Thermo Scientific 310, 320, 310 Series, 311, 321 User Manual

...
Page 1
Visit us online to register your warranty www.thermoscientific.com/labwarranty
User Manual
Forma Direct Heat CO2 Incubator
Model 310 Series
Operatign and Maintenance Manual 7010310 Rev. 16
Page 2
Thermo ScientificThermo Scientific
Model 310 Series DH Incubator i
Model CO2 Sensor* Voltage**
310 T/C 115
311 T/C 230
320 IR 115
321 IR 230
350† T/C 115
351† T/C 230
360† IR 115
361† IR 230
* T/C is a thermal conductivity sensor.
IR is an infrared sensor.
** All units are 50/60 Hz.
† Copper interior
MANUAL NUMBER 7010310
16 31343/IN-4592 10/22/14 New magnetic door switch (exploded drawing, electrical schematics) RoHS ccs
15 29174/IN-4469 3/21/14 Solenoid change - update 203 drawing and schematics (pgs 8-5 & 9-1 to 9-3) ccs
14 30649/IN-4446 1/28/14 Updated electrical schematics - display board ccs
13 29704/IN-4310 3/29/13 Changed wingnut in blower assembly to hex nut (310-202-1-D) ccs
12 28116/IN-4201 5/29/12 Chg’d harness from 190522 to 1900089 - pg 8-6 ccs
11 27990/IN-4179 5/3/12 Added Intended Use to pg ii, updated Specs uniformity to 0.3°C from 0.2 ccs
10 27292/IN-4073 9/28/11 Switcher from 400199 to 400201 kit - pg 8-1 ccs
-- -- 7/13/10 Re-installed “Specs are based on...” on Specification page. Unintentional removal ccs
Page 3
Thermo Scientificii Model 310 Series DH Incubator
Important Read this instruction manual. Failure to read, understand and follow the instructions in this manual may result in damage to the unit, injury to operating personnel, and poor equipment performance.
s
Warning If this unit is not used in the manner specified in this operating manual, the protection provided by the equipment design may be impaired. s
Intended Use: The 310 Series Direct Heat Incubators are designed to maintain an optimal environment for the incubation of tissue and cell samples. These models are designed to maintain temperature and Carbon Dioxide levels as set by the operator as well as monitoring chamber relative humidity. CO2 is controlled using either a thermo conductivity cell, or an infra-red sensor. T
The incubators are approved for general purpose use as a Microbiological Incubator (21 CFR 866.2540) and are also approved use in assisted reproductive procedures for the incubation of ova and embryos (21 CFR
884.6120).
Non-intended Use:
The 310 Series Direct Heat Incubators are not intended for use where electrical or physical
contact with the patient is established.
These incubators are not intended to be operated in potentially explosive environments and are not intended for use with flammable materials.
Caution All internal adjustments and maintenance must be performed by qualified service personnel.
s
Material in this manual is for information purposes only. The contents and the product it describes are subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
©2003 Thermo Fisher Scientific. All rights reserved.
CAUTION
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
Page 4
Thermo Scientific Model 310 Series DH Incubator iii
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be damaged by elevated temperatures.
Marking of electrical and electronic equipment, which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market after 13 August 2005.
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State European Country, and this product should be disposed of or recycled through them. Further information on Thermo’s compliance with this directive, the recyclers in your country and information on Thermo Scientific products will be available at www.thermo.com.
4 Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4 Always dissipate extreme cold or heat and wear protective clothing.
4 Always follow good hygiene practices.
4 Each individual is responsible for his or her own safety.
Page 5
Thermo Scientificiv Model 310 Series DH Incubator
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct 1-800-438-4851 Toll Free, U.S. and Canada 1-877-213-8051 FAX http://www.thermoscientific.com Internet Worldwide Web Home Page service.led.marietta@thermofisher.com Tech Support Email Address www.unitylabservices.com
Our staff can provide information on pricing and give you quotations. We can
Sales Support
take your order and provide delivery information on major equipment items or make arrangements to have your local sales representative contact you. Our products are listed on the Internet and we can be contacted through our Internet home page.
Certified Service Web Page
Our staff can supply technical information about proper setup, operation or
Service Support
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or provide you with on-site service. We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service call.
When more extensive service is necessary, we will assist you with direct factory trained technicians or a qualified service organization for on-the-spot repair. If your service need is covered by the warranty, we will arrange for the unit to be repaired at our expense and to your satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by telephone or fax. If you wish to write, our mailing address is:
Thermo Fisher Scientific 401 Millcreek Road, Box 649 Marietta, OH 45750
International customers, please contact your local Thermo Scientific distributor.
Page 6
Thermo Scientific Model 310 Series DH Incubator vThermo Scientific
Warranty Notes
Information You Should Know Before Requesting Warranty Service
Locate the model and serial numbers. A serial tag is located on the unit itself.
•For equipment service or maintenance, or with technical or special application inquiries, contact Technical
Services at 1-800-438-4851 or 1-740-373-4763 (USA and Canada). Outside the USA, contact your local distributor.
Repairs NOT Covered Under Warranty
Calibration of control parameters. Nominal calibrations are performed at the factory; typically ±1°C for
temperature, ±1% for gases, and ±5% for humidity. Our service personnel can provide precise calibrations as a billable service at your location. Calibration after a warranty repair is covered under the warranty.
• Damage resulting from use of improper quality water, chemicals or cleaning agents detrimental to equipment materials.
Service calls for improper installation or operating instructions. Corrections to any of the following are billable services:
1) electrical service connection
2) tubing connections
3) gas regulators
4) gas tanks
5) unit leveling
6) room ventilation
7) adverse ambient temperature fluctuations
8) any repair external to the unit
• Damage resulting from accident, alteration, misuse, abuse, fire, flood, acts of God, or improper installation.
Repairs to parts or systems resulting from unauthorized unit modifications.
Any labor costs other than that specified during the parts and labor warranty period, which may
include additional warranty on CO
2 sensors, blower motors, water jackets, etc.
Page 7
Page 8
Model 310 Series DH Incubator viThermo Scientific
Table of Contents
Installation and Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Control Panel Keys, Displays & Indicators . . . . . . . . . . . . . . . . . . . .1-2
Keypad Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Installing the Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Incubator Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Calibration . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Calibrating the Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Calibrating the T/C CO
2 System . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Calibrating the Infrared CO2 System . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Calibrating Relative Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Turn the Audible Alarm ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Set an Access Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Set a Low Temp Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Enable the Low Temp Alarm to Trip Contacts . . . . . . . . . . . . . . . . .3-2
Set a Low CO
2 Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Setting a High CO2 Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Enable CO
2 Alarms to Trip Contacts . . . . . . . . . . . . . . . . . . . . . . . .3-4
Set New Zero Number for T/C CO2 Sensors . . . . . . . . . . . . . . . . . .3-4
Set New Span Number for T/C CO2 Sensors . . . . . . . . . . . . . . . . . .3-5
Set a Low RH Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Enable RH Alarms to Trip Contacts . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Enable Temp/RH to be Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Select Primary Tank w/ Gas Guard Option . . . . . . . . . . . . . . . . . . . .3-7
Disable the Gas Guard System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Communications Address for RS485 . . . . . . . . . . . . . . . . . . . . . . . .3-8
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . .4-1
Temp Control Failure Alarm TMP CNTR ERR . . . . . . . . . . . . . . .4-2
Sensor Fault Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
CO2 SNSR ERR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
IR AUTOZ ERR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Section 1
Section 2
Section 3
Section 4
Page 9
vii Model 310 Series DH Incubator Thermo Scientific
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Disinfecting the Incubator Interior . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Cleaning the Cabinet Exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Cleaning the Glass Doors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Cleaning the Humidity Pan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Reversing the Door Swing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
HEPA Filter (Factory Installed Option) . . . . . . . . . . . . . . . . . . . . . . .5-8
Replacing the Power Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
Major Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Electronics Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Replacing the Sample Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Factory Options . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Connect Remote Alarm Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Connect RS485 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Connecting the Analog Output Boards . . . . . . . . . . . . . . . . . . . . . . .6-2
CO
2 Gas Guard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Humidity Readout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Uninterruptible Power Supply Connections . . . . . . . . . . . . . . . . . . .6-7
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Spare Parts .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Electrical Sch
ematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Warranty Information . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
Table of Contents
Section 5
Section 6
Section 8
Section 7
Section 9
Section 10
Page 10
Model 310 Series DH Incubator 1-1Thermo Scientific
Section 1 Installation and Start-Up
Incubator components are described below.
Figure 1-1. Model 310 Series Direct Heat Incubator
• Outer Door - Reversible to opposite swing, see Section 5
• Inner Door - Reversible to opposite swing, see Section 5
• Chamber Gas Sample Port - Used for sampling chamber CO
2 content
using a FYRITE or similar instrument.
• Main Power Switch
• Control Panel - Keypad, Displays and indicators (See Figure 1-2)
• Leveling Legs - Used to level the unit
Note The incubators are stackable. Information follows.
s
Page 11
1-2 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-up
Silence - Silences the audible alarm.
Alarm Indicator - Light pulses on/off during an alarm condition in the
cabinet.
Mode Select Switch - Used to select Run, Setpoints, Calibration and
System Configuration Modes.
Message Center - Displays system status.
Mode Select Indicators -
Run: Run Menu
Set: Set Points Menu
Cal: Calibrate Menu
Config: Configuration Menu
Up and Down Arrows - Increases or decreases the number values, toggles
between choices.
Enter - Stores the value into computer memory.
Heat Indicator - Lights when power is applied to the heaters.
Temp Display - Program to display temperature continuously, RH con-
tinuously (with RH option), or toggle between temperature and humid­ity (with RH option). See Section 3, Configuration.
Scroll for Parameters Arrows - Moves the operator through the choices
of the selected mode.
CO
2 Inject Indicator - Lights when CO2 is being injected into the incu-
bator.
% CO2 Display - Displays CO
2 percentage continuously.
Control Panel Keys,
Displays & Indicators
Figure 1-2. Control Panel
Page 12
Model 310 Series DH Incubator 1-3Thermo Scientific
Section 1
Installation and Start-up
The Model 310 Series direct heat incubator has four basic modes which allow incubator setup: Run, Setpoints, Calibration and System Configuration.
