Thermo Fisher Scientific Temp 340 User Manual

Page 1
Instruction Manual
INSTRUCTION MANUAL
Temp-340 Thermistor
Oakton® Temp-340
Datalogging Thermometer
Thermistor
Datalogging Thermometer
Part of Thermo Fisher Scientific
68X536601 Rev1 April 2010
Page 2
Page 3
T
A
BLE OF CONTENTS
1. INTRODUCTION .......................................1
2. SAFET
3. SPECIFICATIONS ....................................3
4. BATTER
5. INSERTING
6. ASSEMBLING OPTIONAL HANDSFREE
7. CON
8. K
9. DISPLA
10. MEASUREMEN
11. HOLD FUN
12. MIN, MAX, and
13. DA
14. SETUP MODE
15. GE
16. CALIBRATION SCREEN
17. A
18. DA
19. CALIBRATION REPORT SCREEN
20. CLEAR / RESET SCREEN .....................28
21. MAINTENANCE
22. CLEANING
23. BATTERIES
24. TROUBLE SH
25. ACCESSORIES
26. WARRANTY
27. PRODUCT RETURN
28. INNOCAL® CAL
Y PRECAUTIONS.......................... 2
REPLACEMENT........................................ 6
RUBBER ARMOUR................................... 8
ACCESORIES
SERVICES............................................... 33
Y INSTALLATION AND
AND REMOVING OPTIONAL
........................................... 9
NECTING A THERMISTOR ............ 10
EY FUNCTIONS ...................................11
Y OVERVIEW ............................. 12
T MODE ......................... 14
CTON .................................... 15
AVE FUNCTION............. 15
TA LOGGING .....................................15
......................................... 16
NERAL SETUP SCREEN .................. 17
......................... 19
LARMS SCREEN .................................. 21
TA LOGGING SCREEN ..................... 24
......... 27
....................................... 29
............................................... 29
............................................. 29
OOTING............................ 30
....................................... 31
............................................ 32
............................... 32
IBRATION AND REPAIR
Page 4
1. INTRODUCTION
his versatile hand-held
T in
strument accu measurements instrument for includes features:
provides highly
rate te
. The
is designed
easy oper
the follo
mperature
ation
and
wing
• Menu driven setup and operation
• Datalogging for up to 2000 points
• USB output
• Operator selection of Celsius or Fahrenheit scale
• Resolution of up to 0.01°C/ °F
• Large backlit LED and dot-matrix
graphic LCD display
• Hold feature for temporarily retain a reading
• Displays minimum and maximum and average readings
• Field calibration capability
• Disabling of Auto-Off function
• Low battery warning
• Bayonet connector input
- 1 -
Page 5
2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to accept low level signals supplied by standard Thermistors. Under NO circumstances should the input voltage exceed the specified 50V RMS.
2. To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be non-hazardous.
CAUTION:
1. Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas.
2. Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
DANGER:
1. Voltages present at the Thermistor
may also be present at the battery terminals. Always disconnect the Thermistor when changing batteries.
- 2 -
Page 6
3. SPECIFICATIONS
Thermistor Thermometers
Type
400 series
Temperature range
-40 to 150°C (-40 to 302°F)
Out of range display:
OPEN
Reso
lution
0.01 or 0.1 °F/°C; auto-ranging to 0.1°
above +99.99°
Accuracy
From –40.00 to 99.99°F (–40.00 to
99.99°C): ±0.06°F (±0.03°C), From 100.0
to 257.0°F (100.0 to 125.0°C): ±0.1°F (±0.1°C), From 257.0 to 302.0°F (125.0 to
150.0°C): ±0.9°F (±0.5°C)
Displa
Backlit Dot-matrix 50mm X 37.2mm
y
Data Logging
2000 points
Logging Interval
2 sec to 60 min
Min/Max/Avg Function
Yes
- 3 -
Page 7
Auto Off (adjustable time)
Enable/Disable option available
Stability Criteria
Yes, upon stability of 5 seconds
Display update
0.6
sec per up
rate
date.
Input
One bayonet connector
.
Input Protection 50V
rms
Storage
–40°C to 65°C (–40°F to 149°F)
Humidity
10% to 90% (non-condensing)
Battery Life Size: Three AA, 1.5V; Alkaline Life: 400 hours continuous, typical,
(without backlighting)
Dimensions Without Armor:
175mm (L) X 97mm (W) X 42mm (H)
With Armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with batteries Without Armor: 267g With Armor: 362g
- 4 -
Page 8
Ingress protection:
Meets IEC-529 IP-54 for dust and water resistant enclosures (probe attached)
CE Compliance
EN61326-1/A1: 1998 (EU EMC Directive)
- 5 -
Page 9
4. BATTERY INSTALLATION AND
REPLACEMENT
The total battery life without backlighting is about 400 hours. Remaining battery power is indicated but the battery life indicator.
