Thermo Scientific AI 3000, AS 3000, AS 3000 II Operating Manual

Page 1
Thermo Scientific
Home
AI 3000 / AS 3000 AI 3000 / AS 3000 II
3000 Series Automatic Sampling System
Operating Manual
P/N 31709392, Tenth Edition, March 2012
Page 2
rmo Fisher Scientific Inc., and its subsidiaries.
The
Published by Thermo Fisher Scientific S.p.A., Strada Rivoltana 20090 Rodano-Milan, Italy Tel: +39 02 95059303; Fax: +39 02 95059388
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
d
uct operation. This document is copyright protected and any reproduction of the whole or any part of this
pro document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
cument is for reference purposes only. System configurations and specifications in this document supersede
do all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
ee and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
fr result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
cument shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
do Sale shall govern all conflicting information between the two documents.
Release history: First edition, January 2002; Second Edition, May 2003; Third Edition, September 2003
o
urth Edition, April 2004; Fifth Edition, September 2004; Sixth Edition, May 2007; Seventh Edition,
F December 2007 (see the Note (*); Eighth Edition, April 2009; Ninth Edition, May 2010; Tenth Edition, March 2012
(*) Note: The seventh edition of this manual considers the AI 3000/AS 3000 II automatic sampling system.
For Research Use Only. Not for use in diagnostic procedures.
Page 3
AI 3000 / AS 3000 II Operating Manual, PN 31709392, Tenth Edition
Reader’s Survey
fold
fold
Strongly
Agree
Agree Neutral Disagree
Strongly
Disagree
The manual is well organized. 1 2 3 4 5 The manual is clearly written. 1 2 3 4 5 The manual contains all the info rmation I need. 1 2 3 4 5 The instructions are easy to follow. 1 2 3 4 5 The instructions are complete. 1 2 3 4 5 The technical information is easy to understand. 1 2 3 4 5 Examples of operation are clear and useful. 1 2 3 4 5 The figures are helpful. 1 2 3 4 5 I was able to operate the system using this manual. 1 2 3 4 5
If not, please comment below. At
__________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________ __________________________________________________________ __________________________________________________________________
tach additional sheets if necessary.
Customer Registration Card
Register now…and receive all the privileges associated with being a Thermo Fisher Scientific product user including customer support, application reports, and technical reports.
MY ORGANIZATION IS: (Che
Commercial (for profit) lab Analytical Government lab Biomedical Hospital/Clinic Clinical/Toxicology Industrial lab Energy Research Institute Environmental University/College Food/Agricultural Veterinary Forensic/Toxicology Other______________________ Pharmaceutical
MY PRIMARY JOB FUNCTION IS:
Administration Lab management Operator Other______________________
c
k only one) MY PRIMARY APPLICATION IS: (Check only one)
Research/Education
(Check only one) Other______________________
Name __________________________________________________ Title__________________________________________________________________ Company __________________________________________________ __________________________________________________________________ Address ___________________________________________________ __________________________________________________________________ City/State _________________________________________ Postal Code__________________________________________________________________ Country ___________________________________________________ __________________________________________________________________ Telephone______________ _________________________ ________ Ext. __________ __________________________ _________________________ _____ Serial Number __________________________________ Date purchased __________________________________________________________________
Fold and mail or e-mail to:
Editor, Technical Publications
Thermo Fisher Scientific S.p.A. Strada Rivoltana km 4 20090 Rodano (MI) Italy
Editor, Technical Publications
Thermo Fisher Scientific SID GC-GC/MS 2215 Grand Avenue Parkway Austin TX 78728-3812 Unites States of America
Page 4
Page 5
Declaration
Manufacturer: Thermo Fisher Scientific
Thermo Fisher Scientific is the manufacturer of the instrument described in this manual and, as such, is responsible
r the instrument safety, reliability and performance only if:
fo
•installation
•re-calibration
•changes and repairs
have been carried out by authorized personnel and if:
the local installation complies with local law regulations
the instrument is used according to the instructions provided and if its operation is only entrusted to qualified rsonnel
trained p
Thermo Fisher Scientific is not liable for any damages derived from the non-compliance with the aforementioned
ommendations.
rec
e
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with applicable domestic and international regulations.
When the system is delivered to you, it meets all pert standards.
inent electromagnetic compatibility (EMC) and safety
Safety
This device complies with EN 61010-1:2001, EN 61010-2:2002 according to Low Voltage Directive 2006/95/EC.
Electromagnetic Compatibility
This device complies with EN 61326-1:2006 according to directive 2004/108/EC.
FCC Compliance Statement
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED, INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained inside this manual pertaining to the safe use and operation of this product before using the device.
Page 6
Notice on Lifting and Handling of
Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/ or bulky for one person alone to handle safely.
Notice on the Proper Use of
Thermo Scientific Instruments
In compliance with international regulations: Use of this instrument in a manner not specified by Thermo Fisher Scientific could impair any protection provided by the instrument.
Notice on the Susceptibility
to Electromagnetic Transmissions
Your instrument is designed to work in a controlled electromagnetic environment. Do not use radio frequency transmitters, such as mobile phones, in close proximity to the instrument. For manufacturing location, see the label on the instrument.
Page 7
WEEE/RoHs Compliance
This product is in compliance with the EU 2002/95/EC (RoHs) Directive on restriction of hazardous substances in electrical and electronic equipment. This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU) Member State, and these companies should dispose of or recycle this product. See www.thermo.com/WEEERoHS information on Thermo Fisher Scientific’s compliance with these Directives and the recyclers in your country.
for further
WEEE/RoHs Konformität
Dieses Produkt ist in Befolgung mit der EU-2002/95/EG (RoHs) Richtlinie zur Beschränkung der Verwendung bestimmter gefährlicher Stoffe in Elektro- und Elektronikgeräten. Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-Mitgliedsstaaten getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Mehr Information über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, über die Verwerter, und weitere Hinweise, die nützlich sind, um die Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter www.thermo.com/WEEERoHS
.
Conformité DEEE/RoHs
Ce produit est conformement à la directive européenne 2002/95/CE (RoHs) relative à la limitation de l’utilisation de certaines substances dangereuses dans les équipements électriques et électroniques. Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS
.
Page 8
Conformità RAEE/RoHs
Questo prodotto è conforme alla direttiva europea 2002/95/CE (RoHS) sulla restrizione dell’uso di determinate sostanze pericolose nelle apparecchiature elettriche ed elettroniche. Questo prodotto è marcato con il seguente simbolo in conformità alla direttiva europea 2002/96/CE (RAEE) sui rifiuti di apparecchiature elettriche ed elettroniche:
Thermo Fisher Scientific si è accordata con una o più società di riciclaggio in ciascun Stato Membro della Unione Europea (EU), e queste società dovranno smaltire o riciclare questo prodotto. Per maggiori informazioni vedere il sito www.thermo.com/
WEEERoHS.
IMPORTANTE Gli utenti italiani devono riferirsi alla dichiarazione di Conformità alle Direttive RAEE allegata allo strumento.
Conformidad RAEE/RoHs
Este producto es in conformidad a la Directiva 2002/95/CE de la Unión Europea sobre restricciones a la utilización de determinadas substancias peligrosas en aparatos eléctricos y electrónicos. Este producto es marcado con el siguiente símbolo en conformidad a la Directiva 2002/96/CE de la Unión Europea sobre los residuos de aparatos eléctricos y electrónicos:
Thermo Fisher Scientific ha contratado una o más empresas de reciclo para tratar residuos en cada Estado Miembro de la Unión Europea, y estas empresas deberían reciclar o eliminar este producto. Referirse a www.thermo.com/WEEERoHS una mayor información sobre la conformidad de Thermo Fisher Scientific con estas Directivas y para las empresas de reciclaje en su país.
IMPORTANT
The symbol indicates that the product must not be disposed o of this product prevents any potentially negative impact on the environmental and human health that could arise from any inappropriate handling of the product itself.
WEEE and RoHS rules, while laid down at European level, are put into national law at national level. When ex­porting to Europe, it is essential to comply with national law in each relevant country. The EU law simply serves
l
as a temp
Each EU Member State has own regulations regarding the application of these directives. Please refer to the regulations in force in your country.
ate for national laws, which may differ considerably.
f with the normal household wastes. Correct disposal
para
Page 9
C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
About Your System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Rating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Safety Alerts and Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Safety Symbols and Signal Words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xvi
Instrument Markings and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvii
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
The AI 3000/AS 3000 II Sampling System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Sampling Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Configurations of the 3000 Series Automatic Sampling System. . . . . . . . . . . . . . 5
AI 3000 Auto-injector Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
AS 3000 Auto-sampler Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Gemini Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Chapter 2 Sampling Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Base. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Turret . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Injection Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Vial Capture Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Turret Movements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Washing and Waste Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Rear of the Sampling Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Sample Trays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8-Position Sample Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
105-Position Sample Tray Assemble . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Vials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Centering Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Centering Plate for TRACE 1300/TRACE 1310 . . . . . . . . . . . . . . . . . . . . . 18
Centering Plate for TRACE GC Ultra and FOCUS GC. . . . . . . . . . . . . . . . 19
Status LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual ix
Page 10
Contents
Chapter 3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Who Performs the Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
How to Lift and Carry the Sampling Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Sampling System Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Sampling System Support for TRACE 1300/TRACE 1310 . . . . . . . . . . . . . 23
Sampling System Support for TRACE GC Ultra. . . . . . . . . . . . . . . . . . . . . . 26
Sampling System Support for FOCUS GC . . . . . . . . . . . . . . . . . . . . . . . . . . 28
References for the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Installing the Sampling System Support on the TRACE 1300/TRACE
1310 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Assembling of the Sampling System Support on the Front Injector
Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Assembling of the Sampling System Support on the Back Injector
Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Assembling of the Sampling System Support for the Gemini
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Assembling of the Sampling System Support When a TSQ Quantum
is Coupled with the GC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Installing the Sampling System Support on the TRACE GC Ultra . . . . . . . . . . 43
Installing the Spacer Inside the GC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Assembling of the Sampling System Support on the GC Right Side . . . . . . . 44
Assembling of the Sampling System Support on the GC Left Side. . . . . . . . . 44
Installing the Sampling System Support on the FOCUS GC. . . . . . . . . . . . . . . 46
Assembling of the Sampling System Support on the GC right side . . . . . . . . 46
Assembling of the Sampling System Support on FOCUS GC MS
Version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Space Saving Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Installation of the AI 3000/AS 3000 II on the GC . . . . . . . . . . . . . . . . . . . . . . 49
Installation of the Sampling Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Installation of the Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Electrical Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Instrument Start-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Chapter 4 AI 3000/AS 3000 II Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Control from PC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Washes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Info Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Status Line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
x AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 11
Contents
Control Through the TRACE GC Ultra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Configuring the AI 3000/AS 3000 II. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
AI 3000/AS 3000 II Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Extended Control Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
When No Vial Abort Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Control Through the FOCUS GC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Configuring the AI 3000/AS 3000 II. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
AI 3000/AS 3000 II Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Sequence of Samples with the AS 3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Guidelines for Programming with Different Injectors . . . . . . . . . . . . . . . . . . . . 70
SSL Inlet (TRACE GC Ultra/FOCUS GC) . . . . . . . . . . . . . . . . . . . . . . . . . 70
Nano-Volumes Injections in S/SL Inlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Packing a SSL Liner. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
S/SL Inlet with a Merlin Microseal™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
PTV Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
PTV Inlet with a Merlin Microseal™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
PKD and PPKD Inlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Chapter 5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Current Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Emptying of the Waste Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Cleaning of the Sample Tray Accessory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Cleaning the Instrument Externally. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Replacing the Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Cleaning the Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Automated Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Manual Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Preparing a New Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Moving Away the Sampling Unit from the Injector Module . . . . . . . . . . . . . . . 80
Abbreviations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual xi
Page 12
Contents
xii AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 13
P

Preface

This manual contains descriptions of the features and components of the AI 3000/AS 3000 II sampling systems. Inside, you will find all of the information necessary for routine operation of your sampling system. This includes operating procedures, sample injection techniques, and diagrams and descriptions of the major components.
This manual is organized as follows:
Chapter 1, “Introduction,” provides a basic overview of the features and options of the ailable instrument configurations.
AI 3000/AS 3000 II System. It also describes the
Chapter 2, “Sampling Unit,” provides a description of the components of the sampling
unit of the AI 3000/AS 3000 II System.
Chapter 3, “Installation,” contains the instructions for the installation of the
AI 3000/AS 3000 II on the TRACE 1300/TRACE 1310, TRACE GC Ultra and
US GC gas chromatographs, the syringe and the electrical connections with the
FOC different units of the gas chromatographic system.
Chapter 4, “AI 3000/AS 3000 II Control,” provides the information to control the
AI 3000/AS 3000 II from the computer or the GC. The chapter contains also the
rk
ing procedures with different injectors.
wo
av
Chapter 5, “Maintenance,” provides guidelines for the maintenance of the AI 3000/AS
3000 II.
Abbreviations contains definitions of terms used in this manual. This also includes
abbreviations, acronyms, metric prefixes, and symbols.
Index contains an alphabetical list of key terms and topics in this guide, including cross
references and the corresponding page numbers.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual xiii
Page 14
Preface

