“Analyze. Detect. Measure. Control” is a trademark of Thermo Electron Corporation. All other product names and
logos are the property of their respective owner.
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult
your local sales representative for details.
Page 3
WEEE Compliance
This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is
marked with the following symbol:
Thermo Electron has contracted with one or more recycling/disposal
companies in each EU Member State, and this product should be disposed
of or recycled through them. Further information on Thermo Electron’s
compliance with these Directives, the recyclers in your country, and
information on Thermo Electron products which may assist the detection of
substances subject to the RoHS Directive are available at:
www.thermo.com/WEEERoHS.
Read Coils for 42i Low Source .................................................................C-8
Read Registers for 42i Low Source ...........................................................C-9
Write Coils for 42i Low Source ..............................................................C-10
Operating Status for Model 42i Low Source ...........................................D-7
Error Status for Model 42i Low Source ................................................... D-7
Thermo Electron CorporationModel 42i Low Source Instruction Manualxiii
Page 14
Tables
xivModel 42i Low Source Instruction ManualThermo Electron Corporation
Page 15
About This Manual
This manual provides information about operating, maintaining, and
servicing the analyzer. It also contains important alerts to ensure safe
operation and prevent equipment damage. The manual is organized into the
following chapters and appendices to provide direct access to specific
operation and service information:
•Chapter 1 “Introduction” provides an overview of product features,
describes the principles of operation, and lists the specifications.
•Chapter 2 “Installation” describes how to unpack, setup, and startup the
analyzer.
•Chapter 3 “Operation” describes the front panel display, the front panel
pushbuttons, and the menu-driven software.
•Chapter 4 “Calibration” provides the procedures for calibrating the
analyzer and describes the required equipment.
•Chapter 5 “Preventive Maintenance” provides maintenance procedures
to ensure reliable and consistent instrument operation.
•Chapter 6 “Troubleshooting” presents guidelines for diagnosing analyzer
failures, isolating faults, and includes recommended actions for restoring
proper operation.
•Chapter 7 “Servicing” presents safety alerts for technicians working on
the analyzer, step-by-step instructions for repairing and replacing
components, and a replacement parts list. It also includes contact
information for product support and technical information.
•Chapter 8 “System Description” describes the function and location of
the system components, provides an overview of the software structure,
and includes a description of the system electronics and input/output
connections.
•Chapter 9 “Optional Equipment” describes the optional equipment that
can be used with this analyzer.
•Appendix A “Warranty” provides a copy of the warranty statement.
Thermo Electron CorporationModel 42i Low Source Instruction Manualxv
Page 16
Preface
Safety
•Appendix B “C-Link Protocol Commands” provides a description of the
C-Link protocol commands that can be used to remotely control an
analyzer using a host device such as a PC or a datalogger.
•Appendix C “MODBUS Protocol” provides a description of the
MODBUS Protocol Interface and is supported both over RS-232/485
(RTU protocol) as well as TCP/IP over Ethernet.
SafetyReview the following safety information carefully before using the analyzer.
This manual provides specific information on how to operate the analyzer,
however if the analyzer is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
Safety and Equipment
Damage Alerts
This manual contains important information to alert you to potential safety
hazards and risks of equipment damage. Refer to the following types of
alerts you may see and the actual alerts that appear in this manual.
Safety and Equipment Damage Alert Descriptions
AlertDescription
DANGERA hazard is present that will result in death or serious
personal injury if the warning is ignored.
WARNINGA hazard is present or an unsafe practice can result in
serious personal injury if the warning is ignored.
CAUTIONThe hazard or unsafe practice could result in minor to
moderate personal injury if the warning is ignored.
Equipment DamageThe hazard or unsafe practice could result in property
damage if the warning is ignored.
xviModel 42i Low Source Instruction ManualThermo Electron Corporation
Page 17
Safety and Equipment Damage Alerts in this Manual
AlertDescription
WARNINGIf the equipment is operated in a manner not specified by
the manufacturer, the protection provided by the
equipment may be impaired.
The service procedures in this manual are restricted to
qualified service personnel only.
The Model 42i Low Source is supplied with a three-wire
grounding cord. Under no circumstances should this
grounding system be defeated.
CAUTIONIf the LCD panel breaks, do not to let the liquid crystal
contact your skin or clothes. If the liquid crystal contacts
your skin or clothes, wash it off immediately using soap
and water.
Avoid contact with converter heated components. Allow
converter to cool to room temperature before handling
converter components.
Preface
Safety
Equipment DamageDo not attempt to lift the analyzer by the cover or other
external fittings.
This adjustment should only be performed by an
instrument service technician.
Some internal components can be damaged by small
amounts of static electricity. A properly grounded
antistatic wrist strap must be worn while handling any
internal component.
Handle all printed circuit boards by the edges only.
Do not point the photomultiplier tube at a light source.
This can permanently damage the tube.
Do not remove the panel or frame from the LCD module.
The LCD module polarizing plate is very fragile, handle it
carefully.
Do not wipe the LCD module polarizing plate with a dry
cloth, it may easily scratch the plate.
Do not use alcohol, acetone, MEK or other Ketone based
or aromatic solvents to clean the module, use a soft
cloth moistened with a naphtha cleaning solvent.
Do not place the LCD module near organic solvents or
corrosive gases.
Do not shake or jolt the LCD module.
Thermo Electron CorporationModel 42i Low Source Instruction Manualxvii
Page 18
Preface
Where to Get Help
WEEE SymbolThe following symbol and description identify the WEEE marking used on
the instrument and in the associated documentation.
SymbolDescription
Marking of electrical and electronic equipment which applies to
electrical and electronic equipment falling under the Directive
2002/96/EC (WEEE) and the equipment that has been put on the market
after 13 August 2005.
Where to Get HelpService is available from exclusive distributors worldwide. Contact one of
the phone numbers below for product support and technical information.
866-282-0430 Toll Free
508-520-0430 International
Thermo Electron CorporationModel 42i Low Source Instruction Manualxviii
Page 19
Chapter 1 Introduction
The Model 42i Low Source Chemiluminescence NO-NO2-NOx Analyzer
combines proven detection technology, easy to use menu-driven software,
and advanced diagnostics to offer unsurpassed flexibility and reliability. The
Model 42i Low Source has the following features:
•320 x 240 graphics display
•Menu-driven software
•Field programmable ranges
•User-selectable single/dual/auto range modes
•Multiple user-defined analog outputs
•Analog input options
•High sensitivity
•Fast response time
•Linearity through all ranges
•Independent NO-NO
•Replaceable NO
2
-NOx ranges
2
converter cartridge
•User-selectable digital input/output capabilities
•Standard communications features include RS232/485 and Ethernet
•C-Link, MODBUS, and streaming data protocols
For details of the analyzer’s principle of operation and product
specifications, see the following topics:
•“Principle of Operation” on page 1-2 describes the principles by which
your analyzer operates.
Thermo Electron Corporation42i Low Source Instruction Manual1-1
Page 20
Introduction
Principle of Operation
•“Specifications” on page 1-3 is a list of the analyzer’s performance
specifications.
Thermo Electron is pleased to supply this chemiluminescence
NO-NO2-NOx analyzer. We are committed to the manufacture of
instruments exhibiting high standards of quality, performance, and
workmanship. Service personnel are available for assistance with any
questions or problems that may arise in the use of this analyzer. For more
information on servicing, see Chapter 7, “Servicing”.
Principle of OperationThe Model 42i Low Source operates on the principle that nitric oxide (NO)
and ozone (O3) react to produce a characteristic luminescence with an
intensity linearly proportional to the NO concentration. Infrared light
emission results when electronically excited NO2 molecules decay to lower
energy states. Specifically,
υ
hONOO NO
++→+
223
Nitrogen dioxide (NO2) must first be transformed into NO before it can be
measured using the chemiluminescent reaction. NO2 is converted to NO by
a stainless steel NO2-to-NO converter heated to about 625 °C (the optional
molybdenum converter is heated to 325 °C).
The ambient air sample is drawn into the Model 42i Low Source through
the sample bulkhead, as shown in Figure 1-1. The sample flows through a
capillary, and then to the mode solenoid valve. The solenoid valve routes the
sample either straight to the reaction chamber (NO mode) or through the
NO2-to-NO converter and then to the reaction chamber (NOx mode). A
flow sensor prior to the reaction chamber measures the sample flow.
Dry air enters the Model 42i Low Source through the dry air bulkhead,
passes through a flow switch, and then through a silent discharge ozonator.
The ozonator generates the ozone needed for the chemiluminescent
reaction. At the reaction chamber, the ozone reacts with the NO in the
sample to produce excited NO
molecules. A photomultiplier tube (PMT)
2
housed in a thermoelectric cooler detects the luminescence generated during
this reaction. From the reaction chamber, the exhaust travels through the
ozone (O3) converter to the pump, and is released through the vent.
The NO and NO
concentrations calculated in the NO and NOx modes
x
are stored in memory. The difference between the concentrations is used to
calculate the NO2 concentration. The Model 42i Low Source outputs NO,
1-242i Low Source Instruction ManualThermo Electron Corporation
Page 21
Introduction
Specifications
NO2, and NOx concentrations to the front panel display, the analog
outputs, and also makes the data available over the serial or ethernet
connection.
Figure 1-1. Model 42i Low Source Flow Schematic
SpecificationsTable 1-1. Model 42i Low Source Specifications
Custom ranges0-0.2 to 100 ppm (0-1 to 500 ppm in extended ranges)
0-0.5 to 150 mg/m
Zero noise0.005 ppm RMS (60 second averaging time)
Lower detectable limit0.01 ppm (60 second averaging time)
Zero drift (24 hour)≈ 0.005 ppm
Span drift (24 hour)±
Response time
(NO/NO
mode)
x
1% full-scale
15 sec (10 second averaging time)
85 sec (60 second averaging time)
305 sec (300 second averaging time)
3
(0-2 to 750 mg/m3 in extended ranges)
3
3
Thermo Electron Corporation42i Low Source Instruction Manual1-3
Page 22
Introduction
Specifications
Response time
(NO mode)
15 sec (10 second averaging time)
65 sec (60 second averaging time)
305 sec (300 second averaging time)
Linearity±
1% full-scale
Sample flow rate≈ 25 cc/min. measured at atmospheric pressure
Operating temperature
15–35 °C (may be safely operated over the range of 0–45 °C)
*
Power requirements100 VAC @ 50/60 Hz
115 VAC @ 50/60 Hz
220
–240 VAC @ 50/60 Hz
300 watts
Physical dimensions16.75” (W) X 8.62” (H) X 23” (D)
WeightApproximately 55 lbs.
