TriPlus RSH, TRACE 1300, TRACE 1310, TRACE GC Ultra, and FOCUS GC are trademarks of Thermo
Fisher Scientific Inc., and its subsidiaries.
Published by Thermo Fisher Scientific S.p.A., Strada Rivoltana 20090 Rodano-Milan, Italy
Tel: +39 02 95059303; Fax: +39 02 95059388
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history:
First edition, released September 2011 “Original Instructions”
Second edition, October 2011; Third edition, March 2012; Fourth edition, September 2012; Fifth edition,
March 2013; Sixth edition, October 2013; Seventh edition, June 2014; Eighth Edition, February 2015; Ninth
Edition, December 2015
For Research Use Only. Not for use in diagnostic procedures.
The manual is well organized.12345
The manual is clearly written.12345
The manual contains all the info rmation I need.12345
The instructions are easy to follow.12345
The instructions are complete.12345
The technical information is easy to understand.12345
Examples of operation are clear and useful.12345
The figures are helpful.12345
I was able to operate the system using this manual.12345
If not, please comment below. Attach additional sheets if necessary.
Register now…and receive all the privileges associated with being a Thermo Fisher Scientific product user including customer
support, application reports, and technical reports.
MY ORGANIZATION IS: (Check only one)MY PRIMARY APPLICATION IS: (Check only one)
Name __________________________________________________Title__________________________________________________________________
Company __________________________________________________ __________________________________________________________________
Address ___________________________________________________ __________________________________________________________________
City/State _________________ ___________________ _____Postal Code____________________ ______________________________________________
Country ___________________________________________________ __________________________________________________________________
Telephone_______________________________________________ Ext. __________________________________________________________________
Serial Number __________________________________ Date purchased __________________________________________________________________
Fold and mail or e-mail to:
Editor, Technical Publications
Thermo Fisher Scientific S.p.A.
Strada Rivoltana km 4
20090 Rodano (MI)
Italy
Editor, Technical Publications
Thermo Fisher Scientific SID GC-GC/MS
2215 Grand Avenue Parkway
Austin TX 78728-3812
Unites States of America
Declaration
Manufacturer: Thermo Fisher Scientific
Thermo Fisher Scientific is the manufacturer of the instrument described in this manual and, as such, is responsible
for the instrument safety, reliability and performance only if:
•installation
•re-calibration
•changes and repairs
have been carried out by authorized personnel and if:
•the local installation complies with local law regulations
•the instrument is used according to the instructions provided and if its operation is only entrusted to qualified
trained personnel
Thermo Fisher Scientific is not liable for any damages derived from the non-compliance with the aforementioned
recommendations.
Thermo Fisher Scientific S.p.A.
Strada Rivoltana, 20090 Rodano - Milan - Italy — Tel: +39 02 950591 - Fax: +39 02 9505276
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable domestic and international regulations.
Thermo Fisher Scientific declares, under sole responsibility, that the product as originally delivered complies with
the requirements of the following applicable European Directives and carries the CE marking accordingly:
•Low Voltage Directive:2006/95/EC
•EMC Directive:2004/108/EC
•Machinery Directive: 2006/42/EC
… and conforms with the following product standards:
• EN 61000-6-2:2005 | IEC 61000-6-2:2nd Edition| IEC 61000-6-3:2nd Edition am1 | EN 61000-6-3:2007 +
A1:2011
• Conducted Emission, Subpart B. FCC part 15, §15.107(a) and §15.109(a)
Laser Class 1
The selected Class 1 Laser for the TriPlus RSH module Barcode Reader complies with the following regulations:
•21 CFR1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001
•EN60825-1:1994 + A1:2002 + A2:2001
•IEC60825-1:1993 + A1:1997 + A2:2001
FCC Compliance Statement
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO THE
FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED,
INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained
inside this manual pertaining to the safe use and operation of this product before using the device.
Notice on Lifting and Handling of
Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher
Scientific instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/
or bulky for one person alone to handle safely.
Notice on the Proper Use of
Thermo Scientific Instruments
In compliance with international regulations: Use of this instrument in a manner not specified by Thermo Fisher
Scientific could impair any protection provided by the instrument.
