Thermo Scientifi? TSX Series, TSX2304BD, TSX2304BV, TSX2304BA Installation And Operation Manual

Blood Bank Refrigerators
TSX Series
Installation and Operation
327930H01 Rev. A August 2016
Visit us online to register your warranty www.thermofisher.com/labwarranty
IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result in damage to the unit, injury to operating personnel, and poor equipment performance.
Material in this manual is for informational purposes only. The contents and the product it describes are subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising from or related to the use of this manual.
© 2016 Thermo Fisher Scientific Inc. All rights reserved.
For your future reference and when contacting the factory, please have the following information readily available. It can be found on the dataplate attached to your unit.
Model Number:
Serial Number:
The following information, if available, is helpful for contacting the factory.
Date Purchased:
Purchase order number:
Source of Purchase:
(manufacturer or specific agent/rep organization)
Table of Contents
Table of Contents
Models.......................................................................................1
Safety Precautions....................................................................2
Unpacking.................................................................................4
Packing List..............................................................................5
General Recommendations.....................................................6
Temperature Monitoring ......................................................6
Intended Use ........................................................................6
Operating Standards ...............................................................7
Unit Specifications ...............................................................7
Installation................................................................................8
Location ...............................................................................8
Leveling ...............................................................................8
Wiring ..................................................................................8
Shelves .................................................................................9
Drawers ................................................................................9
Baskets ...............................................................................13
Anti-Tip Brackets ...............................................................13
Temperature Sensor Bottles ...............................................16
Door Operation ..................................................................16
Remote Alarm (Optional) ..................................................16
Final Checks .......................................................................17
Startup ....................................................................................18
Initial Startup ......................................................................18
Product Loading and Unloading Guidelines ......................18
Operation................................................................................19
Control Panel ......................................................................19
Display Functions ..............................................................21
Programming Functions .....................................................22
Service Parameters .............................................................22
Temperature Settings .........................................................24
Safety Trip ..........................................................................24
Alarms ................................................................................24
Chart Recorder......................................................................26
Set Up and Operation .........................................................26
Changing Chart Paper ........................................................27
Power Supply .....................................................................27
Calibration and Adjustments ..............................................28
Temperature Transmitter (Optional)..................................29
Maintenance...........................................................................30
Table of Contents
Cleaning the Cabinet Interior .............................................30
Cleaning the Condenser Filter ............................................30
Cleaning the Condenser .....................................................30
Automatic Defrost ..............................................................31
Gasket Maintenance ...........................................................31
Alarm Battery Maintenance ...............................................31
Troubleshooting.....................................................................32
End of Life Care.....................................................................34
Warranty ................................................................................35
Models

1 Models

The table below shows the units covered in this operation and installation manual by model number. Refer
Section 6.1
for details on the voltage
specifications for the various models.
Table 1. Applicable Models
Unit Model (*)
Thermo Scientific - TSX2304B* A/D/V
Thermo Scientific TSX Blood Bank Refrigerators 1
Safety Precautions

2 Safety Precautions

In this manual, the following symbols and conventions are used:
This symbol when used alone indicates important operating instructions which reduce the risk of injury or poor performance of the unit.
WARNING: This symbol indicates potentially hazardous situations which, if not avoided, could result in serious injury or death.
WARNING: This symbol indicates situations where dangerous voltages exist and potential for electrical shock is present.
WARNING: This symbol indicates potentially hazardous situations, which if not avoided could result in fire.
CAUTION: This symbol, in the context of a CAUTION, indicates a potentially hazardous situation which if not avoided could result in minor to moderate injury or damage to the equipment.
CAUTION: This indicates a situation which may result in property damage.
This symbol indicates surfaces which may become hot during use and may cause a burn if touched with unprotected body parts.
Before installing, using or maintaining this product, please be sure to read the manual and product warning labels carefully. Failure to follow these instructions may cause the product to malfunction, which could result in injury or damage.
This symbol indicates possible pinch points which may cause personal injury.
This symbol indicates the need to use gloves during indicated procedures. If performing decontamination procedures, use chemically resistant gloves.
2 TSX Blood Bank Refrigerators Thermo Scientific
Safety Precautions
Below are important safety precautions that apply to this product:
Use this product only in the way described in the product literature and in this manual. Before using it, verify that this product is suitable for its intended use. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Do not modify system components, especially the controller. Use OEM exact replacement equipment or parts. Before use, confirm that the product has not been altered in any way.
WARNING: Your unit must be properly grounded in conformity with national and local electrical codes. Do not connect the unit to overloaded power sources.
WARNING: Disconnect the unit from all power sources before cleaning, troubleshooting, or performing other maintenance on the product or its controls.
WARNING: This unit is not for storage of flammable materials.
WARNING: This unit is charged with hydrocarbon refrigerants. Only
qualified service personnel should service this unit.
WARNING: Unauthorized repair of your refrigerator will invalidate your warranty. Contact Technical Service at 1-800-438-4851 for additional information.
Thermo Scientific TSX Blood Bank Refrigerators 3
Unpacking

3 Unpacking

At the time of delivery, be sure to inspect the unit packaging for damage before signing for the shipment. If packaging damage is present, request immediate product inspection and file a claim with the carrier.
Note Packaging damage does not denote that unit damage exists.
If concealed damage is found (damage that is not apparent until the item has been unpacked), stop further unpacking and save all packing for carrier's inspection. Make a written request for inspection to delivering carrier. This must be done within 15 days after delivery. Then file a claim with the carrier.
Do not return goods to the manufacturer without written authorization
.
4 TSX Blood Bank Refrigerators Thermo Scientific
Packing List

4 Packing List

Inside the refrigerator cabinet is a bag containing:
USB drive with this and other user manuals, including translated versions
Certificate of Conformance
•Warranty Card
Safety Data Sheet for Glycerin
Quick start guide
Two control panel keys (silver)
Two cabinet door keys (gold)
Chart recorder pamphlet
Extra chart recorder paper
If the unit was ordered a unit with shelves, the bag will also include:
Small bag with shelving clips
If specified on the order, the bag may also include:
QC temperature graph and test log
Calibration information
Other items with your unit include:
Power cord
Anti-Tip Bracket Kit
Baskets, Shelves or Drawers
Thermal Bottles
Thermo Scientific TSX Blood Bank Refrigerators 5

