Thermo Scientific Lindberg/Blue M VO Series, Lindberg/Blue M VO914, Lindberg/Blue M VO1218, Lindberg/Blue M VO1824 Installation And Operating Manual

Page 1
f
\
Thermo Fisher Scientific
LindbergIBlue M Vacuum Ovens
V0914 I V01218 I V01824
Installation and Operating Manual
34598H19
E
11/2011
\
/
Part of Thermo Fisher Scientific
Therrno
SCIENTIFIC
Page 2
© 2
011 Thermo Fisher Scientific Inc. All rights reserved.
These operating instructions are protected by copyright. Rights resulting thereof, particularly reprint, photomechanical or digital
post processing or reproduction, even in part, are only allowed with the written con
sent of Thermo Fisher Scientific Inc. This regulation does not apply to reproductions for in-plant use. The contents of this operating instructions manual may change at any time and without a
ny prior notice. In case of conflicting translations into foreign languages the German-language version of these operating instructions shall be binding.
Trademarks
Thermo Scientific is a brand owned by Thermo Fisher Scientific, Inc. Lindberg Blue M® and a
ll other trademarks
mentioned in the operating instructions are the exclusive property of the
respective manufacturers. Thermo Fisher Scientific Inc.
275 Aiken Road Asheville, NC 28804 USA
Thermo Fisher Scientific Inc. 8
1 Wyman Street Waltham, MA 02454 USA
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this d
ocument is
strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc. The contents of this document are subject to change without notice. All technical information in this document is for
reference purposes only. System con
figurations and specifications in this document supersede all previous
information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-free and assumes no responsibility and w
ill not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc
. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
Page 3
Page 4
Laboratory Vacuum
Ovens
6
Operation
.
UP150
Controller ("P" Models only)
........................................
8
6.1 Normal Controller Operation
................................................................
8
6.2 Single Setpoint Operation
..................................................................
9
6.2.1 Setting High Temperature Alarm Setpoint:
...............................................
9
6.2.2 AccessingLocalMode
..............................................................
9
6.2.3 ExitingLocalMode
.................................................................
9
..................................................
6.3 Programming Operation: Entering a Program
9
6.3.1 Entering Programming Mode
.........................................................
9
6.3.2 Entering Program Parameters
.......................................................
10
6.3.3
RunningaProgram
................................................................
11
.................................................................
6.3.4 EndingaProgram
11
6.3.5 ChangingaProgram
...............................................................
11
6.4 Auto Tuning the UP1 50 Controller
..........................................................
11
6.5 Temperature Offset Procedure
.............................................................
12
6.6 Changing Temperature Scale Between
C
and
T
..............................................
12
7
Maintenance
.....................................................................
13
7.1 GeneralMaintenance
....................................................................
13
7.2
GasketMaintenance
.....................................................................
13
7.3
DoorAlignment
.........................................................................
13
..................................................................
8
Troubleshooting
14
9
Replacementparts
................................................................
15
Page 5
Installation and Operation
Laboratory Vacuum Ovens
1
Introduction
The LindbergIBlue M VO
Series is a family of laboratory vacuum ovens designed for drying, curing, outgassing, process control, and applications which require elevated temperature in reduced atmospheres or
vacuudpurge with non-flammable
gases. Refer to Table
1
on page 2 for specifications.
1.1
Features and Benefits
Single setpoint digital electronic control (models with suffix
P
have programmable controls).
Built
-
in overtemperature protection.
Fully flexible
vacuudpurge/release system.
Vacuum system includes inert gas injection valve, fresh air inlets, vacuum port to exhaust vacuum atmosphere, and controllable vacuum release vent port.
1.2
Available Door Gaskets
In addition to the standard silicone gasket provided with the oven, the following gaskets are available from
LindbergIBlue
M:
Buna-N gasket for use with solvents up to
160°C.
Viton-B gasket for applications involving acids.
Refer to Table
7
on page 15 for gasket ordering information.
Page 6
Installation and Operation
Laboratory Vacuum Ovens
1.3
Specifications
Table
1.
Lindbergmlue M VO Series Laboratory Vacuum Ovens
2
Safety
Considerations
Model
V0914(P)A
VO914(P)C
V01218(P)A
VO1218(P)C
V01824(P)A
VO1824(P)C
V0914S(P)A
V0914S(P)C
V01218S(P)A
VOIZI8S(P)C
V0 1824S(P)A
V01824S(P)C
WARNING!
Do not modify or change system
components. Replacement parts must be
OEM.
exact replacement equipment. Modification or use of the equipment in a manner other than expressly intended may cause death or serious injury. This includes use of user
-
supplied components and materials not specifically designed for the oven. Reconfiguring the controller may cause death or serious injury.
LindbergIBiue M shall not be liable for any damages,
including incidental
and/or consequential damages,
regardless of the legal
theoty asserted, including
negligence
and/or strict liability.
WARNING1 Before maintaining this equipment,
read the applicable MSDS (Material Safety Data Sheets) provided with your oven.
Chamber
9x14~9
(22.9 X 35.6 X 22.9)
12x 18x 12
(30.5 x 45.7 x 30.5)
18 x 24 x 18
(45.7 X 61 X 45.7)
9x14~9
(22.9 X 35.6 X 22.9)
12x18~12
(30.5 x 45.7 x 30.5)
18 x 24 x 18
(45.7
X
61 X 45.7)
WARNING! This unit is not intended for use in processing hazardous work loads. Fatal injuries and property damage can result from processing combustible volatile fluids or materials which emit explosive vapors.
Voltage
120 VAC, 50160
Hz,
Single Phase
240
VAC,
50160
H~,
Single Phase
120 VAC, 50160
Hz,
Single Phase
240 VAC, 50160
Hz,
Single Phase
120 VAC, 50160
Hz,
Single Phase
240
VAC,
50160
H~,
Single Phase
120 VAC, 50160
Hz,
Single Phase
240
VAC,
50160
H~,
Singie Phase
120 VAC, 50160
Hz,
Single Phase
240 VAC, 50160
Hz,
Single Phase
120 VAC, 50160
Hz,
Single Phase
240
VAC,
50160
H~,
Single Phase
Do not use combustible gases in this oven.
