Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please
consult your local sales representative for details.
Thermo Fisher Scientific
Air Quality Instruments
27 Forge Parkway
Franklin, MA 02038
1-508-520-0430
www.thermo.com/aqi
Page 3
WEEE Compliance
This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is
marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more
recycling/disposal companies in each EU Member State, and this product
should be disposed of or recycled through them. Further information on
Thermo Fisher Scientific’s compliance with these Directives, the recyclers
in your country, and information on Thermo Fisher Scientific products
which may assist the detection of substances subject to the RoHS Directive
are available at: www.thermo.com/WEEERoHS.
Thermo Fisher Scientific WEEE Compliance
Page 4
Page 5
About This Manual
This manual provides information about installing, operating, maintaining,
and servicing the Model 15i analyzer. It also contains important alerts to
ensure safe operation and prevent equipment damage. The manual is
organized into the following chapters and appendices to provide direct
access to specific operation and service information.
● Chapter 1 “Introduction” provides an overview of product features,
describes the principles of operation, and lists the specifications.
● Chapter 2 “Installation” describes how to unpack, setup, and startup
the analyzer.
● Chapter 3 “Operation” describes the front panel display screens, the
front panel pushbuttons, and the menu-driven firmware.
● Chapter 4 “Calibration” provides the procedures for calibrating the
to ensure reliable and consistent instrument operation.
● Chapter 6 “Troubleshooting” presents guidelines for diagnosing
analyzer failures, isolating faults, and includes recommended actions for
restoring proper operation.
● Chapter 7 “Servicing” presents safety alerts for technicians working on
the analyzer, step-by-step instructions for repairing and replacing
components, and a replacement parts list. It also includes contact
information for product support and technical information.
● Chapter 8 “System Description” describes the function and location of
the system components, provides an overview of the firmware structure,
and includes a description of the system electronics and input/output
connections.
● Chapter 9 “Optional Equipment” describes the optional equipment
that can be used with this analyzer.
● Appendix A “Warranty” is a copy of the warranty statement.
● Appendix B “C-Link Protocol Commands” provides a description of
the C-Link protocol commands that can be used to remotely control an
analyzer using a host device such as a PC or datalogger.
Thermo Fisher Scientific Model 15i Instruction Manual i
Page 6
About This Manual
Safety
● Appendix C “MODBUS Protocol” provides a description of the
MODBUS Protocol Interface and is supported both over RS-232/485
(RTU protocol) as well as TCP/IP over Ethernet.
● Appendix D “Geysitech (Bayern-Hessen) Protocol” provides a
description of the Geysitech (Bayern-Hessen or BH) Protocol Interface
and is supported both over RS-232/485 as well as TCP/IP over
Ethernet.
Safety
Safety and Equipment
Damage Alerts
Review the following safety information carefully before using the analyzer.
This manual provides specific information on how to operate the analyzer,
however, if the analyzer is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
This manual contains important information to alert you to potential safety
hazards and risks of equipment damage. Refer to the following types of
alerts you may see in this manual.
Safety and Equipment Damage Alert Descriptions
Alert Description
DANGER A hazard is present that could result in death or serious
WARNING A hazard or unsafe practice could result in serious
CAUTION A hazard or unsafe practice could result in minor to
Equipment Damage A hazard or unsafe practice could result in property
personal injury if the warning is ignored.
personal injury if the warning is ignored.
moderate personal injury if the warning is ignored.
damage if the warning is ignored.
▲
▲
▲
▲
Safety and Equipment Damage Alerts in this Manual
Alert Description
ii Model 15i Instruction Manual Thermo Fisher Scientific
WARNING The Model 15i is supplied with a three-wire grounded
If the equipment is operated in a manner not specified by
The service procedures in this manual are restricted to
power cord. Under no circumstances should this
grounding system be defeated.
the manufacturer, the protection provided by the
equipment may be impaired.
qualified service personnel only.
▲
▲
▲
Page 7
Alert Description
About This Manual
FCC Compliance
CAUTION If the LCD panel breaks, do not let the liquid crystal
Equipment Damage Do not attempt to lift the analyzer by the cover or other
Some internal components can be damaged by small
Disconnect the serial cable before changing the RS-232
Do not use solvents or other cleaning products to clean
Do not remove the LCD panel or frame from the LCD
The LCD polarizing plate is very fragile, handle it
Do not wipe the LCD polarizing plate with a dry cloth, as
contact your skin or clothes. If the liquid crystal contacts
your skin or clothes, wash it off immediately using soap
and water.
external fittings.
amounts of static electricity. A properly grounded
antistatic wrist strap must be worn while handling any
internal component.
and RS-485 selection to prevent damage to any
equipment that the analyzer is currently connected to.
▲
the outside case.
module.
carefully.
it may easily scratch the plate.
▲
▲
▲
▲
▲
▲
▲
FCC Compliance
Do not use alcohol, acetone, MEK or other ketone based
or aromatic solvent to clean the LCD module, but rather
use a soft cloth moistened with a naphtha cleaning
solvent.
Do not place the LCD module near organic solvents or
corrosive gases.
Do not shake or jolt the LCD module.
▲
▲
▲
Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
Note This equipment has been tested and found to comply within the
limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference, in
Thermo Fisher Scientific Model 15i Instruction Manual iii
Page 8
About This Manual
W
W
WEEE Symbol
which case the user will be required to correct the interference at his or her
own expense. ▲
EEE Symbol
here to Get Help
The following symbol and description identify the WEEE marking used on
the instrument and in the associated documentation.
Symbol Description
Marking of electrical and electronic equipment which applies to waste
electrical and electronic equipment falling under the Directive 2002/96/EC
(WEEE) and the equipment that has been put on the market after 13 August
2005.
▲
Service is available from exclusive distributors worldwide. Contact one of
the phone numbers below for product support and technical information
or visit us on the web at www.thermo.com/aqi.
1-866-282-0430 Toll Free
1-508-520-0430 International
iv Model 15i Instruction Manual Thermo Fisher Scientific
Table B–11. Power up Mode Values ................................................................B-40
Table B–12. Set Layout Ack Values..................................................................B-41
Table B–13. Analog Current Output Range Values..........................................B-42
Table B–14. Analog Voltage Output Range Values .........................................B-42
Table B–15. Default Analog Output Channel Assignments.............................B-44
Table C–1. Read Coils for 15i.............................................................................C-8
Table C–2. Read Registers for 15i......................................................................C-9
Table C–3. Write Coils for 15i..........................................................................C-11
Table D–1. Operating Status for Model 15i...................................................... D-7
Table D–2. Error Status for Model 15i .............................................................. D-7
xviii Model 15i Instruction Manual Thermo Fisher Scientific
Page 23
Chapter 1
Introduction
The Model 15i HCl Analyzer measures HCl concentrations using Gas
Filter Correlation (GFC). The Model 15i combines proven technology,
easy to use menu-driven firmware, and advanced diagnostics to offer
unsurpassed flexibility and reliability. The Model 15i has the following
features:
● 320 x 240 graphics display
● Menu-driven firmware
● Field programmable ranges
● User-selectable single/dual/auto range modes
● Multiple user-defined analog outputs
● Analog input options
● High sensitivity
● Fast response time
● Linearity through all ranges
● Highly specific to HCl
● Self-aligning optics
● Automatic temperature and pressure compensation
● User-selectable digital input/output capabilities
● Standard communications features include RS-232/485 and Ethernet
● C-Link, MODBUS, Geysitech (Bayern-Hessen), streaming data, and
NTP (Network Time Protocol) protocols. Simultaneous connections
from different locations over Ethernet.
For details of the analyzer’s principle of operation and product
specifications, see the following topics:
● “Principle of Operation” on page 1-2
● “Specifications” on page 1-3
Thermo Fisher Scientific Model 15i Instruction Manual 1-1
Page 24
Introduction
Principle of Operation
Thermo Fisher Scientific is pleased to supply this HCl analyzer. We are
committed to the manufacture of instruments exhibiting high standards of
quality, performance, and workmanship. Service personnel are available for
assistance with any questions or problems that may arise in the use of this
analyzer. For more information on servicing, see Chapter 7, “Servicing”.
Principle of Operation
The Model 15i operates on the principle that HCl absorbs infrared
radiation. Because infrared absorption is a non-linear measurement
technique, it is necessary to transform the basic analyzer signal into a linear
output. The Model 15i uses an internally stored calibration curve to
accurately linearize the instrument output over any range up to a
concentration of 5000 ppm.
The sample is drawn into the Model 15i through the sample bulkhead, as
shown in Figure 1–1. The sample flows through the optical bench.
Radiation from an infrared source is chopped and then passed through a
gas filter alternating between HCl and N2. The radiation then passes
through a narrow bandpass interference filter and enters the optical bench
where absorption by the sample gas occurs. The infrared radiation then
exits the optical bench and falls on an infrared detector.
The HCl gas filter acts to produce a reference beam which cannot be
further attenuated by HCl in the sample cell. The N2 side of the filter
wheel is transparent to the infrared radiation and therefore produces a
measurement beam, which can be absorbed by HCl in the cell. The
chopped detector signal is modulated by the alternation between the two
gas filters with an amplitude related to the concentration of HCl in the
sample cell. Other gases do not cause modulation of the detector signal
since they absorb the reference and measure beams equally. Thus, the GFC
system responds specifically to HCl.
The Model 15i outputs the HCl concentration to the front panel display,
the analog outputs, and also makes the data available over the serial or
Ethernet connection.
1-2 Model 15i Instruction Manual Thermo Fisher Scientific
Page 25
Introduction
Specifications
Specifications
Figure 1–1. Model 15i Flow Schematic
Table 1–1 lists the specifications for the Model 15i.
Zero noise 0.1 ppm RMS (300 second averaging time)
Lower detectable limit 0.20 ppm
Zero drift (24-hour) ±0.2 ppm
Span drift (24-hour) ±0.2% of full-scale
Response time 120 seconds (30 second averaging time)
Sample flow rate 0.5-2 liters/min
3
3
Operating temperature 5–45 °C*
Thermo Fisher Scientific Model 15i Instruction Manual 1-3
Page 26
Introduction
Specifications
Power requirements 100 VAC @ 50/60 Hz
115 VAC @ 50/60 Hz
220
–240 VAC @ 50/60 Hz
275 Watts
Physical dimensions 16.75” (W) X 8.62” (H) X 23” (D)
Weight Approximately 49 lbs.
Analog outputs 6 voltage outputs; 0–100 mV, 1, 5, 10 V (user selectable),
5% of full-scale over/under range (user selectable), 12 bit
resolution, measurement output user selectable per
channel
Digital outputs 1 power fail relay Form C, 10 relays Form A, user
selectable alarm output, relay logic, 100 mA @ 200 VDC
Digital inputs 16 digital inputs, user select programmable, TTL level,
pulled high
Serial Ports 1 RS-232 or RS-485 with two connectors, baud rate 1200–
115200, data bits, parity, and stop bits, protocols: C-Link,
MODBUS, Geysitech (Bayern-Hessen), and streaming data
(all user selectable)
Ethernet connection RJ45 connector for 10 Mbs Ethernet connection, static or
dynamic TCP/IP addressing, up to 3 simultaneous
connections per protocol
*In non-condensing environments
1-4 Model 15i Instruction Manual Thermo Fisher Scientific
Page 27
Chapter 2
Installation
Installation of the Model 15i includes lifting the instrument, unpacking
and inspection, connecting sample, zero, span, and exhaust lines, and
attaching the analog and/or digital outputs to a recording device. The
installation should always be followed by instrument calibration as
described in the “Calibration” chapter of this manual.
