Thermo BioMate 3 series, 335904, 335904P, 335905, 335905P Operator's Manual

BioMate™ 3 Series
Spectrophotometers
Operator’s Manual
Figure 1 BioMate 3 spectrophotometer
Figure 2 Back panel of the spectrophotometer
Key/
c A/C power connector d RS232C port
Figure 4 Keypad of the BioMate 3 spectrophotometer
Key
c Cell position keys
d Function keys
Key/
c Paper roll holder
d Icon for paper direction
e Paper entry slot
f Finger tab
Figure 5 Installing the Single-Position Cell Holder (335916)
Figure 6 Installing the printer (335988)
Key/
c Captive screw on lamp door
d Removing door from hinge
e Hinge
f Connecting wires to printer
Figure 7 Removing the cover of the fuse compartment
Figure 8 Removing the fuse holder
Figure 9 Removing and replacing the fuses
Declaration of Conformity
(For instruments manufactured after July 1, 2001)
Model: GENESYS™ 10 Series Catalog Nos.: 335900, 335900A, 335900P, 335900AP, 335901, 335901A, 335901P, 335901AP 335902, 335902A, 335902P, 335902AP, 335903, 335903A, 335903P, 335903AP 335906, 335906P, 335907, 335907P
Model: BioMate™ 3 Catalog Nos.: 335904, 335904P, 335905, 335905P
Model: GENESYS™ 6 Catalog Nos.: 335908, 335908P
North American Plug (NEMA 5-15) Numbers above European Plug (CEE 7/7 Schuko) Add a –02 suffix United Kingdom Plug (BS 1363/A) Add a –04 suffix
Thermo Electron Scientific Instrument Corporation certifies that the GENESYS 10, GENESYS 6, and BioMate 3 Spectrophotometers have been tested according to the instrumentation standards listed in this section in compliance with IEC directives and other regulatory requirements. The equipment under test (EUT) consisted of a sample instrument and applicable accessories, which are manufactured by Thermo Electron Scientific Instrument Corporation. The EUT was configured to ensure that both the worst case condition,
and that all of the accessories were tested. This
equipment
has been tested for use in non-residential environments.
IEC Directives
89/336/EEC Electromagnetic Compatibility Directive 73/23/EEC Low Voltage Directive
Electromagnetic Compatibility Test Standards
IEC 61326-1 1998, Electrical Equipment for Measur
ement, Control, and Laboratory use
- EMC Requirements. Class A Limits
IEC 61000-4-2 1999, Electrostatic Discharge Immunity Test (Test level: 4KV Air Discharge and 4 KV Contact Discharge)
IEC 61000-4-3 1998, Radiated, Radio Frequency, Electromagnetic Field Immunity Test (Test level: 3V/m)
IEC 61000-4-4 1995, Electrical Fast Transient/Burst Immunity Test (Test level: 1KV on the supply lines)
IEC 61000-4-5 1995, Surge Immunity Test (Test level: 0.5KV line to line and 1KV line to earth on the supply lines)
IEC 61000-4-6 1996, Immunity to Conducted Disturbances, Induced by Radio-Frequency Fields (Test level: 3V on the supply lines)
IEC 61000-4-11 1994, Voltage Dips, Short Interruptions, and Voltage Variations Test Level: 1 cycle/100%
CISPR 16-2 1999, Specification for Radio Disturbance and
Immunity Measuring Apparatus and Methods
Methods of Measurement of Disturbances and Immunity
Safety Test Standards
IEC 61010-1 1990 + A1 1992 +A2 1995. Safety requirements for El
ectrical Equipment for
Measurement, Control and
Laboratory use.
CSA C22.2 No. 1010.1, plus Am 2
IEC 61010-1 1997, Safety requirements for Electrical Equipment for Measuremen
t, Control and Laboratory use;
Part 1: General Requirements Test level: Installation Category II, Pollution Degree 2
Authorized signature:
_______________________________________ ___ Date: ____27May03____ Brenda Wilcox Vice President, Molecular Spectroscopy
i BioMate 3 Operator’s Manual
The information in this publication is provided for reference only. All information contained in this publica­tion is believed to be correct and complete. Thermo Electron Scientific Instruments Corporation shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the fur­nishing, performance or use of this material. All product specifications, as well as the information con­tained in this publication, are subject to change without notice.
This publication may contain or reference information and products protected by copyrights or patents and does not convey any license under our patent rights, nor the rights of others. We do not assume any liabil­ity arising out of any infringements of patents or other rights of third parties.
We make no warranty of any kind with regard to this material, including but not limited to the implied war­ranties of merchantability and fitness for a particular purpose.
Copyright © 2003 by Thermo Electron Scientific Instruments Corporation, Madison, WI 53711. Printed in the United States of America. All world rights reserved. No part of this publication may be stored in a retrieval system, transmitted, or reproduced in any way, including but not limited to photocopy, photograph, magnetic, or other record, without the prior written permission of Thermo Electron Scientific Instruments Corporation.