Dimensions W x
F-B x H
Exterior
17.5 x 20 x 24.7
(44.5
X
50.8 X 62.7)
20.5 x 23 x 28.4
(52 x 58.4 x 72.1)
26.3
x
30.4 x 34.3
(66.8
X
77.2 X 87.1)
17.5 x 20 x 24.7
(44.5
X
50.8 X 62.7)
20.5 x 23 x 28.4
(52 x 58.4 x 72.1)
26.3 x 30.4 x 34.3
(66.8
X
77.2 X 87.1)
Avoid combustible products which generate toxic or hazardous vapor or fumes. Work should only be done in a properly vented environment.
Exterior
Enameled
Steel
Stainless
in.
(cm)
Shipping
(Approximate)
27.5 x 30 x 34.7
(69.8
X
76.2 X 88.1)
30.5 x 33 x 38.4 (77.5 x 84
x
98)
36x41
x45
(91.5 X 104.1 X 114.3)
27.5 x 30 x 34.7
(69.8
X
76.2 X 88.1)
30.5 x 33 x 38.4 (77.5 x 84 x 98)
27.5 x 30 x 34.7
(69.8
X
76.2 X 88.1)
Before using, user shall determine the suitability and
integrity of the product for the intended use and that the unit has not been altered in any way. User assumes all risk and liability whatsoever therewith.
Heater
Power
(kw)
1.05
.55
1.55
1.05
.55
1.55
Shipping
Weight
Ibs
(kg)
150
(68)
250
(114)
325
(148)
150
(68)
250
(114)
325
(1 48)
Page 7
Installation and Operation
Laboratory Vacuum Ovens
Figure
1.
Vacuum and Gas Connections
114" Gas
inlet
07
vacuum por~
114" Vent port
I
Vent Gas Vacuum
.
1
-
.
-
-
1
-
I
1
-
-
1
-
.
1
-
I
-
-
-
-
.
-
CAUTION!
This product contains fiberglass wool or
other refractories which can result in the following:
1" NPTvacuum port
-
May be irritating to skin, eyes, and respiratory tract. May be harmful if inhaled.
Possible cancer hazard based on tests with
laboratoly animals. Animal studies to date are
inconclusive. No human exposure studies with this product have been reported.
-
-
-
-
-
-
I
WARNING! Before maintaining this equipment, read the applicable MSDS (Material Safety Data Sheets) at the back of this manual.
Dl
.
-
1
I
-
-
1
1
.
1
-
-
Power cord
?,-;
-
.
Overtemperature reset button
WARNING! When installing, maintaining, or removing the refractory insulation. the following precautions will minimize airborne dust and fiber:
Keep personnel not involved in the installation out of the area. Use a good vacuum to clean area and equipment. Use a dust suppressant if sweeping is necessary. Do
not
use compressed air. Use disposable mask suitable for nuisance dust. Wear long sleeve clothing, gloves, hat, and eye protection to minimize skin and eye contact. Do not wear contact lenses. Thoroughly wash self after work is complete. Launder work clothing separate from other clothes and thoroughly clean laundering equipment after use. If clothing contains a large amount of dust
andlor fiber, dispose of rather than clean.
Promptly place used fiberglass parts and dust in plastic bags and dispose of properly.
Page 8
Note:
Since
the
front
panel
display
reads
gauge
pressure
the
value
is
relative
to the
local
atmospheric
pressure.
Use
in a location
with
reduced
atmospheric
pressure
will
result
in
the
gauge
displaying
diminished
vacuum
levels.
Page 9
Installation and Operation
Laboratory Vacuum Ovens
4
Start
Up
Table
2.
UT150 Parameter Functions
CAUTION! Observe the following precautions when operating the oven:
Wear protective eyeware. Wear protective gloves.
Use tongs to insert and remove oven load. Do not allow the load to touch the oven walls.
CAUTION! Do not attempt to operate the oven above the maximum rated temperature of
260°C.
Operation above
260°C
can cause damage to the
oven.
4.1
Oven Start Up
To start the oven, complete the following steps:
1.
Turn
on main power switch. Wait until the controller runs
initial diagnostics.
2.
Verify the vacuum pump connection (see Section 3.3 on
page
4).
3. Select the desired setpoints (refer to Section 5.3 on page 6 if you have a
UT150 controller, Section 6 if you have a UP150
controller).
5
Operation
-
UT150
Controller
The oven temperature controller is configured and tuned at the factory to function well for most applications. Occasionally, it may be advisable to configure the temperature controller differently to suit a particular working environment or process.
Table
3.
Pushbutton Keypad
CAUTION!
Before reconfiguring the controller, read
this chapter and the
"TI50
operation manual.
Reconfiguring the controller can change the unit
SC
dr
characteristics and design parameters, which can
hamper performance and make the equipment
dangerous to use.
Super Control
Directheverse action
This chapter provides brief instructions on how to perform the following configuration changes:
Setting the temperature Setting the Overtemperature Protection Temperature Changing between celsius and fahrenheit Setting the ramp to setpoint time
Autotuning the controller
Detailed instructions on configuring the temperature controller are found in the
UT150 operation manual.
5.1
Normal Controller Operation
The Temperature Controller senses the chamber air temperature
of the oven (the
PV,
or process value) and supplies the heat
necessary to achieve the desired setpoint. The controller includes
an
LED
display and a pushbutton keypad. Refer to Table 2 and
Table 3 for lists of displayed parameters and keypad functions.
Button
A
v
Function
Pressing and holding the SETIENT for three seconds advances the display to the Operation Parameters Menu.
