Thermo Scientific AquaSensors™
AV38 Local Display
and Controller
User Guide
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink,
ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ,
AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product
Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir
Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture
Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY,
Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of
Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent 6,168,707.
PerpHecT ROSS electrodes are protected by U.S. patent 6,168,707.
ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.
ionplus electrodes and Optimum Results solutions are protected by U.S. patent 5,830,338.
ROSS Ultra electrodes are protected by U.S. patent 6,793,787.
ORP standard is protected by U.S. patent 6,350,367.
No Cal electrodes are protected by U.S. patent 7,276,142.
The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Preface
This instruction manual serves to explain the use of the Thermo Scientific AquaSensors AV38 Local Display
and is written to cover as many applications as possible. Please do not hesitate to contact Thermo Fisher
Scientific or an authorized representative with questions or concerns.
The information presented in this instruction manual is subject to change without notice as improvements are
made, and does not represent any commitment whatsoever on the part of Thermo Fisher Scientific.
Thermo Fisher Scientific cannot accept any responsibility for damage or malfunction of the product due to
improper use.
Contact Information
To contact Thermo Scientific AquaSensors Technical Support:
Within the United States call 1.800.225.1480 or fax 978-232-6015.
Outside the United States call 978.232.6000 or fax 978.232.6031.
In Europe, the Middle East and Africa, contact your local authorized dealer.
Visit us on the web at www.thermo.com/processwater
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 1
Safety Information
The Thermo Scientific AquaSensors AV38 Local Display shall be
installed and operated only in the manner specified. Only a skilled,
trained or authorized person should carry out installation, setup and
operation of the sensor system.
Before using the display, make sure that the sensor cable is
connected as specified. Failure to do so may result in permanent
damage to the system of its components.
Protection against electric shock will be achieved only by observance
of the corresponding installation rules.
1.1.GENERAL INFORMATION .................................................................................................................................6
1.2.INTENDED USE .................................................................................................................................................7
2.2.BASE PART NUMBERS......................................................................................................................................9
7.1.SETTING DATASTICK STATION NUMBER ......................................................................................................57
7.2.SETTING DATA TYPE .....................................................................................................................................58
17. TERMS AND CONDITIONS.......................................................................................................... 77
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 4
TABLE OF FIGURES
Figure 1: The elements of the AV38 user interface. .............................................................................................. 18
Figure 2: The "H" indicates that hold mode is active. ............................................................................................ 19
Figure 3: The initial measure screen when the DataStick is absent. .......................................................................20
Figure 4: The initial measure screen when the Sensor Head is absent. .................................................................. 20
Figure 5: The complete calibrate menu for pH sensors. ........................................................................................27
Figure 6: The complete calibrate menu for D.O sensors........................................................................................ 27
Figure 7: The complete calibrate menu for conductivity and turbidity sensors.......................................................... 27
Figure 8: The pH sensor value is dynamically updated during a 1-point buffer calibration......................................... 28
Figure 9: The pH sensor value is dynamically updated during a 2-point buffer calibration......................................... 30
Figure 10: Dynamic display of the conductivity value during calibration. .................................................................35
Figure 11: The conductivity calibration edit screen................................................................................................ 35
Figure 12: Entering the conductivity calibration value............................................................................................ 35
Figure 13: Dynamic display of the pH value during calibration. ..............................................................................37
Figure 14: The pH calibration edit screen............................................................................................................. 37
Figure 15: Entering the pH calibration value......................................................................................................... 38
Figure 16: The complete calibrate menu for conductivity....................................................................................... 39
Figure 17: Zero is Calibrating…. ......................................................................................................................... 40
Figure 18: The user is prompted to place the sensor in air. ...................................................................................41
Figure 19: The temperature calibration value is presented for editing. .................................................................... 43
Figure 20: The complete configure menu for conductivity...................................................................................... 47
Figure 21: The complete configure menu for pH................................................................................................... 47
Figure 22: The complete configure menu for D.O., Ozone and Chlorine. ................................................................47
Figure 23: The complete configure menu for turbidity............................................................................................ 47
Figure 24: The complete DataStick Comms menu................................................................................................ 57
Figure 25 The wash cycle showing when the relay is activated and the outputs are held.......................................... 72
TABLE OF TABLES
Table 1: AV38 base part numbers......................................................................................................................... 9
Table 2: Main menu items. ................................................................................................................................. 21
Thank you for purchasing the Thermo Scientific AquaSensors AV38 Local Display.
