Thermo Scientific HERAcell 150i, HERAcell 240i Operating Instructions Manual

Page 1
Operating Instructions
CO2 Incubator HERAcell
®
150i / 240i
50115191 A / 08.2009
Page 2
Copyright 2009 Thermo Fisher Scientific. All rights reserved.
These operating instructions are protected by copyright. Rights resulting thereof, particularly reprint, photomechanical or digital postprocessing or reproduction, even in part, are only allowed with the written consent of the Thermo Electron LED GmbH. This regulation does not apply to reproductions for in-plant use. The contents of the operating instructions are subject to change without further notice. Concerning translations into foreign languages, the German version of these operating instructions is binding.
Trademarks
HERAcell® and iCan™ are registered trademarks of Thermo Scientific. Thermo Scientific is a brand of Thermo Fisher Scientific Inc. All trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers.
Thermo Electron LED GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
Thermo Electron LED GmbH is a subsidiary company of: Thermo Fisher Scientific Inc. 81 Wyman Street Waltham, MA 02454 USA
Page 3
Operating Instructions | HERAcell® 150i / 240i
Contents
3
50115191 A / 08.2009
1 General notes 9
1.1 Identification of the device and of the documentation....................................... 10
1.2 Instruction of the operating personnel................................................................ 10
1.3 Applicability of the instructions ........................................................................... 11
1.4 Warranty ............................................................................................................... 11
1.5 Explanation of safety information and symbols ................................................. 12
1.5.1 Safety information and symbols used in the operating instructions........... 12
1.5.2 Additional symbols for safety informations: ................................................ 12
1.5.3 Symbols on the device .................................................................................. 13
1.6 Intended purpose of the device........................................................................... 14
1.6.1 Correct use ..................................................................................................... 14
1.6.2 Incorrect use................................................................................................... 14
1.7 Standards and directives ..................................................................................... 15
1.8 Safety notes on gases.......................................................................................... 16
1.8.1 Safety notes on carbon dioxide (CO
2
) .......................................................... 16
1.8.2 Safety notes on oxygen (O
2
) ......................................................................... 17
1.8.3 Safety notes on nitrogen (N
2
)........................................................................ 17
2 Delivery of the device 18
2.1 Packaging ............................................................................................................. 18
2.2 Acceptance inspection......................................................................................... 18
2.3 Standard equipment components HERAcell
®
150i............................................ 19
2.4 Optional equipment components HERAcell
®
150i............................................. 19
2.5 Standard equipment components HERAcell
®
240i............................................ 20
2.6 Optional equipment components HERAcell
®
240i............................................. 21
3 Installation of the device 22
3.1 Ambient conditions.............................................................................................. 22
3.2 Room ventilation .................................................................................................. 22
3.3 Space requirements ............................................................................................. 23
3.4 Transport .............................................................................................................. 24
3.5 Stacking ................................................................................................................ 24
3.6 Retrofitting/Modifications .................................................................................... 25
4 Description of the device 26
4.1 Front view of HERAcell® 150i.............................................................................. 26
4.2 Rear view HERAcell
®
150i ................................................................................... 28
4.3 Front view HERAcell
®
240i .................................................................................. 32
4.4 Rear view HERAcell
®
240i ................................................................................... 34
4.5 Safety devices ...................................................................................................... 38
4.6 Work space atmosphere ...................................................................................... 38
4.7 Door switch .......................................................................................................... 40
4.8 Sensor system...................................................................................................... 41
4.9 Supply interface ................................................................................................... 42
4.10 Work space components ................................................................................... 44
4.10.1 Interior container.......................................................................................... 44
4.10.2 Optional split gas-tight screens ................................................................... 44
4.10.3 Water reservoir............................................................................................. 45
4.10.4 Gas humidification (optional, only for O
2
control)...................................... 45
4.10.5 Heating system............................................................................................. 45
Page 4
Operating Instructions | HERAcell® 150i / 240i
Contents
4
50115191 A / 08.2009
4.10.6 Rear panel openings .................................................................................... 46
4.10.7 Shelf system................................................................................................. 47
4.10.8 Bottle turning device for HERAcell® 240i (optional) .................................. 47
4.10.9 Water pump.................................................................................................. 48
5 Start-up 49
5.1 Preparing the work space .................................................................................... 49
5.2 Installing the shelf system ................................................................................... 49
5.2.1 Installation/removal of the support rail ......................................................... 49
5.2.2 Installing the shelf supports: ......................................................................... 50
5.2.3 Levelling the device ....................................................................................... 50
5.3 Installing the gas humidification (optional, only for O
2/N2
-control)................... 50
5.4 Installing the optional bottle turning devices HERAcell
®
240i ........................... 51
5.5 Installing the optional center strut HERAcell
®
240i for split inserts................... 54
5.6 Gas connection..................................................................................................... 55
5.6.1 Installing gas pressure hoses ........................................................................ 55
5.6.2 Installing devices without the optional gas monitoring system .................. 56
5.6.3 CO
2
connection with gas monitoring system (optional) ..............................57
5.6.4 Combined CO
2/O2/N2
connection with
gas monitoring system (optional) ................................................................. 57
5.7 Power supply connection .................................................................................... 59
5.8 Connecting the RS 232 interface ......................................................................... 60
5.9 Connecting the USB interface (optional)............................................................. 60
5.10 Connecting the alarm contact............................................................................ 61
6 Operation 63
6.1 Preparing the device ............................................................................................ 63
6.2 Starting operation ................................................................................................ 64
7 Handling and control (iCan™ touchscreen controller) 65
7.1 Power switch ........................................................................................................ 65
7.2 Operating panel and operating structure............................................................ 65
7.3 Factory presettings of the iCan™ touchscreen controls..................................... 68
7.4 Heat-up phase of the control loop sensors......................................................... 68
7.5 Event actions for settings .................................................................................... 69
7.6 Setting the temperature set value ....................................................................... 70
7.7 Setting the CO
2
set value..................................................................................... 71
7.8 Setting the O
2
set value ....................................................................................... 72
7.9 Auto-start routine ................................................................................................. 73
7.9.1 Activating auto-start....................................................................................... 75
7.9.2 Interrupting the auto-start routine ................................................................. 76
7.10 Running the contra-con routine......................................................................... 76
7.11 User configuration.............................................................................................. 76
7.11.1 Settings......................................................................................................... 77
7.11.2 Event logging ............................................................................................... 83
7.11.3 Options ......................................................................................................... 85
7.11.4 Icon description............................................................................................ 90
7.11.5 Enabling/disabling the keypad lock............................................................. 92
7.11.6 Software versions ........................................................................................ 93
7.12 Scaling the trend display ................................................................................... 94
7.13 Error messages .................................................................................................. 95
Page 5
Operating Instructions | HERAcell® 150i / 240i
Contents
5
50115191 A / 08.2009
7.13.1 Response to an error message event ......................................................... 95
7.13.2 Troubleshooting........................................................................................... 95
8 Shut-down 99
8.1 Shutting the device down.................................................................................... 99
9 Cleaning and disinfection 100
9.1 Cleaning.............................................................................................................. 100
9.2 Decontamimation procedures ........................................................................... 101
9.3 Wipe/spray disinfection ..................................................................................... 101
9.4 contra-con decontamination routine................................................................. 104
9.4.1 Activating the contra-con routine................................................................ 107
9.4.2 Interrupting contra-con ................................................................................ 107
9.4.3 contra-con interruption due to error ........................................................... 108
9.4.4 Completing contra-con ................................................................................ 109
10 Maintenance 110
10.1 Inspections and checks.................................................................................... 110
10.2 Service intervals............................................................................................... 110
10.3 Preparing the temperature calibration ............................................................ 111
10.4 Temperature calibration procedure ................................................................ 112
10.5 Preparing the CO
2
calibration.......................................................................... 113
10.6 CO
2
calibration procedure ............................................................................... 115
10.7 Replacing the gas inlet filter............................................................................. 116
10.8 Replacing the device fuses .............................................................................. 116
10.9 Replacing the door seal ................................................................................... 117
11 Disposal 118
12 Technical data 120
13 Annex: Data communication 124
13.1 Interfaces .......................................................................................................... 124
13.1.1 RS 232 interface ......................................................................................... 124
13.1.2 USB port (optional) .................................................................................... 124
13.1.3 Installing the USB port driver .................................................................... 125
13.2 Structure of the command sequences for data communication ................... 127
13.2.1 Description of the protocol ........................................................................ 127
13.3 Overview of general parameters (addresses 0xxx)........................................ 128
13.3.1 Reading general parameters ..................................................................... 128
13.4 Overview of incubator parameters (addresses 2xxx)..................................... 129
13.4.1 Reading basic parameters ......................................................................... 129
13.4.2 Reading internal function parameters....................................................... 130
13.5 Error memory structure ................................................................................... 132
13.5.1 Error memory data set structure scheme:................................................ 133
13.5.2 Overview of the possible error messages in hex coding......................... 133
Page 6
Operating Instructions | HERAcell® 150i / 240i
Contents
6
50115191 A / 08.2009
13.5.3 Overview of the possible error messages in bit coding........................... 134
13.6 Data logger structure ....................................................................................... 136
13.6.1 Data logger data set structure scheme:.................................................... 137
13.6.2 Overview of possible event entries in bit coding ..................................... 138
13.7 Examples of data logger codes ....................................................................... 140
13.7.1 Functions for data logger queries ............................................................. 140
13.7.2 Example of a code for a a data logger query............................................ 140
13.8 Program Heracell 150i & 240i........................................................................... 145
13.8.1 Installing the Heracell 150i & 240i.............................................................. 146
13.8.2 Operating the Heracell 150i & 240i ............................................................ 147
14 Device log 154
15 Certificate of decontamination 155
Page 7
Betriebsanleitung | HERAcell® 150i / 240i
Figures
7
50115191 A / 08.2009
Fig. 1: Device dimensions.................................................................................... 23
Fig. 2: Lift points................................................................................................... 24
Fig. 3: Stacking ..................................................................................................... 24
Fig. 4: Front view HERAcell
®
150i ....................................................................... 27
Fig. 5: Rear view HERAcell
®
150i ........................................................................ 29
Fig. 6: Looped-through gas supply HERAcell
®
150i........................................... 31
Fig. 7: Front view HERAcell
®
240i ....................................................................... 33
Fig. 8: Rear view HERAcell
®
240i ........................................................................ 35
Fig. 9: Looped-through gas supply HERAcell
®
240i........................................... 37
Fig. 10: Door switch ............................................................................................... 40
Fig. 11: Temperature, CO
2
,and O2 sensors .......................................................... 41
Fig. 12: Supply interface ........................................................................................ 42
Fig. 13: Work space components .......................................................................... 44
Fig. 14: Water reservoir.......................................................................................... 45
Fig. 15: Rear panel openings ................................................................................. 46
Fig. 16: Bottle turning device................................................................................. 47
Fig. 17: Water pump............................................................................................... 48
Fig. 18: Support rail installation/removal .............................................................. 49
Fig. 19: Shelf support installation .......................................................................... 50
Fig. 20: Installing gas humidification ..................................................................... 50
Fig. 21: Installing the bottle turning devices ......................................................... 51
Fig. 22: Bottle turning speed.................................................................................. 53
Fig. 23: Center strut installation............................................................................. 54
Fig. 24: Installing gas pressure hoses ................................................................... 55
Fig. 25: Gas connection without gas monitoring system..................................... 56
Fig. 26: Gas connection with gas monitoring system (optional) ......................... 57
Fig. 27: Combined connection with gas monitoring system (optional)............... 58
Fig. 28: Combined gas connection with gas monitoring system......................... 59
Fig. 28: Power supply connection ......................................................................... 59
Fig. 29: Connection example ................................................................................. 62
Fig. 30: Power switch ............................................................................................. 65
Fig. 31: Pressure-sensitive areas ........................................................................... 65
Fig. 32: iCan™ touchscreen withoutO
2/N2
gas supply ..........................................66
Fig. 33: iCan™ touchscreen with combined gas connection ............................... 66
Fig. 34: Menu structure overview.......................................................................... 67
Fig. 35: Heat-up phase display .............................................................................. 68
Fig. 36: Setting the temperature set value ............................................................ 70
Fig. 37: Setting the CO
2
set value.......................................................................... 71
Fig. 38: Setting the O
2
set value ............................................................................ 72
Fig. 39: Activating the auto-start routine............................................................... 75
Fig. 40: Interrupting the auto-start routine............................................................ 76
Fig. 41: Changing the code .................................................................................... 77
Fig. 42: Date / time setting ..................................................................................... 78
Fig. 43: Display contrast setting ............................................................................ 79
Fig. 44: Key tone setting ........................................................................................ 79
Fig. 45: RS 232 interface baud rate setting ........................................................... 80
Fig. 46: USB interface baud rate setting................................................................ 80
Fig. 47: Language setting....................................................................................... 81
Fig. 48: Reminder interval setting.......................................................................... 82
Fig. 49: Event display ............................................................................................. 83
Page 8
Betriebsanleitung | HERAcell® 150i / 240i
Figures
8
50115191 A / 08.2009
Fig. 50: Logging cycle setting................................................................................ 83
Fig. 51: Error table display ..................................................................................... 84
Fig. 52: Alarm relay setting.................................................................................... 85
Fig. 53: Low humidity setting ................................................................................ 86
Fig. 54: Gas tight screen setting ............................................................................ 87
Fig. 55: Water level setting .................................................................................... 87
Fig. 56: Alarm relay setting.................................................................................... 88
Fig. 57: Bottle turning device speed setting.......................................................... 88
Fig. 58: Switching the O
2
control on and off......................................................... 89
Fig. 59: Icon description overview ........................................................................ 90
Fig. 60: Gas monitoring icons................................................................................ 91
Fig. 61: Enabling/disabling the keypad lock.......................................................... 92
Fig. 62: Software versions ..................................................................................... 93
Fig. 63: Trend display scaling ................................................................................ 94
Fig. 64: Event error messages ............................................................................... 95
Fig. 65: Decontamination routine phases ........................................................... 106
Fig. 66: Activating contra-con.............................................................................. 107
Fig. 67: Interrupting / cancelling contra-con ....................................................... 108
Fig. 68: contra-con completion............................................................................ 109
Fig. 69: Preparing the temperature calibration ................................................... 111
Fig. 70: Temperature calibration procedure ....................................................... 112
Fig. 71: Preparing CO
2
calibration ....................................................................... 113
Fig. 72: Gas-tight screen measurement opening............................................... 114
Fig. 73: CO
2
calibration procedure ...................................................................... 115
Fig. 74: Gas inlet filter installation........................................................................ 116
Fig. 75: Device fuse replacement ........................................................................ 116
Fig. 76: Magnetic door seal replacement............................................................ 117
Page 9
Operating Instructions | HERAcell® 150i / 240i
9
50115191 A / 08.2009
1 General notes
Overview of the Thermo Fisher international sales organisations
Europe, Middle East, Africa Postal address Germany:
Thermo Electron LED GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold
Enquiries from Germany:
Phone Sales 0800 1 536376 Phone Service 0800 1 112110 Fax Sales/Service 0800 1 112114 E-Mail info.labequipment.de@thermofisher.com
Enquiries from Europe, Middle East, and Africa:
Phone + 49 (0) 6184 / 90-6940 Fax + 49 (0) 6184 / 90-6772 E-Mail info.labequipment.de@thermofisher.com
America Postal address USA:
Thermo Fisher Scientific 275 Aiken Road Asheville, NC 28804 USA
Enquiries from North America:
Phone +1 800-879 7767 Fax +1 828-658 0363 E-Mail info.labequipment@thermofisher.com
Enquiries from Latin America:
Phone +1 828-658 2711 Fax +1 828-645 9466 E-Mail info.labequipment@thermofisher.com
Enquiries from Asia Pacific:
Phone +852-2711 3910 Fax +852-2711 3858 E-Mail info.labequipment@thermofisher.com
Page 10
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
10
50115191 A / 08.2009
1.1 Identification of the device and of the documentation
Identification of the device
Identfication data Device name: CO
2
incubator
Type designation: HERAcell
®
150i
HERAcell
®
240i
Serial No.: 40830469 or higher
Allocation of product documentation
User information: Operating instructions 50113925 D Valid: 08.2009
Certifications and quality audit:
Certification: CE Certification Test mark: GS by VDE Germany
CSA/UL by VDE Germany
1.2 Instruction of the operating personnel
These operating instructions describe the CO2 incubator HERAcell® 150i / 240i. The CO
2
incubator has been manufactured in keeping with the latest technological developments and is operationally safe. However, the device may present potential hazards, particularly if it is operated by inadequately trained personnel or if it is not used in accordance with the intended purpose. Therefore, the following must be observed to prevent accidents:
Target group •The CO
2
incubator must be operated only by trained and authorized
personnel.