While in the Operation Parameters Menu, use
SETIENT to move from one parameter to the next, and to register changes you have made in
setpoint and
parameter values. Holding SETIENT for three seconds exits either the
Operation or Setup Parameters menu. Use the Up Arrow button to increase the temperature
setpoint display and to change parameter values in the
Operation and Setup Parameter menus. Whenever you change the value of a
setpoint or parameter, the
decimal point flashes to remind you to register the changed value with SETIENT.
Use the Down Arrow button to decrease the temperature
setpoint display and to change parameter
values in the Operation and Setup Parameter menus.
Whenever you change the value of a
setpoint or
parameter, the decimal point flashes to remind you to
register the changed value with SETIENT.
Page 10
Installation and Operation
Laboratory Vacuum Ovens
5.2
Setting the Temperature
To set the temperature to the desired setpoint, complete the
following steps:
1. Press
a
or 7 until the desired setpoint is indicated on the
bottom line of the display.
2. Press
SETIENT
to register the new setpoint.
5.3
Setting the Overtemperature Protection (OTP) Temperature
The high limit alarm system with the temperature controller
disables the heater output. To set the alarm on the temperature controller (typically 5°C above the desired main temperature setpoint), complete the following steps:
1. Press and hold
SETENT
for 3 seconds, until
A1
is displayed
on the upper line.
2. Press
A
or V until the desired overtemperature limit
setpoint shows on the bottom line of the display.
3. Press
SETENT
to register the new overtemperature alarm
setpoint.
4. Press and hold
SETENT
for 3 seconds to return to the normal
display.
5.4
Changing Between Celsius and Fahrenheit
The controller is factory-set to operate with degrees Celsius. To
change the display modes and parameter settings to the Fahrenheit scale, you will need to change the Input Type parameter
In
and also the values of various scale-dependent parameters. If during this procedure the buttons are inactive for more than two minutes, the controller will return to the standard display.
To change from Celsius to Fahrenheit:
1. With the controller operating, access the Operating
Parameters menu by pressing and holding
SETIENT
for 3
seconds.
2. Press and release
SETIENT
repeatedly until the upper display
reads
LoC.
3. Press V until the displayed value of LoC is -1; then press
SETIENT
to access the Setup Parameters menu (refer to
Table 2 on page
5).
4. The first setup parameter displayed is Input 'Qpe (In). Press to change its value from
5
to 35. After making this adjustment (and all following parameter adjustments) be sure to press and release
SETIENT
again to register the change.
5. Press and release
SETIENT
to advance to the SPH parameter
and change its value to 500.
6. Press and release
SETIENT
to advance to the SPL parameter
and change its value to 32.
7.
Press and release
SETENT
to advance to the HY1 parameter
and change its value to
1.
8.
Press and hold
SETENT
for 3 seconds to exit the Setup
Parameters Menu.
9.
Press and hold
SETIENT
for 3 seconds to enter the Operating
Parameters Menu and display the
A1 parameter.
10.Use the Abutton to set the
A1
parameter to the desired
overtemperature limit in
OF.
11. Press and release
SETIENT
to advance to the P parameter and
change its value to 16.2.
12. Press and hold
SETIENT
for 3 seconds to exit the Operating
Parameters Menu.
13. The new temperature units are now effective. Follow the instructions in Section 5.2 to reset the temperature
setpoint in
"
E
14. Apply the OF label over the "C label on the control panel.
5.5
Setting the Ramp to Setpoint Rate
The Ramp Rate feature allows the chamber to be heated or cooled at any rate slower than the maximum capability of the unit. To fine tune ramp rates, you may need to test using loads with similar mass and thermal properties to loads you intend to use in oven applications.
To set the ramp to
setpoint time, complete the following steps. If
during this procedure the buttons are inactive for more than two minutes, the controller will return to the standard display.
1.
With the controller operating, press and hold
SETIENT
for
3
seconds to enter the Operating Parameters menu.
2. Press and release
SETIENT
until the LoC parameter is on the
upper display.
3. Press the down arrow button to show
'-I1, and press
SETIENT
once to enter the Setup Parameters menu.
4.
Press and release
SETIENT
until the Upr parameter is on the
upper display.
5. Press the arrow buttons to select the new Up Ramp Rate value, in
OC per minute or OF per minute, or 'OFF'. Press and
release
SETENT
to register the value change.
6. Press and release
SETIENT
until the dnr parameter is on the
upper display.
7.
Press the arrow buttons to select the new Down Ramp Rate value, in
OC per minute or OF per minute, or 'OFF'. Press and
release
SETIENT
to register the value change.
8.
Press and hold
SETENT
for three seconds to exit the Setup
Parameters menu.
9.
The new Ramp Rates are now effective.
10. Follow the instructions in Section 5.2 to reset the temperature setpoint.
Note:
The ramp rate begins when the
SET/ENT
button is
pressed after the target
setpoint is selected. The setpoint
display on the controller will show the changing setpoint at the selected ramp rate.
To view the target
setpoint during the ramp rate, press and release
an arrow button. The lower display will show the
setpoint next to
the selected target setpoint. Return to the ramping
setpoint
display by pressing and releasing the other arrow button.
Page 11
Installation and Operation
Laboratory Vacuum Ovens
5.6
Auto Tuning the Controller
Factory settings are provided for general purposes, but your process can be enhanced through the auto tune feature. For a given process temperature and product load, auto tuning maximizes the performance of the chamber by operating with the quickest response and minimal temperature overshoot.
To auto tune the controller.
1.
Load the chamber with materials that have the same mass and thermal characteristics as a typical product load.
2.
Operate the chamber to the process temperature.
3.
Press and hold
SETlENT
for 3 seconds to display the A1
parameter of the Operating Parameter menu.
4.
Press and release
SETIENT
to show the At parameter.
5.
Press and release the arrow buttons to show on in the lower display.
6.
Press
SETIENT
once to enter the auto tune mode and exit the
Operating Parameters menu.
