
Visit us online to register your warranty
www.thermoscientific.com/labwarranty
User Manual
Locator and Locator Plus
Cryogenic Storage Vessel with or without
Ultrasonic Level Monitor
Operation Manual and Parts List LT509X9 (7018141) Rev. 4

Thermo Scientific
MANUAL NUMBER LT509X9 (7018141)
4 40578 2/23/16 Clarified Do Not Roll warning ccs
3 -- 10/27/15 Updated symbols and labels, added cautions ccs
2 31623 5/15/15 Added caution for moving vessel in Section 3 ccs
1 31528 9/15/14 Updated conformity specs and declaration ccs
0 27564 1/31/14 Original - new PCBA design ccs
REV ECR/ECN DATE DESCRIPTION By
Preface
Cryogenic Storage Vessel i
Locator Catalog Number
Locator
Model
Name
CY50925 w/o Monitor 8141 Locator Jr
CY50935 w/o Monitor 8143 Locator 4
CY50985 w/o Monitor 8197 Locator 6
CY50945 w/o Monitor 8145 Locator 8
CY509105 w/ Monitor 8201 Locator Jr.
CY509105CN w/ Monitor 8202 Locator Jr. China
CY509107 w/ Monitor 8246 Locator 4
CY509107CN w/ Monitor 8247 Locator 4 China
CY509113 w/ Monitor 8198 Locator 6
CY509110 w/ Monitor 8211 Locator 8
CY509110CN w/ Monitor 8212 Locator 8 China
Locator Plus Catalog Number
Locator Plus
Model
Name
CY50925-70 w/o Monitor 8142 Locator Jr Plus
CY50935-70 w/o Monitor 8144 Locator 4 Plus
CY50985-70 w/o Monitor 8147 Locator 6 Plus
CY50945-70 w/o Monitor 8146 Locator 8 Plus
CY509106 w/ Monitor 8237 Locator Jr Plus
CY509106CN w/ Monitor 8238 Locator Jr Plus China
CY509108 w/ Monitor 8207 Locator 4 Plus
CY509108CN w/ Monitor 8208 Locator 4 Plus China
CY509109 w/ Monitor 8209 Locator 6 Plus
CY509109CN w/ Monitor 8210 Locator 6 Plus China
CY509111 w/ Monitor 8213 Locator 8 Plus
CY509111CN w/ Monitor 8214 Locator 8 Plus China

Thermo Scientificii Cryogenic Storage Vessel
Preface
Important Read this instruction manual. Failure to read, understand and follow instructions in this manual may
result in damage to the unit, injury to operating personnel, and poor equipment performance.
s
Caution All internal adjustments and maintenance must be performed by qualified service personnel. s
Caution It is recommended to have this vessel tested by the manufacturer or qualified cryovessel service
technician every 7-10 years, regardless of any problems (or lack thereof) you may have had in the past.
This will help insure your samples against sudden loss of liquid nitrogen due to vacuum failure.
s
Warning Nitrogen gas can cause suffocation without warning. Store and use liquid nitrogen only in a
well-ventilated place. As the liquid evaporates, the resulting gas displaces the normal air in the area.
(The cloudy vapor that appears when liquid nitrogen is exposed to the air is condensed moisture, not
the gas itself. The issuing gas is invisible.) In closed areas, excessive amounts of nitrogen gas reduces the
concentration of oxygen and can result in asphyxiation. Because nitrogen gas is colorless, odorless, and
tasteless, it cannot be detected by the human senses. Breathing an atmosphere that contains less than
19.5% oxygen can cause dizziness and quickly result in unconsciousness and death. Therefore, the use
of oxygen monitoring equipment is strongly recommended.
s
Warning Contact of liquid nitrogen with the skin or eyes may cause serious (freezing) injury. s
Warning Do not roll. Always keep liquid nitrogen vessel in an upright position. Exercise caution to
prevent spilling and splashing.
s
Material in this manual is for information purposes only. The contents and the product it describes are subject
to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.
Intended Use:
The Thermo Fisher Scientific cryogenic storage vessels are intended to store biological samples at cryogenic
temperatures. These products are comprised of a vessel and a lid, with or without a level monitor.
Unintended Use:
These products are not intended for use within the patient environment.
When this manual is translated into other languages, the US English version is binding.
© 2010 Thermo Fisher Scientific. All rights reserved. Printed in U.S.A.
CAUTION
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)

