Visit us online to register your warranty
www.thermoscientific.com/labwarranty
Important installer and user information:
A redundant temperature sensing device has been included in this ULT freezer. This
device is a type “T” thermocouple. For convenient access, the thermocouple (Figure 1-
3) terminates in an interconnect jack (Figure 1-5) behind the base front cover. (May be
located differently in chests. See Section 1.) It is strongly recommended that this thermocouple be attached to a redundant 24 hour 7 day monitoring system with alarm
capabilities. Connecting the sensor to a monitoring and alarm system separate from the
freezer provides the utmost in product safety, should the integral system fail.
4 40544 12/9/15 BUS information (preventive maintenance and BUS control panel and operation) ccs
Thermo Scientific
8600 Seriesi
Preface
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
CAUTION
Important Read this operating manual. Failure to read, understand and follow the instructions in this manual
may result in damage to the unit, injury to operating personnel, and poor equipment performance.
s
Caution All internal adjustments and maintenance must be performed by qualified service personnel. s
Material in this manual is for information purposes only. The contents and the product it describes are subject
to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related
to the use of this manual.
The -40C and -86C freezers (see page i for model numbers) described in this manual are high performance
units which can be used for research and in situations that directly support medical applications. When these
products are used to support a medical application, it is an accessory to a medical device and is therefor
considered as a medical device in its own right by the regulatory body ( e.g. FDA).
This product is intended for use:
• as cold storage in research use
• as a medical device for diagnostic use (storage of samples not intended to be re-introduced to the human
body).
Registration: This medical application is considered a Class I device by the FDA. This product is classified as
product code - JRM, regulation number 862.2050 and is considered a Class I device, 510(K) exempt.
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this
symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Extreme temperature hazards, hot or cold. Use special handling equipment or wear special, protective
clothing.
WEEE Compliance: Thermo Fisher Scientific has contracted with companies for recycling/disposal in
each EU Member State. For further information, send an email to weee.recycle@thermofisher.com.
4Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4 Always dissipate extreme cold or heat and wear protective clothing.
4 Always follow good hygiene practices.
4 Each individual is responsible for his or her own safety.
8600 SeriesiiiThermo Scientific
Preface
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763Direct
1-800-438-4851Toll Free, U.S. and Canada
1-877-213-8051FAX
http://www.thermofisher.comInternet Worldwide Web Home Page
service.led.marietta@thermofisher.comTech Support Email Address
Certified Service Web Pagewww.unitylabservices.com
Ourstaff can provide information on pricing and give you quotations. We canSales Support
take your order and provide delivery information on major equipment items or make
arrangements to have your local sales representative contact you. Our products are listed on the
Internet and we can be contacted through our Internet home page.
Ourstaff can supply technical information about proper setup, operation orService Support
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or
provide you with on-site service. We can also provide you with a quotation on our Extended
Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your
applications. If you are experiencing technical problems, working together, we will help you
locate the problem and, chances are, correct it yourself...over the telephone without a service
call.
When more extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If your service need is
covered by the warranty, we will arrange for the unit to be repaired at our expense and to your
satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you
Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. If you wish to write, our mailing address is:
Figures 1-1 and 1-2 show the front view of the freezer and indicate the
following freezer components:
• Control Panel - keypad, displays and indicators.
• BUS (Optional Back Up System) panel.
• Optional temperature recorder (7 day, one pen) or datalogger.
• Keylock - keyed lid lock.
Figure 1-1. Front View 3 cu ft Models
Figure 1-2. Front View Remaining Models
8600 Series1-1Thermo Scientific
Section 1
Optional BUS battery
(just behind this panel)
Optional
BUS Connections
RS-232 or RS-485
Interface
Remote Alarm Contacts
and Analog Output
Power Switch
(mains disconnect)
Power Inlet
Threaded holes for
wall bumpers
Access Port
Threaded holes for
wall bumpers
Remote Alarm Contacts
and Analog Output
Optional
BUS Connections
Power Inlet
Power Switch
(mains disconnect)
RS-232 or RS-485
Interface
Installation and Start-up
Figures 1-3 and 1-4 display the rear view of the freezer and indicate the
following freezer components:
• Remote alarm contacts and selectable analog output connection: 01V, 4-20mA (default), 0-5V.
