Thermo Scientific 9609BNWP Instruction Manual

User Guide
Fluoride Ion Selective Electrode
Ross and the COIL trade dress are trademarks of Thermo Fisher Scientifi c, Inc. and its subsidiaries.
AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, ROSS, ROSS Ultra, Sage, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientifi c, Inc. and its subsidiaries.
A+, All in One, Aplus, AUTO-BAR, AUTO-CAL, Auto-ID, AUTO­READ, AUTO-STIR, Auto-Test, AutoTration, CISA, digital LogR, DuraProbe, EZ Startup, ISEasy, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, Optimum Results, Orion Dual Star, Orion Star, SAOB, SMART AVERAGING, SMART STABILITY, Star LogR, Star Navigator 21, Stat Face, Triode are trademarks of Thermo Fisher Scientifi c, Inc. and its subsidiaries.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientifi c, Inc. and its subsidiaries.
© 2011 Thermo Fisher Scientifi c Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientifi c, Inc. and its subsidiaries.
The specifi cations, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Electrode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Electrode Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Checking Electrode Operation (Slope) . . . . . . . . . . . . . . . . . . . . . 6
Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Sample Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Measuring Hints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Electrode Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Electrode Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Analytical Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Direct Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Low-Level Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Single Known Addition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Titrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Fluoride in Acid Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Fluoride in Alkaline Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrode Characteristics . . . . . . . . . . . . . . . . . . . . . . . . .33
Electrode Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Reproducibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Limits of Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Temperature Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Interferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
pH Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Complexation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Troubleshooting Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Specifi cations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Introduction
This user guide contains information on the preparation, operation and maintenance for the fl uoride ion selective electrode (ISE). General analytical procedures, electrode characteristics and electrode theory are also included in this user guide. Fluoride electrodes measure free fl uoride ions in aqueous solutions quickly, simply, accurately and economically.
The measurement of fl uoride in drinking water and wastewater using an ion selective electrode is an EPA approved test procedure. EPA-approved ISE test procedures for fl uoride in drinking water are ASTM D1179-04 (B) and Standard Methods 4500-F- C (21st edition). EPA-approved ISE test procedures for fl uoride in wastewater are ASTM D1179-93, 99 (B) and Standard Methods 4500-F- C-97 (online) or 4500-F- B (18, 19, 20th editions). Approved test method lists are updated periodically. Check with your regulatory agency for latest approve method revisions.
Technical Support Chemists can be consulted for assistance and troubleshooting advice. Within the United States call
1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermoscientifi c.com/water.
Fluoride ionplus® Sure-Flow® Solid State Combination ISE
The reference and sensing electrodes are built into one electrode, which decreases the amount of required solution and reduces waste. The built-in Sure-Flow reference junction prevents electrode clogging and provides fast and stabile readings. The fl uoride ionplus combination ISE is available with a waterproof BNC connector, Cat. No. 9609BNWP. Electrodes with a waterproof BNC connector can be used on any ISE meter with a BNC connection.
Fluoride Solid State Half-Cell ISE
The fl uoride half-cell electrode must be used with the single junction reference electrode, Cat. No. 900100. The fl uoride half-cell is available with a BNC connector, Cat. No. 9409BN; and a screw cap connector, Cat. No. 9409SC. Electrodes with a screw cap connector require a separate cable.
1Fluoride Ion Selective Electrode User Guide
inner
body
filling solution
chamber
outer body
reference
element
cap
spring
cable
O-ring
filling hole
inner cone
sensing crystal
Figure 1 9609BNWP Fluoride Combination Electrode
2 Fluoride Ion Selective Electrode User Guide
Required Equipment
1. Thermo Scientifi c Orion ISE meter, such as the Orion 4-Star pH/ISE meter, Orion 5-Star pH/ISE/DO/conductivity meter or Orion DUAL STAR meter.
Fluoride electrodes can be used on any ISE meter with
a BNC connection. The electrodes can also be used on meters with a variety of inputs when an adapter cable is used. Visit www.thermoscientifi c.com/water for details.
2. Thermo Scientifi c Orion fl uoride ion selective electrode.
The 9409BN and 9409SC fl uoride half-cell electrodes
require a separate reference electrode, Cat. No. 900100.
3. Magnetic stirrer or Orion stirrer probe, Cat. No. 096019. The Orion stirrer probe can be used with 3-Star, 4-Star, 5­Star benchtop meters, and Dual Star meter.
