Thermo Scientific AQ3010 Instruction Manual

Page 1
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter
User Guide
Page 2
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. US patent 6,793,787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
© 2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
The specific
ations, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Page 3
Table of Content
Chapter 1
Chapter 2
Chapter 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Information . AQ3010 Meter Overview .
Unpacking the Meter and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . 2
isplay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
D Keys and Functions. Battery Installation .
Turbidity Calibration .
Calibration Overview . Indexing . Calibration Standards . Preparing and Filling the Calibration Standard Vials . Calibration Check. Refilling the Calibration Standard Vials with Fresh Standard. Calibration Procedure . Restoring Factory Calibration .
Turbidity Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
General Information . Turbidity Measurement Procedure .
Preparation of Sample Vial. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
M
ingle-Shot or Continuous Measurement. . . . . . . . . . . . . . . . . . . . . .19
S
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
easurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
. . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
. . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Chapter 4
Chapter 5
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Routine Maintenance .
Vials – Handling, Cleaning and Care .
Customer Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Troubleshooting Guide Assistance . Warranty . Notice of Compliance . WEEE Compliance. Declaration of Conformity . Ordering Information . Specifications.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
. . . . . . . . . . . . . . . . . . . . . . . . . 21
I
Page 4
Appendix 1
Turbidity Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Definition . Why Is It Important . Measurement Principle . Nephelometric Turbidity Units (NTU) . Indexing a Vial .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . 30
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Appendix 2
Guide to Good Measurement Technique .
Maintain Sample Vials in Good Condition. Match Sample Vials. Degassing . Timeliness of Sample. Other Important Sampling Techniques . Calibration Techniques . Dilution Techniques .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
. . . . . . . . . . . . . . . . . . . . . . . . 33
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
. . . . . . . . . . . . . . . . . . . . . . . . . 31
. . . . . . . . . . . . . . . . . . . . . 31
II Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 5
Chapter 1 Introduction
General
Information
This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion AQUAfast® AQ3010 turbidity meter. help you operate the waterproof AQ3010 turbidity meter and serves as a handy r
Follow the basic instructions contained in this user guide during the operation, car provided by this equipment may be impaired if it is used in a manner not described in this user guide read this user guide prior to working with this meter.
The manufacturer will not accept any responsibility for damage or malfunction to the meter caused b
The information presented in this user guide is subject to change without notice as improvements are made, and does not represent a commitment on the par
Note: The manufacturer reserves the right to make improvements in design, constr
eference guide.
This user guide provides both a step by step guide to
e and maintenance of the meter. The safety protections
. It is recommended that all operators should
y improper use of the instrument.
t of the manufacturer.
uction and appearance of products without notice.
AQ3010 Meter
Overview
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Thank you for selecting the waterproof portable AQ3010 turbidity meter.
The AQ3010 meter allows you to measure turbidity of an aqueous sample in the field. This instrument operates on the nephelometric principle of turbidity measurement and is designed to meet the criteria specified in ISO 7027 and DIN 27027 standar Principle section
.
ds. See the Measurement
1
Page 6
Unpacking
the Meter and
Accessories
The list below indicates the items that you should find in your AQ3010 turbidity meter shipment.
Item Quantity
AQ3010 portable turbidity meter with 4 AAA batteries 1
User guide 1
Instrument carrying case 1
Light shield cover 1
Calibration set with 0.02, 20.0,100 and 800 NTU standards 1
Empty calibration vials 4
Empty sample vials 3
Silicone oil 1
Lint free cloth 1
Vial cleaning brush 1
Remove the AQ3010 turbidity meter from the carrying case. Carefully inspect all items to ensure that no visible damage has occurred during shipment
. If the items you received do not match your order, please
contact your dealer immediately.
Warning: Extra care should be taken when unpacking, opening and handling the calibration standar
ds and sample vials. Surface scratches or finger smudges on the vial surface may cause measurement errors. Handle these items by their caps only.
The batteries provided with the meter package are to be installed prior to use
. Refer to the Battery Installation section
Figure 1 sho
ws the meter. The three main components of the instrument
.
are the sample well, the display and the keypad. The following sections describe the functionality of the display and the keypad. The proper use of the instrument and the sample well are discussed in later sections.
2 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 7
Figure 1
Sample Well
Alignment Mark
Liquid Crystal Display
Keypad
Sample Vial
AQ3010 Turbidity
Meter Components
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
3
Page 8
Display
CAL 13 CAL 24
mg/l NTU ppm %pH
TAvg
14-segment Liquid Crystal Display
Battery Indicator
Calibration Standards
Units of Measurement
Figure 2
Customized LCD with
All Annunciators
All the LCD segments and annunciators that can appear on the display are shown in Figure 2.
The display is used for reporting the turbidity reading and to provide guidance for the operation of the instrument. In addition, the display has several other annunciators that are used to communicate err
or messages and provide user guidance.
4 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 9
read
enter
cal
Turbidity Meter Chlorine Colorimeter
Keys and
CAL
CAL
CAL
READ
ENTER
CAL
CAL
Functions
The keypad has five keys: ON/OFF, CAL, , and READ/ENTER.
Key Functions
Powers on and off the meter. The meter automatically shuts off 20 minutes after last key press to conserve the battery life.
Initiates the calibration mode of the meter. When pressed, the meter is set to accept the first calibration standard.
It is also used to exit the calibration mode if the user does not want to follow the complete calibration procedure.