Analog outputs6 voltage outputs; 0
–100 mV, 1 V, 5 V, 10 V (User selectable), 5% of
full-scale over/under range, 12 bit resolution, user selectable for
measurement input
Digital outputs1 power fail relay Form C, 10 digital relays Form A, user selectable
alarm output, relay logic, 100 mA @ 200 VDC
Digital inputs16 digital inputs, user select programmable, TTL level, pulled high
Serial Ports1 RS-232 or RS-485 with two connectors, baud rate 1200
–115200,
data bits, parity, and stop bits, protocols: C-Link, MODBUS, and
streaming data (all user selectable)
Ethernet connectionRJ45 connector for 10Mbs Ethernet connection, static or dynamic
TCP/IP addressing
*
In non condensing environments. Performance specifications based on operation in 15–35 °C range.
1-442i Low Source Instruction ManualThermo Electron Corporation
Page 23
Chapter 2 Installation
Installing the Model 42i Low Source includes the following
recommendations and procedures:
•“Lifting” on page 2-1
•“Unpacking and Inspection” on page 2-1
•“Setup Procedure” on page 2-3
•“Startup” on page 2-5
LiftingWhen lifting the instrument, use procedure appropriate to lifting a heavy
object, such as, bending at the knees while keeping your back straight and
upright. Grasp the instrument at the bottom in the front and at the rear of
the unit. Although one person can lift the unit, it is desirable to have two
persons lifting, one by grasping the bottom in the front and the other by
grasping the bottom in the rear.
Unpacking and
Inspection
Equipment Damage Do not attempt to lift the instrument by the cover or
other external fittings.
The Model 42i Low Source is shipped complete in one container. If there is
obvious damage to the shipping container when you receive the instrument,
notify the carrier immediately and hold for inspection. The carrier is
responsible for any damage incurred during shipment.
Use the following procedure to unpack and inspect the instrument.
1. Remove the instrument from the shipping container and set it on a table
or bench that allows easy access to both the front and rear.
2. Remove the cover to expose the internal components.
Thermo Electron CorporationModel 42i Low Source Instruction Manual2-1
Page 24
Installation
Unpacking and Inspection
3. Remove the packing material.
Remove Packing
(2 pieces)
Remove Packing
(2 pieces)
Units without Optional I/O Board
Units with Optional I/O Board
Figure 2-1. Remove the Packing Material
4. Remove the three shipping screws (Figure 2-2).
Shipping Screws (3)
Figure 2-2. Remove the Shipping Screws
5. Check for possible damage during shipment.
6. Check that all connectors and circuit boards are firmly attached.
2-2Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 25
7. Re-install the cover.
Setup ProcedureUse the following procedure to setup the instrument.
1. Connect the sample line to the SAMPLE bulkhead on the rear panel
(Figure 2-3). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of FEP
Teflon®, 316 stainless steel, borosilicate glass, or similar tubing with an
OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the
tubing should be less than 10 feet.
Note Gas must be delivered to the instrument at atmospheric pressure. It
may be necessary to use an atmospheric bypass plumbing arrangement as
shown in Figure 2-4 if gas pressure is greater than atmospheric pressure.
2. Connect the air dryer to the DRY AIR bulkhead.
Installation
Setup Procedure
3. Connect the EXHAUST bulkhead to a suitable vent or ozone scrubber.
The exhaust line should be 1/4-inch OD with a minimum ID of
1/8-inch. The length of the exhaust line should be less than 10 feet.
Verify that there is no restriction in this line.
4. Connect a suitable recording device to the rear panel connector. See the
“Operation” chapter for more information about the rear panel
pin-outs.
5. Plug the instrument into an outlet of the appropriate voltage and
frequency.
WARNING The Model 42i Low Source is supplied with a three-wire
grounding cord. Under no circumstances should this grounding system
be defeated.
Thermo Electron CorporationModel 42i Low Source Instruction Manual2-3
Page 26
Installation
Setup Procedure
Figure 2-3. Model 42i Low Source Rear Panel
Instrument Bulkhead
SAMPLE
Figure 2-4. Atmospheric Dump Bypass Plumbing
Vent to Exhaust Line
at Atmospheric Pressure
Input Gas Line
2-4Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 27
StartupUse the following procedure when starting the instrument.
1. Turn the power ON.
2. Allow 90 minutes for the instrument to stabilize.
Note It is best to turn the ozonator on and let the instrument run overnight
before calibration in order to obtain the most accurate information.
3. Set instrument parameters such as operating ranges and averaging times
to appropriate settings. For more information about instrument
parameters, see the “Operation” chapter.
4. Before beginning actual instrument monitoring, perform a multipoint
calibration as described in the “Calibration” chapter.
Installation
Startup
Thermo Electron CorporationModel 42i Low Source Instruction Manual2-5
Page 28
Installation
Startup
2-6Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 29
Chapter 3 Operation
This chapter describes the front panel display, front panel pushbuttons, and
menu-driven software.
•“Display” on page 3-2 describes the LCD graphics display.
•“Pushbuttons” on page 3-3 describes the various front panel
pushbuttons and the expected key actions for each.
•“Software Overview” on page 3-4 describes the menu-driven software
and submenus.
•“Range Menu” on page 3-8 describes the gas units, NO- NO2 -NOx
ranges, and custom ranges.
•“Averaging Time” on page 3-17 describes the averaging period applied
to NO, NO2, and NOx measurements.
•“Calibration Factors Menu” on page 3-18 describes the calibration
factors used to correct NO, NO2, and NOx measurement readings.
•“Calibration Menu” on page 3-23 describes calibration of zero and span.
•“Instrument Controls Menu” on page 3-30 describes the instrument
hardware control and configuration.
•“Diagnostics Menu” on page 3-65 describes the diagnostic information
and functions.
•“Alarms Menu” on page 3-73 describes a list of items that are monitored
by the analyzer.
•“Service Menu” on page 3-82 describes service related menu items.
•“Password” on page 3-97 describes how to enter/change a password,
lock and unlock the instrument.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-1
Page 30
Operation
Display
DisplayThe 320 x 240 graphics liquid-crystal display (LCD) shows the sample
concentrations, instrument parameters, instrument controls, help, and error
messages. Some menus contain more items than can be displayed at one
time. For these menus, use and to move the cursor up and
down to each item.
Figure 3-1. 42i Low Source Front Panel Display
CAUTION If the LCD panel breaks, do not let the liquid crystal contact your
skin or clothes. If the liquid crystal contacts your skin or clothes, wash it off
immediately using soap and water.
3-2Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 31
Pushbuttons
PushbuttonsThe Pushbuttons allow the user to traverse the various screens/menus.
Figure 3-2. Front Panel Pushbuttons
Operation
Table 3-1. Front Panel Pushbuttons
= Soft KeysThe soft keys are used to provide shortcuts that
allow the user to jump to user-selectable menu screens. For
more information on processing soft keys, see “Soft Keys”
below.
= RunThe is used to display the Run screen. The Run
screen normally displays the NO, NO
concentrations.
= MenuThe is used to display the Main Menu when in the
Run screen, or back up one level in the menu system. For
more information about the Main Menu, see “Main Menu”
later in this chapter.
= HelpThe is context-sensitive, that is, it provides
additional information about the screen that is being
displayed. Press for a brief explanation about the
current screen or menu. Help messages are displayed using
lower case letters to easily distinguish them from the
operating screens. To exit a help screen, press or
return to the Run screen.
, and NOx
2
to return to the previous screen, or to
= Up, Down
= Left, Right
= EnterThe is used to select a menu item,
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-3
The four arrow pushbuttons (, , ,
and ) move the cursor up, down, left, and right or
change values and states in specific screens.
accept/set/save a change, and/or toggle on/off functions.
Page 32
Operation
Software Overview
Soft KeysThe Soft Keys are multi-functional keys that use part of the display to
identify their function at any moment. The function of the soft keys allow
immediate access to the menu structure and most often used menus and
screens. They are located directly underneath the display and as the keys'
functions change this is indicated by user-defined labels in the lower part of
the display, so that the user knows what the keys are to be used for.
To change a soft key, place the menu cursor “>” on the item of the selected
menu or screen you wish to set. Press followed by the selected soft key
within 1 second of pressing the right-arrow key. The edit soft key prompt
will be displayed for configuration for the new label.
Note Not all menu items may be assigned to soft keys. If a particular menu
or screen item cannot be assigned, the key assignment screen will not come
up upon entering right-arrow-soft key combinations. All items under the
Service menu (including the menu itself) cannot be assigned soft keys.
EDIT SOFT KEY PROMPT:
CURRENTLY: RANGE
RANGE
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Software OverviewThe Model 42i Low Source utilizes the menu-driven software illustrated by
the flowchart in Figure 3-3. The Power-Up screen, shown at the top of the
flowchart, is displayed each time the instrument is turned on. This screen is
displayed while the instrument is warming up and performing self-checks.
After the warm-up period, the Run screen is automatically displayed. The
Run screen is the normal operating screen. It displays the NO, NO
NO
concentrations, depending on operating mode. From the Run screen,
x
the Main Menu can be displayed by pressing . The Main Menu
contains a list of submenus. Each submenu contains related instrument
parameters and/or functions. This chapter describes each submenu and
screen in detail. Refer to the appropriate sections for more information.