The controller will cycle three times through a heating and cooling pattern, measuring the characteristics of the load and chamber temperature controls. During the auto tuning, At will alternately flash with the measured temperature
(PV)
to indicate
that the auto tuning is in progress. The length of time for the auto
tune varies with the load, chamber size and temperature selected.
The auto tune is completed when the regular display of the
measured temperature is shown. The chamber should now
operate to the process temperature with the given product load,
with the quickest response and minimal temperature overshoot.
If the process temperature or load changes significantly, another auto tune session may be necessary to optimize the chamber performance.
Page 12
Installation and Operation
Laboratory Vacuum Ovens
6
Operation - UP150 Controller
Table
4.
UP150 Parameter Functions
("P"
Models only)
RUN
m
m
HOLD
Table 5. Pushbutton Keypad
Figure
2. UP150
Control Panel
parameter
'Ode
The oven temperature controller is configured and tuned at the factory to function well for most applications. Occasionally, it
may be advisable to configure the temperature controller
differently to suit a particular working environment or process.
CAUTION1
Before reconfiguring the controller, read
this chapter carefully.
Reconfiguring the controller can
change the unit characteristics and design
parameters, which can hamper performance and
make the equipment dangerous to use.
Fag:'y
Value
For more detailed instructions, refer to the Yokogawa UP150 manual.
Description
Operating parameters (access
by
holding the SETENT key)
6.1
Normal Controller Operation
HoLd AdV CtL
At
P
I
d
Ct FL
bS LoC
The Temperature Controller senses the chamber air temperature
of the oven (the
PV,
or process value) and supplies the heat necessary to achieve the desired setpoint. The controller includes an
LED
display and a pushbutton keypad. Refer to Table 2 and
Table
3
for lists of displayed parameters and keypad functions.
The UP150 controller will accept a single program of up to
16
segments. This controller includes an automatic tuning feature
and
"
Super" control to improve the performance. Refer to the
UP150 manual for detailed information.
OFF OFF Pid
OFF
32.4 1384 346
1
OFF
0.0 0
Program Hold Segment Advance Control mode Auto tuning Proportional band
QF=ll)
Integral time Derivative time Heat cycle time Sensor filter PV bias (offset) Key lock
Setup parameters (access
by
setting LoC=-1)
In
SC
Button
@
a
v
Function
Pressing and holding the SETIENT for three seconds advances the display to the Operation Parameters Menu.
While in the Operation Parameters Menu, use SETIENT to move from one parameter to the next, and to register changes you have made in
setpoint and
parameter values. Holding SETIENT for three seconds exits either the
Operation or Setup Parameters menu. Use the Up Arrow button to increase the temperature
setpoint display and to change parametervalues in the
Operation and Setup Parameter menus. Whenever you change the value of a
setpoint or parameter, the
decimal point flashes to remind you to register the changed value with SETIENT.
While in operating mode, pressing this key stops (resets) program operation.
Use the Down Arrow button to decrease the temperature
setpoint display and to change parameter
values in the Operation and Setup Parameter menus. Whenever you change the value
of
a setpoint or parameter, the decimal point flashes to remind you to register the changed value with SETIENT.
While in operating mode, pressing this key starts (runs) a program
5 ON
Input type
(J
thermocouple in *C; *F=35)
Super function
Page 13
Installation and Operation
Laboratory Vacuum Ovens
6.2
Single Setpoint Operation
The following sections describe how to operate the controller in single
setpoint (local) mode. Use this mode when you only need
to run the oven with a specific
setpoint and do not require a
programmed sequence of steps.
6.2.1
Setting High Temperature Alarm Setpoint:
1. Press and HOLD for three seconds the 'SETENT' button to display
"modE rES".
2.
Press and release the 'SETENT' button to display "PrG
0".
3.
Press the 'UPIRESET' Button to show the lower display value of
"1"
.
4.
Press and release the 'SETENT' button to select this new value and advance to the
"
SSP
25
"
display.
5.
Press and release the 'SETENT" button until the High Temperature Alarm
Setpoint value is displayed as
"Al"
.
6.
Select an alarm setpoint 10°C above the target setpoint to be selected.
7.
Press and release the 'SETIENT' button to place this new value in the controller memory.
8.
Press and HOLD for three seconds the 'SETIENT' button to exit this menu.
6.2.2
Accessing Local Mode
1.
Press and hold for three seconds the 'SETIENT' button to display
"modE rES".
2.
Press and release the 'UP' button twice to select the display
"modE LCL,".
3.
Press and release the 'SETENT' button once to select Local Mode. This selection causes the red indicator to illuminate beside
"L"
on the control panel.
4.
Use the
'UP'
and 'DOWN' buttons to select the desired
operating temperature setpoint.
5.
Press and release the 'SETENT' button once to register the
setpoint value.
6.
The display will then show measured temperature in the upper display, the present temperature
setpoint in the lower
display.
7.
This display and the buttons will remain active as long power continues to the control module. Power interruptions will cause the controller to enter reset or standby mode in which no actions are made to operate the heating equipment.
You may use the arrow buttons to adjust the
setpoint (lower)
value to be adjusted in this display mode. The
'SETJENT' button
will register
setpoint value changes, until these values are
changed again.
6.2.3
Exiting Local Mode
To exit Single Setpoint or Local Mode and turn off the energy to the heaters:
1.
Press and hold for three seconds the 'SETENT' button to display
"modE LCL".
2.
Press and release the 'DOWN' button twice to select the display
"modE rES".
3.
Press and release the 'SETIENT' button once to select the Reset Mode. This selection causes the red indicator to extinguish beside the display label
"L"
that had indicated the Local Mode.
4.
This will change the display showing the measured temperature in the upper display, with the lower display showing the Start Set Point (SSP) temperature
setpoint of the
program.
6.3
Programming Operation: Entering a Program
This section describes how to enter a simple program that is designed to:
direct the controller to ramp to a higher temperature; stabilize; ramp to a lower-temperature; end with
an
indefinite dwell.
