Thermo Scientific Heratherm IMH 180, Heratherm IMH 100, Heratherm IMH 180-S, Heratherm IMH 400-S, Heratherm IMH 750-S Operating Instructions Manual

...
©
2012 Thermo Fisher Scientific Inc. All rights reserved. These operating instructions are protected by copyright. Rights resulting thereof, particularly reprint, photomechanical or digital postprocessing or reproduction, even in part, are only allowed with the written consent of Thermo Fisher Sci­entific. This regulation does not apply to reproductions for in-plant use. The contents of this operating instructions manual may change at any time and without any prior notice. In case of con­flicting translations into foreign languages the German-language version of these operating instructions shall be binding.
Trademarks
Heratherm™ is a registered trademark of Thermo Scientific. Thermo Scientific is a brand owned by Thermo Fisher Scientific, Inc. All other trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers.
Thermo Fisher Scientific Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
Thermo Electron LED GmbH is an affiliate to: Thermo Fisher Scientific Inc. 81 Wyman Street Waltham, MA 02454 USA
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product oper­ation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc. The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
i-ii Heratherm Thermo Scientific
C
Contents
Chapter 1 Safety Notes ..................................................................................................................... 1-1
Basic Operating Precautions ............................................................................................................... 1-1
Operational Safety Rules .................................................................................................................... 1-2
Warranty ............................................................................................................................................ 1-2
Explanation of Safety Information and Symbols ................................................................................. 1-3
Safety Notes and Symbols Used Throughout These Operating Instructions ................................... 1-3
Additional Symbols for Safety Information..................................................................................... 1-4
Symbols on the Incubator............................................................................................................... 1-5
Intended Purpose of the Incubator ..................................................................................................... 1-6
Intended Purpose of the Incubator ................................................................................................. 1-6
Incorrect Use .................................................................................................................................. 1-6
Standards and Directives .................................................................................................................... 1-6
Chapter 2 Delivery of the Incubator................................................................................................. 2-1
Packaging ........................................................................................................................................... 2-1
Acceptance Inspection ........................................................................................................................ 2-1
Scope of Supply .................................................................................................................................. 2-2
Chapter 3 Installation........................................................................................................................ 3-1
Ambient Conditions ........................................................................................................................... 3-1
Location Requirements ................................................................................................................... 3-1
Intermediate Storage........................................................................................................................... 3-2
Room Ventilation............................................................................................................................... 3-2
Space Requirement ......................................................................................................................... 3-3
Installation Built-in Units............................................................................................................... 3-3
Table-top incubators ...................................................................................................................... 3-4
Floor Stand Incubators ................................................................................................................... 3-5
Transport ........................................................................................................................................... 3-7
Stacking kit ........................................................................................................................................ 3-9
Installing the Stacking Feet................................................................................................................. 3-9
Installing the Stacking Adapter ......................................................................................................... 3-10
Installing the Anti-tilt Anchor .......................................................................................................... 3-11
Floor stand unit spacers .................................................................................................................... 3-13
Chapter 4 Product Description ........................................................................................................ 4-1
Heratherm IMH Series Incubator Overview ....................................................................................... 4-1
Heratherm IMH-S Incubator Overview ............................................................................................. 4-4
Safety Devices................................................................................................................................... 4-11
Work Space Atmosphere .................................................................................................................. 4-11
Door Switch ..................................................................................................................................... 4-11
Sensing and Control System ............................................................................................................. 4-12
Thermo Scientific Heratherm i
Contents
Data Communications & Alarm Interface ........................................................................................ 4-13
RS-232 Interface........................................................................................................................... 4-13
Alarm Contact .............................................................................................................................. 4-13
AC Power Socket.......................................................................................................................... 4-13
Fuses............................................................................................................................................. 4-13
Work Space Components ................................................................................................................. 4-14
Inner Chamber ............................................................................................................................. 4-14
Connecting Nozzles for Fresh-air Filters (Accessory Items) ........................................................... 4-14
Access Port ................................................................................................................................... 4-15
Tube Access Ports............................................................................................................................. 4-17
Chapter 5 Installation procedures ................................................................................................... 5-1
Table-top incubators .......................................................................................................................... 5-1
