Thermo Scientific 9512BNWP Instruction Manual

TABLE OF CONTENTS
GENERAL INFORMATION 1
Introduction 1 Required Equipment 1 Required Solutions 2
USING THE ELECTRODE 3
Set Up 3
Electrode Assembly 3 Electrode Preparation 4 Checking Electrode Operation (Slope) 7
Before Analysis 8
Units of Measurement 8 Sample Requirements 8 Measuring Hints 8 Sample Storage 9
Analytical Procedures 10
Analytical Techniques 10 Direct Calibration 11 Low Level Measurements By Direct Calibration 14 Known Addition 16
Electrode Storage 19
TROUBLESHOOTING 22
Troubleshooting Checklist 22 Troubleshooting Guide 26 Assistance 28
ELECTRODE CHARACTERISTICS 29
Electrode Response 29 Reproducibility 30 Temperature Effects 30 Interferences 30 pH Effects 31 Complexation 31 Effect of Dissolved Species 31 Electrode Life 31 Theory of Operation 32
ORDERING INFORMATION 37
SPECIFICATIONS 38
1
GENERAL INFORMATION
Introduction
The Orion 95-12 Ammonia Electrode allows fast, simple, economical, and accurate measurements of dissolved ammonia in aqueous solutions. This gas-sensing electrode can also be used to measure the ammonium ion after conversion to ammonia, or organic nitrogen after Kjeldahl digestion of the sample. Sample color and turbidity do not affect the measurement, and samples need not be distilled. Almost all anions, cations, and dissolved species, other than volatile amines, do not interfere.
All apparatus and solutions required for measurement, electrode characteristics and electrode theory are discussed in this manual. General analytical procedures, low-level procedures and a method for measuring ammonia in solutions that wet the membrane are given.
The Orion 95-12 comes with the following:
• forty loose membranes
• reusable membrane cap
• tweezers for handling membranes
• dispensing cap
• electrode filling solution
• electrode instruction manual.
Operator instructions for Orion Meters are outlined in the individual meter’s instruction manual. Our Technical Service Chemists can be consulted for assistance and troubleshooting advice. Please refer to Troubleshooting Section for information on contacting Technical Services.
2
Required Equipment
Meter
The easiest type of meters to use is direct concentration readout specific ion meter, such as Orion EA940, 920A, 920Aplus, 720A, 720Aplus, 710A, 710Aplus, 290A, 290Aplus or 370. If unavailable, a pH/mV meter with readability to 0.1 mV could be used.
Magnetic Stirrer
Highly recommended for laboratory measurements.
Graph Paper
4-cycle semilogarithmic paper for preparing calibration curves (for use with digital pH/mV laboratory meters)
Required Solutions
Distilled or Deionized Water
Water MUST be ammonia-free. Pass distilled or deionized water through an ion-exchange column containing a strong acidic cation exchange resin, such as Dowex 50W-X8.
Standard Solutions Orion
Ionic Strength Adjustor (ISA) 951211 Electrode Filling Solution 951202
0.1 M Ammonium Chloride Standard 951006 1000 ppm as Nitrogen Standard 951007
Solutions Prepared by Customer
For inner body check:
To check the operation of the electrode inner body.
Solution 1 — pH 4.01 Buffer with 0.1M NH
4
Cl (or 0.1 M NaCl)
Take 200 mL of the pH 4.01 buffer solution, Orion 910104, add 1.07 g reagent-grade NH
4
Cl (or 1.16g reagent-grade NaCl), stir to mix and label bottle as Solution 1. Store the buffer for repeated use. Discard buffer if turbidity develops.
Solution 2 — pH 7.00 Buffer with 0.1M NH
4
Cl (0.1 M NaCl)
Take 200 mL of pH 7.00 buffer solution, Orion 910107, add 1.07 g reagent-grade NH
4
Cl (or 1.16g reagent-grade NaCl), stir to mix and label bottle as Solution 2. Store the buffer for repeated use. Discard bottle if turbidity develops.
3
USING THE ELECTRODE
Set Up
Electrode Assembly
NOTE: Soak inner body in electrode filling solution for at least two hours before assembling new electrode. For best results, soak inner body overnight.
Assemble electrode according to instructions below. The electrode is shipped dry with loose membranes and a loose membrane cap. Avoid excessive handling of the membrane during assembly; this may affect the membrane’s hydrophobic properties, causing shortened membrane life. Use the tweezers provided. A membrane will last from one week to several months depending on usage.
NOTE: Membrane failure is characterized by a shift in electrode potential, drift, and poor response. See Troubleshooting section. Membrane failure may be apparent on visual inspection as dark spots or discoloration of the membrane.
4
Electrode Preparation
NOTE: Soak inner body in electrode filling solution for at least two hours before assembling electrode. For best results, soak inner body overnight.