Indicator Voltage
Cell + 3 bars
More than 4.1 V
Cell + 2 bars
Cell + 1 bars
Cell + empty
bars flashing
Adaptor power
supply
USB power
supply
More than 3.6 V
More than 3.0 V
More than 2.85 V
Main power supply
connected
USB power connected
- 6 -
Page 10
Selected settings
will remain in memory even
and power is turned are being replac
are
stored in
off, or
ed.
1. Before instrume Thermistor
2.
Loosen screw and back Remove
3.
4. Insert three pol
5.
Install
nt off
of case.
the t
arity.
cover and
changing
.
hree new
tighten
lift battery
AA batteries.
batteries
memory
after
while
batteries
battery, tur
and disconnect
cover
observing
screw
.
n
off
- 7 -
Page 11
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOUR
1. To insert thermometer into the optional rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for bench top applications if necessary.
2. To remove thermometer from armor, push out from the bottom edges of meter until it is completely out of boot.
- 8 -
Page 12
6. ASSEMBLING OPTIONAL
HANDSFREE ACCESORIES
You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations.
- 9 -
Page 13
7. CONNECTING A THERMISTOR
Use the correct 400 series Thermistor for your instrument. Using an incorrect probe type will result in erroneous readings. Insert the bayonet plug into the mating connector on the top of the instrument.
If no probe is connected the display will read “open”.
Thermistor sensing element measurements, allow
o
stabilize. Multip
t of the approx
s are sensitive
probe by 5 will give you
imate time
lying
required
at the tip or
. W
hen
time for the
the time constant
.
taki
reading
ng
the
- 10 -
Page 14
8. KEY FUNCTIONS
F1
F2
F3
hold Freeze display
on/off
light
recall
log
Note: Function keys change in setup mode to provide advanced operation flexibility. Text above key will indicate function.
Step through Min, Max and Avg readings.
Toggle between F and C display
Toggle between menu and measure mode
Turns meter on and off (press and hold for 3 seconds to turn off)
Press momentarily to turn on backlight
Recalls and steps through stored readings
Stores current measured value to memory
- 11 -
Page 15
9. DISPLAY OVERVIEW
The dot matrix display features a large primary display, smaller secondary displays for channel info or min/max/ave, and helpful annuciators for added measurement data
- 12 -
Page 16
1 Power supply indicator
(Battery/Main adaptor/USB) 2 Date (format of mmm – dd) 3 Time (hour:min 4 Time format (am/pm/hrs) 5 Measurement mode 6 Hold function indicator 7 Stable indicator 8 Data logging indicator 9 PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min/Max/Avg elapsed time 13 Current Min/Max/Avg reading
14 Meter logging memory location 15 Measurement alarm active
16 Countdown time indicator with
17 Min/Max/Avg Display 18 Min/Max/Avg unit indicator 19 Function keys
indicator
indicator
countdown time
- 13 -
Page 17
10. MEASUREMENT MODE
On initial start-up the meter will display the measured value for in the primary display.
F2
Pressing the
key will toggle reading
between F and C display.
F1
Pressing the
key initiates and toggles through Minimum, Maximum, and Average reading modes.
F3
Pressing
enter accesses Setup mode.
- 14 -
Page 18
11. HOLD FUNCTON
Press the hold key to retain the reading on the display. Press the hold key again for normal operation.
12. MIN, MAX, and AVE FUNCTION
Press the F1 key to toggle between the minimum, maximum, and average readings. The minimum and maximum reading function is ideal for monitoring unattended operations while continually displaying every temperature change that occurs. The minimum and maximum values are sensed and automatically stored.
To exit and clear this function, press the F3 to access the Menu functions. See the Clear Reset menu section for more details.
13. DATA LOGGING
Press the
log
key to store the current reading to memory. The memory indicator M = 1234 shows the memory location for the next stored reading.
Press the
recall
key to review stored
readings.
See section on Data Logging for timed logging, and logging to a computer.
See section on Clear/Reset for information on clearing stored readings.
- 15 -
Page 19
14. SETUP MODE
To access the setup mode from measurement mode press the F3 key (Menu).