, and
2
3-4Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 33
2CUUYQTF
5GV2CUUYQTF
%JCPIG2CUUYQTF
7PNQEM+PUVTWOGPV
.QEM+PUVTWOGPV
4GOQXG2CUUYQTF
5GTXKEG
6GOR%CN
#PCNQI1WVRWV%CN
%QPX5GV6GOR
2/68QNVCIG#FL
4CPIG/QFG5GNGEV
2TGUUWTG%CN
#PCNQI+PRWV%CN
+PRWV$F%CN
1\QPCVQT5CHGV[
'ZVGPFGF4CPIGU
&KNWVKQP4CVKQ
&KURNC[2KZGN6GUV
4GUVQTG7UGT&GHCWNVU
#NCTOU
#NCTOU&GVGEVGF
%JCODGT6GOR
%QQNGT6GOR
+PVGTPCN6GOR
&KCIPQUVKEU
5ETGGP
5ETGGP
5GNH6GUV
2QYGTÄWR
4WP5ETGGP
/CKP/GPW
+PUVTWOGPV
%CNKDTCVKQP
6GORU
8QNVCIGU
2TQITCO8GTUKQP
2TGUUWTG
%QPVTQNU
#WVQ/CP/QFG
1\QPCVQT
2/65WRRN[
&CVCNQIIKPI5GVVKPIU
:
%CN01$MI
%CN01$MI
%CN01%QGH
%CN01%QGH
%QPX6GOR
(NQY
%QOO5GVVKPIU
:
%CN01%QGH
#PCNQI+PRWV4GCFKPIU
+1%QPHKI
%CN2TGUUWTG
2TGUUWTG
1\QPCVQT(NQY
#PCNQI+PRWV8QNVCIGU
&KIKVCN+PRWVU
6GOR%QORGPUCVKQP
2TGU%QORGPUCVKQP
<GTQ5RCP%JGEM
<GTQ%JGEM
4GNC[5VCVGU
5ETGGP%QPVTCUV
%CN01$MI
5RCP%JGEM
6GUV#PCNQI1WVRWVU
5GTXKEG/QFG
%CN01$MI
<GTQ#WVQECN
:
:
5RCP#WVQECN
01%QPE
01%QPE
01%QPE
/$5VCVWU
+PV$F5VCVWU
+1'ZR$F5VCVWU
%QPVCEV+PHQ
+PUVTWOGPV%QPHKI
&CVG6KOG
:
:
%CN*K01%QGH
%CN*K01%QGH
%CN*K01%QGH
%CN.Q01%QGH
%CN.Q01%QGH
%CN.Q01%QGH
%CN2TGUUWTG
(CEVQTU
%CNKDTCVKQP
:
:
01%QGH
%CN2TGUUWTG
01%QGH
01%QGH
4GUGV7UGT%CN&GHCWNVU
01$MI
01$MI
:
01$MI
01$MI
*K01%QGH
*K01%QGH
*K01%QGH
%CN2TGUUWTG
.Q01%QGH
.Q01%QGH
.Q01%QGH
4GUGV7UGT%CN&GHCWNVU
Figure 3-3. Flowchart of Menu-Driven Software
:
:
6KOG
#XGTCIKPI
#XI6KOG
4CPIG
)CU7PKVU
014CPIG
014CPIG
*K#XI6KOG
.Q#XI6KOG
(QT&WCNQT#WVQ4CPIG1PN[
:
:
:
5GV%WUVQO4CPIGU
014CPIG
)CU7PKVU
*K014CPIG
*K014CPIG
.Q014CPIG
.Q014CPIG
5GV%WUVQO4CPIGU
*K014CPIG
.Q014CPIG
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-5
Page 34
Operation
Software Overview
Power-Up ScreenThe Power-Up screen is displayed on power up of the Model 42i Low
Source. The Self-Test is displayed while the internal components are
warming up and diagnostic checks are performed.
Please wait while booting...
Run ScreenThe Run screen displays the NO, NO
status bar displays time and the status of the remote control interface, and
optional zero/span solenoid valves, if installed. The word “SAMPLE” on the
bottom left of the display indicates the analyzer has the span/zero valve
option and is in “SAMPLE” mode. Other modes appear in the same area of
the display as “ZERO” or “SPAN”. For more information about the
optional solenoid valves, see Chapter 9, “Optional Equipment”.
When operating in dual or auto range mode, two sets of coefficients are used
to calculate the NO-NO2-NOx “LOW” and “HIGH” concentrations. Also,
two averaging times are used—one for each range. The title bar indicates
which range concentrations are displayed. The words “LOW RANGE
CONCENTRATION” on the top of the display indicates that the low
concentration is displayed. In dual range mode, pressing the
arrows will toggle between high and low concentrations. The
example below shows the Run screen in single range mode.
, and NOx concentrations. The
2
and
3-6Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 35
Operation
XCONCENTRATIONW
NO62.7 PPB
NO225.5 PPB
NOx88.2 PPB
SAMPLE 12:34 ALARMx
RANGE AVG DIAGS ALARM
Software Overview
Main MenuThe Main Menu contains a number of submenus. Instrument parameters
and features are divided into these submenus according to their function.
The concentration appears above the main menu and submenus in every
screen, as shown below. The Service menu is visible only when the
instrument is in service mode. For more information on the service mode,
see “Service Mode” later in this chapter.
•Use and to move the cursor up and down.
•Press to select a choice.
•Press to return to the Main Menu or to return to the Run
screen.
XXLO WCONCENTRATIONXXLO W
NO 62.7 PPB
NO2 25.5 PPB
NOx 88.2 PPB
SAMPLE 12:34ALARMx
MAIN MENU:
>RANGE
AVERAGING TIME
CALIBRATION FACTORS
CALIBRATION
INSTRUMENT CONTROLS
DIAGNOSTICS
ALARMS
RANGE AVG DIAGS ALARM
ª
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-7
SERVICE
PASSWORD
Page 36
Operation
Range Menu
Range MenuThe Range menu allows the operator to select the gas units, NO-NO
ranges, and to set the custom ranges. The screens below show the range
menu in single range mode and dual/auto range modes. For more
information about the single, dual, and auto range modes, see “Single Range
Mode”, “Dual Range Mode”, and “Auto Range Mode” below.
•In the Main Menu, choose Range.
RANGE:RANGE:
>GAS UNITS PPB >GAS UNITS PPB
NO RANGE 200 HI NO RANGE 1000
NO2 RANGE 200 LO NO RANGE 200
NOx RANGE 200 HI NO2 RANGE 1000
SET CUSTOM RANGES LO NO2 RANGE 200
HI NOx RANGE 1000
LO NOx RANGE 200
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
SET CUSTOM RANGES
-NOx
2
ª
Single Range ModeIn the single range mode, the NO, NO
range, one averaging time, and one span coefficient.
By default, the three analog outputs are arranged on the rear panel
connector as shown in Figure 3-4. See Table 3-2 for channels and pin
connections. Single range mode may be selected from the
“Single/Dual/Auto Select” in the “Service Menu”, later in this chapter.
, and NOx channels each have one
2
3-8Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 37
Operation
Range Menu
Figure 3-4. Pin-Out of Rear Panel Connector in Single Range Mode
Table 3-2. Default Analog Outputs in Single Range Mode
ChannelPinDescription
114NO Analog Output
233NO
315NO
434None
517None
636None
Ground16, 18, 19, 35, 37Signal Ground
Analog Output
2
Analog Output
x
Note All channels are user definable. If any customization has been made to
the analog output configuration, the default selections may not apply.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-9
Page 38
Operation
Range Menu
Dual Range ModeIn the dual range mode, there are two independent channels defined for
each compound being measured. These are labeled simply as the “High
Range” and the “Low Range”. Each channel has its own analog output
range, averaging time and span coefficient.
This enables the sample concentration readings to be sent to the analog
outputs using two different ranges. For example, the low NO analog output
can be set to output concentrations from 0 to 200 ppb and the high NO
analog output set to output concentrations from 0 to 500 ppb.
In addition to each channel having two ranges, each channel has two span
coefficients. There are two span coefficients so that each range can be
calibrated separately. This is necessary if the two ranges are not close to one
another. For example, if the low NO range is set to 0–200 ppb and the high
NO range is set to 0–20,000 ppb.
By default, in the dual range mode, the analog outputs are arranged on the
rear panel connector as shown in Figure 3-5. See Table 3-3 for channels and
pin connections. Dual range mode may be selected from the
“Single/Dual/Auto Select” in the “Service Menu”, later in this chapter.
Figure 3-5. Pin-Out of Rear Panel Connector in Dual Range Mode
3-10Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 39
Table 3-3. Default Analog Outputs in Dual Range Mode
ChannelPinDescription
114NO High Range
Operation
Range Menu
233NO
315NO
434NO Low Range
517NO
636NO
Ground16, 18, 19, 35, 37Signal Ground
High Range
2
High Range
x
Low Range
2
Low Range
x
Note All channels are user definable. If any customization has been made to
the analog output configuration, the default selections may not apply.
Auto Range ModeThe auto range mode switches the NO, NO
between high and low ranges, depending on the NOx concentration level,
except when continuous NO mode is active. The high and low ranges are
defined in the Range menu.
For example, suppose the low ranges are set to 200 ppb and the high ranges
are set to 1000 ppb (Figure 3-6). Sample concentrations below 200 ppb are
presented to the low ranges analog outputs and sample concentrations
above 200 ppb are presented to the high ranges analog outputs. When the
low ranges are active, the status output is at 0 volts. When the high ranges
are active, the status output is at half of full-scale.
and NOx analog outputs
2
When the high ranges are active, the NOx concentration must drop to 95%
of the low NO
range for the low ranges to become active.
x
In addition to each channel having two ranges, each channel has two span
coefficients. There are two span coefficients so that each range can be
calibrated separately. This is necessary if the two ranges are not close to one
another. For example, if the low NO range is set to 0–200 ppb and the high
NO range is set to 0–20,000 ppb.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-11
Page 40
Operation
Range Menu
Figure 3-6. Analog Output in Auto Range Mode
By default, in the auto range mode, the analog outputs are arranged on the
rear panel connector as shown in Figure 3-7. See Table 3-4 for channels and
pin connections. Auto range mode may be selected from the
“Single/Dual/Auto Select” in the “Service Menu”, later in this chapter.
3-12Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 41
Operation
Range Menu
Figure 3-7. Pin-Out of Rear Panel Connector in Auto Range Mode
Table 3-4. Default Analog Outputs in Auto Range Mode
ChannelPinDescription
114NO Analog Output
233NO
315NO
434Range Status:
517None
636None
Ground16, 18, 19, 35, 37Signal Ground
Analog Output
2
Analog Output
x
half scale = high range
zero scale = low range
Note All channels are user definable. If any customization has been made to
the analog output configuration, the default selections may not apply.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-13
Page 42
Operation
Range Menu
Gas UnitsThe Gas Units screen defines how the NO, NO
readings are expressed. Gas units of parts per billion (ppb), parts per million
(ppm), micrograms per cubic meter (µg/m3), or milligrams per cubic meter
(mg/m3) are available. The µg/m3 and mg/m3 gas concentration modes are
calculated using a standard pressure of 760 mmHg and a standard
temperature of 20 °C.
