Thermo Scientifi? Cryostat Microm Cryo-Star HM 560 Instruction Manual

Thermo Scientific Cryostat Microm Cryo Star HM 560
INSTRUCTION MANUAL
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
1
Thermo Scientific Cryostat Microm Cryo Star HM 560
2
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
CERTIFICATION
Thermo Fisher Scientific Microm International GmbH certifies that this instrument has been tested and checked carefully. Its technical data was verified before shipment to be in accordance with the published specifications.
The instrument complies with applicable international safety regulations.
WARRANTY
This Thermo Fisher Scientific product is warranted against defects in material and workmanship for a period of 1 year. Parts which prove to be defective during the warranty period will be repaired or replaced free of charge by the manufacturer. No other warranty is expressed or implied. Unauthorized modification or repair by third party persons will void the warranty. The warranty will expire in case of improper or wrong use of the instrument and in case the warning and precautionary messages are not observed. Thermo Fisher Scientific is not liable for any occurring damage.
Errors and omissions excepted. Subject to amendment and improvement without further notice.
This instruction manual will be supplied together with each instrument. Further copies can be ordered at the nearest Thermo Fisher Scientific sales office by giving the serial number of the instrument, the number of the instruction manual and the date of issue.
This instruction manual is available in the following languages:
Cat. No. German: 387 490 English: 387 500 French: 387 510
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
3
Thermo Scientific Cryostat Microm Cryo Star HM 560
INTENDED USE
Dear valued Customer,
Thank you for buying this Thermo Fisher Scientific instrument.
Before putting the instrument into operation, please read these operating instructions carefully to familiarize you with its proper operation and functions.
Only skilled or specially trained personnel must operate the microtome cryostat HM 560, i.e. clamping the specimen, trimming, sectioning and taking off the sections from the instrument. The listed and marked safety measures as well as the regulations and hygiene measures of your respective lab must strictly be observed.
Serial No.: __________________
Please check the Serial No. on the type plate, which is placed on the rear side of your instrument and enter this number here. This way, questions and service can be handled faster.
Instruction Manual No. 387500
Issued on February 03rd, 2009
Microm International GmbH part of Thermo Fisher Scientific Robert-Bosch-Strasse 49
69190 Walldorf Germany
Phone: +49.(0)6227.836-0 Fax: +49.(0)6227.836-111 Email: info.dxd.dewal@thermofisher.com Internet: www.ap-walldorf.com
4
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
Intended Use
Table of Contents
Safety Precautions
PART 1 INTRODUCTION
1-1 Description of the Cryo-Star HM 560 1-2 Technical data Cryo-Star HM 560
PART 2 OPERATING INSTRUCTIONS
2-1 Setting up the cryostat
2-2 Initial turn-on
2-3 Basic operational rules
2-4 Setting section thickness and trimming thickness
2-5 Specimen and knife cooling 2-5-1 Actual and set value of the specimen temperature 2-5-2 Actual and set value of the knife temperature 2-5-3 Function fast freezing 2-5-4 Programming temperature parameters
2-6 Cutting movement and retraction
2-7 Feed 2-7-1 Knife carrier coarse feed 2-7-2 Speed of the knife carrier coarse feed 2-7-3 Automatic approach system 2-7-4 Trimming and first cuts 2-7-5 Fine feed
2-8 Setting the cutting window
2-9 Hand wheel brake
2-10 Detaching the specimen
2-11 Indication of cutting processes 2-11-1 Section counter 2-11-2 Section thickness sum 2-11-3 Remaining travel to front end position 2-11-4 Real time
2-12 Setting the real time, WAKE time and ACTIVE time 2-12-1 Setting the real time 2-12-2 Setting the WAKE time 2-12-3 Setting the ACTIVE time
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
5
Thermo Scientific Cryostat Microm Cryo Star HM 560
2-13 Defrosting 2-13-1 Setting the defrost time 2-13-2 Defrosting cycle 2-13-3 Interrupting a defrosting cycle
2-14 Turning off the function retraction
2-15 Customer-specific settings 2-15-1 Setting the contrast 2-15-2 Selecting the language
2-16 Turning off the cooling function
2-17 Illumination of the cooling chamber
2-18 Adjusting the operating panel
2-19 Specimen holder, freezing-on and specimen orientation 2-19-1 Specimen holder, freezing-on 2-19-2 Specimen orientation