Run is the default mode which the incubator will normally be in during
operation.
Set is used to enter system setpoints for incubator operation.
Calibration is used to calibrate various system parameters.
Configuration allows for custom setup of various options.
The chart below shows the selections under each of the modes.
1
T/C units only Base Unit Displays
2
IR units only Option Displays
*T/C units only
Scroll for Parameters Arrows: Steps the operator through the parameters
of SET, CAL and CONFIG Modes. The right arrow goes to the next parameter, the left arrow returns to the previous parameter.
Up Arrow: Increases or toggles the parameter value that has been selected
in the SET, CAL, and CONFIG Modes.
Enter: Must press Enter key to save to memory all changed values.
Down Arrow: Decreases or toggles the parameter values that have been
selected in the SET, CAL and CONFIG Modes.
Silence Key: Press to silence the audible alarm. See Section 4 for alarm
ringback times.
Keypad Operation
RUN SETPOINT CALIBRATION CONFIGURATION Default Mode Temperature Temp Offset Audible
Overtemp CO2 Cal
1
Access Code
CO2 IR Cal
2
Temp Lo Alarm
RH Cal Temp Relay
CO2 Lo Alarm CO2 Hi Alarm CO2 Relay CO2 Z & S #'s *
RH Lo Alarm RH Relay Display Temp Display Rh Tank Select Gas Guard
RS485 Address
Page 13
Message Center: Displays the system status (Mode) at all times. Displays
SYSTEM OK during normal operation, or alarm messages if the system detects an alarm condition. See Section 4, Alarms.
Upper and Lower Displays: These 7 segment displays vary depending
upon the options present and the configuration chosen. The upper dis­play can display temp or RH, or toggle between them. The bottom dis­play shows CO
2 continuously.
Warning Single and stacked units must be installed against a wall or similar structure. Maintain a three-inch clearance behind the incubator for electrical and gas hook-ups.
s
1. Locate the unit on a firm level surface capable of supporting the unit’s weight of 205 lbs.
2. Locate the unit away from doors and windows and heating and air conditioning ducts.
3. Lift the unit only by the sides of the cabinet base . Do not attempt to lift it by the front and back . This places stress on the outer door hinges.
Caution When stacking incubators, the direct heat incubator must be the top unit. Never stack a water-jacketed incubator on top of a Model 310 Series unit.
s
Warning With incubators in a stacked configuration, do not leave both exterior doors open at the same time.
s
Warning If the units have been in operation, shut them both off and disconnect from the power source before beginning any service work. s
Note Two stacking brackets (shown at right) are included in the parts bag shipped with each incubator.
s
1-4 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-Up
Displays
Installing the
Incubator
Stacking the Incubators
Figure 1-3.
Stacking Brackets
Page 14
1. Remove and discard the slotted head screws on the top of the bottom incubator (Figure 1-4). Replace with 5/16 HH bolts and flatwashers included in the stacking kit. Do not tighten.
2. Remove the kickplate from below the door on the top incubator (Figure 1-5). Two Phillips screws secure the plate. Note the notches at the base of the incubator frame.
3. Disconnect the door cable from the connector behind the plate. Lift the door off the hinge pins and set it aside. The hinge pins are not attached to the brackets and may fall out when the door is lifted off.
4. Unscrew and remove the leveling feet from the top unit and lift it onto the bottom unit, offsetting the base of the top unit approximately 2-3 inches behind the screws identified by the arrows in Figure 1-4.
Warning This incubator weighs 205 lbs. Have sufficient personnel available when lifting. Lift the unit only by the sides of the cabinet base. Do not attempt to lift it by the front and back as this places stress on the outer door hinge.
s
5. Align the sides of the top unit with the bottom unit and slide the top unit forward until the notches in the base of the top unit align with the bolts in the top of the bottom unit. The flatwashers should slide over the base frame. Tighten bolts.
6. Remove the four nylon plugs on the lower back of the upper incubator.
7. Insert the stacking brackets into the slots on the back of the control panel of the bottom unit as shown in Figure 1-6.
Model 310 Series DH Incubator 1-5Thermo Scientific
Section 1
Installation and Start-Up
Stacking (continued)
Figure 1-4. Slotted Head Screws
Figure 1-5. Align
Figure 1-6. Brackets Installed
Page 15
8. Align the slots in the brackets with the mounting holes on the back of the top incubator. Secure the brackets with the screws provided in the stacking kit.
9. Make sure the hinge pins are in the hinge brackets. Slide the door onto the hinge pins.
10. Secure the door heater cable to the connector below the door and replace the kick plate on the top unit, taking care not to crimp cable.
11. The stacked incubators are ready for service.
Remove vinyl from shelf channels, duct sheets, and air duct, if present. Using a suitable laboratory disinfectant, disinfect all interior surfaces including shelves and shelf supports, door gaskets, blower wheel and CO
2
sensor. Refer to Section 5.
Caution Before using any cleaning or decontamination method except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. Accidental spills of hazardous materials on or inside this unit are the responsibility of the user.
s
1. Install the large sheet metal shelf supports with the tabs facing into the center of the chamber with their slots up. There are no right side or left side supports. Simply turn one of them to fit the opposite side. Tilt the shelf supports as they are placed in the chamber so the tops fit into the top air duct, then guide them into the vertical position. Figure 1-7 shows the support as it would be oriented for right side of chamber.
1-6 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-Up
Preliminary Cleaning and
Disinfecting
Installing the Shelves
Figure 1-7. Shelf Channels and Ductwork
Stacking (continued)
Page 16
2. Note there is no difference in the left side and right side shelf channels.
3. Install the shelf channels by placing the channel’s rear slot over the appropriate rear tab on the shelf support. Pull the shelf channel forward and engage the channel’s front slot into the shelf support’s appropriate forward tab. Refer to Figure 1-8.
Locate the opening in the top left corner on the inside of the chamber. Remove the tape covering the opening on the outside of the unit. Locate the stopper with filter in the hardware bag. Install the assembly in the opening inside the chamber. See Figure 1-9.
Figure 1-9. Filter and Sensor Locations
Caution Be careful when handling the filter. The media can be damaged if
it is mishandled. To avoid damage to the incubator, do not operate the unit without the HEPA filter in place.
s
1. Remove the filter from the shipping box.
2. Remove the plastic coating from the filter, using caution not to touch the filter media.
3. Install the filter as shown in Figure 1-9. Refer to Section 5 for HEPA filter maintenance.
Model 310 Series DH Incubator 1-7Thermo Scientific
Section 1
Installation and Start-Up
Figure 1-8. Channel Ends
Installing (optional) HEPA
Filter
Installing Access Port
Filter
Installing the Shelves
(continued)
HEPA Filter
Temperature
sensor
Access port filter
Sample Air Filter
Page 17
Leveling the Unit
1. Remove the filter from the shipping bag.
2. Install the air sample filter assembly to the black hose barb behind the top duct.
3. Insert the other end of the filter assembly into the larger hole in the air intake to the blower. Refer to Figure 1-10.
4. Verify that the temperature sensor is installed into the smaller hole in the intake, as in Figure 1-10.
Check for level by placing a bubble style level on one of the shelves. Turn the hex nut on the leveler counterclockwise to lengthen the leg or clockwise to shorten it. Level the unit front-to-back and left-to-right.
See the serial tag on the side of the unit for electrical specifications or refer to the electrical schematics at the end of this manual.
Caution Connect the incubator to a grounded, dedicated circuit. The power cord connector is the mains disconnect device for the incubator. Position the incubator so the unit can be easily disconnected.
s
1-8 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-Up
Figure 1-10. View of Chamber Ceiling
Installing the Air Sample
Filter
Connecting the Unit to
Electrical Power
Page 18
Plug the provided power cord into the power inlet connector on the back of the cabinet (See Figure 1-11), then into the grounded, dedicated, electrical circuit.
Figure 1-11. Connections on Unit Back
Fill the humidity pan with sterile pure water to within 1/2” of the top. Place the pan directly in the center of the incubator floor to ensure best humidity and temperature response.
For applications requiring abnormally high humidity, the pan should be placed against the left side wall of the incubator. The ductwork has been modified to allow the pan to be placed against the wall of the incubator. In some ambient conditions, this may cause condensation to form in the chamber.
For best operation of the incubator, sterilized distilled, demineralized or de-ionized water should be used in the humidity pan. Water purity should be in the resistance range of 50K to 1M Ohm/cm, or a conductivity range of 20.0 to 1.0 uS/cm. Refer to ASTM Standard D5391-93 or D4195-88 for measuring water purity.
Distillation systems, as well as some types of reverse osmosis water purity systems, can produce water in the quality range specified. Tap water is not recommended as it may contain chlorine, which can deteriorate the stainless steel. Tap water may also have a high mineral content, which would produce a build-up of scale in the pan. High purity or ultra pure water is not recommended as it is an extremely aggressive solvent and will deteriorate the stainless steel. High purity water has a resistance of above 1M to 18M Ohm. Even high purity water can contain bacteria and organic contaminants. Water should always be sterilized or treated with a decontaminant, safe for use with stainless steel as well as safe for the product, prior to being introduced into the humidity pan.
Model 310 Series DH Incubator 1-9Thermo Scientific
Section 1
Installation and Start-Up
Filling the Humidity Pan
Connecting the Unit to
Electrical Power (cont.)
Page 19
Caution Distilled or de-ionized water used in the humidity pan must be within a water quality resistance range of 50K to 1M Ohm/cm to protect and prolong the life of the stainless steel. Use of chlorinated water, or decontamination products containing chlorine, will deteriorate the stainless steel and cause rust, voiding the warranty.
s
Check the level and change the water frequently to avoid contamination. Do not allow the water level to fluctuate significantly. “Dry-outs” will have an adverse effect on the humidity level and CO
2 calibration of the T/C
units.
