Thermo Fisher MAN0014293_Countess_II_Automated_Cell_Counter_UG MUSE NX Cardiology Information System Centricity® Clinical Gateway Integration Manual 2102027-008A

Page 1
Countess™ II automated cell counter
USER GUIDE
Catalog Number AMQAX1000
Publication Number MAN0014293
Revision C.0
For Research Use Only. Not for use in diagnostic procedures.
Page 2
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. MAN0014293
Revision Date Description
C.0 07 June 2019 Updated trypan blue handling instructions. Document converted to XML. B.0 30 June 2017 Add info about edited profile indicator, save profile from results screen,
A.0 01 September 2015 New user guide
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2019 Thermo Fisher Scientific Inc. All rights reserved.
dilution calculator, and reports.
Page 3

Contents

CHAPTER 1 About this guide ............................................ 7
Audience ....................................................................... 7
User documentation ............................................................. 7
Text and keyboard conventions .................................................... 7
User attention words ............................................................ 8
Safety alert words ............................................................... 8
CHAPTER 2 Product information ....................................... 9
Product description ............................................................. 9
Product contents ........................................................... 9
Countess™ II automated cell counter .......................................... 9
Upon receiving the instrument ................................................ 9
Register your instrument ................................................... 10
Instrument exterior components ................................................. 10
CHAPTER 3 Getting started ............................................ 11
Installation .................................................................... 11
Operating environment ..................................................... 11
Install the instrument ...................................................... 11
Turn ON the instrument .................................................... 11
Load profile ................................................................... 12
Profiles screen ............................................................ 12
Edit profile screen ......................................................... 13
Load a profile ............................................................. 14
Add/edit a profile .......................................................... 14
CHAPTER 4 Cell count and cell viability assays ...................... 16
Prepare sample ................................................................ 16
Recommendations ......................................................... 16
Load Countess™ chamber slide .............................................. 16
Count cells .................................................................... 17
Count procedure ........................................................... 17
Next steps ................................................................ 18
Countess™ II automated cell counter User Guide
3
Page 4
Contents
CHAPTER 5 Results .................................................... 19
View results ................................................................... 19
Results screen ............................................................ 19
Identify the cells counted ....................................................... 19
Advanced screen .......................................................... 19
Identify live and dead cells .................................................. 20
Graph count results ............................................................ 21
View graph ................................................................ 21
Gate count results .............................................................. 21
Adjust screen ............................................................. 21
Gate count results ......................................................... 21
Save as new protocol ........................................................... 23
Edit and save profile as new protocol ......................................... 23
CHAPTER 6 Dilution calculator ........................................ 25
Dilution calculator ............................................................. 25
Calculate dilution .............................................................. 25
CHAPTER 7 Save results ............................................... 28
Save count results ............................................................. 28
Save screen ............................................................... 28
Save procedure ............................................................ 28
Report ........................................................................ 31
Report file ................................................................ 31
CHAPTER 8 Instrument settings ...................................... 32
Overview ...................................................................... 32
Instrument settings screen ................................................. 32
Software update ............................................................... 32
Guidelines for software update .............................................. 32
Update the Countess™ II/II FL software ....................................... 33
Date/Time ..................................................................... 34
Set the date and time ....................................................... 34
4
Countess™ II automated cell counter User Guide
Page 5
CHAPTER 9 Maintenance .............................................. 36
Instrument care ............................................................... 36
General guidelines for care ................................................. 36
Power supply .............................................................. 36
Clean the Countess™ II automated cell counter .................................... 37
Introduction ............................................................... 37
Clean the touch-screen ..................................................... 37
Clean the instrument case .................................................. 37
Decontaminate the instrument .............................................. 37
Set nominal focus .............................................................. 38
Overview ................................................................. 38
Set nominal focus .......................................................... 38
APPENDIX A Troubleshooting ......................................... 40
APPENDIX B Product specifications .................................. 43
Contents
Technical specifications ......................................................... 43
Physical characteristics .................................................... 43
Technical specifications .................................................... 43
Optics .................................................................... 43
Analysis slide ............................................................. 43
APPENDIX C Ordering information .................................... 44
Countess™ II automated cell counter ............................................. 44
Accessory products ............................................................ 44
APPENDIX D CSV file format ........................................... 45
CSV file format, explained ....................................................... 45
Overview ................................................................. 45
APPENDIX E Safety ..................................................... 47
Safety alert words .............................................................. 47
Electrical symbols ............................................................. 48
Safety symbols ................................................................ 48
Environmental symbols ......................................................... 49
Safety labels on instruments .................................................... 50
General instrument safety ...................................................... 50
Operating the instrument ................................................... 50
Safety precautions ......................................................... 51
Cleaning or decontaminating the instrument .................................. 51
Removing covers or parts of the instrument ................................... 51
Countess™ II automated cell counter User Guide
5
Page 6
Contents
Chemical safety ................................................................ 52
Chemical waste safety .......................................................... 52
Chemical waste hazard ..................................................... 52
Chemical waste safety guidelines ............................................ 53
Waste disposal ............................................................ 53
Electrical safety ................................................................ 53
Biological hazard safety ......................................................... 54
Safety and electromagnetic compatibility (EMC) standards .......................... 55
Documentation and support ............................................. 56
Customer and technical support ................................................. 56
Limited product warranty ....................................................... 56
6
Countess™ II automated cell counter User Guide
Page 7
1

Audience

This user guide is for laboratory sta operating, maintaining, and analyzing data using the Countess™ II Automated Cell Counter.

User documentation

The guides listed below are available for the Countess™ II Automated Cell Counter.
About this guide
Countess™ II Automated Cell Counters User Guide
Countess™ II and Countess™ II FL Automated Cell Counters Quick Reference Card (QRC)
Additional resources are available on the Countess™ Technical Resources page. Go to www.thermosher.com/countess to access protocols, application notes, and tutorials.

Text and keyboard conventions

Text and keyboard conventions Text and keyboard conventions used in this user guide are listed below. For safety alert words and symbols used in this document, see “Safety alert words“ on page 8.
Convention
Bold Bold text indicates user action. For example: Press More.
Right arrow symbol () indicates a menu choice, and separates
successive commands you execute or select from a drop-down or shortcut menu. For example: Select More4Adjust.
Guide
Pub. no.
MAN0014293
MAN0010826
Use
Countess™ II automated cell counter User Guide
7
Page 8
Chapter 1 About this guide
1

User attention words

User attention words
Two user aention words appear this document. Each word implies a particular level of observation or action as described below.
Note: Provides information that may be of interest or help but is not critical to the use of the product.
IMPORTANT! Provides information that is necessary for proper instrument
operation, accurate installation, or safe use of a chemical.

Safety alert words

Four safety alert words appear in this document at points where you need to be aware of relevant hazards. Each alert word—IMPORTANT, CAUTION, WARNING, DANGER—implies a particular level of observation or action, as dened below:
IMPORTANT! – Provides information that is necessary for proper instrument
operation, accurate installation, or safe use of a chemical.
CAUTION! – Indicates a potentially hazardous situation that, if not avoided,
may result in minor or moderate injury. It may also be used to alert against unsafe practices.
WARNING! – Indicates a potentially hazardous situation that, if not avoided,
could result in death or serious injury.
DANGER! – Indicates an imminently hazardous situation that, if not avoided,
will result in death or serious injury. This signal word is to be limited to the most extreme situations.
Except for IMPORTANT! safety alerts, each safety alert word in this document appears with an open triangle gure that contains a hazard symbol. These hazard symbols are identical to the hazard symbols that are axed to the instruments (see Safety symbols in Appendix E).
8
Countess™ II automated cell counter User Guide
Page 9
2

