Thermo Scientific AQ3010 Instruction Manual

Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter
User Guide
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. US patent 6,793,787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
© 2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
The specific
ations, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Table of Content
Chapter 1
Chapter 2
Chapter 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Information . AQ3010 Meter Overview .
Unpacking the Meter and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . 2
isplay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
D Keys and Functions. Battery Installation .
Turbidity Calibration .
Calibration Overview . Indexing . Calibration Standards . Preparing and Filling the Calibration Standard Vials . Calibration Check. Refilling the Calibration Standard Vials with Fresh Standard. Calibration Procedure . Restoring Factory Calibration .
Turbidity Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
General Information . Turbidity Measurement Procedure .
Preparation of Sample Vial. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
M
ingle-Shot or Continuous Measurement. . . . . . . . . . . . . . . . . . . . . .19
S
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
easurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
. . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
. . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Chapter 4
Chapter 5
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Routine Maintenance .
Vials – Handling, Cleaning and Care .
Customer Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Troubleshooting Guide Assistance . Warranty . Notice of Compliance . WEEE Compliance. Declaration of Conformity . Ordering Information . Specifications.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
. . . . . . . . . . . . . . . . . . . . . . . . . 21
I
Appendix 1
Turbidity Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Definition . Why Is It Important . Measurement Principle . Nephelometric Turbidity Units (NTU) . Indexing a Vial .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . 30
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Appendix 2
Guide to Good Measurement Technique .
Maintain Sample Vials in Good Condition. Match Sample Vials. Degassing . Timeliness of Sample. Other Important Sampling Techniques . Calibration Techniques . Dilution Techniques .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
. . . . . . . . . . . . . . . . . . . . . . . . 33
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
. . . . . . . . . . . . . . . . . . . . . . . . . 31
. . . . . . . . . . . . . . . . . . . . . 31
II Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Chapter 1 Introduction
General
Information
This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion AQUAfast® AQ3010 turbidity meter. help you operate the waterproof AQ3010 turbidity meter and serves as a handy r
Follow the basic instructions contained in this user guide during the operation, car provided by this equipment may be impaired if it is used in a manner not described in this user guide read this user guide prior to working with this meter.
The manufacturer will not accept any responsibility for damage or malfunction to the meter caused b
The information presented in this user guide is subject to change without notice as improvements are made, and does not represent a commitment on the par
Note: The manufacturer reserves the right to make improvements in design, constr
eference guide.
This user guide provides both a step by step guide to
e and maintenance of the meter. The safety protections
. It is recommended that all operators should
y improper use of the instrument.
t of the manufacturer.
uction and appearance of products without notice.
AQ3010 Meter
Overview
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Thank you for selecting the waterproof portable AQ3010 turbidity meter.
The AQ3010 meter allows you to measure turbidity of an aqueous sample in the field. This instrument operates on the nephelometric principle of turbidity measurement and is designed to meet the criteria specified in ISO 7027 and DIN 27027 standar Principle section
.
ds. See the Measurement
1
Unpacking
the Meter and
Accessories
The list below indicates the items that you should find in your AQ3010 turbidity meter shipment.
Item Quantity
AQ3010 portable turbidity meter with 4 AAA batteries 1
User guide 1
Instrument carrying case 1
Light shield cover 1
Calibration set with 0.02, 20.0,100 and 800 NTU standards 1
Empty calibration vials 4
Empty sample vials 3
Silicone oil 1
Lint free cloth 1
Vial cleaning brush 1
Remove the AQ3010 turbidity meter from the carrying case. Carefully inspect all items to ensure that no visible damage has occurred during shipment
. If the items you received do not match your order, please
contact your dealer immediately.
Warning: Extra care should be taken when unpacking, opening and handling the calibration standar
ds and sample vials. Surface scratches or finger smudges on the vial surface may cause measurement errors. Handle these items by their caps only.
The batteries provided with the meter package are to be installed prior to use
. Refer to the Battery Installation section
Figure 1 sho
ws the meter. The three main components of the instrument
.
are the sample well, the display and the keypad. The following sections describe the functionality of the display and the keypad. The proper use of the instrument and the sample well are discussed in later sections.
2 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Figure 1
Sample Well
Alignment Mark
Liquid Crystal Display
Keypad
Sample Vial
AQ3010 Turbidity
Meter Components
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
3
Display
CAL 13 CAL 24
mg/l NTU ppm %pH
TAvg
14-segment Liquid Crystal Display
Battery Indicator
Calibration Standards
Units of Measurement
Figure 2
Customized LCD with
All Annunciators
All the LCD segments and annunciators that can appear on the display are shown in Figure 2.
The display is used for reporting the turbidity reading and to provide guidance for the operation of the instrument. In addition, the display has several other annunciators that are used to communicate err
or messages and provide user guidance.
