TROUBLESHOOTING GUIDES PAGE 2-5
Page 14
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-6
Error code
“T”
This error code occurs if the ambient temperature
is out of range.
The temperature must be between -40°C and 60°C
to be valid.
AMBIENT TEMPERATURE ERROR TROUBLESHOOTING FLOW CHART
Note:
1. The voltage at this test point is related to the
Replace cable
R&P P/N 51-005633
No
Continuity OK?
Check continuity of cable
between P54 and ambient
temperature sensor back
panel connector (P52) (see
note 2).
Yes
Check continuity of
cable between P4
(interface board) and
P54 (see note 3).
Replace cable
R&P P/N 51-006183
No
Continuity OK?
Yes
Replace ambient
temperature sensor cable
R&P P/N 51-004352-0001
2. Pins for continuity check of cable P54 to P52:
3. Pins for continuity check of cable P54 to P4:
temperature according to the following equation:
1°K = 10mVDC (nominal)
P54 P52
Pin 1 Pin 1 Pin 2 Pin 2
----- Pin 3
P54 P4
Pin 1 Pin 1 Pin 2 Pin 2
Service Manual, Partisol Model 2000 Air Sampler
2.2.1. A
MBIENT TEMPERATURE ERROR (T) FLOW CHART
Ensure ambient temperature probe is properly attached to
the back of the instrument.
Go to the statistics screen
and observe the ambient
temperature reading.
Place mult imeter ac ross test
points AMBTEMP (blue) and
AGND (black) on interface
board (see note 1).
Yes
Voltage =
~+15V?
Place multimeter across
test point +15V (red) and
AGND (black) on interface
board.
No
Voltage
nominal?
Yes
No
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Revision B.000
Page 15
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-7
Display blank or
incomplete
Power s ystem
off then on.
Blank or
incomplete?
Incomplete
Check continuity of cable P16A (interfac e board) to
P1 (display) (see note 1
on she et B).
Continui ty OK?
Yes
Install a working display and recheck. R&P P/N 10-002254
Display n ow
work?
No
Replace interface/CPU
board assembl y R&P P/N 50 -005496
& 10-002407
Hit several keys
Blank
on the keypad.
Replace cable
No
R&P P/N 51-005637
Original display was
Yes
A
defective.
DISPLAY TROUBLESHOOTING FLOW CHART-A
Is backlight on?
No
Did pum p run
at all when p ower was
turned on?
No
Is the electronics compartment fan
running?
No
Ensure th e proper
voltage is present at
the AC outlet.
AC voltage OK?
No
Correct problem.
Display contrast
Yes
Yes
may need
adjustment.
Yes
Go to B
electronics
Yes
Check the continuity of all the
compartment
temperature
Replace interface/CPU
board as sembly
R&P P/N 50-005496
& 10-002407
AC components between the
power cord and P2 on th e
interface board.
>65°C?
Adjust contrast with
Is
No
knob on fro nt of
display.
Yes
Resume normal
operation.
Yes
Able to adjust
contrast?
Replace assembly
R&P P/N 51-005668
Place hand near (d o NOT
touch! ) th e heater at t he
top of the electronics
compartment.
Heat coming from heater?
Compone nts
OK?
No
Replace defective
component.
Yes
No
No
Wait for inst rument to
NoYes
Replace interface/CP U
R&P P/N 50-005496
Check contin uity of
contrast adjustment
pot assembly.
Continui ty OK?
Yes
Perform system
reset.
Still Blank
Go to A
cool do wn.
board assembly
& 10-002407
Service Manual, Partisol Model 2000 Air Sampler
2.2.2. D
ISPLAY TROUBLESHOOTING FLOW CHARTS — A AND B
Revision B.000
Page 16
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-8
Service Manual, Partisol Model 2000 Air Sampler
B
Is pump still
running?
Yes
Is
electronics
compartment
temperature
>2°C?
Yes
Place m ulti meter across
test point +24V (red) and
PGND (black) on interface
board.
Voltage =~
24V?
Yes
Place multimeter across
test po int +6 V (red) and
PGND (black) on interface
board.
Perform system
No
No
reset.
Placehandnear(doNOT
touch!) the heater at the
No
top of the electronics
Check the 4A fuse used
for the 2 4.5V@3. 6A
voltage from
transformer
Voltage =~ 6V?
Perform system
compartment.
Yes
reset.
DISPLAY TROUBLESHOOTING FLOW CHART-B
Resume normal
operation.
Yes
Problem
corrected?
Heat co ming from heater?
Wait for instr ument to
warm up.
Replace fuse
R&P P/N 04-005510-00 04
No
Fuse OK?
Check the 5A fuse used
for the 8.3V@5.4A
No
voltage from transform er.
Yes
No
Go to A
No
Replace fuse
R&P P/N 04-003681
Check continuity
between the heater
terminals and P11 on
the interface board.
Yes
Fuse OK?
No
Yes
R&P P/N 50-005496 & 10-00 2407
Note:
1. Pins for continuity check of cable P16A to P1:
P16A P1 P16A P1
Pin1 Pin1 Pin12 Pin12 Pin2 Pin2 Pin13 Pin13 Pin3 Pin3 Pin14 Pin14 Pin5 Pin5 Pin15 Pin15 Pin6 Pin6 Pin16 Pin16 Pin7 Pin7 Pin17 Pin17 Pin8 Pin8 Pin18 Pin18 Pin9 Pin9 Pin19 Pin19 Pin10 Pin1 Pin20 Pin20 Pin 11 Pin 1 1
Continuit y OK? No
Yes
Replace interface/CPU
board assembly
R&P P/N 50 -005496
& 10-002407
Replace interface/CP U board
assembly with known good one
Replace heate r
R&P P/N 56-005909-012 0
or 56-005909-02 40
Replace transforme r R&P P/N 03-002402
No
Problem
corrected?
Yes
Original boar ds were
defective.
Revision B.000
Resume normal
operation.
Yes
Problem
corrected?
No
Contact R&P.
Page 17
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-9
Backlight not on
Power system
off then on.
DISPLAY BACKLIGHT TROUBLESHOOTING FLOW CHART
Note:
1. Pins for continuity check of cable P16B to display:
P16B Display
Pin 1 A Pin 2 K
Service Manual, Partisol Model 2000 Air Sampler
2.2.3. D
ISPLAY BACKLIGHT TROUBLESHOOTING FLOW CHART
Hit several keys on
the keypad.
Is backlight on?
Yes
Resume normal
operation.
Go to diagnostics
screen and turn on
No
backlight.
Key on keypad pressed
before may be bad.
Yes
CheckstatusofLED “BLITE” on interface
NoIs backlight on?
Perform a s ystem reset and recheck
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
board.
Check continuity of cable
P16B (interface board)
Is LED on?
No
the LED.
LED on? Recheck backlight.
No
Yes
and display connection
Yes
(see note 1).
Replace display
R&P P/N 10-002254
Yes
Continuity OK?
No
Replace cable
R&P P/N 51-004665
Revision B.000
Page 18
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
2.2.4. E
board assembly
R&P P/N 50-005496
Replace interface/CPU
No
Yes
Replace heater
or 56-005909-0240
R&P P/N 56-005909-0120
present?
AC line voltage
heater term inals.
Place multimeter
across line voltage
LECTRONICS COMPARTMENT ERROR (I) FLOW CHART
temperature.
Recheck instr ument
Yes
temperature.
Recheck instrument
No
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
& 10-002407
No
LED off?
LED status.
Perform system
reset and recheck
No
Yes
LED off?
temperature.
Recheck instrument
Yes
on?
Electronics
heater status
heater status.
Perfo rm a syst em
reset and recheck
No
Yes
LED on?
No
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
No
Perform system
reset and recheck.
LED should be
Observe t he LED
Yes
LED on?
is off.
on when heater
interface board.
labeled “HEAT” on the
Yes
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
No
Yes
is o n.
LED should be
off when heater
Yes
interface board.
Observe t he LED
labeled “HE AT” on the
on?
Electronics
heater status
heater.
Go to the diagnostics
screen and observe the
status for the instrument
No
temperature.
Recheck instrument
OFF?
Electronics
heater status
No
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
No
heater status.
Perfo rm a syst em
reset and recheck
OFF?
Electronics
heater status
compartment heater.
Go to the diagnostics
screen and observe the
status for the electronics
Yes
to run properly.
too hot for instrument
Ambient temperature may be
Yes
60°C?
Electronics
temperature >=
Yes
ELECTRONICS COMPARTMENT TEMPERATURE ERROR TROUBLESHOOTING FLOW CHART
1°K = 10mVDC
temperature acc ording to the following
equation:
Note:
1. The voltage at this test point is related to the
“I”
Error code
the electronics
is out of range.
This error code occurs if
compartment temperature
electronics may not be
be valid. Outside of this
This temperature must be
bet ween -5 °C and 60° C to
range, the operation of the
Yes
to warm up.
Wait for instrument
reliable.
Heat coming
from heater?
Place hand near (do NOT
Go to the statistics screen
and observe the reading for
the instrumenttemperature.
Place multimeter across
Yes
compartment.
top of the electronics
touch!) the heater at the
(see note 1).
interface board
and AGND (black) on
test points BDTEMP (blue)
No
Electronics
temp < 15°C?
Yes
Voltage
nominal?
compartment.
of the electronics
Place hand near (do NOT
touch!) the heater at the top
No
& 10-002407
board assembly
R&P P/N 50-005496
Repla ce inte rface /CPU
No
Heat coming
from heater?
operating?
Electronics
compartment fan
No
compartment fan.
Replace electronics
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-10
Page 19
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
2.2.5. F
Contact R&P.
Yes
voltage
Voltage
displayed?
measured ~
No
Observe t he value
LOW OUT OF RANGE ERROR (F) FLOW CHARTS — A AND B
No
& 10-002407
board assembly
R&P P/N 50-005496
Replace interfac e/CPU
displ ayed fo r FLOW .
LED On?Yes
Recheck f low.
pump.
Hub1 va lve and t he
Go to the diagnostics
screen and turn on the
LED status.
Turn on valveand recheck
interface board.
Place theleads of a
AGND (black) on the
multimeter acrosstest
points MFS (blue) and
No
No
reset.
Perform asystem
Replace valve
R&P P/N 32-002379
LED On?
assembly (P5,6,7 or 8 to
Check cont inuity o f valve
No
ContinuityO K?
If the LED is on,
correspondingv alve).
Yes
board assembly
Replace interface/CPU
B
be on.
the valve should
R&P P/N 50-005496
Chec k the
& 10-002407
Yes
on interfaceboard.
correspondingv alve’s
LED (Hub1, Sat2, etc.)
Remove the hosethat
No
Flow rate
adjusted?
change as
to the valve manifold.
attaches the back side
of the large in-line filter
NoYes
Flow rate
adjusted?
change as
Go to
the flow rate.
Press the <Decrflw> or
<Incrflw> key and observe
Able to
cali brate?
the mass flow sensor
inter face b oard (P 21A)
connector (P60) and the
(see note 1 onsheet B).