For technical assistance, please contact:
THERMO ELECTRON CORPORATION
(North America, Asia Pacific, Middle East, Africa, India and Latin America) 5225 Verona Road, Madison, WI 53711-4495, USA Telephone: (800) 642-6538 or (608) 276-6373 Fax: (608) 273-5045 E-Mail: careplan.techsupport@thermo.com
THERMO ELECTRON CORPORATION
(Europe) Mercers Row, Cambridge CB5 8HY, UK Telephone: Int +44 (0) 1223 446655 Fax: Int +44 (0) 1223 446644 E-Mail: careplan.techsupport@thermo.com
SPECTRONIC is a registered trademark and BioMate is a trademark of Thermo Electron Scientific Instruments Corporation.
Patent No.: 6414753
Item No.: 335904-10001, Rev. G 06/03 Component No.: 335904-10063, Rev. G 06/03
Thermo Electron Corporation ii
GENERAL SAFETY NOTES USED IN THIS MANUAL
This symbol alerts you to important information about using the instrument. Be sure to read and follow the associated instructions carefully.
This symbol alerts you to potential electrical hazards. Be sure that only qualified personnel perform the related procedures.
This symbol alerts you to hot surfaces. Be sure to read and follow the associated instructions carefully.
This symbol alerts you to potential UV radiation exposure, which can cause eye damage. Wear UV-opaque eye protection..
NEW PRODUCT WARRANTY
Thermo Electron Scientific Instruments Corporation instrumentation and related accessories are warranted against defects in material and workmanship for a period of one (1) year from the date of delivery. This warranty is provided only if the warranty registration card is returned to Thermo Electron Scientific Instruments Corporation within fifteen (15) days after delivery.
This warranty covers parts (except those specified below) and labor, and applies only to equipment which has been installed and operated in accordance with the operator's reference guide and which has been serviced only by authorized Thermo Electron Scientific Instruments Corporation dealers or service personnel. This warranty does not apply to equipment and accessories that have been modified or tampered with in any way, misused, or damaged by accident, neglect, or conditions beyond Thermo Electron Scientific Instruments Corporation's control.
This warranty does not apply to lamps, glassware, and similar expendable components. However, such parts and components may be warranted by their manufacturer.
Thermo Electron Scientific Instruments Corporation is not responsible under this warranty for loss in operating performance due to environmental conditions.
THIS WARRANTY IS IN LIEU OF ALL WARRANTIES EXPRESSED, IMPLIED, OR STATUTORY, INCLUDING, BUT NOT LIMITED TO, WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY OR OTHERWISE, and states Thermo Electron Scientific Instruments Corporation's entire and exclusive liability and the Customer's exclusive remedy for any claim in connection with the sale or furnishing of services, goods, or parts, their design, suitability for use, installation, or operations. Thermo Electron Scientific Instruments Corporation will in no event be liable for any direct, indirect, special, or consequential damages, whatsoever, including loss of goodwill, whether grounded in tort (including negligence), strict liability or contract, and Thermo Electron Scientific Instruments Corporation's liability under no circumstances will exceed the contract price for the goods and/or services for which liability is claimed.
iii BioMate 3 Operator’s Manua
SOFTWARE PASSWORD
This password allows you to enter the security section of the software used on the Thermo Electron Corporation spectrophotometer. Through the security section, you can "lock" test setups (test parameters) so that they may not be altered. The password also allows you to remove the security so that you may edit the test parameters. Please refer to the appropriate section in this Operator's Manual for more information on locking a test.
PASSWORD: 4 3 6 3 7 9 7
Thermo Electron Corporation iv
CHAPTER 1 - Setting Up the Instrument
Setting up the instrument.................................................................................................................1-1
Setting utility parameters ...........................................................................................................1-1
Selecting the language ..............................................................................................................1-1
Setting the date and time ..........................................................................................................1-2
Selecting standby settings.........................................................................................................1-2
Setting baseline expiration time ................................................................................................1-2
Setting the screen contrast........................................................................................................1-2
Loading paper in the internal printer.........................................................................................1-2
Setting the utility parameters for the printer .............................................................................1-3
Selecting and positioning glassware...............................................................................................1-3
Z-dimensions ....................................................................................................................................1-4
CHAPTER 2 - Using “Biotests” Software
Overview ...........................................................................................................................................2-1
Table of parameters ...................................................................................................................2-1
Entering information & commands..................................................................................................2-1
Types of parameter entries........................................................................................................2-1
Keypad layout.............................................................................................................................2-1
SmartStart feature ............................................................................................................................2-2
Setting up a single-test SmartStart ...........................................................................................2-2
Setting up a multiple-test SmartStart ........................................................................................2-2
Nucleic acid measurements ............................................................................................................2-2
DNA (260/280 and DNA 260/230) ............................................................................................2-3
DNA with Scan (260/280) and DNA with Scan (260/230).......................................................2-4
dsDNA, ssDNA, RNA and Oligos (entered factor) Direct or UV Measurements...................2-5
Protein Measurements.....................................................................................................................2-8
Bradford (standard & micro), Lowry (standard & micro), BCA (standard & micro) and Biuret
measurements ...........................................................................................................................2-8
Direct UV (280) and Direct UV (205)........................................................................................2-9
Warburg-Christian....................................................................................................................2-10
Cell Growth .....................................................................................................................................2-11
Setting test parameters ...........................................................................................................2-11
Measuring the sample .............................................................................................................2-11
Oligo Calculator..............................................................................................................................2-12
Using the oligo calculator ........................................................................................................2-12
CHAPTER 3 - Using “General Tests” Software
General information..........................................................................................................................3-1
Editing and Loading Saved Tests .............................................................................................3-1
Specifying names for tests ........................................................................................................3-1
Specifying concentration units...................................................................................................3-1
Using the SmartStart feature.....................................................................................................3-2
Running the cell correction program.........................................................................................3-3
Taking measurements ...............................................................................................................3-5
Saving tests ................................................................................................................................3-5
Basic Absorbance/%T measurements............................................................................................3-6
Setting the wavelength ..............................................................................................................3-6
Table of Contents
v BioMate 3 Operator’s Manua
Table of Contents
Measuring a blank......................................................................................................................3-6
Measuring unknowns.................................................................................................................3-6
Basic Concentration measurements...............................................................................................3-6
Setting the wavelength & mode................................................................................................3-7
Measuring a blank......................................................................................................................3-7
Measuring a standard................................................................................................................3-7
Entering a factor.........................................................................................................................3-8
