Thermo Scientific INNOVA Series Instruction Manual

Page 1
INNOVA Series
Instruction Manual
Part Number 71-6417 5Aug2009
Page 2
© 2007 Thermo Fisher Scientific Inc. All rights reserved.
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin, MA 02038 1-508-520-0430 www.thermo.com/aqi
Page 3
WARNING
THIS INSTRUMENT IS DESIGNED TO DETECT ONE OR MORE OF
THE
FOLLOWING:
FLAMMABLE VAPORS, OXYGEN CONTENT, AND/OR TOXIC GAS AND TO GIVE
WARNING BEFORE THEY REACH HARMFUL CONDITIONS. IN ORDER
TO
ENSURE THAT IT WILL WARN OF DANGEROUS CONCENTRATIONS, IT
IS
ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL, PARTICULARLY
THOSE MAINTENANCE
CONCERNING START UP, OPERATION, CALIBRATION, AND
, BE READ, UNDERSTOOD, AND FOLLOWED.
NOTATION CONVENTIONS
Notices are used in this operator's manual to alert you to hazardous condi­tions to person or instrument and to notify you of additional information. This operator's manual uses the following notices.
WARNING
Notifies you of potential danger that can result in personal injury or death.
CAUTION
Notifies you of potential damage to equipment.
NOTE
Notifies you of additional or critical information.
Page 4
SERVICE LOCATIONS
For additional assistance, service is available from exclusive distributors worldwide. Contact one of the phone numbers below for product support and technical information or visit us on the web at www.thermo.com/aqi.
1-866-282-0430 Toll Free
1-508-520-0430 International
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Fisher Scientific’s compliance with these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at: www.thermo.com/WEEERoHS.
Page 5
Warranty
Seller warrants that the Products will operate or perform substantially in conformance with Seller's published specifications and be free from defects in material and workmanship, when subjected to normal, proper and intended usage by properly trained personnel, for the period of time set forth in the product documentation, published specifications or package inserts. If a period of time is not specified in Seller’s product documentation, published specifications or package inserts, the warranty period shall be one (1) year from the date of shipment to Buyer for equipment and ninety (90) days for all other products (the "Warranty Period"). Seller agrees during the Warranty Period, to repair or replace, at Seller's option, defective Products so as to cause the same to operate in substantial conformance with said published specifications; provided that (a) Buyer shall promptly notify Seller in writing upon the discovery of any defect, which notice shall include the product model and serial number (if applicable) and details of the warranty claim; (b) after Seller’s review, Seller will provide Buyer with service data and/or a Return Material Authorization (“RMA”), which may include biohazard decontamination procedures and other product-specific handling instructions; and (c) then, if applicable, Buyer may return the defective Products to Seller with all costs prepaid by Buyer. Replacement parts may be new or refurbished, at the election of Seller. All replaced parts shall become the property of Seller. Shipment to Buyer of repaired or replacement Products shall be made in accordance with the Delivery provisions of the Seller’s Terms and Conditions of Sale. Consumables, including but not limited to lamps, fuses, batteries, bulbs and other such expendable items, are expressly excluded from the warranty under this warranty.
Thermo Fisher Scientific Warranty
Notwithstanding the foregoing, Products supplied by Seller that are obtained by Seller from an original manufacturer or third party supplier are not warranted by Seller, but Seller agrees to assign to Buyer any warranty rights in such Product that Seller may have from the original manufacturer or third party supplier, to the extent such assignment is allowed by such original manufacturer or third party supplier.
In no event shall Seller have any obligation to make repairs, replacements or corrections required, in whole or in part, as the result of (i) normal wear and tear, (ii) accident, disaster or event of force majeure, (iii) misuse, fault or negligence of or by Buyer, (iv) use of the Products in a manner for which they were not designed, (v) causes external to the Products such as, but not limited to, power failure or electrical power surges, (vi) improper storage
Page 6
and handling of the Products or (vii) use of the Products in combination with equipment or software not supplied by Seller. If Seller determines that Products for which Buyer has requested warranty services are not covered by the warranty hereunder, Buyer shall pay or reimburse Seller for all costs of investigating and responding to such request at Seller's then prevailing time and materials rates. If Seller provides repair services or replacement parts that are not covered by the warranty provided in this warranty, Buyer shall pay Seller therefor at Seller's then prevailing time and materials rates. ANY INSTALLATION, MAINTENANCE, REPAIR, SERVICE, RELOCATION OR ALTERATION TO OR OF, OR OTHER TAMPERING WITH, THE PRODUCTS PERFORMED BY ANY PERSON OR ENTITY OTHER THAN SELLER WITHOUT SELLER'S PRIOR WRITTEN APPROVAL, OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER, SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS.
THE OBLIGATIONS CREATED BY THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A DEFECTIVE PRODUCT. EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT, SELLER DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT.
Thermo Fisher Scientific Warranty
Page 7
TABLE OF CONTENTS
Chapter 1
Introduction
About the Innova Series. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Sensor Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 2
Start Up & Operation
Preparing for Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Starting Up the Innova . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Turninging Off the Innova . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Performing a Bump Test on Power-Up . . . . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 3
Alarms
Gas Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Battery Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Sensor Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Pump Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Datalog Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Temperature Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Chapter 4
User Program
About the User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Display Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Accessing Password-Protected Menus. . . . . . . . . . . . . . . . . . . . . . . . . 19
Advanced Features Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Datalog Features Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
71-6417 vii
Page 8
Innova Series Operator’s Manual
Chapter 5
Calibration
Preparing the Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Calibrating the Innova . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Calibration Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Chapter 6
Maintenance
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Trouble Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Recharging Ni-Cd Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Replacing Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Appendix A
Parts List and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Appendix B
Interference Factors
High Concentration of Combustible Gas . . . . . . . . . . . . . . . . . . . . . . . 57
Relative Combustible Response Table . . . . . . . . . . . . . . . . . . . . . . . . 59
Appendix C
Supplemental Instructions
Innova Land Surveyor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Innova Transformer Gas Tester . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Soil and Fuel Vapor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Calibrating the Innova SV and FV . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Innova Soil and Fuel Vapor Monitors Specific Parts List . . . . . . . . . . 75
viii 71-6417
Page 9
Chapter
1
INTRODUCTION
About the Innova Series
The Innova Series (shown in Figure 1-1) is a line of portable gas monitors capable of detecting from one to four gases. It continuously computes average readings for Short Term Exposure Limit (STEL) and Time Weighted Average (TWA) levels. You can display Peak readings, STEL, and TWA levels on command. The Alarm circuit alerts you to dangerous gas conditions, low battery conditions, sensor failure, datalog full, calibration reminder, and sample-draw pump failures.
Figure 1-1 Innova Series Gas Monitor
The Innova is powered by four “D” size alkaline or nickel-cadmium (NiCd) batteries. A jack is provided on the Innova so you can connect the external NiCd battery charger.
71-6417 1
Page 10
Innova Series Operator’s Manual
Specifications
Table 1-1 lists specifications for the Innova Series gas monitor.
Table 1-1 Specifications for the Innova Series Gas Monitor
Intrinsic Safety Rating (Area Classification)
Sampling Method Sample-drawing
Response Time Initial: 5 seconds average (with 5-foot hose)
Operating Temperature Storage Temperature
Humidity 0 to 95% relative humidity (RH), non-condensing
Regulatory Approvals UL Classified; CSA Certified, Cenelec
Alarms Audible/visible, coded for gas and trouble. Also a
Alarm Actions Low flow; low battery; rising gas reading (rising or
Alarm Functions Alarm levels are user-selectable in the Edit Alarms
Display Digital liquid crystal display (LCD). Displays up to
Power Source Four “D” size alkaline or NiCd batteries.
Battery life Alkaline - 24 hours @ 68° F(20° C)
Controls AIR, ON/OFF (power), and RANGE buttons.
Dimensions 10 in. L x 5 in. W x 6 in. H
Weight 5 pounds (2.25 kg)
Case High-impact, chemical resistant polycarbonate-
Standard Accessories
1
Innova models for transformer testing include the following standard accessories: dilution fitting, and gas collection bag.
1
Class I, Division 1, Groups A, B, C, and D
90% complete within 30 seconds (except NH 90% complete within 150 seconds (NH
)
3
)
3
-4° F to 113° F (-20° C to 45° C)
-22° F to 149° F (-30° C to 65° C)
comfort beep that can be turned off.
falling gas readings for oxygen); rising TWA and STEL reading (toxic versions only); sensor failure; temperature; datalog.
Screen.
four different gases at a time. A back light is available on demand by momentarily pressing any button.
Ni-Cd - 12 hours @ 68° F(20° C) Due to the nature of alkaline cells, battery life is greatly reduced at low temperatures and may be less than two hours at -15° C. NiCd batteries are recommended for low temperature applications.
(254 mm L x 152 mm W x 127 mm H)
polyester plastic with RF-resistant coating.
Shoulder strap; hose; probe (with filter); operator’s manual; and quick reference card.
2 71-6417
Page 11
Introduction
Sensor Specifications
Table 1-2 Innova Sensor Specifications and Factory Default Settings
71-6417 3
T arget Gas Display Increments Range Low Alarm High Alarm TWA Alarm STEL Alarm
HC 1% LEL 0 to 100% LEL 10% LEL 50% LEL N/A N/A
HC 0.1% VOL 0 to 5.0% VOL 0.5% VOL 2.5% VOL N/A N/A
HC 20 PPM 0 to 10,000 PPM 1000 PPM 5000 PPM N/A N/A
10 PPM 0 to 1,000 PPM 100 PPM 500 PPM N/A N/A
1
HC
1% VOL 0 to 100% VOL 10% VOL 50%VOL N/A N/A
2
CH
0.1% VOL 0 to 30.0% VOL 19.5% VOL 23.5% VOL N/A N/A
4
S 1 PPM 0 to 200 PPM 10 PPM 15 PPM 10 PPM 15 PPM
2
O
H
2
CO 1 PPM 0 to 250 PPM 25 PPM 200 PPM 25 PPM 200 PPM
0.1 PPM 0 to 9.9 PPM 0.5 PPM 1.0 PPM 0.5 PPM 1.0 PPM
2
Cl
HCN 1 PPM 0 to 30 PPM 5 PPM 10 PPM 5 PPM 10 PPM
1 PPM 0 to 100 PPM 25 PPM 35 PPM 25 PPM 35 PPM
3
NH
NO 1 PPM 0 to 100 PPM 25 PPM 75 PPM 25 PPM 75 PPM
0.1 PPM 0 to 9.9 PPM 3.0 PPM 5.0 PPM 3.0 PPM 5.0 PPM
2
NO
0.01 PPM 0 to 3.0 PPM 0.3 PPM 1.0 PPM 0.3 PPM 1.0 PPM
3
PH
0.1 PPM 0 to 9.9 PPM 2.0 PPM 5.0 PPM 2.0 PPM 5.0 PPM
2
SO
1
Soil Vapor and Fuel Vapor only.2Land Surveyor version only.
Page 12
Innova Series Operator’s Manual
Optional Accessories
Table 1-3 lists the optional accessories available for the Innova Series gas monitor. Part numbers for all accessories are in Appendix A, Parts List and Accessories.
Table 1-3 Innova Series Optional Accessories
Accessory Description
NiCd battery charger Charges NiCd batteries while still in the monitor.
Optional hoses and probes Lengths of hose up to 100 feet are available (except Cl
Dilution fitting Used to provide sufficient oxygen to allow proper
Moisture trap Glass-bodied with a pleated paper filter that collects
Auxiliary hydrophobic filter
Remote Buzzer Repeats all audible alarms of the Innova. Plugs into the
Carrying case A case that holds the Innova and the most commonly
Confined space kit This carrying case can contain all equipment and the
Calibration kit A kit consisting of a carrying case containing gas
Data retrieval kit (Innova View)
Provides a full charge over an 8-hour period, then drops to a sustaining rate. Includes an alkaline recognition feature that prevents you from attempting to recharge alkaline batteries.
versions). Two optional 30-inch probes (aluminum with dust filter or fiberglass with hydrophobic filter) are available.
response of the hydrocarbon (LEL/ppm) sensor when sampling inert environments.
excess water that is drawn into or condensed in the sample hose.
A filter with a water-impervious membrane that connects between the sampling hose and the Innova’s inlet fitting.
remote alarm jack (CHGR). Includes clip to attach to lapel or other convenient place.
used accessories. The case also has space to hold the NiCd battery charger.
most commonly used accessories necessary to safely and accurately calibrate and use the Innova.
cylinders, valves, and appropriate fittings to calibrate the Innova.
Windows 95/98/NT compatible software necessary to access the data stored in the Innova (includes serial cable).
2
4 71-6417
Page 13
Chapter
2
START UP & OPERATION
Preparing for Start Up
1. To remove the battery compartment cover, place the Innova upside down, turn the large screw counterclockwise until it is loose in its socket, then pull up.
2. Install four “D” size batteries according to the diagram inside the battery compartment. Make sure that the battery polarities are correctly oriented.
NOTE
Make sure the slide switch at the bottom of the battery compartment is
set to “ALK” for alkaline or “NI-CAD” for NiCd batteries, see
Chapter 6, Maintenance, Figure 6-1, Innova Battery Compartment.
3. Replace the battery compartment cover and turn the screw clockwise. Tighten snugly to compress the gasket and seal the battery compartment.
CAUTION
Some toxic gas sensors require up to 15 minutes to stabilize, after you install the batteries. If your Innova includes a toxic gas sensor(s), do not turn on and use the Innova during this period. If the unit is started up immediately after installing batteries, the “BIAS” message may appear, see Chapter 6, Maintenance, for instructions on installing or recharging batteries.