Check continuity between
No
ContinuityO K?
Yes
FLOW OUT OF RANGE ERROR TROUBLESHOOTING FLOW CHART - A
Replace mass flow
controller assembly
Replace cable
R&P P/N 51-004656-0018
& 10-002407
board assembly
R&P P/N 55-002663
R&P P/N 50-005496
Replace interface/CPU
ambient
pressur e
Refer to th e
Yes
Recheck fl ow.
No
Yes
LED On?
ambient
flow chart.
Refer to th e
troubleshooting
temperature
troubleshooting
Yes
Yes
Recheck flow.
pump.
Go to the diagnostics
screen and turn on the
flow chart.
reset.
Perfo rm a syst em
No
Perform a flow
cont roller c alibrati on.
Yes
Check LED “Pump” on
occur?
Did “ consta nt
out of range” error
Yes
ambient
pressure
flow chart.
Refer to th e
troubleshooting
No
LED On?
on, t he pum p shoul d be of f).
interface board (if the LED is
AC line
voltage present
No
board assembly
Repla ce inte rface/ CPU
No
Perform an ambi ent
pres sure ca librati on.
board)?
at P10 (interface
Replace pump
& 10-002407
or 54-005949-0240
R&P P/N 54-005949-0120
R&P P/N 50-005496
occur?
Did “ consta nt
out of range” error
Yes
ambient
flow chart.
Referto the
temperature
troubleshooting
Yes
between P10 (interface
board) and P53 (pump)
Check continuity of cable
Yes
ContinuityO K?
Perform an ambi ent
temperature calibration.
(see note 2 on sheet B).
No
Replace cable
R&P P/N 51-005910
Go to the calibration screen,
problem station.
turn on the pump and the
valve corresponding to the
hub or satellite.
Ensure that nothing is
blocking the inlet to the
filter is not clogged.
Ensure that the sample
clogged.
in-lin e filte r is no t
Ensure that the large
“F”
Error code
setpoint for 1 minute.
rate through the filter has not
Thiserrorcodeoccursiftheflow
been maintained within 7% of the
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-11
No
present?
Is s tatus “ P”
No
present?
Is status “T”
Go to the statistics scree n and
Sat2, etc.) was sampling when
determine which station (Hub1,
pump .
error occurred.
Go to the diagnostics
screen and turnon the
Yes
Is pump oper ating? No
Turn off the pump.
properlyc onnected
between pump and
back of enclosure).
Ensure that hoses are
accumulator (located on
assembly.
mass flow control
properlyc onnected
Ensure that hoses are
between accum ulator and
on mass flow controller
Ensure that manual valves
assembly are open
flow).
(handles in direction of
Page 20
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
P60 P21A
Pin 1 Pin 1 {MAS Out}
Pin 2 ----- {Not Used}
Pin 3 Pin 3 {MAS Gnd}
Pin 4 Pin 4 {MAS V+}
P10 P53
Pin 1 Pin 1 {Line}
Pin 2 Pin 2 {Neutral}
Pin 3 ----- {Not Used}
Pin 4 ----- {Not Used}
P61 P21B
Pin 1 Pin 1 {+24V}
Pin 2 Pin 2 {-}
Note:
1. Pins forcontinuitycheck of cable P60 to P21A:
2. Pins forcontinuitycheck of cable P10 to P53:
3. Pins forcontinuitycheck of cable P61 to P21B:
Recheck fl ow.Recheck flow.
Yes
No
Voltage vary as
Replace mass flow
controller assembly
R&P P/N 55-002663
FLOW OUT OF RANGE ERROR TROUBLESHOOTING FLOW CHART - B
flow is adjusted?
Yes
No
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
rech eck vol tage.
Reset the system and
No
Voltage
adjusted?
var y as fl ow is
Able to adjust?
Yes
Recheck fl ow.
Yes
Recheck fl ow.
interface board.
and AGND (black) on the
across test points MFS (blue)
Plac e the l eads of a mult imeter
No
recheck voltage.
Reset the s ystem and
Voltage vary as
flow is adjusted?
No
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
Able to adjust?Yes
Recheck fl ow.
No
recheck voltage.
Reset the system and
Voltage2.5V if motor is not
No
the correct voltage.
the interfac e board to obtain
Adjust potentiometer R53 on
setflowratetozero.
Turn off the pump and
Yes
Voltage vary as
flow is adjusted?
depends on direction).
movi ng, vol tage = 5V or 0
when motor is moving (voltage
Voltage = 2.5V
Place the leads of a multimeter
across test points 2.5V (red) and
Yes
±0.002?
board.
PWM0 (green) on the interface
& 10-002407
board assembly
R&P P/N 50-005496
Replace interface/CPU
the interface board.
across test points PWM0
(green) and DGND (black) on
Place the leads of a multimeter
Turn on pump.
Yes
B
Turn off the pum p.
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-12
should be heard as it turns.
As flow is adjusted, the motor
No
turning?
Hear motor
on the interfac e board.
across test points MOTORA
(green) and MOTORB (green)
Place the leads of a multimeter
turning.
Voltage 0 V if motor off,
voltag e ≈ ±8.4 to 14V when
No
Yes
8.4 and 14V?
Voltage between
(P21B) (see note 3).
(P61) to the interface board
from the MFC motor connection
Check the co ntinuit y of th e cable
Replace mass flow
controller assembly
R&P P/N 55-002663
No
Conti nuity OK ? Yes
Replace cable
R&P P/N 51-005904
Page 21
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-13
Service Manual, Partisol Model 2000 Air Sampler
KEYPAD TROUBLESHOOTING FLOW CHART
No response from keypad
Perform system
reset.
If onl y one key is bad,
keypad is probably
defective.
Respo nse
from keys on
keypad?
Only one or
several keys bad
Replace keypad with
known g ood one
R&P P/N 55-002345
New k eypad fix pro blem ?
Yes
Return defective k eypad
for rep lacem ent.
No res ponse
Check conti nuity of cable between P15
(interface board) and
P70 (keypad).
No
Replac e cab le
R&P P/N 51-002376
No
Contin uity OK?
Yes
Replace inte rface/CPU
board assem bly
R&P P/N 50-005496
& 10-0 02407
AMBIENT PRESSURE TROUBLESHOOTING FLOW CHART
Error code
“P”
This error code occurs if
the ambient pressure is
out of range.
The pressur e mus t be greater than 0.4 atm to
be valid.
Go to th e statist ics screen
and ob serve the ambient
pressure reading.
Place mul timeter across test
points AMBP RESS (blue) a nd
AGND (black) on interface
board (see note 1).
Voltage
nomina l?
No
Replace int erface/CPU
board assembly
R&P P/ N 50- 005496
& 10- 002407
Yes
Note :
1. Th e vol tage at this test point is rel ated to the pressure according to the following equation:
V * 0.2 194 + 0.1042 15 = P (atm )
Resume normal
operat ion.
Yes
Perform system
reset.
No
Proble m
correct ed?
2.2.6. K
2.2.7. AMBIENT PRESSURE ERROR (P) FLOW CHART
EYPAD TROUBLESHOOTING FLOW CHART
Revision B.000
Page 22
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-14
Service Manual, Partisol Model 2000 Air Sampler
115 / 230 VAC
Corcom
Power switch /
circuit breaker (5A)
P2
Thermostat (TH103)
opens if > 150F (66C)
POWER SUPPLY FLOW CHART (115 / 230 VAC)
Jumper
23
Pump
relay (RL1)
AC voltage selector
(SW1)
Electronics
compartment
heater relay
(RL2)
P10
P3
P11
24.5V @ 3.6A
24.5 V/8.3V transformer
8.3V @ 5.4A
Line voltage electronics
compartment heater
2.2.8. P
OWER SUPPLY FLOW CHARTS
PumpP53
Interface Board
P101
Electronics
compartment fan
Revision B.000
Page 23
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-15
Service Manual, Partisol Model 2000 Air Sampler
24.5V/8.3V transformer
24.5V @ 3.6A
Fuse(4A)
POWER SUPPLY FLOW CHART (24.5 VAC / 24VDC)
P3
Thermostat (TH2)
opens if > 140F (60C)
Thermostat (TH1)
closes if > 35F (2C)
Jumper
12
24VDC voltage
regulator (VREG1)
Heater
relay (RL2)
24VDC
P11
Pump
relay control
(RL1)
Relay #3
control
+15VDC voltage
regulator
(VREG101)
P5
P6
+15V
Low voltage electronics
compartment heater
(Field retrofit only)
Hub valve
Satellite 2 valve
Interface Board
P7
P8
P9
Satellite 3 valve
Satellite 4 valve
Pump compartment
fan
Revision B.000
Page 24
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-16
Service Manual, Partisol Model 2000 Air Sampler
POWER SUPPLY FLOW CHART (15VDC / 10VDC / 5VDC / 5V REF)
24VDC
+15VDC voltage
regulator
(VREG101)
Interface Boar d
10 VDC voltage
regulator (VREG3)
5 VDC voltage
regulator (VREG4)
5V +/- 2.5mV A/D reference
(U501)
P18
P4 P52
10VDC P60
5VDC
P21A
Ambient pressure
tran sducer (PT1)
Vacuum transducer
(PT5 02)
P17
5V REF
P54
P50
Mass flow sensor
CPU Board
Anemometer
Ambient temperature
sensor
Revision B.000
Electronics compartment temperature sensor (U4)
Page 25
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-17
24.5V/8.3V
tran sf ormer
8.3V @ 5.4A
Fuse(5A)
P3
Relay #3
POWER SUPPLY FLOW CHART (8.3 VAC / 2.5VDC / 6VDC)
Interface Board
2.5VDC voltage regulator
(VREG601)
2.5 VDC
Service Manual, Partisol Model 2000 Air Sampler
6VDC voltage
regulator (VREG2)
Mass flow controller
6VDC
motor drive
D2
P1B
5.2 VDC
P21B P61
P16B
Mass flow controller
motor
LDC backlight
(control)
CPU Board
Revision B.000
Page 26
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-18
Service Manual, Partisol Model 2000 Air Sampler
SYSTEM LEAK TROUBLESHOOTING FLOW CHART
System fails
leak test
Install a new filter c asset te
and sam ple filter and
repeat leak test.
Leak test still fails
Pinch the hose downstream
of the sam ple f ilter in the hub
or satellite and repeat leak
test.
Leak test pass?
No
Pinch the hose
upstre am of the in-line
filter in the hub and
repeat leak test.
Yes
If system still leaks,
contact R&P.
Check fittings on
underside of the
sample filter.
Ensure that the fl ow
audit adapte r is not
damage d.
Ensure the downtube
is pr operly and f ully
installed into the hub
or satellite.
Checkfittingsonmassflow
controller assembly. If no problem
is foun d, r eplace the assembly.
No
POWER FAILURE ERROR FLOW CHART
Error C ode
“E”
The power has been
interrupted to t he sys tem
during sampling.