Process procedures for personnel • For any operation of this device, the operator must prepare clear and
concise written instructions in the language of the operating and cleaning personnel based on these operating instructions, applicable safety data sheets, plant hygiene guidelines, and technical regulations, in particular:
– which decontamination measures are to be applied for the CO
2
incubator and accessories, – which protective measures apply when specific agents are used, – which measures are to be taken in the case of an accident.
• Repairs to the device must be carried out only by trained and authorized expert personnel.
Page 11
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
11
50115191 A / 08.2009
1.3 Applicability of the instructions
Original document for translation • The contents of the operating instructions are subject to change
without further notice.
• Concerning translations into foreign languages, the German version of these operating instructions is binding.
• Keep these operating instructions close to the device so that safety instructions and important information are always accessible.
Should you encounter problems that are not detailed adequately in these operating instructions, please contact Thermo Scientific immediately for your own safety.
1.4 Warranty
Requirements for warranty Thermo Scientific warrants the operational safety and functions of the CO
2
incubator only under the condition that:
• the device is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions,
• the device is not modified,
• only original spare parts and accessories that have been approved by Thermo Fisher Scientific are used,
• inspections and maintenance are performed at the specified intervals.
The warranty is valid from the date of delivery of the device to the operator.
Page 12
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
12
50115191 A / 08.2009
1.5 Explanation of safety information and symbols
1.5.1 Safety information and symbols used in the operating
instructions
WARNING: Personal injuries
Indicates a hazardous situation which, if not avoided, could result in death or serious injuries.
CAUTION: Personal injuries
Indicates a hazardous situation which, if not avoided, could result in minor or moderate injuries.
NOTICE: Property damage
Indicates a situation which, if not avoided, could result in property damage.
NOTE: Useful information
Is used for applicational hints and useful information.
1.5.2 Additional symbols for safety informations:
Wear safety gloves!
Wear safety goggles!
Harmful liquids!
Electric shock!
Hot surfaces!
Fire hazard!
Explosion hazard!
Suffocation hazard!
Page 13
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
13
50115191 A / 08.2009
1.5.3 Symbols on the device
CE conformity mark: confirms conformity according to EU Guidelines
VDE test mark
Mark of conformity USA/Canada
Observe operating instructions!
H
Page 14
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
14
50115191 A / 08.2009
1.6 Intended purpose of the device
1.6.1 Correct use
Use The CO
2
incubator HERAcell® 150i / 240i is a laboratory device for preparing and cultivating cell and tissue cultures. The device allows the simulation of the special physiological ambient conditions for these cultures due to the exact control of:
• temperature,
•CO
2
content,
•O
2/N2
content,
• and increased relative humidity.
Fields of application HERAcell
®
150i / 240i has been designed for installation and operation in
the following fields of application:
• Laboratories for cytobiological and biotechnological experiments of safety levels L1, L2, and L3.
• Medical-microbiological laboratories in accordance with DIN 58 956.
• Laboratories in the central area of clinics and hospitals.
Gas supply system The gases required for the incubator (CO
2
and / or O2/N2) are supplied to the device from a separate gas supply system, either from gas cylinders or from a central pressurized gas container. The layout of the gas supply system must ensure that the operating pressure of the gas supply lines can be set to a range between 0.8 bar (min.) to 1 bar (max.) and that the pressure cannot be changed. If the device is equipped with the optional gas monitoring system, up to four devices can be interconnected in line, independent of the of the gas supply system capacity. The CO
2
incubator is suitable for continuous operation.
1.6.2 Incorrect use
Incorrect use Do not use cell or tissue cultures in the device that are not in accordance
with the regulations of safety levels L1, L2, and L3. Do not use tissues, substances or liquids that:
• are easily ignitible or explosive,
• release vapors that form combustible or explosive mixtures when exposed to air,
• release poisons.
Page 15
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
15
50115191 A / 08.2009
1.7 Standards and directives
The device complies with the following standards and guidelines:
• DIN EN 61010 - 1: 2002, DIN EN 61010 - 2 - 010 : 2004
• Low Voltage Guideline 2006/95/EG
• EMC Guideline 2004/108/EG
• UVV BGV D4 Revised Version (UVV VBG 20, continues to be valid for old stock)
The following safety regulations must be observed if the device is operated within the territory of the Federal Republic of Germany:
•BGR 104
•BGR 120
•BGI 629
•BGI 630
• TRG 280
• EC Official Gazette, L 374
• Safety data sheets of the gas suppliers with regard to the particular characteristics of CO
2
, O2, and N2.
• Principles of good microbiological proceedings, notice of the trade association of the German chemical industry.
For other countries, the applicable national regulations are binding.
Page 16
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
16
50115191 A / 08.2009
1.8 Safety notes on gases
Installation work:
Any work to supply lines and pressurized gas containers, cylinders or containers used for storing CO
2
or O2/N2 must only
be carried out by expert personnel using the appropriate tools.
Instruction of the personnel:
Personnel operating devices with CO
2
supply must be
instructed about the particularities in the handling of CO
2
before starting their work:
• Correct operation of pressurized gas containers and gas supply systems,
• obligation to report damages and shortcomings in CO
2
supply lines,
• measures to be taken in case of accidents or failures.
These instructions must be repeated at appropriate intervals and must comprise the particular operating instructions of the gas supplier.
1.8.1 Safety notes on
carbon dioxide (CO
2
)
Carbon dioxide Since CO
2
is rated as a harmful gas, certain safety instructions must be
observed when the CO
2
incubator is started up and when the device is
operated.
Suffocation hazard!
Large amounts of CO
2
released into the room atmosphere may cause suffocation. If CO
2
is released, initiate safety measures immediately!
• Leave the room immediately and do not allow others to
enter the room!
• Inform security service or fire department!
Page 17
Operating Instructions | HERAcell® 150i / 240i
1 General notes1 General notes
17
50115191 A / 08.2009
1.8.2 Safety notes on oxygen (O2)
Oxygen O
2
is a gas that promotes combustion and may explode in combination
with grease-containing materials.
Oxygen explosion!
O
2
may explode in combination with oils, greases, and lubricants. If highly compressed oxygen comes in contact with grease- or oil-containing substances, the mixture may explode!
• For cleaning these device components, use only oil- and
grease-free lubricants.
• Keep all connections and components of the oxygen system
free from substances that contain oil, grease, or lubricant!
Fire hazard!
Released oxygen promotes combustion. Do not use open flames in the vicinity of oxygen-operated systems!
• Do not smoke in the vicinity of oxygen systems.
• Do not expose the components of an oxygen system to
excessive heat.
1.8.3 Safety notes on nitrogen (N
2
)
Nitrogen Nitrogen mixes easily with air. High concentrations of nitrogen reduce the
oxygen content in the air.
Suffocation hazard!
Large amounts of N
2
released into the room atmosphere may cause suffocation. If N
2
is released, initiate safety measures immediately!
• Leave the room immediately and do not allow others to
enter the room!
• Inform security service or fire department!
Page 18
Operating Instructions | HERAcell® 150i / 240i
18
50115191 A / 08.2009
2 Delivery of the device
2.1 Packaging
The CO2 incubator HERAcell® 150i / 240i is delivered in a stable packaging box. All packaging materials can be separated and are reusable:
Packaging materials •Packaging carton:
Recycled paper
• Foam elements: Styrofoam (CFC-free)
•Pallet: Untreated wood
•Packaging film: Polyethylene
• Packaging ribbons: Polypropylene
2.2 Acceptance inspection
After the device has been delivered, check the delivery immediately for:
• completeness,
• possible damage.
Damage report If components are missing or damage is found on the device or the
packaging, esp. damages caused by humidity and water, please contact the linehauler as well as the Technical Support of Thermo Scientific immediately.
Page 19
Operating Instructions | HERAcell® 150i / 240i
2 Delivery of the device2 Delivery of the device2 Delivery of the device
19
50115191 A / 08.2009
2.3 Standard equipment components HERAcell® 150i
2.4 Optional equipment components HERAcell® 150i
*1) O2 sensor with gas humidification set is packed in a separate carton and placed into
the work space during transport.
Solid glass door and continuous shelves
3-door gas tight screen and continuous shelves
Shelf 3 3
Support rail for shelf 4 4
Shelf supports for shelf 6 6
Insert for pressure compensation opening 1 1
Plug for pipe channel 1 1
Power supply cable 1 1
Connector, potential-free contact 1 1
Spare caps, set 1 1
CO
2
connecting hose set 1 1
Open-end wrench, 24 mm 1 1
Allen wrench 2 mm for blower wheel 1 1
Allen wrench 3 mm for blower wheel cover 1 1
Operating instructions (CD) 1 1
Water pump 1 1
Quantity of the delivered components (in piece)
HERAcell
®
150i
CO
2
incubator or CO2/O2 incuba tor
Solid glass door and continuous shelves
3-door gas tight screen and continuous shelves
O2 connecting hose set
11
O2 sens or wit h set for gas humidific ation *1)
11
CO2 connec ting hos e set, gas monitoring (optional)
11
O2 / N2 connec ting hos e set, gas monitoring (optional)
11
Quantity of the delivered components (in piece)
HERAcell
®
150i
CO
2
incubator or CO2/O2 incuba tor
Page 20
Operating Instructions | HERAcell® 150i / 240i
2 Delivery of the device2 Delivery of the device2 Delivery of the device
20
50115191 A / 08.2009
2.5 Standard equipment components HERAcell® 240i
Solid glass door and continuous shelves (standard version)
Solid glass door and split shelves (optional)
6-door gas tight screen and continuous shelves (standard version)
6-door gas tight screen and split shelves (optional)
Shelf 3 6 3 6
Support rail for shelf 4 6 4 6
Shelf supports for shelf 6 12 6 12
Insert for pre ssure compensation opening
111 1
Plug for pipe channel 1 1 1 1
Power supply cable 1 1 1 1
Connector, potential-free contact
111 1
Spare caps, set 1 1 1 1
CO
2
connecting hose set 1 1 1 1
Open-end wrench, 24 mm 1 1 1 1
Allen wrench 2 mm for blower wheel
111 1
Allen wrench 3 mm for blower wheel cover
111 1
Operating instructions (CD) 1 1 1 1
Water pump 1 1 1 1
Quantity of the delivered components (in piece)
HERAcell
®
240i
CO
2
incuba tor or CO2/O2 incubator
Page 21
Operating Instructions | HERAcell® 150i / 240i
2 Delivery of the device2 Delivery of the device2 Delivery of the device
21
50115191 A / 08.2009
2.6 Optional equipment components HERAcell® 240i
*1) O2 sensor with gas humidification set is packed in a separate box that is placed into
the work space during transport. *2) For bottle turning devices, the version with split shelves is not available. *3) Each level of a bottle turning device is packed in a separate box. The boxes (1 to 4)
are delivered with the device.