4. Verify that the hydrophobic filter and cotton ball are in good condition and installed properly in the probe body.
NOTE
For transformer versions, connect the dilution fitting directly to the disconnect coupler fitting on the front of the monitor, then connect the 10-inch probe to the dilution fitting.
5. Attach the probe to the female disconnect coupler fitting on the sample hose.
6. Attach the other end of the hose to the female disconnect coupler on the front of the instrument.
71-6417 5
Page 14
Innova Series Operator’s Manual
WARNING
Operation of instrument without probe/filter assembly attached will result in pump damage and possible impaired performance. Do not operate without probe/filter assembly attached.
Starting Up the Innova
Perform the following steps to start up the Innova and adjust internal circuits to fresh air readings (demand zero). Please read this entire section before turning on your Innova.
WARNING
Perform the following procedure in a “fresh air” environment (environment known to be free of combustible and toxic gases and of normal oxygen content).
1. If you are using NiCd batteries, make sure the batteries are fully charged before you continue this procedure.
2. Press and hold the ON/OFF button for one second. The Innova begins
the warm-up period. Several messages appear during the warm-up period, starting with the model and software version. Other screens displayed are:
• Battery capacity on a scale from E (empty) to F (full capacity) for
new alkaline or a fully charged battery pack.
• Datalogging time left.
• Settings for High, Warning, STEL, and TWA alarms.
• The CAL DUE # DAYS indicates when the Innova is due for
calibration according to the schedule you accepted in the user program when enabled.
NOTE
During the warm-up period the Innova determines if the toxic sensors are properly biased. If the sensors require additional biasing, the screen
continuously displays “BIAS” with alarms instead of the gas reading.
Once the readings fall within acceptable limits, a fixed time-out is initiated. During this time-out, the display will alternate between
“BIAS” and the sensor(s) reading. If an air adjust is performed during
this time-out period, it will not adjust the sensor(s) that are experiencing the ‘bias time-out’ but will adjust the other sensors.
6 71-6417
Page 15
Start Up & Operation
3. When the warm-up is complete, the buzzer will beep, lights will flash
several times, and display will read “WARM UP COMPLETE”.
NOTE
If the bump test is set to “BUMP TST ENABLED” the display will cue the operator to perform an “AIR ADJUST” and then apply gas,
see Performing a Bump Test on Power-Up instructions.
4. Hold down the AIR button for 3 seconds until the display counts down and the bar graph starts to scroll, then the display will read “DONE”
once the “fresh air” reading for all active sensors is complete (20.9% for oxygen, 0 for all other sensors).
NOTE
After you “AIR ADJUST”, a “XXX” sensor failure may appear on the
display screen, see Chapter 3, Alarms, Sensor Alarms, to respond to these messages.
at the end of this chapter for
Normal Operation
Normal operation is defined as any time after the warm-up period is complete and the Innova is not alerting you to an alarm or failure.
After warm up is complete, the Innova simultaneously displays the current gas concentration for all active sensors and the NORM icon. This display screen is defined as the normal screen. The sensor labels will alternate between the gas type and unit of measure.
During normal operation, the Innova Series gas monitor simultaneously monitors for all target gases for your configuration of the Innova.
Monitoring Gases
To monitor for the target gas(es), expose the probe to the area to be monitored. You can leave the Innova monitoring for an entire workday. The Innova monitors and displays all applicable gas conditions at the same time.
CAUTION
If your Innova includes an Ammonia (NH3), Chlorine (Cl2), Sulfur Dioxide (SO Nitrogen Dioxide (NO accumulation. The target gas is absorbed by moisture. Inspect the probe’s hydr ophobic filter and cotton ball frequently if you use the Innova in an environment where moisture might accumulate.
71-6417 7
), Hydrogen Cyanide (HCN), Nitric Oxide (NO),
2
), Phosphine (PH3) sensor, avoid moisture
2
Page 16
Innova Series Operator’s Manual
While in normal operation, press any button to illuminate the display screen back light.
NOTE
If the Comfort Alert is enabled, the Innova will beep and/or blink every 3 minutes to remind the operator that the unit is turned on and functioning.This feature can be disabled or enabled by the operator, see Chapter 4, Alarms, Comfort Alert, for activating/deactivating this feature.
Changing Combustible Ranges
The RANGE button is used to change the range of combustible gas detected.
The available ranges depend on the version of the instrument, see Chapter 1, Introduction, Table 1-2. When changing to ppm range the reading is displayed in the bottom row. If toxic sensor(s) are present the combustible ppm reading will overwrite the toxic readings. If a toxic sensor(s) goes into an alarm condition, the instrument will auto range back to the next combustible range, and allow the toxic sensor(s) to be displayed.
To change the combustible gas range, press and release the RANGE button.
The unit will beep once and switch to the alternate range (e.g. % LEL to ppm).
See Appendix C, Innova Land Surveyor, for operation of the RANGE button
for Land Surveyor models.
If, while operating in the ppm range, the gas concentration exceeds the full scale range, the Innova will automatically switch to the % LEL (or % VOL if so configured) range.
Turning Off the Innova
Press and hold the ON/OFF button for 3 seconds. A power off message on the
display screen counts down from 3 seconds. The Innova sounds a pulsing tone, and the backlight automatically turns on during the power off sequence.
To cancel the power off command, release the ON/OFF button before the
display screen goes blank.
NOTE
The STEL, TWA, maximum, and minimum (for O2) readings are automatically reset on power down unless the Lunch Break mode is enabled, see Chapter 4, User Program, Lunch Break Option for more detail.
8 71-6417
Page 17
Start Up & Operation
CAUTION
If the Innova will not be used for more than 4 weeks, remove the batteries from the instrument. Failure to remove the batteries prior to storage could result in damage to the instrument.
If your Innova uses rechargeable NiCd batteries, the batteries must be fully charged before each use. When using alkaline batteries, it is best to install fresh batteries before each use, see Chapter 6, Maintenance, to recharge NiCd batteries or replace alkaline batteries.
Performing a Bump Test on Power-Up
When the bump test option is set to “BUMP TST ENABLED” the display
alerts the operator to perform a bump test after the warm-up cycle is complete, see Chapter 4, User Program, Bump Test on Power-Up Option.
To perform the bump test:
1. Assemble the calibration kit, see Chapter 5, Calibration.
2. Turn on the Innova and wait for it to complete the warm-up cycle. The
display will read “AIR ADJUST NOW”. The screen will blink on and
off with alternating lights, prompting the operator to perform a fresh air adjust.
3. Press and hold the AIR button for three seconds until “DONE”
appears. Once the Innova has been air adjusted, the display will begin
counting down, and prompt the operator to “APPLY GAS”, alternating with “BUMP TEST”. An audible beep will sound.
NOTE
If an air adjust is not performed within 30 seconds, a “BUMP TEST INCOMPLETE” message appears and the buzzer beeps on and off for
three seconds. The Innova will return to the normal operating mode.
4. Attach the tubing from the regulator to the probe tube, the Innova will draw gas from the cylinder.
5. The Innova should respond to the detected gas and beep to indicate that a Bump Test is being performed. All gas levels for standard gases are displayed as in the normal operating mode.
NOTE
If no gas is detected during the “APPLY GAS” countdown, a “BUMP TEST INCOMPLETE” message appears and the buzzer beeps on and
off for 3 seconds. The Innova will return to normal operating mode.
71-6417 9
Page 18
Innova Series Operator’s Manual
Once gas is detected a countdown is initiated. During this single beep countdown the user should visually verify the sensors are responding correctly.
NOTE
Bump test peak readings are recorded. Events are recorded as “BUMP TEST COMPLETE” if the Innova detected gas, and “BUMP TEST ABORTED” if the Innova did not detect gas.
A “BUMP TEST COMPLETE” message will be displayed. Then the screen will alternate between “REMOVE GAS” and the gas readings.
6. Remove the tubing from the probe.
Once the instrument detects that the gas has been removed the screen
display an “ALARMS SILENCED” message. The alarms will be
silenced for a blind 30 second count.
After the 30 second count has elapsed an “ALARMS ACTIVE”
message will display briefly, audible alarms will be enabled and the Innova will return to the normal operating mode.
NOTE
If the Innova does not respond properly during the bump test, follow the instructions in Chapter 5, Calibration, or Chapter 6, Troubleshooting sections of this manual.
NOTE
For instruments having a toxic sensor other than CO or H2S, calibration gases are only available in single-gas cylinders. Therefore a multi-step bump process has been programmed into the bump test routine. First, the standard sensors are bumped using a multi-gas mixture. Then the first Toxic sensor is bumped using a single-gas cylinder and repeated if your unit has a second toxic sensor.
WARNING
Avoid using a gas mixture containing H2S for bump testing an instrument with Cl2, NH3, NO2 or PH3 sensors.
10 71-6417
Page 19
Chapter
3
ALARMS
This chapter describes Innova alarm indications, display messages, and response for the following types of alarms:
• Gas Alarms
• Battery Alarms
• Sensor Alarms
• Pump Alarms
• Datalog Alarms
• Temperature Alarms
Gas Alarms
This section describes alarm indications, display messages, and response to Low, High, STEL, and TWA gas alarms. The backlight automatically turns on during all gas alarms.
Gas Alarm Indications
Table 3-1 lists the indications for each type of gas alarm.
Table 3-1 Alarm Indications (Gas Alarms)
Alarm Display Light/Buzzer
Low Gas Alarm Reading flashes slow flashing/pulsing
(once/second)
High Gas Alarm Reading flashes fast flashing/pulsing
(twice/second)
Decreasing O
Increasing O
STEL Gas Alarm STEL icon flashing/pulsing
TWA Gas Alarm TWA icon flashing/pulsing
Alarm Reading flashes fast flashing/pulsing
2
Alarm Reading flashes fast flashing/pulsing
2
71-6417 11
Page 20
Innova Series Operator’s Manual
Responding to Gas Alarms
1. Follow the established procedure for the type of gas alarm (Low, High, STEL or TWA). If a procedure is not in place, establish one that is appropriate for your application.
2. Reset the alarm circuit. The Innova resets the circuit according to the
alarm reset setting “AUTO RESET” or “MANUAL RESET”, see
Chapter 4, User Program, Alarm Latch option to enter the program and display and/or update the setting.
• If the alarm reset setting is auto reset, the Innova resets the
applicable alarm circuit after the gas reading falls below the alarm setting (or rises above for decreasing O
).
2
• If the alarm reset setting is manual reset, press theON/OFF
button to reset the alarm circuits after the gas reading falls below the alarm setting (or rises above for decreasing O
).
2
Battery Alarms
This section describes alarm indications, display messages, and response for the following battery alarms: Low Battery and Replace Battery. The backlight automatically turns on during all battery alarms
Battery Alarm Indications.
Table 3-2 Alarm Indications (Battery Alarms)
Alarm Display Light/Buzzer
Low Battery LOW BATTERY double flash/beep (every two minutes)
battery icon flashes
Battery Failure (ALK)
Battery Failure (NiCd)
REPLACE BATTERY constant flash/beep
battery icon steady
RECHARGE BATTERY constant flash/beep
battery icon steady
Responding to Battery Alarms
This section describes response to Low Battery and Replace/Recharge Battery alarms.
1. For low battery alarm, proceed to step 2 as soon as possible. The Innova will continue to function properly for approximately 1/2 hour until the Replace/Recharge Battery screen and alarm sounds. For battery failure alarm, turn off the Innova immediately. It is not functioning as a gas detection device.
12 71-6417
Page 21
Alarms
2. You must replace the alkaline or recharge the NiCd batteries within your monitor before putting it back into operation, see Chapter 6, Maintenance, for procedures to replace or recharge your batteries.
WARNING
Always replace or recharge batteries in a “fresh air” environment (environment known to be free of combustible and toxic gases and of normal oxygen content).
Sensor Alarms
This section describes alarm indications, display messages, and response to sensor failure alarms. The backlight automatically turns on during all failure alarms.
Sensor Alarm Indications
Table 3-3 lists the indications for each type of failure alarm.
Table 3-3 Alarm Indications (Failure Alarms)
Alarm Display Light/Buzzer
Sensor Failure XXX Continuous
Responding to Sensor Alarms
A sensor failure alarm can occur during normal operation or calibration. Alarms that occur during calibration are described in Chapter 5, Calibration.
1. Examine the display screen to determine which sensor is failing. The
failing sensor displays “XXX” instead of the gas reading for a sensor
that has failed an air adjust, Calibration, or gas missing sensor.
2. Perform an air adjust, if the sensor(s) continues to fail, perform a calibration, see Chapter 5, Calibration.
3. If the sensor(s) continue to fail, replace the sensor(s), see Chapter 6, Replacing Sensors.
4. You can enter the Innova user program and change the sensor status
setting to “OFF”. All information for an absent sensor is removed
from the normal screen and recorded as XXX in the datalog.
WARNING
The Innova does not display gas readings, initiate alarms, compute averages, or store data for “OFF” sensor(s).
71-6417 13
Page 22
Innova Series Operator’s Manual
Pump Alarm
This section describes alarm indications, display messages, and response to sensor, memory, and pump failure alarms. The backlight automatically turns on during all failure alarms.
Pump Alarm Indications
Table 3-4 lists the indications for each type of failure alarm.
Table 3-4 Alarm Indications (Failure Alarms)
Alarm Display Light/Buzzer
Pump Failure PUMP FAILED
alternating with PRESSTO CLEAR alternating with gas readings. X in place of spinning icon
Responding to a Pump Alarm
1. Check the hose, probe, and filter for debris.
2. Press the ON/OFF button to reset the alarm circuit and return to the
normal screen with the pump icon spinning.
