Thermo Fisher Scientific Temp 360 User Manual

Instruction Manual
RTD Data logging Thermometer
Temp-360
Part of Thermo Fisher Scientific
68X526601 Rev 2 06/1
T
A
BLE OF CONTENTS
1. INTRODUCTION ............................... 1
2. SAFETY PRECAUTIONS.................. 2
3. SPECIFICATIONS ............................ 3
4. BATTERY INSTALLATION AND
REPLACEMENT................................ 6
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOUR........ 7
6. ASSEMBLING OPTIONAL HANDSFREE ACCESORIES
7. CONNECTING A RTD....................... 9
8. KEY FUNCTIONS ........................... 10
9. DISPLAY OVERVIEW ..................... 11
10. MEASUREMENT MODE................. 13
11. HOLD FUNCTON ............................ 14
12. MIN, MAX, AND AVE FUNCTION... 14
13. DATA LOGGING............................. 14
14. SETUP MODE................................. 15
15. GENERAL SETUP SCREEN .......... 16
16. CALIBRATION SCREEN
17. ALARMS SCREEN.......................... 20
18. DATA LOGGING SCREEN............. 25
19. CALIBRATION REPORT SCREEN
20. CLEAR / RESET SCREEN ............. 28
21. MAINTENANCE
22. CLEANING
23. BATTERIES
24. TROUBLESHOOTING..................... 30
25. ACCESSORIES
26. WARRANTY
27. PRODUCT RETURN
28. INNOCAL® CALIBRATION AND
REPAIR SERVICES ........................ 33
............................... 29
...................................... 29
..................................... 29
............................... 31
.................................... 32
....................... 32
............. 8
................. 18
.27
1. INTRODUCTION
his versatile hand-held
T in
strument accu measurements instrument easy oper includes features:
provides highly
rate te
mperature
is designed
ation
the follo
. The
and
wing
• Menu driven setup and operation
• Data logging for up to 2000 points
• USB output
• Operator selection of Celsius or Fahrenheit scale
• Resolution of up to 0.01°C/ °F
• Large backlit LCD and dot-matrix graphic LCD display
• Hold feature for temporarily retain a reading
• Displays minimum, maximum and average readings
• Field calibration capability
• Disabling of Auto-Off function
• Low battery warning
• Three-pin Din connector input
• Operates with a wide selection of probes
for
- 1 -
2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to accept low level signals supplied by standard RTDs. Under NO circumstances should the input voltage exceed the specified 50V RMS.
2. To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be non­hazardous.
CAUTION:
1. Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas.
2. Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
DANGER:
1. Voltages present at the RTD may also be present at the battery terminals. Always disconnect the RTD when changing batteries.
- 2 -
3. SPECIFICATIONS
RTD Thermometers
Temperature Range:
-201 to 1210°C (-330 to 2210°F)
Out of
ra
nge display:
- - - -
Reso
lution:
0.1 °C/°F: From -330.0 to -100 °C/°F
0.01 °C/°F: From -99.99 to 99.99 °C/°F
0.1 °C/°F: From 100.0 to 999.9 °C/°F 1 °C/°F: Above 1000 °C/°F
Accuracy:
±0.1 °C/ ±0.2 °F: From -330.0 to -100 °C/°F ±0.03 °C/±0.06 °F: From -99.99 to 99.99 °C/°F ±0.1 °C/ ±0.2 °F: From 100.0 to 999.9 °C/°F ±1 °C/ ±2 °F: Above 1000 °C/°F
Display:
Backlit Dot-matrix 50mm X 37.2mm
Data Logging:
2000 points
Logging Interval:
2 sec to 60 min
Min/Max/Avg Function:
Yes
- 3 -
Auto Off (Adjustable Time):
Enable/Disable option available
Stability Criteria:
Yes, upon stability of 5 seconds
Display Update
0.5
sec per up
Rate:
date
Input:
One three-pin DIN connector
Input Protection:
rms
50V
Storage:
– 40°C to 65°C (– 40°F to 149°F)
Humidity:
10% to 90% (non-condensing)
Battery Life: Size: Three AA, 1.5V; Alkaline Life: 400 hours continuous, typical,
(without backlighting and buzzer on)
Dimensions: Without Armor:
175mm (L) X 97mm (W) X 42mm (H)
With Armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with Batteries: Without Armor: 267g With Armor: 362g
- 4 -
Ingress Protection:
Meets IEC-529 IP-54 for dust and water resistant enclosures (with probe attached)
CE Compliance:
EN61326-1/A1: 1998 (EU EMC Directive)
- 5 -
4. BATTERY INSTALLATION AND REPLACEMENT
The total battery life without backlighting is about 400 hours. Remaining battery power is indicated by the battery life indicator.