Solid glass door and continuous shelves (standard version)
Solid glass door and split shelves (optional)
6-door gas tight screen and continuous shelves (standard version)
6-door gas tight screen and split shelves (optional)
O
2
connecting
hose set
111 1
O
2
gas humidification system connecting hose set *1)
111 1
Bottle turning device (motor with four rollers, optional) *2), *3 )
1—1 —
CO
2
connecting hose set,
gas monitoring (optional)
111 1
O
2
/ N2 connecting hose set, gas monitoring (optional)
111 1
Quantity of the delivered components (in piece)
HERAcell
®
240i
CO
2
incuba tor or CO2/O2 incubator
Page 22
Operating Instructions | HERAcell® 150i / 240i
22
50115191 A / 08.2009
3 Installation of the device
3.1 Ambient conditions
The device must only be operated at locations that meet the particular ambient conditions listed below:
Requirements:
Location • Draft-free and dry location.
• The minimal distance to adjacent surfaces must be observed on all sides (see Section 3.3.).
• The operating room must be equipped with appropriate room ventilation.
• Solid, level, fire-proof surface.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the weight of the device and of accessories (particularly if several devices are stacked).
• The electrical system of the device has been designed for an operating height of up to 2000 m above sea level.
• To ensure a constant incubation temperature of 37 °C, the ambient temperature must be within a range of +18 °C to +33 °C.
• Relative humidity up to 80% (max.).
• No direct exposure to sunlight.
• Devices that produce excessive heat must not be placed near the location of the HERAcell
®
150i / 240i.
3.2 Room ventilation
When CO2/O2/N2 is supplied, the work space of the incubator is slightly pressurized. The pressure is released through the pressure compensation opening into the operating room. Pressure compensation and any opening of the glass door/gas-tight screen during the operation of the device will release very small quantities of CO
2
/
O
2/N2
into the operating room; the room ventilation must be capable of carrying the released gas safely off into the open. In addition, heat dissipating from the device during continuous operation may cause a change in the room climate.
• Therefore, the HERAcell
®
150i / 240i must only be installed in rooms
with sufficient ventilation.
• Do not install the device in room recesses without ventilation.
Technical ventilation • The room ventilation should be a technical ventilation that complies
with the requirements of BGR 120 (Guidelines for laboritories, Germany) or some other suited ventilation system with appropriate capacity.
Page 23
Operating Instructions | HERAcell® 150i / 240i
3 Installation of the device3 Installation of the device3 Installation of the device
23
50115191 A / 08.2009
3.3 Space requirements
Fig. 1: When installing the device, make sure that the
installation and supply connections are freely accessible. The control box on the rear panel of the device may serve as a spacer to adjacent objects. The side distances given are minimal distances. To protect the CO
2
incubator against contamination, use a floor stand even if the device is installed near the floor. The height of the floor stand should not fall below 200 mm.
Accessibility of the devices:
To ensure the accessibility for care and maintenance works, keep larger side and rear distances.
Fig. 1: Device dimensions
Page 24
Operating Instructions | HERAcell® 150i / 240i
3 Installation of the device3 Installation of the device3 Installation of the device
24
50115191 A / 08.2009
3.4 Transport
Fig. 2: For transport, do not lift the device using the
doors or components attached to the device (e.g. control box on rear panel) as lift points.
Lift points:
Lift the device only using the lift points.
3.5 Stacking
Fig. 3: HERAcell® 150i / 240i is suited for stacking up
to two devices of the same device type. For thermal separation, an adapter plate [3] can be placed between the two devices.
HERAcell
®
150i
• Place the adapter plate with the receptacles [4] at the bottom onto the stacking elements [5] on top of the device. Make sure that the adapter plate is seated safely on the stacking elements.
• Place the device to be stacked with the device stands [1] onto the stacking elements [2] at the top of the adapter plate.
HERAcell
®
240i
• Place the adapter plate with the receptacle recesses [8] at the bottom onto the stacking elements [9] on top of the device. Make sure that the adapter plate is seated safely on the recesses.
• Place the device to be stacked with the device stands [6] onto the stacking elements [7] at the top of the adapter plate.
The devices are secured to one another by their own weight.
Transporting stacked devices!
The stacking elements are not locking elements. Therefore, it is not permitted to transport stacked devices.
Fig. 2: Lift points
Fig. 3: Stacking
Page 25
Operating Instructions | HERAcell® 150i / 240i
3 Installation of the device3 Installation of the device
25
50115191 A / 08.2009
Mobile racks Fastening on mobile racks:
If the devices are installed onto mobile racks, make sure that the rollers [10] are secured with a locking brake during the operation of the incubators and that the rollers are oriented toward the front for increased stability.
Condensate formation during the operation of stacked devices:
If stacked devices of type HERAcell
®
150i are operated at an ambient temperature of more than 26 °C, condensate formation may occur on the upper device while the contra-con decontamination routine is run for the lower device. To prevent condensate formation, an adapter plate should be placed between the two devices for thermal separation.
3.6 Retrofitting/Modifications
Retrofitting options The following components can be retrofitted to the standard version:
HERAcell
®
150i
• 3-segment gas-tight screen (the gas-tight screen is installed in place of the glass door),
• door hinge replacement on outer door and glass door (RH / LH door opening),
• lockable outer door,
• split inserts,
• infrared (IR) sensor,
• PCB for USB interface.
HERAcell
®
240i
• 6-segment gas-tight screen (the gas-tight screen is installed in place of the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• split inserts,
• infrared (IR) sensor,
• PCB for USB interface.
Modifications:
Retrofittings and modifications must only be performed by the Technical Service of Thermo Electron LED.
Page 26
Operating Instructions | HERAcell® 150i / 240i
26
50115191 A / 08.2009
4 Description of the device
4.1 Front view of HERAcell® 150i
Front view [1] Stacking elements
[2] Plug caps [3] Glass door [4] Measuring cell with blower wheel and sensors [5] Door switch [6] Oxygen sensor (optional) [7] Pressure compensation opening with insert [8] Measuring opening
[9] Outer door [10] Outer door seal, replaceable [11] Stand, height-adjustable [12] Gas humidification (optional) [13] Water level sensor [14] Nameplate [15] Power switch [16] Support rail [17] Shelf [18] Latch, glass door [19] Support hook for shelf [20] Access port with plug [21] Glass door seal, replaceable [22] iCan™ Touchscreen (Operating panel) [23] Door handle
Page 27
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
27
50115191 A / 08.2009
Fig. 4: Front view HERAcell® 150i
Page 28
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
28
50115191 A / 08.2009
4.2 Rear view HERAcell® 150i
Rear view [1] Access port, Ø 42 mm
[2] Pressure compensation opening
[3] Control box with supply interface for combined gas connection
(optional) CO
2
and O2/N2 without gas monitoring [4] Condensate drain gutter [5] Diagram: Gas connection CO
2
and O2/N2 without gas monitoring
[6] O
2/N2
gas cylinder
[7] CO
2
gas cylinder
[8] Secondary CO
2
gas supply B with optional gas monitoring
[9] Primary CO
2
gas supply A with optional gas monitoring
[10] Secondary O
2/N2
gas supply B with optional gas monitoring
[11] Primary O
2/N2
gas supply A with optional gas monitoring
[12] Control box with supply interface for combined gas connection
(optional) CO
2
and O2/N2 with gas monitoring (optional)
[13] Distributor connection for looping through the CO
2
gas supply for
up to three additional devices (independent of device type)
[14] Distributor connection for looping through the O
2/N2
gas supply for
up to 3 additional devices (independent of device type)
[15] Diagram: Gas connection CO
2
and O2/N2 with optional gas
monitoring
Page 29
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
29
50115191 A / 08.2009
Fig. 5: Rear view HERAcell® 150i
Page 30
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
30
50115191 A / 08.2009
Looped-through gas supply for HERAcell® 150i:
[1] Control box with supply interface for combined gas connection
(optional) CO
2
and O2/N2 with gas monitoring (optional)
[2] Distributor connection for looping through the CO
2
gas supply for up
to three additional devices (independent of device type)
[3] Distributor connection for looping through the O
2/N2
gas supply for
up to 3 additional devices (independent of device type) [4] Tee joint for fitting the gas pressure hoses together [5] Gas pressure hose for looping-through the gas supply [6] Diagram: Gas connection CO
2
and O2/N2 with optional gas
monitoring [7] HERAcell
®
150i with looped-through gas supply; therefore device
must only be equipped with combined gas connection (optional) CO
2
and O
2/N2
Page 31
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
31
50115191 A / 08.2009
Fig. 6: Looped-through gas supply
HERAcell® 150i
Page 32
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
32
50115191 A / 08.2009
4.3 Front view HERAcell® 240i
Front view [1] Stacking elements
[2] Plug caps [3] Glass door [4] Measuring cell with blower wheel and sensors [5] Door switch [6] Oxygen sensor (optional) [7] Receptacle for bottle turning device (optional) [8] Measuring opening [9] Plug for bottle turning device (optional)
[10] Outer door [11] Outer door seal, replaceable [12] Stand, height-adjustable [13] Drive roller for bottle turning device (optional) [14] Gas humidification (optional) [15] Water level sensor [16] Insert for bottle turning device (optional) [17] Nameplate [18] Cylinders (optional) [19] Power switch [20] Support rail [21] Shelf [22] Latch, glass door [23] Support hook for shelf [24] Pressure compensation opening with insert [25] Access port with plug [26] Glass door seal, replaceable [27] iCan™ Touchscreen (Operating panel) [28] Door handle
Page 33
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
33
50115191 A / 08.2009
Fig. 7: Front view HERAcell® 240i
Page 34
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
34
50115191 A / 08.2009
4.4 Rear view HERAcell® 240i
Rear view [1] Pressure compensation opening
[2] Access port, Ø 42 mm [3] Condensate drain gutter [4] Control box with supply interface for combined gas connection
(optional) CO
2
and O2/N2 without gas monitoring
[5] Diagram: Gas connection CO
2
and O2/N2 without gas monitoring
[6] O
2/N2
gas cylinder
[7] CO
2
gas cylinder
[8] Secondary CO
2
gas supply B with optional gas monitoring
[9] Primary CO
2
gas supply A with optional gas monitoring
[10] Secondary O
2/N2
gas supply B with optional gas monitoring
[11] Primary O
2/N2
gas supply A with optional gas monitoring
[12] Control box with supply interface for combined gas connection
(optional) CO
2
and O2/N2 with gas monitoring (optional)
[13] Distributor connection for looping through the CO
2
gas supply for
up to three additional devices (independent of device type)
[14] Distributor connection for looping through the O
2/N2
gas supply for
up to 3 additional devices (independent of device type)
[15] Diagram: Gas connection CO
2
and O2/N2 with optional gas
monitoring
Page 35
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
35
50115191 A / 08.2009
Fig. 8: Rear view HERAcell® 240i
Page 36
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
36
50115191 A / 08.2009
Looped-through gas supply for HERAcell® 240i:
[1] Control box with supply interface for combined gas connection
(optional) CO
2
and O2/N2 with gas monitoring (optional)
[2] Distributor connection for looping through the CO
2
gas supply for up
to three additional devices (independent of device type) [3] Distributor connection for looping through the O
2/N2
gas supply for
up to 3 additional devices (independent of device type) [4] Tee joint for fitting the gas pressure hoses together [5] Gas pressure hose for looping-through the gas supply [6] Diagram: Gas connection CO
2
and O2/N2 with optional gas
monitoring [7] HERAcell
®
240i with looped-through gas supply; therefore device
must only be equipped with combined gas connection (optional) CO
2
and O
2/N2
Page 37
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
37
50115191 A / 08.2009
Fig. 9: Looped-through gas supply
HERAcell® 240i
Page 38
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
38
50115191 A / 08.2009
4.5 Safety devices
The device is equipped with the following safety devices:
• A door switch interrupts the CO
2/O2/N2
supply and the work space
heating when the glass door is opened.
• The optional gas detector switches the gas supply over to a full gas cylinder.
• An independent thermal protection protects the samples from harmful overheating in case of failures.
• A pressure compensation opening ensures pressure compensation in the device work space.
• The alarm relay circuit uses audible and visual alarms to indicate errors during operation.
4.6 Work space atmosphere
In the work space of the incubator, the particular physiological ambient conditions for the preparation and cultivation of cell and tissue cultures are simulated. The work space atmosphere is determined by the following factors:
• Temperature
• Relative humidity
•CO
2
concentration
•O
2
concentration (optional)
Temperature:
Temperature conditions To ensure undisturbed operation, the temperature in the operating room
must be at least 18 °C and the incubation temperature must exceed this temperature by at least 3 °C. The heating system controls the incubation temperature from this temperature threshold up to 55 °C. The principle of air jacket heating and the additional, separate heating of the outer door and glass door/gas-tight screen minimize the risk of condensate formation at the side walls, at the ceiling of the work space, and at the glass door/gas-tight screen.
