Thermo Scientific FINNPIPETTE PDP Instructions For Use Manual

1
2 5 .0
Labsystems
GRIPPY
HANDLE
DIGITAL DISPLAY
OPERATING BUTTON
FINNPIPETTE PDP
INSTRUCTIONS FOR USE
Thermo Electron Oy P.O.Box 100, Fin-01621 Vantaa, Finland Tel. +358-9-329 100, fax -358-9-3291 0414 www.thermo.com/finnpipette
2
PRODUCT DESCRIPTION
The FINNPIPETTE PDP is a continuously adjustable, positive displacement pipette specially designed for PCR (Polymerase Chain Reaction) [1] work. It operates on the positive displacement principle (i.e. no air interface) and uses detachable, disposable tips with a pre-assembled plunger. FINNPIPETTE PDP pipettes cover a volume range of 0.5 µl to 200 µl.
RAW MATERIALS
The FINNPIPETTE PDP is made of mechanically durable and chemically resistant materials:
Operating button PA Handle polyamide Tip ejector polyamide Tip cone PVDF
THE DIGITAL DISPLAY (Fig.1)
The adjusted delivery volume is clearly indicated by the digital display on the pipette handle. [1] The Polymerase Chain Reaction process PCR, is covered by US patents issued to the Cetus Corporation.
DESCRIPTION OF FINNTIP PDP (Fig.4)
FINNPIPETTE PDP tips are supplied in racks of 96 or bags of 960. The product code numbers are as follows:
Cat.No. Volume Qty
9403070 0.5-25µl 960/bag 9403080 20-200µl 960/bag
PIPETTE OPERATION
SETTING THE DELIVERY VOLUME (Fig.2)
A) The delivery volume is set using the operating button
on the top of the pipette. To increase the delivery volume turn the operating button counterclockwise. To decrease, turn it clockwise.
B) Make sure that the desired delivery volume clicks into
place and that the digits are completely visible on the handle display.
C) Do not set volumes outside of the pipette´s specified
volume range!
ATTACHING A TIP (Fig.5)
Attach a Finntip PDP from a rack as follows:
1. Push the pipette firmly onto the tip until the tip and cone snap.
2. Without removing the tip from the tray press the operating button to the first stop to engage the plunger.
3. The tip and plunger are now attached for pipetting.
3
TIP EJECTION (Fig.3)
1. To remove the tip and plunger from the pipette simply press the operating button all the way down.
2. Then release the operating button.
3. This action simutaneously ejects the tip and the plunger.
PIPETTING TECHNIQUE
Figure 6
A = Ready Position B = Pipetting Stop C = Tip Ejection
Ensure that clean tips are firmly pushed on to the tip cones of the pipette. The operation of the pipette is controlled by the operating button. For the best possible accuracy:
Operate the button slowly and with steady speed at all times, particularly when working with liquids having high viscosity. Never let the operating button snap back. When working with serum or biological fluids prime the tips with the liquid before pipetting. This is done by filling and emptying the tips.
Hold the pipette vertically (10 degrees maximum from vertical) during the liquid intake. For maximum hand control, hold the pipette tightly in the hand with the "Grippy" resting on the index finger. Fill a clean reagent reservoir with the liquid to be dispensed.
1. Depress the operating button to the pipetting stop.
2. Dip the tip just under the surface of the liquid in the reservoir (2-3 mm) and slowly release the operating button. Withdraw the tip from the liquid touching them against the edge the reservoir to remove excess liquid.
3. Deliver the liquid by gently depressing the operating button to the pipetting stop.
4. Release the operating button to the ready position.
If necessary, change the tips and continue with the pipetting.
4
MAINTENANCE
When the FINNPIPETTE PDP is not in use make sure it is safely stored in a vertical position, preferably on a Finnpipette Stand.
SHORT TERM CHECKING
The pipette should be checked at the beginning of each day for dust and dirt on outside surfaces of the pipette. Particular attention should be paid to the tip cones. Solvents should not be used for cleaning the pipette.
LONG TERM MAINTENANCE
If the pipette is used daily it should be checked at least twice a year. The servicing procedure starts with disassembly of the pipette.
1. To remove tip cone cover: Insert the tooth of the service tool into the opening at the base of the tip cone cover.
2. Pull away the tip cone cover
3. Remove the tip cone by turning it counterclockwise with the service tool.
4. Pull out the piston.
5. Clean the piston and the piston spring with a dry napless cloth.
6. Grease the cleaned parts with lubricant provided in the package.
7. Reassemble the parts.
CALIBRATION
All Finnpipettes are factory calibrated and adjusted to give the volumes as specified with distilled or deionized water. Normally, the pipettes do not need adjustment, but they are constructed to permit recalibration and adjustment for liquids of different temperature and viscosity.
DEVICE REQUIREMENTS AND TEST CONDITIONS
An analytical balance must be used. The scale graduation value of the balance should be chosen according to the selected test volume of the pipette:
Volume range readable graduation
under 10µl 0.001mg 10-100µl 0.01mg above 100µl 0.1mg
Loading...
+ 8 hidden pages