Thermo Products REDOX-ORP 9678BNWP User Manual

User Guide
Redox/ORP Electrodes
Introduction
This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion ORP electrodes. These ORP electrodes combine a platinum redox sensing electrode and a silver/silver chloride reference electrode in one body.
The redox/ORP electrodes can be used to make direct measurements of redox potentials or perform potentiometric titrations. The ORP Triode electrodes have a built-in thermistor for measuring temperature.
Epoxy Sure-Flow Combination Redox/ORP Electrodes, Cat. No. 9678BNWP and 967800
The Sure-Flow junction prevents the electrode from clogging in dirty, viscous or wastewater samples. The epoxy body is break­resistant for rugged environments.
Glass Combination Redox/ORP Electrodes, Cat. No. 9778BNWP and 9778SC
The glass body enables the electrode to be used in solutions that contain organic solvents. Use the electrode in water, wastewater, metal plating and organic solvent samples.
Epoxy Refillable ORP/ATC Triode, Cat. No. 9180BNMD and 9180BN
The electrode includes a built-in temperature probe for measuring ORP and temperature with one electrode. The refillable design allows the electrode filling solution to be changed in case of contamination or clogging.
Epoxy Low Maintenance ORP/ATC Triode, Cat. No. 9179BNMD and 9179BN
The electrodes include a built-in temperature probe for measuring ORP and temperature with one electrode. The electrodes contain a gel reference solution, so no filling solution is required.
Note: Redox/ORP electrodes are not designed for constant current or amperometric titrations, such as Karl Fischer moisture analysis.
1Redox/ORP Electrode User Guide
Required Equipment
1. Thermo Scientific Orion pH/mV meter, such as the 3-Star pH meter, 4-Star pH/ISE meter, 4-Star pH/DO meter, 4-Star pH/conductivity meter or 5-Star pH/ISE/DO/conductivity meter.
ORP electrodes can be used on any pH/mV meter with a
BNC or U.S. standard connection. The electrodes can also be used on meters with a variety of inputs when an adapter cable is used. Visit www.thermo.com/water for details.
Refillable ORP Triodes have temperature connectors that
are compatible with specific meters, refer to the list below.
2. Thermo Scientific Orion ORP electrode.
The 9180BNMD and 9179BNMD ORP/ATC Triode
electrodes have a temperature connector that is compatible with the Star Series meters.
The 9180BN and 9179BN ORP/ATC Triode electrode
have a temperature connector that is compatible with the A+ Series meters.
3. ORP electrode filling solution. Use Cat. No. 900011 filling solution for samples that have a total ionic strength greater than 0.2 M. Use Cat. No. 900001 filling solution for samples that have a total ionic strength less than 0.2 M.
ORP electrode filling solution is not required if using a
9179BNMD or 9179BN low maintenance ORP/ATC Triode
4. ORP standard, Cat. No. 967901 or 967961.
ORP standard is highly recommended to ensure accurate
ORP measurements.
5. Beakers, plastic or glass.
6. Magnetic stirrer or Orion stirrer probe, Cat. No. 096019. The stirrer probe can be used with 3-Star, 4-Star and 5-Star benchtop meters.
7. Distilled or deionized water.
2 Redox/ORP Electrode User Guide
Sample Requirements
Electrodes with an epoxy body should only be used in aqueous solutions.
Electrodes with a glass body may be used in non-aqueous solutions and solutions that contain organic solvents.
The ORP electrodes contain a silver/silver chloride (Ag/AgCl) reference that may become clogged in solutions that contain silver complexing or binding agents such as TRIS, proteins and sulfides. Frequent cleaning may be required when measuring solutions that contain these agents. Proteins cause the additional problem of coating the electrode, so extra care should be taken to keep the electrode clean while measuring samples that contain proteins.
3Redox/ORP Electrode User Guide
Electrode Preparation
Electrode Preparation for Refillable Electrodes
1. Remove the protective shipping cap from the sensing element and save the cap for storage.
2. Clean any salt deposits from the exterior of the electrode by rinsing with distilled water.
3. Uncover the filling hole by removing the tape and then add electrode filling solution, Cat. No. 900011, to the electrode. To maintain an adequate flow rate, the level of filling solution must always be above the reference junction and at least one inch above the sample level. The filling hole should be open whenever the electrode is in use.
4. For Sure-Flow electrode only, wet the junction by pressing down on the electrode cap to let a few drops of the filling solution flow out of the electrode and then replenish any lost solution.
5. Shake the electrode downward (similar to a clinical thermometer) to remove air bubbles.
6. Connect the electrode to the meter.
Note: The fill hole must be open when the electrode is in use.
Electrode Preparation for Gel-Filled Electrodes
1. Remove the protective shipping cap from the sensing element and save the cap for storage.
2. Clean any salt deposits from the exterior of the electrode by rinsing with distilled water.
3. Shake the electrode downward (similar to a clinical thermometer) to remove air bubbles.
4. Connect the electrode to the meter.
4 Redox/ORP Electrode User Guide
filling hole
Ag/AgCl
internal
reference
filling solution chamber
ceramic reference junction
sensing element
Figure 1– Refillable ORP Electrode
filling hole
Ag/AgCl
internal
reference
filling solution chamber
ceramic reference junction
sensing element
fill hole
AgAgCl
internal
reference
reference filling solution chamber
Sure-Flow junction
sensing element
Figure 2– Sure-Flow Refillable ORP Electrode