Measuring unknowns.................................................................................................................3-8
Advanced A/%T/C - Absorbance & %Transmittance measurements ..........................................3-8
Recalling a test...........................................................................................................................3-8
Setting up test parameters ........................................................................................................3-9
Taking measurements ...............................................................................................................3-9
Advanced A/%T/C - Concentration measurements.......................................................................3-9
Recalling a test.........................................................................................................................3-10
Setting up test parameters ......................................................................................................3-10
Measuring a standard..............................................................................................................3-10
Entering a factor .......................................................................................................................3-11
Measuring unknowns...............................................................................................................3-11
Standard Curve...............................................................................................................................3-11
Recalling a standard curve......................................................................................................3-11
Setting the parameters for a standard curve..........................................................................3-12
Measuring the standards for a standard curve......................................................................3-12
Measuring unknowns...............................................................................................................3-13
Editing a standard curve..........................................................................................................3-13
Absorbance Ratio...........................................................................................................................3-13
Recalling a test.........................................................................................................................3-14
Setting up test parameters ......................................................................................................3-15
Measuring unknowns...............................................................................................................3-15
Absorbance Difference ..................................................................................................................3-16
Recalling a test.........................................................................................................................3-16
Setting up test parameters ......................................................................................................3-16
Measuring unknowns...............................................................................................................3-16
Kinetics............................................................................................................................................3-17
Recalling a test.........................................................................................................................3-17
Setting up test parameters ......................................................................................................3-17
Measuring unknowns...............................................................................................................3-17
Rescaling & recalculating kinetics results ..............................................................................3-18
Rescaling & recalculating tabular kinetics results..................................................................3-19
Survey Scan/Scanning ..................................................................................................................3-20
Recalling a test.........................................................................................................................3-20
Setting up test parameters ......................................................................................................3-20
Collecting a baseline scan.......................................................................................................3-20
Scanning an unknown.............................................................................................................3-21
Viewing & manipulating scan data..........................................................................................3-21
3-Point Net......................................................................................................................................3-24
Recalling a test.........................................................................................................................3-24
Setting up test parameters ......................................................................................................3-24
Thermo Electron Corporation vi
Taking measurements .............................................................................................................3-24
Multiple Wavelengths.....................................................................................................................3-25
Recalling a test.........................................................................................................................3-25
Setting up test parameters ......................................................................................................3-26
Taking measurements .............................................................................................................3-26
CHAPTER 4 - Using the Performance Validation Program
Overview ...........................................................................................................................................4-1
Accessing the Performance Validation tests ..................................................................................4-1
Troubleshooting checklist.................................................................................................................4-1
Wavelength Accuracy - Internal.......................................................................................................4-1
Wavelength Accuracy - Standards..................................................................................................4-2
Adding wavelengths...................................................................................................................4-2
Deleting wavelengths.................................................................................................................4-3
Photometric Accuracy ......................................................................................................................4-3
Selecting the mode ....................................................................................................................4-3
Adding standards .......................................................................................................................4-3
Deleting standards .....................................................................................................................4-4
Running the test .........................................................................................................................4-4
Noise Measurement.........................................................................................................................4-4
Stray Light.........................................................................................................................................4-5
Running the test .........................................................................................................................4-5
Internal Printer Test ..........................................................................................................................4-5
RS232C Test ....................................................................................................................................4-6
CHAPTER 5 - Connecting & Using Accessories
General information..........................................................................................................................5-1
Cell holders & cell holder accessories ............................................................................................5-1
Changing cell holders ................................................................................................................5-2
Internal printer...................................................................................................................................5-3
Installing the internal printer.......................................................................................................5-3
Loading paper in the internal printer.........................................................................................5-3
External computers ..........................................................................................................................5-3
CHAPTER 6 - Performing Maintenance Procedures
Routine care .....................................................................................................................................6-1
Cleaning............................................................................................................................................6-1
Cleaning and maintenance of cells...........................................................................................6-1
Cleaning the windows of the sample compartment.................................................................6-2
Changing the fuse ............................................................................................................................6-2
Replacement parts ...........................................................................................................................6-3
Table of Contents
vii BioMate 3 Operator’s Manua
Table of Contents
Appendix A - Specifications
Appendix B - Parameters
Appendix C - Calculations
Index
Thermo Electron Corporation viii
Setting up the instrument
1. Carefully unpack the shipping carton and verify
that you have received all the items listed below.
