Ion Trap series refers to ITQ and PolarisQ instruments. DSQ series refers to all DSQ and DSQ II instruments.
Information about the TRACE GC and FOCUS GC instruments is included in this document.
NSTRUMENT USAGE: Thermo Scientific systems operate safely and reliably under carefully controlled
I
environmental conditions. If the equipment is used in a manner not specified by the manufacturer, the
protections provided by the equipment may be impaired. If you maintain a system outside the specifications
listed in this guide, failures of many types may occur. The repair of such failures is specifically excluded from
the standard Warranty and service contract coverage.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Rev A, April 29, 2008, Rev B, November 30, 2009.
For Research Use Only. Not regulated for medical or veterinary diagnostic use by U.S. Federal Drug
Administration or other competent authorities.
Reader’s Survey
Ion Trap Series Hardware Manual, 120425-00HW, Revision B
Free gift for your returned Reader’s Survey. Please help us improve the quality of our documentation by completing
and returning this survey. Circle one number for each statement:
fold
fold
Strongly
Agree
AgreeNeutralDisagree
Strongly
Disagree
The manual is well organized.12345
The manual is clearly written.12345
The manual contains all the information I need.12345
The instructions are easy to follow.12345
The instructions are complete.12345
The technical information is easy to
understand.
12345
Examples of operation are clear and useful.12345
The figures are helpful.12345
I was able to operate the system using this
manual.
12345
If not, please comment below. Attach additional sheets if necessary.
Register now…and receive all the privileges associated with being a Thermo Fisher Scientific product user including customer
support, application reports, and technical reports.
MY ORGANIZATION IS: (Check only one)MY PRIMARY APPLICATION IS: (Check only one)
Name __________________________________________________Title__________________________________________________________________
Company __________________________________________________ __________________________________________________________________
Address ___________________________________________________ __________________________________________________________________
City/State _________________ ________________________Postal Code____ ______________________________________________________________
Country ___________________________________________________ __________________________________________________________________
Telephone_______________________________________________ Ext. __________________________________________________________________
Serial Number __________________________________ Date purchased __________________________________________________________________
Fold and mail or email to: techpubsaustin@thermofisher.com
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable domestic and international regulations. When the system is delivered to you, it meets all pertinent
electromagnetic compatibility (EMC) and safety standards as described below.
EMC Directive 89/336/EEC
EMC compliance has been evaluated by Professional Testing.
• PolarisQ, ITQ, and Ion Trap Series standards: EMC EN 55011:1998 + EN 50082-1:1998,
Safety EN 61010-1:1990 + A1:1992 + A2:1995
• DSQ standards: EMC EN 61326-1:1998 + A1:1998. Safety EN 61010-1:1990 + A1:1992 + A2:1995
• DSQ II standards: EMC EN 61326-1:1997 + A1:1998 + A2:2001. Safety EN 61010-1:2001
• Direct Probe Controller (DPC) standards: EMC EN 55011:1991 + EN 50082-1:1992. Safety EN 61010-1:1994
Low Voltage Safety Compliance
This device complies with Low Voltage Directive 73/23/EEC and harmonized standard EN 61010-1:2001. Changes
that you make to your system may void compliance with one or more of these EMC and safety standards. Changes to
your system include replacing a part or adding components, options, or peripherals not specifically authorized and
qualified by Thermo Fisher Scientific. To ensure continued compliance with EMC and safety standards, replacement
parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific or one of its
authorized representatives.
FCC Compliance Statement
Certifications, FCC part 15, Class A
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates, uses,
and can radiate radio frequency energy. If it is not installed and used in accordance with the instruction
manual, it may cause harmful interference to radio communication. Operation of this equipment in a
residential area is likely to cause harmful interference. In this case, users will be required to correct the
interference at their own expense. Detailed installation requirements are in the respective instrument’s
preinstallation guide.
WARNING Read and understand the various precautionary notes, signs, and symbols
contained inside this manual pertaining to the safe use and operation of this product
before using the device.
Notice on Lifting and Handling of Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific
instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/or bulky for one
person alone to handle safely.
Notice on the Proper Use of Thermo Scientific Instruments
In compliance with international regulations: Use of this instrument in a manner not specified by Thermo Fisher
Scientific could impair any protection provided by the instrument.