2-20 Knife carriers 2-20-1 Standard knife carrier SM 2-20-2 Disposable blade holder SE 2-20-3 Modification from standard knife carrier SM to disposable blade holder SE
2-21 Error code indication 2-21-1 Definition of the error codes
2-22 Accessories 2-22-1 Standard equipment 2-22-2 Additional equipment
PART 3 Theory of operation
3-1 Specimen and knife carrier cooling, defrost cycle
3-2 Cutting movement
3-3 Knife carrier coarse feed and trimming stages
3-4 Automatic approach system
PART 4 Working with the cryostat
4-1 Preparations on the microtome and inside the cooling chamber
4-2 Freezing on the specimen
4-4 Orientation and trimming of specimens
4-5 Sectioning and taking off sections
4-6 How to avoid malfunctions
6
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
PART 5 Maintenance of the cryostat
5-1 Shutting-off for cleaning
5-2 Cleaning and care of the microtome
5-3 Changing the fluorescent lamp
5-4 Disposal of the instrument after final shutdown
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
7
Thermo Scientific Cryostat Microm Cryo Star HM 560
SAFETY PRECAUTIONS
WARNING SIGNALS AND SYMBOLS
The installation and routine use of the Cryo-Star HM 560 is easy and safe if the instructions in this manual are being observed. However, the situations which might be risky for the personnel or equipment are specially marked in this manual with the following symbols and messages.
Note:
Special instructions regarding operation of the instrument.
Caution: Special precautionary measures to prevent damage to equipment. For a long lifetime of the equipment, please observe these instructions carefully.
Danger:
Special warning messages to prevent harm to persons and/or serious damage to equipment. For your own safety, please observe these instructions carefully.
Biohazard:
Warning of biological danger.
Separate taking back of electrical and electronic instruments in the countries of the European Union:
This is to be applied in the countries of the European Union and other European countries with a separate collecting system within the waste management. This product, being an electro and/or electronic instrument, must be treated separately within the waste management process (WEEE).
8
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
SAFETY PRECAUTIONS
A T T E N T I O N !
The operator's safety is affected, when the instrument is not operated in accordance with this instruction manual. Please observe the following general precautions during operation of this instrument. Failure to comply with these precautions violates safety standards and the intended use of the instrument. Thermo Fisher Scientific is not liable for misuse of the instruments and failure to comply with basic safety requirements.
INSTRUMENT GROUNDING
To avoid injury from electrical current, the instrument must be connected with the safety ground. The instrument is equipped with a three wire ground plug. The power outlet must be connected to the safety ground and must meet the International Electrotechnical Commission (IEC) regulations.
CAUTION: MAINS VOLTAGE
Never remove instrument covers during operation. Component replacements as well as adjustments must only be made by trained service personnel. Unplug the unit before removing or opening the covers.
DANGER IN EXPLOSIVE ENVIRONMENT
The instrument must not be operated in the presence of flammable gases.
HAZARD OF FROSTBITE
Avoid permanent touching of metal parts inside the cryostat microtome chamber as frostbite may occur at unprotected hands and arms.
HAZARD OF RADIOACTIVE RADIATION
When working with radioactive specimens observe all applicable radiation safety procedures. When working with radioactive contaminated material, appropriate safety and disinfection measures must be carried out. According to the rules and regulations concerning the handling of radioactive contaminated material of the respective laboratory, safety clothing (e.g. particle mask, gloves, protective shoe covers) must be worn. Radioactive contaminated waste must be disposed of according to the respective regulations.
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
9
HAZARD OF INFECTION
Thermo Scientific Cryostat Microm Cryo Star HM 560
Use the appropriate safety and disinfection measures when working with infectious specimens. According to the rules and regulations concerning the handling of infectious/radioactive contaminated material of the respective laboratory, safety clothing (e.g. particle mask, gloves, protective shoe covers) must be worn.