Relative humidity:
Water quality The heating of the work space promotes the condensation of the water,
thereby ensuring a constant humidity within the work space. For the running operation, always keep a sufficient quantity of processed water of the following quality available:
•HERAcell
®
240i: quantity 4.5 l,
•HERAcell
®
150i: quantity 3.0 l,
Water quality recommendation:
For trouble-free operation of the equipment, fill the water reservoir with sterilized destilled water or equivalent. The acceptable conductivity should be within the range of 1 to 20 μS (resistivity within the range of 50 k to 1 MΩ).
Page 39
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
39
50115191 A / 08.2009
Termination of warranty!
Using chlorinated tap water or additives that contain chlorine will void the manufacture warranty. Similarly, the use of ultrapure water whose conductivity is out of the range of 1 to 20 μS (and whose resistivity is out of the range of 50 k to 1 M) will void the manufacture warranty. If you should have any questions, please contact Technical Support of Thermo Fisher Scientific.
Under normal operating conditions and at the usual incubation temperature of 37 °C, a constant relative humidity of approx 93% is achieved in the work space. If dew formation occurs on the culture containers due to elevated relative humidity, the humidity in the work space can be adapted to a lower value by enabling the Low Humidity feature. The relative humidity in the work space is lowered from approx. 93% to approx. 90%. The modification requires an extended adaption phase. To ensure that it effectively prevents dew formation on culture containers, it must be used as a permanent setting.
CO
2
supply:
CO2-Work space atmosphere control To ensure the growth conditions for the cell and tissue cultures, the work
space is supplied with CO
2
. The pH of the bicarbonate-buffered culture media largely depends on the CO
2
content of the work space atmosphere.
The CO
2
content of the work space atmosphere can be controlled within a range of 0 to 20%. The supplied CO
2
must have one of the following quality characteristics:
• Purity 99.5% min.,
• medical gas quality.
O
2
supply:
O2-Work space atmosphere control If the CO
2
incubator is to be operated with more than 21% oxygen, the
work space is supplied with oxygen. The O
2
content of the work space
atmosphere can be controlled within a range of 0 to 21%.
N
2
supply:
N2-Work space atmosphere control If the oxygen content during operation is to be lowered to less than 21%
(air oxygen content), the work space is supplied with nitrogen. The O
2
content of the work space atmosphere can be controlled depending on the configuration of the sensor within a range of 1% to 21% and 5% to 90%.
Page 40
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
40
50115191 A / 08.2009
4.7 Door switch
A door switch [1] is installed at the upper edge of the work space opening. If the door switch is activated by opening the glass door, the gas supply and the heating of the work space are interrupted and the display shows a corresponding message. If the door remains open for more than 30 seconds, a short audible alarm sounds. If the door remains open for more than 10 minutes, an audible alarm sounds and the alarm relay responds.
The outer door can only be closed after the glass door has been latched properly.
Versions with gas-tight screen:
For devices with the optional gas-tight screen, the door switch function described above is activated when the outer door is opened.
Fig. 10: Door switch
Page 41
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
41
50115191 A / 08.2009
4.8 Sensor system
The blower wheel and two sensor modules are integral to the baseplate [1] of the measuring cell:
• Sensor [2] for the acquisition of the work space temperature and the thermal protection,
•CO
2
sensor [3] for the acquisition of the CO2
content in the work space atmosphere.
The O
2
sensor (optional) [4] for the acquisition of the
O
2
content in the work space atmosphere is installed
at various locations next to the measuring cell:
•HERAcell
®
150i in the sidewall in the upper area [5],
•HERAcell
®
240i in the ceiling of the work space [4].
The sensor for the acquisition of the work space temperature as well as the CO
2
sensor and the O2 sensor are integral to the control system of the device. Their measured values are compared to the selected set values. Based on this data, the control system controls heating and CO
2/O2/N2
supply. The blower intermixes the supplied gases and ensures an even temperature distribution within the work space.
The thermal protection has been preprogrammed at the factory and cannot be changed. It protects the stored cultures from overheating. If the set temperature is exceeded by more than 1°C, the thermal protection responds and the work space temperature is automatically reduced to the selected set value so that the incubation process can be continued even in case of a failure. Any response of the thermal protection will simultaneously trigger a visual alarm. If the thermal protection is enabled:
• an error message is issued,
• the alarm relay responds. If the error message is accepted, the display shows the Overtemperature icon to indicate the response of the thermal protection, and the temperature display is highlighted in red.
Fig. 11: Temperature, CO
2
,and
O
2
sensors
Page 42
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
42
50115191 A / 08.2009
4.9 Supply interface
All supply connections are installed in the supply interface (control box [1]) in the rear of the device.
Gas connection:
The gas supply lines between the device and the gas supply system are connected using the supplied connecting hoses. O
2/N2
is supplied to the device
through connecting sleeve [2]. CO
2
is connected to sleeve [3]. All process gases must be supplied to the device at a fixed pressure that has been preset within a range of
0.8 to 1.0 bar and must remain unchanged. Before the gases are fed into the work space, they flow through a gas inlet filter with a separation rate of 99.97 % for a particle size of 0.3 μm (HEPA filter quality). The illustration shows the optional combined gas connection without the optional gas monitoring system.
Label:
The label [4] contains information about gas supply, an alarm contact terminal legend, and notes about the electrical fusing of the device.
RS 232 interface:
Via the RS 232 interface [5], the incubator can be connected to the serial interface of a PC. This connection allows the computer-aided acquisition and documentation of the major operating parameters (temperature, CO
2/O2/N2
concentration, error codes,
etc.).
USB interface (optional):
As an alternative to the RS 232 interface [5], the incubator can be connected to a PC via the optional USB interface [6]. This connection - USB 1.1 / USB 2.0 full speed compatible - allows the computer­aided acquisition and documentation of the major operating parameters (temperature, CO
2/O2/N2
concentration, error codes, etc.).
Fig. 12: Supply interface
Page 43
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device
43
50115191 A / 08.2009
Alarm contact:
External alarm system Fig. 12: The device can be connected to an on-site, external alarm system
(e.g. telephone system, building monitoring system, visual or audible alarm system). For this purpose, a potential-free alarm contact [7] is preinstalled in the device.
Alarm contact:
The alarm contact responds for all errors reported by the control loops (see chap. 7.13‚ pg. 95).
Power supply connection:
Fig. 12: The connection to the power supply system is established by connecting a cable with plug for IEC connector [9] to the socket [8]. The receptacle for the two device fuses is integral to the socket.
Page 44
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
44
50115191 A / 08.2009
4.10 Work space components
The work space of the incubator has only a minimum of surface, thereby supporting both the prevention of contamination and easy, effective decontamination.
4.10.1 Interior container
All components of the work space are made of stainless steel and have a high-gloss polished, absolutely smooth and easy-to-clean surface. Any embossings have a large radius. As an option, the interior container, the shelf system, and the blower wheel with its cover can be made of copper material.
Oxidation of copper components:
When exposed to heat and humid air, the copper material of the interior container will oxidize. This results in a discoloration of the copper components during the test run for checking the device functions. Do not remove the oxide layer during routine cleaning as the antimicrobial effect of the copper material is based on it.
The components of the shelf system can be removed easily so that only the easily treatable, surface­reduced interior container [1] remains in the work space for cleaning and manual disinfection works.
4.10.2 Optional split gas-tight screens
Equipment versions:
•HERAcell
®
150i with 3-segment gas-tight screen
[2],
•HERAcell
®
240i with 6-segment gas-tight screen
[3].
For devices that are equipped with the optional gas­tight screen, the contamination hazard is considerably lower and the recovery times of the incubation parameters are shorter for:
• work space temperature,
•CO
2
concentration,
•O
2/N2
concentration,
• relative humidity.
Advantage of the gas-tight screen: The cross sections
Fig. 13: Work space components
Page 45
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
45
50115191 A / 08.2009
of the openings are smaller when samples are accessed.
4.10.3 Water reservoir
The water reservoir [1] is integral to the interior container floor and inclines toward the rear. The water level is monitored by a water level sensor [2] that issues an alarm message at the display and an audible signal when the water falls below the minimal level. The embossings [3] in the water tray are used as indicators for the maximal level.
4.10.4 Gas humidification (optional, only for O
2
control)
A hose [5] connects the gas humidification [6] to the device-integral oxygen or nitrogen supply line [4]. The inflowing oxygen or nitrogen is fed to the heated water. This ensures humidification of the gases as they enter the work space and prevents an undesired drop of the work space humidity.
4.10.5 Heating system
An air jacket heating is used for heating the work space. The arrangement of the heating elements ensures that condensate formation above the water reservoir is prevented as far as possible. The outer door of the device is also heated. The heat radiated onto the interior glass door/gas-tight screen prevents condensate formation. The work space of the device always remains visible, despite high humidity.
Fig. 14: Water reservoir
Page 46
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
46
50115191 A / 08.2009
4.10.6 Rear panel openings
A pressure compensation opening with insert in the rear panel of the device allows a compensation between the pressures in the work space and in the operating room. A sealable access port allows cables, hoses or additional sensors to be routed into the work space of the device.
HERAcell
®
150i: – access port, Ø 42 mm [1] – pressure compensation opening [2]
HERAcell
®
240i: – access port, Ø 42 mm [3] – pressure compensation openingt [4]
Operating conditions:
When accessories are to be operated in the work space of the CO
2
incubator, the ambient condition requirements must be observed (see table below). The energy introduced into the work space affects the beginning of the temperature control range. When additional heating sources are introduced into the work space, condensation (e.g. at the glass door) may occur.
Fig. 15: Rear panel openings
Introduced
energy
General
Example: RT* = 21° C
0 W RT + 3° C 24° C
5 W RT + 6,5° C 27,5° C
10 W RT + 9,5° C 30,5° C
15 W RT + 13° C 34° C
20 W RT + 16° C 37° C
Control range of the temperature
*RT = Ambient temperature
Page 47
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
47
50115191 A / 08.2009
4.10.7 Shelf system Fig. 16: The support rails [1] of the shelf system are
perforated every 42 mm so that the support hooks [8] can be inserted variably for any culture container size required. The shelves [2] have an integrated tilt protection and withdrawal stop. For details of the shelf system, see Section „Start-up“.
4.10.8 Bottle turning device for HERAcell® 240i
(optional)
Fig. 16: HERAcell
®
240i may be equipped with up to four separate bottle turning devices. Each unit consists of a roller insert [6] with a drive roller [4] and four idle rollers [7] that are also rotated due to the transmission function of the inserted cylinders [5]. Each unit is controlled separately as the drive roller is connected to the device control unit by a connector [3] at the right side of the device. This modular layout also allows the operation of only one bottle turning device in combination with several shelves. The arrangement of the idle rollers in the roller insert frame depends on the configuration of the cylinders used. The drive roller is always inserted into the right roller guide. The position and the number of the idle rollers in the frame then depend on the cylinder diameter and on the number of cylinders. The figure shows three examples of different idle roller positions in dependence of the cylinder diameter. The arrangement of the roller inserts in the work space is designed alphabetically as a, b, c, and d with a being the lowest and d being the uppermost roller insert. The connections of the drive rollers are assigned correspondingly to the device control unit. The drive roller of the switching insert a must only be connected to the lower receptacle a of the corresponding control unit.
Connection assignment:
If the drive roller is not properly connected to the control unit, the parameters cannot be set correctly on the operating panel of the device.
Fig. 16: Bottle turning device
Page 48
Operating Instructions | HERAcell® 150i / 240i
4 Description of the device4 Description of the device4 Description of the device
48
50115191 A / 08.2009
Receptacles:
To prevent corrosion due to moisture in the work space, unassigned receptacles of the bottle turning device must always be plugged with the protective cap.
The rotating speed of the bottle turning devices is set depending on two parameters:
• Diameter (of the cylinders used),
• drive roller set value.
The value is determined using a diagram (see chap.
5.4‚ pg. 51).
4.10.9 Water pump Fig. 17: The water pump serves for suction cleaning of
the remainig water in the water reservoir. The water drain works through gravity.
1. Place the water pump [1] on lowest shelf [5].
2. Feed the inlet hose [4] through a whole [7] of the lowest the shelf to the water reservoir [8].
3. Prepare a bucket [2].
4. Suck the water, therefore pump the water pump using the hand grip [6] about four times until water flows in the outlet hose.
5. Let the water run into the bucket [2] through gravity.
6. Wipe out the rests of the water at the floor plate of the water reservoir.
Fig. 17: Water pump
Page 49
Operating Instructions | HERAcell® 150i / 240i
49
50115191 A / 08.2009
5 Start-up
5.1 Preparing the work space
7. Upon delivery, the CO2 incubator is not in a sterile state. Before the initial start-up, the device must be decontaminated.
Before the decontamination is performed, the following work space components must be cleaned:
•Support rails,
• support hooks,
•shelves,
• bottle turning assembly (optional),
• gas humidification (optional),
• work space surfaces,
• work space seals and gaskets,
• glass door/gas-tight screen.
Decontamination:
For details about the cleaning and disinfection of the device, (see chap. 9‚ pg. 100).
5.2 Installing the shelf system
Tools are not required for the installation of the shelf system. The support rails are secured using spring pressure. After the support hooks have been inserted into the rail, the shelves are pushed onto the support hooks.
5.2.1 Installation/removal of the support rail
The support rails are held at the sides by the embossings [2] and [5] and secured by the embossings [1] and [6]. The support rails marked with the diamond shape are inserted at the rear panel of the device with the locksprings [3] facing upward.