General Recommendations

5 General
Recommendations

5.1 Temperature Monitoring

5.2 Intended Use

This section includes some general recommendations for your unit.
IMPORTANT NOTE We recommend the use of a redundant and independent temperature monitoring system so that the refrigerator can be monitored continuously for performance commensurate with the value of product stored.
The 4°C refrigerators described in this manual are intended for the storage of blood and blood products. All materials stored in these products must be in sealed blood bags. Product must not be stored in open containers. These products are intended for storing samples or inventory at operating temperatures between 3°C and 7
These products should only be used by an authorized and adequately trained personnel.
°C.
It is considered a medical device and has therefore been registered with the medical device regulatory body FDA and is considered a Class 2 medical device by the FDA. This product is classified as product code
864.9700 and is considered 510(K) exempt.
This unit is not intended for use in classified hazardous locations, nor to be used for the storage of flammable or corrosive inventory.
CAUTION: Storage of unsealed corrosive substances may cause the interior of the unit to corrode.
KSE, regulation number
6 TSX Blood Bank Refrigerators Thermo Scientific

Operating Standards

6 Operating
Standards
The refrigerators described in this manual are classified for use as stationary equipment in a Pollution Degree 2 and Overvoltage Category II environment.
These units are designed to operate under the following environmental conditions:
Indoor use
Altitude up to 2000 m (6512 feet)
Maximum relative humidity 60% for temperatures from 15 to 32°C (59 to 90°F).
For V models: Maximum relative humidity 75% for temperatures
from 10°C to 35°C (50 to 95°F).
Main supply voltage fluctuations not to exceed 10% of the nominal voltage.
The refrigerator must not be connected to a GFCI (Ground Fault Circuit Interrupter) protected outlet as it may subject to nuisance tripping.

6.1 Unit Specifications

The last character in the model number listed on the refrigerator identifies the electrical specifications for your unit.
The dataplate is located on the upper left side of the unit.
Table 2. Unit Specifications
Model
A 115V 4.5 A 60Hz/1 D 208-230V 2.5 A 50/60Hz/1 227 (500) V 208-230V 2.5 A 50/60Hz/1 227 (500)
Rated Voltage
Rated Current
Frequency / Phase
Power Module Plug
IEC C19
Drawers/ 450ml Blood Bags
7 / 385 (55 per drawer)
Weight Kg (lbs)
236 (521)
Exterior Dimensions (H x D x W)
199.7 x 94.0 x 71.1 cm (78.6 x 37.0 x 28.0 cm)
Thermo Scientific TSX Blood Bank Refrigerators 7
Installation

7 Installation

WARNING: Do not exceed the electrical rating printed on the dataplate
located on the upper left side of the unit.

7.1 Location

7.2 Leveling

7.3 Wiring

Install the unit on a level area free from vibration with a minimum of 6 inches of space on the sides and rear and 12 inches at the top. Do not position the equipment in direct sunlight or near heating diffusers, radiators, or other sources of heat.
Make sure that the floor is level. The unit must be level both front to back and side to side. If the unit is out of level, you may need to shim the corners or casters with thin sheets of metal. Be sure to set the brakes for units equipped with casters.
The wiring diagrams are attached on the back of the cabinet.
CAUTION: Connect the equipment to the correct power source. Incorrect voltage can result in severe damage to the equipment.
CAUTION: For personal safety and trouble-free operation, this unit must be properly grounded while in use. Failure to ground the equipment may cause personal injury or damage to the equipment. Always conform to the National Electrical Code and local codes. Do not connect the unit to overloaded power lines.
CAUTION: Do not position the unit in a way that impedes access to the disconnecting device or circuit breaker in the back of the unit.
CAUTION: Always connect the unit to a dedicated (separate) circuit. Each unit is equipped with a service cord and plug designed to connect it to a power outlet which delivers the correct voltage. Supply voltage must be within ±10% of the unit rated voltage. If cord becomes damaged, replace with a properly rated power supply cord.
Table 3. Power Cord Specifications
Model Power Cord Specifications
A 3-G 12 AWG, NEMA 5-15P, 15A/125V D 3-G 14 AWG, NEMA 6-15P, 15A/250V V 3-G 1.5 mm
2
, CEE 7/7, 16A/250V
8 TSX Blood Bank Refrigerators Thermo Scientific
Installation
CAUTION: Never cut the grounding prong from the service cord plug. If the prong is removed, the warranty is invalidated.
CAUTION: In an emergency, the power cord is a disconnected device.

7.4 Shelves

7.5 Drawers

Blood Bank Refrigerators have the option to be ordered with 4 wire shelves. Additional shelves are offered as available options.
Maximum shelf capacity is 45 kg (100 lbs).
For safety in shipping, the shelves are packaged and secured inside the cabinet. Insert the shelf support hangers (included with the manual inside the unit) into the built-in shelf supports (located on the inside walls of the cabinet interior) at the desired locations. Position the shelves on the flat supports (refer to
Figure 1. Shelf Support Hanger
Blood Bank Refrigerators come standard with 7 drawers.
Figure 1
).
Drawers are shipped in a factory installed position that promotes uniform spacing between the drawers. The drawer slides are adjustable to higher and lower positions in the cabinet. You can position these slides in the vertical slots which are spaced at one-inch intervals. The drawers must be removed in order to adjust the position of the drawer slides.
The maximum drawer capacity is 27 kg (60 lbs).