About Your System

About Your System
WARNING Thermo Fisher Scientific systems operate safely and reliably under carefully
controlled environmental conditions. If the equipment is used in a manner not specified by the manufacturer, the protections provided by the equipment may be impaired. If you maintain a system outside the specifications listed in this guide, failures of many types, including personal injury or death, may occur. The repair of instrument failures caused by operation in a manner not specified by the manufacturer is specifically excluded from the Standard Warranty and service contract coverage.
WARNING When, for technical reasons, it is necessary to work on instrument parts which may involve an hazard (moving parts, components under voltage, etc.) the authorized Technical Service must be contacted. This type of situations can be identified because access to these parts is possible only by using a tool. The removable protective covers bear a warning symbol suggesting to refer to the documentation accompanying the instrument. Should an operator perform a maintenance operation, he/she must have received proper training to carry out that specific action.
WARNING Before using dangerous substances (toxic, harmful, etc.), please read the hazard indications and information reported in the Safety Sheet supplied by the manufacturer referring to the relevant CAS (Chemical Abstract Service) number. The AI 3000/AS 3000 II sampling system requires the use of several chemical products, which are present in vials and syringes, having different hazard characteristics. Before using these substances or replacing the syringe, please read the hazard indications and information reported in the Safety Sheet supplied by the manufacturer referring to the relevant CAS (Chemical Abstract Service) number.

Environmental Conditions

•Internal use
• Up to 2000 meters altitude
• Temperature 18 to 30 °C
• Maximum relative humidity between 30% and 85%
• Voltage variations must not exceed the nominal voltage by ± 10%
• Transient overloads in compliance wi
th installation categories II
• Pollution degree according to IEC 664 (3.7.3) 2
• Protection degree IP00
xiv AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 15

Rating

Contacting Us

Preface
Rating
AI 3000/AS 3000 II Automatic Sampling System
• 24 Vdc through a portable external power supply
input 90-264 Vac; 47/63 Hz — output 24 Vdc; 70 W
Thermo Fisher Scientific provides comprehensive technical assistance worldwide and is dedicated to the quality of our customer relationships and services.
Use http://www.thermoscientific.com address for products information.address for products information.
Use http://www.gc-gcms-customersupport.com/WebPage/Share/Default.aspx address to contact your local Thermo Fisher Scientific office or affiliate GC-GC/MS Customer Support.

Safety Alerts and Important Information

Make sure you follow the precautionary notices presented in this guide. Safety and other special notices appear in boxes and include the following:
WARNING This is the general warning safety symbol and safety alert word to prevent actions that could cause personal injury. It highlights hazards to humans or the environment. Each WARNING safety alert is preceded with this safety symbol and another appropriate safety symbol (see “Safety Alerts and Important Information” on page xv) Then it is followed with an appropriate safety precautionary message. When you see a safety alert on your instrument or in the publications, please carefully follow the safety instructions before proceeding.
CAUTION This is the safety alert word to prevent actions that may cause personal injury or instrument damage. We use it to highlight information necessary to prevent personal injury or damage to software, loss of data, or invalid test results; or might contain information that is critical for optimal system performance. A CAUTION safety alert is always preceded with an appropriate safety symbol (see “Safety Alerts and Important
Information” on page xv) Then it is followed with an appropriate safety precautionary
message. When you see a safety alert on your instrument or in the publications, please carefully follow the safety instructions before proceeding.
IMPORTANT Highlights information necessary to prevent damage to software, loss of data, or invalid test results; or might contain information that is critical for optimal performance of the system.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual xv
Page 16
Preface
Safety Alerts and Important Information
Note Emphasizes important information about a task.
Tip Helpful information that can make a task easier.

Safety Symbols and Signal Words

All safety symbols are followed by WARNING or CAUTION, which indicates the degree of risk for personal injury and/or instrument damage. Cautions and warnings are following by a descriptor, such as BURN HAZARD. A WARNING is intended to prevent improper actions that could cause personal injury. Whereas, a CAUTION in intended to prevent improper actions that may cause personal injury and/or instrument damage. The following safety symbols may be found on your instrument and/or in this guide:
BIOHAZARD. This symbol indicates that a biohazard will, could, or may occur.
BOOTS REQUIRED. This symbol indicates that you must wear boots when
performing a task or else physical injury could or may occur.
BURN HAZARD. This symbol indicates a hot surface could, or may incur burn injuries.
CLOTHING REQUIRED. This symbol indicates that you should wear a work clothing when performing a task or else physical injury could or may occur.
ELECTRICAL SHOCK HAZARD. This symbol indicates that an electrical shock hazard could, or may occur.
EXPLOSION HAZARD. This symbol indicates an explosion hazard. This symbol indicates this risk could or may cause physical injury.
FIRE HAZARD. This symbol indicates a risk of fire or flammability, or that fire/flammability damage will, could, or may occur.
FLAMMABLE GAS HAZARD. This symbol alerts you to gases that are compressed, liquefied or dissolved under pressure and can ignite on contact with an ignition source. This symbol indicates this risk could or may cause physical injury.
GLOVES REQUIRED. This symbol indicates that you must wear gloves when performing a task or else physical injury could or may occur.
HAND AND CHEMICAL HAZARD. This symbol indicates that chemical damage or physical injury will, could, or may occur.
HARMFUL. This symbol indicates that the presence of harmful material will, could, or may occur.
INSTRUMENT DAMAGE. This symbol indicates that damage to the instrument or module may occur. This damage may not be covered under the standard warranty.
xvi AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 17
Preface
3

Instrument Markings and Symbols

LIFTING HAZARD. This symbol indicates two or more people are required to lift the object to prevent a physical injury, which could or may occur.
MATERIAL AND EYE HAZARD. This symbol indicates that eye damage will, could, or may occur. Eye protection must be worn.
RADIOACTIVE HAZARD. This symbol indicates that the presence of radioactive material could, or may occur.
READ MANUAL. This symbol alerts you to carefully read your instrument’s operational instructions before usage to ensure your safety and the instrument’s operational ability. Failing to carefully read the instructions could or may put you at risk for a physical injury.
TOXIC SUBSTANCES HAZARD. This symbol indicates that exposure to a toxic substance will, could, or may cause personal injury or death.
This is the general warning symbol that the ISO 3864-2 standard refers to as the general warning signal to prevent personal injury. It is a triangle with an exclamation mark that precedes the WARNING safety alert word. In the vocabulary of ANSI Z535 signage, this symbol indicates a possible personal injury hazard exists if the instrument is improperly used or if unsafe actions occur. We use this symbol and another appropriate safety symbol to alert to an imminent or potential hazard that could cause personal injury.
Instrument Markings and Symbols
Tab le 1 explains the symbols used on Thermo Fisher Scientific instruments. Only a few of
them are used on the AI 3000/AS 3000 II. See the asterisk.
Table 1. Instrument Marking and Symbols (Sheet 1 of 2)
Symbol Description
Direct Current
* Alternating Current
Both direct and alternating current
Three-phase alternating current
Earth (ground) terminal
Protective conductor terminal
Frame or chassis terminal
Equipotentiality
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual xvii
Page 18
Preface
+-
Instrument Markings and Symbols
Table 1. Instrument Marking and Symbols (Sheet 2 of 2)
Symbol Description
* On (Supply)
* Off (Supply)
Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION (Equivalent to Class II of IEC 536)
* Instruction manual symbol affixed to product. Indicates that the user must
refer to the manual for specific WARNING or CAUTION information to avoid personal injury or damage to the product.
Caution, risk of electric shock
* Caution, hot surface
* Caution, biohazard
In-position of a bistable push control
Out-position of a bistable push control
* Jack socket
* Symbol in compliance to the Directive 2002/96/EC on Waste Electrical
and Electronic Equipment (WEEE) placed on the european market after August, 13, 2005.
xviii AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 19
1

Introduction

This chapter provides a basic overview of the features and options of the AI 3000/AS 3000 II. It also describes the available instrument configurations.
Contents
The AI 3000/AS 3000 II Sampling System
Configurations of the 3000 Series Automatic Sampling System
Technical Specifications
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 1
Page 20
1
Turret
Safety Door
Injection Unit
Washing and Waste Tray
8-position Sample Tray
Introduction

The AI 3000/AS 3000 II Sampling System

The AI 3000/AS 3000 II Sampling System
The 3000 Series Automatic Sampling System, available as AI 3000 or AS 3000, consists of the following parts, as respectively shown in Figure 1 and Figure 2.
Figure 1. AI
3000 Auto-injector
2 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 21
Figure 2. AS 3000 Auto-sampler
Turret
Safety Door
Injection Unit
Centering Plate
Washing and Waste Tray
Base
105-position Sample Tray
1
Introduction
The AI 3000/AS 3000 II Sampling System

Sampling Unit

The sampling unit consists of the following major components:
Base
This is the structure bearing the sampling unit components.
Turret
It consists of a vertical moving structure that accommodates the injection unit and accomplishes the necessary movements to carry out the three main operating steps: sample drawing, injection, syringe washing. The turret is provided with a safety door allowing access to the syringe.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 3
Page 22
1
Introduction
The AI 3000/AS 3000 II Sampling System
Injection Assembly
Housed inside the rotating turret, it consists of a moving support (sliding plate) on which the syringe is installed.
Washing and Waste Tray
It allows to accomodate up to four 4 mL vials of solvents to wash the syringe, and a 50 mL plastic container to collect the washing solvents after their use.
Centering Plate
It consists of a plate provided with circular guides to center the sampling unit on the injector nut.
Sample Tray
It contains the seats for the sample vials. Each position is numbered to make the sample identification easy. The following options are available:
AI 3000 Auto-injector — sample tr
AS 3000 Auto-sampler
mobile tray mounted on a dedicated support plate.
— ample holding assembly consisting of a 105-position turnable
ay consisting of an 8-position fixed support.
Sampler Support
It constitutes the supporting base allowing to install the autoinjector/sampler on the gas chromatograph.
User Interface
The functions of the AI 3000/AS 3000 II can be controlled through:
• a data processing system for PC with dedicated software.
• the keypad of the TRACE GC Ultra.
For further information, please refer to:
“Configurations of the 3000 Series Automatic Sampling System” on pag
Chapter 2, “Sampling Unit.” on page 9
Chapter 4, “AI 3000/AS 3000 II Control,” on page 55
e 5
4 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 23

Configurations of the 3000 Series Automatic Sampling System

8-position Sample Tray
Configurations of the 3000 Series Automatic Sampling System
The AI 3000/AS 3000 II can be installed on the TRACE 1300/TRACE 1310, TRACE GC Ultra, or FOCUS GC gas chromatographs for sample introduction into the following injectors:
• TRACE 1300/TRACE 1310: S/SL, PTV, PTVBKF
• TRACE GC Ultra: S/SL, PTV, PKD, PPKD
• FOCUS GC: S/SL
The autoinjector/sampler is available in the following options:
AI 3000 Auto-injector Configuration
AS 3000 Auto-sampler Configuration
Gemini Configuration
1
Introduction

AI 3000 Auto-injector Configuration

In this configuration the 8-position sample tray is used. See Figure 3.
Figure 3. AI 300
0 Auto-injector Configuration
As a function of the configuration of installation, right or left, the support will be introduced into the relevant side housing on the base of the sampling unit. On the opposite side housing, the centering plate will be positioned.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 5
Page 24
1
105-position Sample Tray
Introduction
Configurations of the 3000 Series Automatic Sampling System

AS 3000 Auto-sampler Configuration

In this configuration the 105-position sample tray is used, mounted on the hub provided on the dedicated support plate.
Figure 4. AS 3
The support plates of the tray and the sampler are electrically connected to each other by means of
the proper connecting cable.
000 Auto-sampler Configuration
As a function of the configuration of installation, right or left, the support will be introduced into the relevant housing on the base of the sampling unit.