Notice on the Susceptibility to Electromagnetic Transmissions
Your instrument is designed to work in a controlled electromagnetic environment. Do not use radio frequency
transmitters, such as mobile phones, in close proximity to the instrument.
For manufacturing location, see the label on the instrument.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE)
Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union
(EU) Member State, and these companies should dispose of or recycle this product. See www.thermo.com/
WEEERoHS for further information on Thermo Fisher Scientific’s compliance with these Directives and the recyclers
in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen. Das
Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-Mitgliedsstaaten
getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Mehr Information
über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, über die Verwerter, und weitere Hinweise, die
nützlich sind, um die Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter
www.thermo.com/WEEERoHS
.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et
Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de
l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la
conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les produits
Thermo Fisher Scientific qui peuvent aider la détection des substances sujettes à la directive RoHS sont disponibles sur
xiiIon Trap Series Hardware ManualThermo Scientific
Preface
This guide contains instructions to operate and maintain the instrument. It is also designed to
increase understanding and knowledge of the mass spectrometer assembly and theory of
operations.
About Your System
Thermo Fisher Scientific systems provide the highest caliber gas chromatography/mass
spectrometry (GC/MS) instrumentation available on today’s market.
P
GC/MS represents a combination of two powerful analytical techniques: GC, which acts as a
separation technique and MS, which acts as a detection technique. Complex mixtures of
individual compounds can be injected into the GC, either manually or through the use of an
optional autosampler, and then separated for presentation to the MS. The MS will then
generate a mass spectrum of the GC eluent and its components, which can be used for
qualitative identification as well as accurate and precise quantification of the individual
compounds present in the sample.
IMPORTANT Thermo Fisher Scientific systems are designed to optimize both the
separation and detection capabilities of GC/MS techniques and combine them in a
synergistic fashion to provide high performance analytical capabilities for both research
and routine applications. More information on the use of this system can be found in
related documentation sources and through the provided contact information.
WARNING Thermo Fisher Scientific systems operate safely and reliably under carefully
controlled environmental conditions. If the equipment is used in manner not specified by
the manufacturer, the protections provided by the equipment may be impaired. If you
maintain a system outside the specifications listed in this guide, failures of many types,
including personal injury or death, may occur. The repair of instrument failures caused by
operation in a manner not specified by the manufacturer is specifically excluded from the
Standard Warranty and service contract coverage.
Thermo ScientificIon Trap Series Hardware Manualxiii
Preface
Power Ratings
Power Ratings
Mass Spectrometer (MS)
• 120 V ac +6/-10%, 50/60 Hz, 15 A max
• 230 V ac ±10%, 50/60 Hz, 10 A max
Gas Chromatograph (GC)
• 120 V ac +6%/-10%, 50/60 Hz, 16 A max
• 230 V ac ±10%, 50/60 Hz, 16 A max
Detailed instrument specifications are in the Product Specification or Product Brochure.
Safety Alerts and Special Notices
Make sure you follow the precautionary notices presented in this guide. Safety and other
special notices appear in boxes and include the following:
WARNING A warning safety alert is used to prevent actions that could cause personal
injury. It highlights hazards to humans or the environment. When you see a safety alert on
your instrument or in the documentation, carefully follow the safety instructions before
proceeding.
CAUTION A cautionary safety alert is used to prevent actions that may cause personal
injury or instrument damage. We use it to highlight information necessary to prevent
personal injury or damage to software, loss of data, or invalid test results; or might contain
information that is critical for optimal system performance. When you see a safety alert on
your instrument or in the documentation, carefully follow the safety instructions before
proceeding.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Emphaizes important information about a task.
Tip Helpful information that can make a task easier.
xivIon Trap Series Hardware ManualThermo Scientific
Safety Symbols and Signal Words
All safety symbols are followed by WARNING or CAUTION, which indicates the degree of risk
for personal injury and/or instrument damage. Cautions and warnings are following by a
descriptor, such as BURN HAZARD. A WARNING is intended to prevent improper actions that
could cause personal injury. Whereas, a CAUTION in intended to prevent improper actions
that may cause personal injury and/or instrument damage. The following safety symbols may
be found on your instrument and/or in this guide:
BURN HAZARD. This symbol indicates a hot surface that could or may cause
burn injuries.