When switching the selected units from ppb or ppm to µg/m3 or mg/m3,
the analog ranges all default to the highest range in that mode. For example,
when switching from mg/m3 to ppm, all the ranges default to 100 ppm.
Therefore, whenever you change units, you should also check the range
settings.
•In the Main Menu, choose Range > Gas Units.
•Use and to scroll through a list of choices.
•Press to save the new units.
Note If the units change from ppb/ppm to µg/m3/mg/m3 or vice-versa, the
instrument should be recalculated, particularly if the user’s standard
temperature is different from 20 °C. A display warning will appear that
ranges will be defaulted and calibration parameters reset.
, and NOx concentration
2
GAS UNITS:
CURRENTLY: PPB
SET TO: UG/M3 ?
AND SET RANGES AND CAL TOX
DEFAULT SETTINGSXXXXXXXXXX
, and NOx Ranges define the concentration range of the
2
analog outputs. For example, an NO2 range of 0–200 ppb restricts the NO2
analog output to concentrations between 0 and 200 ppb.
The display shows the current NO, NO
, or NOx range. The next line of
2
the display is used to change the range. The range screen is similar for the
single, dual, and auto range modes. The only difference between the screens
are the words “High” and “Low” to indicate which range is displayed. The
example below shows the NO range screen in single mode. For more
3-14Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 43
Operation
Range Menu
information about the single, dual, and auto range modes, see “Single
Range Mode”, “Dual Range Mode”, and “Auto Range Mode” earlier in this
chapter.
Table 3-5 lists the standard ranges. Table 3-6 lists the extended ranges.
When switching from standard to extended ranges, the PMT voltage must
be readjusted. For more information about readjusting the PMT voltage,
see Chapter 7, “Servicing”.
•In the Main Menu, choose Range > NO, NO2, or NOx Range.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-15
Page 44
Operation
Range Menu
Table 3-6. Extended Ranges
ppbppmµg/m
1,0001.02,0002.0
2,0002.05,0005.0
5,0005.010,00010.0
10,00010.020,00020.0
20,00020.050,00050.0
50,00050.0100,000100
100,000100200,000200
200,000200500,000500
500,000500750,000750
C1C1C1C1
C2C2C2C2
C3C3C3C3
3
mg/m
3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “Set Custom Ranges” below.
Set Custom RangesThe Set Custom Ranges menu lists three custom ranges: C1, C2, and C3.
Custom ranges are user-defined ranges. In the standard range mode, any
value between 200 ppb (0.2 ppm) and 100,000 ppb (100 ppm) can be
specified as a range. In the µg/m3(mg/m3) mode, any value between 500
µg/m3 (0.5 mg/m3) and 150,000 µg/m3 (150 mg/m3) can be specified as a
range. In the extended range mode, any value between 1,000 ppb (1.0 ppm)
and 500,000 ppb (500 ppm) can be specified as a range. In the
3
µg/m
750,000 µg/m
•In the Main Menu, choose Range > Set Custom Ranges.
(mg/m3) mode, any value between 2,000 µg/m3 (2.0 mg/m3) and
3
(750 mg/m3) can be specified as a range.
CUSTOM RANGES:
>CUSTOM RANGE 1 255.6
CUSTOM RANGE 2 275.0
CUSTOM RANGE 3 1125.0
RANGE AVG DIAGS ALARM
3-16Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 45
Custom RangesThe Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display is
used to set the range. To use the custom full-scale range, be sure to select it
(Custom range 1, 2, or 3) in the NO, NO2, or NOx Range screen. For
more information about selecting ranges, see “NO, NO
, and NOx Ranges”
2
above.
•In the Main Menu, choose Range > Set Custom Ranges > Custom range 1, 2, or 3.
•Use and to increment or decrement the numeric value.
•Press to save the new range.
CUSTOM RANGE 1:
CURRENTLY: 255.6
SET TO: 000255.7 ?
Averaging TimeThe averaging time defines a time period (10 to 300 seconds) over which
NO, NO2, and NOx measurements are averaged. The average
concentration of the NO, NO2, and NOx readings are calculated for that
time period. The front panel display and analog outputs are updated every
10 seconds with the calculated averages. An averaging time of 10 seconds,
for example, means that the average concentration of the last 10 seconds
will be output at each update. An averaging time of 300 seconds means that
the moving average concentration of the last 300 seconds will be output at
each 10-second update. Therefore, the lower the averaging time, the faster
the front panel display and analog outputs respond to concentration
changes. Longer averaging times are typically used to smooth output data.
The Averaging Time screen for the single range mode is shown below. In the
dual and auto range modes, an averaging time menu is displayed before the
averaging time screen. This additional menu is needed because dual and
auto range modes have two averaging times (high and low). The averaging
time screens function the same way in the single, dual and auto range
modes. The following averaging times are available: 10, 20, 30, 60, 90, 120,
180, 240, and 300 seconds. Additional averaging times are available in
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-17
Page 46
Operation
Calibration Factors Menu
manual NO and manual NOx modes: 1, 2, and 5 seconds. For more
information about the manual mode, see “Auto/Manual Mode” later in this
chapter.
•In the Main Menu, choose Averaging Time.
•Use and to scroll through a list of choices.
•Press to save the averaging time.
AVERAGING TIME:
CURRENTLY: 30 SEC
SET TO: 10 SEC ?
Calibration factors are used to correct the NO, NO2, and NOx
concentration readings that the instrument generates using its own internal
calibration data. The Calibration Factors menu displays the calibration
factors. The screens below show the calibration factors menu in single mode
and dual/auto range modes.
Normally, the instrument is calibrated automatically using the commands
listed in the Calibration menu described later in this chapter. However, the
instrument can also be calibrated manually using the Calibration Factors
menu.
To manually calibrate the instrument, see “NO and NO
Backgrounds”,
x
“NO, NO2, and NOx Span Coefficients”, and “Calibration Pressure” below.
•In the Main Menu, choose Calibration Factors.
3-18Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 47
Operation
Calibration Factors Menu
CALIBRATION FACTORS:CALIBRATION FACTORS:
>NO BKG 0.0 >NO BKG 0.0
NOx BKG 0.0 NOx BKG 0.0
NO COEF 1.000 HI NO COEF 1.000
NO2 COEF 1.000 HI NO2COEF 1.000
NOx COEF 1.000 HI NOx COEF 1.000
CAL PRESSURE 150.0 mmHg LO NO COEF 1.000
RESET USER CAL DEFAULTS
LO NO2 COEF 1.000
ª
RANGE AVG DIAGS ALARM
NO and NOx BackgroundsThe NO and NO
calibration. The NO background is the amount of signal read by the
analyzer in the NO channel while sampling zero air. The NOx background
is the amount of signal read by the analyzer in the NOx channel while
sampling zero air. Although the background is expressed in terms of
concentration, the background signal is actually the combination of
electrical offsets, PMT dark current, and trace substances undergoing
chemiluminescence. Before the analyzer sets the NO and NOx readings to
zero, it stores these values as the NO and NOx background corrections,
respectively. The NO2 background correction is determined from the NO
and NOx background corrections and is not displayed. NO and NOx
background corrections are typically below 15 ppb.
The NO and NO
adjustment of the instruments zero background. Before performing a
background adjustment, allow the analyzer to sample zero air until stable
readings are obtained. The NO channel should be adjusted first. Both the
NO and NOx Background screens operate the same way. Therefore, the
following description of the NO background screen applies to the NOx
background screen as well. The first line of the display shows the current
NO reading. The second line of the display shows the NO background
correction that is stored in memory. The NO background correction is a
value, expressed in the current gas units, that is subtracted from the NO
reading to produce the NO reading that is displayed.
RANGE AVG DIAGS ALARM
LO NOx COEF 1.000
CAL PRESSURE 150.0 mmHg
RESET USER CAL DEFAULTS
background corrections are determined during zero
x
Background screens are used to perform a manual
x
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-19
Page 48
Operation
Calibration Factors Menu
In the example below, the analyzer is displaying 4.4 ppb of NO while
sampling zero air. A background correction of 0.0 ppb means that 0 ppb is
being subtracted from the NO concentration being displayed. Therefore,
the background correction must be increased to 4.4 ppb in order for the
NO reading to be at 0 ppb, i.e., a NO reading of 4.4 ppb minus a NO
background reading of 4.4 ppb gives the corrected NO reading of 0 ppb.
To set the NO reading in the example below to zero, use to
increment the NO background correction to 4.4 ppb. As the NO
background correction is increased, the NO concentration is decreased.
Note that at this point, pressing and however, has no affect on
the analog outputs or the stored NO background correction of 0.0 ppb. A
question mark following both the NO reading and the NO background
correction indicates that these are proposed changes as opposed to
implemented changes. To escape this screen without saving any changes,
press to return to the Calibration Factors menu or to return
to the Run screen. Press to actually set the NO reading to 0 ppb and
store the new background correction of 4.4 ppb. Then the question mark
prompt beside the NO reading disappears.
•In the Main Menu, choose Calibration Factors > NO or NOx
Background.
•Use and to increment or decrement the proposed
background value.
3-20Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 49
Operation
Calibration Factors Menu
NO, NO2, and NO
Coefficients
The NO, NO2, and NOx span coefficients are usually calculated by the
x
instrument processor during calibration. The span coefficients are used to
correct the NO, NO2, and NOx readings. The NO and NOx span
coefficients normally have values near 1.000. The NO2 span coefficient
normally has a value between 0.95 and 1.050.
The NO, NO2, and NOx Coefficient screens allow the NO, NO2, and
NOx span coefficients to be manually changed while sampling span gas of
known concentration. The NO, NO2, and NOx coefficient screens operate
the same way. Therefore, the following description of the NO coefficient
screen applies to the NO2 and NOx coefficient screens as well.
The display shows the current NO concentration reading. The next line of
the display shows the NO span coefficient that is stored in memory and is
being used to correct the NO concentration. Notice that as the span
coefficient value is changed, the current NO concentration reading on the
above line also changes. However, no real changes are made to the value
stored in memory until is pressed. Only proposed changes, as
indicated by a question mark prompt, are displayed until is pressed.