3. If the pump restarts, and the monitor functions normally, the problem was momentary. If the indications remain, turn the monitor off, then investigate the probe, hose, or internal flow system for obstructions, dirty filter, or loose sensor(s) inside instrument.
alternating/beeping (every second)
WARNING
Operation of instrument without probe/filter assembly attached will result in pump damage and possible impaired performance. Do not operate without probe/filter assembly attached.
14 71-6417
Page 23
Alarms
Datalog Alarm
This section describes the alarm indication, display message, and response for the datalog full alarm. A datalog full alarm occurs when the datalog becomes
full and the datalog full command was accepted as “STOP ON FULL”.
Datalog Alarm Indications
Table 3-5 lists the indications for the datalog full alarm.
Table 3-5 Alarm Indications (Datalog Full Alarm)
Alarm Display Light/Buzzer
Datalog Full LOGGING STOPPED flashing/pulsing (every 30 seconds)
Responding to the Datalog Alarm
Perform one of the following procedures to reset the datalog alarm.
• Download the data, save and clear the log when prompted.
• Enter the Datalog Features Menu, and update the datalog full
command to “OVRWRITE ON FULL”, see Chapter 4, User
Program, Log Memory Manager option.
• Enter the Datalog Features Menu and clear the datalog, see Chapter 4, User Program, Clear Log Memory.
Temperature Alarm
The Innova has an alarm to alert the user that the unit is being operated at a temperature that exceeds its operating specification of -20°C to +45°C (-4°F to +113°F). If the temperature is outside this range, the following message
appears every 30 seconds and the sensor readings may be inaccurate: “TEMP WARNING” alternating with the gas readings.
The buzzer beeps and lights flash twice every 30 seconds, and the condition will clear automatically when the temperature returns to within the operating limits.
NOTE
The temperature alarm is user enabled or disabled, with the default setting “disabled”, see Chapter 4, User Program, Temperature Alarm.
.
71-6417 15
Page 24
Innova Series Operator’s Manual
16 71-6417
Page 25
Chapter
4
USER PROGRAM
This chapter includes an introduction to the Innova User Program and step­by-step procedures to run the program.
About the User Program
The Innova User Program allows you to view and update factory-set parameters and display several items, such as the battery voltage, datalog status, current temperature, time, and date.
The menu-driven program has three main menus:
• Display Menu
• Advanced Features Menu
• Datalog Features Menu
The user navigates through the user program by using the three buttons:
ON/OFF
AIR
↑ ↓
RANGE.
• The right arrow (→) is used to accept within the menu.
•The up arrow ( ) and down arrow ( ) are used to scroll either
forward or backwards through the menus.
NOTE
The Innova has a time-out feature. If no buttons are pressed for 10 seconds while in the Display Menu, Advanced Features Menu or Datalog Features Menu the display will flash and buzzer will beep for 10 seconds. If no buttons are pressed during this time, the Innova will return to the normal operating mode.
71-6417 17
Page 26
Innova Series Operator’s Manual
Display Menu
In the Display Menu the following may be viewed:
• Battery Capacity
• Peak Readings (Option to Clear)
• STEL Levels (Toxic Sensors Only)
• TWA Levels (Toxic Sensors Only)
• Temperature
• Date and Time
• Operator ID (Option to Change)
• Location ID (Option to Change)
• Exit
To enter the Display Menu:
1. Press and hold the RANGE button for 3 seconds. The display will show “DISPLAY MENU” and count down from 3 seconds. Your first screen will display the “BATTERY CAPACITY”.
2. To scroll to the next screen, use the button. To scroll backwards through the menu, press the button. Continue to scroll through the screens to display the desired information, clear Peak Readings, or change an Operator or User ID.
3. To exit the program, press either or ↓ until you get to the ‘EXIT’ display, then press to exit to the normal operating mode.
NOTE
When Peak values are displayed, after 3 seconds the display reads
“PRESS TO CLEAR”, so the user can clear peak values if desired.
To change a User or Operator ID:
NOTE
Operator and Location must first be entered before a selection can be made, see Edit Operator, Location, and Instrument ID’s Chapters.
1. For the Operator ID and Location ID screens, to change the selection press the button, and then use or to designate a new selection.
2. Press the button to accept the selected ID.
18 71-6417
Page 27
User Program
Accessing Password-Protected Menus
A password may be used to add extra security to the following menus:
• Advanced Features protected by “PASSWORD”
• Datalogging Menu protected by “LOGWORD”
• Calibration Menu protected by “CALWORD”
See Create/Edit CalWord/PassWord/LogWord for details on creating a password. If enabled, a code must be entered before the user may enter these menus.
To enter a PassWord, LogWord, or CalWord:
1. Press and hold and for 3 seconds. The message “FEATURES MENU” appears on the display and counts down from 3 seconds, and then displays “BEGIN CAL”. Press to scroll to “ADVANCED FEATURES” or “DATALOG MENUS”.
2. Stop scrolling at the desired screen and press to accept. The display now reads “PASSWORD”, “LOGWORD”, or “CALWORD” and “* * *” (if protection has been enabled). The cursor is on the first asterisk.
NOTE
If the “***” does not appear when entering any of the three menus, security protection is not enabled.
3. Press or to scroll to the correct character.
4. Press to accept. The cursor moves to the next character.
5. Repeat steps 4 and 5 until all three characters are accepted. If your entry is correct, you will advance to the next screen. If your entry is incorrect, you will return to the original screen. Begin at step 2 to re­enter your password.
71-6417 19
Page 28
Innova Series Operator’s Manual
Advanced Features Menu
The Advanced Features Menu allows you to change the following parameters of the instrument.
The menu is as follows:
• Edit Alarms
• O2 Alarm Configuration
• Alarm Latch Option
• Alarm Delay Selection
• Lunch Break Option
• Backlight Delay Selection
• Calibration Interval Selection
• Bump Test Option
• Sensor Select
• Setting Auto Cal Gas Values
• Comfort Alert Options
• Temperature Alarm
• Change Cal Word
• Change Password
• Change Logword
• Exit
To enter the Advanced Features Menu:
1. Press and hold both the AIR and RANGE buttons for 3 seconds. The screen displays “FEATURES MENU” and counts down from 3
seconds.
2. The first screen displayed after the countdown is “BEGIN CAL”. Press the button, then press the button. If prompted by “PASSWORD” and “***”, enter a password as described in the
previous section, Accessing Password-Protected menus. When entered correctly, or if PassWord is not enabled, you are now at the first option
“EDIT ALARMS” in the Advanced Features Menu.
NOTE
The Innova has a time-out feature. While in the Advanced Features Menu, if a button is not pressed within 10 seconds the display will flash and the buzzer will beep for 10 seconds. If no button is pressed during the 10 seconds, the Innova will return to normal operating mode.
20 71-6417
Page 29
User Program
DIT ALARMS
E
This allows the user to increase or decrease the High and Low Alarms for each gas. See Chapter 1, Introduction, Table 1-2, Sensor Specifications for factory default settings.
1. Press to enter the HC screen. The high alarm reading will flash.
2. Press ↑ or to increase or decrease the alarm setting. If you choose to keep the alarm as is, press to move to the next alarm setting.
3. Continue step 2 until you have set all gas alarms. After the last gas alarm
is set, the display automatically returns to the “EDIT ALARMS”
screen.
NOTE
You will scroll through the following gas alarms, appropriate to your version of Innova: HC, O2, TOXIC1, TOXIC2.
4. Press ↑ or ↓ to continue scrolling through the Advanced Features Menu.
O2 A
LARM CONFIGURATION
This option allows you to set the oxygen alarms (High and Low) to either increasing or decreasing.
Options you can choose are:
ONE INC/ONE DEC - The warning alarm is set to decreasing at
19.5%VOL and high alarm increasing at 23.5%VOL.
TWO ALRM/DECREASE - Both alarms are set to decrease. Warn at
19.5%VOL and High at 18.0%VOL.
TWO ALRM/INCREASE - Both alarms are set to increase. Warn at
22.5%VOL and high at 23.5%VOL.
The factory default setting is “ONE INC/ONE DEC”.
To enter this menu from the Advanced Features Menu:
1. Press to accept “O2 ALARM CONFIG”.
2. Press or to scroll through the settings. Press the to accept the setting.
3. Press to continue scrolling through the Advanced Features Menu.
71-6417 21
Page 30
Innova Series Operator’s Manual
LARM LATCH OPTION
A
This option changes the alarm setting to either “AUTO RESET” for automatically resting the alarm or “MAUNAL RESET”, in which case an alarm must be manually reset. The factory default setting is “AUTO RESET”.
1. Press to accept “ALARM LATCH”.
2. Press or to scroll between settings. Press to accept the setting.
3. Press to continue scrolling through the Advanced Features Menu.
LARM DELAY SELECTION
A
This selection allows the user to select a time-delay period. When a delay other than 0 seconds is selected, all gas alarms are delayed until the alarm condition has persisted for the length of the delay period. This feature can be useful if the instrument is exposed to high intensity RFI (radio frequency interference) or EMI (electromagnetic interference), which could cause momentary false alarms. Selection options are 0 seconds (no delay), 1, 3, or
5 seconds. The default setting is “0 SECOND DELAY”.
1. Press to accept “ALARM DELAY”.
2. Press or to scroll between settings (“0, 1, 3, OR 5 SECOND DELAY”). Press → to accept the setting.
3. Press to continue scrolling through the Advanced Features Menu.
WARNING
The Innova does not, initiate gas alarms until the selected Alarm Delay period has been exceeded. Select an Alarm Delay period as short as possible to avoid unnecessary delay of alarms.
UNCH BREAK OPTION
L
The lunch break option allows the user to either reset averages (STEL and TWA) or retains averages when the Innova is turned off. The factory default
setting is “RESET AVERAGES”.
NOTE
If the lunchbreak menu is set to “RET AIN AVERAGES”, the Innova
will retain the STEL and TWA averages when turned off. If the
lunchbreak menu is set to “RESET AVERAGES”, the Innova will
reset the averages when turned off.
To change the lunch break menu from the Advanced Features Menu:
1. Press at the screen “LUNCH BREAK”.
22 71-6417
Page 31
User Program
2. Press or to scroll through the settings. Press to accept the setting.
3. Then press to continue scrolling through the Advanced Features Menu.
ACKLIGHT DELAY SELECTION
B
The backlight delay feature sets the number of seconds the backlight is illuminated after any button is pressed. The factory default is setting is 5 seconds. Your options are in 5 second increments up to 60 seconds, and 2, 3, 5, or 10 minutes.
To change the backlight delay setting from the Advanced Features Menu:
1. Press at the screen “BKLIGHT DELAY”.
2. Press or to scroll through the settings. Press to accept the setting.
3. Then press to continue scrolling through the Advanced Features MenuCal Interval Selection
This sets the number of days between calibrations.
NOTE
The number of days left until calibration is displayed during the warm-
up cycle. When the Innova is due for calibration, display reads “CAL OVERDUE” and buzzer and lights, pulse and flash after the warm-up
cycle is complete.
Your options are no cal reminder, 1 to 6 one-day increments and 1 to 26 one-
week increments. The factory default setting is “NO CAL REMINDER”,
which means the calibration reminder function is turned off.
NOTE
For optimum performance, calibrate the Innova before every use.
To change the calibration reminder setting from the Advanced Features Menu:
1. Press at the screen “CAL INTERVAL”.
2. Press or to scroll through the settings.
3. Press to accept the setting.
4. Then press to continue scrolling through the Advanced Features Menu.
UMP TEST ON POWER-UP OPTION
B
71-6417 23
Page 32
Innova Series Operator’s Manual
The bump test option enables or disables the Innova to cue the user to apply gas for a bump test immediately after the Innova warm-up cycle is complete.
The factory default setting is “BUMP TST DISABLED”. A bump test is
sometimes necessary, to ensure the Innova is detecting gases before each use.
To change the bump test setting from the Advanced Features Menu:
1. Press at the screen “BUMP TEST”.
2. Press ↓ or ↑ to scroll between “BUMP TST DISABLED” to “BUMP TST ENABLED”.
3. Press to accept the setting.
4. Then press to continue scrolling through the Advanced Features Menu.Sensor Select
The sensor select function turns the HC and O2 sensor(s) on or off. Both toxic 1 and toxic 2 sensor(s) may be turned off or changed to monitor another gas type provided the corresponding sensor(s) are installed in the proper location.
WARNING
The Innova does not display gas readings, initiate alarms, compute averages, or store data for “OFF” sensor(s).
To change the sensor select setting from the Advanced Features Menu:
1. Press at the screen “SENSOR SELECT”. All four gases and their
settings will display. The cursor and reading will flash for HC.
2. Press ↑ or to change the setting to “ON” or “OFF”. If you choose to keep the setting, as is, press to move on to the O2.
3. Repeat step 2 for O2, press to move on to the Toxic 1 sensor.
4. For the toxic sensors press ↑ or ↓ to select or change the sensor type, press to move on to Toxic 2.
5. Continue step 4 for Toxic 2.
6. Press to continue scrolling through the Advanced Featuring Menu.
NOTE
You will scroll through the following sensor selections, appropriate to your version of Innova: HC, O2, Toxic 1, Toxic 2.
S
ETTING AUTOCAL GAS VALUES
This function allows you to set the auto calibration values to the concentration of the cylinders you are using.
24 71-6417
Page 33
User Program
NOTE
The factory default settings for a standard four gas instrument are HC 50%LEL, O2 12.0%VOL, H2S 25 PPM, and CO 50 PPM. These are
the values that match Thermo GasTech’s multi-gas cylinders.
To change the value(s) from the Advanced Features Menu:
1. Press at the screen “AUTOCAL GAS VAL”. All four gases and
their values will display. The cursor and reading will flash for HC (if installed and ON).