This chapter provides the following recommendations and procedures for
installing the instrument:
● “Lifting” on page 2-1
Lifting
Unpacking and
Inspection
● “Unpacking and Inspection” on page 2-1
● “Setup Procedure” on page 2-3
● “Connecting External Devices” on page 2-5
● “Startup” on page 2-10
When lifting the instrument, use a procedure appropriate to lifting a heavy
object, such as, bending at the knees while keeping your back straight and
upright. Grasp the instrument at the bottom in the front and at the rear of
the unit. Although one person can lift the unit, it is desirable to have two
persons lifting, one by grasping the bottom in the front and the other by
grasping the bottom in the rear.
Equipment Damage Do not attempt to lift the instrument by the cover or
other external fittings.
The Model 15i is shipped complete in one container. If there is obvious
damage to the shipping container when the instrument is received, notify
the carrier immediately and hold for inspection. The carrier is responsible
for any damage incurred during shipment.
▲
Use the following procedure to unpack and inspect the instrument.
1. Remove the instrument from its shipping container and set it on a table
or bench that allows easy access to both the front and rear.
Thermo Fisher Scientific Model 15i Instruction Manual 2-1
Page 28
Installation
Unpacking and Inspection
2. Remove the cover to expose the internal components.
3. Remove the packing material in the analyzer (Figure 2–1).
Remove Packing
(2 pieces)
Remove Packing
(2 pieces)
Units without Optional I/O Board
Figure 2–1. Remove the Packing Material
4. Remove the three shipping screws from the pump (Figure 2–2).
Shipping Screws (3)
Units with Optional I/O Board
Figure 2–2. Removing the Shipping Screws
2-2 Model 15i Instruction Manual Thermo Fisher Scientific
Page 29
Installation
Setup Procedure
5. Check for possible damage during shipment.
6. Check that all connectors and circuit boards are firmly attached.
7. Re-install the cover.
8. Remove any protective plastic material from the case exterior.
Setup Procedure
Use the following procedure to setup the instrument:
1. Connect the sample line to the SAMPLE bulkhead on the rear panel
(Figure 2–3). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of
FEP Teflon, 316 stainless steel, borosilicate glass, or similar tubing with
an OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the
tubing should be less than 10 feet.
Figure 2–3. Model 15i Rear Panel
Note Gas must be delivered to the instrument free of particulates. It may
be necessary to use the Teflon particulate filter as described in “Teflon
Particulate Filter” on page 9-2.
Thermo Fisher Scientific Model 15i Instruction Manual 2-3
▲
Page 30
Installation
Setup Procedure
Note Gas must be delivered to the instrument at atmospheric pressure. It
may be necessary to use an atmospheric bypass plumbing arrangement as
shown in Figure 2–4 if gas pressure is greater than atmospheric pressure. ▲
SAMPLE
Instrument Bulkhead
Vent to Exhaust Line at
Atmospheric Pressure
Input Gas Line
Figure 2–4. Atmospheric Dump Bypass Plumbing
2. Connect the EXHAUST bulkhead to a suitable vent. The exhaust line
should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of
the exhaust line should be less than 10 feet. Verify that there is no
restriction in this line.
3. If the optional zero/span solenoid valves are installed, connect a source
of HCl-free air to the ZERO bulkhead, and connect a source of HCl
span gas to the SPAN bulkhead.
4. Connect a suitable recording device to the rear panel connector. For
detailed information about connecting to the instrument, refer to:
“Connecting External Devices” on page 2-5
“Instrument Controls Menu” on page 3-23
“I/O Configuration” on page 3-39
“External Device Connection Components” on page 7-6
“Terminal Block and Cable Kits” on page 9-3
For detailed information about troubleshooting a connection, refer to
“Analog Output Testing” on page 7-28.
2-4 Model 15i Instruction Manual Thermo Fisher Scientific
Page 31
Connecting External Devices
5. Plug the instrument into an outlet of the appropriate voltage and
frequency.
WARNING The Model 15i is supplied with a three-wire grounded power
cord. Under no circumstances should this grounding system be defeated. ▲
Installation
Connecting
External Devices
Terminal Board PCB
Assemblies
Several components are available for connecting external devices to iSeries
instruments.
For detailed information on the optional connection components, refer to
the “Optional Equipment” chapter. For associated part numbers, refer to
“External Device Connection Components” on page 7-6.
The terminal board PCB assemblies are circuit boards with a D-Sub
connector on one side and a series of screw terminals on the other side.
These assemblies provide a convenient mechanism for connecting wires
from a data system to the analyzer’s I/O connectors.
The following terminal board PCB assemblies are available for iSeries
instruments:
● 25-pin terminal board PCB assembly, (included with optional I/O
expansion board)
I/O Terminal Board
Figure 2–5 shows the recommended method for attaching the cable
(customer supplied) to the terminal board using the included tie-down and
spacer. Table 2–1 identifies the connector pins and associated signals.
Note Not all of the I/O available in the instrument is brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. See optional “Terminal Block and Cable Kits”.
Thermo Fisher Scientific Model 15i Instruction Manual 2-5
▲
Page 32
Installation
Connecting External Devices
See Detail “A”
See Detail “B”
Component Side Viewed from Top of Board
Assembled Connector
Figure 2–5. I/O Terminal Board Views
Table 2–1. I/O Terminal Board Pin Descriptions
Screw Signal Description Screw Signal Description
1 Analog1 13 Power_Fail_NC
2 Analog ground 14 Power_Fail_COM
3 Analog2 15 Power_Fail_NO
4 Analog ground 16 TTL_Input1
Detail “A”
Detail “B”
5 Analog3 17 TTL_Input2
6 Analog ground 18 TTL_Input3
7 Analog4 19 TTL_Input4
8 Analog ground 20 Digital ground
9 Analog5 21 TTL_Input5
10 Analog ground 22 TTL_Input6
11 Analog6 23 TTL_Input7
12 Analog ground 24 Digital ground
The analog outputs are arranged on the rear panel connector as shown in
Figure 2–6. See Table 2–2 for channels and pin connections.
2-6 Model 15i Instruction Manual Thermo Fisher Scientific
Note All channels are user definable. If any customization has been made
to the analog output configuration, the default selections may not apply.
▲
Figure 2–7 shows the recommended method for attaching the cable
(customer supplied) to the terminal board using the included tie-down and
spacer. Table 2–3 identifies the connector pins and associated signals.
Thermo Fisher Scientific Model 15i Instruction Manual 2-7
Page 34
Installation
Connecting External Devices
See Detail “A”
See Detail “B”
Component Side Viewed from Top of Board
Assembled Connector
Figure 2–7. D/O Terminal Board Views
Table 2–3. D/O Terminal Board Pin Descriptions
Screw Signal Description Screw Signal Description
1 Relay1_ContactA 13 Relay7_ContactA
2 Relay1_ContactB 14 Relay7_ContactB
3 Relay2_ContactA 15 Relay8_ContactA
4 Relay2_ContactB 16 Relay8_ContactB
Detail “A” Detail “B”
5 Relay3_ContactA 17 Relay9_ContactA
6 Relay3_ContactB 18 Relay9_ContactB
7 Relay4_ContactA 19 Relay10_ContactA
8 Relay4_ContactB 20 Relay10_ContactB
9 Relay5_ContactA 21 (not used)
10 Relay5_ContactB 22 +24V
11 Relay6_ContactA 23 (not used)
12 Relay6_ContactB 24 +24V
25-Pin Terminal Board
The 25-pin terminal board is included with the optional I/O Expansion
Board. Figure 2–8 shows the recommended method for attaching the cable
(customer supplied) to the terminal board using the included tie-down and
spacer. Table 2–4 identifies the connector pins and associated signals.
2-8 Model 15i Instruction Manual Thermo Fisher Scientific
Page 35
Installation
Connecting External Devices
See Detail “A”
Component Side Viewed from Top of Board
See Detail “B”
Assembled Connector
Figure 2–8. 25-Pin Terminal Board Views
Detail “A” Detail “B”
Table 2–4. 25-Pin Terminal Board Pin Descriptions
Screw Signal Description Screw Signal Description
1 IOut1 13 Analog_In1
2 Isolated ground 14 Analog_In2
3 IOut2 15 Analog_In3
4 Isolated ground 16 Ground
5 IOut3 17 Analog_In4
6 Isolated ground 18 Analog_In5
7 IOut4 19 Analog_In6
8 Isolated ground 20 Ground
9 IOut5 21 Analog_In7
10 Isolated ground 22 Analog_In8
11 IOut6 23 Ground
12 Isolated ground 24 Ground
Thermo Fisher Scientific Model 15i Instruction Manual 2-9
Page 36
Installation
Startup
Startup
Use the following procedure when starting the instrument.
1. Turn the power ON.
2. Allow 90 minutes for the instrument to stabilize.
3. Set instrument parameters such as operating ranges and averaging times
to appropriate settings. For more information about instrument
parameters, see the “Operation” chapter.
4. Before beginning the actual monitoring, perform a multipoint
calibration as described in the “Calibration” chapter.
2-10 Model 15i Instruction Manual Thermo Fisher Scientific
Page 37
Chapter 3
Operation
This chapter describes the front panel display screens, front panel
pushbuttons, and menu-driven firmware. For details, see the following
topics:
● “Display” on page 3-1
● “Pushbuttons” on page 3-2
● “Firmware Overview” on page 3-4
● “Range Menu” on page 3-8
Display
● “Averaging Time” on page 3-15
● “Calibration Factors Menu” on page 3-16
● “Calibration Menu” on page 3-18
● “Instrument Controls Menu” on page 3-23
● “Diagnostics Menu” on page 3-53
● “Alarms Menu” on page 3-61
● “Service Menu” on page 3-69
● “Password Menu” on page 3-80
The 320 x 240 graphics liquid-crystal display (LCD) shows the sample
concentrations, instrument parameters, instrument controls, help, and error
messages. Some menus contain more items than can be displayed at one
time. For these menus, use and to move the cursor up and
down to each item.
Thermo Fisher Scientific Model 15i Instruction Manual 3-1
Page 38
Operation
Pushbuttons
Pushbuttons
Figure 3–1. Front Panel Display
CAUTION If the LCD panel breaks, do not let the liquid crystal contact
your skin or clothes. If the liquid crystal contacts your skin or clothes, wash
it off immediately using soap and water. ▲
The Pushbuttons allow the user to traverse the various screens/menus.
Figure 3–2. Front Panel Pushbuttons
3-2 Model 15i Instruction Manual Thermo Fisher Scientific
Page 39
Table 3–1 lists the front panel pushbuttons and their functions.
Table 3–1. Front Panel Pushbuttons
Key Name Function
= Soft Keys The (soft keys) are used to provide shortcuts that
allow the user to jump to user-selectable menu screens. For
more information on processing soft keys, see “Soft Keys”
below.
Operation
Pushbuttons
Soft Keys
= Run The
= Menu The is used to display the Main Menu when in the
= Help The is context-sensitive, that is, it provides
= Enter The is used to select a menu item, accept/set/save
= Up, Down
= Left, Right
screen normally displays the current HCl concentration.
Run screen, or back up one level in the menu system. For
more information about the Main Menu, see “Main Menu”
later in this chapter.
additional information about the screen that is being
displayed. Press
current screen or menu. Help messages are displayed using
lower case letters to easily distinguish them from the
operating screens. Press
screen, or any other key to exit a help screen.
The four arrow pushbuttons (
change values and states in specific screens.
a change, and/or toggle on/off functions.
is used to display the Run screen. The Run
for a brief explanation about the
to return to the Run
, , , and
) move the cursor up, down, left, and right or
The soft keys are multi-functional keys that use part of the display to
identify their function. The function of the soft keys is to provide a
shortcut to the most often used menus and screens. They are located
directly underneath the display, and user-defined labels in the lower part of
the display indicate the function of each key at that time.
To change a soft key, place the menu cursor “>” on the item of the selected
menu or screen you wish to set. Press followed by the selected soft
key within 1 second of pressing the right-arrow key. The “edit soft key
prompt” will be displayed for configuration of the new label.