7.5.1 Removing the Drawers

Thermo Scientific TSX Blood Bank Refrigerators 9
To remove the drawers, complete the following steps (refer to Figure 2
1. Pull the drawer towards you until the slides are fully extended.
2. Lift the back of the drawer to disengage the mounting tabs from the slots on the slides.
Note The drawers fit snugly between the slides. Push the back to the drawer from underneath to remove the drawer.
3. Raise the back of the drawer almost to a vertical position and disengage the front mounting clips from the sides.
):
Installation
Figure 2. Drawer Removal
Label Description
1Tabs

7.5.2 Reinstalling the Drawers

7.5.3 Changing Drawer Position

2 Drawer slide 2Slot
To reinstall the drawers, complete the following steps (refer to
Figure 2
):
1. Pull both the slides of the drawer towards you until the slides are fully extended.
2. Position the drawer between the slides and with the back facing end raised at 45 degree angle, insert the mounting clips at the front into the slots on front of the slides.
3. Push the back of the drawer down between the slides and insert the basket tabs into the slots at the back.
Note The drawers fit snugly between the slides. Push on the back of the drawer from the inside to insert the drawer tabs completely into the slots. Make sure both the drawer tabs are aligned with the slots on the slides before pushing the drawer down between the slides.
Be careful when reinstalling the drawers to avoid possible pinching.
Drawer slides have a small wire safety clip at the front pilaster which prevent the slides from falling when the drawer is removed. To change the position of the drawer slides, complete the following steps:
1. Locate the safety clip.
2. Slip a small screwdriver under the bottom of the wire clip and pry the clip towards the inside of the refrigerator.
10 TSX Blood Bank Refrigerators Thermo Scientific
Installation
Figure 3. Clip Removal
3. Lift up the slide at the front. The slide is free to move from the front pilaster.
Figure 4. Front of slide Removal
4. The drawer slide must be removed from the rear pilaster at approximately a 45 degree angle towards the center of the cabinet.
Figure 5. Slide Rotation
Thermo Scientific TSX Blood Bank Refrigerators 11
Installation
5. Pull the slide towards the front of the cabinet.
6. Determine desired location for the slide and insert the slide into the rear pilaster at a 45 degree angle towards the center of the cabinet.
7. Once the slide is seated in the rear pilaster, rotate the slide until it is parallel with the unit wall and insert slide into front pilaster.
Figure 6. Slide Installation
CAUTION: Ensure the slide is level. If the slide is not seated in a level
orientation, the baskets will not function as intended. This could cause injury or product damage.
8. Replace the safety clip by hooking the clip under the front pilaster slide bracket. Then rotate the wire safety clip to position the other bend of the wire safety clip to position the other bend of the clip over the top of the same bracket.
Figure 7. Clip Replacement
CAUTION: Drawer slides do not require lubrication. Additional lubricant
could impede movement of the drawers when lubricant is cold.
12 TSX Blood Bank Refrigerators Thermo Scientific
Installation

7.6 Baskets

7.7 Anti-Tip Brackets

7.7.1 Installation Instructions

Blood Bank Refrigerators have the option to order 6 wire baskets. Baskets are shipped in a factory installed position that promotes uniform spacing
between the baskets. The basket slides are adjustable to higher and lower positions in the cabinet. You can position these slides in the vertical slots which are spaced at one-inch intervals. The baskets must be removed in order to adjust the position of the slides. The procedure for removing, reinstalling, and changing position of the baskets can be found in
7.5.3
.
Section 7.5.1, Section 7.5.2
, and
Section
The maximum basket capacity is 27 kg (60 lbs).
WARNING: The refrigerator must be secured by the anti-tip bracket supplied.
Unless properly installed, the refrigerator could tip when
drawers are loaded. Personal injury and damage to the equipment and contents may result from the refrigerator tipping.
This refrigerator has been designed to meet all recognized industry tip standards for all normal conditions when anti-tip bracket is installed and properly engaged.
Installation instructions are provided for wood and concrete floors. Any other type of construction may require special installation techniques as deemed necessary to provide adequate fastening of the anti-tip bracket to the floor. For installation on floors other than wood and concrete, please contact technical support.
The use of this bracket does not prevent the tipping of the refrigerator when not properly installed.
Figure 8. Materials Supplied
Label Description
1 Bracket 2 Bolts 3 Anchors 4 Instructions and Installation Template
Thermo Scientific TSX Blood Bank Refrigerators 13
Installation
Table 4. Tools required
Wood Floor Concrete Floor
Flashlight
Tape Measure
1/2
"(13 mm) Wrench
3/4
" (19 mm) Wrench
Drill Hammer Drill
15/64
" (6 mm) Drill Bit 1/2" (13 mm) Masonry Bit
1. Locating the Bracket
a. Determine where you want the centerline of the refrigerator to be. b. Place the included template on the floor lined up with the centerline
of the refrigerator and keep 6"-12" between the wall and the back of the unit.
c. On the floor, mark the location of Hole #1 & Hole #2.
2. Anti-Tip Bracket Installation
Wood Construction
a. Drill 15/64" (6 mm) pilot holes in locations marked in step 1. b. Place bracket on floor aligned with holes.
c. Use supplied lag bolts to attach bracket to floor.
Concrete Construction
a. Drill 1/2" (13 mm) holes in locations marked in step 1 with
masonry bit.
b. Slide lag screw anchors into holes to be flush with floor surface.
c. Place bracket on floor aligned with holes.
d. Use supplied lag bolts to attach bracket to floor.
3. Adjusting Bolt in Refrigerator
a. Locate 1/2" bolt attached to bottom of cabinet. b. Unscrew 1/2" bolt until there is 1/2" clearance between floor and
head of bolt as shown in Figure 9 below.
c. Tighten lock nut against bottom of unit.
14 TSX Blood Bank Refrigerators Thermo Scientific
Figure 9. Bolt Location
Label Description
Installation
1 Back of Unit 2 1/2" Bolt Location 3 Detailed View 4 Lock Nut 5 1/2" Bolt 6 Floor 7 1/2" Clearance
4. Refrigerator Positioning
a. Line up 1/2" bolt installed in step 3 with anti-tip bracket. b. Roll or slide refrigerator into position until bolt stops against
bracket.
c. Lock the casters.
5. Checking the Installation
Check to see if the anti-tip bracket is installed properly by shining light under cabinet and confirming bolt in cabinet is secured by bracket on floor.
Thermo Scientific TSX Blood Bank Refrigerators 15
Installation

