Thermo Scientifi? Installation And Operation Manual
Specifications and Main Features
Frequently Asked Questions
User Manual
Blood Bank Refrigerators
TSX Series
Installation and Operation
327930H01 Rev. AAugust 2016
Visit us online to register your warranty
www.thermofisher.com/labwarranty
IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result
in damage to the unit, injury to operating personnel, and poor equipment performance.
CAUTION All internal adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for informational purposes only. The contents and the product it describes are
subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with
respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental,
arising from or related to the use of this manual.
For your future reference and when contacting the factory, please have the following information readily
available. It can be found on the dataplate attached to your unit.
Model Number:
Serial Number:
The following information, if available, is helpful for contacting the factory.
The table below shows the units covered in this operation and installation
manual by model number. Refer
Section 6.1
for details on the voltage
specifications for the various models.
Table 1. Applicable Models
UnitModel (*)
Thermo Scientific - TSX2304B* A/D/V
Thermo Scientific TSX Blood Bank Refrigerators1
Safety Precautions
2 Safety Precautions
In this manual, the following symbols and conventions are used:
This symbol when used alone indicates important operating instructions which
reduce the risk of injury or poor performance of the unit.
WARNING: This symbol indicates potentially hazardous situations which,
if not avoided, could result in serious injury or death.
WARNING: This symbol indicates situations where dangerous voltages
exist and potential for electrical shock is present.
WARNING: This symbol indicates potentially hazardous situations, which
if not avoided could result in fire.
CAUTION: This symbol, in the context of a CAUTION, indicates a
potentially hazardous situation which if not avoided could result in minor to
moderate injury or damage to the equipment.
CAUTION: This indicates a situation which may result in property
damage.
This symbol indicates surfaces which may become hot during use and may cause
a burn if touched with unprotected body parts.
Before installing, using or maintaining this product, please be sure to read the
manual and product warning labels carefully. Failure to follow these instructions
may cause the product to malfunction, which could result in injury or damage.
This symbol indicates possible pinch points which may cause personal injury.
This symbol indicates the need to use gloves during indicated procedures. If
performing decontamination procedures, use chemically resistant gloves.
2 TSX Blood Bank RefrigeratorsThermo Scientific
Safety Precautions
Below are important safety precautions that apply to this product:
Use this product only in the way described in the product literature and in this
manual. Before using it, verify that this product is suitable for its intended use. If
the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
Do not modify system components, especially the controller. Use OEM exact
replacement equipment or parts. Before use, confirm that the product has not
been altered in any way.
WARNING: Your unit must be properly grounded in conformity with
national and local electrical codes. Do not connect the unit to overloaded
power sources.
WARNING: Disconnect the unit from all power sources before cleaning,
troubleshooting, or performing other maintenance on the product or its
controls.
WARNING: This unit is not for storage of flammable materials.
WARNING: This unit is charged with hydrocarbon refrigerants. Only
qualified service personnel should service this unit.
WARNING: Unauthorized repair of your refrigerator will invalidate your
warranty. Contact Technical Service at 1-800-438-4851 for additional
information.
Thermo Scientific TSX Blood Bank Refrigerators3
Unpacking
3 Unpacking
At the time of delivery, be sure to inspect the unit packaging for damage before
signing for the shipment. If packaging damage is present, request immediate
product inspection and file a claim with the carrier.
Note Packaging damage does not denote that unit damage exists.
If concealed damage is found (damage that is not apparent until the item has
been unpacked), stop further unpacking and save all packing for carrier's
inspection. Make a written request for inspection to delivering carrier. This must
be done within 15 days after delivery. Then file a claim with the carrier.
Do not return goods to the manufacturer without written authorization
.
4 TSX Blood Bank RefrigeratorsThermo Scientific
Packing List
4 Packing List
Inside the refrigerator cabinet is a bag containing:
•USB drive with this and other user manuals, including translated
versions
•Certificate of Conformance
•Warranty Card
•Safety Data Sheet for Glycerin
•Quick start guide
•Two control panel keys (silver)
•Two cabinet door keys (gold)
•Chart recorder pamphlet
•Extra chart recorder paper
If the unit was ordered a unit with shelves, the bag will also include:
•Small bag with shelving clips
If specified on the order, the bag may also include:
•QC temperature graph and test log
•Calibration information
Other items with your unit include:
•Power cord
•Anti-Tip Bracket Kit
•Baskets, Shelves or Drawers
•Thermal Bottles
Thermo Scientific TSX Blood Bank Refrigerators5
General Recommendations
5 General
Recommendations
5.1 Temperature
Monitoring
5.2 Intended Use
This section includes some general recommendations for your unit.
IMPORTANT NOTE We recommend the use of a redundant and
independent temperature monitoring system so that the refrigerator can be
monitored continuously for performance commensurate with the value of
product stored.