Note Not all menu items may be assigned to soft keys. If a particular menu
or screen item cannot be assigned, the key assignment screen will not come
up upon entering right-arrow-soft key combinations. All items under the
Service menu (including the menu itself) cannot be assigned soft keys. ▲
Thermo Fisher Scientific Model 15i Instruction Manual 3-3
Page 40
Operation
Firmware Overview
EDIT SOFT KEY PROMPT:
CURRENTLY: RANGE
RANGE
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Alphanumeric Entry
Screen
Entry Line
Keyboard
ANGE ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The alphanumeric entry screen is used to enter strings consisting of letters,
numbers, and other characters. The cursor may be positioned within the
entry line using the and keys. If a character is entered over an
existing character, that character will be overwritten. Use the and
keys to switch between the entry line and the keyboard as well as to
move within the keyboard. To select a character to add to the string, use
the cursor keys to position the cursor over the desired character, and then
press the key to add that character to the entry line.
On the right side of the keyboard are special functions. BKSP is used to
move the cursor in the entry line one place to the left, deleting the
character that was to the left of the cursor and moving any character at or
to the right of the cursor one place to the left. PAGE is used to change the
keyboard character page. For the English language, this switches between
upper and lower-case alphabetic characters. SAVE stores the string from the
entry line into the parameter. Alternately, if the active cursor is moved to
the entry line, may be pressed to store the string from the entry line
into the parameter.
ANGEabcdefghijklmn BKSP
opqrstuvwxyz PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Firmware
The Model 15i utilizes the menu-driven firmware as illustrated by the
flowchart in Figure 3–3. The Power-Up screen, shown at the top of the
Overview
flowchart, is displayed each time the instrument is turned on. This screen is
displayed while the instrument is warming up and performing self-checks.
After the warm-up period, the Run screen is automatically displayed. The
Run screen is the normal operating screen. It displays the HCl
concentration, depending on operating mode. From the Run screen, the
Main Menu can be displayed by pressing
. The Main Menu contains
a list of submenus. Each submenu contains related instrument settings.
This chapter describes each submenu and screen in detail. Refer to the
appropriate sections for more information.
3-4 Model 15i Instruction Manual Thermo Fisher Scientific
Page 41
Operation
Firmware Overview
Password
Set Password
Change Password
Lock Instrument
Remove Password
Unlock Instrument
Service
Alarms
Diagnostics
Screen
Screen
Self Test
Power−up
Run Screen
Main Menu
Controls
Instrument
Calibration
Initial S/R Ratio
Range Mode Select
Pressure Cal
Flow Cal
Alarms Detected
Intern al Temp
Bench Temp
Pressure
Temps
Voltages
Program Version
Pressure
Temp Compensation
Comm Settings
I/O Config
Datalogging Settings
Zero/Span Check
Cal Bkg
Cal Coef
Temp Cal
Analog Output Cal
Hi Multipoint Cal
AGC Intensity
AGC Intensity
Service Mode
Cal Hi Coef
Motor Speed
Date/Time
Cal Lo Coef
Analog Input Cal
Pre−amp Board Cal
Dilution Ratio
Display Pixel Test
Restore User Defaults
Zero Check
Span C heck
Zero Autocal
Span A utocal
Motor Speed
Conc
MB Status
Interface Status
I/O Exp Status
Test Analog Outputs
Analog Input Readings
Analog Input Voltages
Timezone
Contact Info
Instrument Config
Digital Inputs
Relay St ates
Multipoint Cal
Lo Multipoint Cal
Sample Flow
Bias Voltage
Sample/Ref Ratio
Flow
Screen C ontrast
Pres Co mpensation
Cal Bkg
Factors
Calibration
Coef
Bkg
Reset User Cal Defaults
Time
Averaging
Avg Time
Range
Set Custom Ranges
Gas Units
Range
Hi Coef
Bkg
Lo Coef
Reset User Cal Defaults
Hi Avg Time
Lo Avg Time
For Dual or Auto Range Only
Set Custum Ranges
Gas Units
Hi Range
Lo Range
Figure 3–3. Flowchart of Menu-Driven Firmware
Thermo Fisher Scientific Model 15i Instruction Manual 3-5
Page 42
Operation
Firmware Overview
Power-Up Screen
Run Screen
The Power-Up screen is displayed when power is applied to the Model 15i.
This screen is displayed while the internal components are warming up and
diagnostic checks are being performed.
Please wait while booting...
The Run screen displays the HCl concentrations. The status bar displays
the time (24-hour format), the password (lock) icon, service (wrench) icon,
alarm (bell) icon, and optional zero/span sample solenoid valve status, if
installed.
Status Bar Icons
The password (lock) icon indicates that
no parameter changes can be made from
the front panel.
The service (wrench) icon indicates that the
instrument is in the service mode.
The alarm (bell) icon indicates
that an alarm is active.
The word “SAMPLE” on the left of the status bar indicates the analyzer has
the span/zero valve option and is in “SAMPLE” mode. Other modes
appear in the same area of the display as “ZERO” or “SPAN”. For more
information about the optional solenoid valves, see Chapter 9, “Optional
Equipment”.
3-6 Model 15i Instruction Manual Thermo Fisher Scientific
Page 43
Operation
Firmware Overview
When operating in dual or auto range mode, two sets of coefficients are
used to calculate the HCl “High” and “Low” concentrations. Also, two
averaging times are used, one for each range. The title bar indicates which
range concentrations are displayed. The words “LOW RANGE
CONCENTRATION” on the top of the display indicates that the low
concentration is displayed. In dual range mode, pressing the
arrows will toggle between high and low concentrations. The
and
following example shows the Run screen in single range mode.
Status Bar
Main Menu
Title Bar
X CONCENTRATION W
28.7 PPM
Gas Type
HCl 25.5 PPM
Gas Concentration
Gas Units
29.2 PPM
SAMPLE 12:34 ALARMx
RANGE AVG DIAGS ALARM
Programmable Soft Key Labels
The Main Menu contains a number of submenus. Instrument parameters
and settings can be read and modified within the submenus according to
their function. The concentration appears above the Main Menu and
submenus in every screen. The Service menu is visible only when the
instrument is in service mode. For more information on the service mode,
see “Service Mode” later in this chapter.
Status Bar Icons
● Use and to move the cursor up and down.
● Press to make a selection.
● Press to return to the Main Menu or to return to the Run
screen.
Thermo Fisher Scientific Model 15i Instruction Manual 3-7
Page 44
Operation
Range Menu
XXLO WCONCENTRATIONXXLO W
HCl 25.5 PPM
xSAMPLE 12:34 x
MAIN MENU:
>RANGE
AVERAGING TIME
CALIBRATION FACTORS
CALIBRATION
INSTRUMENT CONTROLS
DIAGNOSTICS
ALARMS
RANGE AVG DIAGS ALARM
SERVICE
PASSWORD
ª
Range Menu
RANGE:
>GAS UNITS PPM
RANGE 5000
SET CUSTOM RANGES
RANGE AVG DIAGS ALARM
Single Range Mode
The Range menu allows the operator to select the gas units, HCl ranges,
and to set the custom ranges. The screens below show the range menu in
single range mode and dual/auto range modes. The only difference between
the screens are the words “HI” and “LO” to indicate which range is
displayed. For more information about the single, dual and auto range
modes, see “Single Range Mode”, “Dual Range Mode”, and “Auto Range
Mode” below.
● In the Main Menu, choose Range
RANGE:
>GAS UNITS PPM
LO RANGE 1000
HI RANGE 5000
SET CUSTOM RANGES
RANGE AVG DIAGS ALARM
In the single range mode, there is one range, one averaging time, and one
span coefficient.
By default, the two HCl analog outputs are arranged on the rear panel
connector as shown in Figure 3–4. See Table 3–2 for channels and pin
connections. Single range mode may be selected from the “Range Mode
Select” in the “Service Menu”, later in this chapter.
3-8 Model 15i Instruction Manual Thermo Fisher Scientific
Page 45
Figure 3–4. Pin-Out of Rear Panel Connector in Single Range Mode
Operation
Range Menu
Dual Range Mode
Table 3–2. Default Analog Outputs in Single Range Mode
Note All channels are user definable. If any customization has been made
to the analog output configuration, the default selections may not apply. ▲
In the dual range mode, there are two independent analog outputs. These
are labeled simply as the “High Range” and the “Low Range”. Each
channel has its own analog output range, averaging time, and span
coefficient.
This enables the sample concentration reading to be sent to the analog
outputs at two different ranges. For example, the low HCl analog output
can be set to output concentrations from 0 to 50 ppm and the high HCl
analog output set to output concentrations from 0 to 100 ppm.
Thermo Fisher Scientific Model 15i Instruction Manual 3-9
Page 46
Operation
Range Menu
In addition to each channel having two ranges, each channel has two span
coefficients. There are two span coefficients so that each range can be
calibrated separately. This is necessary if the two ranges are not close to one
another. For example, the low HCl range is set to 0–5 ppm and the high
HCl range is set to 0–5000 ppm.
By default, in the dual range mode, the analog outputs are arranged on the
rear panel connector as shown in Figure 3–5. See Table 3–3 for channels
and pin connections. Dual range mode may be selected from the “Range
Mode Select” in the “Service Menu”, later in this chapter.
Figure 3–5. Pin-Out of Rear Panel Connector in Dual Range Mode
Table 3–3. Default Analog Outputs in Dual Range Mode
Note All channels are user definable. If any customization has been made
to the analog output configuration, the default selections may not apply. ▲
3-10 Model 15i Instruction Manual Thermo Fisher Scientific
Page 47
Operation
Range Menu
Auto Range Mode
The auto range mode switches the HCl analog outputs between high and
low ranges, depending on the concentration level. The high and low ranges
are defined in the Range menu.
For example, suppose the low range is set to 50 ppm and the high range is
set to 100 ppm (Figure 3–6). Sample concentrations below 50 ppm are
presented to the low ranges analog outputs and sample concentrations
above 50 ppm are presented to the high ranges analog outputs. When the
low range is active, the status output is at 0 volts. When the high range is
active, the status output is at half of full-scale.
When the high range is active, the concentration must drop to 85% of the
low HCl range for the low range to become active.
In addition to each channel having two ranges, each channel has two span
coefficients. There are two span coefficients so that each range can be
calibrated separately. This is necessary if the two ranges are not close to one
another. For example, the low HCl range is set to 0–5 ppm and the high
HCl range is set to 0–5000 ppm.
Figure 3–6. Analog Output in Auto Range Mode
Thermo Fisher Scientific Model 15i Instruction Manual 3-11
Page 48
Operation
Range Menu
By default, in the auto range mode, the analog outputs are arranged on the
rear panel connector as shown in Figure 3–7. See Table 3–4 for channels
and pin connections. Auto range mode may be selected from the “Range
Mode Select” in the “Service Menu”, later in this chapter.
Figure 3–7. Pin-Out of Rear Connector in Auto Range Mode
Table 3–4. Default Analog Outputs in Auto Range Mode
Note All channels are user definable. If any customization has been made
to the analog output configuration, the default selections may not apply. ▲
3-12 Model 15i Instruction Manual Thermo Fisher Scientific
Page 49
Operation
Range Menu
Gas Units
The Gas Units screen defines how the HCl concentration reading is
expressed. Gas units of parts per million (ppm) and milligrams per cubic
meter (mg/m3) are available. The mg/m3 gas concentration mode is
calculated using a standard pressure of 760 mmHg and a standard
temperature of 20 °C.
When switching the selected units, the analog ranges all default to the
highest range in that mode. For example, when switching from mg/m3 to
ppm, all the ranges default to 5000 ppm. Therefore, whenever you change
units, you should also check the range settings.
● In the Main Menu, choose Range > Gas Units.