Product description

Product information

Product contents

Countess™ II automated cell counter

The Countess™ II Automated Cell Counter is shipped with the components listed below.
Component
Countess™ II Automated Cell Counter (Cat. no. AMQAX1000) 1 each
Power Cord with 4 adaptor cords (for U.S./Canada/Taiwan/Japan, Europe, or UK)
Countess™ Cell Counting Chamber Slides (50 slides/box) 1 box
Countess™ II Disposable Slide Holder 1 each
Countess™ II USB drive 1 each
Countess™ II Automated Cell Counter Quick Reference Card 1 each
The Countess™ II Automated Cell Counter is a fully automated cell counter and assay platform that uses state-of-the-art optics and image analysis algorithms to analyze trypan blue-stained cells in suspension.
• The Countess™ II Automated Cell Counter oers an intuitive user interface, and provides the option to save data and generate a report, which can then be transferred to a PC using the USB drive supplied with the instrument or available separately.
• The cells to be counted are loaded into the instrument in disposable Countess Cell Counting Chamber Slides (“Prepare sample“ on page 16). Each chamber slide contains two enclosed chambers to hold the sample to allow you to measure two dierent samples or perform replicates of the same sample.
• The Countess™ II Automated Cell Counter takes 10 seconds per sample for a typical cell count and is compatible with a wide variety of eukaryotic cells. In addition to cell count and viability, the instrument also provides information on cell size.
Quantity
1 each

Upon receiving the instrument

Countess™ II automated cell counter User Guide
Examine the instrument carefully for damage incurred during transit. Ensure that all parts of the instrument, including accessories listed above, are included with the product. Damage claims must be led with the carrier; the warranty does not cover in-transit damage.
See “Installation“ on page 11 for instructions on installing the instrument.
9
Page 10
Chapter 2 Product information
2

Instrument exterior components

Register your instrument

Visit www.thermosher.com/registercountess to register your instrument. You will be asked to supply the serial number, your name, and your contact details. Registering your instrument ensures that you will receive notications of software upgrades and information on new assays for use with the Countess™ II Automated Cell Counter.
Instrument exterior components
Touch-screen display: The 7-inch capacitive touch-screen display is the main user interface of the Countess™ II Automated Cell Counter. It contains the buons for all instrument functions and displays data from the cell count.
②⑤ USB ports: The USB ports allow you to transfer and save the cell count data and image to an external computer for record keeping and printing purposes. You can use the USB drive supplied with the instrument or any other standard, FAT32-formaed USB drive for data transfer. If desired, you can plug in a USB mouse into the rear USB port for instrument control.
Note: The USB ports located in the front and the back of the instrument function the same. However, the rst USB drive connected will be the preferred saving location and both USB drives cannot be accessed at the same time.
Slide port: The slide port is used to insert the Countess™ Cell Counting Chamber Slide containing the sample into the instrument.
Power switch: The ON/OFF rocker switch is the main power switch. It is not necessary to use the power switch for day-to-day operation of the instrument.
Power input jack: The power input jack connects the instrument to an electrical outlet through the supplied power cord and the appropriate plug, based on the electrical outlet conguration in your country.
Note: The Countess™ II Automated Cell Counter (Cat. no. AMQAX1000) uses brighteld illumination only and does not support the EVOS™ LED light cubes.
10
Countess™ II automated cell counter User Guide
Page 11
3

Installation

Getting started

Operating environment

Install the instrument

• Place the instrument on a level surface away from vibrations emanating from other pieces of equipment.
• Allow at least 5 cm (2 in) free space at the back of the instrument to allow for proper ventilation and prevent overheating of electronic components.
• Set up the instrument away from direct light sources, such as windows. Ambient room lighting can enter the imaging path and aect the image quality.
• Operating temperature range: 4°–32 (40°–90°F).
• Relative humidity range: 30–90%.
IMPORTANT! Do not position the instrument so that it is dicult to turn o the main
power switch located on the back of the instrument (see “Instrument exterior components“ on page 10). In case of an instrument malfunction, turn the main power switch to the OFF position and disconnect the instrument from the wall outlet.
Unpack the instrument and place the instrument on a at, level, dry surface.
1.
Remove the thin plastic protector lm from the touch-screen display.
2.
Plug one end of the power cord appropriate for your region into the instrument.
3.
Plug the power cord into the electrical outlet. Be sure to use only the power cord
4.
supplied with your instrument. Powering the instrument with an unapproved power cord may damage the instrument.
Turn on the instrument by ipping the power switch on the back of the

Turn ON the instrument

Countess™ II automated cell counter User Guide
1.
instrument (“Instrument exterior components“ on page 10) to the ON position.
The instrument initializes and displays the Home screen.
11
Page 12
Chapter 3 Getting started
3
Load profile
From the Home screen, you can proceed immediately to the assays by inserting a
2.
slide (Chapter 4, “Cell count and cell viability assays“).
Alternatively, you can change or add a prole (Step 3 on page 12) or change instrument seings (Step 4 on page 12).
To change the current prole or to add a new prole to the instrument, press
3.
the Proles buon in the upper left corner.
Proles allow you to create customized count preferences (i.e., gate counts based on cell size, brightness, and/or circularity) (“Load prole“ on page 12).
Load profile
Profiles
screen
To change instruments seings, press the Instrument Seings buon in
4.
upper right corner.
Instrument seings allow you to update the Countess™ II software and change the date and time (Chapter 8, “Instrument seings“).
Proles screen allows you to create and save up to 9 customized proles. Each custom prole denes the count parameters (size, brightness, and circularity) and the Auto
Lighting function for a consistent and streamlined workow.
the
12
• You can access the Proles screen from the Home, Capture, Results, Advanced, or Adjust screens.
• The current prole is displayed on the upper left corner of the Home, Capture, Results, Advanced, or Adjust screens .
• The Auto Lighting function and count parameters are set in the “Edit prole screen“ on page 13.
• The Default prole contains default count seings and cannot be edited.
• The count parameters specied in the selected prole are applied to all new cell counts.
• If you have already performed a count, loading a new prole from the Results screen applies the count preferences to the current counts results (total cells, viability etc.) and to all new counts.
• If you change any seing that is saved as part of the protocol (size, brightness, or circularity) on the Results screen, the prole name is appended with the (*) symbol.
Countess™ II automated cell counter User Guide
Page 13
Chapter 3 Getting started
Load profile
3
Edit
profile screen
The Auto Lighting function and count parameters are set in the Edit prole screen.
Auto lighting
Auto Lighting function automatically illuminates the sample in the brighteld channel for increased sample-to-sample consistency and decreased user-to-user variability. By default, the Auto Lighting function is turned on; to toggle it on and o, use the Auto Lighting checkbox (see “Add/edit a prole“ on page 14).
Count parameters
Count parameters are set using the parameter sliders, which correspond to a single channel that is selected via the channel selection radio buons.
Size: As you move the slider up, the algorithm includes larger objects in the
count. As you move the slider down, only the smaller objects are counted. = larger objects
Brightness: As you move the slider up, the algorithm includes the brighter
objects in the count. As you move the slider down, only the dimmest of objects are counted.
= brighter objects = dimmer objects
Circularity: As you move the slider up, the algorithm includes more objects with
shapes other than circular in the count. As you move the slider down, only the objects that are perfect circles are counted.
= less circular = more circular
Size and brightness sliders are range sliders.
To adjust the upper and lower boundaries without changing the data range, drag the slider by its middle section (i.e., the slider bar).
To adjust only the upper or the lower boundary, move the upper or the lower handle in the desired direction. This will also change the range of values within which the cells are counted.
• The circularity slider only sets a single threshold value; cells that fall below the
set value are counted, and cells that are beyond this range are excluded. To adjust the threshold for circularity, drag the slider in the desired direction.
= smaller objects
Countess™ II automated cell counter User Guide
13
Page 14
3
Load a
Chapter 3 Getting started
Load profile
profile
1.
2.
Press the Proles buon located on the upper left corner of the screen to open the Proles screen.
Press the desired prole to select, and then press Load.
The instrument will load the count parameters specied in the selected prole and return to the previous screen.
Add/edit a profile
To return to the previous screen without loading the new prole, press
3.
previous buon.
The instrument will keep the saved prole, but return to the previous screen without loading it.
Press the Proles buon located on the upper left corner of the screen to
1.
navigate to the Proles screen.
To add or edit a new prole, select an empty or an existing prole, and then
2.
press Edit. The Edit screen for the selected prole opens.
Note: The Default prole contains default count seings and cannot be edited.
the
14
Select or deselect the Auto Lighting checkbox to turn the Auto Lighting function
3.
ON or OFF.
To dene the new count parameters, select Live or Dead from the Trypan Blue
4.
radio buons.
Adjust the size, brightness, and circularity thresholds using the appropriate
5.
parameter slider (“Count parameters“ on page 13).
Note: Although the Auto FL Threshold checkbox and the FL channel radio
buons are visible on the Edit prole screen, they are not functional on Countess™ II Automated Cell Counters, which only feature a single brighteld channel.
Countess™ II automated cell counter User Guide
Page 15
Chapter 3 Getting started
To assign a name to the new prole or to change the name of the existing prole,
6.
press the Prole name text box. The alpha-numeric keypad opens.
Type in the desired prole name using the alpha-numeric keypad.
7.
To enter symbols, press the symbol (@%&) key. To return to the alpha-numeric keypad, press ABC.
Press Enter to save the name and return to the Edit prole screen.
8.
To return to the Edit prole screen without saving the name, press the close buon.
Load profile
3
Press Save to save the new prole, and then press Close in the conrmation
9.
screen to return to the Proles screen.
To return to the Proles screen without saving, press Cancel.
On the Proles screen, press Load. The instrument will load the count
10.
parameters specied in the selected prole and return to the previous screen.
To return to the previous screen without loading the new prole, press the
11.
previous previous screen without loading it.
buon. The instrument will keep the saved prole, but return to the
Countess™ II automated cell counter User Guide
15
Page 16
4