• Power inlet for power cord connection.
• Optional BUS connections for probe and solenoid.
• RS-232 or RS-485 interface.
• Power Switch (mains disconnect).
1-28600 SeriesThermo Scientific
Figure 1-3. Rear View 3 cu ft Models
Figure 1-4. Rear View Remaining Models
Access Port
Section 1
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
aaaaaaa
To remove
filter
Thermocouple
interconnect jack
O
Battery
Switch
Condenser
Installation and Start-up
The probe cover houses the control, optional recorder, datalogger, 1535
alarm, or BUS probes.
Figure 1-5. Probe Cover
Figures 1-6, 1-7, and 1-8 indicate the following components:
• Freezer filter location
• Battery power switch (freezer and BUS)
• Thermocouple receptacle
• Battery mounting bracket
• Freezer and optional BUS battery
Figure 1-6. All Models Except 3 cu ft
8600 Series1-3Thermo Scientific
Section 1
micro
board
battery
bracket
battery
switch
probe
access
port
O
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
aaaaaaa
a
O
System
battery
Battery
switch
Thermocouple
interconnect jack
To
remove filter
Installation and Start-up
Figure 1-7. Freezer Left Side - Sidecar panel removed
1-48600 SeriesThermo Scientific
Figure 1-8. 3 cu ft Models
Section 1
SYSTEM OK
Mode
Temp Display
Visual Alarm
Indicator
Silence
Low Battery
Indicator
Hot Condenser
Indicator
Message
Center
Scroll for Parameters
Arrows
Up and Down
Arrows
Enter
Installation and Start-up
Control Panel Keys,
Display, Indicators
Mode Select Switch - Used to select Run, Settings, Calibrate and System
Configuration Modes.
Mode Select Indicators -
Run: Run Menu
Settings: Set Points Menu
Calibrate: Calibrate Menu
Configuration: Configuration Menu
Temperature Display - Displays temperature in degrees Celsius.
Alarm Indicator - Light pulses on/off during an alarm condition.
Silence - Mutes the audible alarm.
Low Battery - indicates a low battery condition of the freezer battery.
Hot Condenser - indicates a hot condenser condition.
Message Center - displays system status and alarms.
Scroll for Parameters Arrows - moves through the choices of the selected
mode.
Up and Down Arrows - Increases or decreases values, toggles between
choices.
Enter - Stores the value into computer memory.
Figure 1-9. Control Panel Keys, Display and Indicators
8600 Series1-5Thermo Scientific
Section 1
Installation and Start-up
Panel Keys, Display,
Indicators (cont.)
8600 Series freezers have four basic modes which allow freezer setup: Run,
Settings, Calibrate and Configuration.
Run is the default mode which the freezer will normally be in during
operation.
Settings is used to enter system set points for freezer operation.
Calibrate is used to calibrate various system parameters.
Configuration allows for custom setup of various options.
The chart below shows the selections under each of the modes.
RunSettingsCalibrateConfiguration
Default Mode
System Ok
Line Voltage High Alarm Set Point
Control Set PointControl ProbeHigh Alarm Test
Optional Sample
Probe
Low Alarm Test
Compensated VoltageLow Alarm Set PointSystem Battery Test
* HSHX Temperature
Optional Back Up
System Set Point
BUS Battery Test
Display Temperature
* Clear High Stage
Alarm
Set Access Code
RS485 Address
BUS type CO2 or LN2
Cold Excursion
Warm Excursion
Reset Excursion
Table 1-1. Modes of Operation
* -86C units only
1-68600 SeriesThermo Scientific
Section 1
Installation and Start-up
Panel Keys, Display,
Indicators (cont.)
Displays
Scroll for Parameters Arrows: Steps the operator through the parameters
of Settings, Calibrate and Configuration Modes. The right arrow goes
to the next parameter, the left arrow returns to the previous parameter.