2. Press or to change value up or down.
3. Once you’ve reached the desired value, press to move on to the HC ppm. If you choose to keep the setting, as is, press to move on to the next sensor.
4. Continue this process through all gases.
NOTE
You will scroll through the following autocal gas values, appropriate to your version of Innova: HC, O2, TOXIC1, TOXIC2.
71-6417 25
Page 34
Innova Series Operator’s Manual
OMFORT ALERT OPTIONS
C
The comfort alert is a friendly reminder that the instrument is on. The reminder begins 3 minutes after the last button press and will sound every 3 minutes while in the normal operating mode. The alert has four options:
• Beep Only - Every 3 minutes the buzzer sounds.
• Blink Only - Every 3 minutes the lights flash.
• Blink and Beep- Both buzzer beeps and lights flash every 3 minutes.
• Both Off - Option disabled.
The factory default setting is “BLINK AND BEEP”.
NOTE
The 3 minute comfort alert timer is reset after any button is pressed by the user.
To change the setting from the Advanced Features Menu:
1. Press at the screen “COMFORT ALERT”.
2. Press or to scroll through the options.
3. Press → to accept the setting. Press to continue scrolling through the Advanced Features Menu.
T
EMPERATURE ALARM
The Innova has an alarm to alerts the user that the Innova is being operated at a temperature that exceeds its operation specification of -4ºF to 113ºF (20ºC to 45ºC). When the temperature alarm is activated, the buzzer and
lights alternate every 30 seconds. The factory default setting is “TEMP ALR DISABLED”.
To change the setting from the Advanced Features Menu:
1. Press at the screen “TEMP ALARM”.
2. Press or to change to either “TEMP ALR DISABLED” or “TEMP ALR ENABLED”.
3. Press to accept the setting.
4. Press to continue scrolling through the Advanced Features Menu.
26 71-6417
Page 35
User Program
NTER/EDIT CALWORD, PASSWORD, OR LOGWORD
E
These options add extra security to the calibration process (CALWORD), Advanced Features Menu (PASSWORD), and Datalog Features Menu
(LOGWORD). By changing the default factory setting of “***”, you
automatically turn this option on, and password protect that area of the user program. If turned on, a code must be entered before one may continue a process. If “***” is entered, the password security is turned off. The factory default setting for each password is “***”, or disabled.
To add or change the code from the Advanced Features Menu,
1. Press at the appropriate screen “CHANGE CALWORD”, “CHANGE PASSWORD” or “CHANGE LOGWORD”. The
screen will display either three asterisks “***”, or the three digit code, with the cursor on the first character.
2. Press or to change the first character of the code. You have the option of “*, 0 through 9”.
NOTE
When entering a CalWord, PassWord, or LogWord all three characters must be a number, 0 through 9. A password containing one or more “*” is not valid, no security protection is implemented if a password contains an “*”.
3. Press to accept the setting, and move on to the second character. If you do not want to change the first character, press to move on to the second character.
4. Repeat the process through the third character.
5. Press to continue scrolling through the Advanced Features Menu.
XIT
E
The ‘EXIT’ screen exits you from the Advanced Features Menu. Press at “EXIT” to return to the normal operating mode.
71-6417 27
Page 36
Innova Series Operator’s Manual
Datalog Features Menu
The Datalog Features Menu allows you to change the following parameters of the instrument.
The menu is as follows:
• Adjust Date and Time
• Edit Operator IDs
• Edit Location IDs
• Edit Instrument ID
• Log Memory Manager Option
• Log Averages Selection
• Log Interval Selection
• View Log Status
• Clear Log Memory
• View Log Began
• Exit
To enter the Datalog Features Menu:
1. Press and hold both the AIR and RANGE buttons for 3 seconds. The screen displays “FEATURES MENU” and counts down from 3
seconds.
2. The first screen displayed after the countdown is “BEGIN CAL”. Press the button twice. The display will read “DATALOG MENUS”.
3. Press the button. If prompted by “LOGWORD” and “***”, enter
a LogWord as described earlier in this chapter. When entered correctly, or if LogWord is not enabled, you are now at the first option
“ADJUST DATE AND TIME”.
NOTE
The Innova has a time-out feature. While in the Advanced Features Menu, if a button is not pressed within 10 seconds the display will flash and the buzzer will beep for 10 seconds. If no button is pressed during the 10 seconds, the Innova will return to normal operating mode.
28 71-6417
Page 37
User Program
DJUST DATE AND TIME
A
The date and time are adjustable by the user. The default factory settings are Pacific Standard Time. The innova uses a 24-hour format to display time. For example, 21:30:00 is 9:30 PM.
To change the date and time from the Datalog Menu:
1. Press at the screen “ADJ DATE AND TIME”. The date and time
will appear, with the cursor on the month that is flashing.
2. Press or to change to the correct month. Once you’ve reached the correct value, press to move to the day of the month. If you choose to keep the setting, as is, press to move to the year.
3. Continue this process through remaining date and time settings. Once
you’ve completed and pressed past the seconds, the “ADJ DATE AND TIME” screen displays, and your selection has been accepted.
DIT OPERATOR, LOCATION, AND INSTRUMENT IDS
E
The IDs are used to allow the user to identify the Operator and Location of personnel operating the Innova. The information is stored in the datalogging memory for retrieval at a later time. The Innova has three IDs the user can update:
• The Operator ID screen allows the Innova to reference logged data
with 16 different operator codes.
• The Location ID screen allows the Innova to reference logged data
with 64 different location codes.
• The Instrument ID screen allows you to uniquely identify the Innova.
The Operator and location IDs are a two row, 8 characters per row, field. Each character accepts A-Z, a-z, 0-9, asterisk (*), and blank space. The Innova User Program stores up to 16 Operator and 64 Location IDs.
Each Operator ID is uniquely referenced by a number from 1 to 16. Each Location ID is referenced by a number from 1 to 64. You can accept (and record in the datalog) an Operator and/or Location ID by its reference number.
To add or change the Operator and Location ID from the Datalog Menu:
NOTE
The factory default setting for Operator, Location, and Instrument IDs are blank.
71-6417 29
Page 38
Innova Series Operator’s Manual
1. Press at the appropriate screen “EDIT OPER ID”, “EDIT LOC ID”, or “INST ID”. The first ID screen will appear, you can change
this screen by pressing to accept and start editing this ID or you can press orto scroll through the other ID screens. Then press to accept.
2. Press ↓ or ↑ to change the first character of the ID. You have the option of “*, 0 - 9, A-Z, a-z, and blank”.
3. Press to accept the character or space, and move to the next character. If you do not want to change the first character, press → to move to the next character.
4. Repeat the process through both rows of characters.
5. Press to continue scrolling through the Datalog Menu.
NOTE
For the Operator ID and Location ID screens, the ID’s reference number is displayed on the bottom line. Press the button to continue scrolling through the various operators and locations you want to display or update. Then press to edit or update the screen.
OG MEMORY MANAGER
L
This instructs the datalog memory to either overwrite the oldest data,
“OVRWRITE ON FULL”, or stop recording data, “STOP ON FULL”, when the memory is full. The factory default setting is “OVRWRITE ON FULL”.
NOTE
If you accept “STOP ON FULL”, the Innova alerts you with a datalog
full alarm when the Innova is no longer recording data, see Chapter 3,
Alarms. If you accept “OVRWRITE ON FULL”, the Innova begins
writing over the oldest data when the datalog is full. The overwritten data is not retrievable.
To change the setting, from the Datalog Features Menu:
1. Press at the screen “LOG MEM MANAGER”.
2. Then press ↑ or ↓ to display the correct setting.
3. Press to accept the setting.
4. Then press to continue scrolling the Datalog Features Menu.
30 71-6417
Page 39
User Program
OG AVERAGES
L
This parameter controls which averages (STEL or TWA) the Innova stores in the datalog. Available selections for each toxic sensor that is active are:
• LOG ALL - Log both STEL and TWA
• LOG TWA ONLY - Log TWA only
• LOG STEL ONLY - Log STEL only
• NO AVE LOGGING - Neither STEL or TWA are logged
The factory default setting is “LOG ALL”.
To change the setting for logging STEL and TWA from the Datalog Features Menu:
1. Press at the screen “LOG AVERAGES”.
2. Press or to scroll between options until you get to the appropriate screen.
3. Press to accept the setting.
4. Press to continue scrolling through the Datalog Features Menu.
L
OG INTERVAL
This parameter instructs the Innova how often to record data in the datalog. You have the option of 5, 15, 30, 60, 90, 300, or 900 seconds. The factory
default setting is “LOG EVRY 60 SEC”.
To change the setting for log interval from the Datalog Features Menu:
1. Press at the screen “LOG INTERVAL”.
2. Press or to scroll between options until you get to the appropriate screen.
3. Press to accept the setting.
4. Press to continue scrolling through the Datalog Features Menu.
Table 4-1 illustrates approximately how long it takes to fill the datalog for the various datalog interval settings.
Table 4-1 Datalog Capacity Specifications
(accumulated on time)
Sample Interval Capacity (No Alarm Activity)
5 seconds 18 hours
15 seconds 55 hours
30 seconds 111 hours
60 seconds 9 days
90 seconds 13 days
300 seconds 36 days
900 seconds 108 days
71-6417 31
Page 40
Innova Series Operator’s Manual
OG STATUS
L
This is a display screen only, which indicates the number of datalog hours or days left in memory that is not currently occupied.
To display the number of hours or days left, from the Datalog Features Menu:
1. Press at the screen “LOG STATUS”. The screen displays reads “XX LOG HRS LEFT” or “XX LOG DAYS LEFT” (“XX” for
number of hours or days available in memory). The display will
automatically go back to the “LOG STATUS” screen.
2. Press or to continue scrolling the Datalog Features Menu.
C
LEAR LOG
To clear the data in the memory, from the Datalog Features Menu:
1. Press → at the screen “CLEAR LOG”. The screen displays AND TO CLEAR”.
2. Press and hold ↑ and ↓ for 3 seconds and release. The display reads “CLEARING LOG” and beeps for several seconds. When complete the display indicates “LOG CLEARED”.
NOTE
If you do not want to clear the log, press the button and continue scrolling through the Datalog Features Menu.
CAUTION
You cannot recover data after you clear the log. Download all required data before you clear the log.
L
OG BEGAN
This screen displays the date and time when the log was last cleared.
To display this data and time, from the Datalog Features Menu:
1. Press at the screen “LOG BEGAN”. The screen displays the data
and time. For example, “7/12/99 10:00:00”. The display automatically
goes back to the “LOG BEGAN” screen.
2. Press or to continue scrolling on the Datalog Features Menu.
E
XIT
The ‘EXIT’ screen exits you from the Datalog Features Menu. Press to exit the menu and return to the normal operating mode.
32 71-6417
Page 41
71-6417 33
Chapter
5
CALIBRATION
This chapter contains instructions to prepare the calibration kit and calibrate the Innova Series gas monitor. For Land Surveyor, Transformer Gas Tester, Soil and Fuel Vapor, see Appendix C for supplemental instructions. Note: Refer to Appendix C for information on calibrating and operating the Transformer Gas Tester (XT), and Soil Vapor (SV), and Fuel Vapor (FV) models.
WARNING
Accurate calibration of the Innova Series gas monitor is essential to ensure correct readings of gas and oxygen concentrations. Incorrect or improper calibration can impair the performance of the instrument and place you in potential danger if hazardous conditions exist.
Preparing the Calibration Kit
Perform the following steps to prepare the Innova calibration kit.
WARNING
LEL response will vary for different gases. For best accuracy, LEL calibration should be done using the target gas of the LEL/ ppm sensor. (Expected relative combustible responses, listed in Appendix B, Interference Factors, are not verified by UL.)
1. Verify that the calibrating area contains a level surface to set the Innova and calibration kit accessories upon.
2. Turn on the Innova instrument. Enter the Display Menu and verify that the Battery Capacity screen displays at least three bars (see “Display Menu” at the beginning of Chapter 4, User Program for a detailed description).
3. Carefully screw the threaded end of the regulator into the gas cylinder.
4. Attach the sample tubing over the fitting on the regulator.
Page 42
Innova Series Operator’s Manual
Calibrating the Innova
The Innova offers two calibration options, automatic and manual.
• The automatic calibration (AUTO CAL) displays all active
sensors on the display screen at the same time. You can calibrate all active sensors simultaneously using a multi-gas cylinder available
from Thermo Fisher Scientific. The concentrations set in the Auto
Cal Gas Value screen must match the multi-gas mixture, see Chapter 4, User Program, Setting Auto Cal Gas Values.
• During the manual calibration (MAN CAL) only one active sensor
is displayed on the screen at a time, only one sensor is calibrated during a manual calibration. It is recommended to use a single gas cylinder for manual calibrations. During a manual calibration you may manually change the concentration of the gas before the gas is applied. The default value is that set in the Auto Cal Gas Values screen, see Chapter 4, User Program, Setting Auto Cal Gas Values.
Auto Calibration
WARNING
Calibrate the Innova in a “fresh air” environment (environment known to be free of combustible and toxic ga ses and of normal oxygen content).
NOTE
If you use the automatic adjustment method to calibrate the sensor, make sure the gas cylinder contains the same concentrations of the
target gas as the values set in the “AUTOCAL GAS VAL” menu, see
Chapter 4, User Program, Setting Auto Cal Gas Values.
1. Hold down the AIR button for 3 seconds until the display counts down and the bar graph starts to scroll, then the display will read “DONE”
once the “fresh air” reading for all active sensors is complete (20.9% for oxygen, 0 for all other sensors).
NOTE
After you complete the fresh air adjustment, a “XXX” may appear on
the display screen in place of a sensor reading. The buzzer and lights will be on steady, see Calibration Alarms section at the end of this chapter to respond to this message.