Warning High concentrations of CO
2 gas can cause asphyxiation! OSHA
Standards specify that employee exposure to carbon dioxide in any eight­hour shift of a 40-hour work week shall not exceed the eight-hour time weighted average of 5000 PPM (0.5% CO
2). The short term exposure
limit for 15 minutes or less is 30,000 ppm (3% CO2). Carbon dioxide monitors are recommended for confined areas where concentrations of carbon dioxide gas can accumulate.
s
Warning This incubator is designed to be operated with CO2 gas only. Connecting a flammable or toxic gas can result in a hazardous condition. Gases other than CO
2 should not be connected to this equipment. CO2
gas cylinders have a UN1013 label on the cylinder and are equipped with a CGA 320 outlet valve. Check the gas cylinder for the proper identification labels. The CO
2 gas supply being connected to the incubator should be
industrial grade, 99.5% pure. Do not use CO2 gas cylinders equipped with siphon tubes. A siphon tube is used to extract liquid CO2 from the cylinder which can damage the pressure regulator. Consult with your gas supplier to ensure that the CO
2 cylinder does not contain a siphon tube.
Gas cylinders should also be secured to a wall or other stationary object to prevent them from tipping. A two-stage CO
2 pressure regulator is required to be installed on the outlet
valve of the gas cylinder. Input pressure to the incubator must be maintained at 15 psig (103.4 kPa), ±5 psig, for proper performance of the CO
2 control system. A single stage CO2 pressure regulator will not
maintain 15 psig (103.4 kPa) to the incubator as the pressure in the CO
2
cylinder decreases; therefore, a two stage regulator is recommended. If higher purity CO2 is desired inside the incubator (greater than 99.5% pure), the pressure regulator should be constructed with a stainless steel diaphragm, along with specifying the purity of the CO
2 from the gas
supplier. Follow the manufacturer’s instructions to ensure proper and safe installation of the pressure regulator on the gas cylinder. Consult your facility safety officer to ensure that the equipment is installed in accordance with the codes and regulations applicable in your area.
s
1-10 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-Up
Connecting the CO2 Gas
Supply
Filling the Humidity Pan
(cont.)
Page 20
The CO2 gas supply being connected should be industrial grade 99.5% pure and should not contain siphon tubes. Install a two-stage pressure regulator at the cylinder outlet. The high pressure gauge at the tank should have 0-2000 psig range. The low pressure gauge, at the incubator inlet, should have a 0-30 psig range. Input pressure to the incubator must be maintained at 15 psig (103.4 kPa).
The incubator has serrated fittings on the back of the cabinet to connect the gas supply. Refer to Figure 1-11. The fitting is labeled CO2 Inlet #1 Tank. Make sure that the connections are secured with clamps. Check all fittings for leaks.
For units having the CO
2 Gas Guard option, refer to Section 6.
With the incubator properly installed and connected to power, the humidity pan filled, and the unit connected to gas supplies, system setpoints can be entered. The following setpoints can be entered in Set mode: Temperature, Overtemperature and CO
2. To enter Set mode, press
the Mode key until the Set indicator lights. Press the right and/or left arrow keys until the proper parameter appears in the message center. See Chart 1-1 for more detail.
Model 310 Series incubators have an operating temperature range of 10°C to 50°C, depending on ambient temperature. The incubator is shipped from the factory with a temperature setpoint of 10°C. At this setting, all heaters are turned off. To change the operating temperature setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “Temp XX.X” is displayed in the message center.
3. Press the up/down arrow key until the desired temperature setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights for Run mode or press the right/left arrow keys to go to next/previous parameter.
Model 310 Series DH Incubator 1-11Thermo Scientific
Section 1
Installation and Start-Up
Incubator Start-Up
Setting the Operating
Temperature
Connecting the CO
2 gas
Supply (continued)
Page 21
Caution The independent overtemp system is designed as a safety to protect the incubator only. It is not intended to protect or limit the maximum temperature of the cell cultures or customer’s equipment inside the incubator if an overtemp condition occurs.
s
Model 310 Series incubators are equipped with a secondary temperature monitoring system to monitor the air temperature inside the cabinet. This system is designed as a safety device to turn off all heaters in the event of a temperature control failure. Temperature control in the incubator will be ± 1° of the overtemp setpoint.
The factory overtemperature setpoint (default) is 40°C. However, the overtemp can be set up to 55°C in 0.5° increments.
If the incubator’s operating temperature setpoint is set above the overtemp setpoint, the overtemp setpoint will automatically update to 1°C above the temperature setpoint. It is recommended that the overtemp setpoint be maintained at 1°C over the operating temperature setpoint.
To set the Overtemp setpoint:
1. Press the Mode key until the Set indicator lights
2. Press the right arrow until Otemp XX.X is displayed in the message center
3. Press the up or down arrow key until the desired Overtemp setpoint is displayed
4. Press Enter to save the setting
5. Press the Mode key until the Run indicator lights or press the right or left arrow to go to the next or previous parameter.
1-12 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-Up
Setting the Overtemp
Setpoint
Page 22
All T/C CO2 cells are calibrated at the factory at 37°C, high humidity, and 10% CO
2. Therefore, if a temperature setpoint of 37°C has been entered,
the humidity pan has been filled and the CO
2 control is to run between 0-
10% with a T/C CO
2 sensor, the CO2 setpoint may be entered
immediately. Otherwise, it is important to allow the unit 12 hours to stabilize at the temperature setpoint before entering the CO
2 setpoint.
All models of the incubator have a CO
2 setpoint range of 0.0% to 20.0%.
The incubator is shipped from the factory with a CO
2 setpoint of 0.0%.
At this setting, all CO
2 control and alarms are turned off. To change the
CO
2 setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “CO2 XX.X” is displayed in the message center.
3. Press the up/down arrows until the desired CO
2 setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights to go to Run mode or press the right/left arrow keys to go to next/previous parameter.
Model 310 Series DH Incubator 1-13Thermo Scientific
Section 1
Installation and Start-Up
Setting the CO2 Setpoint
Page 23
1-14 Model 310 Series DH Incubator Thermo Scientific
Section 1
Installation and Start-Up
Page 24
Model 310 Series DH Incubator 2-1Thermo Scientific
Section 2 Calibration
After the unit has stabilized, several different systems can be calibrated. In the calibration mode, the air temperature, CO2 and RH levels can be calibrated to reference instruments. To access the calibration mode, press the Mode key until the Cal indicator lights. Press the right and/or left arrow until the proper parameter appears in the message center. See Chart 2-1 at the end of this section for more detail.
Calibration frequency is dependent on use, ambient conditions and accuracy required. A good laboratory practice would require at least an annual calibration check. On new installations, all parameters should be checked after the stabilization period.
Prior to calibration, the user should be aware of the following system functions. While the unit is in the calibration mode, all system control functions are stopped so the unit remains stable. Readout of the system being calibrated will appear on the message center. If no keys are pressed for approximately five minutes while in calibration mode, the system will reset to Run mode so control functions are reactivated.
Caution Before making any calibration or adjustments to the unit, it is imperative that all reference instruments be properly calibrated.
s
Before calibration, allow the cabinet temperature to stabilize. Place the calibrated instrument in the center of the chamber. The instrument should be in the airflow, not against the shelf.
Temperature Stabilization Periods
Startup - Allow 12 hours for the temperature in the cabinet to stabilize
before proceeding.
Already Operating - Allow at least 2 hours after the display reaches setpoint for temperature to stabilize before proceeding.
Calibrating the
Temperature
Page 25
2-2 Model 310 Series DH Incubator Thermo Scientific
Section 2
Calibration
1. Press the Mode key until Cal indicator lights.
2. Press the right arrow until “TEMPCAL XX.X” appears in the message center.
3. Press up/down arrow to match display to calibrated instrument.
4. Press Enter to store calibration.
5. Press the Mode key to return to Run or the right/left arrow to go to next/previous parameter.
Models 310, 311, 350 and 351 have a thermal conductivity (T/C) CO
2
sensor. Thermal conductivity of the incubator atmosphere is not only effected by the quantity of CO
2 present, but also by the air temperature
and water vapor present in the incubator atmosphere. In monitoring the effects of CO
2, air temperature and absolute humidity must be held
constant so any change in thermal conductivity is caused by a change in CO
2 concentration.
Changing temperature or changing from elevated humidity to room ambient humidity levels will necessitate a re-calibration of the CO
2
control.
T/C CO
2 Sensor Stabilization Periods
Some T/C CO
2 sensors go through an aging period, especially on new
installations. Calibration should be checked on a weekly basis, and adjusted as necessary. When stabilization occurs, checks may become less frequent.
Start -Up - The CO
2 sensor has been calibrated at the factory for 37°C
and elevated humidity. Allow the temperature, humidity, and CO
2 levels
in the chamber to stabilize at least 12 hours before checking the CO
2
concentration with an independent instrument.
Presently Operating - Make sure the chamber doors are closed. Allow at least 2 hours after the temperature and CO
2 displays reach their setpoints
for chamber atmosphere stabilization.
Calibrating the T/C
CO
2 System
Calibrating Temp
(continued)
Page 26
Model 310 Series DH Incubator 2-3Thermo Scientific
Section 2
Calibration
1. Make sure the stabilization periods outlined above are followed.
2. Sample the chamber atmosphere through the sample port with an independent instrument. Sample the atmosphere at least 3 times to ensure accuracy of the instrument.
3. Press the Mode key until the Cal indicator lights.
4. Press the right arrow until “CO2 CAL XX.X” is displayed in the message center.
5. Press the up /down arrows to change the display to match the independent instrument.
6. Press Enter to store the calibration.
7. Press the Mode key to return to Run or the right or left arrows to go to the next/ previous parameter.
Models 320, 321, 360 and 361 have an infrared (IR) CO
2 sensor. Infrared
CO
2 sensors are not effected by chamber atmosphere temperature or
humidity. However, the light detector in the sensor is effected by wide temperature changes. Therefore, changing temperature setpoints could necessitate a recalibration of the CO
2. Chamber temperature should be
allowed to stabilize before checking CO
2 concentrations with an
independent instrument, especially on start-up.
All models equipped with an IR/CO
2 sensor have an automatic calibration
which occurs every 24 hours and lasts for 5 to 6 minutes. During automatic calibration, the CO
2 display is blanked out and HEPA filtered
room air is pumped through the CO
2 sensor. A new CO2 calibration value
is stored in memory for use as the 0.0% CO
2 reference point. The keypad/
control panel is locked during calibration, with the message center reading “CO2 AUTO CAL”.
