Thermo Scientific Orion
AQUAfast AQ3010 Turbidity Meter
User Guide
Page 2
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. US patent 6,793,787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink,
ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ,
AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product
Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir
Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture
Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY,
Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of
Thermo Fisher.
Maintain Sample Vials in Good Condition.
Match Sample Vials.
Degassing .
Timeliness of Sample.
Other Important Sampling Techniques .
Calibration Techniques .
Dilution Techniques .
IIThermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 5
Chapter 1 Introduction
General
Information
This user guide contains information on the preparation, operation
and maintenance of the Thermo Scientific Orion AQUAfast® AQ3010
turbidity meter.
help you operate the waterproof AQ3010 turbidity meter and serves as a
handy r
Follow the basic instructions contained in this user guide during the
operation, car
provided by this equipment may be impaired if it is used in a manner not
described in this user guide
read this user guide prior to working with this meter.
The manufacturer will not accept any responsibility for damage or
malfunction to the meter caused b
The information presented in this user guide is subject to change without
notice as improvements are made, and does not represent a commitment
on the par
Note: The manufacturer reserves the right to make improvements in
design, constr
eference guide.
This user guide provides both a step by step guide to
e and maintenance of the meter. The safety protections
. It is recommended that all operators should
y improper use of the instrument.
t of the manufacturer.
uction and appearance of products without notice.
AQ3010 Meter
Overview
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Thank you for selecting the waterproof portable AQ3010 turbidity
meter.
The AQ3010 meter allows you to measure turbidity of an aqueous
sample in the field. This instrument operates on the nephelometric
principle of turbidity measurement and is designed to meet the criteria
specified in ISO 7027 and DIN 27027 standar
Principle section
.
ds. See the Measurement
1
Page 6
Unpacking
the Meter and
Accessories
The list below indicates the items that you should find in your AQ3010
turbidity meter shipment.
ItemQuantity
AQ3010 portable turbidity meter with 4 AAA batteries1
User guide1
Instrument carrying case1
Light shield cover1
Calibration set with 0.02, 20.0,100 and 800 NTU standards1
Empty calibration vials4
Empty sample vials3
Silicone oil1
Lint free cloth1
Vial cleaning brush1
Remove the AQ3010 turbidity meter from the carrying case. Carefully
inspect all items to ensure that no visible damage has occurred during
shipment
. If the items you received do not match your order, please
contact your dealer immediately.
Warning: Extra care should be taken when unpacking, opening and
handling the calibration standar
ds and sample vials. Surface scratches
or finger smudges on the vial surface may cause measurement errors.
Handle these items by their caps only.
The batteries provided with the meter package are to be installed prior to
use
. Refer to the Battery Installation section
Figure 1 sho
ws the meter. The three main components of the instrument
.
are the sample well, the display and the keypad. The following sections
describe the functionality of the display and the keypad. The proper use
of the instrument and the sample well are discussed in later sections.
2Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 7
Figure 1
Sample Well
Alignment Mark
Liquid Crystal Display
Keypad
Sample Vial
AQ3010 Turbidity
Meter Components
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
3
Page 8
Display
CAL 13 CAL 24
mg/l
NTU
ppm
%pH
TAvg
14-segment Liquid Crystal Display
Battery Indicator
Calibration
Standards
Units of
Measurement
Figure 2
Customized LCD with
All Annunciators
All the LCD segments and annunciators that can appear on the display
are shown in Figure 2.
The display is used for reporting the turbidity
reading and to provide guidance for the operation of the instrument.
In addition, the display has several other annunciators that are used to
communicate err
or messages and provide user guidance.
4Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 9
read
enter
cal
Turbidity Meter Chlorine Colorimeter
Keys and
CAL
CAL
CAL
READ
ENTER
CAL
CAL
Functions
The keypad has five keys: ON/OFF, CAL, ▲, ▼ and READ/ENTER.
KeyFunctions
Powers on and off the meter. The meter automatically
shuts off 20 minutes after last key press to conserve the
battery life.
Initiates the calibration mode of the meter. When pressed,
the meter is set to accept the first calibration standard.
It is also used to exit the calibration mode if the user does
not want to follow the complete calibration procedure.
When in measurement mode, the READ/ENTER key is used
to perform a measurement.
Single-shot measurements: When the key is pressed and
released immediately (a quick key stroke of less than 0.3
seconds), the display will blink [-Rd-] ten times and display
the measured value.
Continuous measurements: If the READ/ENTER key is
pressed and held, the instrument will perform a continuous
measurement during which the display is updated every
2 seconds. This can be used for indexing vials. (This
function is not available in calibration mode). After
the READ/ENTER key is released, the instrument will
automatically perform a single-shot measurement.
Active only during calibration mode. These scroll keys are
used to select the next calibration point.
Note: After the successful calibration of one point, the
meter automatically selects the next calibration point, or
automatically exits the calibration mode after the fourth
calibration point.
Figure 3
Keypad
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
5
Page 10
Battery
Installation
Four AAA batteries are included in the carrying case.
Note: Use caution when loosening the screws, as some of the screws may
be tightly seated
1. Use a Phillips screwdriver to loosen the four screws holding the
battery cover. See Figure 4.
attached to the battery cover.
2. Remove the battery cover.
.