HAZARD OF BIOLOGICAL DANGER
Specimens used during the intended operation of the instrument might potentially be infectious. For this reason, it is recommended to observe the general laboratory regulations concerning protection against danger of infection. Information on decontamination media, their use, dilution and effective range of application can be read in the Laboratory Biosafety Manual : 1984 of the World Health Organization.
CAUTION: DANGER OF SQUEEZING
When placing the microtome cryostat HM 560 onto the height adjustable stage, it is absolutely necessary that the stage is in its upper position or in case of a mechanical stage that the rear wall will be removed. Not observing this leads to the risk of squeezing your hands between the rear carrying handles and the stage housing.
HAZARD OF MALFUNCTION
To avoid the hazard of malfunction of an instrument, it must only be operated in a controlled electromagnetic environment. This means, that transmitters such as mobile phones must not be operated in their close vicinity.
CARE IN USING MICROTOME KNIFE
To diminish the danger of being injured by the knife or blade, use the knife guard when adjusting specimen and knife. If possible, the specimen should be clamped in before the knife is inserted into the knife holder. Before changing the knife holder, always remove blade or knife! Unused
knives should always be kept in a knife case. Never place the knife with the cutting edge upwards. Never try to catch a dropping knife!! Never check the sharpness of the cutting edge with your fingers. The cutting edge is extremely sharp!
WASTE DISPOSAL
All debris, waste as well as infectious and radioactive contaminated material from operation must be disposed of in accordance with the respective regulations of the lab. Disinfection and cleaning liquids as well as section waste must be disposed of according to the respective regulations for special waste!
10
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
Fig. 1
Microm International GmbH
part of Thermo Fisher Scientific
Robert-Bosch-Str. 49
69190 Walldorf / Germany
387 500 - English
11
Thermo Scientific Cryostat Microm Cryo Star HM 560
1 INTRODUCTION
1-1 DESCRIPTION OF THE CRYO-STAR HM 560
Open top cryostat. Modular design. Stainless steel cooling chamber. Temperature regulation variable down to -50°C for specimen holder and down to -35°C for knife carrier. Graphic LC-display of preset and actual temperatures, electronic control with user-oriented touchpad keyboard, battery-buffered memory and self­diagnostic system. Standby and sleep status for noise reduction and energy savings. Automatic evaporator defrosting and in addition a manual defrost cycle which can be activated when needed. Defrost interrupt on keystroke request. Integrated rapid freezing device with controlled cooling element down to -60°C. Freezing device temperature independent of cryo-chamber temperature. Detach function to remove used specimen from specimen chucks. Sliding window with heater and integrated, position adjustable fluorescent lamp for cryo­chamber. Storage space inside the cryo­chamber. Flat storage space on top of the housing.
HM-Rotary Microtome with backlash and maintenance-free cross roller bearings in stainless steel. Electro-motorized feed system. Section thickness setting from 0,5 to 100 microns; up to 2 µm in 0,5 µm-increments, up to 10 µm in 1 µm-increments, up to 20 µm in 2 µm-increments, up to 50 µm in 5 µm-increments and up to 100 µm in 10 µm-increments.
Trimming thickness setting from 5 to 500 microns; up to 10 µm in 5 µm-increments, up to 100 µm in 10 µm-increments, up to 200 µm in 20 µm-increments and up to 500 µm in 50 µm-increments.
Specimen retraction during the return stroke with optical indication. Horizontal feed range 48 mm. Vertical specimen stroke 60 mm. Max. specimen size 75 x 55 mm. Motorized coarse feed in two directions with three speed selections. Limit indication and automated switch off at front and rear limits of horizontal travel. Automatic approach system for exact and safe approach of specimen towards the knife edge.
Section counter and indication of sum of section thicknesses with reset-button. Indication of remaining travel. Hand wheel brake in any position. Fine orientation of specimen on two axes. Rotatable on Z-axis, 360°.
Basic outfit with three specimen chucks, cryo­compound, 100 ml cryostat oil, brush shelf and section waste tray.
A multi-use knife carrier is available. The standard knife carrier is designed so the knives as well as the disposable blade holder can be easily clamped in place and adjusted. Depending on the version, the disposable blade holders SE take up low or high profile blades. The corresponding anti-roll guides facilitate taking off sections.
