Thermo Scientific Barnstead GenPure xCAD Plus System Quick Reference Manual

Barnstead GenPure xCAD Plus System
Quick Reference Guide
Thermo Scientific
Quick Reference Guide
© 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
BRWPQGGENXCAD 0416
Final Filter
Dispense Button
UV Lamp
Ultrafilter
Ultrapure Cartridge
GenPure xCAD Plus Consumables Replacement Information
GenPure xCAD Plus System
Consumable Part #
Estimated
Life Time *
How You Will Know
To Change
Ultrapure Cartridge
09.2005
Up to
12 months
• Purity drops below acceptable purity level
• “Lim. val. pure w.”displayed
• “Change cartridge” message displayed – cartridge usage exceeded maximum time in system
0.2 µm Final Filter
09.1003
Up to
12 months
• Dispensing water flow is reduced
• 09.2005 ultrapure cartridge is replaced
Ultrafilter 50133980
When
applicable –
up to 2 years
• Dispensing water flow is reduced
• Replace at least every 2 years
UV Lamp 09.2002
When
applicable –
up to 2 years
• “UV Intensity” alert (units with TOC only)
• “UV Duration” alert is displayed
Cleaning Solution
CMX25 n/a
• Bacteria present in water
• When 09.2005 ultrapure cartridge is replaced
• Requires reusable cleaning cartridge
Note: The actual life time of the consumables depend on the quality of the feed water and the amount of water used by the system.
Please refer to operation manual for complete instructions
Display Functionality Keys and Display Readout
Operation
• Nonstop mode during working hours for highest purity water.
Pump and UV runs continuously.
• Volumetric Dispense: In Nonstop mode only. Press Enter or
Dispense button, display will show last volume dispensed, press button again to dispense water. While dispensing, pressing Enter or Dispense button will stop water flow. Change Volume: Press Enter, then i or h to change volume, UV button to move over to next digit.
• Manual Dispense: Press dispense button 2X to start dispensing.
Press again to stop. (Change volume on display as necessary to dispense larger volumes of water).
• Interval: Standby mode. Unit enters interval upon start up and after
period of non-use. – In interval mode, pump recirculates water for 5 min (adjustable) out of every 30 min. – Nights and Weekends: Keep system power on and allow it to automatically switch to interval mode. – Displayed purity may change when system is not recirculating. – Manually put system in Interval mode by pressing On/Off button twice.
• Rinse: Manually activate rinse if unit has been off or unused for a
long weekend or holiday. – Activate Rinse by pressing menu button until “Rinse” is displayed and press Enter. System will rinse for 30 seconds. – Interval Rinse Time: System will activate a 0.5 second rinse when system switches from Interval to Nonstop mode and when system initiates recirculation in Interval mode.
Common Alerts
Check operation manual for full list of alerts.
• Screen may turn red indicating an alert condition.
• Lim. Val. Pure W.: Displayed when purity has reached ultrapure
water limit value. Replace ultrapure polishing cartridge.
• Limit Value Feed: Displayed when feed water purity limit is out of
range. Check feed water quality.
• Change Cartridge: Cartridge exceeds maximum time in the system.
Disinfect system and replace ultrapure cartridge. Enter serial number for new cartridge using user code.
• UV Duration: UV light has exceeded maximum time in the system.
Replace UV lamp.
• UV Intensity: Displayed in TOC units only, red screen alert if below
50 % intensity. Replace UV lamp.
Maintenance
• Regular maintenance is important for consistent purity. Change ultrapure cartridge and final filter when purity has dropped to limit value or a minimum of once a year.
• Clean system before changing the ultrapure cartridge using resuable, 09.2201 cleaning cartridge.
• Change UV light and/or ultrafilter on a regular set schedule.
Current operating mode: NonstoporInterval
TC is displayed if temperature compensation is turned on. (Typical operation TC turned on)
Real time resistance of ultrapure water. Display can show conductivity when option selected in menu.
Interval button will decrease a numeral value.
Enter confirms and saves selections.
Menu key brings you to the following
system parameters:
• Feed water resistivity and limit value
• Ultrapure water limit value
• UV lamp timer and UV intensity
• Cartridge serial number
• Rinse
• Disinfection
• Error history
• Registering new xCAD station
• Display system faults
• Code – Unlock user settings by code (see manual)
1.00 L – Indicates current volume
dispense setting.
ON/OFF
UV button turns UV light on. (Where
applicable) UV light should be on during working hours. Also moves cursor to the right.
Nonstop button switches unit into Nonstop mode for dispensing. Increases numerical values.
TOC content of the product water in systems equipped with TOC monitoring. If TOC feature is not included it will read “– –”.
Ultrapure water temperature.
Displays UV if the UV lamp is turned on for use.