These captive screws will remain
Figure 4
Battery Installation
3. Insert the batteries. Follow the diagram inside the cover for
ect polarity.
corr
4. Replace the battery cover in its original position and tighten the four
screws removed earlier to secure the battery cover to the meter.
5. The meter is now ready for use.
Note: D
ispose of used batteries in accordance with your local regulations.
6Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 11
Chapter 2 Turbidity Calibration
Calibration
Overview
Indexing
The AQ3010 turbidity meter was calibrated and tested prior to leaving
the factory. The meter kit includes a set of prepared primary standards
that can be used to verify the meter calibration daily and to recalibrate the
, if required. Before measuring samples for the first time, verify the
meter
meter calibration with the prepared primary standards.
The SDVB primary standards in the meter kit are in plastic bottles for
impr
oved precision and shelf life. Four vials are included and labeled for
each standard level.
The instrument includes a light shield cover to keep stray light from
changing the calibration and sample measur
environments, the light shield is not required. The vial cap keeps stray
light from entering the sample well, saving you many steps and reducing
analysis time
Due to the high quality of the glass vials provided, indexing is not
r
equired. You only need to align the mark on the vial with the mark
on the meter. However, in order to achieve a better precision of the
measurement, you can proceed with indexing of the vials. See the
Indexing a Vial section for mor
.
e information.
ement readings. In many
Calibration
Standards
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
We recommend that you use the following materials during calibration to
achieve the precision stated in this user guide:
•
CAL 1: 800 NTU calibration standard
• CAL 2: 100 NTU calibration standard
• CAL 3: 20.0 NTU calibration standard
• CAL 4: 0.02 NTU calibration standard
It is well known that diluted Formazin is unstable. If you choose to use
Formazin to calibrate the instrument, ensure that you are using a fresh
stock suspension of Formazin to achieve the precision quoted for the
instr
ument. It is very difficult to accurately pipet Formazin for low level
NTU standards, due to the non-homogeneous nature of Formazin and
because these standar
to use good laboratory technique and accurate Class A pipets and refer
to ASTM Method 6855-03 for low level standard vial preparation when
using F
ormazin.
ds have very limited stability at low levels. Be sure
7
Page 12
The four bottles of calibration standards supplied with your meter are
more stable than Formazin, do not need to be shaken and have a shelf life
of 12 months
the instrument, review the expiration date (indicated on the bottle label)
to ensur
Note: It is important that the calibration standards are not violently
shaken or agitated because air entrapment in the fluid introduces an error
factor during calibration which subsequently will lead to an inaccurate
measurement. Also, do not store in freezing temperatures which
causes irreversible shrinkage of the standards’ particles thus resulting to
inaccurate calibration and measur
. If you use the supplied calibration standards to calibrate
e that the standards have not expired.
ement.
Preparing and
Filling the
Calibration
Standard
Vials
Before using the meter for the first time, rinse the inside of the vial with a
small amount of the same NTU standar
1. Gently pour about 5 mL of the standard into the vial.
2. Cap and swirl the vial so the inside of the vial is fully rinsed with
the standar
3. Pour the standard out of the vial.
4. Repeat steps 1 through 3.
5. Gently pour the standard into the vial up to the mark on the vial and
cap the vial.
6. Repeat this rinse and filling procedure for the other three vials.
Note: The vial rinsing is most critical for precision with the lower
NTU standar
It is recommended to use the prepared calibration standard vials daily as
check standar
d.
ds.
ds prior to the measurement of samples.
d that will be used to fill that vial.
8Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 13
“-Rd-” flashes about
10 times
CAL STD
100
NTU
Place standard
vial into the
sample well
NTU
READ
ENTER
Measurement Mode
Press key and
release immediately
100
Calibration
1. Place the AQ3010 turbidity meter on a flat and level surface.
Check
2. Insert the CAL 1 standard (800 NTU) into the sample well, aligning
the mark on the vial with the alignment mark on the meter.
See Figure 11.
3. Press down the vial until it slides fully into the instrument.
4. Cover the vial using the light shield cover, if necessary. For many
environments this is not required. The vial cap seals the sample well.
5. Press the
ON/OFF key to turn on the meter.
The meter goes to
measurement mode after the self test power-up sequence.
6. Press the
READ/ENTER key.
7. The meter will display the results. Record the reading.
8. Repeat the calibration check for CAL 2, CAL 3 and CAL 4
calibration standards.
9. If the displayed results are within 10% of the nominal NTU value of
the standard or the precision criteria required by your method, the
calibration check passed and the meter is no
w ready for measurement.
Figure 5
Calibration Check
Sequence
Note: If a standard no longer reads within 10% of the nominal NTU
v
alue for the standard, the standard has most likely degraded. The
lower level NTU standards will be less stable when stored in glass than
the higher NTU standar
ds. The lower level standards may need to be
replaced periodically, especially if the value falls outside of the tolerance
. See the Refilling the Calibration Standard Vials with Fresh
range
Standard section
.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
9
Page 14
Refilling the
1. Pour the old standard out of the vial.
Calibration
Standard
Vials with
Fresh
Standard
2. Wash the vial with laboratory glassware detergent and water. Rinse
the vial with turbidity-free water to remove all laboratory detergent
fr
om the vial.
3. Rinse the vial with about 5 mL of the standard from the plastic bottle
with the matching NTU value.