Thermo Scientific Cryogenic Storage Vessel iii
Preface
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this
symbol.
Extreme temperature hazards. Only qualified persons should perform procedures associated with this
symbol.
Potential biological hazards. Proper protective equipment and procedures must be used.
Marking of electrical and electronic equipment, which applies to electrical and electronic equipment
falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market
after 13 August 2005.
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Fisher
Scientific has contracted with one or more recycling/disposal companies in each EU Member State
European Country, and this product should be disposed of or recycled through them. Further
information on Thermo’s compliance with this directive, the recyclers in your country and
information on Thermo Scientific products will be available at www.thermoscientific.com.
4 Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4 Always dissipate extreme cold or heat and wear protective clothing.
4 Always follow good hygiene practices.
4 Each individual is responsible for his or her own safety.

Thermo Scientificiv Cryogenic Storage Vessel
Preface
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct
1-800-438-4851 Toll Free, U.S. and Canada
1-877-213-8051 FAX
http://www.thermoscientific.com Internet Worldwide Web Home Page
service.led.marietta@thermofisher.com Tech Support Email Address
Certified Service Web Pagewww.unitylabservices.com
Our staff can provide information on pricing and give you quotations. We canSales Support
take your order and provide delivery information on major equipment items or make
arrangements to have your local sales representative contact you. Our products are listed on the
Internet and we can be contacted through our Internet home page.
Our staff can supply technical information about proper setup, operation orService Support
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or
provide you with on-site service. We can also provide you with a quotation on our Extended
Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your
applications. If you are experiencing technical problems, working together, we will help you
locate the problem and, chances are, correct it yourself...over the telephone without a service
call.
When more extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If your service need is
covered by the warranty, we will arrange for the unit to be repaired at our expense and to your
satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you
Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. If you wish to write, our mailing address is:
Thermo Fisher Scientific (Asheville) LLC
401 Millcreek Road, Box 649
Marietta, OH 45750
International customers, please contact your local Thermo Scientific distributor.

Cryogenic Storage Vessel vThermo Scientific
Table of Contents
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Please Read . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
General Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Measuring Liquid Nitrogen Quantity . . . . . . . . . . . . . . . . . . . . . . . .4-1
General Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Storing Samples in Vapor Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Liquid Nitrogen Level Monitor Operation . . . . . . . . . . . . . . . . . . . . . . . .6-1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Remote Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Alarm Contact Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Microprocessor Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Level Monitor Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Safety Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Audible and Visual Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
LN2 Level Monitor Troubleshooting and Parts List . . . . . . . . . . . . . . .7-1
Ordering Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Section 1
Section 2
Section 3
Section 4
Section 5
Section 6
Section 7

Cryogenic Storage Vessel 1-1Thermo Scientific
Please Read
Section 1 Safety Information
This manual contains important operating and safety information. The
user must carefully read and understand the contents of this manual prior
to the use of this equipment.
Your Thermo Scientific cryogenic vessel has been designed with function,
reliability, and safety in mind. It is the user’s responsibility to install it in
conformance with local electrical codes. For safe operation, please pay
attention to the warnings and cautions throughout the manual.
All cryopreservation vessels, regardless of who manufactures them, use a
vacuum to provide the super insulative properties needed to keep nitrogen
in a liquid form. They are very similar, in fact, to the Thermos® vessels you
may have used to store coffee, soup or milk.
Remember how fragile they were? Since no vacuum is perfect or will last
forever, we suggest that you monitor the consumption of liquid nitrogen
used by your vessel on a regular basis, i.e. every 1-2 days. By monitoring,
you may be able to anticipate subtle changes in consumption and possibly
react to problems before they arise. You may opt to use a level monitor for
this purpose (see page 6-1 for level monitor operation.).
Also, if your samples are super critical, one-of-a-kind, or irreplaceable,
consider allocating your samples to 2 separate vessels. Remember the old
adage about placing all your eggs into one basket….
Finally, because no vacuum will last forever, have your vessel tested by the
manufacturer or a qualified cryovessel service technician every 7-10 years,
regardless of any problems (or lack thereof) you may have had in the past.
This will help insure your samples against sudden loss of nitrogen due to
vacuum failure.
Do not use this product for anything other than its intended usage.
Use of the equipment in a manner not specified by the manufacturer may
impair the protection provided by the equipment.
General Usage