1. Position the support rail [4] onto the lower embossing [6] and tilt it toward the work space side wall so that the rail is positioned over the two embossings [5] and [2].
2. Clamp the lockspring [3] behind the upper embossing [1].
3. To remove the support rails, pull the lockspring tab down out of the embossing and remove the rail.
Fig. 18: Support rail installation/removal
Page 50
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
50
50115191 A / 08.2009
5.2.2 Installing the shelf supports:
1. Fig. 19: Insert the shelf supports [3] into the perforations [1] of the support rail with the bar facing down.
2. Make sure that the two vertical elements [2] of the shelf support are flush with the support rail.
Installing the shelves:
1. Push the shelf [4] onto the shelf supports with the tilt protection [5] facing the rear panel of the device. The tilt protection [5] is also used as a guide for the shelf.
2. Slightly raise the shelf so that the withdrawal stop [6] can be routed over the shelf supports.
3. Make sure that the shelf support is positioned in the two tilt protections in a way that allows it to move freely.
5.2.3 Levelling the device
1. Position a bubble level onto the center shelf or onto the roller holder.
2. Rotate the adjustable device stands using the supplied 24 mm wrench until the shelf is positioned horizontally in all directions. Perform the adjustment of the device stands from left to right and from rear to front.
5.3 Installing the gas humidification (optional, only for O2/N2-control)
Fig. 20: The gas humidification [1] is installed to the
water tray parallel to the rear panel of the device. The position to the right side wall is predetermined by the hose length.
1. Install hose [2] to the sleeve of the gas humidification and then to the sleeve [3] of the device-integral oxygen or nitrogen supply line.
2. Place gas humidification immediately to the rear panel of the device and align it parallel to the rear panel.
Fig. 19: Shelf support installation
Fig. 20: Installing gas humidification
Page 51
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
51
50115191 A / 08.2009
5.4 Installing the optional bottle turning devices HERAcell® 240i
1. Align the roller insert so that the roller holder [8] and the crossmember [7] are flush and facing the work space opening.
2. Slide the roller insert with both guide rails [5] onto the shelf supports.
3. Slightly raise the roller insert so that the withdrawal stop can be routed over the shelf support.
4. Insert the drive roller [6] into the right receptacle and connect the cable [4] to the corresponding control unit a.
5. Insert the four idle rollers [3] the into receptacles and position the rollers in the receptacles in accordance with the bottle diameter.
6. Place the bottles onto the rollers. To ensure that the bottles will not be shifted against each other by the rotation, the bottoms of the bottles should contact the stop disks [2] of the rollers and the cylinder necks [1] should contact the rubber stop rings [8]. The stop ring can be positioned accordingly on the roller.
Set speed of the bottle turning device:
7. The rotating speed of the bottle turning devices is set depending on two parameters: – Diameter (of the bottles used), – drive roller set value.
Use of diagram:
8. The diagram shows the common bottle diameters as different graphs (ascending straight). – Set the desired speed at the Y axis. Read the
corresponding set value (in %) from the X axis at the point of intersection of the Y value with the specific graph.
Fig. 21: Installing the bottle turning devices
Page 52
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up
52
50115191 A / 08.2009
Example:
– Bottle diameter: 56 mm – Desired speed: 3.5 rpm – Determined set value: 48%
In the dialog box S
PEED BOTTLE TURN. DEVICE, the value 48% is set (see chap.
7.11.3‚ pg. 85).
Page 53
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
53
50115191 A / 08.2009
Fig. 22: Bottle turning speed
Page 54
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
54
50115191 A / 08.2009
5.5 Installing the optional center strut HERAcell® 240i for split inserts
If HERAcell® 240i is equipped with the optional 6­segment insert, two center struts [1] with perforations at each side must be installed. In this case the shelf support rails must be installed to the support rails at the sides and at the left and right side of the strut. The support rails marked with the diamond shape are installed at the rear panel of the device.
1. A pressure spring keeps the upper rectangular guide [3] of the center strut under tension. Insert this guide into the receptacle [2] in the work space ceiling und push it slightly upward so that it is pushed into the center strut.
2. Then, position the lower rod guide [5] of the center strut into the hole [6] in the work space floor and remove spring tension.
3. The center strut is stabilized by the spring pressure.
Configuration with split inserts:
For versions with split inserts, bottle turning devices cannot be installed.
Fig. 23: Center strut installation
Page 55
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
55
50115191 A / 08.2009
5.6 Gas connection
Gas quality:
The gases must have one of the following qualities:
• Purity 99.5 % min,
• medical gas quality.
Overpressure!
The operating pressure of the gas applied to the device must not exceed 1 bar. If the gas is supplied at a higher pressure, the valves integral to the device may not close correctly and the gas supply control may be impaired. Set the gas supply to a range between 0.8 bar min. and 1.0 bar max. and make sure that this pressure setting cannot be changed!
5.6.1 Installing gas pressure hoses
The gas supply from the gas supply system to the device is achieved using the supplied flexible gas pressure hoses:
1. Connect the gas pressure hose [2] to the sleeve of the gas supply system.
2. Remove the protective cap [3] from the sterile filter.
3. Slide the hose clamp [1] onto the gas pressure hose and connect hose to the sleeve [4] of the sterile filter [5].
4. Secure the gas pressure hose to the sleeve of the sterile filter using the hose clamp.
Pressure compensation opening:
To ensure permanent pressure compensation, the pressure compensation opening must not be connected to an exhaust air system. The pipe of the
Fig. 24: Installing gas pressure hoses
Page 56
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
56
50115191 A / 08.2009
pressure compensation opening must not be extended or redirected.
5.6.2 Installing devices without the optional gas monitoring system
The gas supply from the gas supply system to the device is established using the supplied flexible gas pressure hoses (see chap. 5.6.1‚ pg. 55):
CO
2
connection:
• For a device with CO
2
connection, connect the gas supply to the sterile filter [2] in accordance with the connection diagram [1].
Combined CO
2
and O2/N2 connection (optional):
For a combined CO
2/O2/N2
connection, proceed as follows in accordance with the connection diagram [3]:
• Connect the O
2/N2
supply line to the upper gas
inlet filter [4],
• connect the CO
2
supply line to the lower gas inlet
filter [5].
Manual gas monitoring:
On devices without the optional gas monitoring system, there is no automatic gas monitoring. Therefore, the filling level of the gas bottles must be checked every day.
Fig. 25: Gas connection without gas monitoring system
Page 57
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
57
50115191 A / 08.2009
5.6.3 CO2 connection with gas monitoring
system (optional)
The gas supply from the gas supply system to the device is established by the supplied flexible gas pressure hoses. Fig. 26: Devices with CO
2
connections that are equipped with the optional gas monitoring system are connected according to the connection diagram [7].
CO
2
connection:
For a device that is operated with process gas CO
2
and equipped with the optional gas monitoring system, connect the gas supply lines as follows:
• Distributor connection [5] for the CO
2
gas supply for up to three additional devices. For looping the gas supply through, interconnect the devices with a gas pressure hose.
• Connect the gas container B to the upper connection [2] of the gas monitoring system,
• connect the gas container A to the lower connection [3] of the gas monitoring system.
• The outlet of the gas monitoring system [4] is connected at the factory to the gas inlet filter [5] by a short gas pressure hose [6].
Looping the gas supply through:
By looping the gas supply through, up to 4 devices can be supplied by a CO
2
gas
supply.
5.6.4 Combined CO
2/O2/N2
connection with gas
monitoring system (optional)
The gas supply from the gas supply system to the device is established by the supplied flexible gas pressure hoses (see chap. 5.6.1‚ pg. 47). Fig. 27: Devices with the combined CO
2/O2/N2
connection that are equipped with the optional gas monitoring system are connected according to connection diagram [13] and have the following gas connections:
O
2/N2
supply:
• Distributor connection [1] for the O
2/N2
gas supply for up to three additional devices. For looping the gas supply through, interconnect the devices with a gas pressure hose.
• Connect the gas container B [2] to the upper
Fig. 26: Gas connection with gas monitoring system (optional)
Page 58
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
58
50115191 A / 08.2009
connection of the gas monitoring system,
• connect the gas container A [3] to the lower connection of the gas monitoring system,
• The outlet of the gas monitoring system [4] is connected at the factory to the gas inlet filter [9] by a short gas pressure hose [11].
CO
2
supply:
• Distributor connection [5] for the CO
2
gas supply for up to three additional devices. For looping the gas supply through, interconnect the devices with a gas pressure hose.
• Connect the gas container B to the upper connection [6] of the gas monitoring system,
• connect the gas container A to the lower connection [7] of the gas monitoring system.
• The outlet of the gas monitoring system [8] is connected at the factory to the gas inlet filter [10] by a short gas pressure hose [12].
Looping the gas supply through:
By looping the gas supply through, up to 4 devices (see annex at the end of this section) cyn be supplied by a CO
2
gas
supply or O
2/N2
gas supply.
Fig. 27: Combined connection with
gas monitoring system (optional)
Page 59
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
59
50115191 A / 08.2009
5.7 Power supply connection
Electric shock!
Contact with current-carrying components may cause a lethal electric shock. Before connecting the device to the power supply, check plug and power supply cable for damage. Do not use damaged components when connecting the device to the power supply!
The device must be connected only to a correctly installed and grounded power supply source:
•Fusing T 10 A
• Circuit breaker G 16
Connection to the power supply system:
1. Before connecting the device to the power supply, check to see if the voltage of the power supply corresponds with the specifications on the nameplate on the front of the device. If the ratings given for voltage (V) and current (A) are not correct, the device must not be connected to the power supply.
2. Connect the IEC connector [2] to the socket [1] of the control box of the device.
3. Connect the grounding plug [3] of the power supply cable to a correctly grounded and fused socket.
4. Make sure the power supply cable is not subjected to tensile or compressive force.
Fig. 28: Combined gas connection with gas monitoring system
Fig. 28: Power supply connection
Page 60
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up
60
50115191 A / 08.2009
5.8 Connecting the RS 232 interface
Data communication via RS 232 The RS 232 data communication interface has been designed for a cable
connection with 9-pin connector and a contact assignment of 1:1. Data exchange is accomplished by a predefined structure of command sequences (see annex at the end of this section).
Data communication diagram:
The command sequence syntax as well as the RS 232 interface data communication diagram are described in the annex at the end of this section.
Connecting the devices:
1. Turn the PC off.
2. Connect the serial interface cable connector (not included in the standard equipment) to the socket at the device supply interface at the rear of the device.
3. Connect the second connector to an unassigned serial port COM 1 / COM 2 etc. at the PC.
4. Turn the PC on.
5.9 Connecting the USB interface (optional)
Data communication via USB As an alternative to data communication via RS 232 interface, the devices
can be equipped with the optional USB interface. The USB interface complies with Standard USB 1.1 and is compatible with Standard USB 2.0 (full speed). The interface stepping rate can be changed within the defined baud rates (9,600, 19,200, 38,400, 57,600 baud). Data exchange is accomplished via a defined command sequence structure (frames). The command sequences correspond with the RS 232 interface installation diagram (see annex at the end of this section). Using the USB interface for data communication between a PC and the device a driver must be installed first (see annex data communication).
Page 61
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up
61
50115191 A / 08.2009
5.10 Connecting the alarm contact
Skilled work:
Thermo Scientific warrants the operational safety and the operativeness of the device only if installation and repairs are performed properly. The connection of the device to an external alarm system must only be carried out by adequately trained and authorized expert electrical/telecommunication personnel!
Function:
When system errors and failures occur in the temperature or gas control circuits, an alarm message is issued to the connected alarm/monitoring system. The potential-free contact (1 changeover contact) has been designed for the following circuits:
Alarm relay:
Switching structure:
The alarm relay switches for all failures reported by the control loops (see chap. 7.13‚ pg. 95).
Circuit Voltage External fusing
Circuits with sy ste m voltage max. 250 V ~ max. 6 A
25 V ~ max. 2 A
60 V = max. 1 A
50 V ~ max. 1 A
120 V = max. 0.5 A
SELV circuits (cf. VDE 0100, Part 410)
SELV-E circuits (cf. VDE 0100, Part 410)
Operating states
Contact
4 - 1
Contact
4 - 3
Operating state power failure "off" X O
Operating state power failure "on" O X Failure: power failure "off"
OX
Failure: power failure "on"
XO
X: Contact closed / O: Contact open
Page 62
Operating Instructions | HERAcell® 150i / 240i
5 Start-up5 Start-up5 Start-up
62
50115191 A / 08.2009
Connection example:
The connector [5] for the connecting cable is included in the standard equipment. The values for the operating voltage of the external circuits and of the fusing of the alarm system are given in the table on the previous page.
1. Connect the individual conductors [1] to [4] of the connecting cable as shown in the wiring diagram.
2. Connect the connector [5] of the alarm system connecting cable to the interface [6] at the control box at the rear panel of the device.
Fig. 29: Connection example
Page 63
Operating Instructions | HERAcell® 150i / 240i
63
50115191 A / 08.2009
6 Operation
6.1 Preparing the device
The device must only be released for operation after all major measures for the start-up have been taken (see Section 5, page 45).
Device check:
Device check Prior to starting operation, the following device components must be
checked for their correct function
• Gas hoses must be seated tight on the connecting filter and must be secured using a hose clamp.
• The access port must be capped.
• The pressure compensation opening must be permeable, its insert must be installed in the work space.
