Thermo Scientific 9156APWP, 9104APWP, 9107APMD, 9115APWP, 9135APWP Instruction Manual

...
User Guide
AquaPro and Double Junction pH Electrodes
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent 6,168,707.
PerpHecT ROSS are protected by U.S. patent 6,168,707.
ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.
ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338.
ROSS Ultra electrodes are protected by US patents 6,793,787.
Orion ORP Standard is protected by US Patent 6,350,367.
Orion NoCal electrodes are protected by US Patent 7,276,142.
© 2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Introduction
This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion AquaPro and double junction pH electrodes.
AquaPro Professional pH Electrodes
All AquaPro pH electrodes feature a patented double junction reference design and are mercur y-free. These electrodes use an isolated Ag/AgCl internal reference and a polymer external reference with an open junction for fast, accurate readings in complex sample matrices without the sample coming in contact with silver ions. The AquaPro electrodes are easy to use and require minimal electrode preparation due to the low maintenance polymer gel reference.
Double Junction pH Electrodes
Double junction pH electrodes are for general purpose samples and feature an isolated Ag/AgCl internal reference and an external reference that is either refillable or gel-filled. These electrodes offer a mercury-free alternative to hazardous calomel pH electrodes.
KNIpHE® Double Junction pH Electrodes
The KNIpHE double junction pH electrode is designed for meat, cheese and other solid or semi-solid food samples. The pH electrode is surrounded by a stainless steel blade that protects the electrode and cuts through solid or tough materials. This electrode is mercury-free and environmentally safe.
1AquaPro and Double Junction pH Electrode User Guide
Cat. No. Description
9102AP AquaPro combination pH electrode with
glass body and BNC connector
9103APWP AquaPro combination pH electrode with
glass body, semi-micro tip, and waterproof BNC connector
9104APWP AquaPro combination pH electrode with
glass body, rugged bulb, and waterproof BNC connector
9107APMD AquaPro pH/ATC Triode with epoxy body
and waterproof BNC and 8 pin MiniDIN connectors
9115APWP AquaPro combination pH electrode with
epoxy body, semi-micro tip, and waterproof BNC connector
9135APWP AquaPro combination pH electrode with
epoxy body, flat surface tip, and waterproof BNC connector
9156APWP AquaPro combination pH electrode with
epoxy body and waterproof BNC connector
9102DJWP
9110DJWP Double junction refillable combination pH
9156DJWP Double junction gel-filled combination pH
9120APWP KNIpHE double junction refillable
Double junction refillable combination pH electrode with glass body and waterproof BNC connector
electrode with glass body, semi-micro tip, and waterproof BNC connector
electrode with epoxy body and waterproof BNC connector
combination pH electrode with stainless steel cutting blade and waterproof BNC connector
• All electrodes with a waterproof BNC and BNC connector have one meter cable lengths.
• Electrodes with a waterproof BNC connector can be used on any pH meter with a BNC connection.
2 AquaPro and Double Junction pH Electrode User Guide
Required Equipment
1. Thermo Scientific Orion pH meter, such as the 3-Star pH meter, 4-Star pH/ISE meter, 4-Star pH/DO meter, 4-Star pH/ conductivity meter or 5-Star pH/ISE/DO/conductivity meter.
AquaPro and double junction pH electrodes can be used
on any pH meter with a BNC connection. The electrodes can also be used on meters with a variety of inputs when an adapter cable is used. Visit www.thermo.com/water for details.
The AquaPro Triode has a temperature connector that is
compatible with specific meters, refer to the list below.
2. Thermo Scientific Orion AquaPro or double junction pH electrode.
The 9107APMD pH/ATC Triode has a temperature
connector that is compatible with the Star Series pH meters.
3. Electrode filling solution (for 9102DJWP, 9110DJWP and 9120APWP electrodes only), Cat. No. 910008.
4. pH electrode storage solution, Cat. No. 910001.
5. pH buffers, at least two pH buffers are recommended for precise measurements. One buffer should be near pH 7 and buffers should be one to three pH units apart.
6. Beakers, plastic or glass.
7. Magnetic stirrer or Orion stirrer probe, Cat. No. 096019. The stirrer probe can be used with 3-Star, 4-Star and 5-Star benchtop meters.
8. Distilled or deionized water.
3AquaPro and Double Junction pH Electrode User Guide
Electrode Preparation
1. AquaPro electrodes ship in a storage bottle that contains storage solution. Remove the electrode from the storage bottle and save the storage bottle for later storage. Refer to the figure below.
For double junction electrodes, carefully remove the
protective shipping cap from the sensing element and save the cap for storage.
2. Clean any salt deposits from the exterior of the electrode by rinsing with distilled water.
3. If using a refillable electrode, uncover the filling hole by removing the tape and then add electrode filling solution, Cat. No. 910008, to the electrode. To maintain an adequate flow rate, the level of filling solution must always be above the reference junction and at least one inch above the sample level. The filling hole should be open whenever the electrode is in use.
4. Soak electrode in pH electrode storage solution, Cat. No. 910001, for 30 to 60 minutes.
5. Connect the electrode to the meter.
4 AquaPro and Double Junction pH Electrode User Guide