ELECTRICAL SHOCK HAZARD. This symbol indicates that an electrical shock could or may occur.
FIRE HAZARD. This symbol indicates a risk of fire or flammability, or that
fire/flammability damage could or may occur.
FLAMMABLE GAS HAZARD. This symbol alerts you to gases that are
compressed, liquefied or dissolved under pressure and can ignite on contact
with an ignition source. This symbol indicates this risk could or may cause
physical injury.
GLOVES REQUIRED. This symbol indicates that you must wear gloves when
performing a task or else physical injury could or may occur.
HAND AND CHEMICAL HAZARD. This symbol indicates that chemical
damage or physical injury could or may occur.
INSTRUMENT DAMAGE. This symbol indicates that damage to the
instrument or module may occur. This damage may not be covered under the
standard warranty.
LIFTING HAZARD. This symbol indicates two or more people are required to
lift the object to prevent a physical injury that could or may occur.
MATERIAL AND EYE HAZARD. This symbol indicates that eye damage could
or may occur.
RADIOACTIVE. This symbol indicates the presence of radioactive material
could or may occur.
READ MANUAL. This symbol alerts you to carefully read your instrument’s
operational instructions before usage to ensure your safety and the
instrument’s operational ability. Failing to carefully read the instructions could
or may put you at risk for a physical injury.
Preface
Safety Alerts and Special Notices
TOXIC SUBSTANCES HAZARD. This symbol indicates that exposure to a
toxic substance will, could, or may cause personal injury or death.
This is the general warning symbol that the ISO 3864-2 standard refers to as
the general warning signal to prevent personal injury. It is a triangle with an
exclamation mark that precedes the WARNING safety alert word. In the
vocabulary of ANSI Z535 signage, this symbol indicates a possible personal
injury hazard exists if the instrument is improperly used or if unsafe actions
occur. We use this symbol and another appropriate safety symbol to alert to
an imminent or potential hazard that could cause personal injury.
Thermo ScientificIon Trap Series Hardware Manualxv
Preface
Contacting Us
Contacting Us
There are several ways to contact Thermo Fisher Scientific.
Y To contact Technical Support
Phone800-532-4752
Fax561-688-8736
E-mailUS.Techsupport.Analyzer@thermofisher.com
Find software updates and utilities to download at http://mssupport.thermo.com.
Y To contact Customer Service for ordering information
Phone800-532-4752
Fax561-688-8731
Web sitehttp://www.thermo.com/com/cda/resources/resources_
detail/1,,12512,00.html
Y To suggest changes to documentation or to Instrument Help
• Fill out a reader survey online at www.thermo.com/lcms-techpubs.
• Send an e-mail message to the Technical Publications Editor at
techpubsaustin@thermofisher.com.
Related Documentation
In addition to this guide, Thermo Scientific provides the following documents for the ITQ
Series. These documents are also available on a “Print-By-Request” basis.
Ion Trap Series Document Set, PN 120425-0ALL
• ITQ Preinstallation Guide, PN 120425-00PI
• Ion Trap Series Hardware Manual, PN 120425-00HW
• Ion Trap Series User’s Guide, PN 120425-USER
Instrument Help is available from within the software.
xviIon Trap Series Hardware ManualThermo Scientific
Routine Operations
This chapter describes how to use the Ion Trap Series mass spectrometer. This mass
spectrometer uses ion trap technology that gives analysts, technicians, and technical directors
a powerful tool for mass spectrometry and complete control when using the Xcalibur
software.
Contents
• “Installing a GC Column” on page 2
• “Removing a GC Column” on page 13
• “Starting Up” on page 14
1
• “Shutting Down” on page 16
Thermo ScientificIon Trap Series Hardware Manual1
1
Routine Operations
Installing a GC Column
Installing a GC Column
To install a column, you will need:
• Capillary column, 5MS, 30 m, 0.25 mm i.d., 0.25 μm (PN A0012-08143), or similar
• Injector ferrule, for 0.25 mm column (PN 290 134 88)
• Leak detector (Thermo Scientific GLD Pro PN 66002-001 or equivalent)
• Lint-free tissue
• Magnifying glass
• Methanol or other suitable solvent
• Scoring wafer (or sapphire scribe) to cut capillary column
• Transfer Line ferrule, 0.4 mm i.d. (PN A0101-18100)
• Wrench, open-ended, 5/16-in.