In dual or auto range mode, “High” or “Low” is displayed to indicate the
calibration of the high or low coefficient. The example below shows the
coefficient screen in single mode.
Note The concentration value will show ERROR if the measured
concentration is not a valid span value (either higher than the selected
range, or 0 or lower).
•In the Main Menu, choose Calibration Factors > NO, NO2 or NOx
Coef.
•Use and to increment or decrement the coefficient value.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-21
Page 50
Operation
Calibration Factors Menu
Calibration PressureThe Calibration Pressure screen reports the reactor pressure at which current
calibration was performed. The screen also allows this value to be edited by
the operator. This value should not be changed unless a complete manual
calibration is being performed. The calibration pressure is set equal to the
reactor pressure at the time of calibration.
Note Turning the pressure compensation on and off can produce significant
artificial jumps in the apparent sample concentration. If the pressure
compensation feature is to be used, the instrument must be calibrated with
the pressure compensation feature on. For more information about
calibration, see Chapter 4, “Calibration”.
•In the Main Menu, choose Calibration Factors > Calibration Pressure.
•Use and to increment or decrement the calibration
pressure value.
•Press to save the new calibration pressure.
CALIBRATION PRESSURE:
PRESSURE: 151.6
SET CAL PRES TO: 150.0
Reset User Calibration DefaultThe Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
•In the Main Menu, choose Service > Calibration Factors > Reset User Cal Defaults.
•Press to warn user and enable restore with .
•Use to reset the calibration pressure reference value when pressed
after .
3-22Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 51
RESTORE DEFAULT CAL:RESTORE DEFAULT CAL:
RESTORE RESTORE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESTORE
Operation
Calibration Menu
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Calibration MenuThe Calibration menu is used to calibrate the analyzer, including zero
backgrounds, the NO, NO2, and NOx coefficients, and the calibration
pressure. The screens below show the calibration menu in single mode and
dual/auto range modes.
The calibration procedure is the same in dual, auto, or single range,
however, there are two sets of gas coefficients in dual or auto range (i.e. low
and high coefficients). This enables each range to be calibrated separately.
When calibrating the instrument in dual or auto range, be sure to use a low
span gas to calibrate the low range and a high span gas to calibrate the high
range.
Whether dual or single range, the first step in calibrating the instrument is
to assign the calibration pressure. The calibration pressure is used to
account for slight pressure fluctuation brought on by changing weather. The
calibration pressure is set equal to the reactor pressure at the time of
calibration. For more information about calibration, see Chapter 4,
“Calibration”.
•In the Main Menu, choose Calibration.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-23
Page 52
Operation
Calibration Menu
CALIBRATION:CALIBRATION:
>CAL NO BACKGROUND>CAL NO BACKGROUND
CAL NOx BACKGROUND CAL NOx BACKGROUND
CAL NO COEFFICIENT CAL HI NO COEFFICIENT
CAL NO2 COEFFICIENT CAL HI NO2 COEFFICIENT
CAL NOx COEFFICIENT CAL HI NOx COEFFICIENT
CALIBRATE PRESSURE CAL LO NO COEFFICIENT
ZERO/SPAN CHECK
CAL LO NO2 COEFFICIENT
ª
RANGE AVG DIAGS ALARM
Calibrate NO and NO
Background
The Calibrate NO and NOx Background screens are used to adjust the
x
background, or perform a “zero calibration”. Be sure the analyzer samples
zero air until the readings stabilize. The display shows the current NO or
NOx reading.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more precise the calibration will be. For the most precise
calibration, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “Calibration”.
•In the Main Menu, choose Calibration > Calibrate NO or NOx
•Press to set the new reading to zero.
RANGE AVG DIAGS ALARM
CAL LO NOx COEFFICIENT
CALIBRATE PRESSURE
Background.
NO BACKGROUND:
NO: 1.2
CURRENTLY: 0.0 ?
SET NO TO ZERO
RANGE AVG DIAGS ALARM
3-24Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 53
Operation
Calibration Menu
Calibrate NO, NO2, and
NOx Coefficients
The Calibrate NO Coefficient screen is used to adjust the NO span
concentrations while sampling span gas of known concentration. All
calibration screens operate the same way. Therefore, the following
description of the NO calibration screen applies to the NO2 and NOx
calibration screens as well.
The display shows the current NO concentration reading and the current
NO range. The next line of the display is where the NO calibration gas
concentration is entered.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more precise the calibration will be. For the most precise
calibration, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “Calibration”.
•In the Main Menu, choose Calibration > Cal NO, NO2, or NOx
Coefficents.
•Use and pushbuttons to move the cursor left or right.
•Use and to increment or decrement the numeric value.
•Press to calculate and save the new coefficient based on the
entered span concentration.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-25
Page 54
Operation
Calibration Menu
Calibrate PressureThe Calibrate Pressure screen is used to set the calibration pressure.
The calibration pressure should be set only when calibrating the instrument.
•In the Main Menu, choose Calibration > Calibrate Pressure.
•Press to save the current chamber as the new calibration pressure.
CALIBRATE PRESSURE:
CUR CAL PRES: 150.0 mmHg
CURRENT PRES: 151.6 mmHg ?
SAVE CURRENT PRESSURE
AS CAL PRESSURE
RANGE AVG DIAGS ALARM
Zero/Span CheckThe Zero/Span Check menu is available with the zero/span valve option. It
is used to program the instrument to perform fully automated zero and span
check or adjustments. Total Duration Hour is the sum of zero, span, and
purge duration minutes. Zero and Span Calibration Reset are toggle items
that change between yes or no when selected, and displayed if auto
calibration is installed.
•In the Main Menu, choose Calibration > Zero/Span Check.
ZERO/SPAN CHECK:
>NEXT TIME 01Jan05 12:00
PERIOD HR 24
TOTAL DURATION HR 1.5
ZERO DURATION MIN 30
SPAN DURATION MIN 30
PURGE DURATION MIN 30
ZERO/SPAN AVG SEC 60
RANGE AVG DIAGS ALARM
ZERO CAL RESET
SPAN CAL RESET OFF
ZERO/SPAN RATIO 1
ª
3-26Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 55
Next TimeThe Next Time screen is used to view and set the next zero/span check date
and time. Once the initial zero/span check is performed, the date and time
of the next zero/span check is calculated and displayed.
•In the Main Menu, choose Calibration > Zero/Span Check > Next Date/Time.
•Use , , and to move and change the value of
the date and time.
•Press to accept a change.
NEXT DATE AND TIME:NEXT DATE AND TIME:
01 JAN 2005 12:00 02 JAN 2005 12:00 ?
PRESS
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-27
Page 56
Operation
Calibration Menu
Zero/Span/Purge Duration
Minutes
The Zero Duration Minutes screen defines how long zero air is sampled by
the instrument. The span and purge duration screens look and function the
same way as the zero duration screen, and are used to set how long the span
gas and sample gas are sampled by the instrument. Durations between 0 and
60 minutes are acceptable. Each time a zero/span check occurs the zero
check is done first, followed by the span check. To perform just a zero check,
set the span and purge duration screen to 0 (off). The same applies to
perform just a span or purge check.
•In the Main Menu, choose Calibration > Zero/Span Check > Zero, Span or Purge Duration Min.
•Use and to increment and decrement the numeric value.
Zero/Span Averaging TimeThe Zero/Span Averaging Time screen allows the user to adjust the
zero/span averaging time. The zero/span averaging time is used by the
analyzer only when performing an automatic zero or span check. The
analyzer’s averaging time is used for all other functions. The following
averaging times are available: 1, 2, 5, 10, 20, 30, 60, 90, 120, 180, 240, and
300 seconds.
•In the Main Menu, choose Calibration > Zero/Span Check > Zero/Span Avg Sec.
•Use and to scroll through a list of choices.
•Press to save the averaging time.
3-28Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 57
Operation
Calibration Menu
ZERO/SPAN AVERAGING TIME:
CURRENTLY: 60 SEC
SET TO: 90 SEC ?
Zero/Span RatioThe Zero/Span Ratio screen is used to adjust the ratio of zeros to spans. For
example, if this value is set to 1, a span check will follow every zero check. If
this value is set to 3, there will be two zero checks between each zero/span
check. This value may be set from 1 to 10, with 1 as default.
•In the Main Menu, choose Calibration > Zero/Span Check >
Zero/Span Ratio.
•Use and to increment and decrement the numeric value.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-29
Page 58
Operation
Instrument Controls Menu
Instrument Controls
Menu
The Instrument Controls menu contains a number of items. The software
controls listed in this menu enable control of the listed instrument
functions.
•In the Main Menu, choose Instrument Controls.
INSTRUMENT CONTROLS:
>OZONATOR
PMT SUPPLY
AUTO/MANUAL MODE
DATALOGGING SETTINGS
COMMUNICATION SETTINGS
I/O CONFIGURATION
TEMPERATURE COMPENSATION
RANGE AVG DIAGS ALARM
PRESSURE COMPENSATION
SCREEN CONTRAST
SERVICE MODE
DATE/TIME
ª
OzonatorThe Ozonator screen is used to turn the internal ozonator on or off. The
display shows the status of the control line that turns the ozonator on or off.
The next line of the display shows the user-specified ozonator setting. Under
most conditions, the control line status and ozonator set status are the same.
However, as a safety precaution, the microprocessor can override the
user-specified ozonator setting. This occurs only if the ozonator flow doesn’t
indicate any flow or if the NO2 converter temperature is below the
minimum alarm limit. In this case, an alarm is activated and the ozonator is
turned off. This is done to prevent the ozonator from overheating, which
will result in permanent damage to the ozonator, or if the converter
temperature drops below the minimum limit, which reduces the
effectiveness of the ozone destruct.
It is possible, however, to defeat this feature, by setting the ozonator safety
“OFF” in the “Service Menu”.
Note The ozonator must be “ON” to obtain NO, NO
As an additional safety precaution, a lit LED mounted on the measurement
interface board indicates that the ozonator is on.
•In the Main Menu, choose Instrument Controls > Ozonator.
, and NOx readings.
2
•Press to toggle and set the ozonator on or off.
3-30Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 59
Operation
Instrument Controls Menu
OZONATOR:
CURRENTLY: OFF
SET TO: ON ?
RANGE AVG DIAGS ALARM
TOGGLE VALUE
PMT SupplyThe PMT Supply screen is used to turn the PMT power supply on or off.