Prepare sample

Cell count and cell viability assays

Recommendations

Load Countess
chamber slide
To obtain the best results, follow these recommendations:
• Ensure that the cell sample is homogeneously mixed.
• The measurement range extends from 1 × 104–1 × 107 cells/mL, but the optimal range is 1 × 105–4 × 106 cells/mL.
• For accurate results in cell viability assays, ensure that the counting area is covered with the cell suspension and count the cells immediately after staining per the assay protocol.
• Do not press the optical surfaces of the chamber slides. Hold the slides by the edges.
• Take care to avoid forming bubbles in the sample.
• Sterile ltering and centrifugation can be used to remove precipitates common within trypan blue solutions. Alternatively, avoid mixing and vortexing trypan blue stock solutions to allow precipitates to remain at the boom of tube, thereby promoting more accurate cell counts. Also, precipitates can be reduced by gentle heating at 37°C for 10 minutes.
Prepare the sample by adding 10 µL of your cell suspension to 10 µL of 0.4%
1.
trypan blue stain. Mix the sample mixture well by pipeing up and down a few times.
Gently pipet 10 µL of the trypan blue-stained sample into the half moon-shaped
2.
sample loading area. The sample is loaded into the chamber through capillary action.
16
Note: Each chamber in the Countess™ Cell Counting Chamber Slide has a 10-µL
sample capacity. Do not overll the slide chambers.
Let the sample mixture sele in the chamber for 30 seconds, and then insert the
3.
slide into the slide port (see “Instrument exterior components“ on page 10). You will hear a soft click, if the slide is pushed in correctly.
Countess™ II automated cell counter User Guide
Page 17

Count cells

Count procedure

Chapter 4 Cell count and cell viability assays
To remove the slide, push the slide gently into the instrument until it “clicks”
4.
and a spring pushes the slide out. Grasp the slide and pull it out the rest of the way.
Note: After using the Countess™ Cell Counting Chamber Slides, appropriately dispose of them as biohazardous waste. Do not reuse the disposable chamber slides.
Load the Countess™ Cell Counting Chamber Slide with your trypan blue-stained
1.
sample and insert it into the slide port of the instrument as described in “Prepare sample“ on page 16. You will hear a soft click, if the slide is pushed in correctly.
When the slide is inserted, the instrument automatically illuminates the sample,
2.
sets the intensity of brighteld illumination, and auto focuses on the cells.
Note: To turn o the Auto Lighting function, see “Auto lighting“ on page 13.
Count cells
4
Optional: Press the Proles buon and load the desired prole as
3.
described in “Load a prole“ on page 14.
Optional: If desired, press the Focus buon, and then use the Focus slider to
4.
manually bring your sample into focus (see Note on page 17).
If needed, Zoom in on the image while adjusting the focus.
Note: The auto focus algorithm is designed to highlight the dierences between live and dead cells. Therefore, the optimal focus level is where the “live” cells have a light colored center and the “dead” cells are dark throughout. To enable optimal auto focus functionality, you may need to initially rene the focus by adjusting it manually and then seing it as the nominal focus. For more information, see “Set nominal focus“ on page 38.
Optional: Manually adjust exposure using the light source slider.
5.
The light source slider controls the LED intensity, camera gain, and exposure time and it is used for adjusting the image brightness.
Countess™ II automated cell counter User Guide
17
Page 18
Chapter 4 Cell count and cell viability assays
4
Count cells
Press Capture. The instrument temporarily captures the image and displays the
6.
results (total concentration, percentage and concentration of live and dead cells). For more information, see “Chapter 5, “Results“”.

Next steps

• To identify the objects (i.e., cells) counted as live or dead, press More to navigate to the Advanced™ screen (“Identify the cells counted“ on page 19).
• To see the distribution of live and dead cells in a graphical format, press the Graph buon (see “Graph count results“ on page 21).
• To gate the results by object size, brightness, or circularity, rst press More to open the Advanced™ screen, and then press Adjust to navigate to the Adjust screen (“Gate count results“ on page 21).
• To permanently save the results, press Save (see Chapter 7, “Save results“).
• To perform a new count, remove the slide and reinsert it into the instrument to count the sample in the second chamber.
18
Countess™ II automated cell counter User Guide
Page 19
5

View results

Results

Results screen

The Results screen for cell count and cell viability assays displays a composite image of the objects counted and the results of the cell count and cell viability calculations (total concentration, percentage and concentration of live and dead cells).
Note: When performing brighteld (BF) cell counts and cell viability assays, the counting algorithm assumes that you have diluted your cells 1:1 in trypan blue and takes this dilution into account when calculating the total cell concentration. The cell concentration displayed in the Results screen is the original cell concentration before dilution into trypan blue.

Identify the cells counted

Advanced screen

Countess™ II automated cell counter User Guide
The Advanced screen allows you to identify the objects (i.e., cells) counted in each channel and included in the count results for further review. After reviewing the marked objects, you can adjust the threshold for size, brightness, and/or circularity as desired for your application.
19
Page 20
Chapter 5 Results
5
Identify the cells counted

Identify live and dead cells

On the Results screen, click More. The Advanced screen opens.
1.
To identify the cells that are included in the count as “"live"”, press the Live
2.
buon. "“Live"” cells will be circled in green on the screen.
To identify the cells that are included in the count as "“dead"”, press the Dead buon. “"Dead"” cells will be circled in red on the screen.
Note: You may select either or both options. In the example below, both Live and Dead buons are pressed and "“live"” and “"dead"” cells are marked with green and red circles around them, respectively.
20
To unmark the cells identied as "“live"” (green) or "“dead"” (red) on the screen,
3.
press the Live or the Dead buon again, respectively.
Countess™ II automated cell counter User Guide
Page 21

Graph count results

Chapter 5 Results
Graph count results
5

View graph

For cell count and cell viability assays, you can view the distribution of cells (live and/or dead) based on size in a graphical format.
Note: You can view the Graph on Results, Advanced™, and Adjust screens.
To view the graph showing the distribution live and/or dead cells based on cell
1.
size, press the Graph buon.
To view the distribution of only the live or dead cells, check the corresponding
2.
Live or Dead check box on the graph.
The graph will automatically update and display the distribution of cells based on size only in the selected population.
3.
4.