4. Cap and swirl the vial so the inside of the vial is fully rinsed with the
fresh standard.
5. Pour the standard out of the vial.
6. Pour the fresh standard into the vial up to the mark on the vial and
cap the vial.
7. Perform the procedure in the
containing the fresh standard.
If the fresh standard does not read with the 10% range of the nominal
value, or to the tolerance of your method, follow the steps in the
Calibration Procedure section.
Calibration Check sectionwith the vial
10Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 15
Calibration
Procedure
Note: Before switching on the AQ3010 turbidity meter, a sample vial
must be placed in the sample well. You can use any of the calibration
standards for this purpose. If a sample vial is not placed in the sample
well, the turbidity meter will display the Err 8 error message. To clear the
error message, power off the meter., insert a sample vial into the sample
well and power on the meter.
1. Place the AQ3010 turbidity meter on a flat and level surface.
2. Press the
measurement mode after the self test power-up sequence.
3. Select the calibration function of the instrument by pressing the
CAL key once
the meter will prompt for the first calibration standard by displaying
“CAL 1 800 NTU”
4. Insert the CAL 1 standard (800 NTU) into the sample well, aligning
the mark on the vial with the alignment mark on the meter.
See Figure 11.
5. Press down the vial until it slides fully into the instrument.
6. Cover the vial using the light shield cover, if necessary. For many
environments this is not required. The vial cap seals the sample well.
7. Press the
8. The “CAL 1 800 NTU” annunciator will blink for about 12 seconds
while the instrument performs the calibration of CAL 1 point.
When the instrument has completed calibration for this point, it will
prompt you to insert the next calibration standard into the sample
well by displaying “CAL 2 100 NTU”.
ON/OFF key to turn on the meter.
. The “CAL” annunciator will blink momentarily and
.
READ/ENTER key.
The meter goes to
9. Repeat steps 4 through 8 for the CAL 2, CAL 3 and CAL 4
calibration standards.
10. After you successfully calibrate the CAL 4 standard (0.02 NTU), the
meter will display “STbY”.
11. The meter is now ready for measurement.
Figure 6 sho
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
ws the complete calibration sequence.
11
Page 16
Calibration
CAL 1
800 NTU
NTU
CA L 1
NTU
NTU
CAL 3
CA L 3
NTU
CAL 2
NTU
CA L 2
RE AD
EN TER
NTU
CA L 1
CA L
Pla ce t he C AL 1
sta ndar d int o the
sam ple wel l
Pla ce t he C AL 3
sta ndar d int o the
sam ple wel l
Fla shes abo ut 5 time s
Fla shes ab out 1 0 tim es
Fl ash es ab out 10 ti me s
Fla she s abo ut 1 0 tim es
Fla she s abo ut 1 0 tim es
CAL 4
CAL 3
20.0 NTU
0.02 NTU
NTU
NTU
CA L 4
CA L 4
RE AD
EN TER
RE AD
EN TER
Pla ce th e C AL 2
sta ndar d int o the
sa mple wel l
Pla ce t he C AL 4
sta ndar d int o the
sam ple wel l
Measurement Mode
CAL 2
100 NTU
RE AD
EN TER
Measurement Mode
Calibration saved
Notes
• To exit the calibration mode, press the CAL key at the end of any
step. The meter accepts only the values calibrated prior to exiting.
Figure 6
Calibration
Sequence
• You can skip a calibration point by pressing the
▲ or ▼ key and move
on to the next calibration point.
• After a successful calibration of one point, the meter automatically
selects the next calibration point. The meter automatically exits
calibration mode after the fourth calibration point.
• If an error occurs during calibration, the display shows an error
message. The meter aborts the calibration and returns to the
measurement mode without saving the last calibration value.
• For an error list, refer to the
Troubleshooting Guide section
.
12Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 17
Restoring
Factory
Calibration
The AQ3010 turbidity meter allows you to reset the meter back to the
factory default calibration values. This feature is extremely useful when
there are errors in calibration or when you have new calibration standards.
1. With the meter switched off, press and hold the
▲ key.
2. Switch on the meter by pressing the ON/OFF key.
when “URST” (User-Reset) appears on the display.
3. The display will show “URST” for about 2 seconds and then “No”.
4. Use the
• YES = Restore the meter back to factory calibration values
• No = Retain the last calibrated values
5. Press the
performs the reset if “YES” is selected.
6. The display will flash “-Rd-” about ten times and the meter will go to
the measurement mode.
Figure 7 sho
▲ or ▼ key to scroll between “YES” or “No”.
READ/ENTER key to confirm y
ws the sequence for restoring factory calibration values.
our selection. The meter
Release the ▲ key
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
13
Page 18
Figure 7
READ
ENTER
READ
ENTER
“-R d-” f lash es a bo ut 10 tim es
Se lect YE S an d
to res tore fact ory
ca libra tion
the n
pr ess
ke y
RE AD /ENT ER
Se lect No
to ret ain las t
ca lib ratio n v alue s
wi tho ut ret riev ing
fac tory ca libra tion
an d th en
pr ess
ke y
RE AD /EN TE R
Instrument Off
Measurement Mode
Pr es s & ho ld k ey and
sw itch on t he me ter u sing
ke yON /OF F
Re lea se key wh en y ou
se e ‘U RST ’ in the di spla y
Pr es s or key to
se lec t o r YE S No
Restoring Factory
Calibration
14Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 19
Chapter 3 Turbidity Measurement
CA L 13 C AL 24
mg /l
NTU
ppm
%p H
TAvg
NTU
"-R d-” f las hes ab out 10
tim es ind icat ing tha t the
me ter g oes t o
m eas ure men t mo de.