IR CO
2 Sensor Stabilization Times
Start-Up- Allow the temperature and the CO
2 of the cabinet to stabilize at
least 12 hours before proceeding.
Presently Operating - Allow CO
2 to stabilize at least 2 hours at setpoint
before proceeding.
Calibrating the
Infrared CO
2 System
Calibrating T/C CO2
System (continued)
Page 27
Calibrating Relative
Humidity
1. Measure the CO2 concentration in the chamber through the gas sample port with a FYRITE or other independent instrument. Several readings should be taken to ensure accuracy.
2. Press the Mode key until the Cal indicator lights.
3. Press the right arrow until “IR CAL XX.X” appears in the message center.
4. Press the up/down arrow to adjust the display to match the independent instrument reading.
5. Press Enter to store the calibration.
6. After Enter is pressed, the unit will go into a calibration cycle that will last 5 to 6 minutes. The control panel is locked during this cycle.
7. Press the Mode key to return to Run mode.
All 310 Series incubators can be equipped with an optional direct readout relative humidity (RH) sensor. This is a readout only of the chamber relative humidity level. It does not provide any control of the relative humidity in the cabinet.
Relative Humidity Stabilization Times
Start-Up - Allow 12 hours for the relative humidity and temperature in the
chamber to stabilize before proceeding.
Already Operating - Allow at least 2 hours after temperature display reaches setpoint for relative humidity to stabilize before proceeding.
2-4 Model 310 Series DH Incubator Thermo Scientific
Section 2
Calibration
Calibrating the IR CO2
System (continued)
Page 28
1. Place an accurate independent instrument in the center of the chamber. Allow at least 30 minutes for RH to stabilize.
2. Press the Mode key until the Cal indicator lights.
3. Press the right arrow key until “RH CAL XX” appears in the message center.
4. Press the up/down arrow to match the display to the independent instrument.
5. Press Enter to store the calibration.
6. Press the Mode key to return to Run mode.
If a reliable RH measuring device is not available, the display may be calibrated to a typical level.
1. Follow the RH stabilization periods outlined previously.
2. With a full humidity pan and stable temperature, the relative humidity in the chamber will be 95%.
3. Using Step 3-5 of the relative humidity sensor adjustment, adjust the display to 95%.
This calibration method should be accurate to within 5%.
Model 310 Series DH Incubator 2-5Thermo Scientific
Section 2
Calibration
Calibrating Relative
Humidity (continued)
Page 29
2-6 Model 310 Series DH Incubator Thermo Scientific
Section 2
Calibration
Page 30
Model 310 Series DH Incubator 3-1Thermo Scientific
Section 3 Configuration
Several features available in Configuration Mode allow custom setup of the incubator. These features are listed and described below. All features may not be necessary in all applications, but are available if needed. To enter Configuration mode, press the Mode key until the Config indicator lights. Press the right and/or left arrow until the appropriate parameter appears in the message center. See Chart 3-1 for more detail.
The audible alarm can be turned on or off. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Audible XXX is displayed in the message center.
3. Press up/down arrow to toggle Audible ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to run mode or right/left to go to next/previous parameter.
A 3-digit Access Code can be entered to avoid unauthorized personnel from changing the setpoints, calibration, or configuration. A setting of 000 will bypass the access code. The factory setting is 000.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Acc Code XXX is displayed in the message center.
3. Press up/down arrow to change the access code.
4. Press Enter to save the access code.
5. Press the Mode key to return to the Run mode or right/left to go to next/previous parameter.
Set an Access Code
Turn the Audible
Alarm ON/OFF
Page 31
Enable the Low Temp
Alarm to Trip Contacts
3-2 Model 310 Series DH Incubator Thermo Scientific
Section 3
Configuration
Set a Low Temp
Alarm Limit
The low temp alarm limit (tracking alarm) is the deviation from the temperature setpoint which will cause a low temp alarm. The low temp alarm is variable from 0.5° below setpoint to 5° below setpoint. The factory setting is 1° below setpoint. A minus sign (-) in the display indicates that the alarm setting is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Temp Lo -X.X is displayed in the message center.
3. Press up/down arrow to change the low temp alarm limit.
4. Press Enter to save the low temp alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
The low temperature alarm can be programmed to trip the remote alarm contacts. A setting of ON allows it, a setting of OFF will not allow the temperature alarm to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until TMP RLY XXX is displayed.
3. Press the up/down key to toggle the setting ON/OFF.
4. Press Enter to save the setting
5. Press the Mode key to return to Run or the right/left arrow key to go to next/previous parameter.
Page 32
Model 310 Series DH Incubator 3-3Thermo Scientific
Section 3
Configuration
The low CO2 alarm limit (tracking alarm) is the deviation from the CO2 setpoint which will cause a low CO2 alarm. The setpoint is variable from
0.5% CO
2 below setpoint to 5.0% CO2 below setpoint. The factory
setting is 1.0% CO
2 below setpoint. A minus sign (-) in the display
indicates that the alarm setting is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 Lo -X.X is displayed in the message center.
3. Press up/down arrow to change the low CO
2 alarm limit.
4. Press Enter to save the low CO
2 alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
The high CO
2 alarm limit (tracking alarm) is the deviation from the CO2
setpoint which will cause a high CO2 alarm. The setpoint is variable from
0.5% CO
2 above setpoint to 5.0% CO2 above setpoint. The factory
setting is 1.0% CO
2 above setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 Hi X.X is displayed in the message center.
3. Press up/down arrow to change the high CO
2 alarm limit.
4. Press Enter to save the high CO
2 alarm limit.
5. Press the Mode key to return to run mode or right/left to go to next/previous parameter.
Set a Low CO2 Alarm
Limit
Setting a High CO
2
Alarm Limit
Page 33
Set New Zero Number
for T/C CO
2 Sensors
High and Low CO2 alarms can be programmed to trip the remote alarm contacts. A setting of ON allows it, a setting of OFF will not allow the CO
2 alarms to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 Rly XXX is displayed in the message center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run Mode or right/left to go to next/previous parameter.
If a new T/C CO
2 sensor is being installed, the two numbers on the
factory installed sticker on the T/C cell must be entered to calibrate the CO
2 in the unit.
Note For the technician’s convenience, a label containing the two numbers on the T/C cell is affixed inside the electronics drawer. s
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until T/CZ# XXXX is displayed in the message center.
3. Press up/down arrow to change the zero number to match the sticker.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
3-4 Model 310 Series DH Incubator Thermo Scientific
Section 3
Configuration
Enable CO2 Alarms to
Trip Contacts
Page 34
Set a Low RH Alarm
Limit
If a new T/C CO2 sensor is being installed, the two numbers on the factory installed sticker on the T/C cell must be entered to calibrate the CO
2 in the unit.
Note For the technician’s convenience, a label containing the two numbers on the T/C cell is affixed inside the electronics drawer.
s
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until T/CS# XXXX is displayed in the message center.
3. Press up/down arrow to change the span number to match the sticker.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
On units that have the RH option installed, a low RH alarm limit may be entered. The low RH alarm limit is the %RH in the cabinet which will cause a low RH alarm. The setpoint is variable from setpoint 0 to 90% RH The factory setting is 0% RH which disables the alarm.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until RH Lo XX is displayed in the message center.
3. Press up/down arrow to change the RH low alarm limit.
4. Press Enter to save the RH low alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
Model 310 Series DH Incubator 3-5Thermo Scientific
Section 3
Configuration
Set New Span Number
for T/C CO
2 Sensors
Page 35
The low RH alarm can be programmed to trip the remote alarm contacts. A setting of ON will cause this, a setting of OFF will not allow the RH alarm to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until RH Rly XXX is displayed in the message center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
On units that are equipped with the RH option, the upper seven segment display on the control panel can be configured to display Temp continuously, RH continuously, or toggle between Temp and RH. If the units does not have RH, the upper display will always display temperature. If temperature is set to ON and the RH is set OFF, temperature will be displayed continuously. If temperature is set to OFF and RH is set to ON, RH will be displayed continuously. If both are turned ON, the display will toggle between the two. The factory setting will default to toggle mode if the RH option is present.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Disp Tmp XXX or Disp RH XXX is displayed in the message center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
3-6 Model 310 Series DH Incubator Thermo Scientific
Section 3
Configuration
Enable Temp/RH to be
Displayed
Enable RH Alarms to
Trip Contacts
Page 36
On units equipped with the Gas Guard option, a primary tank can be selected. The primary tank will be either Tank 1 or 2. The factory setting is Tank1.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Tnk Sel X is displayed in the message center.
3. Press up/down arrow to toggle the setting between 1 and 2.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
On units equipped with the Gas Guard option, the Gas Guard system may be turned ON, or OFF if not in use. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Gas Grd XX is displayed in the message center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
Model 310 Series DH Incubator 3-7Thermo Scientific
Section 3
Configuration
Select Primary Tank
w/ Gas Guard Option
Disable the Gas
Guard System
Page 37
On units that have the RS485 option, direct communication with the Model 1535 alarm system can be established. Each piece of equipment connected to the 1535 must have a unique address. An address of 0-24 can be entered for the incubator. A setting of 0 is an invalid address that the 1535 will ignore. The factory setting for the RS485 address is 0.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until RS485 XX is displayed in the message center.
3. Press up/down arrow to move the RS485 address.
4. Press Enter to save the RS485 address.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
3-8 Model 310 Series DH Incubator Thermo Scientific
Section 3
Configuration
Communications
Address for RS485
Page 38
Model 310 Series DH Incubator 3-9Thermo Scientific
Section 3
Configuration
Page 39
3-10 Model 310 Series DH Incubator Thermo Scientific
Section 3
Configuration
Page 40
Model 310 Series DH Incubator 3-11Thermo Scientific
Section 3
Configuration
Page 41
Page 42
Table 4-1. Alarms and Descriptions
Model 310 Series DH Incubator 4-1Thermo Scientific
Section 4 Alarms
The Model 310 Series incubator alarm system is shown in the table below. When an alarm is active, the message appears in the LED message center. Pressing Silence disables the audible alarm for the ringback period. However, the visual alarm continues until the incubator returns to a normal condition. The alarms are momentary alarms only. When an alarm condition occurs and then returns to normal, the incubator automatically clears the alarm condition and the message center.
