TriPlus is a trademark and TSQ Quantum, TSQ Quantum XLS, TSQ Quantum GC, and TRACE GC Ultra
are registered trademarks of Thermo Fisher Scientific Inc. in the United States.
The following are registered trademarks in the United States and other countries: Swagelok is a registered
trademark of the Crawford Fitting Company. Dranetz is a registered trademark of Dranetz Technologies, Inc.
Styrofoam is a registered trademark of Dow Chemical Company.
All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A, March 2010.
Software version: Xcalibur 2.1, TSQ 2.3.
For Research Use Only. Not for use in diagnostic procedures.
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable domestic and international regulations. When the system is delivered to you, it meets all pertinent
electromagnetic compatibility (EMC) and safety standards as described in the next section or sections by product name.
Changes that you make to your system may void compliance with one or more of these EMC and safety standards.
Changes to your system include replacing a part or adding components, options, or peripherals not specifically
authorized and qualified by Thermo Fisher Scientific. To ensure continued compliance with EMC and safety standards,
replacement parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific
or one of its authorized representatives.
TSQ Quantum XLS
USA and European Union Regulatory Compliance requirements concerning product safety and electromagnetic
compliance of this product are clearly stated in the hard copy of the Declaration of Conformity included with
this system.
TSQ Quantum GC
EMC Directive 2004/108/EC
EMC compliance has been evaluated by TUV Rheinland of North America Inc.
EN 61326-11998, 2001, 2003EN 61000-4-61996, A1; 2003
EN 61000-4-22001EN 61000-4-111994, A1; 2001
CISPR 111998; 1999; 2002
FCC Class A, CFR 47 Part 15: 2007
Low Voltage Safety Compliance
Compliance with safety issues is declared under Thermo Fisher Scientific sole responsibility.
This device complies with Low Voltage Directive 73/23/EEC and harmonized standard EN 61010-1:2001.
Changes that you make to your system may void compliance with one or more of these EMC and safety standards.
Changes to your system include replacing a part or adding components, options, or peripherals not specifically
authorized and qualified by Thermo Fisher Scientific. To ensure continued compliance with EMC and safety standards,
replacement parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific
or one of its authorized representatives.
FCC Compliance Statement
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO
THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTErfERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTErfERENCE RECEIVED,
INCLUDING INTErfERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained inside
this manual pertaining to the safe use and operation of this product before using the device.
Notice on Lifting and Handling of
Thermo Fisher Scientific San Jose Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific
San Jose instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/or
bulky for one person alone to handle safely.
Notice on the Proper Use of
Thermo Fisher Scientific San Jose Instruments
In compliance with international regulations: Use of this instrument in a manner not specified by Thermo Fisher
Scientific San Jose could impair any protection provided by the instrument.
Notice on the Susceptibility
to Electromagnetic Transmissions
Your instrument is designed to work in a controlled electromagnetic environment. Do not use radio frequency
transmitters, such as mobile phones, in close proximity to the instrument.
IMPORTANT PREINSTALLATION INFORMATION... PLEASE READ
TSQ Quantum XLS and TSQ Quantum GC Installation
Request Form
Dear User:
Read the TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide, and then print and
complete the following installation request form. After all items on the form are fulfilled, sign and date the form.
Then, mail or fax this form to your local Thermo Fisher Scientific sales/service office. The address and fax
number for your local office are located on the following pages.
1. All laboratory remodeling has been completed.
2. TSQ Quantum XLS or TSQ Quantum GC
system is on site.
3. Principal operator will be available during the
installation / certification period.
4. Doorways, hallways, and so on are a minimum
width of 94 cm (37 in.).
5. Available floor area is sufficient and flooring will
support the load.
6. Sufficient bench space is available for all of the
equipment. List the following:
Width: ____________________________
Depth: ____________________________
Height: ____________________________
7. Workbench can support the load of the system
[191 kg (420 lbs)] and is free from vibration.
8. Lighting is adequate.
9. Main power is installed and is in compliance with
local electrical codes.
10. Power for test and cleaning equipment is
installed.
12. Voltage of power outlet has been measured.
Note measured voltage: ______________
13. Power is free from fluctuations due to slow
changes in the average voltage or changes due
to surges, sags, or transients.
14. Air conditioning is adequate for temperature,
humidity, and particulate matter control. The
laboratory can be maintained at a constant
temperature, between 15 and 27 °C (59 and
81 °F).
15. Relative humidity is between 40% and 80% with
no condensation.
16. System work area is free from magnetic
disruption and electrostatic discharge.
17. All required gases (argon, helium, and chemical
ionization reagent gas) are on site and the
appropriate gas regulators and fittings to the
regulators are available.
List gases and their purity: ______________
18. There is a suitable exhaust system.
19. One voice telephone line is installed near the
system.
20. All relevant safety regulations are complied with.
Have any special acceptance specifications been agreed to in the contract?Yes No
If YES, attach full details of specifications.
Is there any additional equipment that needs to be interfaced to the system?Yes No
If YES, attach full details of additional equipment
Note: We reserve the right to invoice against the engineer’s time if the installation requirements are not met on the
date of the installation.
.
Print your name, company name, and company address clearly below:
Name _______________________________________________________
Company _____________________________________________________ Telephone ___________________
Notes: The country code is enclosed in square brackets [ ]. The city code or area code is enclosed in parenthesis ( ). For countries other than the U.S.A., when you
are dialing from within the specified country, dial the 0 of the city code. For countries other than Italy, when you are dialing from outside the country, do not dial the 0
of the city code.
Offices for Thermo Fisher Scientific San Jose Products - Continued
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE)
Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union
(EU) Member State, and these companies should dispose of or recycle this product. See www.thermo.com/
WEEERoHS for further information on Thermo Fisher Scientific’s compliance with these Directives and the
recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen.
Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-Mitgliedsstaaten
getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Mehr Information
über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, über die Verwerter, und weitere Hinweise,
die nützlich sind, um die Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter
www.thermo.com/WEEERoHS
.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et
Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de
l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la
conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les
produits Thermo Fisher Scientific qui peuvent aider la détection des substances sujettes à la directive RoHS sont
disponibles sur www.thermo.com/WEEERoHS
The Thermo Scientific TSQ Quantum XLS™ and TSQ Quantum GC™ systems are member
of the TSQ™ Quantum family of mass spectrometers.
This Preinstallation Requirements Guide provides information to assist in planning and
preparing your lab site for the system prior to delivery and installation. Read each section
carefully to be sure that your laboratory is ready for the system installation.
Contents
• Related Documentation
• Safety and Special Notices
• Contacting Us
To suggest changes to documentation or to Help
• Complete a brief survey about this document by clicking the link below.
Thank you in advance for your help.
Related Documentation
In addition to this guide, Thermo Fisher Scientific provides the following PDF document and
Help for the TSQ Quantum XLS and TSQ Quantum GC systems:
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION
notice is accompanied by an appropriate CAUTION symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Highlights information of general interest.
Contacting Us
Tip Highlights helpful information that can make a task easier.
There are several ways to contact Thermo Fisher Scientific for the information you need.
To contact Technical Support
Phone800-532-4752
Fax561-688-8736
E-mailus.techsupport.analyze@thermofisher.com
Knowledge basewww.thermokb.com
Find software updates and utilities to download at mssupport.thermo.com.
To contact Customer Service for ordering information