1-2 Cryogenic Storage Vessel Thermo Scientific
Section 1
Safety Information
• This unit is intended to be used by trained personnel in a wellventilated environment.
• Liquid nitrogen is extremely cold; it boils at -196°C.
• Follow all necessary precautions and warnings relevant to applicable
hazards such as frostbite, contact burn, asphyxiation, etc. in cryogenic
area.
• To avoid injury due to frostbite or ruptured vials, use extreme care
whenever handling liquid nitrogen, liquid nitrogen storage or transfer
vessels, or any objects which have come in contact with liquid
nitrogen.
• Leave no areas of skin exposed.
• Wear appropriate PPE where necessary; face shield, cryogenic gloves,
cryogenic apron.
• Do not tightly seal liquid nitrogen containers or prevent nitrogen gas
from escaping.
• Always handle ampules with tongs. They may explode when removed
from the vessel.
• Use extreme care to prevent spilling and splashing liquid nitrogen
during transfer and removal of storage contents and holders.
• Immediately remove any clothing or safety attire on which liquid
nitrogen has been spilled or splashed.
• Get immediate medical attention for any frostbite injuries due to
liquid nitrogen.
Warnings

• Only use the power supply provided by the unit.
• Never overfill liquid nitrogen vessels. Liquid nitrogen level should
never be above the tops of the racks inside the vessel. The excess
weight may crack the neck and result in vacuum failure.
• When inserting or removing racks, be careful not to come in contact
with the neck tube area of the vessel. Remove or insert racks slowly
in a vertical manner. Scratches on the neck tube area can cause
premature vacuum failure.
• Do not tamper with or remove vacuum port (covered by a black
plastic cap on side of vessel); this will void warranty.
• Do not spill liquid nitrogen on vacuum port - this can cause vacuum
failure.
• Always keep liquid nitrogen vessel in an upright position.
• Do not use this Locator vessel for transport.
Cryogenic Storage Vessel 1-3Thermo Scientific
Section 1
Section title
Cautions

Cryogenic Storage Vessel 2-1Thermo Scientific
Section 2 General Specifications
Operating temperature: 17°C - 27°C
20% to 80% relative humidity, non-condensing.
Installation Category II (overvoltage) in accordance with IEC 664.
Pollution Degree 2 in accordance with IEC 664.
Altitude limit: 2,000 meters.
Storage: -25°C to 65°C; 10% to 85% relative humidity.
Locator and Locator Plus units are intended for indoor use only.
Locator 4 Locator 4 Plus Locator 6 Locator 6 Plus Locator 8 Locator 8 Plus Locator JR Locator JR Plus
Height
(including lid)
1
37.5”
(95.2cm)
39.5” (100.3cm)
39.5”
(100.3cm)
39.5” (100.3cm)
37.5”
(95.2cm)
39.5” (100.3cm)
26.5”
(67.3cm)
28.9” (73.4cm)
Diameter
22”
(55.9cm)
22” (55.9cm) 26” (66cm) 26” (66cm)
22”
(55.9cm)
22” (55.9cm) 22” (55.9cm) 22” (55.9cm)
LN2 Capacity,
Liquid Phase
Storage
111L 121L 184L 184L 111L 121L 60L 71L
Static LN2
Evaporation
rate - L/Day
3
.99 .99 .99 .99 .60 .60 .85 .85
Liquid Phase
Capacity (2mL
Vial2actually
submerged in
liquid nitrogen)
3600 4000 6000 6000 1800 2000 1600 2000
1
The liquid nitrogen level monitor will add 1 inch to these heights. The transportation cart will add 4 inches to these heights.
2
Using Nalgene System 100 cryoboxes with a 10 x 10 vial configuration (Nalgene System 100 vial or equivalent are needed).
3
Actual working time may vary due to current atmospheric conditions, container history, manufacturing tolerances, and any individual patterns
of use.
Environmental
Conditions

