Thermo Scientific 1403 Series, 1404 Series, D14031, D14041 Operating Manual

Thermo Scientific
Barnstead TII Water Purification System Series 1403 and 1404
Operating Manual and Parts List
50128664 1/7/11
These operating instructions are protected by copyright. Rights resulting thereof, particularly reprint, photomechanical or digital postprocessing or reproduction, even in part, are only allowed with the written consent of Thermo Fisher Scientific. This regulation does not apply to reproductions for in-plant use.
Trademarks
All trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers.
Thermo Fisher Scientific Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error­free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
Thermo Scientific TII Water System i
C
Chapter 0 Preface..............................................................................................................................................0-1
Models covered in this manual ...........................................................................................................0-1
Chapter 1 Safety Information..........................................................................................................................1-1
Warnings............................................................................................................................................1-1
Chapter 2 Product Information.......................................................................................................................2-1
TII Water Specifications.....................................................................................................................2-1
Technical Specifications ..................................................................................................................... 2-3
Declaration of Conformity ............................................................................................................ 2-3
Chapter 3 Installation .......................................................................................................................................3-1
Unpacking.......................................................................................................................................... 3-2
Locating Unit ..................................................................................................................................... 3-3
Bench Mounting ................................................................................................................................3-3
Wall mounting ...................................................................................................................................3-3
Push-to-Connect Fitting Tubing Install.............................................................................................. 3-4
Push-To-Connect Tubing Removal....................................................................................................3-4
Tubing Adapter Fittings ..................................................................................................................... 3-5
Quick Disconnect Fittings..................................................................................................................3-5
UV Lamp Installation.........................................................................................................................3-6
Cartridge / Membrane Installation ..................................................................................................... 3-7
MPS Cartridge Installation.................................................................................................................3-7
Carbon Filter Installation ................................................................................................................... 3-8
RO Membrane Installation.................................................................................................................3-8
12 LPH Units (1 Membrane) ......................................................................................................... 3-9
24 LPH Units (2 Membranes)......................................................................................................3-10
Cleaning Preparation........................................................................................................................ 3-11
Feed Water Connections .................................................................................................................. 3-12
Atmospheric Drain...........................................................................................................................3-12
RO Membrane Rinse........................................................................................................................3-13
Reservoir Connection .......................................................................................................................3-14
System Cleaning Procedure ..............................................................................................................3-15
Cartridge Pack Installation ............................................................................................................... 3-17
Cartridge Pack Rinse-Up..................................................................................................................3-18
Control Panel Wall Mount Bracket..................................................................................................3-18
Optional Accessories.........................................................................................................................3-19
Optional Calibration of the Resistivity Meter with a N.I.S.T. Module ........................................ 3-19
Resending the N.I.S.T. module for validation ............................................................................. 3-21
Dispense Overflow Cutoff Float ..................................................................................................3-22
Computer/Printer Setup...................................................................................................................3-22
Contents
Contents
ii TII Water System Thermo Scientific
One Hyperterminal ..................................................................................................................... 3-23
Procomm..................................................................................................................................... 3-23
Connecting and Starting the Printer............................................................................................ 3-24
Chapter 4 Using your TII .................................................................................................................................4-1
Normal Operation..............................................................................................................................4-1
Control Panel .....................................................................................................................................4-1
LED System Lights.............................................................................................................................4-2
Buttons.......................................................................................................................................... 4-2
Idle Mode...........................................................................................................................................4-3
DI Run Mode ....................................................................................................................................4-3
DI Standby Mode............................................................................................................................... 4-4
Automatic Flush (RO Membrane)................................................................................................. 4-4
Manual Flush (RO Membrane) ......................................................................................................4-5
Dispensing Water...............................................................................................................................4-5
Automatic Dispensing ................................................................................................................... 4-5