A re adin g is val id o nly i f yo u h ave pla ced a
via l or a ny o f the cal ibra tion stan dar ds
pro vide d in to th e sa mp le we ll.Oth erwi se
the rea ding has no m ean ing o r an err or
me ssag e " Err 8" app ears .
Instrument Off
Pla ce a sam ple vial or a ny of th e
ca libra tion s tand ards pro vid ed i nto
the sam ple w ell of th e m ete r, be fore
sw itch on.
X.x
“X .X” is s oftw are r evisi on
nu mb er
Place light shield cap over vial.
General
Information
Figure 8
Power-up Sequence
The AQ3010 waterproof turbidity meter allows you to measure the
turbidity of a grab sample. The turbidity is reported in Nephelometric
Turbidity Units (NTU). Readings above 1000 NTU are outside the
range of this instrument.
Note: Before switching on the AQ3010 turbidity meter, a sample vial
must be placed in the sample w
ell. You can use any of the calibration
standards for this purpose.
The instrument includes a light shield cover to keep stray light from
changing the calibration and sample measur
ement readings. In many
environments, the light shield is not required. The vial cap keeps stray
light from entering the sample well, saving you many steps and reducing
analysis time
.
When the ON/OFF key is pressed to switch on the meter, it goes through
the power-up sequence as shown in Figure 8.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
15
Page 20
Turbidity
Measurement
Procedure
An accurate turbidity measurement depends on good measurement
techniques. Factors such as clean sample vials, positioning of vial in the
sample well, covering the vial with the light shield cover (if needed),
meter calibration, handling of meter and others have to be taken into
consideration
Appendix 2 Guide to Good Measurement Technique sections for more
information
. Refer to the Vials – Handling, Cleaning and Care and
.
Preparation of
Sample Vial
1. Obtain a clean and dry sample vial. See Figure 9.
handle the sample vial by the top cap.
2. Rinse the vial with approximately 10 mL of the sample water, capping
the vial with the black screw cap and gently inverting it several times.
Discard the used sample and repeat the rinsing procedure two times.
4. Fill the rinsed vial with the remaining portion (approximately 10 mL)
of the grab sample up to the mark indicated in the vial. Cap the vial
with the supplied black screw cap.
5. Wipe the vial with the soft, lint-free cloth supplied. Ensure that the
outside of the vial is dry, clean and free from smudges.
6. Apply a thin film of silicone oil (supplied) on the sample vial. See
Figure 10.
7. Wipe the vial with a soft cloth to obtain an even distribution over
the entire vial surface. The purpose of oiling the vial is to fill small
scratches and to mask the imperfection in the glass. Do not apply
large quantity of oil as this may collect dirt and dust.
Take care to
8. The sample vial is now ready to be inserted into the sample well of
the meter for measurement.
Figure 9
Sample Vial
16Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 21
Figure 10
Align the alignment
mark on the sample
vial with the
alignment mark on
the AQ3010 turbidity
meter and push until
it is fully snapped in
Applying the Silicon Oil
Measurement
Procedure
Figure 11
Aligning the Sample
Vial Alignment Mark
with the Meter
Alignment Mark
1. Place the AQ3010 turbidity meter on a flat and level surface.
Note: Do not hold the meter in your hand while operating it, as this may
cause inaccurate measur
ements.
2. Place the sample vial inside the sample well and align the vial’s
alignment mark with the meter’s alignment mark. See Figure 11.
. Push the vial down until it slides fully in.
3
4. Cover the vial using the light shield cover, if necessary. For many
environments this is not required. The vial cap seals the sample well.
5. Turn on the meter by pressing the
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
ON/OFF key.
17
Page 22
6. After the power-up sequence, the meter will go to the measurement
“-Rd-” flashes about
10 times
SAMPLE
Unknown
NTU
Place sample
vial into the
sample well
NTU
REA D
EN TE R
Measurement Mode
and cover with
the light shield cover,
if needed
mode and the display will blink “-Rd-” about ten times. See
Figure 12.
Figure 12
Reading a
Turbidity Value
. The measured reading will appear in the display.
7
8. If necessary, place the second sample vial into the sample well. Align
the vial alignment mark with the meter alignment mark.
Measurement Notes
9. Press the
READ/ENTER key.
The display will blink “-Rd-” several
times and measured reading will appear.
10. Repeat this procedure for all of your samples.
• Never pour liquid directly into the sample well of the instrument.
Always use a 25 mm vial. The instrument will only accurately
measure the turbidity of a sample when 25 mm vials sealed with the
black caps ar
e used. The black cap serves to both seal the vial and
block stray light.
• Never attempt to clean the sample well. The optics may be damaged.
• For battery conservation, the instrument automatically powers off 20
minutes after the last key pressed.
18Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 23
Single-Shot
“-Rd-” flashes about
10 times
SAMPLE
Unknown
NTU
Place sample
vial into the
sample well
NTU
R EA D
EN TER
Measurement Mode
Press key and
release immediately
and cover with
the light shield
cap
or Continuous
Measurement
You can use the AQ3010 turbidity meter to take a single reading or
perform continuous measurement. The latter is only used for indexing
the vials. See the Indexing a Vial section for mor
e information.
For Single-Shot
Measurement
Figure 13
Single-Shot
Measurement
1. Make sure the meter is sitting on a flat and level surface and is in
measurement mode. The display will show the last measured value or
“STbY” after exiting calibration mode.
2. Place the sample vial in the sample well.
3. Cover the vial using the light shield cover, if necessary. For many
environments this is not required. The vial cap seals the sample well.
4. Press the
0
.3 seconds). See Figure 13.
READ/ENTER key and immediately release it (less than
5. The display will blink “-Rd-” about ten times and then display the
measured value.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
19
Page 24
For Continuous
Reading flashes and
updates every 2 seconds
SAMPLE
Unknown
NTU
Place sample
vial into the
sample well
NTU
R EA D
EN TER
Measurement Mode
Press key and hold
and cover with
the light shield
cap
Measurement
1. Make sure the meter is sitting on a flat and level surface and is in the
measurement mode. The display will show the last measured value or
“STbY”.
2. Place the sample vial in the sample well.
3. Cover the vial using the light shield cover, if necessary. For many
environments this is not required. The vial cap seals the sample well.
Figure 14
Continuous
Measurement
4. Press the
READ/ENTER key and hold it
. See Figure 14.
5. Wait for the reading to stabilize before rotating the sample vial. You
can rotate the sample vial for indexing purpose. See the
Vial section for mor
e information.
Indexing a
Note: During continuous measurement, the display is updated every
2 seconds
. The displayed reading may not be the actual turbidity
value. For accurate measurement, use the single-shot measurement.
6. Once you release the
per
forms a single-shot measurement.
READ/ENTER key, the meter automatically
20Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Continuous
Measurement
Notes
• After a measurement is completed and the display is updated, a 4
second recovery time occurs before the meter can perform another
function. If any key is pressed during this time, the meter will
perform the corresponding action at the end of the recovery period.
• When performing a measurement, if the meter detects stray light
exceeding the amount equivalent to 0.02 NTU in the low range, the
measurement is immediately aborted and an error message [ERR 8] is
ed. Ensure that the vial sits properly into the sample well and
display
the light shield cover is on. Press the ENTER/READ key to retake
the measurement.
Page 25
Chapter 4 Routine Maintenance
The supplied carrying case is optimal for protecting the instrument. If
you do not plan on leaving the instrument in the supplied carrying case,
when it not in use ensure that the instrument has been turned off and
that a clean sample vial fitted with a black cap has been placed in the
sample w
will be able to settle on the optics of the instrument. Keep the light shield
cover with the meter.
ell. This will ensure that a minimal amount of dust and debris
Vials –
Handling,
Cleaning
and Care
Proper measurement of the turbidity of a sample requires the use of a
vial that is fr
Therefore, sample vials must be handled with absolute care to avoid
contamination or damage, which might change the optical characteristics
of the glass
vial or inside the sample well can cause stray light interference leading to
inaccurate r
Cleaning the vial is accomplished by washing the interior and exterior of
the vial in a detergent solution
thoroughly eight to ten times with clean distilled or deionized water to
eliminate the possibility of detergent buildup and str
Vials can also be acid washed periodically and coated with a special
silicone oil to fill small scratches and mask the imper
Since the silicone oil required for this application should have the same
refractive characteristics as glass, it is recommended that only the Thermo
Scientific O
Care should be taken not to apply excessive oil that could attract dirt or
contaminate the sample w
to the vial, the excess oil should be removed with a lint-free cloth. The
result should be a sample vial surface with a dry appearance, but with all
imper
ee of marks, smudges, scratches and any bacterial growth.
. Scratches, fingerprints and water droplets on the sample
eadings.
. Once cleaned, the vial should be rinsed
eaking.
fections in the glass.
rion AQUAfast® silicone oil, Cat. No. AC3SIL, be used.
ell of the meter. Once the oil has been applied
fections filled with oil.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
21
Page 26
Sample vials should always be handled from the top or by the cap to
avoid fingerprints or smudges. After a vial has been filled with a sample
and capped, the outside surface should be wiped with a clean, lint-
ee absorbent cloth until it is dry. Cleaned and dried vials should be
fr
stored with the black caps on. The vials can be stored in the carrying
case. During normal operation, you may use any typical laboratory glass
cleaner along with a lint free cloth or tissue to clean the outside of the
vials
.
Condensation may appear on the vial when your sample is very cold
and the r
elative air humidity is high. When this happens, the turbidity
that you read may be higher than the actual turbidity due to the light
ed by the condensate on the vial. If this occurs, you can alleviate
scatter
the problem by either coating the vial with an anti-fogging agent or by
running warm water over the vial for a short period of time to warm the
sample prior to measur
ement.
22Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 27
Chapter 5 Customer Services
Troubleshooting
Guide
The AQ3010 turbidity meter routinely performs self-diagnostics and will
automatically generate messages to provide you with specific diagnostic
information
. These messages are for your use and do not indicate a
reduction in the performance of the instrument or a failure of any
component in the instr
LCD
Message
ERR 1
ERR 2
ERR 3
DescriptionCorrective Actions
Calibration error – the meter is
unable to recognize the 800 NTU
calibration standard.