• The glass door/gas-tight screen seal must not be damaged.
• The glass door measurement opening must be capped.
• The shelf system components must be installed safely.
• The gas humidification (optional) must be connected to the device­integral gas supply and aligned parallel to the rear panel.
• The drive rollers of the bottle turning device (optional) must be properly connected to the corresponding control unit.
• The idle rollers of the bottle turning device (optional) must be positioned in accordance with the cylinder diameters.
Decontaminating the device work space:
Decontamination Run the contra-con decontamination routine (see chap. 9.4‚ pg. 104) or
decontaminate the work space according to the hygiene regulations set forth by the operator.
Hygiene regulations:
Prior to any operation, the user must clean and disinfect the work space in accordance with the hygiene regulations set forth by the operator to protect the cultures. The “Principles of good microbiological proceedings” at the annex of these instructions can be used as safety information for personnel working with the device.
Water supply:
A sufficient quantity of treated water of the following quality must always be available for the running operation:
• distilled, fully demineralized, and
• autoclaved, sterile. When the filling level drops below a minimum value, it must be possible to top up water during the work process. Capacities:
•HERAcell
®
150i: 3.0 l
•HERAcell
®
240i: 4.5 l
Page 64
Operating Instructions | HERAcell® 150i / 240i
6 Operation6 Operation
64
50115191 A / 08.2009
6.2 Starting operation
1. Fill the water tray with a sufficient quantity of processed water. Do not exceed the upper level mark.
2. Make sure that the CO
2/O2/N2
supply system valves are open.
3. Turn the device on using the power switch.
4. Set the set values for temperature and CO
2
/ O2 content on the iCan™
touchscreen.
5. Start contra-con.
Starting the device:
Starting with auto-start 6. Start the device with the auto-start routine (see chap. 7.9.1‚ pg. 75).
7. The progress indicator auto-start appears on the display, and the automatic start routine is run.
8. The temperature control adjusts the temperature to the selected set value, humidity rises.
9. When temperature and relative humidity are constant, the automatic adjustment of the CO
2
/ O2 measuring system is performed
10. The CO
2
/ O2 control supplies the set amount of CO2 / O2.
11. If the auto-start routine has been completed, the progress indicator is hidden, and the main menu is displayed. The device is operational.
Loading the device:
Loading the work space 12. Load the work space with cultures or place bottles onto the bottle
turning device rollers.
Duration of the auto-start routine:
When the device is cold and when the ambient temperature is low, the auto-start routine may take up to 10 hours.
Loading:
To ensure sufficient air circulation and even heating of the samples, the loading surface within the work space should be used up to 70% max. Voluminous objects in the work space that dissipate heat may impair heat distribution.
Page 65
Operating Instructions | HERAcell® 150i / 240i
65
50115191 A / 08.2009
7 Handling and control (iCan™ touchscreen controller)
7.1 Power switch
Fig. 30: Depending on which side the door hinges are
installed, the power switch [1] is integral to the front cover [2] of one of the front device stands.
• To turn the device on: Press the power switch [1]; the switch illumination comes on. – After a short audible signal and a short dark
phase, the touchscreen display appears.
– The control loop sensors pass the heat-up
phase (see chap. 7.4‚ pg. 68)
• To turn the device off: Press the power switch; the switch illumination goes off
7.2 Operating panel and operating structure
Fig. 31: The control panel works as a touchscreen
(iCan™ touchscreen) and can be controlled performing light pressure on the pressure-sensitive areas of the screen using a finger or a blunt pin:
• Temperature display [1],
•CO
2
display [3],
•O
2
display [5] (optional).
•contra-con key [2]
• menu key [4]
•auto-start key [6],
Extended pressure-sensitive area:
To accept a failure message, the entire touchscreen can be used as a pressure­sensitive area.
.
Fig. 30: Power switch
Fig. 31: Pressure-sensitive areas
Page 66
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
66
50115191 A / 08.2009
Versions without O2/N2 control: Fig. 32: Function keys and value displays of the
operating panel for a device version without O
2/N2
gas
supply:
[1] Type designation of the device [2] Temperature display [3] CO
2
display
[4] Key for starting the contra-con
decontamination routine [5] Key for opening the menu navigation [6] Key for starting the auto-start routine [7] Actual value CO
2
[8] Set value CO
2
[9] Temperature actual value
[10] Temperature set value [11] Current date and time
Version with combined CO
2/O2/N2
control
(optional): Fig. 33: Function keys and value displays of the
operating panel of a device version with combined CO
2/O2/N2
gas supply: [1] Type designation of the device [2] Temperature display [3] CO
2
display
[4] CO
2
set value
[5] CO
2
actual value
[6] Key for starting the contra-con
decontamination routine [7] Key for opening the menu navigation [8] Key for starting the auto-start routine [9] O
2
display
[10] O
2
actual value
[11] O
2
set value [12] Temperature actual value [13] Temperature set value [14] Current date and time
Structure of the operating levels: Fig. 34: Operation is divided into three levels:
A: Direct access to the control loop settings: Temperature, CO
2
, O2 set value and bottle turing
device (optional),
B: Start of the device routines contra-con or auto­start,
C: Navigation through the submenus for device configuration.
Fig. 32: iCan™ touchscreen withoutO2/N2gas supply
Fig. 33: iCan™ touchscreen with combined gas connection
Page 67
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
67
50115191 A / 08.2009
Fig. 34: Menu structure overview
Page 68
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
68
50115191 A / 08.2009
7.3 Factory presettings of the iCan™ touchscreen controls
Upon delivery of the device, the following set values have been preset:
• Temperature: 37 °C
•CO
2
content: 0.0%
•O
2
content (optional): 21.0%
• Optional bottle turningdevice speed a, b, c, and d: 0%.
CO
2/O2
control:
Since the CO
2
concentration of the air is
nearly 0%, the CO
2
control and the control loop error monitoring system are disabled at a set value of 0%. Since the O
2
concentration of the air is
nearly 21%, the O
2
control and the control loop error monitoring system are disabled at a set value of 21%.
7.4 Heat-up phase of the control loop sensors
When the device has been switched on, the control loop sensors pass through a heat-up phase of varying duration during the start process:
Temperature control loop 10 s CO
2
control loop 10 s
CO
2
control loop with IR measuring cell 3 min
O
2
control loop 5 min
The start process is indicated by an audible signal. During the heat-up phase, the displays show dots (...) instead of values:
• Temperature display [2],
•CO
2
display [1] and
•O
2
display [3] After the heat-up phase has been completed, the control loop actual values are indicated.
Fig. 35: Heat-up phase display
Page 69
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
69
50115191 A / 08.2009
Exposure to CO2 gas:
During the 5-minute heat-up phase of the O
2
control loop, the
work space is not exposed to CO
2
gas and the CO2 control and
is disabled.
7.5 Event actions for settings
Key response:
Pressing a key can increase or reduce a value gradually:
• When the – key [4] or the + key [6] is kept depressed, the system switches to quick run,
• after the key has been kept depressed for more than 3 seconds, the quick run speed increases.
Saving the settings:
To save changed values, press the Enter key.
Resetting the settings:
Unless a user action (contact with the pressure-sensitive areas and keys) occurs within 30 seconds, the system automatically exits the menu and the most recently confirmed setting defaults.
Page 70
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
70
50115191 A / 08.2009
7.6 Setting the temperature set value
1. Press the TEMPERATURE DISPLAY key [1].
> The Temperature menu is displayed.
2. To exit the Temperature menu:
> Press the E
ND key [2].
3. To set the temperature set value:
> Press the S
ET VALUE key [3].
To increase the set value: > Press the + key [6].
To reduce the set value: > Press the - key [4].
4. To accept and save the set value:
> Press the E
NTER key [5].
> The system returns to the main menu. The
temperature displays shows the actual value currently measured in the work space.
Fig. 36: Setting the temperature set value
Page 71
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
71
50115191 A / 08.2009
7.7 Setting the CO2 set value
1. Press the CO2 DISPLAY key [1]. > The CO
2
menu is displayed.
2. To exit the CO
2
menu:
> Press the E
ND key [2].
3. To set the CO
2
set value:
> Press the S
ET VALUE key [3].
To increase the set value: > Press the + key [6].
To reduce the set value: > Press the - key [4].
4. To accept and save the set value:
> Press the E
NTER key [5].
> The system returns to the main menu. The CO
2
display shows the actual value currently measured in the work space.
Deactivating the CO
2
control loop:
To deactivate the CO
2
control, the set value is set to 0%. If the control loop is deactivated, error monitoring is enabled as well.
Gas monitoring:
The switching state of the CO
2
control loop does not affect the function of the optional gas monitoring system. The gas monitoring system remains active even when the CO
2
control is switched off. For set values of 21% O
2
, the O2 control is switched off and error monitoring is disabled as well. The control loop error monitoring system (optional) is enabled.
Fig. 37: Setting the CO
2
set value
Page 72
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
72
50115191 A / 08.2009
7.8 Setting the O2 set value
This setting is possible only on versions with the optional O
2/N2
control.
1. Press the O
2
DISPLAY key [1].
> The O
2
menu is displayed.
2. To exit the O
2
menu:
> Press the E
ND key [2].
3. To set the O
2
set value:
> Press the S
ET VALUE key [3].
To increase the set value: > Press the + key [6].
To reduce the set value: > Press the - key [4].
4. To accept and save the set value:
> Press the E
NTER key [5].
> The system returns to the main menu. The O
2
displays shows the actual value currently measured in the work space.
Factory settings:
Depending on the type of the O
2
sensor,
one of the two O
2
control ranges has been preset at the factory: Control range I: 1% to 21% Control range II: 5% to 90%
Usage of the process gases:
For set values below 21% O
2
, the device must be connected to a nitrogen supply system. For set values above 21% O
2
, the device must be connected to an oxygen supply system. Since the O2 concentration of the air is nearly 21%, the O2 control and the control loop error monitoring system are disabled at a set value of 21%. The control loop error monitoring system (optional) is enabled.
Fig. 38: Setting the O
2
set value
Page 73
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
73
50115191 A / 08.2009
7.9 Auto-start routine
The auto-start function is an automated routine for the start and the subsequent adjustment of the CO
2
measuring system. After the start, the device control adjusts the temperature to the set value while humidity is generated. When temperature and relative humidity have reached constant values, the CO
2
measuring system is automatically adjusted to these
values, and the work space is supplied with the preset quantity of CO
2
.
Application of the auto-start routine:
To ensure that the specified accuracy of the CO
2
measuring system is maintained, the device should always be started using the auto-start routine if
• a difference of more than 1 °C is entered upon setting the
temperature set value,
• the Low Humidity function is enabled/disabled,
• the device is restarted after an extended interruption of
operation. The auto-start routine should be run at least every three months on the occasion of cleaning and maintenance works.
Duration of the routine:
Running the routine usually takes 5 to 7 hours. At low room temperatures and when the device is cold, it may take up to 10 hours until the auto-start routine has been completed. If the glass door is opened or if the power supply of the device is interrupted while the routine is running, the routine is interrupted and rerun after the glass door has been closed or the power supply has been reestablished.
Start conditions for the auto-start routine:
Prior to running the auto-start routine, set the CO
2
and O2 set values to the desired values and make sure that the atmosphere in the work space only consists of ambient air. The floorpan of the device must be filled with a sufficient quantity of water.
Conditions that prevent the start of the auto-start routine:
The auto-start routine cannot be started if one of the following failure conditions exists. Temperature control loop:
• Sensor breakage,
• actual value above set value (excessive deviation),
• actual value below set value (excessive deviation),
Page 74
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
74
50115191 A / 08.2009
• set value not plausible,
• calibration values too high or too low,
• sensor communication failure,
• sensor parameter not plausible,
• no communication with LM 75.
Control loop for CO
2
gas supply:
• No communication with sensor. In this case, the auto-start key is dimmed and its function is not available.
Faulty cancellation of the auto-start routine:
The auto-start is cancelled if:
• an error is detected in the temperature control loop,
• an error is detected in the CO
2
control loop,
• the water filling level is insufficient.
auto-start-Dry execution of auto-start routine:
If the auto-start routine is to be run dry (without water in the floorpan of the work space), the water level sensor must be disabled prior to its start (see chap. 7.11.3‚ pg. 85).
Page 75
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
75
50115191 A / 08.2009
7.9.1 Activating auto-start Preparations for the start:
1. Make sure that the CO
2
-/O2-/N2 gas supply system
valves are open.
2. Fill the floorpan of the work space with a sufficient quantity of water. The filling level must not exceed the upper filling level mark.
3. Set the set values for temperature, CO
2
and O2 on
the touchscreen.
auto-start routine activation:
1. Press the auto-start key [2]. > The auto-start-instruction menu [1] is
displayed.
2. To exit the auto-start-instruction menu and cancel auto-start:
> Press the E
ND key [4].
3. Activate the auto-start routine:
> Press the S
TART key [3].
4. To air the work space, open both device doors.
5. When the audible alarm sounds after 30 seconds, close both device doors. > The trend status indicator [5] is displayed.
6. During the progress, the status indicator displays the following information: – Temperature, – start time.
Cancellation:
The auto-start routine can be cancelled at any time! Press the Stop key [6].
Automatic restart:
The auto-start routine is restarted automatically if the routine is cancelled due to one of the following events:
• The glass door is opened,
Fig. 39: Activating the auto-start routine
Page 76
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
76
50115191 A / 08.2009
• the outer door is opened on a device with optional gas-tight screen,
• the power supply is interrupted.
