This manual functions in two ways: first as a step by step guide to help you operate the
waterproof TN-100/ T-100 turbidimeter; second, it serves as a handy reference guide.
It contains basic instructions that you must follow during the operation, care and
maintenance of the instrument. The safety protection provided by this equipment may
be impaired if it is used in a manner not described in this manual. It is recommended
that all operators should read this manual prior to working with this instrument.
Eutech Instruments/Oakton Instruments cannot accept any responsibility for damage or
malfunction to the meter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as
improvements are made, and does not represent a commitment on the part of Eutech
Instruments Pte Ltd/Oakton Instruments.
Note: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make
improvements in design, construction, and appearance of products without
notice.
11. Return of Items .............................................................................. 24
Instruction Manual TN-100/ T-100
1. Overview
Thank you for selecting the waterproof portable TN-100/ T-100 turbidimeter. The TN100/ T-100 allows you to measure turbidity of an aqueous sample in the field. This
instrument operates on the nephelometric principle of turbidity measurement and is
designed to meet the criteria specified in ISO 7027 and DIN 27027 standards (see
Measurement Principle on page 19).
1.1 Unpacking TN-100/ T-100 Meter and Accessories
The table below indicates the items that you should find in your turbidimeter shipment.
Item Quantity
1. Portable TN-100/ T-100 Turbidimeter with 4 “AAA” batteries 1
2. Instruction Manual 1
3. Instrument Carrying Case 1
4. Light Shield Cap 1
5. Calibration Set (0.02, 20.0,100, 800 NTU Standards, 60 ml each)
6. Empty Vials
7. Silicone Oil 1
8. Lint free cloth 1
1
7
Remove TN-100/ T-100 turbidimeter from the packing carton. Carefully inspect all
items to ensure that no visible damage has occurred during shipment. If the items you
received do not match your order, please contact your nearest distributor immediately.
WARNING: Extra care should be taken when unpacking, opening, and handling
the calibration standards and sample vials. Surface scratches or finger smudges
on the vial surface may cause measurement errors. Handle these items by their
caps only.
Batteries provided with the meter package are to be installed prior to use. See Section
1.4 - Battery Installation on page 4.
Figure 1 depicts the meter. The three main components of the instrument are the
sample well, the display, and the keypad. The following sections describe the
functionality of the display and the keypad. The proper use of the instrument and the
sample well are discussed in later sections.
1
Instruction Manual TN-100/ T-100
Sample Vial
Sample Well
Index Mark
Liquid Crystal Display
Keypad
CAL 13 CAL 24
mg/l
NTU
ppm
%pH
TAvg
14-segment Liquid Crystal Display
Battery Indicator
Calibration
Standards
Units of
Measurement
Figure 1: Parts of TN-100/T-100 Turbidimeter
1.2 Display
All the LCD segments and annunciators that can appear on the display are shown in
Figure 2. The display is used for reporting the turbidity reading and to provide guidance
for the operation of the instrument. In addition, the display has several other
annunciators that are used to communicate error messages and provide user
guidance.
Figure 2: Customized LCD with all annunciators
2
Instruction Manual TN-100/ T-100
Active only during calibration mode. These keys are used to select the
READ
ENTER
READ
ENTER
CAL
OFF
ON
WATERPROOF
1.3 Keys and Functions
The keypad has five keys: ON/OFF, CAL, ▲, ▼, and READ/ENTER.
Key Functions
ON
OFF
CAL
Powers on and shuts off the meter. With the auto switch-off feature the
meter automatically shuts off 20 minutes after last key press.
Initiates the calibration mode of the meter. When pressed, the meter is set
to accept the first calibration standard.
It is also used to exit the calibration mode if the user does not want to
follow the complete calibration procedure.
When in measurement mode, the READ/ENTER key is used to perform a
measurement.
Single-shot measurements: When the key is pressed and released
immediately (a quick key stroke of less than 0.3 seconds), the display will
blink [-Rd-] for 10 times and then display the measured value.
Continuous measurements
the instrument will perform a continuous measurement during which the
display is updated every 2 seconds. This can be used for indexing vials.
(This function is not available in calibration mode).
After the READ/ENTER key is released, the instrument will automatically
perform a single-shot measurement.
: If the READ/ENTER key is pressed and held,
calibration point in an incremental/ decremental manner.
NOTE: This meter is fitted with an auto incrementing feature, i.e. after
the successful calibration of one point it auto selects the next calibration
point, or automatically exits the calibration mode after the fourth
calibration point.
Figure 3: Keypad
3
Instruction Manual TN-100/ T-100
Philips Screws
Battery cover
Four 'AAA'-sized batteries
1.4 Battery Installation
Four AAA-sized batteries are included in the packaging:
1. Use a Philips screwdriver to remove the two screws holding the battery cover.
See Figure 4.
2. Remove the battery cover.
3. Insert the batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover onto its original position using the two screws
removed earlier.
5. The meter is now ready to operate.
Figure 4: Battery Installation
NOTE: Dispose used batteries in accordance with your local regulations.
4
Instruction Manual TN-100/ T-100
2. Turbidity Calibration
Calibration Overview
The TN-100/T-100 turbidity meter was calibrated and tested prior to leaving the factory.
The meter kit now includes a set of prepared SDVB primary standards in plastic bottles
which are EPA-approved for meter calibration. This keeps the stock solutions for the
0.02, 20.0, 100 and 800 NTU standards ready to use and offers improved precision
and shelf life. A set of 4 empty vials for the standards are included in the meter kit.
Before using the meter for the first time, verify the meter calibration by reading back the
value for the prepared primary standard at the appropriate NTU value. This check is
recommended daily to check the meter prior to measuring a sample. When the
standard value drifts, retest with a vial of fresh standard solution. The 0.02 NTU
standard needs to read <0.1 NTU for acceptance.
A replacement set of the 4 standards in plastic bottles is available. These 60mL bottles
are identical to the bottles in the meter kit, so they can easily be carried in the case to
where the measurement needs to be taken. These standards are specifically made for
the TN-100/T-100 portable turbidity meters. The standards will not read to the stated
NTU value in other manufacturer’s turbidity meters.
2.1 Calibration Standards
We recommend that you use the following materials during calibration to achieve the
accuracy stated in this manual:
CAL 1: 800 NTU Calibration Standard
CAL 2: 100 NTU Calibration Standard
CAL 3: 20.0 NTU Calibration Standard
CAL 4: 0.02 NTU Calibration Standard
It is well known that diluted Formazin is unstable. If you choose to use Formazin to
calibrate the instrument, ensure that you are using a fresh stock suspension of
Formazin to achieve the accuracy quoted for the instrument. Calibration standards
offered are more stable than Formazin. For expiry date of calibration standard, please
refer to the expiration date indicated on the cap label.
It is important that the calibration standards are not violently shaken or agitated
because air entrapment in the fluid introduces an error factor during calibration
which subsequently will lead to an inaccurate measurement. Also, do not store in
freezing temperatures which causes irreversible shrinkage of the standards’
particles thus resulting to inaccurate calibration and measurement.
2.2 Indexing
Due to the high quality of the glass vials provided, indexing is not required. You only
need to align the mark on the vial with the mark on the meter. However, in order to
achieve a better accuracy of the measurement, you can proceed with indexing of the
vials. See Section 8.5 - Indexing a Vial on page 20 for more information.
5
Loading...
+ 21 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.