12
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
1-2 TECHNICAL DATA CRYO-STAR HM 560
Specimen temperature control: .......................................................................... +10°C to -50°C
Knife carrier control: ................................................................................ -10°C to -35°C
Fast freezing station: ................................................................................. down to -60°C
Defrosting: automatic defrosting, adjustable with temperature control manual immediate defrosting
Microtome: Section thickness range: ............................................. 1 - 500 µm
Fine section thickness range: .................................... 0,5 - 100 µm
Resolution: .................................................. 0,5 µm for 0,5 - 2 µm
.......................................................................... 1 µm for 2 - 10 µm
........................................................................ 2 µm for 10 - 20 µm
........................................................................ 5 µm for 20 - 50 µm
.................................................................... 10 µm for 50 - 100 µm
Trimming thickness range: ........................................... 5 - 500 µm
Resolution: ........................................................ 5 µm for 5 - 10 µm
.................................................................... 10 µm for 10 - 100 µm
.................................................................. 20 µm for 100 - 200 µm
.................................................................. 50 µm for 200 - 500 µm
Knife carrier retraction during return travel ...........................40 µm
Horizontal knife carrier range ............................................ 48 mm
Vertical specimen stroke ................................................... 60 mm
Read-outs: Graphical LC display: section thicknesses, section counter, sum of section thicknesses, remaining travel to front end position, size of the cutting window
Specimen approach: .......................................................................... automatic function
........................................ alternatively manual with variable speed
Temperature parameter: ............................... temperature set values can be stored in pairs
Size of chucks: .......................................................................................30, 40 mm
............................................................. special sizes upon request
Specimen orientation: x - and y - axes: ......................................................... universal 8°
z - axis: ......................................................................... up to 360°
Coarse feed: ........................................... motorized, graduated and continuous
Cooling chamber illumination: ................................................ with variable illumination position
Sliding window: ............................................................................................. heated
Cutting speed: .................................................................................. 0 - 250 mm/s
Hand wheel brake: ................................................................................. in any position
All temperatures refer to an ambient temperature of +20°C!
Cont'd on page -2-
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
13
Thermo Scientific Cryostat Microm Cryo Star HM 560
1-2 TECHNICAL DATA CRYO-STAR HM 560
Storage
temperature range: ......................................................................... -20°C up to +50°C
Operating conditions: ....................... +5°C up to +35°C (at a max. rel. humidity of 60%)
........................................................... altitude up to 2000 m M.S.L.
.......................................................................... for indoor use only
Power requirements: ....................................................... 100 V 12 A +/-10% 50...60 Hz
............................................................... 115 V 12 A +/-10% 60 Hz
................................................. 220...230 V 6 A +/-10% 50...60 Hz
................................................................. 240 V 6 A +/-10% 50 Hz
Pollution degree: ...................................................................................................... 2
Overvoltage category: ...................................................................................................... II
Acoustic pressure: ......................................................................................... 45 dB(A)
............................... measured with 1 m distance to the instrument
Dimensions: .. Wide (w/o Hand wheel): 769 mm, deep: 932 mm, high: 578 mm
Weight: ............................................................................................. 141 kg
All temperatures refer to an ambient temperature of +20°C!
page -2-
14
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
1 = On indication (sleep status) 2 = Button specimen temperature 3 = Button knife temperature 4 = Button TRIM/FEED 5 = Button UP 6 = Button DOWN 7 = LC display 8 = Button fast freezing
Fig. 2 9 = Button options 10 = Button detaching 11 = Scroll button 12 = Button RESET 13 = Button PROG 14 = Info line 15 = Status line
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
15
Thermo Scientific Cryostat Microm Cryo Star HM 560
1 = Button cutting window 2 = Button ACA 3 = Button coarse feed, forwards 4 = Button coarse feed, backwards 5 = Button TRIM
16
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
Fig. 3
Thermo Scientific Cryostat Microm Cryo Star HM 560
1 = Symbol Hand wheel 2 = Button to activate the Hand wheel brake 3 = Button to loosen the Hand wheel brake 4 = Brake LED
Fig. 4
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
17
Thermo Scientific Cryostat Microm Cryo Star HM 560
2 OPERATING INSTRUCTIONS
2-1 SETTING UP THE CRYOSTAT
Note:
If the microtome cryostat is supplied with an extra height-adjustable stage, first unpack and set up this stage
according to the separate instruction manual.