• Wrench, two, open-ended, 7/16-in.
• Wrench, open-ended, 6 mm
Refer to the documentation supplied with your GC for additional setup information.
WARNING - BURN HAZARD. The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
1. Connect the column to the injector, as shown in Figure 1.
Figure 1. GC Injector (Front)
6
1
5
4
Distance
2
3
4
1Injector4Column
2Injector Ferrule5Column Outlet
3Injector Nut6Transfer Line
Note Wear clean, lint- and powder-free gloves when you handle the column and
injector ferrule.
2Ion Trap Series Hardware ManualThermo Scientific
1
Routine Operations
Installing a GC Column
a.Unwind about half a turn of the column.
b.Wipe about 100 mm (4 in). of the column with a tissue soaked in methanol.
c.Insert the column through the injector nut and ferrule (open end up).
Note Sliding a septum on the column before the injector nut will help you
measure the proper distance between the nut and the end of the column.
d. Wipe the column again with a tissue soaked in methanol.
e.Score and then break the column about 2.5 cm (1 in). from the end with a scoring
wafer. With the magnifying glass, check for an even, flat cut. Repeat if necessary.
f.Insert the column into the injector so that the end of the column is the proper
distance from the back of the injector nut. Proper distances are as follows:
splitless = 64 mm, split = 40 mm, PTV = 30 mm.
g.Tighten the injector nut by hand and then turn it an additional quarter turn with the
wrench.
h. Turn on the gas chromatograph.
2. Set up the gas chromatograph.
a.Set the oven and injector temperatures to 40 °C.
b.Set the injector flow to 1.0 mL/min.
c.Turn off vacuum compensation (under the right or left carrier menu).
d. Dip the column outlet in a small vial of methanol. Bubbles indicate there is flow
through the column.
e.Allow the column
to purge for at least 10 minutes.
3. Perform column characterization.
a.Raise the oven and injector temperatures to 50 °C and allow them to stabilize. Then
confirm that the carrier gas is on.
CAUTION - INSTRUMENT DAMAGE. Do not allow the clear plastic component to
exceed 80 °C. Otherwise, it will melt and damage the instrument.
Thermo ScientificIon Trap Series Hardware Manual3
1
Routine Operations
Installing a GC Column
b. Carefully push the capillary column end into the flowmeter section of the
column-flow meter connector, as shown in Figure 2.
Figure 2. Flow Measure
c.Connect the flowmeter to the dedicated fitting on the column flow meter connector.
Note The available options (Right column or Left column) will differ depending
on the configuration of your GC.
d.Press Column Eval to open the following menu:
COLUMN INFORMATION
Right column
Left column <
Note Make sure the Vacuum Compensation parameter is set to OFF in the
Carrier menu, which is accessed by pressing the Left Carrier or Right Carrier
button. Otherwise an accurate column evaluation cannot be performed.
e.Scroll to the Right or Left column to evaluate and press Enter. The following menu
appears if you select the left column and and use the settings below to set the physical
characteristics of the column.
LEFT1 COLUMN INFO
Length (m) 15.00
ID (mm) 0.25
Film th. (um) 0.25
1. These settings could also apply to a right
column.
4Ion Trap Series Hardware ManualThermo Scientific
1
Routine Operations
Installing a GC Column
f.Select Pre/post column? yes (Y) or no (N).
• If a pre-/post-column is not present, select N.
• If a pre-/post-column is present, select Y. You must set the length and nominal
internal diameter of the pre-/post-column in the same valid ranges for the
column. The following two lines are added to the menu.
Pre/post column? Y
P/p col. L 10.00
P/p col. ID 0.530
g.According to the physical characteristics of the column, the system calculates and
displays the relevant Column K-factor.
1
LEFT
COLUMN INFO
Column K= (0.8087)
Calc’d ID (0.0000)
Run Column eval?
1. These settings could also apply to a right
column.
h. To evaluate the column, scroll to Run column eval.? and press ENTER.
The system will set the Column K factor by pressurizing the column to obtain a
nominal carrier flow of 5 mL/min. The display shows the pressure.
i.Using the flowmeter, measure the carrier gas flow at the outlet of the column. Scroll
to Measured Flow and set the value of the flow.