Gate count results

Adjust screen

Gate count results

The Adjust screen contains the controls for gating count results based on size, brightness, and circularity. You can adjust the count parameters before or after performing a count.
1.
2.
Optional: Using the size, brightness, and circularity sliders, adjust the count parameters. As you adjust the count parameters, the count results and the graph will be automatically updated.
To close the graph, press the Graph buon again.
On the Results screen, press More to open the Advanced screen.
Optional: Press the Live and/or the Dead buon to identify the cells in the selected population (see “Identify the cells counted“ on page 19).
Countess™ II automated cell counter User Guide
21
Page 22
Chapter 5 Results
5
Gate count results
On the Advanced screen, press Adjust to open the Adjust screen.
3.
Optional: Press the Graph buon to view the distribution of cells (live and/or
4.
dead) based on size as you gate the count results (see “Graph count results“ on page 21).
Select the channel (Live or Dead) you wish to gate.
5.
Using the size, brightness, and circularity sliders, adjust the count parameters.
6.
Note: For a description of the count parameters and count parameter controls
(i.e., parameter sliders), see “Count parameters“ on page 13.
When nished, press Done to save the changes to count parameters and return
7.
to the Advanced screen.
Press Cancel to return to the Results screen without saving the changes.
On the Advanced screen, press Count to recalculate your results with the new
8.
count parameters.
To save the changes to size, brightness, or circularity parameters to the current
9.
prole or to create a prole with the new count parameters, see “Save as new protocol“ on page 23.
To permanently save your results to an external computer using a USB ash
10.
drive, see Chapter 7, “Save results“.
22
Countess™ II automated cell counter User Guide
Page 23

Save as new protocol

If you have made any changes to the count parameters before or after performing
Edit and save profile as new protocol
1.
a count, the displayed prole name is appended with the (*) symbol and the Advanced results screen displays the New protocol buon, which allows you to save the changes to the current prole or as a separate protocol.
To save the changes to the count parameters to the current prole or to create a
2.
new prole with the edited parameters, press the New protocol buon. The Select prole to edit screen opens.
Chapter 5 Results
Save as new protocol
5
Note: By default, the current prole buon is selected on the Select prole screen. If you are using the Default prole for the count, no prole buon is selected on this screen, because the Default prole cannot be edited.
Select the prole you wish to edit, then press Import seings.
3.
Note: You can select only a previously saved or an empty prole. The Default
prole cannot be edited.
Countess™ II automated cell counter User Guide
23
Page 24
Chapter 5 Results
5
Save as new protocol
The Edit prole screen opens and displays the edited count parameters from the
4.
Adjust screen (21).
Note: If you have selected a prole that had been previously saved, the name of that prole populates the Prole name text box by default. Otherwise, the textbox remains empty.
To change the name of the selected prole, press the Prole name text box and
5.
enter the desired name using the alpha-numeric keypad.
(Optional): If desired, make additional changes to the prole and the count
6.
parameters as described in “Add/edit a prole“ on page 14.
Click Save to save the new prole seings and return to the Results page for the
7.
last count. The prole name will be displayed without the “(*)” symbols.
Click Cancel to return to the Results page for the last count without saving the changes to the prole. The prole name will be displayed with the (*) symbol, indicating that the count parameters for the selected prole had been altered, but not yet saved.
24
Countess™ II automated cell counter User Guide
Page 25
6

Dilution calculator

Dilution calculator function allows you to calculate the volume of cell sample and buer needed to reach a desired concentration using the count results.

Calculate dilution

1.
Dilution calculator
On the Results screen, press Dilution Calculator to open the Dilution calculator screen.
Press the cell type buon located to the right of the current concentration, then
2.
select the count result you wish to use for the dilution calculation from the dropdown. The current concentration changes to reect the results for the cell type selected.
Available Cell type options are Live cells or Total cells. By default, Live cells is selected.
Countess™ II automated cell counter User Guide
25
Page 26
Chapter 6 Dilution calculator
6
Calculate dilution
Enter the desired cell concentration (“What is your desired cell concentration?”):
3.
Press the value text box, then enter the value using the number pad.
a.
You can enter one digit to the left of the decimal separator (integer part) and one to the right (fractional part). If you do not enter the fractional part, the software enters a 0 by default.
Press Enter or touch anywhere outside the number pad to close it.
Press the exponent buon, then select the exponent value for the desired
b.
concentration.
By default, the exponent is (n–1), where n = the exponent value of the current count. The maximum selectable exponent is same exponent value as the current count.
Pressing an exponent buon selects that exponent and closes the window. Pressing anywhere outside the window keeps the previously selected exponent and closes the window.
Press the total volume text box (“How many mL do you need?”), then enter the
4.
total volume of the sample you wish to have at the new concentration using the number pad.
By default, the total volume box is blank. The maximum volume you can enter is
999.9 mL and you can use only a single decimal. Press Enter or touch anywhere outside the number pad to close the number pad.
26
Countess™ II automated cell counter User Guide
Page 27
Chapter 6 Dilution calculator
When you have made valid entries for cell type, desired cell concentration, and
5.
total volume, and closed the last popup window, the boom line of the dilution calculator displays the volumes of cell solution and buer needed.
If you enter a combination of values that is not valid (e.g., desired concentration is greater than the current concentration), the results line remains on (or return to) the blank state, and screen displays a warning message.
Calculate dilution
6
If you make any changes to any of the input areas above, the results are recalculated automatically upon closing the popup window.
Press Done or the Back buon to return to the main Results screen.
6.
Countess™ II automated cell counter User Guide
27
Page 28
7

Save count results

Save results

Save screen

Save procedure

The Countess™ II FL Automated Cell Counter allows you to save your data and images using a USB ash drive. To save your experiment, you can choose from the following options, in any combination:
Result: Saves the Results screen as it is displayed on the instrument, with or without the Graph, in the selected image format (JPEG, BMP, PNG, or TIFF).
Images: Saves only the raw captured image in the selected image format (JPEG, BMP, PNG, or TIFF).
Data: Saves the data from the experiment as a CSV le (comma separated values). The CSV format allows for processing or re-displaying results with any third party software or spreadsheet program. For more information on the CSV le format, see “Appendix D, “CSV le format“”.
Report: Saves a printer-friendly report of the results, graph(s), and image in the selected format (PDF, PNG, or JPEG). For more information, see “Report“ on page 31.
Note: If you wish to save your results with the Graph showing the distribution of cells based on cell size or uorescence intensity, make sure that the desired graph is displayed on the Results screen.
To save your data, insert the Countess™ II USB drive (or equivalent) into an
1.
available USB port on the instrument (see “Instrument exterior components“ on page 10).
Note: The USB ports located in the front and the back of the instrument function the same. However, the rst USB drive connected will be the preferred saving location.
28
On the Results screen, press Save to navigate to the Save screen.
2.
Countess™ II automated cell counter User Guide
Page 29
Chapter 7 Save results
Save count results
To assign a name to your experiment, press the File name text eld. The alpha-
3.
numeric keypad opens.
Enter the le name using the alpha-numeric keypad.
4.
To enter symbols, press the symbol (@%&) key. To return to the alpha-numeric keypad, press ABC.
Press Enter to save the name and return to the Save screen.
5.
To return to the Save screen without saving the name, press the close buon.
7
Select the desired mode(s) to save your experiment (Result, Images, Data,
6.
Report). You can select an individual mode (e.g., Result only) or any combination of modes (e.g., Result, Images, Data, and / or Report).
In the example below, Data and Report are selected.
By default, Result and Images are saved as JPEG les, and Report is saved as
7.
PDF.
To choose a dierent le format, press the le type buon. The Choose le type screen opens.
Note: Data can only be saved as a CSV le.
Countess™ II automated cell counter User Guide
29
Page 30
Chapter 7 Save results
7
Save count results
Press to select the desired le type. Available options are JPEG, BMP, PNG, and
8.
TIFF.
After you make your selection, the instrument returns to the Save screen. To return to the Save screen without changing the le format, press the close
buon.
Press Save to save your experiment in the selected mode(s) in the USB drive.
9.
30
Press Close and then transfer the USB drive to the desired location.
10.
Countess™ II automated cell counter User Guide
Page 31

Report

Chapter 7 Save results
Report
7
Report file
The Report function allows you to save a printer-friendly report of the results, graphs, and images in the selected format (PDF, PNG, or JPEG).
You can create reports using the Report dropdown as described in “Save count results“ on page 28.
1
The top section of the Report contains a table with the results as displayed on the Countess™ Results screen, showing the concentration of the sample, and the percentage and number for the total, Live, and Dead channels.
2
Below the results table, the report contains the “number of cells vs. cell size” graph.
3
Under the graph, the report contains the brightfield count image, with the live and dead cells identified by the green and red circles, respectively.
4
At the bottom, the report displays the profile information used to gate these images.
Countess™ II automated cell counter User Guide
31
Page 32
8

Overview

Instrument settings

Instrument settings screen

To access the Instrument Seings screen, press the Instrument Seings buon on the Home page (“Turn ON the instrument“ on page 11).
In the Instrument Seings screen, you can:
• perform software update (“Software update“ on page 32)
• set the date and time (“Date/Time“ on page 34)
Note: Although the Change Light Cube buon is visible on the Instrument Seings screen, this buon is not functional on the Countess™ II Automated Cell Counter, which is a brighteld-only instrument and does not support EVOS™ light cubes.