Gemini Configuration

Consists of the installation of two AI 3000/AS 3000 II on the same GC.
Note Gemini configuration is not possible with the FOCUS GC.
• With the TRACE 1300/TRACE 1310 — the samplers are installed on the right side, one
on
the front injector module and the other on the back injector module.
• With the TRACE GC Ultra — one sampler is installed on the left side of the GC, and
her is installed on the right side.
the ot
Each autosampler operates simultaneously on its own relevant injector. In this configuration, both autosampler are ready to inject the sample simultaneously, according to the programmed method that is the same for both.
6 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 25

Installation

Right configuration
1
Introduction
Configurations of the 3000 Series Automatic Sampling System
The AI 3000/AS 3000 II can be installed on the right or left side of the GC according to the following indications:
The right configuration, see Figure 5, is the default configuration for all the GC units.
Note The right configuration is the only one possible on the TRACE 1300/TRACE 1310 and FOCUS GC.
Left Configuration
Figure 5. Righ
The left configuration, see Figure 6, is only possible with the TRACE GC Ultra. The installation of the AI 3000/AS 3000 II requires a dedicated metal support on the base of the sampler support.
Figure 6. Le
t Configuration
f
t Configuration
WARNING If an existing AI 3000 auto-injector will be upgraded to AS 3000 auto-sampler, it is very important to switch off the AI 3000 before to install the 105 sample tray and plug-in its electrical cable. This recommendation will prevent a possible driver failure on AI 3000’s pcb.
Gemini Configuration
Two AI 3000/AS 3000 II are installed on the same TRACE 1300/TRACE 1310 or TRACE GC Ultra by using their own dedicated supports. Refer to “Sampling System Support” on
page 23.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 7
Page 26
1
Introduction

Technical Specifications

Technical Specifications
Technical specification of the AI 3000/AS 3000 II are listed in Tabl e 1.
Table 1. Technical Specifications
Sample tray capacity • 8 vials in the AI 3000 configuration
Vials capacity 0.3 and 2 mL (standard) with both 8- and 105-position
Flushing solvents 4 vials of 4 mL
Waste container capacity 50 mL
Types of syringes • 10 μL syringe (standard)10 μL syringe for Merlin
• 105 vials in the AS 3000 configuration
trays
Microseal Valve (optional)
•5 μL syringe (optional)
•0.5 μL syringe for Nano-volumes injections (optional)
Volumes programming (μl) • 0 to 5 μL (10 μL syringe) with increments of 0.1 μL
• 0 to 2.5 μL (5 μL syringe) with increments of 0.1 μL
• 0 to 0.5 μL (0.5 μL syringe) with increments of 0.01 μL
Programmability Remote
External interface Serial line RS 232
Power supply rating 24 Vdc through a portable external power supply
Electrical characteristics of the supply
• input 90-264 Vac; 47/63 Hz
• output 24 Vdc; 70 W
Dimensions (height x width x length)
Mass About 6 kg
Sampling unit:
• AI Configuration: approx. 437 x 230 x 250 mm
• AS Configuration: approx. 437 x 400 x 250 mm
8 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 27
2

Sampling Unit

This chapter provides a description of the components of the sampling unit of the AI 3000/AS 3000 II.
Contents
Base
Turret
Injection Assembly
Washin g an d Wast e Tray
Rear of the Sampling Unit
Sample Trays
Centering Plate
Status LED
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 9
Page 28
2
Front Section
Right Side Section
Left Side Section
Rear Section
Sampling Unit

Base

Base
It is the supporting structure of the instrument on which the sampling unit components and accessories are positioned (sample tray, centering plate). It consists of four sections. See
Figure 7.
e
Figure 7. Th
Sections of the Base
Front Section — It includes the housings for solvents vials and waste container.
On the lower front part there is a LED indicating the instrument status. For details, refer to “Status LED” on pag
Right and Left Side Sections
e 20.
It can accommodate the, 8-position sample tray or
105-position sample tray and the centering plate for the injector.
The position of the above described accessories depends on the configuration of the
njector/auto-sampler installation.
auto-i
See “Configurations of the 3000 Series Automatic Sampling System” on
10 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
page 5.
Page 29

Turret

2
Sampling Unit
Rear Section — It includes the connections of the instrument to the mains, the gas chromatograph, the data system and the sample tray (only for AS 3000), and for the Gemini configuration.
The turret consists of a moving structure pivoted on a hub. The rotating movements are accomplished by a motor located on the base. An optical sensor determines the turret travel ends.
The front side is provided with a safety door allowing access to the injection assembly. The door opening immediately cuts off the power supply to the sampling system.
he T
Figure 8. T
urret
Turret
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 11
Page 30
2
5
Safety Door
Injection Device
Sliding Plate
Syringe Plunger Control
Syringe
Vial Capture Device
Sampling Unit

Injection Assembly

Injection Assembly
It consists of a sliding plate which supports and guides the vertical movements of the injection device constituted by the syringe plunger control, the vial capture device and the syringe. See Figure 9.
Figure 9. In
je
ction Assembly
The vertical movements of the injection device are controlled by motors housed in the turret. The travel ends of the movable parts is defined by a series of sensors.
Inside the turret two electronic boards are also present, one for the motors control and one for the instrument functions control, respectively.
12 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 31

Vial Capture Device

The lower part of the injection device consists of the vial capture device. See Figure 10.
The function of this device are:
Figure 10. Vial Capture Device
• to guide the syringe needle
• to acknowledge the vial presence
• to acknowledge the injector presence
2
Sampling Unit
Injection Assembly

Syringe

The standard syringe has 10 μL capacity with 50 mm needle length.
Note 5 and 0.5 μL syringes are optionally available.
A thorough washing of the syringe can be done thanks to the possibility of using up to four solvents. See Wash ing a nd Wa ste Tra y .
To install or replace the syringe, please refer to the Chapter 3, “Installation,” and to the
Chapter 5, “Maintenance.”
Figure 11 shows the syringe housing.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 13
Page 32
2
Syringe Housing
Syringe
Sampling Unit
Injection Assembly
Figure 11. Syringe Housing

Turret Movements

The sequence of the turret movements allows the autoinjector/sampler to perform the following functions:
Sample drawing
tray, to allow the syringe to draw the sample it contains.
Sample injection
injection.
Syringe washing
washing. The solvent is automatically withdrawn and discharged into the waste container.
14 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
— the turret positions itself on the selected vial, housed in the sample
the turret positions itself on the injector to perform the sample
the turret positions itself on the solvents vials for the syringe
Page 33

Washing and Waste Tray

Waste Container
Vials for Solvent A and B
Vials for Solvent C and D
24 Vdc
-
+
TRAY
TWIN SYNC
RS 232 GC
The washing and waste tray can contain up to four 4 mL vials and a 50 mL plastic container. See Figure 12.
2
Sampling Unit
Washing and Waste Tray
Figure 12. Wa
shing and Waste Tray
The vials contain the solvent (solvents) necessary for the syringe washing, which can take place
both before and after the injection. The washing solvent is then collected in the waste
container.
The four vials containing solvents for the syringe washing are introduced into the appropriate seats A, B and C, D, on the left and on the right of the waste container respectively, marked WA ST E.

Rear of the Sampling Unit

The rear side of the sampling unit includes the following components. See Figure 13:
Figure 13. Rear
•Jack marked 24 Vdc for the instrument power supply through the external portable
feeder.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 15
of the Sampling Unit
Page 34
2
Sampling Unit

Sample Trays

• 15-pin female connector (double density) marked TRAY for the communication between
the sampling unit and the motor actuating the hub positioned on the supporting plate of the 105-position sample tray.
Sample Trays
• 9-pin female connector marked RS232 fo
autoinjector/sampler and the GC (connector marked AUTOSAMP L ER) through a proper cable.
• 6-pin female connector marked GC fo
(connector marked SAMPLER SIGNAL) through the proper cable.
• 4-pin male connector marked TWIN S
AI 3000/AS 3000 II installed in Gemini configuration through the proper cable.
The AI 3000/AS 3000 II can accept two types of sample trays which are accommodated in one of the two side housings on the base.
The type of sample tray accessory installed on the instrument (8 or 105 positions) is automatically acknowledged by the autoinjector/sampler when switched on, and it determines the instrument configuration.
I 3000 uses the 8-position sample tray.
A
AS 3000
uses the 105-position sample tray.
r the communication between the
r connecting the autoinjector/sampler to the GC
YNC for the synchronism between two

8-Position Sample Tray

The 8-position tray, see Figure 14, consists of a fixed arc-shaped support and can contain up to 8 vials numbered 1 to 8. The slot present on the tray determines the position 1 to inspect the level of the liquid contained in the vial.
Figure 14. 8-
position Sample Tray

105-Position Sample Tray Assemble

The assemble, see Figure 15, consists of a dedicated supporting plate provided with a hub and a 105-position removable sample tray.
and allows
16 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 35
2
Supporting Plate
Hub
105-position Sample Tray
Sampling Unit
Sample Trays
The rotation of the tray is accomplished by a motor, located under the supporting plate, which actuates the hub thanks to suitable commands coming from the sampling unit via an interfacing cable.
An optical sensor defines the rotation travel end and the vial position.
ppo
Figure 15. Su
rting Plate and 105-Position Sample Tray

Vials

The tray can contain up to 10
•Vials 01 to 40 are pl
•Vials 41 to 80 are pl
•Vials 81 to 105 are place
5 vials numbered 01 to 105 arranged on three circular rows.
aced into the external row.
aced into the middle row.
d into the internal row.
Vials of 0.3 and 2 mL capacity can be used.
The 0.3 mL vial is exactly the same as the 2 mL vial, provided inside with a 0.3 mL calibrated insert.
CAUTION Vials must be accurately closed using appropriate septa and ring nuts. Septa must be those recommended by Thermo Fisher Scientific. The use of septa with different characteristics might damage or bend the syringe needle. During the vials preparation, it is recommended to comply with applicable safety regulations, specially as far as the conditions of the workplace ventilation are concerned.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 17
Page 36
2
CORRECT
WRONG
Sampling Unit

Centering Plate

Centering Plate
The centering plate consists of a plate provided with circular guides to center the sampling unit on the injector nut.

Centering Plate for TRACE 1300/TRACE 1310

The centering plate is provided with two circular guides to center the sampling unit on the PTV and SSL injector nuts.
• To center the sampling unit on the SSL injector nut, use the circular guide for SSL only.
See B of Figure 16.
• To center the sampling unit on the PTV injector nut, use the circular guide for PTV only.
See A of Figure 17.
Figure 16. Circular Guides for SSL
CAUTION The circular guide for SSL placed in the centering plate must be aligned as
shown below:
18 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 37
Figure 17. Circular Guides for PTV
Circular Guide
2
Sampling Unit
Centering Plate

Centering Plate for TRACE GC Ultra and FOCUS GC

Two types of centering plates are available with the GC according to the injector mounted.
• A standard centering plate to center the sampling unit on the SSL, PKD and PPKD
inj
ector nut. This centering plate is used also with the FOCUS GC.
• A centering plate to center the sampling unit on the PTV injec
a circular slot present on the circular guide. This centering plate is not used with the FOCUS GC.
Figure 18. Centering Plate for TRACE GC Ultra and FOCUS GC
tor nut. It is recognized by
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 19
Page 38
2

Status LED

Sampling Unit
Status LED
Status LED
The status LED, located on the front side of the sampling unit, see Figure 19, provides the operator with indications on the instrument operating conditions showing a continuous or intermittent light. See Tab le 2 for details.
Figure 19. The St
Table 2. Status LED Conditions
Condition LED
St-by Lit with continuous green light
Position zeroing, sample preparation, washing, injection
Run (waiting GC Ready) Intermittent green-orange lights; 0.5 seconds green, 0.5 seconds
Firmware upload Intermittent orange light; 0.5 seconds ON, 0.5 seconds OFF
atus LED
Intermittent green light; 0.5 seconds ON, 0.5 seconds OFF
orange
Error Lit with continuous orange light
Buzzer
Two beep monotone; 0.5 seconds ON, 0.5 seconds OFF, 0.5 seconds ON indicates that the start signal is received.
Five consecutive beep, 1 second with interval of 0.5 seconds OFF in case of error condition.
20 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 39
3

Installation

This chapter contains the instructions for the installation of the AI 3000/AS 3000 II on the TRACE 1300/TRACE 1310, TRACE GC Ultra, and FOCUS GC gas chromatographs, the syringe, and the electrical connections with the different units of the gas chromatographic system.
Contents
Who Performs the Installation
Electrical Requirements
How to Lift and Carry the Sampling Unit
Sampling System Support
Installing the Sampling System Support on the TRACE 1300/TRACE
1310
Installing the Sampling System Support on the TRACE GC Ultra
Installing the Sampling System Support on the FOCUS GC
Installation of the AI 3000/AS 3000 II on the GC
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 21
Page 40
3
Installation

Who Performs the Installation

Who Performs the Installation
The AI 3000/AS 3000 II is installed by authorized Thermo Fisher Scientific field service engineers (FSE), who will check its correct operation. For more details, please contact Thermo Fisher Scientific local representatives. Should the instrument not be installed by Thermo Fisher Scientific personnel, please strictly adhere to the following instructions.