Up Arrow: Increases or toggles the parameter value that has been selected
in the Settings, Calibrate, and Configuration Modes.
Enter: Must press Enter key to save to memory all changed values.
Down Arrow: Decreases or toggles the parameter values that have been
selected in the Settings, Calibrate, and Configuration Modes.
Silence Key: Press to mute the audible alarm. See Section 4 for alarm
ringback times.
Message Center: Displays the system status (Mode) at all times. Displays
SYSTEM OK during normal operation, or alarm messages if the system
detects an alarm condition. See Section 4 - Alarms.
There are two displays on the control panel. The temperature display
shows the temperature in degrees Celsius. The message center displays the
system status (Mode) at all times. The message SYSTEM OK displays
during normal operation. Alarm messages are displayed if the system
detects an alarm condition. See Section 4 - Alarms.
Install Freezer
To remove the freezer from the pallet, use a 7/16" wrench to remove all
the bolts securing the shipping bracket to the pallet.
Note If tipped more than 45°, allow the unit to set upright for 24 hours
before start up.
Remove the shipping bracket. Remove the ramp boards from the pallet
and place the slotted end over the ramp brackets on the pallet. The
support blocks on the ramps will be facing down. Before moving the
freezer, make sure the casters are unlocked and moving freely. Align the
caster with the ramp boards. Use adequate personnel to roll the freezer off
the pallet.
The freezer can be easily pushed to the desired approved location, as
described previously. When the freezer is in position, set the front caster
brakes.
Note Do not move the freezer with the product load inside.
s
s
8600 Series1-7Thermo Scientific
Section 1
Installation and Start-up
Choose Location
Install Wall Bumpers
Locate the freezer on a firm, level surface in an area with an ambient
temperature between 18°C and 32°C. Provide ample room to reach the
mains disconnect switch (power switch) located on the rear of the freezer.
Note For proper ventilation and airflow, a minimum clearance of 5” at the
rear and front and a clearance of 8” on the side of the freezer is required.
Allow adequate space for lid opening. If ambient increases above 36°C,
clearance at the rear of the cabinet must be increased to 8”.
s
The parts bag, located inside the cabinet, contains the following parts.
QuantityStock #DescriptionPurpose
2510016 1/4-20 x 5-1/2” Bolt Wall Bumper
2380520 Neoprene CapCap Protector
Install the bolts into the pre-tapped holes on the back of the compressor
section. Install a neoprene cap on each bolt. Refer to Figure 1-2 for the
locations of the pre-tapped holes.
RS-232 Communications
8600 Series freezers have a data communications interface. The factory
default setting is RS-232.
The wiring identification for the interface is
shown in Figure 1-10. One nine pin, sub
"D" style connector is located on the back
of the freezer. See Figure 1-2 for the
location of the connector on the freezer.
Figure 1-10. RS-232 Interface
The freezer transmits temperature information every 60 minutes. A
standard DB9 serial extension cable can be used to connect the freezer to a
serial device. Some serial devices may require a null modem adapter.
In the event of a CNTRLFAIL, Er07, or the control probe is out of range
error, the numerical temperature data (XXX) in the transmission would be
replaced by T_ERR.
If no alarm condition exists, spaces will be sent. A total of 20 characters
will be sent.
SP - Space LF - Line feed
CR - Carriage return EOT - End of text (4)
NUL - Null character (00)
If an alarm condition does exist, “Error Message” in the protocol will be
replaced by the following:
UNDERTEMP (temperature above the low alarm setpoint)
OVERTEMP (temperature below the high alarm setpoint)
Note The RS-232 is not compatible with 1535 Monitor/Alarm System.
s
* -86C units only
8600 Series freezers have remote
alarm contacts and analog output.
See Figures 1-3 and 1-4 for the
location of the remote alarm
contacts. The remote alarm
connector is located in the parts bag
provided with the manual. It must
be installed if connecting the freezer
to an alarm system. After installing
Figure 1-11. Remote Alarm Contacts
the wiring from the alarm system to
the connector, install the connector to the freezer microboard and secure
with the two screws provided. The remote alarm provides a NO (normally
open) output, a NC (normally closed) output and COM (common). The
contacts will trip on a power outage, high temperature alarm or low
temperature alarm. They will also trip on high stage (-86C units only),
control probe and microboard failures. Figure 1-11 shows the remote
contacts in alarm state.