Software update

Guidelines for software update

32
• The USB drive used for transferring the software update le must be FAT32 formaed; verify this before proceeding. If necessary, reformat the USB drive to
FAT32 following the recommended procedure for your operating system.
Note: Reformaing the USB drive will result in the loss of all les. Back up the les in the USB drive prior to reformaing.
• The software update le must be saved on the top level of the USB drive, not within a folder or a subfolder.
• The software update le must be uncorrupted during transfer. Do not rename, zip, or compress the software update le.
Countess™ II automated cell counter User Guide
Page 33
Update the Countess™ II/II FL software
Chapter 8 Instrument settings
Go to www.thermosher.com/countessupdate, and download the latest
1.
Countess™ II/II FL software version to your desktop.
Note: The software update le has a version-specic name followed by the extension .lft (e.g., Countess™_II_v_1_0_202.lft for software version 1.0.202).
Copy the software update le onto the USB drive, making sure that it is saved on
2.
the top level and not hidden within a folder.
Insert the USB drive into one of the USB ports of the instrument (see “Instrument
3.
exterior components“ on page 10).
Press the Instrument Seings buon on the Home page (“Turn ON the
4.
instrument“ on page 11) to open (“Overview“ on page 32).
Select Software Update from the Instrument Seings menu. The instrument
5.
scans the USB drive for the latest software version.
the Instrument Seings screen
Software update
8
When prompted, select Update Now.
6.
Once the update has completed, restart the instrument.
7.
Countess™ II automated cell counter User Guide
33
Page 34
Chapter 8 Instrument settings
8
Date/Time
Date/Time

Set the date and time

Press the Instrument Seings buon on the Home page (“Turn ON the
1.
instrument“ on page 11) to open the Instrument Seings screen.
Press Date/Time on the Instrument Seings menu to open the Date/Time screen.
2.
Select the Date format you wish to use.
3.
Press any Date text box (MM, DD, or YYYY) to open the Edit Date keypad.
4.
Using the keypad, enter the date into Month, Day, and Year text boxes, pressing
5.
Enter after each entry.
After you are nished entering the date, press the close buon to return
6.
Date/Time screen.
to the
34
Countess™ II automated cell counter User Guide
Page 35
Chapter 8 Instrument settings
Select the Time format you wish to use. Available options are 12 Hour and
7.
24 Hour formats.
Press any Time text box (Hours or Minutes) to open the Edit Time keypad.
8.
Date/Time
8
Using the keypad, enter the time into Hours and Minutes text boxes, pressing
9.
Enter after each entry.
After you are nished entering the time, press the close buon to return to
10.
the Date/Time screen.
If you have selected the 12 Hour format, select AM or PM.
11.
Press Save to set the Time and Date and return to the Instrument Seings screen.
12.
Press Cancel to return to the Instrument Seings screen without saving your changes.
Countess™ II automated cell counter User Guide
35
Page 36
9

Instrument care

Maintenance

General guidelines for care

Power supply

• Use the appropriate cleaning solutions for each component, as indicated in the cleaning procedures in “Clean the Countess™ II automated cell counter“ on page 37.
• If liquid spills on the instrument, turn o the power immediately and wipe dry.
Always use the correct power supply. The power adaptor specications appear on the serial number label (boom of the instrument) and in the Technical specications section of this user guide (“Technical specications“ on page 43). Damage due to an incompatible power adaptor is not covered by warranty.
CAUTION! Never disassemble or service the instrument yourself. Do not
remove any covers or parts that require the use of a tool to obtain access to moving parts. Operators must be trained before being allowed to perform the hazardous operation. Unauthorized repairs may damage the instrument or alter its functionality, which may void your warranty. Contact your local distributor to arrange for service.
IMPORTANT! If you have any doubt about the compatibility of decontamination or
cleaning agents with parts of the equipment or with material contained in it, contact Technical Support () or your local distributor for information.
36
Countess™ II automated cell counter User Guide
Page 37

Clean the Countess™ II automated cell counter

Clean the Countess™ II automated cell counter
Chapter 9 Maintenance
9

Introduction

Clean the touch­screen
We recommend cleaning the Countess™ II Automated Cell Counter periodically to prevent the buildup of dust and dirt that might reduce its performance and cause contamination.
CAUTION! To avoid electrical shock, always turn o the Countess
Automated Cell Counter and unplug the power cord before cleaning or decontaminating the instrument.
II
CAUTION! All biological samples and materials that come into contact with
them have the potential to transmit infectious diseases and are considered biohazardous. Follow all applicable local, state/provincial, and/or national regulations. Wear appropriate protective eyewear, clothing, and gloves.
IMPORTANT! Using a cleaning or decontaminating method other than that specied
by the manufacturer may result in damage to the instrument.
• Wipe the touch-screen of the Countess™ II Automated Cell Counter using a soft, lint-free cloth moistened with an LCD cleaning solution. Do not apply excessive force during cleaning. Wipe the touch-screen dry immediately after cleaning.
• Ensure that the cleaning solution does not enter the power buon, the power inlet, the slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical shock when the instrument is plugged in.
• Do not use abrasive cleaning solutions or material to prevent the touch-screen from geing scratched.

Clean the instrument case

Decontaminate the instrument

Countess™ II automated cell counter User Guide
• Wipe the instrument case of the Countess™ II Automated Cell Counter using a soft, lint-free cloth moistened with distilled water. Wipe the instrument dry immediately after cleaning.
• Ensure that water or other cleaning solutions do not enter the power buon, the power inlet, the slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical shock when the instrument is plugged in.
• Wipe the instrument case of the Countess™ II Automated Cell Counter using a soft, lint-free cloth moistened with 70% alcohol. Wipe the instrument dry immediately after cleaning.
• Avoid using a bleach solution, because it may leave a residue of bleach crystals on the instrument.
• Ensure that water or other cleaning solutions do not enter the power buon, the power inlet, the slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical shock when the instrument is plugged in.
37
Page 38
Focus is not optimal Focus is optimal
Chapter 9 Maintenance
9

