Thermo Scientific HAAKE RotoVisco 1, HAAKE RheoStress 1 Instruction Manual

Instruction Manual
HAAKE RotoViscoHAAKE RotoVisco
HAAKE RotoViscoHAAKE RotoVisco
HAAKE RotoVisco
®®
®®
®
1 1
1 1
1
HAAKE RheoStress
®®
®®
®
1 1
1 1
1
version 2.3
“Translation of the original instruction manual“
Table of Contents
1
1. Key to Symbols 3...........................
1.1 Symbols used in this manual 3.................
1.2 Symbols used on the unit 3....................
1.3 Information concerning the CE sign 4...........
1.4 WEEE Compliance 5.........................
2. Quality Assurance 6........................
3. Your Contacts at Thermo Fisher Scientific 6..
4. Warranty 7.................................
5. Safety Notes and Warnings 8................
6. Unpacking / Ambient conditions 11...........
6.1 Transportation damage? 11....................
6.2 Contents of Delivery 11........................
6.2.1 Standard Delivery Rheometer 11...........
6.2.2 Sensor Systems 13......................
6.2.3 Accessories for the Temperature Control Units 13
6.2.4 Application software 13...................
6.3 Space Requirements 14.......................
6.4 Mains supply 14..............................
6.5 Cooling air for the measuring head
(HAAKE RotoVisco 1) 14.......................
6.6 Requirements for the air supply 15..............
6.7 Pipes in the building 15........................
6.8 Air compressors 15............................
6.9 Ambient conditions according to EN 61010 15....
7. Unit Description 16..........................
7.1 HAAKE RotoVisco1 17........................
7.2 HAAKE RheoStress1 18.......................
7.3 Temperature control: 19........................
7.4 Main features of the HAAKE rheometers: 20......
8. Installation 21...............................
8.1 Setting up the Rheometer 21...................
8.2 Connecting up 22.............................
8.2.1 PT100 connection 22.....................
8.3 Hose connections 23..........................
8.3.1 Temperature control unit (with liquid): 23....
8.3.2 Air bearing (measuring unit) 24............
9. Functional Elements 25......................
9.1 Temperature control units 25...................
9.2 External filter 26..............................
9.3 Measuring Instrument with no
temperature control unit. 27....................
9.4 Measuring Instrument with
TCL/Z -- temperature control unit. 29.............
9.5 Measuring Instrument with TCL/P / TCE/P / TCP/P / TCP/PE
-- temperature control unit. 31...................
9.6 Measuring unit with TCE/PC-temperature
stabilisation unit and cone heater TC1 33........
9.7 Measuring Instrument with
SHRP -- temperature control unit. 35.............
9.8 Display unit (optional) 37......................
9.9 Menu tree of the display unit 38.................
Table of Contents
2
10. Operating 39................................
10.1 Switching on 39...............................
10.2 Working with the display unit 39.................
10.3 Starting the software 45........................
10.4 The ”Upload Mode” for the Display Unit
(RheoWin 2.6 or higher) 46.....................
10.5 Quick Cut-off 55..............................
11. Temperature control units 56.................
11.1 Temperature control unit TCO 56................
11.2 Temperature control unit TCL/Z 57..............
11.3 Temperature control unit TCL/P 59..............
11.4 Temperature control unit TCE/P 60..............
11.5 Temperature control unit TCP/P and TCP/PE 61..
11.6 Temperature stabilisation unit TCE/P with
cone heater TC1 67..........................
11.7 Cone heater TC1 67...........................
11.7.1 Correct application of the ceramic rotors 67.
11.7.2 Operation 68............................
11.7.3 Compressed air distributor for TC1 69......
11.8 Temperature control unit SHRP 70..............
12. Sensor Systems 78..........................
12.1 Cylinder Sensor Systems 82....................
12.2 Cylinder Sensor Systems Z-DIN 82..............
12.3 Cylinder Sensor System Z 88...................
12.4 Double Gap Cylinder Sensor DG43
according to DIN 53544 93.....................