Note If the units change from ppm to mg/m3 or vice versa, the instrument
should be re-calibrated, particularly if the user’s standard temperature is
different from 20 °C. A display warning will appear that ranges will be
defaulted and calibration parameters reset. ▲
GAS UNITS:
CURRENTLY: PPM
SET TO: MG/M3 ?
AND SET RANGES AND CAL TOX
DEFAULT SETTINGSXXXXXXXXXX
The Range screen defines the concentration range of the analog outputs.
For example, a HCl range of 0–50 ppm sets the analog output to
concentrations between 0 and 50 ppm by default.
The display shows the current HCl range. The next line of the display is
used to change the range. The range screen is similar for the single, dual,
and auto range modes. The only difference between the screens are the
words “High” and “Low” to indicate which range is displayed. The
following example shows the HCl range screen in single mode. For more
information about the dual and auto range modes, see “Single Range
Mode”, “Dual Range Mode”, and “Auto Range Mode” earlier in this
chapter.
Table 3–5 lists the standard ranges.
●In the Main Menu, choose Range > Range.
Thermo Fisher Scientific Model 15i Instruction Manual 3-13
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “Set Custom Ranges” below.
The Set Custom Ranges menu lists three custom ranges, which are userdefined. In the ppm mode, any value between 5 ppm and 5000 ppm can be
specified as a range. In the mg/m3 mode, any value between 10 mg/m3 and
3
7500 mg/m
● In the Main Menu, choose Range > Set Custom Ranges.
can be specified as a range.
CUSTOM RANGES:
>CUSTOM RANGE 1 55.6
CUSTOM RANGE 2 75.0
CUSTOM RANGE 3 125.0
RANGE AVG DIAGS ALARM
3-14 Model 15i Instruction Manual Thermo Fisher Scientific
Page 51
Operation
Averaging Time
Custom Ranges
Averaging Time
The Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display is
used to set the range. To use the custom full-scale range, be sure to select it
(Custom range 1, 2, or 3) in the Ranges screen. For more information
about selecting ranges, see “Range” above.
●In the Main Menu, choose Range > Set Custom Ranges > Custom
Range 1, 2, or 3.
CUSTOM RANGE 1:
CURRENTLY: 55.6
SET TO: 000055.7 ?
The Averaging Time screen allows the user to change the averaging time,
which is the time period (1 to 300 seconds) over which HCl measurements
are taken. The average concentration of the readings is calculated for that
time period. The front panel display and analog outputs are updated every
second for averaging times 1, 2, and 5, and every 10 seconds for averaging
times between 10 and 300 seconds. An averaging time of 10 seconds, for
example, means that the average concentration of the last 10 seconds will
be output at each update. An averaging time of 300 seconds means that the
moving average concentration of the last 300 seconds will be output at each
update. Therefore, the lower the averaging time the faster the front panel
display and analog outputs respond to concentration changes. Longer
averaging times are typically used to smooth output data.
The Averaging Time screen for the single range mode is shown below. In
the dual and auto range modes, an Averaging Time Menu is displayed
before the averaging time screens. This additional menu is needed because
the dual and auto range modes have two averaging times (high and low).
The Averaging Time screen functions the same way in the single, dual, and
auto range modes. The following averaging times are available: 1, 2, 5, 10,
20, 30, 60, 90, 120, 180, 240, and 300 seconds.
● In the Main Menu, choose Averaging Time.
Thermo Fisher Scientific Model 15i Instruction Manual 3-15
Page 52
Operation
Calibration Factors Menu
AVERAGING TIME:
CURRENTLY: 30 SEC
SET TO: 10 SEC ?
CALIBRATION FACTORS:
>BKG 0.0
COEF 1.000
RESET USER CAL DEFAULTS
RANGE AVG DIAGS ALARM
The Calibration Factors menu displays the calibration factors, which are
used to correct the HCl concentration readings that the instrument
generates using its own internal calibration data. The screens below show
the calibration factors menu in single mode and dual/auto range modes.
The only difference between the screens are the words “HI” and “LO” to
indicate which range is displayed.
Normally, the calibration factors are calculated automatically using the
functions described in “Calibration Menu” later in this chapter. However,
the calibration factors can also be set manually using the functions in this
menu.
● In the Main Menu, choose Calibration Factors.
CALIBRATION FACTORS:
>BKG 0.0
HI COEF 1.000
LO COEF 1.000
RESET USER CAL DEFAULTS
RANGE AVG DIAGS ALARM
HCl Background
The background correction is determined during zero calibration. The HCl
background is the amount of signal read by the analyzer while sampling
zero air. The background signal is electrical noise. Before the analyzer sets
the HCl reading to zero, it stores this value as the HCl background
correction.
The Background screen is used to perform a manual adjustment of the
instrument’s zero reading. Before performing a manual zero adjustment,
allow the analyzer to sample zero air until stable readings are obtained.The
first line of the display shows the current HCl reading. This reading is the
HCl background signal. The second line of the display shows the HCl
background correction that is stored in memory and is being used to
correct the HCl reading.
3-16 Model 15i Instruction Manual Thermo Fisher Scientific
Page 53
Operation
Calibration Factors Menu
In the following example, the analyzer displays 15.80 ppm of HCl while
sampling zero air. The HCl background correction is 0.0 ppm. That is, the
analyzer is not applying a zero background correction. The question mark
is used to indicate that the background value has been changed. In this case
the background correction must be increased to 15.80 ppm in order for the
HCl reading to be at 0 ppm.
To set the HCl reading in the following example to zero, use to
increment the HCl background correction to 15.80 ppm. As the HCl
background correction is increased, the HCl concentration is decreased.
Note that at this point, pressing and have no effect on the
analog outputs or the stored HCl background correction of 0.0 ppm. A
question mark following the HCl background correction indicates that
these are proposed changes as opposed to implemented changes. To escape
this screen without saving any changes, press to return to the
Calibration Factors menu or to return to the Run screen. Press
to actually set the HCl reading to 0 ppm and store the new
background correction of 15.80 ppm. Then the question mark prompt
beside the HCl reading disappears.
HCl Coefficient
● In the Main Menu, choose Calibration Factors > Bkg.
The span coefficient is usually calculated by the instrument processor
during calibration. The span coefficients are used to correct the HCl
readings and normally have a value near 1.000.
The Coefficient screen allows the HCl span coefficient to be manually
changed while sampling span gas of known concentration.
The display shows the current HCl concentration reading. The next line of
the display shows the HCl span coefficient that is stored in memory and is
being used to correct the HCl concentration. Notice that as the span
coefficient value is changed, the current HCl concentration reading on the
above line also changes. However, no actual changes are made to the value
stored in memory until is pressed. Only proposed changes, as
indicated by a question mark prompt, are displayed until
is pressed.
Thermo Fisher Scientific Model 15i Instruction Manual 3-17
Page 54
Operation
Calibration Menu
In dual or auto range modes, “HIGH” or “LOW” is displayed to indicate
the calibration of the high or low coefficient. The following example shows
the coefficient screen in dual/auto range mode.
Note The screen will show “SPAN CONC NOT VALID” if the
coefficient change causes the concentration to be either higher than the
selected range or less than or equal to zero. ▲
● In the Main Menu, choose Calibration Factors > Hi Coef.
HIGH HCl COEFFICIENT:
HCl: 15.80
SET COEF TO: 1.000
The Reset User Calibration Defaults screen allows the user to reset the
calibration factors to factory defaults.
●In the Main Menu, choose Calibration Factors > Reset User Cal
Defaults.
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
The Calibration menu is used to automatically set the zero background and
the span coefficient. The calibration menu is similar for the single, dual,
and autorange mode as shown below. The dual and autorange modes have
two HCl span factors (high and low). This allows each range to be
calibrated separately. This is necessary if the two ranges used are not close
to one another. For example, a low HCl range of 0-5 ppm and a high HCl
range of 0-5000 ppm. For more information about calibration, see Chapter
4, “Calibration”.
●In the Main Menu, choose Calibration.
3-18 Model 15i Instruction Manual Thermo Fisher Scientific
Page 55
Operation
Calibration Menu
CALIBRATION:
>CAL BACKGROUND
CAL COEF
ZERO/SPAN CHECK
RANGE AVG DIAGS ALARM
Calibrate HCl
Background
CALIBRATION:
>CAL BACKGROUND
CAL HI COEF
CAL LO COEF
RANGE AVG DIAGS ALARM
The Calibrate HCl Background screen is used to adjust the instrument
zero, or background. Before making an adjustment, be sure the analyzer
samples zero air until the readings stabilize. The display shows the current
HCl reading and the current background value. Press to set the HCl
reading to zero.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “Calibration”.
● In the Main Menu, choose Calibration > Cal Background.
Calibrate HCl
Coefficient
HCl BACKGROUND:
HCl: 15.80
CURRENTLY: 0.00 ?
SET HCl TO ZERO
RANGE AVG DIAGS ALARM
The Calibrate HCl Coefficient screen is used to adjust the HCl coefficient
and enter the span concentration. The display shows the current HCl
concentration reading and the current HCl range. The next line of the
display is where the HCl calibration gas concentration is entered.
The HCl span coefficient is calculated, stored, and used to correct the
current HCl reading. For more information about calibration, see Chapter
4, “Calibration”. In dual or auto range modes, “High” or “Low” is
displayed to indicate the calibration of the high or low coefficient.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “Calibration”.
Thermo Fisher Scientific Model 15i Instruction Manual 3-19
Page 56
Operation
Calibration Menu
● In the Main Menu, choose Calibration > Cal Coef.
The Zero/Span Check menu is available only if the zero/span valve option
is installed. It is used to program the instrument to perform fully
automated zero and span checks or adjustments.
● In the Main Menu, choose Calibration > Zero/Span Check.
ZERO/SPAN CHECK:
>NEXT TIME 01Jan06 12:00
PERIOD HR 24
TOTAL DURATION HR 1.5
ZERO DURATION MIN 30
SPAN DURATION MIN 30 ª
RANGE AVG DIAGS ALARM
PURGE DURATION MIN 30
ZERO/SPAN AVG SEC 60
ZERO CAL RESET ON
SPAN CAL RESET OFF
ZERO/SPAN RATIO 1
The Next Time screen is used to view and set the initial date and time (24hour format) of the zero/span check. Once the initial zero/span check is
performed, the date and time of the next zero/span check is calculated and
displayed.
●In the Main Menu, choose Calibration > Zero/Span Check > Next
Time.
NEXT DATE AND TIME:
19 MAR 2007 12:34
PRESS TO EDIT
RANGE AVG DIAGS ALARM
3-20 Model 15i Instruction Manual Thermo Fisher Scientific
The Period Hours screen defines the period or interval between zero/span
checks. Periods between 0 and 999 hours are acceptable. To turn the
zero/span check off, set the period to 0.
●In the Main Menu, choose Calibration > Zero/Span Check > Period
HR.
ZERO/SPAN PERIOD:
CURRENTLY: 024 HRS
SET TO: 025 HRS ?
The Total Duration Hour line is the sum of the zero, span, and purge
duration minutes, and is for viewing only.
The Zero Duration Minutes screen defines how long zero air is sampled by
the instrument. The Span and Purge Duration Minutes screens look and
function the same way as the zero duration screen. The span duration
screen is used to set how long the span gas is sampled by the instrument.
The purge duration screen is used to set how long the purge period will be
after doing a zero or span check. This gives the instrument time to flush
out the zero and span gas before any meaningful data is taken. Logged data
is flagged as taken during a purge to show that the data is suspect.
Durations between 0 and 99 minutes are acceptable. Each time a zero/span
check occurs the zero check is done first, followed by the span check. To
perform just a zero check, set the span duration to 0 (off). The same applies
to perform just a span check.