34 71-6417
Page 43
Calibration
2. Press and hold and for 3 seconds. The message “FEATURES MENU” appears on the display and counts down from 3 seconds, and
then displays “BEGIN CAL”.
3. Press to accept and begin a calibration. If a CalWord has not enabled, the “AUTO CAL” screen is displayed. Skip the next section and proceed with Calibrating with the Auto Calibration (cont.).
NOTE
If prompted by “CALWORD” and “***”, enter a password as
described in the next section.
Entering a CalWord
A password may be used to add extra security to the Calibration with a
“CALWORD”, see Chapter 4, User Program, Create/Edit CalWord/
PassWord/LogWord. If turned on, a code must be entered before the user may perform a calibration.
To enter a CalWord (if activated):
1. Press and hold and for 3 seconds. The message “FEATURES MENU” appears on the display and counts down from 3 seconds, and then displays “BEGIN CAL”.
2. Press to accept. The display reads “CAL WORD” and “* * *”. The
cursor is on the first asterisk.
3. Press or to scroll to the correct character.
4. Press to accept. The cursor moves to the next character.
5. Repeat steps 3 and 4 until all three characters are accepted. If correct,
the next display reads “AUTOCAL”. If the CalWord is incorrect, the display reads “BEGIN CAL” message. Begin at step 2 to start over.
Auto Calibration (cont.)
4. Press to accept “AUTO CAL”. The screen displays “BEGIN AUTOCAL” alternating with “APPLY GAS” counting down from
30 seconds. If no gas is detected during this period, the screen returns
to “BEGIN CAL”.
5. Attach the tubing from the regulator to the probe tube, the Innova will draw gas from the gas cylinder.
First, the display will read “CAL IN PROCESS”. Then the screen displays “CAL TIME LEFT” (counting down from 75 seconds)
alternating with the gas readings.
71-6417 35
Page 44
Innova Series Operator’s Manual
NOTE
During calibration, datalogging is disabled but calibration events are recorded in the log.
6. When calibration is complete, the display reads “AUTOCAL PASSED” and then alternates between “>ACCEPT, ABORT” and the gas readings. The gas labels alternate with ‘PASS”. Press → to accept. The display will show “CAL ACCEPTED”, and then
alternate between “REMOVE GAS” and the gas readings.
If you choose to abort, press to scroll to “ABORT” then → to accept
the abort option.
The display will read “CAL ABORTED” and then alternate between “REMOVE GAS” and the gas readings.
NOTE
A “WEAK” or “FAIL” flag may appear in place of the “PASS” flag
in the case of a weak or failed sensor. See the instructions in the Calibration Alarms section, later in this chapter for corrective action.
7. Remove the tubing from the probe. After fresh air has been detected,
the display will alternate between “ALARMS SILENCED”, and the
readings, and then return to the normal operating mode, after an
“ALARMS ACTIVE” message is displayed.
8. Unscrew the regulator from the cylinder, leave the components of the calibration kit assembled for convenience and place in the storage case.
Calibration of T oxic Gases other than H
S & CO or for 0-100% V ol
2
TC Sensor
These calibration gases are only available in single-gas cylinders. Therefore a multi-step calibration process has been programmed into the Auto Calibration routine. First, the standard sensors are calibrated using a multi-gas mixture. Then the first sensor is calibrated using a single-gas cylinder, and repeated if your unit has a second sensor.
1. Begin the calibration process by performing an Auto Calibration. The non-standard sensor(s) will be ignored during the multi-gas calibration.
36 71-6417
Page 45
2. When the multi-gas Auto Calibration is complete, a double beep prompts the user to begin the non-standard sensor calibration(s). The default gas concentration is displayed with the ADJ icon. Change the concentration if the gas cylinder does not match the concentration displayed.
3. Once the concentrations match, Press to begin the calibration.
4. Apply the calibration gas and follow the prompts to complete the calibration process. When the first non-standard sensor is complete, and a second non-standard sensor is present, a double beep prompts the user to begin the second sensor calibration. Repeat steps 2, 3 and 4.
WARNING
Avoid using a gas mixture containing H2S for calibrating an instrument with Cl2, NH3, NO2 or PH3 sensors.
Manual Calibration
WARNING
Calibrate the Innova in a “fresh air” environment (environment known to be free of combustible and toxic ga ses and of normal oxygen content).
Calibration
1. Hold down the AIR button for 3 seconds until the display counts down and the bar graph starts to scroll, then the display will read “DONE”
once the “fresh air” reading for all active sensors is complete (20.9% for oxygen, 0 for all other sensors).
NOTE
After you complete the fresh air adjustment, a “XXX” may appear on
the display screen in place of a sensor reading. The buzzer and lights will be on steady, see the Calibration Alarms section at the end of this chapter to respond to this message.
2. Press and hold ↑ and ↓ for 3 seconds. The “FEATURES MENU ”
appears on the display and counts down from 3 seconds. Then displays
“BEGIN CAL”.
3. Press → to accept and begin calibration. If a Calword has not enabled, the “AUTO CAL” screen is displayed.
NOTE
If prompted by “CALWORD” and “***”, enter a password, see
Entering CalWord earlier in this chapter.
71-6417 37
Page 46
Innova Series Operator’s Manual
4. With the “AUTO CAL” message displayed. Press the to advance to the “MAN CAL” screen. Press to accept “MAN CAL”.
5. The screen displays the first active sensor. Press the to accept or press to advance to the next gas depending on what sensor you want to calibrate. To accept the desired selection, press →.
6. The default gas concentration is displayed with the ADJ icon and a double beep to indicate you may change the default concentration. The default concentration is set in the Auto Cal Gas Value, see Chapter 4, User Program, Setting Auto Cal Gas Values.
7. Press or to adjust the value to match the concentration of the calibration gas as read on the cylinder label.
8. Press to accept the displayed concentration. The display reads “BEGIN MAN CAL” alternating with “APPLY XX GAS” (where XX is the gas selected) and counts down from 45 seconds. If no gas is detected during this period, the screen returns to “BEGIN CAL”
9. Attach the tubing from the regulator to the probe tube, the Innova will draw gas from the gas cylinder.
When gas is detected, the display will read “CAL IN PROCESS” for XX seconds, with the SPAN icon on. Then the screen displays “CAL TIME LEFT” (counting down) alternating with the gas reading.
NOTE
During calibration, logging is disabled but calibration events are recorded in the log.
10. When calibration is complete, the display reads “MAN CAL PASSED” and then alternates between “>ACCEPT , ABOR T” and the gas reading. The gas label alternates with “PASS”.
11. Press to accept. The display will show “CAL ACCEPTED”, and then alternate between “REMOVE GAS” and the all active sensors gas
readings.
NOTE
If the calibration gas used interferes with any of the other sensors a non-zero reading may result, this does not affect the calibration of those sensor(s).
12. If you choose to abort, press to scroll to “ABORT” then → to accept the abort option. The display will read “CAL ABORTED” and then alternate between “REMOVE GAS” and all active sensor gas
readings.
38 71-6417
Page 47
Calibration
NOTE
The “WEAK” or “FAIL” flag may appear in place of the “PASS” flag
in the case of a weak or failed sensor. See the instructions in the Sensor Failed Alarms section later in this chapter for corrective action.
13. Remove the tubing from the probe. After fresh air has been detected, the
display will alternate between “ALARMS SILENCED”, and the
readings, and then return to the normal operating mode, after an
“ALARMS ACTIVE” message is displayed.
14. Unscrew the regulator from the cylinder, leave the components of the calibration kit assembled for convenience and place in the storage case.
Calibration Alarms
The Innova may alert you to one or more of the following alarms during calibration:
• Sensor Failed Alarm
• Weak Sensor
• Calibration Incomplete
Sensor Failed Alarm
If you attempt to air adjust or calibrate a faulty sensor, the Innova alerts you with a Sensor Failed Alarm.
I
NDICATIONS
• The backlight turns on automatically.
• The gas reading is replaced by XXX.
• Lights and Buzzer are steady.
R
ESPONDING TO A SENSOR FAILED ALARM
1. Complete the calibration procedure for the remaining sensors (if applicable), then return to the normal screen. The normal screen will continue to notify you of the sensor failure with a message.
2. Turn off the Innova, then verify that the sensor is installed properly, see Chapter 6, Maintenance, Replacing Sensors. Verify that the calibration gas value is correct in the Auto Cal Gas Valve Screen.
3. If the calibration gas value is correct, and the sensor is installed properly, turn the instrument on and repeat the air adjust and calibration procedures.
4. If the Fail condition persists, replace the sensor, see Chapter 6, Replacing Sensors, then calibrate the new sensor.
71-6417 39
Page 48
Innova Series Operator’s Manual
NOTE
You can enter the Innova User Program, Advanced Features Menu, and
change the sensor status setting to “OFF”. All information for an
“OFF” sensor is removed from the normal screen and recorded as a blank in the datalog.
Weak Sensor Warning
If the sensor that you are calibrating is nearing the end of its useful life, the Innova alerts you with a Weak Sensor message.
The gas flag will show “WEAK” at the conclusion of the calibration period.
R
ESPONDING TO A WEAK SENSOR
1. If you are manually calibrating the Innova, verify that you are correctly adjusting the gas concentration to match the concentration of the calibration gas. If using the Auto Cal method, verify that the gas concentration in the calibration gas cylinder matches the Auto Cal Gas Value, see Chapter 5, User Program, Setting Auto Cal Gas Values.
NOTE
A weak sensor is still functioning properly and can continue to be used. However, it is near the end of its useful life, and should be replaced as soon as possible, see Chapter 6, Maintenance, Replacing Sensors, for replacement instructions.
40 71-6417
Page 49
Chapter
6
MAINTENANCE
This chapter contains maintenance information to ensure proper operation and reliability of the Innova Series gas monitor.
WARNING
Perform all procedures in a “fr esh air” environment (environment known to be free of combustible and toxic ga ses and of normal oxygen content).
Preventive Maintenance
The following are daily, monthly, quarterly, and “as required” preventive maintenance suggestions to ensure the reliability of the Innova Series gas monitor.
NOTE
In instances where the Innova is in continual or everyday use, Thermo Fisher Scientific recommends that you perform preventive maintenance procedures more often than suggested in the following sections.
Daily
B
ATTERIES
The Innova should always contain fully-charged NiCd batteries or sufficiently powered alkaline batteries before each day’s use. You can verify the capacity of the batteries using the Display Menu. To verify battery capacity:
1. Press and hold the RANGE button, for three beeps, then release the
button.
If the display shows less than three bars, recharge the NiCd batteries or replace the alkaline batteries as described later in this chapter.
2. Press the ↑ or ↓ buttons to scroll to exit.
3. Press to return to the normal operation mode.
71-6417 41
Page 50
Innova Series Operator’s Manual
ALIBRATION
C
For optimum efficiency of the monitor, calibrate the Innova before each use.
If multiple calibrations over a period of days indicate that only a minimum of adjustments are required, the frequency of calibration can be changed to weekly or monthly, depending on how often the monitor is used, and how demanding the monitoring environment is.
NOTE
At the very least, “challenge” the normal operation of the oxygen (O2) sensor (if applicable) before every use.
Exhale over the inlet of the probe as you watch the display. The O
reading should decrease. When the O
reading decreases to 19.5%,
2
2
the alarm should activate. This confirms normal operation of the O2
AMPLE-DRAW SUBCOMPONENTS
S
Verify the proper operation of the flow alarm circuit by holding your finger
over the inlet of the probe for a few seconds. The pump shuts off, “PUMP FAILED”, and “PRESS → ΤΟ CLEAR” alternate with the gas readings display, the spinning icon changes to an “X”. The audible alarm sounds every second alternating/beeping. Press to restart the pump circuit. The icon will
change from an “X” or failed condition, and the pump will restart.
Monthly/Quarterly
ALIBRATION
C
Calibrate the sensors at least every one to three months. Calibration frequency depends on the frequency of use and also the environmental conditions in which you use the Innova.
As Required
A
LARM CIRCUITS
Periodically verify that all visual and audible alarms function properly.
WARNING
Verify alarm circuits in a “fresh air” environment (environment known to be free of combustible and toxic ga ses and of normal oxygen content).
To verify the alarm circuits, use a concentration of the proper gas sample that is greater than the preset warn or alarm levels. Verify that the LEDs illuminate and the buzzer sounds. Also verify that the display reading in alarm flashes during the alarm sequence.
42 71-6417
Page 51
Maintenance
AMPLE-DRAW SUBCOMPONENTS
S
Periodically check the probe, hoses, internal filter, and tubing for obstructions
that can accumulate over time. This is especially important if you use the Innova in a dusty or dirty environment. Replace the cotton and hydrophobic
filter elements if they become contaminated or discolored.
WARNING
Operation of instrument without probe/filter assembly attached will result in pump damage and possible impaired performance. Do not operate without probe/filter assembly attached.
Troubleshooting
NOTE
See Chapter 3, Start Up & Operation, for any of the various alarm indications that you may encounter while the Innova Series gas monitor is in normal use.
The troubleshooting section on the following pages describes symptoms and probable causes and recommends action for problems you may encounter with the Innova.
71-6417 43
Page 52
Innova Series Operator’s Manual
charger (same type). If the condition clears, the battery charger
is defective.
1. Verify that the battery selector switch is set to NI-CAD.
2. Verify that NiCd batteries are installed.
3. Replace the batteries.
4. Attempt to recharge the batteries with a different battery
for further instruction.
5. If the difficulty continues, contact Thermo Fisher Scientific,
adapter for longer than an 8-hour period, disconnect then
1. Verify that the battery selector switch is set to NI-CAD.
2. Verify that NiCd batteries are installed.
reconnect the battery charger to restart the charging timer.