Installing the Shelf System .................................................................................................................. 5-1
Initial Installation ........................................................................................................................... 5-1
Installing the Perforated Shelves ......................................................................................................... 5-2
Preparing the useable space................................................................................................................. 5-2
Installation or Removal of the Support Rails................................................................................... 5-3
Installing and Uninstalling the Air Baffle ........................................................................................ 5-4
Levelling the Incubator ................................................................................................................... 5-4
Commissioning of floor stand incubators ........................................................................................... 5-5
Installing/Removing air baffles (IMH-S Series)............................................................................... 5-5
Commissioning, general ..................................................................................................................... 5-6
Installing the Shelf Support Brackets .............................................................................................. 5-6
Installing the Perforated Shelves ..................................................................................................... 5-7
Mains connection ............................................................................................................................... 5-7
Connecting to the Power Supply Source......................................................................................... 5-8
Connecting the RS-232 Interface ....................................................................................................... 5-9
Interconnecting the Incubator with a Computer............................................................................. 5-9
Wiring the Alarm Contact................................................................................................................ 5-11
Functional Description ................................................................................................................. 5-11
Alarm Relay Specifications............................................................................................................ 5-11
Connection Example .................................................................................................................... 5-12
Chapter 6 Operation.......................................................................................................................... 6-1
Preparing the Incubator ...................................................................................................................... 6-1
Starting Operation.............................................................................................................................. 6-1
Chapter 7 Handling and Control ...................................................................................................... 7-1
Powering Up ...................................................................................................................................... 7-6
Switching the Incubator Off / Powering Down .................................................................................. 7-6
Temperature Set Value ....................................................................................................................... 7-7
Timer ................................................................................................................................................. 7-8
Setting a Fixed-time On or Off Timer .......................................................................................... 7-10
Setting a Weekly Timer ................................................................................................................ 7-11
Stopping a Timer.......................................................................................................................... 7-13
Power Outlet (table-top incubators only).......................................................................................... 7-15
Decontamination.............................................................................................................................. 7-16
Fan ................................................................................................................................................... 7-19
Table-top incubators ........................................................................................................................ 7-19
ii Heratherm Thermo Scientific
Contents
Floor stand incubators .......................................................................................................................7-20
Settings .............................................................................................................................................7-22
Error Log ......................................................................................................................................7-22
Calibration ....................................................................................................................................7-23
Date and Time ..............................................................................................................................7-24
Temperature Display Unit ............................................................................................................7-27
Chapter 8 Shut-down .........................................................................................................................8-1
Shutting the Incubator Down .............................................................................................................8-1
Chapter 9 Cleaning and Disinfection ...............................................................................................9-1
Cleaning..............................................................................................................................................9-1
Cleaning Exterior Surfaces...............................................................................................................9-1
Wipe / Spray Disinfection...................................................................................................................9-1
Preparing the Manual Wipe/Spray Disinfection ..............................................................................9-2
Predisinfection ................................................................................................................................9-3
Cleaning..........................................................................................................................................9-4
Final Disinfection............................................................................................................................9-4
Starting the Decontamination Routine............................................................................................9-4
Chapter 10 Maintenance....................................................................................................................10-1
Inspections and Checks .....................................................................................................................10-1
Regular Checks..............................................................................................................................10-1
Semi-annual Inspection .................................................................................................................10-1
Service Intervals.................................................................................................................................10-2
3-monthly service ..........................................................................................................................10-2
Annual Service...............................................................................................................................10-2
Preparing Temperature Calibration...................................................................................................10-3