● In the Main Menu, choose Calibration > Zero/Span Check > Zero,
Thermo Fisher Scientific Model 15i Instruction Manual 3-21
Page 58
Operation
Calibration Menu
Zero/Span Averaging
Time
Zero/Span Calibration
Reset
The Zero/Span Averaging Time screen allows the user to set the zero/span
averaging time. The zero/span averaging time is used by the analyzer only
when performing an automatic zero or span check or adjustment. The
analyzer’s averaging time is used for all other functions. The following
averaging times are available: 1, 2, 5, 10, 20, 30, 60, 90, 120, 180, 240,
and 300 seconds.
● In the Main Menu, choose Calibration > Zero/Span Check >
Zero/Span Avg Sec.
ZERO/SPAN AVERAGING TIME:
CURRENTLY: 60 SEC
SET TO: 90 SEC ?
Zero and Span Calibration Reset are toggle items that change between yes
or no when selected, and are displayed only if Auto Calibration is set to
YES in the Instrument Configuration screen in the Diagnostics menu.
Zero/Span Ratio
If the zero calibration reset line is set to yes, then not only is a zero check
done, but a zero adjustment is made. If the span calibration reset line is set
to yes, then not only is a span check done, but a span adjustment is made.
(This is how to set up a scheduled, recurring auto calibration.)
The Zero/Span Ratio screen is used to set the ratio of zero checks to span
checks. For example, if this value is set to 1, a span check will follow every
zero check. If this value is set to 3, there will be three zero checks between
each span check. This value may be set from 1 to 10, with 1 as default.
● In the Main Menu, choose Calibration > Zero/Span Check >
3-22 Model 15i Instruction Manual Thermo Fisher Scientific
Page 59
Operation
Instrument Controls Menu
Instrument Controls
Menu
Datalogging Settings
The Instrument Controls menu contains a number of items that may be
selected to control various instrument operational parameters. The
firmware controls listed in this menu enable control of the listed
instrument functions.
● In the Main Menu, choose Instrument Controls.
INSTRUMENT CONTROLS:
>DATALOGGING SETTINGS
COMMUNICATION SETTINGS
I/O CONFIGURATION
TEMPERATURE COMPENSATION
PRESSURE COMPENSATION ª
RANGE AVG DIAGS ALARM
SCREEN CONTRAST
SERVICE MODE
DATE/TIME
TIMEZONE
The iSeries instruments include a built-in datalogging capability as a
standard feature. The operator is allowed to create two different types of
records, which for historical reasons, are named lrecs and srecs. Each record
can contain up to 32 different fields, or data items, and records can be
created at user-defined intervals ranging from 1 to 60 minutes.
Record generation is tied to the instrument’s real-time clock. For example,
if the logging period for srecs is set to 30 minutes, a new srec will be
generated on every hour and every half hour (10:00, 10:30, 11:00 …).
Lrecs and srecs function independent of each other. So, for example, an
srec containing just the current HCl level could be generated every 5
minutes while an lrec containing a full set of diagnostic data could be
generated once every hour.
The analyzer's computer system includes three megabytes of flash memory
which is enough to store a full lrec containing 32 data items and a full srec
containing 32 items once each minute for a week (>20,000 total records).
If logging is limited to the minimum content of date, time, HCl
concentration and error flags, the analyzer can store data once each minute
for 4 months (>190,000 total records).
The Datalogging Settings menu allows the user flexibility in how data is
stored and recorded.
●In the Main Menu, choose Instrument Controls > Datalogging
Settings.
Thermo Fisher Scientific Model 15i Instruction Manual 3-23
The Select Srec/Lrec screen is used to select the log record type for other
operations in this menu.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Srec/Lrec.
SELECT LOG TYPE:
CURRENTLY: SREC
SET TO: LREC ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The View Logged Data screen is used to select the starting point to view
the logged data by the number of records or by date and time. Note that
both types of records cannot be viewed at the same time, only the selected
record type.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
The Number of Records screen is used to select the number of records to
view, ending with the most recent. It also shows the total number of
records that have been logged for the selected record type.
3-24 Model 15i Instruction Manual Thermo Fisher Scientific
The Date and Time screen is used to set a start date and time for which to
view logged data. For example, if “20 Jan 2007 10:00” is entered, then the
first logged data record that is displayed is the first record after this time. If
set to one minute logging, this would be at “20 Jan 2007 10:01”.
Thermo Fisher Scientific Model 15i Instruction Manual 3-25
Page 62
Operation
Instrument Controls Menu
Erase Log
The Erase Log screen is used to erase all saved data for the selected record
type only (not both srecs and lrecs).
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Erase Log.
ERASE LREC LOG FILE DATA?
ERASE
RANGE AVG DIAGS ALARM
Select Content
The Select Content submenu displays a list of 32 record fields to use to
configure the data to be logged. When a field is selected, a submenu list
appears of the items to choose from to assign to those fields. Item types are
Concentrations, Other Measurements, and Analog Inputs (if the I/O
expansion board is installed). These record fields comprise a temporary list
of items for the selected record type that must be committed via the
datalogging menu before the changes will apply. Note that committing any
changes to this list will erase all currently logged data for the selected record
type only, as the format of the stored data is changed.
ERASE LREC LOG FILE DATA?
ERASE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM ERASURE
RANGE AVG DIAGS ALARM
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content.
LREC FIELDS:
>FIELD 1 HCL
FIELD 2 HIHCL
FIELD 3 INTT
FIELD 4 CHT
FIELD 5 PRES ª
RANGE AVG DIAGS ALARM
Choose Field Data
The Choose Field Data submenu displays a list of the types of data that can
be logged for the current field. Choices are Concentrations, Other
Measurements, and Analog Inputs (if the I/O expansion board is installed).
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > Field 1–32.
3-26 Model 15i Instruction Manual Thermo Fisher Scientific
Page 63
Operation
Instrument Controls Menu
DATA IN SREC FIELD 1:
>CONCENTRATIONS
OTHER MEASUREMENTS
ANALOG INPUTS
RANGE AVG DIAGS ALARM
Concentrations
Other Measurements
The Concentrations screen allows the user to assign one of the
concentrations to the selected record field. The currently selected item is
shown by “<--” after it. Note that at this point, pressing indicates
that these are proposed changes as opposed to implemented changes. To
change the selected record format and erase record log file data, see
“Commit Content” below. In dual or auto range mode, “HI” or “LO” is
displayed to indicate high or low range concentrations.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > select Field > Concentrations.
CONCENTRATIONS:
>NONE
RANGE STATUS
HCl <- LO HCl
HI HCl
RANGE AVG DIAGS ALARM
The Other Measurements screen allows the user to assign one of the other
available measurement types to the selected record field. The currently
selected item is shown by “<--” after it. Items displayed are determined by
the options installed. Note that at this point, pressing indicates that
these are proposed changes as opposed to implemented changes. To change
the selected record format and erase record log file data, see “Commit
Content” below.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > select Field > Other Measurements.
Thermo Fisher Scientific Model 15i Instruction Manual 3-27
Page 64
Operation
Instrument Controls Menu
OTHER MEASUREMENTS:
>NONE
S/R
LO S/R
HI S/R
INT TEMP ª
RANGE AVG DIAGS ALARM
BENCH TEMP
BENCH PRES
SAMPLE FLOW
INTENSITY
MOTOR SPEED
BIAS SUPPLY
Analog Inputs
Commit Content
The Analog Inputs screen allows the user to select the parameter (none or
analog inputs 1–8) to the selected record field. The selected item is shown
by “<--” after it. Note that at this point, pressing indicates that these
are proposed changes as opposed to implemented changes. To change the
selected record format and erase record log file data, see “Commit
Content” below.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Select Content > select Field > Analog Inputs.
ANALOG INPUTS:
>NONE
ANALOG IN 1
ANALOG IN 2
ANALOG IN 3
ANALOG IN 4 ª
RANGE AVG DIAGS ALARM
The Commit Content screen is used to save any changes that have been
made to any of the record fields for the selected record type. Saving changes
will erase record log file data for that record type. If no changes have been
made “NO CHANGES TO RECORD LIST!” will appear. For more
information about selecting the content of logged data fields, see “Select
Content” above.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Commit Content.
3-28 Model 15i Instruction Manual Thermo Fisher Scientific
Page 65
Operation
Instrument Controls Menu
CHANGE LREC DATA AND
ERASE LREC LOG FILE DATA?
COMMIT
RANGE AVG DIAGS ALARM
Reset to Default Content
RESET LREC DATA AND
ERASE LREC LOG FILE DATA?
RESET
RANGE AVG DIAGS ALARM
CHANGE LREC DATA AND
ERASE LREC LOG FILE DATA?
COMMIT
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM ERASURE
RANGE AVG DIAGS ALARM
The Reset to Default Content screen is used to reset all of the datalogging
record field items to default values for the selected record type. For more
information about selecting the content of logged data fields, see “Select
Content” above.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Reset to Default Content.
RESET LREC DATA AND
ERASE LREC LOG FILE DATA?
RESET
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESETxx
RANGE AVG DIAGS ALARM
Configure Datalogging
The Configure Datalogging menu deals with datalogging configuration for
the currently selected record type.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging.
DATALOGGING SETTINGS:
>LOGGING PERIOD MIN 60
MEMORY ALLOCATION % 50
DATA TREATMENT AVG
RANGE AVG DIAGS ALARM
Logging Period Min
The Logging Period Min screen is used to select the logging period in
minutes for the selected record format (srec or lrec). List of choices include:
off, 1, 5, 15, 30, and 60 minutes (default).
Thermo Fisher Scientific Model 15i Instruction Manual 3-29
Page 66
Operation
Instrument Controls Menu
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Logging Period Min.
SET PERIOD FOR SREC:
CURRENTLY: 60 MIN
SET TO: OFF ?
The Memory Allocation Percent screen is used to select the percentage of
total memory that may be used by the selected record type (lrecs or srecs).
Percentages between 0 and 100% are available in increments of 10.
Changing this value results in log erasure for both types of records, and
changing the percent allocated to one record type will automatically change
the other.
● In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Memory Allocation %.
SET PERCENT LRECS:
CURRENTLY: 50%
SET TO: 60% ?
THIS WILL ERASE ALL LOGS!x
PRESS ¨ TO CONFIRM CHANGE
RANGE AVG DIAGS ALARM
The Data Treatment screen is used to select the data type for the selected
record type: whether the data should be averaged over the interval, the
minimum or maximum measured during the interval, or the current value
(last value measured). Data treatment doesn’t apply to all data, just to the
concentration measurement. All other data points log the current value at
the end of the interval.
Note This feature is found in all iSeries instruments, but it is
recommended that the data type be set to ONLY the current value (cur), as
the datalogging averaging is done in addition to the normal concentration
averaging. ▲
3-30 Model 15i Instruction Manual Thermo Fisher Scientific
Page 67
Operation
Instrument Controls Menu
● In the Main Menu, choose Instrument Controls > Datalogging Settings
The Communication Settings menu is used for communications control
and configuration.
●In the Main Menu, choose Instrument Controls > Communication
Settings.
COMMUNICATION SETTINGS:
>SERIAL SETTINGS
INSTRUMENT ID
COMMUNICATION PROTOCOL
STREAMING DATA CONFIG
TCP/IP SETTINGS
RANGE AVG DIAGS ALARM
The Serial Setting submenu is used for serial communications control and
configuration.
● In the Main Menu, choose Instrument Controls > Communication
Settings > Serial Settings.
SERIAL SETTINGS:
>BAUD RATE 9600
DATA BITS 8
PARITY NONE
STOP BITS 1
RS-232/485 SEL RS-232
RANGE AVG DIAGS ALARM
Baud Rate
The Baud Rate screen is used to set the RS-232/RS-485 interface baud rate.
Baud rates of 1200, 2400, 4800, 9600, 19200, 38400, 57600, and 115200
are available. The analyzer’s default baud rate is set to 9600 to provide
backwards compatibility with the older C-series analyzers.