The product is designed for continuous use in industrial process applications and complies with safety
regulations currently in force. Improper use could lead to hazards for the user or a third-party, and/or adverse
effects to the plant or other equipment.
Thermo Fisher Scientific does not accept any liability for damage that may arise if information in this
manual is not followed. Therefore, the operating instructions and specifications must be read and
understood by all persons involved in installation and operation of this equipment.
This manual identifies safety instructions and additional information by means of the following symbols:
This symbol draws attention to safety instructions and warnings of potential danger,
which if neglected, could result in injury to persons and/or damage to property.
This symbol identifies additional information and instructions, which if neglected, could
lead to inefficient operation and possible loss of production.
It is recommended that this manual be made accessible to everyone who may need it as a reference.
Please contact Thermo Fisher Scientific or an authorized representative with any questions.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 6
1.2. Intended use
The Thermo Scientific AquaSensors AV38 Local Display is used to monitor, configure, calibrate and diagnose
any DataStick measurement system configured with a Modbus communications adapter in accordance with
the technical product specifications in Section 2 of this manual.
Data is reported through a scalable 4-20 milliamp analog output or through one of several digital protocols.
Options include a second analog output and up to 2 relays. PID control is a standard function when pH or
conductivity measurements are connected. Hypochlorous acid (HOCl) and free chlorine concentrations are
automatically available when a chlorine sensor is connected.
Any other use, or use not mentioned here, that is incompatible with the technical specifications is deemed
inappropriate. The operator is solely responsible for any damage arising from such use.
Other prerequisites for appropriate use include:
• Observing the instructions, notes and requirements set out in this instruction manual.
• Observing all local safety regulations.
• Observing all warnings and cautions in the documentation regarding all products used in this
measurement system, including the sensor, mounting hardware, AV38 electronics and cabling.
• Observing the prescribed environmental and operational conditions.
• Observing chemical compatibility with all wetted materials.
1.3. Safety Instructions
The AV38 Local Display should be installed and operated only by personnel familiar
with the sensor and qualified for such work.
A defective AV38 should be returned to Thermo Fisher Scientific for repair or
replacement. Contact Thermo Fisher Scientific to obtain a Return Material
Authorization (RMA) number.
No modifications to the AV38 are allowed. The manufacturer/supplier accepts no
responsibility for damage caused by unauthorized modifications. The risk is borne
entirely by the user.
1.4. Removal from Service / Correct Disposal of the AV38 Local Display
Removal from Service
• Disconnect the cable wiring from the controller terminal block.
• Remove the AV38 from the mounting hardware.
Correct Disposal of Unit
When the AV38 is taken out of service, observe the local environmental regulations for correct disposal.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 7
2. PRODUCT DESCRIPTION
2.1. System Overview
The AV38 is a universal display interface for DataStick measurement systems. The enclosure has ¼-DIN
dimensions for easy mounting and is rated NEMA 4X for outdoor use. It uses a liquid crystal display (LCD) with
chlorine sensor is connected. The AV38 supports up to two 4-20 analog outputs, two alarm/control/wash relays
and network communications to a host computer.
a high-contrast backlight for best readability and is powered with 24
volts DC. A line-powered version is also available.
Up to 247 DataStick measurement systems can be connected to a
single AV38 on a Modbus RTU communications network. Up to two of
the DataSticks connected can be mapped to “Channel 1” or “Channel
2” for direct local display, calibration, configuration and diagnosis. All
DataSticks can be accessed remotely when the host communications
hardware is present.