7.9.2 Interrupting the auto-start routine If the S
TOP key in the status display is depressed (see
above), the auto-start routine is interrupted and the auto-start stop dialog box is displayed for a safety scan. The routine can now be permanently cancelled or resumed.
1. To resume the auto-start routine:
> Press the B
ACK key [1].
> The system returns to the status display, and
the auto-start routine is resumed.
2. To cancel the auto-start routine:
> Press the E
ND key [2].
> The warning triangle [3] is displayed as a failure
message together with the audible signal.
3. To accept the failure message:
> Press any position on the display. > The E
RROR dialog box with the description of
the error is displayed.
4. To end the auto-start routine:
> Press the E
ND key [4].
> The system returns to the main menu.
7.10 Running the contra-con routine
contra-con is an automatic decontamination routine for disinfecting the device work space. For a detailed description of this device function, (see chap. 9.4‚ pg. 104).
7.11 User configuration
The user configuration settings allow the user interface and the additional device functions to be adjusted to the requirements of everyday operation. The user configuration menu is split into six categories:
• Settings,
•Event logging,
•Options,
• Icon description,
Fig. 40: Interrupting the auto-start routine
Page 77
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
77
50115191 A / 08.2009
•Keypad lock,
• Software versions. To make a user-specific setting in a dialog box, navigate through the submenus listed in the illustrations and open the dialog box.
7.11.1 Settings The input dialogs of the S
ETTINGS category comprise
all settings for the individual configuration of the user display screen and the user interface:
• Keypad lock coding,
• Date/time setting,
• Contrast setting,
• Key tone setting,
• Interface configuration,
• User display screen language setting,
• Reminder interval setting.
Changing the keypad lock code:
The keypad lock prevents the unauthorized changing of the operational settings. Only those keys are locked where values can be entered. The keypad is locked by entering four numbers.
• Upon delivery, the default code is: 0000. This default can be changed into a user-defined code that is then enabled using the
KEYPAD LOCK dialog box
(see chap. 7.11.5‚ pg. 92).
1. To enter the default 0000:
> Press the corresponding numeric keys. > The number combination is displayed hidden in
the input box [1].
2. To delete the input numbers:
> Press the D
ELETE key [2] .
3. To exit the menu:
> Press the B
ACK key [4] .
4. To confirm the input:
> Press the E
NTER key [3].
> The system changes to the C
ODE NEW menu.
Entering the new 4-digit code:
> Press the corresponding numeric keys. > The number combination is displayed in the
input box [1].
Fig. 41: Changing the code
Page 78
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
78
50115191 A / 08.2009
Fig. 41: To set the cursor to the left to overwrite a value:
5. Press the B
ACKSPACE key (<<) [6].
6. To accept and save the input value:
> Press the E
NTER key [5].
> The system returns to the Settings menu.
Changing the user-defined code:
The user-defined code can be changed as often as required using the same procedure:
• Activate the recoding function by entering the valid code,
• enter the new code and confirm it.
Setting date / time:
The input dialog allows date and time to be set to the required time zone.
1. Fig. 42: To enter the date:
> Press the numeric key [2]. > The input numbers are displayed in the input
box [1].
2. To set the cursor to the left to overwrite a value:
> Press the B
ACKSPACE key (<<) [4].
3. To accept and save the input value:
> Press the E
NTER key [3].
4. The system returns to the Date/Time menu.
5. Enter the time accordingly.
6. To accept and save the input value:
> Press the E
NTER key [3].
> The system returns to the Date/Time menu.
Fig. 42: Date / time setting
Page 79
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
79
50115191 A / 08.2009
Setting the contrast:
The input dialog allows the color contrast of the operating panel to be set within the value range of 48% to 80%.
1. To increase the value: > Press the + key [4].
2. To reduce the value: > Press the - key [2].
3. The value change appears in the display [1].
4. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Settings/Setup
menu.
Setting the key tone:
The input dialog allows the loudness of the key tone that sounds whenever a key is depressed to be set. The value range is 0 to 100. The change occurs in increments of 5%.
1. To increase the value: > Press the + key [4].
2. To reduce the value: > Press the - key [2].
3. The value change appears in the display [1].
4. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Settings/Setup
menu.
Fig. 43: Display contrast setting
Fig. 44: Key tone setting
Page 80
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
80
50115191 A / 08.2009
Setting the interface baud rate:
The input dialog allows the setting of the stepping rate for data communication of the:
• RS232 interface,
• USB interface (optional).
The stepping rate of both interfaces can be changed within the defined baud rates (9.600, 19.200, 38.400,
57.600 baud). The standard stepping rate of both interfaces is 57.600 baud.
1. To increase the value: > Press the + key [4].
2. To reduce the value: > Press the - key [2].
3. The value change appears in the display [1].
4. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Interface menu.
5. To activate the new settings:
> Browse back to the main menu. > Wait for approx. 10 seconds and perform a
reset by switching the device off and on using the power switch.
Fig. 45: RS 232 interface baud rate setting
Fig. 46: USB interface baud rate setting
Page 81
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
81
50115191 A / 08.2009
Setting the user display screen language:
The input dialog allows the language of the display screen to be set. Five languages are available:
•German,
• English,
•Spanish,
• French,
•Italian.
1. To browse upward in the selection: > Press the + key [4].
2. To browse downward in the selection: > Press the - key [2].
3. The new language appears in the display [1].
4. To accept and save the selection:
> Press the E
NTER key [3].
> The system returns to the Settings menu.
Fig. 47: Language setting
Page 82
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
82
50115191 A / 08.2009
Setting the reminder intervals:
The reminder intervals are integral components of the alarm and monitoring system of the device control. For the two essential functions contra-con and auto­start as well as for routine service work, the user can set dates that trigger an alarm whenever they occur. The counting begins at 00:00 hrs of the day on which the previously set reminder interval has elapsed.
1. To increase the number of days: > Press the + key [4].
2. To reduce the number of days: > Press the - key [2].
3. The value change appears in the display [1].
4. To deactivate the reminder interval:
> Set the value to O
FF.
> Press the - key [2].
5. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Reminder Interval
menu.
On the due date, the display shows a reminder message [5] for the activated reminder interval.
contra-con: Please run contra-con.
auto-start: Please run auto-start. Is displayed after the contra-con decontamination routine has been run successfully.
Service interval: Request service. The service message can be confirmed. In this case, the Request Service icon is displayed.
After the routines have been run successfully, the reminder messages are hidden.
Factory settings:
contra-con decontamination routine 90 days auto-start routine Off Service interval 365 days
Fig. 48: Reminder interval setting
Page 83
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
83
50115191 A / 08.2009
7.11.2 Event logging
The input dialogs of the category
EVENT LOGGING
comprise all settings for logging and outputting events during the operation of the device:
• Event display,
• time interval (of the logging cycle),
•error table.
Event display:
The event display [1] uses short, single-line entries specifying date and time to report the events that were logged during the operation of the device. The entries are listed in chronological order with the most recent event at the top position. The list can be displayed but not edited. If the event display consists of several pages, the user can browse through the list. The status indicator [2] indicates which page of the total number of pages is currently being displayed.
1. To browse forward in the list:
> Press the C
ONTINUE key [5].
2. To browse backward in the list:
> Press the P
REVIOUS key [3].
3. To exit the display:
> Press the E
ND key [4].
> The system returns to the Event Logger menu.
Changing the logging cycle time:
Due to limited memory resources, the oldest entries are deleted as new entries are logged. The period from which the displayed entries originate depends largely on the selected logging cycle time.
The setting controls the logging cycle in sections of seconds during which the control loop measured values are logged during the operation of the device and displayed by the trend display (see chap. 7.12‚ pg. 94).
Fig. 49: Event display
Fig. 50: Logging cycle setting
Logging cy cle time Minimal displaye d pe riod
10 s 22.5 hours
30 s 2.8 days
60 s 5.6 days
120 s 11.2 days
180 s 16.8 days
Page 84
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
84
50115191 A / 08.2009
Fig. 50: The settings can be made within the value range of 10 seconds to 3600 seconds.
1. To increase the value: > Press the + key [4].
2. To reduce the value: > Press the - key [2].
3. The value change appears in the display [1].
4. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Event Logger menu.
Event logger interval time:
The logging cycle time does not affect the entries of the error table.
Displaying the error table:
The error table lists the errors detected by the device­integral monitoring system in descending chronological order. The most recently detected error is listed at the top position of 22 possible entries. An entry consists of the control loop in which the error occurred, the date, the time, and an error description. The error table can be displayed but not edited. If the event display consists of two pages, the user can browse through the list. The status indicator [2] indicates which page of the two is currently being displayed.
1. To browse forward in the error table:
> Press the C
ONTINUE key [4].
2. To browse backward in the list:
> Press the P
REVIOUS key.
3. To exit the display:
> Press the E
ND key [3].
> The system returns to the Event Logger menu.
Fig. 51: Error table display
Page 85
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
85
50115191 A / 08.2009
Troubleshooting:
For a detailed overview of causes for errors and their correction, please refer to the end of this chapter!
7.11.3 Options The input dialogs of the O
PTIONS category comprise all
settings of the functional device options:
• Alarm relay,
• Low humidity,
• Gas-tight screen,
• Water level sensor,
•Audible alarm,
• Bottle turning device (optional),
•O
2
(optional).
Setting the alarm relay:
The alarm relay is the interface for the connection of the device-integral monitoring system to an external monitoring system. Depending on the required input signal of the external monitoring system, network monitoring can be enabled or disabled. If network monitoring is enabled, a power failure is detected as an error. The relay cannot be switched off (see chap.
5.10‚ pg. 61).
1. To toggle between two states [1]: > Press the + key [4]. or > Press the - key [2].
2. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Options menu.
Fig. 52: Alarm relay setting
Page 86
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
86
50115191 A / 08.2009
Setting low humidity:
If condensation occurs on the culture containers due to high relative humidity, the humidity in the work space can be set to a lower level. At the factory, the device control is preset to high humidity (approx. 93% relative humidity). By enabling the Low Humidity feature, the relative humidity in the work space is lowered from approx. 93% to approx. 90%. The modification requires an extended adaption phase. To effectively prevent condensation on culture containers, it must be used as a permanent setting.
1. To toggle between two states [1]: > Press the + key [4]. or > Press the - key [2].
2. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Options menu.
Low humidity:
The enabling/disabling of the Low humidity function is entered into the event list. If the Low humidity function is enabled, the corresponding icon [5] is displayed in the main menu.
Fig. 53: Low humidity setting
Page 87
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
87
50115191 A / 08.2009
Setting the gas-tight screen:
As the aperture cross-sections are smaller when accessing the samples, devices equipped with the optional gas-tight screen achieve shorter recovery times for the incubation parameters:
• Work space temperature,
•CO
2
concentration,
•O
2
concentration,
• relative humidity.
After conversion of the device, the device control system must be adjusted to the gas-tight screen option.
Malfunction:
The conversion to gas-tight screen causes a change of the control parameters. Failure to set the gas-tight screen function in accordance with the actually installed door version may impair the incubation performance.
1. To toggle between two options [1]: > Press the + key [4]. or > Press the - key [2].
2. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Options menu.
Switching the water level sensor on/off:
For incubation operation with ambient humidity or if the auto-start routine is to be run dry (without water), the water level sensor can be switched off. This prevents alarm messages for the water level sensor by the device-integral monitoring system from being issued.
1. To toggle between two states [1]: > Press the + key [4]. or > Press the - key [2].
2. To accept and save the change:
Fig. 54: Gas tight screen setting
Fig. 55: Water level setting
Page 88
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
88
50115191 A / 08.2009
> Press the ENTER key [3]. > The system returns to the Options menu.
Switching the audible alarm on/off:
If the device-integral monitoring system detects an error:
• an audible alarm sounds
• in addition to the visual error message and the switching of the alarm relay.
The audible alarm can be permanently disabled:
1. To toggle between two states [1]: > Press the + key [4]. or > Press the - key [2].
2. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Options menu.
Setting the bottle turning device speed (option available only on HERAcell
®
240 i):
If the device has been configured at the factory for the bottle turning device function, the main menu displays the bottle turning device icon. The speeds of the drive rollers can be set separately for each level within a value range of 0% to 100% (see chap. 4.10.8‚ pg. 47). The drive roller of a bottle turning insert is activated as soon as a value> 0 for the speed has been entered and saved.
1. To increase the value: > Press the + key [4].
2. To reduce the value: > Press the - key [2].
3. To switch the drive roller of one level off: > Set the value to 0%.
4. The value change appears in the display [1].
5. To accept and save the change:
> Press the E
NTER key [3].
> The system returns to the Bottle Turning Device
menu.
Fig. 56: Alarm relay setting
Fig. 57: Bottle turning device speed setting
Page 89
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
89
50115191 A / 08.2009
Quick access to bottle turning device levels:
To directly access the Bottle turning device submenu, touch the Bottle turning device icon in the main menu.
Switching the O
2
control on and off:
Depending on the requirements to the work process, the O
2
control can be switched on and off. This setting
is possible only on versions with the optional O
2/N2
control.
1. To toggle between two states of the O
2
control: > Press the + key [4]. or > Press the - key [2].
2. The value change appears in the display [1].
3. To accept and save the setting:
> Press the E
NTER key [3].
> The system returns to the Options menu.
O
2
value display:
If the O
2
control has been switched off, the
O
2
display does not show an actual value (-
- -). This procedure provides protection for the O
2
sensor.