Program will continue after power has been
restored.
The “E” error code will be
displayedonthemain
screen unti l the instrument is
put in to st op mo de.
Resume normal
operation.
2.2.9. S
2.2.10. POWER FAIL ERROR (E) FLOW CHART
YSTEM LEAK TROUBLESHOOTING FLOW CHART
Leak test pass?
Yes
Check fit tings and
hoses on va lve ma nifol d. If
no prob lem is fou nd, r eplace
the corresponding valve.
Pinc h the hose
downst ream of t he in-
No
line filter in the hub and
repeat leak test.
Leak test pass?
Yes
Proble m is in hose attac hment s to
the in-line filter or the filter itself.
Revision B.000
Page 27
S
ECTION 2: TROUBLESHOOTING GUIDES PAGE 2-19
Service Manual, Partisol Model 2000 Air Sampler
User analog output
errors
Refer to Appendix C of the
Operating Manual for
wiring the analog outputs
Is the
analog output wired
properly?
Yes
Go to the setup screen and
ensure the desired analog
output type ("Flow" or "Stat")
has been set.
Analog output
type set correctly?
USER ANALOG OUTPUT TROUBLESHOOTING FLOW CHART
Note:
1. Pins for continuity check of cable P51 to P18:
P51 P18
Pin 1 ----­Pin 2 Pin 9 Pin 3 Pin 10 Pin 4 -----
No Correct the wiring error
Perform a system
Yes
reset.
Replace cable
R&P P/N 51-005915
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Continuity OK?Yes
Check continuity between
the user output connector on
the back of the enclosure
(P51) and the interface
board (P18) (see note 1).
Voltage = 1.0 V?
No
Voltage = 0.5 V?No Yes
Set run type to
manual on setup
Yes
screen.
No
Begin sampling by
Return to the main
instrument into run
Call R&P.
pressing <Shift>
followed by
<F6:Hub1>.
screen and press
<F4:Run/Stp> to
place the
mode.
2.2.11. U
SER ANALOG OUTPUT TROUBLESHOOTING FLOW CHART
No
Correct analog output setting.
Problem
corrected?
Yes
Resume normal
operation.
Ensure the instrument is in the
stop operating mode. If not,
No
returntothemainscreenand
press <F4:Run/Stp>
Place the leads of a
multimeter across test
points PWM1 (green) and
AGND (black) on the
interface board.
Go to the setup screen
and set the analog
outputtypeto"STAT."
Revision B.000
Page 28
This page left intentionally blank.
Page 29
Service Manual, Partisol Model 2000 Air Sampler
Section 3: Service Procedures
This section describes the service procedures required to maintain consistent opera­tion of the Partisol Sampler. Additional maintenance procedures (such as cleaning the PM-10 inlet) are contained in the Operating Manual. This section provides recom­mended intervals for maintenance and calibration procedures. However, when determining your own intervals for maintenance and calibration, you must take into account the quality assurance guidelines of your specific monitoring organization and any special local considerations.
Maintenance Procedures Interval Reference
Exchange large in-line filter 6 months Section 3.1.1 Exchange fuses As needed Section 3.1.2 Clean air intake filters 6 months Section 3.1.3 Inspect “V” seals during filter exchange Section 3.1.4 System leak test 3 months Section 3.1.5 Test batteries 6 months Section 3.1.6 Rebuild vacuum pump 18 months Rebuild kit
Revision B.000
Calibration Procedures Interval Reference
Interface board calibration 1 year Section 3.2.1 Analog input calibration 1 year Section 3.2.2 Ambient air temperature calibration 1 year Section 3.2.3 Ambient pressure calibration 1 year Section 3.2.4 Mass flow controller calibration 1 year Section 3.2.5
3.1. MAINTENANCE PROCEDURES
This section contains maintenance procedures for the Partisol Sampler. Additional maintenance procedures (such as cleaning the PM-10 inlet) are contained in the Operating Manual.
3.1.1. EXCHANGING THE LARGE IN-LINE FILTER
The large in-line filter prevents particulate matter from entering the sampler’s flow controller and causing damage.
Tools Needed: Phillips screwdriver
Needle nose pliers
Materials: Large in-line filter (R&P P/N 32-002643)
SECTION 3: SERVICE PROCEDURES PAGE 3-1
Page 30
Service Manual, Partisol Model 2000 Air Sampler
Pump
compartment
cover
Electronics compartment door
Flow controller access door
Contrast knob
Large in-line filter
Black filter
holder
Follow these steps to exchange the large in-line filter:
1) Turn off the sampler and remove the pump compartment cover (Figure 3-1).
Figure 3-1. Partisol Sampler with door open.
Revision B.000
2) Locate the large in-line filter (inside the black filter holder) to the right of the valve manifold (Figure 3-2).
Figure 3-2. Location of the large in-line filter.
SECTION 3: SERVICE PROCEDURES PAGE 3-2
Page 31
Service Manual, Partisol Model 2000 Air Sampler
Hose clip
Large in-line filter
3) Remove the in-line filter from its holder and examine the hose clips which secure the hoses to each end of the filter (Figure 3-3).
Figure 3-3. Large in-line filter removed from its black filter holder.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
4) Using a pair of needle nose pliers, squeeze the tabs on one of the hose clips and remove the corresponding hose from the end of the filter. Repeat this procedure for the other hose clip and hose.
5) Place the hoses and clips onto the new filter using the same method used in step 4. Replace the filter into its holder.
6) Replace the pump compartment cover.
7) Turn on the sampler and perform a system leak check (Section
3.1.5).
3.1.2. EXCHANGING THE FUSES
The Partisol Sampler contains two fuses: a 5A and a 4A fuse. Both fuses are installed into in-line fuse holders located between the transformer and the interface board. The 5A fuse is used for the 8.3V @ 5.4A voltage from the transformer and the 4A fuse is used for the 24.5V @ 3.6A voltage from the transformer. The fuse must be removed from the fuse holder to be identified.
Tools Needed: Phillips screwdriver Materials: 5A fuse (R&P P/N 04-003681)
4A fuse (R&P P/N 04-005510-0004)
SECTION 3: SERVICE PROCEDURES PAGE 3-3
Page 32
Service Manual, Partisol Model 2000 Air Sampler
Fuse
Fuse holders
Follow these steps to exchange the 5A and/or 4A fuse:
1) Turn off the sampler and disconnect the unit’s power cord from its AC source.
2) Remove the three screws and open the electronics compartment door (Figure 3-1).
3) Locate the in-line fuse holder (Figure 3-4). Grasp the holder on both sides and unscrew it until the two halves come apart (Figure 3-5).
4) Remove the fuse from the fuse holder and determine its rating. The fuse rating should be stamped on one of the ends of the fuse.
Figure 3-4. Inside the electronics compartment.
Revision B.000
Figure 3-5. Inside the electronics compartment with an open fuse holder highlighted.
SECTION 3: SERVICE PROCEDURES PAGE 3-4
Page 33
Service Manual, Partisol Model 2000 Air Sampler
Fan
5) Place a new fuse with the proper rating into the holder and screw the two halves of the fuse holder together to secure.
6) Close the electronics compartment door and secure it with the three screws that you previously removed.
7) Connect the power cord to the AC source and resume normal operation.
Revision B.000
Figure 3-6. Location of fan inside the pump compart­ment.
3.1.3. C
LEANING THE AIR INTAKE FILTER
The Partisol Sampler contains an air intake filter attached to the fan in the pump compartment. This filter cleans the air flow drawn into the instrument enclosure by the fan.
Tools Needed: Phillips screwdriver
Mild soap solution (if required)
Materials: Air intake filter (R&P P/N 10-002799)
Follow these steps to clean the air intake filter:
1) Turn off the sampler and remove the pump compartment cover (Figure 3-1).
2) Locate the air intake fan on the left side of the pump compartment (Figure 3-6).
SECTION 3: SERVICE PROCEDURES PAGE 3-5
Page 34
Service Manual, Partisol Model 2000 Air Sampler
3) Snap off the fan cover on the outside of the enclosure. This cover contains the air intake filter and a screen.
4) Take out the filter and screen and brush any dirt off. If they are covered with a sticky coating, wash both the filter and with a mild soap solution and water. If either part is damaged, they should be replaced.
5) If the filter and/or screen was washed in the liquid solution, allow them to dry before reinstalling them into the cover.
6) Reinstall the filter and screen into the cover and remount the assembly.
7) Replace the pump compartment cover.
8) Turn on the sampler and resume normal operation.
Revision B.000
3.1.4. INSPECTING “V” SEALS
There are two “V” seals used in the hub unit of the Partisol sampler: one mounted above and one mounted below the filter cassette. Also, there are two “V” seals used in each satellite unit. These seals ensure that the sample flow passes through the sample filter and seals the filter from leaks. These seals must be kept clean and inspected frequently for damage.
Tools Needed: Clean dry cloth Materials: Upper “V” seal (R&P P/N 22-002680)
Lower “V” seal (R&P P/N 22-005957)
Follow these steps to check the “V” seals:
1) Open the filter exchange mechanism as if to perform a filter ex­change. Remove any sample filter cassette installed.
2) Locate the two “V” seals, one mounted above and one mounted below the filter cassette. Examine each for damage. Wipe seals with a clean dry cloth, if required.
3) Replace damaged “V” seal, if necessary.
4) Perform a system leak test (Section 3.1.5).
SECTION 3: SERVICE PROCEDURES PAGE 3-6
Page 35
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure 3-7. Audit screen.
Audit Screen
Station: 1
Flow Rate: 0.0 l/min
Ambient Temp: 22.5 C
Ambient Pres: 0.987 atm
FTS Pres: 0.00 inH2O FTS Flow: 0.0 l/min
Edit Hub1 Sat2 Sat3 Pump >
Soft Function Keys in View Data Mode
Edit Hub1 Sat2 Sat3 Pump > Edit Sat4 Pump <
Soft Function Keys in Edit Mode
-- - + ++ Bksp >
+/- <
3.1.5. LEAK CHECK
Tools Needed: None Materials: Flow audit adapter (R&P P/N 57-000618)
New 47mm filter
Unshifted
Shifted
Unshifted
Shifted
Follow these steps to perform a leak test on the hub and satellite units:
1) Install a new 47mm filter into the sampling station (hub or satellite) being tested.
2) From the Partisol Main screen, press <F5:Setup> to enter the Setup screen. Press <F5:Audit> from the Setup screen to enter the Audit screen (Figure 3-7).
3) Remove the size-selective inlet from the sampling station being checked. Refer to the Operating Manual for the maintenance procedure for the PM-10 inlet, if needed.
4) Install the supplied flow audit adapter on the end of the sample tube of the sampling station being checked. Ensure that the flow audit adapter’s valve is in the open position (Figure 3-8).
5) Press either <F2:Hub1>, <F3:Sat2>, <F4:Sat3>, or <F9:Sat4> (<Shift><F4>) depending on which sampling station is currently being checked.