When multiple alarm conditions occur, active messages are displayed in the message center one at a time, updating at 5 second intervals. Pressing Silence during multiple alarms causes all active alarms to be silenced and to ring back in 15 minutes.
The temperature alarms are disabled when the Temp set point is 10°C. The CO
2 alarms are disabled when the CO2 set point is 0.0%.
Description Message Delay Ringback Relay
No alarm condition exists SYSTEM OK ----- ----- -----
CO2 System Auto Calibrating CO2 AUTO CAL ----- ----- -----
Temp > Otemp Set point SYS IN OTEMP 0 min. 15 min. Yes
Air Temp Sensor Fault (See Section 4) TMP SNSR ERR 0 min. 15 min. No
Temperature Controller Failure (See Section 4) TMP CTRL ERR 0 min. 15 min. YES
CO2 Sensor Fault (See Section 4) CO2 SNSR ERR 0 min. 15 min. No
CO2 Auto Zero Fault (IR option only, See Section 4) IR AUTOZ ERR 0 min. 15 min. No
CO2 Sensor Cannot be Calibrated (IR option only, See Section 4) REPL IR SNSR 0 min. 15 min. No
Inner Door is Open DOOR OPEN 15 min. 15 min. No
CO2 is higher than CO2 High Tracking Alarm CO2 IS HIGH 15 min. 15 min. Programmable
CO2 is lower than CO2 Low Tracking Alarm CO2 IS LOW 15 min. 15 min. Programmable
TEMP is lower TEMP Low Tracking Alarm TEMP IS LOW 15 min. 15 min. Programmable
RH is lower than RH Low Limit Alarm (RH option) RH IS LOW 30 min. 15 min. Programmable
Tank 1 is low, switch to Tank 2 (Gas Guard only) TANK1 LOW 0 min. ---- No
Tank 2 is low, switch to Tank 1 (Gas Guard only) TANK2 LOW 0 min. ---- No
Both tanks are low (Gas Guard only) TANK 1 and 2 LOW 0 min. 15 min. No
- All alarm delays and ringback times are ±30 seconds -
Page 43
4-2 Model 310 Series DH Incubator Thermo Scientific
Section 4
Alarms
Temp Control Failure
Alarm TMP CNTR ERR
In addition to other safety features designed into Model 310 Series incubators, a thermostat is also provided to monitor the cabinet’s temperature. In the unlikely event of a temperature control failure, the thermostat will turn off all heaters at a cabinet temperature of 65°C, ±10%. This is intended to be a safety feature to protect the incubator, and is not intended to protect the cell cultures or the equipment inside the chamber should a temperature control failure occur.
The microprocessor in Model 310 Series incubators continually scans all available sensors to ensure that they are operating properly. Should an error be detected, the incubator sounds an alarm and displays the appropriate message. If such an alarm occurs, contact your local distributor or the Technical Services department at 740-373-4763 or 1-800-438-4851 (USA and Canada) or fax 740-373-4189.
If the cables or connectors between the main microprocessor board and the CO
2 sensor or between the CO2 board and the sensor head on the I/R
CO
2 units become loose or disconnected, the CO2 SENS ERR alarm
occurs.
I/R Units
On I/R incubators, the red light on the I/R module (see Figure 5-8 for module circuit board location) will be lit continuously if communication is lost between the CO
2 board and the sensor head. The CO2 display will
also be locked at 00.0 without injection. Turning the incubator Off and On will not clear the alarm. Only proper connection of all components will correct the alarm.
On incubators equipped with I/R CO
2 control, calibration is done
automatically using an auto zero system. Auto zero occurs once every 24 hours. If, during the auto zero cycle, a CO
2 correction of more than 0.45%
is detected, the IR AUTOZ ERR alarm occurs.
Sensor Fault Alarms
CO2 SNSR ERR
IR AUTOZ ERR
Page 44
Model 310 Series DH Incubator 4-3Thermo Scientific
Section 4
Alarms
Possible problems which cause this alarm are:
• Auto Zero pump, orifice, filter or tubing will not let air to the sensor.
Possibilities are:
Defective or electrically disconnected air pump
Kinked auto zero vinyl tubing
Disconnected tubing between the air pump and the sensor
Plugged filter or orifice on the auto zero assembly
Defective auto zero circuit
• Cabinet temperature has been increased significantly from a previous setpoint. For example, the unit was calibrated and operating at 30°C and the setpoint is increased to 50°C. In this example, calibrating the CO
2 will correct the alarm.
• High background CO
2 in the laboratory. This could be from leaks in
the tank, regulator or vinyl CO
2 tubing.
• High CO
2 sensor calibration drift. This will require replacement of the
sensor.
• Calibration of the CO
2 system to less than 3% actual CO2. If this has
occurred, contact the Technical Services Department.
IR AUTOZ ERR
(continued)
Page 45
4-4 Model 310 Series DH Incubator Thermo Scientific
Incubators
PREVENTIVE MAINTENANCE
sensor.
2
, distilled or demineralized water.
Do NOT use bleach or any disinfectant that has high chloros
Use sterile
Avoid spraying cleaner on the CO
Do not use powdered gloves for tissue cultures.
Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your unit
functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum performance and
efficiency, it is recommended the unit be checked and calibrated periodically by a qualified service technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the operating manual for further
details.
Thermo Fisher Scientific has qualified service technicians, using NIST traceable instruments, available in many areas. For more information
on Preventive Maintenance or Extended Warranties, please contact Technical Services.
Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.
Tips for all incubators:
Page 46
Model 310 Series DH Incubator 4-5Thermo Scientific
Years
Yearly 2
Months
Between experiments
depending on use and environmental conditions
More frequent decontamination may be required,
Preventive Maintenance for Direct Heat Incubators
Action Daily Weekly Monthly 6
, humidity and temperature calibration, as
2
, O
2
Inspect door latch, hinges and door gasket seal.
Check water level in the humidity pan, ½” from top. See “Filling the
Humidity Pan”.
Verify and document CO
applicable. See Calibration section
Perform a complete decontamination procedure. Wipe down interior,
shelves and side ducts with disinfectant. Change or clean blower wheel and
scroll. Clean top duct. Clean humidity pan. Rinse everything well with
sterile distilled water. See “Cleaning the Interior, “Cleaning the Humidity
Pan”.
Change HEPA and gas filters, if applicable (or as needed). See “Installing
Air Sample Filter”, “Installing HEPA Filter”.
Replace AquaTec water cleaning cell, if applicable.
Page 47
Page 48
Model 310 Series DH Incubator 5-1Thermo Scientific
Section 5 Routine Maintenance
Warning If the unit has been in service, turn it off and disconnect the
power cord connector before proceeding with any maintenance. s
Caution Before using any cleaning or decontamination method except those recommended by the manufacturer, users must check with the manufacturer that the proposed method will not damage the equipment.
s
Use an appropriate disinfectant. All articles and surfaces to be disinfected must be thoroughly cleaned, rinsed with sterile water, and rough dried.
Warning Alcohol, even a 70% solution, is volatile and flammable. Use it only in a well ventilated area that is free from open flame. If any component is cleaned with alcohol, do not expose the component to open flame or other possible hazard. Do not spray T/C sensor with flammable solutions. The internal temperature of the CO
2 sensor is approximately 150°C when the unit is in
operation. Allow sufficient time for the sensor to cool before cleaning.
s
Caution Do not use strong alkaline or caustic agents. Stainless steel is corrosion resistant, not corrosion-proof. Do not use solutions of sodium hypochlorite (bleach) as they may also cause pitting and rusting. s
1. Remove the shelves, HEPA filter (a factory installed option), access port filter, sample air filter, left and right duct sheets and the top duct. The top duct requires the two wing nuts to be removed.
2. Set the HEPA filter aside. Wash the shelves, ducts, wing nuts and stopper with disinfectant. Optional: The shelves, ducts and wing nuts may be autoclaved.
3. Remove the blower scroll held on by four screws. Remove the blower wheel and the access port filter. If a new blower wheel and scroll are going to be used, discard the old ones. If the old ones are being reused, wash and rinse all parts with disinfectant.
Disinfecting the
Incubator Interior
Page 49
4. The CO2 sensor may be lowered by unscrewing the retaining ring around the sensor. This will allow the sensor gasket to be removed, cleaned and replaced. WIPE the CO
2 sensor, do not spray or soak.
5. Wash the inner door gasket with disinfectant. The inner door gasket may be removed and/or replaced.
6. Wash the cabinet interior with disinfectant, starting at the top and working down. Refer to the disinfectant directions for length of time needed before rinsing. Wash the inner door both inside and out. The cabinet and door must be rinsed with sterile water until the disinfectant has been removed. After the cabinet has been rinsed, spray with 70% alcohol.
7. If the CO
2 sensor gasket was removed, spray it with 70% alcohol and
install it. Install the sample filter.
8. Install the top duct, left and right ducts, inner door gasket, and access port stopper spraying each with 70% alcohol.
9. Install new HEPA Filter.
10. Install the shelves and spray with 70% alcohol.
Clean the incubator exterior with a damp sponge or soft, well-wrung cloth and mild detergent dissolved in water. Dry with a soft cloth.
The chamber glass door and the optional independent inner glass doors may be cleaned using the same disinfectant used on the incubator interior. It is imperative that they be rinsed with sterile distilled water to remove the disinfectant residue. The doors should then be dried with a soft cloth.
Precautions in the cleaning and care of the incubator glass doors:
Moisture leaches alkaline materials (sodium, Na) from the surface of the glass. Evaporation of the moisture concentrates the alkaline and may produce a white staining or clouding of the glass surface. Cleaning chemicals with a pH above 9 and heat (autoclaving) accelerate the corro­sion process. Therefore, it is very important to rinse and dry the glass doors after cleaning. Autoclaving the glass doors should be avoided.
There is no simple method for repairing corroded glass. In most cases, the glass must be replaced.
5-2 Model 310 Series DH Incubator Thermo Scientific
Section 5
Routine Maintenance
Cleaning the Glass
Doors
Cleaning the Cabinet
Exterior
Disinfecting the
Interior (cont.)