(CE models only) We hereby declare under our sole responsibility that this
product conforms with the technical requirements of the following
standards:
EMC:
EN 61000-3-2: Limits for Harmonic Current Emissions
EN 61000-3-3: Limits for Voltage Fluctuations and Flicker
EN 61326-1 Electrical Equipment for Measurement, Control, and
Laboratory Use; Part I: General Requirements
Safety:
EN 61010-1 Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use; Part I: General
Requirements
per the provisions of the Electromagnetic Compatibility Directive
2004/108/EC.
Medical Device Directive 93/42/EEC as amended by 2007/47/EC
Copies of the Declaration of Conformity are available upon request.
2-2 Cryogenic Storage Vessel Thermo Scientific
Section 2
General Specifications
Declaration of
Conformity
Electrical
(cover with monitors only)
Inputs . . .100-240VDC, .08 - .04 amps, single phase
Frequency . . . . . . . . . . . . . . . . . . . . . . . .47-63 Hz
Output . . . . . .15VDC (±10%), 0.2 amps, 3.0 watts
0543

Cryogenic Storage Vessel 3-1Thermo Scientific
Section 3 Unpacking
Before using your new Locator or Locator Plus vessel, carefully inspect the
vessel prior to use. Check for signs of damage which may have occurred in
shipment. It is advisable to fill (see filling instructions) all new units with
liquid nitrogen and watch liquid nitrogen loss rate for a few days. If there
are any problems, call Customer Service as soon as possible.
The warranty registration card enclosed with the vessel must be completed
and returned to the factory within 30 days to make warranty effective.
This information must include the serial number, which is located on the
label on the vessel.
Note The most prevalent cause of failure of liquid nitrogen storage vessels
is mechanical. The vessel necktube supports the full weight of the inner
shell and all the liquid nitrogen it contains. A side blow to the vessel causes
the inner shell to swing in a pendulum-like motion causing the necktube
to be damaged. Any storage vessel which has been in an accident, has been
dropped, or lowered to hit on its side, will tend to fail more rapidly than
one that has not.
Caution Exercise caution when moving your Locator or Locator Plus
vessel. Locator and Locator Plus cryobiological storage systems are not
transportation vessels. Transport carts are designed for mobility within the
lab, or lab to lab only. Moving full vessels long distances, over cracks in
floor, thresholds, on inclined ramps or in elevators can cause premature
vacuum failure. Avoid using excessive force when moving vessel; this can
cause tipping, resulting in damage and/or spilling of liquid nitrogen.
s
Note If samples must be transported under cryogenic conditions, consider
Arctic Express line of Dry Shippers available from Thermo Scientific.

Cryogenic Storage Vessel 4-1Thermo Scientific
Section 4 Operation
Caution Never overfill your Locator or Locator Plus vessel with liquid
nitrogen. The liquid nitrogen level in your tank (with racks inserted)
should never be above 20 inches (50cm) for Locator 4 and Locator 8, 10
inches (25cm) for Locator Jr., 22 inches (55.8cm) for Locator 4, 6, and 8
Plus and 12 inches for Locator Jr. Plus. Filling the tank up to or above the
bottom of the necktube may cause immediate or premature vacuum failure
to occur.
s
Caution When inserting or removing racks, be careful not to come in
contact with the necktube area of the vessel. Remove or insert racks slowly
in a vertical manner. Scratches on the necktube area can cause premature
vacuum failure.
s
Caution Do not spill liquid nitrogen on vacuum port (covered by a black
plastic cap on side of vessel). This can cause vacuum failure. s
To avoid damage to your Locator or Locator Plus cryogenic storage vessel
which may result in premature vacuum loss, it is important that the
following procedure be used when adding liquid nitrogen to a warm vessel.
1. Add only a small amount of liquid nitrogen (5-10 liters) to a new or
warm vessel.
2. Allow this small amount of liquid nitrogen to sit in the covered vessel
for a minimum of 2 hours. This will limit stress caused by the sudden
temperature change associated with adding liquid nitrogen to a warm
vessel.
3. Add an additional 15 liters of liquid nitrogen to the vessel.
4. Allow vessel to sit for 48 hours and monitor liquid nitrogen
consumption with a wooden yarsdstick, optional dip measurer (part
#180143) or Level Monitor.
5. Fill Locator or Locator Plus as indicated (see Caution in this section).
Allow for displacement of liquid nitrogen when racks and boxes are
inserted.
6. Insert and remove racks slowly. Allow liquid nitrogen to run out of
boxes and off racks.
Filling Instructions