SECTION 3: SERVICE PROCEDURES PAGE 3-7
Page 36
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-8. Flow audit adapter installed.
Revision B.000
6) Turn on the pump by pressing <F5:Pump> or <F10:Pump> (<Shift><F5>).
7) Close the valve on the flow audit adapter.
8) Open the mass flow controller access door on the pump compart­ment cover (Figure 3-1).
9) Close the flow of the flow controller assembly by turning the manual shut off valve that is attached to the large air filter on the left side of the mass flow controller assembly (labeled “A” in Figure 3-9).
10) Record the reading on the vacuum gauge on the mass flow con­troller assembly.
11) Shut off the flow to the pump by turning the other manual shut off valve (the “pump” valve), which is located on the bottom of the mass flow controller assembly (labeled “B” in Figure 3-9).
12) Record the reading on the vacuum gauge 10 seconds after the pump valve is closed. This reading should not drop below half of the original reading (recorded in step 10) during this 10 second period. If the reading does drop below half of the original read­ing, go to Section 2.
13) Slowly open the two valves on the mass flow controller that were closed.
SECTION 3: SERVICE PROCEDURES PAGE 3-8
Page 37
Service Manual, Partisol Model 2000 Air Sampler
A
B
Vacuum Gauge
Figure 3-9. Mass flow controller.
Revision B.000
14) Slowly open the valve of the flow audit adapter assembly and remove it from the sampling station being checked. Replace the size-selective inlet.
15) Remove the 47mm filter from the sampling station being checked and discard it.
16) Perform steps 1-14 for each sampling station.
17) When all sampling stations have been checked, close the mass flow controller access door.
3.1.6. TESTING THE BATTERIES
There are two batteries used in the Partisol Sampler: a round coin cell battery and a battery-backed RAM socket battery. Both batteries are located on the CPU board.
3.1.6.1. TESTING THE ROUND COIN CELL BATTERY (BT1)
The round coin cell battery has an approximate 6-month storage capacity when the unit is unpowered and stored at a temperature of 20° C.
SECTION 3: SERVICE PROCEDURES PAGE 3-9
Page 38
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure 3-10. Setup screen.
Setup
Stations: 3 Std Temp: 25.0 Run Type: BASIC Std Pres: 1.000 Set Flow: 16.7 Ave Temp: 99.9 A/O Type: STAT Ave Pres: 9.999 Curr Date: 03-Dec-99 Basic Cyc: 0 Curr Time: 16:21:36 Intv Mode: SAMP
Edit Calib Diag RS232 Audit >
Soft Function Keys in View Data Mode
Edit Calib Diag RS232 Audit >
LCD + LCD - <
Soft Function Keys in Edit Mode
-- - + ++ Bksp >
+/- <
Follow these steps to check the round battery:
1) From the Partisol Main screen, press <F5:Setup> to enter the Setup screen (Figure 3-10).
Unshifted
Shifted
Unshifted
Shifted
2) Press <F1:Edit> to enter the Edit Mode.
3) Change the time or date to a new value and press <Enter>.
4) Turn off the instrument, wait approximately 20 seconds and turn the power back on.
5) From the Partisol Main screen, press <F5:Setup> to enter the Setup screen.
6) If the time or date is the same value as entered in step 3, the battery is functioning properly. If the time or date shows a differ­ent value than the value entered in step 3, the battery must be replaced. Contact R&P.
7) Restore the time or date to the correct value.
3.1.6.2. TESTING THE BATTERY-BACKED RAM SOCKET BATTERY
The battery-backed RAM socket battery has an approximate 2-1/2 year storage capacity if the unit is unpowered and stored at temperatures between 0° and 40° C.
SECTION 3: SERVICE PROCEDURES PAGE 3-10
Page 39
Service Manual, Partisol Model 2000 Air Sampler
Follow these steps to check the socket battery:
1) From the Partisol Main screen, press <F3:Storage> to enter the Filter Data screen.
2) Scan the records on the Filter Data and Interval Data screens for random characters or incorrect data. If any discrepancies are found, contact R&P.
3.2. CALIBRATION PROCEDURES
This section contains calibration procedures for the Partisol Sampler. Calibrations should be performed according to the table on page 3-1.
IMPORTANT:
R&P strongly advises that these calibrations be performed in the order presented.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
3.2.1. INTERFACE BOARD CALIBRATION
The interface board is located on the bottom panel of the electronics compartment (the area behind the keypad in the hub unit). See Figure 3-11 for interface board test point locations.
Tools Needed: Phillips screwdriver
Multimeter with four-digit resolution
Materials: None
Follow these steps to calibrate the interface board:
1) Turn off the sampler.
2) Remove the three screws and open the electronics compartment door (Figure 3-1).
3) Locate the test point labeled “+6V” (red) near the back of the interface board (Figure 3-11). Place the positive lead of a multim­eter on the +6V test point.
4) Locate the ground test point labelled “PGND” (black) to the left of the +6V test point (Figure 3-11). Place the ground lead on the PGND test point.
SECTION 3: SERVICE PROCEDURES PAGE 3-11
Page 40
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-11. Interface board test point locations.
Revision B.000
SECTION 3: SERVICE PROCEDURES PAGE 3-12
Page 41
Service Manual, Partisol Model 2000 Air Sampler
5) Ensure that the sampler is turned on and that the display backlight is on. The backlight must be on during the +6V calibration. If the backlight is off, press any key on the keypad to turn it on.
6) Locate the R21 pot on the rear section of the interface board. It is located tp the right of the +6V test point. R21 is a blue vertical pot, and its designation, “R21,” is silk-screened on the board.
7) Adjust the R21 pot with a potentiometer adjustment tool until the multimeter reads 6.00 VDC ± 0.05 V.
8) Locate the test point labeled “+10V” (red) in the center of the interface board. Place the positive lead of the multimeter on the +10V test point.
9) Locate the ground test point labelled “AGND” (black) near the front right corner of the interface board. Place the ground lead on the AGND test point.
Revision B.000
10) Locate the R44 pot in the center of the interface board. R44 is a blue vertical pot, and its designation, “R44,” is silk-screened on the board.
11) Adjust the R44 pot with a potentiometer adjustment tool until the multimeter reads 10.000 VDC ± 0.002 V.
12) Locate the test point labelled “+2.5V” (red) near the right side of the interface board. Place the positive lead of the multimeter on the +2.5V test point.
13) Locate the test point labelled “PWM0” (green) to the right of the +2.5V test point. Place the ground lead on the PWM0 test point.
14) Locate the R53 pot below the PWM0 test point on the interface board. R53 is a blue vertical pot, and its designation, “R53,” is silk-screened on the board.
15) Adjust the R53 pot with a potentiometer adjustment tool until the multimeter reads 0.000 VDC ± 0.002 V.
16) Go to Section 3.2.2.
SECTION 3: SERVICE PROCEDURES PAGE 3-13
Page 42
Service Manual, Partisol Model 2000 Air Sampler
3.2.2. ANALOG INPUT CALIBRATION
4
Always wear appropriate antistatic devices when working with the system’s electronics.
Please note that the following procedure must be performed after the interface board calibration and before the temperature, pressure and flow calibrations.
Tools Needed: Multimeter with four-digit resolution Materials: Analog calibration cable (R&P P/N 51-002604)
Follow these steps to calibrate the analog input:
1) If the Main screen is not currently displayed, press <Esc> until the Main screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an analog input calibration. If the hardware is currently in the Run Operating Mode, press <F4: Run/Stp> to enter the Stop Operating Mode. The current operating mode is displayed in the upper right-hand corner of the Main screen.
Revision B.000
Figure 3-12. Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib> to access the Calibration screen (Figure 3-13).
4) Plug the 6-pin end of the Analog Input Calibration Cable supplied with the sampler into the socket labelled “ANEMOMETER” on the back of the Partisol hub unit (Figure 3-14).
Basic Partisol Main Screen Stop
16:21 03 - Feb - 00 Curr:2 Status:OK
Stat Std Vol Valid Total Date H1: DONE 24.143 24:00 24:00 02-Feb S2: ON 16.342 16:21 16:21 03-Feb S3: OFF 0.000 0:00 0:00 04-Feb S4:
Edit Stats Storage Run/Stp Setup
Soft Function Keys in View Data Mode
Edit Stats Storage Run/Stp Setup
Soft Function Keys in Edit Mode
Daily Today+1 + ++ Bksp >
-- - + ++ Bksp <
Unshifted
Shifted
Unshifted
Shifted
SECTION 3: SERVICE PROCEDURES PAGE 3-14
Page 43
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure 3-13. Calibration screen.
Figure 3-14. Back of Partisol enclosure.
Basic Calibration Screen Stop
Calc Act Offset Span A/I: 1.003 0.000 0.0000 1.0000 Temp: 27.9 0.0 1.0000 Pres: 0.998 0.000 1.0000 Flow: 16.7 0.0 0.0000 1.0000 A/O: 1.00 Pot: 3.752 Flow Adj: Hold
Edit Hub1 Sat2 Sat3 Sat4 >
Soft Function Keys in View Data Mode
Edit Hub1 Sat2 Sat3 Sat4 >
FTS Pump DecrFlw HoldFlw IncrFlw <
Soft Function Keys in Edit Mode
-- - + ++ Bksp >
+/- <
Unshifted
Shifted
Unshifted
Shifted
5) Plug the 4-pin end of the Analog Input Calibration Cable into the socket labelled “USER OUTPUT” on the back of the Partisol hub unit.
6) Ensure that the electronics compartment door (Figure 3-1) is open.
SECTION 3: SERVICE PROCEDURES PAGE 3-15
Page 44
Service Manual, Partisol Model 2000 Air Sampler
7) Locate the test point labelled “PWM1” (green) on the middle-right side of the interface board (Figure 3-11). Attach the positive lead from the multimeter on the PWM1 test point.
8) Locate the ground test point labeled “AGND” (black) near the front right corner of the interface board. Place the ground lead on the AGND test point.
9) Using the arrow key, position the cursor so that it is in the location labelled “A/O” on the Calibration screen.
10) Press <F1:Edit> to enter the Edit Mode. Type in a number between
0.050 and 0.150 volts and press <ENTER>.
11) Observe the number displayed in the row labelled “A/I” in the column labelled “Calc.” Ensure that this number does not vary more than ±0.005 volts after watching it for 5 seconds. If this number is not stable, choose a new number for “A/O” between
0.050 and 0.150 volts.
Revision B.000
12) Read the voltage displayed on the multimeter.
13) Using the arrow keys, position the cursor so that it is in the row labelled “A/I” and the column labelled “Act” (actual).
14) Press <F1:Edit> to enter the Edit Mode. Type the voltage read from the multimeter (to a 3-digit accuracy; i.e., “0.xxx”) in this position and press <ENTER>. This calculates the “Offset.”
15) Ensure that the number now displayed in the row labelled “A/I” and the column labelled “Calc” matches the number displayed on the multimeter within ±0.005 volts.