BioMate 3 spectrophotometer
Power cord
Operator's Manual - BioMate 3
spectrophotometer
Dust cover
Single Cell Holder
Models shipped with internal printer installed will also include:
1 roll of printer paper
1 printer holder set
2. Place the instrument on a flat, even surface that is:
As far as possible from any strong electric
or magnetic fields and from any electrical device that may generate high-frequency fields
Free of dust, corrosive gases and strong
vibrations
3. Remove any obstructions or materials that could hinder the flow of air under, behind and around the instrument.
4. Connect the female end of the power cord into the connector labeled A/C power on the back panel of the instrument (see Figure 2).
5. Plug the other end of the power cord into a grounded outlet with the appropriate voltage.
6. Ensure that the sample compartment does not contain any samples, and that the sample compartment door is closed.
7. Snap the paper roll holders (if needed) into place as shown in Figure 3. They will fit flush with the top of the instrument.
8. Turn on the instrument by pressing the power switch to ON (1=ON, 0=OFF). The power-on sequence will appear on the display and the instrument will go through its self-diagnostics. The instrument performs these diagnostics in the sequence shown:
a) Logo
b) Initializing
c) Calibrating filter wheel
d) Finding zero order e) Finding energy peak
f) Calibrating grating
Setting utility parameters
You should set up some of the utility parameters after the instrument is first powered up. These parameters include the date and time, standby setting, language, screen contrast settings and parameters needed to set up the printer.
You can set up the other utility parameters, or change the utility settings, at any time except when an entry screen is displayed or when the instrument is carrying out a task (such as setting the blank).
Press the UTILITY key. The Utility screen appears on the display.
From this screen, you can set the date and time, select the standby settings, select the language, reset the lamp hours, select the contrast settings and select the parameters for a printer.
Selecting the language
The instrument supports English, Spanish, French, German and Italian as the language options.
With the Utility screen displayed, press the arrow keys to highlight Language and press ENTER.
Press the arrow keys to highlight the language you want to select and press ENTER.
Setting Up the Instrument
1-1 BioMate 3 Operator’s Manual
Setting the date and time
To access the date and time settings:
With the Utility screen displayed, press the arrow keys to highlight Date/Time Setup and press ENTER. The screen displays the three date/time options that you can modify - date, time format and time.
To set the date:
Press the arrow keys to highlight Set Date and press ENTER.
Press Set Day, type the date, then press ENTER.
Press Set Month, highlight the correct month, then press ENTER.
Press Set Year, type the year, then press ENTER.
When the date is correct, press ESC to save the settings and return to the Utility screen.
To select the time format:
You can set up the spectrophotometer to display the time in either am/pm format or in 24-hour format.
To change the display format for the time, press the arrow keys to highlight Time Format and press ENTER until the format that you want to use (AM/PM or 24 hour) appears.
To set the time:
Press the arrow keys to highlight Set Time and press ENTER.
To set the hour, press Set Hour, type in the hour and press ENTER.
To set the minutes, press Set Minute, type in the minute and press ENTER.
To select between AM and PM, press Set
AM/PM until the appropriate setting appears.
Note: Any changes you make are saved
automatically (even during power down) by battery backup.
Selecting standby settings
To prolong xenon lamp life, your spectrophotometer has been pre-set at the factory to automatically go into standby mode after 15 minutes. To change Standby Mode time:
With the Utility screen displayed, press the arrow keys to highlight Standby and press ENTER.
Press the arrow keys to highlight the length of time you want the instrument will wait before entering standby mode and press ENTER.
Setting baseline expiration time
If you will be performing scans on your samples, you can set a time limit for a collected baseline.
To set the baseline expiration time:
With the Utility screen displayed, press the arrow keys to highlight Baseline Expiration (hr:min) and press ENTER.
Enter the desired time into the Entry baseline expiration time field. Press ENTER.
Setting the screen contrast
To make it easier to read the display, you can adjust the screen contrast on the spectrophotometer.
With the Utility screen displayed, press the arrow keys to highlight Screen Contrast and press ENTER.
Press the arrow keys to adjust the screen contrast.
When the screen contrast is correct, press ESC.
Setting up the internal printer
To set up the printer properly, you need to load the paper and set the utility parameters for the printer. Before setting up the parameters for the printer, be sure that the printer is installed. If you have ordered the internal printer as a separate item, you will need to install it. Refer to the section Connecting & using accessories for instructions on installing the printer.
Loading paper in the internal printer
Note: Make sure that the paper roll holders are in
place as shown in Figure 3. When installed correctly, they will fit flush with the top of the instrument.
Cut the paper so the edge is even.
Note: Arrows on the paper roll holders indicate
the direction of the paper feed (see Figure 3).
Setting Up the Instrument
Thermo Electron Corporation 1-2
Feed the paper straight into the paper entry slot. The printer grabs the end of the paper and pulls it in.
In Basic Absorbance/%Transmittance (%T), when the paper stops, press ENTER to continue advancing the paper until the paper comes out of the paper exit slot.