Set nominal focus

Set nominal focus

Overview

Set nominal focus
Nominal focus is the Z-point (i.e., depth) around which the auto focus function searches to provide ne focus to the sample.
The auto focus algorithm of the Countess™ II FL Automated Cell Counter is designed to highlight the dierences between live and dead cells in the brighteld channel. The optimal focus level is where the “live” cells have a light colored center and the “dead” cells are dark throughout (see examples below).
To enable optimal auto focus functionality, you may need to initially rene the brighteld focus by adjusting it manually and then seing the nominal focus. This allows the auto focus function to have a set point from which to focus on the cells in subsequent samples.
Prepare the sample by adding 10 µL of cell suspension to 10 µL of 0.4% trypan
1.
blue stain. Mix the sample mixture well by pipeing up and down a few times.
Load 10 µL of the sample mixture into the Countess™ Cell Counting Chamber
2.
Slide (“Load Countess™ chamber slide“ on page 16). Let the sample mixture sele for 30 seconds to ensure a uniform focal plane.
Insert the sample slide into the slide port of the instrument, making sure that the
3.
side containing the sample is inserted completely.
When the slide is inserted, the instrument automatically illuminates the sample,
4.
sets the intensity of the brighteld light source, and auto focuses on the cells.
To manually adjust the focus, press the Focus buon.
5.
38
Countess™ II automated cell counter User Guide
Page 39
Chapter 9 Maintenance
Set nominal focus
Use the Focus™ slider or the plus and minus buons to rene the brighteld
6.
focus.
Note: If needed, Zoom in on the image to adjust focus or lighting.
After nding the optimal focus, press Set to set the nominal focus.
7.
Once the focus has been set, the Set buon on the focus slider becomes inactive, conrming that the focus seing has been stored.
9
Countess™ II automated cell counter User Guide
39
Page 40
A
Troubleshooting
Note: The software for the Countess™ II is updated regularly. If you are having any
issues with your experiments, rst check the website to see if a new software version is available. You can download the most recent version of the software from www.thermosher.com/countessupdate. You can also register your Countess™ II instrument at www.thermosher.com/registercountess to be informed of any future software updates.
Problem
Autofocus does not seem to focus on the cells very well
Some cells appear on the image but are not included in the count
Images are very bright and washed out Enable Auto Lighting from the Profiles menu, or decrease
• Make sure there are no bubbles or debris visible on the screen that could interfere with the autofocus and make it more difficult to get the sample in the correct focal plane.
• Ideally, the live cells should have bright centers compared to the dead cells, which are dark throughout (“Count procedure“ on page 17).
• Setting the nominal focus will improve autofocus consistency with future slides. To set the nominal focus, see “Set nominal focus“ on page 38.
• Adjust the size, brightness, and circularity gates for both live and dead cells to include all of the cells in the count (“Graph count results“ on page 21).
• After including all of the cells in the count, you can narrow the count criteria, if you wish to exclude cells of a certain size or certain brightness.
• When the gates are fully maximized, the CSV should indicate 0–60 for cell size and 0–255 for brightness.
the bright field light intensity before counting the cells.
Possible solutions
Getting incorrect concentration for the Countess™ test beads
40
• The beads can settle quickly in solution, which will affect the concentration reading.
• Vortex the bead stock on high for a full 30 seconds to resuspend, and add 10 µL of the bead suspension to 10 µL of trypan blue without delay.
• Pipet the bead and trypan blue mixture up and down several times to make sure it is well mixed, and immediately load 10 µL into the slide.
Countess™ II automated cell counter User Guide
Page 41
Appendix A Troubleshooting
Set nominal focus
Problem Possible solutions
A
Variable counts for the same sample of cells
Variable counts when performing replicate counts of the same slide
• If you are pipetting different samples from the same cell sample, the variability could be due to pipetting or mixing.
• Use recently calibrated pipettors and make sure that the cells are well suspended by pipetting up and down several times before adding trypan blue.
• Pipet the bead and trypan blue mixture up and down several times to make sure it is well mixed, and load 10 µL into the slide without delay.
• If you are counting replicates of the exact same slide, visually inspect that all cells are counted correctly in the image.
• There may be a slightly different field of view each time a slide is inserted. Depending on the concentration and uniformity of the cells, this will cause some variability when performing replicate counts of the same slide, although it should be less than 10%.
• When counting fewer cells, a small field of view change for only a small number of cells can have a larger affect. Count cells at a higher concentration to reduce variability.
• Make sure that you do not shake or agitate the slide between counts.
Abnormally high percentage of dead cells or live cells counted as dead
USB drive not recognized by the instrument
• Ensure that the cells are focused correctly so that live cells have bright centers and dead cells are dark throughout (see “Count procedure“ on page 17). If the cells are not well focused and look dark on the screen, the Countess™ II will count them as dead cells.
• If cells are well focused, have bright centers, and are being counted as dead, confirm that they are within the appropriate cell size range and try adjusting the settings.
• If cells are exposed to trypan blue for a long period of time, viability could be affected so slide should be prepared and counted fresh each time for best results.
• Gate out the debris using the size, brightness, and circularity sliders.
• The USB drive must be FAT32 formatted to be recognized by the instrument. If it is not, reformat the USB drive to FAT32 (“Software update“ on page 32).
• Try another correctly formatted USB drive.
Countess™ II automated cell counter User Guide
41
Page 42
Appendix A Troubleshooting
A
Set nominal focus
Problem Possible solutions
Unable to update the Countess™ software
• Make sure the USB drive is formatted to FAT32. If it is not, reformat the USB drive to FAT32 before transferring the files onto the USB drive for software update.
• The update file must sit on the top level of the USB drive, not within a folder or a subfolder.
• File cannot be renamed in any way.
• File cannot be zipped or compressed during distribution. It must be uncorrupted during transfer and have a .lft suffix.
• If needed, check that the USB port is functional by testing a USB mouse.
42
Countess™ II automated cell counter User Guide
Page 43
B
Technical specifications
Product specifications

Physical characteristics

Technical
specifications
Instrument type
Instrument dimensions 9 (W) × 5½ (D) × 9 inches (H)
Weight 8 lbs
Operating power 100–240 VAC, 0.58 A MAX
Frequency 50/60 Hz
Electrical input 12 VDC, 2 A
Installation site Indoor use only, Class A Environments (i.e., non-residential
Operating temperature 4–32°C (40–90°F)
Operating humidity <80% (non-condensing)
Processing time
Sample concentration range 1 × 104–1 × 107 cells/mL
Particle/cell diameter range 4–60 µm (particles); 7–60 µm (cells)
Required sample volume 10 µL
Benchtop cell counter and suspension cell-based assay
platform
or light industrial); Pollution degree 2.
~15 seconds
Firmware Countess™ Automated Cell Counting Platform Software
USB Drive 4 Gigabyte

Optics

Analysis slide

Countess™ II automated cell counter User Guide
Optics
Camera 5 Mega pixels, 2.5× Optical Magnification
Material
Dimensions 25 mm (W) × 75 mm (D) × 1.7 mm (H)
Chamber volume 10 µL
1 channel (brightfield)
Poly(methyl methacrylate) (PMMA)
43
Page 44
C

Countess™ II automated cell counter

The Countess™ II instrument and instrument accessories are available from Thermo Fisher Scientic. For more information, visit www.thermosher.com or contact Technical Support (“Customer and technical support“ on page 56).
Ordering information
Countess™ II Automated Cell Counter 1 each AMQAX1000
Countess™ II power cord with four adapter cords 1 each AMEP4716
Countess™ II USB drive 1 each A25751
Countess™ II Disposable Slide Holder 1 each AMEP4745

Accessory products

The following products can be used with the Countess™ II Automated Cell Counter and are available separately from Thermo Fisher Scientic. For more information, visit www.thermosher.com or contact Technical Support (“Customer and technical support“ on page 56).
Countess™ Cell Counting Chamber Slides, 50 Slides (100 counts)
Countess™ Cell Counting Chamber Slides, 500 Slides (1000 Counts)
Product
Product
Quantity Cat. no.
Quantity Cat. no.
[1]
1 box
10 boxes
[1]
C10228
C10312
44
Countess™ Cell Counting Chamber Slides, 1250 Slides (2500 Counts)
Countess™ Cell Counting Chamber Slides, 2500 Slides (5000 Counts)
Countess™ Cell Counting Chamber Slides, 5000 Slides (10,000 Counts)
Countess™ Test Beads (1 × 106 beads/mL) 1 mL C10284
Trypan blue stain (0.4 %) 2 × 1 mL T10282
[1]
Each box of Countess™ Cell Counting Chamber Slides contains 50 slides and 2 × 1 mL vials of trypan blue (0.4%), sufficient for 100 counts.
Countess™ II automated cell counter User Guide
25 boxes
50 boxes
100 boxes
[1]
[1]
[1]
C10313
C10314
C10315
Page 45
D
CSV file format, explained
CSV file format

Overview

Category
General A Number Sequential sample run number
Trypan
Blue/Brightfield
A comma-separated values (CSV) le stores tabular data (numbers and text) in plain­text form. Plain text means that the le is a sequence of characters, with no data that has to be interpreted as binary numbers. A CSV le can be opened with any third party software or spreadsheet program. The table below describes the categories of the Countess™ II data saved as a CSV le and opened with a spreadsheet program.
Column Name Description
B File Name Name of file
C Date & Time Date and time of sample run
D Mode BF-Brightfield
E Total Concentration Concentration of the entire sample
F Total cells counted Total number of cells counted in the sample
G Live concentration Concentration of just the "live" portion of the sample
H Live cells counted Total number of "live" cells counted
I Dead concentration Concentration of just the "dead" portion of the sample
J Dead cells counted Total number of "dead" cells counted
K Viability (%)
L Average size (um) Average cell size in microns
General Details X Focus value Focal position number
Y BF Light intensity Brightfield light intensity value from 0–100%
Trypan
Blue/Brightfield
Count Parameters
Countess™ II automated cell counter User Guide
Z Live Size min Minimum size of "live" cells in microns
AA Live Size max Maximum size of "live" cells in microns
AB Live Brightness min "Live" adjustment slider value for minimum brightness
AC Live Brightness max "Live" adjustment slider value for maximum brightness
AD Live Circularity "Live" adjustment slider value for circularity
AE Dead Size min Minimum size of "dead" cells in microns
Percent viability of the sample based on trypan blue staining
45
Page 46
Appendix D CSV file format
D
CSV file format, explained
Category Column Name Description
Trypan
Blue/Brightfield
Count Parameters
AF Dead Size max Maximum size of "dead" cells in microns
AG Dead Bright min "Dead" adjustment slider value for minimum brightness
AH Dead Bright max "Dead" adjustment slider value for maximum brightness
AI Dead Circ "Dead" adjustment slider value for circularity
46
Countess™ II automated cell counter User Guide
Page 47
E