16) Using the arrow key, position the cursor so that it is in the location labelled “A/O.”
17) Press <F1: Edit> to enter the Edit Mode. Type in a number be­tween 4.800 and 4.900 volts and press <ENTER>.
18) Observe the number displayed in the row labelled “A/I” in the column labelled “Calc.” Ensure that this number does not vary more than ±0.005 volts after watching it for 5 seconds. If this number is not stable, choose a new number for “A/O” between
4.800 and 4.900 volts.
19) Read the voltage displayed on the multimeter.
20) Using the arrow keys, position the cursor so that it is in the row labelled “A/I” and the column labelled “Act.”
SECTION 3: SERVICE PROCEDURES PAGE 3-16
Page 45
Service Manual, Partisol Model 2000 Air Sampler
21) Press <F1:Edit> to enter the Edit Mode. Type the voltage read from the multimeter (to a 3-digit accuracy; i.e., “0.xxx”) in this position and press <ENTER>. This calculates the “Span.”
22) Ensure that the number now displayed in the row labelled “A/I” and the column labelled “Calc” matches the number displayed on the multimeter within ±0.005 volts.
23) Remove the multimeter leads from the interface board and the Analog calibration cable from the back of the Partisol hub.
24) After the analog input calibration has been performed success­fully, the temperature, pressure and flow calibrations must be performed.
NOTE: If the instrument has been reset and you have recorded the value of “Offset” and “Span” for the flow, you may enter these numbers directly when in the Edit Mode.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
25) Close the electronics compartment door and secure with the three screws previously removed.
26) Go to Section 3.2.3.
3.2.3. AMBIENT AIR TEMPERATURE CALIBRATION
Tools Needed: Temperature measurement device Materials: None
Follow these steps to calibrate the ambient air temperature:
1) If the Main screen is not currently displayed, press <Esc> until the screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an analog input calibration. If the hardware is currently in the Run Operating Mode, press <F4: Run/Stp> to enter the Stop Operating Mode. The current operating mode is displayed in the upper right-hand corner of the Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib> to access the Calibration screen (Figure 3-13).
4) Determine the current temperature (°C) at the ambient temperature sensor mounted on the sample tube of the hub unit using an external thermometer, [°C = 5/9 x (°F - 32)].
SECTION 3: SERVICE PROCEDURES PAGE 3-17
Page 46
Service Manual, Partisol Model 2000 Air Sampler
5) Press <F1: Edit> to enter the Edit Mode, and move the cursor to the “Act” (actual) column in the row labelled “Temp.”
6) Enter the current ambient temperature (°C), using the <F6: +/-> key to enter negative temperatures if necessary. Press <ENTER> to leave the Edit Mode. Check that the temperature shown in the “Calc” column now matches the current temperature within ±2° C.
7) Upon receiving the actual temperature, the system’s microproces­sor automatically computes “Span” for the ambient temperature. Note this number for future reference.
NOTE: If the instrument has been reset and you have recorded the value of “Span” for the ambient temperature, you may enter it directly in the “Span” column when in the Edit Mode.
8) Go to Section 3.2.4.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
3.2.4. AMBIENT PRESSURE CALIBRATION
Tools Needed: None Materials: None
Follow these steps to calibrate the ambient pressure:
1) If the Main screen is not currently displayed, press <Esc> until the screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an ambient pressure calibration. If the hardware is currently in the Run Operating Mode, press <F4: Run/Stp> to enter the Stop Operating Mode. The current operating mode is displayed in the upper right-hand corner of the Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib> to access the Calibration screen (Figure 3-13).
4) Determine the current ambient station pressure in Atmospheres (absolute pressure, not corrected to sea level).
• To convert from mm Hg @ 0 °C to Atmospheres, multiply by 0.001316.
• To convert from millibars to Atmospheres, multiply by 0.000987.
• To convert from inches Hg @ 32 °F to Atmospheres, multiply by 0.03342.
SECTION 3: SERVICE PROCEDURES PAGE 3-18
Page 47
Service Manual, Partisol Model 2000 Air Sampler
5) Press <F1: Edit> to enter the Edit Mode, and move the cursor to the “Act” (actual) column in the row labelled “Pres.”
6) Enter the current ambient pressure (Atmospheres) and press <ENTER> to leave the Edit Mode. Check that the pressure shown in the “Calc” column now matches the current temperature within ± 0.02 atm.
7) Upon receiving the actual pressure, the system’s microprocessor automatically computes “Span” for the ambient pressure. Note this number for future reference.
NOTE: If the instrument has been reset and you have recorded the value of “Span” for the ambient pressure, you may enter it directly in the “Span” column when in the Edit Mode.
8) Go to Section 3.2.5.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
3.2.5. MASS FLOW CONTROLLER CALIBRATION
The temperature and pressure calibrations described above must be performed before the mass flow controller calibration is undertaken. In addition, the leak check discussed in Section 3.1.5 must also be undertaken before the following instructions are executed.
Tools Needed: Flow measuring device such as the
Streamline FTS (R&P P/N 57-004506-0001)
Materials: Flow audit adapter (if required) (R&P P/N 57-000618)
New 47mm filter
Follow these steps to calibrate the mass flow controller:
1) If the Main screen is not currently displayed, press <Esc> until the screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an analog input calibration. If the hardware is currently in the Run Operating Mode, press <F4: Run/Stp> to enter the Stop Operating Mode. The current operating mode is displayed in the upper right-hand corner of the Main screen.
3) Remove the size-selective inlet from the hub.
4) Install a filter cassette containing a new 47 mm filter into the filter holder of the hub unit. This filter will be thrown away at the end of this flow calibration.
SECTION 3: SERVICE PROCEDURES PAGE 3-19
Page 48
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure 3-15. FTS Flow Calibration screen.
Basic FTS Flow Calibration Stop
Cur Flow: 0.1 l/min Act Flow: 0.1 l/min Valve FTS Pres: 0.00 in/H2O
Flow Adj: Hold Offset: -0.0156 FTS Const m: Span: 1.0096 FTS Const b:
Edit Hub1 Sat2 Sat3 Sat4 >
Soft Function Keys in View Data Mode
Edit Hub1 Sat2 Sat3 Sat4 >
Calib Pump DecrFlw HoldFlw IncrFlw <
Soft Function Keys in Edit Mode
-- - + ++ Bksp >
+/- <
5) While in the Main screen, press <F5: Setup> and then <F2: Calib> to enter the Calibration screen. From the Calibration screen, press <F6: FTS> (<SHIFT> <F1>) to enter the FTS Flow Calibration screen (Figure 3-15).
Unshifted
Shifted
Unshifted
Shifted
6) If you are using the Streamline FTS Flow Transfer Standard to calibrate the flow of the Partisol sampler, enter the FTS calibra­tion constants “m” and “b” on the FTS Flow Calibration screen. Execute the following steps to enter these constants:
• Press <F1:Edit>
• Scroll down to the “FTS Const m” field and enter the m constant from the FTS
calibration sheet
• Scroll down to the “FTS Const b” field and enter the b constant from the FTS
calibration sheet
• Press <Enter>
7) Attach the Streamline FTS Flow Transfer Standard to the hub’s sample tube (Figure 3-16). Other flow meters may require the use of the flow audit adapter to attach the meter to the sample tube. The flow audit adapter should be installed with its valve open (Figure 3-8).
8) Do not turn the pump on. If it is currently on, press <F7: Pump> (<SHIFT> <F2>) to turn it off.
SECTION 3: SERVICE PROCEDURES PAGE 3-20
Page 49
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-16. Streamline FTS Flow Transfer Standard attached to the Partisol.
Revision B.000
9) Press <F1: Edit> to enter the Edit Mode, and move the cursor to the row labelled “Act Flow.”
10) Enter a zero (“0”) in this field and press <ENTER> to leave the Edit Mode. This causes the microprocessor to compute the “Offset,” which is the zero offset for the mass flow sensor. Record this number for future reference.
11) Turn on the pump by pressing <F7: Pump> (<SHIFT> <F2>), and then press <F2: Hub1> to allow the sample flow to pass through the sample tube of the hub. The valve column should display “H1.”
12) Use the <F8: DecrFlw> (<SHIFT> <F3>) (continuously decrease flow), <F9: HoldFlw> (<SHIFT> <F4>) (maintain current flow rate), and <F10: IncrFlw> (<SHIFT> <F5>) (continuously increase flow) keys to control the flow rate. The flow rate displayed in the row labeled “Cur Flow” should be approximately 16.7 l/min.
NOTE: Pressing <F8:DecrFlw> (or <F10:IncrFlw>) will cause the flow to continuously decrease (or increase) until <F9:HoldFlw> is pressed. If the flow is not stopped, it will decrease (or increase) to a maximum point and will begin to increase (or decrease) because of the circular gear used on the flow controller.
13) Determine the flow in actual l/min by using the external flow meter. If you are using the Streamline FTS Flow Transfer Standard, enter the pressure (inches H2O) in the FTS Pres field. Press <F1:Edit>,
SECTION 3: SERVICE PROCEDURES PAGE 3-21
Page 50
Service Manual, Partisol Model 2000 Air Sampler
enter the value in the FTS Pres field using the keypad, and then press <Enter>. The sampler will calculate the volumetric flow in l/ min.
14) If you are using the Streamline FTS Flow Transfer Standard, the sampler will automatically enter the calculated flow into the “ACT FLOW” field and compute “Span,” which is the span slope for the mass flow sensor. Record this number for future reference.
15) If you are using a different flow measurement device, you must manually enter the volumetric flow in the “ACT FLOW” field by pressing <F1:Edit> to enter the Edit Mode, moving the cursor to the “ACT FLOW” field and entering the flow determined by the external flow meter, followed by <Enter>. This causes the micro­processor to compute “Span,” which is the span slope for the mass flow sensor. Record this number for future reference.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
NOTE: If the instrument has been reset and you have recorded the value of “Offset” and “Span” for the flow, you may enter these numbers directly when in the Edit Mode.
15) Return to the Main screen by pressing <ESC> three times.
16) Restore the sampling hardware to its original state by removing the flow metering hardware and re-installing the size-selective inlet on the sample tube of the hub. Remove the filter cassette from the hub unit, and discard the filter installed in it.
3.3. REMOVE AND REPLACE PROCEDURES
3.3.1. REMOVING/REPLACING THE DISPLAY
Tools Needed: Phillips screwdriver
Slotted screwdriver
Materials: Display (R&P P/N 10-002254)
Static-shielded bag
Follow these steps to remove the display from the sampler:
1) Turn off the instrument and disconnect the power cord from its AC source.
2) Remove the three screws and open the electronics compartment door (Figure 3-1).
SECTION 3: SERVICE PROCEDURES PAGE 3-22
Page 51
Service Manual, Partisol Model 2000 Air Sampler
Display cover
Contrast
knob
RS232 connector
Circuit breaker
Ground wire
Figure 3-17. Back of electronics compartment door.