Pull out on the finger tabs on the paper roll holders and secure the roll of paper onto the paper roll holder.
Setting the utility parameters for the printer
If you wish to use a printer you can output to:
Internal printer.
External RS232 printer.
To ensure that the spectrophotometer can output information correctly to the printer, you need to select the appropriate device.
Press the UTILITY key. The Utility screen appears on the display.
Press the arrow keys to highlight Printer and press ENTER.
Select the printer that you want to use and press ENTER until On appears.
Press ESC to save the settings and return to the Utility screen.
Selecting and positioning glassware
The wavelength range for different types of cells varies depending on the manufacturer:
Glass - From 320 to 360nm, up to 1100nm.
Quartz - From 190 to 230nm, up to 1100nm.
Disposable - Refer to manufacturer's specifications and ensure that you work within the recommended range.
Test tubes vs. cuvettes
Square cuvettes that are carefully matched (see “Correcting for cell variability” below) yield very precise results. Matched test tubes, when properly handled, can show as little as 1-2% deviation between readings.
The pathlength of test tubes is not as well defined as in square cuvettes. However, constructing a standard curve eliminates the need for great accuracy in knowing the pathlength, provided that the same pathlength cell is used for all blanks, standards and samples.
Other guidelines
Be sure to position cuvettes and test tubes so that the clear sides face the light beam. This means that one clear side should face the front of the instrument and the other clear side should face the back of the instrument.
Note: Test tubes should always be placed in the
instrument in exactly the same orientation in the light beam. A fiducial mark on the test tube helps you orient the test tubes consistently and correctly.
When using small aperture cells:
Always use masked cells.
Use the same cell (or cuvette) for your
blank and your samples.
Setting Up the Instrument
1-3 BioMate 3 Operator’s Manual
Z-dimensions
The figure below illustrates the position of the light beam in the spectrophotometer.
The specifications for the sample compartment, including the dimensions of the beam are:
Z-dimension
Square cuvettes/Test tubes 8.5mm
Beam size 2mm (wide) x 7mm (high)
Setting Up the Instrument
Thermo Electron Corporation 1-4
Overview
The BioMate instrument allows you to run an assortment of tests used to characterize biological and biochemical substances. These tests fall into the following categories:
Nucleic acid measurements
Protein measurements
Cell growth analysis
Oligonucleotide calculations
All of the parameters for the BioMate applications described in this chapter are factory-set. This means that if you want to change the parameters, you will need to specify a different name to save the new test parameters.
Turn on your spectrophotometer. After the power­on sequence is completed, the list of BioMate test categories appears.
Table of parameters
Parameters used in the spectrophotometer are located in Appendix B.
You can use this list as a reference when you are setting up tests.
Entering information & commands
The keypad on the spectrophotometer includes a numeric keypad as well as certain special keys that you can use to enter information and commands.
Types of parameter entries
Different parameters require different types of entries - you can select from a list for some parameters, while you need to enter a value for others. While using the spectrophotometer, you will
notice the following types of entries for parameters:
Type-in entries are numeric values that you can enter using the numeric keypad.
Toggle entries offer two options for a parameter. You can press ENTER to switch between the two options, then press one of the arrow keys to move to another parameter.
Pop-up windows list multiple options for a parameter or display messages. You can press the arrow keys to select the value you want, then press ENTER to select it and move to another parameter.
Cursor selections are used on graphical displays and the character list you use to name tests and files. On graphical displays, the cursor appears as a vertical line; on the character list, the cursor highlights the character where it is positioned. You use the cursor control keys to move the cursor to the character or position on the graph you want to select.
Keypad layout
The keypad on the spectrophotometer (shown in Figure 4) includes a numeric keypad as well as the following special keys:
Function keys - These four keys allow you to select a particular task you want to perform. The tasks you can perform appear on the display screen above each key and vary from one screen to another. On some screens, all four keys will function, while on other screens only some of the function keys will work.
Esc - In general, when you press ESC, the program clears the current entry but does not change any values you have already accepted. The program may also return you to the previous screen when you press ESC.
Clear - When you press CLEAR, the program deletes any entry you have made but does not change any values you have already accepted. The program does not return you to the previous screen when you press CLEAR.
Enter - Typically, when you press ENTER, the program accepts any highlighted or selected values and advances to the next parameter or screen. Specific instructions later in this guide indicate any special instructions about using
ENTER.
Using “Biotests” Software
2-1 BioMate 3 Operator’s Manua
Arrow keys - These keys allow you to control the position of the cursor so you can select values from lists or select an option from a screen.
Cell position keys - These keys allow you to select which cell position the instrument will use for a measurement. If you have the optional turret installed, one position is reserved for your blank and you can select from five cell positions for your samples.
Utility - When you press UTILITY, the Utility screen appears on the display.
Test - When you press TEST, the Test Types screen appears on the display.
Print - When you press PRINT, the instrument prints the information that appears on the display.
SmartS
tart feature
BioMate’s SmartStart feature enables you to select the test methods you use most frequently and have them appear when you start up your instrument. If your laboratory runs only a single test, you can use the SmartStart feature to select it and it will appear each time you start up your instrument. Similarly, if you have a set of tests you run, you can use SmartStart to select them so the list appears when you start up the instrument.