12.5 Solvent trap for Z10, Z20, Z31,
Z34, Z38, Z41 und DG43 97....................
12.6 Solvent trap for Z43 98........................
12.7 Cone-Plate and Plate-Plate Sensor Systems 99...
12.8 Cone-Plate Sensor Systems 99.................
12.9 Plate-Plate Sensor Systems 115.................
12.10Platter--platter measuring equipment with
profiled surface. 119............................
12.11 High Shear Cylinder Sensor System HS 122.......
12.12SHRP Plate-Platte Sensor System PP 1 mm 126...
12.13Immersion Sensor System FL 129................
13. Optional sensor systems 132..................
14. Hints on Measurement 133....................
14.1 Temperature programs with Series 1 units 133.....
14.2 Range Limits in Oscillation 134..................
14.3 Correction of Dynamic Measurement 135..........
14.4 Determination of the massmoment of inertia 137...
15. Maintenance 138..............................
15.1 Maintenance Instructions HAAKE RheoStress1 139
15.2 Maintenance Instructions HAAKE Rotovisco1 140..
15.3 Filter unit 141..................................
15.4 Flat sieve 142.................................
15.5 Repairs 143...................................
16. Pin Wiring 144................................
17. External Connections 146.....................
18. Technical Specifications 147...................
19. Terms of Rheological Measurements 149.......
!
Key to Symbols
3
1. Key to Symbols
1.1 Symbols used in this manual
Warns the user of possible damage to the unit, draws attention to the risk of injury or contains safety notes and warnings.
Denotes an important remark.
1
Indicates the next operating step to be carried out and
what happens as a result thereof.
Draws attention to the risk of injury.
1.2 Symbols used on the unit
Caution: Read the instruction manual!
Caution: danger of injury your hands
Caution: Unit becomes hot
Connection for cooling air support
Display operationa
Display heating
Information concerning the CE sign / WEEE compliance
4
1.3 Information concerning the CE sign
Thermo Scientific electrical equipment for measurement, control and laboratory use bears the CE marking.
The CE marking attests the compliance of the product with the EC-Directives which are necessary to apply and con­firms that the apparatus meets all relevant essential requi­rements of the directive, the defined relevant protection requirements.
The conformity assessment procedures were performed following a defined methodology according to each appli­cable directive.
The council decision 93/465/EEC shall be authoritative concerning the modules of the various phases of the con­formity assessment procedures and the rules for the affi­xing and use of the CE marking, which are intended to be used in the technical harmonization directives.
To confirm compliance with the EC-Directive 2004/108/EC Electromagnetic Compatibility (EMC) our product was te­sted according to the EMC requirements for emission and immunity for electrical equipment for measurement, con­trol and laboratory use.
Compliance with the protection requirements areas (do­mestic establishments and establishments directly con­nected to a low voltage power supply network which sup­plies buildings used for domestic purposes) and industrial areas is ensured.
Our strict standards regarding operating quality and resul­ting considerable amount of time and money spent on de­velopment and testing reflect our commitment to guaran­tee the high level of quality of our products even under ex­treme electromagnetic conditions.
Practice however also shows that even electrical equip­ment which bears the CE marking such as monitors or analytical instruments can be affected if their manufactu­res accept an interference (e.g. the flickering of a monitor) as the minimum operating quality under electromagnetic compatibility phenomena. For this reason we recommend you to observe a minimum distance of approx. 1 m from such equipment.
Information concerning the CE sign / WEEE compliance
5
1.4 WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following sym­bol:
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Fisher Scientific compliance with these Directives, the recyclers in your country, and information on Thermo Fisher Scienti­fic products which may assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/WEEERoHS
Quality Assurance/Contacts at Thermo Fisher Scientific
6
2. Quality Assurance
Dear customer, Thermo Fisher Scientific implements a Quality Manage­ment System certified according to ISO 9001:2008. This guarantees the presence of organizational structures which are necessary to ensure that our products are deve­loped, manufactured and managed according to our cu­stomers expectations. Internal and external audits are car­ried out on a regular basis to ensure that our QMS is fully functional. We also check our products during the manufacturing pro­cess to certify that they are produced according to the specifications as well as to monitor correct functioning and to confirm that they are safe. The results are recorded for future reference. The “Final Test” label on the product is a sign that this unit has fulfilled all requirements at the time of final manufacturing. Please inform us if, despite our precautionary measures, you should find any product defects. You can thus help us to avoid such faults in future.
