COMPARISON SHEET
Orion 8010cX Ammonia Analyzer
specifications comparison
The table below shows a comparison between the new Thermo Scientific™ Orion™ 8010cX and the existing product,
the Thermo Scientific™ Orion™ 2210AM Ammonia Analyzer.
Product specifications Orion 8010cX analyzer Orion 2210AM analyzer
Interface 7 inch color touchscreen Monochrome screen, keypad entry
Measuring range 0.02–500 mg/L 0.03–30 mg/L ammonia as N user programmable
Accuracy error 3% ± 0.04 mg/L of reading for whichever is
greater, from 0.02–2 mg/L for 5–40° C
Response time ≤28 min (two consecutive readings) Less than 15 minutes per analysis
Repeatability error 3% ± 0.02 mg/L of reading for whichever is
greater, from 0.02–2 mg/L for 5–40° C
Limit of detection 0.02 mg/L 0.03 mg/L
Reagent consumption Reagents every 100 days with 2 hour cycle time 1 L of reagents every 60 days with 15 minute
Method Salicylate colorimetric method Salicylate colorimetric method
Ambient operating
temperature
Maximum humidity 95% RH noncondensing 90% at 40° C (104° F)
Sample flow 50–1000 mL/min 50–1000 mL/min
Sample pressure 3–73.5 psi (0.2–5 bar, for flow-cell) 5 psig max
Sample supply Continuous, interval (configurable start time) Continuous
Sample temperature 0–50° C (32–122° F) with no ice 5–40° C (41–104° F)
Total dissoved solids < 600 mg/L -
Suspended solids < 30 mg/L after filtration by pre-fliter < 60 microns
Sample inlet/outlet
connections
Drain tubing 1/8 inch OD flexible tubing 3/8 inch OD flexible tubing—polypropylene or
Sample streams One One
5–40° C (41–104° F) 5–45° C (41–113° F)
Metric fittings: G1/2 external thread to 10 mm
outer diameter, stainless steel tubing G1/2
external thread to 19.5 mm pagoda-shape joint,
PVC (over flow)
Less than 5% of reading or ± 0.03 mg/L,
whichever is greater, from 0.03–30 mg/L
for 5–40° C
Less than ± 2% of reading or ± 0.05 mg/L,
whichever is greater from 0.03–30 mg/L
cycle time
1/4 inch OD flexible tubing—polypropylene or
similar material
similar material
Product specifications 8010cX 2210AM
Enclosure integrity IP65 NEMA12 (Electronics cabinet),
NEMA4X (Fluidics cabinet)
Enclosure dimensions 450 mm x 737 mm x 322 mm (17.7 inches x 29
inches x 12.7 inches)
Shipping weight 40 kg (88 lbs) 18 kg (40 lbs)
Electrical power
requirements
Current loops Two 0/4–20 mA—Direct or Reverse Acting
Relays Two relays, 2 A @ 250 VAC Four SPDT, 2 A @ 250 VAC—Programmable
Digital comms RS485, MODBUS (standard),
Safety cTUVus, CE: EN/IEC 61010-1 CE: EN/IEC 61010-1, cTUVus
EMC FCC (Class A), ICES, CE: EN/IEC 61326-1,
Environmental safety CE-RoHS, REACH, China RoHS, Prop65 -
Certification CPA (China) N/A
Warranty 1 year 1 year
Sample handling Intelligent sample awareness; automatic on
Autovalidation function Yes No
100–240 VAC, 100 W, 50/60 Hz 100–240 VAC, 110 W, 50/60 Hz
(Isolated). Maximum 900 Ω load
PROFIBUS (optional)
RCM, KC
schedule or manual
27.5 inches x 16 inches x 6.4 inches (698 mm x
406 mm x 162 mm)
Two 0/4–20 mA—Direct or Reverse Acting
(Isolated). Maximum 900 Ω load
RS232 ASCII protocol for data reporting
FCC (Class A), CE: EN/IEC 61326-1
Intelligent sample awareness
For more details, please visit thermofisher.com/orion8000
For process water analysis and monitoring. © 2020 Thermo Fisher Scientific Inc. All rights reser ved. All trademarks are the property of
Thermo Fisher Scientific and its subsidiaries unless otherwise specified. DS-RAINBOW-E 0620