Comparison Measurement Procedure................................................................................................10-3
Temperature Calibration Procedure ..................................................................................................10-4
Replacing the Door Seal ....................................................................................................................10-4
Replacing the Power Cord.................................................................................................................10-5
Returns for Repair .............................................................................................................................10-5
Chapter 11 Disposal...........................................................................................................................11-1
Overview of Materials Used ..............................................................................................................11-1
Chapter 12 Error Codes.....................................................................................................................12-1
Chapter 13 Technical Data ................................................................................................................13-1
Chapter 14 Spare Parts and Accessories ........................................................................................14-1
Chapter 15 Device Log ......................................................................................................................15-1
Chapter 16 Contact ............................................................................................................................16-1
Thermo Scientific Heratherm iii
Contents
iv Heratherm Thermo Scientific
L
List of Figures
Figure 3-1 Table-top incubators, dimensions and required clearances .................................................................... 3-4
Figure 3-2 Floor stand incubators, dimensions and required clearances.................................................................. 3-5
Figure 3-3 Floor stand incubators, dimensions and required clearances.................................................................. 3-6
Figure 3-4 Lift Points............................................................................................................................................. 3-7
Figure 4-1 Heratherm IMH 60/ IMH 100/ IMH 180 Front View ........................................................................ 4-2
Figure 4-2 Heratherm IMH 60/ IMH 100/ IMH 180 Rear View.......................................................................... 4-3
Figure 4-3 Heratherm IMH 60-S/ IMH 100-S/ IMH 180-S Series Front View .................................................... 4-5
Figure 4-4 Heratherm IMH 60-S / IMH 100-S / IMH 180-S Series Rear View .................................................... 4-6
Figure 4-5 Heratherm IMH 400-S Series Front View ............................................................................................ 4-7
Figure 4-6 Heratherm IMH 400-S Rear View ....................................................................................................... 4-8
Figure 4-7 Heratherm IMH 750-S Front View...................................................................................................... 4-9
Figure 4-8 Heratherm IMH 750-S Rear View ....................................................................................................... 4-10
Figure 4-9 Sensor System (for table-top incubators)............................................................................................... 4-12
Figure 4-10 Sensor System (for floor stand incubators) .......................................................................................... 4-12
Figure 4-11 Signal Interfaces and Power Socket..................................................................................................... 4-13
Figure 4-12 Connecting Nozzle for Fresh-air Filters (Accessory Item).................................................................... 4-15
Figure 4-13 Shelf System ....................................................................................................................................... 4-16
Figure 4-14 Tube Access Ports............................................................................................................................... 4-17
Figure 5-1 Sliding the Retaining Spring into the Support Rail ............................................................................... 5-1
Figure 5-2 Installing the Shelving .......................................................................................................................... 5-2
Figure 5-3 Support Rail Installation....................................................................................................................... 5-3
Figure 5-4 Removing the Air Baffle ....................................................................................................................... 5-4
Figure 5-5 Removing the bottom plate .................................................................................................................. 5-5
Figure 5-6 Removing the left and right support profiles......................................................................................... 5-5
Figure 5-7 Removing the rear air baffle IMH 400-S / 750-S.................................................................................. 5-6
Figure 5-8 Shelf Support Bracket Installation ........................................................................................................ 5-6
Figure 5-9 Perforated Shelf .................................................................................................................................... 5-7
Figure 5-10 AC Power Supply Socket .................................................................................................................... 5-8
Figure 5-11 Alarm Contact Connection Example .................................................................................................. 5-12
Figure 7-1 Control Panel for Heratherm IMH Series and IMH-S Series Incubators. ............................................. 7-1
Figure 10-1 Door Seal Replacement ...................................................................................................................... 10-4
Thermo Scientific Heratherm i
List of Figures
ii Heratherm Thermo Scientific
1
Safety Notes
Basic Operating Precautions
These operating instructions describe Heratherm incubators.
Heratherm incubators have been manufactured to the latest state of the art and have been tested thoroughly for flawless functioning prior to shipping. However, the incubator may present potential hazards, particularly if it is operated by inadequately trained personnel or if it is not used in accordance with the intended purpose. Therefore, the following must be observed for the sake of accident prevention:
• Heratherm incubators must be operated by adequately trained and authorized professional personnel.
• Heratherm incubators must not be operated unless these operating instructions have been fully read and understood.
• The present operating instructions, applicable safety data sheets, plant hygiene guidelines and the corresponding technical rules issued by the operator shall be used to create written procedures targeted at personnel working with the subject matter device, detailing:
• the decontamination measures to be employed for the incubator and the accessories used with it,
• the safety precautions to be taken when processing specific agents,
• the measures to be taken in case of accidents.
• Repair work on the incubator must be carried out only by trained and authorized expert personnel.
• The contents of these operating instructions are subject to change at any time without further notice.
• Concerning translations into foreign languages, the German version of these operating instructions is binding.
• Keep these operating instructions close to the incubator so that safety instructions and important information are always accessible.
• Should you encounter problems that are not detailed adequately in these operating instructions, please contact Thermo Electron LED GmbH immediately for your own safety.
Thermo Scientific Heratherm 1-1
Safety Notes Operational Safety Rules
Operational Safety Rules
The following rules must be heeded when working with Heratherm incubators:
• Observe the sample weight limits specified for your Heratherm incubator as a whole and its shelving in particular; see “Technical Data” on page 13-1.