Notice on the Susceptibility
to Electromagnetic Transmissions
Do not use radio frequency transmitters, such as mobile phones, in close proximity to the instrument.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive
2012/19/EU. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU)
Member State, and these companies should dispose of or recycle this product. See www.thermoscientific.com/rohsweee
further information on Thermo Fisher Scientific’s compliance with these Directives and the recyclers in your country.
for
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2012/19/EU erfüllen. Das Produkt
ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-Mitgliedsstaaten getroffen,
damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Mehr Information über die Einhaltung
dieser Anweisungen durch Thermo Fisher Scientific, über die Verwerter, und weitere Hinweise, die nützlich sind, um die
Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter www.thermoscientific.com/rohsweee
.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2012/19/EU) des Déchets d'Equipements Electriques et
Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union
européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la conformité de Thermo
Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les produits Thermo Fisher Scientific qui
peuvent aider la détection des substances sujettes à la directive RoHS sont disponibles sur www.thermoscientific.com/rohsweee
.
Conformità RAEE
Questo prodotto è marcato con il seguente simbolo in conformità alla direttiva europea 2012/19/EU (RAEE) sui rifiuti di
apparecchiature elettriche ed elettroniche:
Thermo Fisher Scientific si è accordata con una o più società di riciclaggio in ciascun Stato Membro della Unione Europea
(EU), e queste società dovranno smaltire o riciclare questo prodotto. Per maggiori informazioni vedere il sito
www.thermoscientific.com/rohsweee
.
Conformidad RAEE
Este producto es marcado con el siguiente símbolo en conformidad a la Directiva 2012/19/EU de la Unión Europea sobre los
residuos de aparatos eléctricos y electrónicos:
Thermo Fisher Scientific ha contratado una o más empresas de reciclo para tratar residuos en cada Estado Miembro de la
Unión Europea, y estas empresas deberían reciclar o eliminar este producto. Referirse a www.thermoscientific.com/rohsweee
para una mayor información sobre la conformidad de Thermo Fisher Scientific con estas Directivas y para las empresas de
reciclaje en su país.
IMPORTANT
The symbol indicates the product must not be disposed of with the normal household wastes. Correct disposal of
this product prevents any potentially negative impact on the environmental and human health that could arise
from any inappropriate handling of the product itself.
WEEE and RoHS rules, while laid down at European level, are put into national law at national level. When exporting to Europe, it is essential to comply with national law in each relevant country. The EU law simply serves
as a template for national laws, which may differ considerably.
Each EU Member State has own regulations regarding the application of these directives. Please refer to the
regulations in force in your country.
This manual details informations for installing, defining object, interfacing, maintaining, and
troubleshooting the TriPlus RSH.
This manual is organized as follows:
• Chapter 1, “Specifications,” provides general specifications of the TriPlus RSH.
• Chapter 2, “Installation of the TriPlus RSH System,” provides informations for installing
the TriPlus RSH System.
• Chapter 3, “TriPlus RSH Defining Object Positions,” provides the instructions for
defining the teaching position of he TriPlus RSH modules, the referencing position and
the calibration of the tools. The definition of these terms, and the Row and Column
sample positions, are also explained.
• Chapter 4, “TriPlus RSH Firmware,” provides instructions for creating, copying, and
restoring backup files, and for upgrading and downgrading the Control Software.
• Chapter 5, “Maintenance,” provides guidelines for maintaining the TriPlus RSH and its
modules.
• Chapter 6, “Troubleshooting,” gives a quick overview of possible causes and
recommended actions which can be taken to eliminate an erratic behavior.
Thermo ScientificTriPlus RSH Hardware Manualxiii
Preface
About Your System
About Your System
Thermo Fisher Scientific systems operate safely and reliably under carefully controlled
environmental conditions. If the equipment is used in a manner not specified by the
manufacturer, the protections provided by the equipment may be impaired. If you maintain a
system outside the specifications listed in this guide, failures of many types, including personal
injury or death, may occur. The repair of instrument failures caused by operation in a manner
not specified by the manufacturer is specifically excluded from the Standard Warranty and
service contract coverage.
Power Rating
Tr iP l us RSH
• 100/240 Vac +/-10%; 50/60 Hz; 5 A max; 200 VA (400 VA when two Power Modules
are required).
Detailed instrument specifications are in the Product Specification or Product Brochure.
Contacting Us
Thermo Fisher Scientific provides comprehensive technical assistance worldwide and is
dedicated to the quality of our customer relationships and services.
Use http://www.thermoscientific.com address for products information.address for products
information.
Use http://www.gc-gcms-customersupport.com/WebPage/Share/Default.aspx address to
contact your local Thermo Fisher Scientific office or affiliate GC-GC/MS Customer Support.
Related Documentation
In addition to this guide, Thermo Scientific provides the following documents for the TriPlus
RSH.