This is useful in a troubleshooting situation.
•In the Main Menu, choose Instrument Controls > PMT Supply.
•Press to toggle and set the PMT supply on or off.
PMT SUPPLY:
CURRENTLY: OFF
SET TO: ON ?
RANGE AVG DIAGS ALARM
TOGGLE VALUE
Auto/Manual ModeThe Auto/Manual Mode screen allows selection of the automatic mode
(NO/NOx), NO mode (manual NO), or NOx mode (manual NOx). The
auto mode switches the mode solenoid valve automatically on a 10 second
cycle so that NO, NO2, and NOx concentrations are determined. The
manual NO mode puts the mode solenoid valve into the open position so
that the sample gas bypasses the NO2-to-NO converter. Therefore, only the
NO concentration is determined. The manual NOx mode puts the mode
solenoid valve into the closed position so that the sample gas passes through
the NO2-to-NO converter. Therefore, only the NOx concentration is
determined. In the manual modes, additional averaging times of 1, 2, and 5
seconds are available from the Averaging Time screen.
•In the Main Menu, choose Instrument Controls > Auto/Manual Mode.
•Use and to scroll through a list of choices.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-31
Page 60
Operation
Instrument Controls Menu
Datalogging SettingsThe Datalogging Settings menu deals with datalogging.
Date and TimeThe Date and Time screen is used to set a start date and time for which to
view logged data. For example, if “20 JUN 2005 10:00” is entered, then the
first logged data record that is displayed is the first record after this time. If
set to one minute logging, this would be at “20 JUN 2005 10:01”.
•Use and to increment or decrement the selected field.
•Use to advance to next field.
•Press to set the date and time of the first record to be displayed
and continue to the record display screen.
3-34Model 42i Low Source Instruction ManualThermo Electron Corporation
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM ERASURE
RANGE AVG DIAGS ALARM
Select ContentThe Select Content submenu displays a list of 32 record fields to use and a
submenu list of the analog output signal group choices to choose from.
Choices are Concentrations, Other Measurements, and Analog Inputs (if
the I/O expansion board is installed). This is a temporary list of items for
the selected record type that must be committed via the datalogging menu
before the changes will apply. Note that committing any changes to this list
will erase all currently logged data, as the format of the stored data is
changed.
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content.
LREC FIELDS:
>FIELD 1 NO
FIELD 2 NOX
FIELD 3 PRES
FIELD 4 PMTT
FIELD 5 INTT
FIELD 6 CONVT
FIELD 7 NONE
RANGE AVG DIAGS ALARM
ª
RANGE AVG DIAGS ALARM
DATA IN LREC FIELD 1:
>CONCENTRATIONS
OTHER MEASUREMENTS
ANALOG INPUTS
RANGE AVG DIAGS ALARM
3-36Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 65
Operation
Instrument Controls Menu
ConcentrationsThe Concentrations screen allows the user to select the output signal that is
tied to the selected field item. The selected item is shown by “<--” after it.
Note that at this point, pressing indicates that these are proposed
changes as opposed to implemented changes. To change the selected record
format and erase record log file data, see “Commit Content” below. Range
) is visible only in auto range mode.
(NO
x
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > Select Field > Concentrations.
•Use and to move the cursor up and down.
•Press to select a new choice.
CONCENTRATIONS:
>NONE
NO <- NO2
NOx
LO NO
LO NO2
LO NOx
ª
RANGE AVG DIAGS ALARM
HI NO
HI NO2
HI NOx
RANGE (NOX)
Other MeasurementsThe Other Measurements screen allows the user to select the output signal
that is tied to the selected field item. The selected item is shown by “<--”
after it. Note that at this point, pressing indicates that these are
proposed changes as opposed to implemented changes. To change the
selected record format and erase record log file data, see “Commit Content”
below.
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > Select Field > Other Measurements.
•Use and to move the cursor up and down.
•Press to select a new choice.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-37
Page 66
Operation
Instrument Controls Menu
OTHER MEASUREMENTS:
>NONE
INT TEMP
CHAMBER TEMP
COOLER TEMP
NO2 CNV TEMP
CHAMBER PRES
PMT VOLTS
RANGE AVG DIAGS ALARM
OZONATOR FLOW
ª
Analog InputsThe Analog Inputs screen allows the user to select the output signal (none or
analog inputs 1-8) that is tied to the selected field item. The selected item is
shown by “<--” after it. Note that at this point, pressing indicates
that these are proposed changes as opposed to implemented changes. To
change the selected record format and erase record log file data, see
“Commit Content” below.
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > Select Field > Analog Inputs.
•Use and to move the cursor up and down.
•Press to select a new choice.
ANALOG INPUTS:
>NONE
ANALOG IN 1
ANALOG IN 2
ANALOG IN 3
ANALOG IN 4
ANALOG IN 5
ANALOG IN 6
RANGE AVG DIAGS ALARM
ª
Commit ContentThe Commit Content screen is used to save the selected output signal that is
tied to the selected field item. For more information about selecting the
analog output signal group choices, see “Select Content” above.
3-38Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 67
Operation
Instrument Controls Menu
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Commit Content.
•Press to commit changes to selected record format and erase
record log file data.
•Press to confirm changes.
CHANGE LREC FORMAT ANDCHANGE LREC FORMAT AND
ERASE LREC LOG FILE DATA?ERASE LREC LOG FILE DATA?
COMMIT COMMIT
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM CHANGES
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Reset to Default ContentThe Reset to Default Content screen is used to reset all of the datalogging
field items to default values. For more information about selecting the
analog output signal group choices, see “Select Content” above.
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Reset to Default Content.
•Press to reset selected record format and erase record log file
data.
•Press
RESET LREC FORMAT ANDRESET LREC FORMAT AND
ERASE LREC LOG FILE DATA?ERASE LREC LOG FILE DATA?
RESET RESET
to reset.
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESETXX
RANGE AVG DIAGS ALARM
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-39
RANGE AVG DIAGS ALARM
Page 68
Operation
Instrument Controls Menu
Configure DataloggingThe Configure Datalogging menu deals with datalogging configuration for
Logging Period MinThe Logging Period Min screen is used to select the logging period in
the currently selected record type.
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging.
DATALOGGING SETTINGS:
>LOGGING PERIOD MIN OFF
MEMORY ALLOCATION % 50
DATA TREATMENT AVG
RANGE AVG DIAGS ALARM
minutes for the record format (srec or lrec). List of choices include: off, 1, 5,
15, 30, and 60 minutes (default).
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Logging Period Min.
•Use and to scroll through a list of choices.
•Press to set the logging period.
SET PERIOD FOR SREC:
CURRENTLY: OFF MIN
SET TO: 60 MIN ?
Memory Allocation PercentThe Memory Allocation Percent screen is used to select the percentage of
each record type for both short records and long records. Percentages
between 0 and 100% are available in increments of 10. Changing this value
results in log erasure for both short records and long records.
3-40Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 69
Operation
Instrument Controls Menu
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Memory Allocation %.
•Use and to scroll through a list of choices.
•Press to set the percentage for both record types and proceed to
the erasure warning screen.
•Press to confirm change.
SET PERCENT SRECS:SET PERCENT SRECS:
CURRENTLY: 50 % CURRENTLY: 50 %
SET TO: 60 % ? SET TO: 60 % ?
THIS WILL ERASE ALL LOGS!x
PRESS ¨ TO CONFIRM CHANGE
ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM RANGE AVG DIAGS ALARM
Data TreatmentThe Data Treatment screen is used to select the data type for the selected
record: whether the data should be averaged over the interval, the minimum
or maximum used, or the current value logged. Data treatment doesn’t
apply to all data, just to the concentration measurement. All other data
points log the current value at the end of the interval.
•In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Data Treatment.
Instrument IDThe Instrument ID screen allows the operator to edit the instrument ID.
The ID is used to identify the instrument when using the C-Link or
MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more of the same instrument are
connected to one computer over the serial port. Valid Instrument ID
numbers are from 0 to 127. The Model 42i Low Source has a default
3-42Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 71
Operation
Instrument Controls Menu
Instrument ID of 42. For more information about the Instrument ID, see
Appendix B “C-Link Protocol Commands” or Appendix C “MODBUS
Protocol”.
•In the Main Menu, choose Instrument Controls > Communication
Settings > Instrument ID.
Communication ProtocolThe Communication Protocol screen is used to change the instrument
communication protocol for serial communications. This does not affect
communications over Ethernet, because the protocols are processed
simultaneously on different ports.
In the Main Menu, choose Instrument Controls > Communication Settings
> Communication Protocol.
•Use and to scroll through a list of choices.
•Press to save the new protocol.
COMMUNICATION PROTOCOL:
CURRENTLY: CLINK
SET TO: STREAMING ?
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-43
Page 72
Operation
Instrument Controls Menu
Streaming Data ConfigurationThe Streaming Data Configuration menu is used to allow for configuration
of the 18 streaming data output items, streaming interval, current data
format, and current time. The Choose Stream Data submenu displays a list
of the analog signal group choices to choose from. Choices are
Concentrations, Other Measurements, and Analog Inputs (if the I/O
expansion board option is installed).
Note Add Labels and Prepend Timestamp are toggle items that change
between yes or no when selected.
•In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config.
ConcentrationsThe Concentrations screen allows the user to select the output signal that is
tied to the selected streaming data item. The selected item is shown by “<--”
after it. In dual or auto range mode, “High” or “Low” is displayed to
indicate high or low concentrations. Range (NOx) is visible only in auto
range mode.
•In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Concentrations.
•Use and to move the cursor up and down.
•Press to select a new choice.
CONCENTRATIONS:
>NONE
NO <- NO2
NOx
RANGE (NOX)
RANGE AVG DIAGS ALARM
Other MeasurementsThe Other Measurements screen allows the user to select the output signal
that is tied to the selected streaming data item. The selected item is shown
by “<--” after it.
•In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Other Measurements.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-45
Page 74
Operation
Instrument Controls Menu
•Use and to move the cursor up and down.
•Press to select a new choice.
OTHER MEASUREMENTS:
>NONE
INT TEMP
CHAMBER TEMP
COOLER TEMP
NO2 CNV TEMP
CHAMBER PRES
PMT VOLTS
RANGE AVG DIAGS ALARM
Analog InputsThe Analog Inputs screen allows the user to select the output signal (none or
analog inputs 1-8) that is tied to the selected streaming data item. The
selected item is shown by “<--” after it.