When in measurement mode, the READ/ENTER key is used to perform a measurement.
Single-shot measurements: When the key is pressed and released immediately (a quick key stroke of less than 0.3 seconds), the display will blink [-Rd-] ten times and display the measured value.
Continuous measurements: If the READ/ENTER key is pressed and held, the instrument will perform a continuous measurement during which the display is updated every 2 seconds. This can be used for indexing vials. (This function is not available in calibration mode). After the READ/ENTER key is released, the instrument will automatically perform a single-shot measurement.
Active only during calibration mode. These scroll keys are used to select the next calibration point.
Note: After the successful calibration of one point, the meter automatically selects the next calibration point, or automatically exits the calibration mode after the fourth calibration point.
Figure 3
Keypad
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
5
Page 10
Battery
Installation
Four AAA batteries are included in the carrying case.
Note: Use caution when loosening the screws, as some of the screws may be tightly seated
1. Use a Phillips screwdriver to loosen the four screws holding the
battery cover. See Figure 4. attached to the battery cover.
2. Remove the battery cover.
.
These captive screws will remain
Figure 4
Battery Installation
3. Insert the batteries. Follow the diagram inside the cover for
ect polarity.
corr
4. Replace the battery cover in its original position and tighten the four
screws removed earlier to secure the battery cover to the meter.
5. The meter is now ready for use.
Note: D
ispose of used batteries in accordance with your local regulations.
6 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 11
Chapter 2 Turbidity Calibration
Calibration
Overview
Indexing
The AQ3010 turbidity meter was calibrated and tested prior to leaving the factory. The meter kit includes a set of prepared primary standards that can be used to verify the meter calibration daily and to recalibrate the
, if required. Before measuring samples for the first time, verify the
meter meter calibration with the prepared primary standards.
The SDVB primary standards in the meter kit are in plastic bottles for impr
oved precision and shelf life. Four vials are included and labeled for
each standard level.
The instrument includes a light shield cover to keep stray light from changing the calibration and sample measur environments, the light shield is not required. The vial cap keeps stray light from entering the sample well, saving you many steps and reducing analysis time
Due to the high quality of the glass vials provided, indexing is not r
equired. You only need to align the mark on the vial with the mark on the meter. However, in order to achieve a better precision of the measurement, you can proceed with indexing of the vials. See the Indexing a Vial section for mor
.
e information.
ement readings. In many
Calibration
Standards
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
We recommend that you use the following materials during calibration to achieve the precision stated in this user guide:
CAL 1: 800 NTU calibration standard
CAL 2: 100 NTU calibration standard
CAL 3: 20.0 NTU calibration standard
CAL 4: 0.02 NTU calibration standard
It is well known that diluted Formazin is unstable. If you choose to use Formazin to calibrate the instrument, ensure that you are using a fresh stock suspension of Formazin to achieve the precision quoted for the instr
ument. It is very difficult to accurately pipet Formazin for low level NTU standards, due to the non-homogeneous nature of Formazin and because these standar to use good laboratory technique and accurate Class A pipets and refer to ASTM Method 6855-03 for low level standard vial preparation when using F
ormazin.
ds have very limited stability at low levels. Be sure
7
Page 12
The four bottles of calibration standards supplied with your meter are more stable than Formazin, do not need to be shaken and have a shelf life of 12 months the instrument, review the expiration date (indicated on the bottle label) to ensur
Note: It is important that the calibration standards are not violently shaken or agitated because air entrapment in the fluid introduces an error factor during calibration which subsequently will lead to an inaccurate measurement. Also, do not store in freezing temperatures which causes irreversible shrinkage of the standards’ particles thus resulting to inaccurate calibration and measur
. If you use the supplied calibration standards to calibrate
e that the standards have not expired.
ement.
Preparing and
Filling the
Calibration
Standard
Vials
Before using the meter for the first time, rinse the inside of the vial with a small amount of the same NTU standar
1. Gently pour about 5 mL of the standard into the vial.
2. Cap and swirl the vial so the inside of the vial is fully rinsed with the standar
3. Pour the standard out of the vial.
4. Repeat steps 1 through 3.
5. Gently pour the standard into the vial up to the mark on the vial and cap the vial.
6. Repeat this rinse and filling procedure for the other three vials.
Note: The vial rinsing is most critical for precision with the lower NTU standar
It is recommended to use the prepared calibration standard vials daily as check standar
d.
ds.
ds prior to the measurement of samples.
d that will be used to fill that vial.
8 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 13
“-Rd-” flashes about 10 times
CAL STD
100
NTU
Place standard vial into the sample well
NTU
READ
ENTER
Measurement Mode
Press key and release immediately
100
Calibration
1. Place the AQ3010 turbidity meter on a flat and level surface.
Check
2. Insert the CAL 1 standard (800 NTU) into the sample well, aligning the mark on the vial with the alignment mark on the meter. See Figure 11.
3. Press down the vial until it slides fully into the instrument.
4. Cover the vial using the light shield cover, if necessary. For many environments this is not required. The vial cap seals the sample well.
5. Press the
ON/OFF key to turn on the meter.
The meter goes to
measurement mode after the self test power-up sequence.
6. Press the
READ/ENTER key.
7. The meter will display the results. Record the reading.
8. Repeat the calibration check for CAL 2, CAL 3 and CAL 4 calibration standards.
9. If the displayed results are within 10% of the nominal NTU value of the standard or the precision criteria required by your method, the calibration check passed and the meter is no
w ready for measurement.