3. Your Contacts at Thermo Fisher Scientific
Please get in contact with us or the authorized agent who supplied you with the unit if you have any further questions.
International / Germany
Thermo Fisher Scientific
Dieselstraße 4 D-76227 Karlsruhe, Germany Tel. +49(0)721 4094--444 Fax +49(0)721 4094--300
support.mc.de@thermofisher.com www.thermoscientific.com/mc
The following specifications should be given when product enquiries are made:
Unit name printed on the front of the unit and specified on the name plate.
Typ: Order No. (e.g.: 557--3001) Ser.:Nr.:
_ ___________
_ Manufacturing order no.: ( 1--9)
__ Manufacturing year ( e.g. 09)
______
Production order no.: (000001 --99999)
___ Serial--number
Mains voltage in V / power input:
e.g. 115 V/ 50 Hz/ 2 A
Warranty
7
4. Warranty
For the warranty and any potential additional warranty, the user shall have to ensure that the devices are serviced by an expert at the following intervals:
The maintenance is required after approx. 2000 operating hours, at the latest, however, twelve months after the in­itial operation or the last maintenance, respectively.
Two thousand operating hours are achieved:
at an operating period of eight hours daily (five days a week) about once a year
at an operating period of more than eight to sixteen hours daily about every six months
at an operating period of more than sixteen hours daily about every three months
We recommend to have the maintenance carried out by Thermo Fisher Scientific or by staff authorised by Thermo Fisher Scientific as special knowledge and tools are requi­red.
The maintenance and calibration work carried out has to be recorded by certificates in conformity with ISO 9000 ff.
! !
!
!
!
!
! !
!
!
!
!
Safety Notes and Warnings
8
5. Safety Notes and Warnings
The Rheometer corresponds to the relevant safety regula­tions. However you are solely responsible for the correct handling and proper usage of the instrument.
This instrument exclusively determines the rheological behav­ior of fluid and half-solid materials. These materials may not be tested if people can be hurt or devices be damaged.
Do not lift or move the unit at the ends of the glass pane.
Do not measure / temperature control any materials that may give off dangerous vapours or may be inflammable within the working temperature range of the rheometer.
The device may not be operated if there are any doubts regarding a safe operation due to the outer appearance (e.g. damages).
A safe operation of the instrument cannot be guaran­teed if the user does not comply with this instruction manual.
Ensure that this instruction manual is made readily available to every operator.
Do not bend connection and/or mains cable, do not subject cables to stress or high temperatures (higher than 70 C).
Check cables visually at regular intervals.
The operator must have an uninterrupted view of the machine and its surroundings.
The rheometer must be fully visible from the PC con­trol stand.
This unit should only be used for the applications it was designed for.
The rheometer is designed for use with a rotor. All exist­ing safety devices are based on the correct installation of the rotor. Is the lift (of the measuring table) used with­out a rotor installed injuries may occur when reaching in the lift area.
Make sure that the unit has been switched off before you connect or disconnect the cables. This is to avoid elec­trostatic charging resulting in a defect of the electronic circuit boards.
!
! !
!
!
!
!
!
!
!
Safety Notes and Warnings
9
Once the probe attains appropriate viscosity,cylindrical tube (Order no. 222--1394) protectors that prevent probe’s radial exit must be used when measuring. An operational error when using the measuring device can lead to the probe’s radial exit from the measuring gap; it is therefore recommended to wear protective goggles.
Do not operate the unit with wet or oily hands.
Do not immerse the unit in water or expose it to spray water.
Do not clean the unit using solvents (fire danger!) -- a damp cloth applied with a household cleaning sub­stance is often sufficient.
Complete separation from the mains is required when repairs or maintenance work is about to be carried out.