4. Volumetric fl asks, graduated cylinders and beakers. Plastic labware is highly recommended for fl uoride analysis.
5. Distilled or deionized water.
6. Fluoride electrode fi lling solution.
Use Optimum Results A fi lling solution, Cat. No. 900061,
for the 9609BNWP fl uoride electrodes.
Use single junction reference electrode fi lling solution,
Cat. No. 900001, for the reference electrode used with the 9409BN and 9409SC fl uoride electrodes.
7. Fluoride calibration standards.
Cat. No. Description
940906 0.1 M NaF, 475 mL bottle
940907 100 ppm F¯, 475 mL bottle
040908 10 ppm F¯ with TISAB II, 475 mL bottle
040907 2 ppm F¯ with TISAB II, 475 mL bottle
040906 1 ppm F¯ with TISAB II, 475 mL bottle
3Fluoride Ion Selective Electrode User Guide
8. Total Ionic Strength Adjustment Buffer (TISAB), which provides a constant background ionic strength, decomplexes fl uoride ions and adjusts the solution pH.
Cat. No. Description
940909 TISAB II, 3.8 L bottle
940999 TISAB II, 4 x 3.8 L bottle
940911 TISAB III concentrate, 475 mL bottle
Low-Level TISAB
For use when measuring in samples containing less than 0.4 ppm (2 x 10 as iron or aluminum are present.
To prepare low level TISAB: Place 500 mL of distilled water in a 1 liter beaker. Add 57 mL of glacial acetic acid and 58 g of reagent grade sodium chloride to the beaker. Place the beaker in a water bath for cooling. Immerse a calibrated pH electrode into the solution and slowly add 5 M NaOH until the pH is between 5.0 and 5.5. Cool the solution to room temperature. Pour the solution into a 1 liter volumetric fl ask and dilute to the fl ask mark with distilled water. All reagents must be as pure as possible to keep the fl uoride level low in the buffer.
Use as directed for TISAB II; combine equal volumes of low level TISAB and sample or standard before taking any measurements.
-5
M) fl uoride and no fl uoride complexing agents such
TISAB IV
TISAB IV will complex more than 100 ppm of iron or aluminum in the presence of 1 ppm fl uoride. A measurement of 1 ppm fl uoride will be in error by 5% in the presence of 200 ppm iron or aluminum.
To prepare TISAB IV: Place 500 mL of distilled water in a 1 liter volumetric fl ask. Add 84 mL of concentrated HCl (36 to 38 %), 242 g TRIS (hydroxymethyl) aminomethane and 230 g sodium tartrate (Na and cool the solution to room temperature. Dilute to the fl ask mark with distilled water.
-2H2O) to the fl ask. Stir to dissolve the solids
2C4H406
Use as directed for TISAB II; combine equal volumes of TISAB IV and sample or standard before measurements.
4 Fluoride Ion Selective Electrode User Guide
Electrode Setup
Electrode Preparation
9409BN and 9409SC Fluoride Half-Cell Electrode– Remove the protective shipping cap from the sensing element and save the cap for storage.
900100 Single Junction Reference Electrode– Prepare the reference electrode according to the reference electrode user guide. Fill the reference electrode with single junction reference fi lling solution, Cat. No. 900001.
9609BNWP Fluoride Combination Electrode– Remove the protective shipping cap from the sensing element and save the cap for storage. Fill the electrode with Optimum Results A fi lling solution, Cat. No. 900061.
Note: Optimum Results A fi lling solution minimizes junction potential issues and fl uoride contamination in the sample. The use of any other fi lling solution will void the electrode warranty.
Electrode Filling Instructions
1. Lift the fl ip spout on the fi lling solution bottle to a vertical position.
2. Insert the spout into the fi lling hole on the outer body of the electrode and add a small amount of fi lling solution to the reference chamber. Invert the electrode to moisten the top O-ring and then return the electrode to the upright position.
3. Hold the electrode body with one hand and use your thumb to push down on the electrode cap to allow a few drops of fi lling solution to drain out of the electrode.
4. Release the electrode cap. If the sleeve does not return to its original position, check if the O-ring is moist and repeat steps 2-4 until the sleeve returns to the original position.
5. Add fi lling solution to the electrode up to the fi lling hole.
Note: Add fi lling solution each day before using the electrode. The fi lling solution level should be at least one inch above the level of sample in the beaker to ensure a proper fl ow rate. The fi ll hole should always be open when taking measurements.