CAUTION! Stored product should be protected
by a redundant 24 hour/day monitoring system
with alarm capability. An interconnect jack and
thermocouple are installed for centralized
monitoring, should on-board system fail.
Installation and Start-up
Remote Alarm Contacts
and Analog Output (cont.)
Attach Power Cord
The analog output function allows the
freezer to output signals representing
the temperature of the freezer cabinet.
The factory default setting is 4-20 mA.
Refer to Table 1-2 for output
specifications.
Table 1-2. Analog Output Specifications
4-20 mA0-1V0-5V
Temperature-100 to +50°C-100 to +50°C-100 to +50°C
Insert the power cord into the power inlet module. Place the retaining
bracket (P/N 195763) over the connector. Tighten retaining screws to
secure.
Connect Unit to Electrical
Power
1-108600 SeriesThermo Scientific
Figure 1-12. Power Cord Attachment
Note See the serial tag on the side of the unit for electrical specifications or
refer to the electrical schematics in this manual.
s
The freezer should be operated on a dedicated, grounded service. Check
the voltage rating on the serial tag of the unit and compare it with the
outlet voltage. Then, with the power switch turned off, plug the line cord
into the wall outlet.
First, turn on the freezer power switch. Then open the lower front door by
grasping the bottom left corner. Locate the battery switch (Figure 1-4) and
turn it to Standby mode ( ). During initial freezer start-up, the system
battery may require charging and the Low Battery message may appear in
the message center.
Note Ensure the battery switch is turned to Standby mode ( ). The
rechargeable batteries require 36 hours to charge at initial start-up. A “Low
Battery” alarm may occur until the batteries are fully charged. Should a
power failure occur during the initial start-up period, the electronics will
have limited operation.
s
Section 1
Installation and Start-up
Freezer Start-Up
Set Operating
Temperature
With the freezer properly installed and connected to power, system set
points can be entered. The following set points can be entered in Settings
mode: Control temperature, high temperature alarm set point, low
temperature alarm set point, and (optional) BUS set point. Default
settings are shown in the table below. See Chart 1-1 for more detail.
-86C Units
Control Set Point-80°C
High Temperature Alarm-70°C
Low temperature alarm-90°C
Optional BUS Set Point-60°C
-40C Units
Control Set Point-40°C
High Temperature Alarm-30°C
Low temperature alarm-50°C
Optional BUS Set Point-30°C
Note If the set point is changed and the low temperature and high
temperature alarms are set 10° from the set point, the alarm set points will
be adjusted automatically to maintain a distance of at least 10° from set
point.
s
The freezer has an operating temperature range of -50°C to -86°C (-86C
units) or -10°C to -40°C (-40C units), depending on ambient
temperature. The freezer is shipped from the factory with a temperature
set point of -80°C (-86C units) or -40° (-40C units). To change the
operating temperature set point:
1. Press the Mode key until the Settings indicator lights.
2. Press the right arrow until “SET PT = -XX” is displayed in the message
center.
3. Press the up/down arrow key until the desired temperature set point is
displayed.
4. Press Enter to save the set point.
5. Press the Mode key until the Run indicator lights for Run mode or
press the right/left arrow keys to go to next/previous parameter.
If no control keys are pressed, the freezer will automatically return to RUN
mode after 5 minutes.
8600 Series1-11Thermo Scientific
Section 1
Installation and Start-up
Set High Temperature
Alarm
The high temperature alarm activates an audible/visual warning when the
freezer chamber temp reaches or exceeds the high temp alarm set point.
To set the high temperature alarm set point:
1. Press the Mode key until the Set indicator lights.
2. Press right arrow until “HI ALM = -XX” displays in message center.
3. Press the up or down arrow key until the desired high temperature
alarm set point is displayed.