EVALUATING L COLUMN
Pressure (162)
Measured Flow: 4.90
Use <STOP> to abort
Note To abort column evaluation, you can press STOP.
j.The following message will display if the operation was successful.
L. COL. EVALUATION
COMPLETED
Calc’d ID 0.242
K. 0.9020
Thermo ScientificIon Trap Series Hardware Manual5
1
Routine Operations
Installing a GC Column
k.Expect a K-factor of about 0.7 – 0.9 for a 15 m, 0.25 mm i.d. column (1.3 – 2.0 for
a 30 m, 0.25 mm i.d. column). If the column does not report a K-factor within this
range or within 0.1 units of the previous stored value, check for a leak or broken
column using the leak detector. The K-factor is a measured resistance for the column.
A K-factor that is too low may indicate a leak in the system, while a K-factor that is
too high may indicate a blockage.
4. Perform a column leak check.
a.Carefully push the capillary column end into the column section of the column-flow
meter connector, as shown in Figure 3.
CAUTION - INSTRUMENT DAMAGE. Do not allow the clear plastic component to
exceed 80 °C. Otherwise, it will melt and damage the instrument.
Figure 3. Leak Check
A B C
b.Press Leak Check to access the Leak Check Columns menu.
Note The available options (Right column or Left column) will differ depending
on the configuration of your GC.
LEAK CHECK COLUMN
Right column
Left column <
1. This item appears if the relevant channel is
present and configured.
c.Scroll to a channel to open the left or right column’s leak check menu.
L. COL. LEAK CHECK
Start Leak Check
Leak Check Settings
6Ion Trap Series Hardware ManualThermo Scientific
1
Routine Operations
Installing a GC Column
d.Scroll to Leak Check Settings.
L. COL. SETT.
Check Leak press 200
Sensitivity 5.0
e.In the Check at press line, set the pressure at which the leak check will be performed.
The value must be between 10 and 999 kPa (1.45 and 145 psi).
f.In the Sensitivity line, set the maximum pressure drop that is allowed during the test.
The value must be between 1 and 10 kPa (0.145 and 1.45 psi).
g.Press CLEAR to exit the Leak Check Settings menu and return to the previous menu.
L. COL. LEAK CHECK
Start leak check
Leak Check settings
h. Select Start leak check to begin operation. The split and purge valves of the selected
channel are automatically closed and the channel is pressurized with carrier gas to the
leak check set point.
CHECKING L COLUMN
Pressure (200)
Elapsed time 0.90
Use <STOP> to abort
Note To abort a leak check, you can press STOP.
i.The system monitors the pressure for one minute. If the pressure does not drop more
than the maximum allowed sensitivity value, then the leak check will pass.
R/L.LEAK CHECK
COMPLETED
SUCCESSFULLY
Leak check passed.
If the leak check did not pass, you should use the leak detector to find and fix the
leaks.
Tip Leaks can be caused by not tightening the fitting on the clear plastic
component. We recommend that you check that fitting before looking elsewhere.
j.Repeat the column evaluation and leak check until no leaks are indicated.
Thermo ScientificIon Trap Series Hardware Manual7
1
Routine Operations
Installing a GC Column
5. Disconnect the column from the clear plastic component.
6. Remove the clear plastic component, including its fittings, from the oven and set it aside.
7. Condition the column before inserting it into the mass spectrometer.
CAUTION INSTRUMENT DAMAGE. The material released from the column during the
conditioning (column bleed) will contaminate the ion source if the column is inserted
into the transfer line. The ion source must then be cleaned.
a.Raise the injector temperature to the desired temperature (normally 250 °C).
b.Run the slow temperature program that is recommended by the manufacturer. For
example, hold the column at 40 °C for 15 minutes, then ramp it to 10 °C per minute
up to 10 °C above the maximum temperature at which you will operate the column
(normally 300+10 °C = 310 °C). Hold the column at this temperature for two hours.
CAUTION INSTRUMENT DAMAGE. Never exceed the column manufacturer’s maximum
operating temperature.
8. Connect the column to the transfer line.
a.Shut down and vent the mass spectrometer, as described in “Shutting Down” on
page 16.
b.Lower the oven temperature to 40 °C and allow it to cool before continuing.
WARNING - BURN HAZARD. The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
c.Unwind about one turn of the column (shown in Figure 4) from the column outlet
end.