Suitable personal protective gear, consisting of lab coat, protective eyewear and safety gloves, must be worn at all times when working with the instrument
Repairs, alterations or any work involving opening up the unit should only be carried out by specialist person­nel. Considerable damage can be caused by incorrect repair work. The Thermo Fisher Scientific service depart­ment is at your disposal for any repairs you may require.
Have the unit serviced by specialists at regular inter­vals.
RheoStress1: The pressure of the air supply for the air bearing must not exceed 4 bar. Higher pressure will damage the air bearing permanently!
For applications at elevated temperatures above 250C it is a must to switch the fan for the air bearing cooling on stage 2 (on the rear) and use ceramic shafts only. Furthermore use the cone heater TC1 as thermal shield to prevent the heating up of the air bearing housing!
Severe skin burns can be caused by contact with hot unit parts!
The rheometer can reach temperatures up to 350 C, depen­ding on the temperature control unit installed. This can result in parts of the rheometer heating up to such an extent, even when taking the cooling and insulation into account, that se­rious burns can result if they come into contact with the skin.
Thermo Fisher Scientific recommends shielding the rheome­ter when operating at high temperatures and handling it only with high temperature proof gloves and safety glasses.
A danger symbol on the glass plate warns the user of possi­ble danger (burn injury).
!
Safety Notes and Warnings
10
We do not know which substances you intend to test us­ing this unit. Many substances are
inflammable, easily ignited, explosive
hazardous to health
environmentally unsafe
i.e.: dangerous
You alone are responsible for your handling of these substances!
Our advice:
If in doubt, consult a safety specialist
Read the product manufacturer ’s or supplier’s
“EU SAFETY DATA SHEET”
Read the REGULATIONS CONCERNING DANGER-
OUS MATERIALS
Observe the “Guidelines for Laboratories”
!
!
! !
Unpacking / Ambient conditions
11
6. Unpacking / Ambient conditions
6.1 Transportation damage?
Notify carrier (forwarding merchant, railroad,
post office) etc,
Compile a damage report.
Before return delivery:
Inform dealer or manufacturer
(Small problems can often be dealt with on the spot).
Do not lift or move the unit at t he ends of the glass pane.
Use the transport handles provided for the instru­ment when unpacking.
The instrument must be carried by two persons.
Unpacking the instrument and putting it into ope­ration is part of the installation and is carried out by trained personnel from Thermo Scientific.
6.2 Contents of Delivery
6.2.1 Standard Delivery Rheometer
The Rheometer is delivered in a recyclable package with the following accessories:
HAAKE RotoVisco 1
Part.No. TCO TCL/Z TCL/P TCE/P
TCE/PC
TCP/P
Connection cable conntry specific
1 1 1 1 1
Instruction Manual ger. 003--5212 1 1 1 1 1 Instruction Manual uk. 003--5213 1 1 1 1 1 RS232 cable to PC, 9 pole 222--1490 1 1 1 1 1 Protection ring 222--1394 1 1 1 Splashboard 003--5172 1 1 1 Lever for cylinders 222--1639 1 Fuses 230V/T1.6A 087--0532 2 2 2 Fuses 230V/T3,15A 087--0533 2 2 Fuses 100/115V/T3.15A 087--0533 2 2 2 Fuses 100/115V/T5A 087--3353 2 2 Leveling indicator 002--4696 1 1 1 1
Unpacking / Ambient conditions
12
HAAKE RheoStress 1
Part.No. TCO SHRP TCL/Z TCL/P TCE/P
TCE/PC
TCP/P
Connection cable conntry specific
1 1 1 1 1 1
Instruction Manual ger. 003--5212 1 1 1 1 1 1 Instruction Manual uk. 003--5213 1 1 1 1 1 1 RS232 cable to PC, 9 pole 222--1490 1 1 1 1 1 1 Compressed air hose 10m 082--2451 1 1 1 1 1 1 Protection ring 222--1394 1 1 1 Splashboard 003--5172 1 1 1 Ring for Peltier tempera-
ture control unit
222--1720 1
Lever for cylinders 222--1639 1 Fuses 230V/T1.6A 087--0532 2 2 2 2 Fuses 230V/T3,15A 087--0533 2 2 Fuses 100/115V/T3.15A 087--0533 2 2 2 2 Fuses 100/115V/T5A 087--3353 2 2 Leveling indicator 002--4696 1 1 1 1 1
!
