LABORATORY INCUBATORS
MODELS INCLUDE:
TI-150F TI-150G TI-250F TI-250G
TI-350F TI-500F
& MSP MULTI SET POINT
P a g e | 2
SAFETY INFORMATION
2
ALERT BOXES
2
INTRODUCTION
3
THE MANUAL
3
UNPACKING
3
LOCATION
3
INTENDED USE
3
FUNCTION
4
PRINCIPLE OF OPERATION
4
PERFORMANCE
4
C-TICK
5
INSTALLATION
5
MAINTENANCE
5
TECHNICAL ASSISTANCE
5
EXPLANATION OF CONTROLS
6
MAIN SWITCH
6
CONTROLS
6
ELECTRICAL
7
OVER TEMPERATURE INDICATION
7
OPERATING THE EQUIPMENT
7
OPERATING GUIDELINES FOR 3216 TIMER/PROGRAMMER
8 CHANGING SET TEMPERATURE
8
OPERATING THE 3216 PROGRAMMER
9 TO 12
CALIBRATION
13 & 14
WARRANTY POLICY
15
IMPORTANT INFORMATION
This manual contains operating and safety information. The operator must read and understand the contents of this manual prior to using this equipment. Incorrect operation may cause harm or damage the severity is classified by the following Alert Boxes. Ensure that this manual is saved for future reference:
Warning alerts apply when there
is a Possibility of personal injury.
Caution alerts apply when there is a Possibility of damage to the Equipment .
To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout these operating instructions.
CONTENTS
17/03/11
P a g e | 3
INTRODUCTION
Thank you for selecting this product from the large range of products manufactured in Australia by Thermoline Scientific.
The materials incorporated in this design have been carefully selected for their durability, technical performance, and practicability, the resultant product is one we are proud of.
THE MANUAL
Please familiarise yourself with the information contained in this manual before using the equipment.
Ensure that the manual is kept safe but accessible for current and future operators to refer to.
UNPACKING
Remove the equipment from the packing material, check that it is complete. Assemble the components and inspect for damage.
Retain the packaging materials until the equipment has been thoroughly tested. Notify the detail of any damage without delay to your supplier, or Thermoline
Scientific.
LOCATION
Place the equipment in a well-ventilated area on a firm surface. Confirm that the power supply is suitable for this equipment. Refer to the equipment rating plate for details. To control temperature effectively, the equipment should be located in a stable ambient.
WARNING: Ensure that the electrical outlet is earthed and is of the correct voltage and is capable of carrying the stated current load.
INTENDED USE
This equipment is used to provide a mild temperature controlled environment for use in microbiology, immunology, culture departments, stability storage of veterinary samples, education and many more.
17/03/11
P a g e | 4
CAUTION: The range of equipment described here are not to be used in conjunction with large quantities of water. The introduction of water and the resultant vapour could result in premature component failure.
WARNING: Do not use flammable solvents or combustible materials as fire or explosion may occur. This equipment uses components which may be a source of ignition.
FUNCTION
To provide an accurate controlled temperature between ambient +5C to +60C.
PRINCIPLE OF OPERATION
The range of Laboratory Incubators are ideally suited for laboratory bench top applications or floor. The incubator body is insulated to minimise heat loss. The outer door which is also insulated closes on a magnetic gasket.
An Inner Clear Door for viewing products under evaluation whilst not disturbing the conditions, is standard in each model.
Removable chrome plated open wire shelves are supplied with additional shelf spacing.
The desired operating temperature is set at the temperature controller. A signal from the temperature sensor located inside at the stainless steel liner wall is monitored at the controller. The resultant calculated pulsed output is directed to a solid state relay which switches power to the heating elements which are bonded to the exterior of the liner. The warmed walls of the incubator, gently and uniformly heat the air. The amount of energy is determined by the sensor.
Air movement is either gravity ‘G’ convection, Models: TI-150G and TI-250G or Fan F’ Forced, Models: TI-150F, TI-250F, TI-350F and TI-500F.
PERFORMANCE
The operating range for this equipment is ambient +5C to +60C with a control accuracy of 0.1C.
MAXIMUM TEMPERATURE
For this product is 60C.
17/03/11
P a g e | 5
C-TICK
The equipment referred to here complies with requirement of AS/NZS 1044:1995 electromagnetic interference (EMI).
INSTALLATION (Refer to Unpacking Page 2)
Place the equipment in a well-ventilated area, on a firm level surface and ensure that a suitable power supply is available. Allow sufficient room for the door to open fully.
WARNING: Ensure that the electrical outlet is earthed and is of the correct voltage and is capable of carrying the equipment stated current load (AMPS).
The models TI-150 (Gravity) or (Fan Forced) are suitable for bench top use, where as models TI-250G and F, TI-350F and TI-500F are better suited for floor standing due to their height and weight.
CAUTION:
THE STANDARD Range of incubators are not suitable for stacking.
MAINTENANCE
The Thermoline range of laboratory incubators require no routine maintenance other than normal levels of cleanliness.
The external surface powder coated steel and the inner surface stainless steel may be wiped clean using a damp, soft cloth.
CARING FOR STAINLESS STEEL
Stainless steel is under most conditions extremely resistant to corrosion. This is in part due to the addition of chromium and nickel to the steel and the formation of a durable chromium oxide at the surface during the manufacturing process. There are several chemicals which will attack the surface of stainless steel, plus the lack of oxygen at the surface will cause rusting, corrosion and pitting. Note: Deionised water is particularly corrosive, to minimise, maintain a Neutral pH (7-9). Always have in place a program of regular cleaning using a soft damp cloth with a mild solution of soap and water and allow to dry.
TECHNICAL ASSISTANCE
If you require additional operational or technical information regarding this equipment please contact:
Thermoline Scientific Customer Service Division Telephone: 61 2 9604 3911 Facsimile: 61 2 9725 1706 Email: thermoline@thermoline.com.au