Figure 5
Calibration Check
Sequence
Note: If a standard no longer reads within 10% of the nominal NTU v
alue for the standard, the standard has most likely degraded. The lower level NTU standards will be less stable when stored in glass than the higher NTU standar
ds. The lower level standards may need to be
replaced periodically, especially if the value falls outside of the tolerance
. See the Refilling the Calibration Standard Vials with Fresh
range Standard section
.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
9
Page 14
Refilling the
1. Pour the old standard out of the vial.
Calibration
Standard
Vials with
Fresh
Standard
2. Wash the vial with laboratory glassware detergent and water. Rinse the vial with turbidity-free water to remove all laboratory detergent fr
om the vial.
3. Rinse the vial with about 5 mL of the standard from the plastic bottle with the matching NTU value.
4. Cap and swirl the vial so the inside of the vial is fully rinsed with the fresh standard.
5. Pour the standard out of the vial.
6. Pour the fresh standard into the vial up to the mark on the vial and cap the vial.
7. Perform the procedure in the containing the fresh standard.
If the fresh standard does not read with the 10% range of the nominal value, or to the tolerance of your method, follow the steps in the
Calibration Procedure section.
Calibration Check section with the vial
10 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 15
Calibration
Procedure
Note: Before switching on the AQ3010 turbidity meter, a sample vial
must be placed in the sample well. You can use any of the calibration standards for this purpose. If a sample vial is not placed in the sample well, the turbidity meter will display the Err 8 error message. To clear the error message, power off the meter., insert a sample vial into the sample well and power on the meter.
1. Place the AQ3010 turbidity meter on a flat and level surface.
2. Press the measurement mode after the self test power-up sequence.
3. Select the calibration function of the instrument by pressing the CAL key once the meter will prompt for the first calibration standard by displaying “CAL 1 800 NTU”
4. Insert the CAL 1 standard (800 NTU) into the sample well, aligning the mark on the vial with the alignment mark on the meter. See Figure 11.
5. Press down the vial until it slides fully into the instrument.
6. Cover the vial using the light shield cover, if necessary. For many environments this is not required. The vial cap seals the sample well.
7. Press the
8. The “CAL 1 800 NTU” annunciator will blink for about 12 seconds while the instrument performs the calibration of CAL 1 point. When the instrument has completed calibration for this point, it will prompt you to insert the next calibration standard into the sample well by displaying “CAL 2 100 NTU”.
ON/OFF key to turn on the meter.
. The “CAL” annunciator will blink momentarily and
.
READ/ENTER key.
The meter goes to
9. Repeat steps 4 through 8 for the CAL 2, CAL 3 and CAL 4 calibration standards.
10. After you successfully calibrate the CAL 4 standard (0.02 NTU), the meter will display “STbY”.
11. The meter is now ready for measurement.
Figure 6 sho
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
ws the complete calibration sequence.
11
Page 16
Calibration
CAL 1
800 NTU
NTU
CA L 1
NTU
NTU
CAL 3
CA L 3
NTU
CAL 2
NTU
CA L 2
RE AD
EN TER
NTU
CA L 1
CA L
Pla ce t he C AL 1 sta ndar d int o the sam ple wel l
Pla ce t he C AL 3 sta ndar d int o the sam ple wel l
Fla shes abo ut 5 time s
Fla shes ab out 1 0 tim es
Fl ash es ab out 10 ti me s
Fla she s abo ut 1 0 tim es
Fla she s abo ut 1 0 tim es
CAL 4
CAL 3
20.0 NTU
0.02 NTU
NTU
NTU
CA L 4
CA L 4
RE AD
EN TER
RE AD
EN TER
Pla ce th e C AL 2 sta ndar d int o the sa mple wel l
Pla ce t he C AL 4 sta ndar d int o the sam ple wel l
Measurement Mode
CAL 2
100 NTU
RE AD
EN TER
Measurement Mode
Calibration saved
Notes
To exit the calibration mode, press the CAL key at the end of any step. The meter accepts only the values calibrated prior to exiting.
Figure 6
Calibration
Sequence
You can skip a calibration point by pressing the
or key and move
on to the next calibration point.
After a successful calibration of one point, the meter automatically selects the next calibration point. The meter automatically exits calibration mode after the fourth calibration point.
If an error occurs during calibration, the display shows an error message. The meter aborts the calibration and returns to the measurement mode without saving the last calibration value.
For an error list, refer to the
Troubleshooting Guide section
.
12 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 17
Restoring
Factory
Calibration
The AQ3010 turbidity meter allows you to reset the meter back to the factory default calibration values. This feature is extremely useful when there are errors in calibration or when you have new calibration standards.
1. With the meter switched off, press and hold the
key.
2. Switch on the meter by pressing the ON/OFF key. when “URST” (User-Reset) appears on the display.
3. The display will show “URST” for about 2 seconds and then “No”.
4. Use the
YES = Restore the meter back to factory calibration values
No = Retain the last calibrated values
5. Press the performs the reset if “YES” is selected.
6. The display will flash “-Rd-” about ten times and the meter will go to the measurement mode.
Figure 7 sho
or key to scroll between “YES” or “No”.