7.8 Temperature Sensor Bottles

7.9 Door Operation

The temperature displayed on the control panel and chart recorder are measured by probes inserted in the sensor bottles inside the cabinet. The sensor probes must be inserted in the glycol bottle prior to operation.
There are two sensor bottles at different heights. The upper bottle is positioned on the left wall and the lower bottle is positioned on the back wall.
To install the sensor(s) into the bottles, complete the following steps:
1. Remove the bottle from the bracket and verify that each bottle is full of liquid.
Note To better access the bottom bottle, the bottom two drawers may need to be removed following the instructions in Section 7.5.1.
2. Remove the solid cap from the bottle and save for the next time the unit needs to be moved.
3. Remove the taped lid and temperature sensor(s) from the wall.
4. Insert the temperature probe(s) into the bottle.
5. Secure the cap to the bottle and return the bottle to the bracket.
The doors on all units are designed to stay open if opened 90 degrees. The door spring tension cannot be adjusted.
If the self-closing door does not work properly, make sure the unit is level.

7.9.1 Door Seal

7.10 Remote Alarm (Optional)

CAUTION: Door seal integrity is critical for refrigerator performance. A
loose fitting gasket allows moist air to be drawn into the cabinet, resulting in quicker frost buildup on the evaporator coil, longer running time, poor temperature maintenance, and increased operation cost.
All units have factory-installed local alarm contacts that can be used for remote alarm systems. The maximum distance between a refrigerator and a remote alarm depends on the wire gauge used. Refer
Table 5
below:
Table 5. Wire Gauges and Distance to Remote Alarm
Wire Gauge
20 530 265 18 840 420 16 1330 665 14 2120 1060 12 3370 1685
Total Wire Length (feet)
Distance to Alarm 1/2 Wire Length (feet)
16 TSX Blood Bank Refrigerators Thermo Scientific
Installation
Remote alarm terminals are located at the rear of the machine compartment. The three terminals are: COMMON, OPEN ON FAIL (Normally Closed), and CLOSE ON FAIL (Normally Open).

7.11 Final Checks

Figure 10. Remote Alarm Diagram
To install the remote alarm, make the following connections:
1. Connect the COMMON terminal on the cabinet switch to the COMMON wire on the alarm.
2a. To get an alarm when the switch contacts open, connect the OPEN ON
FAIL terminal on the cabinet to the OPEN ON FAIL wire on the alarm.
2b. To get an alarm when the switch contacts close, connect the CLOSE
ON FAIL terminal on the cabinet to the CLOSE ON FAIL wire on the alarm. The COMMON and CLOSE ON FAIL wires must be tied together in this application.
3. Plug the alarm system service cord into an electrical outlet.
The contacts will trip in the event of a power outage, high temperature alarm or low temperature alarm.
Before start up, be sure to complete the following steps:
1. Make sure that the unit is free of all wood or cardboard shipping materials, both inside and outside.
2. Check the positions of the shelves, drawers and baskets. If you want to adjust the positions, see instructions in Section 7.4, Section 7.5 and Section 7.6.
3. Verify that the unit is connected to a dedicated circuit.
Thermo Scientific TSX Blood Bank Refrigerators 17
Startup

8 Startup

8.1 Initial Startup

To start up the refrigerator, complete the following steps:
1. Verify that the sensor probe(s) have been installed in the bottle(s) as described in Section 7.8.
2. Ensure the double pole circuit breaker switch located next to the power inlet is in “ON” position (i.e. “I” position).
3. Insert the silver colored key in the switch and turn to the Power On position. The display will show the actual cabinet temperature and the compressor and evaporator should start within 10 minutes.
4. Allow the unit to reach operating temperature before loading it with any product. To stabilize the temperature profile, a 24-hour waiting period is recommended.
5. If you desire to enable the alarms, turn the three position key switch one turn further clockwise to the Alarm On position. To avoid nuisance alarms, wait until the unit has pulled down to the desired operating temperature (4°C default).
6. If you have a remote alarm, hook it up at this point (refer Section 7.10).
7. If desired, lock the cabinet door using the gold colored key. Place duplicate key copies in a safe place.

8.2 Product Loading and Unloading Guidelines

All controls should now be fully operational, the alarm active (if enabled), and all visual indicators active.
When loading your blood bank refrigerator, take care to observe the following guidelines:
Never load the blood bank beyond capacity. The number of 450ml bags that are recommended for your unit are 385.
Distribute the load as evenly as possible. Temperature uniformity depends on air circulation, which could be impeded if drawers are overfilled, particularly at the top of the cabinet.
For critical applications such as blood storage, be sure that the alarm systems are working and active before you load any product.
Stacking the load higher than 8 in (20.3 cm) from the bottom of the drawer may affect unit performance or impede operation of the drawers.
For initial product loading and after removing the drawers for cleaning, be sure to load the unit from the bottom drawer up to ensure the drawers are seated properly.
18 TSX Blood Bank Refrigerators Thermo Scientific

9 Operation



 
 