If you intend to use the program features of the controller, it is advisable to go through all the steps in this sample program to familiarize yourself with the elements of programming mode.
Note:
If
the controller buttons are NOTpushed for 2 minutes, the controller will return to the regular operator
mode/menu.
6.3.1
Entering Programming Mode
To access the programming menu:
1.
Make sure the indicators beside "RUN and
"L"
on the controller face are off. If either indicator is on, press and hold the 'SETENT' buttonuntil the display shows
'modE'. Select
'rES' in the lower display with the 'arrow' buttons. Press and
release the 'SETENT' button once.
2.
Press the 'SETENT' button for 3 seconds to display "modE in the upper display and "rES" Reset) in the lower display.
3.
Press and release 'SETENT' until "LoC" is displayed. Make sure the display below
"LoC" is
"0"
(zero). If it is not
"0",
use 'DOWN ARROW' to select
"0"
and press and release
"
SETENT' button to register the change to
"0".
4.
Press and release the "SETENT' button until "PrG" is displayed.
5.
At "PrG display, press the 'UP ARROW' to make the lower display
"1"
.
6.
Press and release the 'SETIENT' button once to enter the programming menu.
Page 14
Installation and Operation
Laboratory
Vacuum
Ovens
6.3.2
Entering Program Parameters
The first display is the Start Set Point parameter, shown as "SSP
in the upper display. The value assigned to SSP is usually the
current room temperature,
25°C.
On the next page is
an
illustration of the program profile and a
table of the parameters entered.
Basic Ramp and
Dwell
Parameters:
1. Use the arrow buttons to select "25" in the lower display, then press and release the 'SETENT' button twice to enter this new value and to advance to the
"StC" display.
If the value for
"
SSP" is correct and does not need to be changed, press and release the 'SETENT' button once to advance to the
"StC" display.
2.
Next is the Start Code parameter, shown as "StC" in the
upper display. The value assigned to
StC is usually
"0".
This
will instruct the program to follow the Start Set Point. Press
the SETENT button to advance to the next display.
3. The next parameter,
"SPl", is the first setpoint value that is
desired in the chamber and is normally a ramp segment. Select this target temperature
setpoint value with the arrow
buttons then press and release the
'SETIENT' button twice to
enter this value and to advance to the
"tM1" display. If the
value for
"SP1" is correct and will not be changed, press and
release the
'SET/ENT' button once to advance to the "tM1"
display.
4. The next parameter,
"tMl", represents the first time period
for the unit to reach the target temperature
setpoint selected
in
"SP1". This selection can be a value ranging from 0.00 to
99.59, which represents hours and minutes. Select this time value with the arrow buttons and enter it by pressing and releasing the 'SETENT' button twice.
5. Press and release the 'SETENT' button to advance to the next display of
"SP2", this is normally the dwell segment.
Select the same target
setpoint temperature value as "SP1"
with the arrow buttons. Press and release the 'SETIENT' button twice to enter this value and to advance to the next display.
6.
Thenext parameter, "tM2", represents the second time period
used to maintain or dwell at the target
setpoint selected in
"
SPY. This selection can be a value ranging from 0.00 to
99.59, which represents hours and minutes. Select this time value with the arrow buttons and enter it by pressing and releasing the 'SETENT' button twice.
7.
Next, "SP3" is the third setpoint value desired in the chamber. Select this target temperature
setpoint with the
arrow buttons and press and release the
'SET/ENT' button
twice to enter this value and to advance to the
"tM3" display.
If this value is correct and not changed, press and release the 'SETENT' button once to advance to the
"tM3" display.
8.
"tM3" represents the third time period for the unit to reach
the target
setpoint selected in "SP3". This selection can be a
value ranging from 0.00 to 99.59, which represents hours and minutes. Select this value with the arrow buttons and enter it by pressing and releasing the
'SETIENT' button.
9. The next parameter,
"SP4" is normally the dwell segment.
Select the same target temperature as
"SP3" with the arrow
buttons, then press and release the 'SETENT' button twice to enter this new value and to advance to the next display.
Additional Program Parameters
"
10.The next parameter displayed is "tM4". Select a lower
display value of
"OFF"
with the 'arrow' buttons, then press and release the 'SETENT' button twice to enter this value change and advance to the next display.
11 .The next display shows
"EV1" in the upper display. The
lower value should always be
"0"
(zero). Press and release
the
'SETIENT' button once to go to the next display.
12."ALlW should always have a lower value of
"9"
.
Press and
release the
'SETIENT' button once to advance to the next
display.
13.The next parameter, "Al", is used to select the high
temperature alarm trip setpoint. Use the 'arrow' buttons to select a value 10°C (or 20°F) HIGHER than the
highest
target setpoint to be used. Select the High Temperature Alarm value with the arrow buttons then press and release the
'SETIENT' button twice to enter this new value and to
advance to the
"
HY 1"
display. If the value for "Al" is correct
and not changed, press and release the
'SETIENT' button
once to advance to the
"HY1" display.
14."HY17' is used to select the amount of temperature change
below the high temperature alarm
setpoint where the alarm
relay will reset. This value is usually
"1". Select "1" with the
arrow buttons and press the
'SET/ENT' button six times to
enter the correct value and advance to the
"JC"
display. Or if
the value is correct, press the
'SETLENT' button five times to
advance to the
"JC"
display.
15.For the parameter displayed as
"JC",
select "1" with the arrow buttons, then press and release the 'SETENT' button twice to display
"WTZ".
Selecting the value of "1" will cause
the program to hold the
setpoint at this last segment. A value
of
"0"
would cause the program to reset and stop running the
program and stop the power to the heaters. A value of
"2"
will cause the program to repeat 'continuously'.
16.When the display shows
"WTZ",
select a lower display
value of
"OFF" with the 'arrow' buttons. Press and HOLD the
'SETENT' button for 3 seconds to return to the Reset or standby display.