READ/ENTER key to confirm y
ws the sequence for restoring factory calibration values.
our selection. The meter
Release the key
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
13
Page 18
Figure 7
READ
ENTER
READ
ENTER
“-R d-” f lash es a bo ut 10 tim es
Se lect YE S an d
to res tore fact ory
ca libra tion
the n pr ess ke y
RE AD /ENT ER
Se lect No
to ret ain las t
ca lib ratio n v alue s
wi tho ut ret riev ing
fac tory ca libra tion
an d th en
pr ess
ke y
RE AD /EN TE R
Instrument Off
Measurement Mode
Pr es s & ho ld k ey and sw itch on t he me ter u sing
ke yON /OF F
Re lea se key wh en y ou se e ‘U RST ’ in the di spla y
Pr es s or key to
se lec t o r YE S No
Restoring Factory
Calibration
14 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 19
Chapter 3 Turbidity Measurement
CA L 13 C AL 24
mg /l NTU ppm %p H
TAvg
NTU
"-R d-” f las hes ab out 10
tim es ind icat ing tha t the
me ter g oes t o
m eas ure men t mo de.
A re adin g is val id o nly i f yo u h ave pla ced a
via l or a ny o f the cal ibra tion stan dar ds
pro vide d in to th e sa mp le we ll.Oth erwi se
the rea ding has no m ean ing o r an err or
me ssag e " Err 8" app ears .
Instrument Off
Pla ce a sam ple vial or a ny of th e
ca libra tion s tand ards pro vid ed i nto
the sam ple w ell of th e m ete r, be fore
sw itch on.
X.x
“X .X” is s oftw are r evisi on
nu mb er
Place light shield cap over vial.
General
Information
Figure 8
Power-up Sequence
The AQ3010 waterproof turbidity meter allows you to measure the turbidity of a grab sample. The turbidity is reported in Nephelometric Turbidity Units (NTU). Readings above 1000 NTU are outside the range of this instrument.
Note: Before switching on the AQ3010 turbidity meter, a sample vial must be placed in the sample w
ell. You can use any of the calibration
standards for this purpose.
The instrument includes a light shield cover to keep stray light from changing the calibration and sample measur
ement readings. In many environments, the light shield is not required. The vial cap keeps stray light from entering the sample well, saving you many steps and reducing analysis time
.
When the ON/OFF key is pressed to switch on the meter, it goes through the power-up sequence as shown in Figure 8.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
15
Page 20
Turbidity
Measurement
Procedure
An accurate turbidity measurement depends on good measurement techniques. Factors such as clean sample vials, positioning of vial in the sample well, covering the vial with the light shield cover (if needed), meter calibration, handling of meter and others have to be taken into consideration Appendix 2 Guide to Good Measurement Technique sections for more information
. Refer to the Vials – Handling, Cleaning and Care and
.
Preparation of
Sample Vial
1. Obtain a clean and dry sample vial. See Figure 9. handle the sample vial by the top cap.
2. Rinse the vial with approximately 10 mL of the sample water, capping the vial with the black screw cap and gently inverting it several times. Discard the used sample and repeat the rinsing procedure two times.
4. Fill the rinsed vial with the remaining portion (approximately 10 mL) of the grab sample up to the mark indicated in the vial. Cap the vial with the supplied black screw cap.
5. Wipe the vial with the soft, lint-free cloth supplied. Ensure that the outside of the vial is dry, clean and free from smudges.
6. Apply a thin film of silicone oil (supplied) on the sample vial. See Figure 10.
7. Wipe the vial with a soft cloth to obtain an even distribution over the entire vial surface. The purpose of oiling the vial is to fill small scratches and to mask the imperfection in the glass. Do not apply large quantity of oil as this may collect dirt and dust.
Take care to
8. The sample vial is now ready to be inserted into the sample well of the meter for measurement.
Figure 9
Sample Vial
16 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 21
Figure 10
Align the alignment mark on the sample vial with the alignment mark on the AQ3010 turbidity meter and push until it is fully snapped in
Applying the Silicon Oil
Measurement
Procedure
Figure 11
Aligning the Sample
Vial Alignment Mark
with the Meter
Alignment Mark
1. Place the AQ3010 turbidity meter on a flat and level surface.
Note: Do not hold the meter in your hand while operating it, as this may cause inaccurate measur
ements.
2. Place the sample vial inside the sample well and align the vial’s alignment mark with the meter’s alignment mark. See Figure 11.
. Push the vial down until it slides fully in.
3
4. Cover the vial using the light shield cover, if necessary. For many environments this is not required. The vial cap seals the sample well.
5. Turn on the meter by pressing the
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
ON/OFF key.
17
Page 22
6. After the power-up sequence, the meter will go to the measurement
“-Rd-” flashes about 10 times
SAMPLE
Unknown
NTU
Place sample vial into the sample well
NTU
REA D
EN TE R
Measurement Mode
and cover with the light shield cover, if needed
mode and the display will blink “-Rd-” about ten times. See Figure 12.
Figure 12
Reading a
Turbidity Value
. The measured reading will appear in the display.
7
8. If necessary, place the second sample vial into the sample well. Align the vial alignment mark with the meter alignment mark.
Measurement Notes
9. Press the
READ/ENTER key.
The display will blink “-Rd-” several
times and measured reading will appear.
10. Repeat this procedure for all of your samples.
Never pour liquid directly into the sample well of the instrument. Always use a 25 mm vial. The instrument will only accurately measure the turbidity of a sample when 25 mm vials sealed with the black caps ar
e used. The black cap serves to both seal the vial and
block stray light.
Never attempt to clean the sample well. The optics may be damaged.
For battery conservation, the instrument automatically powers off 20 minutes after the last key pressed.