Unpacking / Ambient conditions
13
6.2.2 Sensor Systems
Various sensor systems are available for the Rheome­ter which differ also in their temperature control specifi­cations.
6.2.3 Accessories for the Temperature Control Units
The temp. control units for liquid temp. control may be operated with different hoses:
For the temperature control units, the open-bath circu­lators and heating circulators the necessary tubing hoses are not part of the standard accessories. They have to be ordered separately.
Temperature range up to 150C:
Viton-hoses with quick coupling 222--0610 Hose nozzle set 222--1492
Temperature range of 100 up to 350C:
Metal hoses (150 cm each) 333-0294 Hose nozzle set 222--1492
Secure all hose connections using hose clamps!
6.2.4 Application software
HAAKE RheoWin Rheometer software for Windows XP
Software RheoWin
for HAAKE RotoVisco1 098--5003 for HAAKE RheoStress1 098--5004
!
!
230 V
115 V
230V
115V
6
!
Unpacking / Ambient conditions
14
6.3 Space Requirements
Good working conditions for a complete installation require an area of about 1 x 0.6 meters. The two benches should be rigid with a level surface and easy to clean. The circulator used for temperature control should be located below the rheometer on a separate bench to avoid backflow of thermal fluid and possible mechanical oscillation.
A suitable laboratory bench must be provided for the installation.
The circulator has to be located below the level of the rheometer, otherwise the thermal fluid flows back in the cup holder and flushes the instrument.
6.4 Mains supply
Only attach the units to mains sockets with a grounded earth. Compare the local mains voltage with the specifications written on the name plates of the measuring instrument and the control unit. Voltage deviations of 10 % are permissible.
Mains cable and unit fuses:
1
Use the mains cable according to your local mains volt­age ( see chapter “Contents of delivery”).
2
Insert the fuses according to your local mains voltage below the mains connection 6.
3
For TCO, SHRP, TCL/Z, TCL/P: 230 V 2 x T1.6 A 100/115 V 2 x T3.15 A
For TCE/P, TCE/P, TCE/PC, TCP/P: 230 V 2 x T3.15 A 100/115 V 2 x T5A
4
Insertion of the fuses: Pull out the fuse holder from the mains socket and
insert the fuses. Reinsert according to the marked vol­tage.
6.5 Cooling air for the measuring head (HAAKE RotoVisco 1)
Compressed air is connected at nozzle 9 to avoid overheat-
ing under extreme load (high torque, high temperature (200--350 C)) and at use of the measuring instrument at temperatures > 100 C in the continuous operation respec­tively. maximum conduction pressure 0.5 bar. (RheoStress are cooled by the outgoing air of the air bea­ring.)
1.0 m
0.85
Rheometer
Circulator
!
Unpacking / Ambient conditions
15
6.6 Requirements for the air supply HAAKE RheoStress RS1 / RW1F
Pure air pressure is connected at nozzle for the air bearing
with a pressure by 2.5 bar (ideal way). The air supply must have the following conditions:
-- max. 4 bar,
-- consumption approx. 10 dm3/minute (1 dm3= 1 thousandth of a cubic meter),
-- no synthetic oil in the line,
-- max. oil contents 3mg/m3air,
-- dry air with a residual moisture < 40 %
Use the filter unit (order no. 222-1211) or the HAAKE compressor (order no. 222-1434 for 230 V and 222-1435 for 115 V).
HAAKE RheoStress1: The pressure of the air supply for the air bearing must not exceed 4 bar. Higher pressure will dam­age the air bearing permanently!