Revision B.000
3) Examine the back of the electronics compartment door. Remove the screw securing the ground wire to the corner of the display cover (Figure 3-17).
4) Remove the remaining screws that secure the display cover.
5) Note the orientation of the display as it is mounted on the back of the electronics compartment door (the display connector is on the hinge side of the door). You will need this information to install a new display.
6) Disconnect the cable connecting the contrast knob to the display.
7) Disconnect the cable connecting P16A of the interface board and the display at the display connector.
8) Disconnect the cable connecting P16B of the interface board and the display at the interface board connector.
9) Remove the four round standoffs from the corners of the display.
10) Place display in a static-shielded bag.
SECTION 3: SERVICE PROCEDURES PAGE 3-23
Page 52
Service Manual, Partisol Model 2000 Air Sampler
Follow these steps to install the display into the sampler:
1) Mount the display into position on the electronics compartment door. The ribbon cable connector should be located on the hinge side of the door.
2) Replace the four round standoffs onto the corners of the display.
3) Connect the cable connecting P16B of the interface board and the display.
4) Connect the cable connecting P16A of the interface board and the display.
5) Connect the cable connecting the contrast knob to the display (Figure 3-17).
6) Replace the display cover onto the standoffs and secure with three of the screws, leaving the lower left corner loose.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
7) Place the ground wire on the loose corner of the display cover and secure with the remaining screw.
8) Close and secure the electronics compartment door.
9) Connect the unit’s power cord to an AC source.
3.3.2. R
EMOVING/REPLACING THE FAN
Tools Needed: Slotted screwdriver
Phillips screwdriver
Materials: Fan (R&P P/N 10-002251)
Follow these steps to remove the fan from the sampler:
1) Turn off the instrument and disconnect the power cord from the AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Remove the fan cover, screen and filter from the outside of the fan (on the outside of the enclosure).
4) Disconnect the fan connector (P9) from the underside of the interface board.
3) Remove the four screws which secure the fan to the enclosure and remove the fan assembly.
SECTION 3: SERVICE PROCEDURES PAGE 3-24
Page 53
Service Manual, Partisol Model 2000 Air Sampler
Follow these steps to install the fan into the sampler:
1) Hold the replacement fan and outside guard mounting frame in the proper position on the enclosure. The fan should be positioned so that the air is blown INTO the enclosure. There is an arrow on the side of the fan that indicates the air flow direction.
2) Secure the fan using the four screws previously removed.
3) Replace the screen, filter and fan cover by pressing it onto the fan mounting frame from the outside of the enclosure.
4) Connect the fan connector (P9) to the underside of the interface board.
5) Replace the pump compartment cover.
6) Connect the power cord to the AC source and resume normal operation.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
3.3.3. REMOVING/REPLACING THE MASS FLOW CONTROLLER ASSY
Tools Needed: Adjustable wrench
Phillips screwdriver Pliers
Materials: Mass flow controller assembly
(R&P P/N 55-002663)
Follow these steps to remove the mass flow controller assembly from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from its AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Locate the hose that attaches the mass flow sensor to the large in­line filter. Using a pair of pliers, squeeze the clip that secures the hose to the mass flow sensor and remove the hose (Figure 3-3).
4) Locate the hose attached to the manual valve on the bottom of the assembly. Examine the hose clip that secures the hose to the valve.
5) Using a pair of pliers, squeeze the clip and remove the hose from valve.
SECTION 3: SERVICE PROCEDURES PAGE 3-25
Page 54
Service Manual, Partisol Model 2000 Air Sampler
6) Locate the wires that are attached to the mass flow controller motor. Follow the wires until you find a Molex connector. Unplug this connector.
7) Disconnect the mass flow sensor connector (P21A) from the interface board.
8) Remove the four nuts that secure the mass flow controller assem­bly to the enclosure. Remove the assembly.
Follow these steps to install the mass flow controller assembly into the sampler:
1) Place the mass flow controller assembly in its proper position in the enclosure. Ensure that the assembly is oriented so that the vacuum gauge is visible.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
2) Replace the four nuts that secure the mass flow controller assem­bly to the enclosure.
3) Reconnect the mass flow sensor connector (P21A) on the inter­face board.
4) Reconnect the mass flow controller motor connector.
5) Using a pair of pliers, squeeze the clip that secures the hose to the mass flow sensor and replace the hose onto the mass flow sen­sor fitting.
6) Using a pair of pliers, squeeze the clip that secures the hose to the manual valve and replace the hose onto the valve.
7) Perform a system leak test (Section 3.1.5).
8) Perform a mass flow controller calibration (Section 3.2.5).
3.3.4. REMOVING/REPLACING THE ELECTRONICS HEATER/FAN ASSY
Tools Needed: Phillips screwdriver Materials: Electronics heater/fan assy
(R&P P/N 56-005909-0120 or 56-005909-0240 )
Follow these steps to remove the electronics compartment heater/fan assembly from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from its AC source.
SECTION 3: SERVICE PROCEDURES PAGE 3-26
Page 55
Service Manual, Partisol Model 2000 Air Sampler
2) Remove the three screws and open the electronics compartment door (Figure 3-1).
3) Locate the electronics compartment heater/fan assembly and note its orientation. You will need this information later when install­ing a new heater/fan assembly.
4) Disconnect the cable connecting the electronics heater to the interface board (P11).
5) Disconnect the cable connecting the electronics fan to the inter­face board (P101).
6) While supporting the assembly, remove the two screws that se­cure the heater/fan assembly to the back of the electronics compartment.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
Follow these steps to install the electronics compartment heater/fan assembly into the sampler:
1) While supporting the assembly in the proper position on the back of the electronics compartment, secure the two screws. Ensure that the orientation is correct (the fan should be on top).
2) Reconnect the cable connecting the electronics heater to the interface board (P11).
3) Reconnect the cable connecting the electronics fan to the inter­face board (P101).
4) Close the electronics compartment door and secure it with its three screws.
5) Connect the unit’s power cord to its AC source and resume nor­mal operation.
3.3.5. REMOVING/REPLACING THE INTERFACE/CPU BOARD ASSY
Inside the electronics compartment of the Partisol Sampler, the interface board is mounted on top of the CPU board. These boards are attached so that they can be installed and removed as a unit. The user should not attempt to separate these boards. The boards should be returned to R&P as a set if service or replacement is required.
SECTION 3: SERVICE PROCEDURES PAGE 3-27
Page 56
Service Manual, Partisol Model 2000 Air Sampler
Tools Needed: Phillips screwdriver
Materials: CPU board (R&P P/N 50-002407)
Follow these steps to remove the interface/CPU board assembly from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from its AC source.
2) Remove the three screws and open the electronics compartment door (Figure 3-1).
Revision B.000
Slotted screwdriver Labeling materials
Interface board (R&P P/N 50-005496) Static-shielded bag
3) Remove all wires and cables connected to the interface and CPU boards.
NOTE: While removing cables, be sure to label each connector with its location for easy re-assembly.
4) Remove the 1/8-inch silicone rubber hose from the pressure transducer on the right side of the interface board.
5) Locate the screws that secure the assembly to the floor of the electronics compartment. Remove the screws.
6) Lift up the assembly approximately 2 inches from the floor of the electronics compartment and remove it from the compartment.
7) Place the boards assembly in a static-shielded bag.
Follow these steps to install the interface/CPU board assembly into the sampler:
1) Position the board assembly over the standoffs on the floor of the electronics compartment.
2) Replace the screws and secure the assembly to the floor of the electronics compartment.
3) Replace the 1/8-inch silicone rubber hose from the pressure transducer on the right side of the interface board.
4) Replace all wires and cables connected to the CPU and interface boards. Be sure to follow the location labels that you made earlier when you removed the cables.
SECTION 3: SERVICE PROCEDURES PAGE 3-28
Page 57
Service Manual, Partisol Model 2000 Air Sampler
5) Close the electronics compartment door and secure it with the tree screws.
6) Connect the unit’s power cord to an AC source.
7) Perform the calibration procedures listed in Section 3.2.
3.3.6. REMOVING/REPLACING THE KEYPAD
The keypad of the Partisol Sampler is integrated into the door to the electronics compartment. Therefore, to replace the keypad, the entire electronics compartment door must be replaced.
4
Always wear appropriate antistatic devices when working with the system’s electronics.
Tools Needed: Adjustable wrench
Materials: Keypad (R&P P/N 55-002345)
Revision B.000
Phillips screwdriver
Static-shielded bag
Follow these steps to remove the keypad from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from its AC source.
2) Remove the three screws and open the electronics compartment door (Figure 3-1).
3) Remove the display from the electronics compartment door (Sec­tion 3.3.1).
4) Remove the two screws that secure the RS232 connector to the electronics compartment door (Figure 3-17). Remove the connec­tor.
5) Remove the four wires attached to the inside of the circuit breaker/ power switch by pulling them straight off.
6) Depress the tabs on both sides of the circuit breaker/power switch and push through the hole in the electronics compartment door.
7) Remove the contrast adjustment knob from the electronics com­partment door.
8) Disconnect the keypad cable from the mating cable on the right side wall of the electronics compartment.
9) Remove the three nuts securing the electronics compartment door to the enclosure.
SECTION 3: SERVICE PROCEDURES PAGE 3-29
Page 58
Service Manual, Partisol Model 2000 Air Sampler
10) Place the defective keypad into a static-shielded bag for return to R&P.
Follow these steps to install the keypad into the sampler:
1) Place the electronics compartment door in its proper position on the enclosure. Tighten the three nuts that secure the door to the enclosure.
2) Connect the keypad cable to the mating cable on the right side wall of the electronics compartment.
3) Replace the contrast adjustment knob onto the electronics com­partment door.
4) Push the circuit breaker/power switch through the hole in the electronics compartment door.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
5) Replace the four wires onto the circuit breaker/power switch.
6) Replace the two screws that secure the RS232 connector to the electronics compartment door.
7) Replace the display onto the electronics compartment door (Sec­tion 3.3.1).
8) Close the electronics compartment door and secure.
9) Connect the unit’s power cord to an AC source and resume nor­mal operation.
3.3.7. REMOVING/REPLACING THE PUMP
Tools Needed: Adjustable wrench
Phillips screwdriver
Materials: Pump
(R&P P/N 54-005949-0120 or 54-005949-0240)
Follow these steps to remove the pump from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from its AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Remove the black, pneumatic tubing from the fittings of the pump. Depress the collar on the fitting surrounding each tube with a flat
SECTION 3: SERVICE PROCEDURES PAGE 3-30
Page 59
Service Manual, Partisol Model 2000 Air Sampler
blade screwdriver and then pull the tubing straight out of its fitting (Figure 3-18).
4) Remove the pump’s ground wire from the left side of the enclo­sure.
5) Disconnect the cable connecting the pump to the interface board (P10). The cable is connected on the underside of the interface board.
6) Remove the four nuts securing the pump to the floor of the pump compartment. You will need to lift or tip the unit sideways to access these nuts.
7) Lift the pump and remove it from the sampler.