Electrical Requirements

The instrument must be electrically supplied as indicated in the “Technical Specifications” on
page 8.
CAUTION The power line and the connections between the instruments must maintain good electrical grounding. Poor grounding represents a danger for the operator and may seriously affect the instrument performance. Do not connect the AI 3000/AS 3000 II to lines feeding devices of a heavy duty nature, such as motors, UV lamps, refrigerators and other devices that can generate disturbances. Pay attention not to leave any cable connecting the sampling unit and the chromatographic system or the power cord close to the GC hot air vents.

How to Lift and Carry the Sampling Unit

The AI 3000/AS 3000 II must be lifted and carried by hand. See Figure 20.
Figure 20. How
to lift and carry the Sampling Unit
22 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 41

Sampling System Support

The AI 3000/AS 3000 II is installed on the GC using the appropriate support provided.
According to your GC, please refer to:
Sampling System Support for TRACE 1300/TRACE 1310
Sampling System Support for TRACE GC Ultra
Sampling System Support for FOCUS GC
Note The FOCUS GC MS Version, may accept the AI 3000/AS 3000 II also on the left side. Please refer to “Assembling of the Sampling System Support on FOCUS GC MS
Ver si on” on page 46.
To reduce the FOCUS GC working area, the AS 3000 may be installed as described in
“Space Saving Installation” on page 47.
3
Installation
Sampling System Support

Sampling System Support for TRACE 1300/TRACE 1310

The AI 3000/AS 3000 II can be installed only on the right side of the GC on the front or back injector module or on both when in Gemini configuration by using the proper sampling
system support.
The sampling unit support consists of a bottom and a top plates.
sh
1. The bottom plate can be L- Gemini installation). The plate is provided with a series of holes for the corresponding fixing screws and must be fixed on the GC.
2. The movable top plate is mounted over the bottom plate and it is provided with a guide
or the accommodation and centering of the sampling unit. The movable top plate
pivot f can slides over the bottom plate through a sliding slot and a retainer.
This allows moving away the injection device of the sampling unit from the injector
le when the latter will require maintenance.
modu
Note After the installation of the sampling unit on the proper sampling system support, it is possible locking its plates if desired.
To use the proper sampling system support on the TRACE 1300/TRACE 1310 please refer to:
ape (see both front/back installation) or double (see
Support for the Installation on the Front/Back Injector Module
Support for the Installation in Gemini Configuration
Support for the Installation When a TSQ Quantu
the GC
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 23
m Mass Spectrometer is Coupled with
Page 42
3
Guide Pivot
Retainer
Sliding Slot
L-shape Bottom Plate (Oriented on Front)
Plates Locking Screw
Top Plate
Guide Pivot
Retainer
Sliding Slot
Top Plate
Plates Locking Screw
L-shape Bottom Plate (Oriented on Back)
Installation
Sampling System Support
Support for the Installation on the Front/Back Injector Module
The same sampling unit support is used. The L-shape bottom plate must be oriented accordingly before fixing it on the GC. See Figure 21 and Figure 22.
Figure 21. Su
pport Oriented for the Installation on the Front Injector Module
Figure 22. Support Oriented for the Installation on the Back Injector Module
24 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 43
3
Double Bottom Plate
Retainer
Sliding Slot
Guide Pivot
Top Plate
Double Bottom Plate
Retainer
Sliding Slot
Guide Pivot
Top Plate
Installation
Sampling System Support
Support for the Installation in Gemini Configuration
This configuration uses a dedicated double bottom plate and two top plates. See Figure 23.
Figure 23. Suppo
rt for the Installation in Gemini Configuration
Support for the Installation When a TSQ Quantum Mass Spectrometer is Coupled with the GC
This configuration uses a dedicated double bottom plate and two top plates. See Figure 24.
Figure 24. Suppo
rt for the Installation When a TSQ Quantum is Coupled with the GC
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 25
Page 44
3
1
2
3
5
4
1 = Semi-circular Plate 2 = Guide Pivot 3 = Spacers 4 = Slots 5 = Fixing Screws
Installation
Sampling System Support

Sampling System Support for TRACE GC Ultra

The support for the installation of the AI 3000/AS 3000 II on right or left side of the TRACE GC Ultra consists of a semi-circular plate resting on three spacers non adjustable in height, see Figure 25. The top surface of the plate is provided with two slots for corresponding fixing screws. In the middle the guide pivot allows the accommodation and centering of the sampling unit.
rt for the Installation of the AI 3000/AS 3000 II on the TRACE GC Ultra
Figure 25. Su
ppo
The AI 3000/AS 3000 II is installed on the rig
ht or left side of the TRACE GC Ultra.
Installation on the Right Side
When installing the AI 3000/AS 3000 II on the TRACE GC Ultra right side, the support is directly mounted on the GC top panel previous installation of the spacer inside the GC. See
Figure 25.
Note For details about the installation of the spacer, please refer to “Installing the Spacer
Inside the GC” on page 43 for details.
Installation on the Left Side
When installing the AI 3000/AS 3000 II on the TRACE GC Ultra left side, before to mount the sampling system support, the support bracket must be placed and fixed on the GC top cover. See Figure 26.
26 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 45
Figure 26. Bracket for the Installation of the AI 3000/AS 3000 II Support
1
3
3
3
2
2
2
4
1 = Support Bracket 2 = Holes on the TRACE GC Ultra Top Cover 3 = Fixing Screws 4 = Sampling Unit Support
3
Installation
Sampling System Support
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 27
Page 46
3
1
2
3
4
5
1 = Upper Semi-circular Plate 2 = Lower Semi-circular Plate 3 = Guide Pivot 4 = Slots 5 = Fixing Screws
Installation
Sampling System Support

Sampling System Support for FOCUS GC

The support for the installation of the AI 3000/AS 3000 II on the right side of the FOCUS GC, consists of two semi-circular plates. See Figure 27. The top surface of the upper plate is provided with two slots for corresponding fixing screws. In the middle the guide pivot allows the accommodation and centering of the sampling unit.
ppo
Figure 27. Su
rt for the Installation of the AI 3000/AS 3000 II on the FOCUS GC
Note The FOCUS GC MS Version, may accept the AI 3000/AS 3000 II also on the left side. Please refer to “Assembling of the Sampling System Support on FOCUS GC MS
Ver si on” on page 46.
To reduce the FOCUS GC working area, the AS 3000 may be installed as described in
“Space Saving Installation” on page 47.
28 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 47

References for the Installation

For the installation of the AI 3000/AS 3000 II, please refer to:
“Installing the Sampling System Support on the TRACE 1300/TRACE 1310” on
page 30
3
Installation
Sampling System Support
“Installing the Sampling System Support on the TRACE GC Ultra” on pag
“Installing the Sampling System Support on the FOCUS GC” on page 46
“I
nstallation of the AI 3000/AS 3000 II on the GC” on pag
e 49
e 43
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 29
Page 48
3
Installation Holes for the AI 3000/AS 3000 II on the Front Injector Module.
See Assembling of the Sampling System
Support on the Back Injector Module
Installation Holes for the AI 3000/AS 3000 II on the Back Injector Module.
See Assembling of the Sampling System
Support on the Front Injector Module
.
Installation Holes for Two AI 3000/AS 3000 in Gemini Configuration.
See Assembling of the Sampling System
Support for the Gemini Configuration
Installation Holes for One or Two AI 3000/AS 3000 II when a TSQ Quantum Mass Spectrometer is Coupled with the GC
See Assembling of the Sampling System
Support When a TSQ Quantum is Coupled with the GC
Installation

Installing the Sampling System Support on the TRACE 1300/TRACE 1310

Installing the Sampling System Support on the TRACE 1300/TRACE 1310
To install the sampling system support on the TRACE 1300/TRACE 1310, please refer to the instructions reported below. Figure 28 shows the installation holes on the top of the GC.
Figure 28. Sa
TRACE 1300/TRACE 1310
mpler Unit Support Installation Holes
30 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 49
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
A
B
C

Assembling of the Sampling System Support on the Front Injector Module

The following is the description of how to mount the sampling system support on the front injector module of the TRACE 1300/TRACE 1310. See Figure 29 and Figure 30.
L-shape Bottom Plate
3
Installation
Figure 29. As
sembling on the Front Injector Module (1)
1. Remove from the GC top cover the three plastic caps covering the corresponding fixing holes. S
ee A, B, and C.
Note In comparison with the installation on the back injector module, the installation on the front injector module requires to upturn the L-shape bottom plate before fixing it on the GC.
2. Orient the L-shape bottom plate to align the three installation holes to the corresponding fixing holes located on the GC top.
3. Insert the provided fixing screws into the holes present on the L-shape bottom plate.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 31
Page 50
3
Retainer
Top Plate
L-shape Bottom Plate
B’
Guide Pivot
A’
C’
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Figure 30. Assembling on the Front Injector Module (2)
4. Place the L-shape bottom plate on the GC top and guide the three fixing screws into the
5. Tighten the three fixing screws.
6. If not already done, mount and fix the guide pivot on the top plate by using the proper
7. Place the top plate over the L-shape bottom plate.
8. Guide the retainer and its fixing screw into the sliding slot present on the top plate.
9. Align the retainer fixing screw into the corresponding fixing hole located on the L-shape
0. Tighten the retaining fixing screw.
1
11. Clamp the two plates by using the plates locking screw.
co
rresponding fixing holes. See A’, B’, and C’.
ng screw.
fixi
bottom plate.
32 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 51
3
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Note Removing the plates locking screw, the top plate is free to move outwards. See the figure below.
This allows the sampling unit, when installed, to be moved away from the injector module. See “Moving Away the Sampling Unit from the Injector Module” on page 80.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 33
Page 52
3
A
B
C
L-shape Bottom Plate
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310

Assembling of the Sampling System Support on the Back Injector Module

The following is the description of how to mount the sampling system support on the back injector module of the TRACE 1300/TRACE 1310. See Figure 31 and Figure 32.
Figure 31. As
sembling on the Back Injector Module (1)
1. Remove from the GC top cover the three plastic caps covering the corresponding fixing holes. See A, B, and C.
Note In comparison with the installation on the front injector module, the installation on the back injector module requires to upturn the L-shape bottom plate before fixing it on the GC.
2. Orient the L-shape bottom plate to align the three installation holes to the corresponding fixing holes located on the GC top.
3. Insert the provided fixing screws into the holes present on the L-shape bottom plate.
34 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 53
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Retainer
Top Plate
L-shape Bottom Plate
A’
B’
C’
Guide Pivot
Figure 32. Assembling on the Back Injector Module (2)
3
Installation
4. Place the L-shape bottom plate on the GC top and guide the three fixing screws into the co
rresponding fixing holes. See A’, B’, and C’.
5. Tighten the three fixing screws.
6. If not already done, mount and fix the guide pivot on the top plate by using the proper
ng screw.
fixi
7. Place the top plate over the L-shape bottom plate.
8. Guide the retainer and its fixing screw into the sliding slot present on the top plate.
9. Align the retainer fixing screw into the corresponding fixing hole located on the L-shape bottom plate.
0. Tighten the retaining fixing screw.
1
11. Clamp the two plates by using the plates locking screw.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 35
Page 54
3
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Note Removing the plates locking screw, the top plate is free to move outwards. See the figure below.
This allows the sampling unit, when installed, to be moved away from the injector module. See “Moving Away the Sampling Unit from the Injector Module” on page 80.
36 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 55
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
A
C
B

