Thermo Fisher Scientific 2-Wire Temperature Transmitter User Manual

EUTECH INSTRUMENTS
2-Wire Temperature Transmitter
Operating Instructions
3 Diagram and Dimensions (in mm)
4 Installation
Operating with one or several Transmitters:
Figure 1 - Operating with 1 Transmitter
1 Preface
Thank you for purchasing the Eutech 2-wire Temperature Transmitter. This manual serves to explain the correct installation and operations of
EC-PH-XTR;
EC-ORP-XTR;
EC-CON-XTR;
EC-DO-XTR; and
EC-TEMP-XTR.
The information presented in this manual is subject to change without prior notice, as improvements are made to the product. While Eutech will endeavor to try its best to inform the market, this does not represent a commitment on the part of Eutech Instruments Pte Ltd.
Eutech Instruments does not accept any responsibility for damage or malfunction of the unit due to improper installation or operation of the Transmitter.
Copyright 1999 Eutech Instruments Pte Ltd. Version 1.0. All rights reserved.
Eutech Instruments Pte Ltd. Blk 55 Ayer Rajah Crescent #04-14/24, Singapore 139 949. Tel: (65) 778 6876; Fax: (65) 773 0836; e-mail:
marketing@eutechinst.com; Home page: http://www.eutechinst.com
2 Introduction
The Transmitter is used for the continuous measurement of pH, ORP (REDOX), Conductivity, Dissolved Oxygen or Temperature, in conjunction with an appropriate sensor. Data output is via 4-20 mA current output.
The Transmitter is housed in a IP65 enclosure, with openings for input and 4-20mA output.
Figure 2 - Operating with several transmitters
5 Electrical Connections
6 Before Calibration
Remove cover of Transmitter and ensure that electrode is correctly connected as above. Connect Amp meter to the 4/20mA output of Transmitter (Appendix 1).
7 Temperature Transmitter Calibration
1) Immerse Temperature probe in water bath at varying temperatures and observe Amp meter display;
2) Amp meter should read corresponding mA output as follows:
3) In 0oC, the output is 4 mA;
4) In 25oC, the output is 8 mA;
5) In 50oC, the output is 12 mA;
6) In 80oC, the output is 16.8 mA;
7) In 100oC, the output is 20 mA;
8) To calibrate, locate “Temp adjustment” trimmer;
9) Using a fine screwdriver, slowly turn trimmer till Amp meter reads correct current output;
NB: For other Temperature values, expected current output is based on:
mA = {(16/100) x Temperature value} + 4mA
8 Specifications
Figure 4 - Method Two
10 Ordering Information
pH
Item Cat Code
pH Transmitter ECPHXTR Combination pH Electrode, General, integral 5m cable ECARTSO05B Combination pH Electrode, w/PMP, integral 5m cable ECARGTSO05B
Dissolved Oxygen
Item Cat Code
Dissolved Oxygen Transmitter ECDOXTR General Purpose D.O. probe, 0.5 – 40 ppm, Industrial,
integral 5m cable General Purpose D.O. probe, 0 – 10 ppm, Industrial,
integral 5m cable
ECDOGEN
ECDOTPll
Model ECTEMPXTR
Range 0 TO 100 oC Accuracy
± 0.5 oC Calibration 1 pt Input 2-wire Pt 100 Output 4-20 mA Operating
DC 12 V to 24 V
Voltage Load 100 Ohms max. for 12 V;
600 Ohms max. for 24V
Housing 77 x 77 x 28 mm, field mountable
9 Appendix 1
Connecting Amp Meter for Calibration
ORP
Item Cat Code
ORP Transmitter ECORPXTR Combination Platinum ORP Electrode, w/PMP,
integral 5m cable Combination Gold ORP Electrode, w/PMP, integral
5m cable
ECHTPTTSO05B
ECHTAUTTSO05B
Conductivity
Item Cat Code
Conductivity Transmitter ECCONXTR 2-Pin SS,k=1.0, integral 3m cable ECCONSEN46
Temperature
Item Cat Code
Temperature Transmitter ECTEMPXTR Pt 100 probe, SS316 ECPT10001M or
ECPT10005M
Power Supply
Item Cat Code
120V Power Supply EC-120-XTR 220V Power Supply EC-220-XTR
Figure 3 - Method One