5Fluoride Ion Selective Electrode User Guide
Checking Electrode Operation (Slope)
These are general instructions that can be used with most meters to check the electrode operation. Refer to the meter user guide for more specifi c information.
This procedure measures electrode slope. Slope is defi ned as the change in millivolts observed with every tenfold change in concentration. Obtaining the slope value provides the best means for checking electrode operation.
1. If the electrode has been stored dry, prepare the electrode as described in the Electrode Preparation section.
2. Connect the electrode to the meter. Set the meter to the mV mode.
3. Place 50 mL of distilled water and 50 mL of TISAB II (Low­level TISAB or TISAB IV) into a 150 mL beaker. Stir the solution thoroughly.
If using TISAB III, place 90 mL of distilled water and
10 mL of TISAB III into a 150 mL beaker. Stir the solution thoroughly.
4. Rinse the electrode with distilled water and place the electrode into the solution prepared in step 3.
5. Select either a 0.1 M sodium fl uoride or 100 ppm fl uoride standard. Pipette 1 mL of the standard into the beaker and stir thoroughly. When a stable reading is displayed, record the electrode potential in millivolts.
6. Pipette 10 mL of the same standard into the same beaker and stir thoroughly. When a stable reading is displayed, record the electrode potential in millivolts.
7. There should be a 54 to 60 mV difference between the two millivolt readings when the solution temperature is between 20 to 25 ˚C. If the millivolt potential is not within this range, refer to the Troubleshooting section.
6 Fluoride Ion Selective Electrode User Guide
Measurement Units
Fluoride concentration can be measured in moles per liter (M), parts per million (ppm) or any convenient concentration unit.
Table 1 Concentration Unit Conversion Factors
Moles/Liter (M) ppm
1.0 19000
-1
10
-2
10
-3
10
-4
10
1900
190
19
1. 9
Sample Requirements
The epoxy body of the fl uoride electrode is resistant to damage by inorganic solutions. The electrode may be used intermittently in solutions that contain methanol or acetone. Contact Technical Support for information on using the electrode for specifi c applications.
Samples and standards should be at the same temperature. The solution temperature must be less than 100 °C.
In all analytical procedures, TISAB must be added to all samples and standards before measurements are taken.
7Fluoride Ion Selective Electrode User Guide
Measuring Hints
• Add 50 mL of TISAB II (Low-level TISAB or TISAB IV) to every 50 mL of standard or sample. Add 10 mL of TISAB III to every 90 mL standard or sample. Once TISAB II (Low­level TISAB or TISAB IV) or TISAB III is selected, it should be added to all samples and standards so the dilution ratio of TISAB to solution remains the same.
• Stir all standards and samples at a uniform rate.
• Always use fresh standards for calibration.
• Always rinse the electrode with deionized water between measurements and shake the electrode to remove the water and prevent sample carryover. Do not wipe or rub the electrode sensing element.
• Allow all standards and samples to come to the same temperature for precise measurements.
Place a piece of insulating material, such as Styrofoam or cardboard, between the magnetic stirrer and beaker to prevent measurement errors from the transfer of heat to the sample.
• Verify the electrode calibration every two hours by placing the electrode in a fresh aliquot of the least concentrated standard used for calibration. If using more than 2 standards, use a mid-level standard. If the value has changed by more than 2%, recalibrate the electrode.
• After immersing the electrode in a solution, check the electrode sensing surface for air bubbles and remove air bubbles by reimmersing the electrode in the solution.
• For high ionic strength samples, prepare standards with a background composition similar to the sample or use manual known addition procedure or a Dual Star meter.
• Adjust highly acidic or highly basic solutions to pH 5 - 6 before adding TISAB.
• Remove the fi lling hole cover during measurements to ensure a uniform fl ow of fi lling solution.
• If the electrode is used in dirty or viscous samples or the electrode response becomes sluggish, empty the electrode completely, hold the junction open and fl ush the junction with deionized water. Empty any water from the electrode and refi ll it with fresh fi lling solution. Press down on the electrode cap to let a few drops of the fi lling solution fl ow out of the electrode and then replenish any lost solution.
8 Fluoride Ion Selective Electrode User Guide
Electrode Storage
9409BN and 9409SC Fluoride Half-Cell Electrode
The fl uoride half-cell electrodes should be rinsed thoroughly with distilled water and stored dry in the air at all times. When storing the electrode for long periods of time, cover the sensing element with the protective shipping cap.