The AV38 automatically recognizes the type of DataStick connected to
it and provides the appropriate calibration, configuration and d
iagnostic
menus. A 10-point user-entered concentration table in the DataStick
can be edited and used to convert measured conductivity into %
concentration. PID control is a standard function when pH or
conductivity DataSticks are connected. Hypochlorous acid (HOCl) and
free chlorine concentrations are automatically available when a
The AV38 can operate in two different modes:
1. Analyzer Mode: Map up to two DataSticks to the display, analog outputs and relays. Switch the
display between the two DataSticks using the RIGHT and LEFT arrow keys. Access all DataSticks on
the network for local calibration, configuration or diagnostics using the menu system. When host
communications hardware is present, a host computer can simultaneously be used to acquire data or
remotely calibrate, configure or diagnose the DataStick that is locally selected for display. This mode is
primarily used when analog output, PID and relay functions are needed with a small network of
DataSticks.
2. Gateway Mode: The host computer has access to all measure, configure, calibrate and diagnose
commands for all connected DataSticks. In this mode the AV38 suspends all local functions including
menu access, data display, analog outputs and relays. This mode is primarily used when the host
computer collects digital data and analog outputs and relays are not needed. The AV38 allows local
configuration and calibration of DataSticks when the application requires it. Gateway mode is switched
to Measure Mode with a RIGHT arrow or LEFT arrow key press. It is not possible to enter Gateway
mode using the RIGHT or LEFT arrow keys when the Measurement is ppmFCL2 (free chlorine
concentration). This is because the AV38 needs to continuously acquire data from the chlorine
DataStick and, if necessary, the pH DataStick, in order to perform the free chlorine concentration
calculation.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 8
When there is no DataStick present at the selected station address, the measure screen will indicate that the
DataStick is absent. When the sensor head is removed from the DataStick, the measure screen will indicate
that the sensor head is absent.
Host communication options include, Modbus, DeviceNet, and others. The host protocol is independent of the
Modbus RTU protocol used by the AV38 to communicate with the DataSticks.
There are seven keys for menu navigation. The MENU key is used to toggle between the menu and the
measure screen. Pressing the MENU key provides options for calibration, configuration, communications,
analog outputs and relays.
2.2. Base Part Numbers
Part numbers can be easily decoded to give descriptive information about included functions. The part number
is printed on the AV38 terminal block overlay where detailed information is provided on the host
communications protocols. The base part number and firmware version are displayed at power-up and in the
HELP menu. For example, if the display shows “AV38-CBYC, V4.00”, the AV38 has two analog outputs, two
relays and host communications, and is operating with version 4.00 firmware.
Base part numbers are decoded as follows:
Table 1: AV38 base part numbers.
Base Part
Number
AV38BB0A1 0 Yes None
AV38BB-Y-A
AV38CB0C2 2 Yes None
AV38CB-Y-C
Analog
Output(s)
1 0 Yes (Y=1) RS-232 ASCII
2 2 Yes (Y=1) RS-232 ASCII
Relays PID Host Comms
Indicated by second to last character in base part number
(Y=4) Modbus RTU
(Y=5) DeviceNet
(Y=7) EtherNet/IP, Modbus TCP, PCCC
(Y=4) Modbus RTU
(Y=5) DeviceNet
(Y=7) EtherNet/IP, Modbus TCP, PCCC
Profibus DPV0/V1 Host communications can be obtained using an AV38 with Modbus RTU Host
communications (Y=4) connected to a Profibus-to-DataStick Communications Adapter (CA26R).
2.3. Sensor Mapping
Multiple DataStick sensors can be connected to a single AV38 and any number of them can be accessed
directly via digital host communications. To simplify local operation only two sensors can be mapped to the
display, the analog outputs and the relays at any one time. This is done by selecting the appropriate station
number for DataStick 1 (DS1) and DataStick 2 (DS2).
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 9
2.4. Mounting
The AV38 can be mounted on a wall or panel mounted in ¼-DIN or ½-DIN forms. A gasket is built into the
display for a NEMA 4X seal. Mounting instructions must be followed to maintain seals. The AV38 can be pipe
mounted and panel mounted with optional mounting hardware.
