ROSS Ultra®,
ROSS®, ROSS®
Sure-Flow®, and
PerpHecT® ROSS®
Electrodes
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion,
pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS
Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of
Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR,
AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test,
BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR,
DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value,
FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low
Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch
Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode,
pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode,
QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART
STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense,
Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of
Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent 6,168,707.
PerpHecT ROSS are protected by U.S. patent 6,168,707.
ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093,
D334,208 and D346,753.
ionplus electrodes and Optimum Results solutions are protected by
US Patent 5,830,338.
ROSS Ultra electrodes are protected by US patents 6,793,787.
Orion ORP Standard is protected by US Patent 6,350,367.
Orion NoCal electrodes with stabilized potential patent pending.
The specifications, descriptions, drawings, ordering information and part numbers
within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Introduction
This user guide contains information on the preparation,
operation and maintenance of the Thermo Scientific Orion ROSS
Ultra, ROSS, ROSS Sure-Flow and PerpHecT ROSS electrodes.
All ROSS electrodes have a pH range of 0 to 14, pH precision of
0.01 and temperature range of 0 to 100 °C.
ROSS electrodes can be used in samples that contain TRIS,
sulfides and proteins since they do not contain silver or mercury.
ROSS electrodes incorporate the patented ROSS internal
reference system that provides superior measurement stability,
faster response, greater accuracy and more reproducible
results than conventional electrodes. ROSS electrodes provide
readings stable to 0.01 pH in less than 30 seconds, even in the
extreme case of samples varying from one another by 50 °C or
more. Results are three to five times more precise than those
obtained with conventional electrodes. Electrode drift is less
than 0.002 pH per day, so restandardization is minimized.
ROSS electrodes have a double junction reference, so the filling
solution can be modified to more closely match the sample
composition in high pH, low pH or non-aqueous solutions.
ROSS Ultra® Electrodes
ROSS Ultra electrodes offer the ultimate measurement stability
and the least long-term drift plus a two year warranty.
ROSS® Electrodes
ROSS electrodes provide precise pH measurements and
measurement stability regardless of temperature changes or
sample composition.
ROSS® Sure-Flow® Electrodes
ROSS Sure-Flow electrodes incorporate a free-flowing junction
that is easy to clean and never clogs. These electrodes are ideal
for dirty or viscous samples that clog traditional electrodes.
PerpHecT® ROSS® Electrodes
PerpHecT ROSS electrodes are designed to be used with
PerpHecT meters to provide both pH and temperature
measurements. The PerpHecT meters use patented LogR
technology that allows temperature to be measured using the
resistance of the PerpHecT ROSS electrode.
ROSS® Electrode User Guide
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
1
Cat. No.Description
8102BNUWPROSS Ultra combination pH electrode with
glass body and waterproof BNC connector
8103BNUWPROSS Ultra combination pH electrode with
glass body, semi-micro tip and waterproof
BNC connector
8104BNUWPROSS Ultra combination pH electrode with
glass body, rugged bulb and waterproof BNC
connector
8115BNUWPROSS Ultra combination pH electrode with
epoxy body, semi-micro tip and waterproof
BNC connector
8135BNUWPROSS Ultra combination pH electrode with
epoxy body, flat surface tip and waterproof
BNC connector
8156BNUWPROSS Ultra combination pH electrode with
epoxy body and waterproof BNC connector
800500UROSS Ultra reference half-cell electrode with
glass body and pin tip connector
8102BN /
810200 /
8102SC
8103BN /
810300 /
8103SC
8104BN /
810400 /
8104SC
8115BN
8115SC
8135BN
8135SC
815600 /
815500 /
8155SC
8162SCROSS combination pH electrode with glass
8163BNWP /
8163SC
ROSS combination pH electrode with glass
body and BNC connector / U.S. standard
connector / screw cap connector
ROSS combination pH electrode with glass
body, semi-micro tip and BNC connector
/ U.S. standard connector / screw cap
connector
ROSS combination pH electrode with glass
body, rugged bulb and BNC connector / U.S.