Page 50
Clean the humidity pan with soap and water and a general use laboratory disinfectant. Rinse with sterile water and spray with 70% alcohol. The humidity pan may be autoclaved.
For side-by-side operation or changing lab layouts, the inner and outer doors are field reversible. The procedure is written from the prospective of changing the door swing from a left-hand to a right-hand swing. All screw holes are pre-drilled for reversing the door. The tools required are a Phillips screwdriver, a 1/8” Allen wrench and a flatblade screwdriver.
The door reversing procedure takes about 30 minutes.
Model 310 Series DH Incubator 5-3Thermo Scientific
Section 5
Routine Maintenance
Reversing the Door
Swing
Cleaning the Humidity
Pan
Figure 5-1. Door Swings
Page 51
1. Remove the kick plate at the bottom of the cabinet by removing the two Phillips screws indicated by the black arrows in the illustration. Disconnect the cable inside the cabinet.
2. Open the outer door and remove it by lifting it off its hinge pins. Place the door on a padded surface to prevent scratches (Figure 5-2).
Note The hinge pins are not connected to the brackets and may fall out when the door is lifted off.
s
3. Refer to the illustration of the outer door in Figure 5-2. With a Phillips screwdriver, remove and save the four nylon screws from the door hinge mounting holes. The black arrows identify the screw locations.
4. Remove the door hinges, rotate them 180° and secure them on the left side of the door. Install the nylon screws into the newly exposed holes.
5. Remove the two plastic clips securing the door heater cable. These clips will be rotated 180° when installed on the left lower edge of the door frame.
5-4 Model 310 Series DH Incubator Thermo Scientific
Section 5
Routine Maintenance
Figure 5-2. Remove Two Screws
Reversing the Door
Swing (continued)
Page 52
6. Rotate the kick plate and attach the door heater cable with the clips as shown in Figure 5-3. The insert in the illustration shows the proper positioning of the clips.
7. The incubator and door should look similar to Figure 5-4.
Model 310 Series DH Incubator 5-5Thermo Scientific
Section 5
Routine Maintenance
Figure 5-3. Rotate Kickplate
Figure 5-4. Unit with Detached Door
Reversing the Door
Swing (continued)
Page 53
8. Refer to Figure 5-5 for the next steps. Remove the nylon screws along the right side of the cabinet. Remove the outer door hinges. Rotate them 180° and install them on the right side of the cabinet. Install the nylon screws in the holes vacated by the hinges.
9. With a 1/8” Allen wrench, remove the door open switch target plate on the top of the inner door.
10. Remove the door latch striker plate.
11. Hold the door securely and remove the upper hinge bracket on the cabinet. This hinge bracket will be rotated 180° and be placed in the lower right corner of the cabinet where indicated.
12. Lift the glass door from the lower hinge and set it aside. Then remove the lower hinge from the cabinet. This hinge bracket will be rotated 180° and be placed in the upper right corner of the cabinet at where indicated.
13. Remove the screws and washers from the new striker plate location. These screws and washers can be threaded into the holes vacated by the striker plate where indicated.
14. Rotate the striker plate 180° from its original position and attach it to the cabinet where indicated.
15. Using a flatblade screwdriver, remove the threaded nylon plugs from the new hinge bracket locations. Install the lower hinge bracket where indicated.
5-6 Model 310 Series DH Incubator Thermo Scientific
Section 5
Routine Maintenance
Figure 5-5. Component Locations
Reversing the Door
Swing (continued)
Page 54
16. Rotate the door 180° and place it in the lower hinge bracket. Install the upper hinge bracket where indicated.
17. Screw the threaded plugs into the holes vacated by the door hinge brackets where indicated in Figure 5-5.
18. Install the door open switch target plate. Refer to Figure 5-6 and 5-7. The plate must be positioned exactly 12.5 inches from the outside edge of the cabinet to the leading edge of the plate. Do not overtighten.
19. Connect the door heater cable and install the kick plate with the door heater cable on the right side.
20. Installing the outer door onto the hinge pins completes the door reversal procedure.
Model 310 Series DH Incubator 5-7Thermo Scientific
Section 5
Routine Maintenance
Figure 5-6. Reversed Components
Figure 5-7. Plate Position
Reversing the Door
Swing (continued)
Page 55
Replace the HEPA filter routinely on an annual basis, or more often if it changes color. The filter should also be changed when the chamber is cleaned.
Warning The electronics section contains hazardous voltages. Only qualified service personnel should access this area.
s
There are only two replaceable fuses in the incubator. To access the fuses:
1. Turn off the incubator’s power switch and unplug the power cord.
2. Open the outer door to the chamber.
3. Remove the two screws located below the control panel.
4. Lift up on the control panel, then pull it out.
5. Refer to Figure 5-8 for the location of the two fuses.
6. Slide the control panel back in, replace the screws and return the unit to service. If the fuse(s) blow after restoring power to the incubator, contact Technical Services Department.
5-8 Model 310 Series DH Incubator Thermo Scientific
Section 5
Routine Maintenance
Replacing the Power
Fuses
Forma
Figure 5-8. Major Components in Electronics Drawer
HEPA Filter (Factory
Installed Option)
CO
CO manifold and
2
supply tank connectors
RS485 and
remote alarm contacts
sensor board
output board
CO sensor
IR CO
Analog
microbiological
IR_T/C
2
2
2
filter
CO gas
solenoid
Humidity
sensor
Overtemp thermostat
Power
transformer,
export
2
Temp probe
CO
injection
port
Micro board
Accessory
outlet
75W max.
Sample port into chamber
Blower
2
motor
Fuses
Power switch
Filtered line cord connector
Main power transformer
Power supply board
Page 56
Electronics Section
Expanded descriptions for some of the components indicated in Figure 5-8 follow.
Power Switch - Located beneath the control panel. The outer door must
be open to access the switch.
RS 485 and Remote Alarm Contacts - Connectors and auxiliary wire
through port
CO
2 manifold and supply tank connectors - Part of the Gas Guard sys-
tem, a factory installed option.
Humidity sensor (optional)
Analog output board - 4-20 mA, 0-1 Volt, 0-5 Volt
1. Turn off the incubator’s power switch and unplug the power cord.
2. Remove the two screws from the front of the cabinet top and the two screws in the uppermost corners of the back panel.
3. Slide the top about an inch to the rear and lift it off.
Note There are small locking tabs in the back which will be bent if the top is lifted off without sliding it first to the back.
s
Model 310 Series DH Incubator 5-9Thermo Scientific
Section 5
Routine Maintenance
Major Components
Fuse Voltage Manufacturers Part #
Amperage
Rating
Rupture Speed
IEC Letter
Code
115 VAC ACC BUSS GMC - 1.0A 1.0 Amp Time-Lag T
230 VAC ACC BUSS BK-GMC-500ma 0.5 Amp Time-Lag T
24 VAC Door Heat Schurter 034.3124 5.0 Amp Time-Lag T
Table 5-1. Fuses
Replacing the Power
Fuses (cont.)
Page 57
1. Connect one end of the sample port filter to the ceiling port.
2. Connect the other end of the filter to the back of the blower fan inlet.
Figure 5-9. Air Sample Filter
5-10 Model 310 Series DH Incubator Thermo Scientific
Section 5
Routine Maintenance
Replacing the Sample
Air Filter
Page 58
Section 6 Factory Options
Descriptions for connections to external equipment follow.
A set of relay contacts is provided to monitor alarms through an RJ11 telephone style connector on the back of the cabinet. Refer to Figure 6-3 for the location of the alarm connector. The 12-foot telephone cord (P/N
190388) and RJ11-to-screw terminal conversion box (190392) are available through our service department.
The remote alarm provides a NO (normally open) output, an NC (normally closed) output and COM (common). Refer to Figure 6-1.
The contacts will trip on a power outage or an overtemperature condition. The contacts may also be programmed to trip or not trip on temperature alarms, CO
2 alarms, and RH alarms. See Section 3, Configuration Mode.
Figure 6-1. Remote Alarm Output
Model 310 Series DH Incubator 6-1Thermo Scientific
Connect Remote
Alarm Contacts
CAUTION
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
Page 59
All incubator models can be purchased with the RS485 communications option (P/N 190523). This option allows the incubator to be directly connected to a Model 1535 alarm system without the use of a communications module. A junction box is provided with each RS485 option. See Figure 6-2 for wiring details. Figure 6-3 shows the RS485 connector location on the back of the incubator cabinet.
To allow the incubator and the 1535 to communicate, an address must be allocated on the 1535. Refer to Section 5.8 of the 1535 operating manual. The same address number must be assigned to the incubator. Refer to Section 3 of this manual.
Warning The electronics section contains hazardous voltages. Only qualified personnel should access this area.
s
The analog output board option (P/N 190512, 190543, 190544) allows the incubator to output analog signals representing the air temperature of the cabinet, CO
2 content, and relative humidity, depending upon which
systems are in the incubator. There are three different analog output board options available : 0-1V, 0-5V, or 4-20mA signals. Negative display readings will output 0V. The outputs do not have isolated grounds. Refer to Table 6-1 for output specifications of the three boards.
For the 0-1V and 0-5V boards, the recording device must supply a load >/= 1000 ohm. For the 4-20mA board, the recording device must supply a load of </= 100 ohm.
6-2 Model 310 Series DH Incubator Thermo Scientific
Section 6
Factory Options
Connect RS485
Interface
Connecting the
Analog Output Boards
Figure 6-2. RS485 Wiring
Page 60
Model 310 Series DH Incubator 6-3Thermo Scientific
Section 6
Factory Options
To wire in the analog output board, a shielded 22 gauge, 3-conductor wire (P/N 73041) is recommended, with the maximum length being 50 feet (15.2m). This wire is readily available from other vendors including Alpha P/N 2403 and Deerborn P/N 972203.
Figure 6-3. Auxiliary Through-Port on Back Panel
Caution Accuracy of the output at the board terminal strip to the
incubator display is ±1 unit. There is no calibration from the incubator. Calibration to the incubator display must be at the instrument connected to the output board.
s
To install the analog board:
1. Turn off the incubator and unplug it from the wall outlet.
2. Remove the top of the electronics section. Refer to Section 5 of this manual.*
3. Locate the Analog Output board, Figure 6-4 and Figure 5-8.
* Alternate access to the electronics is to open the outer door and remove the two screws in the upper corners of the cabinet just under the display top. Lifting up the top section, the electronics drawer can be pulled out to the limit of the internal wiring.