Assembling of the Sampling System Support for the Gemini Configuration

The following is the description of how to mount the sampling system support on the TRACE 1300/TRACE 1310 for the installation of two AI 3000/AS 3000 II sampling systems in Gemini configuration. See Figure 33 and Figure 34.
3
Installation
Figure 33. A
s
sembling In Gemini Configuration (1)
Double Bottom Plate
1. Remove from the GC top cover the five plastic caps covering the corresponding fixing holes. See A, B, and C.
2. Insert the provided fixing screws into the three holes present on the double bottom plate.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 37
Page 56
3
A’
B’
C’
Top Plate Back
Double Bottom Plate
Retainer
Guide Pivot
Guide Pivot
Top Plate Front
Retainer
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Figure 34. Assembling In Gemini Configuration (2)
3. Guide the three fixing screws into the corresponding fixing holes located on the GC top. See A’, B’, and C’.
4. Tighten the three fixing screws.
5. If not already done, mount and fix the guide pi
vot on each top plate by using the proper
fixing screw.
6. Place the two top plates over the double bottom plate.
7. Guide each retainer and its fixing screw into the sliding slot present on each top plate.
8. Align each retainer fixing screw into the corresponding fixing hole located on the double bottom plate.
Tighten each retainer fixing screw to block the corresponding upper plate to the double
9. bottom plate.
38 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 57
3
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Note Loosening the screw that secure each retainer, the corresponding top plate are free to move outwards. See the figure below.
This allows the sampling unit, when installed, to be moved away from the injector module. See “Moving Away the Sampling Unit from the Injector Module” on page 80.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 39
Page 58
3
A
C
B
Double Bottom Plate
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310

Assembling of the Sampling System Support When a TSQ Quantum is Coupled with the GC

The following is the description of how to mount the sampling system support on the TRACE 1300/TRACE 1310 for the installation of one or two AI 3000/AS 3000 II sampling systems when a TSQ Quantum mass spectrometer is coupled with the GC. See Figure 33 and
Figure 35.
s
Figure 35. A
sembling (1)
1. Remove from the GC top cover the five plastic caps covering the corresponding fixing holes. See A, B, and C.
2. Insert the provided fixing screws into the three holes present on the double bottom plate.
40 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 59
Figure 36. Assembling (2)
A’
B’
C’
Top Plate Back
Double Bottom Plate
Retainer
Guide Pivot
Guide Pivot
Top Plate Front
Retainer
3
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
3. Guide the three fixing screws into the corresponding fixing holes located on the GC top.
4. Tighten the three fixing screws.
5. If not already done, mount and fix the guide pi
6. Place the two top plates over the double bottom plate.
7. Guide each retainer and its fixing screw into the sliding slot present on each top plate.
8. Align each retainer fixing screw into the corresponding fixing hole located on the double
9.
See A’, B’, and C’.
vot on each top plate by using the proper
fixing screw.
bottom plate.
Tighten each retainer fixing screw to block the corresponding upper plate to the double bottom plate.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 41
Page 60
3
Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Note Loosening the screw that secure each retainer, the corresponding top plate are free to move outwards. See the figure below.
This allows the sampling unit, when installed, to be moved away from the injector module. See “Moving Away the Sampling Unit from the Injector Module” on page 82.
42 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 61

Installing the Sampling System Support on the TRACE GC Ultra

Installing the Sampling System Support on the TRACE GC Ultra
To install the support on the TRACE GC Ultra, please refer to the instructions reported below. Before installing the sampling system support, the spacer should be installed inside the GC, then assemble the sampling system support on the GC according to the required configuration.
•See Installing the Spacer Inside the GC
•See Assembling of the Sampling System Support on the GC Right Side
•See Assembling of the Sampling System Support on the GC Left Side

Installing the Spacer Inside the GC

To installing the spacer should be installed inside the GC proceed as described below. See
Figure 37.
3
Installation
Figure 37. Inst
allation of the Spacer Inside the GC
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 43
Page 62
3
Installation
Installing the Sampling System Support on the TRACE GC Ultra
1. Remove the cover and the collar of the injector and detector compartment.
2. Open the column oven door. Loosen the two knobs located in the upper parts.
3. Push the cover backwards and remove it.
4. Screw the spacer on the threaded pin.
5. Put the GC cover back in the original position.
6. Tighten the two knobs. Close the column oven door.
7. Mount the spacer fixing Allen screw.
8. Reinstall the cover and the collar of the injector and detector compartment.
Assemble the sampling system support on the GC ac
cording to the required configuration.

Assembling of the Sampling System Support on the GC Right Side

The following is the description of how to mount the sampling system support in the default configuration, i.e. on the right side of the TRACE GC Ultra.
1. Remove from the GC top cover the two plasti holes.
2. Insert into each slot present on the support the provided fixing screw.
3. Insert each screw into the relevant spacer paying attention to keep its largest surface
towards the support base.
turned
4. Holding the spacers in position with your hands, rest the support on the GC top cover, with thei
5. Guide the two fixing screws located on the external spacers into the corresponding fixing holes.
Loosely tighten the screws.
6.
r flat side towards the inside.
c caps covering the corresponding fixing

Assembling of the Sampling System Support on the GC Left Side

The following is the description of how to mount the sampling system support on the left side of the TRACE GC Ultra.
1. Remove from the GC top cover the four plastic caps covering the corresponding fixing holes.
Mount the support bracket on the TRACE GC Ultra and fix by using fixing screws
2.
ided.
prov
3. Insert into each slot present on the support the provided fixing screw.
44 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 63
3
Installation
Installing the Sampling System Support on the TRACE GC Ultra
4. Insert each screw into the relevant spacer paying attention to keep its largest surface turned towards the support base.
5. Holding the spacers in position with your hands, rest the support on the bracket, with
lat side towards the inside.
their f
6. Guide the two fixing screws located on the external spacers into the corresponding fixing
ocated on the bracket.
holes l
7. Loosely tighten the screws.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 45
Page 64
3
A
1
2
1
B
1
4
3
A = Conventional Installation B = Left Side Installation 1 = Right Side Fixing Hole 2 = Support Positioning 3 = Left Side Fixing Holes 4 = Support Positioning
Installation

Installing the Sampling System Support on the FOCUS GC

Installing the Sampling System Support on the FOCUS GC
To install the support on the FOCUS GC, please refer to the instructions reported below.
•See Assembling of the Sampling System Support on the GC right side
•See Assembling of the Sampling System Support on FOCUS GC MS Version
•See Space Saving Installation

Assembling of the Sampling System Support on the GC right side

The following is the description of how to mount the sampling system support in the default configuration, i.e. on the right side of the FOCUS GC.
1. Remove from the GC upper cover the two plastic caps covering the corresponding fixing holes.
Insert into each slot present on the support the provided fixing screw.
2.
3. Guide the two fixing screws into the corresponding fixing holes located on the GC upper panel.
Loosely tighten the screws.
4.

Assembling of the Sampling System Support on FOCUS GC MS Version

The installation on the GC left side, instead the conventional installation on the GC right side, may be carried out to reduce the working area of your FOCUS GC MS Version. Refer to the Figure 38.
mpl
Figure 38. Sa
ing System Support on the FOCUS GC MS Version (1)
46 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Proceed as follows:
Page 65
3
1
1 = Internal Plastic Caps
1
1 = Fixing Screws
Installation
Installing the Sampling System Support on the FOCUS GC
1. Remove from the GC upper cover the two internal plastic caps covering the corresponding fixing holes. See Figure 39.
Figure 39. Sampling System Support on the FOCUS GC MS Version (2)
2. Insert the provided fixing screw into each slot present on the support.
3. Guide the two fixing screws into the corresponding fixing holes located on the GC upper panel. S
ee Figure 40.
Figure 40. Sampling System Support on the FOCUS GC MS Version (3)
4. Loosely tighten the screws.

Space Saving Installation

This type of installation is suggested to reduce the working area of your FOCUS GC equipped with the AS 3000. This is applicable for example when two or more FOCUS GC units are to be placed on the same workbench. See Figure 41.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 47
Page 66
3
A
B
1
2
3
2
1
1
3
4
4
A = Conventional Installation B Space Saving Installation 1 = Fixing Screws 2 = Sampling System Support 3 = AS 3000 4 = Second FOCUS GC
Installation
Installing the Sampling System Support on the FOCUS GC
Figure 41. Two or More FOCUS Placed on the Same Workbench
48 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Proceed as follows:
1. Remove from the GC upper cover the two plastic caps covering the corresponding fixing holes.
2.
Insert into each slot present on the support the provided fixing screw.
3. Guide the two fixing screws into the corresponding fixing holes located on the GC upper panel.
Loosely tighten the screws.
4.
Page 67
Installation of the AI 3000/AS 3000 II on the GC
Sampling Unit
Guide Pivot
Centering Plate for TRACE GC Ultra
Centering Plate for TRACE 1300/1310
PTV
SSL
To install the AI 3000/AS 3000 II on the TRACE 1300/TRACE 1310, TRACE GC Ultra, and FOCUS GC gas chromatographs, please refer to the instructions reported below:
The procedure includes the following steps:
Installation of the Sampling Unit
Installation of the Syringe
Electrical Connections
Instrument Start-up

Installation of the Sampling Unit

Proceed as follows:
1. Lift the sampling unit, see Figure 20 on page 22, and insert it into the guide pivot located on the sampling system support. See Figure 42.
3
Installation

Installation of the AI 3000/AS 3000 II on the GC

Figure 42. Inst
allation of the Sampling Unit
2. Insert the proper centering plate for TRACE GC Ultra or TRACE 1300/TRACE 1310, see Figure 43, into its seat of the sampling unit base paying attention that the guide hole, present on the arm of the centering plate, correctly fits the injector nut.
nt
Figure 43. Ce
ering Plates
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 49
Page 68
3
8-position Sample Tray
Support Plate
105-position Sample Tray
Hub
Installation
Installation of the AI 3000/AS 3000 II on the GC
3. Check the correct alignment of the sampling system support, then fix it by tightening the proper fixing screws.
4. Insert the sample tray into the sampling unit base.
AI 3000
insert the 8-position sample tray into the appropriate housing of the
sampling unit base. See Figure 44.
Figure 44. 8-
position Sample Tray
AS 3000 — insert the dedicated support plate into its appropriate housing of the sampling unit base. Place the 105-position sample tray on the hub located on the support. The system will automatically recognize the sample tray at the instrument power on. See
05-
Figure 45. 1
position Sample Tray
50 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 69

Installation of the Syringe

1
2
3
4
5
6
7
8
9
1 = Syringe Needle 2 = Syringe Body 3 = Flange 4 = Plunger 5 = Plunger Head 6 = Vial Capture Device 7 = Syringe Body Seat 8 = Flange Guide 9 = Plunger Head Guide
To install the syringe follow the instructions reported below.
The installation of the syringe is a simple operation. However, it must be performed with caution to avoid damages to the syringe needle and ensure an optimal performance of the injection device.
The standard syringe has 10 μL capacity with 50 mm needle length. 5 and 0.5 μL optional syringes are optionally available.
1. Open the safety door of the turret.
2. Insert the syringe needle into the vial capture device. See Figure 46.
Figure 46. Sy
nge Installation (1)
ri
3
Installation
Installation of the AI 3000/AS 3000 II on the GC
3. Accommodate the syringe body into its seat paying attention to insert simultaneously the fla
nge and head of the syringe plunger into their relevant guides.
4. Lock the syringe by turning clockwise the lock knob by approximately 180°. See
Figure 47.
5. Close the safety door of the rotating turret.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 51
Page 70
3
1
Installation
Installation of the AI 3000/AS 3000 II on the GC
Figure 47. Syringe Installation (2)

Electrical Connections

The instrument must be powered as indicated in “Technical Specifications” on page 8.
Before performing electrical connections, ensure that the requirements reported in paragraph
“Electrical Requirements” on page 22 have been fulfilled.
This section provides the instruction to connect the AI 3000/AS 3000 II, in standard o Gemini configuration, to a TRACE 1300/TRACE 1310, TRACE GC Ultra, or FOCUS GC.
Refer to:
Standard Configuration
Gemini Configuration
Standard Configuration
Proceed as follows:
1. By using the cable provided, connect the 6-pin female connector marked GC loca the sampling unit back side and the connector, located on the rear panel of the GC, marked:
AUTOSAMPLER HANDSHAKE in the
ted on
case of TRACE 1300/TRACE 1310.
SAMPLER SIGNAL i
52 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
n case of TRACE GC Ultra and FOCUS GC.
Page 71
3
24 Vdc
-
+
TRAY
TWIN SYNC
RS 232 GC
24 Vdc
-
+
TRAY
TWIN SYNC
RS 232 GC
GC
PC
SAMPLER SIGNAL
COM 1
AI/AS 3000 (a)
AI/AS 3000 (b)
Installation
Installation of the AI 3000/AS 3000 II on the GC
2. By using the cable provided, connect the 9-pin female connector marked RS232 located on the sampling unit back side to 9-pin connector marked AUTOSAMPLER located on the GC rear panel.
Gemini Configuration
3. In case of AS 3000, conne
ct the 15-pin female connector of the cable coming from the
support plate of the 105-position sample tray to the connector marked TRAY located on the sampling unit back side.
Figure 48 shows how to connect two AI 3000/AS 3000 II in Gemini Configuration.
Figure 48. Co
nnections in Gemini Configuration
Proceed as follows:
1. By using the cable provided, connect the 4-pin male connector marked TWIN on the rear of the sampling unit of the sampler “a”, to the 4-pin male connector marked TWIN SYNC on the rear of the sampling unit of the sampler “b”.
2. By using the Y shape cable provided, connect:
a. the 6-pin female connector marked GC l
ocated on the sampling unit back side and
the connector, located on the rear panel of the GC, marked:
AUTOSAMPLER HANDSHAKE in the
SAMPLER SIGNAL i
n case of TRACE GC Ultra and FOCUS GC.
b. the first branch of the cable, labelled AI/
case of TRACE 1300/TRACE 1310.
AS 1 to the 6-pin female connector marked
GC on the rear of the sampling unit of the sampler “a”.
c. the second branch of the cable, labelled AI/A
S 2 to the 6-pin female connector
marked GC on the rear of the sampling unit of the sampler “b”.
3. By using the Y shape cable provided, connect:
a. the 9-pin main connector to a PC serial port (e.g. COM 1) or to an USB port,
in
terposing a proper adapter not provided.
SYNC,
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 53
Page 72
3
Installation
Installation of the AI 3000/AS 3000 II on the GC
b. the first branch of the cable, labelled Primary AI/AS to the 9-pin female connector
marked RS232 on the rear of the sampling unit of the sampler “a”.