9.1 Control Panel

Operation
Figure 11. Refrigerator Control Panel
The control panel is located on the top right side of your blood bank refrigerator. You can use the three pushbuttons (#5, #8, and #9 in
Figure 11
) to change the
temperature display (#1) or to adjust temperature and alarm setpoints as given in
Section 9.3
. The thermometer display (#2) provides a quick visual indicator of
current cabinet temperature and alarm conditions.
1. Main temperature display - during normal operation, shows cabinet temperature in degrees Celsius, as measured by the primary sensor inside the cabinet. You can use the buttons to display other values such as setpoints, highest and lowest recorded temperatures. The number in the main display flashes when the value can be modified.
2. Thermometer - shows cabinet temperature and alarm conditions. There are 10 horizontal bars: 9 are displayed during normal operation, the tenth (top) bar indicates a warm alarm condition. The number of bars illuminated indicates approximate cabinet temperature. With the default settings, 5 to 7 bars illuminated indicate that the cabinet is at desired setpoint. For example, suppose that the cabinet temperature setpoint is 4°C and that the warm and cold alarm setpoints are 5.5°C and 2°C. Then the number of bars illuminated indicates cabinet temperature as follows:
Thermo Scientific TSX Blood Bank Refrigerators 19
Operation
Table 6. Thermometer Display on Control Panel (Setpoint
Bars Displayed
Bulb Only 2 (cold alarm) 6 Bars 4.1 1 Bar 2.35 7 Bars 4.45 2 Bars 2.7 8 Bars 4.8 3 Bars 3.05 9 Bars 5.15 4 Bars 3.4 10 Bars 5.5 (warm alarm) 5 Bars 3.75
Temperature (°C)
Bars Displayed
Temperature (°C)
4°C)
When cabinet temperature exceeds the warm alarm setpoint, the top bar of the thermometer flashes. When temperature is lower than the cold alarm setpoint, the bulb flashes. When you are in programming mode (described in
Table 8
) the
thermometer shows the setpoint value you are changing.
3. Power failure - illuminated when the main power supply is interrupted. In this case the audible alarm also sounds.
4. Service required - illuminated when the controller is in service programming mode or when simulated warm or cold alarm conditions are failing to occur during an alarm test.
5. Increase - pushbutton used to increase setpoint values in programming mode and for various display functions.
6. Door ajar - illuminates when the refrigerator door is open longer than 3 minutes (when the alarm is activated and the key switch is turned to the alarm position).
7. Battery low - illuminates when the backup battery is low (Refer Section
12.6).
8. Decrease - pushbutton used to decrease setpoint values in programming mode and for various display functions.
9. Scan - pushbutton used to change the main display and for various other functions.
10. Audible alarm - illuminates during warm and cold alarm conditions.
11. Light Switch (glass doors only) - pushbutton switch to toggle the refrigerator light on and off without opening the door. Light will always come on when the door is opened.
20 TSX Blood Bank Refrigerators Thermo Scientific
12. Key Switch - switch used to turn the power and alarms on and off for the unit.
13. Alarm On - Setpoint Security - when the key switch is in the alarm on position, the alarms are on and the setpoints cannot be modified.
14. Power On - when the key switch is in the on position, the unit power is on with no alarms active.
15. Off - when the key switch is in the off position, the unit is off.
For full descriptions of display, programming and service functions, refer to Tables 7, 8, and 9.

9.2 Display Functions

Table 7. Control Panel Display Functions
Function Meaning Sequence Display
Operation
Normal operation
Coldest logged temperature
Warmest logged temperature
Mute
Reset
Second Sensor
Default display while refrigerator is running.
Show coldest cabinet temperature since last startup or reset.
Show warmest cabinet temperature since last startup or reset.
Silence audible alarm for 6 minutes.
Return to default display after alarm condition, clears temperature log.
Displays second sensor temperature
Press
Press
Press
Press and simultaneously, hold for five seconds.
Press and simultaneously and hold for five seconds
Temperature display and control panel thermometer icon show cabinet temperature.
Display shows coldest logged temperature while button is pressed.
Display shows warmest logged temperature while button is pressed.
Display and thermometer show cabinet temperature, alarm icon continues to flash.
Excursion values are reset; temperature display shows cabinet temperature.
Display shows second probe temperature while bottons are held, returns to primary temperature display when buttons are released. ERR is displayed if no lower sensor probe is detected.
Thermo Scientific TSX Blood Bank Refrigerators 21
Operation

9.3 Programming Functions

You can enter the programming mode by pressing the Scan button ( ) and holding for 5 seconds. The display will then flash "Prg" to indicate that you have entered the programming mode. Use ( ) and ( ) buttons to modify the values and press the scan ( ) button to save the value and go to the next screen.
Table 8. Setpoint Programming Functions
Screen/Button Function Summary
1 Initial Screen Unit Set point
2Press Scan ()
3Press Scan ()
4Press Scan ()
Cold Alarm Temperature
Warm Alarm Temperature
Exit Program Mode
The temperature that the unit is set to run
The temperature that will actuate the cold alarm (if active)
The temperature that will actuate the warm alarm (if active)
Returns to normal operating screen
If the unit is left idle for a period of 30 seconds during program mode, the program mode shall be exited.
If at any point the scan button is pressed and held for a period of 10 seconds, the unit shall enter the service mode.