Calibration error – the meter is
unable to recognize the 100 NTU
calibration standard.
Calibration error – the meter is
unable to recognize the 20.0 NTU
calibration standard.
ument, unless otherwise stated in this list.
Use the correct 800 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
Use the correct 100 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
Use the correct 20.0 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
Calibration error – the meter is
unable to recognize the 0.02 NTU
ERR 4
ERR 5
ERR 6
ERR 7
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
calibration standard.
Calibration error – there is not
sufficient signal to achieve the
appropriate resolution in the 0 to
1000 NTU range.
General calibration failure – there
is not sufficient signal to achieve
the appropriate resolution in the 0
to 100 NTU range.
General calibration failure – there
is not sufficient signal to achieve
the appropriate resolution in the 0
to 20 NTU range.
Use the correct 0.02 NTU
calibration standard. Pour fresh
standard into a clean vial and retry
the calibration. Verify that the
standard has not expired.
Pour fresh standard into a clean
vial and retry the calibration with
all four standards. Verify that the
standards have not expired.
Pour fresh standard into a clean
vial and retry the calibration with
all four standards. Verify that the
standards have not expired.
Pour fresh standard into a clean
vial and retry the calibration with
all four standards. Verify that the
standards have not expired.
23
Page 28
LCD
Message
ERR 8
ERR 9Lamp failure.Contact Technical Support.
DescriptionCorrective Actions
Excessive stray light detected.Verify that the vial is fully in the
sample well and light shield cover
is properly placed over the vial.
Attempting a measurement when
low battery indicator is on.
Replace the batteries.
Assistance
Warranty
Or
Turbidity value is above the
measurement range (>1000 NTU).
Low battery indication – the
batteries need to be replaced.
Dilute sample, see the Dilution
section.
Replace the batteries.
Note: If an error message appears, take the
appropriate corrective action and re-do the
ed procedure. If the problem persists,
desir
contact Technical Support or your dealer.
After troubleshooting all components of your measurement system,
contact
Technical Support. Within the United States call 1.800.225.1480
and outside the United States call 978.232.6000 or fax 978.232.6031.
In Europe, the Middle East and Africa, contact your local authorized
For the most current contact information, visit
dealer.
thermo.com/contactwater.
www.
For the most current warranty information, visit www.
thermo.com/water.
The AQ3010 meter is supplied with a 3 year warranty from
manufacturing defects and calibration standar
ds for 6 months. If
repair or adjustment is necessary and has not been the result of abuse or
misuse within the designated period, please contact Technical Support
our dealer for an RMT number. A repair technician will determine
or y
if the product problem is due to deviations or customer misuse. Out of
warranty products will be repaired on a charged basis.
Exclusions
The instrument warranty shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
•
• Unauthorized modification or misuse
• Operation outside of the environment specifications of the products
Waterproof Seal: Opening the instrument enclosure (excluding the
y compartment) may void the warranty.
batter
24Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 29
Notice of
Compliance
Warning: This meter may radiate radio frequency energy and if
not installed and used properly, that is in strict accordance with
the manufacturer’s instructions, may cause interference to radio
communications
for a Class A computing device pursuant to Subpart J of Part 15 of FCC
Rules, which are designed to provide reasonable protection against such
ference in a commercial environment. Operation of the meter in a
inter
residential area may cause interference in which case the user at his own
expense will be required to take whatever measures may be required to
ect the interference.
corr
The Thermo Scientific Orion AQ3010 turbidity meter is CE and
cTUVus certified, which includes TUV 3-in-1 testing to EMC and US
and Canadian standar
. It has been tested and found to comply with the limits
ds.
WEEE
Compliance
This product is required to comply with the European Union’s Waste
E
lectrical and Electronic Equipment (WEEE) Directive 2002/96/EC. It
is marked with the following symbol:
We have contracted with one or more recycling/disposal companies in
each EU Member State and this product should be disposed of or recycled
ough them. Further information on compliance with these Directives,
thr
the recyclers in your country, and information on Thermo Scientific
Orion products which may assist the detection of substances subject to
oHS Directive are available at www.thermo.com/WEEERoHS.
the R
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
25
Page 30
Declaration of
Conformity
Manufacturer: Thermo Fisher Scientific Inc.
166 Cummings Center
Beverly, MA 01915 USA
The manufacturer hereby declares that the AQ3010 turbidity meter
conforms with the following standards and documents:
Safety EN/IEC 61010-1:2001
EMC EC 89/336/EEC Electromagnetic Compatibility
EN 61326-1:2006
Emissions EN 61326-1:2006 Emissions
CISPR 11:2003 Radiated Emissions
FCC Part 15B Class A
Immunity
This Thermo Scientific Orion product has been manufactured in
compliance with the provisions of the relevant manufacturing and test
documents and pr
as complying with ISO 9001:2008 by QMI, listed as File # 001911.