190512 4-20 mA Output Scaling 4-20 mA Equals
190544 0-1V Output Scaling 0-1 V Equals
190543 0-5V Output Scaling 0-5V Equals
Temperature 0.0-100.0°C 0.0-100.0°C 0.0-100.0°C
RH 0-100 %RH 0-100 %RH 0-100 %RH
CO2 0.0-100.0 %CO2 0-100.0 %CO2 0-100.0 %CO2
Table 6-1. Analog Output Boards
Connect Analog
Output Boards (cont.)
RS 485
ALARM CONTACTS 30V MAX / 1A MAX
#2
#1
CO inject
CO inject
ACCESSORY
75 WATTS MAX.
22
15 p.s.i. Regulated
15 p.s.i. Regulated
(103.4 kPa)
(103.4 kPa)
ALARM CONTACTS 30V MAX / 1A MAX
RS 485
Page 61
Figure 6-4. Connector Terminals
4. Each system monitored (Temp, CO
2, RH) requires two conductors
(refer to Figure 6-4). Feed the wires through the auxiliary wire through-port on the back panel. This port is a circular fitting adjacent to the remote alarm and RS485 connectors as illustrated in Figure 6-3. The wire routing is shown by the black arrows in Figure 6-5.
5. Strip the ends of each conductor and wire it to the appropriate terminals of connectors J2 and/or J3 on the analog board. Refer to Figure 6-4.
6. When wiring is completed, slide in the electronics drawer or the replace the sheet metal cabinet top.
7. Replace the screws removed earlier and return the unit to service.
6-4 Model 310 Series DH Incubator Thermo Scientific
Section 6
Factory Options
Figure 6-5. Wire Routing
Connect Analog
Output Boards (cont.)
Page 62
Warning CO2 can be lethal in high concentrations. Refer to warnings in Section 1 of this manual. s
The Model 310 Series incubators can be equipped with a built-in Gas Guard system (P/N 190640) that will operate with a CO
2 gas supply. The
Gas Guard uses two pressure switches to continuously monitor the pressures of two independent CO
2 supplies and automatically switches
from one supply to the other when the supply pressure drops below 10 psig (0.690 bar). The Gas Guard’s design does not facilitate use by multiple incubators.
The CO
2 gas supplies must be equipped with two-stage pressure regulators
with gauges. The high pressure gauge at the tank should have a 0-2000 psig range and the low pressure gauge should have a 0-30 psig range. The gas supply to the incubator must be maintained at 15 psig (1.034 bar), ±5 psig. Gas pressures below 15 psig will cause nuisance alarms to occur on incubators equipped with the built-in Gas Guard.
The CO
2 inlets for the Gas Guard are located on the back of the cabinet.
Using 1/4” ID tubing, connect one of the CO
2 supply tanks to the fitting
labeled CO2 Inlet #1 Tank. Connect the second CO
2 supply tank to the
fitting labeled CO2 Inlet #2 Tank. Install 3/8” hose clamps to secure the 1/4” ID tubing to the fittings on the back of the drawer. (Refer to Figure 6-6.)
The built-in Gas Guard is turned ON when shipped from the factory. In addition, the Tank Sel for the Gas Guard is specified as Tank 1 when shipped. Refer to Section 3, Configuration, to de-activate the Gas Guard or change the Tank Sel from #1 to #2. If the Gas Guard system is not used, the incubator will function normally by supplying CO
2 from the
supply tank connected to Inlet #1.
Model 310 Series DH Incubator 6-5Thermo Scientific
Section 6
Factory Options
Figure 6-6. Connect Gas Supplies
De-activating the Gas
Guard
Connecting the CO
2 Gas
Supplies
CO2 Gas Guard
Page 63
With the Gas Guard in operation, the incubator will use the gas supplied by the tank connected to Inlet #1 until the pressure drops below 10 psig (0.690 bars). At this time, the Gas Guard automatically switches to gas supplied through CO2 Inlet #2.
In addition, the incubator automatically changes the Tank Sel in Configuration Mode from 1 to 2 to indicate that the incubator is now using gas supplied through Inlet # 2. If the gas supply to Inlet #1 is replenished, the incubator will continue to operate using the gas supplied through Inlet #2 unless the operator changes the Tank Sel from #2 to #1 through Configuration Mode. Refer to Section 3, Configuration.
Audible and visual alarms occur on the control panel when the gas guard switches from one supply to the other. The audible alarm sounds until the operator presses the Silence key on the control panel. The visual alarm in the Message Center reads Tank 1 Low while the audible alarm is sounding, but the message is removed when the operator presses the Silence key. However, the ‘Tank Low’ indicator on the control panel stays lighted until the condition is resolved. The unit operates normally.
Caution Both the audible and visual alarms described above do not ring back once the Silence key is pressed.
s
If the Gas Guard system does not detect an adequate gas supply at CO2 Inlet #1 or Inlet #2, a visual and audible alarm will again occur on the control panel. The visual alarm in the Message Center will read Tank 1&2 Low. The audible alarm will continue to ring until the Silence key is pressed. The audible alarm will ring back every 15 minutes after the alarm is silenced if the Gas Guard continues to detect that both gas supply pressures are below 10 psig (0.690 bars).
The 310 Series incubators can be equipped with a humidity sensor (P/N
190643) to monitor the relative humidity inside the chamber. The sensor is mounted to the top air duct and provides a signal that is displayed in 1% increments on the control panel. The humidity readout can be displayed continuously or toggles with the temperature readout. In addition, a low alarm limit can be set on the humidity readout, which will detect when the humidity pan runs dry. Refer to Section 3, Configuration.
6-6 Model 310 Series DH Incubator Thermo Scientific
Section 6
Factory Options
Humidity Readout
Operation of the CO2 Gas
Guard
Figure 6-7. Silence Key
Page 64
• Water level in the humidity pan
• Frequency of door openings
• The humidity pan located on bottom of chamber versus a shelf.
Caution Incubators equipped with a Thermal Conductivity CO
2 sensor
rely on a constant level of relative humidity in order to accurately measure and control the CO
2 concentration in the incubator. s
The sensor is capable of measuring relative humidity from 10% to 100% with an accuracy of ±5% above 90%. See Section 2, Calibration, for details on calibrating the humidity readout.
Warning The Uninterruptible Power Supply is not rated for life-support applications.
s
The optional Uninterruptible Power Supply ([UPS] P/N 270078 and
270082) provides a back-up power source for the incubator. When the batteries are fully charged and the incubator operating at 37°C in an 18°C ambient temperature, the UPS provides approximately one hour of uninterrupted power during a power failure. A higher operating temperature decreases this back-up time, a higher ambient temperature increases it. This estimated time assumes that the incubator will be already at its operating temperature and that the incubator door will not be opened during the back-up period.
The remote alarm contacts will not trip during a power failure while connected to the UPS until battery power is exhausted. There is however, an audible and visual alarm when the unit switches to battery operation.
The UPS batteries are fully charged when the unit has been connected to utility power for a minimum of four hours. The UPS power switch does not have to be on for the batteries to charge. When battery replacement becomes necessary, refer to the UPS user’s manual.
Caution Use only factory recommended UPS units which produce a pure sine wave output. Other UPS units could damage the incubator and void the warranty. Typical facility back-up systems will not effect the efficiency of the incubator.
s
Model 310 Series DH Incubator 6-7Thermo Scientific
Section 6
Factory Options
Uninterruptible Power
Supply Connections
Humidity Level Factors
Humidity Readout
Accuracy
Page 65
1. Plug the incubator line cord into one of the receptacles on the back of the UPS. Then plug the UPS line cord into a suitable wall outlet.
2. Turn the UPS power switch on.
Use the line cord supplied with the incubator to connect the UPS to a suitable wall outlet, and the UPS coupler to connect the incubator to the UPS.
Warning The Uninterruptible Power Supply contains potentially hazardous voltages even when not connected to line voltage. Do not attempt to disassemble UPS as it contains no user-serviceable parts. Repairs must be performed by factory trained service personnel.
s
6-8 Model 310 Series DH Incubator Thermo Scientific
Section 6
Factory Options
Connecting UPS
P/N 270078, 115V, 50/60Hz
Connecting UPS
P/N 270082, 230V, 50/60Hz
Page 66
Section 7 Specifications
* Specifications are based on nominal voltages of 115V or 230V in ambient of 22°C
to 25°C.
Temperature
Control . . . .±0.1°C Microprocessor PID Control
Setpoint . . . . . . . . . . .Digital - Touch pad, 0.1°C
Range . . . . . . . . . . . .+5°C above ambient to 50°C
Uniformity . . . . . . . . . . . . . . . . . .±0.3°C @ +37°C
Tracking alarm . . User programmable (low) indicator
Overtemp . .Tracking, user programmable, action, and indicator
Display . . . . . . . . .Digital, LED, 0.1°C increments
Temperature Safety
Type . . Extreme temperature safety, action, and indicator
Sensor . . Thermostat, independent of temperature control system
Indicator . . Message center, audible and visual alarms
Relative Humidity
Control . . . . .Humidity pan - natural vaporization
Humidity with pan . . . . . . . . . . .95% RH at 37°C
Display . . . . . . . . . . . . .Optional in 1% increments
Alarm . . . . . . .Low RH with optional RH monitor
CO
2
Control . . . . .±0.1% microprocessor PID control
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . .T/C or IR
Readability . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1%
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 20%
Inlet pressure . . . . . . . . . . . . . . . . .15 psig, ±5 psig
Display . . . . . . . . . .Digital LED, 0.1% increments
Model 310 Series DH Incubator 7-1Thermo Scientific
Page 67
7-2 Model 310 Series DH Incubator Thermo Scientific
Section 7
Specifications
Shelves
Dimensions . . . . . . . .18.5” x 18.5” (47cm x 47cm
Construction . . Stainless steel (belt sanded, both sides)
Surface area . . . . . .2.4 sq. ft. (0.22 sq. m) per shelf
Max. per chamber . . . . . . . . .38.4 sq ft (3.6 sq. m)
Loading . . . .35 lbs (16kg) slide in and out, 50 lbs (23kg) stationary
Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Maximum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Construction
Interior volume . . . . . . . . . . . . . . . . . . . .6.5 cu. ft.
