THERMOLYNE FC2020P, FC2020P-1, FC2024P, FC2024P-1, FC2025P Operation Manual

...
THERMOLYNE
Type 2000 Furnace
OPERATION MANUAL
AND PARTS LIST
Series 326
BARNSTEAD/THERMOLYNE
READ ALL WARNINGS, CAUTIONS, AND INSTRUCTIONS CAREFULLY BEFORE OPERATING THIS THERMOLYNE FURNACE.
Record serial number here
January 1990 LT326X1
MODEL NUMBERS FC2020P & FC2020P-1 FC2024P & FC2024P-1 FC2025P & FC2025P-1 FC2028P & FC2028P-1
January 1990 LT326X1
IMPORTANT INFORMATION
This manual contains important operating and safety information. The user must carefully read and
understand the contents of this manual prior to the use of this equipment.
TABLE OF CONTENTS
SAFETY INFORMATION ....................................................................................................... 1
SIGNAL WORDS.................................................................................................................... 1
WARNINGS.............................................................................................................................. 1
INTRODUCTION ...................................................................................................................... 2
GENERAL SPECIFICATIONS ............................................................................................. 5
UNPACKAGING ..................................................................................................................... 7
INSTALLATION........................................................................................................................ 7
OPERATION............................................................................................................................. 8
PREVENTIVE MAINTENANCE............................................................................................ 9
TROUBLE SHOOTING GUIDE.............................................................................................10
MAINTENANCE AND SERVICING.....................................................................................12
EXPLODED VIEW ...................................................................................................................18
REPLACEMENT PARTS LIST..............................................................................................19
SCHEMATICS .........................................................................................................................21
ORDERING PROCEDURES.................................................................................................22
WARRANTY INFORMATION...............................................................................................22
A
SAFETY INFORMATION
Your Thermolyne furnace has been designed with function, reliability and safety in mind. It is the user’s responsibility to insure conformance with required electrical codes. For safe operation, please observe the following:
SIGNAL WORDS
"DANGER" notes apply when there are hazards which result in severe personal injury or death.
"WARNING" notes apply when there is a possibility of personal injury.
"CAUTION" notes apply when there is a possibility of damage to the equipment.
"NOTES" alert the user of the manual to pertinent facts and conditions.
WARNINGS
WARNING
TO AVOID ELECTRICAL SHOCK, THIS FURNACE MUST:
1) BE INSTALLED BY A COMPETENT, QUALIFIED ELECTRICIAN WHO INSURES COMPATIBILITY AMONG FURNACE SPECIFICATIONS, POWER SOURCE AND GROUNDING CODE REQUIREMENTS.
2) ALWAYS BE DISCONNECTED FROM THE POWER SUPPLY PRIOR TO MAINTENANCE AND SERVICING.
3) HAVE THE DOOR SWITCH OPERATING PROPERLY.
TO AVOID BURNS, THIS FURNACE MUST:
1) NOT BE TOUCHED ON THE EXTERIOR OR INTERIOR FURNACE SURFACES DURING USE OR FOR A PERIOD OF TIME AFTER USE.
TO AVOID PERSONAL INJURY:
1) DO NOT USE IN THE PRESENCE OF FLAMMABLE OR COMBUSTIBLE CHEMICALS. FIRE OR EXPLOSION MAY RESULT; THIS DEVICE CONTAINS COMPONENTS WHICH MAY IGNITE SUCH MATERIALS.
CAUTION: Space unit 3 inches from combustible walls under any condition.