18 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 23
Single-Shot
“-Rd-” flashes about 10 times
SAMPLE
Unknown
NTU
Place sample vial into the sample well
NTU
R EA D
EN TER
Measurement Mode
Press key and release immediately
and cover with the light shield cap
or Continuous Measurement
You can use the AQ3010 turbidity meter to take a single reading or perform continuous measurement. The latter is only used for indexing the vials. See the Indexing a Vial section for mor
e information.
For Single-Shot
Measurement
Figure 13
Single-Shot
Measurement
1. Make sure the meter is sitting on a flat and level surface and is in measurement mode. The display will show the last measured value or “STbY” after exiting calibration mode.
2. Place the sample vial in the sample well.
3. Cover the vial using the light shield cover, if necessary. For many environments this is not required. The vial cap seals the sample well.
4. Press the 0
.3 seconds). See Figure 13.
READ/ENTER key and immediately release it (less than
5. The display will blink “-Rd-” about ten times and then display the measured value.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
19
Page 24
For Continuous
Reading flashes and
updates every 2 seconds
SAMPLE
Unknown
NTU
Place sample vial into the sample well
NTU
R EA D
EN TER
Measurement Mode
Press key and hold
and cover with the light shield cap
Measurement
1. Make sure the meter is sitting on a flat and level surface and is in the measurement mode. The display will show the last measured value or “STbY”.
2. Place the sample vial in the sample well.
3. Cover the vial using the light shield cover, if necessary. For many environments this is not required. The vial cap seals the sample well.
Figure 14
Continuous
Measurement
4. Press the
READ/ENTER key and hold it
. See Figure 14.
5. Wait for the reading to stabilize before rotating the sample vial. You can rotate the sample vial for indexing purpose. See the Vial section for mor
e information.
Indexing a
Note: During continuous measurement, the display is updated every
2 seconds
. The displayed reading may not be the actual turbidity
value. For accurate measurement, use the single-shot measurement.
6. Once you release the per
forms a single-shot measurement.
READ/ENTER key, the meter automatically
20 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Continuous
Measurement
Notes
After a measurement is completed and the display is updated, a 4 second recovery time occurs before the meter can perform another function. If any key is pressed during this time, the meter will perform the corresponding action at the end of the recovery period.
When performing a measurement, if the meter detects stray light exceeding the amount equivalent to 0.02 NTU in the low range, the measurement is immediately aborted and an error message [ERR 8] is
ed. Ensure that the vial sits properly into the sample well and
display the light shield cover is on. Press the ENTER/READ key to retake the measurement.
Page 25
Chapter 4 Routine Maintenance
The supplied carrying case is optimal for protecting the instrument. If you do not plan on leaving the instrument in the supplied carrying case, when it not in use ensure that the instrument has been turned off and that a clean sample vial fitted with a black cap has been placed in the sample w will be able to settle on the optics of the instrument. Keep the light shield cover with the meter.
ell. This will ensure that a minimal amount of dust and debris
Vials –
Handling,
Cleaning
and Care
Proper measurement of the turbidity of a sample requires the use of a vial that is fr Therefore, sample vials must be handled with absolute care to avoid contamination or damage, which might change the optical characteristics of the glass vial or inside the sample well can cause stray light interference leading to inaccurate r
Cleaning the vial is accomplished by washing the interior and exterior of the vial in a detergent solution thoroughly eight to ten times with clean distilled or deionized water to eliminate the possibility of detergent buildup and str
Vials can also be acid washed periodically and coated with a special silicone oil to fill small scratches and mask the imper Since the silicone oil required for this application should have the same refractive characteristics as glass, it is recommended that only the Thermo Scientific O Care should be taken not to apply excessive oil that could attract dirt or contaminate the sample w to the vial, the excess oil should be removed with a lint-free cloth. The result should be a sample vial surface with a dry appearance, but with all imper
ee of marks, smudges, scratches and any bacterial growth.
. Scratches, fingerprints and water droplets on the sample
eadings.
. Once cleaned, the vial should be rinsed
eaking.
fections in the glass.
rion AQUAfast® silicone oil, Cat. No. AC3SIL, be used.
ell of the meter. Once the oil has been applied
fections filled with oil.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
21
Page 26
Sample vials should always be handled from the top or by the cap to avoid fingerprints or smudges. After a vial has been filled with a sample and capped, the outside surface should be wiped with a clean, lint-
ee absorbent cloth until it is dry. Cleaned and dried vials should be
fr stored with the black caps on. The vials can be stored in the carrying case. During normal operation, you may use any typical laboratory glass cleaner along with a lint free cloth or tissue to clean the outside of the vials
.
Condensation may appear on the vial when your sample is very cold and the r
elative air humidity is high. When this happens, the turbidity
that you read may be higher than the actual turbidity due to the light
ed by the condensate on the vial. If this occurs, you can alleviate
scatter the problem by either coating the vial with an anti-fogging agent or by running warm water over the vial for a short period of time to warm the sample prior to measur
ement.
22 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 27
Chapter 5 Customer Services
Troubleshooting
Guide
The AQ3010 turbidity meter routinely performs self-diagnostics and will automatically generate messages to provide you with specific diagnostic information
. These messages are for your use and do not indicate a reduction in the performance of the instrument or a failure of any component in the instr
LCD Message
ERR 1
ERR 2
ERR 3
Description Corrective Actions
Calibration error – the meter is unable to recognize the 800 NTU calibration standard.
Calibration error – the meter is unable to recognize the 100 NTU calibration standard.