Unpacking the instrument:
Loosen and remove the four tightening
straps.
Lift and remove the wooden top of the
packing.
Remove the foam parts upwards.
The four side parts of the packing, which are
connected with each other, can now be removed upwards.
Slightly tilt the cryostat to the side and
remove the right and left lower foam parts one after the other.
Lift the cryostat by the carrying handles (fig.
5.1 and 5.2) from the pallet.
In case the handles are not installed, securely
install the attached carrying handles (fig. 5.1 and 5.2) at the front and rear side of the instrument and carry the instrument to the site of installation (table according to the customer's choice) or place it onto the height­adjustable stage.
Danger:
When placing the microtome cryostat HM 560 onto the height adjustable
stage, it is absolutely necessary that the stage is in its upper position. Not observing this leads to the risk of squeezing your hands between the rear carrying handles and the stage housing.
Remove the handles (fig. 5.1 and 5.2).
cryostat, as when lowering the height­adjustable stage, the handles will hit the stage and cause damage to the stage.
Caution:
Before the initial set-up of the height-
adjustable stage, it is absolutely
necessary to dismount the two rear
carrying handles (fig. 5.2) on the
Fig. 5
18
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
Adjust the handles of the cryostat (fig. 5a.1
and 5a.2) on the rear side of the cryostat stage.
Carefully keep the shipping container for
possible return shipments.
Choose installation site that
enough ventilation for the cooling system is
guaranteed.
the distance between wall and rear panel is
approx. 10 - 15 cm.
the suction areas on either side are kept free.
Moreover, the installation site must be free from:
draught by open doors or by air conditioning
systems.
direct exposure to sunlight into the cooling
chamber.
Note:
Both measures reduce the formation of
frost and therefore result in more
favourable work conditions. A high air moisture as well as high ambient temperatures reduce the maximum performance of the instrument.
In case the instrument is equipped with the height-adjustable stage, turn the screws (fig.
5.5) on the front foot ends to fix the complete unit.
Afterwards:
Install the attached Hand wheel handle (fig.
5.3).
1
2
Fig. 5a
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
19
Thermo Scientific Cryostat Microm Cryo Star HM 560
2-2 INITIAL TURN-ON
As already mentioned in part 2-1, the height­adjustable stage, if equipped, must be set up according to the separate instruction manual before the initial turn-on of the microtome cryostat.
Caution:
Before turning on the instrument for the
first time, please check if the power
requirements indicated on the type plate (fig. 7) correspond to the power supply voltage being used.
No other instruments should be connected to
the circuit used for the cryostat, as the compressor needs high surge currents when started.
Do not use multi-socket power outlets and
extension lines for the supply of the stage.
Connect the power line of the instrument to
the power outlet.
Turn on the power switch (fig. 6.1) on the rear
side of the instrument.
Close the heated sliding window (fig. 5.4) and
cool down the instrument.
Note:
To avoid frost built-up on the
specimen, the specimen temperature
is controlled to a temperature of +10°C until the knife has reached a temperature of -10°C.
The cooling phase will take approx. 1 to 2 h depending on the set knife temperature between -15°C and -30°C.
1 = Power switch 2 = Mains plug 5 = Socket for option Vacutome/Disinfection
MICROM
69190 Walldorf, Germany
GmbH
Type HM 560V
Ser. No.XXXXX
Cat. No. 956320
220...230V 6A 50...60Hz
Cooling System
Refr.: R 404a
Amount: 0,275 kg Max. Press.: 2300 kPa
Made in Germany
I V D
I
0
GmbH
MICROM
69190 Walldorf, Germany
Type HM 560 MV
Ser. No.____
Cat. No. ______
__A / ___ V + 10% __Hz
Fuse 2 x __AT
HP Circuit
Refr.: R 134a Amount: 0,180kg Max. Press.: 2300kPa
LP Circuit
Refr.: R 404a Amount: 0,270kg Max. Press.: 2300kPa
Made in Germany
1
4
GmbH
MICROM
69190 Walldorf, Germany
Type Stage f. HM 560 HV
Ser. No. __ __ Cat. No. ______
_A /___V +10% __Hz
Fuse 2x __AT
Made in Germany
3
Fig. 6
Fig. 7
20
~___V F1/F2 2x __AT
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
I
0
Fig. 8
Thermo Scientific Cryostat Microm Cryo Star HM 560
2-3 BASIC OPERATIONAL RULES
When turning on the instrument, the display shows:
the configuration of the instrument
the section thickness selection
the specimen and knife temperatures with the
corresponding set and actual values
the symbol for the cutting window
This is the display (fig. 9) while the instrument is in its active status.