Instrument Start-up

Proceed as follows:
c. the second branch of the cable, labelled Sec
ondary AI/AS to the 6-pin female
connector marked RS232 on the rear of the sampling unit of the sampler “b”.
1. Plug in the Vdc power cable of the external portable power supply into the jack marked
24 V
dc located on the sampling unit back side.
2. Connect the power cord of the external power su
pply to the mains outlet. The AI 3000/AS 3000 II will automatically run the self-testing routine during which the following automatic checks and settings are carried out:
• alignment between AI 3000/AS 3000 II and the GC injector
• acknowledgement of the installed sample tray
• calculation of the syringe zero
Note The self-test routine is automatically carried out every time that the safety door of the turret is closed.
54 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 73
4

AI 3000/AS 3000 II Control

This chapter provides the information to control the AI 3000/AS 3000 II from the computer or the GC. The chapter contains also the working procedures with different injectors.
Contents
Control from PC
Control Through the TRACE GC Ultra
Control Through the FOCUS GC
Guidelines for Programming with Different Injectors
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 55
Page 74
4
AI 3000/AS 3000 II Control

Control from PC

Control from PC

User Interface

The AI 3000/AS 3000 II is controlled through a Thermo Fisher Scientific data system for gas chromatography installed in the PC and interfaced with the instrument via serial line RS 232. This paragraph describes how to program the instrument operating parameters.
IMPORTANT Before creating an autosampler method you must configure the correct syringe capacity in the AI 3000/AS 3000 II Configuration page of the Data System in use (refer to your Data System Operating Manual – Configuration for details).
The user interface, see Figure 49, is the same for is the same for AI 3000/AS 3000 II and Gemini configurations. In this page the sampler parameters can be set, whereas the sequence of samples to be analyzed and acquired is managed by the data system as a function of the type of sample tray installed.
300
Figure 49. AI
Note The page is visualized previous the selection in the configuration page of the data
system in use referring to the relevant manual.
When in Gemini configuration, the user interface controls both the samplers. The parameters set here will be send to both the samplers as they were a single ones. The hardware must be the same for both the samplers such as type of sampler, tray and samples numbering.
0/AS 3000 II Autosampler Page
IMPORTANT In the Sampler Configuration page of the data system in use, select the function Vial error handling >> Stop on vial missing.
56 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 75

Sampling

This page includes the following fields, menus and messages:
Sampling
Injection
Washe s
Info Line
Status Line
Menus
Messages
Set up the sampling parameters in this field. See Ta b l e 3. for details.
Table 3. Sampling Parameters
4
AI 3000/AS 3000 II Control
Control from PC

Injection

Parameter Description
Sample Volume Enter a value for the injection volume. The range depends on the
size of syringe selected in the Instrument Configuration. For a 5 μl syringe the range is 0.0-2.5 μl and for a 10 μl syringe, 0.0-5.0 μl.
Plunger Strokes Select the number of strokes of the plunger used to eliminate air
bubbles. The range is 1-15.
Viscous Sample Select the check box to indicate that the sample is viscous. Clearing
the check box implements a fast aspiration speed while a selected check box leads to a slow one.
Sampling Depth Choose the position of the needle in the vial from which the
aliquot is drawn. The options are Bottom or Center.
These parameters control the behavior of the syringe when the needle is inserted into the inlet. See Ta bl e 4 for details.
Table 4. Injection Parameters (Sheet 1 of 2)
Parameter Description
Injection Depth Choose an injection depth. The options are Standard or
Minimum. If the pre and post -injection dwell times are set to
zero, the autosampler defaults to the Minimum setting.
Pre-injection Dwell Time
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 57
This is the time that the needle is held in the inlet prior to injection. Enter a value in the range 0-63 s.
Page 76
4
AI 3000/AS 3000 II Control
Control from PC
Table 4. Injection Parameters (Sheet 2 of 2)
Parameter Description

Washes

Pre-Injection
Post-injection Dwell Time
This is the time that the needle is held in the inlet after the injection. Enter a value in the range 0-63 s.
This field contains parameters controlling syringe washing with solvents and the sample.
These parameters are applied to pre-injection washing with a solvent chosen from the vials in the Washing and Waste tray. See Ta b le 5 for details.
Table 5. Washer Parameters
Parameter Description
Pre-Injection These parameters are applied to pre-injection washing with a
solvent chosen from the vials in the Washing and Waste tray. See
Tab le 6 for details.
Sample This is the number of times that the syringe is rinsed with the
sample prior to injection. Select a value in the range 1-15.
Sample
Post-Injection
Table 6. Pre-injection parameters
Item Description
Solvent Choose a solvent from the drop-down list. The options are: A, B,
C, D, A+B and C+D. These settings correspond to the vials in the
Washi n g an d Was te t r ay.
Cycles Select the number of times the solvent is cycled through the
syringe. The range is 1-15.
This is the number of times that the syringe is rinsed with the sample prior to injection. Select a value in the range 1-15.
These parameters are applied to post-injection washing with a solvent chosen from the vials in the Washing and Waste tray. See Ta b le 7 for details.
58 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 77

Info Line

Status Line

4
AI 3000/AS 3000 II Control
Table 7. Post-injection parameters
Item Description
Solvent Choose a solvent from the drop-down list. The options are: A, B,
C, D, A+B and C+D. These settings correspond to the vials in the
Washing and Waste tray.
Cycles Select the number of times the solvent is cycled through the
syringe. The range is 1-15.
It displays the range of the selected parameter.
It displays the instrument status messages and error messages that may occur during operation. See “Messages” on page 60.
Control from PC

Menus

File Menu
The menus of the AI 3000/AS 3000 II are the follows: File, Edit, Command, and Help.
This menu contains the following items:
New — This function creates a new method for the AI 3000/AS 3000 II starting from the data currently in memory or from default parameters previously stored with the function Save
As Default.
Open — This function uploads a pre-existing analytical method for the AI 3000/AS 3000 II
from disk to the computer memory.
Open Recent... — It allows to retrieve the last 5 saved analytical methods.
Note To activate a loaded or modified method, use the command Send Method to Sampler.
Save (Save as...) — This function allows to save an analytical method on the hard disk. With the command Save as... the method can be saved with a new name.
Load From Default — This function allows to upload a default analytical method.
Save As Default — This function allows to save the current analytical method as a default
method for the AI 3000/AS 3000 II ( parameters of this method can be retrieved when selecting the function
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 59
....\tmp.\AI/AS1300 Default method.asm). The
New.
Page 78
4
AI 3000/AS 3000 II Control
Control from PC
Edit Menu
Get Method from Sampler — It allows to transfer the AI 3000/AS 3000 II current method to the GC.
Send Method to Sampler — It transfers the method from the computer to the a AI 3000/AS 3000 II and makes the method active.
Print — This function allows to print the method.
Close — This function allows to close the method. If modifications have been done, the
system will ask where to save them in the method.
This menu gives access to the following functions:
View Audit Trail — It displays the log file of the method. Each change to the method parameters is recorded on the log file with the operator’s name as well as the date and time of the modification. Using the function contents into Windows and then transfer information to other programs if required.
Copy it is possible to copy part or all the log file
Command Menu
Help Menu

Messages

Note Audit trail method log is available only if a protection policy is active in the data
system.
This menu gives access to the following functions:
Start Clean Cycle — It allows to start a washing cycle.
Inject Single Sample — It allows to start the injection of a single sample.
Force Autoinjector to Stand-by — It interrupts any current operation and forces the AI
3000/AS 3000 II to stand-by.
It allows to have access to the on-line guide.
This paragraph lists the status and error messages that the computer may display according to the status of the AI 3000/AS 3000 II.
Status Messages
The status messages are listed below:
• Opening Connection
60 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 79
Error Messages
4
AI 3000/AS 3000 II Control
• No connection (Check the serial line connections)
• The AI 3000/AS 3000 II is in stand-by
• AI 3000/AS 3000 II is checking positions
• AI 3000/AS 3000 II ready to inject sample in vial n but waiting GC Ready
• AI 3000/AS 3000 II is preparing the sample in vial n.
• AI 3000/AS 3000 II has injected the sample in vial n.
• The AI 3000/AS 3000 II front door is open (all operations are paused)
The error messages are listed below:
• AI 3000/AS 3000 II error, solvent not found
• AI 3000/AS 3000 II position error, cannot find the initial zero positions
Control from PC
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 61
Page 80
4
AI 3000/AS 3000 II Control

Control Through the TRACE GC Ultra

Control Through the TRACE GC Ultra
This paragraph describes how to program and control the AI 3000/AS 3000 II through the TRACE GC Ultra keypad.
Note It is not possible to control the 0.5 μL syringe through the GC keypad but only through the data system.
Groups of samples can be automatically analyzed under different analytical conditions by programming a sequence of samples. The sequence includes a set of information for a group of samples and their position on the sample tray. The sequence also specifies the method to be used to run each group of samples.
Note In an analytical method all the AI 3000/AS 3000 II functions can be programmed. To develop an analytical method and activate a sample sequence, please refer to the TRACE GC Ultra Operating Manual.