3. If you are using the battery charger as a continuous operation
4. Replace the batteries.
,
operation adapter. If the condition clears, the battery charger is
defective.
5. Use a different battery charger (same type) as a continuous
for further instruction.
6. If the difficulty continues, contact Thermo Fisher Scientific
is set to ALK.
installed or alkaline
• The battery selector switch
• The batteries are not
not indicate that the
batteries are receiving a
charge.
NiCd batteries
Condition Symptom(s) Probable Cause(s) Recommended Action
Table 6-1 Troubleshooting
• The battery charger does
Unable to recharge
batteries are installed.
• After recharging, the
• The batteries are defective.
• The battery charger is
batteries do not have a
sufficient charge.
defective.
is set to ALK.
installed or alkaline
• The battery selector switch
Innova with the battery
• You cannot operate the
Unable to use the
battery charger as
batteries are installed.
• The batteries are not
charger/continuous
operation adapter in place.
a continuous
operation adapter
• The battery charger is
supplying a voltage that is
too high.
defective.
• The battery charger is
44 71-6417
Page 53
Maintenance
sensor as described later in this chapter.
2
Table 6-1 Troubleshooting (Continued)
supply of the calibrating gas.
1. Verify that the gas cylinder contains an adequate and fresh
or out-dated.
• The gas cylinder is empty
sensor’s “fresh air”
• You cannot set the
Unable to calibrate
a sensor
Condition Symptom(s) Probable Cause(s) Recommended Action
2. Replace the sensor as described later in this chapter.
3. If the difficulties continue, contact Thermo Fisher Scientific,
• The sensor is defective.
reading to zero.
• Sensor shows “xxx”.
sensor)
2
(except O
for further instruction.
for further instruction.
2. If the difficulties continue, contact Thermo Fisher Scientific,
sensor is defective. 1. Replace the O
2
• The O
reading to 20.9.
2
O
• You cannot fresh air the
sensor only)
2
Unable to calibrate
a sensor
(O
71-6417 45
Page 54
Innova Series Operator’s Manual
Recharging NiCd Batteries
WARNING
Recharge the batteries only in a “fresh air” environment (environment known to be free of combustible and toxic gases and of normal oxygen content).
Check NiCd battery capacity by using the Display Menu, see Chapter 4, User Program.
CAUTION
For optimum battery capacity, the batteries should be fully depleted before you recharge them. Repeated recharging of partially discharged batteries will reduce the battery capacity and operating time of the batteries.
Use only the battery chargers listed in Appendix A, Parts Lists, to recharge the batteries.
1. Verify that the battery selector switch is set to NI-CAD.
2. Verify that the voltage listed on the battery charger is the same as the AC outlet (115V or 220V), then plug the charger into the outlet.
3. Plug the other end of the charger into the CHGR jack at the side of the
Innova.
4. Verify that the charger’s amber light goes on. Allow the batteries to charge for at least 8 hours. Once the batteries are fully charged, the green light on the charger goes on (the amber light remains on).
NOTE
If the switch position is in the NI-CAD position, and alkaline batteries are installed, the battery charger will not charge alkaline batteries. If you attempt to recharge alkaline batteries, the amber light on the battery charger will flash.
A completely depleted battery recharges in approximately eight hours. The battery charger automatically cuts the charging current to a sustaining rate when the batteries are fully recharged. You can leave the charger plugged in indefinitely.
46 71-6417
Page 55
Maintenance
Replacing Components
WARNING
Replace components in a “fresh air” environment (environment known to be free of combustible and toxic ga ses and of normal oxygen content).
Replacing Batteries
WARNING
When replacing batteries, use only the batteries supplied by Thermo Fisher Scientific (refer to Appendix A, Parts List). They are a special, high capacity type and are required to maintain agency certification of your Innova Series.
• Duracell MN1300 (Alkaline)
• SAFT 410907-104 (NiCd) To maintain Cenelec certification of your Innova Series monitor,
the hex head screw must be installed to secure the battery lid in place.
1. Turn off the Innova by pressing and holding the ON/OFF button
while the Innova sounds three beeps.
2. Turn the captive screw at the middle of the battery compartment door counterclockwise, then remove the door.
3. Remove all four spent batteries.
4. Verify that the ALK/NI-CAD switch is set to the proper setting, see
Chapter 6, Maintenance, Figure 6-1 Innova Battery Compartment.
WARNING
Never mix NiCd batteries together with alkaline batteries, this can cause internal damage to the monitor.
5. Install four new batteries (refer to the raised picture in the battery compartment for the proper orientation of the batteries).
NOTE
Always replace all four batteries at the same time
6. Install the battery compartment door and tighten screw until snug against gasket.
71-6417 47
Page 56
Innova Series Operator’s Manual
A
lkaline/Ni-Cad
battery switch
Figure 6-1 Innova Battery Compartment
WARNING
If applicable, follow all federal, state, or local regulations regarding the disposal of alkaline and NiCd batteries.
Some toxic gas sensors require up to 15 minutes to stabilize, after you install the batteries. If your Innova includes a toxic gas sensor(s), do not turn on and use the Innova during this period. If the unit is started up immediately after installing batteries, the “BIAS” message may appear, see Chapter 6, Maintenance, for instructions on installing or recharging batteries.‘
48 71-6417
Page 57
Replacing Sensors
WARNING
Recalibrate the Innova whenever you replace a sensor. If applicable, follow all federal, state, or local regulations
regarding the disposal of electrochemical sensors. The toxic sensors contain sulfuric acid. The O
sensor contains an alkali
2
solution.
Maintenance
TOX2 PSW
TOX1
2346R51
PUMPYOXY
T B
W
O
P
B
1
Figure 6-2 Innova Series Sensor Locations and Connection
Assignments
1. Turn off the Innova.
CAUTION
To avoid damage to internal circuitry , r emove the batteries befor e opening the main housing.
2. Open the Innova, then separate the two halves.
3. Locate the sensor you want to replace in the flow block, see Figure 6-2.
71-6417 49
Page 58
Innova Series Operator’s Manual
1
2
3
4
5
6
Figure 6-3 Removing a Sensor from the Innova
NOTE
Figure 6-2 illustrate sensor assignments for the standard Innova. The sensor assignment for your Innova may be different depending on the Innova model you are using. For models that have less than four sensors, empty sensor cavities are filled with sensor plugs.
4. Disconnect or unplug the sensor’s connection to the Innova.
CAUTION
If you use a tool to loosen the sensor from the flow block, be careful not to damage the sensor housing or the flow block.
5. Using your fingers, tilt the sensor from side to side to loosen it, then pull the sensor straight out of its flow block cavity, see Figure 6-3.
6. Insert the replacement sensor into the cavity, then apply downward pressure on top of the sensor until it is firmly seated in the cavity. For toxic sensors, make sure not to bend or damage the sensor pins.
7. Reconnect the sensor’s connection to the Innova main board.
For the combustible gas sensor, connect the purple, white, and red wires to the proper terminals (P, W, R) on the terminal block (TB1).
For the remaining sensors, plug the sensor wire connector into the applicable socket (OXY, TOX1, or TOX2).
50 71-6417
Page 59
8. Close the Innova housing and re-install batteries.
CAUTION
Some toxic gas sensors require up to 15 minutes to stabilize, after you install the batteries. If your Innova includes a toxic gas sensor(s), do not turn on and use the Innova during this period. If the unit is started up immediately after installing batteries, the “BIAS” message may appear, see Chapter 6, Maintenance, for instructions on installing or recharging batteries.Calibrate the Innova before you use it again.
Replacing the Pump
CAUTION
To avoid damage to internal circuitry , r emove the batteries befor e opening main housing.
1. Turn off the Innova, then separate the halves.
2. Locate the pump in the flow block, see Figure 6-2.
3. Disconnect the pump wire connector from the main board socket
labeled PUMP.
4. Remove the nut that secures the pump bracket to the pump, then remove the bracket.
5. Remove the pump from the flow block by rocking it back & forth.
Maintenance
CAUTION
When you insert the new pump, do not place pressure on the pump motor. The pump motor is the dark-colored circular component at the top of the pump.
6. Before installing the new pump, apply a thin film of petroleum jelly lubricant or O-ring grease to the outside of the barbs. Be careful not to get lubricant into the end of the barbs, as this could clog the pump.
CAUTION
Do not use lubricant that contain silicone, as this can poison the combustible sensor.
7. Place the pump bracket in its previous position, then secure the bracket to the pump with the nut you removed in step 4.
8. Reconnect the pump wire connector to the main board socket labeled
PUMP, close and secure the Innova housing, and re-install batteries.
71-6417 51
Page 60
Innova Series Operator’s Manual
Replacing the Internal Filter
CAUTION
To avoid damage to internal circuitry , r emove the batteries befor e opening main housing.
1. Turn off the Innova, remove batteries, and then separate the halves.
2. Locate the internal filter, see Figure 6-2. It is connected to the inlet fitting by a small polyurethane tube.
3. Disconnect the polyurethane tubes from the slip-on barbs on the filter.
4. Push the tubes onto the barbs on the new filter, close and secure the Innova, and re-install batteries.
Replacing the Probe’s Hydrophobic Filter
1. Unscrew the probe body’s two halves. Be careful not to lose the o­rings that seal the filter.
2. Remove the filter and cotton ball from the probe half that mounts to the Innova (you may need to use a small screwdriver to pry the filter loose).
3. Install the new filter and cotton ball in place of the filter and cotton ball you removed.
4. Make sure the O-rings are in place, then reassemble the probe halves.
52 71-6417
Page 61
Appendix
A
PARTS LIST
Table A-1 lists replaceable parts and accessories for parts common to all Innova Series. See Parts List in Appendix C for supplemental parts unique to particular models.
Table A-1 Parts List and Accessories
Part No. Description
06-1255 Tubing, polyurethane, 5/16 in. OD x 3/16 in. ID
10-0491 Screw, hex head
13-0110 Shoulder strap, adjustable
20-0650 Carrying case, Innova Series instrument
20-0651 Case, confined space kit (case only)
30-0042 Pump
30-0044 Kit, rebuild pump (30-0042/30-0042-01)
33-0152-01 Filter, Hydrophobic (for 80-0187 probe)
33-0160 Filter, internal
33-1031 Cotton ball, pkgs. of 24 (for 80-0187 probe)
49-1201 Battery, alkaline, D size
49-1501-01 Battery, NiCd, D size
49-2149 Charger, NiCd batteries, dual-rate, 230 VAC
49-2150 Charger, NiCd batteries, dual-rate, 115 VAC
49-2151 Charger, NiCd batteries, dual-rate, 12 VDC
52-2033 Remote audible alarm, with lapel clip
52-2033-01
61-0123 Sensor, combustibles (LEL/ppm)
65-0612 Sensor, oxygen (O
65-0613 Sensor, oxygen (Fuel Vapor only)
65-2417 Sensor, sulfur dioxide (SO
65-2426-01 Sensor, carbon monoxide (CO)
Remote audible alarm, with 20-foot cord and magnet
)
2
)
2
71-6417 53
Page 62
Innova Series Operator’s Manual
Table A-1 Parts List and Accessories
Part No. Description
65-2426-02 Sensor, hydrogen sulfide (H2S)
65-2431-01 Sensor, chlorine (Cl
65-2431-02 Sensor, hydrogen cyanide (HCN)
65-2431-03 Sensor, nitrogen oxide (NO)
65-2431-07 Sensor, ammonia (NH
65-2431-08 Sensor, phosphine (PH
65-2431-10 Sensor, nitrogen dioxide (NO
71-6416 Quick Reference Card
71-6417 Operator’s Manual
77-0043 Kit, Cenelec (includes allen wrench and hex head screw)
80-0187 Probe with hydrophobic filter
80-0187-04 Probe, fiberglass, 30-inch, with hydrophobic filter
80-0211 Moisture trap
80-0405 Dilution fitting, 50%
80-0504 Hose, teflon, 3 feet (for Cl2/Fuel & Soil Vapor)
)
2
)
3
)
3
)
2
80-0505 Hose, polyurethane, 5 feet
80-0511 Hose, polyurethane, 10 feet
80-0516 Hose, polyurethane, 15 feet
80-0520 Hose, polyurethane, 20 feet
80-0525 Hose, polyurethane, 25 feet
80-0530 Hose, polyurethane, 30 feet
80-0550 Hose, polyurethane, 50 feet
80-0599 Hose, polyurethane, 100 feet
81-0002 Cylinder, 2% H
(17 liter)
2
81-0007 Cylinder, 40% LEL hexane (17 liter)
81-0012 Cylinder, 50% LEL CH
(17 liter)
4
81-0064 Cylinder, 50 PPM CO (17 liter)
81-0078 Cylinder, 100% N2 (17 liter)
81-0151 Cylinder, 25 PPM H
S (58 liter)
2
54 71-6417
Page 63
Table A-1 Parts List and Accessories
Part No. Description
81-0153 Cylinder, 50% LEL CH4/ 40% CO2/10% N2 (17 liter)
81-0154 Cylinder, 50% LEL CH
81-0155 Cylinder, 50% LEL CH
81-0170 Cylinder, 5 PPM SO2 (58 liter)
81-0175 Cylinder, 25 PPM NO (58 liter)
81-0180 Cylinder, 5 PPM NO
81-0190 Cylinder, 5 PPM Cl
81-0191 Cylinder, 25 PPM NH
81-0192 Cylinder, 10 PPM HCN (58 liter)
81-0193 Cylinder, 0.5 PPM PH3 (58 liter)
81-0226-01 Cal kit, O
81-0226-02 Cal kit, CH
2
4
81-0226-03 Cal kit, CO
81-0226-04 Cal kit, CH
(Land Survyor)
4
81-0226-05 Cal kit, Hexane (Soil Vapor)
/25 PPM H2S/50 PPM CO/12% O2 (58 liter)
4
/50 PPM CO/12% O2 (103 liter)
4
(58 liter)
2
(58 liter)
2
(58 liter)
3
Appendix A
81-0226-06 Cal kit, Hexane/O2 (Fuel Vapor)
81-0226-07 Cal kit, H
81-0226-08 Cal kit, H
81-0227-01 Cal kit, H
(Transformer Gas Tester)
2
(Transformer Gas Tester)
2/O2
S
2
81-0227-02 Cal kit, Cl2
81-0227-03 Cal kit, HCN
81-0227-04 Cal kit, NH
3
81-0227-05 Cal kit, NO
81-0227-06 Cal kit, No
81-0227-07 Cal kit, SO
81-0227-08 Cal kit, PH
2
2
3
81-0227-09 Cal kit, CH4/O2/CO
81-0227-10 Cal kit, CH
81-0227-11 Cal kit, CH
4/O2
4/O2
/CO/H2S
/CO/Cl
2
71-6417 55
Page 64
Innova Series Operator’s Manual
Table A-1 Parts List and Accessories
Part No. Description
81-0227-12 Cal kit, CH4/O2/CO/HCN
81-0227-13 Cal kit, CH
81-0227-14 Cal kit, CH4/O2/CO/NO
4/O2
/CO/NH
3
81-0227-15 Cal kit, CH
81-0227-16 Cal kit, CH4/O2/CO/SO
81-0227-17 Cal kit, CH4/O2/CO/PH
4/O2
/CO/NO
2
2
3
81-0227-18 Cal kit, CH4/O2/H2S/HCN
81-0227-19 Cal kit, CH
81-0227-20 Cal kit, CH
4/O2/H2
4/O2/H2
S/NO
S/SO
2
81-0228-01 Cal kit, CH4/O2/Cl2/NO
81-0228-02 Cal kit, CH
81-0228-03 Cal kit, CH4/O2/HCN/NH
4/O2
/Cl2/PH
3
3
81-0228-04 Cal kit, CH4/O2/HCN/NO
81-0228-05 Cal kit, CH
81-0228-06 Cal kit, CH4/O2/NH3/NO
81-0228-07 Cal kit, CH4/O2/NH3/PH
81-0228-08 Cal kit, CH4/O2/NO/NO
81-0228-09 Cal kit, CH4/O2/NO/PH
81-0228-10 Cal kit, CH4/O2/NO2/PH
81-0228-11 Cal kit, CH4/O2/SO2/PH
4/O2
/HCN/PH
3
2
3
2
3
3
3
81-0228-12 Cal kit, CH4/O2/CO (Land Surveyor)
81-0228-13 Cal kit, CH
81-1063 Regulator, demand flow (34/58/103 liter cylinders)
81-1064 Regulator, demand flow (17 liter cylinders)
81-6500 Confined space kit (non H
81-6501 Confined space kit (any H
82-0200 1/8” hex wrench
82-5032 Cable, serial
82-5071 Data retrieval kit
56 71-6417
S (Land Surveyor)
4/O2/H2
S)
2
S version)
2
Page 65
Appendix
B
INTERFERENCE FACTORS
The following factors interfere with the accurate gas measurement capability of the Innova:
• Filament poisoning
• Rich mixtures
• Oxygen-deficient mixtures
• Oxygen-enriched mixtures
• Interfering gases
• Response to various combustible gases
Filament Poisoning
Silicone vapors and chlorine and fluorine compounds, such as chlorinated hydrocarbons, can damage the HC sensor and interfere with accurate LEL readings.