• Tri Pl us R SH Ha rdwar e Ma nu al , PN 31709640
To suggest ways we can improve the documentation, follow this link to complete our
documentation survey.
xivTriPlus RSH Hardware ManualThermo Scientific
Safety Alerts and Important Information
Make sure you follow the precautionary notices presented in this manual. The safety and
other special notices appear in boxes.
Special Notices
Notices includes the following:
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Emphasizes important information about a task.
Tip Helpful information that can make a task easier.
Preface
Safety Alerts and Important Information
Safety Symbols and Signal Words
All safety symbols are followed by WARNING or CAUTION, which indicates the degree of risk
for personal injury, instrument damage, or both. Cautions and warnings are following by a
descriptor, such as BURN HAZARD. A WARNING is intended to prevent improper actions that
could cause personal injury. Whereas, a CAUTION is intended to prevent improper actions
that might cause personal injury, instrument damage, or both. You can find the following
safety symbols on your instrument, or in this guide:
SymbolDescriptor
BIOHAZARD: Indicates that a biohazard will, could, or might occur.
BURN HAZARD: Alerts you to the presence of a hot surface that could or might
cause burn injuries.
ELECTRICAL SHOCK HAZARD: Indicates that an electrical shock could or
might occur.
FIRE HAZARD: Indicates a risk of fire or flammability could or might occur.
EXPLOSION HAZARD. Indicates an explosion hazard. This symbol indicates
this risk could or might cause physical injury.
FLAMMABLE GAS HAZARD. Alerts you to gases that are compressed,
liquefied or dissolved under pressure and can ignite on contact with an
ignition source. This symbol indicates this risk could or might cause physical
injury.
Thermo ScientificTriPlus RSH Hardware Manualxv
Preface
Safety Alerts and Important Information
GLOVES REQUIRED: Indicates that you must wear gloves when performing a
task or physical injury could or might occur.
CLOTHING REQUIRED. Indicates that you should wear a work clothing when
performing a task or else physical injury could or might occur.
BOOTS REQUIRED. Indicates that you must wear boots when performing a
task or else physical injury could or might occur.
MATERIAL AND EYE HAZARD. Indicates you must wear eye protection when
performing a task.
HAND AND CHEMICAL HAZARD: Indicates that chemical damage or physical
injury could or might occur.
HARMFUL. Indicates that the presence of harmful material will, could, or
might occur.
INSTRUMENT DAMAGE: Indicates that damage to the instrument or
component might occur. This damage might not be covered under the
standard warranty.
LIFTING HAZARD. Indicates that a physical injury could or might occur if two
or more people do not lift an object.
MATERIAL AND EYE HAZARD: Indicates that eye damage could or might
occur.
READ MANUAL: Alerts you to carefully read your instrument’s
documentation to ensure your safety and the instrument’s operational ability.
Failing to carefully read the documentation could or might put you at risk for a
physical injury.
TOXIC SUBSTANCES HAZARD: Indicates that exposure to a toxic substance
could occur and that exposure could or might cause personal injury or death.
LASER HAZARD. Indicates that exposure to a laser beam will, could, or might
cause personal injury.
RADIOACTIVE HAZARD. Indicates that the presence of radioactive material
could or might occur.
For the prevention of personal injury, this general warning symbol precedes
the WARNING safety alert word and meets the ISO 3864-2 standard. In the
vocabulary of ANSI Z535 signs, this symbol indicates a possible personal
injury hazard exists if the instrument is improperly used or if unsafe actions
occur. This symbol and another appropriate safety symbol alerts you to an
imminent or potential hazard that could cause personal injury.
xviTriPlus RSH Hardware ManualThermo Scientific
Instrument Markings and Symbols
3
Ta b le 1 explains the symbols used on Thermo Fisher Scientific instruments. Only a few of
them are used on the TriPlus RSH. See the asterisk.
Table 1 . Instrument Marking and Symbols (Sheet 1 of 2)
SymbolDescription
Direct Current
*Alternating Current
Both direct and alternating current
Three-phase alternating current
Earth (ground) terminal
Protective conductor terminal
Preface
Instrument Markings and Symbols
Frame or chassis terminal
Equipotentiality
*On (Supply)
*Off (Supply)
Equipment protected throughout by DOUBLE INSULATION or
REINFORCED INSULATION (Equivalent to Class II of IEC 536)
Fuse
*Instruction manual symbol affixed to product. Indicates that the you must
refer to the manual for specific WARNING or CAUTION information to
avoid personal injury or damage to the product.