9.4 Service Parameters

You can enter the service mode by pressing the Scan button ( ) and holding for 10 seconds while in Programming mode (refer
Section 9.3
). The display will then flash "Ser" to indicate that you have entered the service mode followed by the checksum values flashing on the screen. The service icon ( )will also illuminate. Use up ( ) and down ( ) buttons to modify the values and press the scan ( ) button to save the value and go to the next screen.
Note If the alarms are enabled, all parameters are read only and cannot be edited.
CAUTION: Resetting any of the following parameter values could adversely affect the performance of your blood bank refrigerator. These settings very rarely need to be changed from normal 4°C operation. Be sure to understand your product requirements prior to making any adjustments to the service parameter values. Call technical support if you have any questions prior to making any adjustments to service parameter values.
22 TSX Blood Bank Refrigerators Thermo Scientific
Table 9. Service Parameters
Screen/Button Display Function Summary
Operation
04 Temp of unit
Varies CPU Checksum Shows the current checksum of the CPU board for validation
1 Initial Screen
Varies
0.0 Display Offset
2 Press Scan ( ) Cnt ucl
3 Press Scan ( ) Cnt lcl
4 Press Scan ( ) dEF Int Defrost Interval The time interval between defrost cycles in hours.
5 Press Scan ( ) dEF dur Defrost Duration
6 Press Scan ( ) dEF Hl
7 Press Scan ( ) Qua Quality Mode Normally OFF(00). Used for Production line testing.
Relay Board Checksum
Control Upper Hysteresis
Control Lower Hysteresis
Defrost T emp Out
Allows the user to see what temperature the unit is designed to run
Shows the current checksum of the Relay board for validation
Allows the user to create offset in the temperature displayed. This can be useful to adjust display to correlate with calibration
The degree increase in temperature from the set point that will trigger the unit to begin cooling to prevent the unit from getting too warm.
The degree decrease in temperature from the set point that will trigger the unit to stop cooling to prevent the unit from getting too cold.
The maximum amount of time the defrost cycle will run in minutes.
The maximum temperature that the defrost probe can reach before stopping the defrost cycle.
8 Press Scan ( ) bot BOT Mode
9 Press Scan ( ) Per
10 Press Scan ( ) Cnt Prb
11 Press Scan ( ) Anb Prb
12 Press Scan ( ) d lo Prb
13 Press Scan ( ) dEF Prb
14 Press Scan ( ) -
Perimeter Heater Duty Cycle
Control Probe Temperature
Ambient Probe Tempearature
Lower Display Probe
Defrost Probe Temperature
Enter Defrost cycle
Normally OFF(00). If ON(01), the unit is forced to run at coldest possible temperature for 24 hours. Defrost will still occur as scheduled.
Controls the perimeter heater ON time.
Displays control probe temperature (Read only).
Displays ambient probe temperature in deck (Read only).
Displays the lower bottle probe temperature (Read only).
Displays defrost probe temperature (Read only).
Pressing the Scan button after displaying the defrost probe temperature while not in an alarm mode will force the unit into a defrost cycle. To avoid defrost cycle, leave unit screen idle for 30 seconds.
Thermo Scientific TSX Blood Bank Refrigerators 23
Operation
If the unit is left idle for a period of 30 seconds during service mode, the unit will exit service mode and return to normal operation.
Note If an alarm occurs while in the service mode, it will not show until exiting this mode.
To reset values back to factory settings, perform the following procedure:
1. Turn keyswitch to "OFF" position.
2. Hold the up ( ) and down ( ) arrows while turning the keyswitch to
the "ON" position.
3. Continue to hold buttons for 10 seconds.
4. Release buttons and turn keyswitch to "OFF"position.
5. Turn keyswitch to "ON" position.
CAUTION: Performing this reset will overwrite any changes that were made after the receipt of this unit.

9.5 Temperature Settings

9.6 Safety Trip

9.7 Alarms The alarm system is designed to provide visual and audible warning signals

The unit has been set to maintain a temperature of 4°C. For most applications, you will not need to change the temperature setpoint. However, the setpoint can be changed if desired. To change the factory temperature settings, refer to the instructions in
CAUTION: The refrigerator described in this manual is designed for optimum performance at 4°C. It is advisable to call Technical Service before changing setpoints.
The Blood Bank refrigerators are equipped with a safety trip which prevents product from freezing in the event of loss of temperature control.
The safety trip is designed to turn off the cooling system when the temperature reaches 2°C, according to the sensor located in the bottom of the unit. The safety trip automatically allows the cooling system to be reactivated based on product temperature.
for both power failure and rise in temperature. The alarm is equipped with a battery backup.
Section 9.3
.
Default cold and warm alarm values are 2°C and 5.5°C. These values may be adjusted, following instructions in Section 9.3.
The alarm system is activated only when the key switch is turned to the Alarm On position. The audible warning signal sounds when there is a
24 TSX Blood Bank Refrigerators Thermo Scientific
Operation
power failure, temperature alarm condition, or when the door is ajar for more than 3 minutes.
The Mute function (pressing the () button) allows you to turn off the audio warning without turning off the visual indicators. The alarm will ring
back approximately after 6 minutes if the alarm is still active. During a warm alarm, a flashing of the uppermost bar of the thermometer,
the temperature display, and the speaker icon will occur at the same time. If there has been a warm alarm since the last alarm reset, but the temperature is not currently in an alarm state, the uppermost bar of the thermometer and the speaker icon will slowly alternate in flashing.
During a cold alarm, a flashing of the bulb of the thermometer, the temperature display, and the speaker icon will occur at the same time. If there has been a cold alarm since the last alarm reset, but the temperature is not currently in an alarm state, the bulb of the thermometer and the speaker icon will slowly alternate in flashing.
During a power failure, the power failure icon will illuminate, the thermometer will display without the bulb, and the temperature will flash approximately every 6 seconds. If there has been a power failure since the last alarm reset, the power failure icon and the speaker icon will slowly alternate in flashing.
During a low battery condition, the low battery icon will illuminate.
The alarms can be reset by pressing and holding the and simultaneously.
Thermo Scientific TSX Blood Bank Refrigerators 25
Chart Recorder

10 Chart Recorder

10.1 Set Up and Operation

Panel-mounted six inch recorders are standard and factory-installed on all models. Recorder operation begins when the system is powered on.
To prepare the recorder to function properly, complete the following steps:
1. Open the recorder door to access the recorder.
2. Connect the nine volt DC battery located at the recorder’s upper right corner. This battery provides back-up power.
3. Install clean chart paper (Refer to Section 10.2 below).
4. Close the recorder door.
Note The recorder may not respond until the system reaches temperatures within the recorder’s range.
Figure 12.
Chart Recorder
Label Description
1 Pressure Sensitive Chart Buttons 2 Imprinting Stylus 3 Reference Mark 4 Hub-Nut and Retaining Wire 5Chart
26 TSX Blood Bank Refrigerators Thermo Scientific
Figure 13. Chart Buttons
Chart Recorder

10.2 Changing Chart Paper

To change the chart paper, complete the following steps:
1. Locate the pressure sensitive buttons at the front, upper left of the recorder panel.
2. Press and hold the change chart button (#3 in Figure 13) for one second. The pen will move off the scale.
3. Unscrew the center nut, remove the old chart paper, and install new chart paper. Carefully align the day and time with the reference mark on the recorder panel (a small groove on the left side of the panel, as shown in Figure 12).
4. Replace the center nut and hand tighten. Press the change chart button again (#3) to resume temperature recording.
Note Be careful not to pull up on the chart recorder arm while changing paper. Pulling on the arm may damage the recorder.
CAUTION: Do not use sharp or pointed objects to depress the chart buttons. This may cause permanent damage to the recorder.