Configuring the AI 3000/AS 3000 II

When the GC is switched on, the presence of the autoinjector/sampler and its configuration is automatically acknowledged.
The type of configuration is displayed by pressing CONFIG and selecting
The following non editable message is displayed according to the AI 3000/AS 3000 II installed:
Sample tray 105
for the AS 3000.
or
Sample tray
for the AI 3000.
AUTOSAMPLER.
CONFIG AUTOSAMPLER
CONFIG AUTOSAMPLER
8
Note Should the AI 3000/AS 3000 II not be installed or correctly connected, the TRACE GC Ultra will display the message NO AUTOSAMPLER INSTALLED.
62 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 81

AI 3000/AS 3000 II Menu

To set the parameters of the AI 3000/AS 3000 II method, press Autosampler key to open the relevant menu described in Tabl e 8 .
Table 8. Control Through the TRACE GC Ultra: Autosampler Menu
Menu Range Comment
AUTOSAMPLER This line displays the menu title.
4
AI 3000/AS 3000 II Control
Control Through the TRACE GC Ultra
Sample volume 0–5 μL
with increments of
This parameter allows to set the sample quantity to be injected.
0.1 μL
Sample rinses 0–15 This parameter allows to set the number of
syringe pre-washings with the sample.
Plunger strokes 0–15 This parameter allows to set the number of
plunger strokes to eliminate air bubbles forming during the sample withdrawal.
Pre wash solvent A, B, C, D
A+B C+D
This parameter allows to select the vial, or combination of two vials, containing the washing solvent.
Press Mode/Type to select the solvent vial to be used.
Pre wash cycles 0–15 This parameter allows to set the number of
syringe pre-washings with the selected solvent.
Post wash solvent A, B, C, D
A+B C+D
This parameter allows to select the vial, or combination of two vials, with the washing solvent.
Press Mode/Type to select the vial with the solvent to be used.
Post wash cycles 0–15 This parameter specifies the number of
syringe post-washings with the solvent selected.
Extended control See “Extended
Control Menu” on
This parameter allows to set optional injection parameters.
page 64
Press Mode/Type to enter the submenu.
When no vial abort See “When No Vial
This line shows the menu title.
Abort Menu” on page 65
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 63
Page 82
4
AI 3000/AS 3000 II Control
Control Through the TRACE GC Ultra

Extended Control Menu

In this menu it is possible to set how long the syringe needle must remain inside the injector without injecting. This allows to avoid discriminations caused by evaporation of the sample contained in the syringe needle. In this way the sample is completely drawn into the syringe and the needle left inside the injector and heated for a few seconds before injecting. See
Tab le 9 for details.
Table 9. Control Through the TRACE GC Ultra: Extended Control Menu
Menu Range Comments
EXTENDED
This line shows the menu title.
CONTROL
Viscous sample Yes, No This parameter defines the speed at which the
sample is drawn from the vial as a function of the sample viscosity. Select No (default value) if the sample has low viscosity. Select Yes if the sample has high viscosity.
Sampl. depth Bottom,
Center
This parameter determines the penetration
depth of the syringe needle into the vial. selectable in the submenu
Press Mode/Type or Enter to enter the
submenu.
•Selecting Bottom (default value) the needle goes down to the vial bottom.
•Selecting Center the needle goes down to half vial.
Inj. Depth Standard,
Minimum
This parameter determines the penetration
depth of the syringe needle into the injector. selectable in the submenu
Press Mode/Type or Enter to enter the
submenu.
•Selecting Standard (default value) the needle goes down to the maximum depth allowed.
•Selecting Minimum the needle enters the injector and stops immediately beyond the septum (Cold Needle Technique).
64 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 83

When No Vial Abort Menu

See Table 10 for details.
Table 10. Control Through the TRACE GC Ultra: When No Vial Abort Menu
Menu Comments
ON MISSING AS VIAL This line shows the menu title.
Skip to next If this function is selected, the AI 3000/AS 3000 II skips a
Abort sequence If this function is selected, the sequence is aborted after a
4
AI 3000/AS 3000 II Control
Control Through the TRACE GC Ultra
missing vial and goes to the next one. The sample sequence and the sample table are not affected.
missing vial.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 65
Page 84
4
AI 3000/AS 3000 II Control

Control Through the FOCUS GC

Control Through the FOCUS GC
This paragraph describes how to program and control the AI 3000/AS 3000 II through the FOCUS GC keypad.
The AS 3000 can automatically analyze groups of samples under different analytical conditions when a sequence of samples is programmed.
The sequence includes a set of information for the group of samples and their position on the sample tray. The sequence also specifies the method to be used to process each group of samples. Refer to paragraph Sequence of Samples with the AS 3000 on page 68.

Configuring the AI 3000/AS 3000 II

When the GC is switched on, the presence and type of sampling system installed is automatically acknowledged.
To configure the AI 3000/AS 3000 II:
1. In Main Menu move the cursor until Othe
2. Scroll to select Conf
3. Scroll to Aut
The following message will be displayed according to the AI 3000/AS 3000 II installed:
Syringe Volume Sample tray
for the AS 3000.
or
Syringe Volume Sample tray
for the AI 3000.
The parameter Syringe v parameter Sample tray is not editable.
osampler, then press Enter.
iguration, then press Enter.
CONFIG AUTOSAMPLER
CONFIG AUTOSAMPLER
olume can be modified according to the syringe volume, whereas the
rs is displayed, then press Enter.
10
105
10
8
66 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 85

AI 3000/AS 3000 II Menu

To set the parameters of the AI 3000/AS 3000 II method, In Main Menu move the cursor until Autosampler is displayed, then press Enter to open the menu. See Ta bl e 11 for details.
Table 11. Control Through the FOCUS GC: Autosampler Menu (Sheet 1 of 2)
Menu Range Comment
AUTOSAMPLER This line displays the menu title.
4
AI 3000/AS 3000 II Control
Control Through the FOCUS GC
Sample volume 0–5 μL, 0-2.5 μL
with increments of
This parameter allows to set the sample quantity to be injected.
0.1 μL
Sample rinses 0–15 This parameter allows to set the number of
syringe pre-washings with the sample.
Plunger strokes 0–15 This parameter allows to set the number of
plunger strokes to eliminate air bubbles forming during the sample withdrawal.
Pre wash solvent A, B, C, D,
A+B, C+D
This parameter allows to select the vial, or combination of two vials, containing the washing solvent.
Pre wash cycles 0–15 This parameter allows to set the number of
syringe pre-washings with the selected solvent.
Post wash solvent A, B, C, D
A+B C+D
This parameter allows to select the vial, or combination of two vials, with the washing solvent.
Post wash cycles 0–15 This parameter specifies the number of
syringe post-washings with the solvent selected.
Viscous sample Yes, No This parameter sets the sample drawing
speed from the vial as a function of the sample viscosity.
• Select No (default value) if the sample has low viscosity.
• Select Yes when the sample has high viscosity.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 67
Page 86
4
AI 3000/AS 3000 II Control
Control Through the FOCUS GC
Table 11. Control Through the FOCUS GC: Autosampler Menu (Sheet 2 of 2)
Menu Range Comment
Sample depth Bottom, Center,
selectable in the
This parameters sets the penetration depth of the syringe needle into the vial.
submenu
Bottom (defualt value) is selected, the
If needle goes down to the vial bottom.
Center is selected, the needle goes
If down to half vial.
Inj. depth Standard, Minimum,
selectable in the
This parameter defines the penetration depth of the syringe needle into the injector.
submenu
Standard (defualt value) is selected, the
If needle goes down to the maximum allowed depth.
Minimum is selected, the needle enters
If the injector and stops immediately beyond the septum (Cold Needle Technique).
Pre dwell time 0–63 sec This parameter specifies how long the
syringe needle must remain in the injector before injection (Hot Needle Technique).
Post dwell time 0–63 sec This parameter specifies how long the
syringe needle must remain inside the injector after injection.

Sequence of Samples with the AS 3000

To set the parameters to run a sequence of samples with the AS 3000, in Main Menu move the cursor until Sequence is displayed, then press Enter to open the menu. See Tab le 1 2 for details
Table 12. Control Through the FOCUS GC: Sequence Menu with AS 3000 (Sheet 1 of 2)
Menu Range Comments
SEQUENCE This line shows the menu title.
First sample 1 to 105 Indicates the first sample of the
sequence.
Last sample 1 to 105 Indicates the last sample of the
sequence.
Injection/vial 1 to 99 Indicates how many times the same
sample has to be run.
68 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 87
4
AI 3000/AS 3000 II Control
Control Through the FOCUS GC
Table 12. Control Through the FOCUS GC: Sequence Menu with AS 3000 (Sheet 2 of 2)
Menu Range Comments
When no vial See Table 13.
Start sequence Indicates the sequence start.
Stop Sequence Indicates the sequence end.
Table 13. Control Through the TRACE GC Ultra: When No Vial Abort Menu
Menu Comments
ON MISSING AS VIAL This line shows the menu title.
Skip to next If this function is selected, the AI 3000/AS 3000 II skips a
missing vial and goes to the next one. The sample sequence and the sample table are not affected.
Abort sequence If this function is selected, the sequence is aborted after a
missing vial.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 69
Page 88
4
AI 3000/AS 3000 II Control

Guidelines for Programming with Different Injectors

Guidelines for Programming with Different Injectors
To get the best results from your AI 3000/AS 3000 II, follow these guidelines when programming the autosampler method according to the injector in use. Refer to:
SSL Inlet (TRACE GC Ultra/FOCUS GC)
Nano-Volumes Injections in S/SL Inlets
Packing a SSL Liner
S/SL Inlet with a Merlin Microseal™
PTV Inlet
PTV Inlet with a Merlin Microseal™
PKD and PPKD Inlets

SSL Inlet (TRACE GC Ultra/FOCUS GC)

The AI 3000/AS 3000 II can inject in a hot SSL inlet using two distinct needle insertion modes: “Standard” and “Minimum”. In the first mode the needle penetrates for almost its whole length and a dwell time is programmed to provide needle heating. In the latter the needle penetrates for a minimum depth and for the minimum time allowing to minimize contact of the needle with the hot inlet.
These two needle insertion modes are conceived to optimize respectively the “Hot Needle” and the “Cold Needle” injection techniques.
old needle technique — (M
C
and lets the sample to enter the injector nearly entirely in liquid phase. This injection technique requires the use of packed liner to stop the liquid run (see “S/SL Inlet with a
Merlin Microseal™” on
Hot needle technique — (
heat the needle (usually 3-5 seconds) to achieve a complete expulsion of the sample from the needle in form of a spray. This technique permits the use of empty liners.
Tab le 14 r
tec
hniques. For your reference the type of liner and the column insertion inside the inlet are
also reported.
eports a summary of injection parameters and conditions for the two SSL
inimum depth mode) prevents boil-out from the needle
pag
e 73).
Standard depth mode) requires a minimum pre-dwell time to
70 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 89
Table 14. Summary of Injection Parameters and Conditions
4
AI 3000/AS 3000 II Control
Guidelines for Programming with Different Injectors
Mode Technique
Split Hot Needle 40 mm Tubular Liner
Column Insertion
Liner Type Packing of Liner
Not required Standard Injection Depth
(3 or 5 mm ID)
Cold Needle 40 mm Packing material
at half height
Splitless Hot Needle 64 mm Tapered Liner
Not Required Standard Injection Depth
(3 or 5 mm ID)
Cold Needle 50 mm Packing material
at bottom
S/SL Nano-Volumes
1
Only with 0.5μL syringe. For details, refer to the Nano-Volumes Kit for S/SL Inlet System User Guide.
Cold Needle 40 mm Double Tapered
1
Liner
No Minimum Injection Depth
AI 3000/AS 3000 II Injection Parameters
3-5 sec. Pre-Dwell Time 0 sec. Post-Dwell Time
Minimum Injection Depth Dwell Time are automatically inhibited
3-5 sec. Pre-Dwell Time 0 sec. Post-Dwell Time
Minimum Injection Depth Dwell Time are automatically inhibited
Dwell Time are automatically inhibited

Nano-Volumes Injections in S/SL Inlets

Injections of sample volumes below 0.1-0.2 μL in GC SSL inlets is very attractive when the dilution with a solvent is not desired and/or the main component easily overloads the capillary column even at high split ratios. Examples of such applications are analyses of essential oils or petroleum fractions.
The reduction of the injection volume to the sub microliter range using a standard autosampler, like the AI 3000/AS 3000 II unit, is achievable by exploiting the low capacity syringe (0.5 μL) and the tapered liner provided in the Nano-Volumes Kit for S/SL Inlet System. High accuracy and repeatability injecting nano-volumes are achievable by exploiting the “Cold Needle” injection mode.
For injection volumes higher than 0.2-0.3 μL use a 5 (or 10) μL syringe and a standard split or splitless liner (possibly packed with glass wool whenever you inject in “Cold Needle” mode). In summary, use the following guideline to choose the best syringe capacity for your injections with the AI 3000/AS 3000 II autosampler. See Table 15 for details
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 71
Page 90
4
Tubular Liner
Tapered Liner
GLASS WOOL
AI 3000/AS 3000 II Control
Guidelines for Programming with Different Injectors
Table 15. Syringe Guideline for AI 3000/AS 3000 II Autosampler
Injection Volume Range Suggested Syringe
0.5 – 5.0 μL10 μL Standard – general purpose
0.2 – 1.0 μL5 μL---
0.01 – 0.3 μL0.5 μL Plunger-in-needle
1

Packing a SSL Liner

Proceed as follows:
1. On your TRACE GC Ultra or FOCUS GC keypad, enter the ove
2. Enter the injector menu and
3. Enter carrier M
4. Remove the centering plate of the AI 3000/AS 3000 II (if installed).
5. Wait till the injector has reached room temperature and by means of the outfit tool
1
See your GC Spare Part Catalog for items part number.
AI 3000/AS 3000 II Injection Parameters
n menu and switch off
the oven temperature. This will avoid heating of the analytical column when carrier is off.
switch off the inlet temperature.
enu and switch off the carrier supply.
unscr
ew the septum cap, then remove it together with the septum holder and its plate.
6. Unscrew the liner holder and remove it. Use twee
zers to gently extract the liner from the
injector.
7. Insert a little plug of packing material (i.e. deact
ivated glass wool) filling the liner cavity
about 5 mm in length. Push the plug till it reaches the position. See Figure 50.
Figure 50. Packing a SSL Liner
72 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 91
8. Insert the liner in its location, then correctly close the injector and restore the GC parameters.