Indicator Voltage
Cell + 3 bars
More than 4.1 V
Cell + 2 bars
Cell + 1 bars Cell + empty bars
flashing Adaptor power
supply USB power supply USB power
More than 3.6 V
More than 3.0 V More than 2.85 V
Main power supply connected
Selected settings memory and will remain in memory even
after
batteries
1. Before
are being replac
changing
instrume
2.
Loosen screw and off
back
3.
Remove
4. Insert three pol
arity.
5.
Install
cover and
are
stored in
power is turned
battery, tur
nt off
and disconnect RTD
lift battery
of case.
the t
hree
AA batteries.
new
batteries
tighten
- 6 -
off, or
ed.
observing
screw
while
n
cover
.
.
5. INSERTING AND REMOVING OPTIONAL RUBBER ARMOUR
1. To insert thermometer into the optional rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for bench top applications if necessary.
2. To remove thermometer from armor, push out from the bottom edges of meter until it is completely out of boot.
- 7 -
6. ASSEMBLING OPTIONAL HANDSFREE ACCESSORIES
You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations.
- 8 -
7. CONNECTING A RTD
Use the correct 100 RTD (alpha =
0.003850) for your instrument. Using an incorrect probe type will result in erroneous readings. Insert the 3-pin plug into the mating connector on the top of the instrument.
If no probe is connected the display will read “open”.
s are sensitive
RTD sensing element measurements reading to tim give you required
stabilize. Multip
e
constant of the
the
.
. W
,
allow
approx
at the tip or
hen
taki
ng
time for the
lying
probe by 5 will
imate time
the
- 9 -
8. KEY FUNCTIONS
F1
F2
F3
hold Freeze display
on/off
light
recall
log
Note: Function keys change in setup mode to provide advanced operation flexibility. Text above key will indicate function.
Step through Min, Max and Avg readings
Toggle between F and C display
Toggle between menu and measure mode
Turns meter on and off (press and hold for 3 seconds to turn off)
Press momentarily to turn on backlight
Recalls and steps through stored readings
Stores current measured value to memory
- 10 -
9. DISPLAY OVERVIEW
The dot matrix display features a large primary display, smaller secondary displays for channel info or min/max/ave, and helpful annuciators for added measurement data.
- 11 -
1 Power supply indicator
(Battery/Main adaptor/USB) 2 Date (format of mm – dd) 3 Time (hour:min) 4 Time format (am/pm/hrs) 5 Measurement mode 6 Hold function indicator 7 Stable indicator 8 Data logging indicator 9 PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min/Max/Avg elapsed time 13 Current Min/Max/Avg reading
14 Meter logging memory location 15 Measurement alarm active
16 Countdown time indicator with
17 Min/Max/Avg display 18 Min/Max/Avg unit indicator 19 Function keys
indicator
indicator
countdown time
- 12 -
10. MEASUREMENT MODE
On initial start-up the meter will display the measured value in the primary display.
F2
Pressing the
key will toggle reading
between F and C display.
Pressing the
F1
key initiates and toggles through the Minimum, Maximum, and Average Reading modes.
F3
Pressing the
key accesses the Setup mode.
- 13 -
11. HOLD FUNCTION
Press the on the display. Press the
hold
key to retain the reading
hold
key
again for normal operation.
12. MIN, MAX, AND AVE FUNCTION
Press the F1 key to toggle between the minimum, maximum, and average readings. The minimum and maximum reading function is ideal for monitoring unattended operations while continually displaying every temperature change that occurs. The minimum and maximum values are sensed and automatically stored. To exit and clear this function, press the
F3
key to access the Menu functions. See the CLEAR/RESET SCREEN section for more details.
13. DATA LOGGING
Press the reading to memory. The memory indicator M = 1234 shows the memory location for the next stored reading. Press the readings. See the DATA LOGGING section for timed logging, and logging to a computer. See the CLEAR/RESET SCREEN section for information on clearing stored readings.
log
key to store the current
recall
key to review stored
- 14 -
14. SETUP MODE
To access the setup mode from measurement mode press the “Menu” key (F3).
Press the ▲▼ keys on the meter key pad to scroll through options.
To enter a setup screen press the
(F1)
“Select” key
To return to the measurement mode press the “Meas” key following menu options are listed:
1. General setup
2. User field calibration
3. Alarm settings
4. Data logging settings
5. View user calibration report
6. Clear/Reset options
.
- 15 -
(F3)
. The
15. GENERAL SETUP SCREEN
The first page of the General Setup screens let you set resolution, auto-off time, and password.
Press the F1 key to indicate that you want to change the setting of the current parameter or press the recall key or log key to move to the next parameter.
Press the recall key or log key to change the options.
Press the F2 key to choose the next setting. When the option is set, press the F1 key to accept the choice.