Caution, risk of electric shock
*Caution, hot surface
*Caution, biohazard
*Caution, Laser beam
Thermo ScientificTriPlus RSH Hardware Manualxvii
Preface
Safety Information and Warnings
Table 1 . Instrument Marking and Symbols (Sheet 2 of 2)
SymbolDescription
*Symbol in compliance to the Directive 2012/19/EU on Waste Electrical
and Electronic Equipment (WEEE) placed on the European market after
August, 13, 2005.
Safety Information and Warnings
This safety guide raises awareness of potential safety issues and general points for
consideration for Thermo Fisher Scientific representatives during installation, and repair of
the TriPlus RSH, or parts of it (following the life cycle principle), as well as for the end user
TriPlus RSH in the lab during the learning phase, and in routine work.
IMPORTANT Read this section first before operating the TriPlus RSH.
General Considerations
• Before a unit is put to use, consult the TriPlus RSH User Guide and related documents
under all circumstances.
• Changes or modifications to this unit not expressly approved by the party responsible for
compliance, could void your’s authority to operate the equipment.
• Be aware that if the equipment is used in a manner not specified by the manufacturer, the
protective and safety features of the equipment might be impaired.
• The repair of instrument failures caused by operation in a manner not specified by the
manufacturer is expressly excluded from the standard warranty and service contract
coverage.
• When for technical reasons it is necessary to work on instrument parts which might
involve a potential hazard (moving parts, components under voltage, and so on.) contact
the Thermo Fisher Scientific authorized representative. In general, this type of situation
arises when access to the parts is only possible using a tool. When you perform a
maintenance operation, you must have received proper training to carry out that specific
task.
Electrical Hazards
Every analytical instrument has specific hazards. Be sure to read and comply with the
following pre-cautions. They ensure the safe and long-term use of your TriPlus RSH.
xviiiTriPlus RSH Hardware ManualThermo Scientific
Preface
Safety Information and Warnings
The installation over-voltage category is Level II. The Level II category pertains to equipment
receiving its electrical power from the local level, such as an electrical wall outlet.
Connect the TriPlus RSH only to instruments complying with IEC 61010 safety regulations.
The power line and the connections between the TriPlus RSH and other instruments, used in
the configuration setup of the total analytical system, must maintain good electrical
grounding. Poor grounding represents a danger for the operator, and might seriously affect the
performance of the instrument.
Do not connect the TriPlus RSH to power lines that supply devices of a heavy duty nature,
such as motors, refrigerators and other devices that can generate electrical disturbances.
Use only fuses of the type and current rating specified. Do not use repaired fuses, and do not
short-circuit the fuse holder. The supplied power cord must be inserted into a power outlet
with a protective earth (ground) contact. When using an extension cord, make sure that the
cord also has an earth contact.
If the supplied power cord does not fit the local electrical socket and a replacement or adapter
has to be purchased locally, make sure that only a certified power cord is used. Any power cord
used must be certified by the appropriate local authorities.
Pay attention not to leave any cable connecting the TriPlus RSH and the chromatographic
system, or the power cord close to heated zone, such as the injector or detector heating blocks,
or the GC hot air vents.
Always replace any cable showing signs of damage with another one provided by the
manufacturer. Safety regulations must be respected.
Do not change the external or internal grounding connections. Tampering with or
disconnecting these connections could endanger you and damage the TriPlus RSH. The
instrument is properly grounded in accordance with these regulations when shipped. To
ensure safe operation, you do not must make any changes to the electrical connections or the
instrument's chassis.
Do not turn the instrument on if you suspect that it has incurred any type of electrical
damage. Instead, disconnect the power cord and contact a Thermo Fisher Scientific
representative for a product evaluation. Do not attempt to use the instrument until it has been
evaluated. Electrical damage might have occurred if the TriPlus RSH shows visible signs of
damage, exposure to any liquids or has been transported under severe stress.
Damage can also result if the instrument is stored for prolonged periods under unfavorable
conditions: for example, subjected to heat, moisture, and so on. Ensure that the power
supply/controller unit is always placed in a clean and dry position. Avoid any liquid spills in
the vicinity.
Thermo ScientificTriPlus RSH Hardware Manualxix
Preface
Safety Information and Warnings
Before attempting any type of maintenance work, always disconnect the power cords from the
power supply(ies) if optional devices are installed. Capacitors inside the instrument might still
be charged also if the instrument is turned off.
To avoid damaging electrical parts, do not disconnect an electrical assembly while power is
applied to the TriPlus RSH. After the power is turned off, wait approximately 30 seconds
before you disconnect an assembly.
The instrument includes a number of integrated circuits. These circuits might be damaged if
exposed to excessive line voltage fluctuations, power surges or electrostatic charges, or both.