Setting up a single-test SmartStart
1. With the BioMate Tests Screen displayed,
press the Stored Tests key on the spectrophotometer keypad. A list of all the tests on the instrument appears on the screen.
2. Scroll down through the list until the appropriate test is highlighted.
3. When the appropriate test is highlighted, press Select Test to add the selected test to the
SmartStart menu. An arrow sign “>” will indicate the test has been selected
4. Press Load Test.
5. The parameter screen of the test you selected
will be displayed.
Note: At this point, you can power down the
instrument and then power it back up. When it starts up again, the parameter screen for the selected test will be displayed.
Setting up a multiple-test SmartStart
1. Press the Stored Tests key on the
spectrophotometer keypad.
2. Scroll down through the list until the first appropriate test is highlighted.
3. Press Select Tests to add the selected test to the SmartStart menu.
4. Continue scrolling through the list and adding tests until you’ve made all the appropriate selections.
5. Press ESC until you return to the BioMate Tests screen.
Note: At this point, you can power down the
instrument and then power it back up. When it starts up again, the list of tests you’ve selected will be displayed.
Nucleic acid measurement
s
You can use these tests to determine the concentration and purity of nucleic acid in a given sample.
DNA - measures absorbance at 260 and 280nm; determines concentration and purity based on absorbance ratio and absorbance difference; also calculates protein concentration.
DNA with scan - records absorbance scan between 260 and 280nm or between 260 and 230nm; determines concentration and purity based on absorbance ratio and absorbance concentration; also calculates protein concentration.
dsDNA - measures absorbance at 260nm; calculates concentration based on absorbance and concentration factor.
ssDNA, RNA - measures absorbance at 260nm; calculates concentration based on absorbance and concentration factor.
Oligonucleotides ("Oligos") - measures absorbance at 260nm; calculates concentration based on absorbance and concentration factor or calculates concentration based on absorbance and concentration factor determined by oligo calculator.
Using “Biotests” Software
Thermo Electron Corporation 2-2
Several of these categories include multiple tests that are similar, so the section may not include screen samples for each test. For example, the parameters are the same for the Direct UV measurement of ssDNA and RNA tests, but the factor used to convert absorbance to concentration is different. Similarly, for the Direct UV measurement of oligonucleotides tests, the parameters are also the same, but the factors used to convert absorbance to concentration are different. For a complete list of all parameters and calculations for each test, refer to Appendices B and C.
DNA (260/280) and DNA (260/230)
These tests function almost identically - the only difference is the wavelengths used for the measurements. One test measures absorbance at 260 and 280nm, while the other measures absorbance at 260 and 230nm. Refer to Appendix B for a description of the parameters and Appendix C for the default values.
To get started, with the BioMate Tests screen displayed, move the arrow keys to highlight Nucleic Acid Tests and press ENTER. A list of nucleic acid test appears. Move the arrow keys to highlight DNA (260/280). The DNA (260/280) parameter screen appears.
Note: The following screens show the
parameters for the DNA (260/280) test. For the DNA (260/230) test, Wavelength 2 is set to 230nm.
Note: If Cell Correction is ON, you must run the
Setup Correction program before you can access the Run Test or Measure Samples keys.
Setting up test parameters
1. With the DNA (260/280) or DNA (260/230) screen displayed, use the arrow keys to highlight the name of the parameter you want to set.
2. When the parameters are set, you can press Save Test to save the test or Run Test to
measure the blank and unknowns.
Measuring unknowns
Measuring unknowns automatically (using Auto 6 or Auto 3)
1. Place the blank and unknowns in the correct cell positions.
2. Press ENTER to start the measurements. The instrument automatically measures the blank first, then measures the unknowns and displays the sample measurements on the screen.
Using “Biotests” Software
2-3 BioMate 3 Operator’s Manual
Measuring unknowns manually (using Manual 6 or Single Cell Positioner)
1. With the DNA setup screen displayed, press
Run Test.
2. Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a 6-Position Cell Holder, you can place up to five samples in the cell holder.
3. Press Measure Blank to measure the blank. If your instrument is equipped with a 6-Position Cell Holder, it automatically moves to the B position to measure the blank. When it completes the measurement, it returns to its previous cell position.
4. Press Measure Sample to start the measurement. If your instrument is equipped with a 6-Position Cell Holder, press the cell position buttons on the keypad to re-position the cell holder and measure the rest of the unknowns manually.
DNA with Scan (260/280) and DNA with Scan (260/230)
The DNA measurement with scan tests include two test that function almost identically - the only difference is in the wavelengths used for the measurements. In both cases the scan is measured from 225 to 325nm. One test measures absorbance at 260 and 280nm, while the other measures absorbance at 260 and 230nm. Refer to Appendix B for a description of the parameters and Appendix C for the default values.
To get started, with the BioMate Tests screen displayed, move the arrow keys to highlight Nucleic Acid Tests and press ENTER. A list of
nucleic acid test appears. Move the arrow keys to highlight DNA with Scan (260/280). The DNA with
Scan (260/280) parameter screen appears.
Note: The following screens show the
parameters for the DNA (260/280) test. For the DNA (260/230) test, Wavelength 2 is set to 230nm.