Thermo Fisher Scientific Model 15i Instruction Manual 3-31
Page 68
Operation
Instrument Controls Menu
● In the Main Menu, choose Instrument Controls > Communication
DISCONNECT THE SERIALX X
CABLES BEFORE CHANGING X
THE SELECTION!XXXXXX TO CONTINUE
RANGE AVG DIAGS ALARM
The RS-232/RS-485 Selection screen allows the user to choose between the
RS-232 or RS-485 specification for serial communication.
Equipment Damage Disconnect the serial cable before changing the RS232 and RS-485 selection to prevent damage to any equipment currently
connected to the analyzer. ▲
● In the Main Menu, choose Instrument Controls > Communication
Settings > Serial Settings > RS-232/485 Selection.
RS-232/RS-485 SELECTION:
CURRENTLY: RS-232
SET TO: RS-485 ?
MAKE SURE THAT THE CABLEXX
IS OFF: PRESS ¨ TO CONFIRM
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Instrument ID
The Instrument ID screen allows the operator to edit the instrument ID
number. The ID is used to identify the instrument when using the C-Link
or MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more instruments of the same
model are connected to one computer. Valid instrument ID numbers are
from 0 to 127. The Model 15i has a default instrument ID of 15. For more
information about the instrument ID, see Appendix B “C-Link Protocol”
or Appendix C “MODBUS Protocol”.
● In the Main Menu, choose Instrument Controls > Communication
Settings > Instrument ID.
Thermo Fisher Scientific Model 15i Instruction Manual 3-33
The Communication Protocol screen is used to change the protocol for
serial communications. Possible choices include: C-Link, MODBUS,
Geysitech, and Streaming Data.
● In the Main Menu, choose Instrument Controls > Communication
Settings > Communication Protocol.
COMMUNICATION PROTOCOL:
CURRENTLY: CLINK
SET TO: STREAMING ?
The Streaming Data Configuration menu is used to allow for configuration
of the 8 streaming data output items, streaming interval, current data
format, and current timestamp setting. The Choose Stream Data submenu
displays a list of the analog output signal group choices to choose from.
Choices are Concentrations, Other Measurements, and Analog Inputs (if
the I/O expansion board option is installed).
● In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config.
STREAMING DATA CONFIG
>INTERVAL 10 SEC
ADD LABELS NO
PREPEND TIMESTAMP YES
ADD FLAGS YES
ITEM 1 HCL ª
RANGE AVG DIAGS ALARM
Streaming Data Interval
The Streaming Data Interval screen is used to adjust how frequently a new
record will be generated. The following interval times are available: 1, 2, 5,
10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds.
3-34 Model 15i Instruction Manual Thermo Fisher Scientific
Page 71
Operation
Instrument Controls Menu
● In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Interval.
STREAMING DATA INTERVAL:
CURRENTLY: 10 SEC
SET TO: 20 SEC ?
Note Add Labels, Prepend Timestamp, and Add Flags are toggle items that
change between yes or no when selected.
▲
Choose Stream Data
Concentrations
The Choose Stream Data screen displays a list of the types of data that can
be sent via streaming data. Choices are Concentrations, Other
Measurements, and Analog Inputs (if the I/O expansion board is installed).
● In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Item 1–8.
CHOOSE STREAM DATA:
>CONCENTRATIONS
OTHER MEASUREMENTS
ANALOG INPUTS
RANGE AVG DIAGS ALARM
The Concentrations screen allows the user to assign one of the
concentrations to the selected streaming data item. The currently selected
item is shown by “<--” after it. Once an item is selected, pressing
will save the selected streaming data item. In dual or auto range mode,
“HI” or “LO” is displayed to indicate high or low range concentrations.
● In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > select Item > Concentrations.
Thermo Fisher Scientific Model 15i Instruction Manual 3-35
Page 72
Operation
Instrument Controls Menu
CONCENTRATIONS:
>NONE
RANGE STATUS
HCl <- LO HCl
HI HCl
RANGE AVG DIAGS ALARM
Other Measurements
The Other Measurements screen allows the user to assign one of the other
available measurement types to the selected streaming data item. The
currently selected item is shown by “<--” after it. Once an item is selected,
pressing will save the selected streaming data item.
● In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > select Item > Other
Measurements.
OTHER MEASUREMENTS:
>NONE
S/R
LO S/R
HI S/R
INT TEMP ª
RANGE AVG DIAGS ALARM
BENCH TEMP
BENCH PRES
SAMPLE FLOW
INTENSITY
MOTOR SPEED
BIAS SUPPLY
Analog Inputs
The Analog Inputs screen allows the user to assign an analog input signal
(none or analog inputs 1–8) to the selected streaming data item. The
currently selected item is shown by “<--” after it. Once an item is selected,
pressing will save the selected streaming data item.
● In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > select Item > Analog Inputs.
ANALOG INPUTS:
>NONE
ANALOG IN 1
ANALOG IN 2
ANALOG IN 3
ANALOG IN 4 ª
RANGE AVG DIAGS ALARM
3-36 Model 15i Instruction Manual Thermo Fisher Scientific
Page 73
Operation
Instrument Controls Menu
TCP/IP Settings
Use DHCP
The TCP/IP Settings menu is used for defining parameters that are
required for Ethernet communications.
Note The instrument power must be cycled after any of these parameters
have been changed for the change to take effect. ▲
● In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings.
TCP/IP SETTINGS:
>USE DHCP OFF
IP ADDR 10.209.40.149
NETMASK 255.255.252.0
GATEWAY 10.209.42.1
HOST NAME ISERIESª
RANGE AVG DIAGS ALARM
NTP SVR 10.209.43.237
The Use DHCP screen is used to specify whether to use Dynamic Host
Configuration Protocol (DHCP) or not. When DHCP is enabled, the
network dynamically provides an IP address for the instrument.
IP Address
● In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Use DHCP.
DHCP:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
CYCLE POWER TO CHANGE DHCP
RANGE AVG DIAGS ALARM
The IP Address screen is used to edit the IP address. The IP address can
only be changed when DHCP is off. If DHCP is on, the instrument will
respond with “NOT SETTABLE IF DHCP IS ON”. For more
information on DHCP, see “Use DHCP” above.
● In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > IP Addr.
Thermo Fisher Scientific Model 15i Instruction Manual 3-37
Page 74
Operation
Instrument Controls Menu
IP ADDRESS:
CURRENT: 10.209.40.149
SET TO: 10.209.40.149 1
The Netmask screen is used to edit the netmask. The netmask is used to
determine the subnet on which the instrument can directly communicate
to other devices. The netmask can only be changed when DHCP is off. If
DHCP is on, the instrument will respond with “NOT SETTABLE IF
DHCP IS ON”. For more information on DHCP, see “Use DHCP”
above.
● In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Netmask.
NETMASK:
CURRENT: 255.255.252.0
SET TO: 255.255.252.0 0
The Default Gateway screen is used to edit the gateway address. The
default gateway can only be changed when DHCP is off. If DHCP is on,
the instrument will respond with “NOT SETTABLE IF DHCP IS ON”.
For more information on DHCP, see “Use DHCP” above. Any traffic to
addresses that are not on the local subnet will be routed through this
address.
● In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Gateway.
DEFAULT GATEWAY:
CURRENT: 10.209.42.1
SET TO: 10.209.42.1 1
The Network Time Protocol (NTP) Server screen is used to edit the IP
address of the NTP server. An NTP server may be used to periodically
synchronize the instrument’s real-time clock with a standard. More
information about the NTP servers and a list of public servers may be
found at http://www.ntp.org.
● In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > NTP Svr.
I/O Configuration
NTP SERVER IP ADDRESS:
CURRENT: 10.209.43.237
SET TO: 10.209.43.237 1
The I/O Configuration menu deals with configuration of the analyzer’s
I/O system. The analog input configuration is displayed only if the I/O
expansion board option is installed.
● In the Main Menu, choose Instrument Controls > I/O Configuration.
Thermo Fisher Scientific Model 15i Instruction Manual 3-39
Page 76
Operation
Instrument Controls Menu
I/O CONFIGURATION:
>OUTPUT RELAY SETTINGS
DIGITAL INPUT SETTINGS
ANALOG OUTPUT CONFIG
ANALOG INPUT CONFIG
RANGE AVG DIAGS ALARM
Output Relay Settings
Logic State
The Output Relay Settings menu displays a list of the 10 digital output
relays available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note The digital outputs may take up to one second after the assigned
state occurs to show up on the outputs.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital output pin and ground
is normally open and closes to trigger the digital output action.
● Press to toggle and set the logic state to normally open or
normally closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
3-40 Model 15i Instruction Manual Thermo Fisher Scientific
Page 77
Operation
Instrument Controls Menu
Instrument State
Alarms
The Instrument State submenu allows the user to select the instrument
state that is assigned to the selected relay output. A submenu lists signal
types of either alarms or non-alarm to choose from.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
The Alarms screen allows the user to select an alarm status for the selected
relay output. The currently selected item is shown by “<--” after it. The
I/O board status alarm is only present if the I/O expansion board is
installed. Zero and Span Check Cal are only present if the automated
zero/span check is enabled. See the Alarms Menu section later in this
chapter for more information on alarm items.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
ALARM STATUS ITEMS:
>NONE
GEN ALARM <--
CONC MAX
CONC MIN
INT TEMP ª
RANGE AVG DIAGS ALARM
BENCH TEMP
PRESSURE
SAMPLE FLOW
INTENSITY
MOTOR SPEED
BIAS VOLTAGE
MB STATUS
MIB STATUS
I/O BD STATUS
CONC ALARM
ZERO CHK/CAL
SPAN CHK/CAL
Thermo Fisher Scientific Model 15i Instruction Manual 3-41
Page 78
Operation
Instrument Controls Menu
Non-Alarm
Digital Input Settings
The Non-Alarm screen allows the user to select a non-alarm status for the
selected relay output. The currently selected item is shown by “<--” after it.
The zero, span, and sample modes are only present if the zero/span valve
option is installed.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE
LOCAL/REMOTE
SERVICE
UNITS ª
RANGE AVG DIAGS ALARM
ZERO MODE
SPAN MODE
SAMPLE MODE
PURGE MODE
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note The digital inputs must be asserted for at least one second for the
action to be activated. ▲
Not all of the I/O available in the instrument are brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. (See optional Terminal Block and Cable Kits.)
● In the Main Menu, choose Instrument Controls > I/O Configuration >
▲
Digital Input Settings.
DIGITAL INPUT SETTINGS:
>1 NOP SET BACKGROUND
2 NOP CAL TO LO SPAN
3 NOP AOUTS TO ZERO
4 NOP AOUTS TO FS
5 NOP NONE ª
RANGE AVG DIAGS ALARM
3-42 Model 15i Instruction Manual Thermo Fisher Scientific
Page 79
Operation
Instrument Controls Menu
Logic State
Instrument Action
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital input pin and ground is
normally open and closes to trigger the digital input action.
● Press to toggle and set the logic state to normally open or
normally closed.
DIGITAL INPUT SETUP:
>LOGIC STATE OPEN
INSTRUMENT ACTION
RANGE AVG DIAGS ALARM
The Instrument Action submenu allows the user to choose the instrument
action that is assigned to the selected digital input. The zero and span
modes are only present if the zero/span valve option is installed.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings > select Relay > Instrument Action.
Analog Output Configuration
(Select Channel)
CHOOSE ACTION:
>NONE
ZERO MODE
SPAN MODE
SET BACKGROUND <- CAL TO LO SPAN ª
RANGE AVG DIAGS ALARM
AOUTS TO ZERO
AOUTS TO FS
CAL TO HI SPAN
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, individual voltage channels 1–6, and
individual current channels 1–6 (if the I/O expansion board option is
installed).
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config.