Figure 4. Column Measuring Tool
1
3
2
4
5
1Column Measuring Tool*4Column
2Transfer Line Ferrule5Septum
3Transfer Line Nut
* The Column Measuring Tool is sold separately and can be purchased through
Thermo Fisher Scientific using part number 119640-0550.
8Ion Trap Series Hardware ManualThermo Scientific
1
Routine Operations
Installing a GC Column
Note Wear clean, lint- and powder-free gloves when you handle the column and
transfer line ferrule.
d. Wipe about 300 mm (12 in). of the column with a tissue soaked in methanol.
e.Insert the column through the septum, transfer line nut, and ferrule. Wipe the
column again with a tissue soaked in methanol.
Note Sliding a septum on the column before the transfer line nut will help you
measure the proper distance between the nut and the end of the column. The
column should extend approximately 1–2 mm past the end of the transfer line.
f.Score and break the end of the column with a scoring wafer. With the magnifying
glass, check for an even, flat cut. Repeat, if necessary.
g.Insert the column into the transfer line using one of the following methods: Method
One, Method Two, or Method Three.
Method One
If you have a column measuring tool, follow these steps to insert the column:
1. Screw the transfer line nut onto the column measuring tool
1
.
2. Push the column past the end of the column measuring tool and score and break the end
of the column with a scoring wafer. With the magnifying glass, check for an even, flat cut.
Repeat if necessary.
3. Pull the column back so that it is flush with the end of the column measuring tool and
tighten the transfer line nut.
4. Slide the septum up to the back of the transfer line nut.
5. Remove the column, transfer line nut and ferrule from the column measuring tool, as
shown in Figure 5.
Figure 5. Transfer Line (Front)
1
2
3
4
1
Purchase the Column Measuring Tool PN 119640-0550 from Thermo Fisher Scientific.
Thermo ScientificIon Trap Series Hardware Manual9
1
Routine Operations
Installing a GC Column
Figure 5. Transfer Line (Front)
1Septum5Transfer Line Union
2Transfer Line Nut4Transfer Line
a.Insert the column into the transfer line. Be careful not to move the septum.
b.Tighten the transfer line nut and the transfer line union.
6. Condition the transfer line ferrule. Graphite/vespel ferrules like the transfer line ferrule
require conditioning to ensure a leak-tight seal.
a.Raise the oven temperature to the maximum temperature you will operate the
column (normally 300 °C).
b.Wait 10 minutes.
c.Lower the oven temperature to 40 °C and allow it to cool before continuing.
WARNING - BURN HAZARD. The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
Method Two
d.Re-tighten the transfer line nut and the transfer line union.
7. Set up the gas chromatograph.
a.Make sure the column does not have any sharp bends and that it does not touch any
metal objects or walls inside the oven.
b.Raise the oven temperature to the initial temperature you will use (normally 40 °C).
c.Turn on the vacuum compensation (under the right or left carrier menu).
If you have the optional inlet valve, follow these steps to insert the column:
Tip Slide a septum on the column before adding the transfer line nut to make it easier to
measure the proper distance between the nut and the end of the column. You can also use
a white marker to mark the position of the column and nut.
1. Remove the mass spectrometer front cover so you can get a better view of the column.
2. Using the I/R tool, remove the ion volume, as described in “Cleaning the Ion Source
Components” on page 37.
3. Insert the column into the transfer line and tighten the transfer line nut by hand.
4. Push the column in until you can see it through the inlet valve.
5. Pull the column back just far enough that you cannot see it.
6. Tighten the transfer line nut and transfer line union.
10Ion Trap Series Hardware ManualThermo Scientific
1
Routine Operations
Installing a GC Column
7. Using the I/R tool, replace the ion volume, as described in “Cleaning the Ion Source
Components” on page 37.
8. Condition the transfer line ferrule. Graphite/vespel ferrules like the transfer line ferrule
require conditioning to ensure a leak-tight seal.
a.Raise the oven temperature to the maximum temperature you will operate the
column (normally 300 °C).
b.Wait 10 minutes.
c.Lower the oven temperature to 40 °C and allow it to cool before continuing.
WARNING - BURN HAZARD. The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
d.Re-tighten the transfer line nut and the transfer line union.