1. While holding the probe vertically, unscrew the top cap (a)
and carefully remove the glass electrode inner body (b) from the electrode outer body (c) (See Fig. 1). Set cap with inner body aside carefully. Pour out remaining electrode filling solution from outer body if applicable, new probe will not have any filling solution in the body.
2. Unscrew membrane cap (d) from outer body (See Fig. 2).
Remove old membrane, if applicable.
3. Wearing gloves, use tweezers to carefully grasp corner of
white membrane from between paper separators (See Fig. 3). Do not to touch center of membrane.
(a)
(b)
(c)
(d)
Fig. 1
(d)
Fig. 2
Fig. 3
4. With free hand, grasp electrode outer body with threads
oriented to the top. Align straight edge of membrane against threaded shoulder and hold with thumb (See Fig. 4). With other hand, gently stretch the membrane upwards (See Fig. 5), then across the opening (See Fig. 6), then down to align other edge with the opposite shoulder (See Fig. 7). While holding each edge on both sides, gently stretch each serrated side of the membrane out and down over threads making sure that membrane surface is smooth and without wrinkles (See Fig. 8). Smooth any loose material taking care not to touch center of membrane (See Fig. 9).
5
Fig. 4
Fig. 7
Fig. 8
Fig. 5
Fig. 6
Fig. 9
6
NOTE: Do not overstretch the membrane. The membrane is considered overstretched if the black body material can be seen through the membrane. Discard membrane if overstretched and replace with new membrane. Continue from Step 3.
5. Replace membrane cap being careful not to touch membrane center (See Fig. 10). Screw down half way and tuck any loose membrane material under cap while twisting. Make sure cap is fully screwed on.
6. Fill the outer body with 2.5 ml filling solution, or approximately 50 drops from the bottle (See Fig. 11).
7. Place inner body into outer body containing filling solution and screw on upper cap (See Fig. 12).
8. Carefully shake fully assembled electrode as if it were a clinical thermometer to remove bubbles. Gently pull spring loaded cable back and slowly release to allow filling solution to migrate between membrane and glass electrode inner body (See Fig. 13).
9. Rinse the electrode well and wipe dry.
Fig. 10
Fig. 11
Fig. 12 Fig. 13
Checking Electrode Operation (Slope)
These are general instructions, which can be used with most meters to check electrode operation. See individual meter instruction manuals for more specific information.
This procedure measures electrode slope. Slope is defined as the change in millivolts observed with every tenfold change in concentration. Obtaining the slope value provides the best means for checking electrode operation.
1. If electrodes have been stored dry, prepare the electrodes as described under the section entitled Electrode Preparation.
2. Connect electrodes to the meter. For an electrode with a BNC connector, turn the connector clockwise to attach to the meter. For an electrode with a U.S. Standard connector, insert reference pin-tip connector and the sensing electrode connector into appropriate jacks on the meter. Non-Orion meters may require special adaptors. Consult your meter instruction manual.
3. Place 100 mL distilled water into a 150 mL beaker. Add 2 mL pH-adjusting ISA, Orion 951211. Stir thoroughly. Set the meter to the mV mode.
4. Rinse electrode with distilled water and place in the solution prepared in step 3 above. To prevent air entrapment on the membrane surface, be sure to use an electrode holder that keeps the electrode at a 20˚ angle. If bubbles appear on the sensing membrane, tap electrode gently to remove.
5. Select either 0.1 M or 1000 ppm ammonium chloride standard. Pipet 1 mL of the standard into the beaker, stir thoroughly. When a stable reading is displayed, record the electrode potential in millivolts.
6. Pipet 10 mL of the same standard into the same beaker. Stir thoroughly. When a stable reading is displayed, record the electrode potential in millivolts.
7. Take the first potential reading and subtract it from the second one. The difference should be in the range of
-54 to -60 mV/decade when the solution temperature is between 20-25 ˚C. If the potential is not within this range, refer to Troubleshooting.
7
Before Analysis
Units of Measurement
Ammonia can be measured in units of moles per liter, parts per million as ammonia, parts per million as nitrogen, or any other convenient unit (see Table 1).
Table 1 Concentration Unit Conversion Factors
Moles/Liter ppm as N ppm as NH
3
10
-4
1.40 1.70
10
-3
14.0 17.0
10
-2
140 170
10
-1
1400 1700
Sample Requirements
Samples must be aqueous; they must not contain organic solvents. Consult Our Technical Service Chemists for use of the electrode in unusual applications. Samples and standards should be at the same temperature. A 1 ˚C difference in temperature will give rise to about 2% measurement error. In all analytical procedures, pH­adjusting ISA must be added to all samples and standards immediately before measurement. After addition of the ISA all solutions should fall within a pH 11 to 14 range (solution should be blue in this range) and have a total level of dissolved species below 1 M. If the total level of dissolved species is above 1 M, see section entitled Effect of Dissolved Species.