Timed Dispensing ......................................................................................................................... 4-5
Volumetric Dispensing.................................................................................................................. 4-6
Reservoir Replenishment ....................................................................................................................4-6
Setting the Displayed Languages.........................................................................................................4-6
Setting the Time and Date .................................................................................................................4-6
Setting the Low Product Purity Alarm................................................................................................4-6
Setting the RO % Reject Alarm.......................................................................................................... 4-7
Setting the Feedwater Chlorine Level .................................................................................................4-7
Setting System Options ......................................................................................................................4-7
Auto Standby Timer...................................................................................................................... 4-8
Unit Under Counter (Factory set to “OFF”) ................................................................................. 4-8
Display Contrast............................................................................................................................ 4-8
Set Units........................................................................................................................................ 4-8
Printer Attached ............................................................................................................................ 4-8
Reservoir Type............................................................................................................................... 4-8
Connecting TII to Nanopure System .................................................................................................4-9
Chapter 5 Maintenance and Servicing............................................................................................................5-1
Prefilter Replacement .........................................................................................................................5-1
MPS Replacement .............................................................................................................................. 5-2
Carbon Replacement ..........................................................................................................................5-2
Cartridge Pack Replacement...............................................................................................................5-3
Membrane Replacement..................................................................................................................... 5-4
Micron Final Filter Replacement........................................................................................................5-4
Fuse Replacement...............................................................................................................................5-4
UV Lamp Replacement ...................................................................................................................... 5-5
Resetting Routine Timers...................................................................................................................5-5
Prefilter, MPS, Carbon, Membrane, Pack Timers.......................................................................... 5-5
Reservoir Vent Timers................................................................................................................... 5-5
System UV Timer.......................................................................................................................... 5-5
Reservoir UV Timer ...................................................................................................................... 5-5
Sanitization Timer......................................................................................................................... 5-5
Procedures to Reset Routine Timers................................................................................................... 5-6
Contents
Thermo Scientific TII Water System iii
Troubleshooting5-7
System Alarms and Errors ..................................................................................................................5-8
Routine Care Alerts..........................................................................................................................5-14
Chapter 6 Ordering Information.....................................................................................................................6-1
Ordering Consumables and Supplies..................................................................................................6-1
Basic Consumables.........................................................................................................................6-2
General Maintenance Supplies .......................................................................................................6-2
Safety Stock....................................................................................................................................6-3
TII Accessories ...............................................................................................................................6-4
Chapter 7 Wiring Diagrams.............................................................................................................................7-1
Chapter 8 Maintenance Log.............................................................................................................................8-1
Contents
iv TII Water System Thermo Scientific
Thermo Scientific TII Wa ter System P-1
P
Preface
Models covered in this manual
Material in this manual is for information purposes only. The contents and the product it describes are subject to change without notice. Thermo Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
©2010 Thermo Scientific. All rights reserved.
Model Number Description Voltage
D14031 (7165) 12 LPH 90 V - 240 V
D14041 (7166) 24 LPH 90 V - 240 V
IMPORTANT Read this instruction manual. Failure to read, understand and follow the
instructions in this manual may result in damage to the unit, injury to operating personnel, and poor equipment performance.
CAUTION All internal adjustments and maintenance must be performed by qualified
service personnel.
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this symbol.
Preface
Models covered in this manual
P-2 TII Water System Thermo Scientific
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be damaged by elevated temperatures.
Marking of electrical and electronic equipment, which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market after 13 August 2005.
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Scientific has contracted with one or more recycling/disposal companies in each EU Member State European Country, and this product should be disposed of or recycled through them. Further information on Thermo’s compliance with this directive, the recyclers in your country and information on Thermo products will be available at www.thermofisher.com.
Note
• Always use the proper protective equipment (clothing, gloves, goggles, etc.)
• Always dissipate extreme cold or heat and wear protective clothing.
• Always follow good hygiene practices.
• Each individual is responsible for his or her own safety.
Preface
Models covered in this manual
Thermo Scientific TII Water System P-3
Preface
Models covered in this manual
P-4 TII Water System Thermo Scientific
Thermo Scientific TII Water System 1-1
1
Safety Information
Your Thermo Scientific TII has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. This manual contains important operating and safety information. The user must carefully read and understand the contents of this manual prior to the use of this equipment. For safe operation, pay attention to Notes, Cautions, and Warnings throughout the manual.