9. Set up the gas chromatograph.
a.Make sure the column does not have any sharp bends and that it does not touch any
metal objects or walls inside the oven.
Method Three
b.Raise the oven temperature to the initial temperature you will use (normally 40 °C).
Note Turn on vacuum compensation (under the Right or Left Carrier menu).
If you do not have an inlet valve, follow these steps to insert the column:
1. Remove the front and top covers so you can get a better view of the column.
2. Remove the vacuum manifold cover, as described in “Removing the Vacuum Manifold
Cover” on page 28.
3. Remove the ion source assembly, as described in“Cleaning the Ion Source Components”
on page 37.
4. Adjust the column so that it extends 1–2 mm past the end of the transfer line.
5. Tighten the transfer line nut and transfer line union.
6. Condition the transfer line ferrule. Graphite/vespel ferrules like the transfer line ferrule
require conditioning to ensure a leak-tight seal.
a.Raise the oven temperature to the maximum temperature you will operate the
column (normally 300 °C).
b.Wait 10 minutes.
c.Lower the oven temperature to 40 °C and allow it to cool before continuing.
Thermo ScientificIon Trap Series Hardware Manual11
1
Routine Operations
Installing a GC Column
WARNING - BURN HAZARD. The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
d.Re-tighten the transfer line nut and the transfer line union.
7. Replace the ion source assembly, as described in “Cleaning the Ion Source Components”
on page 37.
8. Replace the vacuum manifold cover.
9. Replace the front and top covers.
10. Condition the transfer line ferrule. Graphite/vespel ferrules like the transfer line ferrule
require conditioning to ensure a leak-tight seal.
a.Raise the oven temperature to the maximum temperature you will operate the
column (normally 300 °C).
b.Wait 10 minutes.
c.Lower the oven temperature to 30 °C and allow it to cool before continuing.
WARNING - BURN HAZARD. The oven may be hot. Allow it to cool to room temperature
before opening it. The injector will still be hot, so do not touch it.
d.Re-tighten the transfer line nut and the transfer line union.
11. Set up the gas chromatograph.
a.Make sure the column does not have any sharp bends and that it does not touch any
metal objects or walls inside the oven.
b.Raise the oven temperature to the initial temperature you will use (normally 40 °C).
Note Turn on vacuum compensation (under the Right or Left Carrier menu).
12Ion Trap Series Hardware ManualThermo Scientific
Removing a GC Column
To remove a GC column, you will need:
• Gloves, clean, lint- and powder-free
• Wrench, open-ended, 5/16-in.
• Wrench, open-ended, 7/16-in.
• Wrench, open-ended, 6 mm
1. Shut down the mass spectrometer.
a.Shut down and vent the mass spectrometer, as described in “Shutting Down” on
page 16.
b.Lower the oven, injector, and transfer line temperatures to 30 °C and allow them to
cool before continuing.
WARNING - BURN HAZARD. The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
c.Once the oven, injector, and transfer line are cool, turn off the gas chromatograph.
1
Routine Operations
Removing a GC Column
2. Remove the column from the transfer line, as shown in Figure 6.
Figure 6. Transfer Line (Front)
1
2
3
4
1Septum3Transfer Line Union
2Transfer Line Nut4Transfer Line
a.Unscrew the transfer line nut.
b.Remove the column from the transfer line.
Thermo ScientificIon Trap Series Hardware Manual13
1
Routine Operations
Starting Up
3. Remove the column from the injector, as shown in Figure 7.
Figure 7. Injector (Front)
6
5
1
Distance
2
3
Starting Up
4
4
1Injector4Column
2Injector Ferrule5Column Outlet
3Injector Nut6Transfer Line
a.Unscrew the injector nut.
b.Remove the column from the injector.
1. Set up the mass spectrometer.
a.Install the GC column, as described in “Installing a GC Column” on page 2.
b.Verify that the gas chromatograph is on and there is carrier gas flowing through the
column into the mass spectrometer.
WARNING - INSTRUMENT DAMAGE. Damage may occur if you turn on the mass
spectrometer without column flow. This forces air to be drawn through the column. This
large air leak into the mass spectrometer may cause the ion source to require cleaning as
well as potentially damaging the column.
c.Plug in the mass spectrometer power cord.
14Ion Trap Series Hardware ManualThermo Scientific
Loading...
+ 220 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.