2.4.1. Conduit Box Wall Mount
The AV38 conduit box is mounted to a wall or plate with #8 or M4 screws through the mounting holes as
shown. There are seven Pg11 and Pg16 knockouts in the conduit box. Two cord grips are supplied and it is
recommended that they be used in the conduit holes on the bottom side of the box to prevent rain or moisture
from settling at the conduit interface. It is also possible to mount the box using interior breakouts. Four plastic
caps are provided to seal the interior mounting screws. NEMA 4X ratings will be maintained only when the #8
mounting screws are used in the holes outside the conduit box gasket.
AV38 conduit box is mounted to a
wall or plate with #8 or M4 screws
through the mounting holes as
shown.
Seven Pg11/16 Conduit breakouts
are available for cord grips or conduit
hardware.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 10
2.4.2. ½-DIN Panel Mount (PN: AV38-1/2DIN)
A ½-DIN panel-mount bezel and gasket can be ordered to fit the AV38 as shown. The AV38 display is
tightened down to the bezel using the four captured screws in the AV38. The bezel is then mounted to a ½-DIN
square opening with the hardware shown in the figure below.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 11
2.4.3. ¼-DIN Panel Mount (PN: AV38-1/4DIN)
The AV38 Display can be mounted directly to a ¼-DIN panel opening using the captive mounting screws in the
AV38 and the ¼-DIN supports as illustrated in the figure below. A ¼-DIN bezel gasket can be provided to
ensure a NEMA 4X seal on the panel.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 12
2.5. Wiring
DC-Powered Version (AV38-XBXXX1)
The DC-powered AV38 Display comes with two terminal blocks when ordered with one or two 4-20mA analog
outputs and/or host communications. If relays are ordered, a third terminal block is present. The figure below
shows an AV38 with the DeviceNet host communications option present.
DC-powered AV38 with DeviceNet host comms
The current loops (analog outputs) should be wired as follows:
Terminal DC-Powered AV38 Current Loop Wiring
1 Current Loop 1 (+) (Power provided by AV38)
2
3 Current Loop 2 (+) (Power provided by AV38)
Thermo Scientific AquaSensors
wiring
Current Loop 1 (–)
Current Loop 2 (–)
™
AV38 Local Display and Controller User Guide 13
The cable from the DataStick should be wired as follows:
Terminal DC-Powered AV38 with DeviceNet Wiring
8 Red wire from DataStick
9 Black wire from DataStick
10 White wire from DataStick
11 Blue wire from DataStick
12 Clear wire from DataStick
The DeviceNet cable from the host should be wired as follows:
Terminal DC-Powered AV38 with DeviceNet Wiring
4 Ground from DeviceNet Power Supply (Black)1
5 CAN_H (White)
6 CAN_L (Blue)
7 Shield (Clear)
Q 24VDC from DeviceNet Power Supply (Red)
The relays should be wired as follows:
Terminal DC-Powered AV38 with DeviceNet Wiring
A Relay B: Normally Closed Contact
B Relay B: Common
C Relay B: Normally Open Contact
D Relay A: Normally Closed Contact
E Relay A: Common
F Relay A: Normally Open Contact
1
Terminal 4, Ground (Black), is electrically the same as terminal P, GND, so the ground from the DeviceNet power supply can be connected to
either terminal.
™
Thermo Scientific AquaSensors
AV38 Local Display and Controller User Guide 14
DC-Powered Legacy Version (AV38-XBXXX)
The DC-powered legacy version of the AV38 Display comes with a single terminal block when ordered with one
4-20mA analog output and/or host communications. If a second analog output or relays are ordered, a second
terminal block is present.
AV38 Terminals: 2 Outputs and Relays AV38 Terminals: 1 Output and Host Comms
One or Two Terminal Blocks Depending on Features
Figure 2.1: AV38 Wiring
In either configuration:
1. 24 VDC Power is wired to Pins 1 and 2 as shown above. The power supply should be rated
to deliver 0.25 amps when the AV38 is used with one DataStick.