4. Press Enter to save the setting.
5. Press the Mode key until the Run indicator lights or press the right or
left arrow to go to the next or previous parameter.
If no control keys are pressed, the freezer will automatically return to RUN
mode after 5 minutes.
Note The high alarm set point must be set at least 5°C from the control
set point.
s
Set Low Temperature
Alarm
Note At initial start-up, the high temperature alarm is disabled until the
cabinet reaches set point or 12 hours elapse.
s
The low temperature alarm activates an audible/visual warning when the
freezer chamber temp reaches or decreases below low temp alarm set point.
To set the low temperature alarm set point:
1. Press the Mode key until the Settings indicator lights.
2. Press right arrow until “LO ALM = -XX” displays in message center.
3. Press the up or down arrow key until the desired low temperature
alarm set point is displayed.
4. Press Enter to save the setting.
5. Press the Mode key until the Run indicator lights or press the right or
left arrow to go to the next or previous parameter.
If no control keys are pressed, the freezer will automatically return to RUN
mode after 5 minutes.
1-128600 SeriesThermo Scientific
Note The low alarm set point must be set at least 5°C from the control set
point.
s
Section 1
Installation and Start-up
Access Code
Run Mode
An access code can be set to prevent unauthorized change of settings in
Calibrate, Configuration and Settings mode. (An access code of 000 is
required to make changes.) If the access code is not at the default 000, you
can not leave RUN mode without entering a code. See Section 3,
Configuration for instructions on modifying the access code.
Run mode is the default mode for the freezer. The run mode displays the
cabinet temperature on the temperature display and ‘SYSTEM OK’ on the
message center under normal operating conditions. In addition, the Run
mode allows display of the following information:
LINE VOLTAGE
COMPENSATED VOLTAGE
* HSHX TEMPERATURE (heat exchanger temperature); -86C units
only
This information is scrolled individually by pressing the right arrow key.
In each case, the message center returns to SYSTEM OK in 10 seconds if
no keys are pressed.
8600 Series1-13Thermo Scientific
Section 1
Press MODE
to light SETTINGS
indicator
M
ode
Mode
SET PT = -XX
Numbers increase
Press MODE to move
to CALIBRATE
mode
Numbers decrease
Press Enter
to save setting
S
croll fo r Paramete rs
Scroll f or Paramet ers
Mode
HI ALM = -XX
Scroll f or Paramet ers
Press to return
to previous parameter
Press to return
to previous parameter
Press to return
to previous parameter
Low Alarm
Set Point
High Alarm
Set Point
Control
Set Point
To Set:
Settings Mode
Numbers increase
Numbers decrease
Press Enter
to save setting
Numbers increase
Numbers decrease
Press Enter
to save setting
Mode
Press to return
to previous parameter
Optional
Back Up System
Set Point
Numbers increase
Numbers decrease
Press Enter
to save setting
SYSTEM OK
LO ALM = -XX
BACKUP = -XX
Installation and Start-up
1-148600 SeriesThermo Scientific
Section 2 Calibrate
Calibrate Mode
Calibrate Control Probe
Once the freezer has stabilized, the control probe may need to be
calibrated. Calibration frequency is dependent on use, ambient conditions
and accuracy required. A good laboratory practice would require at least an
annual calibration check. On new installations, all parameters should be
checked after the stabilization period.
Note Before making any calibration or adjustments to the unit, it is
imperative that all reference instruments be properly calibrated.
Plug a type T thermocouple reader into the receptacle located inside the
lower door (see Figures 1-4 and 1-6). Compare the control temperature
set point to the temperature of the measuring device. See Chart 2-1 at the
end of this section for more detail.
1. Press the Mode key until the Calibrate indicator lights.
2. Press the right arrow until “CONT T = -XX.X” appears in the message
center.
3. Press up/down arrow to match the display to calibrated instrument.
s
Temperature Stabilization Periods
4. Press Enter to store calibration.
5. Press the Mode key to return to Run or the right/left arrow to go to
next/previous parameter.