Water purification technology employs one or more of the following: chemicals, electrical devices, mercury vapor lamps, steam and heated vessels. Care should be taken when installing, operating or servicing Thermo Scientific products. The specific safety notes pertinent to this product are listed below.
Warnings
To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of correct voltage and current handling capacity.
2. Do not locate the Thermo Scientific TII directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located.
3. Replace fuses with those of the same type and rating.
4. Do not disassemble water lines or remove cartridges where spilled water could contact equipment that requires electrical service. Electrical shock hazard could result.
5. Power unit OFF before plugging in or unplugging unit.
6. Disconnect from the power supply prior to maintenance and servicing.
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components that may ignite such materials.
2. Do not use in the presence of highly corrosive substances such as bleach or acid baths; fire may result.
3. This device is to be used with water feeds only. Cleaning agents must be used in compliance with instructions in this manual. Failure to comply with the above could result in explosion and personal injury.
4. Avoid splashing cleaning solutions on clothing or skin.
5. Ensure all piping connections are tight to avoid chemical leakage.
Safety Information
Warnings
1-2 TII Water System Thermo Scientific
6. Ensure adequate ventilation.
7. Carefully follow manufacturer's safety instructions on labels of chemical containers and material safety data sheets.
8. This unit is equipped with an ultraviolet lamp. Ultraviolet radiation is harmful to the eyes and skin. Do not attempt to observe the lamp directly.
9. Refer servicing to qualified personnel.
Do not use this product for anything other than its intended usage. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Note The UV lamp contains mercury. If broken or no longer needed, do not dispose of
the UV lamp in the trash. Recycle or dispose of the UV lamp as hazardous waste.
Thermo Scientific TII Water System 2-1
2
Product Information
The Thermo Scientific TII is a tap-fed water purification system designed to be simple to use and to provide reagent-grade water that exceeds ASTM Type II, and CLSI®-CLRW standards. It uses a thin film composite reverse osmosis membrane with pretreatment to produce RO water that is then polished using a two-stage deionization process combined with UV oxidation and a 0.2 micron final filter.
The % of the rejection between incoming and reverse osmosis product water is monitored and an indication is provided to the user if the RO water quality is unacceptable. Pressure sensors located inside the unit monitor incoming feed water and RO pressure. The DI water purity is continuously measured by a resistivity cell and displayed on a multifunctional display.
RO operation is automatic and works independently of the DI system to fill the external reservoir as determined by sensors in the reservoir.
Read the instructions carefully to ensure that you receive maximum benefit from the Thermo Scientific TII. Also, please fill out and return the enclosed warranty registration card, as it will help us assure you of proper warranty coverage.
TII Water Specifications
Table 2-1 Feed Water Requirements
Note The Thermo Scientific TII is not shipped with pretreatment and DI polishing
cartridges or a final filter. These must be ordered separately. The start-up kit containing these components can be ordered using part number D502138 for all units.
Types Tap
1
(Potable)
Turbidity 1.0 N.T.U. maximum
Pressure Range 2-6.9 bar (30-100psig)
Temperature Range 4-40°C (39-104°F)
Minimum Inlet Flow Requirements 100 lph
pH 3-10
TDS 600 (max. ppm as CaCO3)
Silt Density Index < 5%
Free Chlorine < 3 ppm
Langlier Saturation Index < 1
Product Information
TII Water Specifications
2-2 TII Water System Thermo Scientific
1
1500 ppm NaCl feed water solution @ 25 °C (77 °F), @ 4.5 bar (65 psig), 15% recovery. Flow rate and recovery will decrease with lower water temperature and pressure. We recommend the use of a hot water mixing valve before the RO system for water cooler than 15
°C (59 °F). Feed water
suitability must be determined by a water analysis.
Table 2-2 RO System
Table 2-3 Membrane Performance
2
2
500 ppm NaCl feed water solution @ 25 °C (77 °F), @ 4.5 bar (65 psig), 15% recovery. Flow rate and recovery will decrease with lower water temperature and pressure. The use of a hot water mixing valve before the RO system is recommended for water cooler than 15
°C (59 °F).