• Do not load the bottom of the interior workspace to avoid the risk of overheating any samples placed there and to prevent the temperature sensor from being damaged.
• Arrange the samples evenly throughout the work space, making sure not to place them too closely to the interior walls to ensure a uniform temperature distribution.
• Do not load your Heratherm incubator with substances that exceed the capabilities of the available lab apparatus and Personal Protection Equipment to provide sufficient degrees of protection to users and third parties.
• Check the door seal every six months for proper sealing performance and possible damage.
• Do not process any samples containing hazardous chemical substances that may be released into the ambient air through defective seals or may cause corrosion or other defects on parts of the Heratherm incubator.
Warranty
Thermo Fisher Scientific warrants the operational safety and functions of the Heratherm incubators only under the condition that:
• the incubator is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions,
• the incubator is not modified,
• only original spare parts and accessories that have been approved by Thermo Scientific are used (third-party spares without Thermo Scientific approval void the limited warranty),
• inspections and maintenance are performed at the specified intervals,
• an operation verification test is performed after each repair activity.
The warranty is valid from the date of delivery of the incubator to the customer.
1-2 Heratherm Thermo Scientific
Explanation of Safety Information and Symbols
Explanation of Safety Information and Symbols
Safety Notes and Symbols Used Throughout These Operating Instructions
Indicates a hazardous situation which, if not avoided, will result in death or serious injuries.
Safety Notes
Indicates a hazardous situation which, if not avoided, could result in death or serious injuries.
Indicates a situation which, if not avoided, could result in damage to equipment or property.
Is used for useful hints and information regarding the application.
Thermo Scientific Heratherm 1-3
Safety Notes Explanation of Safety Information and Symbols
Additional Symbols for Safety Information
Wear safety gloves!
Wear safety goggles!
Harmful liquids!
Electric shock!
Hot surfaces!
Fire hazard!
Explosion hazard!
Suffocation hazard!
Biological hazard!
Contamination hazard!
Danger of tipping!
1-4 Heratherm Thermo Scientific
Symbols on the Incubator
Observe operating instructions
Mark of conformity USA/Canada
120 Volts AC power socket
Alarm contact
Safety Notes
Explanation of Safety Information and Symbols
Thermo Scientific Heratherm 1-5
Safety Notes Intended Purpose of the Incubator
Intended Purpose of the Incubator
Intended Purpose of the Incubator
Heratherm incubators are laboratory devices for preparing and cultivating cell and tissue cultures. The devices employ precision temperature control for simulating the specific physiological ambient conditions for these cultures.
Incorrect Use
To avoid the risk of explosion do not load the incubator with tissue, material, or liquids that:
• are easily flammable or explosive,
• release vapor or dust that forms combustible or explosive mixtures when exposed to air,
• release poisons,
• do not pour any liquids on the bottom of the interior surface or into a collecting basin inside the unit.
• release dust
• exhibit exothermic reactions
• are pyrotechnical substances
• refrain also from pouring any liquids onto the internal base plate or inserting bowls filled with liquids into the sample compartment.
Standards and Directives
The incubator complies with the following standards and guidelines:
• IEC EN 61010 - 1, IEC EN 61010 - 2 - 010
• Low Voltage Directive 2006/95/EC
• EMC Directive 2004/108/EC
Additionally, the incubator is in compliance with many other international standards, regulations and directives not listed here. Should you have any questions regarding compliance with national standards, regulations and directives applicable for your country, please contact your Thermo Fisher Scientific sales organization.
1-6 Heratherm Thermo Scientific
2
Delivery of the Incubator
Packaging
Heratherm incubators are delivered in a rugged packaging box. All packaging materials can be separated and are reusable:
Packaging materials
Packaging carton: Recycled paper
Foam elements: Styrofoam (CFC-free)
Pallet: Chemically untreated wood
Packaging film: Polyethylene
Packaging ribbons: Polypropylene
Acceptance Inspection
After the incubator has been delivered, check the delivery immediately for:
• completeness,
• possible damage.
If components are missing or damage is found on the incubator or the packaging, in particular damage caused by humidity and/or water, please notify the carrier as well as Thermo Scientific Technical Support immediately.