Such compounds, even in small proportions, should be avoided. Verification checks on known gas samples are necessary, if the possibility of exposure to these compounds exists.
High Concentration of Combustible Gas
With rich mixtures of combustible gases, the instrument displays the following indications (refer to Table B-1)
Table B-1 Indications with High Combustible Gas
Concentrations
Cause Indication
Mixtures up to 100% LEL Reading displayed.
Mixtures between LEL and Upper Explosive Limit (UEL)
Mixtures above the UEL The reading increases to 100% LEL, then
71-6417 57
100% displayed.
comes back down. Very rich mixtures read close to zero.
Page 66
Innova Series Operator’s Manual
Oxygen-Deficient Mixtures
Samples containing less than 10% oxygen may give low LEL readings by starving the combustible gas reaction in the sensor. Such mixtures will also cause a low oxygen alarm. As a rule, 10% oxygen or more gives a full reading on any combustible gas up to the LEL.
Oxygen-Rich Mixtures
Samples containing combustible gases that have more than normal proportion of oxygen will give a normal reading. However, they should be avoided because the flame arrestor used is not dense enough to arrest flames from combustible gas in oxygen. Such mixtures will also cause a high oxygen alarm.
WARNING
Do not attempt to use the Innova on samples of combustible gas in oxygen; for example, oxy-acetylene mixtures. Instead, use sample mixture in air.
Interfering Gases
The CO sensor responds to some other gases besides CO. The CO sensor has a charcoal filter that prevents interference from H gases. However, continuous exposure may saturate the filter and break through as CO indication. Hydrogen, ethylene, and acetylene will pass quickly through the filter, and thus are strong interferences, while heavier olefins and aromatic hydrocarbons will break through in time, making sensor replacement necessary.
S and some of these other
2
58 71-6417
Page 67
Appendix B
Relative Combustible Response
The Innova is normally calibrated with methane gas, but responds to a variety of hydrocarbons (refer to Table B-2).
Table B-2 Relative Response of New Innova Series LEL/ppm Hydrocarbon Indicator to Various Gases and Vapors
LEL RANGE PPM RANGE
GAS or VAPOR
Acetone 2.5 0.73 1.7 1.6 0.48
Acetylene 2.5 0.63 1.5 1.2 0.34
Benzene* 1.2 1.0 2.3 1.0 0.30
Butadiene, 1, 3- 2.0 0.52 1.21 0.98 0.29
Carbon Monoxide 12.5 0.39 0.9 5.9 1.8
Chloroform* NONE - - 9.8 3.0
Dimethyl Formamide*
Ethyl Acetate* 2.0 0.9 2.1 1.6 0.48
Ethyl Alcohol 3.3 0.70 1.6 2.1 0.64
Heptane 1.05 1.00 2.3 0.99 0.30
Hexane 1.1 1.00 2.3 1.0 0.30
Hydrogen 4.0 0.33 0.7 1.3 0.39
Methane 5.0 0.43 1.0 3.3 1.0
Methyl Chloroform 7.5 0.68 1.6 4.9 1.5
Methyl Ethyl Ketone 1.4 1.09 2.5 1.7 0.52
Pentane 1.5 0.71 1.7 1.0 0.30
Propane 2.1 0.56 1.3 1.3 0.39
Propyl Acetate* 1.7 0.77 1.8 1.4 0.42
Propyl Alcohol, iso- 2.0 0.89 2.1 1.9 0.58
Propyl Alcohol, n- 2.2 0.81 1.9 1.9 0.58
Styrene Monomer 0.9 4.2 9.7 2.7 0.82
Toluene 1.1 1.2 2.9 1.2 0.36
Trichloroethylene 8.0 1.0 2.4 6.8 2.1
Xylenes, isomer mix .9-1.1 3.0 7.0 1.7 0.52
* These response factors are estimates, based on factors for 6 volt Thermo Fisher Scientific sensor adjusted for characteristics of Innova
sensor.
Please note that response varies from one sensor to another and the relative response of a sensor can change with the age of the sensor. These data should be used for estimation purposes only.
Assuming an instrument calibrated directly for hexane or methane, but used to observe a different gas, the e quivalent response in % LEL (or ppm) for that gas is secured by multiplying the observed% LEL (or ppm) reading by the LEL (or ppm) scale conversion factor.
% LEL
VOL
2.2 1.4 3.3 2.7 0.81
CONVERSION FACTOR: CONVERSION FACTOR:
VS. HEXANE
VS.
METHANE
VS. HEXANE
METHANE
VS.
71-6417 59
Page 68
Innova Series Operator’s Manual
60 71-6417
Page 69
Appendix
C
SUPPLEMENTAL INSTRUCTIONS
Innova Land Surveyor (LS)
The Innova LS consists of two different models (PPM/% LEL/% GAS and PPM/%LOW/%HI) that permits measurements of methane gas in three ranges of sensitivity:
Table C-1 Ranges for Innova LS
PPM/%LEL/%GAS PPM/%LOW/%HI
0-10,000 PPM 0-10,000 PPM
0-100% LEL 0-5.0% Gas by VOL [%LOW], equivalent to 0-100% LEL
0-100% GAS by VOL* 0-100% Gas by VOL [%HI]*
* By means of Thermal Conductivity Sensor
Operation
For the PPM/%LOW/%HI, see the text in brackets [ ].
Operation in the first two ranges is exactly as described in the text of this manual, except that if the PPM reading should exceed its range (10,000 PPM) the range will automatically switch to the % LEL [%LOW] range. If the % LEL [%LOW] reading should exceed 100% LEL, equivalent to 5% methane, the Innova LS will automatically switch to the 0-100% GAS range [%HI] range. The display will show the legend “% GAS” [“%HI”] following the gas reading while in this range.
The % GAS [%HI] range may also be entered manually from the % LEL [%LO] range by pressing the RANGE button twice. If it is likely that high concentrations will be encountered, to prolong the life of the LEL/PPM sensor, start out in the % GAS [%HI] range. The Innova LS does not automatically range down. To exit the % GAS [%HI] range and return to the LEL [%LOW] range press the RANGE button twice. There is a 30 second warm-up period after the transition, during this period the display count from W 30 to W 1 before gas readings are displayed. To manually switch between the ppm and % LEL (%LO) ranges, press the RANGE button once.
71-6417 61
Page 70
Innova Series Operator’s Manual
NOTE
If the sample is very high in methane content and also very low in oxygen content, the unit may not auto-range upscale. If during sampling the % LEL [%LOW] indication rises rapidly and then decays, manually switch to the % GAS [%HI] range as described above to determine the methane content.
Precautions
Be aware that the methane/air mixtures in the range 5% to 15% by volume are highly explosive. The Innova is Intrinsically Safe, and is not a source of ignition, but there are many common ignition sources such as static electrical discharges or sparks from metal tools that can ignite such mixtures. The relationship between ranges is correct only for methane. Some rough approximations may be made for gases such as propane and butane, but values will not be accurate, and should not be used for safety considerations.
WARNING
Initiate Fresh Air Zero adjustment only in a non-hazardous environment free of combustible or toxic gas content and consisting of normal oxygen content. T o do othe rwise could cause a serious error in the indicated gas level shown on the display, and result in failure of the Innova LS to properly indicate the degr ee of hazard. Also, use care not to initiate Fresh Air Zero immediately after a range change, as this can cause an error in the “clean air” baseline.
62 71-6417
Page 71
Appendix C
Land Surveyor Calibration (Auto)
Calibration is as described earlier in this manual, with one exception: the % GAS [%HI] sensor is calibrated separately from the Catalytic (PPM/%LEL or PPM/%LO) sensor. Therefore a multi-step calibration process has been programmed into the Auto Calibration routine. First, the standard sensors are calibrated using a multi-gas mixture. Then TC sensor is calibrated using a multi-gas cylinder, containing 50% CH
(Nitrogen).
and N
2
(Methane), CO2 (Carbon Dioxide),
4
1. A separate air adjust is required for the % LEL and % GAS ranges. Perform an air adjust in the % LEL and % GAS ranges before calibration.
2. Begin the calibration process by performing an Auto Calibration. The TC sensor will be ignored during the multi-gas calibration.
3. When the multi-gas Auto Calibration is complete, a double beep prompts the user to begin the TC sensor calibration. The default gas concentration is displayed with the ADJ icon to indicate you may change the concentration. If the concentration on the gas cylinder does not match the concentration displayed.
4. Press to begin the calibration.
5. Apply the calibration gas and follow the prompt to complete the calibration process.
Land Surveyor Calibration (Manual)
Manual calibrations are performed as described in Chapter 5, Calibration. Once in the manual calibration screen, the location of the sensor(s) are as illustrated below (CH
version shown).
4/O2
CH4 O2 TC
1. Press to accept or to advance to the next gas depending on what sensor you want to calibrate.
71-6417 63
Page 72
Innova Series Operator’s Manual
Pump
The Land Surveyor uses a pump capable of drawing a sample from a source that is at a vacuum of up to 60” of water column. If pump replacement is required, the pump listed in the Parts List below must be specified.Parts List
Parts List
See Parts List in Appendix A for a more complete list of replacement parts and accessories.
Replacement Parts
30-0042-01 Pump, Land Surveyor
61-0124 Sensor, Thermal Conductivity
81-0012
81-0226-04 Cal kit, CH
81-0228-12 Cal kit, CH
81-0228-13 Cal kit, CH
Cylinder, 50% LEL CH
4
/CO
4/O2
S
4/O2/H2
(17 liter)
4
81-0153 Cylinder, 50% CH4/40% CO2/10% N2 (% gas range) (17 liter)
81-0154 Cylinder, 50% LEL CH
81-0155 Cylinder, 50% LEL CH
/25 PPM H2S/50 PPM CO/12% O2 (58 liter)
4
/50 PPM CO/12% O2 (103 liter)
4
81-1063 Regulator, demand flow (34/58/103 liter cylinders)
81-1064 Regulator, demand flow (17 liter cylinders)
64 71-6417
Page 73
71-6417 65
Appendix C
Innova Transformer Gas Tester (XT)
Overview
The Innova Transformer Gas Tester is factory configured to monitor for total combustibles from the gas space of power transformers. If transformer faults, such as overheating, hot spots, arcing or corona occur, various combustible gases are generated. If left unchecked, gas can accumulate creating a potential for explosion, failure of the transformer, significant remediation costs and loss of electrical service.