Table 2-4 DI Product Water Quality (dispensed from unit)
Iron (total as Fe) < 0.5 ppm
Silica < 30 ppm
Model RO Product Flow Rate Recovery
D14031 12 lph (3.2gph), ±15% 10-20 %
D14041 24 lph (6.3gph), ±15% 20-40 %
Rejection
Inorganic (minimum) >90 %
Inorganic (typical) >95 %
Particles >99 %
Bacteria >99 %
Organics (>300 MW) >99 %
Resistivity ASTM Type II, >15 Megohm-cm
TOC <15 PPB
Flow Rate Up to 1 LPM with a new D3750 final
filter
Bacteria Less than 0.01 CFU/ml
Volumetric Dispense Accuracy: ±5%, Repeatability: ±3%
Types Tap
1
(Potable)
Product Information
TII Water Specifications
Thermo Scientific TII Water System 2-3
Technical Specifications
Table 2-5 Dimensions and Clearance Requirements
Table 2-6 Environmental Conditions
Declaration of Conformity
We hereby declare under our sole responsibility that this product conforms to the technical requirements of the following standards:
EMC
EN 61000-3-2 Limits for harmonic current emissions EN 61000-3-3 Limits for voltage fluctuations and flicker EN 61326-1 Electrical equipment for measurement, control, and laboratory use - EMC requirements; Part I: General Requirements
Safety
EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use; Part I: General Requirements
Per the provisions of the Electromagnetic Compatibility Directive 2004/108/EC, and per the provisions of the Low Voltage Directive 2006/95/EC.
Copies of the Declaration of Conformity are available upon request.
Thermo Scientific TII Dimensions
13.5" W X 18.5" D X 19.5" H (34.3 cm X 47.0 cm X
49.5 cm)
Clearances Sides - 9 " (22.9 cm) minimum to allow for servicing
Above -3 " (7.6 cm) minimum for removal of top cover
Front - 4.75 " (12.1 cm) minimum for opening front door
Electrical Requirements Thermo Scientific TII is equipped with four power cords
and corresponding fuses taped to each power cord to be plugged into a grounded electrical outlet of the appropri­ate voltage.
All models: 90-240 VAC, +5 % -10 %, 47-63 Hz.
Operating 15°C -30°C; 20% -80% RH, non-condensing.
Installation Category II (over-voltage) in accordance with IEC 664.
Pollution Degree 2 in accordance with IEC 664.
Altitude limit 6562 ft (2,000 meters)
Storage -25 °C -65 °C; 10 % to 85 % relative humidity
Product Information
TII Water Specifications
2-4 TII Water System Thermo Scientific
Thermo Scientific TII Water System 3-1
3
Installation
Figure 3-1 Drain Water Tubing
Figure 3-2 Feed Water Tubing
Figure 3-3 Reservoir Connection Tubing
Figure 3-4 Power Cord with Fuses
Figure 3-5 Tube Removal Tool
Installation
3-2 TII Water System Thermo Scientific
Unpacking
Remove the unit from its shipping container and ensure that the following items are removed before discarding the packaging materials:
1. Thermo Scientific TII unit
2. Drain Water Tubing. Approximately 6 ft. (1.83 m) of 1/4” (0.64 cm) O.D. drain tubing
(TU1190X12) with no fittings (Fig 3-1)
3. Feed Water Tubing. Approximately 10 ft. (3.05 m) of 3/8” (0.95 cm) O.D. feed water tubing
provided with a quick disconnect insert on one end and a 1/4” NPT fitting on other end (TU1119X7) (Fig 3-2)
4. Reservoir Connection Tubing. Two approximately 10 ft. of 3/8” O.D. tubes (TU550X5) for
fluidically connecting the unit to the storage reservoir (Fig 3-3).