•In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Analog Inputs.
•Use and to move the cursor up and down.
•Press to select a new choice.
ANALOG INPUTS:
>NONE
ANALOG IN 1
ANALOG IN 2
ANALOG IN 3
ANALOG IN 4
ANALOG IN 5
ANALOG IN 6
RANGE AVG DIAGS ALARM
ª
RS-232/RS-485 SelectionThe RS-232/RS-485 Selection screen allows the user to choose between the
RS-232 or RS-485 specification for serial communication.
Note Disconnect the serial cable before changing RS-232 and RS-485
selection to prevent damage to the connected equipment.
•In the Main Menu, choose Instrument Controls > Communication
Settings > RS-232/RS-485 Selection.
3-46Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 75
Operation
Instrument Controls Menu
•Press to leave the warning screen and move to next screen.
•Use to confirm and save the new selection change.
RS-232/RS-485 SELECTION:RS-232/RS-485 SELECTION:
XXXXXX** WARNING **XXXXXXX CURRENTLY: RS-232
DISCONNECT THE SERIALX X SET TO: RS-485 ?
CABLES BEFORE CHANGING XMAKE SURE THAT THE CABLEXX
THE SELECTION!XXXXXX
TO CONTINUE
IS OFF: PRESS
TOGGLE VALUE
¨ TO CONFIRM
RANGE AVG DIAGS ALARM
TCP/IP SettingsThe TCP/IP Settings menu is used for defining TCP/IP settings.
•In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings.
TCP/IP SETTINGS:
>USE DHCP OFF
IP ADDRESS 192.168.1.15
NETMASK 255.255.255.0
GATEWAY 192.168.1.1
HOST NAME ISERIES
RANGE AVG DIAGS ALARM
Use DHCPThe Use DHCP screen is used to specify whether to use DHCP or not.
RANGE AVG DIAGS ALARM
Note The instrument power must be cycled after this parameter has been
changed for the change to take effect.
•In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Use DCHP.
•Press to toggle and set DHCP on or off.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-47
Page 76
Operation
Instrument Controls Menu
DHCP:
CURRENTLY: OFF
SET TO: ON ?
CYCLE POWER TO CHANGE DHCP
RANGE AVG DIAGS ALARM
TOGGLE VALUE
IP AddressThe IP Address screen is used to edit the IP address.
•In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > IP Address.
•Use , , and to move and change the value of
the IP address.
•Press to save the new address.
IP ADDRESS:
CURRENT: 192.168.1.15
SET TO: 192.168.1.15 1
Host NameThe host name screen is used to edit the host name. When DHCP is
enabled, this name is reported to the DHCP server.
•In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Host Name.
•Use , , and to move the cursor or change
between the edit field and the alpha page.
•Press to save the new letter in the alpha table or save the new
alpha page.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-49
Page 78
Operation
Instrument Controls Menu
I/O ConfigurationThe I/O Configuration menu deals with configuration of the analyzer’s I/O
HOST NAME:
CURRENT: ISERIES
ISERIES SSSSSS?
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
system. The analog input configuration is displayed if the I/O expansion
board option is installed.
•In the Main Menu, choose Instrument Controls > I/O Configuration.
I/O CONFIGURATION:
>OUTPUT RELAY SETTINGS
DIGITAL INPUT SETTINGS
ANALOG OUTPUT CONFIG
ANALOG INPUT CONFIG
RANGE AVG DIAGS ALARM
Output Relay SettingsThe Output Relay Settings menu displays a list of the 10 analog output
relays available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note The digital outputs may take up to one second after the assigned state
occurs to show up on the outputs.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings.
3-50Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 79
Operation
Instrument Controls Menu
OUTPUT RELAY SETTINGS:
>1 NOP GEN ALARM
2 NOP NONE
3 NOP UNITS
4 NOP CONC ALARM
5 NOP NONE
6 NOP NONE
7 NOP NO MODE
RANGE AVG DIAGS ALARM
ª
Logic StateThe Logic State screen is used to change the I/O relay to either normally
open or normally closed.
•Press to toggle and set the logic state between open or closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
Instrument StateThe Instrument State submenu allows the user to select the instrument state
that is tied to the selected relay output. A submenu lists signal types of
either alarms or non-alarm to choose from.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State.
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
RANGE AVG DIAGS ALARM
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-51
Page 80
Operation
Instrument Controls Menu
AlarmsThe Alarms status screen allows the user to select the alarm status for the
selected relay output. The selected item is shown by “<--” after it. Items
displayed are determined by the options installed.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State > Alarms.
•Use and to scroll through a list of choices.
•Press to save the new selection for the relay.
ALARM STATUS ITEMS:
>NONE
GEN ALARM <- NO CONC MAX
NO CONC MIN
NO2 CONC MAX
NO2 CONC MIN
NOx CONC MAX
ª
RANGE AVG DIAGS ALARM
NOx CONC MIN
INT TEMP
CHAMB TEMP
COOLER TEMP
NO2 CV TEMP
PRESSURE
OZONE FLOW
MB STATUS
MIB STATUS
I/O BD STATUS
CONC ALARM
Non-AlarmThe Non-Alarm status screen allows the user to select the non-alarm status
for the selected relay output. The selected item is shown by “<--” after it.
Items displayed are determined by the options installed.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State > Non-Alarm.
•Use and to scroll through a list of choices.
•Press to save the new selection for the relay.
3-52Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 81
Operation
Instrument Controls Menu
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE (NOx)
SERVICE
UNITS
ZERO MODE
SPAN MODE
NO MODE
RANGE AVG DIAGS ALARM
NOX MODE
SAMPLE MODE
PURGE MODE
ª
Digital Input SettingsThe Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note The digital inputs must be asserted for at least one second for the
action to be activated.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings.
DIGITAL INPUT SETTINGS:
>1 NOP NO MODE
2 NOP NOX MODE
3 NOP SET BACKGROUND
4 NOP CAL TO LO SPAN
5 NOP AOUTS TO ZERO
6 NOP AOUTS TO FS
7 NOP NONE
RANGE AVG DIAGS ALARM
ª
Logic StateThe Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
and closes to trigger the digital input action. If nothing is connected to the
digital input pin, the state should be left at open to prevent the action from
being triggered.
•Press to toggle and set the logic state open or closed.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-53
Page 82
Operation
Instrument Controls Menu
Instrument ActionThe Instrument Action screen allows the user to select the instrument state
DIGITAL INPUT SETUP:
>LOGIC STATE OPEN
INSTRUMENT ACTION
RANGE AVG DIAGS ALARM
that is tied to the selected digital input. The selected item is shown by “<--”
after it.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings > Select Relay > Instrument Action.
•Use and to scroll through a list of choices.
•Press to save the new selection for the relay.
CHOOSE ACTION:
>NONE
ZERO MODE
SPAN MODE
NO MODE <- NOX MODE
SET BACKGROUND
CAL TO LO SPAN
RANGE AVG DIAGS ALARM
AOUTS TO ZERO
AOUTS TO FS
CAL TO HI SPAN
ª
Analog Output ConfigurationThe Analog Output Configuration menu displays a list of the analog output
channels available for configuration. Channel choices include all voltage
channels, all current channels, voltage channels 1-6, and current channels
1-6 (if the I/O expansion board option is installed). Configuration choices
include selecting range, setting minimum/maximum values, and choosing
signal to output.
3-54Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 83
Operation
Instrument Controls Menu
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config.
OUTPUT CHANNELS:
>ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
ª
RANGE AVG DIAGS ALARM
Select Output RangeThe Select Output Range screen is used to select the hardware range for the
selected analog output channel. Possible ranges for the voltage outputs are:
0-100 mV, 0-1, 0-5, 0-10 V.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Select Channel > Select Range.
•Use and to move the cursor up and down.
•Press to save the new range.
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-5V ?
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-55
Page 84
Operation
Instrument Controls Menu
Minimum and Maximum ValueThe MinimumValue screen is used to edit the zero (0) to full-scale (100)
value in percentages for the selected analog output channel. See Table 3-7
for a list of choices. The minimum and maximum output value screens
function the same way. The example below shows the set minimum value
screen.
•In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > Select Channel > Set Minimum or Maximum Val ue .
•Use and to increment or decrement the numeric value.
•Press to validate and save the new minimum value.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V ALL
CURRENTLY: N/A %
SET TO: 0000.5 % ?
Internal TempUser-set alarm min valueUser-set alarm max value
Chamber TempUser-set alarm min valueUser-set alarm max value
Cooler TempUser-set alarm min valueUser-set alarm max value
)Recommend not to change the setting for this output
x
Zero (0)Range Setting
Zero (0)Range Setting
Zero (0)Range Setting
Zero (0)Range Setting
Zero (0)Range Setting
Zero (0)Range Setting
3-56Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 85
Operation
Instrument Controls Menu
Table 3-7. Analog Output Zero to Full-Scale Table
OutputZero % ValueFull-Scale 100% Value
NO2 Converter TempUser-set alarm min valueUser-set alarm max value
Chamber PressureUser-set alarm min valueUser-set alarm max value
PMT Volts700 volts1100 volts
Everything Else0 units10 units
Choose Signal to OutputThe Choose Signal to Output screen displays a submenu list of the analog
output signal group choices. Group choices are Concentrations, Other
Measurements, and Analog Inputs (if the I/O expansion board option is
installed). This allows the user to select the output signal to the selected
output channel. The Concentrations screen is shown below. See Table 3-8
below for a list of items for each signal group choice.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Select Channel > Choose Signal To Output.
•Use and to move the cursor up and down.
•Press to select a choice.
CHOOSE SIGNAL - CONC
SELECTED OUTPUT: V1
CURRENTLY: NONE
SET TO: NONE ?