Risk of injury
Should sharp edges have formed in damaged areas or elsewhere on the device, take all necessary precautions to protect personnel handling the incubator. For example, have them wear protective gloves and other personal protection equipment.
Thermo Scientific Heratherm 2-1
Delivery of the Incubator Scope of Supply
Scope of Supply
Incubators
Quantity of components supplied (pieces)
Perforated shelves 2
Support rail for shelf table-top incubators 2
Shelf support 4
Power cord 1
Connector, potential-free contact 1
Clip springs for table-top incubators 2
Plug 1
Anti-tilt anchor 1
Operating manual 1
Short reference guide 1
IMH Series IMH-S Series
2-2 Heratherm Thermo Scientific
3
Installation
During installation of built-in units, ensure that the escaping air will be safely discharged out of the installation space.
Ambient Conditions
Location Requirements
Built-in incubators can, heating and drying ovens must be operated with an air exhaust system and exhaust hose (only original Thermo accessory should be used).
For safety reasons, the installation space should be made of non-combustible materials, according to DIN 4102.
The incubator must only be operated in a location that meets all of the ambient condition requirements listed below:
• Installation location indoors in dry areas free from drafts.
• The dust burden may not exceed the contamination category 2 based on EN 61010-1. Using the incubator in an atmosphere with electrically conductive dust is prohibited.
• The minimal distance to adjacent surfaces must be observed on all sides (see Section
“Space Requirement” on page 3-3)
• The operating room must be equipped with appropriate ventilation.
• Solid, level, fire-proof surface; no flammable materials opposite to the rear panel of the incubator.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the dead weight of the incubator and its accessories (particularly if two devices are stacked).
• The electrical circuitry of the incubator has been designed for an operating height of up to 2000 m above sea level.
• Relative humidity up to 80% (maximum; preferably 60-70%), non condensing.
• Should condensation exist, wait until the moisture has evaporated completely before connecting the incubator to a power source and powering up.
• The ambient temperature must be within a range of +18 °C to +32 °C (64.4 °F to 89.6 °F).
• Avoid direct exposure to sunlight.
Thermo Scientific Heratherm 3-1
Installation Intermediate Storage
• Devices that produce excessive amounts of heat must not be placed near the incubator.
• Place the incubator on a floor stand (option; to be ordered separately), never on the lab
• Power line voltage variations must not exceed ±10 % of the nominal voltage.
• Transient surges must lie within the range of levels that normally occur in the power
• Consider installing one dedicated upstream circuit breaker per incubator to avoid multiple
floor. This prevents the penetration of dust or dirt into the device.
supply system. The impulse withstand voltage based on surge category II of IEC 60364-4-443 shall be applied at the nominal voltage level.
device failures in case of an electrical fault.
Contamination hazard
Do not place the incubator directly on the lab floor, but mount it on the floor stand or on a lab work surface (option; to be ordered separately). Contaminants, such as bacteria, viruses, fungi, prions, and other biological substances may use the open door to migrate easily from the floor into the incubator’s work space.
Intermediate Storage
When the incubator is placed in intermediate storage, which is permissible for a maximum of four weeks, make sure that the ambient temperature is between 20 °C to 60 °C (68 °F to 140 °F) and the maximum relative humidity does not exceed 90%, non-condensing.
Room Ventilation
Heat dissipating from the incubator during continuous operation may cause a change in the room climate.
• Therefore, the incubator must only be installed in rooms with sufficient ventilation.
• Do not install the incubator in room recesses without ventilation.
• When several devices are to be placed in the same room, additional ventilation may have to be provided as necessary.
• To avoid any impact of the heat dissipated by the incubator on the ambient climate the room must be vented by means of a laboratory-grade ventilation system that complies with applicable local and national health and safety regulations and has sufficient capacity.
• If excessive temperatures tend to occur in the operating room, be sure to provide a thermal protection means that cuts out the power supply to mitigate the impact of overtemperature scenarios.
3-2 Heratherm Thermo Scientific
Space Requirement
Fig. 3
After connecting the built-in unit to the electrical power supply, avoid damaging the power cord, then sliding into the installation position.
For built-in units following clearances should be kept:
Installation
Room Ventilation
A, B, C and D see Operating Instructions of the unit.
E (mm/inch) F (mm/inch) G (mm/inch) H (mm/inch)
100 / 4 50 / 2 80 / 3.2 30 / 1.2
Installation Built-in Units
IGS
IMH/IMH-S
OGS/OMS
OGH/OGH-S OMH/OMH-S
Clearances without exhaust hose.