Safety alert words

Safety
WARNING! GENERAL SAFETY. Using this product in a manner not specied
in the user documentation may result in personal injury or damage to the instrument or device. Ensure that anyone using this product has received instructions in general safety practices for laboratories and the safety information provided in this document.
Before using an instrument or device, read and understand the safety
·
information provided in the user documentation provided by the manufacturer of the instrument or device. Before handling chemicals, read and understand all applicable Safety Data
·
Sheets (SDSs) and use appropriate personal protective equipment (gloves, gowns, eye protection, and so on). To obtain SDSs, see the “Documentation and Support” section in this document.
Four safety alert words appear in this document at points where you need to be aware of relevant hazards. Each alert word—IMPORTANT, CAUTION, WARNING, DANGER—implies a particular level of observation or action, as dened below:
IMPORTANT! – Provides information that is necessary for proper instrument
operation, accurate installation, or safe use of a chemical.
CAUTION! – Indicates a potentially hazardous situation that, if not avoided,
may result in minor or moderate injury. It may also be used to alert against unsafe practices.
WARNING! – Indicates a potentially hazardous situation that, if not avoided,
could result in death or serious injury.
DANGER! – Indicates an imminently hazardous situation that, if not avoided,
will result in death or serious injury. This signal word is to be limited to the most extreme situations.
Except for IMPORTANT! safety alerts, each safety alert word in this document appears with an open triangle gure that contains a hazard symbol. These hazard symbols are identical to the hazard symbols that are axed to the instruments (see Safety symbols in Appendix E).
Countess™ II automated cell counter User Guide
47
Page 48
Appendix E Safety
E

Electrical symbols

Electrical symbols
The following table describes the electrical symbols that may be displayed.
Symbol Description
Indicates the On position of the main power switch.
Indicates the Off position of the main power switch.
Indicates a standby switch by which the instrument is switched on to the Standby condition. Hazardous voltage may be present if this switch is on standby.
Indicates the On/Off position of a push-push main power switch.
Indicates a terminal that may be connected to the signal ground reference of another instrument. This is not a protected ground terminal.

Safety symbols

Indicates a protective grounding terminal that must be connected to earth ground before any other electrical connections are made to the instrument.
Indicates a terminal that can receive or supply alternating current or voltage.
Indicates a terminal that can receive or supply alternating or direct current or voltage.
The following table describes the safety symbols that may be displayed. Each symbol may appear by itself or in combination with text that explains the relevant hazard (see “Safety labels on instruments”). These safety symbols may also appear next to DANGERS, WARNINGS, and CAUTIONS that occur in the text of this and other product-support documents.
Symbol
Caution, risk of danger. Consult the manual for further safety information.
Description
48
Caution, risk of electrical shock.
Countess™ II automated cell counter User Guide
Page 49
Symbol Description
Caution, hot surface or other high-temperature hazard.
Caution, laser.
Caution, moving parts.
Caution, potential biohazard.
Appendix E Safety

Environmental symbols

E
Environmental symbols
The following symbol applies to all Thermo Fisher Scientic electrical and electronic products placed on the European market after August 13, 2005.
Symbol
Caution, ultraviolet light.
Description
Do not dispose of this product as unsorted municipal waste. Follow local municipal waste ordinances for proper disposal provisions to reduce the environmental impact of waste electrical and electronic equipment (WEEE).
European Union customers:
Call your Customer Service representative for equipment pick-up and recycling. See www.thermofisher.com for a list of customer service offices in the European Union.
Countess™ II automated cell counter User Guide
49
Page 50
Appendix E Safety
E

Safety labels on instruments

Safety labels on instruments
The following CAUTION, WARNING, and DANGER statements may be displayed Thermo Fisher Scientic instruments in combination with the safety symbols described in the preceding section.
Symbol English Français
CAUTION! Hazardous chemicals. Read the Safety
Data Sheets (SDSs) before handling.
CAUTION! Hazardous waste. Refer to SDS(s) and local regulations for handling and disposal.
DANGER! High voltage. DANGER! Haute tension.
WARNING! To reduce the chance of electrical
shock, do not remove covers that require tool access. No user-serviceable parts are inside. Refer servicing to Thermo Fisher Scientific qualified service personnel.
DANGER! Class 3B visible and/or invisible laser radiation present when open. Avoid exposure to beam.
CAUTION! Moving parts. Crush/pinch hazard. ATTENTION! Pièces en mouvement, risque de
ATTENTION! Produits chimiques dangereux. Lire
les fiches techniques de sûreté de matériels avant toute manipulation de produits.
ATTENTION! Déchets dangereux. Lire les fiches techniques de sûreté de matériels et la régulation locale associées à la manipulation et l’élimination des déchets.
AVERTISSEMENT! Pour éviter les risques d’électrocution, ne pas retirer les capots dont l’ouverture nécessite l’utilisation d’outils. L’instrument ne contient aucune pièce réparable par l’utilisateur. Toute intervention doit être effectuée par le personnel de service qualifié venant de Thermo Fisher Scientific.
DANGER! Rayonnement visible ou invisible d’un faisceau laser de Classe 3B en cas d’ouverture. Evitez toute exposition au faisceau.
pincement et/ou d’écrasement.

General instrument safety

WARNING! PHYSICAL INJURY HAZARD. Use this product only as specied
in this document. Using this instrument in a manner not specied by Thermo Fisher Scientic may result in personal injury or damage to the instrument.

Operating the instrument

50
Ensure that anyone who operates the instrument has:
• Received instructions in both general safety practices for laboratories and specic safety practices for the instrument.
• Read and understood all applicable Safety Data Sheets (SDSs).
Countess™ II automated cell counter User Guide
Page 51
Appendix E Safety
General instrument safety
E

Safety precautions

• Do not install the instrument in heavy humidity, such as a greenhouse or an incubator, to avoid a danger of electric shock. If water or other material enters the instrument, the adaptor, or power inlet, disconnect the power cord and contact a service person.
• Do not press the main plug or power cord with wet hands.
• Always ensure that the power supply input voltage matches the voltage available in your location.
• Do not install the instrument on a slant or a place prone to vibrations, which induces the risk of instrument malfunction or damage of the instrument.
• Never insert any objects into the air vents of the instrument as this could result in electrical shock, personal injury, and equipment damage.
• Plug the power cord rmly into the wall outlet and the instrument.
• To avoid potential shock hazard, make sure that the power cord is properly grounded.
• Be sure to position the equipment such that it is easy to disconnect the instrument.
• Turn o the instrument before unplugging the power cord and/or moving the instrument.
• If the instrument is broken or dropped, disconnect the power cord and contact a service person. Do not disassemble the instrument.
• Use only authorized accessories (adaptor, power cord, and USB drive).
• If the instrument emits smoke, disconnect the power cord from the wall outlet and contact a service person.