This concludes the steps required to enter a typical ramp-and dwell program.
Page 15
Installation and Operation
Laboratory Vacuum
Ovens
6.3.3 Running a Program
6.4
Auto Tuning the
UP150
Controller
To run a program such-as the one outlined above, press and hold the
'DOWNIRUN' button making the 'RUN' indicator
illuminate. At the end of this program the 'HLD' (hold) indicator is illuminated to indicate this program is in the indefinite dwell at the last target temperature. This hold indicator is caused by the
"JC"
selection of "I", while the
"JC"
selections of
"0"
or
"2"
will
not illuminate the 'HLD' (hold) indicator.
6.3.4 Ending a Program
To end a program while in the 'RUN' or 'HLD" (hold) mode,
press and hold the
'UPIRESET' button to turn off the current
program and extinguish the 'RUN' or 'HLD' indicator.
Turning off the unit's power will also stop the program. When
power is restored, the controller is in the Reset or standby mode with no power to the heaters.
Note: The programmer/controller will not operate the unit's
heaters (to change or maintain a temperature) unless there is a program running or a single setpoint value is selected in the Local Mode.
6.3.5 Changing a Program
To make changes ONLY to the target temperature and segment length times for simple program operation, follow these steps:
1. Make sure the indicators beside
"
RUN" and
"L,"
on the
controller face are off. If either indicator is on, press and hold
the 'SETENT' button until the display shows
'modE'. Select
'rES' in the lower display with the 'arrow' buttons. Press and
release the 'SETENT' button once.
2.
Press the 'SETIENT' button for 3 seconds to display "modE in the upper display and "rES" in the lower display. Press and release 'SETENT' repeatedly to display
"LoC". Make sure
the value below
"LoC"
is
"0"
(zero). If it is not
"O",
use
'DOWN' arrow to select
"0"
and press and release
"
SETENT" button to register the change to
"0".
3.
Press and release the 'SETJENT' button once to show "PrG" on the upper display.
4.
Press the
'UP'
arrow to make the lower value "1".
5.
Press and release 'SETIENT' button twice to display "SP1". Using the arrow buttons to revise the target setpoint.
6.
Press and release 'SETENT' button twice to display "tml". Using the arrow buttons to revise the segment time length needed to get to the target
setpoint 'SP1'.
7.
Press and release 'SETENT' button to display other setpoints and segment time lengths. Use the arrow buttons to change the temperature setpoints and time lengths. Press and release the
'SETIENT' button to register any new values.
8.
Press and HOLD the 'SETENT' button for 3 seconds to exit the program menu and return to the reset or standby display.
Auto tuning maximizes the performance of the chamber at a selected temperature with the product load's characteristics, by operating with the quickest response and minimal temperature overshoot.
Factory settings are for general purposes, but your process can be enhanced through the auto tune feature. To obtain this maximum performance, follow these steps to auto tune the controller.
1. Load the chamber with materials that have the same mass and thermal characteristics as an actual product load.
2.
Operate the chamber to the process temperature.
3.
Start the Auto Tune: Press and hold the 'SETtENT' button for three seconds to display the
"modE" parameter of the
Operating Parameter menu.
4.
Press and release the 'SETENT' button five times to advance to the
"At"
parameter.
5.
Press and release the
'UP'
arrow button to show "on" in the
lower display.
6.
Press the 'SETtENT' button once to enter the auto tune mode and exit the Operating Parameters menu.
7.
The controller will cycle three times through a heating and
cooling pattern, measuring the characteristics of the load and chamber temperature controls. During the auto tuning, 'At' will alternately flash with the measured temperature (PV) display to indicate that the auto tuning is in progress. The
length of time for the auto tune varies with the load, chamber size and temperature selected.
8.
The auto tune is completed when the regular display of the
measured temperature is shown without the
"At"
value
flashing. The chamber should now operate to the process
temperature with the given product load, with the quickest
response and minimal temperature overshoot.
9.
If the process temperature or load changes significantly, another auto tune session may be necessary to optimize the chamber performance.
To
interrupt the auto tune before it is completed, simply turn off
the power to the controller and unit. When the power is restored the auto tune will not be operating.
Page 16
Installation and Operation
Laboratory Vacuum Ovens
6.5
Temperature Offset Procedure
The purpose of this procedure is to create an offset in the
displayed temperature measurement for the Yokogawa model UP150 temperature controller.
1. Operate the oven chamber to your normal stable temperature setpoint, with an independent temperature measurement device located in the center of the chamber. The controller will be 'running' the program or operating in the local mode to maintain the temperature.
2. Note any difference in the controller's measured temperature (upper value) and the independent measurement. If a difference of greater than 1°C is noted proceed with the following steps.
3. Press and hold the
"
SETENT" button for 3 seconds to
display
"modE".
4. Verify the button lockout parameter will give access to make this display offset. Press and release the
"
SETENT" button
twelve times to display
"LoC". The value 0 (zero) displayed
will give full access and is necessary to make the display offset changes desired. If the value displayed is 1 or 2, use the
"
down arrow" button to make 0 (zero) and press and
release the
"
SETENT" button to register this change.
5. Press and release the
"
SETENT" button twelve times to
display
"bS" and the current offset value.
6. Select the offset value with the arrow buttons that is needed to make this controller display correctly. For example, if the independent measurement is
253"C, the
controller Temperature measurement display shows
250°C,
and the current controller offset (bS) is -2, then make the controller display offset
"+I"
[(+3 needed offset) + (-2 current offset) = (+1 new offset)].
7.
Press and release the "SETENT" button once to register this new offset value. Press and hold the
"
SETENT" button for 3
seconds to exit this controller menu.
8. Operate the controller to the same temperature to stabilize the chamber to check for any further variations between the controller and the independent measurement. Repeat steps 2
-
7
as necessary.
9.