Patrick Chiu June 3, 2009
Senior Quality Engineer, Beverly, MA
Regulatory Compliance
EN 61326-1:2006 Immunity
IEC 61000-4-2:2001 Electrostatic Discharge
Immunity
IEC 61000-4-3:2002 RF Radiated I
ocesses. These documents and processes are recognized
mmunity
26Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 31
Ordering
Cat. No.Description
Information
AQ3010AQ3010 waterproof portable turbidity meter with set of 4 calibration
standards (800, 100, 20.0 and 0.02 NTU) in plastic bottles, set of 4
calibration vials, set of 3 sample vials, lint-free cloth, silicone oil, vial
cleaning brush, batteries and rugged carrying case
AC301S
AC3V25Replacement sample vials, pack of 3, 25 mm diameter
AC3SILSilicone oil, 10 mL bottle, and lint-free cloth
AC3CBRReplacement vial cleaning brush
Replacement calibration standards kit, includes 800, 100, 20.0 and 0.02
NTU standards in plastic bottles
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Sample Vials (Cuvettes)Glass vials with screw caps, fill line and
indexing mark, 51 (H) x 25 (Dia) mm ( 2 x 1 in)
Sample Volume Required10 mL (0.33 oz)
Operating Temperature Range0 °C to 50 °C (32 °F to 122 °F)
Sample Temperature Range0 °C to 50 °C (32 °F to 122 °F)
Operating Humidity Range0 to 90% relative humidity,
non-condensing at 30 °C (86 °F)
Power Supply4 x AAA alkaline batteries
Battery Life> 1200 readings
Enclosure Type ABS plastic
Waterproof Rating:IP67
Regulatory CertificationCE, cTUVus, TUV 3-in-1, FCC part 15
Insulation RatingPollution Degree 2
Weight:Meter: 0.23 kg (0.51 lb)
Meter with case: 2.0 kg (4.4 lb)
DimensionsMeter:
7.9 (W) x 17.2 (L) x 5.0 (H) cm
(3.1 x 6.8 x 2.0 in)
Meter & case:
35 (W) x 27 (L) x 15 (H) cm
(13.8 x 10.6 x 5.9 in)
28Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 33
Appendix 1 Turbidity Theory
Definition
Why Is It
Important?
Turbidity is defined as an “expression of the optical property that causes
light to be scattered and absorbed rather than transmitted in straight lines
ough the sample.” That is, turbidity is the measure of relative sample
thr
clarity, not color.
Water with cloudy or opaque appearance will have high turbidity, while
water that is clear or translucent will hav
value is caused by particles such as silt, clay, microorganisms and organic
By definition, turbidity is not a direct measure of these particles
matter.
but rather a measure of how these particles scatter light.
For drinking water applications, a turbidity value may give an indication
of presence of bacteria, pathogens or particles that can shelter harmful
organisms fr
is particularly useful for water treatment plants to ensure cleanliness. For
wastewater effluent testing, turbidity is often required to ensure that the
r
eleased effluent will not harm the environment.
In industrial processes, turbidity can be part of quality control measure to
ensur
om disinfection process. Therefore, turbidity measurement
e efficiency in treatment or manufacturing process.
e low turbidity. High turbidity
Measurement
Principle
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
There are two internationally accepted standard specifications for
turbidity measur
and the US EPA method 180.1.
Basically the ISO 7027 is a more stringent standard and requires the use
of a monochr
meter in the following areas: (1) Light source’s wavelength; (2) Light
sources’ spectral bandwidth; (3) Measuring angle; (4) Aperture angle in
water sample; (5) Distance traversed by incident light and scattered light
within the sample and (6) Calibration standar
The AQ3010 turbidity meter follows the ISO 7027 standard whose
specification allows for greater reproducibility of the measured values and
gr
eater agreement between other measuring instruments.
ement. These are the international standard ISO 7027
omatic light source. It also governs the design of a turbidity
d.
29
Page 34
Figure 15
90°
Lens
Glass Sample Vial
Transmitted
Light
Scattered Light
Lamp
Basic
Nephelometric
Arrangement
for Turbidity
Measurement
Figure 15 shows the AQ3010 waterproof turbidity meter basic optical
system. It includes a light source and a detector to monitor the light
scattered at 90° with respect to the incident beam.
Nephelometric
Turbidity Units
(NTU)
Indexing a
Vial
Nephelometric Turbidity Units (NTU): Unit of measure used when
relating the light scattered by a liquid media to the light scattered by a
known concentration of a standard solution. This unit of measure is
recognized as a measure of the optical clarity of an aqueous sample. NTU
is the accepted unit of measurement for turbidity.
Another unit commonly used to measure turbidity is Formazin Turbidity
Unit (FTU). The two units of measure of turbidity are equivalent:
1 NTU = 1 FTU
.
The United States Environmental Protection Agency (US EPA)
recommends that vials used for turbidity meter calibration or sample
measur
ement be indexed.
To index a sample vial, slowly rotate the vial throughout one complete
r
evolution (360°). While rotating the sample vial, observe the display and
locate the position that the vial is in which provides the lowest turbidity
eading. This position is the indexed position of the vial.
r
Mark this position on the vial (not on the cap) against the mark on
the meter.
After indexing a vial, make sure the vial will always be placed inside the
sample w
30Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
ell in the indexed position.
Page 35
Appendix 2 Guide to Good
Measurement Technique
Turbidity is a very complex analytical measurement that can be affected
by many factors. Some are inherent in the instrument’s design such as
angle of detection, light beam aperture, incident beam wavelength and
color sensitivity of the photocell
However, there are other factors such as stray light, air bubbles and care
of vial, which can be prevented through proper care of equipment and
accessories, and in the operating pr
some points you may want to note.