- 16 -
The screen below is used to reset/change the password. If the user forgets the password, 5586 can be used to reset to a new value.
On the second screen you can set the time and date.
- 17 -
16. CALIBRATION SCREEN
The thermometer is factory calibrated and does not require calibration before use. The Calibration function allows a single point calibration of the thermometer, at any temperature point, to compensate for RTD offset error. It is NOT necessary to perform a field calibration to obtain the specified meter accuracy. Use the field calibration feature to improve thermometer/probe accuracy or to compensate for RTD drift.
Before going into the Calibration mode, the user must enter the password.
F2
Press the digit. (Default Password is 9900)
key to change to the next
- 18 -
There are two calibration options:
Offset – adjusts at a single point. Offset calibration can be performed at any temperature in the offset range of ±10.00 °C or ±18.00 °F.
Slope – adjusts at two points. The two calibration points must be at least
40.00 °C (72.00 °F) apart. The second calibration point should be at a higher value than the first calibration point.
Press the or key to adjust the value to match a known temperature standard. Press the F1 key to accept.
- 19 -
17. ALARMS SCREEN
There are two kinds of alarm settings available in the alarm setting options:
MEASUREMENT ALARM
Disable or enable the alarm by pressing the recall or log key and the F1 key to accept. Increase or decrease the individual limit by pressing the recall or log key.
Meter in alarm mode
- 20 -
COUNTDOWN ALARM
You can enable/disable the countdown alarm and set the countdown time from 5 seconds to 1 hour.
After setting (enabling) the countdown alarm, the measurement screen will look like this:
- 21 -
Press the F2 key to start the countdown:
Press the F2 key in the middle of a countdown to stop the process:
Restart the countdown by pressing the “Start” key (F2).
Alternatively, press the “Menu” key (F3) to go into the timer menu and disable the alarm.
- 22 -
The buzzer will sound for 30 seconds or until the “Reset” key (F2) is pressed at the end of the countdown.
Flashing icon indicates that the countdown has expired
To reset the timer, press the “Reset” key (F2). The display will look like this:
To repeat functions, press the F2 key.
Note: The °C/ °F function is disabled when the countdown function is enabled. To activate the
o
C/ oF function on the measurement screen, see the COUNTDOWN ALARM
- 23 -
section and disable the countdown alarm function.
When activated, the countdown timer temporarily overrides the auto-off function until the countdown is completed or manually stopped. If the meter is manually or automatically shut off, the countdown alarm is automatically set to “Disable” each time the meter is shut off. The “Timer” key will revert back to the “°C/°F” key.
- 24 -
18. DATA LOGGING SCREEN
Press the recall key or log key to choose the logging methods as auto or manual. If auto logging is selected, use the recall key or log key to set the time interval. Its range is from 2 minutes to 60 minutes.
DATA TRANSFER FROM METER TO COMPUTER
- 25 -
Once the USB connection is established with a PC, press the “Select” key (F1) to download data from Meter to PC using HyperTerminal.
***NOTE: METER MUST BE TURNED ON PRIOR TO CONNECTING USB CABLE OR COMPUTER MAY NOT RECOGNIZE THE INSTRUMENT***
- 26 -
19. CALIBRATION REPORT SCREEN
The calibration report will show the time and date along with results of the last user calibration.
- 27 -
20. CLEAR / RESET SCREEN
Press the F1 key to select the data you want to clear or reset. For calibration, logged data and reset all, you will have to enter the password to proceed. (Default Password is 9900)
- 28 -
21. MAINTENANCE
Properly us should maintain calibr and
not occasional cleaning changing
ed, the th
require service
of the
of the batteries.
ermome
ation indefinitely
other than
22. CLEANING
WARNING:
TO PREVENT IGNITION OF A
M
HAZARDOUS AT E
LEC
TROSTATIC DISCHARGE,
OSPHERE BY
CLEAN WITH DAMP CLOTH.
Do
not
solven
ts.
immerse
clean wi
th
abrasives or
Use mild
nor use excessive
detergents,
23. BATTERIES
If there is no thermometer is turned
ition of th
cond Also are clean and install refer to the BATTERY INSTALLATION AND REPLA replacement proced
display when
on,
e three
AA batteries.
check
that the battery ter
batteries
ed. If
replacement is necessary
C
EMENT section for the
ure.
the
are properly
ter
housing and
never
fluid.
check the
minals
,
- 29 -
24. TROUBLESHOOTING
The following chart lists the most probable faults. There are no internal adjustments or user-replaceable parts.
FAULT ACTION
NO Display
Display shows OVER / UNDER
Display Shows OPEN
Display Shows Err
Cannot connect to PC
Check condition of batteries. Check that batteries are inserted properly.