900100 Single Junction Reference Electrode
The single junction reference electrode may be stored in the single junction electrode fi lling solution, Cat. No. 900001, between sample measurements and up to one week. A 0.01 M or 100 ppm fl uoride standard is also an acceptable storage solution. The fi lling solution inside the electrode should not be allowed to evaporate, as crystallization will result.
For storage longer than one week, drain the reference electrode, fl ush the inside with distilled water and store the electrode dry.
9609BNWP Fluoride Combination Electrode
For storage between measurements and up to one week, store the electrode in electrode fi ll solution, Cat. No. 900061, with fl uoride added. The fl uoride concentration of the storage solution should be close to least concentrated fl uoride calibration standard. Do not add TISAB to the storage solution. The fi lling solution should not be allowed to evaporate, as crystallization will result. If fi ll solution is unavailable, store in 2 M potassium nitrate with fl uoride added.
For storage longer than one week, drain the electrode. Flush the chamber with distilled water. Disassemble the electrode as described in the electrode maintenance section. Rinse the inner sleeve with distilled water and blot dry with a Kimwipe. Do not touch the reference element other than to blot dry. Let air dry and then reassemble for storage. Use the protective shipping cap to cover the sensing element.
9Fluoride Ion Selective Electrode User Guide
Electrode Maintenance
Polishing the Sensing Surface of the Fluoride Combination Electrodes and Fluoride Half-Cell Electrodes
The sensing surface of the fl uoride electrode might need restoration over time as deposits coat the sensing surface. The sensing crystal can be restored using the following procedure. Toothpaste that contains fl uoride will be needed for this procedure.
1. Dispense a small amount of fl uoridated toothpaste on a kimwipe. Add a few drops of deionized water. A soft toothbrush can also be used in placed of a kimwipe.
2. Rub the toothpaste on the sensing element on the bottom of the electrode in a circular motion for around one minute.
3. Rinse the electrode thoroughly with DI water, fl ush the electrode fi ll solution out (if using a combination fl uoride electrode) by pressing on the cap of the electrode) then refi ll with fresh fi ll solution.
4. Soak the electrode in 100 ppm fl uoride standard for 2 hours.
Flushing the 9609BNWP and 900100 Single Junction Reference Electrode
If the area between the electrode sleeve and inner cone becomes clogged with sample or precipitate, fl ush the area with fi lling solution or distilled water.
1. Hold the electrode body with one hand and use your thumb to push down on the electrode cap to drain the chamber. Push down on the cap until all the fi lling solution is drained from the chamber.
2. Fill the electrode with distilled water and then push down on the cap until all the water is drained from the chamber.
3. Fill the electrode with fresh fi lling solution up to the fi ll hole. Push down on the cap to allow a few drops of fi lling solution to drain out of the electrode and then refi ll any lost fi lling solution.
10 Fluoride Ion Selective Electrode User Guide
Disassembling the Fluoride Combination Electrodes and Single Junction Reference Electrode
Note: Disassembly is usually not required and should not be done unless a thorough cleaning is required.
1. Tip the electrode so the fi lling solution moistens the O-ring on the electrode body. Hold the electrode body with one hand and use your thumb to push down on the electrode cap to drain the chamber.
2. Unscrew the cap counter clock-wise and then slide the cap and the spring up the cable.
3. Hold the outer sleeve with one hand and fi rmly push down on the threaded portion with the thumb and forefi nger to separate the inner body from the sleeve.
4. Grasp the cone with a clean, lint-free tissue and withdraw the body from the sleeve using a gentle twisting motion. Do not touch the reference electrode element (coiled wire in combination electrode, pellet in single junction reference) above the cone, it will damage to the reference element. Rinse the outside of the electrode body and the entire sleeve with distilled water. Blot dry with a Kimwipe and allow to air dry if needed.
Reassembling the Fluoride Combination Electrodes and Single Junction Reference Electrode
1. Moisten the O-ring on the electrode body with a drop of fi lling solution. Insert the screw-thread end of the electrode body into the tapered, ground end of the sleeve.
2. Push the body into the sleeve using a gentle twisting motion until the bottom surface of the inner cone is fl ush with the tapered end of the sleeve.
3. Place the spring onto the electrode body and screw on the cap. Refi ll the electrode with fi lling solution.
11Fluoride Ion Selective Electrode User Guide
Loading...
+ 33 hidden pages