S/SL Inlet with a Merlin Microseal™

This septum-less device offers an effective solution to get rid of traditional rubber septa side effects (impurities released by excessive warming of septa). Since the temperature around the microseal is lower than the rest of the injector, the most suitable injections technique is the cold needle injection. That means no dwell time to warm up the needle. The adaptor for Merlin Microseal™ also behaves as a spacer between the needle entrance and the liner, so standard injection depth must be used to fulfill correctly the injection.
Note In order to guarantee long-life performance of the Merlin Microseal™ it is recommended to use syringes with 0.63 mm needle diameter and truncated cone shape on tip. Please refer to the GC Spare Part Catalog for part number.
4
AI 3000/AS 3000 II Control
Guidelines for Programming with Different Injectors
Mode Technique
Split Cold Needle 40 mm Tubular Liner
Splitless 50 mm Packing material
Column Insertion
Liner Type Packing of Liner
Packing material
(3 or 5 mm ID)
at half height
at bottom
See “Packing a SSL Liner” on page 72 for instruction.

PTV Inlet

Since PTV is cold during sample introduction the syringe needle will not be heated whenever standard or minimum depth is selected or dwell times are set. However standard injection mode is advisable since the sample evaporation is faster in the deeper part of the liner. Optional 1-2 seconds of dwell times will guarantee stable flow condition after the needle introduction. See Table 16.
Table 16.
Injector Column Insertion
PTV Inlet
AI 3000/AS 3000 II Injection Parameters
Standard Injection Depth 0 sec. Pre-Dwell Time 0 sec. Post-Dwell Time
Standard Injection Depth 0 sec. Pre-Dwell Time 0 sec. Post-Dwell Time
AI 3000/AS 3000 II Injection Parameters
Programmed Temperature Vapor iz i ng I n je c to r (P T V)
30 mm Standard Injection Depth
1-2 sec. Pre-Dwell Time 1-2 sec. Post-Dwell Time
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 73
Page 92
4
AI 3000/AS 3000 II Control
Guidelines for Programming with Different Injectors

PTV Inlet with a Merlin Microseal™

The above PTV conditions apply also when the Merlin Microseal™ is used.
Note In order to guarantee long-life performance of the Merlin Microseal™ it is recommended to use syringes with 0.63 mm needle diameter and truncated cone shape on tip. Please refer to the spare part catalog for part number.

PKD and PPKD Inlets

For injection into packed column the standard insertion depth with 3-5 seconds of Pre-dwell time is suggested. See
Table 17. PKD and PPKD Inlets
Injector AI 3000/AS 3000 II Injection Parameters
Packed Injector (PKD)
Purged Packed Injector (PPKD)
Standard Injection Depth
3-5 sec. Pre-Dwell Time 0 sec. Post-Dwell Time
74 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 93
5

Maintenance

This chapter provides guidelines for the maintenance of the AI 3000/AS 3000 II.
Contents
Current Maintenance
Emptying of the Waste Container
Cleaning of the Sample Tray Accessory
Cleaning the Instrument Externally
Replacing the Syringe
Cleaning the Syringe
Preparing a New Syringe
Moving Away the Sampling Unit from the Injector Module
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 75
Page 94
5
Maintenance

Current Maintenance

Current Maintenance
The AI 3000/AS 3000 II does not generally require maintenance, except emptying of the waste container when full, cleaning of the sample tray and replacement of the syringe. For any other operation, contact Thermo Fisher Scientific Technical Service.

Emptying of the Waste Container

1. If necessary, move the turret to have free access to the washing and waste tray.
2. Take out the container.
3. Remove the cap and empty the container.
4. Put on the cap again and reposition the container into its seat.

Cleaning of the Sample Tray Accessory

The sample tray must be periodically cleaned.
Use a water and soap solution or a household non abrasive product. Dry using a clean cloth.

Cleaning the Instrument Externally

Normal usage of the AI 3000/AS 3000 II can cause the exterior to get dirty.
WARNING It is your responsibility to avoid that dangerous liquids and/or materials seeping inside the AI 3000/AS 3000 II during operation and maintenance.
Solvent should not be used. Do not spray on electrical parts.
To clean the instrument externally:
1. Power off the AI 3000/AS 3000 II.
2. Externally clean the instrument with a soap and water non-abrasive product.
• Pay special attention when cleaning the back but clean using a cloth imbued with the same substance.
solution, or with a household
side of the instrument. Do not spray,
• Carefully avoid seeping of the products used inside the instrument, particularly when
ng the grid of the rear panel.
cleani
76 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 95
• If you just suspect that a substance used for cleaning or a product submitted to analysis has penetrated inside the instrument, immediately shut down the instrument and call an authorized customer support engineer for proper actions. The service engineer must be fully informed on the nature of the concerned substance. In the event that a hazardous material is spilled on or in the instrument, clean the spill according to the procedures reported in the Material Data Sheet for that substance.
3. Dry with a clean cloth.
4. Power on the AI 3000/AS 3000 II.

Replacing the Syringe

The replacement of the syringe is a simple operation. However, it must be performed with caution to avoid damages to the syringe needle and ensure an optimal performance of the injection device. The standard syringe has 10 μL capacity with 50 mm needle length.
Note 5 and 0.5 μL syringes are optionally available.
5
Maintenance
Replacing the Syringe
5 and 0.5 μL syringes are optionally available.
This operation is possible with the instrument either on or off.
1. Open the safety door of the turret.
Note If the instrument is on, power supply is immediately cut off.
2. Turn the syringe locking knob by 180° counterclockwise.
3. Gently take out the syringe plunger head and flange from their relevant guides, then re
move the syringe body from its seat.
4. Completely take out the syringe, paying attention while removing the needle from the
l capture device.
via
5. Take a new syringe.
6. Insert the syringe needle into the vial capture device.
7. Accommodate the syringe body into its seat, paying attention to s introduce the syringe plunger flange and head into their relevant guides.
8. Lock the syringe by turning the lock wheel by 180° clockwise.
9. Close the safety door of the rotating turret.
imultaneously
Note If the instrument is on, when the door is closed the instrument automatically runs the syringe zero routine. If the instrument is off, the operation is performed when it is switched on during the self-test routine.
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 77
Page 96
5
Maintenance

Cleaning the Syringe

Cleaning the Syringe
To get the best performance from the syringe installed in the autosampler and increase the syringe lifetime here are some useful tips in order or cleaning impact:
1. If the sample is not viscous, program in the a
2. Program in the autosampler method one or more post-injection wash cycle with solvent.
3. If the sample is viscous, program in the autosample
4. Perform an automated maintenance cycle every 100 injections (see “Automated Cleaning”
5

Automated Cleaning

The syringe maintenance cycle is always advisable after the injection of pure samples, of undiluted mixes, of liquids containing solids in suspension or, at any rate, after long sequence of analyses.
To perform an automated cleaning cycle proceed as follows:
utosampler method one or more rinse with
sample to maintain the lubrication of the plunger.
r method wash cycles of extra solvents (e.g. A+B, where A is used to clean and B to rinse). The first solvent should have a higher cleaning power than the second solvent. The latter should be chemically similar to the sample scheduled for injection.
ge 78).
on pa
. Perform a manual maintenance routine every 1000 injections (see “Manual Cleaning” on
pag
e 78).

Manual Cleaning

1. Fill a 4 mL vial with fresh solvent with high cleaning power (acetone is recommended) pl
ace it in the D position of the washing tray of the AI 3000/AS 3000 II.
and
2. From your PC, enter the Data System and then access the AI 3000/AS 3000 II method
ge. Select the following parameters:
pa
• Pre-injection Solvent: D
• Pre-injection Cycles: 15
3. Enter the Fil
4. Enter the Comma
5. If necessary, repeat the step 4 of this sequence.
CAUTION This operating sequence is not applicable to the 0.5 μL syringe. For details, refer to the Nano-Volumes Kit for S/SL Inlet System User Guide.
e menu and click on Send Method to Sampler.
nd menu and click on Start Clean Cycle.
78 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
Page 97
5
Maintenance

Preparing a New Syringe

To perform a manual cleaning cycle proceed as follows:
1. Remove the syringe as described in the operating sequence “Replacing the Syringe” on
pag
e 77.
2. Completely take out the syringe plunger and clean it gently using a soft cloth or a paper
ue wetted with high cleaning power solvent (acetone is recommended).
tiss
3. Reintroduce the syringe plunger and gently mov
drawing an organic solvent (e.g. hexane).
4. Reinstall the syringe as described in the operating sequence “Replacing the Syringe” on
e 77.
pag
Preparing a New Syringe
This operating procedure is always advisable before using a new syringe with low capacity (like a 5 or 0.5 μL) or a new 10 μL syringe with Teflon tip plunger. The following procedure will allow to obtain the proper lubrication of the plunger.
To prepare a new syringe proceed as follows:
1. Fill a 4 mL vial with fresh solvent (hexane is recommended) and place it in the D position
f
the washing tray of the AI 3000/AS 3000 II.
o
2. From your PC, enter the Data System and then access the AI 3000/AS 3000 II method
ge. Select the following parameters:
pa
• Pre-injection Solvent: D
• Pre-injection Cycles: 15
e it repeatedly along its whole travel
3. Enter the Fil
4. Enter the Comm
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 79
e menu and click on Send Method to Sampler.
and menu and click on Start Clean Cycle.
Page 98
5
Maintenance

Moving Away the Sampling Unit from the Injector Module

Moving Away the Sampling Unit from the Injector Module
When the maintenance of the injector module is required, it is necessary to move away the sampling system from the injector module.
Front/Back Configuration
1. Power off the AI 3000/AS 3000 II.
2. Remove the centering plate from the injector nut and the sample tray from the sampling
unit
base.
Note It is not necessary to remove the sampling unit from the pivot guide.
3. Remove the plates locking screw. The top plate is now free to move outwards allowing the
m
pling unit to be moved away from the injector module.
sa
4. Carefully push the sampling unit outwards up to reach the end run of the top plate.
5. When the maintenance of the injector module has been performed, restore the position of
the sampling unit on the injector module.
6. Block the two plates by using the plates locking screw.
7. Reinsert the centering plate into its seat of the sampling unit base paying attention that
the guide ho
le, present on the arm of the centering plate, correctly fits the injector nut.
Gemini Configuration
1. Power off the AI 3000/AS 3000 II.
2. Remove the centering plate from the injector nut and from its seat on the sampling unit
base
.
3. Remove the sample tray from the sampling unit base.
Note It is not necessary to remove the sampling unit from the pivot guide.
4. Undo that screw that secure the retainer. The top plate is now free to move outwards
wing the sampling unit to be moved away from the injector module.
allo
5. Carefully push the sampling unit outwards up to reach the end run of the top plate.
6. When the maintenance of the injector module has been performed, restore the position of
the sampling unit on the injector module.
7. Tighten the screw that secure the retainer.
8. Reinsert the centering plate into its seat of the sampling unit base paying attention that
the guide ho
9. Reinsert the sample tray into its seat of the sampling unit base.
80 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific
le, present on the arm of the centering plate, correctly fits the injector nut.
Page 99
A

Abbreviations

This section lists and defines terms used in this guide. It also includes acronyms, metric prefixes, symbols and abbreviations
A B C D E F G H I J K L M N O P Q R S T V V W X Y Z
A
A ampere
ac alternating current
ADC analog-to-digital converter
B
b bit
B byte (8 b)
baud rate data transmission speed in events per second
C
C Carbon
°C Celsius
CIP Carriage and Insurance Paid To
cm centimeter
CPU central processing unit (of a computer)
<Ctrl>
Thermo Scientific AI 3000/AS 3000 Sampling System - Operating Manual 81
control key of the keyboard
Page 100
A
D
d depth
DAC digital-to-analog converter
dc direct current
DS data system
E
EMC electromagnetic compatibility
ESD electrostatic discharge
F
ºF Fahrenheit
FOB Free on board
FSE Field Service Engineer
ft foot
G
g gram
GC gas chromatography- gas chromatograph
GND electrical ground
H
h height
h hour
harmonic distortion A high-frequency disturbance that appe
fundamental sine wave
HV high voltage
ars as distortion of the
Hz hertz (cycles per second)
82 AI 3000/AS 3000 Sampling System - Operating Manual Thermo Scientific