Calibration error – the meter is unable to recognize the 20.0 NTU calibration standard.
ument, unless otherwise stated in this list.
Use the correct 800 NTU calibration standard. Pour fresh standard into a clean vial and retry the calibration. Verify that the standard has not expired.
Use the correct 100 NTU calibration standard. Pour fresh standard into a clean vial and retry the calibration. Verify that the standard has not expired.
Use the correct 20.0 NTU calibration standard. Pour fresh standard into a clean vial and retry the calibration. Verify that the standard has not expired.
Calibration error – the meter is unable to recognize the 0.02 NTU
ERR 4
ERR 5
ERR 6
ERR 7
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
calibration standard.
Calibration error – there is not sufficient signal to achieve the appropriate resolution in the 0 to 1000 NTU range.
General calibration failure – there is not sufficient signal to achieve the appropriate resolution in the 0 to 100 NTU range.
General calibration failure – there is not sufficient signal to achieve the appropriate resolution in the 0 to 20 NTU range.
Use the correct 0.02 NTU calibration standard. Pour fresh standard into a clean vial and retry the calibration. Verify that the standard has not expired.
Pour fresh standard into a clean vial and retry the calibration with all four standards. Verify that the standards have not expired.
Pour fresh standard into a clean vial and retry the calibration with all four standards. Verify that the standards have not expired.
Pour fresh standard into a clean vial and retry the calibration with all four standards. Verify that the standards have not expired.
23
Page 28
LCD Message
ERR 8
ERR 9 Lamp failure. Contact Technical Support.
Description Corrective Actions
Excessive stray light detected. Verify that the vial is fully in the
sample well and light shield cover is properly placed over the vial.
Attempting a measurement when low battery indicator is on.
Replace the batteries.
Assistance
Warranty
Or
Turbidity value is above the measurement range (>1000 NTU).
Low battery indication – the batteries need to be replaced.
Dilute sample, see the Dilution section.
Replace the batteries.
Note: If an error message appears, take the appropriate corrective action and re-do the
ed procedure. If the problem persists,
desir contact Technical Support or your dealer.
After troubleshooting all components of your measurement system, contact
Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized
For the most current contact information, visit
dealer.
thermo.com/contactwater.
www.
For the most current warranty information, visit www.
thermo.com/water.
The AQ3010 meter is supplied with a 3 year warranty from manufacturing defects and calibration standar
ds for 6 months. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please contact Technical Support
our dealer for an RMT number. A repair technician will determine
or y if the product problem is due to deviations or customer misuse. Out of warranty products will be repaired on a charged basis.
Exclusions
The instrument warranty shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of the products
Waterproof Seal: Opening the instrument enclosure (excluding the
y compartment) may void the warranty.
batter
24 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 29
Notice of
Compliance
Warning: This meter may radiate radio frequency energy and if
not installed and used properly, that is in strict accordance with the manufacturer’s instructions, may cause interference to radio communications for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such
ference in a commercial environment. Operation of the meter in a
inter residential area may cause interference in which case the user at his own expense will be required to take whatever measures may be required to
ect the interference.
corr
The Thermo Scientific Orion AQ3010 turbidity meter is CE and cTUVus certified, which includes TUV 3-in-1 testing to EMC and US and Canadian standar
. It has been tested and found to comply with the limits
ds.
WEEE
Compliance
This product is required to comply with the European Union’s Waste E
lectrical and Electronic Equipment (WEEE) Directive 2002/96/EC. It
is marked with the following symbol:
We have contracted with one or more recycling/disposal companies in each EU Member State and this product should be disposed of or recycled
ough them. Further information on compliance with these Directives,
thr the recyclers in your country, and information on Thermo Scientific Orion products which may assist the detection of substances subject to
oHS Directive are available at www.thermo.com/WEEERoHS.
the R
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
25
Page 30
Declaration of
Conformity
Manufacturer: Thermo Fisher Scientific Inc.
166 Cummings Center Beverly, MA 01915 USA
The manufacturer hereby declares that the AQ3010 turbidity meter conforms with the following standards and documents:
Safety EN/IEC 61010-1:2001
EMC EC 89/336/EEC Electromagnetic Compatibility
EN 61326-1:2006
Emissions EN 61326-1:2006 Emissions
CISPR 11:2003 Radiated Emissions FCC Part 15B Class A
Immunity
This Thermo Scientific Orion product has been manufactured in compliance with the provisions of the relevant manufacturing and test documents and pr as complying with ISO 9001:2008 by QMI, listed as File # 001911.