Press ▲▼ keys on the meter key pad to scroll through options.
To enter a setup screen press Select key.
To return to measurement mode press
F3
Meas are listed
key. Following menu options
1. General Setup
2. User field calibration
3. Alarm settings
4. Data logging settings
5. View user calibration report
6. Clear/Reset options
- 16 -
F1
Page 20
15. GENERAL SETUP SCREEN
The first page of the General Setup screens let you set Resolution, auto-off time, and password.
Press F1 to indicate you want to change the setting of the current parameter or recall or log to move to the next parameter.
Press recall or log to change the options.
Press F2 to choose the next setting. Whenever set the options, press F1 for accepting the choice.
- 17 -
Page 21
This screen below is used to reset/change password. In the event if uses forget his/her password, 5586 can be used to reset to a new value
On the second page you can set time and date.
- 18 -
Page 22
16. CALIBRATION SCREEN
The thermometer is factory calibrated and does not require calibration before use. The Calibration function allows single point calibration of the thermometer, at any temperature point to compensate for Thermistor off-set error. It is NOT necessary to perform a field calibration to obtain specified meter accuracy. Use the field calibration feature to improve thermometer/probe accuracy or to compensate for Thermistor drift...
Before go into the calibration mode, must enter the password. Press F2 to change to the next digit. (Default Password is 9900)
- 19 -
Page 23
There are two calibration options:
- adjusts at a single point. Offset
Offset calibration can be performed at any temperature in the offset range of ±5.00 °C or 9.00 °F Slope – adjusts at two points. The two calibration points must be at least 20.00 °C (36.00 °F) apart. The 2nd calibration point should be at a higher value than the 1st calibration point.
Use the or keys to adjust the value to match known temperature standard. Press F1 to accept.
- 20 -
Page 24
17. ALARMS SCREEN
There are two kind of alarm setting is available under the alarm setting options
Measurement alarm
Disable or enable the alarm by pressing recall or logand F1 to accept. Increase or decrease individual limit by pressing recall or log.
Meter in alarm mode
- 21 -
Page 25
Countdown alarm
You can enable/disable the countdown alarm and set the countdown time from 5 sec to 1 hour.
After setting (enabling) the countdown alarm, the measurement screen should look like this:
- 22 -
Page 26
Press F2 key to start the countdown:
Press F2 key in the middle of a countdown will stop the process:
You will need to restart the countdown by selecting the “start” key (F2). Alternatively, you can also choose “Menu” to go into the timer menu and disabling the alarm.
- 23 -
Page 27
The buzzer will sound for 30 seconds, or until the reset (F2) is pressed, at the end of the countdown.
Flashing announciator indicates that the countdown has expired
To reset the timer, select “restart” (F2) it. You display should look like this:
To repeat the functions, select F2 Note: The °C/ °F function is disabled when the count down function is enabled. To activate the oC/ oF function on measurement screen, please go into the
- 24 -
Page 28
“Alarm Setting page 2” and disable the count down alarm function. When activated, the countdown timer temporarily over-rides the Auto-shutoff until the countdown is completed or manually stopped. If the meter is manually or auto shutoff, The Countdown Alarm is automatically reset to “Disable” each time the meter is shut off. The 'Timer' key will revert back to the '°C/°F' key.
- 25 -
Page 29
18. DATA LOGGING SCREEN
Press recall or logto choose the logging methods as auto or manual. If it is auto logging, using recall or logto set time interval. Its range is from 0min to 60min.
Data Transfer from Meter to Computer
Once the USB connection is establish with
- 26 -
Page 30
PC, press the Start button to download data from Meter to PC using HyperTerminal.
***NOTE: METER MUST BE TURNED ON PRIOR TO CONNECTING USB CABLE OR COMPUTER MAY NOT RECOGNIZE THE INSTRUMENT***
19. CALIBRATION REPORT SCREEN
The Calibration report will show the time and date along with results of the last user calibration.
- 27 -
Page 31
20. CLEAR / RESET SCREEN
Press F1 to choose which data you want to clear or reset. For calibration, logged data and reset all, you will have to enter the password to proceed. (Default Password is 9900)
- 28 -
Page 32
21.
MAINTENANCE
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and changing of the batteries.
22. CLEANING
WARNING:
TO PREVENT IGNITION OF A HAZARDOUS ATMOSPHERE BY E
LEC
TROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solvents. Use mild detergents, never immerse nor use excessive fluid.