CAUTION: DISCONNECT FROM SUPPLY IF TEMPERATURE EXCEEDS 2000°F.(For Models FC2020P,
FC2025P & FC2028P)
CAUTION: DISCONNECT FROM SUPPLY IF TEMPERATURE EXCEEDS 2150° F.(For Models FC2020P-1,
FC2025P-l & FC2028P-1)
1 Barnstead/Thermolyne 1-800-553-0039
WARNINGS
WARNING
THIS FURNACE CONTAINS REFRACTORY CERAMIC INSULATION WHICH CAN PRODUCE RESPIRABLE FIBERS AND DUST WHEN HANDLED. THESE FIBERS CAN CAUSE IRRITATION AND CAN AGGRAVATE PRE­EXISTING RESPIRATORY DISEASE. THE INTERNATIONAL AGENCY FOR RESEARCH ON CANCER (IARC) HAS CLASSIFIED REFRACTORY CERAMIC FIBER AS POSSIBLY CARCINOGENIC. AFTER SERVICE REFRACTORY CERAMIC FIBER DUSTS MAY CONTAIN CRYSTALLINE SILICA, WHICH MAY CAUSE LUNG DAMAGE (SILICOSIS) AND WHICH HAS BEEN CLASSIFIED BY IARC AS A PROBABLE CARCINOGEN. CONSULT THE OWNER’S MANUAL FOR THIS EQUIPMENT BEFORE YOU MAKE ANY REPAIRS TO OR REPLACE ANY PARTS, INCLUDING REPAIRS TO OR REPLACEMENTS OF ELECTRICAL COMPONENTS OR INSULATING MATERIALS. THE OWNER’S MANUAL CONTAINS INFORMATION AND WARNINGS RELATING TO THE HAZARDS ASSOCIATED WITH THE REPAIR AND MAINTENANCE OF THIS EQUIPMENT.
WARNING
REFER SERVICING TO QUALIFIED PERSONNEL.
A. Intended Use
The Type 2000 furnaces are general laboratory and heat treating furnaces. Standard models are intended for applications requiring temperatures from 400°F (204°C) to 1800°F (982°C) for continuous use, or temperatures from 1800°F (982°C) to 2000°F (1093°C) for intermittent use. High temperature (-1) models are intended for applications requiring temperatures from 400°F (204°C) to 1950°F (1066°C) for continuous use, or temperatures from 1950°F (1066°C) to 2150°F (1177°C) for intermittent use. Continuous use is operating the furnace for more than 3 hours and intermittent use is operating the furnace for less than 3 hours.
The unit consists of 1) a heating chamber, 2) an automatic proportioning temperature control, 3) a two­position power switch. See Figure 1 for the overall shape and general features of the unit.
B. Principles of Operation
The furnace chamber is heated by four electric resistance heaters which are embedded In a refractory material. The chamber is insulated with a ceramic fiber insulation. The temperature is controlled by an automatic proportioning controller using a chromel/ alumel thermocouple. The control unit is located under the furnace chamber well Insulated from the furnace chamber.
This furnace is equipped with thermocouple break protection, a door safety switch, and cold junction compensation. The silicon controlled rectifiers which control power to the heating elements are controlled by internal circuitry to operate as zero switching devices. This zero switch capability prevents line interference which can effect other sensitive equipment.
INTRODUCTION
2
Figure 1
3 Barnstead/Thermolyne 1-800-553-0039
Figure 2
4
GENERAL SPECIFICATIONS:
FURNACES
Model # FC2020P FC2024P FC2025P FC2028P
Dimensions Overall Width 4
(10)
in. (cm) Height 3-3/4
(10)
Depth 9
(23)
Chamber Width 11
(28)
Height 18-1/4
(47)
Depth 16
(41)
Weight LBS. (KG) 41 (18.6) 41 (18.6) 41 (18.6) 41 (18.6)
Electrical Volts 240 100 120 208
Ratings Amps 9.3 20.7 18.6 10.7
Watts 2240 2070 2240 2240
4
(10)
3-3/4
(10)
9
(23)
11
(28)
18-1/4
(47)
16
(41)
4
(10)
3-3/4
(10)
9
(23)
11
(28)
18-1/4
(47)
16
(41)
4
(10)
3-3/4
(10)
9
(23)
11
(28)
18-1/4
(47)
16
(41)
Freq. 50/60 50/60 50/60 50/60
Phase 1111
Operating
Temperature
Cont.
Inter-
mittent
1800°F
(982°C)
2000°F
(1093°C)
1800°F
(982°C)
2000°F
(1093°C)
1800°F
(982°C)
2000°F
(1093°C)
1800°F
(982°C)
2000°F
(1093°C)
NOTES: Not supplied with cord and plug. Supplied with Chromel/Alumel thermocouples. Continuous use
for standard models is operating at temperatures between 400°F (204°C) and 1800°F (982°C) for more than 3 hours. (-1) models temperatures between 400°F (204°C) and 1950°F (1066°C) for more than 3 hours. Intermittent use for standard models is operating at temperatures between 1800°F (982°C) and 2000°F (1093°C) for less than 3 hours. (-1) models temperatures between 1950°F (1066°C) and 2150°F (1177°C) for less than 3 hours.
5 Barnstead/Thermolyne 1-800-553-0039