Out of range indication
No RTD connected in the connector
If display shows this message other than during the field calibration mode, please return the instrument for servicing
Ensure that the meter is powered on and measuring prior to connecting to USB port. Before and after the driver is installed, your PC may not recognize the Temp­360 if the Temp-360 is not powered on prior to connecting the USB to the computer.
- 30 -
25. ACCESSORIES
ent
Replacem Accessories
Meters and Meter
Item Part
Temp 360 Thermometer 35426-60
Rubber Armour with Stand 35427-80
Number
Handsfree Kit (Two Magnets and a Strap)
General Purpose Probe (immersion into liquids)
Penetration Probe (meat, semi-soft materials)
Surface Probe (direct contact on hot surfaces)
Air/gas Probe 08117-90
35427-85
08117-70
08117-85
08117-78
- 31 -
26. WARRANTY
he Manufac
T product to be free from significant deviations from published specifications ye
ars. If necessary period at
no charge misuse or abuse on your deter
mined by Repair period, or prod
uct
invoiced
turer
warra
nts this
for a
period
repair or adjus
wit
hin
, the
problem will be corrected
if it
the
costs
outside
those misuse or abus to
you
.
tment
the
warranty
is
not
due
Manufacture
the
resulting
e,
of
three
is
to
part as
warr
anty
from
may be
r.
27. PRODUCT RETURN
T
o
limit
charges and dela
seller or Manu
the author
ization
and shipping
in
structions
produ
e warranty peri
th
produc
the for the pack insure loss. The Manufacturer will not be responsible for damage resulting from careless or insufficient packing.
before returning
ct,
either within or
od. W
t,
please
return
. For
the instrument
it
agai
nst
possible damage or
ys, contact
fact
urer
for
outside of
hen
retur
state the
your
protectio
carefully and
the
reason
ning
n,
- 32 -
28. INNOCAL® CALIBRATION AND REPAIR SERVICES
Optimum performance of your temperature-measuring instrument is not a timeless condition. To ensure quality measurements, have your instrument calibrated regularly. Trust InnoCal® to satisfy your calibration and equipment repair needs. With over a decade of service, we've helped thousands of customers meet ISO, FDA, EPA, GLPs/cGMPs and other quality standards.
Conformity*
ISO/IEC 17025:2005 accredited NIST Handbook 150, 2000 Edition ANSI/NCSL Z540-2-1997 NIST Technical Note 1297 ISO 9000:2000
Fast Service
Our substantial inventory of replacement parts ensures a fast turnaround and prevents costly downtime. Most instruments serviced in five business days!
Excellent Value
Get quality at a fair price. Our InnoCal® NIST-traceable certificates offer extensive test data on a broad range of measurement parameters without breaking the bank!
- 33 -
Reliable Support
Trust in our free diagnostic support and troubleshooting advice. Our factory-trained metrologists and technicians have years of experience and extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments functioning properly. Based on your requirements, InnoCal will send you a reminder when it’s time to re-certify or service your instrument.
We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection, measuring, and test equipment.
Certification includes certificate of calibration with test data, including:
description and identification of the item certified
condition of the item
issue date
identification of calibration procedure
calibration date
as found/as left test data (where applicable)
signature of technician
statement of estimated uncertainty
list of equipment used to perform calibration (including their calibration dates)
- 34 -
With today's high quality standards such as ISO 9000, certification is becoming increasingly important. Traceability is not a timeless condition. It must be verified and maintained over the life of the calibration to ensure the highest accuracy possible. When you have your calibration done by InnoCal, we will send you an automatic reminder when it is time to recalibrate your instrument.
Are your calibration certificates good enough?
InnoCal surpasses the competition by providing the most complete certificates as required by NIST. All of our certificates include measured data and point-by-point measurement uncertainty, and by request, we’ll provide test accuracy and test uncertainty ratios at no extra cost. Call us today and see why InnoCal is The Choice of Quality.
*See our Scope of Accreditation for any limitations.
- 35 -
Calibration test points against NIST­traceable standards
Four test points
across range of
instrument.
0, 100, 165,
230 C
(-4, 32, 446,
770F)
Meter only
17000-04 17001-04 17002-04
Probe only
InnoCal—The Choice of Quality
1-866-InnoCal (1-866-466-6225)
InnoCalSolutions.com
System (meter + probe)*
- 36 -
- 37 -
- 38 - - 39 -
TECHNICAL ASSISTANCE
If
you have any questions abo
use
of this
prod
uct, contact the
Manufacturer or authorized se
ut the
ller.
For
more information on
OAKTON Instruments Products, please contact your nearest distributor or visit our web site listed below:
Oakton Instruments
625 E Bunker Court Vernon Hills, IL 60061 USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984
info@4oakton.com www.4oakton.com
Distributed by
:
Loading...