Never try to repair or replace any components of the instrument without the assistance of a
Thermo Fisher Scientific representative. There are no operator-serviceable or replaceable parts
inside the power supply(ies) or in the TriPlus RSH. If a power supply is not functioning,
contact a Thermo Fisher Scientific representative.
The power supplies for the TriPlus RSH, the Temperature Controlled Drawer have the
symbols I/O on the label for the power switch to indicate ON/OFF. If a Temperature
Controlled Drawer is installed in combination with a TriPlus RSH, a second power supply is
active in the complete system. Turning OFF the two power supplies, or pulling the two power
cords in an emergency, stop the entire TriPlus RSH.
It is important that the power supply(ies) is in a location where the power ON/OFF switch is
accessible and easy to operate, and where it is possible to unplug the AC power cord from the
power supply/wall outlet in case of emergency.
Laser Safety Information
Safety Warning for Laser Class 1 Product.
WARNING The installed Laser device is a Class 1 Laser Product.
Class 1 Laser devices are not considered to be hazardous when used for their intended
purpose. The following statement is required to comply with US and international
regulations.
CAUTION Use of controls, adjustments or performance of procedures other than those
specified herein might result in hazardous laser light exposures.
The selected Class 1 Laser for the TriPlus RSH module Barcode Reader complies with the
following regulations:
• 21 CFR1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50,
dated July 26, 2001
xxTriPlus RSH Hardware ManualThermo Scientific
Other Hazards
Preface
Safety Information and Warnings
• EN60825-1:1994 + A1:2002 + A2:2001
• IEC60825-1:1993 + A1:1997 + A2:2001
The software contains a built-in safety time limit such that the laser scanning mechanism
cannot be operated in AIM mode for more than 5 continuous seconds.
To avoid injury and possible infection through contamination during TriPlus RSH operation,
keep your hands away from the syringe.
Do not operate the TriPlus RSH without the safety guard. The safety guard must be installed
for safe operation. Do not place any objects inside the area of the safety guard. Keep away
from the area around the safety guard during operation of the TriPlus RSH.
Danger of crushing to fingers and hands. To avoid injury keep your hands away from moving
parts during operation. Turn off the power to the TriPlus RSH if you must reach inside a
mechanically powered system with moving parts.
To avoid injury, observe safe laboratory practice when handling solvents, changing tubing, or
operating the TriPlus RSH. Know the physical and chemical properties of the solvents you
use. See the MSDS (Material Safety Data Sheets) from the manufacturer of the solvents being
used.
When using the TriPlus RSH, follow the generally accepted procedures for quality control
and method development.
When using the TriPlus RSH in the field of chromatographic analysis, if a change is observed
in the retention of a particular compound, in the resolution between two compounds, or in
the peak shape, immediately determine the reasons for the changes. Do not rely on the
separation results until you determine the cause of a change.
Do not operate on the instrument components that form part of the work area of the TriPlus
RSH when it is in motion.
Use caution when working with any polymer tubing under pressure:
• Always wear eye protection when near pressurized polymer tubing.
• Do not use polymer tubing that has been severely stressed or kinked.
• Do not use polymer tubing, in particular no PEEK or Tefzel tubing when using
tetrahydrofuran (THF), dimethylsulfoxide (DMSO), chlorinated organic solvents,
concentrated mineral acids such as nitric, phosphoric or sulfuric acids, or any related
compounds.
Do not use vials without a sealing cap, or microtiter or deepwell plates without a plate seal.
Vapor phase from organic solvents can be hazardous and flammable. Acidic vapor phase can
cause corrosion to critical mechanical parts.
Thermo ScientificTriPlus RSH Hardware Manualxxi
Preface
Safety Information and Warnings
When sample vials have to undergo heating and agitation, it is important to consider the glass
quality. Use high quality glass only. Remember that depending on the application conditions,
high pressure can build up in the vial. Whenever a temperature greater than 60 °C is applied,
consider the vapor pressure of the solvent used to ensure that no excessive pressure builds up.
This is important when using a temperature above 100°C and especially at the maximum
temperature of 200 °C. Be aware that solid materials can also contain volatile compounds
such as water (humidity) which could cause build-up of excess vapor pressure.
Do not reuse headspace vials. During the process of washing the vial, micro-cracks can form
which will weaken the glass wall and increase the chances of the vial breaking.
In case of a single fault situation where the temperature control of the Agitator fails, there is
the potential danger that the device will heat up in an uncontrolled manner until it reaches
the cut-off temperature of the over temperature fuse, in this case, 240 °C. Based on this single
fault scenario, when working with flammable solvents, ensure that the solvent used has a flash
point which is 25 °C higher than the maximum potential temperature (240 °C) of the
Agitator.