10.3 Power Supply The recorder normally uses AC power when the system is operating. If AC

power fails, the LED indicator on the recorder flashes to alert you to a power failure. The recorder continues sensing cabinet temperature and the chart continues turning for approximately 24 hours with back-up power provided by the nine-volt battery.
The LED indicator glows continuously when main power is functioning and the battery is charged.
When the battery is low, the LED flashes to indicate that the battery needs to be changed.
Thermo Scientific TSX Blood Bank Refrigerators 27
Chart Recorder

10.4 Calibration and Adjustments

This recorder has been accurately calibrated at the factory and retains calibration even during power interruptions. If required, however, adjustments can be made as follows:
1. Run the unit continuously at the control setpoint temperature. Continue steady operation for at least two hours to provide adequate time for recorder response.
2. Measure cabinet center temperature with a calibrated temperature monitor.
3. Compare the recorder temperature to the solution temperature. If necessary, adjust the recorder by pressing the left and right chart buttons (#1 and #2 from Figure 13) for five seconds.
Note The stylus does not begin to move until the button is held for five seconds.
28 TSX Blood Bank Refrigerators Thermo Scientific

Temperature Transmitter (Optional)

11 Temperature
Transmitter
(Optional)
A powered temperature transmitter is an option for all units. This temperature transmitter allows for remote temperature monitoring of your system.
To connect your remote temperature monitoring system to the temperature transmitter, refer to
Figure 14
below.
Figure 14. Temperature Transmitter
Label Description
1 Rear of Unit 2 16.5V Supply Negative 3 Controller and/or Indicator 4 Customer Provided System 5 Temperature T r ansmitter Negative 6 Termin a l St rip
Thermo Scientific TSX Blood Bank Refrigerators 29
Maintenance

12 Maintenance

WARNING: Disconnect equipment from main power before attempting
any maintenance to equipment or its controls unless stated otherwise.

12.1 Cleaning the Cabinet Interior

12.2 Cleaning the
Condenser Filter
To clean the cabinet interior, remove the shelves, drawers and baskets following the instructions in water and a mild detergent for cleaning. Rinse the drawers and wipe them dry with a soft cloth.
Clean the condenser filters every three months. There is one condenser filter located in the back cage of the unit that can be accessed without removing the back cage or disconnecting the power.
To clean the filter, complete the following steps:
1. Remove the filter by pulling upwards through the slot in the back cage.
2. Shake the filter to remove loose dust.
3. Rinse the filter in clean water.
4. Shake the excess water from the filter and let it dry.
5. Reinstall filter by pushing downwards through the slot in the back cage.
CAUTION: Do not pull the filter downwards from the bottom. The condenser has sharp surfaces.
Section 7.4,
Section 7.5
and
Section 7.6
. Use a solution of
12.3 Cleaning the
Condenser
CAUTION: Condensers should be cleaned at least every six months; more
often if the laboratory area is dusty. In heavy traffic areas, condensers load with dirt more quickly. Failure to keep the condenser clean can result in equipment warm-up or erratic temperatures.
CAUTION: Never clean around the condensers with your fingers. Some surfaces are sharp.
The condenser is located in the top rear of the machine compartment. To clean the condenser, complete the following steps:
1. Disconnect the power.
2. Remove the filter.
30 TSX Blood Bank Refrigerators Thermo Scientific
3. Vacuum the condenser and clean up any loose dust.
4. Replace the filter.
5. Reconnect power.
Maintenance

12.4 Automatic Defrost

12.5 Gasket
Maintenance
The defrosting process on all models is primarily accomplished by air circulated during off-cycle periods. This heat-free process ensures that the temperature is not affected by the defrost cycle. The default defrost cycle runs once per hour and terminates once a preset evaporator temperature or timer criteria is reached.
Defrost water is collected in a pan in the rear of the unit and evaporated using system heat. No maintenance is required.
Periodically check the gaskets around the door for punctures or tears. Leaks are indicated by condensation or frost which form at the point of gasket failure. Make sure that the cabinet is level (refer to
Keep the door gaskets clean and frost free by wiping gently with a soft cloth.
To check the door seal, complete the following steps:
1. Open the door.
2. Insert a strip of paper (a couple of inches wide) between the door gasket and the cabinet flange and close the door.
3. Slowly pull the paper strip from the outside. You should feel some resistance.
Section 7.2
for leveling information).
4. Repeat this test at 4-inch intervals around the door. If the door does not seal properly, replace the gasket.

12.6 Alarm Battery Maintenance

Thermo Scientific TSX Blood Bank Refrigerators 31
Have a certified technician replace the alarm battery every twelve months at most or when the alarm is active. The part number for a replacement battery is 322533H01.
Troubleshooting

13 Troubleshooting

WARNING: Troubleshooting procedures involve working with high
voltages which can cause injury or death. Troubleshooting should only be performed by trained personnel.
This section is a guide for troubleshooting equipment problems.
Table 10. Troubleshooting Procedures
Problem Cause Solution
Check that the cord is securely plugged in.
Plug another appliance into the outlet to see if it is live.
Unit does not operate or Power Failure Indicator is on
Temperature fluctuates
Low battery icon is lit
Power supply
Temperature Control
Condenser clogged
Solution bottle
Other causes
12V backup battery needs to be replaced.
Check that the double pole circuit breaker located next to the power inlet is in "ON" position (i.e. "I" position). Try cycling to OFF position (i.e. "O" position) & then bring to ON ("I") position.
Test the voltage and verify that it is correct for your unit (refer to
If the outlet is dead, check the circuit breaker or fuses.
Make sure that the control is set correctly. Refer to
Make sure the condenser and filter are clean. Refer to
Section 9.3.
Section 12.2 and
Table 2).
Section 12.3.
Make sure the solution bottles for the temperature sensors are full. The solution is a 50/50 mixture of glycerin and distilled water.
If the temperature control is set correctly, the condenser is clean, but temperature continues to fluctuate, call an authorized service representative.
Replace the battery. It is located on the top right hand side of the cabinet. Call an authorized service representative.
32 TSX Blood Bank Refrigerators Thermo Scientific
Table 10. Troubleshooting Procedures
Problem Cause Solution
Door is open Make sure the door is completely closed.
Troubleshooting
Unit warms up
"E01" on display
"E02" on display
"E03" on display
"E05" on display
"Err" on display
Door seal Check the door seal, following instructions in Warm product
recently loaded in unit
Power supply
Setpoints need to be adjusted
Faulty Safety Trip Replace Sensor. Communication
error Control Probe
Failure Defrost Probe
Failure Ambient Probe
Failure Upper Bottle Probe
Failure
Allow ample time to recover from loading warm product.
Check for proper voltage to the unit. If there is no voltage to the unit, call an electrician.
To adjust the setpoint, refer to
Call customer service.
Check for loose probe connector. Replace control probe.
Check for loose probe connector. Replace defrost probe.
Check for loose probe connector. Replace ambient probe.
Check for loose probe connector. Replace upper bottle probe.
Section 9.3.
Section 12.5.
“---” on display
Lost Communication
Call customer service.
Thermo Scientific TSX Blood Bank Refrigerators 33
End of Life Care