Clearances, close air port with plug, supplied with the unit.
Clearances, with additional space to operate to air slide, with air exhaust system and exhaust hose Ø 40 mm (1.58 in)/1.5 m (59 in) (original Thermo accessory), shortened to required length.
Clearances, with air exhaust system and exhaust hose Ø 40 mm (1.58 in)/1.5 m (59 in) (original Thermo accessory), shortened to required length. Operating temperature up to max. 250 °C (482 °F).
When installing the incubator, make sure that the installation and supply connections remain freely accessible.
The specified side clearances represent minimum distances.
Thermo Scientific Heratherm 3-3
Installation Room Ventilation
Table-top incubators
Figure 3-1 Table-top incubators, dimensions and required clearances
Table 3-1 Incubator Dimensions
Model A (mm) B (mm) C (mm) D (mm)
IMH 60 530/20.1 565/22.2 720/28.3 540/21.3
IMH 100 640/25.2 565/22.2 820/32.3 650/25.6
IMH 180 640/25.2 738/29.1 920/36.2 650/25.6
IMH 60-S 530/20.1 565/22.2 720/28.3 540/21.3
IMH 100-S 640/25.2 565/22.2 820/32.3 650/25.6
IMH 180-S 640/25.2 738/29.1 920/36.2 650/25.6
* Depth of handle /display (66 mm/2.6 in) not included in overall depth specified; height of adjustable feet (36 mm/1.4 in)
not included in overall height specified.
Table 3-2 Required Clearances
E (mm) F (mm) G (mm) H (mm)
80/3.1 50/2.0 200/7.9 300/11.8
3-4 Heratherm Thermo Scientific
Floor Stand Incubators
400 liter units
Installation
Room Ventilation
Figure 3-2 Floor stand incubators, dimensions and required clearances
Table 3-3 Incubator Dimensions
Model A (mm/inch) B (mm/inch) C (mm/inch) D (mm/inch)
IMH 400-S 755/29.7 770/30.3 1655/65.2 810/31.9
*
Depth of handle /display (66 mm/2.6 in) not included in overall depth specified. Width of hinge (23 mm) not included in
overall width.
Table 3-4 Required Clearances
E (mm/inch) F (mm/inch) H (mm/inch) I (mm/inch)
120/4.7 50 / 2 200 / 8 200/7.9
Thermo Scientific Heratherm 3-5
Installation Room Ventilation
750 liter units
Figure 3-3 Floor stand incubators, dimensions and required clearances
Table 3-5 Incubator Dimensions
Model A (mm/inch) B (mm/inch) C (mm/inch) D (mm/inch)
IMH 750-S 1215 / 47.8 770/30.3 1655 / 65.2 670/26.4
*
Depth of handle /display (66 mm/2.6 in) not included in overall depth specified. Width of hinge (23 mm) not included in
overall width.
Table 3-6 Required Clearances
E (mm/inch) F (mm/inch) H (mm/inch) I (mm/inch)
120/4.7 50/2.0 200/7.9 350/13.8
3-6 Heratherm Thermo Scientific
Transport
Table-top incubators
For transport, do not lift the incubator using the doors or components attached to the incubator as lift points.
Installation
Transport
Figure 3-4 Lift Points
To avoid injury through physical strain, such as strain traumata and slipped discs, do not attempt to lift the incubator alone! To avoid injury through dropped loads, be sure to wear Personal Protection Equipment, such as safety shoes, when lifting the incubator. To avoid crushing your fingers or hands (particularly in a closing door) or damaging the incubator, do not use any other lift points than those indicated in the illustration above.
Heavy loads! Lift with care!
Thermo Scientific Heratherm 3-7
Installation Transport
Floor stand incubators
The floor stand incubators come equipped with four (4) casters. The lever for releasing the caster is located above the locking lever. After positioning the unit in its installation location ensure that the locking levers are pressed down on the casters.
To ensure the degree of stability specified by safety requirements the front casters must be turned so that they are facing forward after the unit has been positioned in its installation location and the locking levers pressed down on these casters.
Danger of tipping when moving!
Before moving the unit, ensure that it has been unplugged.
Move the Heratherm floor stand incubators with caution.
Quick starts and stops can result in tipping!