The air bearing reacts highly sensitive to dirt like particles (dust an lint) or liquids (oil or water which condenses at high humidity lev­els). It is therefore recommended to have new air compressors run for a longer period of time (0,5 hrs.) without actually connecting it to the Rheometer. In the case of internal supply systems polluted air might have collected near the connection nozzle when the line has not been used for some time. Therefore, we also recommend to ”flush” this line for a period of approx. 0.5 hrs.
6.7 Pipes in the building
Oil is generally applied to air pipes in the buildings of material­processing plants in order to prevent rusting in the workshops. By means of complex filters the oil content may be reduced by rarely so much that the air bearing does not suffer damage. Thermo Fisher Scientific air bearings must not be used in plants in which oil is added.
6.8 Air compressors
There are ”oil-free” and ”lubricated” versions of air compressors. Only oil-free compressors (e.g. the HAAKE compressor) are ad­missible for Thermo Fisher Scientific air bearings.
If the hose length is larger then 5 m (between compressor and measurement unit), it must be separately attached pressure con­troller located next to the measurement unit (see instruction manual Air compressor Carat).
6.9 Ambient conditions according to EN 61010
It is recommended to run tests in an air-conditioned room, (T = approx. 23C):
indoors, max. 2000 meters above sea level, ambient temperature 15 ... 40C, relative humidity max. 80%/31C( 50%/40C) excess voltage category II, contamination level 2
Unit Description
16
7. Unit Description
lntroduction
The viscoelastic behavior of a fluid or a soft solid can be char­acterized in two ways; either the fluid is deformed and the re­sulting stress is measured (CR mode), or the stress is ap­plied and the deformation monitored (CS mode). The princi­ple benefits of the CS mode are:
Software
Software control and evaluation with HAAKE RheoWinor OS1 to program the display unit for production laboratories.
Measurements and evaluation
The Rheometers is controlled either by the HAAKE Rheo­WinSoftware or with one of 10 pre--programmable mea­suring procedures by the display unit. The results are avail­able for further processing on a PC with the HAAKE Rheo­WinSoftware. Alternatively up to 50 measurements can be stored internally and be evaluated, displayed or printed out for documentation purposes.
Sensors
For HAAKE RotoVisco1 and HAAKE RheoStress1 plates, cones, coaxial cylinders and immersion systems are avail­able.
Unit Description
17
7.1 HAAKE RotoVisco1
Thixotropy and flow curve
The determination of the flow behavior of a test substance requires a speed ramp up, a holding time and a return curve for the determination of the thixotropy. This hysteresis method can like all other measuring procedures be PC con­trolled or can be loaded as procedure in the display and con­trol unit.
Time curve
Is the viscosity of a test substance changing with time this can be monitored objectively with the HAAKE RotoVisco1. This is especially important for substances which are curing (adhesives, building materials), change their viscosity with storage time or show considerable changes of viscosity after shearing during the production process (cosmetics, paint, food).
Temperature programs
Temperature programs are very time consuming when the rheometer has to be operated and supervised. With the Ro­toVisco1 these tasks can be taken over by the PC which con­trols t he rheometer as well as the temperature control units
-- everything without intervention of the operator.
Determination of the yield point
As soon as substances are strained beyond the Hookean range they start to flow. This can be defined as yield point and is an important characteristic value in quality control. With the HAAKE RotoVisco1 this value can be established relevant to practical applications, fast and easily by applying a small deformation through extremely low speed values.
CD Mode
In the CD mode (speeds lower than 0.1 min
–1
) it is only mea­sured at constant speeds. When ramps or steps are preset, the following speed table is run:
Speed ( min
–1
)
0.0125
0.0250
0.0375
0.0500
0.0630
0.0750
0.0875
0.1000
Unit Description
18
7.2 HAAKE RheoStress1
Flow curves
in CR and CS mode can be recorded as ramp or steps (steady state).
Time curves
for reactions (e.g. curing) at constant temperature, shear rate, shear stress or frequency.
Temperature programs
are software controlled to determine the temperature depen­dence with controlled shear stress, shear rate or frequency.
Yield points
can be determined with creep/recovery tests or with stress controlled ramps.
Viscoelasticity
of a fluid can be quantified by a creep/recovery test or by dy­namic measurements in CD (Controlled Deformation) or CS (Controlled Stress) mode.
