Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A June 2008, Revision B September 2008, Revision C April 2009
For Research Use Only. Not for use in diagnostic procedures.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each
European Union (EU) Member State, and these companies should dispose of or recycle this product.
See www.thermo.com/WEEERoHS for further information on Thermo Fisher Scientific’s compliance
with these Directives and the recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC
erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EUMitgliedsstaaten getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt
werden kann. Mehr Information über die Einhaltung dieser Anweisungen durch Thermo Fisher
Scientific, über die Verwerter, und weitere Hinweise, die nützlich sind, um die Produkte zu
identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter www.thermo.com/
WEEERoHS.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements
Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état
membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage
d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre
pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des
substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS.
This HESI-II Probe User Guide provides you with information on using the
heated-electrospray ionization probe. It also provides procedures for installing and
maintaining the HESI-II probe.
Safety and Special Notices
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
P
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION
notice is accompanied by an appropriate CAUTION symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Thermo ScientificHESI-II Probe User Guidevii
Preface
A
S
Hot surface caution
General caution
Hot surface caution
Avoid touching liquid leaking from
the sample inlet while the mass
spectrometer is in operation.
Precautions for Handling the HESI-II Probe
When operating the mass spectrometer in the heated-electrospray mode, do not touch the
heated surfaces of the HESI-II probe or the Ion Max API source housing. Touching the
heated surfaces of the probe or the source housing can cause burns when the probe is
operating at typical temperatures of 350 to 450 °C.
CAUTION AVO ID B URNS . At typical operating temperatures (350 to 450 °C), the
HESI-II probe can severely burn you. Before removing the probe from the API source
housing, allow the probe to cool to room temperature (for approximately 20 minutes)
before you touch it.
When operating the mass spectrometer in the heated-electrospray mode, if you observe liquid
leaking from the sample inlet, place the mass spectrometer in Standby mode before you
tighten the fitting to eliminate the leak.
CAUTION AVO ID E L ECT RIC S HOCK . Do not tighten the probe sample inlet fitting
to eliminate a liquid leak while the mass spectrometer is in operation. If you touch liquid
leaking from the probe sample inlet while the mass spectrometer is in operation, you
might receive an electric shock.
viiiHESI-II Probe User GuideThermo Scientific
Contacting Us
Preface
There are several ways to contact Thermo Fisher Scientific for the information you need.
Y To contact Technical Support
Phone800-532-4752
Fax561-688-8736
E-mailus.techsupport.analyze@thermofisher.com
Knowledge basewww.thermokb.com
Find software updates and utilities to download at mssupport.thermo.com.
Y To contact Customer Service for ordering information
Go to mssupport.thermo.com and click Customer Manuals in the left margin of the
window.
Y To suggest changes to documentation or to Help
• Complete a brief survey about this document by clicking the link below. Thank you
in advance for your help.
• Send an e-mail message to the Technical Publications Editor at
techpubs-lcms@thermofisher.com.
Thermo ScientificHESI-II Probe User Guideix
Introduction
Heated-electrospray ionization (H-ESI) transforms ions in solution into ions in the gas phase
by using electrospray ionization (ESI) in combination with heated auxiliary gas. You can use
H-ESI to analyze any polar compound that makes a preformed ion in solution.
Basic compounds (for example amines) can form a protonated molecule [M + H]
compounds (for example sulphonic acids) can form a deprotonated molecule [M – H]
positive ion polarity mode, the protonated molecule produces a peak at an m/z value of
M + 1, where M equals the mass of the original molecule. In the negative ion polarity mode,
the deprotonated molecule produces a peak at an m/z value of M – 1, where M equals the
mass of the original molecule.
1
+
, and acidic
-
. In the
Because solution chemistry is the primary factor affecting mass spectra in ESI, other common
adducts include sodium ions (Na
yielding an m/z value of [M + 39]
[M + 18]
This chapter describes the principles of the heated-electrospray ionization technique through
use of the HESI-II probe (see Figure 1) and a Thermo Scientific mass spectrometer.
Figure 1. HESI-II probe
+
.
+
), yielding an m/z value of [M + 23]+; potassium ions (K+),
+
; and ammonium ions (NH
+
), yielding an m/z value of
4
Thermo ScientificHESI-II Probe User Guide1