Note: If Cell Correction is ON, you must run the
Setup Correction program before you can access the Run Test or Measure Samples keys.
Setting up test parameters
1. With the DNA with Scan (260/280) or DNA with Scan (260/230) screen displayed, use the
arrow keys to highlight the name of the parameter you want to set.
2. When the parameters are set, you can press Save Test to save the test or Run Test to
measure the blank and unknowns.
Collecting a baseline scan
Note: If your instrument is equipped with a 6-
Position Cell Holder, be sure to place the blank in the B position. The instrument always uses the B position to collect the baseline.
1. With the DNA with Scan (260/280) or DNA with Scan (260/230) screen displayed, press Run Test. The DNA with Scan measurement
screen appears.
Using “Biotests” Software
Thermo Electron Corporation 2-4
2. Place the blank in the B position.
3. Press Measure Blank to collect the baseline.
When the instrument is finished measuring the blank, the message disappears.
Note: If you want to switch between tabular and
graphical displays, press Graph or
Tabular.
Measuring the sample
1. If your instrument is equipped with a 6-Position Cell Holder, be sure to place the unknown in cell position #1.
Note: The instrument always uses cell position
#1 to measure the sample.
2. With the DNA with Scan measurement screen displayed, press Measure Sample to measure the sample. When the instrument is finished measuring the absorbance scan, it displays a graph of the scan along with the sample ID#, DNA ratio, DNA concentration and protein concentration.
Note: If you want to switch between tabular and
graphical displays, press Graph or
Tabular.
Note: You may need to use the up and down
arrow keys to view all the data for the screen.
dsDNA, ssDNA, RNA and Oligos (entered factor) Direct or UV Measurements
These test measurements are all set up and run using the same types of test parameters. Refer to Appendix B for a description of the parameters and Appendix C for the default values.
To get started, with the BioMate Tests screen displayed, move the arrow keys to highlight Nucleic Acid Tests and press ENTER. A list of nucleic acid test appears. Move the arrow keys to highlight the desired test and press ENTER. The dsDNA, ssDNA, RNA or Oligos (entered factor) parameter screen appears.
Note: The following screens show the
parameters for the dsDNA test.
Note: If Cell Correction is ON, you must run the
Setup Correction program before you can access the Run Test or Measure Samples keys.
Using “Biotests” Software
2-5 BioMate 3 Operator’s Manual
Setting up test parameters
1. With the dsDNA, ssDNA, RNA or Oligos setup screen displayed, use the arrow keys to highlight the name of the parameter you want to set.
2. When the parameters are set, you can press Save Test to save the test or Run Test to
measure the blank and unknowns.
Measuring unknowns
Measuring unknowns automatically (using Auto 6 or Auto 3)
1. Place the blank and unknowns in the correct cell positions.
2. Press ENTER to start the measurements. The instrument automatically measures the blank first, then measures the unknowns and displays the sample measurements on the screen.
Measuring unknowns manually (using Manual 6 or Single Cell Positioner)
1. With the dsDNA setup screen displayed, press
Run Test.
2. Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a 6-Position Cell Holder, you can place up to five samples in the cell holder.
3. Press Measure Blank to measure the blank. If your instrument is equipped with a 6-Position Cell Holder, it automatically moves to the B position to measure the blank. When it completes the measurement, it returns to its previous cell position.
4. Press Measure Sample to start the measurement. If your instrument is equipped with a 6-Position Cell Holder, press the cell position buttons on the keypad to re-position the cell holder and measure the rest of the unknowns manually.
Oligos (calculated factor)
The Oligos (calculated factor) measurement calculates molecular weight, extinction coefficient and a conversion factor for a base sequence that you enter. This conversion factor is used to calculate the concentration of oligos in your sample from the absorbance measurement. Refer to Appendix B for a description of the parameters and Appendix C for the default values.
Using “Biotests” Software
Thermo Electron Corporation 2-6
To get started, with the BioMate Tests screen displayed, move the arrow keys to highlight Nucleic Acid Tests and press ENTER. A list of nucleic acid test appears. Move the arrow keys to highlight the desired test and press ENTER. The
Oligos (calc factor) parameter screen appears.
Note: If Cell Correction is ON, you must run the
Setup Correction program before you can access the Run Test or Measure
Samples keys.
Setting up test parameters
1. With the Oligos (calc factor) setup screen displayed, use the arrow keys to highlight the name of the parameter you want to set.
2. When the parameters are set, you can press Save Test to save the test or Run Test to
measure the blank and unknowns.
Measuring unknowns
Measuring unknowns automatically (using Auto 6 or Auto 3)
1. Place the blank and unknowns in the correct cell positions.
2. Press ENTER to start the measurements. The instrument automatically measures the blank first, then measures the unknowns and displays absorbance, oligo concentration in µg/mL and pmol/µL.
Measuring unknowns manually (using Manual 6 or Single Cell Positioner)
1. With the Oligos (calc factor) screen displayed, press Run Test.
2. Place the blank and unknown in the correct cell positions. If your instrument is equipped with a 6-Position Cell Holder, you can place up to five samples in the cell holder.