5. Model D14041 only: Membrane Reject 1 to Membrane Feed 2 tube
6. Four power cords: one of which is pictured in Figure 3-4 with fuses.
7. UV lamp (LMX13)
8. Tube removal tool (AYX23): shown in Figure 3-5.
9. Wall Bracket - Unit (BC1190X12)
10. Wall Bracket - Remote Display (BC1190X10), shown in Figure 3-6.
11. Display Cable (WH1403X4)
12. Blank Display (DL1190X18), shown in Figure 3-6.
13. RO Membrane (FL1265X1) - Model D14041 has two membranes.
14. Quick Reference Guide (LTX112)
Figure 3-6 Control Panel Accessories:
(Left) Blank Display (DL1190X18), (Right) Control Panel Wall Mount Bracket
Also required for installation (purchased separately):
1. A storage reservoir. If a Thermo Scientific TII storage reservoir (D14061, D14062, D14063, D14064, D14071, D14072, D14073 or D14074) is also purchased, see contents in separate box.
2. Start-up kit (D502138). Contains the Thermo Scientific TII DI Cartridge Pack (with 0.2 μM final filter) (D502137), 1
μM Pre-filter (D502113), MPS Cartridge (D502114), Extruded
Carbon Cartridge (D502115), and Cartridge Block (BK1404X2).
3. Customer-supplied fasteners to mount unit and remote display.
4. Household bleach for the cleaning process.
Installation
Bench Mounting
Thermo Scientific TII Water System 3-3
Locating Unit
Locate storage reservoir approximately 2 - 5 ft. (1.52 m) from the Thermo Scientific TII for electrical connections.
Refer to the Thermo Scientific TII storage reservoir manual or customer-supplied reservoir manual for mounting instructions. The outlet of the Thermo Scientific TII storage reservoir must be above or at the same level as the inlet of the Thermo Scientific TII.
Bench Mounting
Place the Thermo Scientific TII on a bench top that is accessible to electricity and an atmospherically vented drain.
Wall mounting
Wall composition, condition and construction, as well as fastener type, must be considered when mounting this unit. The mounting surface and fasteners selected must be capable of supporting a minimum of 150 lbs. (68kg). Inadequate support and/or fasteners may result in damage to mounting surface and/or equipment. If you are unsure of mounting surface composition, condition and construction or correct fasteners, consult your building maintenance group or contractor. Refer to Section
Technical Specifications, for clearance requirements.
Install the Thermo Scientific TII on a wall in a convenient location that is accessible to an atmospherically vented drain and electricity.
1. Locate the wall bracket packed separately from the unit.
2. Using the wall bracket as a template, locate and drill the mounting holes in the wall. A minimum of four (customer-supplied) fasteners will be required — two on the top and two on the bottom.
3. Attach the wall bracket to the wall using customer-supplied fasteners.
4. Remove the locking screws on each side of the wall bracket.
5. Pull the two locking slides on each side of the wall bracket out as far as they will go.
6. Hang the unit on the wall bracket by sliding the mounting pins into the wall bracket slots.
7. Push the locking slides on each side of the wall bracket in as far as they will go.
8. Replace the locking screws.
WARNING Do not locate the Thermo Scientific TII directly over equipment that
requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located.
Note The outlet of Thermo Scientific TII storage reservoir must be above or at the same
level as the inlet of the Thermo Scientific TII.
Installation
Push-To-Connect Tubing Removal
3-4 TII Water System Thermo Scientific
Push-to-Connect Fitting Tubing Install
The following instructions will apply when you need to attach a piece of tubing to your Thermo Scientific TII during installation, unless otherwise noted in the installation instructions. To make tubing connections:
1. Make sure the tubing is cut off reasonably square and that no plastic burrs or ridges are present.
2. Mark from end of tube an insertion length of 3/4” (1.9 cm).
3. Wet the tube end with water and insert the tube straight into the fitting until it bottoms out on the interior shoulder and the insertion mark is no longer visible.
Figure 3-7 Push-To-Connect Fitting Tubing Installation
Push-To-Connect Tubing Removal
1. Using the tool provided (AYX23), push the collet toward the body while pulling on the tubing to release the tube, as shown in Figure 3-8.