2. The DataStick(s) are wired to Pins 6 through 10.
3. Analog Output 1 is wired to Pins 11 and 12.
4. When Host Comms are used they are wired to Pins 3, 4 and 5. Be sure that the
communications ground is at the same potential as the 24 VDC power ground.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 15
The DC-powered AV38 wiring will vary depending on the options provided.
Model AV38-BB0A1: One 4-20mA Analog Output, No Relays and No Host Comms
Terminal DC-Powered AV38 Wiring
1 24 VDC from Power Supply
2 Ground from Power Supply
3 No connect
4 No connect
5 Earth Ground from Power Supply
6 Red wire from DataStick
7 Black wire from DataStick
8 White wire from DataStick
9 Blue wire from DataStick
10 Shield wire from DataStick
11 Current Loop + (Power provided by AV38)
12 Current Loop –
Model AV38-CB5C1: DC-Powered, DeviceNet Host, Two Relays and Two 4-20mA Analog Outputs
Terminal Wiring
1 24VDC from Power Supply
2 Ground from Power Supply and Network
3 DeviceNet Network CAN_HI: White Wire
4 DeviceNet Network CAN_LO: Blue Wire
5 Shield wire from Host Network
6 Red wire from DataStick cable
7 Black wire from DataStick cable
8 White wire from DataStick cable
9 Blue wire from DataStick cable
10 Shield wire from DataStick cable
11 Current Loop 1 (+)
12 Current Loop 1 (–)
A Relay A: Normally Open Contact
B Relay A: Common
C Relay A: Normally Closed Contact
D Current Loop 2 (+)
E Current Loop 2 (–)
F Relay B: Normally Open Contact
G Relay B: Common
H Relay B: Normally Closed Contact
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 16
AC-Powered (AV38-XBXXX2)
The AC-powered AV38 Display comes with two terminal blocks when ordered with one or two 4-20 mA analog
outputs and/or host communications. If relays are ordered, a third terminal block is present.
AC-powered AV38 wiring:
1. Analog Outputs 1 and 2 are wired to Pins 1, 2 and 3 as shown above.
2. The DataStick(s) are wired to Pins 8 through 12.
3. Relays A and B are wired to Pins A-F.
4. 100-240 VAC, 50/60 Hz power is wired to the wider-spaced Pins G, H and J.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 17
The AC-powered AV38 wiring will vary depending on the options provided.
Model AV38-CB0C2: Two 4-20mA Analog Outputs, two Relays and No Host Comms
Terminal AC-Powered AV38 Wiring
1 Current Loop 1 + (Power provided by AV38)
2 Current Loop –
3 Current Loop 2 + (Power provided by AV38)
8 Red wire from DataStick
9 Black wire from DataStick
10 White wire from DataStick
11 Blue wire from DataStick
12 Shield wire from DataStick
A Relay B, Normally Closed
B Relay B, Common
C Relay B, Normally Open
D Relay A, Normally Closed
E Relay A, Common
F Relay A, Normally Open
G 100-240 VAC, 50/60 Hz
H Neutral
J Earth Ground
2.6. Measure Screen Overview
The AV38 user interface is shown in Figure 1
alphanumeric characters and seven keys to navigate the menu. The contrast of the LCD module can be
adjusted by simultaneously pressing the ESC and UP-arrow keys (for more contrast) or the ESC and DOWNarrow keys (for less contrast).
MENU ESC ENTER ◄
Figure 1: The elements of the AV38 user interface.
For illustration of menu navigation in this manual, a pressed key is identified by a gray background, e.g.,
When the AV38 is equipped with host communications hardware, the measure screens will toggle between
channel 1, channel 2 and Gateway Mode using RIGHT and LEFT arrow keys.