3. Press Measure Blank to measure the blank. If your instrument is equipped with a 6-Position Cell Holder, it automatically moves to the B position to measure the blank. When it completes the measurement, it returns to its previous cell position.
4. Press Measure Sample to start the measurement. If your instrument is equipped with a 6-Position Cell Holder, press the cell position buttons on the keypad to re-position the cell holder and measure the rest of the unknowns manually.
Using “Biotests” Software
2-7 BioMate 3 Operator’s Manual
Protein Measurements
Bradford (standard & micro), Lowry (standard & micro), BCA (standard & micro) and Biuret measurements
You can use these tests to determine the concentration of protein in a given sample, using the following analytical methods:
Bradford - measures absorbance at 595nm; determines concentration for either standard or micro sample concentrations.
Lowry - measures absorbance at 550nm for standard and 770nm for micro; determines concentration for either standard or micro sample concentrations.
Bicinchoninic Acid (BCA) - measures absorbance at 562nm; determines concentration for either standard or micro sample concentrations.
Biuret - measures absorbance at 540nm.
Several of these categories include multiple tests that are similar, so this section includes screen samples for the standard Bradford test only. For a complete list of all parameters for each test, refer to Appendix B; for a list of calculations used for the tests, refer to Appendix C.
To get started, with the BioMate Tests screen displayed, move the arrow keys to highlight Protein Tests and press ENTER. A list of protein tests appears. Move the arrow keys to highlight the desired test and press ENTER. The Bradford-
Standard parameter screen appears.
Note: The following screens show the
parameters for the Bradford-Standard test. All the other protein standard curve methods work in a similar fashion.
Note: If Cell Correction is ON, you must run the
Setup Correction program before you can access the Run Test or Measure Samples keys.
Setting up test parameters for a standard curve
1. With the Bradford-Standard setup screen displayed, use the arrow keys to highlight the name of the parameter you want to set. Set the parameters for measuring the standards. Refer to the list of parameters in Appendix B for a description of the parameters and Appendix C for the default values.
2. When the parameters are set, you can press Save Test to save the test or Run Standards
to measure the blank and standards. The
Standard Measurement screen appears.
3. If you need to edit concentration values, use
the arrow keys to select the standard you want to edit and press Edit Standards. From the Edit Standards window you can edit, add or delete one or all standards.
Measuring the standards for a standard curve
Measuring standards automatically (using Auto 6 or Auto 3)
1. Place the blank and standards in the correct cell positions.
2. When all the standards are correct, press Measure Standards to set up and run the
standards. The instrument automatically measures the blank first, then measures the standards. When the instrument has measured all the standards, the Standards screen appears, showing the absorbance of each
Using “Biotests” Software
Thermo Electron Corporation 2-8
standards, along with the slope, intercept and correlation coefficient of the standard curve.
Measuring standards automatically (using Auto 6 or Auto 3)
1. Place the blank and standards in the correct cell positions.
2. When all the standards are correct, press Measure Standards to set up and run the
standards.
3. Press Measure Blank to measure the blank. If your instrument is equipped with a 6-Position Cell Holder, it automatically moves to the B position to measure the blank. When it completes the measurement, it returns to its previous cell position.
4. Press Measure Standards to measure the standards. If your instrument is equipped with a 6-Position Cell Holder, press the cell position buttons on the keypad to re-position the cell holder and measure the rest of the standards manually.
When the instrument has measured all the standards, the Standards screen appears, showing the absorbance of each standards, along with the slope, intercept and correlation coefficient of the standard curve.
You can use this screen to:
Edit the standards (press Edit Standards)
Display a graph of the standard curve data (press View Graph)
Save the standard curve (press Save Test)
Measure your samples (press Run Test)
Measuring protein samples
Measuring unknowns automatically (using Auto 6 or Auto 3)
1. With the Standards screen displayed, place the blank and unknowns in the correct cell positions.
2. Press ENTER to start the measurements. The instrument automatically measures the blank first, then measures the unknowns and displays the absorbance and concentration of each unknown.
Measuring unknowns manually (using Manual 6 or Single Cell Positioner)
1. With the Standards screen displayed, press Run Test.
2. Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a 6-Position Cell Holder, you can place up to five samples in the cell holder.
3. Press Measure Blank to measure the blank. If your instrument is equipped with a 6-Position Cell Holder, it automatically moves to the B position to measure the blank. When it completes the measurement, it returns to its previous cell position.
4. Press Measure Sample to start the measurement. If your instrument is equipped with a 6-Position Cell Holder, press the cell position buttons on the keypad to re-position the cell holder and measure the rest of the unknowns manually.
Direct UV (280) and Direct UV (205)
The Direct UV methods determine protein concentration based on absorbance at either 280 or 205nm. Refer to Appendix B for a description of the parameters and Appendix C for the default values.
To get started, move the arrow keys on the BioMate tests screen to highlight Protein Tests and press ENTER. Then the move the arrow keys to highlight Direct UV (280) and press ENTER. The Direct UV (280) parameter screen appears.
Note: The following screens show the
parameters for the Direct UV tests at 280nm. For the Direct UV test at 205nm, the wavelength is set to 205nm.
Using “Biotests” Software
2-9 BioMate 3 Operator’s Manual
Loading...
+ 65 hidden pages