Thermo Fisher Scientific Model 15i Instruction Manual 3-43
Page 80
Operation
Instrument Controls Menu
OUTPUT CHANNELS:
>ALLOW OVER/UNDER RANGE
ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2 ª
RANGE AVG DIAGS ALARM
Allow Over/Under Range
Analog Output Configuration
(Select Action)
The Allow Over/Under Range screen is used to select whether or not the
analog outputs are allowed to exceed the maximum selected value of 100
mV, 1 V, 5 V, 10 V, or 20 mA or the minimum selected value of 0 V, 0
mA, or 4 mA. By default, this parameter is set to on and 5% over and
under range is allowed for all analog output channels.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Allow Over/Under Range.
ALLOW OVER/UNDER RANGE:
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Analog Output Configuration menu displays a list of the analog
output configuration choices, from which the user selects the parameter to
adjust for the selected output channel. Configuration choices include
selecting range, setting minimum/maximum values, and choosing the
signal to output.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > All Voltage Channels, All Current
Channels, Voltage Channel 1–6 or Current Channel 1–6.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE AVG DIAGS ALARM
3-44 Model 15i Instruction Manual Thermo Fisher Scientific
Page 81
Operation
Instrument Controls Menu
Select Range
Minimum and Maximum
Value
The Select Range screen is used to select the hardware range for the selected
analog output channel. Possible ranges for the voltage outputs are: 0-100
mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current outputs are: 0-20
mA and 4-20 mA.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > select Channel > Select Range.
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-100mV ?
The Minimum and Maximum Value screens are used to edit the zero (0)
and full-scale (100) values, respectively, in percentages for the selected
analog output channel. See Table 3–6 for a list of choices. The minimum
and maximum output value screens function the same way. The following
example shows the set minimum value screen.
● In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > select Channel > Set Minimum or Maximum
Value.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V ALL
CURRENTLY: N/A %
SET TO: 0000.5 % ?
Thermo Fisher Scientific Model 15i Instruction Manual 3-45
Page 82
Operation
Instrument Controls Menu
Table 3–6. Analog Output Zero to Full-Scale
Output Zero (0)% Value Full-Scale 100% Value
HCl Zero (0) Range Setting
HI HCl Zero (0) Range Setting
LO HCl Zero (0) Range Setting
Range Status Changing the setting for this output is not recommended
S/R Ratio 0.0 1.2
LO S/R Ratio 0.0 1.2
HI S/R Ratio 0.0 1.2
Internal Temp User-set alarm min value User-set alarm max value
Bench Temp User-set alarm min value User-set alarm max value
Pressure User-set alarm min value User-set alarm max value
Choose Signal to Output
Sample Flow User-set alarm min value User-set alarm max value
Intensity User-set alarm min value User-set alarm max value
Motor Speed User-set alarm min value User-set alarm max value
Bias Voltage User-set alarm min value User-set alarm max value
Everything Else 0 Units 10 Units
The Choose Signal to Output screen displays a list of the analog output
signal choices. Choices are grouped into three categories: Concentrations,
Other Measurements, and Analog Inputs (if the I/O expansion board
option is installed). This allows the user to select the output signal that will
be assigned to each analog channel. The Concentrations screen is shown
below. See Table 3–7 for a list of items for each signal group choice.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > select Channel > Choose Signal to Output.
CHOOSE SIGNAL TYPE:
>CONCENTRATIONS
OTHER MEASUREMENTS
ANALOG INPUTS
RANGE AVG DIAGS ALARM
3-46 Model 15i Instruction Manual Thermo Fisher Scientific
Page 83
Operation
Instrument Controls Menu
CHOOSE SIGNAL - CONC
SELECTED OUTPUT: V1
CURRENTLY: HCl
SET TO: NONE ?
HCl (singe range mode) Sample/Reference Analog Input 1 (if the I/O expansion
board is installed)
Analog Input Configuration
LO HCl (dual/auto range
mode)
HI HCl (dual/auto range
mode)
Range Status (auto
range mode)
Bench Temperature Analog Input 5 (if the I/O expansion
Bench Pressure Analog Input 6 (if the I/O expansion
Sample Flow Analog Input 7 (if the I/O expansion
Intensity Analog Input 8 (if the I/O expansion
Motor Speed
Bias Supply
LO Sample/Reference Analog Input 2 (if the I/O expansion
board is installed)
HI Sample/Reference Analog Input 3 (if the I/O expansion
board is installed)
Internal Temp Analog Input 4 (if the I/O expansion
board is installed)
board is installed)
board is installed)
board is installed)
board is installed)
The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O
expansion board option is installed. Configuration includes entering the
Descriptor, Units, Decimal Places, choice of 1–10 points in the table, and
corresponding number of points selected.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config.
Thermo Fisher Scientific Model 15i Instruction Manual 3-47
ANALOG INPUT 01 CONFIG:
>DESCRIPTOR IN1
UNITS V
DECIMAL PLACES 2
TABLE POINTS 2
POINT 1 ª
RANGE AVG DIAGS ALARM
POINT 2
Descriptor
Units
The Descriptor screen allows the user to enter the descriptor, or name, for
the selected analog input channel. The descriptor is used in datalogging
and streaming data to report what data is being sent out. The descriptor
may be from 1 to 3 characters in length, and defaults to IN1 to IN8 (user
input channel number).
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Descriptor.
ANALOG INPUT DESCRIPTOR:
CURRENTLY: IN1
IN1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in
datalogging and streaming data. The units may be from 1 to 3 characters in
length, and defaults to V (volts).
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Units.
3-48 Model 15i Instruction Manual Thermo Fisher Scientific
Page 85
Operation
Instrument Controls Menu
ANALOG INPUT UNITS:
CURRENTLY: V
VN1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
Decimal Places
Number of Table Points
The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal point, from 0 to 6, with a default of 2.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Decimal Places.
The Number of Table Points screen allows the user to select how many
points are used in the analog input conversion table for the selected
channel. The instrument uses linear interpolation between the points in
this table to determine what the reading value is based on the analog input
voltage. Each point in the table consists of an analog input voltage value
(0–10.5 V) and a corresponding reading value. Only two points are
necessary for linear inputs, however a larger number of points may be used
to approximate non-linear inputs. The points range from 2 to 10, with a
default of 2.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Table Points.
Thermo Fisher Scientific Model 15i Instruction Manual 3-49
Page 86
Operation
Instrument Controls Menu
Table Points
Volts
The Table Point submenu allows the user to set up an individual table
point.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Point 1–10.
TABLE POINT 01 CONFIG:
>VOLTS 0.00
USER VALUE 0.00
RANGE AVG DIAGS ALARM
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is
a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U, where U is the previously entered unit of measure.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > select Point > Volts.
User Value
TABLE POINT 01 VOLTS:
CURRENTLY: 0.00
SET TO: 00.00
The User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion
table, from -9999999 to 99999999. The default table is a two-point table
with point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U, where U is
the previously entered unit of measure.
● In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > select Point > User Value.
3-50 Model 15i Instruction Manual Thermo Fisher Scientific
Page 87
Operation
Instrument Controls Menu
TABLE POINT 01 USER VAL:
CURRENTLY: 0.00
SET TO: 00000.00
The Temperature Compensation screen allows the user to turn temperature
compensation on or off. Temperature compensation changes to the
instrument's output signal that occur due to internal instrument
temperature variations. The effects of temperature changes on the analyzer's
subsystems and output have been empirically determined. This empirical
data is used to compensate for any changes in temperature.
When temperature compensation is on, the display shows the current
internal instrument temperature (measured by a thermistor on the Interface
board). When temperature compensation is off, the display shows the
factory standard temperature of 30.0 °C.
●In the Main Menu, choose Instrument Controls > Temperature
Compensation.
TEMPERATURE COMPENSATION:
COMP TEMP: 25.0 oC
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Pressure
Compensation
Pressure compensation provides compensation for any changes to the
instrument's output signal due to reaction chamber pressure variations. The
effects of reaction chamber pressure changes on the analyzer's subsystems
and output have been empirically determined. This empirical data is used
to compensate for any change in reaction chamber pressure.
When pressure compensation is on, the first line of the display represents
the current pressure in the fluorescence chamber. When pressure
compensation is off, the first line of the display shows the factory standard
pressure of 750.0 mmHg.
●In the Main Menu, choose Instrument Controls > Pressure
Compensation.
Thermo Fisher Scientific Model 15i Instruction Manual 3-51
Page 88
Operation
Instrument Controls Menu
PRESSURE COMPENSATION:
COMP PRES: 750.0 mmHg
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Screen Contrast
The Screen Contrast screen is used to change the contrast of the display.
Intensities between 0 and 100% in increments of 5 are available. Changing
the screen contrast may be necessary if the instrument is operated at
extreme temperatures.
Note The optimal contrast will change with changes in temperature.
▲
The optimal contrast will change from one LCD to another. If the LCD is
replaced, the contrast may need to be reset.
▲
If the display contrast is not optimal, but the content on the screen is
visible, select Instrument Controls > Screen Contrast and adjust the screen
contrast. If the content on the screen is not visible, use the “set contrast 10”
C-Link command to set the screen contrast to mid-range, then optimize
the contrast. See “contrast levels” in Appendix B, “C-Link Protocol
Commands” for more information on this command. ▲
● In the Main Menu, choose Instrument Controls > Screen Contrast.
The Service Mode screen is used to turn the service mode on or off.
Turning service mode “ON” locks out any remote actions and allows access
to parameters and functions that are useful when making adjustments or
diagnosing the Model 15i. The service (wrench) icon on the status bar is
shown when service mode is on. For more information about the service
mode, see “Service Menu” later in this chapter.
3-52 Model 15i Instruction Manual Thermo Fisher Scientific
Page 89
Operation
Instrument Controls Menu
Note The service mode should be turned off when finished, as it prevents
remote operation. ▲
● In the Main Menu, choose Instrument Controls > Service Mode.
SERVICE MODE:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Date/Time
The Date/Time screen allows the user to view and change the system date
and time (24-hour format). The internal clock is powered by its own
battery when instrument power is off.
●In the Main Menu, choose Instrument Controls > Date/Time.
DATE AND TIME:
19 MAR 2005 12:34:56
PRESS TO EDIT
RANGE AVG DIAGS ALARM
Timezone
The Timezone screen is used to set the timezone for the Network Time
Protocol (NTP) server. This should be set to the timezone that the
instrument is located in. If the exact timezone is not shown in the list, it
may be entered via the C-LINK “tz” command (see Appendix B). The
selections are: UTC (GMT), EST (GMT+5), CST (GMT+6), MST
(GMT+7), PST (GMT+8), YST (GMT+9), HST (GMT+10), NST
(GMT+11), DLW (GMT+12), CET (GMT-1), EET (GMT-2), BST
(GMT-3), DLT (GMT-4), ECH (GMT-5), FOX (GMT-6), GLF (GMT-
7), CCT (GMT-8), JST (GMT-9), GST (GMT-10), LMA (GMT-11),
DLE (GMT-12), EDT (GMT+5/4), CDT (GMT+6/5), MDT
(GMT+7/6), and PDT (GMT+8/7).
DATE AND TIME:
19 MAR 2005 12:34:56 ?
SETTING: DAYS
The Diagnostics menu provides access to diagnostic information and
functions. This menu is useful when troubleshooting the instrument. The
analog input readings and analog input voltages are only displayed if the
I/O expansion board option is installed.
The Program Versions screen (read only) shows the version numbers of the
programs installed. Prior to contacting the factory with any questions
regarding the instrument, please note the product model name and the
program version numbers.
● In the Main Menu, choose Diagnostics > Program Versions.
3-54 Model 15i Instruction Manual Thermo Fisher Scientific
Page 91
Operation
Diagnostics Menu
PROGRAM VERSIONS:
PRODUCT: MODEL 15i
VERSION: 01.05.02.141
FIRMWARE: 10.13.77
RANGE AVG DIAGS ALARM
Voltages
Motherboard Voltages
The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board item is only
displayed if the I/O expansion board option is installed.