Interior . . . . . .Type 304 stainless steel shiny finish
Exterior . . . . . . . . . . . . . . . .18 gauge cold roll steel
Outer door gasket . . Four-sided molded, magnetic vinyl
Inner door gasket . . . . . . .Feather-edge, removable
Insulation . . . . . . . . . . . . . . . . . . . . . . . .Fiberglass
Fittings
Access port . . . .1-1/4 inch removable silicone plug
CO
2 inlet . . . . . . . . . . . . . . . . . . . . . . .1/4” barbed
Sample port . . . . . . . . . . . . .Front mounted barbed
Electrical
115 Volt models -
90-125VAC, 50/60 Hz, 1PH, 2.4 FLA
230 Volt models -
180-250, 50/60 Hz, 1 PH, 2.0 FLA
Power switch/circuit breaker . . . . .2 Pole, 6.0 Amp
Accessory outlet . . Voltage equal to the cabinet input, 75 Watts maximum, 0.5ma leakage current
Alarm contacts . . Deviation of temperature, CO
2, power, NO and NC
Unit BTU Output
115/230V . . . . . . . . . . . . . .344 BTUH (100 watt)
Dimensions
Exterior . . . . . . . . .26.0” W x 38.5” H x 25.0” F-B
Interior . . . . . . . . .21.4” W x 26.8” H x 20.0” F-B
Page 68
Model 310 Series DH Incubator 7-3Thermo Scientific
Section 7
Specifications
Weight
Net Weight Stainless Steel . . . . . . . . . . . . .205 lbs
Net Weight Copper . . . . . . . . . . . . . . . . .230 lbs.
Options/Accessories
HEPA filter . . Factory installed: Stock no. 190858
Battery back-up . . Free Standing: Stock no. 270078, 120VAC, 60Hz
Stock no. 270082, 220VAC, 50Hz
Humidity . . Factory installed: Stock no. 190463 - provides humidity
readout and low alarm
Inner door kit . . Customer installed: Stock no. 190650 - independent
. . . eight separate glass doors
Built-in Gas Guard . . Factory installed: Stock no. 190640 - CO
2
only, automatically switches supply tanks
Recorders . . Customer installed: Standard recorders used on water-
jacket incubators
Decontamination kits
HEPA filter, blower assembly, inner door gasket, etc. ­ Stock no. 190868 (with HEPA filter),
Stock no. 190869 (without HEPA filter)
Copper Options
Interior ductwork .Solid copper: Stock no. 190725
Perforated shelfCustomer Installed: Stock no. 224166 - solid copper
Humidity PanCustomer Installed: Stock no. 237015 - solid copper
Copper interior kitFactory installed: Replace stainless steel duct work, shelves, and
humidity pan with solid copper Stock no. 190870
Optional Data Outputs
Analog Factory installed: Stock no. 190544 - 0-1 volt
Analog Factory installed: Stock no. 190543 - 0-5 volt
Analog Factory installed: Stock no. 190512 - 4-20mA
Digital Factory installed: Stock no. 190523 - RS-485
(Compatible with Model 1535 Alarm Monitor only)
Certifications
Refer to the Declarations of Conformity at the back of this manual
Page 69
Safety Specifications
Indoor Use Only
Altitude - 2,000 meters
Temperature - 5°C to 40°C
Humidity - 80% RH at or below 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
Installation Category II
1
Pollution Degree 2
2
Class of Equipment I
1
Installation Category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the category used for instruments in installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2
Pollution Degree describes the amount of conductive pollution present in the operating environment. Pollution Degree 2 assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity caused by condensation.
7-4 Model 310 Series DH Incubator Thermo Scientific
Section 7
Specifications
Page 70
Section 8 Spare Parts
Model 310 Series DH Incubator 8-1Thermo Scientific
Part Number Description
190699 Removable feather gasket
190670 Duct sheet and shelf channel kit
190772 Micro board, (main control)
230135 Fuse, 1A, TD, 5 x 20mm
230120 0.5A fuse, TD, 5 x 20mm (accessory outlet)
230161 Fuse, 5A, 5 x 20mm
156126 Motor, 2-pole, 115VAC, 50/60 Hz
190793 Display board
290184 Probe, 2252 Ohm/25°C, 1/8 x2
770001 Disposable filter, 99.97 (CO2 line, air sample)
760175 HEPA filter, 5.5” diameter
760199 HEPA2 VOC Filter Replacement Kit (includes 760200 HEPA2, 117036 silicone plug, 101018 o-ring)
1900067 Filter Replacement Kit (includes [2] 770001 filter and 760175 HEPA)
1900094 HEPA2 VOC Filter Replacement Kit (760200 HEPA2, 190985 access port filter, 770001 filter)
224175 Shelf
190868 Decontamination kit with HEPA filter
190869 Decontamination kit without HEPA filter
130077 Stopper with 3/8” hole
290090 Thermal conductivity type CO2 sensor
190885 Infrared type CO2 sensor w/ auto zero
250087 CO2 valve with fittings
290154 RH sensor assembly (RH display option)
190883 Sample port filter assembly
190661 IR sensor filter
190985 Filtered air exchange assembly
230153 6A, DPDT circuit breaker switch (power)
400201 40W Switcher kit
420096 130VA transformer, int’l, 12/24V S
194021 34 position control to display ribbon cable
420097 43VA transformer (230VAC units only)
360213 Pressure switch (Gas Guard option)
250121 Gas valve (Gas Guard option)
190512 4-20mA output board
190543 0-5V output board
190544 0-1V output board
Page 71
8-2 Model 310 Series DH Incubator Thermo Scientific
Section 8
Spare Parts
Page 72
Model 310 Series DH Incubator 8-3Thermo Scientific
Section 8
Spare Parts
Page 73
8-4 Model 310 Series DH Incubator Thermo Scientific
Section 8
Spare Parts
Assembly Drawing
Model:
310/3110 Series
2
CO Incubator
310-202-1-D Rev. 6
Page 1 of 1
Page 74
Model 310 Series DH Incubator 8-5Thermo Scientific
Section 8
Spare Parts
Page 75
8-6 Model 310 Series DH Incubator Thermo Scientific
Section 8
Spare Parts
Page 76
Model 310 Series DH Incubator 8-7Thermo Scientific
Section 8
Spare Parts
Page 77
Page 78
Model 310 Series DH Incubator 9-1Thermo Scientific
Section 9
Electrical Schematics
Electrical Schematic
Model:
115V Air Jacket
Incubator
310-70-0-D Rev. 6
Page 1 of 3
Page 79
9-2 Model 310 Series DH Incubator Thermo Scientific
Section 9
Electrical Schematics
Electrical Schematic
Model:
115V Air Jacket
Incubator
310-70-0-D Rev. 6
Page 2 of 3
Page 80
Model 310 Series DH Incubator 9-3Thermo Scientific
Section 9
Electrical Schematics
Electrical Schematic
Model:
115V Air Jacket
Incubator
310-70-0-D Rev. 6
Page 3 of 3
Page 81
9-4 Model 310 Series DH Incubator Thermo Scientific
Section 9
Electrical Schematics
Electrical Schematic
Model:
311 & 321
Air Jacket Incubator
311-70-0-D Rev. 5
Page 1 of 3
Page 82
Model 310 Series DH Incubator 9-5Thermo Scientific
Section 9
Electrical Schematics
Electrical Schematic
Model:
311 & 321
Air Jacket Incubator
311-70-0-D Rev. 5
Page 2 of 3
Page 83
9-6 Model 310 Series DH Incubator Thermo Scientific
Section 9
Electrical Schematics
Electrical Schematic
Model:
311 & 321
Air Jacket Incubator
311-70-0-D Rev. 5
Page 3 of 3
Page 84
Model 310 Series DH Incubator 10-1Thermo Scientific
Section 10
Warranty Information
ISO
9001
REGISTERED
Rev. 4 4/09
2 sensor (IR) will be warranted for a second year, parts only. The CO2
THERMO FISHER SCIENTIFIC DIRECT HEAT INCUBATOR WARRANTY USA*
The Warranty Period starts two weeks from the date your equipment is shipped from our facility. This allows shipping time so the warranty will go
into effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner during the first
year warranty period.
During the first year, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo's expense,
including labor. In addition, after the initial warranty period of one year the CO
sensor (T/C) will be warranted for four additional years for a total of five years, parts only. Installation and calibration is not covered by this warran-
ty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs.
Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the com-
ponent part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or
equipment. At Thermo's option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destina-
tion.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for
any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction
manuals carefully detail equipment installation, operation and preventative maintenance.
If equipment service is required, please call your Technical Service Department at l-800-438-4851 (USA and Canada) or 1-740-373-4763. We're
ready to answer your questions on equipment warranty, operation, maintenance, service and special applications. Outside the USA, contact your local
distributor for warranty information.
Page 85
10-2 Model 310 Series DH Incubator Thermo Scientific
Section 10
Warranty Information
ISO
9001
REGISTERED
Rev. 4 4/09
sensor (IR) will be warranted for a second year, parts only. The
2
THERMO FISHER SCIENTIFIC INTERNATIONAL DIRECT HEAT INCUBATOR WARRANTY*
The Warranty Period starts two months from the date your equipment is shipped from our facility. This allows shipping time so the warranty will go
into effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner during the first
year warranty period.
sensor (T/C) will be warranted for four additional years for a total of five years, parts only. Installation and calibration is not covered by this
2
During the first year component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo's expense,
excepting labor. In addition, after the initial warranty period of one year the CO
CO
warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any
repairs. Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the com-
ponent part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or
equipment. At Thermo's option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destina-
tion.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for
any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction
manuals carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA and Canada) or 1-740-373-4763. We're
ready to answer your questions on equipment warranty, operation, maintenance, service and special application. Outside the USA, contact your local
distributor for warranty information.
Page 86
Thermo Fisher Scientific 401 Millcreek Road Marietta, Ohio 45750 United States
www.thermofisher.com
Loading...