If the set value is set to 21%, the O
2
control loop is not monitored. This applies to both O
2
control ranges:
• Control range I: 1% to 21%
• Control range II: 5% to 90% In this case, the O
2
display shows the
actual value.
Airing the work space:
If the device was operated with O
2
or N2 ,
the work space must be aired after the O
2
control has been switched off.
Fig. 58: Switching the O2 control on and of
f
Page 90
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
90
50115191 A / 08.2009
Gas monitoring:
The switching state of the O
2
control loop does not affect the function of the optional gas monitoring system. The gas monitoring system remains active even when the O
2
control is switched off or the O
2
sensor is
disabled.
7.11.4 Icon description
Essential operating states or error messages, e.g. keypad lock [3], low humidity [4] or bottle turning device [5] are displayed as icons in the touchscreen main menu in addition to the entries in the event or in the error table. The Icon Description dialog box [1] explains the meaning of the individual icons.
• To exit the display:
> Press the E
ND key [2].
> The system returns to the Configuration menu.
Function of the individual icons: Bottle turning device:
This function indicator shows that the device has been configured for operation with the optional bottle turning device. The Bottle Turning Device dialog box can also be accessed directly by touching the Bottle Turning Device icon in the main menu.
Low water level:
This error indicator shows that the water level sensor has detected a low water level.
Low humidity:
This function indicator shows that the relative humidity in the work space has been lowered from approx. 93% to approx. 90%.
Keypad lock:
This function indicator shows that the keypad lock has been activated so that currently the settings cannot be changed.
Overtemperature:
This error indicator shows that the device control has activated the overtemperature protection and has switched to backup control.
Fig. 59: Icon description overview
Page 91
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
91
50115191 A / 08.2009
Gas cylinder empty:
Error message, indicating, that the filling level of one or of serveral gas cylinders is too low to ensure regular gas supply. The monitoring function is available only for device versions with the optional gas monitoring system.
Request service:
This indicator shows that the routine service is due. The display of the icon is controlled by the time entry in the R
EMINDER INTERVAL dialog box and it appears
after the reminder message has been confirmed.
Gas cylinder filling level display (optional):
If the device is equipped with the optional gas monitoring system the icons [1] for gas cylinder A and gas cylinder B appear in the corresponding CO
2
display / O
2
display. The icons indicate the gas cylinder filling level (full / empty). The icon with the blue frame indicates the bottle not in use that will be switched over to continue the supply.
• Manual changeover to a full gas cylinder is possible. > Press the cylinder icon with the blue frame.
or
• changeover can occur automatically, if: > the cylinder pressure drops below 0.6 bar.
After a manual or automatic changeover of the gas supply, another changeover is inhibited for 30 seconds. It takes approximately 2 minutes before the filling level of the new cylinder has been determined and can be displayed. The gas monitoring system monitors the filling state of the two connected gas cylinders. If one cylinder is empty:
• no audible alarm sounds but an error message is displayed,
• an entry is written into the event list.
If both bottles are empty:
• an audible alarm sounds and the alarm relay swit­ches,
• an error message is displayed,
• an entry is written into the error table,
• an entry is written into the event list.
Fig. 60: Gas monitoring icons
Page 92
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
92
50115191 A / 08.2009
Cylinder changeover:
A manual or automatic changeover of a gas cylinder is entered into the event display.
7.11.5 Enabling/disabling the keypad lock
This input dialog box allows the keypad lock to be enabled or disabled. At the factory, the keypad lock is preset to the standard code 0000.
1. Enter the 4-digit code using the keypad. The input appears encrypted in the display [1].
2. To delete complete incorrect input:
> Press the D
ELETE key [2].
3. To break off the input:
> Press the B
ACK key[4].
The system returns to the Configuration menu.
4. To confirm the input:
> Press the E
NTER key [3].
> The system returns to the Configuration menu.
Changing an existing code:
The currently valid code can be redifined in the K
EYPAD LOCK CODE dialog box of the
S
ETTINGS/SETUP menu. (see chap. 7.11.3‚
pg. 85).
Resetting the code:
If the keypad lock code is no longer available, the code must be reset to the standard code by the Technical Support of Thermo Fisher Scientific.
Fig. 61: Enabling/disabling the keypad lock
Page 93
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
93
50115191 A / 08.2009
7.11.6 Software versions
This menu shows the unit software versions in the display [1].
• To exit the display:
> Press the E
ND key [2].
> The system returns to the Configuration menu.
Fig. 62: Software versions
Page 94
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
94
50115191 A / 08.2009
7.12 Scaling the trend display
The trend display of the three control loops:
• Temperature,
•CO
2
,
•O
2
,
can be scaled to two different versions.
1. Full screen display
> Press the display, for the value required. > press the S
HOW TREND ICON [1].
2. To display an enlarged section: > Open a rectangular section [2] in the desired
area of the diagram using a finger or a pen. The size of the rectangle is determined by dragging a diagonal from the start (press onto screen at left of upper diagram frame) to the end (release at right of lower diagram frame).
> Press any position within the marked rectangle
area [2]. The section is now enlarged.
> This process can be repeated any number of
times until the section is displayed in the desi­red enlargement or until the maximal enlarge­ment level (max. 30 event logger elements, i.e. 30 min. process at a logging cycle of 60 seconds).
> During zoom mode the trend display can be
scrolled forward and back.
3. To return to the total trend: > Open rectangle over a small diagram section
and press any position outside the marked area.
4. To exit the trend display:
> Press the E
ND key [3].
> The system returns to the main menu.
Logging cycle:
The time interval of the data logging cycle can be redefined in the I
NTERVAL TIME EVENT
L
OGGER dialog box (see chap. 7.11.2‚
pg. 83).
Fig. 63: Trend display scaling
Page 95
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
95
50115191 A / 08.2009
7.13 Error messages
The error detection system is an integral element of the device-internal control system. It monitors the control loops and their sensors. If an error is detected in the system, the alarm relay switches and issues the following signals and messages:
• An audible alarm sounds,
• a flashing warning triangle [1] is displayed in the main menu; the value indicators are no longer updated,
• the detected error is listed in the error table,
• the event is entered into the event display.
7.13.1 Response to an error message event
If the alarm relay was switched due to a user action, the switching state can be reset by accepting the error message (e.g. for the manual cancellation of the contra-con decontamination routine). If the alarm relay was switched due to a techical defect, the switching state remains active until the defect has been corrected (e.g. low water level in work space).
1. To accept the failure message:
> Press any position on the touchscreen. > The Error dialog box [2] apppears and the
detected error is displayed,
> the audible alarm is switched off.
2. To exit the error display:
> Press the E
ND key [3].
> The error message disappears.
7.13.2 Troubleshooting
The error tables indicate source of error, cause of error, and possible corrections. For any communication with the Technical Support of Thermo Fisher Scientific, please have the device data ready.
Fig. 64: Event error messages
Page 96
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
96
50115191 A / 08.2009
Error message overview:
Control
loop
Error message Cause Repair Alarm relay Buzzer
System
Device door open too long
Doors ope n for more than 10 minutes.
Close device doors x x
Failure display
Communication between display and main board faulty
Contact Service when failure is recuring
xx
Failure EEPROM main board
EEPROM on main board is faulty.
Contact Service x x
Failure communication mainboard
Communication between display and main board faulty *1)
Contact Service when failure is recuring
System - Failure data logger
Failure during recording data logger. Device is still working.
Reset device to cancel failure. When failure is recuring contact Service.
x
Failure contra-con Failure in contra-con routine
Reset device to cancel failure. When failure is recuring contact Service.
xx
System - Power down during contra
-
con
Power failure during contra­con routine
Reset device and start contra­con again.
xx
Failure upon auto­start
Failure during auto-start routine
Reset auto-start. When failure is recuring contact Se rv ice.
xx
System - Safety circuit active
Temperature signal plausibility doubtful *2)
Reset device. When failure is listed several times, contact Service.
Failure bottle turning device
Communication between bottle turning device and main board faulty
Contact Service x x
Tempe­rature
Sensor breakage
Detected value is out of limit.
Contact Service x x
Actual value abov e
Act. val. > nom. val. + 1°C *3)
Do not exceed ambient temperature limit
xx
Actual value below
Act. val. < nom. val. - 1°C *4)
Contact Service x x
Calibration values too high / too low
Temperature max. adjustment value exceeded
Contact Service x x
Actual value plausibility doubtful
Temperature signal plausibility doubtful
Contact Service x x
Failure communication\
Measuring cell cannot commuicate with mainboard
Contact Service x x
*1) Error message will be displayed, but not entered in error table. *2) Error message will be displayed, but not entered in error table.
*4) After changing the set value, error time will be set on 159 min., after door opening on 45 min. ( O
2
on 159 min.).
*3) If failure is detected, a special sample protection controlling will be enabled. To indicate the start of this routine, the icon overtemperature is displayed. The routine will will be disabled after a power reset.
Page 97
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
97
50115191 A / 08.2009
Error message overview:
Control
loop
Error message Cause Repair Alarm relay Buzzer
Tempe­ratur
Failure LM75
LM75 sensor cannot not communicate with mainboard
Contact Service x x
Failure EEPROM sensor
NV-RAM sensor faulty Contact Service x x
CO
2
Sensor breakage
Detected value is out of limit.
Reset auto-start. When failure is recuring contact Se rv ice.
xx
Actual value abov e
Act. val. > nom. val. + 1% *4)
Check gas supply. Reduce prepressure to 1 bar max .
xx
Actual value below
Act. val. < nom. val. - 1% *4)
Check gas supply. Change gas cylinder. Raise prepressure to 1 bar max .. Check supply line to device.
xx
Calibration values too high / too low
Max. adjustment value exceeded for CO
2
Contact Service x x
Failure communication
Measuring cell cannot commuicate with mainboard
Contact Service x x
Failure gas cylinder changeover switch
Gas cylinder changeover switch cannot commuicate with mainboard
Contact Service x x
No gas Both CO2 cylinders empty
Replace one or both O2 cylinders.
xx
Gas cylinder A empty
Gas cylinder A is empty. Replace gas cylinder A.
Gas cylinder B empty
Gas cylinder B is empty. Replace gas cylinder B.
Failure EEPROM sensor
NV-RAM sensor error. Contact Se rvice x x
O
2
Sensor breakage
Detected value is out of limit.
Contact Service x x
Actual value abov e
Act. val. > nom. val. + 1% *4)
Check gas supply. Reduce prepressure to 1 bar max .
xx
Actual value below
Act. val. < nom. val. - 1% *4)
Check gas supply. Change gas cylinder. Raise prepressure to 1 bar max .. Check supply line to device.
xx
Failure communication
O
2
sensor cannot
commuicate with mainboard
Contact Service x x
*4) After changing the set value, error time will b e set on 159 min., after door opening on 45 min. ( O2 on 159 min.).
Page 98
Operating Instructions | HERAcell® 150i / 240i
7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)7 Handling and control (iCan™ touchscreen controller)
98
50115191 A / 08.2009
Error message overview:
Control
loop
Error message Cause Repair Alarm relay Buzzer
O
2
Failure gas cylinder changeover switch
Gas cylinder changeover switch cannot commuicate with mainboard
Contact Service x x
No gas Both CO
2
cylinders empty
Replace one or both O
2
cylinders.
xx
Gas cylinder A empty
Gas cylinder A is empty. Replace gas cylinder A.
Gas cylinder B empty
Gas cylinder B is empty. Replace gas cylinder B.
rH No wate r
Water level in the incubator under
Fill in water or if the auto-start routine is to be run dry (without water), the water level sensor can be switched off.
xx
Page 99
Operating Instructions | HERAcell® 150i / 240i
99
50115191 A / 08.2009
8 Shut-down
8.1 Shutting the device down
Contamination hazard!
If the work space surfaces are contaminated, germs can be transferred to the environment of the device. In case of a shut-down, the device must be decontaminated!
1. Remove the culture containers and all accessories from the work space.
2. Pump the water out off the water reservoir (see chap. 9).
3. Fill in 350 ml water and start the contra- con decontamination routine.
4. Turn the device off using the power switch when the contra-con decontamination routine has been completed.
5. Unplug the power connector and protect it against accidental reconnection.
6. Close the CO
2/O2/N2
supply system shut-off valves.
7. Disconnect the gas pressure hoses from the sleeve at the rear of the device.
8. Clean the work space and wipe the device dry.
9. Until the device is shut down, the work space must be continuously ventilated. Leave the glass door and the outer door open and secure them in this state.
Page 100
Operating Instructions | HERAcell® 150i / 240i
100
50115191 A / 08.2009
9 Cleaning and disinfection
9.1 Cleaning
Incompatible cleaners!
Some device components are made of plastic. Solvents can dissolve plastics. Strong acids or caustic solutions can cause to become brittle of the plastic. For cleaning plastic components and surfaces, do not use solvents that contain hydrocarbons, solvents with an alcohol content of more than 10% or strong acids or caustic solutions.
Moisture-sensitive components!
Do not spray cleaning agent onto the touchscreen and the control box at the rear of the device. When wiping the device clean, always make sure that moisture does not enter into these components.
Exterior surfaces Cleaning exterior surfaces:
1. Remove dirt residues and depositions thoroughly using a solution of water and commercial detergent.
2. Wipe the surfaces clean using a clean cloth and clear water.
3. Then, wipe the surfaces dry using a clean cloth.
Cleaning display:
Moisture-sensitive display!
Do not spray or wipe the display with cleaner.
• Clean display using a dry cloth of 100% micro fiber!
Loading...