4-2 Cryogenic Storage Vessel Thermo Scientific
Section 4
Operation
1. Use a wooden yardstick or optional dip measurer (part #180143) to
measure liquid nitrogen level. NEVER use a hollow tube or plastic
dipstick to measure the liquid nitrogen level.
2. Level will be indicated by the frostline which develops when dipstick is
removed.
3. Level Monitor can be used for constant measurements. See Section 6.
1. Remove the 2 bottom-most cryoboxes from each rack.
2. Measure the height of 2 cryoboxes stacked on top of each other.
3. Fill vessel with liquid nitrogen to the height obtained from Step 2,
using a yardstick to measure liquid level in vessel.
4. Place racks into vessel without the 2 bottom-most cryoboxes. Samples
are now above the liquid nitrogen and are stored in the vapor phase.
Important A level monitor should always be used when you store your
samples in the vapor phase because of the already low levels of liquid
nitrogen. Refer to Section 5 for correct level monitor for your system.
s
Wipe exterior surfaces with lightly dampened cloth containing mild soap
solution.
Storing Samples in
Vapor Phase
General Cleaning
Instructions
Measuring Liquid
Nitrogen Quantity

Cryogenic Storage Vessel 5-1Thermo Scientific
Section 5 Replacement Parts
Locator and Locator Plus Covers w/o Monitor
Type Standard Covers
Covers and Rack Kits (for units with
lids made before 3/2010)
Locator Jr. CV509X51 (321746) 1950514
Locator Jr. Plus CV509X51 (321746) 1950515
Locator 4 CV509X51 (321746) 1950516
Locator 4 Plus CV509X51 (321746) 1950517
Locator 6 CV509X50 (321651) 1950513
Locator 6 Plus CV509X50 (321651) 1950513
Locator 8 CV509X23 (317227) CV509X23 (317227)
Locator 8 Plus CV509X23 (317227) CV509X23 (317227)

5-2 Cryogenic Storage Vessel Thermo Scientific
Section 5
Replacement Parts
Catalog # Rack Description
HR509X15A (317050) For Locator Jr (8141, 8201, 8202) with 4 - 2 inch boxes
HR509X2A (317161) For Locator 4 (8143, 8246, 8247) with 9 - 2 inch boxes
HR509X19A (315903) For Locator Jr (8141, 8201, 8202) with 2 - 4 inch boxes
HR509X20A (315905) For Locator 4 (8143, 8246, 8247) with 5 - 4 inch boxes
HR509X15A-70 (317148) For Locator Jr Plus (8142, 8237, 8238) with 5 - 2 inch boxes
HR509X2A-70 (317175) For Locator 4 Plus (8144, 8207, 8208) with 10 - 2 inch boxes
HR509X31A-70 (317192) For Locator 6 and 6 Plus (8147, 8197, 8198, 8209, 8210) with 10 - 2 inch boxes
HR509X1A (317205) For Locator 8 (8211, 8212, 8145) with 9 - 2 inch boxes (25 cell)
HR509X1A-70 (317217) For Locator 8 Plus (8146, 8213, 8214) with 10 - 2 inch boxes (25 cell)
HR509X19A-70 (315904) For Locator Jr. Plus (8142, 8237, 8238) with 2 - 4 inch boxes + 1 - 2 inch box
HR509X20A-70 (315906) For Locator 4 Plus (8144, 8207, 8208) with 5 - 4 inch boxes + 1 - 2 inch box
HR509X32A-70 (315907) For Locator 6 and 6 Plus (8147, 8197, 8198, 8209, 8210) with 5 - 4 inch boxes + 1 - 2 inch box