Multiwave
even allows the determination of the frequency spectra as a function of time or temperature by overlay within shortest time.
!
Unit Description
19
7.3 Temperature control:
For applications at elevated temperatures above 250C it is a must to switch the fan for the air bearing cooling on stage 2 (on the rear) and use ceramic shafts only. Furthermore use the cone heater TC1 as thermal shield to prevent the heating up of the air bearing housing!
TCO Glass plate with a Pt100 temperature sensor
(not temperature controlled).
TCL/Z Liquid temperature controlled receptacle for cylinder
measuring systems with direct contact of the thermal fluidfrom--40upto200C.
TCL/P Liquid temperature controlled measuring plate for
parallel plate or cone & plate measurements (with external circulator from --20 up to 300C).
TCP/P Measuring plate with Peltier temperature control for TCP/PE parallel plate or cone & plate measurements from
--40 up to 180C (external heat transfer required).
TCE/P Electrically temperature controlled measuring plate for
parallel plate or cone & plate measurements from
-- 2 0 Cupto350C.
TCE/PC Electrically temperature controlled measuring plate for
parallel plate or cone & plate measurements with electrical cone heating from --40Cupto350C.
SHRP Special temperature controlled system utilizing
thermal liquid as standardized for asphalts and bitumen. Temperature range: 0C(--25C)... +90C (for HAAKE RheoStress 1)
Unit Description
20
7.4 Main features of the HAAKE rheometers:
Quick fitting sensor with a high precision even if not
perfectly clean.
Remote and manually controlled lift with variable
speed to preserve the fluid’s structure and ensure reproducible test conditions.
Microprocessor controlled positioning of the sensor to
ensure the highest accuracy for routine tests.
Standard temperature range of --40 to 250 C using a
heating circulator; utilizing electrical cone heating extends this up to 350 C.
Controlled rate mode for characterizing rheologically
complicated fluids (HAAKE RotoVisco 1 and HAAKE RheoStress 1).
Controlled stress mode for characterizing sensitive
substances (HAAKE RheoStress1).
Forced oscillation tests at very low frequencies and
very low strains allow destruction free measurements (HAAKE RheoStress 1).
Sophisticated application software packages with
userfriendly window menus and an on-line help key.
Installation
21
8. Installation
8.1 Setting up the Rheometer
Lift the Rheometer out of the package and place onto a stable, level table. Do not lift the unit at the glass pane or at the measuring head ! For sensible measurements a plane table is recommended.
In the base of the measuring instrument, there are four feet which can be screwed in or out for levelling the unit.Upon completion of the preliminary visual levelling, exact precision levelling can be carried out using the spirit level supplied.
1
The spirit level:
is put on the measuring plate at temperature control
unitTCL/P;TCE/P;TCE/PCandTCP/P.
is put into connection with a vessel at temperature con-
trol unit TCL/Z.
is put on the measuring plate in the vessel at tempera-
ture control unit SHRP.
is not used at temperature control unit TCO.
2
Adjust the feet so that the air bubbles remain in the cen­ter of the spirit level.
This adjustment process should be repeated at least once a week.
!
Installation
22
8.2 Connecting up
Make sure that the unit has been switched off be­fore you connect or disconnect the cables.
The cable connections between the sensor system, the con­trol unit, the PC and the printer have to be established ( see. chapter Pin Wiring)
Measuring Instrument
PC
Air
supply
unit
Heating
Bath
and
Circulator
Control
unit
Printer
and/or
Temperature Control Units
RS 232
Centronics
Display
Luft / Air
PT100 ( output )
PT100 ( input )
Shielded RS232 cable with ferrite core
e.g. filter unit 222-1211 or compressor 222-1435
(RheoStress 1)
222-1490
222-0572
8.2.1 PT100 connection
PT100 (input) socket for external PT100 sensor only
when using TCO
PT100 (output) socket for PT100 sensor only when using
TCL/Z and TCL/P
The connections are described in chapter “Pin Wiring”.
!