Follow these steps to install the pump into the sampler:
Revision B.000
Figure 3-18. Pump com­partment.
1) Place the pump into the enclosure in its proper position and secure the pump with the four nuts. You will need to lift or tip the unit sideways to access these nuts.
2) Reconnect the cable connecting the pump to the interface board (P10). The cable is connected on the underside of the interface board.
3) Replace the pump’s ground wire to the left side of the enclosure.
4) Reinsert the black tubing into the pump fittings. Press the tubing firmly into each fitting to insure proper seating.
SECTION 3: SERVICE PROCEDURES PAGE 3-31
Page 60
Service Manual, Partisol Model 2000 Air Sampler
5) Replace the pump compartment cover.
6) Connect the unit’s power cord to an AC source and resume nor­mal operation.
Revision B.000
4
Always wear appropriate antistatic devices when working with the system’s electronics.
3.3.8. R
EMOVING/REPLACING A VALVE
Tools Needed: Flat screwdriver
Phillips screwdriver
Materials: Valve (R&P P/N 59-002456)
Follow these steps to remove the valve from the sampler:
1) Turn the instrument off and disconnect the unit’s power cord from its AC source.
2) Remove the pump compartment cover (Figure 3-1).
3) Locate the valve manifold below the electronics compartment and determine which valve is being replaced (Figure 3-19).
4) Disconnect the cable connecting the valve to the interface board (P5, P6, P7 or P8). The cable is connected on the underside of the interface board.
5) Remove the upper left and lower right screws from the valve. These are the screws that secure the valve to the manifold.
6) Remove the valve from the enclosure.
Follow these steps to install the valve into the sampler:
1) Place the valve onto the manifold in its proper orientation (the valve plug should be toward the pump). Tighten the upper left and lower right screws.
2) Reconnect the cable connecting the valve to the interface board (P5, P6, P7 or P8). The cable is connected on the underside of the interface board.
3) Replace the pump compartment cover.
3) Connect the unit’s power cord to its AC source and resume nor­mal operation.
SECTION 3: SERVICE PROCEDURES PAGE 3-32
Page 61
Service Manual, Partisol Model 2000 Air Sampler
Figure 3-19. Valve mani­fold.
Revision B.000
SECTION 3: SERVICE PROCEDURES PAGE 3-33
Page 62
This page left intentionally blank.
Page 63
Service Manual, Partisol Model 2000 Air Sampler
Section 4: Electrical Schematics
This section contains schematic diagrams and interconnect diagrams of the major electronic components in the Partisol Model 2000 Sampler. Schematic diagrams for the industrial microcomputer system are not included here, because they are propri­etary to the manufacturers of these boards.
4.1. ELECTRICAL SCHEMATICS
Schematic Schematic Schematic Description Part Number Revision
Interface board 81-005496 B Interconnect diagrams 81-005912 C
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-1
Page 64
Service Manual, Partisol Model 2000 Air Sampler
4.1.1. INTERFACE BOARD
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-2
Page 65
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-3
Page 66
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-4
Page 67
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-5
Page 68
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-6
Page 69
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-7
Page 70
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-8
Page 71
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-9
Page 72
Service Manual, Partisol Model 2000 Air Sampler
4.1.2. INTERCONNECT DIAGRAMS
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-10
Page 73
Service Manual, Partisol Model 2000 Air Sampler
3-9-99
5-14-99
8-2-99
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-11
Page 74
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-12
Page 75
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-13
Page 76
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-14
Page 77
Service Manual, Partisol Model 2000 Air Sampler
TWISTED
PAIRS
Revision B.000
10 OHM 1/2W RESISTOR
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-15
Page 78
Service Manual, Partisol Model 2000 Air Sampler
5-14-99
8-2-99
Revision B.000
TO GROUND STUD
TOP VIEW
TOP VIEW
OPTIONAL
TOP VIEW
TOP VIEW
TWISTED
PAIRS
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-16
Page 79
Service Manual, Partisol Model 2000 Air Sampler
8-2-99
8-2-99
TWISTED
PAIRS
Revision B.000
SECTION 4: ELECTRICAL SCHEMATICS PAGE 4-17
Page 80
This page left intentionally blank.
Page 81
Service Manual, Partisol Model 2000 Air Sampler
Section 5: R&P Return Policy
Refer to this section when shipping items to R&P. Always contact the R&P Service Department before returning any unit components.
Service Department, R&P Co. Phone: 518-452-0065 Fax: 518-452-0067 Email: SUPPORT@RPCO.COM
5.1. R&P RETURN POLICY — INSTRUMENT COMPONENTS
There are three types of returns:
· Warranty return for exchange
· Billable return for exchange
· Billable return for repair.
Revision B.000
Warranty Return for Exchange If an instrument component becomes defective during the warranty period, follow this procedure:
1. Contact the R&P Service Department. Have the following information available:
· System serial number
· Nature of the problem
· Troubleshooting steps taken and results.
2. R&P service personnel will determine whether a component exchange is necessary and whether the component is covered under warranty.
3. If R&P service personnel determine that a component exchange is necessary and that it is covered under warranty, you will be issued a return number.
4. A replacement component will be shipped as soon as possible.
5. Once you have received the replacement component and installed it into your unit, you should ship the defective component back to R&P in the same box that the replacement component arrived in, according to the instructions in Section 5.3 or 5.4. Be sure to write the return number on the box or in the return papers.
Billable Return for Exchange If an instrument component becomes defective after the warranty period, follow this procedure:
1. Contact the R&P Service Department. Have the following information available:
· System serial number
· Nature of the problem
· Troubleshooting steps taken and results.
SECTION 5: R&P RETURN POLICY PAGE 5-1
Page 82
Service Manual, Partisol Model 2000 Air Sampler
2. R&P service personnel will determine whether a component exchange is necessary and whether the component is covered under warranty.
3. If the component exchange is billable, R&P service personnel will inform you of the exchange price.
4. You will be issued a return number.
5. As soon as R&P receives your purchase order and appropriate invoice, the company will ship a replacement component to you.
6. Once you have received the replacement component and installed it into your unit, you should ship the defective component back to R&P in the same box that the replacement component arrived in, according to the instructions in Section 5.3 or 5.4. Be sure to write the return number on the box or in the return papers.
Billable Return for Repair If an instrument component becomes defective after the warranty period, the compo­nent can be returned to R&P for repair. This procedure is generally less expensive than a billable return for exchange but can result in a longer downtime for the instrument. If you would like to return your instrument component for repair, follow this procedure:
1. Contact the R&P Service Department. Have the following information available:
· System serial number
· Nature of the problem
· Troubleshooting steps taken and results.
2. R&P service personnel will determine whether the component is defec­tive and whether the component is covered under warranty.
3. If R&P service personnel determine that the component is billable and you wish to return it for repair, they will inform you of the exchange price. The billable repair will not exceed this price.
4. You will be issued a return number.
5. Ship the defective component back to R&P, according to the instructions in Section 5.3 or 5.4. Be sure to write the return number on the box or in the return papers. Please include the name and phone number of the person who should be contacted with the final repair price.
6. R&P service personnel will call this person to tell them the final repair price, once the repair is complete.
7. As soon as R&P receives your purchase order for the repair, the company will ship your repaired component to you.
Revision B.000
SECTION 5: R&P RETURN POLICY PAGE 5-2
Page 83
Service Manual, Partisol Model 2000 Air Sampler
5.2. R&P RETURN POLICY — SYSTEMS
There are two types of returns:
· Warranty return for evaluation/repair
· Billable return for evaluation/repair.
Warranty Return for Evaluation/Repair If an instrument becomes defective during the warranty period, follow this procedure:
1. Contact the R&P Service Department. Have the following information available:
· System serial number
· Nature of the problem
· Troubleshooting steps taken and results.
2. R&P service personnel will determine whether it is necessary to ship the system to R&P and whether the system is covered under warranty.
3. If R&P service personnel determine that it is necessary to ship the system to R&P and that the system is covered under warranty, you will be issued a return number.
4. R&P service personnel will specify which items should be returned. These items should be returned to R&P according to the instructions in Section 5.3 or 5.4. NOTE: Send the system back to R&P in the original shipping boxes. If you discarded the original shipping boxes and the system becomes damaged during a return shipment to R&P, the warranty could be voided.
5. Be sure to write the return number on the box or in the return papers.
Revision B.000
Billable Return for Evaluation/Repair If an instrument becomes defective after the warranty period, follow this procedure:
1. Contact the R&P Service Department. Have the following information available:
· System serial number
· Nature of the problem
· Troubleshooting steps taken and results.
2. R&P service personnel will determine whether it is necessary to ship the system to R&P and whether the system is covered under warranty.
3. If R&P service personnel determine that it is necessary to ship the system to R&P, you will be issued a return number.
4. R&P service personnel will specify which items should be returned. These items should be returned to R&P according to the instructions in Section 5.3 or 5.4. NOTE: Send the system back to R&P in the original shipping boxes. If you discarded the original shipping boxes and the system becomes damaged during a return shipment to R&P, additional costs could be levied.
SECTION 5: R&P RETURN POLICY PAGE 5-3
Page 84
Service Manual, Partisol Model 2000 Air Sampler
5. Be sure to write the return number on the box or in the return papers. Please include the name and phone number of the person who should be contacted with the final repair price.
6. When R&P service personnel receive your instrument, they will repair and calibrate the unit. They will determine a final repair price and contact you with this information.
7. As soon as R&P receives your purchase order for the repair, R&P service personnel will return your system.
5.3. CONTINENTAL US SHIPPING INSTRUCTIONS
The following instructions explain how to return merchandise to Rupprecht & Patashnick Co., Inc. when shipping within the Continental US. It is important that R&P receives and repairs defective exchanged items in a timely manner.
Revision B.000
NOTE: You are responsible for the shipping costs of returned goods.
Please read this before returning any material:
1. Do not ship your order collect.
2. Clearly mark your return merchandise paperwork and packaging with the return number to receive credit for your return.
5.4. INTERNATIONAL SHIPPING INSTRUCTIONS
The following instructions explain how to return merchandise to Rupprecht & Patashnick Co., Inc. when shipping from places outside the Continental US (interna­tionally). It is important that R&P receives and repairs defective exchanged items in a timely manner. Please read this before returning any material:
NOTE: To avoid customs delays, it is very important that you clearly
mark your shipping paperwork with the following statement:
COUNTRY OF ORIGIN: USA
You are responsible for the shipping costs of returned goods. Do not ship your order collect.
SECTION 5: R&P RETURN POLICY PAGE 5-4
Page 85
Service Manual, Partisol Model 2000 Air Sampler
1. Clearly mark your return merchandise paperwork and packaging with the return number to receive credit for your return. Mark your paperwork:
2. R&P will instruct you on the method of return, as follows: — For small items with no rush; ship regular air mail - parcel post. — For intermediate or rush shipments using UPS, FEDEX or DHL,
have the shipper handle customs. List “Rupprecht & Patashnick Co., Inc.” as the “cosignee” and yourself as the “shipper.”