This completes the display offset procedure for the Yokogawa model UP150 temperature controller. If the button lockout parameter
"LOP was originally on a value of 1 or 2,
repeat steps 3
&
4 to return to this original value.
Contact Technical Service at 1
-
800-438-4851 if you have any
questions.
6.6
Changing Temperature Scale Between
'C
and 'F
To change the temperature scale in the UP150 controller to operate on
OF
instead of the factory setting of
"C,
or from
OF
to
"C, follow these steps.
These changes will alter the controller input type and associated scale
-
dependant parameters,
AND
ERASE
the storedprogram
to default values.
Please document the stored program in the
controller
BEFORE
proceeding.
1.
Make sure the indicators beside
"
RUN
and
"L"
on the
controller face are off. If they are on, press and hold the
'UP/RESET' button until the RUN or L indicators are off.
2. To access the Operating Parameters menu, press and HOLD the 'SETENT' button for at least 3 seconds to display
"modE".
3. Press and release the 'SETENT' button until the display shows
"
LOC" in the upper display. Make sure the value
below
"LoC" is
"0"
(zero). If it is not
"0"
use 'down arrow'
to make
"0"
and press and release "SETENT button to
register change to
"0".
4. At "LoC" display, press the 'down arrow' to make the lower value
"-1".
5. Press and release the
"
SETENT' button to enter the Setup
Parameters menu and show
"In"
on the upper display and a
numerical value in the lower display.
6. See table below for the STANDARD values for this parameter and the others needed in the following steps.
7.
Select the appropriate value for the "In" parameter. Press the 'UP' or 'DOWN' arrow buttons to make the lower display to the new value, then press and release the 'SETENT' button TWICE to register the new value and advance to the next parameter.
8.
"
SPH" is the next parameter displayed. Select and enter the new value, then press and release the 'SETENT' button TWICE.
9.
"SPL" is the next parameter displayed. Select and enter the
new value, then press and release the 'SETENT' button ONCE.
10.Press and HOLD the 'SETENT' button for at least 3
seconds to exit.
11.Press and HOLD the 'SETtENT' button for at least 3
seconds to enter the Operating Parameter menu and show
"modE" in the upper display.
12.Press and release the 'SETENT' button until the upper
display shows
"P". Select the value in the table and adjust the
lower display accordingly. Press and release the
'SETtENT'
button TWICE.
13."I" is the next parameter displayed. Select and enter the new
value, then press and release the 'SETENT' button TWICE.
14,"d" is the next parameter displayed. Select and enter the new
value, then press and release the 'SETENT' button ONCE.
15.Press and HOLD the 'SETENT' button for at least 3
seconds to exit.
16.Reenter or create a program using the new temperature
scale.
The P, I and
D
parameters may be altered through auto tuning
(refer to Section 6.4 on page 11).
If during this procedure the buttons are inactive for more than two minutes, the controller will return to the standard display.
Page 17
Installation and Operation
Laboratory Vacuum Ovens
7
Maintenance
7.1
General Maintenance
CAUTION!
Maintenance should only be performed
by trained personnel.
WARNING!
Disconnect oven from main power
A
before attempting any maintenance to oven or its
controls.
CAUTION!
This product contains fiberglass wool or
other refractories which can result in the following:
May be Irritating to skin, eyes, and respiratoty tract. May be harmful if inhaled. Possible cancer hazard based on tests with
laboratory animals.
Animal studies to date are
inconclusive. No human exposure studies with this product have been reported.
WARNING!
Before maintaining this equipment, read
A
the applicable MSDS (Material Safety Data Sheets)
provided with this unit.
WARNING!
When installing, maintaining, or
A
removing the refractory insulation, the foliowing precautions will minimize airborne dust and fiber:
Keep personnel not involved in the installation out of the area. Use
a
good vacuum to clean area and equipment.
Use
a
dust suppressant if sweeping is necessaly
Do
not
use compressed air.
Use
disposable
mask suitable for nuisance dust.
Wear
long sleeve clothing, gloves, hat, and eye
protection to minimize skin and eye contact. Do not
wear contact lenses. Thoroughly wash self after work is complete. Launder work clothing separate from other clothes and thoroughly
cleanlaundering equipment after
use. If
clothina contains a larae amount of dust andlor fiber, $spose of ratheFthan clean.
Promptly place used fiberglass parts and dust in
plastic bags and dispose of properly.
WARNING!
Disconnect oven from main power
A
before attempting any maintenance to oven or its controls.
Regular maintenance is required to keep the oven running at
optimum levels.
1.
Clean the oven interior as necessary with acetone, alcohol, or ether.
WARNING!
When using cleaning materials, follow
A
ail precautions iisted on the cleaning containers. Always use cleaning materials
In well ventilated areas
as inadequate ventilation can be fatal.
Note:
Frequency of cleaning depends on the oven workload.
If
the oven becomes contaminated, the ultimate vacuum level will not be attained.
2.
Check hose connections for leaks weekly.
3.
Check the vacuum pump oil level.
4.
Check the condition of the vacuum pump oil. Replace the oil
if
it is contaminated.
7.2
Gasket Maintenance
WARNING1
Disconnect oven from main power
A
before attempting any maintenance to oven or its controls.
The door gasket is designed for easy removal and installation. This gasket may come off when the door
is
opened after high
temperature operation or processing of workloads containing
resin. To prolong gasket usage, clean the oven to remove any
resin.
7.3
Door Alignment
A
WARNING!
Disconnect oven from main oower before attempting any maintenance to oveh or its controls.
The oven may not pump down properly
if
the door is out of alignment. To realign the oven door, complete the following steps:
1.
Lay the oven on its back.
2.
Loosen top and bottom hinges.
3.
Square the door. Check that the glass comes
in
contact with
the gasket.
4.
Place a small amount of pressure (approximately one pound)
on the hinges.
Note:
This pressure is required so that the door seals when closed. The
glassfloats on springs to achieve a complete
seal.
5.
Tighten the hinges.