23. BATTERIES
If there is no display when the thermometer is turned on, check condition of the three AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary, refer to the BATTERY INSTALLATION AND REPLACEMENT section for replacement procedure.
- 29 -
Page 33
24. TROUBLE SHOOTING
The following chart lists the most probable faults. There are no internal adjustments or user-replaceable parts.
FAULT ACTION
Check condition of batteries.
NO
Display
Check that batteries are inserted properly.
Display
Shows
OPEN
Display
Shows
Err
Cannot
connect
to PC
No Thermistor connected in the Connector
If display shows this message other than during the field calibration mode, please return the instrument for servicing
Ensure that the meter is powered on and measuring prior to connecting to USB port. Before and after the driver is installed, your PC may not recognize the Temp-340 if the Temp-340 is not powered on prior to connecting the USB to the computer.
- 30 -
Page 34
25.
ACCESSORIES
Replacement Meters and Meter Accessories
Item Part
Temp 340 thermometer 35426-50
Rubber Armour with Stand 35427-80
Hands-free Kit (Two magnets and a strap)
General Purpose Probe (Immersion into liquids)
Flexible General Purpose Probe (Vinyl sheath)
USB cable 35427-86
Adapter, 100-240 91427-99
Number
91426-50
35427-85
08491-06
08491-02
- 31 -
Page 35
26. WARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifications for a period of three years. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
27. PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the carefully and insure it against possible damage or loss. The Manufacturer will not be responsible for damage resulting from careless or insufficient packing.
- 32 -
Page 36
28. INNOCAL® CALIBRATION AND REPAIR SERVICES
Optimum performance of your temperature-measuring instrument is not a timeless condition. To ensure quality measurements, have your instrument calibrated regularly. Trust InnoCal® to satisfy your calibration and equipment repair needs. With over a decade of service, we've helped thousands of customers meet ISO, FDA, EPA, GLPs/cGMPs and other quality standards.
Conformity*
ISO/IEC 17025:2005 accredited NIST Handbook 150, 2000 Edition ANSI/NCSL Z540-2-1997 NIST Technical Note 1297 ISO 9000:2000
Fast Service
Our substantial inventory of replacement parts ensures a fast turnaround and prevents costly downtime. Most instruments serviced in five business days!
Excellent Value
Get quality at a fair price. Our InnoCal® NIST-traceable certificates offer extensive test data on a broad range of measurement parameters without breaking the bank!
- 33 -
Page 37
Reliable Support
Trust in our free diagnostic support and troubleshooting advice. Our factory­trained metrologists and technicians are armed with years of experience and extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments functioning properly. Based on your requirements, InnoCal will send you a reminder when it’s time to re-certify or service your instrument.
We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection, measuring, and test equipment.
Certification includes certificate of calibration with test data, including:
description and identification of the
item certified
condition of the item
issue date
identification of calibration procedure
calibration date
as found/as left test data (where
applicable)
signature of technician
statement of estimated uncertainty
list of equipment used to perform
calibration (including their calibration dates)
With today's high quality standards such
- 34 -
Page 38
as ISO 9000, certification is becoming increasingly important. Traceability is not a timeless condition. It must be verified and maintained over the life of the calibration to ensure the highest accuracy possible. When you have your calibration done by InnoCal, we will send you an automatic reminder when it is time to recalibrate your instrument.
Are your calibration certificates good enough?
InnoCal surpasses the competition by providing the most complete certificates as required by NIST. All of our certificates include measured data and point-by-point measurement uncertainty, and by request, we’ll provide test accuracy and test uncertainty ratios at no extra cost. Call us today and see why InnoCal is The Choice of Quality.
*See our Scope of Accreditation for any limitations.
- 35 -
Page 39
Calibration test points against NIST­traceable standards
Four test points
across range of
instrument.
0, 100, 165,
230 C
(-4, 32, 446,
770F)
Meter only
17000-06 17001-06 17002-06
Probe only
InnoCal—The Choice of Quality
866-InnoCal (866-466-6225)
InnoCalSolutions.com
System (meter + probe)*
- 36 -
Page 40
TECHNICAL ASSISTANCE
If
you have any questions abo
prod
of this
uct, contact the
Manufacturer or authorized se
ut the
ller.
For
more information on
OAKTON Instruments Products, please contact your nearest distributor or visit our web site listed below:
Oakton Instruments
625 E Bunker Court Vernon Hills, IL 60061 USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984
info@4oakton.com www.4oakton.com
Distributed by
:
use
Loading...