Cleaning or decontaminating the instrument

Removing covers or parts of the instrument

CAUTION! Using cleaning or decontamination methods other than those
recommended by the manufacturer may compromise the safety or quality of the instrument.
CAUTION! PHYSICAL INJURY HAZARD The instrument is to be serviced
only by trained personnel or vendor specied in the user guide.
Countess™ II automated cell counter User Guide
51
Page 52
Appendix E Safety
E

Chemical safety

Chemical safety
WARNING! GENERAL CHEMICAL HANDLING. To minimize hazards,
ensure laboratory personnel read and practice the general safety guidelines for chemical usage, storage, and waste provided below. Consult the relevant SDS for specic precautions and instructions:
Read and understand the Safety Data Sheets (SDSs) provided by the
·
chemical manufacturer before you store, handle, or work with any chemicals or hazardous materials. To obtain SDSs, see the "Documentation and Support" section in this document. Minimize contact with chemicals. Wear appropriate personal protective
·
equipment when handling chemicals (for example, safety glasses, gloves, or protective clothing). Minimize the inhalation of chemicals. Do not leave chemical containers open.
·
Use only with sucient ventilation (for example, fume hood). Check regularly for chemical leaks or spills. If a leak or spill occurs, follow
·
the manufacturer cleanup procedures as recommended in the SDS. Handle chemical wastes in a fume hood.
·
Ensure use of primary and secondary waste containers. (A primary waste
·
container holds the immediate waste. A secondary container contains spills or leaks from the primary container. Both containers must be compatible with the waste material and meet federal, state, and local requirements for container storage.) After emptying a waste container, seal it with the cap provided.
·
Characterize (by analysis if needed) the waste generated by the particular
·
applications, reagents, and substrates used in your laboratory. Ensure that the waste is stored, transferred, transported, and disposed of
·
according to all local, state/provincial, and/or national regulations.
IMPORTANT! Radioactive or biohazardous materials may require special
·
handling, and disposal limitations may apply.
WARNING! HAZARDOUS WASTE (from instruments). Waste produced by
the instrument is potentially hazardous. Follow the guidelines noted in the preceding General Chemical Handling warning.
WARNING! 4L Reagent and Waste Bole Safety. Four-liter reagent and waste
boles can crack and leak. Each 4-liter bole should be secured in a low-density polyethylene safety container with the cover fastened and the handles locked in the upright position.

Chemical waste safety

Chemical waste hazard

52
CAUTION!
local regulations for handling and disposal.
HAZARDOUS WASTE. Refer to Safety Data Sheets (SDSs) and
Countess™ II automated cell counter User Guide
Page 53
Appendix E Safety

Electrical safety

E

Chemical waste safety guidelines

Waste disposal

To minimize the hazards of chemical waste:
• Read and understand the Safety Data Sheets (SDSs) provided by the manufacturers of the chemicals in the waste container before you store, handle, or dispose of chemical waste.
• Provide primary and secondary waste containers. (A primary waste container holds the immediate waste. A secondary container contains spills or leaks from the primary container. Both containers must be compatible with the waste material and meet federal, state, and local requirements for container storage.)
• Minimize contact with chemicals. Wear appropriate personal protective equipment when handling chemicals (for example, safety glasses, gloves, or protective clothing). For additional safety guidelines, consult the SDS.
• Minimize the inhalation of chemicals. Do not leave chemical containers open. Use only with adequate ventilation (for example, fume hood). For additional safety guidelines, consult the SDS.
• Handle chemical wastes in a fume hood.
• After emptying the waste container, seal it with the cap provided.
• Dispose of the contents of the waste tray and waste bole in accordance with good laboratory practices and local, state/provincial, or national environmental and health regulations.
If potentially hazardous waste is generated when you operate the instrument, you must:
• Characterize (by analysis, if necessary) the waste generated by the particular applications, reagents, and substrates used in your laboratory.
• Ensure the health and safety of all personnel in your laboratory.
• Ensure that the instrument waste is stored, transferred, transported, and disposed of according to all local, state/provincial, and/or national regulations.
IMPORTANT! Radioactive or biohazardous materials may require special handling,
and disposal limitations may apply.
Electrical safety
Part
General DANGER! ELECTRICAL SHOCK HAZARD. Severe electrical shock
Fuses WARNING! FIRE HAZARD. For continued protection against the
Countess™ II automated cell counter User Guide
Symbol Description
can result from operating the Countess™ II Automated Cell Counter or Countess™ II FL Automated Cell Counter without its instrument panels in place. Do not remove instrument panels. High-voltage contacts are exposed when instrument panels are removed from the instrument.
risk of fire, replace fuses only with fuses of the type and rating specified for the instrument.
53
Page 54
Appendix E Safety
E

Biological hazard safety

Part
Power DANGER! ELECTRICAL HAZARD. Grounding circuit continuity is
Overvoltage rating The Countess™ II Automated Cell Counter and Countess™ II FL
Biological hazard safety
WARNING! Biological samples such as tissues, body uids, and blood of
humans and other animals have the potential to transmit infectious diseases. Follow all applicable local, state/provincial, and/or national regulations. Wear appropriate protective eyewear, clothing, and gloves. Read and follow the guidelines in these publications.
Symbol Description
vital for the safe operation of equipment. Never operate equipment with the grounding conductor disconnected.
DANGER! ELECTRICAL HAZARD. Use properly configured and approved line cords for the voltage supply in your facility.
DANGER! ELECTRICAL HAZARD. Plug the system into a properly grounded receptacle with adequate current capacity.
Automated Cell Counter have an installation (overvoltage) category of II, and are classified as portable equipment.
ATTENTION! BIOHAZARD. Les échantillons biologiques tels que les tissus, les
uides corporels et le sang des humains et d’autres animaux ont la possibilité de transmere des maladies infectieuses. Suivre tous les règlements municipaux,
provinciaux/provincial et / ou nationales en vigueur. Porter des lunees de protection approprié, des vêtements et des gants.
In the U.S.:
• U.S. Department of Health and Human Services, Biosafety in Microbiological and Biomedical Laboratories (BMBL), 5th Edition, HHS Publication No. (CDC) 21-1112, Revised December 2009; found at: www.cdc.gov/labs/pdf/CDC-
BiosafetymicrobiologicalBiomedicalLaboratories-2009-P.pdf
• Occupational Safety and Health Standards, Bloodborne Pathogens (29 CFR§1910.1030; www.access.gpo.gov/nara/cfr/waisidx_01/29cfr1910a_01.html)
• Your company’s/institution’s Biosafety Program protocols for working with/handling potentially infectious materials.
• Additional information about biohazard guidelines is available at: www.cdc.gov
54
Countess™ II automated cell counter User Guide
Page 55

Safety and electromagnetic compatibility (EMC) standards

Appendix E Safety
In the EU:
• Check your local guidelines and legislation on biohazard and biosafety precaution, and the best practices published in the World Health Organisation (WHO) Laboratory Biosafety Manual, third edition
www.who.int/csr/resources/publications/biosafety/WHO_CDS_CSR_LYO_2004 _11/en/
Safety and electromagnetic compatibility (EMC) standards
E
Symbol
Description
U.S. and Canadian safety standards.
The CSA C/US Mark signifies that the product meets applicable U.S. and Canadian standards, including those from CSA, CSA America, ANSI, ASME, ASSE, ASTM, NSF and UL.
European safety and EMC standards.
The CE Mark symbolizes that the product conforms to all applicable European Community provisions for which this marking is required. Operation of the instrument is subject to the conditions described in this manual.
The protection provided by the instrument may be impaired if the instrument is used in a manner not specified by Thermo Fisher Scientific.
Australian EMC standards
The C-Tick Mark indicates conformity with Australian and New Zealand standards for electromagnetic compatibility.
Countess™ II automated cell counter User Guide
55
Page 56

Documentation and support

Customer and technical support

Visit thermosher.com/support for the latest service and support information.
• Worldwide contact telephone numbers
• Product support information – Product FAQs
– Software, patches, and updates
– Training for many applications and instruments
• Order and web support
• Product documentation – User guides, manuals, and protocols
Certicates of Analysis
– Safety Data Sheets (SDSs; also known as MSDSs)
Note: For SDSs for reagents and chemicals from other manufacturers, contact the manufacturer.

Limited product warranty

Life Technologies Corporation and/or its aliate(s) warrant their products as set forth in the Life Technologies' General Terms and Conditions of Sale at www.thermosher.com/us/en/home/global/terms-and-conditions.html. If you have any questions, please contact Life Technologies at www.thermosher.com/support.
56
Countess™ II automated cell counter User Guide
Page 57
Page 58
thermofisher.com/support | thermofisher.com/askaquestion
thermofisher.com
7 June 2019