6.
Pump the oven down.
Page 18
8
Troubleshooting
WARNING!
Troubleshooting procedures involve working with high voltages which can cause injury or death.
Troubleshooting should only be petformed by trained personnel.
This section
is
a
guide to troubleshooting oven problems. Refer to Table 6 for troubleshooting procedures.
Table
6.
Troubleshooting
Problem
Oven shelves warp and discolor. Door gasket comes off.
System ultimate pressure is high.
No vacuum in oven.
Pump stalls.
Long drying time.
Probable Causes
Oven temperature is above
260°C.
Gasket is damaged from high temperature operation or from resin damage.
Oven is contaminated with high vapor pressure material.
Pump oil is contaminated.
Leaks in vacuum piping or ports.
Vacuum pump seal is worn. Discharge valve is worn. Internal parts are worn. Vacuum vent valve is opened. Oven does not pump down properly due to
misaligned door. Power to pump is off. Use of extension cord results in insufficient
amperage. Pump is damaged. The discharge line is blocked. No pump oil or lost viscosity (overheating). The piping is too narrow.
The
distance
between
the
pump
and
the
oven is too large.
Solution
Do not operate the oven above
260%.
Do not Operate the oven above 260°C.
Clean oven to remove resin. Clean the oven. Refer to Section 7.1 on page
13.
Change the oil. Use inlet vapor trap andlor gas ballast. Check for leaks in vacuum line, oven port, door. Replace shaft seal. Replace valve. Install new vanes. Close vent valve. Realign the door. Refer to Section
7.3
on page
13
for
realignment information. Check the power switch and outlet. Move the oven closer to power source and do not use extension
cord. Return the pump to the factory for repair. Change the demister and clear the blockage. Change the oil. Install piping the same size as the pump inlet.
Move the pump closer to the oven or use a larger pump.
Page 19
Installation and Operation
Laboratory Vacuum Ovens
9
Replacement Parts
All
quantities are one each unless noted.
Table
7.
Replacement Parts
Silicone Gasket Door Pull Handle, SA and SC
Units
Pull Handle, A and C Units Door Magnet Catch Levelling Feet Aluminum Full Shelf Aluminum Half Shelf
11 8975 380231-101 38024H01 38025H01
(4) 38026H01
38028H01 38028H04
Optional Gaskets
38027H01 38023H01 38024H01
38025H01
(4) 38026H01 (2) 38028H02
38028H05
Buna-N Gasket Viton Gasket
34636H01 38023H01 38024H01
38025H01
(4) 38026H01 (3) 38028H03
-
11 8974 (9 x 9)
118973 (9
x
9)
118977 (12
X
12)
118976 (12 X 12)
34637H01 (1 8 x 18)
34638H01 (1 8 x 18)
Page 20
Over Tenperatwe Reset Thernostot
103508
-
T
NEMA
5-15?
UP150
YOKOGAWA
UT150
CONTROL
NOT
TO
SCALE
"
"ON
'g05008
I
IRE
1
"""'"
Reference
-
-
V0914 - 1218 - 1824
al
l
a
dlrlmTE
5/18/99
T~ERANCE
UNLES
S OTKRWISE SPKIFIEO
1M
VAC
Eb
RA~IU~
.03zn
UNLESS OTHERWISE MTEO
.
XX
=
f
,032
.XXX
=
f
,015
I20
VAC
M914 240
VAC
HI
L
E
DIA.
f
,005
l
MVAC
TP ELEHENT
Page 21
WEEE Compliance
WEEE Compliance. This products is required to comply with the
European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96EC. It is marked with the following symbol. Thermo Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this products should be disposed of or recycling through them. Further information on Thermo Scientific’s compliance with these Directives, the recyclers in your country, and information on Thermo Scientific products which may assist the detec­tion of substances subject to the RoHS Directive are available at
www.thermo.com/WEEERoHS
WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical &
Electronic Equipment (WEEE) Richtlinie 2002/96EC erfüllen. Das
Produkt ist durch folgendes Symbol gekennzeichnet. Thermo Scientific hat Vereinbarungen getroffen mit Verwertungs-/Entsorgungsanlagen in allen EU-Mitgliederstaaten und dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden. Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo Scientific, die Verwerter und Hinweise die Ihnen nützlich sein können, die Thermo Scientific Produkte zu identizfizieren, die unter diese RoHS Anweisung fallen, finden Sie unter www.thermo.com/WEEERoHS
Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’Unione Europea 2002/96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici (WEEE). È marcato col seguente simbolo. Thermo Scientific ha stipulato contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli Stati Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste medesime. Ulteriori informazioni sulla conformità di Thermo Scientific con queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informa­zioni sui prodotti Thermo Scientific che possono essere utili alla
rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili
sul sito www.thermo.com/WEEERoHS
Great Britain
Deutschland
Italia
Conformité WEEE. Ce produit doit être conforme à la directive euro-
péenne (2002/96EC) des Déchets d’Equipements Electriques et Electroniques (DEEE). Il est merqué par le symbole suivant. Thermo Scientific s’est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d’informations sur la conformité de Thermo Scientific à ces directives, les recycleurs dans votre pays et les informations sur les produits Thermo Scientific qui peuvent aider le détection des substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS
France
Page 22
Important
For your future reference and then contacting the factory, please have the following information readily available:
Model Number:
___________________________
Serial Number:
___________________________
Date Purchased: __________________
_________
The above information can be found on the
data plate attached to the equipment. If available, please provide the date purchased, the source of purchase (manufacturer or specific agent/rep organization), and purchase order number.
IF YOU NE
ED ASSISTANCE
:
SALES DIVISION
Phone:
1-
866-984-3766
1-
866
-9-
THERMO
LABORATORY PARTS and SERVICE
Phone:
1-
800-438-4851
TECHNICAL SUPPORT
Phone:
1-
800-438-4851
Page 23
Page 24
www.thermoscientific.com
Loading...