Thermo Scientific HESI-II Probe User Manual

HESI-II Probe
User Guide
70005-97000 Revision C April 2009
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
Ion Max is a trademark of Thermo Fisher Scientific Inc. in the United States. Xcalibur is a registered trademark of Thermo Fisher Scientific Inc. in the United States.
PEEK is a trademark of Victrex PLC.
The following are registered trademarks in the United States and other countries: KEL-F is a registered trademark of 3M Corporation. Upchurch Scientific is a registered trademark of Upchurch Scientific Inc.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error­free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
Release history: Revision A June 2008, Revision B September 2008, Revision C April 2009
For Research Use Only. Not for use in diagnostic procedures.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU) Member State, and these companies should dispose of or recycle this product. See www.thermo.com/WEEERoHS for further information on Thermo Fisher Scientific’s compliance with these Directives and the recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU­Mitgliedsstaaten getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Mehr Information über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, über die Verwerter, und weitere Hinweise, die nützlich sind, um die Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter www.thermo.com/
WEEERoHS.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS.

Contents

C
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Precautions for Handling the HESI-II Probe . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Chapter 2 Functional Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Chapter 3 Removing and Installing the HESI-II Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Removing the HESI-II Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installing the HESI-II Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Adjusting the Probe Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Chapter 4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Flushing the Sample Transfer Line, Sample Tube, and HESI-II Probe . . . . . . . 19
Replacing the Needle Insert. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Chapter 5 Replaceable Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Thermo Scientific HESI-II Probe User Guide v

Preface

This HESI-II Probe User Guide provides you with information on using the
heated-electrospray ionization probe. It also provides procedures for installing and maintaining the HESI-II probe.

Safety and Special Notices

Make sure you follow the precautionary statements presented in this guide. The safety and other special notices appear in boxes.
P
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION notice is accompanied by an appropriate CAUTION symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of data, or invalid test results; or might contain information that is critical for optimal performance of the system.
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Thermo Scientific HESI-II Probe User Guide vii
Preface
A
S
Hot surface caution
General caution
Hot surface caution
Avoid touching liquid leaking from the sample inlet while the mass spectrometer is in operation.

Precautions for Handling the HESI-II Probe

When operating the mass spectrometer in the heated-electrospray mode, do not touch the heated surfaces of the HESI-II probe or the Ion Max API source housing. Touching the heated surfaces of the probe or the source housing can cause burns when the probe is operating at typical temperatures of 350 to 450 °C.
CAUTION AVO ID B URNS . At typical operating temperatures (350 to 450 °C), the HESI-II probe can severely burn you. Before removing the probe from the API source housing, allow the probe to cool to room temperature (for approximately 20 minutes) before you touch it.
When operating the mass spectrometer in the heated-electrospray mode, if you observe liquid leaking from the sample inlet, place the mass spectrometer in Standby mode before you tighten the fitting to eliminate the leak.
CAUTION AVO ID E L ECT RIC S HOCK . Do not tighten the probe sample inlet fitting to eliminate a liquid leak while the mass spectrometer is in operation. If you touch liquid leaking from the probe sample inlet while the mass spectrometer is in operation, you might receive an electric shock.
viii HESI-II Probe User Guide Thermo Scientific

Contacting Us

Preface
There are several ways to contact Thermo Fisher Scientific for the information you need.
Y To contact Technical Support
Phone 800-532-4752
Fax 561-688-8736
E-mail us.techsupport.analyze@thermofisher.com
Knowledge base www.thermokb.com
Find software updates and utilities to download at mssupport.thermo.com.
Y To contact Customer Service for ordering information
Phone 800-532-4752
Fax 561-688-8731
E-mail us.customer-support.analyze@thermofisher.com
Web sit e www.thermo.com/ms
Y To copy manuals from the Internet
Go to mssupport.thermo.com and click Customer Manuals in the left margin of the window.
Y To suggest changes to documentation or to Help
• Complete a brief survey about this document by clicking the link below. Thank you in advance for your help.
• Send an e-mail message to the Technical Publications Editor at
techpubs-lcms@thermofisher.com.
Thermo Scientific HESI-II Probe User Guide ix

Introduction

Heated-electrospray ionization (H-ESI) transforms ions in solution into ions in the gas phase by using electrospray ionization (ESI) in combination with heated auxiliary gas. You can use H-ESI to analyze any polar compound that makes a preformed ion in solution.
Basic compounds (for example amines) can form a protonated molecule [M + H] compounds (for example sulphonic acids) can form a deprotonated molecule [M – H] positive ion polarity mode, the protonated molecule produces a peak at an m/z value of M + 1, where M equals the mass of the original molecule. In the negative ion polarity mode, the deprotonated molecule produces a peak at an m/z value of M – 1, where M equals the mass of the original molecule.
1
+
, and acidic
-
. In the
Because solution chemistry is the primary factor affecting mass spectra in ESI, other common adducts include sodium ions (Na yielding an m/z value of [M + 39] [M + 18]
This chapter describes the principles of the heated-electrospray ionization technique through use of the HESI-II probe (see Figure 1) and a Thermo Scientific mass spectrometer.
Figure 1. HESI-II probe
+
.
+
), yielding an m/z value of [M + 23]+; potassium ions (K+),
+
; and ammonium ions (NH
+
), yielding an m/z value of
4
Thermo Scientific HESI-II Probe User Guide 1