The Innova Transformer Gas Tester is designed as an inspection tool to determine if unacceptable levels of gas exist within the transformer. A sample of gas is extracted from the headspace of the transformer by the Innova Transformer Gas Tester with a 1:1 dilution fitting in place. If gas levels exceed acceptable levels, additional inspection, including performing a dissolved gas analysis, is required to determine fault condition and appropriate action. If oxygen detection is required, an optional oxygen range can be provided.
Detection Range
The Innova Transformer Gas Tester has two detection ranges: 0-100% LEL and 0-10,000 ppm combustibles. An optional range of 0-30% oxygen can be provided if requested.
Alarm Points
The alarm points for the combustibles (LEL) range are preset at 10% LEL (WARN) and 50% LEL (ALARM). PPM combustibles alarms are set at 1000 and 5000 ppm. All alarms are completely user adjustable if alternate alarm points are required. Refer to Chapter 4, Edit Alarms, for alarm adjustment instructions.
The Innova Transformer Gas Tester can also be provided with an optional oxygen channel. The oxygen channel has two user adjustable alarms, one that activates at 19.5% decreasing oxygen and one that activates at 23.5% increasing oxygen.
Page 74
66 71-6417
Innova Series Operator’s Manual
Dilution Fitting
The Innova Transformer Gas Tester includes a 1:1 dilution fitting to provide the necessary dilution ratio to mix one part sample with one part air.
Description
The catalytic detection method requires oxygen in order to function properly. In an inert gas environment, it is necessary to provide oxygen by the use of the 1:1 dilution attachment on the inlet fitting to oxidize the combustible gas.
The dilution fitting consists of a small tee fitting which plugs into the inlet fitting of the instrument. It includes two small drilled orifices. One orifice connects to the sample source and the other to atmospheric air. These are proportioned to give a dilution of one part air to one part of sample.
If testing for combustible gas in a inert environment is attempted without the dilution fitting, the Innova will respond briefly as the available oxygen is exhausted in the flow block and the combustible readings will then return to zero, providing an incorrect reading.
For testing oxygen content in transformers, the dilution fitting can be removed. This will allow you to determine if there is oxygen present. This test however, can only be performed with the Innova Transformer Gas Tester with an active oxygen channel.
Operation
Before operating the Innova Transformer Gas Tester, it is important that you read and understand the Innova Series Operator’s Manual. It is also extremely important that you are familiar with the specific transformer that you are testing and know how to safely extract a gas sample.
1. Turn on the Innova Transformer Gas Tester and allow the instrument to warm up.
2. Perform fresh air adjustment as described in Chapter 2, Start Up & Operation.
3. Connect the probe directly to the end of the Innova Transformer Gas Tester.
4. Open the valve on the Tedlar bag by turning the top of the valve counter clockwise two full turns.
5. To remove excess air from the Tedlar bag, connect the tubing from the Tedlar bag to the end of the probe allowing the internal pump to draw the air out of the gas collection bag. When the Innova alarms and indicates “PUMP FAILED”, turn the valve clockwise until it is completely closed, then remove the tubing from the end of the probe.
Page 75
71-6417 67
Appendix C
6. Press the button to restart the pump.
7. Remove the probe and attach the dilution fitting directly to the inlet fitting of the Innova.
8. Attach the probe to the dilution fitting.
9. Attach the Tedlar bag tubing to the gas sample port of the transformer.
10. Open the valve on the bag two full turns then fill the bag with gas from the head space of the transformer, closing the valve when the bag is full.
11. Remove the gas collection bag.
12. Open the valve on the collection bag and attach the bag directly to the inlet of the probe on the Innova Transformer Gas Tester.
13. View the Innova Transformer Gas Tester display to determine the concentration of combustible gas in the transformer.
14. Remove the gas collection bag from the probe when complete and close valve on bag.
NOTE
When testing for oxygen content, either remove the dilution fitting from the Innova Transformer Gas Tester or place your finger over the small orifice on the dilution fitting before performing test.
Calibration
The Calibration procedure for the transformer gas tester version is identical to the procedure as outlined in the calibration section of this manual. The calibration gas cylinder is 25% LEL Hydrogen, 12% Oxygen, balance Nitrogen. Users should enter the Autocal gas value for the LEL range as 50% LEL, to compensate for the dilution probe and to provide the correct readings when the 1:1 dilution fitting is used. The dilution fitting
must be
removed for calibration.
The Innova Transformer Gas Tester is calibrated with 1% hydrogen (25% LEL) in the LEL range.
NOTE
The Innova XT is calibrated to allow for correct readings when used with the 1:1 dilution probe in %LEL mode or without the dilution probe in PPM mode.
Page 76
68 71-6417
Innova Series Operator’s Manual
Innova Transformer Gas Tester (XT) Parts List
See Parts List in Appendix A for a more complete list of Innova replacement parts and accessories common to all models.
80-0405 Dilution fitting, 50%
81-0013 Cylinder, 1% H2/12% O2 in N
2
81-0227-21 Cal kit, H2/O2 (Transformer Gas Tester)
81-1064 Regulator, demand flow (17 liter cylinders)
81-1155 Tedlar bag w/tubing
Page 77
71-6417 69
Appendix C
Soil Vapor (SV) and Fuel Vapor (FV)
Operation
The Soil Vapor Monitor has a combustible sensor only; the Fuel Vapor Monitor has an oxygen sensor in addition to the combustible sensor. The Soil and Fuel Vapor Monitors have two operating modes, “CH4” mode and “ME” mode. When in the “CH4” mode, the Innova functions as a normal instrument, and the unit will respond to all combustible gases, including methane. When in the “ME” (Methane Elimination) mode, the voltage to the combustible sensor is changed, which causes the sensor to respond only to non-methane combustible gases. This allows the user to discriminate between methane and non-methane hydrocarbons.
Table C-2 Sensor Specifications
Sensor
Sensor
Typ e
Detection
Range
Warn
Setting
Alarm
Setting
Accuracy
(of full-scale
range)
Repeatability
(of calibration
span value)
Hydrocarbon Catalytic 0 to 100
% LEL
10
% LEL
50
% LEL
+ 5%* + 2%
Hydrocarbon Catalytic 0 to 1,000
ppm
100
ppm
500
ppm
+ 25% + 5%
Specifications apply with 40% hexane calibration.
*The + 5% specification in %LEL applies when in “ME” (methane elimination) mode. Readings taken in “CH4” mode have
+ 10% accuracy.
NOTE
The Innova FV and SV instruments provide a 90% measurement value within 30 seconds. For the most accurate readings, allow the unit to measure for 2 minutes.
Page 78
70 71-6417
Innova Series Operator’s Manual
Range Mode and Operating Mode Selection
After the Innova has completed its warm-up cycle, the letters “CH4” and “ME” will appear under the gas reading on the display, indicating the mode of operation. If desired, the mode may be changed by pressing the RANGE button twice.
To change the hydrocarbon range from LEL to PPM, press the RANGE button once.
CAUTION
Note that if the mode of operation is changed to “ME” operating mode, the unit will no longer respond to methane or natural gas. If it is suspected that methane or natural gas may pose a hazard, the unit should first be operated in the “CH4” operating mode to verify that no methane or natural gas is present.
Page 79
71-6417 71
Appendix C
Calibrating the Innova SV and FV
The Innova automatic calibration (AUTO CAL) displays all active sensors on the display screen at the same time. You can calibrate all active sensors simultaneously using a gas cylinder available from Thermo Fisher Scientific.
WARNING
The instruments is calibrated in the methane elimination (ME) mode using the %LEL range mode. The instrument is calibrated with hexane. To measure other gases, use Table B-2 in Appendix B.
The concentrations set in the Auto Cal Gas Value screen must match the multi-gas mixture, see Chapter 4, User Program, Setting Auto Cal Gas Va lu e s.
NOTE
The Innova Soil and Fuel Vapor Monitors use a 3-foot, teflon-lined hose to minimize absorption of hydrocarbon vapors. Use of longer hose lengths is not recommended.
Auto Calibration
WARNING
Calibrate the Innova in a “fresh air” environment (environment known to be free of combustible and toxic gases and of normal oxygen content).
NOTE
If you use the automatic adjustment method to calibrate the sensor, make sure the gas cylinder contains the same concentrations of the target gas as the values set in the “AUTOCAL GAS VAL” menu, see Chapter 4, User Program, Setting Auto Cal Gas Values.
Page 80
72 71-6417
Innova Series Operator’s Manual
The Innova FV and SV models must be zeroed in both ME mode and CH4 mode. If you plan to measure in the %LEL range mode, zero the instrument in %LEL mode. If you plan to measure in the PPM range mode, zero the instrument in PPM mode.
1. Enter the methane measurement (CH4) operating mode. Allow the system to run for 2 minutes drawing ambient air.
2. Hold down the AIR button for 3 seconds until the display counts down and the bar graph starts to scroll, then the display will read “DONE” once the “fresh air” reading for all active sensors is complete.
3. Enter the methane elimination (ME) operating mode. Allow the system to run for 2 minutes drawing ambient air.
4. Hold down the AIR button for 3 seconds until the display counts down and the bar graph starts to scroll, then the display will read “DONE” once the “fresh air” reading for all active sensors is complete.
NOTE
After you complete the fresh air adjustment, a “XXX” may appear on the display screen in place of a sensor reading. The buzzer and lights will be on steady, see Calibration Alarms section at the end of this chapter to respond to this message.
5. Press and hold and for 3 seconds. The message “FEATURES
MENU” appears on the display and counts down from 3 seconds, and then displays “BEGIN CAL”.
6. Attach the tubing from the regulator to the probe tube, the Innova will draw gas from the gas cylinder. Allow the gas to run for 1 minute.
7. Press to accept and begin a calibration. If a CalWord has not enabled, the “AUTO CAL” screen will display and you can proceed with the calibration. If CalWord, is enabled, refer to the following section “Entering a CalWord” then continue with the calibration procedure.
NOTE
If prompted by “CALWORD” and “***”, enter a password as described in the next section.
Page 81
71-6417 73
Appendix C
8. Press to accept “AUTO CAL”. The screen displays “BEGIN
AUTOCAL” alternating with “APPLY GAS” counting down from
30 seconds. If no gas is detected during this period, the screen returns to “BEGIN CAL”.
9. First, the display will read “CAL IN PROCESS”. Then the screen displays “CAL TIME LEFT” (counting down from 75 seconds) alternating with the gas readings.
NOTE
During calibration, datalogging is disabled but calibration events are recorded in the log.
10.When calibration is complete, the display reads “AUTOCAL PASS ED” and then alternates between “>ACCEPT, ABORT” and the gas readings. The gas labels alternate with ‘PASS”. Press → to accept. The display will show “CAL ACCEPTED”, and then alternate between “REMOVE GAS” and the gas readings.
If you choose to abort, press to scroll to “ABORT” then → to accept the abort option.
The display will read “CAL ABORTED” and then alternate between “REMOVE GAS” and the gas readings.
NOTE
A “WEAK” or “FAIL” flag may appear in place of the “PASS” flag in the case of a weak or failed sensor. See the instructions in the Calibration Alarms section, later in this chapter for corrective action.
11.Remove the tubing from the probe. After fresh air has been detected, the display will alternate between “ALARMS SILENCED”, and the readings, and then return to the normal operating mode, after an “ALARMS ACTIVE” message is displayed.
12.Unscrew the regulator from the cylinder, leave the components of the calibration kit assembled for convenience and place in the storage case.
NOTE
After the calibration, refer to the “Conversion Factors” section for information on measuring methane (CH4 operating mode) after the instrument has been calibrated with hexane.
Page 82
74 71-6417
Innova Series Operator’s Manual
Entering a CalWord
A password may be used to add extra security to the Calibration with a “CALWORD”, see Chapter 4, User Program, Create/Edit CalWord/ PassWord/LogWord. If turned on, a code must be entered before the user may perform a calibration.
To enter a CalWord (if activated):
1. If the CalWord function is enabled, the display reads “CAL WORD” and “* * *” as the calibration procedure begins. The cursor is on the first asterisk.
2. Press or to scroll to the correct character.
3. Press to accept. The cursor moves to the next character.
4. Repeat steps 3 and 4 until all three characters are accepted. If correct, the next display reads “AUTOCAL”. If the CalWord is incorrect, the display reads “BEGIN CAL” message. Begin at step 1 to start over.
Conversion Factors
Because the instrument is calibrated with hexane, a conversion factor must be applied when measuring other gases. Refer to Table B-2 for the conversion factors. Thermo Fisher Scientific provides estimates of these values, but the response varies from instrument to instrument and these data should be used for estimation purposes only.
CAUTION
Note that the response varies from one sensor to another and the relative response of a sensor can change with the age of the sensor. These data should be used for estimation purposes only.
Page 83
71-6417 75
Appendix C
Innova Soil and Fuel Vapor Monitors Specific Parts List
See Parts List in Appendix A for a more complete list of all replacement parts and accessories.
61-0123 Sensor, Combustible (LEL/PPM)
65-0613 Sensor, Oxygen (Fuel Vapor only)
80-0504 Hose, 3’ teflon-lined
81-0007 Cylinder, 40% LEL Hexane
81-0078 Cylinder, 100% Nitrogen
81-0226-05 Cal kit, 40% LEL Hexane (Soil Vapor)
81-0226-06 Cal kit, 40% LEL Hexane/100% Nitrogen (Fuel Vapor)
Loading...