When filling-up a standard reservoir or replacing a solvent such as a washing solvent, remove
the solvent reservoir bottle from the system to avoid a possible spill over the instrument.
Depending on the physical, chemical or hazardous properties of the solvent, use the
appropriate protective measures for handling.
Working with Toxic or other Harmful Compounds
WARNING Before using hazardous substances (toxic, harmful, and so on), please read the
hazard indications and information reported in the applicable Material Safety Data Sheet
(MSDS). Use personal protective equipment according to the safety requirements.
Before using dangerous substances (toxic, harmful, and so on) read the hazard indications and
information reported in the Material Safety Data Sheet (MSDS) supplied by the
manufacturer, referring to the relevant CAS (Chemical Abstract Service) number. The TriPlus
RSH requires the use of several chemical products with different hazard characteristics, which
are present in vials and syringes. Before using these substances or replacing the syringe, please
read the hazard indications and information reported in the MSDS supplied by the
manufacturer referring to the relevant CAS number.
When preparing the samples, please refer to local regulations for the ventilation conditions of
the work room.
All waste materials must be collected and eliminated in compliance with the local regulations
and directives in the country where the instrument is used.
xxiiTriPlus RSH Hardware ManualThermo Scientific
Biological Hazards
Preface
Safety Information and Warnings
In laboratories where samples with potential biological hazards are handled, you must label
any equipment or parts thereof which might become contaminated with biohazardous
material. The appropriate warning labels are included with the shipment of the instrument.
It is your responsibility to label the relevant parts of the instrument.
When working with biohazardous materials, it is your responsibility to fulfill the following
mandatory requirements:
• Instructions on how to safely handle biohazardous material must be provided.
• Operators must be trained and made aware of the potential dangers.
• Personal protective equipment must be provided.
• Instructions must be provided on what to do in case operators are exposed to aerosols or
vapors during normal operation (within the intended use of the equipment) or in case of
single fault situations such as a broken vial.
The protective measures must consider potential contact with the skin, mouth, nose
(respiratory organs), and eyes.
Maintenance
• Instructions for decontamination and safe disposal of the relevant parts must be provided.
It is your responsibility to handle hazardous chemicals or biological compounds (including,
but not limited to, bacterial or viral samples and the associated waste), safely and in
accordance with international and local regulations.
Any external cleaning or maintenance must be performed with the TriPlus RSH turned off
and the power cord disconnected. Avoid using solvents and spraying on electrical parts. For
the removal of potentially dangerous substances (toxic, harmful, and so on) read the hazard
indications and information reported in the MSDS (Material Safety Data Sheet) supplied by
the manufacturer referring to the relevant CAS (Chemical Abstract Service) number. Use
proper protective gloves.
When working with hazardous materials such as radioactive, biologically hazardous material,
and so on, it is important to train all operators how to respond in case of spills or
contamination.
Depending on the class of hazardous material, the appropriate measures have to be taken
immediately. Therefore, the chemicals or solvents needed for decontamination have to be on
hand.
Any parts of the equipment which can potentially be contaminated, such as the sample vial
rack, syringe tool, wash module, and so on, must be cleaned regularly. The waste solvent from
cleaning and any hardware which requires to be disposed of has to be properly eliminated
with all the necessary precautions, abiding by national and international regulations.
Thermo ScientificTriPlus RSH Hardware Manualxxiii
Preface
Safety Information and Warnings
Disposal
When preparing for decontamination, ensure that the solvent or chemical to be used will not
damage or react with the surface, dye (color) of the instrument, table or other nearby objects.
If in doubt, please contact your Thermo Fisher Scientific representative to verify the
compatibility of the type or composition of solvents with the TriPlus RSH.
Do not dispose of this equipment or parts thereof unsorted in municipal waste.
Follow local municipal waste regulations for proper disposal provisions to reduce
the environmental impact of waste electrical and electronic equipment (WEEE).
European Union customers: Call your local customer service representative responsible for the
TriPlus RSH for complimentary equipment pick-up and recycling.
WARNING The customer has to ensure that the TriPlus RSH has not been contaminated
by any hazardous chemical or biological compounds including (but not limited to)
bacteria or viruses.
Any part which had direct contact with the analytical sample must be identified and must
undergo an appropriate decontamination procedure prior to shipping for disposal.
Potentially dangerous components are: Syringes, Vials and Well Plates. Any critical parts
sent for disposal must be handled according to national laws for hazardous compounds.