NO (single/auto range mode)Internal TempAnalog Input 1
NO
(single/auto range mode) Chamber TempAnalog Input 2
2
NO
(single/auto range mode)Cooler TempAnalog Input 3
x
LO NO (dual mode)NO
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-57
Converter TempAnalog Input 4
2
Page 86
Operation
Instrument Controls Menu
Analog Input ConfigurationThe Analog Input Configuration menu displays a list of the 8 analog input
Table 3-8. Signal Type Group Choices
ConcentrationsOther MeasurementsAnalog Inputs
LO NO2 (dual mode)Chamber PressureAnalog Input 5
LO NO
(dual mode)PMT VoltsAnalog Input 6
x
HI NO (dual mode)Analog Input 7
HI NO
(dual mode)Analog Input 8
2
HI NO
(dual mode)
x
Range (NO
) (auto range mode)
x
channels available for configuration. This screen is displayed if the I/O
expansion board option is installed. Configuration includes entering
descriptor, units, decimal places, choice of 1-10 points in the table, and
corresponding number of points selected.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config.
ANALOG INPUT 01 CONFIG:
>DESCRIPTOR IN1
UNITS V
DECIMAL PLACES 2
TABLE POINTS 2
POINT 1
POINT 2
ª
RANGE AVG DIAGS ALARM
3-58Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 87
Operation
Instrument Controls Menu
DescriptorThe Descriptor screen allows the user to enter the descriptor for the selected
analog input channel. The descriptor is used in datalogging and streaming
data to report what data is being sent out. The descriptor may be from 1 to
3 characters in length, and defaults to IN1 to IN8 (user input channel
number).
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Descriptor.
•Press to save the new descriptor.
ANALOG INPUT DESCRIPTOR:
CURRENTLY: IN1
IN1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
UnitsThe Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in
datalogging and streaming data. The units may be from 1 to 3 characters in
length, and defaults to V (volts).
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Units.
•Press to save the new value.
ANALOG INPUT UNITS:
CURRENTLY: V
VSS
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-59
Page 88
Operation
Instrument Controls Menu
Decimal PlacesThe Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal, from 0 to 6, with a default of 2.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Decimal Places.
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-61
Page 90
Operation
Instrument Controls Menu
User ValueThe User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion
table from -9999999 to 99999999. The default table is a two-point table
with point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U.
•In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Table Point > User Value.
•Use and to move the cursor up and down.
•Press to save the new value.
TABLE POINT 01 USER VAL:
CURRENTLY: 0.00
SET TO: 00000.00
Temperature compensation provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on the
analyzer's subsystems and output have been empirically determined. This
empirical data is used to compensate for any changes in temperature. This
compensation can be used for special applications, or when operating the
instrument outside the recommended temperature range, even though the
Model 42i Low Source does not require temperature compensation for EPA
equivalency.
When temperature compensation is on, the display shows the current
internal instrument temperature (measured by a thermistor on the Interface
board). When temperature compensation is off, the display shows the
factory standard temperature of 30 °C.
•In the Main Menu, choose Instrument Controls > Temperature Compensation.
•Press to toggle and set the temperature compensation on or off.
3-62Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 91
Instrument Controls Menu
TEMPERATURE COMPENSATION:
COMP TEMP: 30.0
CURRENTLY: OFF
SET TO: ON ?
RANGE AVG DIAGS ALARM
TOGGLE VALUE
o
C
Pressure CompensationPressure compensation provides compensation for any changes to the
instrument's output signal due to reaction chamber pressure variations. The
effects of reaction chamber pressure changes on the analyzer's subsystems
and output have been empirically determined. This empirical data is used to
compensate for any change in reaction chamber pressure. This
compensation can be used even though the Model 42i Low Source does not
require pressure compensation for EPA equivalency.
Operation
When pressure compensation is on, the first line of the display represents
the current pressure in the reaction chamber. When pressure compensation
is off, the first line of the display shows the factory standard pressure of 150
mmHg.
•In the Main Menu, choose Instrument Controls > Pressure Compensation.
•Press to toggle and set the pressure compensation on or off.
PRESSURE COMPENSATION:
COMP PRES: 150.0 mmHg
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-63
Page 92
Operation
Instrument Controls Menu
Screen ContrastThe Screen Contrast screen is used to change the contrast of the display.
Intensities between 0 and 100% in increments of 10 are available. Changing
the screen contrast may be necessary if the instrument is operated at extreme
temperatures.
•In the Main Menu, choose Instrument Controls > Screen Contrast.
•Use and to increment or decrement the screen contrast.
Service ModeThe Service Mode screen is used to turn the service mode on or off. The
service mode locks out any remote actions and includes parameters and
functions that are useful when making adjustments or diagnosing the Model
42i Low Source. For more information about the service mode, see “Service
Menu” later in this chapter.
Note The service mode should be turned off when finished, as it prevents
remote operation.
•In the Main Menu, choose Instrument Controls > Service Mode.
•Press to toggle and set the service mode on or off.
SERVICE MODE:
CURRENTLY: OFF
SET TO: ON ?
RANGE AVG DIAGS ALARM
TOGGLE VALUE
3-64Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 93
Date/TimeThe Date/Time screen and allows the user to view and change the system
date and time (24-hour format). The internal clock is powered by its own
battery when instrument power is off.
•In the Main Menu, choose Instrument Controls > Date/Time.
•Use , , and to move and change the value of
the date and time.
•Press to edit and save the new date and time.
DATE AND TIME:
19 MAR 2005 12:34:56DATE AND TIME:
PRESS
Diagnostics MenuThe Diagnostics menu provides access to diagnostic information and
functions. This menu is useful when troubleshooting the instrument. The
analog input readings and analog input voltages are displayed if the I/O
expansion board option is installed.
•In the Main Menu, choose Diagnostics.
DIAGNOSTICS:
>PROGRAM VERSION
VOLTAGES
TEMPERATURES
PRESSURE
FLOW
ANALOG INPUT READINGS
ANALOG INPUT VOLTAGES
RANGE AVG DIAGS ALARM
ª
DIGITAL INPUTS
RELAY STATES
TEST ANALOG OUTPUTS
INSTRUMENT CONFIGURATION
CONTACT INFORMATION
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-65
Page 94
Operation
Diagnostics Menu
Program VersionThe Program Version screen (read only) shows the version number of the
program installed. Prior to contacting the factory with any questions
regarding the instrument, please note the program version number.
•In the Main Menu, choose Diagnostics > Program Version.
PROGRAM VERSION:
PRODUCT: MODEL 42iLS
VERSION: 01.00.01.074
RANGE AVG DIAGS ALARM
VoltagesThe Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board is displayed if
the I/O expansion board option is installed.
•In the Main Menu, choose Diagnostics > Vo lt ag es .
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
RANGE AVG DIAGS ALARM
Motherboard VoltagesThe Motherboard screen (read only) is used to display the current voltage
readings on the motherboard.
•In the Main Menu, choose Diagnostics > Voltages > Motherboard Vo lt a ge s.
3-66Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 95
Operation
Diagnostics Menu
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
Interface Board VoltagesThe Interface Board screen (read only) is used to display the current voltage
readings on the interface board.
•In the Main Menu, choose Diagnostics > Voltages > Interface Board Vo lt a ge s.
INTERFACE BOARD VOLTAGES:
PMT SUPPLY 785.5 V
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
P15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-15.0 SUPPLY -15.0 V
RANGE AVG DIAGS ALARM
I/O Board VoltagesThe I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This screen is displayed if the I/O
expansion board option is installed.
•In the Main Menu, choose Diagnostics > Voltages > I/O Board Vo lt a ge s.
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15 0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-3.0 SUPPLY -3.0 V
RANGE AVG DIAGS ALARM
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-67
Page 96
Operation
Diagnostics Menu
TemperaturesThe Temperatures screen (read only) displays the internal temperature,
reaction chamber temperature, cooler temperature, and converter
temperatures. The internal temperature is the air temperature measured by a
sensor located on the interface board.
•In the Main Menu, choose Diagnostics > Temperatures.
PressureThe Pressure screen (read only) displays the reaction chamber pressure. The
pressure is measured by a pressure transducer at the reaction chamber.
•In the Main Menu, choose Diagnostics > Pressure.
CHAMBER PRES: 151.6 mmHg
RANGE AVG DIAGS ALARM
FlowThe Flow screen (read only) displays the ozone flow rate. An internal flow
sensor measures the ozone flow. For more information, see Chapter 1,
“Introduction”.
•In the Main Menu, choose Diagnostics > Flow.
3-68Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 97
Diagnostics Menu
OZONATOR FLOW: OK
RANGE AVG DIAGS ALARM
Analog Input ReadingsThe Analog Input Readings screen (read only) displays the 8 current
user-scaled analog readings (if the I/O expansion board option is installed).
•In the Main Menu, choose Diagnostics > Analog Input Readings.
Operation
ANALOG INPUT READINGS:
>CO 10.2 PPM
SO2 18.2 PPB
CO2 250 PPM
FL1 20.42 LPM
WND 9.86 V
FL2 1.865 LPM
IO7 0.0 V
RANGE AVG DIAGS ALARM
ª
Analog Input VoltagesThe Analog Input Voltages screen (read only) displays the 8 raw analog
voltage readings (if the I/O expansion board option is installed).
•In the Main Menu, choose Diagnostics > Analog Input Voltages.
ANALOG INPUT VOLTAGES:
>ANALOG IN 1 6.24 V
ANALOG IN 2 4.28 V
ANALOG IN 3 0.00 V
ANALOG IN 4 0.00 V
ANALOG IN 5 0.00 V
ANALOG IN 6 0.00 V
ANALOG IN 7 0.00 V
ª
RANGE AVG DIAGS ALARM
Thermo Electron CorporationModel 42i Low Source Instruction Manual3-69
Page 98
Operation
Diagnostics Menu
Digital InputsThe Digital Inputs screen (read only) displays the state of the 16 digital
inputs.
•In the Main Menu, choose Diagnostics > Digital Inputs.
Test Analog OutputsThe Test Analog Outputs menu contains a number of digital to analog
converter (DAC) calibration items. Channel choices include all analog
outputs, 6 voltage channels, and 6 current channels (if the I/O expansion
board option is installed).
•In the Main Menu, choose Diagnostics > Test Analog Outputs.
3-70Model 42i Low Source Instruction ManualThermo Electron Corporation
Page 99
Operation
Diagnostics Menu
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
RANGE AVG DIAGS ALARM
ª
Set Analog OutputsThe Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-scale
voltage, zero sets the analog outputs to 0 volts, and normal operation. The
example below shows the selected output state “ALL” is set to normal.
•In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL, Voltage Channel 1-6, or Current Channel 1-6.
•Use and to move the cursor up and down.
•Press to select a choice.
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL