Thermo 49i Instruction Manual

Page 1
Model 49i Primary Standard
Instruction Manual
UV Photometric O Part number 102474-00 11Jan2008
Calibrator
3
Page 2
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin, MA 02038 1-508-520-0430 www.thermo.com/aqi
Page 3
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Fisher Scientific’s compliance with these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at: www.thermo.com/WEEERoHS.
Thermo Fisher Scientific WEEE Compliance
Page 4
Page 5
Contents
Chapter 1 Introduction......................................................................................1-1
Principle of Operation ............................................................1-2
Specifications ..........................................................................1-3
Chapter 2 Installation .......................................................................................2-1
Lifting.....................................................................................2-1
Unpacking and Inspection ......................................................2-1
Setup Procedure......................................................................2-3
Connecting External Devices ..................................................2-4
Terminal Board PCB Assemblies..........................................2-4
Startup ....................................................................................2-8
Chapter 3 Operation ..........................................................................................3-1
Display....................................................................................3-2
Pushbuttons............................................................................3-3
Soft Keys..............................................................................3-4
Software Overview ..................................................................3-4
Power-Up Screen .................................................................3-6
Run Screen...........................................................................3-6
Main Menu..........................................................................3-7
Range Menu ...........................................................................3-7
Gas Units.............................................................................3-8
Range.............................................................................3-8
O
3
Set Custom Ranges ............................................................3-11
Averaging Time.....................................................................3-12
Calibration Factors Menu .....................................................3-12
O3 Background..................................................................3-13
O
Span Coefficients .........................................................3-13
3
Instrument Controls Menu ...................................................3-15
Custom Levels....................................................................3-15
Pump .................................................................................3-16
Temperature Compensation...............................................3-16
Pressure Compensation......................................................3-17
Datalogging Settings ..........................................................3-17
Communication Settings....................................................3-27
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual v
Page 6
Contents
I/O Configuration.................................................................3-36
Screen Contrast..................................................................3-48
Service Mode .....................................................................3-49
Date/Time .........................................................................3-49
Diagnostics Menu .................................................................3-50
Program Version ................................................................3-51
Voltages .............................................................................3-51
Temperatures .....................................................................3-53
Pressure..............................................................................3-53
Flows .................................................................................3-53
Cell A/B O3.......................................................................3-54
Intensities...........................................................................3-54
Analog Input Readings.......................................................3-55
Analog Input Voltages........................................................3-55
Digital Inputs.....................................................................3-55
Relay States ........................................................................3-56
Test Analog Outputs..........................................................3-56
Instrument Configuration..................................................3-57
Contact Information..........................................................3-58
Alarms Menu ........................................................................3-58
O3 Lamp Temperature.......................................................3-59
Lamp Temperature ............................................................3-60
Bench Temperature............................................................3-61
Pressure..............................................................................3-62
Flow A and B .....................................................................3-63
Intensity A and B ...............................................................3-64
O3 Concentration ..............................................................3-65
Service Menu ........................................................................3-66
Pressure Check...................................................................3-67
Lamp Setting......................................................................3-69
Detector Calibration ..........................................................3-69
Intensity Check..................................................................3-70
Response Coefficient ....................................................3-71
O
3
Manual O3 Lamp Setting...................................................3-71
Pressure Calibration ...........................................................3-72
Flow A and B Calibration ..................................................3-74
Temperature Calibration....................................................3-76
Analog Output Calibration ................................................3-77
Analog Input Calibration ...................................................3-79
Display Pixel Test ..............................................................3-81
Restore User Defaults.........................................................3-81
vi Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 7
Contents
Password Menu.....................................................................3-82
Set Password ......................................................................3-82
Lock Instrument ................................................................3-82
Change Password ...............................................................3-83
Remove Password...............................................................3-83
Unlock Instrument ............................................................3-84
Chapter 4 Preventive Maintenance ...............................................................4-1
Safety Precautions ...................................................................4-2
Replacement Parts...................................................................4-2
Outside Case Cleaning............................................................4-2
Optical Bench Cleaning ..........................................................4-2
Lamp Replacement .................................................................4-3
Monitoring Detector Frequencies and Noise...........................4-4
Increasing Lamp Output......................................................4-4
Monitoring Lamp Noise ......................................................4-5
Capillary Service......................................................................4-5
Pump Rebuilding....................................................................4-6
Leak Test and Pump Checkout ...............................................4-7
External Leaks......................................................................4-7
Solenoid Leaks ....................................................................4-8
Solenoid Leak Test...............................................................4-9
Chapter 5 Troubleshooting ..............................................................................5-1
Safety Precautions ...................................................................5-1
Troubleshooting Guides..........................................................5-2
Board-Level Connection Diagrams .........................................5-6
Connector Pin Descriptions....................................................5-8
Service Locations...................................................................5-21
Chapter 6 Servicing...........................................................................................6-1
Safety Precautions ...................................................................6-3
Firmware Updates...................................................................6-4
Accessing the Service Mode.....................................................6-4
Replacement Parts List............................................................6-4
Cable List................................................................................6-6
External Device Connection Components ..............................6-6
Removing the Measurement Bench and
Lowering the Partition Panel...................................................6-8
Fuse Replacement ...................................................................6-9
Pump Replacement...............................................................6-10
Fan Replacement...................................................................6-11
Analog Output Testing .........................................................6-12
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual vii
Page 8
Contents
Analog Output Calibration ...................................................6-14
Analog Input Calibration ......................................................6-15
Calibrating the Input Channels to Zero Volts ....................6-15
Calibrating the Input Channels to Full Scale......................6-16
Optical Bench Temperature Calibration ...............................6-17
I/O Expansion Board (Optional) Replacement .....................6-18
Digital Output Board Replacement.......................................6-19
Motherboard Replacement....................................................6-20
Measurement Interface Board Replacement ..........................6-21
Front Panel Board Replacement............................................6-22
LCD Module Replacement ...................................................6-23
Optical Bench Replacement..................................................6-25
Photometer Board Replacement............................................6-26
Photometer Lamp Replacement ............................................6-27
Photometer Lamp Voltage Adjustment .................................6-28
Sample/Reference Solenoid Replacement ..............................6-29
Detector Replacement...........................................................6-30
Pressure Transducer Replacement .........................................6-31
Pressure Transducer Calibration ...........................................6-32
Flow Transducer Replacement ..............................................6-34
Flow Transducer Calibration ................................................6-35
Ozonator Lamp Replacement................................................6-36
Ozonator Lamp Heater Replacement....................................6-37
Ozonator Replacement..........................................................6-39
Ozonator Board Replacement ...............................................6-39
Service Locations...................................................................6-40
Chapter 7 System Description........................................................................ 7-1
Hardware ................................................................................7-1
Optical Bench with Photometer Lamp.................................7-2
Detector System...................................................................7-2
Flow Transducers.................................................................7-2
Pressure Transducer .............................................................7-2
Ozonator Assembly ..............................................................7-3
Ozonator Lamp Power Supply .............................................7-3
Photometer Lamp power Supply..........................................7-3
Optical Bench Temperature Thermistor ..............................7-3
Pump ...................................................................................7-3
Sample/Reference Solenoid Valves .......................................7-3
Software..................................................................................7-3
Instrument Control..............................................................7-4
Monitoring Signals...............................................................7-4
Measurement Calculations ...................................................7-4
Output Communication......................................................7-5
viii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 9
Contents
Electronics ..............................................................................7-5
Motherboard........................................................................7-5
Measurement Interface Board ..............................................7-6
Digital Output Board...........................................................7-7
I/O Expansion Board (Optional) .........................................7-7
Front Panel Connector Board ..............................................7-8
I/O Components ....................................................................7-8
Analog Voltage Outputs.......................................................7-8
Analog Current Outputs (Optional) ....................................7-9
Analog Voltage Inputs (Optional)........................................7-9
Digital Relay Outputs ..........................................................7-9
Digital Inputs.......................................................................7-9
Serial Ports.........................................................................7-10
RS-232 Connection ...........................................................7-11
RS-485 Connection ...........................................................7-11
Ethernet Connection..........................................................7-12
External Accessory Connector ............................................7-12
Chapter 8 Optional Equipment ........................................................................8-1
Zero Air Source.......................................................................8-1
Teflon Particulate Filter ..........................................................8-1
I/O Expansion Board Assembly...............................................8-1
Terminal Block and Cable Kits ...............................................8-2
Cables .....................................................................................8-2
25-Pin Terminal Board Assembly............................................8-3
Mounting Options..................................................................8-3
Appendix A Warranty ......................................................................................... A-1
Appendix B C-Link Protocol Commands.........................................................B-1
Instrument Identification Number......................................... B-2
Entering Units in PPB ........................................................... B-2
Accessing Streaming Data ...................................................... B-2
Commands ............................................................................ B-2
Measurements........................................................................ B-8
Alarms.................................................................................. B-12
Diagnostics .......................................................................... B-17
Datalogging ......................................................................... B-17
Calibration........................................................................... B-25
Keys/Display........................................................................ B-28
Measurement Configuration ................................................ B-29
Hardware Configuration...................................................... B-34
Communications Configuration .......................................... B-36
I/O Configuration................................................................ B-40
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual ix
Page 10
Contents
Record Layout Definition .................................................... B-44
Format Specifier for ASCII Responses............................... B-45
Format Specifier for Binary Responses .............................. B-45
Format Specifier for Front-Panel Layout .......................... B-45
Appendix C MODBUS Protocol.........................................................................C-1
Serial Communication Parameters .........................................C-2
TCP Communication Parameters ..........................................C-2
Application Data Unit Definition .......................................... C-2
Slave Address.......................................................................C-2
MBAP Header ....................................................................C-2
Function Code.................................................................... C-3
Data.................................................................................... C-3
Error Check ........................................................................C-3
Function Codes...................................................................... C-3
(0x01/0x02) Read Coils / Read Inputs ............................... C-3
(0x03/0x04) Read Holding Registers / Read Input RegistersC-5
(0x05) Force (Write) Single Coil.........................................C-7
MODBUS Commands Supported......................................... C-8
Appendix D Geysitech (Bayern-Hessen) Protocol ........................................D-1
Serial Communication Parameters .........................................D-1
TCP Communication Parameters ..........................................D-2
Instrument Address................................................................D-2
Abbreviations Used ................................................................D-2
Basic Command Structure .....................................................D-2
Block Checksum <BCC> .......................................................D-3
Geysitech Commands ............................................................D-3
Instrument Control Command (ST)...................................D-3
Data Sampling/Data Query Command (DA) .....................D-4
Measurements Reported in Response to
DA Command for 49i and 49i PS.......................................D-7
Operating and Error Status .................................................D-8
x Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 11
Figures
Figures
Model 49i Primary Standard Flow Schematic ..........................1-3
Remove the Packing Material ..................................................2-2
Remove the Shipping Screws ...................................................2-2
Model 49i Primary Standard Rear Panel ..................................2-4
I/O Terminal Board Views ......................................................2-5
D/O Terminal Board Views ....................................................2-6
25-Pin Terminal Board Views .................................................2-7
49i Primary Standard Front Panel Display ..............................3-2
Front Panel Pushbuttons .........................................................3-3
Flowchart of Menu-Driven Software .......................................3-5
Pin-Out of Rear Panel Connector .........................................3-10
Capillary Location ...................................................................4-6
Rebuilding the Pump ..............................................................4-7
Board-Level Connection Diagram - Common Electronics .......5-6
Board-Level Connection Diagram - Measurement System ......5-7
Properly Grounded Antistatic Wrist Strap ...............................6-4
Component Layout .................................................................6-7
Removing the Measurement Bench and
Lowering the Partition Panel ...................................................6-8
Replacing the Pump ..............................................................6-10
Replacing a Fan .....................................................................6-11
Rear Panel Analog Input and Output Pins ............................6-13
Replacing the I/O Expansion Board (Optional) ....................6-19
Rear Panel Board Connectors ................................................6-19
Replacing the Measurement Interface Board .........................6-22
Replacing the Front Panel Board and the LCD Module ........6-23
Replacing the Optical Bench .................................................6-26
Replacing the Detectors .........................................................6-31
Ozonator Board Location ......................................................6-37
Hardware Components ...........................................................7-2
Rack Mount Option Assembly ................................................8-4
Bench Mounting .....................................................................8-5
EIA Rack Mounting ................................................................8-6
Retrofit Rack Mounting ..........................................................8-7
Flag Status ............................................................................ B-12
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual xi
Page 12
Figures
xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 13
Tables
Tables
Model 49i Primary Standard Specifications .............................................. 1-3
I/O Terminal Board Pin Descriptions ...................................................... 2-5
D/O Terminal Board Pin Descriptions .................................................... 2-7
25-Pin Terminal Board Pin Descriptions ................................................. 2-8
Front Panel Pushbuttons .......................................................................... 3-3
Standard Ranges ....................................................................................... 3-9
Default Analog Outputs ......................................................................... 3-10
Analog Output Zero to Full-Scale Table ................................................ 3-42
Signal Type Group Choices ................................................................... 3-44
Troubleshooting - General Guide ............................................................ 5-2
Troubleshooting - Alarm Messages ........................................................... 5-4
Motherboard Connector Pin Descriptions ............................................... 5-8
Front Panel Board Connector Pin Diagram ........................................... 5-13
I/O Expansion Board (Optional) Connector Pin Descriptions ............... 5-15
Digital Output Board Connector Pin Descriptions ................................ 5-16
Measurement Interface Board Connector Pin Descriptions .................... 5-17
Ozonator Power Supply Connector Pin Descriptions ............................. 5-19
Photometer Power Supply Connector Pin Descriptions ......................... 5-20
Model 49i Primary Standard Replacement Parts ...................................... 6-4
Model 49i Primary Standard Cables ......................................................... 6-6
External Device Connection Components ............................................... 6-6
Analog Output Channels and Rear Panel Pin Connections .................... 6-13
Analog Input Channels and Rear Panel Pin Connections ....................... 6-14
RS-232 DB Connector Pin Configurations ............................................ 7-11
RS-485 DB Connector Pin Configuration ............................................. 7-11
Cable Options .......................................................................................... 8-2
Color Codes for 25-Pin and 37-Pin Cables .............................................. 8-3
Mounting Options ................................................................................... 8-3
C-Link Protocol Commands .................................................................... B-3
Averaging Times ......................................................................................B-9
Alarm Trigger Values .............................................................................B-16
Record Output Formats ......................................................................... B-21
Stream Time Values ...............................................................................B-24
Range Settings ........................................................................................B-30
Contrast Levels .......................................................................................B-34
Reply Termination Formats ................................................................... B-38
Analog Current Output Range Values ...................................................B-40
Analog Voltage Output Range Values ....................................................B-41
Default Output Assignment ................................................................... B-42
Read Coils for 49i Primary Standard ........................................................C-8
Read Registers for 49i Primary Standard ..................................................C-9
Write Coils for 49i Primary Standard .....................................................C-10
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual xiii
Page 14
Tables
Operating Status for Model 49i and 49i Primary Standard ..................... D-8
Error Status for Model 49i and 49i Primary Standard ............................. D-8
xiv Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 15
About This Manual
This manual provides information about operating, maintaining, and servicing the analyzer. It also contains important alerts to ensure safe operation and prevent equipment damage. The manual is organized into the following chapters and appendices to provide direct access to specific operation and service information:
Chapter 1 “Introduction” provides an overview of product features, describes the principles of operation, and lists the specifications.
Chapter 2 “Installation” describes how to unpack, setup, and startup the analyzer.
Chapter 3 “Operation” describes the front panel display, the front panel pushbuttons, and the menu-driven software.
Chapter 4“Preventive Maintenance” provides maintenance procedures to ensure reliable and consistent instrument operation.
Chapter 5 “Troubleshooting” presents guidelines for diagnosing analyzer failures, isolating faults, and includes recommended actions for restoring proper operation.
Chapter 6 “Servicing” presents safety alerts for technicians working on the analyzer, step-by-step instructions for repairing and replacing components, and a replacement parts list. It also includes contact information for product support and technical information.
Chapter 7 “System Description” describes the function and location of the system components, provides an overview of the software structure, and includes a description of the system electronics and input/output connections.
Chapter 8 “Optional Equipment” describes the optional equipment that can be used with this analyzer.
Appendix A “Warranty” is a copy of the warranty statement.
Appendix B “C-Link Protocol Commands” provides a description of the C-Link protocol commands that can be used to remotely control an analyzer using a host device such as a PC or a datalogger.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual xv
Page 16
Preface
Appendix C “MODBUS Protocol” provides a description of the MODBUS Protocol Interface and is supported both over RS-232/485 (RTU protocol) as well as TCP/IP over Ethernet.
Appendix D “Geysitech (Bayern-Hessen) Protocol” provides a description of the Geysitech (Bayern-Hessen) Protocol Interface and is supported over RS-232/485 and TCP/IP over Ethernet.
Safety Review the following safety information carefully before using the analyzer.
This manual provides specific information on how to operate the analyzer, however if the analyzer is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Safety and Equipment
Damage Alerts
This manual contains important information to alert you to potential safety hazards and risks of equipment damage. Refer to the following types of alerts you may see in this manual.
Safety and Equipment Damage Alert Descriptions
Alert Description
DANGER A hazard is present that will result in death or serious
personal injury if the warning is ignored.
WARNING A hazard is present or an unsafe practice can result in
serious personal injury if the warning is ignored.
CAUTION The hazard or unsafe practice could result in minor to
moderate personal injury if the warning is ignored.
Equipment Damage The hazard or unsafe practice could result in property
damage if the warning is ignored.
xvi Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 17
Safety and Equipment Damage Alerts in this Manual
Alert Description
WARNING If the equipment is operated in a manner not specified by
the manufacturer, the protection provided by the equipment may be impaired.
The service procedures in this manual are restricted to qualified service personnel only.
The Model 49i Primary Standard is supplied with a three-wire grounding cord. Under no circumstances should this grounding system be defeated.
CAUTION If the LCD panel breaks, do not to let the liquid crystal
contact your skin or clothes. If the liquid crystal contacts your skin or clothes, wash it off immediately using soap and water.
Equipment Damage Do not attempt to lift the analyzer by the cover or other
external fittings.
Some internal components can be damaged by small amounts of static electricity. A properly grounded antistatic wrist strap must be worn while handling any internal component.
This adjustment should only be performed by an instrument service technician.
Handle all printed circuit boards by the edges only. Do not remove the panel or frame from the LCD module.
The LCD module polarizing plate is very fragile, handle it carefully.
Do not wipe the LCD module polarizing plate with a dry cloth, it may easily scratch the plate.
Do not use Ketonics solvent or aromatic solvent to clean the LCD module, use a soft cloth moistened with a naphtha cleaning solvent.
Do not place the LCD module near organic solvents or corrosive gases.
Do not shake or jolt the LCD module.
Preface
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual xvii
Page 18
Preface
FCC Compliance Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the equipment.
Note This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
WEEE Symbol The following symbol and description identify the WEEE marking used on
the instrument and in the associated documentation.
Symbol Description
Marking of electrical and electronic equipment which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market after 13 August 2005.
Where to Get Help Service is available from exclusive distributors worldwide. Contact one of
the phone numbers below for product support and technical information or visit us on the web at www.thermo.com/aqi.
1-866-282-0430 Toll Free
1-508-520-0430 International
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual xviii
Page 19
Chapter 1 Introduction
The Model 49i Primary Standard UV Photometric Ozone Calibrator combines proven detection technology, easy to use menu-driven software, and advanced diagnostics to offer unsurpassed flexibility and reliability. The Model 49i Primary Standard has the following features:
320 x 240 graphics display
Menu-driven software
Multiple user-defined analog outputs
Analog input options
Fast internal ozonator
Fast response time
Short lag time
Automatic temperature and pressure compensation
User-selectable digital input/output capabilities
Standard communications features include RS232/485 and Ethernet
C-Link, MODBUS, Geysitech (Bayern-Hessen) protocol, and streaming data protocols.
For details of the calibrator’s principle of operation and product specifications, see the following topics:
“Principle of Operation” on page 1-2 describes the principles by which your calibrator operates.
“Specifications” on page 1-3 is a list of the calibrator’s performance specifications.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 1-1
Page 20
Introduction
Principle of Operation
Thermo Fisher Scientific is pleased to supply this UV photometric ozone calibrator. We are committed to the manufacture of instruments exhibiting high standards of quality, performance, and workmanship. Service personnel are available for assistance with any questions or problems that may arise in the use of this calibrator. For more information on servicing, see Chapter 7, “Servicing”.
Principle of Operation The Model 49i Primary Standard operates on the principle that ozone (O
molecules absorb UV light at a wavelength of 254 nm. The degree to which the UV light is absorbed is directly related to the ozone concentration as described by the Beer-Lambert Law:
I
I
o
where:
K = molecular absorption coefficient, 308 cm-1 (at 0°C and 1 atmosphere)
L = length of cell, 38 cm
C = ozone concentration in parts per million (ppm)
I = UV light intensity of sample with ozone (sample gas)
Io = UV light intensity of sample without ozone (reference gas)
KLC
=e
)
3
Zero air is supplied to the Model 49i Primary Standard through the zero air bulkhead and is split into two gas streams, as shown in Figure 1-1. One gas stream flows through a pressure regulator to the reference solenoid valve to become the reference gas (I
). The second zero air stream flows through a
o
pressure regulator, ozonator, and manifold to the sample solenoid valve to become the sample gas (I). Ozone from the manifold is delivered to the ozone bulkhead. The solenoid valves alternate the reference and sample gas streams between cells A and B every 10 seconds. When cell A contains reference gas, cell B contains sample gas and vice versa.
The UV light intensities of each cell are measured by detectors A and B. When the solenoid valves switch the reference and sample gas streams to opposite cells, the light intensities are ignored for several seconds to allow the cells to be flushed. The Model 49i Primary Standard calculates the
1-2 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 21
Introduction
Specifications
ozone concentration for each cell and outputs the average concentration to the front panel display, the analog outputs, and also makes the data available over the serial or ethernet connection.
Figure 1-1. Model 49i Primary Standard Flow Schematic
Specifications Table 1-1. Model 49i Primary Standard Specifications
Photometer
Preset ranges 50, 100, 200, 500, 1000, 2000, 5000 ppb
100, 200, 500, 1000, 2000, 5000, 10000 μg/m
Custom ranges 50 to 5000 ppb
100 to 10000 μg/m
Zero roise 0.25 ppb RMS (60 second averaging time)
Lower detectable limit 1 ppb
Response time 20 seconds (0-95%)
Lag time 10 seconds
Linearity ±
Flow rate 1
Operating temperature
Power requirements 100 VAC @ 50/60 Hz
1% of full-scale
3 LPM
0–45 °C*
115 VAC @ 50/60 Hz
220
240 VAC @ 50/60 Hz
150 watts
3
3
Physical dimensions 16.75” (W) X 8.62” (H) X 23” (D)
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 1-3
Page 22
Introduction
Specifications
Photometer
Weight Approximately 35 lbs.
Analog outputs 6 voltage outputs; 0
100 mV, 1, 5, 10 V (user selectable), 5% of
full-scale over/under range, 12 bit resolution, user selectable for measurement input
Digital outputs 1 power fail relay Form C, 10 digital relays Form A, user selectable
alarm output, relay logic, 100 mA @ 200 VDC
Digital inputs 16 digital inputs, user select programmable, TTL level, pulled high
Serial Ports 1 RS-232 or RS-485 with two connectors, baud rate 1200
115200,
data bits, parity, and stop bits, protocols: C-Link, MODBUS, and streaming data (all user selectable)
Ethernet connection RJ45 connector for 10Mbs Ethernet connection, static or dynamic
TCP/IP addressing
Ozonator
Output 0.025-1.000 ppm at 3-4 LPM
Response (to 98% or 5
1 minute ppb of final value, whichever is greater)
Stability ±
*In non condensing environments.
4 ppb or ± 1% of reading, whichever is greater
1-4 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 23
Chapter 2 Installation
Installing the Model 49i Primary Standard includes the following recommendations and procedures:
“Lifting” on page 2-1
“Unpacking and Inspection” on page 2-1
“Setup Procedure” on page 2-3
“Connecting External Devices” on page 2-4
“Startup” on page 2-8
Lifting When lifting the instrument, use procedure appropriate to lifting a heavy
object, such as, bending at the knees while keeping your back straight and upright. Grasp the instrument at the bottom in the front and at the rear of the unit. Although one person can lift the unit, it is desirable to have two persons lifting, one by grasping the bottom in the front and the other by grasping the bottom in the rear.
Equipment Damage Do not attempt to lift the instrument by the cover or other external fittings.
Unpacking and
Inspection
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 2-1
The Model 49i Primary Standard is shipped complete in one container. If there is obvious damage to the shipping container when you receive the instrument, notify the carrier immediately and hold for inspection. The carrier is responsible for any damage incurred during shipment.
Use the following procedure to unpack and inspect the instrument.
1. Remove the instrument from the shipping container and set it on a table or bench that allows easy access to both the front and rear.
2. Remove the cover to expose the internal components.
Page 24
Installation
Unpacking and Inspection
3. Remove the packing material (Figure 2-1).
Remove Packing (2 pieces)
Remove Packing (2 pieces)
Units without Optional I/O Board
Units with Optional I/O Board
Figure 2-1. Remove the Packing Material
4. Remove the three shipping screws (Figure 2-2).
Shipping Screws (3)
Figure 2-2. Remove the Shipping Screws
5. Check for possible damage during shipment.
6. Check that all connectors and circuit boards are firmly attached.
2-2 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 25
7. Re-install the cover.
8. Remove any protective plastic material from the case exterior.
Setup Procedure Use the following procedure to setup the instrument.
1. Connect the zero air supply to the ZERO AIR bulkhead on the rear panel (Figure 2-3). All tubing should be constructed of FEP Teflon®, 316 stainless steel, borosilicate glass, or similar tubing with an OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the tubing should be less than 10 feet.
2. Connect the EXHAUST bulkhead to a suitable vent. The exhaust line should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of the exhaust line should be less than 10 feet. Verify that there is no restriction in this line.
Installation
Setup Procedure
3. Connect a suitable recording device to the rear panel connector. See the “Operation” chapter for more information about the rear panel pin-outs.
4. Plug the instrument into an outlet of the appropriate voltage and frequency.
WARNING The Model 49i Primary Standard is supplied with a three-wire grounding cord. Under no circumstances should this grounding system be defeated.
5. Connect the instrument to be calibrated to the OZONE bulkhead. The total flow though the ozonator should be set to at least 1 LPM.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 2-3
Page 26
Installation
Connecting External Devices
Connecting External
Devices
Terminal Board PCB
Assemblies
Figure 2-3. Model 49i Primary Standard Rear Panel
Several components are available for connecting external devices to iSeries instruments.
These connection options include:
Individual terminal board PCB assemblies
Terminal block and cable kits (optional)
Individual cables (optional)
For detailed information on the optional connection components, refer to the “Optional Equipment” chapter. For associated part numbers, refer to “External Device Connection Components” in the “Servicing” chapter.
The following terminal board PCB assemblies are available for iSeries instruments:
I/O terminal board PCB assembly, 37 pin (standard)
2-4 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 27
Installation
Connecting External Devices
D/O terminal board PCB assembly, 37 pin (standard)
25-pin terminal board PCB assembly, (included with optional I/O Expansion Board)
I/O Terminal Board Figure 2-4 shows the recommended method for attaching the cable to the
terminal board using the included tie-down and spacer. Ta b le 2 - 1 identifies the connector pins and associated signals.
Note Not all of the available I/O for the instrument are brought out on this terminal. If more I/O is desired, you must use an alternative connection method.
See Detail “A”
Component Side Viewed from Top of Board
Assembled Connector
See Detail “B”
Figure 2-4. I/O Terminal Board Views
Detail “A”
Detail “B”
Table 2-1. I/O Terminal Board Pin Descriptions
Pin Signal Description Pin Signal Description
1 Analog1 13 Power_Fail_NC
2 Analog ground 14 Power_Fail_COM
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 2-5
Page 28
Installation
Connecting External Devices
Table 2-1. I/O Terminal Board Pin Descriptions, continued
Pin Signal Description Pin Signal Description
3 Analog2 15 Power_Fail_NO
4 Analog ground 16 TTL_Input1
5 Analog3 17 TTL_Input2
6 Analog ground 18 TTL_Input3
7 Analog4 19 TTL_Input4
8 Analog ground 20 Digital ground
9 Analog5 21 TTL_Input5
10 Analog ground 22 TTL_Input6
11 Analog6 23 TTL_Input7
12 Analog ground 24 Digital ground
D/O Terminal Board Figure 2-5 shows the recommended method for attaching the cable to the
terminal board using the included tie-down and spacer. Table 2-2 identifies the connector pins and associated signals.
See Detail “A”
Component Side Viewed from Top of Board
Assembled Connector
See Detail “B”
Detail “A”
Detail “B”
Figure 2-5. D/O Terminal Board Views
2-6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 29
Connecting External Devices
Table 2-2. D/O Terminal Board Pin Descriptions
Pin Signal Description Pin Signal Description
1 Relay1_ContactA 13 Relay7_ContactA
2 Relay1_ContactB 14 Relay7_ContactB
3 Relay2_ContactA 15 Relay8_ContactA
4 Relay2_ContactB 16 Relay8_ContactB
5 Relay3_ContactA 17 Relay9_ContactA
6 Relay3_ContactB 18 Relay9_ContactB
7 Relay4_ContactA 19 Relay10_ContactA
8 Relay4_ContactB 20 Relay10_ContactB
9 Relay5_ContactA 21 Solenoid_Drive_Output1
10 Relay5_ContactB 22 +24V
11 Relay6_ContactA 23 Solenoid_Drive_Output2
Installation
12 Relay6_ContactB 24 +24V
25-Pin Terminal Board The 25-pin terminal board is included with the optional I/O Expansion
Board.
See Detail “A”
Component Side Viewed from Top of Board
See Detail “B”
Detail “A”
Detail “B”
Assembled Connector
Figure 2-6. 25-Pin Terminal Board Views
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 2-7
Page 30
Installation
Startup
Table 2-3. 25-Pin Terminal Board Pin Descriptions
Pin Signal Description Pin Signal Description
1 IOut1 13 Analog_In1
2 GND_ISO 14 Analog_In2
3 IOut2 15 Analog_In3
4 GND_ISO 16 GNDD
5 IOut3 17 Analog_In4
6 GND_ISO 18 Analog_In5
7 IOut4 19 Analog_In6
8 GND_ISO 20 GNDD
9 IOut5 21 Analog_In7
10 GND_ISO 22 Analog_In8
11 IOut6 23 GNDD
12 GND_ISO 24 GNDD
Startup Use the following procedure when starting the instrument.
1. Turn the power ON.
2. Allow 90 minutes for the instrument to stabilize.
3. Set instrument parameters such as operating ranges and averaging times to appropriate settings. For more information about instrument parameters, see the “Operation” chapter.
2-8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 31
Chapter 3 Operation
This chapter describes the front panel display, front panel pushbuttons, and menu-driven software.
“Display” on page 3-2 describes the LCD graphics display.
“Pushbuttons” on page 3-3 describes the various front panel pushbuttons and the expected key actions for each.
“Software Overview” on page 3-4 describes the menu-driven software and submenus.
“Range Menu” on page 3-7 describes the gas units, O3 range, and custom ranges.
“Averaging Time” on page 3-12 describes the averaging period applied to O3 measurements.
“Calibration Factors Menu” on page 3-12 describes the calibration factors used to make minor corrections to the O3 measurement readings.
“Instrument Controls Menu” on page 3-15 describes the items that control the calibrator.
“Diagnostics Menu” on page 3-50 describes the diagnostic information and functions.
“Alarms Menu” on page 3-58 describes a list of items that are monitored by the calibrator.
“Service Menu” on page 3-66 describes service related menu items.
“Password Menu” on page 3-82 describes how to enter/change a password, lock and unlock the calibrator.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-1
Page 32
Operation
Display
Display The 320 x 240 graphics liquid-crystal display (LCD) shows the sample
concentrations, instrument parameters, instrument controls, help, and error messages. Some menus contain more items than can be displayed at one time. For these menus, use and to move the cursor up and down to each item.
Figure 3-1. 49i Primary Standard Front Panel Display
CAUTION If the LCD panel breaks, do not let the liquid crystal contact your
skin or clothes. If the liquid crystal contacts your skin or clothes, wash it off immediately using soap and water.
3-2 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 33
Pushbuttons
Pushbuttons The Pushbuttons allow the user to traverse the various screens/menus.
Figure 3-2. Front Panel Pushbuttons
Operation
Table 3-1. Front Panel Pushbuttons
= Soft Keys The soft keys are used to provide shortcuts that
allow the user to jump to user-selectable menu screens. For more information on processing soft keys, see “Soft Keys” below.
= Run The is used to display the Run screen. The Run
screen normally displays the O
= Menu The is used to display the Main Menu when in the
Run screen, or back up one level in the menu system. For more information about the Main Menu, see “Main Menu” later in this chapter.
= Help The is context-sensitive, that is, it provides
additional information about the screen that is being displayed. Press for a brief explanation about the current screen or menu. Help messages are displayed using lower case letters to easily distinguish them from the operating screens. To exit a help screen, press or
return to the Run screen.
= Up, Down
= Left, Right
The four arrow pushbuttons ( , , , and ) move the cursor up, down, left, and right or change values and states in specific screens.
concentration.
3
to return to the previous screen, or to
= Enter The is used to select a menu item,
accept/set/save a change, and/or toggle on/off functions.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-3
Page 34
Operation
Software Overview
Soft Keys The Soft Keys are multi-functional keys that use part of the display to
identify their function at any moment. The function of the soft keys allow immediate access to the menu structure and most often used menus and screens. They are located directly underneath the display and as the keys' functions change this is indicated by user-defined labels in the lower part of the display, so that the user knows what the keys are to be used for.
To change a soft key, place the menu cursor “>” on the item of the selected menu or screen you wish to set. Press followed by the selected soft key within 1 second of pressing the right-arrow key. The edit soft key prompt will be displayed for configuration for the new label.
Note Not all menu items may be assigned to soft keys. If a particular menu or screen item cannot be assigned, the key assignment screen will not come up upon entering right-arrow-soft key combinations. All items under the Service menu (including the menu itself) cannot be assigned soft keys.
EDIT SOFT KEY PROMPT: CURRENTLY: RANGE
RANGE
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE 0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Software Overview The Model 49i Primary Standard utilizes the menu-driven software
illustrated by the flowchart in Figure 3-3. The Power-Up screen, shown at the top of the flowchart, is displayed each time the instrument is turned on. This screen is displayed while the instrument is warming up and performing self-checks. After the warm-up period, the Run screen is automatically displayed. The Run screen is the normal operating screen. It displays the O concentration, depending on operating mode. From the Run screen, the Main Menu can be displayed by pressing . The Main Menu contains a list of submenus. Each submenu contains related parameters and/or functions. This chapter describes each submenu and screen in detail. Refer to the appropriate sections for more information.
3
3-4 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 35
2CUUYQTF
5GV2CUUYQTF
%JCPIG2CUUYQTF
.QEM+PUVTWOGPV
4GOQXG2CUUYQTF
7PNQEM+PUVTWOGPV
2QYGTÄWR
5ETGGP
5GTXKEG
2TGUUWTG%JGEM
#NCTOU
#NCTOU&GVGEVGF
&KCIPQUVKEU
8QNVCIGU
2TQITCO8GTUKQP
5ETGGP
5GNH6GUV
4WP5ETGGP
/CKP/GPW
+PUVTWOGPV
%QPVTQNU
%WUVQO.GXGNU
2WOR
+PVGPUKV[%JGEM
.COR5GVVKPI
&GVGEVQT%CN
1.COR6GOR
.COR6GOR
$GPEJ6GOR
6GORU
2TGUUWTG
6GOR%QORGPUCVKQP
2TGU%QORGPUCVKQP
14GURQPUG%QGH
/CPWCN1.COR5GVVKPI
2TGUUWTG
(NQY#
%GNN#$1
(NQYU
%QOO5GVVKPIU
&CVCNQIIKPI5GVVKPIU
(NQY#%CN
(NQY$%CN
2TGUUWTG%CN
+PVGPUKV[#
+PVGPUKV[$
(NQY$
#PCNQI+PRWV4GCFKPIU
#PCNQI+PRWV8QNVCIGU
+PVGPUKVKGU
5ETGGP%QPVTCUV
5GTXKEG/QFG
+1%QPHKI
6GOR%CN
#PCNQI1WVRWV%CN
#PCNQI+PRWV%CN
1%QPE
/$5VCVWU
+PV$F5VCVWU
6GUV#PCNQI1WVRWVU
&KIKVCN+PRWVU
4GNC[5VCVGU
&CVG6KOG
&KURNC[2KZGN6GUV
4GUVQTG7UGT&GHCWNVU
+1'ZR$F5VCVWU
%QPVCEV+PHQ
+PUVTWOGPV%QPHKI
%CNKDTCVKQP
#XGTCIKPI
4CPIG
(CEVQTU
6KOG
1$MI
1%QGH
4GUGV7UGT%CN&GHCWNVU
#XI6KOG
)CU7PKVU
5GV%WUVQO4CPIGU
4CPIG
Figure 3-3. Flowchart of Menu-Driven Software
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-5
Page 36
Operation
Software Overview
Power-Up Screen The Power-Up screen is displayed on power up of the Model 49i Primary
Standard. The Self-Test is displayed while the internal components are warming up and diagnostic checks are performed.
Run Screen The Run screen displays the O
sample/cal solenoid valve or internal ozonator, time, and alarm status. The word “ZERO” on the bottom left of the display indicates the calibrator is in zero mode. Other modes appear in the same area of the display as “LEVEL 1”, “LEVEL 2”, LEVEL 3”, “LEVEL 4” and “LEVEL 5”, or blank to signify manual mode, as shown below. In manul mode, use and to increase and decrease the O3 output concentration. Press to scroll through manual, zero or custom levels.
X CONCENTRATION W
NO 28.7 PPB
O3 600 PPB
SET TO: 0
©ª CHANGE VALUE
SAMPLE 12:34 ALARMx
concentration. The status bar displays
3
NO
SAVE
RANGE AVG DIAGS ALARM
3-6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 37
Operation
Range Menu
Main Menu The Main Menu contains a number of submenus. Parameters and features
are divided into these submenus according to their function. The concentration appears above the main menu and submenus in every screen. The Service menu is visible only when the calibrator is in service mode. For more information on the service mode, see “Service Mode” later in this chapter.
Use and to move the cursor up and down.
Press to select a choice.
Press to return to the Main Menu or to return to the Run screen.
MAIN MENU: >RANGE AVERAGING TIME CALIBRATION FACTORS INSTRUMENT CONTROLS DIAGNOSTICS ALARMS SERVICE
ª
RANGE AVG DIAGS ALARM
PASSWORD
Range Menu The Range menu allows the operator to select the gas units, O
to set the custom ranges.
In the Main Menu, choose Range.
RANGE: >GAS UNITS PPB RANGE 500 SET CUSTOM RANGES
RANGE AVG DIAGS ALARM
range, and
3
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-7
Page 38
Operation
Range Menu
Gas Units The Gas Units screen defines how the O
expressed. Gas units of parts per billion (ppb), parts per million (ppm), micrograms per cubic meter (μg/m3), or milligrams per cubic meter (mg/m3) are available. The μg/m3 and mg/m3 gas concentration modes are calculated using a standard pressure of 760 mmHg and a standard temperature of 20 °C.
When switching the selected units from ppb or ppm to μg/m3 or mg/m3, the analog ranges all default to the highest range in that mode. For example, when switching from mg/m3 to ppm, all the ranges default to 5 ppm. Therefore, whenever you change units, you should also check the range settings.
In the Main Menu, choose Range > Gas Units.
•Use and to scroll through a list of choices.
Press to save the new units.
Note If the units change from ppb/ppm to μg/m3/mg/m3 or vice-versa, the instrument should be re-calibrated, particularly if the user’s standard temperature is different from 20 °C. A display warning will appear that ranges will be defaulted and calibration parameters reset.
concentration reading is
3
GAS UNITS: CURRENTLY: PPB SET TO: UG/M3 ? AND SET RANGES AND CAL TOX DEFAULT SETTINGSXXXXXXXXXX
©ª CHANGE VALUE
RANGE AVG DIAGS ALARM
O3 Range The O
Range screen defines the concentration range of the analog outputs.
3
SAVE
For example, an O3 range of 0–500 ppb restricts the analog output to concentrations between 0 and 500 ppb.
The display shows the current O
range. The next line of the display is used
3
to change the range.
Table 3-2 lists the standard ranges.
3-8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 39
In the Main Menu, choose Range > O3 Range.
•Use and to scroll through a list of choices.
Press to save the new range.
O3 RANGE: CURRENTLY: 500.0 SET TO: 1000.0 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Table 3-2. Standard Ranges
ppb ppm μg/m
3
mg/m
3
Operation
Range Menu
50 0.05 100 0.1
100 0.10 200 0.2
200 0.20 400 0.4
500 0.50 1,000 1
1,000 1 2,000 2
2,000 2 5,000 5
5,000 5 10,000 10
C1 C1 C1 C1
C2 C2 C2 C2
C3 C3 C3 C3
C1, C2, and C3 are custom ranges. The analog outputs are arranged on the rear panel connector as shown in Figure 3-4. See Table 3-3 for channels and pin connections. For more information about custom ranges, see “Set Custom Ranges” below.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-9
Page 40
Operation
Range Menu
Figure 3-4. Pin-Out of Rear Panel Connector
Table 3-3. Default Analog Outputs
Channel Pin I/O Terminal Pin Description
114 1 O
233 3 O
3 15 5 None
4 34 7 None
5 17 9 None
6 36 11 None
Ground 16, 18, 19, 35, 37 2, 4, 6, 8, 10, 12 Signal Ground
Analog Output
3
Analog Output
3
Note All channels are user definable. If any customization has been made to the analog output configuration, the default selections may not apply.
3-10 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 41
Operation
Range Menu
Set Custom Ranges The Set Custom Ranges menu lists three custom ranges: C1, C2, and C3.
Custom ranges are user-defined ranges. In the standard range mode, any value between 50 ppb (0.05 ppm) and 5,000 ppb (5 ppm) can be specified as a range. In the μg/m3(mg/m3) mode, any value between 100 μg/m3 (0.1 mg/m3) and 10,000 μg/m3 (10 mg/m3) can be specified as a range.
In the Main Menu, choose Range > Set Custom Ranges.
CUSTOM RANGES: >CUSTOM RANGE 1 550.0 CUSTOM RANGE 2 650.0 CUSTOM RANGE 3 750.0
RANGE AVG DIAGS ALARM
Custom Ranges The Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display is used to set the range. To use the custom full-scale range, be sure to select it (Custom range 1, 2, or 3) in the O3 Range screen. For more information about selecting ranges, see “O3 Range” above.
In the Main Menu, choose Range > Set Custom Ranges > Custom range 1, 2, or 3.
•Use and to move the cursor left or right.
Use and to increment or decrement the numeric value.
Press to save the new range.
CUSTOM RANGE 1: CURRENTLY: 550.0 SET TO: 000550.1 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-11
SAVE
Page 42
Operation
Averaging Time
Averaging Time The averaging time defines a time period (10 to 300 seconds) over which
O3 measurements are taken. The average concentration of the readings are calculated for that time period. The front panel display and analog outputs are updated every 10 seconds with the calculated averages. An averaging time of 10 seconds, for example, means that the average concentration of the last 10 seconds will be output at each update. An averaging time of 300 seconds means that the moving average concentration of the last 300 seconds will be output at each 10-second update. Therefore, the lower the averaging time, the faster the front panel display and analog outputs respond to concentration changes. Longer averaging times are typically used to smooth output data.
The following averaging times are available: 10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds.
In the Main Menu, choose Averaging Time.
Use and to scroll through a list of choices.
Calibration Factors
Menu
Press to save the averaging time.
AVERAGING TIME: CURRENTLY: 30 SEC SET TO: 10 SEC ?
RANGE AVG DIAGS ALARM
©ª
CHANGE VALUE

SAVE VALUE
Calibration factors are used to make minor compensations to the O3 concentration, when the Model 49i Primary Standard is being precisely matched to a NIST traceable O
calibrator. The Calibration Factors menu
3
displays the calibration factors.
In the Main Menu, choose Calibration Factors
3-12 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 43
.
CALIBRATION FACTORS: >O3 BKG 0.0 O3 COEF 1.000 RESET USER CAL DEFAULTS
RANGE AVG DIAGS ALARM
Operation
Calibration Factors Menu
O3 Background The O
This adjustment should only be made when the Model 49i Primary Standard is being precisely matched to a NIST traceable O Normally, the Model 49i Primary Standard does not require any background compensation.
In the Main Menu, choose Calibration Factors > O3 Background.
•Use and to increment or decrement the proposed background value.
Press to save the new background.
O3 BACKGROUND: O3: 4.4 SET BKG TO: 0.0
RANGE AVG DIAGS ALARM
Background is used to make small adjustments to the zero level.
3
calibrator.
3
©ª INC/DEC
SAVE VALUE
O3 Span Coefficients The O
span coefficients is s used to make small adjustments to the output
3
level. This adjustment should only be made when the Model 49i Primary Standard is being precisely matched to a NIST traceable O3 calibrator. Normally, the Model 49i Primary Standard does not require any compensation.
Note The concentration value will show ERROR if the measured concentration is not a valid span value (either higher than the selected range, or 0 or lower).
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-13
Page 44
Operation
Calibration Factors Menu
In the Main Menu, choose Calibration Factors > O3 Coef.
Use and to increment or decrement the coefficient value.
Press to save the new coefficient.
O3 COEFFICIENT: O3: 600 SET COEF TO: 1.000
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Reset User Calibration Default The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
In the Main Menu, choose Service > Calibration Factors > Reset User Cal Defaults.
Press to warn user and enable restore with .
Use to reset the calibration pressure reference value when pressed after .
RESTORE DEFAULT CAL: RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
ARE YOU SURE YOU WANT TO?x PRESS ¨ TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
RESTORE
3-14 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 45
Operation
Instrument Controls Menu
Instrument Controls
Menu
Custom Levels The Custom Levels menu contains four custom levels: 1, 2, 3, 4 and 5. In
The Instrument Controls menu contains items that control the calibrator.
In the Main Menu, choose Instrument Controls.
INSTRUMENT CONTROLS: >CUSTOM LEVELS PUMP TEMPERATURE COMPENSATION PRESSURE COMPENSATION DATALOGGING SETTINGS COMMUNICATION SETTINGS I/O CONFIGURATION
RANGE AVG DIAGS ALARM
SCREEN CONTRAST SERVICE MODE DATE/TIME
ª
the Run screen, use to scroll through Zero, Level 1, Level 2, Level 3, Level 4, Level 5. Press enter to select a custom level zero.
In the Main Menu, choose Instrument Controls > Custom Levels.
CUSTOM O3 LEVELS: >LEVEL 1 425 LEVEL 2 325 LEVEL 3 215 LEVEL 4 125
LEVEL 5 50
RANGE AVG DIAGS ALARM
Custom Levels The Custom Levels screen is used for setting user-defined concentration
levels.
In the Main Menu, choose Instrument Controls > Custom Levels > Custom Level 1, 2, 3, 4 or 5.
Use and to increment or decrement the numeric value.
•Press to save the new level.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-15
Page 46
Operation
Instrument Controls Menu
CUSTOM LEVEL 1: CONC SETTING: 450 ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Pump The Pump screen allows the user to toggle the pump on and off.
In the Main Menu, choose Instrument Controls > Pump.
•Press to toggle and set the pump on or off.
Temperature
Compensation
PUMP: CURRENTLY: OFF SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Temperature compensation provides compensation for any changes to the instrument's output signal due to variations in sample gas temperature. The Model 49i Primary Standard can be operated with or without temperature compensation.
When temperature compensation is on, the display shows the sample gas temperature (measured by a thermistor on the Interface board). When temperature compensation is off, the display shows the standard temperature of 0.0 °C.
In the Main Menu, choose Instrument Controls > Temperature Compensation.
•Press to toggle and set the temperature compensation on or off.
3-16 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 47
Operation
Instrument Controls Menu
TEMPERATURE COMPENSATION: COMP TEMP: 0.0 CURRENTLY: OFF SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
o
C
Pressure Compensation Pressure compensation provides compensation for any changes to the
instrument's output signal due to variations of sample pressure. The Model 49i Primary Standard can be operated with or without pressure compensation.
When pressure compensation is on, the first line of the display represents the current sample pressure. When pressure compensation is off, the first line of the display shows the standard pressure of 760 mmHg.
In the Main Menu, choose Instrument Controls > Pressure Compensation.
•Press to toggle and set the pressure compensation on or off.
PRESSURE COMPENSATION: COMP PRES: 760.0 mmHg CURRENTLY: OFF SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Datalogging Settings The Datalogging Settings menu deals with datalogging.
In the Main Menu, choose Instrument Controls > Datalogging Settings.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-17
Page 48
Operation
Instrument Controls Menu
Select SREC/LREC The Select SREC/LREC is used to select short record or long record format
DATALOGGING SETTINGS: >SELECT SREC/LREC SREC VIEW LOGGED DATA ERASE LOG SELECT CONTENT COMMIT CONTENT RESET TO DEFAULT CONTENT CONFIGURE DATA LOGGING
RANGE AVG DIAGS ALARM
for other operations in this menu.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select SREC/LREC.
•Press to toggle and set to either short record or long record format.
SELECT LOG TYPE: CURRENTLY: SREC SET TO: LREC ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
View Logged Data The View Logged Data screen is used to select the start point to view the
logged data by number of records or date and time.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select SREC or LREC > View Logged Data.
Use and to toggle between number of records or date and time.
•Press to accept.
3-18 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 49
Operation
Instrument Controls Menu
SELECT START POINT BY:
SET TO: # OF RECS
©ª
CHANGE
RANGE AVG DIAGS ALARM
ACCEPT
Number of Records The Number of Records screen is used to select the starting point to display
the number of records back to view.
•Use and to move the cursor left or right.
Use and to increment or decrement the numeric value.
•Press to set the starting record and continue to the record display screen.
SET # BACK FROM CURRENT: 0000000 TOTAL LRECS: 20
§¨
MOVE CURSOR
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM
SAVE
The Record Display screen (read only) displays the selected records.
•Use
and to view all the items in the record.
Use and to view records at different times.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-19
Page 50
Operation
Instrument Controls Menu
time date flags 10:00 06/20/05 FC0088900 11:00 06/20/05 FC0088900 12:00 06/20/05 FC0088900 13:00 06/20/05 FC0088900
©ª
PGUP/DN
RANGE AVG DIAGS ALARM
§¨
PAN L/R
Date and Time The Date and Time screen is used to set a start date and time for which to
view logged data. For example, if “20 JUN 2005 10:00” is entered, then the first logged data record that is displayed is the first record after this time. If set to one minute logging, this would be at “20 JUN 2005 10:01”.
Use and to increment or decrement the selected field.
•Use to advance to next field.
•Press to set the date and time of the first record to be displayed and continue to the record display screen.
DATE AND TIME: 20 JUN 2005 10:00
©ª
CHG DAYS
¨
SET CURSOR TO MONTHS ACCEPT AS SHOWN
RANGE AVG DIAGS ALARM
The Record Display screen (read only) displays the selected records.
•Use and to view all the items in the record.
Use and to view records at different times.
3-20 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 51
Operation
Instrument Controls Menu
time date flags 10:01 06/20/05 FC0088900 10:02 06/20/05 FC0088900 10:03 06/20/05 FC0088900 10:04 06/20/05 FC0088900
©ª
PGUP/DN
RANGE AVG DIAGS ALARM
§¨
PAN L/R
Erase Log The Erase Log is used to erase all saved data for the selected record type (not
all short records and long records).
In the Main Menu, choose Instrument Controls > Datalogging Settings > Erase Log.
•Press to erase all data.
•Press to confirm erasure.
ERASE LREC LOG FILE DATA? ERASE LREC LOG FILE DATA?
ERASE
RANGE AVG DIAGS ALARM
ARE YOU SURE YOU WANT TO?x PRESS ¨ TO CONFIRM ERASURE
RANGE AVG DIAGS ALARM
ERASE
Select Content The Select Content submenu displays a list of 32 record fields to use and a
submenu list of the analog output signal group choices to choose from. Choices are Concentrations, Other Measurements, and Analog Inputs (if the I/O expansion board is installed). This is a temporary list of items for the selected record type that must be committed via the datalogging menu before the changes will apply. Note that committing any changes to this list will erase all currently logged data, as the format of the stored data is changed.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select Content.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-21
Page 52
Operation
Instrument Controls Menu
LREC FIELDS: >FIELD 1 O3 FIELD 2 CELLAI FIELD 3 CELLBI FIELD 4 BNCHT FIELD 5 LMPT FIELD 6 O3LT FIELD 7 FLOWA
RANGE AVG DIAGS ALARM
ª
Choose Item Type The Choose Item Type submenu displays a list of data that can be logged for
the current field. Choices and Concentrations, Other Measurements, and Analog Inputs (if the I/O expansion board is installed).
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select Content > Field 1-32.
DATA IN LREC FIELD 1: >CONCENTRATIONS OTHER MEASUREMENTS ANALOG INPUTS
RANGE AVG DIAGS ALARM
Concentrations The Concentrations screen allows the user to select the output signal that is
tied to the selected field item. The selected item is shown by “<--” after it. Note that at this point, pressing indicates that these are proposed changes as opposed to implemented changes. To change the selected record format and erase record log file data, see “Commit Content” below.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select Content > Select Field > Concentrations.
Use and to move the cursor up and down.
Press to select a new choice.
3-22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 53
Operation
Instrument Controls Menu
CONCENTRATIONS: >NONE
O3 <--
RANGE AVG DIAGS ALARM
Other Measurements The Other Measurements screen allows the user to select the output signal
that is tied to the selected field item. The selected item is shown by “<--” after it. Items displayed are determined by the options installed. Note that at this point, pressing indicates that these are proposed changes as opposed to implemented changes. To change the selected record format and erase record log file data, see “Commit Content” below.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select Content > Select Field > Other Measurements.
Use and to move the cursor up and down.
Press to select a new choice.
OTHER MEASUREMENTS: >NONE INTENSITY A INTENSITY B NOISE A NOISE B FLOW A FLOW B
RANGE AVG DIAGS ALARM
PRESSURE BENCH TEMP LAMP TEMP O3 LAMP TEMP
ª
Analog Inputs The Analog Inputs screen allows the user to select the output signal (none
or analog inputs 1-8) that is tied to the selected field item. The selected item is shown by “<--” after it. Note that at this point, pressing indicates
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-23
Page 54
Operation
Instrument Controls Menu
that these are proposed changes as opposed to implemented changes. To change the selected record format and erase record log file data, see “Commit Content” below.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Select Content > Select Field > Analog Inputs.
Use and to move the cursor up and down.
Press to select a new choice.
ANALOG INPUTS: >NONE ANALOG IN 1 ANALOG IN 2 ANALOG IN 3 ANALOG IN 4 ANALOG IN 5 ANALOG IN 6
ª
RANGE AVG DIAGS ALARM
Commit Content The Commit Content screen is used to save the selected output signal that is
tied to the selected field item. If no changes have been made “NO CHANGES TO RECORD LIST!” will appear. For more information about selecting the analog output signal group choices, see “Select Content” above.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Commit Content.
•Press to commit changes to selected record format and erase record log file data.
•Press to confirm changes.
3-24 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 55
Operation
Instrument Controls Menu
CHANGE LREC DATA AND CHANGE LREC FORMAT AND ERASE LREC LOG FILE DATA? ERASE LREC LOG FILE DATA?
COMMIT
ARE YOU SURE YOU WANT TO?x PRESS¨ TO CONFIRM CHANGES
COMMIT
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Reset to Default Content The Reset to Default Content screen is used to reset all of the datalogging
field items to default values. For more information about selecting the analog output signal group choices, see “Select Content” above.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Reset to Default Content.
•Press to reset selected record format and erase record log file data.
•Press to confirm reset.
RESET LREC FORMAT AND RESET LREC FORMAT AND ERASE LREC LOG FILE DATA? ERASE LREC LOG FILE DATA?
RESET
ARE YOU SURE YOU WANT TO?x PRESS ¨ TO CONFIRM RESETXX
RESET
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Configure Datalogging The Configure Datalogging menu deals with datalogging configuration for
the currently selected record type.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Configure Datalogging.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-25
Page 56
Operation
Instrument Controls Menu
Logging Period Min The Logging Period Min screen is used to select the logging period in
DATALOGGING SETTINGS: >LOGGING PERIOD MIN OFF MEMORY ALLOCATION % 50 DATA TREATMENT AVG
RANGE AVG DIAGS ALARM
minutes for the record format (srec or lrec). List of choices include: off, 1, 5, 15, 30, and 60 minutes (default).
In the Main Menu, choose Instrument Controls > Datalogging Settings > Configure Datalogging > Logging Period Min.
Use and to scroll through a list of choices.
•Press to set the logging period.
SET PERIOD FOR SREC: CURRENTLY: OFF MIN SET TO: 60 MIN ?
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM

SAVE
Memory Allocation Percent The Memory Allocation Percent screen is used to select the percentage of
each record type for both short records and long records. Percentages between 0 and 100% are available in increments of 10. Changing this value results in log erasure for both short records and long records.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Configure Datalogging > Memory Allocation %.
Use and to scroll through a list of choices.
•Press to set the percentage for both record types and proceed to the erasure warning screen.
3-26 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 57
Operation
Instrument Controls Menu
•Press to confirm change.
SET PERCENT SRECS: SET PERCENT SRECS: CURRENTLY: 50 % CURRENTLY: 50 % SET TO: 60 % ? SET TO: 60 % ?
THIS WILL ERASE ALL LOGS!x PRESS ¨ TO CONFIRM CHANGE
ª CHANGE VALUE
RANGE AVG DIAGS ALARM RANGE AVG DIAGS ALARM
SAVE
Data Treatment The Data Treatment screen is used to select the data type for the selected
record: whether the data should be averaged over the interval, the minimum or maximum used, or the current value logged. Data treatment doesn’t apply to all data, just to the concentration measurement. All other data points log the current value at the end of the interval.
In the Main Menu, choose Instrument Controls > Datalogging Settings > Configure Datalogging > Data Treatment.
Use and to scroll through a list of choices.
•Press to save the data type.
SET LREC DATA TYPE: CURRENTLY: AVG SET TO: CUR ?
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM
SAVE
Communication Settings The Communication Settings menu is used with communications control
and configuration.
In the Main Menu, choose Instrument Controls > Communication Settings.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-27
Page 58
Operation
Instrument Controls Menu
COMMUNICATION SETTINGS: >BAUD RATE INSTRUMENT ID COMMUNICATION PROTOCOL STREAMING DATA CONFIG RS-232/RS-485 SELECTION TCP/IP SETTINGS
RANGE AVG DIAGS ALARM
Baud Rate The Baud Rate screen is used to set the RS-232/RS-485 interface baud rate.
Baud rates of 1200, 2400, 4800, and 9600, 19200, 38400, 57600, and 115200 are available.
In the Main Menu, choose Instrument Controls > Communication Settings > Baud Rate.
Use and to scroll through a list of choices.
•Press to save the new baud rate.
BAUD RATE: CURRENTLY: 9600 SET TO: 19200 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Instrument ID The Instrument ID screen allows the operator to edit the instrument ID.
The ID is used to identify the calibrator when using C-Link or MODBUS protocols to control the calibrator or collect data. It may be necessary to edit the ID number if two or more of the same calibrator are connected to one computer. Valid Instrument ID numbers are from 0 to 127. The Model 49i Primary Standard has a default Instrument ID of 49. For more information about the Instrument ID, see Appendix B “C-Link Protocol Commands” or Appendix C “MODBUS Protocol”.
In the Main Menu, choose Instrument Controls > Communication Settings > Instrument ID.
3-28 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 59
Operation
Instrument Controls Menu
Use and to increment or decrement the ID value.
•Press to save the new instrument ID.
INSTRUMENT ID: CURRENTLY: 49 SET TO: 50 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Communication Protocol The Communication Protocol screen is used to change the communication
protocol for serial communications.
In the Main Menu, choose Instrument Controls > Communication Settings > Communication Protocol.
Use and to scroll through a list of choices.
•Press to save the new protocol.
COMMUNICATION PROTOCOL: CURRENTLY: CLINK SET TO: STREAMING ?
RANGE AVG DIAGS ALARM
©ª
CHANGE VALUE

SAVE VALUE
Streaming Data Configuration The Streaming Data Configuration menu is used to allow for configuration
of the 8 streaming data output items, streaming interval, current data format, and current timestamp setting. The Choose Item Signal submenu displays a list of the analog output signal group choices to choose from. Choices are Concentrations, Other Measurements, and Analog Inputs (if the I/O expansion board option is installed).
In the Main Menu, choose Instrument Controls > Communication Settings > Streaming Data Config.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-29
Page 60
Operation
Instrument Controls Menu
STREAMING DATA CONFIG: >INTERVAL 10 SEC ADD LABELS NO PREPEND TIMESTAMP YES ITEM 1 O3 ITEM 2 CELLAI ITEM 3 CELLBI ITEM 4 NONE
RANGE AVG DIAGS ALARM
CHOOSE STREAM DATA: >CONCENTRATIONS OTHER MEASUREMENTS ANALOG INPUTS
ª
RANGE AVG DIAGS ALARM
Streaming Data Interval The Streaming Data Interval screen is used to adjust the streaming data
interval. The following interval times are available: 1, 2, 5, 10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds.
In the Main Menu, choose Instrument Controls > Communication Settings > Streaming Data Config > Streaming Data Interval.
Use and to scroll through a list of choices.
Press to save the new streaming data period.
STREAMING DATA INTERVAL: CURRENTLY: 10 SEC SET TO: 20 SEC ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
3-30 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 61
Operation
Instrument Controls Menu
Concentrations The Concentrations screen allows the user to select the output signal that is
tied to the selected streaming data item. The selected item is shown by “<--” after it.
In the Main Menu, choose Instrument Controls > Communication Settings > Streaming Data Config > Select Item > Concentrations.
Use and to move the cursor up and down.
Press to select a new choice.
CONCENTRATIONS: >NONE O3 <--
RANGE AVG DIAGS ALARM
Other Measurements The Other Measurements screen allows the user to select the output signal
that is tied to the selected streaming data item. The selected item is shown by “<--” after it.
In the Main Menu, choose Instrument Controls > Communication Settings > Streaming Data Config > Select Item > Other Measurements.
Use and to move the cursor up and down.
Press to select a new choice.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-31
Page 62
Operation
Instrument Controls Menu
OTHER MEASUREMENTS: >NONE INTENSITY A INTENSITY B NOISE A NOISE B FLOW A FLOW B
RANGE AVG DIAGS ALARM
PRESSURE BENCH TEMP LAMP TEMP O3 LAMP TEMP
ª
Analog Inputs The Analog Inputs screen allows the user to select the analog input signal
(none or analog inputs 1-8) that is tied to the selected streaming data item. The selected item is shown by “<--” after it.
In the Main Menu, choose Instrument Controls > Communication Settings > Streaming Data Config > Select Item > Analog Inputs.
Use and to move the cursor up and down.
Press to select a new choice.
ANALOG INPUTS: >NONE ANALOG IN 1 ANALOG IN 2 ANALOG IN 3 ANALOG IN 4 ANALOG IN 5 ANALOG IN 6
RANGE AVG DIAGS ALARM
ª
RS-232/RS-485 Selection The RS-232/RS-485 Selection screen allows the user to choose between the
RS-232 or RS-485 specification for serial communication.
Note Disconnect the serial cable before changing RS-232 and RS-485 selection to prevent damage to the connected equipment.
In the Main Menu, choose Instrument Controls > Communication Settings > RS-232/RS-485 Selection.
3-32 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 63
Operation
Instrument Controls Menu
•Press to leave the warning screen and move to next screen.
Use to confirm and save the new selection change.
RS-232/RS-485 SELECTION: RS-232/RS-485 SELECTION:
XXXXXX** WARNING **XXXXXXX CURRENTLY: RS-232
DISCONNECT THE SERIALX X SET TO: RS-485 ? CABLES BEFORE CHANGING X MAKE SURE THAT THE CABLEXX THE SELECTION!XXXXXX
TO CONTINUE
IS OFF: PRESS
¨ TO CONFIRM
TOGGLE VALUE
RANGE AVG DIAGS ALARM
TCP/IP Settings The TCP/IP Settings menu is used for defining TCP/IP settings.
Note The instrument power must be cycled after this parameter has been
changed for the change to take effect.
In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings.
TCP/IP SETTINGS: >USE DHCP OFF IP ADDRESS 192.168.1.151 NETMASK 255.255.255.0 GATEWAY 192.168.1.1 HOST NAME iSeries
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Use DHCP The Use DHCP screen is used to specify whether or not to use DHCP.
When DHCP is enabled, the network dynamically provides an IP address for the instrument.
In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings > Use DCHP.
•Press to toggle and set DHCP on or off.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-33
Page 64
Operation
Instrument Controls Menu
DHCP: CURRENTLY: OFF SET TO: ON ?
CYCLE POWER TO CHANGE DHCP
RANGE AVG DIAGS ALARM
 TOGGLE VALUE
IP Address The IP Address screen is used to edit the IP address. The IP address can only
be changed when DHCP is OFF. For more information on DHCP, see “Use DHCP.”
In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings > IP Address.
Use , , and to move and change the value of the IP address.
•Press to save the new address.
IP ADDRESS: CURRENT: 192.168.1.151 SET TO: 192.168.1.151 1
§¨
MOVE CURSOR
©ª
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Netmask The Netmask screen is used to edit the netmask. The netmask is used to
determine the subnet the instrument uses to directly communicate with other devices. The netmask can only be changed when DHCP is OFF. For more information on DHCP, see “Use DHCP.”
In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings > Netmask.
Use , , and to move and change the value of the netmask.
3-34 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 65
Operation
Instrument Controls Menu
•Press to save the new netmask.
NETMASK: CURRENT: 255.255.255.0 SET TO: 255.255.255.000
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Defalt Gateway The Default Gateway screen is used to edit the gateway address. Any traffic
to addresses that are not on the local subnet will be routed through this address. The default gateway can only be changed when DHCP is OFF. For more information on DHCP, see “Use DHCP.”
In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings > Gateway.
Use , , and to move and change the value of the gateway address.
•Press to save the new address.
DEFAULT GATEWAY: CURRENT: 192.168.1.1 SET TO: 192.168.1.1 1
§¨
MOVE CURSOR
©ª
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Host Name The host name screen is used to edit the host name. When DHCP is
enabled, this name is reported to the DHCP server.
In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings > Host Name.
Use , , and to move the cursor or change between the edit field and the alpha page.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-35
Page 66
Operation
Instrument Controls Menu
I/O Configuration The I/O Configuration menu deals with configuration of the calibrator’s
•Press to save the new letter in the alpha table or save the new alpha page.
HOST NAME: CURRENT: ISERIES
ISERIES SSSSSS?
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE 0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
I/O system. The analog input configuration is displayed only if the I/O expansion board option is installed.
In the Main Menu, choose Instrument Controls > I/O Configuration.
I/O CONFIGURATION: >OUTPUT RELAY SETTINGS DIGITAL INPUT SETTINGS ANALOG OUTPUT CONFIG ANALOG INPUT CONFIG
RANGE AVG DIAGS ALARM
Output Relay Settings The Output Relay Settings menu displays a list of the 10 analog output
relays available, and allows the user to select the logic state or instrument parameter for the relay selected.
Note The digital outputs may take up to one second after the assigned state occurs to show up on the outputs.
In the Main Menu, choose Instrument Controls > I/O Configuration > Output Relay Settings.
3-36 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 67
Operation
Instrument Controls Menu
OUTPUT RELAY SETTINGS: >1 NOP CONC ALARM 2 NOP NONE 3 NOP UNITS 4 NOP GEN ALARM 5 NOP ZERO MODE 6 NOP OZ LEVEL 1 7 NOP OZ LEVEL 2
RANGE AVG DIAGS ALARM
ª
Logic State The Logic State screen is used to change the I/O relay to either normally
open or normally closed.
•Press to toggle and set the logic state open or closed.
OUTPUT RELAY SETUP: >LOGIC STATE OPEN INSTRUMENT STATE
RANGE AVG DIAGS ALARM
Instrument State The Instrument State submenu allows the user to select the instrument state
that is tied to the selected relay output. A submenu lists signal types of either alarm and non-alarm to choose from.
In the Main Menu, choose Instrument Controls > I/O Configuration > Output Relay Settings > Select Relay > Instrument State.
CHOOSE SIGNAL TYPE: >ALARMS NON-ALARM
RANGE AVG DIAGS ALARM
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-37
Page 68
Operation
Instrument Controls Menu
Alarms The Alarms status screen allows the user to select the alarm status for the
selected relay output. The selected item is shown by “<--” after it. The I/O board status alarm is present only if the I/O expansion board is installed. The zero and span check/calibration alarms are present only if the zero/span valve option is installed and the instrument is operating in manual mode.
In the Main Menu, choose Instrument Controls > I/O Configuration > Output Relay Settings > Select Relay > Instrument State > Alarms.
Use and to scroll through a list of choices.
Press to save the new selection for the relay.
ALARM STATUS ITEMS: >NONE GEN ALARM CONC MAX CONC MIN BENCH TEMP BNCH LMP TEMP OZ LAMP TEMP
ª
RANGE AVG DIAGS ALARM
PRESSURE FLOW A FLOW B INTENSITY A INTENSITY B CONC ALARM <-­ MB STATUS IB STATUS I/O BD STATUS
Non-Alarm The Non-Alarm status screen allows the user to select the non-alarm status
for the selected relay output. The selected item is shown by “<--” after it.
In the Main Menu, choose Instrument Controls > I/O Configuration > Output Relay Settings > Select Relay > Instrument State > Non-Alarm.
Use and to scroll through a list of choices.
Press to save the new selection for the relay.
3-38 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 69
Operation
Instrument Controls Menu
NON ALARM STATUS ITEMS: >NONE SERVICE UNITS ZERO MODE SPAN MODE SAMPLE MODE
OZ LEVEL 1 ª
RANGE AVG DIAGS ALARM
OZ LEVEL 2 OZ LEVEL 3 OZ LEVEL 4 OZ LEVEL 5 PURGE MODE
Digital Input Settings The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument parameter for the relay selected.
Note The digital inputs must be asserted for at least one second for the action to be activated.
In the Main Menu, choose Instrument Controls > I/O Configuration > Digital Input Settings.
Use and to move the cursor up and down.
Press to select a choice.
DIGITAL INPUT SETTINGS: >1 NOP SET BACKGROUND 2 NOP OZONATOR LVL 1 3 NOP OZONATOR LVL 2 4 NOP AOUTS TO ZERO 5 NOP AOUTS TO FS 6 NOP NONE 7 NOP NONE
RANGE AVG DIAGS ALARM
ª
Logic State The Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a relay connected between the digital input pin and ground is normally open
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-39
Page 70
Operation
Instrument Controls Menu
Instrument Action The Instrument Action screen allows the user to choose the instrument
and closes to trigger the digital input action. If nothing is connected to the digital input pin, the state should be left at open to prevent the action from being triggered.
•Press to toggle and set the logic state open or closed.
DIGITAL INPUT SETUP: >LOGIC STATE OPEN INSTRUMENT ACTION
RANGE AVG DIAGS ALARM
action that is tied to the selected digital input.
In the Main Menu, choose Instrument Controls > I/O Configuration > Digital Input Settings > Select Relay > Instrument Action.
Use and to scroll through a list of choices.
Press to save the new selection for the relay.
CHOOSE ACTION: >NONE ZERO MODE SPAN MODE OZONATOR LVL 1 OZONATOR LVL 2 OZONATOR LVL 3 OZONATOR LVL 4
RANGE AVG DIAGS ALARM
OZONATOR LVL 5
SET BACKGROUND <-­ AOUTS TO ZERO AOUTS TO FS
ª
Analog Output Configuration The Analog Output Configuration menu displays a list of the analog output
channels available for configuration. Channel choices include all voltage channels, all current channels, voltage channels 1-6, and current channels
3-40 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 71
Operation
Instrument Controls Menu
1-6 (if the I/O expansion board option is installed). Configuration choices include selecting range, setting minimum/maximum values, and choosing signal to output.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Output Config.
Use and to scroll through a list of choices.
Press to select a choice.
OUTPUT CHANNELS: >ALL VOLTAGE CHANNELS ALL CURRENT CHANNELS VOLTAGE CHANNEL 1 VOLTAGE CHANNEL 2 VOLTAGE CHANNEL 3 VOLTAGE CHANNEL 4 VOLTAGE CHANNEL 5
ª
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CONFIG: >SELECT RANGE SET MINIMUM VALUE SET MAXIMUM VALUE CHOOSE SIGNAL TO OUTPUT
RANGE AVG DIAGS ALARM
Select Output Range The Select Output Range screen is used to select the hardware range for the
selected analog output channel. Possible ranges for the voltage outputs are: 0-100 mV, 0-1, 0-5, 0-10 V.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Output Config > Select Channel > Select Range.
Use and to scroll through a list of choices.
Press to save the new range.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-41
Page 72
Operation
Instrument Controls Menu
SELECT OUTPUT RANGE: SELECTED OUTPUT: V ALL CURRENTLY: 0-10V SET TO: 0-5V ?
©ª CHANGE VALUE  SAVE
RANGE AVG DIAGS ALARM
Minimum and Maximum Value The MinimumValue screen is used to edit the zero (0) to full-scale (100)
value in percentages for the selected analog output channel. See Table 3-6 for a list of choices. The minimum and maximum output value screens function the same way. The example below shows the set minimum value screen.
In the Main Menu, choose Instrument Controls > IO Configuration > Analog Output Config > Select Channel > Set Minimum or Maximum Val ue .
Use and to increment or decrement the numeric value.
Press to validate and save the new minimum value.
MINIMUM OUTPUT PERCENT: SELECTED OUTPUT: V ALL CURRENTLY: N/A % SET TO: 0000.5 % ?
©ª CHANGE VALUE  SAVE
RANGE AVG DIAGS ALARM
Table 3-4. Analog Output Zero to Full-Scale Table
Output Zero % Value Full-Scale 100% Value
O
3
Range Status Recommend not to change the setting for this output
Zero (0) Range Setting
Intensity A User-set alarm min value User-set alarm max value
Intensity B User-set alarm min value User-set alarm max value
3-42 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 73
Operation
Instrument Controls Menu
Table 3-4. Analog Output Zero to Full-Scale Table
Output Zero % Value Full-Scale 100% Value
Noise A Zero (0) Range Setting
Noise B Zero (0) Range Setting
Flow A User-set alarm min value User-set alarm max value
Flow B User-set alarm min value User-set alarm max value
Pressure User-set alarm min value User-set alarm max value
Bench Temp User-set alarm min value User-set alarm max value
Lamp Temp User-set alarm min value User-set alarm max value
O
Lamp Temp User-set alarm min value User-set alarm max value
3
Choose Signal To Output The Choose Signal Type To Output screen displays a submenu list of the
analog output signal group choices. Group choices are Concentrations, Other Measurements, and Analog Inputs (if the I/O expansion board option is installed). This allows the user to select the output signal to the selected output channel. The Concentrations screen is shown below. See Table 3-7 below for a list of items for each signal group choice.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Output Config > Select Channel > Choose Signal To Output.
Use and to move the cursor up and down.
Press to select a choice.
CHOOSE ITEM SIGNAL: CONCENTRATIONS OTHER MEASUREMENTS ANALOG INPUTS
RANGE AVG DIAGS ALARM
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-43
Page 74
Operation
Instrument Controls Menu
CHOOSE SIGNAL - CONC SELECTED OUTPUT: V1 CURRENTLY: O3 SET TO: NONE ?
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM
SAVE
Table 3-5. Signal Type Group Choices
Concentrations Other Measurements Analog Inputs
None None None
O
3
Intensity A Analog Input 1
Intensity B Analog Input 2
Noise A Analog Input 3
Noise B Analog Input 4
Flow A Analog Input 5
Flow B Analog Input 6
Pressure Analog Input 7
Bench Temp Analog Input 8
Lamp Temp
O
Lamp Temp
3
Analog Input Configuration The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O expansion board option is installed. Configuration includes entering descriptor, units, decimal places, choice of 1-10 points in the table, and corresponding number of points selected.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config.
3-44 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 75
ANALOG INPUT CONFIG: >CHANNEL 1 IN1 CHANNEL 2 IN2 CHANNEL 3 IN3 CHANNEL 4 IN4 CHANNEL 5 IN5 CHANNEL 6 IN6 CHANNEL 7 IN7
RANGE AVG DIAGS ALARM
ANALOG INPUT 01 CONFIG: >DESCRIPTOR IN1 UNITS V DECIMAL PLACES 2 TABLE POINTS 2 POINT 1 POINT 2
Operation
Instrument Controls Menu
ª
RANGE AVG DIAGS ALARM
Descriptor The Descriptor screen allows the user to enter the descriptor for the selected
analog input channel. The descriptor is used in datalogging and streaming data to report what data is being sent out. The descriptor may be from 1 to 3 characters in length, and defaults to IN1 to IN8 (user input channel number).
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Descriptor.
Press to save the new descriptor.
ANALOG INPUT DESCRIPTOR: CURRENTLY: IN1
IN1
ABCDEFGHIJKLMN BKSP OPQRSTUVWXYZ PAGE 0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-45
Page 76
Operation
Instrument Controls Menu
Units The Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in datalogging and streaming data. The units may be from 1 to 3 characters in length, and defaults to V (volts).
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Units.
Press to save the new value.
ANALOG INPUT UNITS: CURRENTLY: V
VSS
ABCDEFGHIJKLMN BKSP OPQRSTUVWXYZ PAGE 0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Decimal Places The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal, from 0 to 6, with a default of 2.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Decimal Places.
Use and to increment or decrement the value.
Press to save the new value.
DECIMAL PLACES: CURRENTLY: 2 SET TO: 3 ?
©ª
INC/DEC SAVE VALUE
RANGE AVG DIAGS ALARM
Number of Table Points The Number of Table Points screen allows the user to select how many
points are used in the conversion table. The points range from 2 to 10, with a default of 2.
3-46 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 77
Operation
Instrument Controls Menu
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Table Points.
Use and to move the cursor up and down.
Press to save the new value.
NUMBER OF TABLE POINTS: CURRENTLY: 2 SET TO: 10 ?
©ª
INC/DEC SAVE VALUE
RANGE AVG DIAGS ALARM
Table Point The Table Point submenu allows the user to set up an individual table
point.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Point 1-10.
TABLE POINT 01 CONFIG: >VOLTS 0.00 USER VALUE 0.00
RANGE AVG DIAGS ALARM
Volts The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Select Point > Vo lt s.
•Use and to move the cursor left or right.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-47
Page 78
Operation
Instrument Controls Menu
Use and to move the cursor up and down.
Press to save the new value.
TABLE POINT 01 VOLTS: CURRENTLY: 0.00 SET TO: 00.00
§¨
MOVE CURSOR
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM
SAVE
User Value The User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion table, from -9999999 to 99999999. The default table is a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U.
In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Table Point > User Value.
•Use and to move the cursor left or right.
Use and to move the cursor up and down.
Press to save the new value.
TABLE POINT 01 USER VAL: CURRENTLY: 0.00 SET TO: 00000.00
§¨
MOVE CURSOR
©ª
CHANGE VALUE
RANGE AVG DIAGS ALARM
SAVE
Screen Contrast The Screen Contrast screen is used to change the contrast of the display.
Intensities between 0 and 100% in increments of 10 are available. Changing the screen contrast may be necessary if the calibrator is operated at extreme temperatures.
3-48 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 79
Operation
Instrument Controls Menu
In the Main Menu, choose Instrument Controls > Screen Contrast.
Use and to increment or decrement the screen contrast.
•Press to accept a change.
SCREEN CONTRAST: CURRENTLY: 50 % SET TO: 60 % ?
RANGE AVG DIAGS ALARM
©ª CHANGE VALUE
 SAVE VALUE
Service Mode The Service Mode screen is used to turn the service mode on or off. The
service mode locks out any remote actions and includes parameters and functions that are useful when making adjustments or diagnosing the Model 49i Primary Standard. For more information about the service mode, see “Service Menu” later in this chapter.
Note The service mode should be turned off when finished, as it prevents remote operation.
In the Main Menu, choose Instrument Controls > Service Mode.
•Press to toggle and set the service mode on or off.
SERVICE MODE: CURRENTLY: OFF SET TO: ON ?
RANGE AVG DIAGS ALARM
 TOGGLE VALUE
Date/Time The Date/Time screen allows the user to view and change the system date
and time (24-hour format). The internal clock is powered by its own battery when instrument power is off.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-49
Page 80
Operation
Diagnostics Menu
In the Main Menu, choose Instrument Controls > Date/Time.
Use , and to move and change the value of the date and time.
•Press to edit and save the new date and time.
DATE AND TIME: 19 MAR 2005 12:34:56 DATE AND TIME: PRESS
TO EDIT
19 MAR 2005 12:34:56 ?
SETTING: DAYS
¨ SET MONTHS
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Diagnostics Menu The Diagnostics menu provides access to diagnostic information and
functions. This menu is useful when troubleshooting the instrument. The analog input readings and analog input voltages are only displayed if the I/O expansion board option is installed.
In the Main Menu, choose Diagnostics.
DIAGNOSTICS: >PROGRAM VERSION VOLTAGES TEMPERATURES PRESSURE FLOWS CELL A/B O3 INTENSITIES
RANGE AVG DIAGS ALARM
ª
ANALOG INPUT READINGS ANALOG INPUT VOLTAGES DIGITAL INPUTS RELAY STATES TEST ANALOG OUTPUTS INSTRUMENT CONFIGURATION CONTACT INFORMATION
3-50 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 81
Operation
Diagnostics Menu
Program Version The Program Version screen (read only) shows the version number of the
program installed. Prior to contacting the factory with any questions regarding the instrument, please note the program version number.
In the Main Menu, choose Diagnostics > Program Version.
PROGRAM VERSION: PRODUCT: MODEL 49iPS VERSION: 01.04.02.137
FIRMWARE: 00.06.19
RANGE AVG DIAGS ALARM
Voltages The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating voltages without having to use a voltage meter. The I/O board is only displayed if the I/O expansion board option is installed.
In the Main Menu, choose Diagnostics > Vo lt ag e s.
VOLTAGES: >MOTHERBOARD INTERFACE BOARD I/O BOARD
RANGE AVG DIAGS ALARM
Motherboard Voltages The Motherboard screen (read only) is used to display the current voltage
readings on the motherboard.
In the Main Menu, choose Diagnostics > Voltages > Motherboard Vo lt a ge s.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-51
Page 82
Operation
Diagnostics Menu
Interface Board Voltages The Interface Board screen (read only) is used to display the current voltage
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
readings on the interface board.
In the Main Menu, choose Diagnostics > Voltages > Interface Board Vo lt a ge s.
INTERFACE BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
-15.0 SUPPLY -15.0 V
24.0 SUPPLY 24.0 V PHOTO LAMP 9.6 V O3 LAMP 17.2 V
RANGE AVG DIAGS ALARM
I/O Board Voltages The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This menu is only displayed if the I/O expansion board option is installed.
In the Main Menu, choose Diagnostics > Voltages > I/O Board Vo lt a ge s.
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
3-52 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 83
Temperatures The Temperatures screen (read only) displays the current bench
temperature, bench lamp temperature, and O3 lamp temperature (if the ozonator option is installed).
In the Main Menu, choose Diagnostics > Temperatures.
Operation
Diagnostics Menu
TEMPERATURES: >BENCH 32.3 BENCH LAMP 55.2 O3 LAMP 68.8
RANGE AVG DIAGS ALARM
o
C
o
C
o
C
Pressure The Pressure screen (read only) displays the current bench pressure. The
pressure is measured by a pressure transducer.
In the Main Menu, choose Diagnostics > Pressure.
PRESSURE: 753.4 mmHg
RANGE AVG DIAGS ALARM
Flows The Flows screen (read only) displays the current flow rate through Cell A
and Cell B. These flows are measured by internal flow sensors. For more information, see Chapter 1, “Operations”.
In the Main Menu, choose Diagnostics > Flows.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-53
Page 84
Operation
Diagnostics Menu
FLOWS: CELL A 0.608 LPM CELL B 0.612 LPM
RANGE AVG DIAGS ALARM
Cell A/B O
The Cell A/B O3 screen (read only) displays the current O3 concentration in
3
each cell, as well as the O3 reading displayed in the Run screen (the average of the two cells).
In the Main Menu, choose Diagnostics > Cell A/B O3.
O3 PPB: 600 CELL A 598 CELL B 602
RANGE AVG DIAGS ALARM
Intensities The Intensities screen (read only) displays the current intensity in Cell A
and Cell B in Hertz. These intensities are read by detectors A and B, respectively.
In the Main Menu, choose Diagnostics > Intensities.
INTENSITIES: CELL A 100425 Hz CELL B 100465 Hz
RANGE AVG DIAGS ALARM
3-54 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 85
Operation
Diagnostics Menu
Analog Input Readings The Analog Input Readings screen (read only) displays the 8 current
user-scaled analog readings (if the I/O expansion board option is installed).
In the Main Menu, choose Diagnostics > Analog Input Readings.
ANALOG INPUT READINGS: >CO 10.2 PPM SO2 18.2 PPB CO2 250 PPM FL1 20.42 LPM WND 9.86 V FL2 1.865 LPM IO7 0.0 V
RANGE AVG DIAGS ALARM
ª
Analog Input Voltages The Analog Input Voltages screen (read only) displays the 8 raw analog
voltage readings (if the I/O expansion board option is installed).
In the Main Menu, choose Diagnostics > Analog Input Voltages.
ANALOG INPUT VOLTAGES: >ANALOG IN 1 6.24 V ANALOG IN 2 4.28 V ANALOG IN 3 0.00 V ANALOG IN 4 0.00 V ANALOG IN 5 0.00 V ANALOG IN 6 0.00 V ANALOG IN 7 0.00 V
RANGE AVG DIAGS ALARM
ª
Digital Inputs The Digital Inputs screen (read only) displays the state of the 16 digital
inputs.
In the Main Menu, choose Diagnostics > Digital Inputs.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-55
Page 86
Operation
Diagnostics Menu
DIGITAL INPUTS: >INPUT 1 1 INPUT 2 1 INPUT 3 1 INPUT 4 1 INPUT 5 1 INPUT 6 1 INPUT 7 1
RANGE AVG DIAGS ALARM
ª
Relay States The Relay States screen displays the state of the 10 digital outputs and
allows toggling of the state to either on (1) or off (0). The relays are restored to their original states upon exiting this screen.
In the Main Menu, choose Diagnostics > Relay States.
RELAY STATE: >OUTPUT 1 0 OUTPUT 2 0 OUTPUT 3 0 OUTPUT 4 1 OUTPUT 5 0 OUTPUT 6 0 OUTPUT 7 0
RANGE AVG DIAGS ALARM
ª
Test Analog Outputs The Test Analog Outputs menu contains a number of digital to analog
converter (DAC) calibration items. Channel choices include all analog outputs, 6 voltage channels, and 6 current channels (if the I/O expansion board option is installed).
In the Main Menu, choose Diagnostics > Test Analog Outputs.
TEST ANALOG OUTPUTS: >ALL VOLTAGE CHANNEL 1 VOLTAGE CHANNEL 2 VOLTAGE CHANNEL 3 VOLTAGE CHANNEL 4 VOLTAGE CHANNEL 5 VOLTAGE CHANNEL 6
ª
RANGE AVG DIAGS ALARM
3-56 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 87
Operation
Diagnostics Menu
Set Analog Outputs The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-scale voltage, zero sets the analog outputs to 0 volts, and normal operation. The example below shows the selected output state “ALL” is set to normal.
In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL, Voltage Channel 1-6, or Current Channel 1-6.
•Press or to set output.
Press to reset to normal.
SET ANALOG OUTPUTS: SETTING: ALL OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
RANGE AVG DIAGS ALARM
Instrument Configuration The Instrument Configuration screen displays information on the hardware
configuration of the calibrator.
Note If the calibrator is in service mode, pressing on the item will toggle it yes or no (with the exception of purchased options such as dilution and auto calibration).
In the Main Menu, choose Diagnostics > Instrument Configuration.
•Press to toggle instrument configuration (in service mode only).
INSTRUMENT CONFIGURATION: >I/O EXPANSION BOARD YES
RANGE AVG DIAGS ALARM
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-57
Page 88
Operation
Alarms Menu
Contact Information The Contact Information screen displays the customer service information.
In the Main Menu, choose Diagnostics > Contact Information.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB: WWW.THERMO.COM
RANGE AVG DIAGS ALARM
Alarms Menu The alarms menu displays a list of items that are monitored by the
calibrator. If the item being monitored goes outside the lower or upper limit, the status of that item will go from “OK” to either “LOW” or “HIGH”, respectively. If the alarm is not a level alarm, the status will go from “OK” to “FAIL”. The number of alarms detected is displayed to indicate how many alarms have occurred. If no alarms are detected, the number zero is displayed.
To see the actual reading of an item and its minimum and maximum limits, move the cursor to the item and press .
Items displayed are determined by the options installed. The zero or span check is visible only if the zero/span check or auto calibration options are enabled. The O3 lamp temperature and ozonator level 1, 2, 3, 4, and 5 checks are visible only if the ozonator option is installed. The motherboard status, interface board status, and I/O expansion board status (if installed) indicates that the power supplies are working and connections are successful. There are no setting screens for these alarms.
In the Main Menu, choose Alarms.
3-58 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 89
ALARMS: ALARMS DETECTED 0 >O3 LAMP TEMP OK LAMP TEMP OK BENCH TEMP OK PRESSURE OK FLOW A OK FLOW B OK
RANGE AVG DIAGS ALARM
INTENSITY A OK INTENSITY B OK O3 CONCENTRATION OK MOTHERBOARD STATUS OK INTERFACE STATUS OK I/O EXP STATUS OK
Operation
Alarms Menu
ª
O3 Lamp Temperature The O
temperature and sets the minimum and maximum alarm limits. Acceptable alarm limits range from 65 to 75 °C. If the O3 lamp temperature reading goes beyond either the minimum or maximum alarm limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > O3 Lamp Temp.
O3 LAMP TEMPERATURE: ACTUAL 68.8 >MIN 60.0 MAX 80.0
RANGE AVG DIAGS ALARM
Min and Max O
Temperature Limits
Lamp
3
The Minimum O3 Lamp Temperature alarm limit screen is used to change the minimum O maximum O
Lamp Temperature screen displays the current ozonator lamp
3
o
C
o
C
o
C
Lamp temperature alarm limit. The minimum and
3
Lamp temperature screens function the same way.
3
In the Main Menu, choose Alarms > O3 Lamp Temp > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-59
Page 90
Operation
Alarms Menu
O3 LAMP TEMPERATURE: ACTUAL MIN: 60.0 SET MIN TO: 65.0
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
o
C
o
C ?
Lamp Temperature The Lamp Temperature screen displays the current lamp temperature and
sets the minimum and maximum alarm limits. Acceptable alarm limits range from 50 to 60 °C. If the lamp temperature reading goes beyond either the minimum or maximum limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Lamp Temp.
LAMP TEMPERATURE: ACTUAL 55.2 >MIN 50.0 MAX 60.0
o
C
o
C
o
C
Min and Max Lamp Temperature
Limits
RANGE AVG DIAGS ALARM
The Minimum Lamp Temperature alarm limit screen is used to change the minimum lamp temperature alarm limit. The minimum and maximum lamp temperature screens function the same way.
In the Main Menu, choose Alarms > Lamp Temp > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
3-60 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 91
Operation
Alarms Menu
LAMP TEMPERATURE: ACTUAL MIN: 50.0 SET MIN TO: 52.0
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
o
C
o
C ?
Bench Temperature The Bench Temperature screen displays the current bench temperature and
sets the minimum and maximum alarm limits. Acceptable alarm limits range from 5 to 50 °C. If the bench temperature reading goes beyond either the minimum or maximum limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Bench Temp.
BENCH TEMPERATURE: ACTUAL 32.3 >MIN 15.0 MAX 40.0
o
C
o
C
o
C
Min and Max Bench Temperature
Limits
RANGE AVG DIAGS ALARM
The Minimum Bench Temperature alarm limit screen is used to change the minimum bench temperature alarm limit. The minimum and maximum bench temperature screens function the same way.
In the Main Menu, choose Alarms > Bench Temp > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-61
Page 92
Operation
Alarms Menu
BENCH TEMPERATURE: ACTUAL MIN: 15.0 SET MIN TO: 16.0
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
o
C
o
C ?
Pressure The Pressure screen displays the current pressure and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 200 to 1,000 mmHg. If the pressure reading goes beyond either the minimum or maximum limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Pressure.
PRESSURE: ACTUAL 753.4 mmHg >MIN 200.0 mmHg MAX 1000.0 mmHg
RANGE AVG DIAGS ALARM
Min and Max Pressure Limits The Minimum Pressure alarm limit screen is used to change the minimum
pressure alarm limit. The minimum and maximum pressure screens function the same way.
In the Main Menu, choose Alarms > Pressure > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
3-62 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 93
Operation
Alarms Menu
PRESSURE: ACTUAL MIN: 200.0 mmHg SET MIN TO: 250.0 mmHg ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Flow A and B The Flow A screen displays the current sample flow reading in cell A and
sets the minimum and maximum alarm limits. Acceptable alarm limits range from 0.4 to 1.6 LPM. If the pressure reading goes beyond either the minimum or maximum limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu. The Flow B screen functions the same way.
In the Main Menu, choose Alarms > Flow A or Flow B.
FLOW A: ACTUAL 0.608 LPM >MIN 0.400 LPM MAX 1.400 LPM
RANGE AVG DIAGS ALARM
Min and Max Flow Limits The Minimum Flow A alarm limit screen is used to change the minimum
flow A alarm limit. The minimum and maximum flow A and flow B screens function the same way.
In the Main Menu, choose Alarms > Select Flow > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-63
Page 94
Operation
Alarms Menu
FLOW A: ACTUAL MIN: 0.400 LPM SET MIN TO: 0.500 LPM ?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Intensity A and B The Intensity A screen displays the current lamp intensity reading in Cell A
and sets the minimum and maximum alarm limits. Acceptable alarm limits range from 45,000 to 150,000 Hertz. If the intensity A reading goes beyond either the minimum or maximum limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Intensity A or Intensity B.
INTENSITY A: ACTUAL 98425 Hz >MIN 45000 Hz MAX 150000 Hz
RANGE AVG DIAGS ALARM
Min and Max Intensity Limits The Minimum Intensity A alarm limit screen is used to change the
minimum intensity A alarm limit. The minimum and maximum intensity A and intensity B screens function the same way.
In the Main Menu, choose Alarms > Select Intensity > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
3-64 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 95
INTENSITY A:
ACTUAL MIN: 45000 HZ SET MIN TO: 55000 HZ ?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Operation
Alarms Menu
O3 Concentration The O
sets the minimum and maximum alarm limits. Acceptable alarm limits range from 0 to 5,000 ppb. The minimum alarm may be programmed as a floor trigger (alarm is triggered when the concentration falls below the minimum value) or a ceiling trigger (alarm is triggered when the concentration goes above the minimum value). If the O3 concentration goes beyond either the minimum or maximum limit, an alarm is activated. The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > O3 Concentration.
O3 CONCENTRATION: ACTUAL 600 >MIN 0.0 MAX 5000 MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
Concentration screen displays the current O3 concentration and
3
Min and Max O
Concentration
3
Limits
The Minimum O3 Concentration alarm limit screen is used to change the minimum O
concentration alarm limit. The minimum and maximum O3
3
concentration alarm limit screens function the same way.
In the Main Menu, choose Alarms > O3 Concentration > Min or Max.
•Use and to move the cursor left or right.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-65
Page 96
Operation
Service Menu
O3 CONCENTRATION: ACTUAL MIN: 0.0 SET MIN TO: 000020.0 ?
§¨ MOVE CURSOR
©ª INC/DEC
RANGE AVG DIAGS ALARM
SAVE VALUE
Min Trigger The Minimum Trigger screen allows the user to view and set the O3
concentration alarm trigger type to either floor or ceiling. The minimum alarm may be programmed as a floor trigger (alarm is triggered when the concentration falls below the minimum value) or a ceiling trigger (alarm is triggered when the concentration goes above the minimum value).
In the Main Menu, choose Alarms > O3 Concentration > MinTrigger.
•Press to toggle and set the minimum trigger to floor or ceiling.
MIN TRIG(CEILING/FLOOR): ACTUAL TRIGGER: CEILING SET TRIGGER TO: FLOOR ?
TOGGLE AND SAVE VALUE
RANGE AVG DIAGS ALARM
Service Menu The Service menu appears only when the calibrator is in the service mode.
To put the calibrator into service mode:
In the Main Menu, choose Instrument Controls > Service Mode.
Advanced diagnostic functions are included in the service mode. Meaningful data should not be collected when the calibrator is in the service mode.
In the Main Menu, choose Service.
3-66 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 97
Operation
Service Menu
SERVICE: >PRESSURE CHECK LAMP SETTING DETECTOR CALIBRATION INTENSITY CHECK O3 RESPONSE COEFFICIENT MANUAL O3 LAMP SETTING PRESSURE CALIBRATION
RANGE AVG DIAGS ALARM
FLOW A CALIBRATION FLOW B CALIBRATION TEMPERATURE CALIBRATION ANALOG OUT CALIBRATION ANALOG INPUT CALIBRATION DISPLAY PIXEL TEST RESTORE USER DEFAULTS
ª
Pressure Check The Pressure Check menu is used to manually control the flow of reference
or sample gas through Cell B. This enables the pressure reading of Cell B, with either sample or reference gas, to be determined. Pump pressure is used to test the pump. Selecting any of these menu items will disturb the analog outputs.
In the Main Menu, choose Service > Pressure Check.
PRESSURE CHECK: >SAMPLE PRESSURE REFERENCE PRESSURE PUMP PRESSURE
RANGE AVG DIAGS ALARM
Sample Pressure The Sample Pressure screen (read only) displays the pressure of the sample
gas in Cell B.
In the Main Menu, choose Service > Pressure Check > Sample Pressure.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-67
Page 98
Operation
Service Menu
SAMPLE PRESSURE:
753.4 mmHg
RANGE AVG DIAGS ALARM
Reference Pressure The Reference Pressure screen (read only) displays the pressure of the
reference gas in Cell B.
In the Main Menu, choose Service > Pressure Check > Reference Pressure.
REFERENCE PRESSURE:
753.4 mmHg
RANGE AVG DIAGS ALARM
Pump Pressure The Pump Pressure screen (read only) is used to test the pump. The
solenoids are energized such that there is no flow in Cell B. The flow rate for Cell B drops to zero and the pressure reading should drop below 390 mmHg in less than 20 seconds. This indicates the effectiveness of the internal pump. After 20 seconds, if the pressure is greater than 390 mmHg or the flow is greater than 0.010 LPM, the words “PUMP PROBLEM DETECTED” appears. After 20 seconds, if the pressure is less than 390 mmHg and the flow is less than 0.010 LPM, the words “PUMP OK” appears.
In the Main Menu, choose Service > Pressure Check > Pump Pressure.
3-68 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Page 99
Operation
Service Menu
PUMP PRESSURE: 753.4 mmHg
0.010 LPM
TESTING, WAIT 20 SECS
RANGE AVG DIAGS ALARM
Lamp Setting The Lamp Setting screen is used to adjust the detector intensities. The
display shows the intensities of Cell A and Cell B. The next line of the display shows the current lamp setting. Adjust the lamp setting until the intensities are about 100 kHz.
Note This adjustment should only be performed by an instrument service technician.
In the Main Menu, choose Service > Lamp Setting.
•Use and to increment or decrement the numeric value.
•Press to save the new lamp setting.
BENCH LAMP SETTING: CELL A INT: 98425 Hz CELL B INT: 97465 Hz LAMP SETTING: 72.9 %
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Detector Calibration The Detector Calibration screen is used to set the calibration factors for
detector A and detector B. Calibration factors are computed so that both detectors read 100,000 Hz on zero air. If either detector reads lower than 75,000 Hz or higher than 125,000 Hz, this calibration will not be successful. It is crucial that the lamp intensity be set such that both detector intensities fall inside this range before performing this calibration.
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-69
Page 100
Operation
Service Menu
Note This adjustment should only be performed by an instrument service technician.
In the Main Menu, choose Service > Detector Calibration.
DETECTOR CALIBRAITON: CELL A RAW INT: 76988 HZ CELL B RAW INT: 67079 HZ CELL A FACTOR: 1.000 CELL B FACTOR: 1.000
CALC/SAVE NEW RECORDS
RANGE AVG DIAGS ALARM
Intensity Check The Intensity Check menu is used to manually control the flow of reference
or sample gas through either Cell A or Cell B. This enables the intensity and noise reading of each detector to be determined with either reference or sample gas flow. Selecting any of these menu items will disturb the analog outputs.
In the Main Menu, choose Service > Intensity Check.
INTENSITY CHECK: >INT A REFERENCE GAS INT A SAMPLE GAS INT B REFERENCE GAS INT B SAMPLE GAS
RANGE AVG DIAGS ALARM
Intensity A/B Check The Intensity A Reference Gas screen (read only) switches the solenoid
valves so that reference gas is flowing through Cell A. The intensity and noise reading are displayed. The Intensity B Reference Gas screen functions the same way. Also, the Intensity A and Intensity B Sample Gas screens function the same way (only with sample gas instead of reference gas).
In the Main Menu, choose Service > Intensity Check > Int A or B Ref or Sample.
3-70 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Loading...