14 End of Life Care

Be sure to follow local regulations when disposing of an old unit. Some suggestions are listed in the following:
1. Remove items and defrost unit. Be sure to clean up any biological safety hazards.
2. Remove the cabinet door to help prevent entrapment inside of a unit.
3. Have a certified technician remove the refrigerant and compressor, then drain the compressor and oil from the system. Dispose of components following local regulations.
34 TSX Blood Bank Refrigerators Thermo Scientific
Warranty

15 Warranty

Domestic Warranty • 24 Months Full Warranty Parts and Labor
International Warranty • 24 Months Full Warranty Parts Only
During the first twenty four (24) months from shipment, Thermo Fisher Scientific Inc, through its authorized Dealer or service organizations, will at its option and expense repair or replace any part found to be non-conforming in material or workmanship. Thermo Fisher Scientific Inc reserves the right to use replacement parts, which are used or reconditioned. Replacement or repaired parts will be warranted for only the unexpired portion of the original warranty.
This warranty does not apply to damage caused by (i) accident, misuse, fire, flood or acts of God; (ii) failure to properly install, operate or maintain the products in accordance with the printed instructions provided, (iii) causes external to the products such as, but not limited to, power failure or electrical power surges, (iv) improper storage and handling of the products, (v) use of the products in combination with equipment or software not supplied by Thermo Fisher; or (vi) installation, maintenance, repair, service, relocation or alteration of the products by any person other than Thermo Fisher or its authorized representative. To obtain proper warranty service, you must contact the nearest authorized service center or Dealer. ThermoFisher Scientific, Inc’s own shipping records showing date of shipment shall be conclusive in establishing the warranty period. At Thermo Fisher’s option, all non-conforming parts must be returned to Thermo Fisher postage paid and replacement parts are shipped FOB Thermo Fisher’s location.
Limitation of Liability
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL, OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. THERMO FISHER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT.
THERMO FISHER SHALL NOT BE LIABLE FOR ANY INDIRECT OR CONSEQUENTIAL DAMAGES INCLUDING, WITHOUT LIMITATION, DAMAGES TO LOST PROFITS OR LOSS OF PRODUCTS.
Thermo Scientific TSX Blood Bank Refrigerators 35
WEEE Compliance
Great Britain
Deutschland
Italia
France
WEEE Compliance. This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol. Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on our compliance with these Directives, the recyclers in your country, and information on Thermo Scientific products which may assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/
WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2012/19/EU erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet. Thermo Fisher Scientific hat Vereinbarungen getroffen mit Verwertungs-/Entsorgungsanlagen in allen EU-Mitgliederstaaten und dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden. Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo Scientific, dieVerwerter und Hinweise die Ihnen nützlich sein können, die Thermo Fisher Scientific Produkte zu identizfizieren, die unter diese RoHS. Anweisungfallen, finden Sie unter www.thermo.com/
Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’ Unione Europea 2012/19/EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici (WEEE). È marcato col seguente simbolo.Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli Stati Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste medesime. Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito www.thermo.com/
Conformité WEEE. Ce produit doit être conforme à la directive euro-péenne (2012/19/EU) des Déchets d’Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant. Thermo Fisher Scientific s’est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devraitêtre collecté ou recyclé par celles-ci. Davantage d’informations sur laconformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider le détection des substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/
IF YOU NEED ASSIST ANCE:
Thermo Fisher Scientific products are backed by a global technical support team ready to support your applications. We offer cold storage accessories, including remote alarms, temperature recorders, and validation services. Visit www.thermofisher.com or call:
Phone Numbers Sales Service
North America
India
China
+1 866 984 3766
1800 22 8374, +91 22 6716 2200 +91 22 6716 2200
+800 810 5118, +400 650 5118 +8621 68654588
Japan +81 3 5826 1616
Australia +61 39757 4300
Austria
+43 1 801 40 0 +43 1 801 40 0
Belgium +32 53 73 42 41
France +33 2 2803 2180
Germany
0800 1 536 376, +49 6184 90 6000 0800 1 536 376
Italy +32 02 95059 552 +39 02 95059 552, 432 254 375
Netherlands +31 76 579 55 55 +31 76 571 4440
Nordic/Baltic
Russia/CIS
+358 9 329 10200 +358 9 329 100
+7 812 703 4215 +7 812 703 4215
Spain/Portugal +34 93 223 09 18 +34 93 223 09 18
Switzerland +41 44 454 12 22 +41 44 454 12 12
UK/Ireland +44 870 609 9203 +44 870 609 9203
New Zealand +64 9 980 6700 +64 9 980 6700
Other Asian countries +852 2885 4613 +852 2885 4613
Countries not listed +49 6184 90 6000 +49 6184 90 6000
(800) 438-4851
+81 3 3816 3355
1 300 735 292
+32 2 482 30 30
+33 2 2803 2180
Thermo Fisher Scientific Inc.
327930H01 Rev. A
275 Aiken Road Asheville, NC 28804 United States
www.thermofisher.com
Loading...