The customer and the service engineer are fully responsible for enforcing these
requirements. Thermo Fisher Scientific will hold the representative, customer responsible,
or both, if these regulations are not observed.
xxivTriPlus RSH Hardware ManualThermo Scientific
1
Specifications
This section provides general specifications of the TriPlus RSH.
Contents
• Electrical Specifications
• Physical Specifications
• Operating and Environmental Specifications
• Sound Pressure Level
• Hardware and Software Requirements
• TriPlus RSH System Software Requirements
Thermo ScientificTriPlus RSH Hardware Manual1
1
Specifications
Electrical Specifications
Electrical Specifications
The electrical specifications and the various protection classes are provided in Ta bl e 1 and
Ta b le 2 .
Table 1 . General Electrical Specifications
ParameterRequirements
Voltage36 VDC
Current3.2 A
FuseT6.3 A/250 V
Table 2 . Handheld Controller Power Supply
ParameterRequirements
Input line voltageGrounded AC, 100 V to 240 V
Input line frequency50/60 Hz
Input power4 A
Output voltage36 VDC
Output current5.55 A (total for two Outlets)
• Protection Class I — Describes the insulating scheme used in the instrument to protect
the user from electrical shock. Class I identifies a single level of insulation between live
parts (wires) and exposed conductive parts (metal panels), in which the exposed
conductive parts are connected to a grounding system. In turn this grounding system is
connected to the third pin (ground pin) on the electrical power plug.
• Over Voltage Category II — Pertains to instruments that receive their electrical power
from a local level such as an electrical wall outlet.
• Pollution Degree 2 — Measure of pollution on electrical circuits that may produce a
reduction of the dielectric strength or surface resistivity. Degree 2 normally refers only to
non-conductive pollution.Occasionally, however, a temporary conductivity caused by
condensation is to be expected.
• Moisture Protection — Normal (IPXO) – PIXY means that there is NO Ingress
Protection against any type of dripping or sprayed water. The X is a place holder to
identify protection against dust, if applicable.
2TriPlus RSH Hardware ManualThermo Scientific
Physical Specifications
The physical specification for the TriPlus RSH with Standard are listed in Tab l e 3.
Table 3 . Space and Load Requirements
1
Specifications
Physical Specifications
Instrument
TriPlus RSH standard X axis 46
TriPlus RSH extended X-axis 46
TriPlus RSH standard X axis
Depth (Y-axis)Width (X-axis)Height (Z-axis)Mass
cmin.cmin.cmin.kglbs
80
1
1
1
1
18
18
1
8533.55421.313.3
120.5475421.315.3
31.599
2
39
2
(Working Range)
TriPlus RSH extended X-axis
80
1
31.5135
2
53
2
(Working Range)
1
About 20 cm (about 8-in.) of the orthogonal crossrail (Y-axis) are protruding the back of the GC.
2
Dimension of the entire working range including Handheld bracket
3
Dimension including the support legs [about 22 cm (8.7 in.) height]
4
Mass without accessories.
Operating and Environmental Specifications
These specifications are listed in Ta b le 4 .
53 (74)
53 (74)
3
3
20.9 (29)
20.9 (29)
3
3
4
29.3
4
33.7
2555
2760
4
4
Table 4 . Operating and Environmental Specifications
ParameterRequirements
Operating temperature range5 °C to 40 °C (41 °F to 104 °F)
Non-operating temperature range (transport or storage conditions) -25 °C to +65 °C (-13 °F to + 149 °F)
Maximum relative humidity80%, non-condensing
Work environmentIndoor use
AltitudeUp to 3000 m above sea level
Mains voltage variation±10%
Bench spaceAt least 30 cm (12 in.) at the back.
Access to power switches and power cords.Clean, level, and smooth surface.
Clean, level and smooth surface.Solid bench plate.
VibrationNegligible
Static electricityNegligible
Thermo ScientificTriPlus RSH Hardware Manual3
1
Specifications
Sound Pressure Level
Sound Pressure Level
The sound pressure level is listed in Ta b le 5 .
Table 5 . Sound Pressure Level
ParameterRequirements
Sound Pressure Level Measured value:
• TriPlus RSH (basic unit): 64 dBA
• TriPlus RSH with Agitator: 66 dBA
• TriPlus RSH with Vortexer without sample vial: 67
dBA
• TriPlus RSH with Vortexer with 20 ml sample vial
inserted: 78 dBA.
One meter from the equipment in the direction of
maximum sound pressure level.
According to UL 610107A-1, 1st Edition, clause 12.5. Limit < 85 dBA
dBA = A weighted sound pressure level
Hardware and Software Requirements
This section details the hardware and software requirements.