To save energy and extend the lifetime of some instrument components, the instrument has two more operative status:
Standby status
Sleep status
The instrument switches to the standby status when it is turned on, but not operated for more than a pre-selected active time (see part 2-12-
3) between 1 and 9 h. The chamber illumination goes off and the display (fig. 10) switches to a large indication of the knife actual temperature. This is the clear sign for the standby status and gives the operator information on the knife temperature, also from a greater distance.
When the instrument is not operated for another hour, the cryostat switches to the sleep status. The display illumination goes off. However, the equipment-on indicator (fig. 2.1) then lights up and shows that the instrument is still on.
Note:
While in the sleep status, the
temperature control of the knife
temperature features a reduced control accuracy. The specimen temperature in the sleep status is always kept at -15°C. The set temperature of the knife carrier is kept constant. This does not only save energy, but also reduces the formation of frost on the specimen clamping.
Fig. 9
Fig. 10
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
21
Thermo Scientific Cryostat Microm Cryo Star HM 560
To return to the active operating state, press
any key on the control panels, move the Hand wheel or wait until the WAKE time (see 2-12-2) has been reached.
The WAKE time is used to trigger the instrument back into the active state.
UP/DOWN buttons
These two buttons (fig. 11.5 and 11.6) are used to change pre-selected values of different kinds in connection with the chosen function.
section thickness
trimming thickness
specimen temperature
knife temperature
time settings
Note:
To operate the instrument easily and
comfortably with a few buttons only,
please note the following:
After having chosen a function, its pre-selected values can be changed via the UP or DOWN buttons. However, if no entry is made within 2 seconds, the operating control switches back into its basic status.
RESET-button
This button (fig. 12.4) is used to reset added values (e.g. number of sections, sum of section thicknesses) to zero. In some cases, this button is used as additional button for double­button functions (see part 2-8 and 2-13-3).
Keyboard lock
To lock the keyboards and to block the feed movement, press the buttons (fig. 12.4 and
12.1) for approx. 2 sec. The display (fig. 12a) shows LOCKED. To unlock the keyboards press the buttons (fig.
12.4 and 12.1) again.
Fig. 11
Fig. 12
Fig. 12a
Note:
When the keyboards have been
locked, this leads automatically after 5 sec. to the standby status and after another 5 sec. to the sleep status.
22
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
Thermo Scientific Cryostat Microm Cryo Star HM 560
2-4 SETTING SECTION THICKNESS AND TRIMMING THICKNESS
Basic status of the operating control: To choose between section thickness and trimming thickness
press button (fig. 14.4).
The selected setting is shown inverted on the
display.
Then enter the desired section or trimming
thickness via button UP (fig. 14.5) or DOWN (fig. 14.6).
The changed values are shown on the
display.
The graduation of the section thicknesses (which can be pre-selected) is divided into 5 ranges:
range graduation
from 0,5 µm to 2 µm 0,5 µm from 2 µm to 10 µm 1 µm from 10 µm to 20 µm 2 µm from 20 µm to 50 µm 5 µm from 50 µm to 100 µm 10 µm
The graduation of the trimming thicknesses (which can be pre-selected) is divided into 4 ranges:
range graduation
from 5 µm to 10 µm 5 µm from 10 µm to 100 µm 10 µm from 100 µm to 200 µm 20 µm from 200 µm to 500 µm 50 µm
Fig. 13
Fig. 14
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387 500 - English
23
Loading...
+ 53 hidden pages