— For large items (rush or otherwise) use an air cargo service terminat-
ing at JFK International Airport. Shipping to a different airport will result in significant lost time and added expense. Prepare your “Bill of Lading” as follows:
List “Rupprecht & Patashnick Co., Inc.” as the “cosignee” and yourself as the “shipper.” Address the boxes as follows:
Revision B.000
“COUNTRY OF ORIGIN: USA.”
Rupprecht & Patashnick Co., Inc. c/o Air Express International Phone: (718) 244-4861 J. F. Kennedy International Airport Building 89 Jamaica, NY 11430 Attn.: Julia Harris
— Avoid shipping directly to the Albany Airport. If you must ship
directly to the Albany Airport, prepare your “Bill of Lading” as follows:
List “Rupprecht & Patashnick Co., Inc.” as the “cosignee” and yourself as the “shipper.” List the following customs broker as “notify”:
Notify: James E. Curran & Co., Inc.
29 Dartmouth St. Albany, NY 12209
SECTION 5: R&P RETURN POLICY PAGE 5-5
Page 86
Service Manual, Partisol Model 2000 Air Sampler
3. For non-parcel post shipments over $2,500, a customs declaration letter must accompany the shipment. It should be worded as follows:
“I, __________, declare that to the best of my knowledge and belief, the articles herein specified are products of the USA; that they were exported from the USA, from Albany, New York; and that they are returned without having been advanced in value or improved in condition by any process of manufacture or other means.”
Indicate on the declaration letter the items being returned and the purpose for return (“For Credit,” “For Replacement” or “For Repair”).
Revision B.000
SECTION 5: R&P RETURN POLICY PAGE 5-6
Page 87
Service Manual, Partisol Model 2000 Air Sampler
Appendix A: Hardware Connections
The Partisol Model 2000 Air Sampler contains three connectors for communicating analog and serial information with external devices:
User Output This 4-pin connector, located on the back panel of the
Revision B.000
sampler, transmits an analog voltage level in accordance with the currently active sampling station and existence of any status conditions. The pin layout, as viewed from the back of the enclosure, is shown in Figure A-1 and is defined below:
Pin 1 Cable shield Pin 2 Ground Pin 3 Analog voltage (0-5 VDC) Pin 4 Not used.
Anemometer This 6-pin connector, located on the back panel of the
sampler, connects the sampler to a wind vane/anemom­eter or to an external device, such as a data logger that provides a 0 to 5 VDC control signal to pin 1. The pin layout, as viewed from the back of the enclosure, is shown in Figure A-1 and is defined below:
Pin 1 Wind direction (or control) analog input
(0-5 VDC: 0-50 m/sec (0-180 km/h))
Pin 2 Wind speed analog input
(0-5 VDC: 0-360°) Pin 3 +24 VDC output Pin 4 Ground Pin 5 Cable shield Pin 6 Static ground.
RS-232 This 9-pin connector, located on the keypad panel of the
sampler, allows two-way serial communication between the hub and an external device. The 9 pins in the RS-232 connector are defined in the following manner:
Pin 1 CD Pin 6 DSR Pin 2 RXD Pin 7 RTS Pin 3 TXD Pin 8 CTS Pin 4 DTR Pin 9 RI. Pin 5 GND
APPENDIX A: HARDWARE CONNECTIONS PAGE A-1
Page 88
Service Manual, Partisol Model 2000 Air Sampler
Figure A-1. Pin assign­ments.
Revision B.000
15
6
42
3
View looking at back of
enclosure
13
2
View looking at back of
enclosure
4
Each hub unit is supplied with one cable connector for the 4-pin user output connector (R&P part number 06-002453) and one cable connector for the 6-pin anemometer connector (R&P part number 06-002454). The internal construction of these two connectors is displayed in Figure A-2.
To connect a personal computer (PC) with your sampler, you will need a straight 9­to-9 pin cable, which may be purchased from R&P as optional equipment (R&P part number 07-000587).
APPENDIX A: HARDWARE CONNECTIONS PAGE A-2
Page 89
Service Manual, Partisol Model 2000 Air Sampler
Revision B.000
Figure A-2. Internal construction of 4- and 6-pin cable connectors.
APPENDIX A: HARDWARE CONNECTIONS PAGE A-3
Page 90
This page left intentionally blank.
Page 91
Service Manual, Partisol Model 2000 Air Sampler
Index
A
air intake filter 3-5 ambient pressure
calibration 3-1, 3-18 error code 2-1 troubleshooting 2-13
ambient temperature
calibration 3-1, 3-17 connector 2-4 error code 2-1 troubleshooting 2-6
analog input
calibration 3-1, 3-14 calibration cable 3-14
anemometer
connector 2-4, A-1
Audit screen 3-7
Revision B.000
B
batteries 3-9
maintenance 3-1, 3-9 RAM 3-10 round 3-9
C
cables
checking continuity 2-2
calibration procedures 1-1, 3-11
ambient pressure 3-1, 3-18 ambient temperature 3-1, 3-17 analog input 3-1, 3-14 flow 3-1, 3-19 interface board 3-1, 3-11 intervals 3-1
mass flow controller 3-1, 3-19 Calibration screen 3-15 circuit breaker 3-23 connectors
3-pin, round 2-4
4-pin, round 2-4, A-1
6-pin, round 2-4, A-1
anemometer 2-4, A-1
D-type 2-3, A-1
finding pin 1 2-2, A-1
molex 2-3
ribbon cable 2-3
INDEX I
Page 92
Service Manual, Partisol Model 2000 Air Sampler
Index (continued)
round 2-4, A-1
RS-232 2-3, 3-23, A-1
user analog output 2-4, A-1 contrast knob 3-2, 3-23 CPU board
removing 3-28
replacing 3-28
D
D-type connectors
finding pin 1 2-3, A-1
RS-232 2-3, A-1 display
backlight troubleshooting 2-9
cover 3-23
removing 3-22
replacing 3-24
troubleshooting 2-7
Revision B.000
E
electronics compartment door 3-2 electronics heater/fan assembly
removing 3-26
replacing 3-27
troubleshooting 2-10 electronics temperature
error code 2-1
troubleshooting 2-10 error codes 2-1
ambient pressure (P) 2-1, 2-13
ambient temperature (T) 2-1, 2-6
electronics temperature (I) 2-1, 2-10
flow out of range (F) 2-1, 2-11
no errors (OK) 2-1
power failure (E) 2-1, 2-18
F
fan
removing 3-24
replacing 3-25 filter
air intake 3-1, 3-5
large in-line 3-1, 3-2 flow
calibration 3-1, 3-19
error code 2-1
troubleshooting 2-11
INDEX II
Page 93
Service Manual, Partisol Model 2000 Air Sampler
Index (continued)
flow audit adapter 3-7, 3-8, 3-19 FTS Flow Calibration screen 3-20 fuses 3-1, 3-3
G
guides, troubleshooting 1-1, 2-1, 2-5
ambient pressure 2-13
ambient temperature 2-6
display 2-7
display backlight 2-9
electronics temperature 2-10
flow out of range 2-11
keypad 2-13
power failure 2-18
power supply 2-14
system leak 2-18
user analog output 2-19
using 2-1
Revision B.000
H
hardware connections A-1
I
interconnect diagrams 2-5, 4-10 interface board 2-5
calibration 3-1, 3-11
removing 3-28
replacing 3-28
schematic diagram 4-2
test points 3-12, 4-2 interface/CPU board
removing 3-28
replacing 3-28
K
keypad
removing 3-29
replacing 3-30
troubleshooting 2-13
L
large in-line filter 3-1
location 3-2
INDEX III
Page 94
Service Manual, Partisol Model 2000 Air Sampler
Index (continued)
M
Main screen 3-14 maintenance procedures 1-1, 3-1
air intake filters 3-1, 3-5
batteries 3-1, 3-9
battery-backed RAM 3-1, 3-10
fuses 3-1, 3-3
intervals 3-1
large in-line filter 3-1
PM-10 inlet 3-1
pump 3-1
round battery 3-1, 3-9
system leak test 3-1, 3-7
V-seals 3-1, 3-6 manual, overview 1-1 mass flow controller 3-9
access door 3-2
calibration 3-1, 3-19
removing 3-25
replacing 3-26 molex connectors
finding pin 1 2-3 multimeter 3-11, 3-14
checking cable’s continuity 2-2
required resolution 2-1
Revision B.000
P
PM-10 inlet maintenance 3-1 power failure
error code 2-1
troubleshooting 2-18 pump
compartment 3-31
compartment cover 3-2
maintenance 3-1
removing 3-30
replacing 3-31
R
remove and replace procedures 3-22
CPU board 3-27
display 3-22
electronics heater/fan assembly 3-26
fan 3-24
interface board 3-27
interface/CPU board assembly 3-27
INDEX IV
Page 95
Service Manual, Partisol Model 2000 Air Sampler
Index (continued)
keypad 3-29
mass flow controller 3-25
pump 3-30
valve 3-32 return policy 1-1, 5-1
instrument components, billable exchange 5-1
instrument components, billable repair 5-2
instrument components, warranty 5-1
shipping, international 5-4
shipping, US 5-4
systems, repair 5-3
systems, warranty 5-3 ribbon cable connectors
finding pin 1 2-3 round connectors
3-pin 2-4
4-pin 2-4, A-1
6-pin 2-4, A-1
ambient temperature 2-4
anemometer 2-4, A-1
finding pin 1 2-4, A-1
user analog output 2-4, A-1 RS-232 connector 3-23, A-1
finding pin 1 2-3, A-1
Revision B.000
S
schematic diagrams 1-1, 4-1
interconnect diagrams 4-10
interface board 4-2 screens, software
Audit screen 3-7
Calibration screen 3-15
FTS Flow Calibration screen 3-20
Main screen 3-14
Setup screen 3-10 service department
email address 5-1
fax number 5-1
phone number 5-1 Setup screen 3-10 Streamline FTS Flow Transfer Standard 3-19, 3-21 system leak
testing for 3-1, 3-7
troubleshooting 2-18
INDEX V
Page 96
Service Manual, Partisol Model 2000 Air Sampler
T
test points, electronic board 2-2, 3-12, 4-2 troubleshooting 2-1
How do I check a cable’s continuity? 2-2
How do I find pin 1? 2-2
How do I locate a test point? 2-2
What equipment do I need? 2-1
Where do I start? 2-1 troubleshooting guides 1-1, 2-5
ambient pressure 2-13
ambient temperature 2-6
display 2-7
display backlight 2-9
electronics temperature 2-10
flow out of range 2-11
keypad 2-13
power failure 2-18
power supply 2-14
system leak 2-18
user analog output 2-19
using 2-1
Revision B.000
U
user analog output
connector 2-4, A-1
troubleshooting 2-19
V
V-seals 3-1, 3-6 valve
manifold 3-33
removing 3-32
replacing 3-32
INDEX VI
Loading...