Startup - Allow 12 hours for the temperature in the cabinet to stabilize
before proceeding.
Already Operating - Allow at least 2 hours after the display reaches set
point for temperature to stabilize before proceeding.
Note During calibration, the temperature display will not be available.
If no keys are pressed for approximately five minutes while in calibration
mode, the system will reset to Run mode.
8600 Series2-1Thermo Scientific
Section 2
Press MODE
to light CALIBRATE
indicator
Mode
CONT T = -XX.X
Numbers increase
Press MODE to move
to CONFIGURATION
mode
Numbers decrease
Press Enter
to save setting
Scroll for Parameters
Mode
SAMPT = -XX.X
Scroll for Parameters
Pressto return
to previous parameter
Pressto return
to previous parameter
Optional
Sample Probe
Control
Probe
To Calibrate:
Calibrate Mode
Numbers increase
Numbers decrease
Press Enter
to save setting
Chart 2-1
SYSTEM OK
Calibrate
2-28600 SeriesThermo Scientific
Section 3Configuration
Configuration Mode is used for testing and custom setup of the freezer.
The configuration functions listed and described below may not be
necessary in all applications, but are available if needed. See Chart 3-1 for
more detail.
High Alarm Test
Low Alarm Test
The high alarm test is used to verify that the high alarm will activate,
should the freezer temperature equal or exceed the high alarm set point.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until HI ALRM TEST is displayed in the
message center.
3. Press Enter to initiate the test.
The temperature on the display will begin to increase until the high alarm
set point has been reached. The audible alarm will sound and the alarm
indicator will flash. Press the Silence key to mute the alarm.
The low alarm test is used to verify the that low alarm will activate, should
the freezer temperature equal or become less than the low alarm set point.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until LO ALRM TEST is displayed in the
message center.
3. Press Enter to initiate the test.
The temperature on the display will begin to decrease until the low alarm
set point has been reached. The audible alarm will sound and the alarm
indicator will flash. Press the Silence key to mute the alarm.
8600 Series3-1Thermo Scientific
Section 3
Configuration
System Battery Test
BUS Battery Test
To test the charge of the freezer battery:
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until SYS BAT TEST is displayed in the message
center.
3. Press Enter to initiate the test.
TESTING BATT will display during the testing period. Upon completion
of the test the message center will display BATT GOOD or BATT FAIL
When a test is failed, the audible alarm will sound, the alarm indicator and
the Low Battery indicator will light. Press the Silence key and the alarm
indicator will go off. The Low Battery light will stay on until a future
battery test is performed and passed.
To test the charge of the BUS battery:
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until BUS BAT TEST is displayed in the message
center.
3. Press Enter to initiate the test.
TESTING BATT will display during the testing period. Upon completion
of the test the message center will display BBAT GOOD or BBAT FAIL If
this test fails, the audible alarm will sound, the alarm indicator and the
Low Battery indicator will light. Press the Silence key. The audible alarm
and alarm indicator will go off. The Low Battery light will stay on. If the
test fails, it is recommended to replace the BUS battery.
3-28600 SeriesThermo Scientific
Section 3
Configuration
Display Temperature
Clear High Stage Alarm
(-86C units only)
This function, only available on freezers with the optional sample probe,
allows the user to select which temperature is displayed in the temperature
display window. The options are CONTROL or SAMPLE.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until DISP CONTROL or DISP SAMPLE is
displayed in the message center.
3. Press up/down arrow to toggle between the two display selections.
4. Press Enter to save.
If control probe is selected, the temperature display will be on
continuously. If sample probe is selected, the temperature display will be
preceded with a letter ‘S’.
Should a high stage alarm have occurred, it may become necessary to clear
the alarm condition after the condition has been corrected.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until CLR HS ALARM is displayed in the
message center.
Set Access Code
3. Press Enter to clear the alarm.
To set the Access Code:
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until “SET ACC CODE” is displayed in the
message center.
3. Press Enter.
4. The message center will display ACC CODE = 000. Press the up or
down arrow key until the desired access code is displayed (000 - 999).
Press the left or right arrow key to select digit 1, 2, 3.