Always ensure that the doors are closed when moving the unit.
3-8 Heratherm Thermo Scientific
Stacking kit
The stacking adapter is available only for table-top incubators.
Material Number Description
Installation
Stacking kit
50126665
50126666
50126667
Stacking adapter Heratherm 60L
Stacking adapter Heratherm 100L
Stacking adapter Heratherm 180L
Scope of delivery:
1 Stacking adapter
1 Anti-tilt anchor
1 Plastic bag with 2 stacking feet and 2 M4x16 Torx screws.
Required Tools:
Slotted screwdriver 5,5x100 or Torx screwdriver 20x100.
Installing the Stacking Feet
Remove the left and right blank plugs at the top.
Install the stacking feet with the enclosed screws using a slotted or Torx screwdriver.
Thermo Scientific Heratherm 3-9
Installation
1
Anti-tilt anchor
Installing the Stacking Adapter
Installing the Stacking Adapter
Stack as follows when using a stacking adapter (numbers denote incubator volumes in liters):
• for 60 l or 100 l on 180 l use stacking adapter Heratherm 180L,
• for 60 l on 100 l use stacking adapter Heratherm100L.
To prevent the top device from slipping and dropping down, the following requirements must be fulfilled before devices may be stacked:
• Only two units may be stacked together. When stacking devices with the same type of enclosure, yet with a different footprint the device with the smaller footprint must be placed on top of the one with the larger footprint at all times.
• The bottom device must be correctly levelled.
• Place the corresponding stacking adapter.
• The levelling feet on the top device must be screwed in all the way.
• The levelling feet of the top device must be aligned with and placed exactly on the stacking pads of the stacking adapter.
• The anti-tilt anchor must be installed at the top device.
3-10 Heratherm Thermo Scientific
Installing the Anti-tilt Anchor
Table-top incubators
The anti-tilt anchor secures the top device in a stack to a solid part of a building. The anti-tilt anchor is to be mounted on the side opposite of the door hinges.
Bend the fixing tabs of the anti-tilt anchor up on one side and down on the other by an angle of approx. 90°.
Installation
Installing the Anti-tilt Anchor
1. Do not use this position if the door is hinged on this side. Right-hand hinges represent the standard configuration.
2. Preferred position.
3. Alternative position. Do not use if the door is hinged on this side.
Remove the bracket screws. Use the preferred position, if possible.
Fix the anti-tilt anchor with the bracket side down to the unit.
Position the unit with the anti-tilt anchor to in an angle of approx. 90° +/- 20%.
Take care that the stacking feet of the unit are still in correct place on the lower unit or on the stacking adapter.
Fix the anti-tilt anchor to a solid part of the building.
Floor stand incubators
Heratherm floor stand incubators must always be attached to the wall using two (2) retaining brackets on the outer left and right side on the back of the unit.
Thermo Scientific Heratherm 3-11
Installation
Unsafe part of the building!
Install the anti-tilt anchor to a solid part of the building, which is able for shoring loads.
The installation has to be carried out by qualified personnel only.
The connection to the building must be carried out with appropriate screws and dowels according to the consistence of the building part.
Risk of overheating with stacked devices
To avoid the risk of electrical components and the outer enclosure overheating or temperature control failing due to insufficient ventilation, do not exceed the specified stacking height!
Risk of tipping and dropping of stacked devices
You should be aware at all times that stacked devices do not form a stable unit, even when the stacking pads and frames are correctly used. The top device may tip over and drop down when being transported in a stack. To avoid injury to persons and damage to equipment, do not attempt to move stacked devices as a unit! Separate and move each device one by one, then restack them.
Installing the Anti-tilt Anchor
Remove the screws.
Attach the end of the retaining bracket that is facing downward to the unit.
Align the device at roughly 90°, +/-20° to the retaining bracket.
Affix the retaining bracket to the wall.
Additionally, the following caution notes must be heeded at all times:
3-12 Heratherm Thermo Scientific
Floor stand unit spacers
The spacer on the electrical module must be pulled out and fixed in place before installing the floor stand unit at its installation location.
Installation
Floor stand unit spacers
1 Loosen the 2 screws
2 Pull out the spacer and slide the screws into the recesses
3 Tighten the two screws
Thermo Scientific Heratherm 3-13
Installation Floor stand unit spacers
3-14 Heratherm Thermo Scientific