standard connector / screw cap connector
ROSS combination pH electrode with epoxy
body, semi-micro tip and BNC connector /
screw cap connector
ROSS combination pH electrode with epoxy
body, flat surface tip and BNC connector /
screw cap connector
ROSS combination pH electrode with epoxy
body and BNC connector / U.S. standard
connector / screw cap connector
body, 14/15 standard taper and screw cap
connector
ROSS combination pH electrode with
glass body, spear tip and waterproof BNC
connector / screw cap connector
2
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
ROSS® Electrode User Guide
Cat. No.Description
8101BNWP /
8101SC
800300ROSS Sure-Flow reference half-cell
8165BNWP /
8165DN
8172BNWP /
8172DN
8175BNWPROSS Sure-Flow combination pH electrode
8202BNPerpHecT ROSS combination pH electrode
8203BNPerpHecT ROSS combination pH electrode
8220BNWPPerpHecT ROSS combination pH electrode
8235BNPerpHecT ROSS combination pH electrode
8256BNPerpHecT ROSS combination pH electrode
8272BNPerpHecT ROSS Sure-Flow combination
ROSS pH half-cell electrode with glass body
and waterproof BNC connector / screw cap
connector
electrode with glass body and pin tip
connector
ROSS Sure-Flow combination pH electrode
with epoxy body and waterproof BNC
connector / E DIN waterproof connector
ROSS Sure-Flow combination pH electrode
with glass body and waterproof BNC
connector / E DIN waterproof connector
with epoxy body, semi-micro tip and
waterproof BNC connector
with glass body and BNC connector
with glass body, semi-micro tip and BNC
connector
with glass body, micro tip and waterproof
BNC connector
with epoxy body, flat surface tip and BNC
connector
with epoxy body and BNC connector
pH electrode with glass body and BNC
connector
• All electrodes with a waterproof BNC, BNC and U.S.
standard connector have one meter cable lengths.
• All electrodes with a screw cap connector require a
separate cable.
• Electrodes with a waterproof BNC connector can be used
on any pH meter with a BNC connection.
• The 800500U ROSS Ultra and 800300 ROSS reference
half-cells can be used with the 8101BNWP / 8101SC ROSS
pH half-cell or 8411BN ROSS sodium half-cell.
ROSS® Electrode User Guide
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
3
Required Equipment
1. Thermo Scientific Orion pH meter, such as the 3-Star pH
meter, 4-Star pH/ISE meter, 4-Star pH/DO meter, 4-Star pH/
conductivity meter or 5-Star pH/ISE/DO/conductivity meter.
ROSS electrodes can be used on any pH meter with a BNC
or U.S. standard connection. The electrodes can also be
used on meters with a variety of inputs when an adapter
cable is used. Visit www.thermo.com/water for details.
2. Thermo Scientific Orion ROSS Ultra, ROSS, ROSS
Sure-Flow or PerpHecT ROSS electrode.
3. ROSS filling solution, Cat. No. 810007. Do not use any
filling solution that contains silver, as silver will damage the
electrode.
4. ROSS pH electrode storage solution, Cat. No. 810001, or
standard pH electrode storage solution, Cat. No. 910001.
For enhanced electrode performance, use ROSS pH
electrode storage solution.
5. pH buffers, at least two pH buffers are recommended for
precise measurements. One buffer should be near pH 7
and buffers should be one to three pH units apart.
6. Beakers, plastic or glass.
7. Magnetic stirrer or Orion stirrer probe, Cat. No. 096019.
The stirrer probe can be used with 3-Star, 4-Star and 5-Star
benchtop meters.
8. Distilled or deionized water.
4
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
ROSS® Electrode User Guide
Sample Requirements
A benefit of the ROSS pH electrodes is that the filling
solution composition may be changed depending on sample
requirements.
The ROSS pH electrode filling solution, Cat. No. 810007, is 3 M
KCl. For solutions that precipitate in the presence of chloride
ions, the electrode can be filled with 10% KNO3, Cat. No.
900003.
Electrodes with an epoxy body should only be used in aqueous
solutions.
Electrodes with a glass body may be used in non-aqueous
solutions and solutions that contain organic solvents. A
minimum of 20% water must be present in the sample for
the best measurement results. If there is a great deal of drift
when using the ROSS electrode filled with ROSS filling solution,
try filling the electrode with a mixture of methanol and water
saturated with KCl. The ratio of methanol to water should be
similar to the sample composition.
If using the 8220BNWP PerpHecT ROSS micro pH electrode,
great care must be taken to thoroughly rinse out the reference
chamber and junction before changing the reference solutions.
ROSS® Electrode User Guide
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
5