Patrick Chiu June 3, 2009 Senior Quality Engineer, Beverly, MA Regulatory Compliance
EN 61326-1:2006 Immunity IEC 61000-4-2:2001 Electrostatic Discharge
Immunity
IEC 61000-4-3:2002 RF Radiated I
ocesses. These documents and processes are recognized
mmunity
26 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 31
Ordering
Cat. No. Description
Information
AQ3010 AQ3010 waterproof portable turbidity meter with set of 4 calibration
standards (800, 100, 20.0 and 0.02 NTU) in plastic bottles, set of 4 calibration vials, set of 3 sample vials, lint-free cloth, silicone oil, vial cleaning brush, batteries and rugged carrying case
AC301S
AC3V25 Replacement sample vials, pack of 3, 25 mm diameter
AC3SIL Silicone oil, 10 mL bottle, and lint-free cloth
AC3CBR Replacement vial cleaning brush
Replacement calibration standards kit, includes 800, 100, 20.0 and 0.02 NTU standards in plastic bottles
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
27
Page 32
Specifications
Parameter Specification
Measurement Method ISO 7027 compliant nephelometric
method (90°)
Measurement Range 0 to 1000 NTU
Automatic Range Selection 0.01 to 19.99 NTU
20.0 to 99.9 NTU 100 to 1000 NTU
Resolution 0.01 NTU (0 to 19.99 NTU)
0.1 NTU (20 to 99.9 NTU) 1 NTU (100 to 1000 NTU)
Precision ±2% of reading ± 1 LSD for 0 to 500 NTU;
±3% of reading ± 1 LSD for 501 to 1000 NTU
Calibration Standards 0.02 NTU, 20 NTU, 100 NTU, 800 NTU
Standardization EPA approved polymer-based primary
standards
Light Source Infrared-emitting diode (850 nm wavelength)
Light Source Life > 1,000,000 tests
Detector Silicon photovoltaic
Stray Light < 0.02 NTU
Display 4 digit 14 segments customized liquid crystal
display with annunciators
Sample Vials (Cuvettes) Glass vials with screw caps, fill line and
indexing mark, 51 (H) x 25 (Dia) mm ( 2 x 1 in)
Sample Volume Required 10 mL (0.33 oz)
Operating Temperature Range 0 °C to 50 °C (32 °F to 122 °F)
Sample Temperature Range 0 °C to 50 °C (32 °F to 122 °F)
Operating Humidity Range 0 to 90% relative humidity,
non-condensing at 30 °C (86 °F)
Power Supply 4 x AAA alkaline batteries
Battery Life > 1200 readings
Enclosure Type ABS plastic
Waterproof Rating: IP67
Regulatory Certification CE, cTUVus, TUV 3-in-1, FCC part 15
Insulation Rating Pollution Degree 2
Weight: Meter: 0.23 kg (0.51 lb)
Meter with case: 2.0 kg (4.4 lb)
Dimensions Meter:
7.9 (W) x 17.2 (L) x 5.0 (H) cm (3.1 x 6.8 x 2.0 in)
Meter & case:
35 (W) x 27 (L) x 15 (H) cm (13.8 x 10.6 x 5.9 in)
28 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 33
Appendix 1 Turbidity Theory
Definition
Why Is It
Important?
Turbidity is defined as an “expression of the optical property that causes light to be scattered and absorbed rather than transmitted in straight lines
ough the sample.” That is, turbidity is the measure of relative sample
thr clarity, not color.
Water with cloudy or opaque appearance will have high turbidity, while water that is clear or translucent will hav value is caused by particles such as silt, clay, microorganisms and organic
By definition, turbidity is not a direct measure of these particles
matter. but rather a measure of how these particles scatter light.
For drinking water applications, a turbidity value may give an indication of presence of bacteria, pathogens or particles that can shelter harmful organisms fr is particularly useful for water treatment plants to ensure cleanliness. For wastewater effluent testing, turbidity is often required to ensure that the r
eleased effluent will not harm the environment.
In industrial processes, turbidity can be part of quality control measure to ensur
om disinfection process. Therefore, turbidity measurement
e efficiency in treatment or manufacturing process.
e low turbidity. High turbidity
Measurement
Principle
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
There are two internationally accepted standard specifications for turbidity measur and the US EPA method 180.1.
Basically the ISO 7027 is a more stringent standard and requires the use of a monochr meter in the following areas: (1) Light source’s wavelength; (2) Light sources’ spectral bandwidth; (3) Measuring angle; (4) Aperture angle in water sample; (5) Distance traversed by incident light and scattered light within the sample and (6) Calibration standar
The AQ3010 turbidity meter follows the ISO 7027 standard whose specification allows for greater reproducibility of the measured values and gr
eater agreement between other measuring instruments.
ement. These are the international standard ISO 7027
omatic light source. It also governs the design of a turbidity
d.
29
Page 34
Figure 15
90°
Lens
Glass Sample Vial
Transmitted
Light
Scattered Light
Lamp
Basic
Nephelometric
Arrangement
for Turbidity
Measurement
Figure 15 shows the AQ3010 waterproof turbidity meter basic optical system. It includes a light source and a detector to monitor the light scattered at 90° with respect to the incident beam.
Nephelometric
Turbidity Units
(NTU)
Indexing a
Vial
Nephelometric Turbidity Units (NTU): Unit of measure used when relating the light scattered by a liquid media to the light scattered by a known concentration of a standard solution. This unit of measure is recognized as a measure of the optical clarity of an aqueous sample. NTU is the accepted unit of measurement for turbidity.
Another unit commonly used to measure turbidity is Formazin Turbidity Unit (FTU). The two units of measure of turbidity are equivalent: 1 NTU = 1 FTU
.
The United States Environmental Protection Agency (US EPA) recommends that vials used for turbidity meter calibration or sample measur
ement be indexed.
To index a sample vial, slowly rotate the vial throughout one complete r
evolution (360°). While rotating the sample vial, observe the display and
locate the position that the vial is in which provides the lowest turbidity
eading. This position is the indexed position of the vial.
r
Mark this position on the vial (not on the cap) against the mark on the meter.
After indexing a vial, make sure the vial will always be placed inside the sample w
30 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
ell in the indexed position.