Measuring Hints
Minimize NH3loss from sample by following the recommendations below:
Store samples according to procedure in Sample Storage.
Use beakers that minimize the ratio of surface area to volume.
Keep beakers containing standards and samples covered between measurements.
8
Add 2 mL of pH-adjusting ISA to 100 mL of sample or standard immediately before making measurements, and make sure the blue color is observed.
Between measurements, rinse the electrode with distilled water.
Be sure samples, standards, and electrodes are at the same temperature.
Samples and standards should be stirred using a magnetic stirrer. Some magnetic stirrers generate enough heat to change solution temperature. Placing a piece of insulating material such as cork or styrofoam between the beaker and the stirring plate can minimize this effect.
Verify calibration every two hours by placing electrodes in a fresh aliquot of the first standard solution used for calibration. If value has changed, recalibrate the electrode.
Always use fresh standards for calibration.
After immersion in solution, check electrode for any air bubbles on membrane surface and remove by shaking the electrode as you would with a thermometer or lightly tapping of the electrode.
If electrode response is slow, the membrane may contain a surface layer of contaminant. Restore performance by soaking electrode in distilled water for about 5 minutes, then rinse and soak in a standard solution for about 1 hour before use.
Sample Storage
If possible, alkaline samples should be measured at once. The rate of ammonia loss at room temperature from a stirred 100 mL basic solution in a 100 mL beaker is about 50% in six hours. If samples must be stored, make them slightly acidic (pH 6) by adding 0.5 mL of 1 M HCl to each liter of sample, and place them in tightly capped vessels. Make stored samples basic with pH-adjusting ISA immediately before measurement.
9
Analytical Procedures
Analytical Techniques
A variety of analytical techniques are available to the analyst. The following is a description of these techniques.
Direct Calibration is a simple procedure for measuring a large number of samples. Only one meter reading is required for each sample. Calibration is performed in a series of standards. The concentration of the samples is determined by comparison to the standards. ISA is added to all solutions to ensure that samples and standards have similar ionic strength.
Incremental Techniques provide a useful method for measuring samples, since calibration is not required. As in direct calibration, any convenient concentration unit can be used. The different incremental techniques are described below. They can be used to measure the total concentration of a specific ion in the presence of a large (50-100 times) excess of complexing agents.
Known Addition is an alternate method useful for measuring dilute samples, checking the results of direct calibration (when no complexing agents are present), or measuring the total concentration of an ion in the presence of an excess complexing agent. The electrodes are immersed in the sample solution and an aliquot of a standard solution containing the measured species is added to the sample. From the change in potential before and after the addition, the original sample concentration is determined.
Known Subtraction is useful as a quick version of a titration, or for measuring species for which stable standards do not exist. It is necessary to know the stoichiometric ratio between standard and sample. For known subtraction, an electrode sensing the sample species is used. Stable standards of a species reacting completely with the sample in a reaction of known stoichiometry are necessary.
Analate Addition is often used to measure soluble solid samples, viscous samples, small or very concentrated samples, to diminish the effects of complex sample matrices, or to diminish the effects of varying sample temperatures. This method is not suitable for dilute or low concentration samples. Total concentration is measured even in the presence of complexing agents. The electrodes are immersed in a standard solution containing the ion to be measured and an aliquot of the sample is added to the standard. The original sample concentration is determined from the change in potential before and after the addition.
10
Analate Subtraction is used in the measurement of ions for which no ion-selective electrode exists. The electrodes are immersed in a reagent solution which contains a species that the electrode senses, and that reacts with the sample. It is useful when sample size is small, or samples for which a stable standard is difficult to prepare, and for viscous or very concentrated samples. The method is not suited for very dilute samples. It is also necessary to know the stoichiometric ratio between standard and sample. Specific instructions for known subtraction, analate addition and analate subtraction, can be found in the Orion meter instruction manuals. Titrations are quantitative analytical techniques for measuring the concentration of a species by incremental addition of a reagent (titrant) that reacts with the sample species. Sensing electrodes can be used for determination of the titration end point. Ion-selective electrodes are useful as end point detectors, because they are unaffected by sample color or turbidity. Titrations are approximately 10 times more precise than direct calibration, but are more time-consuming.
Direct Calibration
Set Up
1. Prepare the electrode as described in Electrode Assembly and Electrode Preparation.
2. Connect electrode to meter.
3. Prepare two standards which bracket the expected sample range and differ in concentration by a factor of ten. Standards can be prepared in any concentration unit to suit the particular analysis requirement. All standards should be at the same temperature as the samples. (For details on temperature effects on electrode performance, refer to Temperature Effects.)
11
Loading...
+ 27 hidden pages