4 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
read
enter
cal
Turbidity Meter Chlorine Colorimeter
Keys and
CAL
CAL
CAL
READ
ENTER
CAL
CAL
Functions
The keypad has five keys: ON/OFF, CAL, , and READ/ENTER.
Key Functions
Powers on and off the meter. The meter automatically shuts off 20 minutes after last key press to conserve the battery life.
Initiates the calibration mode of the meter. When pressed, the meter is set to accept the first calibration standard.
It is also used to exit the calibration mode if the user does not want to follow the complete calibration procedure.
When in measurement mode, the READ/ENTER key is used to perform a measurement.
Single-shot measurements: When the key is pressed and released immediately (a quick key stroke of less than 0.3 seconds), the display will blink [-Rd-] ten times and display the measured value.
Continuous measurements: If the READ/ENTER key is pressed and held, the instrument will perform a continuous measurement during which the display is updated every 2 seconds. This can be used for indexing vials. (This function is not available in calibration mode). After the READ/ENTER key is released, the instrument will automatically perform a single-shot measurement.
Active only during calibration mode. These scroll keys are used to select the next calibration point.
Note: After the successful calibration of one point, the meter automatically selects the next calibration point, or automatically exits the calibration mode after the fourth calibration point.
Figure 3
Keypad
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
5
Battery
Installation
Four AAA batteries are included in the carrying case.
Note: Use caution when loosening the screws, as some of the screws may be tightly seated
1. Use a Phillips screwdriver to loosen the four screws holding the
battery cover. See Figure 4. attached to the battery cover.
2. Remove the battery cover.
.
These captive screws will remain
Figure 4
Battery Installation
3. Insert the batteries. Follow the diagram inside the cover for
ect polarity.
corr
4. Replace the battery cover in its original position and tighten the four
screws removed earlier to secure the battery cover to the meter.
5. The meter is now ready for use.
Note: D
ispose of used batteries in accordance with your local regulations.
6 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Chapter 2 Turbidity Calibration
Calibration
Overview
Indexing
The AQ3010 turbidity meter was calibrated and tested prior to leaving the factory. The meter kit includes a set of prepared primary standards that can be used to verify the meter calibration daily and to recalibrate the
, if required. Before measuring samples for the first time, verify the
meter meter calibration with the prepared primary standards.
The SDVB primary standards in the meter kit are in plastic bottles for impr
oved precision and shelf life. Four vials are included and labeled for
each standard level.
The instrument includes a light shield cover to keep stray light from changing the calibration and sample measur environments, the light shield is not required. The vial cap keeps stray light from entering the sample well, saving you many steps and reducing analysis time
Due to the high quality of the glass vials provided, indexing is not r
equired. You only need to align the mark on the vial with the mark on the meter. However, in order to achieve a better precision of the measurement, you can proceed with indexing of the vials. See the Indexing a Vial section for mor
.
e information.
ement readings. In many
Calibration
Standards
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
We recommend that you use the following materials during calibration to achieve the precision stated in this user guide:
CAL 1: 800 NTU calibration standard
CAL 2: 100 NTU calibration standard
CAL 3: 20.0 NTU calibration standard
CAL 4: 0.02 NTU calibration standard
It is well known that diluted Formazin is unstable. If you choose to use Formazin to calibrate the instrument, ensure that you are using a fresh stock suspension of Formazin to achieve the precision quoted for the instr
ument. It is very difficult to accurately pipet Formazin for low level NTU standards, due to the non-homogeneous nature of Formazin and because these standar to use good laboratory technique and accurate Class A pipets and refer to ASTM Method 6855-03 for low level standard vial preparation when using F
ormazin.
ds have very limited stability at low levels. Be sure
7
The four bottles of calibration standards supplied with your meter are more stable than Formazin, do not need to be shaken and have a shelf life of 12 months the instrument, review the expiration date (indicated on the bottle label) to ensur
Note: It is important that the calibration standards are not violently shaken or agitated because air entrapment in the fluid introduces an error factor during calibration which subsequently will lead to an inaccurate measurement. Also, do not store in freezing temperatures which causes irreversible shrinkage of the standards’ particles thus resulting to inaccurate calibration and measur
. If you use the supplied calibration standards to calibrate
e that the standards have not expired.
ement.
Preparing and
Filling the
Calibration
Standard
Vials
Before using the meter for the first time, rinse the inside of the vial with a small amount of the same NTU standar
1. Gently pour about 5 mL of the standard into the vial.
2. Cap and swirl the vial so the inside of the vial is fully rinsed with the standar
3. Pour the standard out of the vial.
4. Repeat steps 1 through 3.
5. Gently pour the standard into the vial up to the mark on the vial and cap the vial.
6. Repeat this rinse and filling procedure for the other three vials.
Note: The vial rinsing is most critical for precision with the lower NTU standar
It is recommended to use the prepared calibration standard vials daily as check standar
d.
ds.
ds prior to the measurement of samples.
d that will be used to fill that vial.
8 Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Loading...
+ 27 hidden pages