Page 35
Appendix 2 Guide to Good
Measurement Technique
Turbidity is a very complex analytical measurement that can be affected by many factors. Some are inherent in the instrument’s design such as angle of detection, light beam aperture, incident beam wavelength and color sensitivity of the photocell
However, there are other factors such as stray light, air bubbles and care of vial, which can be prevented through proper care of equipment and accessories, and in the operating pr some points you may want to note.
.
ocedure for measurement. Here are
Maintain
Sample Vials
in Good
Condition
Match
Sample Vials
Sample vials must be meticulously clean and free from significant scratches silicone oil. This is to mask minor imperfections and scratches that may contribute to stray light. Sample vials should be handled only by the top to avoid dirt accumulation (or deposits) and fingerprints that might inter
More information can be found in the Vials – Handling, Cleaning and Care section
Best precision and repeatability of turbidity measurement are achieved using a single, index vials can be used for measurement provided their readings with the same solution ar the specified repeatability and precision of the meter.
Select a few vials. After the sample vials are cleaned, fill them with ultra­low turbidity water. Allow the sample vials to stand and for air bubbles to rise. Polish sample vials with silicone oil and take turbidity measurement at several points while rotating it in the sample well. Find the position where turbidity reading is the lowest and index it for each vial. Whenever these sample vials are used, use the indexed mark to position each vial into the sample w
. It should be treated on the outside with a thin coat of
fere with the light path.
.
ed vial. However, for more convenience, different
e matched. That is, the meter gives identical readings or within
ell. Choose those vials that match the readings.
Not all vials can be matched due to some manufacturing variations.
Note:
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
31
Page 36
Degassing
Air or other trapped gases should be removed before measurement. Degassing is recommended even if no bubbles are visible. There are three methods commonly used for degassing:
Addition of a surfactant: This involves adding a surfactant to the
• water samples to lower the surface tension of the water, thereby r
eleasing trapped gasses.
Application of a partial vacuum: A partial vacuum can be created by using simple syringe or vacuum pump. This is only recommended for ultra-low turbidity measurements.
Use of an ultrasonic bath: This may be effective in severe conditions or in viscous samples, but not recommended for ultra-low measur
Each of the method above has its own advantages and disadvantages. For instance, under certain sample conditions, the use of vacuum pump or ultrasonic bath may actually incr
ements.
ease the presence of gas bubbles.
Timeliness
of Sample
Measurement
Samples should be measured immediately to prevent changes in particle characteristics due to temperatur affect particles by changing their behavior or creating new particles if
ecipitates are formed. Dilution water may dissolve particles or change
pr their characteristics. It is recommended to take samples only when the turbidity meter is ready to be operated. Samples should not be drawn and allowed to sit while the instrument warms up or is being readied.
e and settling. Temperature can
32 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 37
Other
Important
Sampling
Techniques
1. Samples should not be violently shaken or agitated as particles can be broken apart or air may be entrapped into the fluid. Gentle agitation such as swirling the sample vial is advisable to reduce particle settling.
2. Sample vials should be used only with the instruments for which they were intended. Do not mix and match vials. The AQ3010 uses 25 mm sample vials.
3. Perform a visual observation of the sample vial every time a measurement is made. Ensure that there are no visible bubbles in the sample and the vial is clean and free of scratches.
4. Samples entering the turbidity meter should be at the same temperature as the process flow samples. Changes in temperature can cause precipitation of soluble compounds and affect readings.
5. Sample vials should be evaluated with a low turbidity water (after cleaning) to determine if cells remain matched. If the evaluation determines that a cell is corrupted, discard the vial. It is recommended to conduct this evaluation weekly.
6. When in doubt about whether a sample vial is too scratched or stained, throw it away.
Calibration
Techniques
1. Keep the calibration standard plastic bottles and vials tightly capped and ensure that they are not allowed to freeze.
2. Check that the standards have not expired.
3. Make sure that the calibration vials are free of dust, smudges and scratches before use.
4. Conduct the calibration in the same manner each time. Variations in how the calibration is performed could yield inaccurate measur
5. It is very important that the users who perform calibration have been trained to do so. Creating a Standard Operating Procedure (SOP) for the users to read, learn, and practice may help to ensure precision.
ements.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
33
Page 38
Dilution
Techniques
This dilution procedure is necessary only when your turbidity measurement is above 1000 NTU.
1. To measure the turbidity above 1000 NTU, dilute the sample with turbidity-free water.
2. Turbidity-free water can be obtained by filtering deionized water through a < 0.2 μm filter membrane with precision-sized pores.
3. Measure the volume of the sample (V
) before dilution and record the
s
value in mL.
4. Take a known volume (V
) of dilution water and add it to the
d
sample.
5. Pour 10 mL of the diluted sample in a clean vial and measure the turbidity of the diluted sample. Record this value in NTU (T
).
d
6. Calculate the true turbidity (T) of the original sample in NTU using the following formula:
T =
• (Vs + Vd) / V
T
d
s
Example:
Dilute 20 mL of the original sample (with turbidity above 1000 NTU) with 50 mL of dilution water.
Measure the turbidity of the diluted sample.
If the reading is 300 NTU, the turbidity of the original sample is 1050 NTU.
In this example Td = 300 NTU, Vs = 20 mL, Vd = 50 mL,
T = 300 • (20+50) / 20 = 300 • 70/20 = 21000/20 = 1050.
so
34 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 39
Water Analysis Instruments
9 0 0 1 : 2 0 0 8
North America
166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836
www.thermo.com/water
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
257503-001 Rev. B 06-09
Loading...