Section 6 Liquid Nitrogen Level
Monitor Operation
The liquid nitrogen (LN2) level monitor provides a constant indication of
the LN2 level and notifies you to low level conditions through audible and
visual alarms. It alerts you when your vessel needs filling. The level
monitor is shipped mounted directly to the appropriate Locator or Locator
Plus cover. It can be wired to a remote alarm system to alert you to
problems when you are not in your laboratory. This equipment was
intended to be operated with an external power supply provided by the
manufacturer. It is designed for use with Thermo Scientific LN2 vessels
only.
Caution Use this product for its intended usage only.
s
Warning Potential electrical hazards. Only qualified persons should
perform procedures associated with this symbol. s
Note If the level monitor was purchased as part of a Locator or Locator
Plus system, skip to Step 3. If it is a retrofit for an existing Locator or
Locator Plus, begin with Step 1.
s
1. Remove the level monitor and Locator or Locator Plus cover from box.
2. Remove present cover from your Locator or Locator Plus and replace
with your new level monitor and cover.
3. Check the electrical specifications on the power supply for electrical
data and plug in to appropriately grounded receptacle.
Cryogenic Storage Vessel 6-1Thermo Scientific
Installation

1. Disconnect the liquid nitrogen level monitor from the power supply.
Disconnect power to your alarm system.
2.
Remove the three screws securing the liquid nitrogen level monitor’s
cover. Pull the cover up and out of the way (see Figure 6-1).
When connected to your remote alarm system, the LN2 level monitor will
alert you to a problem even when your laboratory is unoccupied. The
unit’s alarm contacts are activated at the same time the audible alarm
sounds (see Microprocessor Functions).
3. Remove the black plug from the side of the cover. Insert the wires for
the remote alarm through this hole.
4. Connect the remote alarm wires as shown in Figures 6-1, 2 and 3.
5. Replace the LN2 level monitor’s cover. Reconnect power to the LN2
level monitor and to your remote alarm system.
Note The LN2 level monitor can be configured with either a normally
open contact or a normally closed contact. Which mode is currently
operative depends on the placement of a single jumper on the circuit
board. For a normally open contact, the jumper should be placed on J6.
For a normally closed contact, the jumper should be placed on J4 (see
Figure 6-2). Verify that the jumper is set appropriately for your remote
alarm system.
s
The LN2 level monitor’s alarm contact is a Photomos type relay.
Contact Form: Type 1 form B
Load Voltage Maximum: 24VDC Max
Load Current Maximum: 120mA
Typical On Resistance: Maximum 26 Ω
6-2 Cryogenic Storage Vessel Thermo Scientific
Section 6
LN2 Level Monitor Operation
Alarm Contact
Specifications
Level Monitor
Remote Alarm
Connection to Remote
Alarms

Cryogenic Storage Vessel 6-3Thermo Scientific
Section 6
LN2 Level Monitor Operation
Figure 6-1. Connect to Remote Alarm
Figure 6-2. Jumper Location
Remote Alarm wires connect here
(Note polarity.)
- +
J3
J4 J6
1
1
For normally open contact, place jumper here.
For normally closed contact, place jumper here.