Note The left and right arrow keys are used to move from the first through
the third digits within the access code. s
5. Press Enter to save the setting
6. Press the Mode key until the Run indicator lights.A 3-digit Access
Code can be entered to avoid unauthorized personnel from changing
the set points, calibration, or configuration. A setting of 000 will
bypass the access code. The factory setting is 000.
8600 Series3-3Thermo Scientific
Section 3
Configuration
RS485 Address
Back Up System Type
If the freezer is configured for RS-485 communications, it will need to
have a unique identification address. This address is set through the
Configuration mode.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until RS485ADDR is displayed in the message
center.
3. Press Enter. The message center will display 485 ADDR XX.
4. Press up/down arrow to select the appropriate address for the freezer
(1 - 24).
5. Press Enter to save.
This function, which is only available on freezers with the optional back
up system (BUS), allows the user to select which type of gas is injected into
the freezer chamber. The options are CO
1. Press the Mode key until the Configuration indicator lights.
2 and LN2.
Cold Excursion
Warm Excursion
Reset Excursion
2. Press the right arrow until BUS TYPE CO2 or BUS TYPE LN2 is
displayed in the message center.
3. Press up/down arrow to toggle between the two display selections.
4. Press Enter to save.
This function displays the coldest temperature recorded by the control
probe.
This function displays the warmest temperature recorded by the control
probe.
This function resets the cold and warm excursions.
3-48600 SeriesThermo Scientific
Press MODE
to light
CONFIGURATION
indicator
Mode
Mode
SYS BAT TEST
LOALRM TEST
Press MODE to move
to RUN mode
Press Enter
to initiate
test
Scroll for Parameters
Scroll for Parameters
Mode
HIALRM TEST
Scroll for Parameters
Press to return
to previous parameter
Press to return
to previous parameter
Press to return
to previous parameter
LowAlarm
Test
HighAlarm
Test
System Battery
Test
To Configure:
Mode
Scroll for Parameters
Chart 3-1
Press Enter
to initiate
test
Press Enter
to initiate
test
Press Enter
to initiate
test
Press to return
to previous parameter
Optional
BUS Battery
Test
BUS BAT TEST
Press Enter
to initiate
test
Configuration Mode
SYSTEM OK
Section 3
Configuration
8600 Series3-5Thermo Scientific
Section 3
Press Mode to exit.
Press right or left arrow
to move to next digit.
Numbers increase
Numbers decrease
Mode
CLR HSALARM
Press Enter to
clear the alarm
condition
Scroll for Parameters
Press to return
to previous parameter
refer to previous page
Configuration Mode, Chart 3-1, Page 2 of 3
Clear High
StageAlarm
To Configure:
D
ISPCONTROL
CONTROLprobe
Press Enter to
save the setting
M
ode
Scroll for Parameters
Press to return
to previous parameter
Display
Temper at ur e
SAMPLE probe
M
ode
Scroll for Parameters
M
ode
Press Enter to
save the setting
Scroll for Parameters
Mode
BUS TYPE CO2
Press Enter to
save the setting
Scroll for Parameters
Press to return
to previous parameter
Optional
Back Up
System Type
Mode
RS485ADDR
Scroll through the
various options.
Press Enter to
save the setting
S
croll for Parameters
Press to return
to previous parameter
RS-485
Address
Type CO2
Type LN2
Access
Code
SETACC CODE XXX
Configuration
3-68600 SeriesThermo Scientific
Section 3
Mode
WARM EX = -XX
S
croll for Parameters
Press to return
to previous parameter
refer to previous page
Configuration Mode, Chart 3-1, Page 3 of 3
Warm
Excursion
To Configure:
COLD EX = -XX
Mode
Scroll for Parameters
P
ress to return
to previous parameter
Cold
Excursion
M
ode
S
croll for Parameters
M
ode
Press Enter to
reset the cold
and warm
excursion values
S
croll for Parameters
Reset
Excursions
RST EXCUR
Press to return
to previous parameter
Configuration
8600 Series3-7Thermo Scientific
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.