●In the Main Menu, choose Diagnostics > Voltages.
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
RANGE AVG DIAGS ALARM
The Motherboard screen (read only) is used to display the voltage readings
on the motherboard.
●In the Main Menu, choose Diagnostics > Voltages > Motherboard.
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.1 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
Interface Board Voltages
The Interface Board screen (read only) is used to display the voltage
readings on the measurement interface board.
●In the Main Menu, choose Diagnostics > Voltages > Interface Board.
Thermo Fisher Scientific Model 15i Instruction Manual 3-55
Page 92
Operation
Diagnostics Menu
INTERFACE BOARD VOLTAGES:
> 3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-15.0 SUPPLY -15.0 Vª
RANGE AVG DIAGS ALARM
18.0 IR SUPPLY 18.0 V
18.0 MOT SUPPLY 18.0 V
BIAS SUPPLY -110.0 V
I/O Board Voltages
Temperatures
The I/O Board screen (read only) is used to display the voltage readings on
the I/O expansion board. This menu is only displayed if the I/O expansion
board option is installed.
● In the Main Menu, choose Diagnostics > Voltages > I/O Board.
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
The Temperatures screen (read only) displays the current internal
instrument temperature and bench temperature. The internal temperature
is the air temperature measured by a sensor located on the measurement
interface board.
●In the Main Menu, choose Diagnostics > Temperatures.
TEMPERATURES:
INTERNAL 34.6 oC
BENCH 49.7 oC
RANGE AVG DIAGS ALARM
Pressure
The Pressure screen (read only) displays the current optical bench pressure.
The pressure is measured by a pressure transducer.
●In the Main Menu, choose Diagnostics > Pressure.
3-56 Model 15i Instruction Manual Thermo Fisher Scientific
Page 93
Operation
Diagnostics Menu
PRESSURE: 753.0 mmHg
RANGE AVG DIAGS ALARM
Flow
Sample/Reference
Ratio
The Flow screen (read only) displays the flow rate. The flow is measured by
internal flow sensors. For more information, see Figure 1–1.
●In the Main Menu, choose Diagnostics > Flow.
SAMPLE FLOW: 1.108 LPM
RANGE AVG DIAGS ALARM
The Sample/Reference Ratio screen (read only) displays the ratio of the
intensities of the light source through the sample (HCl) side and reference
(N2) side of the correlation wheel. Normally, when zero air is being
sampled, the sample/reference ratio is between 1.05 and 1.15. A ratio
outside may indicate that the correlation wheel is dirty or leaking gas.
● In the Main Menu, choose Diagnostics > Sample/Ref Ratio.
SAMPLE REFERENCE RATIO:
S/R: 1.100000
RANGE AVG DIAGS ALARM
AGC Intensity
The AGC Intensity screen (read only) displays the intensity (in Hertz) of
the reference channel Automatic Gain Control (AGC) circuit. The AGC
circuit optimizes the noise and resolution levels of the analyzer. The AGC
intensity reading should be about 250,000 Hertz.
● In the Main Menu, choose Diagnostics > AGC Intensity.
Thermo Fisher Scientific Model 15i Instruction Manual 3-57
Page 94
Operation
Diagnostics Menu
AGC INTENSITY:
INTENSITY: 250000 Hz
RANGE AVG DIAGS ALARM
Motor Speed
Analog Input
Readings
The Motor Speed screen (read only) displays the status of the chopper
motor. A reading of 100.0% means that the motor speed is correct. A
reading other than 100.0% indicates that there is a problem with the
chopper motor or power supply.
● In the Main Menu, choose Diagnostics > Motor Speed.
MOTOR SPEED:
SPEED: 100.00 %
RANGE AVG DIAGS ALARM
The Analog Input Readings screen (read only) displays the 8 user-scaled
analog readings (if the I/O expansion board option is installed).
● In the Main Menu, choose Diagnostics > Analog Input Readings.
ANALOG INPUT READINGS:
>HCL 10.2 PPM
SO2 18.2 PPM
CO2 250 PPM
FL1 20.42 LPM
IO5 0.0 Vª
RANGE AVG DIAGS ALARM
Analog Input
Voltages
The Analog Input Voltages screen (read only) displays the 8 raw analog
voltage readings (if the I/O expansion board option is installed).
● In the Main Menu, choose Diagnostics > Analog Input Voltages.
3-58 Model 15i Instruction Manual Thermo Fisher Scientific
Page 95
Operation
Diagnostics Menu
ANALOG INPUT VOLTAGES:
>ANALOG IN 1 6.24 V
ANALOG IN 2 4.28 V
ANALOG IN 3 0.00 V
ANALOG IN 4 0.00 V
ANALOG IN 5 0.00 Vª
RANGE AVG DIAGS ALARM
Digital Inputs
Relay States
The Digital Inputs screen (read only) displays the state of the 16 digital
inputs. Pull-ups are provided on all of the inputs, so if nothing they will
read (1), if an input is brought to ground, it will read (0).
● In the Main Menu, choose Diagnostics > Digital Inputs.
DIGITAL INPUTS:
>INPUT 1 1
INPUT 2 1
INPUT 3 1
INPUT 4 1
INPUT 5 1ª
RANGE AVG DIAGS ALARM
The Relay States screen displays the state of the 10 digital outputs and
allows toggling of the state to either on (1) or off (0). The relays are
restored to their original states upon exiting this screen.
● In the Main Menu, choose Diagnostics > Relay States.
● Press to toggle and set the relay state to open or closed for the
The Test Analog Outputs menu is used to set each of the analog output
channels to zero or full-scale. Channel choices include all analog outputs, 6
voltage channels, and 6 current channels (if the I/O expansion board
option is installed).
● In the Main Menu, choose Diagnostics > Test Analog Outputs.
Thermo Fisher Scientific Model 15i Instruction Manual 3-59
Page 96
Operation
Diagnostics Menu
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
RANGE AVG DIAGS ALARM
Set Analog Outputs
Instrument Configuration
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the fullscale voltage, zero sets the analog outputs to 0 volts, and reset returns the
output to normal operation. The analog outputs are returned to normal
operation upon exiting this screen. The following example shows the screen
when all analog outputs are set to “normal” operating mode.
●In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL,
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
Note If the analyzer is in service mode, pressing will toggle the
selected item between yes and no. Zero/span valves needs to be set to ON
in order to do zero and span checks, and auto calibration needs to be set to
ON in order to do zero and span calibration resets.
● In the Main Menu, choose Diagnostics > Instrument Configuration.
▲
3-60 Model 15i Instruction Manual Thermo Fisher Scientific
Page 97
Operation
Alarms Menu
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
ZERO/SPAN VALVES YES
ZERO AIR SCRUBBER YES
WHEEL PURGE YES
DILUTION NO
RANGE AVG DIAGS ALARM
AUTO CALIBRATION NO
Contact Information
Alarms Menu
The Contact Information screen displays the customer service information.
● In the Main Menu, choose Diagnostics > Contact Information.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB: WWW.THERMO.COM
RANGE AVG DIAGS ALARM
The alarms menu displays a list of items that are monitored by the analyzer.
If the item being monitored goes outside the lower or upper limit, the
status of that item will go from “OK” to either “LOW” or “HIGH”,
respectively. If the alarm is not a level alarm, the status will go from “OK”
to “FAIL”. The number of alarms detected is displayed to indicate how
many alarms have occurred. If no alarms are detected, the number zero is
displayed.
To see the actual reading of an item and its minimum and maximum
limits, move the cursor to the item and press
.
Items displayed are determined by the options installed. The zero/span
check and auto calibration screens are visible only if the zero/span check or
auto calibration options are enabled. The motherboard status, interface
board status, and I/O Expansion board status (if installed) indicate that the
power supplies are working and connections are successful. There are no
setting screens for these alarms.
If any alarms are active, the alarm (bell) icon is displayed on the right side
of the status bar.
●In the Main Menu, choose Alarms.
Thermo Fisher Scientific Model 15i Instruction Manual 3-61
Page 98
Operation
Alarms Menu
ALARMS:
ALARMS DETECTED 0
>INTERNAL TEMP OK
BENCH TEMP OK
PRESSURE OK
SAMPLE FLOW OKª
RANGE AVG DIAGS ALARM
BIAS VOLTAGE OK
AGC INTENSITY OK
MOTOR SPEED OK
ZERO CHECK OK
SPAN CHECK OK
ZERO AUTOCAL OK
SPAN AUTOCAL OK
CONCENTRATION OK
MOTHERBOARD STATUS OK
INTERFACE STATUS OK
I/O EXP STATUS OK
Internal
Temperature
Min and Max Internal
Temperature Limits
The Internal Temperature screen displays the current internal temperature
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 8 to 47 °C. If the internal temperature reading goes beyond
either the minimum or maximum limit, an alarm is activated and the alarm
(bell) icon appears in the status bar on the Run screen and in the Main
Menu.
● In the Main Menu, choose Alarms > Internal Temp.
INTERNAL TEMPERATURE:
ACTUAL 30.1 oC
>MIN 8.0 oC
MAX 47.0 oC
RANGE AVG DIAGS ALARM
The Minimum Internal Temperature alarm limit screen is used to change
the minimum internal temperature alarm limit. The minimum and
maximum internal temperature screens function the same way.
● In the Main Menu, choose Alarms > Internal Temp > Min or Max.
3-62 Model 15i Instruction Manual Thermo Fisher Scientific
Page 99
Operation
Alarms Menu
INTERNAL TEMPERATURE:
ACTUAL MIN: 8.0 oC
SET MIN TO: 10.0 oC ?
The Bench Temperature screen displays the current bench temperature and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 40 to 59 °C. If the bench temperature reading goes beyond
either the minimum or maximum limit, an alarm is activated and the alarm
(bell) icon appears in the status bar on the Run screen and in the Main
Menu.
● In the Main Menu, choose Alarms > Bench Temp.
BENCH TEMPERATURE:
ACTUAL 48.4 oC
>MIN 40.0 oC
MAX 59.0 oC
RANGE AVG DIAGS ALARM
The Minimum Bench Temperature alarm limit screen is used to change the
minimum bench temperature alarm limit. The minimum and maximum
Bench Temperature screens function the same way.
● In the Main Menu, choose Alarms > Bench Temp > Min or Max.
BENCH TEMPERATURE:
ACTUAL MIN: 40.0 oC
SET MIN TO: 44.0 oC ?
The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 250 to 1,000 mmHg. If the pressure reading goes beyond either
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
Thermo Fisher Scientific Model 15i Instruction Manual 3-63
Page 100
Operation
Alarms Menu
● In the Main Menu, choose Alarms > Pressure.
PRESSURE:
ACTUAL 746.9 mmHg
>MIN 250.0 mmHg
MAX 1000.0 mmHg
RANGE AVG DIAGS ALARM
Min and Max Pressure
Limits
Sample Flow
The Minimum Pressure alarm limit screen is used to change the minimum
temperature alarm limit. The minimum and maximum pressure screens
function the same way.
● In the Main Menu, choose Alarms > Pressure > Min or Max.
PRESSURE:
ACTUAL MIN: 250.0 mmHg
SET MIN TO: 500.0 mmHg?
The Sample Flow screen displays the current sample flow reading and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from .350 to 1.500 LPM. If the sample flow reading goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
● In the Main Menu, choose Alarms > Sample Flow.
SAMPLE FLOW:
ACTUAL 0.750 LPM
>MIN 0.350 LPM
MAX 1.500 LPM
RANGE AVG DIAGS ALARM
Min and Max Sample
Flow Limits
The Minimum Sample Flow alarm limit screen is used to change the
minimum sample flow alarm limit. The minimum and maximum sample
flow screens function the same way.
3-64 Model 15i Instruction Manual Thermo Fisher Scientific
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.