The 4°C refrigerators described in this manual are intended for the storage of
blood and blood products. All materials stored in these products must be in
sealed blood bags. Product must not be stored in open containers. These
products are intended for storing samples or inventory at operating temperatures
between 3°C and 7
These products should only be used by an authorized and adequately trained
personnel.
°C.
It is considered a medical device and has therefore been registered with the
medical device regulatory body FDA and is considered a Class 2 medical device
by the FDA. This product is classified as product code
864.9700 and is considered 510(K) exempt.
This unit is not intended for use in classified hazardous locations, nor to be used
for the storage of flammable or corrosive inventory.
CAUTION: Storage of unsealed corrosive substances may cause the interior
of the unit to corrode.
KSE, regulation number
6 TSX Blood Bank RefrigeratorsThermo Scientific
Operating Standards
6 Operating
Standards
The refrigerators described in this manual are classified for use as stationary
equipment in a Pollution Degree 2 and Overvoltage Category II environment.
These units are designed to operate under the following environmental
conditions:
•Indoor use
•Altitude up to 2000 m (6512 feet)
•Maximum relative humidity 60% for temperatures from 15 to 32°C
(59 to 90°F).
•For V models: Maximum relative humidity 75% for temperatures
from 10°C to 35°C (50 to 95°F).
•Main supply voltage fluctuations not to exceed 10% of the nominal
voltage.
•The refrigerator must not be connected to a GFCI (Ground Fault
Circuit Interrupter) protected outlet as it may subject to nuisance
tripping.
6.1 Unit Specifications
The last character in the model number listed on the refrigerator identifies the
electrical specifications for your unit.
The dataplate is located on the upper left side of the unit.
WARNING: Do not exceed the electrical rating printed on the dataplate
located on the upper left side of the unit.
7.1 Location
7.2 Leveling
7.3 Wiring
Install the unit on a level area free from vibration with a minimum of 6 inches of
space on the sides and rear and 12 inches at the top. Do not position the
equipment in direct sunlight or near heating diffusers, radiators, or other sources
of heat.
Make sure that the floor is level. The unit must be level both front to back and
side to side. If the unit is out of level, you may need to shim the corners or casters
with thin sheets of metal. Be sure to set the brakes for units equipped with
casters.
The wiring diagrams are attached on the back of the cabinet.
CAUTION: Connect the equipment to the correct power source. Incorrect
voltage can result in severe damage to the equipment.
CAUTION: For personal safety and trouble-free operation, this unit must
be properly grounded while in use. Failure to ground the equipment may
cause personal injury or damage to the equipment. Always conform to the
National Electrical Code and local codes. Do not connect the unit to
overloaded power lines.
CAUTION: Do not position the unit in a way that impedes access to the
disconnecting device or circuit breaker in the back of the unit.
CAUTION: Always connect the unit to a dedicated (separate) circuit. Each
unit is equipped with a service cord and plug designed to connect it to a
power outlet which delivers the correct voltage. Supply voltage must be
within ±10% of the unit rated voltage. If cord becomes damaged, replace
with a properly rated power supply cord.
Table 3. Power Cord Specifications
ModelPower Cord Specifications
A3-G 12 AWG, NEMA 5-15P, 15A/125V
D3-G 14 AWG, NEMA 6-15P, 15A/250V
V3-G 1.5 mm
2
, CEE 7/7, 16A/250V
8 TSX Blood Bank RefrigeratorsThermo Scientific
Installation
CAUTION: Never cut the grounding prong from the service cord plug. If
the prong is removed, the warranty is invalidated.
CAUTION: In an emergency, the power cord is a disconnected device.
7.4 Shelves
7.5 Drawers
Blood Bank Refrigerators have the option to be ordered with 4 wire shelves.
Additional shelves are offered as available options.
Maximum shelf capacity is 45 kg (100 lbs).
For safety in shipping, the shelves are packaged and secured inside the cabinet.
Insert the shelf support hangers (included with the manual inside the unit) into
the built-in shelf supports (located on the inside walls of the cabinet interior) at
the desired locations. Position the shelves on the flat supports (refer to
Figure 1. Shelf Support Hanger
Blood Bank Refrigerators come standard with 7 drawers.
Figure 1
).
Drawers are shipped in a factory installed position that promotes uniform
spacing between the drawers. The drawer slides are adjustable to higher and
lower positions in the cabinet. You can position these slides in the vertical slots
which are spaced at one-inch intervals. The drawers must be removed in order to
adjust the position of the drawer slides.
The maximum drawer capacity is 27 kg (60 lbs).
7.5.1 Removing the Drawers
Thermo Scientific TSX Blood Bank Refrigerators9
To remove the drawers, complete the following steps (refer to Figure 2
1. Pull the drawer towards you until the slides are fully extended.
2. Lift the back of the drawer to disengage the mounting tabs from the
slots on the slides.
Note The drawers fit snugly between the slides. Push the back to the drawer
from underneath to remove the drawer.
3. Raise the back of the drawer almost to a vertical position and disengage
the front mounting clips from the sides.
):