991.3 µS/cm AB
30.8 °C DS1: 6
. It consists of an LCD module that contains two lines of 16
991.3 µS/cm AB
30.8 °C DS1: 6
▲
►
▼
7.57 pH AB
30.8 °C DS2: 7
Gateway Mode
<> to exit [ 1]
ESC
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 18
The sensor measurement is displayed on the top line and the temperature measurement is displayed on the
bottom line. The “AB” indicators on the top right indicate activation of relays A and B. An underscore indicates
that the corresponding relay is not activated. The “DS1: 6” indicates that Channel 1 is connected to the
DataStick with a Modbus address of 6. The “DS2: 7” indicates that Channel 2 is connected to the DataStick
with a Modbus address of 7. Calibration and configuration functions are accessed by pressing the MENU key
from the appropriate measure screen. For example, the DataStick with a Modbus address of 6 can be
calibrated and configured if the MENU key is pressed while the screen shows “DS1: 6”. Correspondingly, the
DataStick with a Modbus address of 7 can be calibrated and configured if the MENU key is pressed while the
screen shows “DS2: 7”. The Gateway Mode screen shows a bracketed number in the lower right corner to
indicate which DataStick is being accessed by the host. The MENU key does not present menu functions while
in the Gateway mode.
When the AV38 is not equipped with host communications hardware, the Gateway mode is not available. The
two measure screens are accessed with the RIGHT and LEFT arrow keys.
991.3 µS/cm AB
30.8 °C DS1: 1
7.57 pH AB
30.8 °C DS2: 3
When a sensor head of the same type is replaced in a DataStick, analog output and relay settings are
maintained. If a sensor head of a new type is installed, analog output and relay settings are automatically set to
default conditions.
The example in Figure 2
illustrates HOLD MODE as indicated in the measure screen. The “H” in front of the
“DS1: 6” in the lower right corner of the display indicates that outputs and relays are on hold.
991.3 µS/cm AB
30.8 °C HDS1: 6
▲
MENU ESC ENTER ◄ ►
Figure 2
▼
Figure 2: The "H" indicates that hold mode is active.
, the measure screen, also shows the sensor value, sensor units, temperature value, and temperature
units. Relays A and B are activated. The measurements are on hold and the Modbus station address for the
DataStick is 6.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 19
When the DataStick is absent the display will read “DataStick Absent” as shown in Figure 3
. If the sensor is
wired correctly and the “DataStick Absent” message appears, be sure that the DataStick Sensor has the correct
station number as indicated by the number in the lower right corner of the measure screen. If that is correct,
also check that baud rate and parity are set up correctly in the DataStick.
DataStick __
Absent DS1: 6
▲
MENU ESC ENTER ◄ ►
▼
Figure 3: The initial measure screen when the DataStick is absent.
When the sensor head is absent, the condition will be indicated by “Sensor Head Absent” on the measure
screen as shown in Figure 4
.
Sensor Head __
Absent DS1: 6
▲
MENU ESC ENTER ◄ ►
▼
Figure 4: The initial measure screen when the Sensor Head is absent.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 20
3. MENU STRUCTURE AND NAVIGATION
The Main Menu is accessed by pressing the MENU key when the AV38 is in Analyzer Mode (measure screen).
The menu structure is navigated using ENTER, UP/DOWN arrows, and ESC keys.
¾ Use the UP/DOWN arrow keys to scroll through the list of Menu options.
¾ Use the ENTER key to select an option.
¾ Use the ESC key to move up in the structure.
The end of the list of menu options is always the EXIT function. Pressing ENTER on the EXIT screen will bring
up the MEASURE screen.
The (≡) symbol is used to indicate that the user is in the menu structure. At the leaf of each branch of the menu
structure, the user is able to enter a value or make a choice. The value to be entered or the choice to be made
will be surrounded by parentheses (()). The name of the value being entered or choice being made will appear
on the upper line of the screen and be appended with a question mark (?) to indicate that the user is to provide
input.
The RIGHT arrow symbol (>) on the screen indicates that downward navigation through the menu structure is
possible. The LEFT arrow symbol (<) on the screen indicates that upward navigation through the menu
structure is possible.
Calibration and configuration of a specific sensor is accessed by pressing the MENU key from the measure
screen for that sensor. Main menu items are listed as follows:
Table 2: Main menu items.
Main Menu Item Function
Calibrate
(Station Number)
Configure
(Station Number)
DataStick Comms Set the station address, baud rate and parity for the Modbus DataStick of interest.