.
ocedure for measurement. Here are
Maintain
Sample Vials
in Good
Condition
Match
Sample Vials
Sample vials must be meticulously clean and free from significant
scratches
silicone oil. This is to mask minor imperfections and scratches that may
contribute to stray light. Sample vials should be handled only by the
top to avoid dirt accumulation (or deposits) and fingerprints that might
inter
More information can be found in the Vials – Handling, Cleaning and Care section
Best precision and repeatability of turbidity measurement are achieved
using a single, index
vials can be used for measurement provided their readings with the same
solution ar
the specified repeatability and precision of the meter.
Select a few vials. After the sample vials are cleaned, fill them with ultralow turbidity water. Allow the sample vials to stand and for air bubbles to
rise. Polish sample vials with silicone oil and take turbidity measurement
at several points while rotating it in the sample well. Find the position
where turbidity reading is the lowest and index it for each vial. Whenever
these sample vials are used, use the indexed mark to position each vial
into the sample w
. It should be treated on the outside with a thin coat of
fere with the light path.
.
ed vial. However, for more convenience, different
e matched. That is, the meter gives identical readings or within
ell. Choose those vials that match the readings.
Not all vials can be matched due to some manufacturing variations.
Note:
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
31
Page 36
Degassing
Air or other trapped gases should be removed before measurement.
Degassing is recommended even if no bubbles are visible. There are three
methods commonly used for degassing:
Addition of a surfactant: This involves adding a surfactant to the
•
water samples to lower the surface tension of the water, thereby
r
eleasing trapped gasses.
• Application of a partial vacuum: A partial vacuum can be created by
using simple syringe or vacuum pump. This is only recommended
for ultra-low turbidity measurements.
• Use of an ultrasonic bath: This may be effective in severe conditions
or in viscous samples, but not recommended for ultra-low
measur
Each of the method above has its own advantages and disadvantages. For
instance, under certain sample conditions, the use of vacuum pump or
ultrasonic bath may actually incr
ements.
ease the presence of gas bubbles.
Timeliness
of Sample
Measurement
Samples should be measured immediately to prevent changes in particle
characteristics due to temperatur
affect particles by changing their behavior or creating new particles if
ecipitates are formed. Dilution water may dissolve particles or change
pr
their characteristics. It is recommended to take samples only when the
turbidity meter is ready to be operated. Samples should not be drawn
and allowed to sit while the instrument warms up or is being readied.
e and settling. Temperature can
32Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
Page 37
Other
Important
Sampling
Techniques
1. Samples should not be violently shaken or agitated as particles can be
broken apart or air may be entrapped into the fluid. Gentle agitation
such as swirling the sample vial is advisable to reduce particle settling.
2. Sample vials should be used only with the instruments for which they
were intended. Do not mix and match vials. The AQ3010 uses 25
mm sample vials.
3. Perform a visual observation of the sample vial every time a
measurement is made. Ensure that there are no visible bubbles in the
sample and the vial is clean and free of scratches.
4. Samples entering the turbidity meter should be at the same
temperature as the process flow samples. Changes in temperature can
cause precipitation of soluble compounds and affect readings.
5. Sample vials should be evaluated with a low turbidity water
(after cleaning) to determine if cells remain matched. If the
evaluation determines that a cell is corrupted, discard the vial. It is
recommended to conduct this evaluation weekly.
6. When in doubt about whether a sample vial is too scratched or
stained, throw it away.
Calibration
Techniques
1. Keep the calibration standard plastic bottles and vials tightly capped
and ensure that they are not allowed to freeze.
2. Check that the standards have not expired.
3. Make sure that the calibration vials are free of dust, smudges and
scratches before use.
4. Conduct the calibration in the same manner each time. Variations
in how the calibration is performed could yield inaccurate
measur
5. It is very important that the users who perform calibration have been
trained to do so. Creating a Standard Operating Procedure (SOP) for
the users to read, learn, and practice may help to ensure precision.
ements.
Thermo Scientific Orion AQUAfast AQ3010 Turbidity Meter User Guide
33
Page 38
Dilution
Techniques
This dilution procedure is necessary only when your turbidity
measurement is above 1000 NTU.
1. To measure the turbidity above 1000 NTU, dilute the sample with
turbidity-free water.
2. Turbidity-free water can be obtained by filtering deionized water
through a < 0.2 μm filter membrane with precision-sized pores.
3. Measure the volume of the sample (V
) before dilution and record the
s
value in mL.
4. Take a known volume (V
) of dilution water and add it to the
d
sample.
5. Pour 10 mL of the diluted sample in a clean vial and measure the
turbidity of the diluted sample. Record this value in NTU (T
).
d
6. Calculate the true turbidity (T) of the original sample in NTU using
the following formula:
T =
• (Vs + Vd) / V
T
d
s
Example:
• Dilute 20 mL of the original sample (with turbidity above 1000
NTU) with 50 mL of dilution water.
• Measure the turbidity of the diluted sample.
• If the reading is 300 NTU, the turbidity of the original sample is
1050 NTU.
• In this example Td = 300 NTU, Vs = 20 mL, Vd = 50 mL,