Computer Hardware Requirements
The computer hardware minimal requirements depends on the Microsoft Windows
Operating System and, if applicable, on the MS.NET Framework software used. See
recommendations from Microsoft.
It is also important to consider the minimum requirements for the operation of the
Chromatography Data System Software (CDS).
For operation of the TriPlus RSH, both an Ethernet (TCP/IP) and a USB communication
port are required. The computer in use must be equipped accordingly: LAN or Ethernet IEEE
802.3 Industry Standard 10/100 Base-T. The network interface of the TriPlus RSH supports
the Auto-MDI(X) protocol.
The USB Device/Client Type B communication port of the TriPlus RSH is currently not
activated.
4TriPlus RSH Hardware ManualThermo Scientific
USB Stick (Memory Drive)
Use a commercially available standard USB memory stick with at least 100 MB memory size.
No special formatting of the drive (stick) is necessary; on the contrary only one partition must
be defined. If two partitions are defined, subsequent errors occur, and the access to the system
is denied.
Note Commercially available USB sticks with encrypting software installed, and secure
USB drives, typically have two partitions. This type of USB drive cannot be utilized in
combination with the TriPlus RSH system.
The following software types of the USB memory drive are compatible with the TriPlus RSH:
•FAT
•FAT32
• The software NTFS is NOT supported. If a USB Stick with this software is used, the stick
will not be recognized by the RSH Firmware.
1
Specifications
Hardware and Software Requirements
• Starting with Firmware version 1.5.0 a USB Stick with software version 3.0 can be
operated. With Firmware versions lower than 1.5.0, USB sticks can be operated with
software version 2.X.X only.
• It is recommended to adjust the Allocation Unit Size as large as possible. There is a
dependency observed between the Capacity Size and the size of the Allocation Unit.
Examples: USB Stick Capacity 4 GB with 4 MB Allocation Unit, or 16 GB Capacity
with 64 MB Allocation Units. These combinations are valid; however a 16 GB Capacity
with 8 MB Allocation Units is not working. The observed effect is that the USB Stick is
not recognized within the 10 seconds time limit.
Note The Recognition Time is time used from the moment the stick is connected to
the TriPlus RSH System until the USB symbol is displayed on the handheld controller
(terminal).
Recognition Time of the USB Stick is:
• Firmware version < 1.5.0:Time-out after 10 s.
• Firmware version ≥ 1.5.0: Time-out after 30 s.
PC Software Requirements
All the software functionality such as Backup, Restore, or Update for the RSH Firmware are
performed directly via USB interface using the USB stick. For these purposes, no specific
software on a PC is required.
Thermo ScientificTriPlus RSH Hardware Manual5
1
Specifications
Hardware and Software Requirements
For the operation of the TriPlus RSH with a Thermo Scientific Chromatography Data System
(CDS), for example: Chromeleon, Xcalibur, ChromQuest, or Chrom-Card, a software driver
is used. The specific requirements for operation of the integrated TriPlus RSH System in the
CDS are outlined by the supplier of the CDS software.
For the operation of the TriPlus RSH in combination with Sample Control, the software from
Microsoft MS.net framework version 4.0 (or higher) is required.
Configuring Windows for Firmware and for the CDS
When configuring the Windows Operating software for the TriPlus RSH, it is important to
note that the CDS to be used in conjunction with the TriPlus RSH might require specific
settings as well. These settings have to be matched for reliable operation of the analytical
system. See the corresponding sections in the installation manual of the specific CDS.
Firewall or Virus Scanner Settings
In order to reliably operate the TriPlus RSH and the selected Chromatography Data System
(CDS), it is necessary to turn OFF any third party Firewall or Virus Scanner software.
If for any reason a firewall must be activated; the TCP Ports 80, 64000, 64001, and 8194
must be unlocked.
Screen Savers
It is highly recommended NOT to use any Screen Savers.
Disabling Power Management
Depending on the Windows Operating System version and the default settings, the computer
may enter standby mode automatically after a specified time if no mouse or keyboard activity
is detected. Using this Power Save mode in the environment with laboratory instruments
where unattended operation day and night is expected, will likely result in loss of data,
interruptions of specific functions, or whole operations.
To disable Windows power management
1. Select the Windows Control Panel
2. Select System and Security
3. Select Power Options
4. From Power Plan select the menu item Put the Computer to sleep: Never
Note Depending on the Windows version, the path or wording may differ slightly.
The example given above is based on Windows 7.
6TriPlus RSH Hardware ManualThermo Scientific
Loading...
+ 156 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.