Chlorine Configure
Hold Outputs Holds the 4-20 mA and Relay outputs
Analog Output 1 Set 4 mA and 20 mA values. Calibrate and test the analog output with a milliammeter.
Analog Output 2 (option) Set 4 mA and 20 mA values. Calibrate and test the current output with a milliammeter.
PID Controller Set up PID control functions for analog outputs. Available with pH or Conductivity.
Relay A (option) Set timer, alarm, or control functions. (Optional)
Relay B (option) Set timer, alarm, or control functions. (Optional)
Diagnose (pH DataStick only) Indicates the pH Sensor Head slope
Help Indicates AV38 firmware version
Reset AV38 Resets AV38 configuration to default analog output, relay and PID settings.
Exit Returns to the measure screen
Sensor calibration choices are automatically available for installed sensor head.
Enter the MENU from the measure display of interest – station in lower right.
Configuration choices are automatically available for the installed sensor head.
Enter the MENU from the measure display of interest – station in lower right.
Select free chlorine or HOCL
or active. (Only when Chlorine DataStick Present).
concentration for output. Select pH compensation – fixed
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 21
MAIN MENU NAVIGATION
¾ Enter the Main Menu by pressing the MENU key.
¾ Scroll the listing with the UP/DOWN arrow keys.
¾ To select an item, press the ENTER key.
If there is no keypad activity for 15 minutes and the display is left on a menu screen, the measure screen will
automatically re-appear. The exception to this rule is when the manual mode screen in the PID controller is
displayed. In this case, there will be no function time-out. The AV38 remembers the last menu option used and
will return to that menu option when re-entering the main menu from the measure screen. The list order is
always the same, so, if pressing MENU displays analog output and the desired operation is calibration; the UP
arrow will be used to move to the Calibrate menu.
≡Main Menu
►Calibrate 1
Select CALIBRATE with the ENTER key
¾ To calibrate the main parameter or temperature of
the sensor previously displayed in the measure
▲
screen. In this example, the “1” in the lower right of
the display indicates the sensor at Modbus Station 1.
MENU ESC ENTER ◄ ►
▼
Press the DOWN ARROW for next selection.
≡Main Menu
►Configure 1
Select CONFIGURE with the ENTER key
¾ To display or change filters, units of measure,
constants, tables and buffer standards of the sensor
▲
MENU ESC ENTER ◄ ►
previously displayed in the measure screen. In this
example, the “1” in the lower right of the display
indicates the sensor at Modbus Station 1.
▼
Press the DOWN ARROW for next selection.
≡Main Menu
► DataStick Comms
Select DATASTICK COMMS with the ENTER key
¾ To set up station address for the DataStick that is to
▲
be displayed in the AV38 measure screen.
MENU ESC ENTER ◄ ►
▼
Press the DOWN ARROW for next selection.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 22
≡Main Menu
►Chlorine Config
Select Chlorine Config with the ENTER Key
¾ Set Measurement: Free chlorine or HOCL
E
MENU ESC ENTER ◄ ►
¾ Set pH Source: Fixed or active
¾ Set Fixed pH Value
▼
Press the DOWN Arrow for next selection.
≡Main Menu
►Hold Outputs
MENU ESC
▲
▲
◄ ►
▼
Select Hold Outputs with the ENTER key
¾ Put outputs and relays on hold. There is no
timeout with this function.
Press the DOWN ARROW for next selection.
≡Main Menu
►Analog Output 1
Select ANALOG OUTPUT1 with the ENTER key
¾ To set up 4 to 20 mA Analog Output for sensor,
▲
temperature or PID.
MENU ESC ENTER ◄ ►
▼
Press the DOWN ARROW for next selection.
≡Main Menu
► Analog Output 2
Select ANALOG OUTPUT2 with the ENTER key
¾ To set up 4 to 20 mA Analog Output for sensor,
▲
temperature or PID.
MENU ESC ENTER ◄ ►
▼
Press the DOWN ARROW for next selection.
Thermo Scientific AquaSensors
™
AV38 Local Display and Controller User Guide 23
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.