!
Installation
23
8.3 Hose connections
8.3.1 Temperature control unit (with liquid):
The temperature control unit with liquid temperature
control is connected with hoses to a heating bath and circulator.
While fastening the hose connections at the con­necting nozzles (14) of the temperature control unit the connecting nozzles must be held up by wrenches.
Direction of flow of the liquid
Temperature range up to 150C:
Viton hoses
Hoes nozzle -Set 222--1492
Temperature range of 100Cupto350 C:
Metal hoses 333-0294
Hoes nozzle -Set 222--1492
Liquid temperature control:
Temperature range up to 100C:
Thermal liquid in the temperature range between --50 C and 30C: water with anti-freeze. Temperature range from 5Cto90 C: distilled water.
Distilled water
Normal mains water causes cable deposits and frequently requires the unit to be descaled.
In principle, water up to 95C can be used. However, above 80C, so much water evaporates that it requires frequent topping--up.
Water with anti--freeze
If you intend working at temperatures below 5C, anti--freeze must be added to the water. The added amount of anti--freeze should be sufficient for a temperature that is about 10C lower than the intended working temperature. This prevents the freezing of water on the evaporator coil of the cooling circuit, the surface of which is always much colder than the
Installation
24
working temperature. However, too much anti--freeze worsens the temperature constancy due to its high viscosity.
Temperature range from 100Cto200 C:
Silicone oils or other suitable liquids are used as ther­mal liquid.
8.3.2 Air bearing (measuring unit)
The air bearings of the measuring units HAAKE
RheoStress1 require cooling air at port 9 (see “Requirements for the air supply”).
Functional Elements
25
9. Functional Elements
9.1 Temperature control units
Rheometer Temperature control units
TCO TCL/Z TCL/P TCE/P
TCE/PC
TCP/P SHRP
HAAKE RotoVisco1 X X X X X HAAKE RheoStress1 X X X X X X
Functional Elements
26
9.2 External filter
An external filter is part of the standard range of items supplied with the temperature control units for Series 1 TCP/P (Peltier-Temperature control system), TCE/P (electrical temperature control unit), TCL/PO (liquid temperature control unit for HAAKE RheoScope) and the UTCP/P (Peltier-Temperature control system for HAAKE RSXXX units). It consists of the filter itself, a plastic cone contained inside it, and clamps. Using this filter enables the interior of the relevant temperature control unit to be protected against any impurities carried in by the temperature liquid. The external filter represents an additional protection. On the standard model a sieve is built in to the intake of each temperature control unit.
As shown in Fig. X, the filter is installed in the intake hose leading into the temperature control unit between the thermostat and the measurement instrument. The intake and outlet directions must be noted (IN and OUT on the filter). It is fastened with the clamps that are supplied with it. Hoses with a diameter of 8 mm can be installed and fastened with the clamps supplied. The plastic cone on the inside of the filter serves as a direction of flow indicator, and its movements show the through-flow of the tempering liquid. If there are impurities in the filter it can be taken out, cleaned, and used again. The use of the external filter is recommended because any impurities can be recognised and removed more readily than with the sieve that is built in to the temperature control unit. The screw-in sieve (internal) must of course be checked for impurities from time to time in order to ensure t he functional capability of the temperature control unit.
Order numbers: 003-5266 Screw-in sieve (internal) for TCL/P, TCP/P,
TCE/P, TCL/Z
222-1667 External filter for TCP/P, TCE/P, TCL/PO
Functional Elements
27
9.3 Measuring Instrument with no
temperature control unit.
Front
1
4
3
1. Measuring unit
3. Quick cut-off switch
4. Green LED display: operational
Functional Elements
28
Rear
6
12
13
11
7 8
10
A
9
A. Caution: Read the instruction manual
6. Mains switch with mains socket and fuses
7. RS 232 interface (PC)
8. Connection for printer
9. Connection for cooling air (RheoStress1 for the air bearing)
10. PT 100 Connection
11. Connection for display unit
12. Reset switch
13. Switch for bootstrap loader
Loading...
+ 125 hidden pages