The innovative Thermo LN2 Level Monitor utilises ultrasonic sound
waves to sense the level of liquid nitrogen inside a Locator or Locator Plus
cryogenic storage vessel. A microprocessor control provides continuous
indication of the liquid nitrogen level inside the Locator or Locator Plus,
with accuracy to ±2”. Audible and visual alarms also provide an early
warning of low level conditions.
Ultrasonic sound waves sense the level of liquid nitrogen inside the Locator
or Locator Plus vessel.
By using ultrasonic sound waves to sense the level of liquid nitrogen, no
physical probe is required to sense the level of liquid nitrogen.
Conventional devices use probes which must be in physical contact with
the liquid. This physical contact increases liquid nitrogen consumption.
There is NO increase in liquid nitrogen consumption by using the
Thermo level monitor.
The LED display is graduated in 8 segments and provides quick and
reliable visual indication of the level of liquid nitrogen inside the Locator
or Locator Plus vessel, from Full to Empty, similar to a fuel gauge on an
automobile.
6-4 Cryogenic Storage Vessel Thermo Scientific
Section 6
LN2 Level Monitor Operation
Principles of
Operation
Microprocessor
Functions
Figure 6-3. Remote Alarm Connections Schematic
U4
1
2
36
The position of J4 and J6 indicate
whether the relay is normally open
or normally closed.
4
5
J3-1
J3-2
Load
V
L
Load Voltage
(DC)

1. “LOW LEVEL” indicator light illuminates continuously when liquid
nitrogen level is below the approximate 2" level. In this situation, the
level of liquid nitrogen is low, and you are alerted to refill the vessel as
soon as possible.
2. An AUDIBLE and remote alarm activates continuously when liquid
nitrogen is below the approximate 1” level inside the Locator or
Locator Plus vessel, additionally, “LOW LEVEL” light will continue to
flash. In this situation, the liquid nitrogen is dangerously low, and you
are alerted to refill the vessel immediately.
3. The AUDIBLE and remote alarms activate continuously, the 8segment LED display and “Low LEVEL” lights flash, to indicate an
emergency event. This alarm event occurs when the Locator or Locator
Plus vessel is empty of liquid nitrogen, or the monitor has an error and
is incapable of detecting LN2 levels.
The microprocessor must be reset each time the cover is removed from the
Locator or Locator Plus. Before removing the cover, the RESET button
must be pressed, illuminating the RESET light. While in this Reset/Hold
mode, the monitor is not sensing. After five minutes in the Reset/Hold
mode, an audible and remote alarm will activate, alerting you to deactivate
the Reset/Hold mode when the cover is replaced on the vessel.
Once the cover with the monitor has been placed on the vessel and
plugged in, wait about 15 minutes for the system to equilibrate before
pressing the Reset button for an accurate measurement of the liquid
nitrogen to be indicated. This equilibration process will occur each time
the cover is removed. Each time a vessel is opened or moved, the stabilized
chamber temperature and/or liquid level is disrupted. The system requires
a stable temperature and liquid level to determine an accurate liquid
nitrogen level.
If you have just purchased a new Locator or Locator Plus along with your
monitor, allow at least 1 day for the liquid nitrogen to equilibrate the
vessel before plugging in the unit. This will avoid false alarms.
Note The Level Monitor may take up to 10 minutes to establish a reading
after the unit is placed on the cryovessel, or the cryovessel is moved.
Remember to press “Reset” after placing unit on the vessel. The green
LED on the switch should be “OFF.”
s
Cryogenic Storage Vessel 6-5Thermo Scientific
Section 6
LN2 Level Monitor Operation
Audible and Visual
Alarms
Safety Feature
Level Monitor
Operation