Figure 3-8 Push-To-Connect Tubing Removal
Installation
Quick Disconnect Fittings
Thermo Scientific TII Water System 3-5
Tubing Adapter Fittings
1. Completely disassemble the fitting. Refer to Figure 16 to familiarize yourself with the names of the component parts.
2. Make sure the tubing is cut off reasonably square and that no plastic burrs or ridges are present.
3. Place the grab ring and back-up ring in the hex nut in the order and orientation shown in Figure
16. Thread the nut onto the adapter. DO NOT use the o-ring at this time.
4. Push the tubing through the nut until it bottoms out in the adapter.
5. Remove the adapter nut and tubing. Place the o-ring over the tubing. Be careful not to push the back-up ring or grab ring further back on the tubing when installing the o-ring.
Figure 3-9 Typical Polypropylene Tubing Adapter Installation
Quick Disconnect Fittings
These fittings are found on the water inlet. To insert the inlet tubing:
1. Press on the metal thumb pad on the coupling body to ensure the fitting is open.
2. Wet the o-ring on the coupling insert and push into the coupling body until you hear a click. Gently pull on the tubing to ensure it is secure.
3. To remove, press the metal thumb pad to release and pull coupling insert out.
CAUTION Do not tighten tube fitting hex nut with a wrench. Tight connections can be
easily made by hand.
Installation
UV Lamp Installation
3-6 TII Water System Thermo Scientific
UV Lamp Installation
Locate the UV chamber inside the left door of the Thermo Scientific TII (Figure 3-10). Install the UV Lamp as follows:
Figure 3-10 UV Chamber
1. Disconnect the power cord from the unit.
2. Access the left side of the unit by unlatching the screw securing the door with a flathead screwdriver. Open the door.
3. Remove the UV lamp from its packaging. DO NOT TOUCH THE GLASS PORTION OF THE LAMP!
4. Remove black cap by sliding it off UV chamber.
5. Insert the lamp halfway into the UV lamp chamber and plug it in noting proper rectangular orientation, then fully insert the lamp into the chamber, and replace black cap.
6. Make sure the UV lamp cable is tucked behind the chamber before closing and relatching the door.
7. Reset the UV lamp timer as directed in the “Resetting Routine Timers” section of this manual.
CAUTION Do not touch the lamp with fingers! It is recommended that lintfree gloves
be worn when handling the lamp. The glass portion must be free of fingerprints, perspiration, etc. Even a light coating of perspiration will reduce the effectiveness of the lamp. If the glass portion of the lamp is touched, clean it with a damp, lint-free cloth. Use isopropyl alcohol as required.
WARNIN G The UV lamp contains mercury. Do not discard in trash. Recycle or
dispose of UV lamp as hazardous waste.
Installation
MPS Cartridge Installation
Thermo Scientific TII Water System 3-7
Cartridge / Membrane Installation
Your Thermo Scientific TII has been shipped with one RO membrane for the 12 lph unit, or two RO membranes for the 24 lph unit, which you will install. The Startup Kit (D502138) includes a prefilter, MPS cartridge, carbon filter, and cartridge pack, and must be purchased separately. To prepare your Thermo Scientific TII for operation, disconnect the unit from the power supply if connected.
Figure 3-11 Cartridge Installation: (left to right) prefilter, MPS, carbon
Particulates can damage your membranes, resulting in premature membrane failure. Therefore, your Thermo Scientific TII uses a prefilter to remove particulates from your feed water. Install the prefilter as follows:
1. Open the front left door.
2. Remove the prefilter (Part Number D502113) from its packaging and wet the o-rings with water.
3. Insert the prefilter upright into the left position upper and lower quick disconnects and push until firmly connected. For easier installation, connect bottom first.
MPS Cartridge Installation
The MPS cartridge contains a slow dissolving anti-scalant material. The anti-scalant material combines with water to prevent scale buildup on the membrane surface. Install the MPS cartridge as follows:
1. Remove the MPS cartridge (Part Number D502114) from its packaging and wet the o-rings with water.
2. Insert the MPS cartridge upright into the middle position upper and lower quick disconnects and push until firmly connected. For easier installation, connect bottom first.
Note The cartridges must be installed in the correct order to operate properly.