Section 7 LN2 Level Monitor
Troubleshooting and Parts List
Cryogenic Storage Vessel 7-1Thermo Scientific
Problem Cause Solution
Level monitor alarm sounds
intermittently. Unit measuring liquid
level incorrectly.
Ice or frost buildup on the
transducer.
Remove the cover w/ monitor from vessel and place in room for approximately two hours, or until no ice is visible on the transducer or tube.
NOTE: Do not use an external heat source (i.e. heat gun) to accelerate
the thawing process as it can damage the cork. Make sure lid is standing upright on the cork to allow moisture to drain out of cover w/ monitor. Wipe any moisture away from the inside of the tube and face of the
transducer with a clean soft cloth. Place cover w/ monitor back on unit.
(Remember to press “Reset” button.)
Level monitor measuring “Full” after
removing samples and replacing lid.
Fog in top part (neck area) of
cryogenic vessel.
The density of the fog that is formed after removing the lid and accessing samples is dependent on the humidity of the room where the cryovessel is located. The fog dissipates after about fifteen minutes and at
this time, the level monitor will take an accurate reading of the liquid
nitrogen level.
Ordering Procedures
Refer to the Specification Plate for the complete model number, serial
number, and series number when requesting service, replacement parts or
in any correspondence concerning this unit.
All parts listed herein may be ordered from the Thermo Scientific dealer
from whom you purchased this unit or can be obtained promptly from the
factory. When service or replacement parts are needed, check first with
your dealer. If the dealer cannot process your request, then contact
Technical Services.
Prior to returning any materials, contact our Technical Services
Department for a “Return Materials Authorization” number (RMA).
Material returned without an RMA number will be refused.
Caution - Preventative Maintenance Depending on environmental
conditions and the number of lid openings, the lid should be periodically
inspected for ice and frost buildup. Refer to chart below.
s

7-2 Cryogenic Storage Vessel Thermo Scientific
PART LOCATOR JR LOCATOR JR PLUS LOCATOR 4
LOCATOR 4 PLUS
LOCATOR 6 LOCATOR 6 PLUS LOCATOR 8 LOCATOR 8 PLUS
Cover with Monitor
and Power Supply *
CN509X16
(316306)
CN509X16-70
(316308)
CN509X17
(316374)
CN509X17-70
(316376)
CN509X15
(316304)
CN509X15
(316304)
CN509X7
(316137)
CN509X7-70
(316138)
Cover with Monitor
and Power Supply
(China) **
CN509X16CN
(316307)
CN509X17CN
(316309)
CN509X18CN
(316375)
CN509X19CN
(316377)
CN509X15CN
(316305)
CN509X15CN
(316305)
CN509X7CN
(316139)
CN509X20CN
(316390)
Cover and Rack
Kits* (for units with
lids made before
03/2010)
1950503 1950505 1950507
1950509
1950501 1950501
CN509X7
(316137)
CN509X7-70
(316138)
Cover and Rack Kits
(China) ** for units
with lids made
before 03/2010
1950504 1950506
1950508
1950510
1950502
1950502
CN509X7CN
(316139)
CN509X20CN
(316390)
Cover CV509X13 CV509X13 CV509X13
CV509X13
CV509X13 CV509X13 CV509X13 CV509X13
Power Supply w/
connectors*
90-240VAC
TN509X2 (317133) *includes US, EU, UK and AUS connectors
Power Supply**
90-264VAC (China)
TNX136 (317143)
Vent Port Cap 1950343
Table 7-1. Parts List By Model

Cryogenic Storage Vessel 8-1Thermo Scientific
ISO
9001
REGISTERED
Rev. 0 5/10
(LN2 Vacuum)
THERMO FISHER SCIENTIFIC STANDARD PRODUCT WARRANTY
2 Vacuum Integrity is covered for three years. Installation and calibration are not
The Warranty Period starts two weeks from the date your equipment is shipped from our facility. This allows for shipping
time so the warranty will go into effect at approximately the same time your equipment is delivered. The warranty protec-
tion extends to any subsequent owner during the first year warranty period.
During the first year, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced
at Thermo's expense, labor included. LN
covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and
direction prior to performance of any repairs. Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the
equipment or to the component part beyond the original warranty period. The Technical Services Department must give prior
approval for return of any components or equipment. At Thermo's option, all non-conforming parts must be returned to
Thermo postage paid and replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR
IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY.
Thermo shall not be liable for any indirect or consequential damages including, without limitation, damages relating to lost
profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment
arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA and Canada) or
1-740-373-4763. We're ready to answer your questions on equipment warranty, operation, maintenance, service and
special application. Outside the USA, contact your local distributor for warranty information.

thermoscientific.com
© 2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its
subsidiaries. Specifications, terms
and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative
for details.
Thermo Fisher Scientific (Asheville) LLC
401 Millcreek Road
Marietta, Ohio 45750
United States
A Thermo Fisher Scientific Brand